Upload
religious-brainwash
View
234
Download
0
Embed Size (px)
Citation preview
8/3/2019 Observations of Directivity in Transducers
1/48
1
Observations ofDirectivity in Transducers
Richard LittleDirector of Advanced Design
Tymphany HK Ltd.
13-14 November 2010Acoustic Block 2010
The Sound of Modern Design
8/3/2019 Observations of Directivity in Transducers
2/48
2
Tymphany HK Ltd.
13-14 November 2010Acoustic Block 2010
The Sound of Modern Design
8/3/2019 Observations of Directivity in Transducers
3/48
Abstract
The causes of transducer directivity arereviewed. The basic theoretical performance of apiston radiating from an infinite baffle is discussed,including calculations of directivity and total radiatedpower. Measurements of transducer directivity and
radiated power are reviewed for a range of transducersizes. The divergences of these measurement results,from the expectations for a piston, are examined anddiscussed.
Confidential 3
8/3/2019 Observations of Directivity in Transducers
4/48
Confidential 4
8/3/2019 Observations of Directivity in Transducers
5/48
Introduction to Directivity A transducers diaphragm moves back and forth during operation,
pushing and pulling on the surrounding air and creating an outgoing
pressure wave.
At low frequencies, the wavelength of the outgoing sound pressure wave
is large relative to the diaphragm, and the diaphragm operates much likea point source, radiating sound pressure waves evenly in all directions.
At higher frequencies, however, the wavelength of the pressure wave issized similarly to the diaphragm size, or even smaller. At different
observation positions, at these frequencies, there is a path lengthdifference from the observation position to different parts of thetransducers diaphragm. This means that the summed contributions tothe total pressure observed may increase or decrease, according to theeffects of the path length differences.
Typically a transducer diaphragm is symmetric around its central axis, sothe effects of the path length differences are observed to be differentaccording to the direction that the transducer is being observed, relativeto the central axis. We call this effect DIRECTIVITY.
Confidential 5
8/3/2019 Observations of Directivity in Transducers
6/48
Confidential 6
8/3/2019 Observations of Directivity in Transducers
7/48
Directivity of a Piston in an Infinite Baffle
In this paper, were going to limit measurements and calculationsto that of transducers and pistons operating mounted to an infinitebaffle. Were going to neglect reflections off of loudspeaker
cabinet boundaries. The figure below illustrates how the pressure at the observation
position, created by a radiating piston mounted in an infinite baffle,is calculated by integrating over the radiating surface.
Confidential 7
Infinitebaffle
piston
Observation position
8/3/2019 Observations of Directivity in Transducers
8/48
Confidential 8
Infinitebaffle
piston
Observation position
8/3/2019 Observations of Directivity in Transducers
9/48
Example: radiation pattern on and off axis
This example shows the radiation pattern for a transducer, on and offaxis, at different frequencies. The output of the transducer is the same
in all directions at low frequencies, but at high frequencies this is nolonger the case.
Confidential 9
8/3/2019 Observations of Directivity in Transducers
10/48
Confidential 10
8/3/2019 Observations of Directivity in Transducers
11/48
Calculating directivity for a infinite-baffle piston
The Directivity Index DI is a measure of the directivity of atransducer, and can be calculated from the following formulae:
Here a refers to the diameter of the diaphragms radiating area,and c is the speed of sound (345 m/s).
These formulae are approximations to more precise Besselfunction relations.
Confidential 11
8/3/2019 Observations of Directivity in Transducers
12/48
8/3/2019 Observations of Directivity in Transducers
13/48
Calculating directivity from measurements
The directivity factor Q is the ratio of the on axis frequencyresponse of a transducer, to the on axis frequency response for a
point source radiator producing the same amount of total radiatedpower. Q and DI are related:
Q is calculated by calculating the total radiated power, in a ratiowith the on-axis frequency response:
For a transducer which is symmetric around its central axis, thiscan be re-stated as:
Confidential 13
8/3/2019 Observations of Directivity in Transducers
14/48
Q QDI
Q
Q
Confidential 14
8/3/2019 Observations of Directivity in Transducers
15/48
Calculating directivity from measurements (2)
Typically, when conducting measurements, we measure off-axisin small increments of . The previous relation for Q can then be
approximated as follows:
We can then calculate DI from Q.
The results of these calculations of the DI can be comparedagainst the piston theoretical curve. Good agreement betweenthe measurement results, vs. the theoretical curve, implies thatthe transducers total radiation pattern is pistonic.
Confidential 15
8/3/2019 Observations of Directivity in Transducers
16/48
(2)
0180Q
DI
DIDI
Confidential 16
8/3/2019 Observations of Directivity in Transducers
17/48
Example: measured directivity vs. piston
theoretical directivity In this example, the transducers radiation pattern is pistonic, for its
radiating area, until ~ 6 kHz. Above that frequency, the driver is
effectively functioning as if its radiating area were smaller than it really is.
Confidential 17
8/3/2019 Observations of Directivity in Transducers
18/48
6
Confidential 18
8/3/2019 Observations of Directivity in Transducers
19/48
Break-up modes and on-axis measurements
The effect of diaphragm resonances(break-up modes) on the on-axis
frequency response can be studiedby mirroring the transducers
impedance curve onto the on-axisfrequency response curve, startingat the impedance minimum
frequency. Large deviationsbetween the impedance-projectedcurve and the on-axis responsecurve suggest mechanical
resonances have altered the
frequency response. The example to the right shows two
possible resonances: a shallowbroad resonance at 7 kHz, and asharp high-Q resonance at 19 kHz.
Confidential 19
8/3/2019 Observations of Directivity in Transducers
20/48
7KHzQ19KHzQ
Confidential 20
8/3/2019 Observations of Directivity in Transducers
21/48
On and off axis measurements
A transducers radiation pattern will become directional at frequenciesabove ka=1 (from our previous equation). The graph below highlightsthis point in the frequency response (the radiation dispersion frequency).
Confidential 21
8/3/2019 Observations of Directivity in Transducers
22/48
(ka=1)()
Confidential 22
8/3/2019 Observations of Directivity in Transducers
23/48
Break-up modes and directivity
A case study Were going to compare the radiation patterns of two drivers, with
essentially the same diaphragm, but two different voice coil diameters(one smaller, one bigger).
A sketch of the diaphragm and voice coil are shown below. Because ofthe difference in coil sizes, the break-up frequency of the diaphragmshould be different in the two cases.
Confidential 23
voice coil
diaphragm
surround
voice coil
diaphragm
surround
Small coil Large coil
8/3/2019 Observations of Directivity in Transducers
24/48
Confidential 24
voice coil
diaphragm
surround
voice coil
diaphragm
surround
Small coil Large coil
8/3/2019 Observations of Directivity in Transducers
25/48
Case study comparison of on-axis response
and break-up mode behaviorSmall coilBreak-up modes small until 19 kHz
Large coilBreak-up modes affect response above 4 kHz
Confidential 25
8/3/2019 Observations of Directivity in Transducers
26/48
-
19KHz 4KHz
Confidential 26
8/3/2019 Observations of Directivity in Transducers
27/48
Case study: comparison of off-axis frequency
responseSmall coilResponse off-axis remains roughly pistonic untilthe effects of the resonance at 19 kHz enhance
the off-axis response
Large coilOff-axis performance kept close to on-axisperformance due to effects of ~12 kHzresonance
Confidential 27
8/3/2019 Observations of Directivity in Transducers
28/48
-
19KHz 12KHz
Confidential 28
8/3/2019 Observations of Directivity in Transducers
29/48
Case study: Directivity index comparison
Small coilDriver performs roughly like a piston until above10 kHz
Large coilDriver shows large diversions from pistonicbehavior above 10 kHz, effectively losingradiating area due to mechanical resonances.
Confidential 29
8/3/2019 Observations of Directivity in Transducers
30/48
-
10KHz 10KHz
Confidential 30
8/3/2019 Observations of Directivity in Transducers
31/48
Case study: radiation pattern comparison
Small coil Large coil
Confidential 31
8/3/2019 Observations of Directivity in Transducers
32/48
-
Confidential 32
8/3/2019 Observations of Directivity in Transducers
33/48
Notes from the case study
Small coil
Smaller coil = cheaper motor
Lower coil inductance
Higher on-axis frequencyresponse bandwidth
Low resonance design more repeatable response inproduction
Higher directivity at higherfrequencies
Large coil
Large coil = expensive motor
Higher coil inductance
Lower on-axis frequencyresponse bandwidth
Design depends uponresonances
Lower directivity at higherfrequencies
Confidential 33
8/3/2019 Observations of Directivity in Transducers
34/48
=
=
Confidential 34
8/3/2019 Observations of Directivity in Transducers
35/48
Typical radiation pattern 25 mm Vifa NE silk
dome tweeter NE25VTS-04
Confidential 35
25 Vif NE
8/3/2019 Observations of Directivity in Transducers
36/48
25 mm Vifa NE
NE25VTS-04
Confidential 36
8/3/2019 Observations of Directivity in Transducers
37/48
Typical radiation pattern 25 mm ring
radiator tweeter XT25SC90-04
Confidential 37
25
8/3/2019 Observations of Directivity in Transducers
38/48
25 mm
XT25SC90-04
Confidential 38
T i l di ti tt NXT BMR46 f ll
8/3/2019 Observations of Directivity in Transducers
39/48
Typical radiation pattern NXT BMR46 full-
range driver
Confidential 39
8/3/2019 Observations of Directivity in Transducers
40/48
NXT BMR46
Confidential 40
T i l di ti tt 180 Vif NE
8/3/2019 Observations of Directivity in Transducers
41/48
Typical radiation pattern 180 mm Vifa NE
woofer NE180W-04
Confidential 41
180 Vif NE
8/3/2019 Observations of Directivity in Transducers
42/48
180 mm Vifa NE
NE180W-04
Confidential 42
Typical radiation pattern 315 mm Vifa NE
8/3/2019 Observations of Directivity in Transducers
43/48
Typical radiation pattern 315 mm Vifa NE
woofer NE315W-04
Confidential 43
315 mm Vifa NE
8/3/2019 Observations of Directivity in Transducers
44/48
315 mm Vifa NE
NE315W-04
Confidential 44
8/3/2019 Observations of Directivity in Transducers
45/48
Conclusions The method of calculating the directivity index and radiated power was
reviewed.
The effects of resonances (break-up modes etc.) in the frequencyresponse of a transducer can be identified by comparing the on-axis
response curve to the predicted fall-off curve, due to the increase inimpedance with frequency.
The directivity index function for a piston of a particular size can beused as an approximate directivity index function for a transducer of
same-sized diaphragm, up to the point where the diaphragm goes intobreak-up and the effective radiating area of the transducer decreases.
Transducer off-axis frequency response and directivity can beimproved through mechanical resonances in the diaphragm, but thereare trade-offs:
It is difficult to have a well-behaved polar radiation pattern.
Mechanical resonances in the diaphragm can also produce undesiredlarge peaks in the on-axis frequency response.
Mechanical resonances are often poorly dampedhard to control.
Confidential 45
8/3/2019 Observations of Directivity in Transducers
46/48
Confidential 46
8/3/2019 Observations of Directivity in Transducers
47/48
8/3/2019 Observations of Directivity in Transducers
48/48
Acoustics, Leo Beranek, 1993 edition.
Confidential 48