ImmunoCAP® ISAC slgE 112
Product Characteristics
IMMUNOCAP® ISAC 112
CONTeNTSIntended Use 1Principle of test procedure 1Clinical utility 1Sample information 2Measuring range 2Precision 2Sensitivity 3Specificity 3Linearity 3Healthy controls 4Comparison with ImmunoCAP Allergen Component 4Ordering information 5Examples of published ImmunoCAP ISAC 103 studies 5
Product Characteristics 1
IMMUNOCAP ISAC slgE 112
INTeNDeD USe
ImmunoCAP ISAC sIgE 112 is an in vitro semi-quantitative assay for the measurement of allergen specific IgE antibodies in human serum or plasma. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories.
PRINCIPLe OF TeST PROCeDURe
ImmunoCAP ISAC sIgE 112 is a solid-phase immunoassay. Allergen components are immobilized on a solid substrate in a microarray format and react with the specific IgE in the patient sample. After washing away non-specific IgE, fluorescence-labeled anti-human IgE antibody is added to form a complex. After incubation unbound fluorescence-labeled anti-human IgE antibodies are removed by washing. The procedure is followed by fluorescence measurement using an appro-priate microarray scanner. The higher the response value the higher concentration of IgE in the specimen. The test results are analyzed with Phadia Microarray Image Analysis (MIA) Software and ISAC Standardized Units for specific IgE (ISU-E) are estimated.
CLINICAL UTILITY
Most allergic patients are multi-sensitized with positive test results to numerous allergens and the true cause of symptoms can be difficult to identify due to an inconclusive medical history regarding the role of different allergens and reactions.
ImmunoCAP ISAC can help to:
· Shed light on the real sensitization profile of multi-sensitized patients.
· Reveal potential risk for severe food-related reactions.
· Identify the IgE antibody profile in patients with unsatisfactory response to treatment.
· Assess patients with idiopathic anaphylaxis.
ImmunoCAP ISAC can also reveal unexpected sensitizations or help to rule out allergy by delivering IgE results for a broad spectrum of allergens.
IMMUNOCAP ISAC slgE 112
Product Characteristics 2
SAMPLe INFORMATION
30 µl of serum and heparin plasma. Venous or capillary blood can be used.
MeASURING RANGe
0.3 – 100 ISU-E.
PReCISION
The mean coefficient of variation per component within and between assays on ImmunoCAP sIgE 112 is <25 % covering the measuring range. Sample level Mean Coefficient of Variation (CV %)
Within Between
0.3 – 1 ISU-E 14 8
1 – 15 ISU-E 10 6
> 15 ISU-E 9 6
Figure 1. Total CV for four samples (covering 105 components) assayed in 3 replicates in 17 runs over a 4-week period on one chip lot.
IMMUNOCAP ISAC slgE 112
Product Characteristics 3
Figure 2. Examples of dilutions for Bet v 1 and Phl p 5.
SeNSITIVITY
The Limit of Detection is < 0.3 ISU-E for all components.
SPeCIFICITY
· The cross-reactivity with other human Immunoglobulins is non-detectable at physiological concentrations of IgA, IgD, IgG and IgM.
· No interferences of Total IgE concentratios up to at least 10 000 kU/l.
· No interferences from 5 % hemolysis or at very high concentrations of triglycerides up to 12 mg/ml.
LINeARITY
ImmunoCAP ISAC sIgE 112 meets the requirements on linearity between dilutions of samples and the calibration curve. See Figure 2 for representative examples of dilution series.
IMMUNOCAP ISAC slgE 112
Product Characteristics 4
HeALTHY CONTROLS
One hundred healthy non-atopic individuals were tested and in total 11 200 values have been analysed of which 99,991 % were undetectable. One single sample was identified with IgE abs to two components at 0.3 ISU-E and to one at 0.4 ISU-E.
COMPARISON WITH IMMUNOCAP ALLeRGeN COMPONeNT
Representative examples of comparison between recombinant ImmunoCAP allergen components and the corresponding component on ISAC.
Figure 3. Left) Negative test result for one healthy control showing only guide dots. Right) Summary of test results for 100 healthy controls x 112 components.
Figure 4. Comparisons of ImmunoCAP allergen component versus ISAC components. (Negative values set to 0.15).
Alt a 1
Can f 1
Phl p 1
Ara h 2
Der p 2
Pru p 3
0 0,05
0,1 0,15
0,2 0,25
0,3 0,35
0,4 0,45
0 20 40 60 80 100
IgE
conc
. for
112
com
pone
nts
(ISU-
E)
Samples
11197 results = 0
IMMUNOCAP ISAC slgE 112
Product Characteristics 5
ORDeRING INFORMATION
eXAMPLeS OF PUBLISHeD IMMUNOCAP ISAC 103 STUDIeS
· The ImmunoCAP ISAC molecular allergology approach in adult multi-sensitized Italian patients with respiratory symptoms. Melioli G, Bonifazi F, Bonini S, Maggi E, Mussap M, Passalacqua G, Rossi ER, Vacca A, Canonica GW; on behalf of the Italian Board for ISAC (IBI). Clin Biochem. 2011Aug; 44(12): 1005–1011
· Use of allergen components begins a new era in pediatric allergology. Magnus P. Borres, Motohiro Ebisawa & Philippe A. Eigenmann. Pediatric Allergy and Immunology 2011; 22; 454–461
· Microarray of allergenic component-based diagnosis in food allergy. Maria L. Sanz, Ana B. Blázquez and Blanca E. Garcia. Current Opinion in Allergy and Clinical Immunology 2011; 11(3): 204–209
· A new tool in the field of in-vitro diagnosis of allergy: preliminary results in the comparison of ImmunoCAP 250 with the ImmunoCAP ISAC. Romy Gadisseur*, Jean-Paul Chapelle and Etienne Cavalier. Clin Chem Lab Med 2011; 49(2): 277–280
· Comparison of conventional and component resolved diagnostics by two different methods (Advia-Centaur/Microarray-ISAC) in pollen allergy. M.T. Lizaso, B.E. García, A.I. Tabar, E Lasa, S. Echechipía, M.J. Álvarez, M. Anda, B. Gómez. Ann. Allergy Asthma Immunol. 2011 Jul; 107(1): 35–41
· Molecular diagnosis in allergy. Sastre J. Clin Exp Allergy. 2010; 40: 1442–1460
ImmunoCAP ISAC 112
Product Size Art.No.
ImmunoCAP ISAC sIgE 112 20 tests 81-1011-01
ImmunoCAP ISAC sIgG4 112 20 tests 81-1012-01
ImmunoCAP ISAC Starter Pack 81-1010-01
ImmunoCAP Washing Solution 6 x 1 l 10-9422-01
thermoscientific.com/phadia
Head office Sweden +46 18 16 50 00 Austria +43 1 270 20 20 Belgium +32 2 749 55 15 Brazil + 55 11 3345 5050 China +86 25 8960 5700 Czech Republic +420 220 518 743 Denmark +45 70 23 33 06 Finland +358 9 8520 2560 France +33 1 61 37 34 30 Germany +49 761 47 8050 India +91 11 4610 7555/56 Italy +39 02 64 163 411 Japan +81 3 5365 8332 Korea +82 2 2027 5400 Norway +47 21 67 32 80 Portugal +351 21 423 5350 South Africa +27 11 792 6790 Spain +34 935 765 800 Sweden +46 18 16 50 00 Switzerland +41 43 343 4050 Taiwan +886 2 2516 0925 The Netherlands +31 30 602 37 00 United Kingdom/Ireland +44 1 908 769 110 USA +1 800 346 4364 Other countries +46 18 16 50 00