9
8/1/2017 1 Comparison of MR and ultrasound thermometry for monitoring tissue temperature during hyperthermia B Cheng, C Bing, R Staruch, J Nofiele, C Futch, T Laetsch, R Chopra Risk of cardiomyopathy (odds ratio) Applying localized drug delivery to pediatric cancer Armenian et al, Lancet Oncology Mar 2015 Doxorubicin is used for the treatment of vast majority of sarcomas Local control is critically important in sarcoma Augmenting the local effect of treatment may increase the ability to deliver or reduce the toxicity of local control. Thermosensitive liposomes containing doxorubicin (i.e. Thermodox) Stable for 1-2 hours in bloodstream, Rapidly release drug when passing through tissue heated above 41 o C, Prolonged heating allows continuous release for enhanced cellular uptake. New Thermodox Phase II, III trials for breast, liver enrolling, HIFU trials planned. Hyperthermia mediated drug delivery

PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

  • Upload
    vodan

  • View
    216

  • Download
    1

Embed Size (px)

Citation preview

Page 1: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

1

Comparison of MR and ultrasound thermometry for monitoring tissue temperature during hyperthermia

B Cheng, C Bing, R Staruch, J Nofiele, C Futch, T

Laetsch, R Chopra

Ris

k o

f ca

rdio

myo

pat

hy

(o

dd

s ra

tio

)

Applying localized drug delivery to pediatric cancer

Armenian et al, Lancet Oncology Mar 2015

Doxorubicin is used for the treatment of vast majority of sarcomas

Local control is critically important in sarcoma

Augmenting the local effect of treatment may increase the ability to deliver or

reduce the toxicity of local control.

4 Confidential

Thermosensitive liposomes containing doxorubicin (i.e. Thermodox)

Stable for 1-2 hours in bloodstream,

Rapidly release drug when passing through tissue heated above 41oC,

Prolonged heating allows continuous release for enhanced cellular uptake.

New Thermodox Phase II, III trials for breast, liver enrolling, HIFU trials planned.

Hyperthermia mediated drug delivery

Page 2: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

2

3 Confidential

Benefits and challenges of mild hyperthermia

Cell death

in < 5 sec

Risk of

damage

in < 5 min Mild HT

Undesired

thermal

damage

No

effect

Risk of

damage

Mild HT (40-43°C, 10-30 minutes)

enhances RT and chemotherapy,

can trigger localized drug release

Higher temperatures or longer durations

damage vasculature and normal tissue,

lower temperatures have reduced effect.

Precise temperature control poses

challenging technical requirements

Thermometry accuracy ± 1°C, >10 min,

overcoming magnetic field drift, motion.

Heat > 5 cm region using 1.5 mm focus,

adapting to spatio-temporally varying

blood flow and energy absorption.

Thermal

ablation

T1 planning and T2-TSE mDIXON water images

H

F

LR

A

P

LR

H

F

AP2 cm

Focal point

Near field

Far field

Heated region

MR-HIFU hyperthermia in rabbit tumor

Temperature in 10 mm diameter region kept at 42oC for 40 minutes (1.2 MHz, 60W).

Temperature-sensitive liposomal doxorubicin infused during first 6 minutes.

1 2 3NF

NF FF

21 3 FF

Page 3: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

3

Heating results: 10, 20, and 40 minutes

10 min

20 min

40 min

DOX

DOX

DOX

Time 41-45oC

Time [s]

Importance of drift correction for long duration heating

Type of correction used for PRF shift thermometry matters▪

Validation required on every platform due to hardware differences▪

Bing et al, International Journal of Hyperthermia, 32(6), 2016

Strategies for dealing with respiratory artifacts

▪ PRF shift thermometry extremely

sensitive to motion and Bo

disturbances

▪ A strategy of ventilated breath-

holds might be a strategy to

achieve long duration heating

A

B

Temp probe

ROI for MR thermometry

Heating complete

Ultrasound Beam

QA Phantom

Heating Target

Moving DirectionMotion Platform

Plastic Rod

Aluminum Disk

A

3

No motion

2

Motion introduced

Motion Amplitude

TimeREFERENCE

MOTION

HIFU ON

MOTION HOLD

HIFU OFF

B

Reference images

Page 4: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

4

Strategies for dealing with respiratory artifacts

PRF shift thermometry extremely ▪

sensitive to motion and Bo

disturbances

A strategy of ventilated breath▪ -

holds might be a strategy to

achieve long duration heating

C

A

Ultrasound Beam

Transverse Slice

40

43

4546

56 ℃

48

Heating ROI

BA

Heating complete

Strain-based US thermometry

11

Thermal induced strains

Speed of sound change

Linearly proportional (37-50 ͦC)

Visa Suomi et al 2016 Phys. Med. Biol. 61 7427 doi:10.1088/0031-9155/61/20/7427

TemperatureChange

Ultrasound echoes shift

US signal analysis

12

US echo shift

US thermal strain

(M-mode)

Page 5: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

5

Is US thermometry feasible for mild hyperthermia in vivo?

Two methods to validate US thermometry

MR thermometry (PRF shift)▪

Gold▪ -standard implanted fiber-optic temperature sensor

Question

13

Ex vivo calibration experiment setup

14

Philips, Ingenia 3T

Ex vivo calibration experiment setup

15

Page 6: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

6

16

Membrane 1

Membrane 2

Mu

scle

Ex vivo calibration data: A-lines in chicken and pork muscle

Timing of interleaved HIFU and pulse-echo acquisition

17

Ultrasound vs Fiber-optic sensor (35-45 C)

18

Chicken muscle

Pork muscle

Strain to T conversion

coefficient:

-4.98±1.0*10-4 / ͦc

Strain to T coefficient:

-4.70±0.8*10-4 ͦC

RMSE < 0.4 ͦC

RMSE < 0.4 ͦC

∆T

Page 7: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

7

In vivo Experiment setup (Rabbit n=4)

19

Philips, Ingenia 3T

3D T1-weighted MRI images

20

Results: MR vs fiber-optic sensor

21

• MR thermometry and fiber-optic sensor had comparable accuracy and precision (< 0.5 ˚C)

• MR ROI: 4×4 mm at the location of the sensor

Page 8: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

8

Results: US at focus vs sensor 4 mm away from focus

22

Calibrated thermal strain to temperature conversion coefficient -0.0005 was applied.

Results: US vs MR vs Fiber-optic sensor

23

0 50 100 150 200 250 300

0

2

4

6

8

10

Time (s)

T

(C

)

US -0.0005

MR at focus

Sensor

▪ Ex vivo ultrasound thermometry has been successfully performed in both chicken and pork

muscle by comparing with implanted fiber-optic temperature sensors.

▪ Simultaneous MR and US thermometry were successfully acquired in stationary in vivo rabbit

muscle under regular respiration and perfusion. In comparison with implanted fiber-optic sensor

readings, we showed that strain-based ultrasound thermometry is feasible in the mild

hyperthermia range.

▪ Although effective, the calibrated strain-to-temperature conversion coefficient has to be

validated in more experiment conditions and more types of tissues.

Conclusion

24

Page 9: PowerPoint Presentation Template 3amos3.aapm.org/abstracts/pdf/127-35751-424554-127803.pdf · ô/ í/ î ì í ó í Comparison of MR and ultrasound thermometry for monitoring tissue

8/1/2017

9

Acknowledgement

25