    Ultrasonic Testing Practical Instructions : General

    1. Linearity of time base:

    General:

    This check may be carried out of a standard calibration block e.g. V

    1

    and a Compression wave

    probe. The Linearity should be checked over a range at least equal to that which is to be used in

    subsequent testing.

    Method:

    1. Place the probe on the 25mm thickness of the IIW V

    1

    Block and adjust the controls to

    display 10 back wall echoes.

    2. Adjust the controls so that the first and last back wall echoes coincide with the scale

    marks at 1 and 10.

    3. Increase the gain to bring successive back wall echoes to 80 full screen Height. The

    leading edge of each echo should line up with the appropriate graticule line.

    4. Record any deviations at approximately half screen height. Deviations should be

    expressed as a percentage of the range between the first and last echoes displayed

    (i.e. 225mm).

    Tolerance: Unless otherwise specified by the testing standard, a tolerance of 2 is considered

    acceptable.

    Frequency of checking: The check shall be carried out at least once per week.

    Document: UT-01 Instruct ion: General Issued 15 7 2 9

  • 7/25/2019 Practical Instructions General

    2/10

    J J J ; ; E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 2 of 10

    Ultrasonic Testing Practical Instructions : General

    2. Linearity of equipment gain:

    General:

    This is a Check on both the Linearity of the amplifier within the set and the calibrated gain control.

    It can be carried out on any calibration block containing a side drilled hole and should be done

    using the probe to be used in subsequent testing. RejecUSuppression controls shall be switched

    off

    Method:

    1. Position the probe on a calibration block to obtain a reflected signal from a small reflector e.g.

    1.6 mm side drilled hole in the V

    1

    block.

    2. Adjust the gain to set this signal to 80 full screen height and note the gain setting (dB).

    3. a) Increase the gain by 2dB and record the amplitude of the signal.

    b) Remove the 2dB and return the signal to 80 FSH.

    c) Reduce the gain by 6dB and record signal amplitude.

    d) Reduce the gain by further 6dB (12 in total) and record signal amplitude

    e) Reduce the gain by further 6dB (18 in total) and record signal amplitude

    f) Reduce the gain by a further 6dB (24 in total) and record signal amplitude.

    Tolerance: (See table below)

    Gain dB

    Expected Amplitude of

    Recorded Amplitude Acceptable Limits

    Full Screen Height

    +2

    101

    66

    Not less than 95

    0

    80

    (Reference Line)

    -6

    40

    3~ /

    37 to 43

    -12

    20

    1 ~

    17 to 23

    -18 10

    i , ; :

    8 to 12

    - 24

    5

    Vis.\:b\e

    Visible, below 8

    Frequency of checking:

    The check shall be carried out at least once per week.

    Document: UT-01 Instruct ion: General Issued 15 7 2 9

  • 7/25/2019 Practical Instructions General

    3/10

    I l T r : 1 E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 3 of 10

    Ultrasonic Testing Practical Instructions : General

    3. Probe index and Beam Alignment:

    General:

    The check for probe index applies only to shear wave probes, and is the first probe characteristic

    to be checked on a shear wave probe. The probe index may be checked on the standard V

    1

    calibration block.

    Method:

    1. Position the probe on the appropriate side of the block to obtain a reflection from the

    quadrant.

    2. Move the probe backwards and forwards to maximize the amplitude of the reflected signal.

    3. When the signal is at maximum, the probe index will correspond to the engraved line on the

    block. Mark this position on the side of the probe.

    Tolerance:

    Tolerance will depend on application but for plotting of defects, it is recommended that probe index

    is accurate to within

    1mm.

    Frequency of checking:

    When a probe is in continuous use, it is recommended that the check be carried out every few

    hours: otherwise, a daily check is recommended.

    Beam Alignment Squint :

    With the probe still in position, a check on a beam alignment can be performed. If the probe beam

    is correctly aligned, the edge of the probe will be parallel to the edge of the block. If this is not the

    case, the squint angle between the two edges. Tolerance depends upon the accuracy of defect

    plotting required. The check should be carried out once per week.

    Document: UT-01 Instruct ion: General Issued 15 07 2009

  • 7/25/2019 Practical Instructions General

    4/10

    f l l E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 4 of 10

    Ultrasonic Testing Practical Instructions : General

    4. Beam Angle

    General:

    Beam angle can be checked on several calibration blocks e.g. V

    1

    or lOW. The beam angle check

    shall preferable be made on a probe in conjunction with the flaw detector to be used in subsequent

    testing.

    Method:

    1. Place the probe in such a position as to receive a reflected signal from the selected transverse

    hole in the calibration block (e.g.19mm deep hole in the AS block).

    2. Maximise the signal from the hole and mark the index point of the probe on the block.

    3. Measure the distance from the marked point on the block to the edge of the block. Knowing the

    position of the drilled hole will allow the beam angle to be calculated (see below).

    *Note. If only a rapid check is required, maximize the signal from the SOmm hole in the V

    1

    block.

    The angle can then be assessed by visual interpolation between the references marking on the

    block.

    Tolerance: The accuracy by the described method is 1.So.

    *The accuracy in this case can only be assumed to be

    3.

    Frequency of checking:

    When a probe is in continuous use, it is recommended that the check be carried out at least every

    few hours: otherwise, a daily check is recommended.

    Document: UT-01 Instruction: General Issued

    15 7 2 9

  • 7/25/2019 Practical Instructions General

    5/10

    I J K ; E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 5

    of

    1

    Ultrasonic Testing Practical Instructions : General

    5. Sensitivity and signal-to-noise ratio:

    General

    The main objective of this check is to provide the operator with a simple method, which will allow

    deterioration in sensitivity of the probe and flaw detector in combination to be identified.

    Method

    1. Place the probe on the A2 calibration block and adjust its position to maximize the signal from

    the 1.5mm diameter hole.

    2. Adjust the gain control to set this signal to 20 FSH, and note the dB setting.

    3. Increase the gain until the overall system noise (electronic noise and grain structure grass) at

    the same range as the target hole reached 20 FSH, and note the new dB setting.

    4. The first gain measurement noted provides a check on the sensitivity of the probe and flaw

    detector, and the difference between the first and second measurements (dB) gives the signal-

    to-noise ratio.

    Note 1: A demonstration of the sensitivity of probe and flaw detector on a calibration block does

    not guarantee that the same size of reflector could be detected in the work piece.

    Note 2: If it is desired to check the sensitivity as a function of range, the use of the standard A5

    block is recommended for longer ranges.

    Tolerance: Tolerance will depend on application. Any deterioration in the sensitivity value will

    indicate a problem with the probe or flaw detector. A low signal-to-noise ratio would be typical of a

    coarse-grained material. The tolerances are usually end user specified.

    Frequency of checking: Unless otherwise agreed, the check shall be carried out once per probe

    per day.

    D ocumen t: U T 1 I ns tr uc tio n: G ene ra l I ss ue d 15 07 2009

  • 7/25/2019 Practical Instructions General

    6/10

    I J l ; t E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 6 of 10

    Ultrasonic Testing Practical Instructions : General

    6 Pulse Duration General:

    This check on the probe and flaw detector in combination, measures the effect on the displayed

    signal of probe damping, amplifier bandwidth, built-in suppression and smoothing circuits. The

    standard V

    calibration block may be used for this check.

    Method:

    1. Calibrate the time base in millimeters to a range that is to be used in subsequent testing.

    2. Maximise the signal from the 1.5mm side drilled hole for shear wave probes or a back wall

    echo for compression wave probes, and set its peak to 100 screen height.

    3. Measure the width of the signal in millimeters at the 10 screen height position.

    4. If desired, the measurement in millimeters can be converted into microseconds by dividing it by

    the relevant sound velocity.

    Tolerance:

    Tolerance will depend upon application. A long pulse duration will limit range resolution and

    indicate the need for a resolution check*, and a short pulse duration may indicate the flaw detector

    has built-in suppression that could prevent the observation of small signals. The tolerances are

    usually end user specified.

    A resolution check is described in clause NO.7

    Frequency of checking: Unless otherwise agreed, the check should be carried out daily.

    Document: UT-01 Instruct ion: General Issued 15 7 2 9

  • 7/25/2019 Practical Instructions General

    7/10

    / l 7 ; ; E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 7 of 10

    Ultrasonic Testing Practical Instructions : General

    7. Resolving Power (Resolution):

    General:

    This check determines the ability of an ultrasonic flaw detection system to give separate indication

    of discontinuities situated close together and simultaneously within the sound beam.

    Method

    1. Calibrate the time base to a range of 0-100 for either the compression wave probe or the shear

    wave probe.

    2. Place the probe so that the axis of the beam impinges upon the 2mm step in the A7 calibration

    block for shear wave probes, or the 3mm step for compression wave probes.

    3. Adjust the position of the probe so that the echoes from the two targets are of the same height

    and approximately half-full graticule height.

    4. The steps are said to be resolved when their echoes are clearly separated at half maximum

    echo height or lower.

    *Note. The 3mm step between the 9mm and 3mm drilled holes in the A6 calibration block may also

    be used when checking compression probes.

    Frequency of checking: The Check shall be carried out monthly, or when too long a pulse

    duration is suspected.

    Document: UT-01 Instruct ion: General Issued

    15 7 2 9

  • 7/25/2019 Practical Instructions General

    8/10

    I l J ; E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 8 of 10

    Ultrasonic Testing Practical Instructions : General

    8 20 dB Beam Spread Profile Plot:

    1. Check and mark index point on the probe

    2

    Calibrate the angle probe for 100 mm range

    3. Select 4 holes at a range of depths from the lOW calibration block (e.g .. 13, 25, 32 43

    mm)

    4. For hole 1 maximize the echo from the hole and adjust to full screen height (100 FSH)

    5. Mark the index point on the block

    6. Move the probe back until the echo is reduced to 10 of FSH 1 10

    th

    or 1 graticule height)

    7. Mark the index point on the block

    8. Move the probe forward through the maximum echo position until the echo again is reduced

    to 10 of FSH 1 10

    th

    or 1 graticule height)

    9. Mark the index point on the block

    10. Measure the standoff distances on the block and record

    11. Repeat steps 4 to 10 for other 3 hole depths

    12. Transfer the measurements for hole depth and standoff distance to a relevant plotting

    devise.

    Document: UT-01 Instruct ion: General Issued 15 7 2 9

  • 7/25/2019 Practical Instructions General

    9/10

    J 7 K 1 E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 9 of 10

    Ultrasonic Testing Practical Instructions : General

    9. Preparation of Distance Amplitude Correction DAC) Curves:

    1. Calibrate the time base to accommodate the full testing range to be used.

    2. Select the target holes to be used for the DAC curve and adjust probe position to obtain

    maximum echo height from the hole giving the highest amplitude. (Not necessarily shortest

    range echo).

    3. Adjust the gain to give an 80 full screen height echo from this hole, and either mark the

    position of the echo tip on the screen or plot the echo height and beam path range on a

    separate graph.

    4. Without altering the equipment gain, adjust the probe position to obtain maximum amplitude

    from each of the holes in turn and either mark the echo tip on the screen or plot on a

    separate graph.

    5. Join the marked points on the screen, or on the graph, to obtain the DAC curve for that

    particular probe. (See Figure 1).

    6. Record the sensitivity at which the DAC curve was plotted by reference to the gain setting

    and echo height from a suitable test block, e.g. the 100mm radius of A2 block. This will

    enable the sensitivity to be easily reset to the DAC level without the use of the special DAC

    test block.

    Document: UT-01 Instruction: General Issued

    15 7 2 9

  • 7/25/2019 Practical Instructions General

    10/10

    f l l E E N A

    INTERNATIONAL

    MEENA INTERNATIONAL

    Page 10 of 10

    Ultrasonic Testing Practical Instructions : General

    10. Penet rat ion Check: Abi li ty to penet rate d if feren t mater ial

    Experiment to be carried out using IIW V1 or BS A 1 block:

    Step-1: Using a normal beam transducer on the 25 mm face of the steel portion calibrate the

    display for 50 mm range by obtaining two back wall echoes.

    Step-2: Now place the normal beam transducer on the face of the 23 mm Perspex without altering

    the calibration.

    Step-3: Observe the reading on the display system

    Note: Sound travels at a lower velocity in Perspex when compared to steel. Hence 23 mm Perspex

    will give reading which is equivalent to that of 50 mm steel and only one back wall echo at 50 mm

    steel equivalent will appear on the screen.