Observations of Directivity in Transducers

Embed Size (px)

Citation preview

  • 8/3/2019 Observations of Directivity in Transducers

    1/48

    1

    Observations ofDirectivity in Transducers

    Richard LittleDirector of Advanced Design

    Tymphany HK Ltd.

    13-14 November 2010Acoustic Block 2010

    The Sound of Modern Design

  • 8/3/2019 Observations of Directivity in Transducers

    2/48

    2

    Tymphany HK Ltd.

    13-14 November 2010Acoustic Block 2010

    The Sound of Modern Design

  • 8/3/2019 Observations of Directivity in Transducers

    3/48

    Abstract

    The causes of transducer directivity arereviewed. The basic theoretical performance of apiston radiating from an infinite baffle is discussed,including calculations of directivity and total radiatedpower. Measurements of transducer directivity and

    radiated power are reviewed for a range of transducersizes. The divergences of these measurement results,from the expectations for a piston, are examined anddiscussed.

    Confidential 3

  • 8/3/2019 Observations of Directivity in Transducers

    4/48

    Confidential 4

  • 8/3/2019 Observations of Directivity in Transducers

    5/48

    Introduction to Directivity A transducers diaphragm moves back and forth during operation,

    pushing and pulling on the surrounding air and creating an outgoing

    pressure wave.

    At low frequencies, the wavelength of the outgoing sound pressure wave

    is large relative to the diaphragm, and the diaphragm operates much likea point source, radiating sound pressure waves evenly in all directions.

    At higher frequencies, however, the wavelength of the pressure wave issized similarly to the diaphragm size, or even smaller. At different

    observation positions, at these frequencies, there is a path lengthdifference from the observation position to different parts of thetransducers diaphragm. This means that the summed contributions tothe total pressure observed may increase or decrease, according to theeffects of the path length differences.

    Typically a transducer diaphragm is symmetric around its central axis, sothe effects of the path length differences are observed to be differentaccording to the direction that the transducer is being observed, relativeto the central axis. We call this effect DIRECTIVITY.

    Confidential 5

  • 8/3/2019 Observations of Directivity in Transducers

    6/48

    Confidential 6

  • 8/3/2019 Observations of Directivity in Transducers

    7/48

    Directivity of a Piston in an Infinite Baffle

    In this paper, were going to limit measurements and calculationsto that of transducers and pistons operating mounted to an infinitebaffle. Were going to neglect reflections off of loudspeaker

    cabinet boundaries. The figure below illustrates how the pressure at the observation

    position, created by a radiating piston mounted in an infinite baffle,is calculated by integrating over the radiating surface.

    Confidential 7

    Infinitebaffle

    piston

    Observation position

  • 8/3/2019 Observations of Directivity in Transducers

    8/48

    Confidential 8

    Infinitebaffle

    piston

    Observation position

  • 8/3/2019 Observations of Directivity in Transducers

    9/48

    Example: radiation pattern on and off axis

    This example shows the radiation pattern for a transducer, on and offaxis, at different frequencies. The output of the transducer is the same

    in all directions at low frequencies, but at high frequencies this is nolonger the case.

    Confidential 9

  • 8/3/2019 Observations of Directivity in Transducers

    10/48

    Confidential 10

  • 8/3/2019 Observations of Directivity in Transducers

    11/48

    Calculating directivity for a infinite-baffle piston

    The Directivity Index DI is a measure of the directivity of atransducer, and can be calculated from the following formulae:

    Here a refers to the diameter of the diaphragms radiating area,and c is the speed of sound (345 m/s).

    These formulae are approximations to more precise Besselfunction relations.

    Confidential 11

  • 8/3/2019 Observations of Directivity in Transducers

    12/48

  • 8/3/2019 Observations of Directivity in Transducers

    13/48

    Calculating directivity from measurements

    The directivity factor Q is the ratio of the on axis frequencyresponse of a transducer, to the on axis frequency response for a

    point source radiator producing the same amount of total radiatedpower. Q and DI are related:

    Q is calculated by calculating the total radiated power, in a ratiowith the on-axis frequency response:

    For a transducer which is symmetric around its central axis, thiscan be re-stated as:

    Confidential 13

  • 8/3/2019 Observations of Directivity in Transducers

    14/48

    Q QDI

    Q

    Q

    Confidential 14

  • 8/3/2019 Observations of Directivity in Transducers

    15/48

    Calculating directivity from measurements (2)

    Typically, when conducting measurements, we measure off-axisin small increments of . The previous relation for Q can then be

    approximated as follows:

    We can then calculate DI from Q.

    The results of these calculations of the DI can be comparedagainst the piston theoretical curve. Good agreement betweenthe measurement results, vs. the theoretical curve, implies thatthe transducers total radiation pattern is pistonic.

    Confidential 15

  • 8/3/2019 Observations of Directivity in Transducers

    16/48

    (2)

    0180Q

    DI

    DIDI

    Confidential 16

  • 8/3/2019 Observations of Directivity in Transducers

    17/48

    Example: measured directivity vs. piston

    theoretical directivity In this example, the transducers radiation pattern is pistonic, for its

    radiating area, until ~ 6 kHz. Above that frequency, the driver is

    effectively functioning as if its radiating area were smaller than it really is.

    Confidential 17

  • 8/3/2019 Observations of Directivity in Transducers

    18/48

    6

    Confidential 18

  • 8/3/2019 Observations of Directivity in Transducers

    19/48

    Break-up modes and on-axis measurements

    The effect of diaphragm resonances(break-up modes) on the on-axis

    frequency response can be studiedby mirroring the transducers

    impedance curve onto the on-axisfrequency response curve, startingat the impedance minimum

    frequency. Large deviationsbetween the impedance-projectedcurve and the on-axis responsecurve suggest mechanical

    resonances have altered the

    frequency response. The example to the right shows two

    possible resonances: a shallowbroad resonance at 7 kHz, and asharp high-Q resonance at 19 kHz.

    Confidential 19

  • 8/3/2019 Observations of Directivity in Transducers

    20/48

    7KHzQ19KHzQ

    Confidential 20

  • 8/3/2019 Observations of Directivity in Transducers

    21/48

    On and off axis measurements

    A transducers radiation pattern will become directional at frequenciesabove ka=1 (from our previous equation). The graph below highlightsthis point in the frequency response (the radiation dispersion frequency).

    Confidential 21

  • 8/3/2019 Observations of Directivity in Transducers

    22/48

    (ka=1)()

    Confidential 22

  • 8/3/2019 Observations of Directivity in Transducers

    23/48

    Break-up modes and directivity

    A case study Were going to compare the radiation patterns of two drivers, with

    essentially the same diaphragm, but two different voice coil diameters(one smaller, one bigger).

    A sketch of the diaphragm and voice coil are shown below. Because ofthe difference in coil sizes, the break-up frequency of the diaphragmshould be different in the two cases.

    Confidential 23

    voice coil

    diaphragm

    surround

    voice coil

    diaphragm

    surround

    Small coil Large coil

  • 8/3/2019 Observations of Directivity in Transducers

    24/48

    Confidential 24

    voice coil

    diaphragm

    surround

    voice coil

    diaphragm

    surround

    Small coil Large coil

  • 8/3/2019 Observations of Directivity in Transducers

    25/48

    Case study comparison of on-axis response

    and break-up mode behaviorSmall coilBreak-up modes small until 19 kHz

    Large coilBreak-up modes affect response above 4 kHz

    Confidential 25

  • 8/3/2019 Observations of Directivity in Transducers

    26/48

    -

    19KHz 4KHz

    Confidential 26

  • 8/3/2019 Observations of Directivity in Transducers

    27/48

    Case study: comparison of off-axis frequency

    responseSmall coilResponse off-axis remains roughly pistonic untilthe effects of the resonance at 19 kHz enhance

    the off-axis response

    Large coilOff-axis performance kept close to on-axisperformance due to effects of ~12 kHzresonance

    Confidential 27

  • 8/3/2019 Observations of Directivity in Transducers

    28/48

    -

    19KHz 12KHz

    Confidential 28

  • 8/3/2019 Observations of Directivity in Transducers

    29/48

    Case study: Directivity index comparison

    Small coilDriver performs roughly like a piston until above10 kHz

    Large coilDriver shows large diversions from pistonicbehavior above 10 kHz, effectively losingradiating area due to mechanical resonances.

    Confidential 29

  • 8/3/2019 Observations of Directivity in Transducers

    30/48

    -

    10KHz 10KHz

    Confidential 30

  • 8/3/2019 Observations of Directivity in Transducers

    31/48

    Case study: radiation pattern comparison

    Small coil Large coil

    Confidential 31

  • 8/3/2019 Observations of Directivity in Transducers

    32/48

    -

    Confidential 32

  • 8/3/2019 Observations of Directivity in Transducers

    33/48

    Notes from the case study

    Small coil

    Smaller coil = cheaper motor

    Lower coil inductance

    Higher on-axis frequencyresponse bandwidth

    Low resonance design more repeatable response inproduction

    Higher directivity at higherfrequencies

    Large coil

    Large coil = expensive motor

    Higher coil inductance

    Lower on-axis frequencyresponse bandwidth

    Design depends uponresonances

    Lower directivity at higherfrequencies

    Confidential 33

  • 8/3/2019 Observations of Directivity in Transducers

    34/48

    =

    =

    Confidential 34

  • 8/3/2019 Observations of Directivity in Transducers

    35/48

    Typical radiation pattern 25 mm Vifa NE silk

    dome tweeter NE25VTS-04

    Confidential 35

    25 Vif NE

  • 8/3/2019 Observations of Directivity in Transducers

    36/48

    25 mm Vifa NE

    NE25VTS-04

    Confidential 36

  • 8/3/2019 Observations of Directivity in Transducers

    37/48

    Typical radiation pattern 25 mm ring

    radiator tweeter XT25SC90-04

    Confidential 37

    25

  • 8/3/2019 Observations of Directivity in Transducers

    38/48

    25 mm

    XT25SC90-04

    Confidential 38

    T i l di ti tt NXT BMR46 f ll

  • 8/3/2019 Observations of Directivity in Transducers

    39/48

    Typical radiation pattern NXT BMR46 full-

    range driver

    Confidential 39

  • 8/3/2019 Observations of Directivity in Transducers

    40/48

    NXT BMR46

    Confidential 40

    T i l di ti tt 180 Vif NE

  • 8/3/2019 Observations of Directivity in Transducers

    41/48

    Typical radiation pattern 180 mm Vifa NE

    woofer NE180W-04

    Confidential 41

    180 Vif NE

  • 8/3/2019 Observations of Directivity in Transducers

    42/48

    180 mm Vifa NE

    NE180W-04

    Confidential 42

    Typical radiation pattern 315 mm Vifa NE

  • 8/3/2019 Observations of Directivity in Transducers

    43/48

    Typical radiation pattern 315 mm Vifa NE

    woofer NE315W-04

    Confidential 43

    315 mm Vifa NE

  • 8/3/2019 Observations of Directivity in Transducers

    44/48

    315 mm Vifa NE

    NE315W-04

    Confidential 44

  • 8/3/2019 Observations of Directivity in Transducers

    45/48

    Conclusions The method of calculating the directivity index and radiated power was

    reviewed.

    The effects of resonances (break-up modes etc.) in the frequencyresponse of a transducer can be identified by comparing the on-axis

    response curve to the predicted fall-off curve, due to the increase inimpedance with frequency.

    The directivity index function for a piston of a particular size can beused as an approximate directivity index function for a transducer of

    same-sized diaphragm, up to the point where the diaphragm goes intobreak-up and the effective radiating area of the transducer decreases.

    Transducer off-axis frequency response and directivity can beimproved through mechanical resonances in the diaphragm, but thereare trade-offs:

    It is difficult to have a well-behaved polar radiation pattern.

    Mechanical resonances in the diaphragm can also produce undesiredlarge peaks in the on-axis frequency response.

    Mechanical resonances are often poorly dampedhard to control.

    Confidential 45

  • 8/3/2019 Observations of Directivity in Transducers

    46/48

    Confidential 46

  • 8/3/2019 Observations of Directivity in Transducers

    47/48

  • 8/3/2019 Observations of Directivity in Transducers

    48/48

    Acoustics, Leo Beranek, 1993 edition.

    Confidential 48