Piezo Lab Amp Use and Applications

Embed Size (px)

Citation preview

  • 8/6/2019 Piezo Lab Amp Use and Applications

    1/43

    Applications of Piezo Film Lab Amp

    2008-10-13R H Brown

  • 8/6/2019 Piezo Lab Amp Use and Applications

    2/43

    Just imagine

    ...the possibilities of piezo film

  • 8/6/2019 Piezo Lab Amp Use and Applications

    3/43

    The need for a lab ampThe need for a lab ampThe need for a lab amp

    Piezo Film: No DC Responsebut howlow can we go?

    Many technologies from MEAS have response down to andincluding DC: piezoresistive MEMS, Microfused, capacitivehumidity, magnetoresistive

    But the piezoelectric effect generates charge within acapacitive source. The charge leaks away, either throughthe dielectric itself, or into the external circuit. This meanswe apply a high-pass filter to every signal

  • 8/6/2019 Piezo Lab Amp Use and Applications

    4/43

    Example sensor: SDT1-028K

    The SDT1-028K has a capacitance of 2.6 nF. With noload, the sensor "tries" to respond down to < 0.1 mHz.When connected to different input resistances, the cut-

    off frequency of the high-pass filter varies as shownhere

    C: 2n6 Load R -3 dB freq Hz

    1T 0.0001

    1G 0.061

    100M 0.612

    10M 6.12

    1M 61.2

    The need for a lab ampThe need for a lab ampThe need for a lab amp

  • 8/6/2019 Piezo Lab Amp Use and Applications

    5/43

    The need for a lab ampThe need for a lab ampThe need for a lab amp

    Influence of Input Impedance selection on low f requency response (voltage mode)

    The table below shows the theoretical -3 dB frequency (in Hz) of the high-pass filter that is formedwhen a sensor of capacitance listed in first column is connected to the specified input impedance.

    Note: the shaded cells have a -3 dB frequency that is below the range of the Low Frequency Selector,and in this case, the Low frequency Selector will determine the performance. In the case of theunshaded cells, the low frequency limit will be determined either by the Low Frequency selector, or by

    the data above, whichever is the higher.

    1 M 10 M 100 M 1 G

    100 p 1592 159 16 1.6

    220 p 723 72 7.2 0.72

    470 p 339 34 3.4 0.34

    1 n 159 16 1.6 0.16

    2.2 n 72 7.2 0.72 0.072

    4.7 n 34 3.4 0.34 0.034

    10 n 16 1.6 0.16 0.016

    22 n 7.2 0.72 0.072 0.0072

    47 n 3.4 0.34 0.034 0.0034

    100 n 1.6 0.16 0.016 0.0016

  • 8/6/2019 Piezo Lab Amp Use and Applications

    6/43

    Instrumentation: digital oscilloscope

    Most conventional instruments offer a 1 M inputresistance as standard. If we connect the SDT1-028Kpiezo film sensor directly to this input, we create a high-

    pass filter at 61 Hz.C: 2n6 Load R -3 dB freq Hz

    1T 0.0001

    1G 0.061

    100M 0.612

    10M 6.12

    1M 61.2

    The need for a lab ampThe need for a lab ampThe need for a lab amp

  • 8/6/2019 Piezo Lab Amp Use and Applications

    7/43

    -4

    -3

    -2

    -1

    0

    1

    2

    3

    4

    0 2 4 6 8 10 12 14 16 18 20

    -4

    -3

    -2

    -1

    0

    1

    2

    3

    4

    0 2 4 6 8 10 12 14 16 18 20

    Effect of the high-pass filter

    The presence of the high-pass filter can affect theobserved waveform in different ways:

    1) Change in waveshape

    2) Change in amplitude

    3) Change in phase

    The need for a lab ampThe need for a lab ampThe need for a lab amp

  • 8/6/2019 Piezo Lab Amp Use and Applications

    8/43

    Simulated waveform examples:

    1) rectangular pulse (1 V pulse high for 1 s, 10 ms r/f time)

    2) Sawtooth (ramp from 0 to 1 V in 1 s, 10 ms fall)

    3) Continuous sine (1 Hz, 2 V pk-pk)

    Example of real-life waveforms:

    4) Chest strap (SDT1-028K in tension, elasticated strap)

    The need for a lab ampThe need for a lab ampThe need for a lab amp

  • 8/6/2019 Piezo Lab Amp Use and Applications

    9/43

    Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse

    Load R pos pk V neg pk V

    1T 1 -0.0004

    1G 0.998 -0.321

    100M 0.981 -0.96

    10M 0.828 -0.828

    1M 0.257 -0.257

  • 8/6/2019 Piezo Lab Amp Use and Applications

    10/43

    Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse

    Load R pos pk V neg pk V

    1T 1 -0.0004

    1G 0.998 -0.321

    100M 0.981 -0.96

    10M 0.828 -0.828

    1M 0.257 -0.257

  • 8/6/2019 Piezo Lab Amp Use and Applications

    11/43

    Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse

    Load R pos pk V neg pk V

    1T 1 -0.0004

    1G 0.998 -0.321

    100M 0.981 -0.96

    10M 0.828 -0.828

    1M 0.257 -0.257

  • 8/6/2019 Piezo Lab Amp Use and Applications

    12/43

    Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse

    Load R pos pk V neg pk V

    1T 1 -0.0004

    1G 0.998 -0.321

    100M 0.981 -0.96

    10M 0.828 -0.828

    1M 0.257 -0.257

  • 8/6/2019 Piezo Lab Amp Use and Applications

    13/43

    Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse

    Load R pos pk V neg pk V

    1T 1 -0.0004

    1G 0.998 -0.321

    100M 0.981 -0.96

    10M 0.828 -0.828

    1M 0.257 -0.257

  • 8/6/2019 Piezo Lab Amp Use and Applications

    14/43

    Case 2: sawtoothCase 2:Case 2: sawtoothsawtooth

    Load R pos pk V neg pk V

    1T 0.999 -0.001

    1G 0.83 -0.17

    100M 0.255 -0.73

    10M 0.026 -0.81

    1M 0.0026 -0.25

  • 8/6/2019 Piezo Lab Amp Use and Applications

    15/43

    Case 2: sawtoothCase 2:Case 2: sawtoothsawtooth

    Load R pos pk V neg pk V

    1T 0.999 -0.001

    1G 0.83 -0.17

    100M 0.255 -0.73

    10M 0.026 -0.81

    1M 0.0026 -0.25

  • 8/6/2019 Piezo Lab Amp Use and Applications

    16/43

    Case 2: sawtoothCase 2:Case 2: sawtoothsawtooth

    Load R pos pk V neg pk V

    1T 0.999 -0.001

    1G 0.83 -0.17

    100M 0.255 -0.73

    10M 0.026 -0.81

    1M 0.0026 -0.25

  • 8/6/2019 Piezo Lab Amp Use and Applications

    17/43

    Case 2: sawtoothCase 2:Case 2: sawtoothsawtooth

    Load R pos pk V neg pk V

    1T 0.999 -0.001

    1G 0.83 -0.17

    100M 0.255 -0.73

    10M 0.026 -0.81

    1M 0.0026 -0.25

  • 8/6/2019 Piezo Lab Amp Use and Applications

    18/43

    Case 2: sawtoothCase 2:Case 2: sawtoothsawtooth

    Load R pos pk V neg pk V

    1T 0.999 -0.001

    1G 0.83 -0.17

    100M 0.255 -0.73

    10M 0.026 -0.81

    1M 0.0026 -0.25

  • 8/6/2019 Piezo Lab Amp Use and Applications

    19/43

    Case 3: sine waveCase 3: sine waveCase 3: sine wave

    Load R pk-pk Vphasedeg

    1T 2 0.004

    1G 1.99 4

    100M 1.71 32

    10M 0.33 81

    1M 0.033 89

  • 8/6/2019 Piezo Lab Amp Use and Applications

    20/43

    Case 3: sine waveCase 3: sine waveCase 3: sine wave

    Load R pk-pk Vphasedeg

    1T 2 0.004

    1G 1.99 4

    100M 1.71 32

    10M 0.33 81

    1M 0.033 89

  • 8/6/2019 Piezo Lab Amp Use and Applications

    21/43

    Case 3: sine waveCase 3: sine waveCase 3: sine wave

    Load R pk-pk Vphasedeg

    1T 2 0.004

    1G 1.99 4

    100M 1.71 32

    10M 0.33 81

    1M 0.033 89

  • 8/6/2019 Piezo Lab Amp Use and Applications

    22/43

    Case 3: sine waveCase 3: sine waveCase 3: sine wave

    Load R pk-pk Vphasedeg

    1T 2 0.004

    1G 1.99 4

    100M 1.71 32

    10M 0.33 81

    1M 0.033 89

  • 8/6/2019 Piezo Lab Amp Use and Applications

    23/43

    Case 3: sine waveCase 3: sine waveCase 3: sine wave

    Load R pk-pk Vphasedeg

    1T 2 0.004

    1G 1.99 4

    100M 1.71 32

    10M 0.33 81

    1M 0.033 89

  • 8/6/2019 Piezo Lab Amp Use and Applications

    24/43

    Case 4: chest strapCase 4: chest strapCase 4: chest strap

    Real-life waveforms: chest strap

    SDT1-028K sensor inserted in series with an

    elasticised chest strap, worn while breathingnormally

  • 8/6/2019 Piezo Lab Amp Use and Applications

    25/43

    Case 4: chest strap (1 G)Case 4: chest strap (1 G)Case 4: chest strap (1 G)

    -4

    -3

    -2

    -1

    0

    1

    2

    3

    4

    0 2 4 6 8 10 12 14 16 18 20

  • 8/6/2019 Piezo Lab Amp Use and Applications

    26/43

    Case 4: chest strap (100 M)Case 4: chest strap (100 M)Case 4: chest strap (100 M)

    -4

    -3

    -2

    -1

    0

    1

    2

    3

    4

    0 2 4 6 8 10 12 14 16 18 20

  • 8/6/2019 Piezo Lab Amp Use and Applications

    27/43

    Case 4: chest strap (10 M)Case 4: chest strap (10 M)Case 4: chest strap (10 M)

    -4

    -3

    -2

    -1

    0

    1

    2

    3

    4

    0 2 4 6 8 10 12 14 16 18 20

  • 8/6/2019 Piezo Lab Amp Use and Applications

    28/43

    Case 4: chest strap (1 M)Case 4: chest strap (1 M)Case 4: chest strap (1 M)

    -4

    -3

    -2

    -1

    0

    1

    2

    3

    4

    0 2 4 6 8 10 12 14 16 18 20

    Case 4: chest strap (1 M rangeCase 4: chest strap (1 MCase 4: chest strap (1 M rangerange

  • 8/6/2019 Piezo Lab Amp Use and Applications

    29/43

    Case 4: chest strap (1 M rangechanged)Case c est st ap (p ( a gegchanged)changed)

    -0.05

    -0.04

    -0.03

    -0.02

    -0.01

    0

    0.01

    0.02

    0.03

    0.04

    0.05

    0 2 4 6 8 10 12 14 16 18 20

    Case 4: chest strap (1 M, 1 Hz HPF,Case 4: chest strap (1 M, 1 Hz HPF,Case 4: chest strap (1 M, 1 Hz HPF,

  • 8/6/2019 Piezo Lab Amp Use and Applications

    30/43

    p ( , ,+40dB )

    p ( , ,p ( , ,+40dB )+40dB )

    -2

    -1.5

    -1

    -0.5

    0

    0.5

    1

    1.5

    2

    0 2 4 6 8 10 12 14 16 18 20

    C

  • 8/6/2019 Piezo Lab Amp Use and Applications

    31/43

    Conclusion low frequenciesConclusionConclusion low frequencieslow frequencies

    The Piezo Film Lab Amp avoids the loss of low-frequency information content from piezo filmsensor signals, by providing controlled load

    resistance of up to 1 G in voltage mode, and LFresponse down to 0.1 Hz

    Pi Fil L b APi Fil L b A

  • 8/6/2019 Piezo Lab Amp Use and Applications

    32/43

    The choice: Voltage or Charge mode

    To detect the signal generated by a piezo sensor, wehave basically two options:

    1) Leave all of the charge on the sensor, leaving the

    voltage exactly as if nothing were connected, or2) Remove all the charge from the sensor and transfer

    it elsewhere, leaving no voltage across the sensor

    We describe option (1) as "voltage mode", and option(2) as "charge mode"

    Piezo Film Lab AmpPiezo Film Lab AmpPiezo Film Lab Amp

    Th d f l bTh d f l bTh d f l b

  • 8/6/2019 Piezo Lab Amp Use and Applications

    33/43

    Charge Mode:

    In a charge mode preamplifier, the input appears to bea short circuit. The potential between the two inputterminals is maintained at exactly zero volts. All of the

    charge generated by the piezo sensor flows into thecircuit, and is transferred to a feedback capacitor. Therate of charge decay is then governed by this selectedcapacitor, and not by the capacitance of the piezosource. Since the input is "shorted", straycapacitance has no effect.

    The need for a lab ampThe need for a lab ampThe need for a lab amp

    Si l h d i itSi l h d i itSi l h d i it

  • 8/6/2019 Piezo Lab Amp Use and Applications

    34/43

    Simple charge mode circuitSimple charge mode circuitSimple charge mode circuit

    Si l h d i itSi l h d i itSimple charge mode circuit

  • 8/6/2019 Piezo Lab Amp Use and Applications

    35/43

    Simple charge mode circuitSimple charge mode circuitSimple charge mode circuit

    Notes:

    Circuit "gain" defined by C1/C2 (with Lab Amp, C2 isswitch-selectable in decades from 100 pF to 100 nF)

    Time constant defined by product R1C2 (with Lab Amp, the

    low freq limit is switch-selectable in decades from 0.1 Hz to1 kHz)

    Additional capacitance across input has no effect (excepton very high frequency response)

    The need for a lab ampThe need for a lab ampThe need for a lab amp

  • 8/6/2019 Piezo Lab Amp Use and Applications

    36/43

    Voltage Mode:

    In a voltage mode preamplifier, the input appears to bean infinite impedance. The potential between the twoinput terminals is maintained at the "open circuit"

    potential of the source network (including all straycapacitance). None of the charge generated by thepiezo sensor flows into the circuit. A bleed resistor isrequired on the input (otherwise circuit would beunstable or saturate), and the low frequency responseis governed by the total capacitance connected to theinput, and this bleed resistance.

    The need for a lab ampThe need for a lab ampThe need for a lab amp

    Simple voltage mode circuitSimple voltage mode circuitSimple voltage mode circuit

  • 8/6/2019 Piezo Lab Amp Use and Applications

    37/43

    Simple voltage mode circuitSimple voltage mode circuitSimple voltage mode circuit

    Simple voltage mode circuitSimple voltage mode circuitSimple voltage mode circuit

  • 8/6/2019 Piezo Lab Amp Use and Applications

    38/43

    Simple voltage mode circuitSimple voltage mode circuitSimple voltage mode circuit

    Notes:Circuit "gain" is unity; separate gain stage can follow (withLab Amp, final gain stage offers 0 to +40 dB in 10 dB steps)

    Time constant defined by product R1C1 (with Lab Amp, R1

    is switch-selectable in decades from 1 M to 1 G)

    Additional capacitance across source reduces the "open-circuit" potential, and thus the detected signal

    Selection of modeSelection of modeSelection of mode

  • 8/6/2019 Piezo Lab Amp Use and Applications

    39/43

    Selection of modeSelection of modeSelection of mode

    Use charge mode when:

    Piezo sensor has long cable attached (or varying cable

    length)

    Only a portion of the active sensor area is impacted orexcited

    "Fundamental" measurement of charge output preferredMinimize triboelectric cable noise

    Use voltage mode when:

    Lab Amp will later be replaced by simple buffer stage

    "End of cable" output voltage is preferred

    Using the low-pass filterUsing the lowUsing the low--pass filterpass filter

  • 8/6/2019 Piezo Lab Amp Use and Applications

    40/43

    Using the low-pass filterUsing the lowUsing the low--pass filterpass filter

    In addition to the control over the low-frequency limit afforded by the inputimpedance selector and the LF filter

    control, the Piezo Film Lab Amp also has alow-pass filter that can block unwantedhigher frequency components.

    Setting this control to the "10 Hz" settingallows effective filtering of 50/60 Hz linefrequency interference, which is commonlyexperienced when experimenting with

    piezo film sensors.

    Using the low-pass filterUsing the lowUsing the low--pass filterpass filter

  • 8/6/2019 Piezo Lab Amp Use and Applications

    41/43

    Using the low-pass filterUsing the lowUsing the low-pass filterpass filter

    With High Freq control set to 1 kHz:

    Using the low-pass filterUsing the lowUsing the low--pass filterpass filter

  • 8/6/2019 Piezo Lab Amp Use and Applications

    42/43

    Using the low pass filterUsing the lowUsing the low pass filterpass filter

    With High Freq control set to 10 Hz:

    Piezo Film Lab AmpPiezo Film Lab AmpPiezo Film Lab Amp

  • 8/6/2019 Piezo Lab Amp Use and Applications

    43/43

    Piezo Film Lab AmpPiezo Film Lab AmpPiezo Film Lab Amp

    Key Features:

    Charge or voltage mode operation0.01 to 1000 mV/pC sensitivity range in charge mode1 M to 1 G input resistance, -40 to 40 dB dB gain, in

    voltage modeMulti-pole low-pass and high-pass filtering, with -3dB frequencies ranging from 0.1Hz to 100kHz

    BNC input and outputInternal battery (2 x 9 V) or external 24 Vdc powersupply operation