35
Prepared by : Mrs. Sohad Noorsaeed, RN,MSN

Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Prepared by : Mrs. Sohad Noorsaeed, RN,MSN

Page 2: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Objectives: Identify central venous catheter and its purposes.

Illustrate the anatomical insertion sites.

Recognize the central venous catheter equipment’s.

Demonstrate line set up, measurement,, and nursing care.

Page 3: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Outlines: Indications of central venous catheter.

Overview for possible insertion sites.

Advantages and disadvantages for each insertion site.

Overview of commonly used equipment.

Manometer and Transducer measurement methods.

Normal range for CVP and reading implications.

Nursing care and considerations.

Contraindications and Complications.

Page 4: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Introduction - Central venous catheter (CVC)is one of the

invasive hemodynamic monitoring through

insertion of the catheter in large vein to right

atrium (tip of catheter should be in the right

atrium) , it is used to provide a good assessment

of right side cardiac function, indirect method to

determining right ventricles filling pressure

(preload).

Page 5: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline
Page 6: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Purposes of central venous catheter : • Serve as a guide for fluid replacement and

medication administration in critically ill patients. • Monitor the pressure and volume in the right

atrium. • Obtain venous access when peripheral vein sites

are inadequate. • Administer blood product(frequent need with

poor peripheral access) and TPN. • Insert pacemaker. • Obtain blood sample.

Page 7: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Insertion Sites: Measuring central venous pressure(CVP) requires

insertion of a catheter in to large central vein:

Internal jugular vein.

Subclavian vein.

Femoral vein.

Page 8: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Internal jugular veins: This site is chosen

frequently as there is a high rate of successful insertion and a low incidence of complications. Internal jugular veins are short, straight and relatively large allowing easy access.

however, catheter occlusion may occur as a result of head movement and may cause irritation in conscious patients.

Page 9: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

:Subclavian veins This site is often chosen as

there are more recognizable anatomical landmarks, making insertion of the device easier.

A subclavian CVC is generally recommended as it is more comfortable for the patient.

Because this site is positioned beneath the clavicle there is a risk of pneumothorax during insertion.

Page 10: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Femoral veins: This site provides rapid

central access during an emergency such as a cardiac arrest.

As the CVC is placed in a vein near the groin there is an increased risk of associated infection.

reported to be uncomfortable and may discourage the conscious patient from moving.

Page 11: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Equipment: Manometer or Monitor with Pressure cable and pressure

bag.

Transducer line (tubing) and holder for the monitor method.

1 liter normal saline (Heparinized).

3-way stopcock (usually comes with the CVP insertion kit).

Wards generally use manometers.

Note: Both CVP measurement methods are reliable

when used correctly.

Page 13: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

The transducer is used to convert the pressure from right

atrium into electrical signal (Accident and Emergency

departments, High Dependency areas and Intensive Care

units use transducers for measuring CVPs).

Tranduced CVP

waveform

Page 14: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Nursing role( Preparatory phase) : Explain procedure to the patient and obtain consent

Evaluate patient PT,PTT,CBC before procedure

Position patient appropriately:

Place patient in trendelenburg position (the body is laid flat on the back

(supine position) with the feet higher than the head by 15-30 degrees ) this position promote the venous return and facilitate the insertion .

Prepare the insertion site(shaving, cleaning)

Establish sterile field on table

Prepare 3 ml syringe and 25 G needle using sterile technique for lidocain injection

Prepare the I.V administration set

Zero transducer & level port with patient right atrium.

Place ECG monitoring.

Page 15: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Trendelenburg position

Page 16: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

:Performance phase ( by physician ) The CVP site is surgically cleaned Ask the patient to do valsalva

maneuver during insertion to increase the intrathoracic pressure and decrease the risk of air embolism.

Monitor for dysrhythmias, tachypnea,

tachycardia. Connect primed IV tubing to catheter and

allow IV solution to flow (regular flushing with NS containing heparin) why?

The catheter should be sutured in place. Place a sterile transparant occlusive

dressing over site. Obtain a chest x-ray.

Page 17: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Line set-up: If the insertion happened while you are

present, you should prime IV line to be

attached to the catheter. Heparinized IV

solution is often used to maintain the patency

of the catheter. “ check the hospital policy”.

Demonstration.

Page 18: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

CVP measurement:

Phlebostatic

axis

CVP is usually recorded at the phlebostaticaxis

(Zero level), This is where the fourth intercostal

space and mid-axillary line cross each other

allowing the measurement to be as close to the

right atrium as possible.

Page 19: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Patient Positioning: Place the patient flat in a supine position if

possible. Alternatively, measurements can

be taken with the patient in a semi-

recumbent position. The position should

remain the same for each measurement

taken to ensure an accurate comparable

result.

Page 20: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Explain the procedure to the patient to gain informed consent.

If IV fluid is not running, ensure that the CVC is patent by flushing the catheter.

1) Using Manometer:

Page 21: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Line up the manometer arm with the

phlebostaticaxis ( Zero level).

Page 22: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Turn the three-way tap off to the

patient and open to the

manometer.

Open the IV fluid bag and slowly fill the manometer to a level higher than the expected CVP.

Page 23: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Turn off the flow from the fluid

bag and open the three-way

tap from the manometer to the

patient.

The fluid level inside the manometer

should fall until gravity equals the

pressure in the central veins.

Page 24: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

When the fluid stops falling the CVP

measurement can be read. If the fluid

moves with the patient's breathing, read

the measurement from the lower number.

Turn the tap off to the manometer.

Page 25: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Document the measurement and report any changes or abnormalities.

Normal Range of CVP:

5-12cm H2O using manometer or 2-6mmHg using transducer

(monitor); when taken from the mid-axillary line at the fourth

intercostalspace.

If increased (hypervolemia) ;decreased (hypovolemia).

Page 26: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

2) Using Transducer: Explain the procedure to the patient to gain informed consent. The CVC will be attached to intravenous fluid within a pressure bag. Ensure that the pressure bag is inflated up to 300mmHg.Place the patient flat in a supine position if possible.

Page 27: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Catheters differ between manufacturers, however, the white or proximal lumen is suitable for measuring CVP.

Tape the transducer to the phlebostaticaxis or as near to the right atrium as possible.

Page 28: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Turn the tap off to the patient and open to the air by removing the cap from the three-way port opening the system to the atmosphere.

Press the zero button on the monitor and wait while calibration occurs.

Page 29: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

When 'zeroed' is displayed on the monitor, replace the cap on the three-way tap and turn the tap on to the patient.

Observe the CVP trace on the monitor.

Page 30: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Post care:

Patient: Return the patient to comfortable position.

Nurse: Hand wash.

Environment: Return equipment.

Documentation: Document the reading.

Page 31: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Contraindications : Patients who have a severe infection or sepsis

Patients having Coagulopathic conditions

Patients who have had thrombolytic or anticoagulant therapy

Patients with superior vena cava syndrome.

Patients with tumor or thrombus in the right atrium.

Page 32: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Complications:

• Bleeding

• Pneumothorax

• Air Embolism

• Thrombus formation

• Infection or sepsis

• Other vessel or organ perforation

Page 33: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Special considerations:

After the initial CVP reading ,reevaluate reading frequently to

establish abase line for the patient

Chang I.V solution every 24 hours according to hospital policy.

Care for the insertion site according to your facility policy, and

observe for signs of infection.

Page 34: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

References: Nettina S. The Lippincott Manual of Nursing

Practice. 8th Edition, Williams &Wilkins Lippincott, 2006

Hand book of Clinical Procedure for Medical Surgical I.

Elaine Cole: Senior lecturer ED/Trauma, City University

Bartsand the London NHS Trust.

Page 35: Prepared by : Mrs. Sohad Noorsaeed, RN,MSNManometer or Monitor with Pressure cable and pressure bag. Transducer line (tubing) and holder for the monitor method. 1 liter normal saline

Thank you