Practical issues in UV, HPLC analysis, and formulation development of Solid Lipid Nanoparticles Vijay Kumar E M.Pharm. Pharmaceutics Email ID: [email protected]

Practical issues in UV, HPLC analysis, formulation development of Solid Lipid Nanoparticles

Embed Size (px)

Citation preview

Practical issues in UV, HPLC analysis, and formulation development of Solid Lipid Nanoparticles

Vijay Kumar E M.Pharm. Pharmaceutics

Email ID: [email protected]

Presentation outline

• Issue 01: Change in Emax and Valley depth

• Issue 02: RP-HPLC chromatograms peak splitting

• Issue 03: Batch to Batch Variability during replication

Issue 01

Change in Emax and Valley depth (UV-Spectrophotometer)

0

Peak 01

Peak 02

Peak 03

Valley 01Valley 02

200nm 400nm

Abso

rban

ce

(AU

)

Wavelength

Day 03Day 02

Day 01

𝐴l=𝑆𝑎𝑚𝑝𝑙𝑒 𝑠𝑜𝑙 .𝑎𝑏𝑠 .l−𝑃𝑢𝑟𝑒 𝑠𝑜𝑙𝑣𝑒𝑛𝑡 𝑎𝑏𝑠 .l

Issue 01

1mg/mL

1.2mg/mL

1.4mg/mL

Day 01

Why do peaks attain a new high Emax as the time progresses?

𝐴=𝑎𝑏𝑐

b

Analyte

Day 02

Day 03

Assume analyte has no chromophoric group that absorbs at 300nm and only responsible agent for absorbance is pure solvent.

Organic

𝑆𝑎𝑚𝑝𝑙𝑒 𝑠𝑜𝑙.𝑎𝑏𝑠 . l 300=1.47

𝐴l 300=1 .47−1 .50

Day 02𝐴l 300=1 .50−1 .50

𝑆𝑎𝑚𝑝𝑙𝑒 𝑠𝑜𝑙.𝑎𝑏𝑠 . l 300=1.50

Day 01

𝑆𝑎𝑚𝑝𝑙𝑒 𝑠𝑜𝑙.𝑎𝑏𝑠 . l 300=1.45

𝐴l 300=1 .45−1 .50

Day 03

Why do valley gets even more deep as the time progresses?

Consider the following case 𝐴l=𝑆𝑎𝑚𝑝𝑙𝑒 𝑠𝑜𝑙 .𝑎𝑏𝑠 .l−𝑃𝑢𝑟𝑒 𝑠𝑜𝑙𝑣𝑒𝑛𝑡 𝑎𝑏𝑠 .l

0

-0.05

300nm

Strategy(s) sought to address the issue

• Store at constant temperature (at 25C) (Temperature ∝ Solubility)

• Adequately tighten the lid of the solution holder

Issue 02

Peak splitting in RP-HPLC chromatograms

AU

-0.002

0.000

0.002

Minutes

0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00

AU

0.000

0.005

0.010

Minutes

0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00

A

B

A-Chromatogram with split peaks; B-Chromatogram with no split peaks

Strategies sought to address the issue

• Mobile phase ratio manipulation• Solvent effect• Guard column replacement

Negative

Negative

Positive

One factor variation at a time

Representation of elution in normal ODS column (guard column) and simultaneous AU-Time graph

Transverse plane view

Longitudinal plane view

P1

P2

P1=P2

Representation of elution in normal ODS column and simultaneous AU-Time graph

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts P1

P2

P1=P2

Representation of elution in normal ODS column and simultaneous AU-Time graph

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts P1

P2

P1=P2

Representation of elution in normal ODS column and simultaneous AU-Time graph

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts P1

P2

P1=P2

Representation of elution in normal ODS column and simultaneous AU-Time graph

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts P1

P2

P1=P2

Peak splitting in RP-HPLC chromatograms due to guard column malfunction

P1

P2

P1=P2

Peak splitting in RP-HPLC chromatograms due to guard column malfunction

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts

P1

P2

P1>P2

Peak splitting in RP-HPLC chromatograms due to guard column malfunction

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts

P1

P2

P1>P2

Peak splitting in RP-HPLC chromatograms due to guard column malfunction

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts P1

P2

P1>>P2

Peak splitting in RP-HPLC chromatograms due to guard column malfunction

Time

Conc

entr

ation

(mcg

/mL)

/ Ab

sorb

ance

uni

ts

P1

P2

Issue 03

Batch to Batch Variability in particle size and PI during replication

Response Replicate I Replicate II Replicate III

Particle size 291.6 513.9 290.9

PI 0.403 0.359 0.364

Replicate II

287.9

0.412

Temp. =50C

50.00Temp.

Setup for replicate 01 & 03

50.00Temp.

Temp. <<<50C

Setup for replicate 02

• Two unequal volumes of water cannot have same temperature when heated for the same time

25mL250mL

Time of heating process = 30min

A B

After heating for 30min; Temp. of A>>Temp. of B

Depth must be kept constant (crucial at low rotation frequency; found to effect PI)

A clamp

Strategy(s) sought to address the issue

• Minimize the background noise (like Constant temperature maintenance, ultra-sonication duration, depth of homogenizer probe immersion )