Upload
biomedicalcombr
View
218
Download
0
Embed Size (px)
Citation preview
7/27/2019 pressure-injectable-cvc-ifu.pdf
1/20
Venous Access | Critical Care
Pressure InjectableCVC Product
7/27/2019 pressure-injectable-cvc-ifu.pdf
2/20
Arrow International, Inc.
2009 Arrow International, Inc. All rights reserved. Printed in the USA.
An issued or revision date for these instructions is included for user information. In the event two years haveelapsed between this date and product use, the user should contact Arrow International, Inc. to see if additionalproduct information is available.
Issued Date: September 2009Rx only.
7/27/2019 pressure-injectable-cvc-ifu.pdf
3/20
PRESSURE INJECTABLE CVC | iii
Table of Contents
Product Description 1Indications / Contraindications . . . . . . . . . . . . . . . . . . . . . . . . 1
Central Venous Catheter Warnings and Precautions 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Catheter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Guidewire / SWG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Tissue Dilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Pressure Injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Possible Complications . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Accessory Component Instructions 4
Arrow Advancer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Arrow Raulerson Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Catheter Clamp and Fastener . . . . . . . . . . . . . . . . . . . . . . . . . 5
Catheter Stabilization Device . . . . . . . . . . . . . . . . . . . . . . . . . 6
Dressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Echogenic Needle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Guidewire / SWG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Luer-Activated Connector . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Maximal Barrier Drape . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Protected Needle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
SharpsAway II Locking Disposal Cup . . . . . . . . . . . . . . . . . . . . . 10
Staple Anchoring Device . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Pre-CVC Insertion & Patient Assessment Activities 10
Procedural Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Preparing for Catheter Insertion . . . . . . . . . . . . . . . . . . . . . . . 11
Prep Puncture Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Flush Catheter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
For reference literature concerning patient assess-
ment, clinician education, insertion techniques
and potential complications associated with this
procedure refer to Arrow International, Inc. website:www.arrowintl.com
For convenience, procedural and general Warnings
and Precautions are listed at the beginning of
the instructions. Please review all content before
performing the procedure.
7/27/2019 pressure-injectable-cvc-ifu.pdf
4/20
iv | PRESSURE INJECTABLE CVC
Catheter Insertion Instructions 11
Gain Initial Venous Access . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Verify Venous Access. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Guidewire / SWG Insertion . . . . . . . . . . . . . . . . . . . . . . . . 12
Catheter Advancement . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Complete Catheter Insertion . . . . . . . . . . . . . . . . . . . . . . . . . 13
Documentation 13
Pressure Injection Instructions 14
Care and Maintenance 14
Dressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Maintain Catheter Patency . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Catheter Removal 15
7/27/2019 pressure-injectable-cvc-ifu.pdf
5/20
| 1
Pressure InjectableCentral Venous Catheter (CVC) Product
Product DescriptionTe Arrow Pressure Injectable CVC is a central venous catheter
manufactured with medical grade, flexible polyurethane. A Pressure
Injectable CVC may vary from two to four noncommunicating
lumens. Te catheter has a soft Blue Flexip that is more pliable
than the catheter body. Lumens are connected to separate color-
coded extension lines which have hubs on the end that are standard
Luer-Lock. Centimeter markings referenced from the tip areplaced along length of indwelling catheter body to facilitate proper
positioning. Te kit components assist the clinician in maintaining
maximal sterile barrier precautions (where provided).
Indications:Te Arrow CVC is indicated to provide short-term (< 30 days)
central venous access for treatment of diseases or conditions requiring
central venous access including, but not limited to:
multiple infusions of fluids, medications, or chemotherapy
infusion of fluids that are hypertonic, hyperosmolar, or have
divergent pH values
frequent blood sampling or blood/blood component infusions
infusion of incompatible medications
central venous pressure monitoring
lack of usable peripheral IV sites
replacement of multiple peripheral sites for IV access
injection of contrast media
When used for pressure injection of contrast media, do not
exceed the maximum indicated flow rate for each catheter
lumen. Te maximum pressure of power injector equipment
used with pressure injectable CVC may not exceed 400 PSI.
Contraindications:None known. See additional labeling for product specific
contraindications.
Central Venous Catheter
Warnings and Precautions:
Do not place the catheter into or allow it to remain in the rightatrium or right ventricle (refer to Figure 1).
Figure 1
General Warnings and PrecautionsWarnings:
1. Read all package insert warnings, precautions, andinstructions prior to use. Failure to do so may result in
severe patient injury or death.
2. Practitioners must be aware of complications associated
with central vein catheters including but not limited to:
cardiac tamponade secondary to vessel wall, atrial or
ventricular perforation, pleural and mediastinal injuries,
air embolism, catheter embolism, catheter occlusion,
thoracic duct laceration, bacteremia, septicemia,thrombosis, inadvertent arterial puncture, nerve damage,
hematoma, hemorrhage, and dysrhythmias.
3. Do not place Central Venous Catheter (CVC) or
Peripherally Inserted Central Catheter (PICC) into or allow
them to remain in the right atrium or right ventricle.
X-ray exam or other method in compliance with hospital/
institutional protocol must show catheter tip located in
right side of mediastinum in the Superior Vena Cava (SVC)
above its junction with right atrium and parallel to vesselwall and its distal tip positioned at a level above either
azygos vein or carina of the trachea, whichever is better
visualized. Although cardiac tamponade secondary
to pericardial effusion is uncommon, there is a high
mortality rate associated with it. Improper advancement
of guidewire into the heart has also been implicated in
causing cardiac perforation and tamponade. For femoral
vein approach, catheter should be advanced into vessel
so catheter tip lies parallel to vessel wall and does notenter right atrium.
7/27/2019 pressure-injectable-cvc-ifu.pdf
6/20
2 |
4. Ensure catheter tip has not entered the heart or no longerlies parallel to vessel wall by performing an x-ray exam orother method in compliance with hospital/institutionalprotocol. If catheter position has changed, immediatelyre-evaluate.
5. Practitioners must be aware of the potential forentrapment of guidewire by any implanted device incirculatory system (i.e., vena cava filters, stents). Reviewpatients history before catheterization procedure toassess for possible implants. Care should be takenregarding length of guidewire inserted. It is recommendedthat if patient has a circulatory system implant, catheterprocedure be done under direct visualization to minimizethe risk of guidewire entrapment.
6. Choose appropriate sized catheter for size of vessel to becannulated.
7. Catheter tip must be located in central circulation whenadministering > 10% glucose solution, total parenteralnutrition, continuous vesicant therapy, infusates withpH less than 5 or greater than 9, and infusates with anosmolality above 600 mOsm/L, or any medication knownto be irritating to vessels proximal to the vena cava.
8. Do not leave open needles or uncapped, unclampedcatheters in central venous puncture site. Air embolismcan occur with these practices.
9. Use only securely tightened Luer-Lock connectionswith any Central Venous Access Device (CVAD) to guardagainst inadvertent disconnect.
10. Use Luer-Lock connectors to help guard against airembolism and blood loss.
11. Pulsatile flow is usually an indicator of inadvertent
arterial puncture.
Cautions:1. The product is designed for single use only.
2. Do not resterilize or reuse.
3. Do not use if package has been previously opened ordamaged.
4. Do not alter the catheter, guidewire, or any other kit/setcomponent during insertion, use, or removal (except as
instructed).5. Procedure must be performed by trained personnel well
versed in anatomical landmarks, safe technique, andpotential complications.
6. Assess patient for heparin sensitivity. Heparin-InducedThrombocytopenia (HIT) has been reported with use ofheparin flush solutions.
7. Do not routinely apply prophylactic topical antimicrobialor antiseptic ointment or cream to the insertion site of
central venous catheters because of the potential risk topromote fungal infections and antimicrobial resistance.
8. Temporarily shut off remaining port(s) through whichsolutions are being infused before blood sampling.
9. Blood aspirate color is not always a reliable indicator ofvenous access.
10. Do not reinsert needle into introducer catheter to reducerisk of catheter embolism.
11. Maintain insertion site with regular meticulous redressingusing aseptic technique.
12. Activate safety feature of safety scalpel (if provided) whennot in use to minimize the risk of sharps injury.
13. Perform hand hygiene before and immediately after all
clinical procedures and before and after donning andremoval of gloves.
14. Properly handle and dispose of sharps in sharps containerin accordance with US OSHA or other governmentalstandards for blood borne pathogens and/or hospital/institutional policy.
15. Hands must remain behind the needle at all times duringuse and disposal.
16. Use universal blood and body-fluid precautions in thecare of all patients due to the risk of exposure to HIV(Human Immunodeficiency Virus) or other blood bornepathogens.
Catheter Warnings and PrecautionsWarnings:1. Only utilize catheters indicated for high pressure injection
applications for such applications. Utilizing catheters notindicated for high pressure applications can result ininter-lumen crossover or rupture with potential for injury.
2. Do not apply excessive force in placing or removingcatheter. Excessive force can cause catheter breakage. Ifplacement or withdrawal cannot be easily accomplished,an x-ray should be obtained and further consultationrequested.
3. Do not secure, staple, and/or suture directly to outside
diameter of catheter body or extension lines to minimizethe risk of cutting or damaging the catheter or impedingcatheter flow. Secure only at indicated stabilizationlocations.
4. Do not cut catheter to alter catheter length.
5. Do not attach catheter clamp and fastener (whereprovided) until either guidewire or placement wire isremoved.
6. Do not use scissors to remove dressing to minimize therisk of cutting catheter.
7/27/2019 pressure-injectable-cvc-ifu.pdf
7/20
PRESSURE INJECTABLE CVC | 3
7. Open catheter clamp prior to infusion through lumen
to minimize the risk of damage to extension line from
excessive pressure.
8. Practitioners should remove slide clamp(s), where
provided, when not in use. Slide clamp(s) may beinadvertently removed and aspirated by children or
confused adults.
9. Do not routinely replace central venous catheters solely
for the purpose of reducing incidence of infection.
10. Do not use guidewire techniques to replace catheters in
patient suspected of having catheter-related infection.
11. Residual catheter track remains an air entry point until
completely sealed, occlusive dressing should remain inplace for at least 24 - 72 hours dependent upon amount
of time catheter was indwelling.
Cautions:1. Check ingredients of prep sprays and swabs before using.
Some disinfectants used at catheter insertion site contain
solvents which can attack the catheter material. Alcohol
and acetone can weaken the structure of polyurethane
materials. These agents may also weaken the adhesivebond between catheter stabilization device and skin.
Acetone: Do not use acetone on catheter surface.
Alcohol: Do not use alcohol to soak catheter surface or
to restore catheter patency.
Take care when instilling drugs containing high
concentration of alcohol. Allow insertion site to dry
completely prior to applying dressing.
2. Ensure catheter patency prior to injection. Donot use syringes smaller than 10 mL (a fluid filled
1 mL syringe can exceed 300 psi), to minimize the risk of
intraluminal leakage or catheter rupture.
3. Remove catheter clamp and fastener (where provided)
prior to attempting a catheter exchange procedure.
4. Do not exert excessive force while removing the catheter,
to minimize the risk of catheter breakage.
5. Continuously monitor indwelling catheter for: desired flow rate
security of dressing
adherence of stabilization device to skin and
connection to catheter
correct catheter position; use centimeter markings to
identify if catheter position has changed
secure Luer-Lock connection(s)
6. Minimize catheter manipulation throughout procedure tomaintain proper catheter tip position.
7. Inject a small amount of radiopaque dye to locatecatheter tip if difficulty is encountered in visualizing thecatheter tip.
Guidewire / SWG Warnings and PrecautionsWarnings:1. Do not cut guidewire to alter length.
2. Do not withdraw guidewire against needle bevel tominimize the risk of possible severing or damaging ofguidewire.
3. Do not use excessive force when introducing guidewireor tissue dilator as this can lead to vessel perforation and
bleeding.4. Passage of guidewire into the right heart can cause
dysrhythmias, right bundle branch block, and aperforation of vessel, atrial, or ventricular wall.
5. Do not apply undue force on guidewire to minimize therisk of possible breakage.
6. Do not apply excessive force in removing guidewire orcatheter. If withdrawal cannot be easily accomplished, avisual image should be obtained and further consultationrequested.
7. Do not cut guidewire with scalpel.
Position cutting edge of scalpel away from guidewire.
Activate safety feature of safety scalpel (whereprovided) once cutaneous puncture site is enlarged,to minimize the risk of cutting the guidewire (refer toFigure 28).
8. Do not aspirate with guidewire in place or air will enter
syringe.Caution:1. Maintain a firm grip on guidewire at all times. Keep
sufficient guidewire length exposed at hub for handlingpurposes. A non-controlled guidewire can lead to wireembolism.
2. If resistance is encountered when attempting to removeguidewire after catheter placement, guidewire maybe kinked about tip of catheter within vessel (refer to
Figure 29).
Tissue Dilator WarningsWarnings:1. Do not leave tissue dilator in place as an indwelling
catheter. Leaving tissue dilator in place puts patient at riskfor possible vessel wall perforation.
2. Do not use excessive force when introducing guidewire
or tissue dilator as this can lead to vessel perforation andbleeding.
7/27/2019 pressure-injectable-cvc-ifu.pdf
8/20
4 |
Pressure Injection Warnings and PrecautionsWarnings:1. Assess each patient for appropriateness of a pressure
injection procedure. Pressure injection procedures mustbe performed by trained personnel well versed in safetechnique and potential complications.
2. Obtain a visual image to confirm catheter tip positionprior to each pressure injection.
3. Ensure patency of each lumen of catheter prior topressure injection to minimize the risk of catheter failureand/or patient complications.
4. Discontinue pressure injections at first sign ofextravasation or catheter deformation. Followhospital/institutional protocol for appropriate medicalintervention.
Cautions:1. Do not exceed ten (10) injections or catheters maximum
recommended flow rate located on product labeling andcatheter luer hub to minimize the risk of catheter failureand/or tip displacement.
2. Warm contrast media to body temperature prior topressure injection to minimize the risk of catheter failure.
3. Pressure limit settings on injector equipment maynot prevent over pressurizing an occluded or partiallyoccluded catheter.
4. Use appropriate administration set tubing betweencatheter and pressure injector equipment to minimize the
risk of catheter failure.5. Follow the contrast media manufacturers specified
instructions for use, contraindications, warnings, andprecautions.
Possible Complications (but not limited to): cardiac tamponade secondary to vessel wall, atrial or
ventricular perforation
pleural injury mediastinal injury
air embolism nerve injury
catheter embolism thoracic duct laceration
bleeding / hemorrhage occlusion
bacteremia septicemia
thrombosis inadvertent arterial puncture
hematoma dysrhythmias
fibrin sheath formation
exit site infection vessel erosion catheter tip malposition
Accessory Component InstructionsReview the list of components that will be utilized before
beginning the Arrow Pressure Injectable CVC insertion
procedure. Kits / Sets may not contain all accessory
components detailed in this section. Become familiar withinstructions for each individual component before beginning
the actual insertion procedure.
Arrow Advancer:Arrow Advancer is used to straighten J Tip of Spring-Wire Guide (SWG) for
introduction of guidewire into Arrow Raulerson Syringe or a needle.
Using your thumb, straighten the J by retracting SWG into theArrow Advancer (refer to Figures 2 and 3).
Figure 2
Figure 3
When tip is straightened, SWG is ready for insertion. Centimetermarks on SWG are referenced from J ip: one band indicates10 cm, two bands 20 cm, and three bands 32 cm.
Place tip of Arrow Advancer with J retracted into the hole
in rear of Arrow Raulerson Syringe plunger (refer to Figure 4)or introducer needle.
7/27/2019 pressure-injectable-cvc-ifu.pdf
9/20
PRESSURE INJECTABLE CVC | 5
Simple Straightening Tube Technique:
Te straightening tube portion of the Arrow Advancer can bedisconnected from the unit and used separately.
Separate Arrow Advancer tip or straightening tube from blueArrow Advancer unit.
Prepare for insertion by sliding plastic tube over J to straighten,if J ip portion of SWG is used.
Advance SWG in routine fashion to desired depth.
Arrow Raulerson Syringe:Arrow Raulerson Syringe is used in conjunction with Arrow Advancer
for SWG insertion.
Warning: Do not withdraw guidewire against needle bevel to
minimize the risk of possible severing or damaging of
guidewire.
Warning: Do not aspirate with guidewire in place or air may enter
syringe.
Caution: Blood aspirate color is not always a reliable indicator of
venous access.Caution: Do not reinfuse blood to minimize the risk of blood leakage
from rear (cap) of syringe.
Caution: Maintain firm grip on guidewire at all times. Keep sufficient
guidewire length exposed at hub for handling purposes. A
non-controlled guidewire can lead to wire embolism.
Insert introducer needle attached to ArrowRaulerson Syringeinto vessel and aspirate.
Vessel may be prelocated with a 22 Ga. locater needle.
For pressure wave form transduction, a separately packagedtransduction probe is available.
Straighten SWG J.
Advance SWG through ArrowRaulerson Syringe into vessel todesired depth.
Hold SWG in place and remove introducer needle and Arrow
Raulerson Syringe.
Catheter Clamp and Fastener:Catheter Clamp and Fastener are used to secure catheter when an additional
securement site other than catheter hub is required for catheter stabilization.
After SWG has been removed and necessary lines have beenconnected or locked, spread wings of rubber clamp and position
on catheter making sure catheter is not moist, as required, tomaintain proper tip location (refer to Figure 7).
Figure 4 Advance SWG into Arrow Raulerson Syringe approximately
10 cm until it passes through syringe valves (refer to Figure 5) orinto introducer needle.
Figure 5
Raise thumb and pull Arrow Advancer approximately 4 - 8 cmaway from Arrow Raulerson Syringe or introducer needle.Lower thumb onto Arrow Advancer and while maintaining afirm grip on SWG, push assembly into syringe barrel to furtheradvance SWG. Continue until SWG reaches desired depth (referto Figure 6).
Figure 6
7/27/2019 pressure-injectable-cvc-ifu.pdf
10/20
6 |
Figure 7
Snap rigid fastener onto catheter clamp (refer to Figure 8).
Figure 8
Secure catheter to patient by stapling or suturing catheter clampand fastener together to skin, using side wings to minimize therisk of catheter migration (refer to Figure 9).
Figure 9
Catheter Stabilization Device:A catheter stabilization device should be used in accordance with manufacturersinstructions for use.
Caution: Minimize catheter manipulation throughout procedure tomaintain proper catheter tip position.
Cleanse and prep anticipated dressing site per hospital/institutional protocol.
Apply skin prep (where provided) to coat skin and maximizeadherence.
Allow to dry thoroughly.
Position catheter stabilization device appropriate distance fromcatheter insertion site to permit ease of insertion site care andmaintenance.
Secure catheter by using the primary suture hub.
Place suture hub wings over posts and press down (refer toFigure 10). Snap retainer wings to closed position to securesuture hub (refer to Figure 11).
Figure 10
Figure 11
Remove paper backing from one half of catheter stabilizationdevice pad and press onto dry, prepared skin. Repeat process forother half.
Complete sterile insertion site dressing according to establishedhospital/institutional protocol.
Document dressing application on patients chart.
Replace dressing per hospital/institutional protocol. Catheter
stabilization device should be replaced at least every 7 days toensure maximum adherence.
7/27/2019 pressure-injectable-cvc-ifu.pdf
11/20
PRESSURE INJECTABLE CVC | 7
Dressing:A transparent dressing should be used in accordance with manufacturers
instructions for use.
Prepare site. Allow all preps to dry completely.
Peel liner from dressing to expose adhesive.
Adhere center of transparent window over insertion site, whileholding notched portion off the skin (refer to Figure 12).
Figure 12
Overlap softcloth tabs under catheter to form a tight seal aroundcatheter hub and lumens (refer to Figure 13).
Figure 13
Press dressing into place.
Slowly remove frame while smoothing down dressing edges.
Smooth dressing from center toward edges, use firm pressure toenhance adhesion (refer to Figure 14).
Figure 14
Use sterile tape strips to secure hub, lumens, and/or tubing (refer
to Figure 15).
Figure 15
Label dressing according to protocol.
Refer to individual manufacturers instructions for more information
and specific detailed instructions for dressing removal (not included).
Echogenic Needle:
An echogenic needle is used to provide greater needle visibility under
ultrasound. The needle tip is enhanced for approximately 1 cm for clinician toidentify exact needle tip location when puncturing the vessel under ultrasound.
8 |
7/27/2019 pressure-injectable-cvc-ifu.pdf
12/20
8 |
Guidewire / SWG:
Kits/Sets are available with a variety of guidewires/SWGs. Guidewires are
provided in different diameters, lengths, and tip configurations for specific
insertion techniques. Become familiar with the guidewire(s) to be used with the
specific technique before beginning the actual CVC insertion procedure.
Image guidance may be used to gain initial venous access.
Luer-Activated Connector:
A Luer-Lock connector is use to minimize the risk of accidental disconnects and
potential for air embolism.
Caution: Do not use needles in connector unless specifically indicated.
Caution: Do not cap connector unless specifically indicated.
Caution: Check for clamping requirement to prevent fluid movement
prior to connector disconnect and use.
General Guidelines for Connector Use:
Use aseptic technique. Remove connector from packaging.
Inspect for integrity.
If applicable:
prime in accordance with hospital/institutional protocol
expel air
Disinfect desired extension set or venous access device; makesure to rotate swab to ensure cleaning of septum and threads.
Attach connector to desired extension set or venous access device.
Secure connector. If using a rotating luer connector, make surethat the spin collar is tightened.
Position connector to achieve the most comfortable position onthe patients skin.
Access: Swab connector with desired disinfectant in accordancewith hospital/institutional and connector manufacturer protocol.
Utilize only fully primed syringe or administration set.
Disconnect from venous access device or administration set,grasp connector/cap and twist until loose.
Flush after each use with normal saline or in accordance withhospital/institutional protocol.
Repeat above steps for subsequent connections.
Refer to individual manufacturers instructions for specific connectordetails, including priming volumes, dead space, and flow rates.
Maximal Barrier Drape:
Drape(s) provide a maximal sterile barrier. Follow the CDC Category 1A
Recommendation.
Unfold the Maximal Barrier Drape:
Peel off fenestration backing (refer to Figure 16).
Figure 16
Position fenestration over intended insertion site (refer to
Figure 17).
Figure 17
Unfold width (refer to Figure 18).
b
a
Figure 18
PRESSURE INJECTABLE CVC | 9
7/27/2019 pressure-injectable-cvc-ifu.pdf
13/20
PRESSURE INJECTABLE CVC | 9
Unfold towards head (refer to Figure 19).
Figure 19
Unfold towards hand (refer to Figure 20).
Figure 20
Perform sterile procedure (refer to Figure 21).
Figure 21
Removal procedure: ear along seam (refer to Figure 22).
TO
REMOVEDRAPE
TEARALONG
SEA
M
Figure 22
Protected Needle:See individual manufacturers instructions for product use, when used as a singleproduct and not as a kit component.
Warning: Hands must remain behind needle at all times during useand disposal.
Caution: Make sure all needles are used in accordance with OSHA and
hospital/institutional safety protocols.
Caution: Do not attempt to override or defeat the safety lockingmechanism of a protected needle.
Caution: Discard in an approved sharps collector in accordance withapplicable regulations and hospital/institutional policy.
General Guidelines for Protected Needle Use: Aspirate medication into syringe using aseptic technique.
If necessary to transport filled syringe to point of administration,use safe, passive recapping technique to cover needle beforetransport to point of use. In accordance with OSHA standards,such recapping must be accomplished by a one-handedtechnique, i.e., do not hold needle shield during recappingprocess.
Administer injection following established technique.
Immediately activate needle protection device upon withdrawalfrom patient by pushing lever arm completely forward until
needle tip is fully covered (see Figure 23).
Figure 23
10 |
7/27/2019 pressure-injectable-cvc-ifu.pdf
14/20
10 |
Visually confirm lever arm has fully advanced and needle tip iscovered. If unable to activate, discard immediately into approvedsharps collector.
Activation of protective mechanism may cause minimal splatter
of fluid that may remain on needle after injection. After single use, discard in an approved sharps collector in
accordance with applicable regulations and hospital/institutionalpolicy. For greatest safety, use a one-handed technique andactivate away from self and others.
SharpsAway II Locking Disposal Cup:The SharpsAway II Locking Disposal Cup is used for disposal of needles(15 Ga. - 30 Ga.).
Using one-handed technique, firmly push needles into disposalcup holes (refer to Figure 24).
Secureneedles for
final disposal
Figure 24
Once placed into disposal cup, needles will be automaticallysecured in place so that they cannot be reused.
Discard entire cup, at completion of procedure, into an approvedsharps container.
Caution: Do not attempt to remove needles that have been placedinto SharpsAway II Locking Disposal Cup. These needlesare secured in place. Damage may occur to needles if theyare forced out of disposal cup.
Where provided, a foam SharpsAway system may be utilized bypushing needles into foam after use.
Caution: Do not re-use needles after they have been placed into thefoam SharpsAway system. Particulate matter may adhereto needle tip.
Staple Anchoring Device:Staples offer an alternative for catheter stabilization.
Warning: Do not secure, staple, and/or suture directly to outsidediameter of catheter body or extension lines to minimizethe risk of cutting or damaging the catheter or impeding
catheter flow. Secure only at indicated stabilizationlocations.
Position thumb and index finger of dominant hand on indentedsurface of staple anchoring device.
Pass staple point through eye of catheter suture hub (refer toFigure 25).
Figure 25
ent skin and position with hub eye between staple opening.
Firmly squeeze anchoring device together to close staple andsecure catheter to skin (refer to Figure 26).
Figure 26
Repeat procedure through other suture eyes, if applicable.
Discard anchoring device upon completion.
Pre-CVC Insertion &Patient Assessment Activities
A Suggested ProcedureWarning: Read all package insert warnings, precautions, and
instructions prior to use. Failure to do so may result in severepatient injury or death.
Clinical assessment of patient must be completed to ensure no contraindicationsexist e.g. allergies. This device is not recommended for use in the presence ofdevice related infections or previous/current thrombosis.
Perform hand hygiene as required.
A procedural checklist is included in many Arrow products. Check individualproduct labeling to see if one is included.
PRESSURE INJECTABLE CVC | 11
7/27/2019 pressure-injectable-cvc-ifu.pdf
15/20
PRESSURE INJECTABLE CVC | 11
Procedural Pause:1. Verify physician order:
Confirm correct patient.
Confirm correct diagnosis. Confirm correct procedure.
Physician order must include post placement assessment ofcatheter tip placement (direct visualization technique or othermethod in compliance with hospital/institutional protocol).
2. Educate patient: Explain procedure to patient. Make sureinformation is presented with respect to patients level ofunderstanding, culture, and language.
3. Have informed consent signed, if required.4. Identify insertion vein:
Identify appropriate vein for insertion.
Use direct visualization technologies, e.g. ultrasound orfluoroscopy, if available; and assess vein health.
5. Position patient as appropriate for insertion site:
Subclavian or jugular approach: Place patient in slightrendelenburg position as tolerated to reduce the risk of air
embolism and enhance venous filling. If femoral approach is used, place patient in supine position.
6. Prepare work area.
Preparing for Catheter Insertion: Perform hand hygiene as required:
before and immediately after all clinical procedures
before and after donning and removal of gloves
Caution: Use universal blood and body-fluid precautions in the careof all patients due to the risk of exposure to HIV (HumanImmunodeficiency Virus) or other blood borne pathogens.
Caution: Properly handle and dispose of sharps in sharps container inaccordance with US OSHA or other governmental standardsfor blood borne pathogens and/or hospital/institutionalpolicy.
Clinicians should use sterile technique, maximal sterile barrierprecautions throughout the procedure, and dress in protective
clothing: mask eye protection
sterile gown sterile gloves
hair cover
Prep Puncture Site:1. Prep puncture site with appropriate antiseptic/agent.
2. Drape puncture site.
See unfolding instructions for Maximal Barrier Drape (where provided) underAccessory Component Instructions section.
3. Perform skin wheal using desired needle (25 Ga. or 22 Ga.) withlocal anesthetic.
4. Dispose of needle.
In kits where provided, the SharpsAway II Locking Disposal
Cup is used for disposal of needles (15 Ga. - 30 Ga.).See SharpsAway II Locking Disposal Cup under Accessory ComponentInstructions section.
Caution: Do not attempt to remove needles that have been placedinto SharpsAway II Locking Disposal Cup. These needlesare secured in place. Damage may occur to needles if theyare forced out of disposal cup.
Caution: Do not re-use needles after they have been placed into thefoam SharpsAway system. Particulate matter may adhere
to needle tip.
Flush Catheter:1. Flush each lumen with sterile saline solution, to establish patency
and prime lumen(s).
2. Clamp or attach injection site cap(s) to extension line(s) tocontain saline within lumen.
3. Leave distal extension line uncapped for SWG passage.
Warning: Do not cut catheter to alter length.
Catheter Insertion Instructions
Gain Initial Venous Access:Warning: Do not leave open needles or uncapped, unclamped
catheters in central venous puncture site. Air embolism canoccur with these practices.
1. Insert introducer needle or catheter/needle with attached syringeor Arrow Raulerson Syringe (where provided) into vein and
aspirate. NOTE: If larger introducer needle is used, vessel may be pre-located with
22 Ga. locater needle and syringe.
See Arrow Raulerson Syringe under Accessory Component Instructions section.
2. Remove locater needle.
Caution: Blood aspirate color is not always a reliable indicator ofvenous access.
Caution: Do not reinsert needle into introducer catheter to reduce
risk of catheter embolism.
Verify Venous Access:Because of the potential for inadvertent arterial placement, one of the followingtechniques should be utilized to verify venous access:
1. Central Venous Waveform: Insert fluid primed blunt tip pressure transduction probe into
rear of plunger and through valves of the Arrow RaulersonSyringe and observe for central venous pressure waveform.
Remove transduction probe if using Arrow RaulersonSyringe.
12 |
7/27/2019 pressure-injectable-cvc-ifu.pdf
16/20
12 |
Observe for central venous pressure waveform obtainedby a calibrated pressure transducer attached directly to theintroducer needle/catheter (refer to Figure 27).
20
0
Figure 27
2. Pulsatile Flow (if hemodynamic monitoring equipment is notavailable):
Use transduction probe to open syringe valving system ofArrow Raulerson Syringe and observe for pulsatile flow.
Disconnect syringe from needle and observe for pulsatileflow.
Warning: Pulsatile flow is usually an indicator of inadvertent arterial
puncture.
Guidewire / SWG Insertion:1. Straighten J of SWG by using a straightening tube or
Arrow Advancer (where provided).
See Arrow Advancer under Accessory Component Instructions section.
2. Insert tip of SWG into plunger of Arrow Raulerson Syringe orinto introducer needle.
3. Advance SWG through Arrow Raulerson Syringe
or introducer needle into vein to desired depth. Advancement of J ip through Arrow Raulerson Syringe
may require a gentle rotating motion.
Advance SWG until triple band mark reaches rear of ArrowRaulerson Syringe plunger.
4. Use centimeter markings on SWG as a reference to assist indetermining how much guidewire has been inserted.
NOTE: If guidewire has three sets of markings, they will be located as
follows: One band - 10 cm from J Tip
Two bands - 20 cm from J Tip
Three bands - 32 cm from J Tip
NOTE: When guidewire is used in conjunction with the Arrow Raulerson
Syringe (fully aspirated) and a 2-1/2 inch introducer needle, the following
positioning references can be made:
20 cm mark entering back of plunger = guidewire tip is at end of needle
32 cm mark entering back of plunger = guidewire tip is approximately10 cm beyond end of needle
Caution: Maintain firm grip on guidewire at all times. Keep sufficientguidewire length exposed at hub for handling purposes. Anon-controlled guidewire can lead to wire embolism.
Warning: Do not aspirate with guidewire in place or air may enter
syringe.Caution: Do not reinfuse blood to minimize the risk of blood leakage
from rear (cap) of syringe.
Warning: Do not withdraw guidewire against needle bevel tominimize the risk of possible severing or damaging ofguidewire.
5. Remove introducer needle and Arrow Raulerson Syringe (orcatheter) while holding SWG in place.
6. Use centimeter markings on SWG to adjust indwelling lengthaccording to desired depth of indwelling catheter placement.
7. Enlarge cutaneous puncture site with cutting edge of scalpel,where provided, positioned away from SWG.
Warning: Do not cut guidewire to alter length.
Warning: Do not cut guidewire with scalpel.
Position cutting edge of scalpel away from guidewire.
Activate safety feature of safety scalpel (where provided)
once cutaneous puncture site is enlarged, to minimize therisk of cutting the guidewire (refer to Figure 28).
Figure 28
8. Use tissue dilator to enlarge puncture site as required.
Warning: Do not leave tissue dilator in place as an indwelling catheter
to minimize the risk of possible vessel wall perforation.
Catheter Advancement:1. Tread tip of catheter over SWG. Sufficient SWG length must
remain exposed at hub end of catheter to maintain a firm grip onSWG.
2. Grasping near skin, advance catheter into vein with slighttwisting motion.
Warning: Do not attach catheter clamp and fastener (where provided)until either guidewire or placement wire is removed.
PRESSURE INJECTABLE CVC | 13
7/27/2019 pressure-injectable-cvc-ifu.pdf
17/20
|
3. Using centimeter marks on catheter as positioning referencepoints, advance catheter to final indwelling position.
NOTE: Centimeter marking symbology is referenced from catheter tip.
numerical: 5, 15, 25, etc.
bands: each band denotes a 10 cm interval, with one band indicating10 cm, two bands indicating 20 cm, etc.
dots: each dot denotes a 1 cm interval
4. Hold catheter at desired depth and remove SWG.
NOTE: Arrow catheters are designed to pass freely over SWG.
Caution: If resistance is encountered when attempting to removeguidewire after catheter placement, guidewire maybe kinked about tip of catheter within vessel (refer to
Figure 29).
Figure 29
In this circumstance, pulling back on SWG may result inundue force being applied resulting in SWG breakage.
If resistance is encountered, withdraw catheter relative toSWG about 2-3 cm and attempt to remove SWG.
If resistance is again encountered remove SWG and cathetersimultaneously.
Warning: Do not apply undue force on guidewire to minimize the risk
of possible breakage.
5. Verify entire SWG is intact upon removal.
Complete Catheter Insertion:1. Check lumen placement by attaching a syringe to each extension
line and aspirate until free flow of venous blood is observed.
2. Flush lumen(s) to completely clear blood from catheter.
3. Connect all extension line(s) to appropriate Luer-Lock line(s)
as required. Unused port(s) may be locked through injectioncap(s) using standard hospital/institutional protocol.
Slide clamps are provided on extension lines to occlude flowthrough each lumen during line and injection cap changes.
Warning: Open catheter clamp prior to infusion through lumento minimize the risk of damage to extension line fromexcessive pressure.
4. Secure catheter to patient.
Where provided, a catheter clamp, fastener, catheterstabilization device, staples or suture may be used.
Use triangular juncture hub with side wings as primarysuture site
Catheter clamp and fastener should be utilized as a secondarysuture site as necessary.
Caution: Minimize catheter manipulation throughout procedure tomaintain proper catheter tip position.
See Catheter Clamp and Fastener, Staple Anchoring Device under AccessoryComponent Instructions section.
5. Ensure insertion site is dry before applying dressing. Apply skinprotectant as needed.
Caution: Do not routinely apply prophylactic topical antimicrobial orantiseptic ointment or cream to the insertion site of central
venous catheters because of the potential risk to promotefungal infections and antimicrobial resistance.
6. Assess placement of catheter tip in compliance with hospital/institutional protocol.
Warning: Do not place Central Venous Catheter (CVC) or PeripherallyInserted Central Catheter (PICC) into or allow them toremain in the right atrium or right ventricle. X-ray examor other method in compliance with hospital/institutionalprotocol must show catheter tip located in right side of
mediastinum in the Superior Vena Cava (SVC) above itsjunction with right atrium and parallel to vessel wall and
its distal tip positioned at a level above either azygos veinor carina of the trachea, whichever is better visualized.Although cardiac tamponade secondary to pericardialeffusion is uncommon, there is a high mortality rateassociated with it. Improper advancement of guidewireinto the heart has also been implicated in causing cardiacperforation and tamponade. For femoral vein approach,catheter should be advanced into vessel so catheter tip lies
parallel to vessel wall and does not enter right atrium.
7. If catheter tip is malpositioned, reposition, redress, and re-verify.
DocumentationHospitals/institutions must establish a permanent medicalrecord that documents the entire procedure, based upontheir policy, procedures, and Best Practices. The actualformat can differ from institution to institution. Report any
product defects/failures to organization risk management,manufacturers, and appropriate regulatory agencies.
Documentation generally includes (but is not limited to)the following information:
1. Device specifics:
type, brand and lot number
length and size of Vascular Access Device (VAD)
internal/external catheter length
antimicrobial or not
14 |
7/27/2019 pressure-injectable-cvc-ifu.pdf
18/20
|
2. Procedure specifics:
time out or procedural pause
informed consent, as required
date, time of insertion, insertion site, number and site
attempts, inserters identification use of visualization and guidance technologies
site preparation and technique
3. Patient assessment and response:
pertinent dx, assessment, vital signs
understanding of procedure, patients response to procedure
complications and barriers to care
4. Terapy specifics:
type of therapy, drug dose, rate, time route and method of administration
laboratory specimen collected
5. Visual confirmation:
verification of appropriate tip location prior to initial use
6. Monitor patient for post catheter insertion complications.
Pressure Injection InstructionsUse sterile technique.
Warning: Obtain a visual image to confirm catheter tip position priorto each pressure injection.
1. Remove injection cap from appropriate extension line ofcatheter.
2. Check for catheter patency:
Attach 10 mL syringe filled with sterile normal saline.
Aspirate catheter for adequate blood return. Vigorously flush catheter.
Warning: Ensure patency of each lumen of catheter prior to pressureinjection to minimize the risk of catheter failure and/orpatient complications.
3. Detach syringe.
4. Attach pressure injection administration set tubing toappropriate extension line of catheter according to manufacturers
recommendations.Caution: Do not exceed ten (10) injections or catheters maximum
recommended flow rate located on product labeling andcatheter luer hub to minimize the risk of catheter failureand/or tip displacement.
5. Inject contrast media in accordance with hospital/institutional protocol.
Caution: Warm contrast media to body temperature prior to pressureinjection to minimize the risk of catheter failure.
6. Disconnect catheter from pressure injector equipment.
7. Flush catheter using 10 mL syringe or larger filled with sterilenormal saline.
8. Disconnect syringe and replace with sterile injection cap oncatheter extension line.
NOTE: Do not exceed ten (10) pressure injections.
Care and Maintenance
Dressing:Replace dressing according to hospital/institutional policies, procedures, andpractice guidelines. Change immediately if the integrity becomes compromisede.g. dressing becomes damp, soiled, loosened, or no longer occlusive.
Caution: Maintain insertion site with regular meticulous redressing
using aseptic technique.
Consult manufacturers recommendations for dressing specifics.
ransparent semipermeable membrane dressing should bechanged every 7 days.
Gauze and tape should be changed every 48 hours.
Label dressing with type, size, and length of catheter; date andtime; and initials of the clinician performing dressing change.
Maintain Catheter Patency:Maintaining central venous catheter patency shall be done in accordance withhospital/institutional policies, procedures, and practice guidelines. All personnelwho care for patients with central venous catheters must be knowledgeableabout effective management to prolong catheters dwell time and prevent injury.
Perform hand hygiene as required.
1. Solution and frequency of flushing a venous access cathetershould be established in hospital/institutional policy.
2. Catheter patency is established and maintained by: flushing intermittently via syringe with heparinized saline or
preservative-free 0.9% sodium chloride continuous drip positive pressure device
3. Te amount of heparin: depends on physician preference, hospital/institutional protocol, patient condition
Caution: Assess patient for heparin sensitivity. Heparin-InducedThrombocytopenia (HIT) has been reported with use ofheparin flush solutions.
4. Te volume of flush solution should be: equal to at least twice the priming volume of the catheter
and any add-on devices
NOTE: Catheter priming volume is printed on product packaging.
Warning: Open catheter clamp prior to infusion through lumen
to minimize the risk of damage to extension line fromexcessive pressure.
PRESSURE INJECTABLE CVC | 15
7/27/2019 pressure-injectable-cvc-ifu.pdf
19/20
5. When using any central venous catheter for intermittent infusiontherapy, proper flushing (heparinization) using a positive-pressure flushing technique will help prevent occlusion.
NOTE: Neutral as well as positive displacement valve systems have also beenshown to help prevent occlusion.
6. Properly cleanse all valves with an appropriate antiseptic beforebeing accessed.
7. Te SASH or SAS method of flushing will help eliminateocclusions due to incompatible solutions:
Saline Administer drug Saline Heparin (if used)
Catheter Removal Procedure1. CVC removal shall be performed:
following order of authorized prescriber
in accordance with organizational policies, procedures, andpractice guidelines
2. A CVC shall be removed immediately upon patient assessmentfor:
suspected contamination i.e. when catheters are insertedduring a medical emergency or if adherence to aseptictechnique cannot be ensured
unresolved complication(s)
discontinuation of therapy
source of infection
Caution: Do not use guidewire techniques to replace catheters inpatient suspected of having catheter-related infection.
3. Place patient in supine position, as clinically indicated tominimize the risk of potential air embolism.
4. Remove dressing.
Warning: Do not use scissors to remove dressing, to minimize the riskof cutting catheter.
5. Remove sutures or staples; or open catheter stabilization deviceretainer wings and remove catheter from catheter stabilizationdevice posts.
6. Place gauze pad over insertion site and catheter.
7. Remove catheter by slowly pulling it parallel to skin. If resistanceis met while removing, catheter should not be forcibly removedand physician should be notified.
Caution: Do not exert excessive force while removing catheter; to
minimize the risk of catheter breakage.
8. Upon removal of catheter:
inspect for intact Blue Flexip or catheter tip
ensure entire catheter length has been removed
9. Apply direct pressure to site until hemostasis is achieved.
10. Apply alcohol swab to catheter stabilization device adhesive andgently lift pad off of skin (if applicable).
11. Dress insertion site. Sterile air occlusive dressing should beapplied and site assessed every 24 hours until site is epithelialized.
Warning: Residual catheter track remains an air entry point until
completely sealed, occlusive dressing should remain in
place for at least 24 - 72 hours dependent upon amount of
time catheter was indwelling.
12. Document catheter removal procedure on patients chart perhospital/institutional protocol.
Include:
catheter condition
presence of intact catheter tip
patients tolerance of the procedure
any interventions needed for removal
For reference literature concerning patient assessment,
clinician education, insertion techniques and potential
complications associated with this procedure refer to Arrow
International, Inc. website: www.arrowintl.com
7/27/2019 pressure-injectable-cvc-ifu.pdf
20/20
2400 Bernville Road, Reading, PA 19605 USA | 1-800-523-8446 1-610-378-0131 | 8 a.m. - 8 p.m. ES | www.arrowintl.com
C-45802-101A (9/09)