Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Embed Size (px)

Citation preview

Page 1: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Proposal for a self-calibrating and Proposal for a self-calibrating and instrument-independent MOS DRSinstrument-independent MOS DRS

Carlo IzzoCarlo Izzo

Page 2: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

MOS arc lamp exposure: MOS arc lamp exposure: FORS2-MXU, GRIS_150I+27FORS2-MXU, GRIS_150I+27

Page 3: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Using first-guess models to Using first-guess models to find reference lines…find reference lines…

Page 4: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Earthquake!Earthquake!

Page 5: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

High ExpectationsHigh Expectations Traditional data reduction techniques are based on first-Traditional data reduction techniques are based on first-

guess distortion models.guess distortion models.

• The instrument is stableThe instrument is stable, , together with its components together with its components (grisms, masks, filter wheels, etc.)(grisms, masks, filter wheels, etc.)

• The code can be kept simpleThe code can be kept simple, because the reference , because the reference patterns in the calibration exposures (flats, arcs) are safely patterns in the calibration exposures (flats, arcs) are safely identifiedidentified

• The procedures will be generalThe procedures will be general, because different , because different distortion models can be stored in appropriate configuration distortion models can be stored in appropriate configuration filesfiles

Page 6: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

The hard realityThe hard reality

• The instrument is not stableThe instrument is not stable, changing in the short and , changing in the short and long time scales, requiring continuous maintenance work long time scales, requiring continuous maintenance work on the configuration fileson the configuration files

• The code cannot be kept simpleThe code cannot be kept simple, because the reference , because the reference patterns are not safely identified (e.g., due to unexpected patterns are not safely identified (e.g., due to unexpected contaminations, or to instrument instabilities)contaminations, or to instrument instabilities)

• The procedures cannot be kept generalThe procedures cannot be kept general, because the , because the ad hocad hoc solutions adding robustness to the DRS are typically solutions adding robustness to the DRS are typically instrument-dependentinstrument-dependent

Page 7: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

VIMOSVIMOS

Page 8: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Interactive systemsInteractive systems

Page 9: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

First-guess vs pattern-recognition

Page 10: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Distortion models parametersDistortion models parameters

Page 11: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

DFO reports a problem…DFO reports a problem…

Page 12: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

… … and the problem is “solved”and the problem is “solved”

Page 13: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

… … but what’s the use of it?but what’s the use of it?

Using first-guess model…Using first-guess model…

• The pipeline may stop with a The pipeline may stop with a genericgeneric ““calibration failedcalibration failed””

• The pipeline may find a The pipeline may find a wrong solutionwrong solution

• The QC1 parameters may The QC1 parameters may show nothing strangeshow nothing strange

Using pattern-recognition…Using pattern-recognition…

• The pipeline always completes The pipeline always completes successfullysuccessfully

• The pipeline always finds the The pipeline always finds the right solutionright solution

• The QC1 parameters report The QC1 parameters report exactly what happenedexactly what happened

Page 14: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Do we need a physical model of the Do we need a physical model of the distortions?distortions?

YES!YES! A physical model of the optical distortions is necessary A physical model of the optical distortions is necessary for comparing the expected distortions with the observed for comparing the expected distortions with the observed ones ones (instrument health monitoring)(instrument health monitoring)

BUT:BUT: We should We should notnot use the model of the expected distortions use the model of the expected distortions as a first-guess as a first-guess (even if we may use it for fitting the data)(even if we may use it for fitting the data)

ALSO:ALSO: A A physicalphysical model of the instrument distortions is model of the instrument distortions is necessary for a meaningful instrument health monitoringnecessary for a meaningful instrument health monitoring

Page 15: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Fix the models, or fix the Fix the models, or fix the instrument?instrument?

• In principleIn principle, the instrument should be fixed., the instrument should be fixed.

• In practiceIn practice, it is often necessary to fix the models because:, it is often necessary to fix the models because:

• To fix the instrument is not always immediate (see for instance To fix the instrument is not always immediate (see for instance the light contaminations in FORS, or the flexures in VIMOS), and the light contaminations in FORS, or the flexures in VIMOS), and in the meantime we must keep reducing the datain the meantime we must keep reducing the data

• Sometimes the real optical distortions are “accepted”, even if Sometimes the real optical distortions are “accepted”, even if they are far from the instrument original designthey are far from the instrument original design

Using a pattern-recognition approach we Using a pattern-recognition approach we would not need to fix models anymore!would not need to fix models anymore!

Page 16: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Looking for patternsLooking for patterns

The pattern: The pattern: wavelengthswavelengths• ……• 5400.5625400.562• 5460.7425460.742• 5764.4195764.419• 5769.5985769.598• 5790.6565790.656• 5852.4885852.488• 5875.6205875.620• 5881.9005881.900• 5944.8305944.830• ……

The data: The data: pixel positionspixel positions• ……• 1220.641220.64• 1253.231253.23• 1299.441299.44• 1304.071304.07• 1339.301339.30• 1400.331400.33• 1450.281450.28• 1457.321457.32• 1471.001471.00• 1496.211496.21• 1520.661520.66• 1522.441522.44• 1549.011549.01• ……

Page 17: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Looking for peaksLooking for peaks

________________________________________

Page 18: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Looking for peaksLooking for peaks

• Any local maximum identifies a peakAny local maximum identifies a peak• A peak positions is determined by parabolic interpolation of A peak positions is determined by parabolic interpolation of

the three nearby pixel valuesthe three nearby pixel values

Page 19: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Looking for peaksLooking for peaks

Page 20: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

A simple case: FORS2-LSS A simple case: FORS2-LSS GRIS_1200RGRIS_1200R_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 21: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Identifying arc lamp linesIdentifying arc lamp lines

Page 22: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Wavelength calibrationWavelength calibration

Mean accuracy: 0.05 pixelMean accuracy: 0.05 pixel

Page 23: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Resampled spectrumResampled spectrum

Mean accuracy: 0.05 pixelMean accuracy: 0.05 pixel

Page 24: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Another Another case: VIMOS case: VIMOS GRIS_HRredGRIS_HRred

Page 25: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Wavelength Wavelength calibrationcalibration

Mean accuracy: 0.07 pixelMean accuracy: 0.07 pixel

Page 26: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Another case: Another case: FORS1-MOS GRIS_300VFORS1-MOS GRIS_300V

Page 27: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Wavelength calibrationWavelength calibration

Mean accuracy: 0.09 pixelMean accuracy: 0.09 pixel

Page 28: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Another case: FORS2-MXU Another case: FORS2-MXU GRIS_150IGRIS_150I

Page 29: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Wavelength calibrationWavelength calibration

Mean accuracy: 0.05 pixelMean accuracy: 0.05 pixel

Page 30: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Identifying the slitsIdentifying the slits

Select the reference wavelength: in this example,Select the reference wavelength: in this example, = 7000.00= 7000.00

Page 31: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Rectified imageRectified image

Page 32: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

AccuracyAccuracy

Page 33: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

AccuracyAccuracy

The accuracy of the extraction mask depends on many factors:The accuracy of the extraction mask depends on many factors:

• Number of fitted points,Number of fitted points,• Accuracy of peaks positions,Accuracy of peaks positions,• Appropriate choice of fitting models,Appropriate choice of fitting models,• Position along the spectral interval,Position along the spectral interval,

… … but, above all, but, above all,

• Correct identification of the detected peaks.Correct identification of the detected peaks.

Inaccuracy comes from misidentification!Inaccuracy comes from misidentification!

Page 34: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

This system is flexibleThis system is flexible

• AnyAny MOS arc lamp exposure can be wavelength calibrated MOS arc lamp exposure can be wavelength calibrated (instrument-independency)(instrument-independency)

• This method can also be directly applied to the scientific This method can also be directly applied to the scientific exposures exposures (if the sky is visible and there are enough sky (if the sky is visible and there are enough sky lines)lines)

• This method may even be applied to intermediate products This method may even be applied to intermediate products from from anyany kind of spectroscopic data kind of spectroscopic data (not just MOS, but also (not just MOS, but also IFU, echelle, etc.).IFU, echelle, etc.).

Page 35: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

GIRAFFE Medusa1_H525.8nmGIRAFFE Medusa1_H525.8nm

Mean accuracy: 0.10 pixelMean accuracy: 0.10 pixel

Page 36: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo

Other issuesOther issues

After the extraction mask is completely defined, the usual After the extraction mask is completely defined, the usual reduction steps can be applied:reduction steps can be applied:

• Object detection,Object detection,• Determination of the sky spectrum,Determination of the sky spectrum,• Optimal extraction,Optimal extraction,• Combining different spectra,Combining different spectra,• Error propagationError propagation

Page 37: Proposal for a self-calibrating and instrument-independent MOS DRS Carlo Izzo