2
Analysis of Sulfur in Oils according to ASTM D2622 Method Thermo Scientific ARL PERFORM'X Series Advanced X-Ray Fluorescence Spectrometers Introduction In order to demonstrate the analysis of sulfur (S) in oil, a calibration curve has been prepared according to the ASTM D2622 norm and repeatability tests done. Instrument Thermo Scientific ARL PERFORM’X series spectrometer used in this analysis was a 4200 watt system. This system is configured with 6 primary beam filters, 4 collimators, up to nine crystals, two detectors, helium purge and our 5GN+ Rh X-ray tube for best performance from ultra-light to heaviest elements thanks to its 50 micron Be window. This new X-ray tube fitted with a low current filament ensures an unequalled analytical stability month after month. The ARL PERFORM'X offers the ultimate in performance and sample analysis safety. Its unique LoadSafe design includes a series of features that prevent any trouble during sample flushing and loading. Liquid cassette recognition prevents any liquid sample to be exposed to vacuum by mistake. Over exposure safety automatically ejects a liquid sample if X-ray exposure time is too long. The Secutainer system protects the primary chamber by vacuum collecting any loose powders in a specially designed container, easily removed and cleaned by any operator. For spectral chamber protection, the ARL PERFORM'X uses a helium shutter designed for absolute protection of your goniometer during liquid analysis under helium operation. In the "LoadSafe Ultra" optional configuration, a special X-ray tube shield provides total protection against sample breakage or liquid cell rupture. Sulfur analysis in oils The ASTM D2622 norm has been followed and six standard samples have been prepared in order to construct a calibration curve. The analytical conditions were set according to the ASTM norms. The S net intensities have been determined with the following formula: S net = (S Ka1, 2 – S background) The intensity of each standard has been correlated with the S content in ppm, which resulted in the calibration curve shown in Figure 1. Table 1 gives the numerical results including the absolute difference between nominal and calculated concentrations, as well as the Standard Error of Estimate. Sample # S net Nom. Calcul. Absolute Conc. % Conc.% diff. % 502 25.031 0.502 0.497 - 0.005 998 48.647 0.998 1.009 0.011 2045 92.567 2.045 2.042 - 0.003 3009 129.271 3.009 2.995 - 0.014 4027 166.437 4.027 4.045 0.018 4986 196.284 4.986 4.979 - 0.007 Standard error of estimate: 0.015% Table 1: Calibration results Figure 1: S calibration curve Key Words • ARL PERFORM'X 4200 W • Oil • X-ray fluorescence • XRF • ASTM D2622 Application Note: 41667

Note: 41667 D2622 Method - Thermo Fisher Scientific · PDF fileAnalysis of Sulfur in Oils according to ASTM D2622 Method Thermo Scientific ARL PERFORM'X Series Advanced X-Ray Fluorescence

Embed Size (px)

Citation preview

Analysis of Sulfur in Oils according to ASTMD2622 MethodThermo Scientific ARL PERFORM'X Series Advanced X-Ray Fluorescence Spectrometers

IntroductionIn order to demonstrate the analysis of sulfur (S) in oil, acalibration curve has been prepared according to theASTM D2622 norm and repeatability tests done.

InstrumentThermo Scientific ARL PERFORM’X series spectrometerused in this analysis was a 4200 watt system. This systemis configured with 6primary beam filters, 4collimators, up to ninecrystals, two detectors,helium purge and our5GN+ Rh X-ray tube forbest performance fromultra-light to heaviestelements thanks to its 50micron Be window. Thisnew X-ray tube fitted witha low current filamentensures an unequalledanalytical stability monthafter month.

The ARL PERFORM'Xoffers the ultimate in performance and sample analysissafety. Its unique LoadSafe design includes a series offeatures that prevent any trouble during sample flushingand loading. Liquid cassette recognition prevents anyliquid sample to be exposed to vacuum by mistake. Overexposure safety automatically ejects a liquid sample ifX-ray exposure time is too long.

The Secutainer system protects the primary chamberby vacuum collecting any loose powders in a speciallydesigned container, easily removed and cleaned by anyoperator. For spectral chamber protection, the ARLPERFORM'X uses a helium shutter designed for absoluteprotection of your goniometer during liquid analysisunder helium operation. In the "LoadSafe Ultra" optionalconfiguration, a special X-ray tube shield provides totalprotection against sample breakage or liquid cell rupture.

Sulfur analysis in oils The ASTM D2622 norm hasbeen followed and six standardsamples have been prepared inorder to construct a calibrationcurve. The analytical conditionswere set according to the ASTMnorms.

The S net intensities have beendetermined with the followingformula:S net = (S Ka1, 2 – S background)

The intensity of each standard has been correlatedwith the S content in ppm, which resulted in thecalibration curve shown in Figure 1. Table 1 gives thenumerical results including the absolute differencebetween nominal and calculated concentrations, as well asthe Standard Error of Estimate.

Sample # S net Nom. Calcul. AbsoluteConc. % Conc.% diff. %

502 25.031 0.502 0.497 - 0.005998 48.647 0.998 1.009 0.0112045 92.567 2.045 2.042 - 0.0033009 129.271 3.009 2.995 - 0.0144027 166.437 4.027 4.045 0.0184986 196.284 4.986 4.979 - 0.007Standard error of estimate: 0.015%

Table 1: Calibration results

Figure 1: S calibration curve

Key Words

• ARL PERFORM'X4200 W

• Oil

• X-ray fluorescence

• XRF

• ASTM D2622

ApplicationNote: 41667

Stability test on a sample of oil residuecontaining sulfurA sulfur stability test has been made on an oil residuesample, loaded into three different liquid cells: A, B and C.

Each cell was analyzed three times; the results areshown in Table 2.

Run # Cell A Cell B Cell CS concentration in % S concentration in % S concentration in %

1 3.60 3.59 3.552 3.61 3.60 3.563 3.59 3.58 3.57Average 3.60 3.59 3.56St.Dev. 0.008 0.011 0.008

Table 2: Stability test on an oil residue sample

Examination of these results reveals the excellentrepeatability of analysis that can be achieved for S in oilwith the ARL PERFORM'X XRF.

Analysis of unknown oil samplesFour unknown samples were prepared and each wasloaded into three different cells: crude oil, heavy oil andtwo different oil residues.

Table 3 shows the repeatability results obtained aswell as a comparison with the certified sulfurconcentration.

Sample Certified Specimen Specimen Specimen Average St.Dev. RSDvalue % #1 % #2 % #3 % % % %

Crude oil 2.05 1.998 2.000 1.999 1.999 0.002 0.121Heavy oil 2.10 2.072 2.063 2.068 2.068 0.004 0.215Oil residue 1 4.6 4.785 4.772 4.771 4.776 0.008 0.169Oil residue 2 3.6 3.612 3.616 3.600 3.609 0.008 0.235

Table 3: Analytical results and comparisons for 4 unknown oil samples

ConclusionGood calibration curves can be obtained with the ARLPERFORM'X XRF for sulfur determination in oil whenapplying the appropriate ASTM standard method.

Thanks to the reproducibility of loading and to theinnovative helium shutter protecting the goniometerchamber from the helium environment, high repeatabilityof analysis can be demonstrated for S analysis.

The results obtained either for concentrations or forrepeatability show that very good accuracy and precisioncan be achieved with the ARL PERFORM'X XRFinstrument. This instrument is well suited for the analysisof S in petrochemical products.

To see our complete X-ray product range, visitwww.thermoscientific.com/xray

Part of Thermo Fisher Scientific

Thermo Fisher Scientific (Ecublens)SARL, Switzerland is ISO certified.

In addition to these

offices, Thermo Fisher

Scientific maintains

a network of represen -

tative organizations

throughout the world.

Africa+27 11 822 4120Australia+61 3 9757 4300Austria+43 1 333 50 34 0Belgium+32 53 73 42 41Canada+1 800 530 8447China+86 10 8419 3588Denmark+45 70 23 62 60 Europe-Other+43 1 333 50 34 0Finland /Norway /Sweden+46 8 556 468 00France+33 1 60 92 48 00Germany+49 6103 408 1014India+91 22 6742 9434Italy+39 02 950 591Japan +81 45 453 9100Latin America+1 561 688 8700Middle East+43 1 333 50 34 0Netherlands+31 76 579 55 55New Zealand+64 9 980 6700Russia/CIS+43 1 333 50 34 0South Africa+27 11 570 1840Spain+34 914 845 965Switzerland+41 21 694 71 11UK+44 1442 233555USA+1 800 532 4752

www.thermoscientific.com©2012 Thermo Fisher Scientific Inc. All rights reserved. Windows is a registered trademark of Microsoft corp. in United States and other countries. ISO is a trademark of the International StandardsOrganisation. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in allcountries. Please consult your local sales representative for details.

AN41667_E 02/12C