270
Multilayered Scaffolds for Osteochondral Tissue Engineering Based on Bioactive Glass and Biodegradable Polymers Mehrlagige Gerüststrukturen für das Knochen-Knorpel Tissue Engineering basierend auf bioaktivem Glas und bioabbaubaren Polymeren Der Technischen Fakultät der Friedrich-Alexander-Universität Erlangen-Nürnberg zur Erlangung des Doktorgrades DOKTOR – INGENIEUR vorgelegt von Patcharakamon Nooeaid aus Trang

Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

Multilayered Scaffolds for Osteochondral Tissue Engineering

Based on Bioactive Glass and Biodegradable Polymers

Mehrlagige Gerüststrukturen für das Knochen-Knorpel Tissue

Engineering basierend auf bioaktivem Glas und bioabbaubaren

Polymeren

Der Technischen Fakultät

der Friedrich-Alexander-Universität

Erlangen-Nürnberg

zur

Erlangung des Doktorgrades DOKTOR – INGENIEUR

vorgelegt von

Patcharakamon Nooeaid

aus Trang

Page 2: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

Als Dissertation genehmigt

von der Technischen Fakultät

der Friedrich-Alexander-Universität Erlangen-Nürnberg

Tag der mündlichen Prüfung: 15.05.2014

Vorsitzende des Promotionsorgans: Prof. Dr.-Ing. habil. Marion Merklein

Gutachter: Prof. Dr.-Ing. habil. Aldo R. Boccaccini

Prof. Dr. rer. nat. Peter Greil

Page 3: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CONTENTS

ZUSAMMENFASSUNG ..........................................................................................................vii

CHAPTER 1 Introduction ........................................................................................................ 1

CHAPTER 2 State of the Art and Literature Review ............................................................ 3

2.1 Characteristics of the osteochondral interface .............................................................. 3

2.2 Scaffolds for osteochondral tissue engineering ............................................................. 6

2.2.1 Scaffold materials ................................................................................................. 6

2.2.2 Scaffold fabrication techniques ........................................................................... 16

2.2.3 Strategies of multilayered scaffold ...................................................................... 26

2.3 Cells and bioactive molecules/growth factors for osteochondral tissue engineering .. 35

CHAPTER 3 Objectives and Outline ....................................................................................39

CHAPTER 4 Preparation and Characterization of Biodegradable Polymer Coated 45S5

Bioglass-Based Scaffolds for Subchondral Bone Tissue Engineering Applications ..43

4.1 Introduction .................................................................................................................. 43

4.2 Materials and methods................................................................................................. 44

4.2.1 Fabrication of 45S5 Bioglass®-based scaffolds ................................................. 44

4.2.2 Preparation of biodegradable polymer coated 45S5 Bioglass®-based scaffolds 45

4.2.3 Characterization and mechanical testing ............................................................ 46

4.2.4 Statistical analysis............................................................................................... 48

4.3 Results and discussion ................................................................................................ 48

4.3.1 Morphology ......................................................................................................... 48

4.3.2 Mechanical properties ......................................................................................... 57

4.3.3 Degradation behavior ......................................................................................... 59

4.3.4 In vitro bioactivity ................................................................................................ 62

4.4 Conclusions ................................................................................................................. 70

CHAPTER 5 Development of 45S5 Bioglass®-based Scaffolds for Controlled Antibiotic

Released in Bone Tissue Engineering via Biodegradable Polymer Layered Coating ...71

5.1 Introduction .................................................................................................................. 71

Page 4: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

iv

5.2 Materials and methods................................................................................................. 72

5.2.1 Fabrication of TCH-loaded layered biodegradable polymer coated Bioglass®-

based scaffolds ............................................................................................................ 72

5.2.2 Characterization and testing ............................................................................... 73

5.2.3 Statistical analysis............................................................................................... 75

5.3 Results and discussion ................................................................................................ 75

5.3.1 Surface property of polymeric coatings .............................................................. 75

5.3.2 Morphology ......................................................................................................... 78

5.3.3 Mechanical properties ......................................................................................... 79

5.3.4 Chemical structure .............................................................................................. 81

5.3.5 In vitro drug release ............................................................................................ 82

5.4 Conclusions ................................................................................................................. 87

CHAPTER 6 Porous Biodegradable Polymer-based Scaffolds for Cartilage Tissue

Engineering Applications .....................................................................................................89

6.1 Introduction .................................................................................................................. 89

6.2 Materials and methods................................................................................................. 91

6.2.1 Fabrication of Alg-foams ..................................................................................... 91

6.2.2 Fabrication of PLLA fibers and Alg-based fibers ................................................ 92

6.2.3 Characterization and testing ............................................................................... 92

6.2.4 Statistical analysis............................................................................................... 95

6.3 Results and discussion ................................................................................................ 96

6.3.1. Effect of processing conditions on the physical and mechanical properties of the

foams ........................................................................................................................... 96

6.3.2 Effect of electrospinning conditions on the properties of fibers ........................ 108

6.4 Conclusions ............................................................................................................... 121

CHAPTER 7 Multilayered Scaffolds Suitable for Osteochondral Tissue Engineering 123

7.1 Introduction ................................................................................................................ 123

7.2 Materials and methods............................................................................................... 125

7.2.1 Fabrication of multilayered scaffolds ................................................................ 125

Page 5: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

v

7.2.2 Characterization and testing ............................................................................. 127

7.3 Results and discussion .............................................................................................. 128

7.3.1 Microstructure ................................................................................................... 128

7.3.2 Interfacial strength of multilayered scaffolds .................................................... 132

7.3.3 Mechanical properties of integrated bilayered scaffolds .................................. 133

7.3.4 In vitro bioactivity .............................................................................................. 134

7.4 Conclusions ............................................................................................................... 139

CHAPTER 8 Biological Response of Osteoblasts Culturing on Bioglass-based

Scaffolds for Bone Regeneration ......................................................................................141

8.1 Introduction ................................................................................................................ 141

8.2 Material and methods ................................................................................................ 142

8.2.1 Fabrication of Bioglass®-based scaffolds ......................................................... 142

8.2.2 In vitro cell culture ............................................................................................. 143

8.2.3 Characterization techniques ............................................................................. 143

8.2.4 Statistical analysis............................................................................................. 146

8.3 Results and discussion .............................................................................................. 146

8.3.1 LDH activity ....................................................................................................... 146

8.3.3 Metabolic activity............................................................................................... 149

8.3.4 Osteoblastic activity .......................................................................................... 150

8.3.5 Cell morphology ................................................................................................ 152

8.4 Conclusions ............................................................................................................... 155

CHAPTER 9 Biological Response of Chondrocytes and Mesenchymal Stem Cells on

Alginate/Chondroitin Sulfate Scaffolds for Cartilage Regeneration ..............................157

9.1 Introduction ................................................................................................................ 157

9.2 Materials and methods............................................................................................... 159

9.2.1 Fabrication of Alg/ChS-foams ........................................................................... 159

9.2.2 Characterization and testing ............................................................................. 159

9.2.3 Release of ChS ................................................................................................. 161

9.2.4 In vitro culturing of primary porcine chondrocytes and human MSCs .............. 162

Page 6: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

vi

9.2.5 Statistical analysis............................................................................................. 165

9.3 Results and discussion .............................................................................................. 165

9.3.1 Characterization of Alg/ChS-foams .................................................................. 165

9.3.2 Release profile of ChS molecules..................................................................... 174

9.3.3 The influence of culturing conditions on the primary porcine chondrocytes

activity ........................................................................................................................ 175

9.3.4 The influence of ChS molecules on chondrogenic differentiation of chondrocytes

and MSCs .................................................................................................................. 178

9.3.5 The influence of chondrogenic induction (TGF-1) on the activity of MSCs .... 183

9.4 Conclusions ............................................................................................................... 185

CHAPTER 10 Summary and Future perspectives ...........................................................187

REFERENCES ......................................................................................................................193

LIST OF FIGURES .................................................................................................................... I

LIST OF TABLES ................................................................................................................... XI

ABBREVIATIONS AND SYMBOLS ..................................................................................... XIII

ACKNOWLEDGEMENTS ..................................................................................................... XXI

LIST OF PUBLICATIONS .................................................................................................. XXV

Page 7: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

ZUSAMMENFASSUNG

In der vorliegenden Dissertation wurden mehrlagige Scaffolds, die für die

Gewebeentwicklung an Grenzflächen geeignet sind, z. B. Knochen-Knorpel-Regeneration,

hergestellt und im Detail diskutiert. Ihre Bauweise, poröse Struktur, physiochemische und

mechanische, sowie biologische Eingeschaften wurden umfassend betrachtet.

Bioglas®-basierte Schäume wurden als Grundlagematerial der Scaffold für den

unterknorpeligen Knochenteil ausgewählt. 3D Bioglas®-basierte poröse Scaffold, die eine

der Spongiosa ähnliche Bauweise und poröse Struktur aufweisen, wurden mit der

Schaumnachbildungsmethode hergestellt. Außerdem konnte die mechanische Festigkeit

und strukturelle Stabilität der Scaffold durch die Beschichtung mit bioabbaubaren Polymeren

verbessert werden. Es wurden verschiedene Polymerbeschichtungen untersucht, dazu

gehören Alginat (Alg), Gelatine (Gel), PDLLA und PHBHHx Beschichtungen, die im

Vergleich zu unbeschichteten Bioglas®-basierten Scaffolds zu einer Erhöhung des E-Moduls

und der Druckfestigkeit führen. Ergänzend dazu haben solche Scaffolds die Bioaktivität

unterstützt, welche durch die Bildung von HA in der SBF Lösung nachgewiesen wurde.

Demzufolge stellen alle in dieser Arbeit entwickelten Scaffolds geeignete Kandidaten für die

Knochenregeneration dar. Darüber hinaus können die polymerbeschichteten Bioglas®-

basierten Scaffolds als Träger von Medikamenten / Biomolekülen, z.B. Überbringer von

Antibiotika, als Anwendung in der Knochengewebeentwicklung dienen. Multifunktionelle

Scaffolds, die auf auf TCH-beladenen Polymerschichten basieren und mit dem Bioglas-

basiertem Scaffolds beschichtet sind, haben im Vergleich zu unbeschichteten Scaffolds eine

verbesserte mechanische Festigkeit aufgewiesen und eine kontrollierte

Medikamentenausscheidung über 14 Tage nach Eintauchen in PBS. Die biologischen

Eigenschaften von Alg-beschichteten Bioglas®-basierten Scaffolds wurden durch die

Züchtung mit Osteoblasten (MG-63) ausgewertet, mit der Zielsetzung ihre Biokompatibilität

und Fähigkeit zur Knochenmineralisierung zu bestätigen. Im Vergleich zu unbeschichteten

und RGD-modifizierten Alg-beschichteten Bioglas®-basierten Scaffolds, weisen Alg-

Page 8: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

viii

beschichtete Scaffolds eine gute Biokompatibilität auf und fördern das Zellwachstum und die

Osteoblastenaktivität.

In Bezug auf die Knorpelphase im Knochen-Knorpel-Gewebe war Alg

höchstinteressant, weil seine chemische Struktur jener von der Hyaluronsäure (HyA) ähnelt,

wenn in Erwägung gezogen wird, dass HyA die Hauptkomponente in knorpligen ECM ist.

Poröse Säulenstrukturen aus 3D-Schäumen wurden erfolgreich hergestellt, um die Migration

und Anordnung von Zellen, und den anschließenden Aufbau von neuem Gewebe durch die

Optimierung der Polymerkonzentration und der Gefriertrocknungsbedingungen zu

unterstützen. Optimierte 3 Gew./Vol.% Alg-Schäume wiesen eine Porengröße im Bereich

von 125 - 325 µm auf, was für die Unterstützung der Besämung und Migration von

Chondrozyten geeignet ist. Ergänzend dazu waren die Schäume in der Lage Wasser in der

gleichen Größenordnung wie der native Knorpel (~ 80 %) zu absorbieren. Die mechanische

Festigkeit und strukturelle Stabilität der Schäume wurde durch den Einsatz ionischer

Vernetzung verbessert. Der E-Modul und die Druckfestigkeit der Schäume lagen

entsprechend bei 0.220 ± 0.009 and 0.14 ± 0.02 MPa, was im Größenbereich des nativen

Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa).

Wesentlich dabei ist, dass die Schäume nicht mineralisiert waren, was bedeutet, dass sie in

Kontakt mit Körperflüssigkeiten keine Knochenbildung hervorrufen können, was aber für die

Knorpelregeneration notwendig ist. Gemäß des Diagramms in Abbildung 10.1, welches die

entsprechenden Kernpunkte bei der Herangehensweise in der Knorpelgewebeentwicklung

zusammenfasst, haben die in der vorliegenden Arbeit angefertigten porösen Schäume die

meisten gerüstbezogenen Kriterien erfüllt. Allerdings mangelt es den Schäumen an

Zelladhäsion, was sich auf die Vermehrung und Differenzierung von Zellen negativ auswirkt.

Daher wurden die Alg-Schäume erstmals durch die Einbindung von biologischen

Signalstoffen, z. B. Chrondroitinsulfat (ChS), modifiziert, mit dem Ziel die Zelladhäsion und

das Verhalten der Zellen zu verbessern. ChS ist eines von natürlichen

Glycosaminoglycanen im Knorpel, welches die Funktion hat den Metabolismums von

Chondrozyten durch die Induzierung der Synthese vom Typ II Kollagen (Col II) und

Proteglykanen (PGs) zu stimulieren. Die Alg/ChS-Schäume haben entweder die

Page 9: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

ix

Chondrozyten oder die MSCs unterstützt und die Zellvermehrung und Zelldifferenzierung.

Die Ausscheidung von Col II und PGs von Chondrozyten- und MSCs, die auf Alg/ChS-

Schäumen besamt wurden, wurden als Marker der Knorpelregeneration charakterisiert.

Diese Ergebnisse haben die wichtigsten zellenbezogenen Anforderungen erfüllt (Abbildung

I). Außerdem haben die Alg/ChS-Schäume, in welchen das ChS als biologischer Signalstoff

dient, assoziiert durch den Einsatz von TGF-β1 eine nennenswerte Steigerung der

Chondrogenese von MSCs begünstigt. Dieses Ergebnis deutete darauf hin, dass die

Einbingung von Biomolekülen (ChS) in Kombination mit Wachstumsförderern (TGF-β1) eine

wichtige Rolle in Hinsicht auf die knorpelige Differenzierung and anschließende

Matrixproduktion spielt. Trotzdem hat das ChS die Zelladhäsion weniger als erwartet

verbessert. Das liegt möglicherweise an der geringen Menge von ChS, welches in die Alg-

Schäume eingebunden wurde. Diese Menge vermag nicht ausreichend genug sein, um von

den Zellen erkannt zu werden. Folglich tendierten die Chondrozyten und MSCs innerhalb

der Poren Klumpen zu bilden, aber haben sich kaum an die Porenwände der Schäume

angehaftet. Deswegen verbleiben einige herausfordernde Fragestellungen hinsichtlich der

drei Ecksteine bei der Vorangehensweise der Gewebeentwicklung offen. Als erstes ist es

notwendig die Schäume zu modifizieren (z. B. durch eine Oberflächenfunktionalisierung),

um die Zelladhäsion zu steigern. Zweitens ist es empfehlenswert die Auswirkung der ChS-

Freisetzungsrate auf die Zellabspaltung und Matrixproduktion weiterhin intensiv zu

untersuchen, in Verbindung mit der Funktion des zugegebenen Wachstumsförderers.

Page 10: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

x

Abbildung I Zusammenfassung der wichtigsten herausfordernden Aspekte im Bereich der

Knorpelgewebeentwicklung, die in der vorliegenden Dissertation untersucht wurden, sowie

Andeutung der Kriterien, welche mit den entwickelten Scaffold erfüllt wurden.

Page 11: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

xi

Mehrlagige Scaffolds wurden entwickelt, die auf optimierten Scaffold für den

Unterknorpelknochen und Knorpel basieren. Da das ideale Scaffold für die Knochen-

Knorpel-Reparatur noch nicht existiert, gewinnt die Entwicklung von Strategien, welche

längerfristig ein ausgezeichnetes Ergebnis bieten, immer mehr an Aufmerksamkeit und

erhalten einen beträchtlichen Forschungsaufwand. Der Schwerpunkt der vorliegenden Arbeit

lag daher, in Hinsicht auf das Material, auf modernen Strategien der zwei- oder mehrlagige

Scaffolds, einschließlich der integrierten zweilagigen und monolitischen zweiphasigen

Scaffolds, die auf dem Alg-Schaum und Alg-beschichteten Bioglas® Scaffolds basieren.

Obwohl es naheliegend ist, dass integrierte zweilagige Scaffolds aufgrund einer möglichen

Delamination an der Grenzfläche zwischen den Schichten einen Schwachpunkt bieten

können, hat unsere Studie nachgewiesen, dass die Delamination durch das Einfügen einer

adhäsiven Zwischenphase, welche als Grenzfläche zwischen dem ausgeprägten Knorpel

und der Knochenschichten dient, überwunden werden kann. Hinsichtlich der

Herstellungsparameter und der später insbesondere an der Grenzfläche auftretenden

porösen Struktur, konnten im Gegensatz dazu monolithische zweiphasige Scaffolds kaum

kontrolliert werden. Zusätzlich wurde im Rahmen von Grenzflächenuntersuchungen mit Hilfe

von Mikrozugversuchen nachgewiesen, dass die Grenzflächenbruchfestigkeit der

integrierten zweilagigen scaffold höher ist im Vergleich zu jener von monolitischen

zweiphasigen Scaffolds. Daher kann im Bereich der Materialien vorläufig zusammengefasst

werden, dass das in der vorliegenden Arbeit entwickelte integrierte zweilagige

Scaffoldsystem ein geeigneter Ansatz ist, um weiter als Knochen-Knorpel-Konstruktion

entwickelt zu werden.

Zusätzlich wurden mit Hilfe des Elektrospinnens Fasergewebenetze (PLLA und

Alg/Gel) hergestellt und als Knorpelphase in zweilagigen Scaffold untersucht. Die

Fasernetze hatten eine Porengröße von 50 µm (mit einer Dicke von bis zu 500 µm), wobei

mit zunehmender Netzdicke eine Abnahme der Porengröße festgestellt wurde. Die kleine

Porengröße der elektroversponnenen Fasern schränkt bekanntlich die Anwendung dieser

Fasern in GewebeentwicklungsScaffolds ein. Die kleine Porengröße der Fasernetze vermag

nach der Implantation die Zellmigration und den Nährstofftransfer hemmen. Die bedingte

Page 12: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

xii

dreidimensionale Struktur könnte für die Regeneration des Neo-Knorpels nicht passend

sein. Die Fasernetze wären dennoch ein interessanter Kandidat für die Anwendung als

kalzifizierte Knorpelschicht an der Knorpel-Knochen-Grenzfläche. Dichte Fasernetze mit

kleiner Porengröße können als Barriere gegen die Infiltration von Knochenzellen vom

Unterknorpel aggieren und in vivo die Gefäßneubildung von der Knorpelphase verhindern.

Daher ist die Strategie der mehrlagige Scaffolds für die Knochen-Knorpel-

Gewebeentwicklung vielversprechend und sollte weiter verfolgt werden, um eine mehr den

nativen Geweben ähnliche biomimetische Struktur zu erreichen.

Wie in Abbildung 10.2 gezeigt wird in der vorliegenden Arbeit ein neuartiges

mehrlagiges Scaffold mit einer anspruchsvolleren Struktur als eine Perspektive aufgezeigt,

mit dem Schwerpunkt auf funktionalen Knochen-Knorpel-Scaffolds. In der gegenwärtigen

Arbeit erfolgte die Gestaltung des Scaffold als eine Kombination von Alg-Schaum, PLLA

Fasern und PDLLA-c-BG Scaffold für Knorpel, Grenzflächen- und Unterknorpel-

Knochenphasen. Kurz dargestellt, wurde die Alg-Lösung auf die PLLA Fasernetze/PDLLa-c-

BG zweilagiges Scaffold angewandt und es kam zur Gelbildung durch den Zusatz von

CaCl2▪H2O. Nach der Gefriertrocknung bildete sich eine poröse Alg-Phase. Diese wurde auf

das Oberteil des Fasernetzes aufgebracht (Abbildung II). In diesem Fall hat das Fasernetz

als intermediäre Schicht zwischen dem Bioglas®-basierten Scaffolds und dem Alg-Schaum

aggiert, welches als Grenzfläche zwischen Knorpel und Unterknorpelknochen gedacht war

und ein dichtes ECM aufwies. Dieses Konzept wurde durch eine frühere Studie von Yunos

et al. [94] inspiriert und unsere in Abschnitt 7 präsentierten Befunde haben aufgezeigt, dass

die Bildung von HA an der Grenzfläche zwischen dem Bioglas®-basierten Scaffold und den

PLLA Fasern die mechanische Stabilität der Grenzfläche weiter verbessern. Es wird

zusätzlich empfohlen osteokonduktive Hybridfasernetze (z. B. PLLA/ Bioglas® Hybridfasern)

als Grenzflächenphase zu verwenden, anstatt von einzelnen Polymerfasern. Dieses Design

imitiert kalzifizierten Knorpel, welcher eine hohe Fähigkeit zur lokalen Mineralisierung an der

Grenzfläche bietet. Anschließend sollte eine starke Grenzfläche gebildet werden, die in der

Lage ist den Knorpel und die knochenähnlichen Schichten während der in vitro Zellzucht

and in vivo Züchtungsbedingungen zu integrieren.

Page 13: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

xiii

Es wird erwartet, dass die Entwicklungen der Scaffold und das Wissen, welches

während der vorliegenden Arbeit gewonnen wurde, hilfreich sein werden und in naher

Zukunft zu Fortschritten im Bereich der Knochen-Knorpel-Geweberegeneration führen.

Abbildung II SEM Abbildung, die ein empfohlenes mehrlagiges Gerüstmodel für

Anwendungen in der Knochen-Knorpel-Gewebeentwicklung aufzeigt, inklusive Alg-Schaum

für die Knorpelphase, PLLA Fasernnetz für die kalzifizierte Grenzflächenphase und PDLLA-

c-BG Gerüst für die Knochenphase.

Page 14: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

xiv

Page 15: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 1

Introduction

The number of research studies in the field of interface tissue engineering,

especially the cartilage-bone (osteochondral) interface [1,2], is continuously increasing given

the need for treatment large sections of the population worldwide suffering from

osteoarthritis (OA). OA, the degeneration of osteochondral tissue at the joint (Fig. 1.1),

affects around 630 million people worldwide and continues to expand as populations age [4].

Osteochondral repair remains a challenge for surgeons and researchers due to the poor

capacity of self-repair of cartilage and the limitations of present surgical techniques. Indeed

current surgical procedures, including debridement, microfracture, mosaicplasty,

periosteal/perichondrial transplantation and autologous chondrocyte implantation (ACI), are

limited in their long-term benefit and usually require second surgery [1,2]. Therefore,

scaffold-based tissue engineering is a promising approach aimed at supporting the

generation of new tissues in order to fulfill unmet clinical demands. Scaffolds based on

tailored combination of biomaterials were investigated in this study targeting the structure

and properties required for their use in osteochondral regeneration. In general,

osteochondral defects affect both the articular cartilage and the underlying subchondral

bone, which are distinct in compositional, biological, physio-chemical and mechanical

properties [3]. Moreover, cartilage tissue exhibits intrinsic complexity along its distinct four

zones (superficial, middle, deep, calcified zones), with each zone being defined by a

particular composition and organization of cells and extracellular matrix (ECM) [3–7].

Therefore, scaffold design in recent reported research are becoming more sophisticated in

terms of the combination of various biomaterials and fabrication techniques in order to

mimic the specific characteristic features of both tissue types (instead of using a single

biomaterial) [5–7]. It has become apparent that in order to design the appropriate

osteochondral scaffold, it is essential to understand well the anatomical structures and

properties of the native tissues to be regenerated. The anatomical structure and

Page 16: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

2

characteristic properties of the tissues to be regenerated will guide the design of the scaffold

and will dictate the specific requirements of ideal scaffolds. The design of the scaffolds will

lead to the selection of suitable biomaterials and fabrication techniques, which will

prominently influence the structural and mechanical properties of the designed scaffolds.

The design of scaffolds and the fabrication techniques were critically considered as a core of

this research. According to the characteriatics of osteochondral scaffold, the research tasks

were devided into two main parts, namely (i) scaffolds for subchondral bone layer and (ii)

scaffolds for articular cartilage layer. Bioactive glass (type 45S5) and alginate are mainly

investigated for used as scaffolds for bone and cartilage, respectively. In addition, an extra

functionality was incorporated by developing an antibiotic drug releasing capability into the

bone scaffold was included. At the same time, biological molecule, i.e. chondroitin sulfate,

was incorporated into the cartilage scaffold in order to enhance cell adhesion, proliferation

and dfiiferentiation. The bi- or multilayered scaffolds was manufactured to replicate the

characteristics of the cartilage-bone tissue interface and the different design strategies were

evaluated and compared. Finally, in vitro cell culture on the scaffolds were reported, in order

to confirm their biological and cellular responses and to assess the relative advantages and

disadvantages of the different concepts proposed, highlighting promising avenues for further

research and the clinical demand.

Figure 1. 1 The scheme of knee osteoarthritis (joint degeneration disease), which is the

result of cartilage wearing out in the load bearing joint (Image courtesy of

Commonsensehealth.com [3]).

Page 17: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 2

State of the Art and Literature Review

2.1 Characteristics of the osteochondral interface

Since the causes leading to osteochondral defect remain elusive, it is necessary to

understand the anatomical structure of the tissues involved in order to gain knowledge about

the mechanisms involved in the disease [8]. Furthermore, the characteristic properties of the

natural tissues (bone and cartilage) are vital aspects to be understood in order to seek a

suitable scaffold for the repair of the defect, according to tissue engineering approach [4,8].

The osteochondral interface involves cartilage and subchondral bone with hyaline cartilage

lying on top of cancellous bone. Cartilage functions as a protector of bone from high

stresses and acts as a reducer of friction at the edge of bone [5], as shown as a common

case of knee articulating joint in Fig. 2.1. Moreover, highly flexible characteristics of articular

cartilage are relevant considering the ability of cartilage to withstand dynamic compressive

loads several times the body weight [6]. However, cartilage has a poor capacity of

regeneration due to its highly organized structure, low chondrocyte numbers, low metabolic

rate, and restricted rate of chondrocytes to divide and migrate due to the cartilage dense

matrix [7]. On the other hand, there are numerous successful reparative approaches

available for bone [9]. Unlike cartilage, bone can be self-repaired involving induction of

vascularity and bone remodeling by osteoblasts and osteoclasts [9]. In bone defect sites,

bone marrow stem cells (BMSCs) can differentiate into bone cells which require the support

of extensive vascularity to provide nutrients and proteins to stimulate bone tissue repair [9].

Bone is a complex tissue consisting of water, type I collagen (Col I) and

hydroxyapatite (HA: Ca10(PO4)6(OH)2) crystals, in which Col I and HA provide the tissue’s

stiffness and compressive strength [10,11]. Subchondral bone (SB) is mainly composed of

bulk of Col I and it has a loosely organized porous structure, these features demonstrate the

characteristics of cancellous bone, as shown in Fig. 2.1 [12]. SB is composed of osteoblasts,

which actively secrete ECM components in order to build up the bone tissue, and

Page 18: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

4

osteoclasts which are indicative of bone resorption activities [12]. In order to engineer a

bone scaffold with an attempt to mimic the natural bone, bone tissue generation and

mineralization process in bone must be understood. Bone matrix maturation involves the

expression of alkaline phosphatase (ALP) and non-collagenous proteins (i.e. osteocalcin,

osteopontin and bone sialoprotein) [12–14]. Calcium and phosphate-binding proteins

regulate deposition of minerals by the regulation of the amount and size of HA crystals

formed [13,14]. Collagen (Col), a major component in the ECM, functions as a

microenvironment for apatite nucleation [12–14]. In general, bone can form by two different

pathways, including endochondral ossification for long bone and intramembrane ossification

for flat bone [12]. Both pathways origin from precursor cells, which follow the condensation

of the mesenchyme (cartilaginous template only occurs in the case of endochondral

ossification) and finally bone formation occurs [12]. Bone formation is an ongoing process

that alters the size and shape of bone by partial resorption of preformed bone tissue

(modeling) and simultaneous deposition of new bone (remodeling) [12]. When bone is

broken, inflammation occurs; blood is supplied to the channels of the broken area causing

swelling and bruising, which is known as hematoma [12]. Dead cells then release cytokines,

which initiate the healing process [12,15]. In concert, osteoclasts remove the dead cells and

fibroblasts form fibrocartilage as spongy material, which is called soft callus formation stage

[12,15]. Afterwards, the hard callus formation stage starts and the soft callus (cartilage)

transforms into woven bone. This stage is guided by the release of minerals such as calcium

and phosphate into the cartilage tissue [15].

Cartilage is composed of four zones, which each zone has different organizations of

chondrocytes and different orientation of Col fibrils (Fig. 2.1) [12,16–19]. First, superficial

zone, in contacting with superficial fluid, is composed of flattened chondrocytes and Col

fibrils, which Col fibrils are parallelly aligned to the articular surface. Second, middle zone

contains rounded chondrocytes and randomly aligned Col fibrils. Third, deep zone is

composed of vertical columns of chondrocytes, while Col fibrils align perpendicularly to the

articular surface. Finally, calcified zone has specific hypertrophic chondrocytes with low cell

density. The hypertrophic chondrocytes have unique ability to synthesize type X collagen

Page 19: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

5

(Col X) and calcified ECM [12,16–19]. The thin layer between non-calcified cartilage and

calcified cartilage is called tidemark, which is believed to act as nutrient diffusion through the

cartilage structure and serve as an attachment of the Col fibrils [20]. Moreover, the series of

interdigitation, which connect the calcified cartilage with SB, support the transformation of

shear stress from the articulation into tensile and compressive stresses [20]. Type II collagen

(Col II) fibrils (up to 60 % dry weight of cartilage) provide high tensile strength and withstand

shear stresses [6]. Moreover, proteoglycans (PGs) embedded within Col II fibrils ( 35 % dry

weight of cartilage) provide the ability to withstand high compressive stress.

Figure 2. 1 Anatomy of the knee joint, which is the most common case found in joint

degeneration disease (according to [21]), demonstrating arrangement of ECM and

organization of chondrocytes along different zones in cartilage (Reproduced from Nooeaid et

al. [19] with the permission of John Wiley and Sons) and showing the structure of cancellous

bone as subchondral bone (Reproduced from Meyer et al. [12] with the permission of

SPRINGER-VERLAG BERLIN/HEIDELBERG).

Page 20: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

6

2.2 Scaffolds for osteochondral tissue engineering

The design of scaffolds for osteochondral regeneration is based on the

consideration of physical and biochemical properties of the cartilage-bone interface [16].

Biochemical properties, including chemical composition, and 3D structure of scaffold

material, mostly affect the cellular behavior [16]. Physical properties, including structural

architecture and geometry, biodegradation behavior and mechanical properties, influence

both cellular activity and mechanical stability [16,22]. In regard to the zonal distinct layers of

osteochondral tissue with complex compositions and property variation, the recent trends of

osteochondral scaffolds are based on bi- or multi-layered structures [5,19,23,24]. Each layer

is customized to closely replicate the features of the specific tissues to be regenerated [24].

Singular scaffold materials, on the other hand, for example single phase either bioceramic or

polymer, have not been reported to successfully support the regeneration of osteochondral

tissue. The reason is that the single phase scaffolds lack the inherent physical structure

required and individual materials cannot achieve all requirements for osteochondral

regeneration [25,26]. The development of osteochondral scaffolds must therefore focus on

the use of composite-based structure based on multilayered and gradient structures

[6,20,27].

2.2.1 Scaffold materials

The current selection of the scaffold material is one of the main factors to be

considered in scaffold-based tissue engineering. The material needs to match the criteria of

the tissues to be regenerated or repaired. In general, suitable materials should primarily

exhibit biocompatibility, controlled biodegradability and sufficient mechanical strength

[28,29]. It should also provide a desirable environment for cell attachment, proliferation and

differentiation [28,29], which are cell functions mainly altered by the intrinsic properties of the

material. In addition, the selected biomaterial should be able to be processed economically

into desired shapes and dimensions [29]. Promising osteochondral scaffold materials will be

based on two different materials, including bioceramics and bioactive glasses, and

biodegradable polymers.

Page 21: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

7

Bioceramics and bioactive glasses are widely used as the potential candidates for

bone tissue engineering applications [30,31], since bioceramics and bioactive glasses (i.e.

synthetic HA, calcium phosphate (CaP) and Bioglass®) are bioactive, leading to bone-like

apatite formation after immersion in body fluids [30,31]. Moreover, these artificial substrates

can bond to natural bone after implantation [28,32,33]. For instance, CaP with different Ca/P

ratio such as HA, tricalcium phosphate (TCP) and biphasic calcium phosphate (BCP)

provide excellent physical properties in terms of stability, degradation rate and processibility.

Moreover, CaP shows biocompatibility and bioactivity, which is the ability to bind to bone by

the release of Ca and P ions, and they also enhance bone tissue formation. However, CaP

has low mechanical strength and brittleness, which limits their application in load bearing

devices. Bioactive glasses are experiencing increasing research efforts due to their high

bioactivity and for having significantly higher mechanical strength in comparison to most

CaP ceramics. According to the literature, bioactive glasses containing Ca- or Si-based are

the most promising for bone scaffolds due to their ions-releasing ability, high

osteoconductive and osteoinductive properties, and hydrophilic behavior [34,35]. On the

other hand, polymer-based scaffolds are not suitable for bone repair because polymers do

not exhibit osteoconductive properties as bioceramics and bioactive glasses do. In addition,

the stiffness and fracture strength of polymers are insufficient for applications in bone tissue

regeneration, compared to those of bioceramics [36].

2.2.1.1 Bioglass® and its composites

with biopolymers as bone scaffolds

45S5 Bioglass® (45 % SiO2, 24.5 % CaO, 24.5 % Na2O and 6 % P2O5 by wt.)

discovered by Hench in 1971 [31], provides excellent osteoconductivity, osteoinductivity,

bioactivity, controlled degradability and ability to deliver cells for bone tissue regeneration

[31,33,34]. 45S5 Bioglass is also able to form interfacial bonding to soft and hard tissues

[31,37]. Especially in bone bonding, 45S5 Bioglass® can bond to bone in vitro, which has

been described by the formation of a dual layer, including a silica rich layer and

hydroxycarbonate apatite (HCA) layer, on its surface in contact with body fluids [31,38]. The

formation of HCA in vitro is suggested to occur also in vivo leading to bone bonding

[31,33,39]. Hench et al. [40] have described the bonding mechanisms at the interface of

Page 22: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

8

Bioglass® substrate and body fluids via HCA formation. The reactions between Bioglass

®

surface and body fluids can be summarized into 5 stages, as described in Table 2.1 (Stage

1-5) [29]. Afterwards, the HCA layer on the Bioglass® surface supports the cellular reactions

in order to form bone (Stage 6-12). At the same time, dissolution products from Bioglass®

surface can up-regulate gene expression of osteoblasts, which this phenomenon controls

osteogenesis, leading to faster bone formation in comparison with synthetic HA [34,35].

Moreover, it has been reported that the ionic dissolution products of 45S5 Bioglass® may

induce an expression of osteoblast genes (i.e. insulin-like growth factor-2 (IGF-2)), leading to

increased cell proliferation and stimulated new bone formation [31,41]. By this fact, 45S5

Bioglass does not only exhibit osteoconductive property, but it is also an active stimulation

of osteoblasts. For instance, alginate/45S5 Bioglass composite scaffolds exhibited better

osteoblastic differentiation of osteosarcoma (MG-63) cells compared to pure alginate

scaffolds because the dissolution products of Bioglass can stimulate osteoblast proliferation

and differentiation, as evidenced by increased osteoblast markers (i.e. ALP, osteocalcin and

osteopontin) [42]. Similarly, El-Gendy et al. [43] have confirmed the osteoblastic

differentiation of human dental pulp stromal cells cultured on 3D porous 45S5 Bioglass®-

based scaffolds.The Bioglass composition and the dissolution products are the key factor

that enables the osteostimulation of osteoblasts and promotes proliferation and

differentiation of the bone cells. Price et al. [44] presented that the surface of Bioglass®

efficiently supports osteoblasts (MG-63 cells) proliferation and their functions in comparison

with the surface of titanium and cobalt chrome. This phenomenon was suggested by the

reason of different surface chemistry (related to chemical composition) of the materials. This

property affects protein adsorption and cell adhesion. In addition, since Bioglass exhibits

fast biological response in culture medium, the ion exchange takes place and induces

alkalinization of the medium, which this phenomenon has an influence on cell metabolism

[45].

Page 23: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

9

Table 2. 1 Mechanisms of bioactivity and bone bonding of Bioglass®, according to

[29,31,34,39,46].

Stages Surface reactions

1 Exchange of Na+ and K

+ with H

+ and H3O

+ from body fluids, leading to hydrolysis of silica

groups and formation of silanol (Si-OH) groups: Si-O-Na+ + H

+ Si-OH + Na

+

2 Network dissolution of silica (SiO2) in the form of silicic acid (Si(OH)4) and continued

formation of Si-OH groups: Si-O-Si + H2O 2Si-OH

3 Condensation and polymerization of silica-gel on glass surface: Si-OH + Si-OH Si-O-Si

4 Further dissolution of glass: chemisorption of amorphous Ca2+

, PO43-

ions from the glass

and the solution through silica-gel, leading to formation of amorphous CaP on the surface

of silica-gel

5 Crystallization of HCA layer: continued dissolution of glass, amorphous CaP incorporates

OH- and CO3

2- from the solution and crystallizes as HCA

6 Adsorption of growth factors on HCA layer

7 Action of macrophages

8 Attachment of osteoprogenitor cells

9 Proliferation and differentiation of osteoprogenitor cells

10 Generation of ECM by osteoblasts

11 Crystallization of ECM, forming nanocrystalline mineral and collagen on the surface of

glass

12 Proliferation of bone

Even though Bioglass has excellent bioactivity and osteoconductivity as required

for bone regeneration, in form of porous structure, Bioglass has limitations such as low

mechanical strength and intrinsic brittleness [29,33,47]. Therefore, Bioglass®-based

scaffolds for load bearing applications are mostly fabricated in composite-form by a

combination with biodegradable polymers [28,48,49]. As reviewed by Chen et al. [28],

Rezwan et al. [48] and Chen et al. [49], biodegradable polymers have been incorporated into

bioactive glass-based scaffolds in order to enhance the mechanical integrity and flexibility in

dynamic environments of injured bone. Aliphatic polyesters such as poly(L-lactide) (PLLA),

poly(D-lactide) (PDLA), poly(D, L-lactide) (PDLLA), poly(lactic-c-glycolic acid) (PLGA),

polycaprolactone (PCL) and poly(hydroxyalcanoate) (PHAs) family, and natural

Page 24: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

10

biodegradable polymers such as chitosan (CS), gelatin (Gel), collagen (Col) and alginate

(Alg) coated Bioglass®-based composite scaffolds have been reported by Boccaccini and

co-workers [50–57]. It has been established that polymer coated Bioglass-based scaffolds

exhibit significant improvement of the mechanical properties [50–57]. In addition, acidic

degradation products from synthetic biodegradable polyesters can be buffered by the

dissolution products from Bioglass®-based scaffolds [28,48,49]. On the other hand, natural-

derived biodegradable polymers provide some additional benefits such as excellent

biocompatibility, non-toxicity and ability to favor cell interaction [58].

Alginate (Alg) is a polysaccharide-based polymer obtained from marine brown

algae [59,60]. Alg is an unbranched binary copolymer consisting of (1→4) linked β-D-

mannuronic acid (M) and -L-guluronic acid (G) residues of varying composition and

sequence [61]. Alg provides useful properties for tissue engineering applications, it is

biocompatible, biodegradable, non-immunogenic, low-toxic, abundant in sources and it can

be obtained at low prices [60]. Alg can be mostly processed in the form of cation-crosslinked

Alg-gel beads/capsules to encapsulate living cells serving as a cell delivery vehicle in vivo

[62–65]. In this study, Alg was chosen as one of the suitable polymer coatings for Bioglass®-

based scaffolds. As indicated above, polymer coating was proposed to improve the

mechanical strength and fracture toughness of Bioglass®-based scaffolds. As presented by

Erol et al. [54], homogeneous Alg coating on Bioglass-based scaffolds can improve the

mechanical properties, while maintaining scaffold bioactivity. Moreover, Alg, which is

compatible for a variety of cells such as chondrocytes and osteoblasts, has been shown to

maintain the phenotype either of seeded cells or encapsulated cells [62–66]. This

characteristic is crucial for specific tissue regeneration, in particular cartilage regeneration, in

order to avoid de-differentiation of cells and subsequently to avoid the formation of

unspecified tissue like fibrocartilage [67,68]. However, Alg has no adhesive sites to cells and

does not adsorb serum proteins due to its high hydrophilicity [61,69]. Therefore, peptides

with a cell adhesive sequence-modified Alg (i.e. Arg-Gly-Asp (RGD) containing peptide)

(RGD-Alg) have been used to enhance cell adhesion on Alg [61,70]. Since amino acid

sequence RGD in fibronectin acts as a primary cell attachment cue, it has been

Page 25: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

11

demonstrated that RGD linear peptide coupling to Alg can enhance osteoblasts adhesion for

4 times when compared to unmodified Alg [71]. In addition, in the form of hydrogels, it has

been previously reported that MC3T3-E1 cells seeded into RGD-modified Alg hydrogels

promoted higher osteoblastic differentiation and mineralization compared to MC3T3-E1 cells

seeded into unmodified hydrogels, as evidenced by higher ALP activity and osteocalcin level

[72].

Gelatin (Gel) is a biomacromolecule derived from Col which is the most abundant

protein in the ECM of connective tissue, such as skin, bone and cartilage [73–75]. Recently,

Gel is used in both soft and hard tissue engineering applications, i.e. in forms of

microcapsule, microsphere, wound dressing and scaffold [73]. The use of Gel-based

scaffolds is an appealing approach in bone tissue engineering due to Gel’s biodegradability,

biocompatibility, non-immunogenic properties and relatively low cost [59,76]. When

compared with Col, Gel does not exhibit antigenicity under physiological conditions [74].

However, the poor mechanical properties and water sensitivity of Gel limit its use to non-

load-bearing applications only [59]. In order to overcome these limitations, Gel has been

used either in combination with synthetic polymers or with the application of chemical

crosslinking, leading to the improvement of the thermal and mechanical properties, and to

increased water resistance [77,78]. In the present work, Gel is one of the degradable

polymers used as a polymer coating on Bioglass-based scaffolds, which was aimed to

improve the mechanical strength of the scaffolds. Metze et al. [53], Erol et al. [74] and

Desimone et al. [79] have presented significantly improved compressive strength and

fracture toughness of bioactive glass-based scaffolds by Gel coating. In addition, Gel coating

does not induce negative effects on bioactivity of bioactive glasses. This result was

confirmed by the formation of HCA after immersion in simulated body fluid (SBF) [53] . In

addition to providing a benefit in the mechanical properties, Gel coated scaffolds (i.e. Gel

coated TCP scaffolds) were reported to support MC3T3-E1 cell adhesion and subsequently

to promote cell proliferation and differentiation compared to uncoated scaffolds, as

presented by Kim et al. [80].

Page 26: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

12

Poly(lactic acid) (PLA) is an aliphatic polyester derived from renewable resources

[81]. PLA is a semi-crystalline polymer exhibiting high tensile strength and elongation

compared to natural polymers [60,81]. This character of PLA makes it suitable for low load

bearing applications [60]. The linear structure of PLA has methyl (-CH3) side groups, leading

to a hydrophobic feature and an ability to be soluble in organic solvents (such as chloroform,

dimethylene chloride (DMC), dichloromethane (DCM), methanol (MeOH), ethanol (EtOH),

benzene, acetone, etc.) [81,82]. PLA can be degraded by the mechanism of homogeneous

hydrolysis erosion [83], leading to lactic acid obtained as a degradation product. The

degradation product helps to reduce the pH of the environment and induce further

degradation [83,84]. The physical properties and degradability of PLA depend on the

racemization of D- and L-isomers. Semi-crystalline PLLA is synthesized from L-lactide, while

amorphous PDLLA is obtained from DL-lactide [81]. As a result, PLLA and PDLLA exhibit

different mechanical properties and degradation rate. PLA is one of the most widely used

polymers in tissue engineering applications due to its biocompatibility, biodegradable control

and suitable mechanical properties [52,56,60,83,84]. Moreover, PLA can be processed

readily and reproducibly [60]. In terms of composite scaffolds, polyesters have been used as

polymer coatings on porous bioceramic and bioactive glass-based scaffolds. For example, it

was shown that PDLLA coating reduced the brittleness of ceramic scaffolds, as presented

by Yunos et al. [52,85,86], Chen et al. [56], Bretcanu et al. [55] and Novak et al. [87].

Moreover, biocompatibility of either PLLA or PDLLA coated Bioglass®-based scaffolds has

been confirmed by in vitro culturing with human osteosarcoma cell line (HOS-TE85) [55], for

example. It has been found that the polymer coating, scaffold microstructure and surface

roughness influenced the cell behavior. In addition, cell differentiation has been confirmed by

culturing mesenchymal stem cells (MSCs) on PDLLA/Bioglass composite scaffolds [88].

Poly(3-hydroxybutyrate-co-3-hydroxyhexanoate (PHBHHx) is a member of PHA

biopolyester family [89]. PHBHHx has higher elastomeric mechanical properties compared

to poly(3-hydroxybutyrate) (PHB) and poly(3-hydroxybutyrate-co-valerate) PHBV [90,91].

PHBHHx, which is synthesized by microorganisms, is a copolymer of hydroxyl butyrate (HB)

and hydroxyl hexanoate (HH) with the adjustable content of HH represented by ‘x’ [92].

Page 27: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

13

Increased content of HHx leads to reduced crystallinity and subsequently the tensile strength

and the elongation at break increased compared to PHB [89,90]. Thus molecular weight

(Mw) and chemical composition of PHBHHx can be tailored to meet the physical properties

required for various tissue-engineered scaffolds [89–91]. Recently, PHBHHx has become a

promising candidate for tissue engineering scaffolds due to good mechanical properties, low

toxicity, biodegradability and biocompatibility with various cell types, i.e. fibroblasts,

osteoblasts, chondrocytes and MSCs [91]. The variety of PHBHHx or PHAs used in

biomedical applications has been reviewed by Chen et al. [92].

2.2.1.2 Composite-based scaffolds as controlled drug-delivery systems

In general, scaffolds are used as a template able to support the growth and repair of

tissues. In recent research, the scaffolds are being enhanced to form multifunctional

systems, which are able to combine tissue regeneration and local drug delivery [93,94].

According to a convenient type of composite scaffolds (biodegradable polymer coated

bioactive glass scaffolds) developed for bone tissue engineering, the polymer coating layer

can act as a carrier of bioactive molecules such as drugs and growth factors [95,96]. At the

same time, such polymer coating can improve the mechanical properties of porous bioactive

glass scaffolds [52,57]. As reported by Yaylaoglu et al. [97], CaP/Gel scaffolds have been

loaded with gentamicin for in-situ drug delivery combined with tissue engineering.

Continuous release of the drug upon 4 weeks in vivo was observed with the release rate

depending on the degradation rate of the Gel component. Kim et al. [98] developed HA-

based scaffolds with controlled tetracycline release function by using PCL/HA hybrid coating.

The scaffolds presented improved mechanical properties due to the presence of PCL hybrid

coating, while the drug entrapped in the polymeric coating exhibited a sustained release

profile. Moreover, improved mechanical properties and sustained drug release function have

been confirmed by developing vancomycin-loaded PHBV coated 45S5 Bioglass-based

scaffolds, as reported by Li et al. [57]. The coated scaffolds provided a lower initial burst

release when compared to the drug release of uncoated scaffolds. In addition, a controlled

drug release over 6 days in phosphate buffer saline (PBS) was measured. Francis et al. [95]

have reported gentamicin-loaded PHB microsphere coated 45S5 Bioglass-based scaffolds,

Page 28: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

14

which not only presented controlled drug release, but also maintained the bioactivity of

Bioglass scaffolds. Similarly, multifunctional scaffolds based on vancomycin-loaded poly(n-

isopropylacryliamide-c-acrylic acid) microgels dispersed in PLGA coated 45S5 Bioglass-

based scaffolds (Olalde et al. [96]) exhibited improved mechanical properties and

maintained bioactivity. In addition, they exhibited controlled release rate from the drug-

loaded microgels. The polymer coatings protect drug molecules from the aqueous

environment and inhibit the fast dissolution of drugs, subsequently the slow release is

achieved [99]. In this case, the dissolution of the drug is caused by the degradation of the

polymer carrier associated with the diffusion of the drug through voids in the carrier [99].

2.2.1.3 Biodegradable polymers as cartilage scaffold

To engineer cartilage-like tissue that mimics the complex and unique structure of

natural cartilage, the focus here is the design of scaffolds with chondroinductive and

chondroconductive properties [100–102]. Recently, it has been shown that cartilage

scaffolds with rather sophisticated 3D architecture can be developed for example starting

from fibrin and agarose-based materials [24]. The ideal scaffold for cartilage tissue

engineering should be biocompatible, biodegradable and show sufficient mechanical

properties in order to resist mechanical forces. Moreover, it should exhibit appropriate

structural and geometrical properties for supporting cell proliferation and differentiation. The

dedifferentiation of cells must be avoided. In addition to these requirements, cartilage

scaffolds should achieve the tissue-like elastic properties, which can tolerate shock

absorption and deformation [24,103].

Biodegradable polymers are widely used in cartilage regeneration, since they can be

fabricated in the forms of hydrogels, porous foams and fibers, which are suitable structures

for scaffolds [104]. Rationale of using polymers as a cartilage scaffold is their intrinsic

elasticity, controlled degradability and sufficient mechanical strength close to the physical

characteristics of native cartilage [105]. In addition, polymers can be customized in terms of

their physical properties by the regulation of Mw and crystallinity [81]. In terms of chemical

design, current research trends focus mostly on natural polymers, i.e. Col and hyaluronan

(HyA), considering that both are components of the cartilaginous ECM [105,106]. Even

Page 29: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

15

though Col and HyA-based scaffolds have been investigated for new cartilage generation,

they exhibit drawbacks concerning the mechanical properties and cost. Thus the

development of alternative, cost-effective materials are of great current interest. For

instance, Gel and Alg exhibit chemical structures similar to Col and HyA, respectively, and

both polymers are inexpensive [107]. Nevertheless, the physical and chemical crosslinking is

crucial for natural polymer scaffolds because they are not mechanically stable in aqueous

environments [65,108,109]. Consequently, biodegradable synthetic polymers represent

another group of materials of choice for cartilage scaffolds. Synthetic biodegradable

polymers are beneficial in terms of mechanical properties, controlled biodegradability and

processibility [104]. Regarding biomimetic approaches to tailor chemical composition and

mechanical stability, researches in last decade have focused on the combination of distinct

polymers, including blending of natural polymers [110–116], synthetic polymers [117,118],

and natural and synthetic polymers [58,119–123].

Alg is a highly interesting polysaccharide, which is widely used as a scaffold in

cartilage regeneration. Alg has a chemical structure similar to HyA and it is cost-effective

compared to HyA [124]. In vitro and in vivo studies [59,69,73,109,125–129] have shown that

Alg is able to support the viability, maintaining the round phenotype of chondrocytes and

promoting the formation of Col II and glycosaminoglycans (GAGs). The production of Col II

and GAGs is an indication of cartilage regeneration [130,131]. In terms of manufacturing, Alg

scaffolds can be easily fabricated via mild gelation via interaction with cations (Ca2+

, Cu2+

,

Zn2+

, Sr2+

and Fe2+

), according to the so-called egg-box mechanism (Fig. 2.2) [109,127,128].

Such crosslinking process via ionic interaction involving anionic chains of Alg and cations

leads to the formation of water-insoluble Ca-Alg gels [109,127,128]. The Alg-gels can be

used as an encapsulation device for living cells [127,132,133]. Since the released cations

exchange with Na+ in the culture medium and in body fluids, ionic crosslinking has been

proved to be non-toxic in vitro and after implantation [125]. Moreover, the Ca-Alg gels can be

transformed to a foam-like structure by the application of a lyophilization process

[63,128,134], which will be detailed in a later section. According to the characteristic

properties stated above, Alg was focused in the present wotk.

Page 30: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

16

Figure 2. 2 Schematic diagram showing the gelation-mechanism of alginate and calcium

cations by the formation of egg-box structure (Image courtesy K. Kashima and M. Imai

[135]).

2.2.2 Scaffold fabrication techniques

The scaffold fabrication technique is another important factor, which affects the

structural architecture and geometry of scaffolds. According to physical considerations for

suitable osteochondral scaffolds, pore size, porosity and interconnectivity are crucial for

supporting cell migration and tissue regeneration. In order to fabricate porous structures

suitable for bone- and cartilage-repair, targeted porosity and pore sizes of scaffolds, specific

for bone and cartilage regeneration, must be taken into account. The appropriate pore size

of bone scaffold is considered to be in the range of 100 - 600 µm, which has been confirmed

to be sufficient for cell migration, nutrient and waste transportation, vascularization, and

tissue ingrowth [136]. In contrast, vascularization does not occur in cartilage, which is

composed of dense connective ECM, thus pore sizes around 50 - 300 µm are sufficient for

chondrocytes proliferation and ECM secretion [6,137]. This requirement is attributed to the

fact that chondrocytes show high tendency of differentiation when the pore size is around 30

times the cell diameter (diameter of chondrocytes 10 - 15 µm) [6]. Highly porous scaffolds

Page 31: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

17

allow for more cell attachment and consequently result in more tissue formation compared to

less porous scaffolds, which is linked to a greater transportation of nutrient and metabolic

waste products [6,73]. Microstructures exhibiting larger surface area are additionally

required for supporting cell attachment and ECM regeneration [103]. A high pore

interconnectivities are required for homogeneous cell seeding, which influences the quality

of the formed tissue [6].

The mechanical properties of scaffolds are influenced by their microstructure as well

as by the intrinsic properties of the material used [6]. As porosity compromises the

mechanical strength of scaffolds, increasing porosity leads to the reduction of strength [6].

Thus, the porous structure of scaffolds should be tailored to achieve sufficient porosity and

pore interconnectivity with the maintenance of suitable mechanical properties. The

mechanical properties of scaffolds should match those of native osteochondral tissues in

order to withstand local loads in the joint by in vivo studies [6]. The mechanical properties of

natural human cartilage-bone tissues are summarized in Table 2.2. In general, cartilage has

the function to transform compressive forces into tension mode and to further transfer loads

to the underlying SB [138]. At the same time, cartilage is able to withstand shear forces by

supplying a intrinsic low friction-surface [8,9,118,139]. On the other hand, the underlying SB

mainly supports compressive and tension loads [9,140].

Table 2. 2 Mechanical properties of natural healthy human osteochondral tissues

[6,12,73,76,141–143].

Mechanical properties (MPa) Articular cartilage Subchondral bone

Compressive modulus

Compressive strength

Young’s modulus (Tension)

Ultimate Tensile strength

Shear modulus

0.24 - 0.85

0.01 - 3

5 - 25

3.7 - 10.5

0.2 - 2

0.05 - 0.6

2 - 12

445

3 - 20

No report

Page 32: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

18

Importantly, the biodegradability of scaffold materials influences the formation and

functionality of new tissues [126]. An appropriate scaffold should exhibit degradation rate

matching the formation of new tissues and must maintain the structural stability until the new

tissue fully assumes the load-bearing function [9,144]. The degradation rate of the scaffold

can be altered by the variation of the material used, namely composition, chemistry and

porous structure [22,105]. In particular, the architecture and topography of scaffolds, which

greatly affected cell attachment, proliferation and differentiation, are partly influenced by the

fabrication techniques. The appropriate fabrication technique needs to be able to generate a

porous scaffold with reproducible architecture and provide mechanical functions for load-

bearing environment [104]. In order to fabricate scaffolds for osteochondral repair, the

combination of different fabrication techniques is crucial for the achievement of sophisticate

structures such as multilayered scaffolds [50,145]. Currently, the variety of fabrication

techniques available, including solvent casting and particle leaching, melt molding, freeze-

drying, thermal induced phase-separation, electrospinning and rapid prototyping techniques

[6,146], are all considered in the fabrication of polymer-based scaffolds. On the other hand,

bioceramic/bioactive glass-based scaffolds are frequently fabricated by foam-replication,

rapid prototyping, fused deposition remodeling, robocasting, stereolithography and 3D-

printing [146]. Rapid prototyping can be used to fabricate both polymer- and ceramics-based

scaffolds [146]. The advantages and disadvantages of current fabrication techniques applied

for manufacturing both bone and cartilage scaffolds are summarized in Table 2.3.

Particularly, freeze-drying, electrospinning and foam-replica techniques, focusing on recent

work, will be discussed. Foam-replication, freeze-drying and electrospinning techniques

were chosen in this study according to the required physical properties, such as porosity,

pore size, architecture and geometry of the scaffolds, and mechanical properties of the

specific tissues. In addition, all techniques are simple cost-effective.

Page 33: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

19

Table 2. 3 Current 3D scaffold fabrication techniques for polymers and ceramics.

Fabrication techniques Pros(+)/Cons(-) References

Polymeric scaffolds

Solvent casting/particle

leaching

+ Pore size 30 - 300 µm

+ Controlled pore sizes by particle size of salt/porogen

- Limit for thin membrane with thin wall section

- Low porosity 20 - 50 % and insufficient pore

interconnectivity

- Required toxic solvents

- Remained salt particles in matrix

- Time consumer

[19,58,119,123

,147,148]

Melt molding + Solvent-free method

+ Pore size 50 - 500 µm

+ Controlled macropore geometry

- Porosity 80 %

- Suitable for thermoplastics

[19,119,149]

Freeze-drying + Can be incorporated in conjunction with thermal

induced phase separation

+ Controlled pore size and pore orientation

+ Porosity 90% and pore size 50 - 400 µm

+ High pore interconnectivity

[110,112,114,1

16,150–156]

Thermal induced phase-

separation (TIP)

+ Extensively applied in the fabrication of

microspheres for drug-delivery system

+ Suitable to fabricate porous polymer/ceramic

composite-based scaffolds

+ High porosity 97 %

+ Pore size 200 µm

+ Obtained high volume of interconnected micro-

pore structure

[51,93,157]

Electrospinning + Can be used to fabricated hybrid fibers (organic-

inorganic mixture)

+ High porosity

- Small pore size

- Limit designed architecture – needs post-

[86,117,158–

174]

Page 34: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

20

Fabrication techniques Pros(+)/Cons(-) References

fabrication techniques

- Insufficient mechanical properties

Rapid prototyping/solid free

form (SFF)

+ Manufactured by computer - generated design

+ Optimized microstructure and mechanical

functions

- Low porosity 60 %

[138,175,176]

Fused deposition

remodeling (FDM)

+ Reproducibility

+ Controlled structure by computer-controlled

method

[102,177]

3D printing + Precise deposition of cells and matrix in layer-by-

layer fashion

+ Highly reproducible architecture

+ Easily tailored porosity

- Limits for used in load-bearing applications,

especially in the cases of natural-derived

polymeric matrices.

[27,178]

Ceramic-based scaffolds

Replication + Conventional technique

+ High porosity 90 %, pore size 100 - 700 µm

+ Achieved architecture similar to that of cancellous

bone

[32,85,150,179

,180]

Rapid prototyping/solid free

form

+ Controlled architecture

+ Designed scaffolds can be fit on the defect site

[138,175,181]

Stereolithography

+ Versatile with respect to the freedom of design

and scale (submicrons-decimeters)

+ Manufactured in layer-by-layer fashion by

computer- controlled method

+ Fabricated gradient scaffolds in porosity and pore

size

[6,182]

3D printing/ robocasting + Achieved thick struts

+ Pore size 500 µm

+ Sufficient compressive strength

- Porosity 60 %

[27,33,93,183]

Page 35: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

21

2.2.2.1 Foam replication technique

The foam fabrication technique was first developed in 1963 for ceramic foam

manufacturing [28]. This technique (Fig. 2.3) involves the production of ceramic foams by

coating a polymer template (i.e. polyurethane (PU) foam) with a ceramic slurry (ceramic

powder/binder/water mixture). Then the sacrificed template is burnt out and the ceramic particles

are sintered by using proper heat treatment. As-sintered foams exhibit high porosity ( 80 %)

and pore size in hundreds microns, depending on the pore size of the used polymer template.

However, highly porous scaffolds are obtained with relatively low mechanical properties,

exacerbated by the intrinsic brittleness of ceramics, which are difficult to handle [28,184].

The foam replication method to fabricate Bioglass-based scaffolds was patented by

Boccaccini group at Imperial College London in 2006 [185]. The technique is currently widely

used to fabricate bioactive glass-based scaffolds in the field of bone tissue engineering. It has

been proved that 45S5 Bioglass-based scaffolds, for example, supported osteoblasts activities

[28]. Cells migrated efficiently and proliferated into entire porous structure [28]. In addition, the

low mechanical properties of Bioglass-based scaffold can be overcome by the incorporation of

polymer phases, forming composite-based scaffolds [48], as mentioned previously. Compared to

other fabrication techniques, such as rapid prototyping, stereolithography, etc., the foam

replication technique is more cost-effective and less time-consuming necessitating simple

equipment [28].

Page 36: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

22

Figure 2. 3 Schematic diagram of the foam replication technique employed to produce 3D

porous bioceramics- and bioactive glass-based scaffolds (according to [28,184]).

2.2.2.2 Freeze-drying technique

Freeze-drying technique is an attractive dehydration method, known as lyophilization

[186]. It is well known in the food industry [187] and it has become widely used in biomedical

applications, in particular scaffolds for tissue regeneration [188]. It basically works by

freezing the solution at a temperature below the freezing point of the solvent used following

by the reduction of the surrounding pressure below atmosphere pressure to allow the frozen

solvent in the material bulk to sublimate directly from solid phase to gas phase [105]. The

basic principle of freeze-drying process can be explained with reference to a simple water

phase diagram, as demonstrated in Fig. 2.4. The process basically consists of three stages.

The first freezing stage involves a fast decrease of material temperature at temperature

underneath the freezing point (TC). The next stage is drying the material by heating below

the triple point (TA) and under vacuum conditions (below PA) to force sublimation, leading to

the formation of an interconnected pore structure. After this stage, the water ( 7 - 8 %) still

bound to the porous material can be desorbed by increasing temperature [189]. Pore size

and orientation of pores are mainly influenced by the freezing temperature [189–191]. If the

freezing temperature is lowered (rapid freezing rate), for example in liquid nitrogen, the

formed nuclei of ice crystallization are small, leading to small pore size of samples after

drying [156,191]. In another case, at - 20 C freezing temperature (i.e. in a freezer), the pore

Page 37: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

23

size of the dried sample is larger compared to the case of rapid freezing rate [156,191]. For

instance, in the case of gelatin foams, which are frozen at - 20 °C, their pore size has been

reported in the range of 250 - 300 µm, while at freezing temperatures - 80 C, smaller pore

sizes are obtained in the range of 45 - 50 µm [156]. In addition, the smaller pore size exhibits

thicker pore walls and subsequently higher mechanical properties [156,192]. Thus the

freezing temperature is the most important factor on determining the microstructure of

freeze-dried samples. The balance between pore size and mechanical properties must

therefore be optimized in each case for specific tissue regeneration. Scaffold architectures

fabricated by using freeze-drying technique are highly interconnected, which is necessary for

tissue ingrowth and regeneration [105]. Moreover, the scaffolds show achievable pore size

up to 300 µm and the porosity up to 97 % [105,193,194]. Freeze-drying also causes less

damage to the material and does not cause shrinkage or toughening of the material being

dried [105].

Figure 2. 4 (A) The schematic diagram of the freeze-drying (lyophilazation) process showing

also the phase diagram of water representing the mechanism of freeze-drying ([186,189]).

Page 38: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

24

3D porous alginate foams as scaffolds for cartilage regeneration are being

extensively researched currently [59,62,63,134,195–201]. Most previous studies have

shown that freeze-dried Alg scaffolds with suitable porous structure can be used for culturing

with chondrocytes and MSCs. The proliferation and differentiation of cells, and the formation

of Col II and GAGs have been confirmed. As reported in the study of Lee et al. [195], Alg

foams promoted the adhesion, proliferation and differentiation of human chondrocytes and

the formation of specific cartilaginous matrices was detected. More recently, Wan et al. [197]

have prepared Alg foams in combination with CS by using freeze-drying technique. In vitro

culture of chondrocytes-seeded scaffolds was developed by on-site gelation (chondrocytes

embedded Alg gelation) in order to promote functional restoration and maintenance of the

round phenotype of chondrocytes. Petrenko et al. [134] have used freeze-drying method

with Ca-Alg hydrogel to develop porous scaffolds with wide pores for culturing with MSCs.

By this approach, cell adhesion and proliferation were not observed because Alg has

basically limited cellular interaction. Another approach has been developed by the

incorporation of Gel as a surface grafting onto the inner pore walls. As a result, the

adhesion, proliferation and differentiation of MSCs were improved [134]. In contrast, Miralles

et al. [202] have confirmed that Alg freeze-dried sponges promoted a favorable environment

for the growth of chondrocytes compared to Alg beads. Consequently, PGs rich matrix was

significantly detected by qualitatively histological evaluation in the case of Alg sponge. This

result has been suggested by the macroporous structure of the sponges indicating that

macro-channels allow better cell seeding and migration, compared to micro-porous gels

[202]. In addition, Wan et al. [196] confirmed that chondrocytes-embedded Alg hydrogels

did not exhibit the organization of synthesized Col in layer-anisotropic manner as in native

cartilage. Chondrocyte-clusters, chondrocyte proliferation and chondrogenic gene

expression (i.e. Col II, transcription factor Sox-9 and aggrecan) were observed in a porous

Alg sponge cultured with chondrocytes after 4 weeks, which indicates cartilage regeneration,

as evaluated by Yen et al. [203].

Page 39: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

25

2.2.2.3 Electrospinning technique

Electrospinning involves the induction of static electric charges on the molecules of

solution at the level that causes the self-repulsion of charges [204]. Once the repulsion force

overcomes the force of surface tension of the solution, a jet of the solution forms, known as

Taylor cone, and stretches fibers toward the grounded collector [170,205]. Therefore, a

typical electrospinning set up consists of syringe containing solution, syringe pump, voltage

generator and collector, as demonstrated in Fig. 2.5. The fiber diameter can be controlled by

the variation of polymer solution concentration, flow rate, applied voltage and distance

between needle tip and collector [170,206]. Electrospinning has gained high popularity in the

field of tissue engineering due to its capability of fabricating fibers in submicron scales (30

nm – 10 µm in diameter). The nanotopographical features of electrospun fibers can for

example mimic Col fibrils in connective tissues [207,208]. Numerous materials, especially

polymers (synthetic and natural-derived polymers), have been successfully electrospun into

porous fibrous scaffolds for tissue engineering applications. Synthetic polymer fibers exhibit

sufficient mechanical properties and controlled biodegradability [117,208,209]. Particularly,

the biodegradation of scaffolds directly relates to the ability of the scaffold to maintain its

structure and subsequently to support cellular activity [208]. As investigated by Li et al.

[208], electrospun PLLA and PCL fibers exhibit higher proliferation of either chondrocytes or

MSCs compared to PLGA and PDLLA fibers. PLLA and PCL fibers maintained their fibrous

architecture over the culture time (21 days), while the degradation of PLGA and PDLLA

fibers was detected after 3 days in culture [208]. However, synthetic polymer fibrous

scaffolds usually lack the appropriate surface properties, i.e. hydrophilicity, to support cell

attachment and proliferation [117]. Natural polymer-based electrospun fibers are another

promising group of materials for scaffold development due to their excellent biocompatibility

[210–212]. CS/PEO fibers [213], for instance, were seen to support the adhesion of

chondrocytes and to maintain the round phenotype throughout the period of study.

Moreover, the study of Skotak et al. [214] proved that crosslinked Gel fibers cultured with

chondrocytes showed cell viability and supported the round phenotype of chondrocytes over

7 days in culture. In addition, Col II was observed in a high ratio to Col I [214].

Page 40: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

26

Figure 2. 5 Schematic diagram of the electrospinning process in horizontal direction, which

is composed of voltage supply, syringe and needle, syringe pump and collector; the SEM

image shows electrospun PLLA fibers.

2.2.3 Strategies of multilayered scaffold

Current scaffold materials and designed strategies being applied in the field of

osteochondral tissue engineering are summarized in Table 2.4. The scaffold strategies can

be categorized into three available systems, namely single phase, bi- or multi-layered and

gradient structures, involving a variety of biocompatible materials and designs. Findings of

previous studies are expected to be a guideline for an improvement of scaffold designs to

get closer to an ideal osteochondral scaffold of relevance for clinical practice. Multilayered

and gradient composite scaffolds are being widely researched for osteochondral repair

instead of using single phase materials due to the fact that they can be designed and

fabricated to mimic the complex zonal structure of the native tissue [152,215–220]. As

reported in several studies, single material-based scaffolds could not provide the formation

of hyaline cartilage, instead fibrocartilage was generated. This is one reason of the

insufficient mechanical stability of the new tissues grown, leading to unsuccessful long term

Page 41: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

27

repair. Moreover, the incorporation of cells and/or growth factors is essential for improved

scaffold performance.

Indeed, each scaffold strategy has its own advantages and disadvantages, which

are the result of either the used scaffold material or the design. The ideal osteochondral

scaffold has not been developed as yet. Additionally, the biological responses of different

engineered scaffold (in vitro and in vivo) are also different, also different animal models

usually show different outcomes. Also, in vitro studies of scaffolds mimicking the complex

structure of natural osteochondral tissue require specific culture systems. For example,

specifically designed bioreactors, i.e. double-chamber bioreactors [221], are required due to

different cell types and different culture media. According with this requirement, it becomes

apparent that osteochondral scaffolds need to avoid the migration of osteoblasts and

vascular tissue from the subchondral phase to the upper chondral phase, which is risky for

osteogenesis in the layer of cartilage. However, it has not been confirmed whether the

intermediate layer between cartilage and SB needs to be dense or porous. Taken into

consideration the anatomical structure of natural tissue, the intermediate layer called

calcified cartilage is very dense due to a mineralized matrix containing hypertropic

chondrocytes and packed extracellular fibrils. Therefore, the available free space is

assumed to be much smaller in comparison with that in cartilage. In addition, calcified

cartilage is connected to the subchondral bone by interdigitation. This might be the natural

barrier to avoid the migration of bone cells and vascular diffusion from SB. Importantly, new

tissues regenerated from implantation of scaffolds must be healthy and durable for long

term, which is initially confirmed by the organization of the new tissue in zonal arrangement

mimicking the structure of natural cartilage tissue.

Page 42: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

28

Table 2. 4 Summary of current strategies in osteochondral tissue engineering.

Scaffold Materials Notable findings References

Cartilage Interface SB

Single phase

PLGA + Cell-free PLGA sponges showed a

possibility to repair full-thickness

defects in rabbits by absorbed local

cells from the erupted underlying bone

marrow.

- Small pore size ( 100 µm), porosity (

83 %) and hydrophobic property of the

scaffolds

Nagura et al.

(2007) [222]

PCL

+ PCL scaffolds in combination with

periosteal grafts provided excellent

integration with SB.

- Deficient development of hyaline

cartilage.

Mrosek et al.

(2008) [101]

PCL/F127 + PCL/F127 scaffolds culturing with

ADSCs and TGF-β1/BMP-7 improved

gross appearance of osteochondral

defect in rabbits.

- By histological results, the growth

factors did not significantly improve the

ability of the scaffold to repair the

defects due to high degree of foreign

body reaction.

Im et al.

(2009) [223]

Page 43: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

29

Scaffold Materials Notable findings References Notable Findings References

Cartilage Interface SB Interface SB

Bi-/multi-layered

Fibrin

Fibrin glue PCL

- Fibrin scaffolds for cartilage phase

exhibited rapid degradation.

- Lack of mechanical support for

cellular development and excretion

of ECM.

Swieszkows-ki

et al. (2007)

[138]

PCL

Fibrin glue PCL/

TCP

+ PCL scaffolds for cartilage phase

prolonged degradation and

enhanced the mechanical

properties in comparison with fibrin

scaffolds.

+ PCL/TCP scaffolds for SB phase

promoted better BMSCs

proliferation compared to pure PCL

scaffolds after implantation in

rabbits.

PCL - PCL/

TCP

+ MSCs-seeded PCL constructs

promoted cartilaginous production

after implantation in pigs.

+ Resurfacing cartilaginous MSCs-

seeded PCL constructs with

electrospun Col I mesh could

protect cell leakage, reduce

generation of fibrocartilage and

enhance content of GAGs.

+ PCL/TCP-SB phase promoted high

mineralization.

Ho et al.

(2010) [224]

Page 44: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

30

Scaffold Materials Notable findings References

Cartilage Interface SB

Col I Activated

plasma

Col I/β-

TCP

+ MSCs-seeded triphasic constructs

promoted cartilaginous production

and osseointegration.

- The constructs did not provide

better outcomes compared to the

conventional treatments such as

osteochondral autograft transfer

system, as confirmed by

histological results.

Marquass et al.

(2010) [225]

Agarose

Agarose/

PLGA/ 45S5

BG

PLGA/

45S5

BG

+ Three-layered scaffolds supported

co-culture of chondrocytes and

osteoblasts, and promoted the

generation of continuously distinct

hyaline cartilage-calcified cartilage-

SB.

+ BG presenting in intermediate and

SB phases enhanced the formation

of CaP, indicating ability of

mineralization.

Jiang et al.

(2010) [226]

Col I - Col I/

HA

+ Layered scaffolds promoted

chondrogenic and osteogenic

differentiation of MSCs.

- Culturing different cell types in the

layered scaffolds at the same time

was limited. Designed double-

chamber bioreactor is thus

required.

Zhou et al.

(2011) [221]

Page 45: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

31

Scaffold Materials Notable findings References

Cartilage Interface SB

PA6 Non-porous

PVA

PVA/H

A

+ In vivo studies of BMSCs-seeded

PA6 and PVA/HA separately in

rabbit muscle pouch showed that

PA6 constructs provided a

cartilaginous marker (i.e. Col II);

and PVA/HA constructs provided

an osteogenic marker (i.e. Col I).

- Non-porous PVA intermediate layer

was expect to function as a native

calcified cartilage but it is not a

clear requirement for ideal

cartilage-bone interface that

whether the interface layer is dense

or porous.

Qu et al.

(2011) [152]

Col I

- Col I + MSCs-seeded Col I scaffolds

promoted both cartilage- and bone-

like tissues during in vitro culture,

respectively.

+ After formed bilayered constructs,

undifferentiated MSCs in the

intermediated layer were

differentiated into hypertrophic

chondrocytes and produced Col X.

+ Zonal organization of osteochondral

tissue was formed by the optimal

designed scaffold and controlled

cell density.

- Mechanical environments must be

taken into account for better zonal

organization of tissues.

Cheng et al.

(2011) [100]

Page 46: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

32

Scaffold Materials Notable findings References

Cartilage Interface SB

CS/Gel Fibrin glue CS/

Gel/HA

+ TGF-β1 and BMP-2 enhanced

chondrogenesis and osteogenesis

of MSCs on CS/Gel and CS/Gel/HA

constructs in vitro, respectively, and

also supported the regeneration of

osteochondral tissue in vivo.

Chen et al.

(2011) [114]

PDLLA - PDLLA/

45S5

BG

+ In vitro culture with ACDC5 showed

cell attachment, proliferation and

migration through PDLLA fibers,

which are suitable for cartilage

regeneration.

Yunos et al.

(2012) [86]

Poly HEMA/

HyA

- Poly

HEMA/

HyA/

nHA

+ HyA incorporated into polyHEMA

layer encouraged chondrogenesis

of chondrocytes and played an

important role in maintenance of

chondrocytes phenotype.

+ nHA in SB phase induced osteo-

conductivity and -inductivity of

MSCs- seeded constructs.

- Small pore size of polyHEMA/HA-

SB phase ( 38 µm) was conflicted

to the ideal pore size suitable for

vascularization and bone ingrowth

( 300 - 500 µm).

Galperin et al.

(2012) [227]

Page 47: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

33

Scaffold materials Notable findings References

Cartilage Interface SB

Gradient/graded

HyA/

Col I

Col I/HA

(40/60)

Col I/HA

(30/70)

+ Gradient in compositions provided

specific proper environment for

specific differentiated cells and

supported the generation of zonal

organized tissues.

+ By using freeze-drying technique, the

obtained scaffold also exhibited

gradient in porous structures.

- MSCs are seemed to be only cell

source suitable for the gradient

strategy because it is extremely

complicated to perform in vitro co-

culture of different cell types in the

single gradient structure.

Tampieri et al.

(2008) [110]

PLGA/

TGF-β1

- PLGA/

HA/

BMP-2

+ Gradient in material compositions

and growth factors promoted the

regeneration of cartilage and SB

after implantation in rabbit knees.

+ Growth factors triggered the

differentiation of progenitor cells into

chondrogenic and osteogenic cells

after implantation.

+ The gradient scaffolds based on the

combination of BMP-2 and nHA

promoted fast bone formation in vivo

and restored new cartilage.

- The manufacture of the gradient in

growth factors is only possible in the

form of microsphere-based scaffolds,

while it is complicated in the form of

sponge-like scaffolds.

Mohan et al.

(2011) [228]

Page 48: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

34

Scaffold materials Notable findings References

Cartilage Interface SB

Agarose/

MSCs in

osteoge-

nic

medium

Agarose/

MSCs in

basic

culture

medium

Agarose/

MSCs in

osteoge-

nic

medium

+ Gradient in MSCs within different

culture media stimulated

differentiation of MSCs into

chondrocytes and osteoblasts on

cartilage and SB phases,

respectively.

+ Gradient in cell types was observed

at intermediate phase.

- Gradient-generating culture device is

required.

Shi et al.

(2012) [229]

Col I Col I/nHA

(40/60)

Col I/nHA

(30/70)

+ After implantation in sheep for 6

months, either cell-free scaffolds or

chondrocytes-seeded scaffolds

provided the formation of

osteochondral tissue and filled the

full-thickness defect.

+ No difference of tissue regeneration

was found in both, cell-free and

chondrocytes-seeded scaffolds, due

to the recruitment of local BMSCs.

+ Cell-free gradient scaffolds were

tested in human patients: at 2 years

follow-up, 70 % of the defects were

completely filled with new tissues.

- Low number of patients and scarcity

of evaluation cases were limited in

the comparison of the results.

Kon et al.

(2010, 2012)

[230,231]

Page 49: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

35

2.3 Cells and bioactive molecules/growth factors for osteochondral

tissue engineering

A current approach in osteochondral tissue engineering is to reconstruct the

functional engineered cartilage–bone interface by co-culturing chondrocytes and osteoblasts

into multilayered scaffolds [232]. Chondrocytes and osteoblasts are an obvious choice for

cell sources because they are found in native cartilage and bone tissues, respectively.

Autologous chondrocytes and osteoblasts have been successfully used in the regeneration

of cartilage [131,209,233–235] and bone [44,171,236–238], respectively. In addition, co-

culture of chondrocytes and osteoblasts has been extensively studied for osteochondral

interface [102,111,232,239]. Cao et al. [102] investigated porous PCL scaffolds co-culturing

with chondrocytes and osteoblasts. Both cell types produced specific ECM in each own

compartment and they also migrated and integrated at the interface of the scaffold. In this

approach, the important aspect is the interaction between osteoblasts and chondrocytes

associating with scaffold material during co-culture, which leads to the proper formation of

osteochondral interface [239]. However, it is critical to control specific osteogenic and

chondrogenic phenotypes during co-culture in order to mimic each tissue zone to native

tissues.

Even though the utilization of autologous cells is conducive to the growth of

functional tissues, they are difficult to isolate and easily change their phenotypes during the

culture process due to their dedifferentiation capacity [25,240]. Therefore, another available

cell source, i.e. progenitor cells (stem cells), is a valid alternative. Stem cells are promising

cells for tissue engineering due to their multipotent nature and self-renewal capacity [241–

243]. Focusing on osteochondral repair, stem cells as single cell source, which can be

differentiated into chondrogenic and osteoblastic cell lines, also overcome the limited supply

of primary cells. Since embryonic stem cells (ESCs) have a limitation in differentiation

capacity [103], bone marrow mesenchymal stem cells (BMSCs) and adipose-derived stem

cells (ADSCs) are the most widely used in current research because they are abundant in

the human body and can be isolated from bone marrow stroma and adipose tissue,

respectively. In particular, BMSCs are promising cells for regeneration of tissues due to

Page 50: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

36

their rapid proliferation and easy differentiation into osteogenic lineage by the use of

supplemented dexamethasone, ascorbic acid and β-glycerophosphate in vitro [25]. It has

been shown in in vitro study that BMSCs can be differentiated into the osteogenic lineage by

the presence of growth factor genes [33]. BMSCs can also undergo chondrogenic

differentiation when cultured in the presence of transforming growth factors (TGF) [4]. The

specific signaling molecules, such as TGF-β family, insulin-like growth factors (IGF), bone

morphogenetic proteins (BMP) and fibroblast growth factors (FGF), are extensively

employed to facilitate tissue growth, by promoting tissue specific proteins and simulating the

differentiation of stem cells [244]. They bind to cell surface receptors and activate

intracellular signaling pathways, which affect cell proliferation, differentiation and ECM

synthesis during tissue regeneration [5,23,244]. The study of Re’em et al. [245] showed that

BMSCs-seeded RGD-modified Alg sponge culturing in the presence of TGF-β1 provided

appropriate progression of BMSCs differentiation. More recently, the combination of TGF-β1

and BMP-2 has been confirmed to activate BMSCs differentiation and to promote the

formation of hyaline cartilage, as shown in the study of Toh et al. [246]. Focusing on

cartilage tissue engineering, growth factors in combination (i.e. TGF-β1/BMP-7 and TGF-

β1/IGF-1) are synergistically high effective for hyaline cartilage regeneration compared to a

single growth factor [244]. For instance, MSCs cultured with the introduction of TGF-β1 were

able to promote an increased cartilaginous ECM synthesis and a decreased Col I synthesis

when TGF-β1 was combined with BMP-7 [244].

Even though BMSCs are widely used in osteochondral repair, the yield of cells from

bone marrow harvest is small because only 10 - 25 ml of bone marrow can be obtained

from a human [5,247–249]. Alternatively, synovial membrane derived cells are gaining

consideration due to their great chondrogenic potential, which is suitable for cartilage tissue

engineering [250–253]. The synovial membrane is the supportive layer of a joint, which is

composed of cellular lining layer [5]. The biopsy of the synovial membrane provides the

accessibility of autologous MSCs [5]. Human synovial MSCs have shown evidence of

chondrogenesis when cultured in 3D Alg scaffolds without the presence of growth factors

[250]. In addition to BMSCs and synovial MSCs, ADSCs obtained from lipoaspirates provide

Page 51: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

37

several advantages related to their abundance, suitable accessibility and relatively low donor

morbidity [223]. However, it was evidenced that ADSCs promoted lower chondrogenic and

osteogenic potential compared to BMSCs [23,254]. This limitation could be overcome by

using the combination of growth factors (i.e. TGF-βs/BMPs) [223,254]. Growth factors have

been used with BMSCs in order to stimulate cell proliferation and the synthesis of

extracellular molecules in vitro and in vivo [244,253]. Since it is commonly found that in vitro

culture of either cell-free scaffolds or cells-seeded scaffolds promotes fibrocartilage

formation instead of the formation of hyaline cartilage, certain growth factors have been

included to induce cell growth into a specific tissue formation [25]. For example, Huang et al.

[177] investigated cartilage formation induced by TGF-β1-loaded fibrin glue incorporated into

PCL scaffold after in vitro culturing with BMSCs and implantation in a lapine model. The

constructs were richly populated with chondrocytes after 4 weeks of implantation, while

immature bone was identified at week 6 of implantation [177]. Therefore, TGF-β1 is

regarded as a cartilage-inducing factor as well as a bone-inducing factor [19].

Generally, growth factors can be introduced to promote the repair of osteochondral

tissue by different methods, either by incorporating them into the scaffold [121,255,256] or

by adding them in the culture medium as a supplement [70,114,245,257]. Cui et al. [118]

investigated the chondrogenic capacity of ADSCs-seeded PGA/PLA fibrous scaffolds

cultured in the presence of TGF-β1 as a medium supplement. Col II and GAGs were

observed after 2 weeks of culture. Similarly, BMSCs-seeded PCL fibrous scaffolds were

cultured in chondrogenic medium (in the presence of TGF-β3) in order to analyze

chondrogenesis and in osteogenic medium in order to analyze mineralization [258]. After 3

weeks in culture, formation of cartilaginous tissue and mineralization were observed in the

newly formed ECM by day 45 [258]. In a recent approach, osteochondral scaffolds have

been designed creating a gradient of encapsulated growth factors, as reported in the study

of Mohan et al. [228]. Chondral PLGA microspheres loaded with TGF-β1 and osseous

PLGA/HA microspheres loaded with BMP-2 were implanted in rabbit knees. It was found

that the gradient scaffolds promoted the complete bone ingrowth and cartilage regeneration

with high GAGs content [228]. Re’em et al. [259] showed that BMSCs seeded onto both,

Page 52: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

38

TGF-β1-loaded porous Alg scaffolds as a chondroinductive phase and BMP-4-loaded

porous Alg scaffold as an osteoinductive phase, promoted the differentiation into

chondrocytes and osteoblasts, respectively. Moreover, biomolecules, for example HyA and

chondroitin sulfate (ChS), have been incorporated into either scaffolds or culture medium to

stimulate chondrogenesis of MSCs [68,198,260].

Page 53: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 3

Objectives and Outline

The main goal of this research is to study and develop biomaterials-based scaffolds

suitable for osteochondral tissue engineering applications, focusing on combination of

inorganic and polymer materials, and including novel fabrication techniques. This work will

include the characterization of the physic-chemical and mechanical properties as well as the

biological response, e.g. in vitro cell culture, of the new scaffolds. The different tasks carried

out to achieve the final thesis objectives are summarized in Fig. 3.1.

The fabrication of an appropriate scaffold for bone regeneration based on

biodegradable polymer coated 45S5 Bioglass®-based scaffolds to satisfy the structural,

physico-chemical, mechanical and biological properties of native bone-tissue. The variety of

biodegradable polymer coated Bioglass®-based scaffolds investigated is reported in

Chapter 4, based on the basic requirements for suitable bone scaffolds such as mechanical

and in vitro biological properties.

The biodegradable polymer coated Bioglass®-based scaffolds were developed

further for imparting a local drug-delivery capability for bone tissue engineering. The

combination of synthetic and natural polymers is carried out as suitable layered coating on

the Bioglass®-based scaffold, which provides a dual functions of improving the mechanical

properties and also acting as a drug carrier. The details are discussed in Chapter 5.

The fabrication and characterization of engineered scaffolds for cartilage

regeneration concerning the selection of biomaterials and fabrication techniques to achieve

the required structural architecture, morphology, physico-chemical and mechanical

properties is detailed in Chapter 6. In this part of the study, alginate-based scaffolds are

fabricated by two different techniques, including freeze-drying and electrospinning

techniques, in order to study the effects of two different architectures and geometries of the

obtained scaffolds. In addition, their morphology, physico-chemical and mechanical

properties are reported and discussed.

Page 54: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

40

Chapter 7 discusses the design and fabrication of bi- or multi-layered scaffolds

suitable for osteochondral tissue engineering based on the optimized scaffold materials and

fabrication methods. Recent strategies discussed previously, considering the biomimetic

approach leading to integrated- and monolithic-layered scaffolds are included in this work to

elucidate the advantages and disadvantages of each design of multilayered scaffolds.

The cellular response and activity on the optimized designed scaffold for bone and

cartilage regeneration are discussed in Chapters 8 and 9, respectively. In detail, MG-63

osteoblast-like cells are seeded on Bioglass®-based composite scaffolds to study cell

proliferation and metabolism. While chondrocytes and MSCs are seeded on 3D

alginate/chondroitin sulfate-based foams to evaluate cell adhesion, proliferation and

differentiation, and cartilaginous matrix formation, which provides basic information about

the suitability of the scaffolds for cartilage tissue regeneration.

In order to improve further this research work and to reach closer to the ideal

osteochondral scaffold highly feasible for in vitro and in vivo culture studies and available for

use in clinical practices, the summary and future perspectives are highlighted in Chapter 10.

Page 55: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

41

Figure 3. 1 Schematic diagram of entire tasks carried out in the dissertation thesis.

Page 56: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

42

Page 57: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 4

Preparation and Characterization of Biodegradable Polymer Coated

45S5 Bioglass-Based Scaffolds for Subchondral Bone Tissue

Engineering Applications

4.1 Introduction

Osteochondral tissue engineering involves the regeneration of both cartilage and SB

tissues, which are connective tissues providing the body with mechanical support and

protection [16,139,140,261]. Morphologically, bone is classified into two forms with different

structural and functional properties: cortical and cancellous bone [12,262]. SB demonstrates

the characteristics of cancellous bone, as described in Chpater 2. In order to engineer

scaffolds for SB regeneration, an attempt to mimic the natural tissue structure is

fundamental. In general terms, biomaterials-based scaffolds, cells and active molecules or

growth factors are three essential components for bone tissue engineering [263]. Especially,

scaffolds as the temporary framework of tissue regeneration, e.g. artificial ECM, are crucial

to achieve the required properties of new bone tissue, including structural and mechanical

properties [34]. By this approach, optimized scaffolds must be developed via the selection of

suitable materials and fabrication methods. Regarding ideal scaffolds for bone regeneration,

3D highly porous templates with tailored porosity, pore size and high interconnectivity are

generally required. In addition, bone tissue engineering requires biocompatible and

biodegradable scaffolds made from biomaterials exhibiting biodegradation matching the rate

of bone tissue formation. The biomaterials should also enable cell attachment, proliferation

and differentiation, ideally exhibiting osteoconductive and osteoinductive properties.

Moreover, the bone tissue scaffolds must have sufficient mechanical properties (mimicking

mechanical properties of natural bone) for load bearing applications and adequate structural

stability for use in clinical practice.

Page 58: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

44

According to these required properties, expecially in relation to bone bonding, 45S5

Bioglass was chosen material for bone scaffold in the present study. Highly porous 45S5

Bioglass®-based scaffolds can be fabricated by foam replication technique [32]. The

architecture of such scaffolds is similar to that of cancellous bone, being also similar to the

structure of the underlying SB in the osteochondral interface tissue. However, porous 45S5

Bioglass®-based scaffolds are characteristically brittle and exhibit low mechanical stability,

also the relatively high porosity of scaffolds results in reduced mechanical properties [34].

Therefore, biodegradable polymer coatings are applied to improve the mechanical properties

and structural stability of scaffolds, while their porosity and bioactivity are maintained [52,56].

Several biodegradable polymers, including natural polymers and synthetic polymers, were

studied in this area as suitable coatings for highly porous bioceramic scaffolds [52–

57,87,96,98,142,264–270]. Different polymers provide different characteristics, degradation

rate, and mechanical properties. Both natural (Alg and Gel) and synthetic (PDLLA and

PHBHHx) biodegradable polymers are therefore considered in this study in order to compare

their performance and efficiency as a coating on Bioglass®-based scaffolds, mechanical

properties and in vitro bioactivity are mainly considered. Improvement of the mechanical

strength and structural stability of scaffolds was aimed at achieving as the polymer would fill

and bridge the microcracks on the struts of Bioglass®-based scaffolds. In addition, the

polymer coating should increase the thickness of the struts without closing the pores.

Another purpose of this work is to optimize the coating conditions of each investigated

polymer and to compare their performances for appropriate scaffolds in osteochondral tissue

engineering.

4.2 Materials and methods

4.2.1 Fabrication of 45S5 Bioglass®-based scaffolds

45S5 Bioglass®-based scaffolds were prepared by the foam replication method, as

described in Chapter 2. In brief, this technique involves coating PU foam with a Bioglass®

slurry. The slurry for the immersion of PU foam was prepared as follows. PVA, purchased

from Merck KGaA, Germany, was dissolved in DI H2O with concentration of 3.5 wt/v %.

Page 59: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

45

Then 45S5 bioactive glass powder, of 45S5 Bioglass® composition (Schott electronic

packaging GmbH, Germany) was added to PVA solution with concentration of 40 wt/v %.

The whole procedure was carried out at 80 C under vigorous magnetic stirring for 2 h. The

“Eurofoam” PU foam with 45 ppi (pore per inch) served as a sacrificial template. It was cut to

size 10 mm × 10 mm × 10 mm. Then PU foams were immersed in the prepared slurry for 1

min. The foams were then removed and the extra slurry was completely squeezed out

manually. The samples were then dried in an oven at 60 C for 12 h. The coating thickness

of the samples (green bodies) was increased by repeating the slurry coating procedure for

three times. Heat treatment was carried out for burning out PU templates and for sintering

45S5 Bioglass® structure. The burning condition of PU templates and sintering condition

were designed to be 450 C for 1 h and 1100 C for 2 h, respectively. The heating and

cooling rates were 2 and 5 C/min, respectively.

4.2.2 Preparation of biodegradable polymer coated 45S5 Bioglass®-based scaffolds

Different types of polymer coated 45S5 Bioglass®-based scaffolds were fabricated.

Natural biodegradable polymers, including Alg and Gel, and biodegradable synthetic

polymers, including PDLLA and PHBHHx, were chosen to coat 3D 45S5 Bioglass®-based

scaffolds by using dipping technique. The characteristics of the various polymer solutions

used are shown in Table 4.1. A simple manual dip coating method was used for infiltrating

polymer into the scaffolds’ structure and to adhere the polymers to the surface of struts. The

procedure of polymer coating of scaffolds is explained for each polymer used in the following

paragraph.

For Alg coating, coating solution was prepared as follows: Alg was dissolved in DI

H2O with the concentration of 2 wt/v % at room temperature, under magnetic stirrer for 2 h.

The 2 wt/v % solution was diluted to 0.5 and 1.5 wt/v % with DI H2O. Then 45S5 Bioglass®-

based foam was immersed in Alg solution for 5 min and dried at room temperature for 24 h.

The coating process of Gel and PDLLA coated 45S5 Bioglass®-based scaffolds was carried

out using the same procedure, following the conditions shown in Table 4.1. In contrast, in

case of PHBHHx coated 45S5 Bioglass®-based scaffolds, PHBHHx was dissolved in

chloroform (Merck KGaA, Germany) with concentration of 1 and 5 wt/v % at 50 C, under

Page 60: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

46

magnetic stirring for 2 h. 45S5 Bioglass®-based scaffolds were soaked in PHBHHx solution

for 10 sec and 5 min, respectively, and dried at room temperature for at least 30 min before

starting another coating layer. In our experiments, the coating process was carried out 30

times in the case of 1 wt/v % solution. Each coating sample was labeled as Alg-c-BG, Gel-c-

BG, PDLLA-c-BG and PHBHHx-c-BG for Alg, Gel, PDLLA and PHBHHx coated scaffolds,

respectively.

Table 4. 1 Polymer coating conditions for polymer coated 45S5 Bioglass®-based scaffolds.

An as-sintered rectangular shaped 45S5 Bioglass®-based scaffold, with the dimensions of 8

mm × 8 mm × 8 mm, was soaked in 5 ml of each polymer solution.

Biodegradable

polymers

Coating conditions

Concentrations

(wt/v %) Soaking times No. of dipping cycles

Alg 0.5, 1.5, 2 5 min 1

Gel 1.5, 3, 5 5 min 1, 3

PDLLA 2, 5, 8 5 min 1

PHBHHx1 1, 5 10 sec, 5 min 30, 1

4.2.3 Characterization and mechanical testing

(i) Porosity

The percent porosity of uncoated (P1) and polymer coated (P2) scaffolds was

calculated from Eq. 4.1 and 4.2 [56]:

P1 (%) = [1 – (Wscaffold/BG/Vscaffold)] × 100 (4.1)

P2 (%) = [1 – (Wscaffold/BG/Vscaffold) + (Wpolymer coating/polymer/Vpolymer coating)] × 100 (4.2)

; where Wscaffold, Wcoated scaffold and Wpolymer coating are the weight of sintered scaffold, polymer

coated scaffold and polymer coating (Wcoated scaffold – Wscaffold), respectively, BG and Polymer

are the density of solid 45S5 Bioglass® (BG = 2.7 g/cm

3) and solid polymer (Alg = 1.02

1 30 coating cycles were done in the case of 1 wt/v % PHBHHx solution and one coating cycle was done in the case

of 5 wt/v % PHBHHx solution.

Page 61: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

47

g/cm3, Gel = 0.98 g/cm

3, PDLLA = 1.26 g/cm

3 and PHBHHx = 1.2 g/cm

3), respectively, and

Vscaffold, Vcoated scaffold and Vpolymer coating are the volume of sintered scaffold, polymer coated

scaffold and polymer coating (Vcoated scaffold – Vscaffold), respectively, determined from the

dimensions of scaffolds.

(ii) Microscopy

The microstructure of the scaffolds was characterized by scanning electron

microscopy (SEM; LEO 435 VP), before and after coating. Samples were sputter coated and

observed at an accelerating voltage of 10 kV. The pore size was measured from images

taken from SEM by using software Image J (Version 1.42S).

(iii) XRD analysis

The scaffolds were characterized by using X-ray diffraction (XRD) analysis in order

to assess the crystallinity after sintering. The scaffolds were ground into powder. Then 0.1 g

of the powder was collected for XRD (Siemens D500) analysis, employing Cu k radiation.

Data were collected over the range of 2 = 15 - 70 using a step size of 0.02 and a counting

time of 25 sec per step.

(iv) In vitro acellular bioactive study

The bioactivity of uncoated and polymer coated scaffolds was investigated by

immersion in SBF solution (pH 7.4 at 37 °C) for 1, 3, 7, 14 and 28 days. The SBF solution

was prepared, following Kokubo et al. [271]. Each sample (with the dimensions of 8 mm × 8

mm × 8 mm) was placed in polystyrene bottle containing 50 ml SBF, then it was incubated in

orbital shaker (IKA RS 4000i) at 37 C with the shaker speed of 90 rpm. The SBF solution

was replaced twice a week in order to avoid the changes in the chemistry of samples. At

each time point, the sample was removed, cleaned with DI water and dried at room

temperature for 24 h. The morphology of scaffolds after immersion in SBF was observed by

SEM. The formation of HCA crystals was characterized with the use of Fourier-transform

infrared (FTIR) spectroscopy and x-ray diffraction (XRD) analysis.

(v) FTIR analysis

The possible presence of HA on the scaffolds upon immersion in SBF was identified

with the use of Fourier-transform infrared spectroscopy (FTIR; Nicolet Nexus 6700, Thermo

Page 62: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

48

Scientific, Waltham, MA). For these measurements, the scaffolds were grinded and mixed

with anhydrous potassium bromide (KBr) powder in the ratio of 1/300 by wt. The mixture was

pressed to a pellet by using an electro-hydraulic press (MAUTHE MASCHINENBAU PE-010;

Wesel, Germany) with a pressure of 10 × 104 N. Then the pellet was analyzed by using

transmission mode with a resolution of 4 cm-1

, in the wavenumber range of 4000 - 400 cm

-1

and applying 64 scans.

(vi) Mechanical testing

The porous scaffolds prepared with dimensions of 8 mm × 8 mm × 8 mm were

tested using a universal testing machine (Zwick Z050) by applying a compression load at a

cross-head speed of 5 mm/min, preload at 0.1 N and maximum load at 1 kN. Stress-strain

curves were recorded to determine the relevant mechanical properties, compressive

strength and to assess the work of fracture. Six specimens were tested for each condition

and data were presented as mean ± standard deviations (SD).

4.2.4 Statistical analysis

Statistical comparisons were carried out using one-way ANOVA method, which p =

0.05 was considered to be a significant difference.

4.3 Results and discussion

4.3.1 Morphology

(i) General characterization

45S5 Bioglass®-based scaffolds fabricated by foam replication technique exhibited

porosity of 92 % with the pore size in the range of 100 - 700 μm. A highly porous

trabecular structure and high pore interconnectivity are obtained and this can be appreciated

in Fig. 4.1 (A). This porosity is desired for facilitating high ability of cell seeding and to enable

efficient cell proliferation and bone ingrowth, as discussed in Chapter 2. Multiple slurry

coatings were carried out to increase the thickness of struts, relative density, and

consequently to improve the strength of scaffolds, compared with that of scaffolds prepared

from only one coating. The scaffolds made by a triple coating process were sufficiently

strong for handling safely in the laboratory by forceps or fingers, while scaffolds made by a

Page 63: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

49

single coating were not. The strut of scaffold shown in Fig. 4.1 (B) exhibits quite a dense

surface indicating good densification achieved at high sintering temperature (1100 C),

which was caused by well bonding of Bioglass® particles. The extensive densification and

the presence of crystalline phase in as-sintered scaffolds are expected to lead to

improvement of the mechanical properties [32].

The crystalline phase of the as-sintered scaffolds was confirmed by the results of

XRD analysis (Fig. 4.2), which shows the sharp crystalline peaks of Na2Ca2Si3O9 compared

to the pattern of as-received 45S5 Bioglass®

powder, matching the same angular location

and the same form of peaks as in a previous study on sintered 45S5 Bioglass® [32].

Figure 4. 1 SEM images of 45S5 Bioglass®-based scaffolds fabricated by foam replication

technique: (A) 3D porous structure and (B) surface of scaffold struts.

Page 64: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

50

Figure 4. 2 X-ray patterns of as-received Bioglass® and as-sintered 45S5 Bioglass

®-based

scaffolds. The major peaks of the phase Na2Ca2Si3O9 are marked by ■.

However, bioceramics and bioactive glasses in a porous form have usually low

strength and fracture toughness because of their intrinsic brittleness, leading to a hindrance

in their clinical applications and commercial uses. As discussed in Chapter 2, polymeric

coating on porous foams is one method being increasingly investigated to improve the

mechanical properties of scaffolds [52,55]. Different polymers, including natural polymers

(Alg and Gel) and synthetic polymers (PDLLA and PHBHHx), were infiltrated in this study

into the porous scaffolds to test the hypothesis that the polymer coating can improve the

mechanical strength and toughness of the porous scaffolds by filling the cracks and open

pores on the struts of scaffolds. For each polymeric coating, the polymer concentration,

immersion times and number of dipping cycles were varied (Table 4.1).

(ii) Alginate coated scaffolds

First, Alg coated Bioglass®-based scaffolds (Alg-c-BG) were studied. At 2 wt/v % Alg

concentration, it was found that some pores were clogged with a polymer membrane, as

shown in Fig. 4.3 (E), which is not satisfactory as it could hinder cell migration during in vitro

cell culture study. Additionally, the surface of coated scaffolds was not homogeneous due to

Page 65: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

51

a too high concentration used and subsequent high viscosity of the coating solution (Fig. 4.3

(F)), leading to hindrance of polymer infiltration. In contrast, coating with a 0.5 wt/v % Alg

solution led to scaffolds, which were not stable due to the fact that the polymer did not fully

cover the surface of the struts (Fig. 4.3 (B)), even though pores were not clogged (Fig. 4.3

(A)). In case of 1.5 wt/v % Alg concentration, the coating was homogeneous and no pores

were seen to be closed by a polymer membrane (Fig. 4.3 (C)). Fig. 4.3 (D) shows the

smooth surface of a coated scaffold using a 1.5 wt/v % Alg solution, the polymer coating

layer covered efficiently the entire surface of the struts. Thus qualitatively optimized Alg-c-

BG scaffolds were obtained using a solution with 1.5 wt/v % Alg concentration. It can be

concluded that a too high Alg concentration ( 2 wt/v % concentration) is not suitable for

infiltrating into the scaffold’s porous structure due to the high viscosity. Therefore, the

viscosity of Alg solution plays a key role in the quality of coating, as also reported in the

previous study [54].

(iii) Gelatin coated scaffolds

Fig. 4.4 (A-F) shows structural and morphological features of Gel coated Bioglass®-

based scaffolds (Gel-c-BG), which were obtained from 1.5, 3 and 5 wt/v % Gel

concentrations, 5 min of immersion time and single dip coating. It was found that 3 and 5

wt/v % Gel concentrations exhibited many clogged pores due to the high solution viscosity

(Fig. 4.4 (C and E)). Gel coatings with lower concentration (1.5 wt/v %) were found to lead to

non-clogging coatings, while the scaffolds obtained were not sufficiently strong because the

coating layer was too thin. Taking into account the thickness of coating layer can be

increased by increasing the number of coating cycles. The morphology of the triple coated

scaffold is shown in Fig. 4.4 (G and H). In this study, three dipping cycles were qualitatively

considered to lead to suitable mechanical properties and to maintain the suitable porous

characteristics of the scaffolds.

Preliminarily, the natural-derived polymers (Alg and Gel) can be recommended to be

used as coating of Bioglass-based scaffolds using relatively low polymer concentrations in

order to obtain homogeneous coatings. In addition, the mechanical strength can be

Page 66: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

52

enhanced by increasing the number of dipping cycles, as discussed in the case of Gel

coatings.

(iv) PDLLA coated scaffolds

For PDLLA coated scaffolds (PDLLA-c-BG), it is seen from Fig. 4.5 that their open

pores were still maintained at PDLLA concentration of 5 wt/v % (Fig. 4.5 (C)). In contrast, the

pores were blocked with the 8 wt/v % coating (Fig. 4.5 (E)), considering that this polymer

solution was highly viscous. This phenomenon is also related to the fact that a polymer

solution with higher concentration takes a longer time for infiltrating through the porous

structure in comparison with lower concentration polymer solutions. Moreover, the used

organic solvent (i.e. DMC) is fast evaporated at room temperature, leading to non-

homogeneous coating on the whole surface area of the struts, as shown in Fig. 4.5 (E and

F). Fig. 4.5 (D) shows that the PDLLA coating integrated well on the strut’s surface, also

filling cracks and pores on the struts’ surface and it leads to a smooth surface. In contrary to

the struts’ surface of 3 wt/v % PDLLA-c-BG scaffold (Fig. 4.5 (A)), some pores on the struts

were observed after coating (Fig. 4.5 (B)).

(v) PHBHHx coated scaffolds

In case of PHBHHx coated scaffolds (PHBHHx-c-BG), to avoid the phenomenon of

closed pores, the optimized conditions should be the use of low polymer concentration and

short immersion time, and the thickness of the struts was increased by multiple coatings.

This hypothesis was thus investigated by coating the scaffolds with 1 wt/v % concentration,

10 sec of immersion time and 30 dipping cycles in comparision with coating the scaffolds

with 5 wt/v % concentration, 5 min of immsion time and one dipping cycle. It was found that

low concentration and short immersion time supported better homogeneous coating surface

in comparison with higher concentration, longer immersion time with one coating cycle

(compare between Fig. 4.6 (A and B) and Fig. 4.6 (C and D)). This result can be described

by the fact that PHBHHx has elastomeric properties (rubber-like behavior), which its

solutions with higher concentration and longer soaking time tend to stick at the intersection

of struts impeding efficient infiltration, leading to clogged pores (Fig. 4.6 (A)). In contrary to

the use of low concentration and short soaking time, the thickness of coating was built up by

Page 67: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

53

30 dipping cycles without closing pores, as shown in Fig. 4.6 (C). This effect thus causes

different coating morphologies.

(vi) Summary of results on polymer coatings

The optimized coating parameters of different biodegradable polymers are

summarized in Table 4.2. By Alg coating, the porosity of scaffolds was reduced from 92 to

84 %. Moreover, Alg formed thin film well adhering on the struts of the scaffold, showing a

smooth surface. The Gel-c-BG scaffolds (porosity 82 %) obtained by using a 1.5 wt/v %

solution, 5 min soaking time and 3 dipping cycles showed the morphology in a similar

manner to the case of Alg coating. These results can be described by the fact that since

both, natural polymer coatings (Alg and Gel) and the strut surface of Bioglass-based

scaffolds, exhibit hydrophilic feature, they are chemically compatible. According to this fact,

the viscosity of natural polymer solutions mainly influences the quality of coating. In contrast,

the optimized conditions of synthetic polymer coatings provided a thinner coating layer on

the scaffolds than that observed in the cases of natural polymer coating. The porosity of

PDLLA-c-BG and PHBHHx-c-BG scaffolds was 85 % and 86 %, respectively. By

increasing concentrations 5 wt/v %, the PDLLA and PHBHHx solutions tended to form as a

membrane covering the pores of the scaffold, instead of coating onto the struts. It seems

that not only the viscosity of polymer solutions plays a key role in the obtained morphology

of synthetic polymer coated scaffolds, but also the different surface chemistries of synthetic

polymer and Bioglass surface influence the quality of coating.

Page 68: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

54

Figure 4. 3 SEM images of Alg coated 45S5 Bioglass®-based scaffolds with variable

concentrations, including (A, B) 1 wt/v %, (C, D) 1.5 wt/v % and (E, F) 2 wt/v %, the scaffolds

were coated once for 5 min soaking time.

Page 69: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

55

Figure 4. 4 SEM images of Gel coated 45S5 Bioglass®-based scaffolds with variable

concentrations, including (A, B) 1.5 wt/v %, (C, D) 3 wt/v %, (E, F) 5 wt/v %, the scaffolds

were coated once on for 5 min soaking time and (G, H) 1.5 wt/v % coated for 5 min and 3

dipping cycles.

Page 70: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

56

Figure 4. 5 SEM images of PDLLA coated 45S5 Bioglass®-based scaffolds with variable

concentrations, including (A, B) 3 wt/v %, (C, D) 5 wt/v % and (E, F) 8 wt/v %, the scaffolds

were coated once for 5 min soaking time.

Figure 4. 6 PHBHHx coated 45S5 Bioglass®-based scaffolds with variable concentrations,

soaking times and number of dipping times, including (A, B) 5 wt/v %, 5 min of soaking and

one coating cycle, and (C, D) 1 wt/v %, 15 sec of soaking time and 30 coating cycles.

Page 71: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

57

Table 4. 2 Optimized polymer coating conditions for different biodegradable polymer coated

45S5 Bioglass®-based scaffolds.

Biodegradable

polymers

Coating conditions

Concentrations (wt/v %) Soaking times No. of dipping cycles

Alg 1.5 5 min 1

Gel 1.5 5 min 3

PDLLA 5 5 min 1

PHBHHx 1 15 sec 30

4.3.2 Mechanical properties

Although as-sintered scaffolds can be handled without crumbling, they are not

particularly strong. As shown in Fig. 4.7 (A), the typical compressive stress-strain curve of

uncoated BG scaffold exhibited a jagged behavior due to the presence of micro-cracks on

the as-sintered struts. The compressive strength and modulus were low, in the average of

0.021 ± 0.003 MPa and 0.05 ± 0.01 MPa, respectively. By coating Bioglass®-based scaffolds

with different polymers, the compressive strength and elastic modulus were significantly

improved (* p 0.05) due to the reduction of micro-cracks on the struts’ surfaces by polymer

filling. The compressive stress-strain curve of all types of polymer coated scaffolds showed

larger area under the curve compared to that of uncoated scaffolds. This behavior means a

higher mechanical stability of the scaffolds (in comparison to uncoated scaffolds) for

handling and further in vitro acellular and cellular evaluations. All biodegradable polymers

led to improvement of the mechanical properties of the base scaffolds in the similar range of

0.1 - 0.3 MPa for the compressive strength. The optimized synthetic polymer coatings

(PDLLA and PHBHHx) led to significantly higher scaffold elastic modulus (# p 0.05) than

the optimized natural polymer coatings (Alg and Gel) (Fig. 4.7 (B)), which is the result of the

intrinsic characteristics of the synthetic polymers concerning higher mechanical properties

compared to natural polymers. Even though the natural polymers, both Alg and Gel,

exhibited an homogeneous coating and also a thicker coating layer onto the scaffolds, the

Page 72: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

58

thin PDLLA and PHBHHx coatings might be more effective in terms of crack-filling behavior

[272]. Generally, natural polymers have a limitation due to their relatively low mechanical

properties and they are not suitable for use in load bearing applications. The suggested

behavior was supported by the appearance of the specimens after the compression strength

test (Fig. 4.7 (C)). Uncoated Bioglass-based scaffolds, which are brittle and have jagged

stress-strain curve, were completely crushed and reduced to powder after compression test,

while the polymers coated Bioglass-based scaffolds did not crumble and showed though

fracture behavior. For example, PDLLA- and PHBHHx-c-BG exhibited higher elasticity

(higher elastic modulus). After compressive load, the scaffolds could recover partially their

initial dimensions (i.e. up to 70 % by height of the specimen). Especially, PHBHHx coated

scaffolds showed elastomeric mechanical behavior, similarly to the study of Deng et al.

[233]. The present PHBHHx-c-BG scaffolds could recover suddenly their initial dimensions

after removing the force. In contrast, natural polymer coated Bioglass®-based scaffolds

showed different fracture behavior, e.g. after loading, the struts were broken and compacted.

This behavior means that these specimens absorbed force but they could not transfer force.

The specimens were separated apart in small pieces in the cases of Alg- and Gel-c-BG

scaffolds. These qualitative results thus clearly confirm that the PDLLA and PHBHHx

coatings in combination with the optimized coating conditions impart better improvement in

terms of mechanical properties of scaffolds due to the characteristics and mechanical

properties of polymers coating layer themselves and the ability to fill the cracks.

Page 73: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

59

Figure 4. 7 (A) Representative compressive stress-strain curves of biodegradable polymer

coated Bioglass®-based composite scaffolds in comparison with the curve of uncoated

Bioglass®-based scaffolds, (B) normalized mechanical properties (compressive modulus and

strength) of coated scaffolds compared to those of uncoated scaffolds (as reference) and (C)

appearance of coated scaffolds after compression load. * indicates significantly different

mechanical properties of coated scaffolds in comparison with those of uncoated scaffolds,

and # indicates significantly different mechanical properties of synthetic polymer coated

scaffolds in comparison with the properties of natural polymer coated scaffolds (p 0.05).

4.3.3 Degradation behavior

The biodegradation behavior of Bioglass®-based scaffolds was investigated. This

parameter is very important because the biodegradation rate of scaffolds has an impact on

the success of tissue regeneration. In this study, the degradation behavior of uncoated and

coated Bioglass®-based scaffolds was monitored by weight loss and pH variation upon

immersion in SBF solution. The weight loss of uncoated Bioglass-based scaffolds (Fig. 4.8

(A)) increased with culturing time. Upon immersion for 7 days, uncoated Bioglass®-based

Page 74: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

60

scaffolds lost gradually their weight up to 25 %. From 7 to 28 days, the rate of weight loss

was reduced to 10 %. This result can be explained that at the initial immersion time, the

dissolution of Bioglass surface took place, leading to the significant weight loss. On the

other hand, HA formed on the surface of Bioglass after longer immersion in SBF, which is

initiated by ionic exchange between Bioglass surface and SBF (as detailed in Chapter 2).

Therefore, the weight loss of Bioglass-based scaffolds was compensated with the formation

of HA after 7 days of immersion. When compared to the pH study in SBF (Fig. 4.8 (B)), the

pH value of the SBF solution after 1 day of immersion was seen to be significant increase

from 7.4 to 8.5, which was attributed to the ionic exchange between the ions from Bioglass

(such as Si4+

and Na+) and H

+ and H3O

+ from SBF. Subsequently, the pH value was reduced

to 7.8 after 7 days of immersion and it remained constant in the range of 7.6 – 7.8 upon 28

days immersion. This pH variation corresponded to the trend of weight loss.

The weight loss of biodegradable polymer coated scaffolds increased with culturing

time and it was seen to vary with the types of polymer coating. All scaffold types could be

classified into two main trends, including water soluble polymers (Alg and Gel) and water

insoluble polymers (PDLLA and PHBHHx). Both Alg- and Gel-c-BG scaffolds showed high

resorption rate in the immersion time investigated. After 14 days immersion in SBF, Gel-c-

BG scaffolds maintained fast resorption rate and lost around 70 % of their weight at day 28,

Fig. 4.8 (A), whereas Alg-c-BG scaffolds maintained constant resorption rate in the period

between 14 days until 28 days immersion, which showed a weight loss of 60 %. This result

demonstrated that Gel coating showed the fastest bioresorption rate among all polymers

investigated. This behavior is likely caused by the hydrophilic nature of Gel, leading to high

ability of water uptake, which is one of the important factors influencing biodegradation rate.

In addition, the incubation conditions at 37 °C can influence the dissolution of gelatin. On the

other hand, Na-Alg can be partly crosslinked with Ca2+

in SBF forming Ca-Alg; therefore, this

phenomenon enhances the Alg network stability and retards the dissolution of Alg coating

during immersion in SBF [273].

In contrast, PDLLA and PHBHHx, which exhibit hydrophobic characteristic, tend to

degrade initially by a hydrolysis process. Both PDLLA and PHBHHx showed lower

Page 75: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

61

degradation rate compared to Alg and Gel. The biodegradation behavior of PDLLA- and

PHBHHx-c-BG scaffolds exhibited the same trend (Fig. 4.8 (A)), indicating that both aliphatic

polyesters provide the same degradation profile. Also, the degradation profile indicated by

weight loss as function of immersion times corresponded to pH variation in all polymers

investigated, as shown in Fig. 4.8 (B). At day 1 in immersion in SBF, pH values of all

investigated coated scaffolds were significant lower compared to the value of uncoated

scaffolds. This phenomenon can be explained that the polymer coating may inhibit the ion

release of Bioglass, and therefore lead to reduced pH value. Furthermore, it is likely that

after 3 days of immersion, acidic groups resulting from degradation of each polymer, in

particular PDLLA and PHBHHx coatings, decreased the pH value of SBF soultion, while Si

and Ca ions released from the Bioglass® surface compensated the pH decrease. The pH

values in the case of both PDLLA- and PHBHHx-c-BG scaffolds maintained in the range of

7.1 – 7.2 after 7 days of immersion. In contrast, Gel and Alg coatings led to higher pH values

than PDLLA and PHBHHx coatings, which is likely related to the different acidicity of each

degradation product resulting from each polymer. Therefore, the pH of the SBF solution was

influenced by the degradation rate of the polymer coating and the dissolution behavior of the

Bioglass®. However, different polymers provided different degradation rates and the range of

pH values of different polymer coated Bioglass® scaffolds was different during the immersion

time.

Page 76: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

62

Figure 4. 8 (A) % weight loss of biodegradable polymer coated 45S5 Bioglass®-based

scaffolds after 1, 3, 7, 14 and 28 days of immersion in SBF and (B) variation of the pH of the

SBF solution.

4.3.4 In vitro bioactivity

(i) Uncoated Bioglass scaffolds

The in vitro bioactivity of as-sintered Bioglass®-based scaffolds was assessed by

investigating the formation of HCA on scaffolds’ surfaces upon immersion in SBF for 1, 3, 14

and 28 days. Fig. 4.9 shows the formation of HCA on struts’ surface after immersion in SBF

for 28 days. It was observed that apatite started to form after immersion in SBF for 1 day

(Fig. 4.9 (A)). HCA formation was more clearly observed on the surface of the scaffolds after

immersion in SBF for 3 days (Fig. 4.10 (B)). After 14 days of immersion in SBF, aggregation

of HCA was observed and the surface of the scaffolds was completely covered after 28 days

in SBF (Fig. 4.9 (C and D)).

Page 77: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

63

Figure 4. 9 SEM images of as-sintered 45S5 Bioglass®-based scaffolds after (A) 1, (B) 3,

(C) 14, and (D) 28 days of immersion in SBF, showing possible formation of HA indicated

qualitatively by the morphology of the deposited structures.

Apatite formation was confirmed by FTIR spectroscopy (Fig. 4.10 (A)). Possible

apatite formation was evidenced by generation of amorphous and crystalline HA phase,

which was expressed as a double peak at 605 cm-1

and 565 cm-1

(P-O bending)

[37,274,275], as depicted by ▲ in Fig. 4.10. Moreover, a double broad peak at 1488 cm-1

and 1420 cm-1

corresponding to amorphous CaP phase was detected after 1 days of

immersion in SBF [37], as depicted by ■ in Fig. 4.10 (A).

In addition, XRD analysis confirmed the formation of HA (Fig. 4.10 (B)). The peak of

HA phase was detected at 2 32 [32,37,275]. The height of the HCA peak increased with

increasing times in SBF, e.g. between 14 and 28 days, indicating that higher amount of HCA

occurred with increasing immersion time in SBF. In contrast, sharp diffraction peaks of

Na2Ca2Si3O9, which is the crystalline phase in as-sintered scaffolds reduced with immersion

time and they nearly disappeared after 28 days of immersion in SBF. This result indicates

the reduction of the crystallinity of as-sintered 45S5 Bioglass®–based scaffolds with

increasing culturing time in SBF, which has been previously reported in literature [32,275].

Page 78: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

64

Figure 4. 10 (A) FTIR spectra of as-sintered 45S5 Bioglass®-based scaffolds after different

immersion times in SBF, in comparison with the scaffolds before immersion. The

characteristic peaks of HA are marked by ▲ and ■, and (B) XRD patterns of as-sintered

45S5 Bioglass®-based scaffolds after different immersion times in SBF, in comparison with

the scaffolds before immersion. The major peak of HA is marked by , while the crystalline

peaks of Na2Ca2Si3O9 were marked by ■.

(ii) Alg coated scaffolds

After coating with biodegradable polymers, each type of polymer coatings showed

slightly different behavior when immersed in SBF. For example, after 1 day immersion in

SBF, the Alg coating started to peel off the struts, as depicted by the arrow in Fig. 4.11 (A).

There was no apatite formation observed after 1 day on the Alg coated scaffolds, which was

confirmed by FTIR spectroscopy, as shown in Fig. 4.12 (A). After 1 day immersion in SBF,

there was no appearance of P-O bending mode (at 560 - 550 cm-1

and 610 - 600 cm-1

),

which are the characteristic peaks of HA. This behavior was not different from that of

uncoated scaffolds, e.g. the possible apatite formation was clearly observed after 3 days

immersion in SBF (Fig. 4.9 (B)). It can be concluded that Alg coating does not affect the

bioactivity of Bioglass®. Moreover, after 3 days immersion in SBF, the HCA formation was

detected by FTIR spectroscopy, as depicted by ▲ in Fig. 4.12 (A), even though it was not

clearly observed by SEM (Fig. 4.11 (B)). Finally, after 28 days, it was found that the Alg

Page 79: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

65

coating still maintained the adherence to the scaffold, which is probably caused by

crosslinking between Alg coating and Ca ions in SBF, and thus retarding the dissolution of

Alg, as previously described in Section 4.3.3. Also, HA covered the struts’ surfaces (Fig.

4.11 (C)). The FTIR spectra of coated scaffolds after 28 days exhibited high intensity of the

HA characteristic peaks at 604 and 563 cm-1

attributed to P-O bending, as marked by ▲ in

Fig. 4.12 (A). In addition, the presence of a double broad peak at 1453 and 1420 cm-1

attributed to amorphous CaP phase [37], confirmed the HA formation after 28 days

immersion in SBF. The characteristic peak of Alg coating, in particular at 1638 cm-1

(-C=O

stretching) as marked by ■, maintained after 28 days in SBF, indicating the lasting of Alg

coating. The results of the in vitro bioactivity study of Alg-c-BG scaffolds agreed with the

data reported by Erol et al. [54] in that Alg coating has no negative effect on the bioactivity of

Bioglass®-based scaffolds.

(iii) Gel coated scaffolds

The Gel coating was dissolved rapidly in SBF due to the high bioresorption rate of

gelatin [276]. Therefore, the Gel coating should be completely degraded after only a few

days in SBF as stated by Bellucci et al. [276]. However, FTIR spectra shows that after 3

days immersion in SBF, Gel was detected by the presence of the characteristic peak of –

C=O and –C-N- stretching at 1637 cm-1

(as marked by ■ in Fig. 4.12 (B)) [78,276]. With

increasing time in SBF, the intensity of the characteristic peak of Gel was reduced with time

and the peaks almost disappeared after 28 days of immersion. At the same time, HA was

formed following the fracture of the coating, which allowed ions release from the Bioglass®

substrate. Apatite crystals were not clearly observed by SEM after immersion for 3 days,

while after 28 days immersion such crystals were visible (Fig. 4.11 (D, E and F)). Moreover,

apatite formation was confirmed by the presence of peaks at 604 and 565 cm-1

attributed to

the P-O bending mode of crystalline apatite and amorphous CaP, respectively. The peaks of

P-O bending were clearly detected on samples after 14 days immersion in SBF,

incorporation with an existence of double broad at 1464 and 1420 cm-1

attributed to

amorphous CaP phase (Fig. 4.12 (B)).

Page 80: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

66

(iv) PDLLA coated scaffolds

Synthetic polymer coated Bioglass®-based scaffolds exhibited different behavior

related to HA formation in comparison to natural polymer coated scaffolds. Due to the fact

that synthetic polymers have slower degradation rate in comparison with natural polymers,

the PDLLA and PHBHHx coatings maintained adherence with the scaffold during the

complete period of 28 days immersion in SBF. This behavior should extensively retard the

rate of ion release from the bioactive phase and consequently HA formation should be

delayed. It is likely that apatite crystals formed on both, PDLLA-c-BG and PHBHHx-c-BG

scaffolds, after 3 days immersion in SBF (Fig. 4.11 (G and J)). The reason for this behavior

could be the hydrolytic (erosion) mechanism [268,277]; active in these polymers, as shown

in Fig. 4.11 (G) in the case of PDLLA-c-BG scaffolds. The polymer coating was perforated in

contact with SBF, hence providing channels for leakage of ions from the Bioglass® substrate

and exposing nucleation sites for HA formation. After 28 days immersion, HA covered the

entire surface of the scaffold, as reported also by Bretcanu et al. [55] for PDLLA coated

scaffolds. The appearance of PDLLA and HCA formation after immersion in SBF for different

time points was confirmed by FTIR spectroscopy (Fig. 4.12 (C)). Upon 28 days in SBF, the

characteristic peaks of PDLLA were maintained, including the peaks at 1750 cm-1

attributed

to -C=O and at 1659 cm-1

attributed to –COO- [278], thus confirming the existence of PDLLA

coating. HCA phase was evidenced by the presence of the characteristic peaks at 603 and

565 cm-1

attributed to P-O bending, as marked by ▲ in Fig. 4.12 (C). In addition, a broad

double peak at 1454 and 1422 cm-1

attributed to amorphous CaP phase was gradually

increased, while the –C=O peak at 1751 cm-1

, as marked by ■, indicating PDLLA coating

decreased in the intensity with increasing time in SBF.

(v) PHBHHx coated scaffolds

The PHBHHx-c-BG scaffolds exhibited possible HA formation similarly to the

PDLLA-c-BG scaffolds. PHBHHx is a polyester-based polymer, which initially degrades by

hydrolytic mechanism [277,279]. In Fig. 4.11 (J), the channels caused by the fracture of the

polymer coating are visible, which were enabled the nucleation of HCA crystals. After 28

days immersion in SBF, the PHBHHx coating maintained adherence to the scaffold, as

Page 81: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

67

showed by an inset in Fig. 4.11 (L). HA fully covered on the PHBHHx coated Bioglass®

surface. The bioactivity of PHBHHx-c-BG was confirmed by FTIR spectroscopy (Fig. 4.12

(D)); the characteristic peaks of HA was detected at 603 and 564 cm-1

, according to P-O

bending mode and a broad peak at around 1480 - 1420 cm-1

, according to amorphous CaP

phase. It was obviously detected after 3 days immersion in SBF and the intensity of the

peaks increased with increasing immersion time. At the same time, the peaks of PHBHHx

coating at 1750 and 1645 cm-1

, attributed to –C=O and –COO-

[280,281], respectively,

broadened with increasing immersion time, indicating partly degradation of PHBHHx coating.

(vi) Summary of results on the bioactivity

All investigated biodegradable polymer coated Bioglass®-based scaffolds were

confirmed to maintain the bioactive properties of Bioglass® by the formation of HA in contact

with SBF, even though they exhibit different behavior among them in terms of degradation

rate and formation of apatite. The PDLLA and PHBHHx coatings showed fast rate of HA

formation due to the fact that these coating were very thin and due to the occurrence of

hydrolytic biodegradation in contact with SBF. On the other hand, the resorption of natural

polymer coatings was initiated by water absorption, subsequent swelling and polymer

detaching from Bioglass® substrate occurred. HA formed by the channels/holes formed after

the polymer coating detached, which resulted in an open surface for ion release. This result

demonstrated that the selection of the polymer coating must be taken into consideration to

tailor degradation kinetics of composite scaffolds. Another factor is the quality of coating, i.e.

inhomogeneity of polymer coating, which initiates the pathway for ion release of Bioglass

substrate. Moreover, a suitable composite scaffold can be selected for controlled drug

release by tailoring the polymer coating layer as a drug carrier-system, thus leading to a

multifunctional scaffold for tissue engineering therapeutics [96,282]. This investigation is

reported in Chapter 5.

Page 82: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

68

Figure 4. 11 SEM images of biodegradable polymer coated 45S5 Bioglass®-based

composite scaffolds after immersion in SBF, showing formation of HA: (A, B, C) Alg-c-BG,

(D, E, F) Gel-c-BG, (G, H, I), PDLLA-c-BG and (J, K, L) PHBHHx-c-BG scaffolds for 1, 3 and

28 days, respectively.

Page 83: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

69

Figure 4. 12 FTIR spectra of biodegradable polymer coated 45S5 Bioglass®-based scaffolds

after 28 days of immersion in SBF: (A) Alg-c-BG, (B) Gel-c-BG, (C) PDLLA-c-BG, and (D)

PHBHHx-c-BG scaffolds. The characteristic peaks of HA are marked by ▲ and the

characteristic peaks of polymer coatings are marked by ■.

Page 84: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

70

4.4 Conclusions

Biodegradable polymer coated 45S5 Bioglass®-based scaffolds were successfully

fabricated in order to develop scaffolds with improved mechanical properties and stability. All

polymer coatings studies led to significant improvement of compressive modulus, strength

and toughness of the bare scaffolds. By comparison between biodegradable natural (Alg

and Gel) and synthetic (PDLLA and PHBHHx) polymers, each coating type showed

differences in the morphology of the scaffolds’ surface, in biodegradation behavior and

subsequently in vitro bioactivity. However, it was proven that all investigated polymer

coatings did not affect the interconnectivity, pore structure and even the in vitro bioactivity of

scaffolds. Coating of scaffolds by using PDLLA and PHBHHx led to structures of higher

mechanical properties in comparison to those of scaffolds coated by Alg and Gel. Therefore,

it can be concluded that the composite scaffolds developed in this study show the potential

for use in bone regeneration based on their physical and mechanical properties, and

bioactivity. In addition, biodegradable polymer coated Bioglass-based composite scaffolds

are attractive for controlled drug release applications, which is another aspect of this project.

Finally, in vitro cell culture study should be performed as reported in Chapter 8, in order to

evaluate the biological properties of the new scaffolds.

Page 85: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 5

Development of 45S5 Bioglass®-based Scaffolds for Controlled

Antibiotic Released in Bone Tissue Engineering via Biodegradable

Polymer Layered Coating

5.1 Introduction

Recent developments in bone tissue engineering provide alternative approaches for

the repair of bone defects caused by trauma and infection [93,283]. Bone repair scaffolds

associated with the application of drugs (i.e. antibiotics and antitumoral medicaments) and

growth factors attract increasing attention to avoid infections, regulate cell growth and

develop bone regeneration [268,284,285]. Basically, bone scaffolds should be

biocompatible, bioresorbable, osteoconductive and possibly they should act as a drug carrier

[283,285–289]. Scaffolds are usually made from tailored combinations of inorganic and

organic phases, which are chosen aiming at replicating the structure and composition of

bone tissue [283,289,290]. Bioactive glasses [95,286–288] and bioceramics [97,216,291–

293], as the inorganic component in composites satisfy the desirable property of bioactivity

for the application in bone tissue engineering approaches. Several natural- and synthetic-

derived biodegradable polymers have been explored as the organic component for

development of composite scaffolds. Specific studies on composite scaffolds with drug

delivery ability have been discussed in Chapter 2.

According to the study presented in Chapter 4, the tailored biodegradable polymer

coated Bioglass-based scaffolds were aimed to develop in as a multifunctional scaffold for

bone tissue engineering applications and were presented in this chapter. By combining the

advantages of synthetic and natural-derived polymers, two polymer coating layers were

applied onto the Bioglass-based scaffold with different purposes, including (i) to improve

the mechanical properties of Bioglass-based scaffolds by first coating with synthetic

PDLLA/P123 copolymer and (ii) at the same time to add extra-functionality on the scaffolds

Page 86: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

72

by second coating with tetracycline antibiotic-loaded natural polymers such as Alg and Gel.

PDLLA was chosen as a coating material on the scaffold due to its good mechanical

properties as shown in Chapter 4. However, PDLLA, which is a hydrophobic polyester, is not

chemically compatible with water soluble antibiotics, thus it exhibits a low drug entrapment

efficiency [294]. Therefore, natural-derived biodegradable polymers, including Alg and Gel,

were selected as the drug carriers. The polymer networks of both Alg and Gel are able to

load a wide range of bioactive substances, cells and drug molecules with minor interaction

between the drug and the polymer matrix [295]. In addition, natural polymers are superior

biocompatible, which facilitate the adhesion and proliferation of cells (i.e. osteoblasts) [289].

Based on the fact mentioned above, Alg and Gel were investigated as a drug carrier in the

present study. However, the layered coating cannot be formed as desired because of the

incompatibility of PDLLA, and Alg and Gel coating layers. Therefore, we aimed to overcome

this problem by modifying surface chemistry of PDLLA coating with an amphiphilic polymer

blending (i.e. P123 copolymer). Furthermore, the mechanical properties and drug release of

the multifunctional scaffold (Alg- vs. Gel-drug carriers) were investigated.

5.2 Materials and methods

5.2.1 Fabrication of TCH-loaded layered biodegradable polymer coated Bioglass®-

based scaffolds

45S5 Bioglass®-based scaffolds were prepared following the same procedure

detailed in Chapter 4 by using the foam replication method. For the preparation of polymer

coatings, PDLLA (Purac Biomaterials, Gorinchem, Netherland) was dissolved in DMC with a

concentration of 5 wt/v % at room temperature while stirred for 2 h. Then PEG-PPG-PEG

triblock copolymer (Pluronic P123, Mn 5800 Da; Sigma) was added into the PDLLA

solution with the PDLLA to P123 ratio of 90/10 wt %. The mixture was continuously stirred

until P123 was completely dissolved. Afterwards, a 45S5 Bioglass® scaffold was immersed

in the polymer solution during 5 min. Subsequently the scaffold was removed from the

solution and dried at room temperature for 24 h. An obtained coated scaffold was labeled as

PL/P123-c-BG.

Page 87: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

73

Tetracycline hydrochloride (TCH)-loaded Alg and Gel solutions were prepared as

follows: Na-Alg (Mw 200000 Da; Sigma) was dissolved in DI H2O with a concentration of

1.5 wt/v % at room temperature while stirred for 2 h. A concentration of 1.5 wt/v % Gel (type

A from porcine skin with 300 g bloom; Sigma) was dissolved in DI H2O at 50 C while stirring

for 1 h. Then 375 µg/ml TCH (C22H24N2O8 · HCl; Appli Chem GmbH, Darmstadt, Germany)

was added into both Alg and Gel solutions. Finally, the PL/P123-c-BG scaffold was

immersed in the TCH-loaded Alg and Gel solutions contained TCH (5 ml of solution/scaffold)

for 5 min and dried at room temperature for 24 h. The drug-loaded scaffolds were labeled as

T-Alg-c-(PL/P123-c-BG) and T-Gel-c-(PL/P123-c-BG) for Alg and Gel as the drug carriers,

respectively. TCH loaded uncoated scaffolds were prepared as control, by dipping uncoated

scaffolds in TCH/DI H2O with a concentration of 375 µg/ml for 5 min. Then the scaffold was

taken out and dried at room temperature for 24 h. These samples were labeled as T-BG.

5.2.2 Characterization and testing

(i) Capillarity test

In order to evaluate the porosity and surface property of the polymeric coatings, a

qualitative capillarity test was performed according to [296]. Briefly, the TCH-loaded

polymeric coating solution, which served as a testing fluid, was prepared as described in

section 5.2.1. In this case, TCH-loaded Alg solution, which exhibits a yellow color, was

added in a petridish. Then a coated scaffold was slowly placed on the surface of the

solution, while the testing time was recorded until the scaffold was completely wet (the fluid

went up through the entire porous network of the scaffold). PDLLA-c-BG and PL/P123-c-BG

scaffolds were tested in order to compare the surface property of the different polymeric

coatings.

(ii) Contact angle measurement

In order to evaluate the hydrophilicity of each polymeric coating, the wettability of

pellets prepared with the same conditions as the 3D scaffolds was measured using a water

contact angle instrument (DSA30, Kruess, Germany). The pellets were prepared as follows:

0.3 g of Bioglass® powder were added in a stainless steel die (diameter: 10 mm) and pellets

were obtained by cold uniaxial pressing using an electro-hydraulic press (MAUTHE

Page 88: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

74

MASCHINENBAU PE-010; Wesel, Germany) working at a load 4 × 104 N. The obtained

pellets were sintered using the same conditions used for porous Bioglass® scaffolds. As-

sintered Bioglass® pellets were coated with TCH drug, PL/P123, and TCH loaded Alg and

Gel following the same procedure described above for the scaffolds.

(iii) Microscopy

The microstructure of the scaffolds was characterized by SEM (LEO 435VP from

Zeiss Leica). The scaffolds were cross-sectioned by using a razor blade. The samples were

then sputter-coated and observed at an accelerating voltage of 10 kV.

(iv) Chemical analysis

The chemical structure of the scaffolds was investigated by using FTIR (Nicolet

6700). Bioglass®-based scaffolds were grinded and the obtained powder was mixed with

potassium bromide (KBr) powder in a weight ratio of 1/300 (scaffold/KBr). The mixture was

pressed into a pellet by using an electro-hydraulic press (MAUTHE MASCHINENBAU PE-

010; Wesel, Germany) at a load of 105 N. Then, pellets were measured by using FTIR in

transmission mode with the resolution of 4 cm-1

in the wavenumber range of 4000 - 400 cm-1

.

(v) Mechanical testing

Polymer coated cubic shaped Bioglass® scaffolds with dimensions of 8 mm × 8 mm

× 8 mm were tested under compression deformation mode by using a universal testing

apparatus (Zwick Z050). The cross-head speed used was 2 mm/min, the preload was 0.1 N

and the maximum load was 50 N. Stress-strain curves were recorded to determine the

mechanical properties. The elastic modulus was calculated from the initial linear slope of the

compressive stress-strain curves, while the compressive strength was obtained from the

maximum stress before the specimens collapse. Eight specimens were tested and the

results are presented as average ± SD.

(vi) In vitro drug release profile

The in vitro drug release behavior of the scaffolds, including T-BG, T-Alg-c-

(PL/P123-c-BG) and T-Gel-c-(PL/P123-c-BG) scaffolds, with dimensions of 8 mm × 8 mm ×

8 mm were evaluated. Each scaffold was immersed for up to 14 days in a glass vial

containing 5 ml of PBS (0.1 M; Sigma) solution at 37 C and pH 7.4. At given interval times,

Page 89: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

75

2 ml of PBS was taken and replaced with fresh PBS. The absorbance of the drug containing

PBS at the wavelength of 362 nm was measured by using a UV spectrophotometer

(Specord 40; Analytikjena, Germany). Then, the amount of drug released was determined by

using a linear relation between absorbance and known concentrations of TCH (2.5 - 100

µg/ml), as given in Eq. 5.1:

Absorbance = [0.0268 x concentration (µg/ml)] – 0.1206, R2 = 0.99. (5.1)

The amount of drug release is reported as a percentage of cumulative drug release ± SD

with respect to the immersion time.

At the same time, in order to confirm the bioactivity of scaffolds after coating with

two polymer layers, T-Alg- and T-Gel-c-(PL/P123-c-BG) scaffolds with dimensions of 8 × 8 ×

8 mm3 were investigated. Each scaffold was placed in a polystyrene bottle containing 50 ml

of SBF solution at 37 C and pH 7.4 [297]. After 14 days of immersion, the scaffold was

taken, washed twice with DI water and dried at room temperature. Afterwards, possible HA

formation and also the morphological changes were analyzed by using SEM.

5.2.3 Statistical analysis

The data were analyzed by using one-way ANOVA analysis and Turkey’s multiple-

comparison test to determine statistical differences. A confidence interval of 95 % (p = 0.05)

was used for all analysis.

5.3 Results and discussion

5.3.1 Surface property of polymeric coatings

A double coating layer based on PDLLA, and Alg and Gel was designed to be

applied on the 45S5 Bioglass®-based scaffolds to impart multifunctionality, e.g. drug delivery

capability. The challenge in developing such synthetic-natural polymer layered coatings is

the difference of surface chemistry between the hydrophobic PDLLA and hydrophilic Alg or

Gel limits their compatibility. As a result of polymer incompatibility, the TCH-loaded Alg and

Gel could not infiltrate the porous structure of unmodified PDLLA-c-BG scaffold. This

observation was qualitatively confirmed by the capillarity test [296] (Fig. 5.1), using the

TCH-loaded Alg solution (yellow color) as a test solution. The test involves the determination

Page 90: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

76

of porosity and capillarity of surface [296], and it revealed the lack of capillarity (wettability) in

the case of PDLLA-c-BG scaffolds. The scaffolds were seen to remain on the surface of the

coating solution (see Fig. 5.1 (Aa and Ba)). Consequently, the TCH-loaded coating solution

could not infiltrate the pore structure and the coating of the struts of PDLLA-c-BG scaffold

was not successful. In order to overcome this problem, a modification of the surface

chemistry of PDLLA-c-BG scaffold was necessary. The approach developed in this study

involved the addition of P123 copolymer in order to increase the hydrophilicity of the PDLLA-

c-BG scaffold, leading to hydrophilicity matching to that of Alg and Gel. The P123 copolymer

was utilized because it contains both hydrophilic (PEG) and hydrophobic (PPG) chains,

which can be homogeneously blended with PDLLA in DMC solution. By using this approach,

the capillarity was obvious in the case of PL/P123-c-BG scaffolds, showing wettability

increase and that the TCH-loaded coating solution ascended through the whole pore

network of the scaffold in few seconds (see Fig. 5.1 (Ab and Bb)).

The increase in the hydrophilicity of PDLLA-c-BG scaffolds was also confirmed by

the water contact angle values, as shown in Fig. 5.2. After coating with PDLLA/P123 blend,

the contact angle value of the scaffolds was significantly decreased (from 74.3 ± 0.2 for

pure PDLLA-c-BG to 31 ± 2 for PDLLA/P123-c-BG) to nearly the values of T-Alg and T-Gel

coatings (38.0 ± 0.5 and 39 ± 0 , respectively). These results prove that the blend of

PDLLA and P123 copolymer can drastically modify the surface chemistry of pure PDLLA

and thus the T-Alg and T-Gel solutions can efficiently infiltrate the porous structure of

Bioglass-based scaffolds, forming layered polymer coatings on the Bioglass-based

scaffolds, as desired in this study.

Page 91: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

77

Figure 5. 1 (A) Scheme of the capillarity test of Bioglass®-based scaffolds, showing the

effect of surface chemistry on the permeability of the porous scaffolds and (B) photographs

representing the coated scaffolds during the capillarity test.

Figure 5. 2 Contact angles of Bioglass®-based scaffolds showing the surface wettability of

different coatings. * indicates the significant difference (p 0.05) of the modified coatings on

the Bioglass-based scaffolds in comparison with PL-c-BG scaffolds.

Page 92: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

78

5.3.2 Morphology

Fig. 5.3 shows SEM micrographs at different magnifications, of different coated

scaffolds. The morphology of the scaffolds after coating with PDLLA/P123 blend is shown in

Fig. 5.3 (A and B). The surface of the coated scaffold was homogeneous and smooth

compared to the surface of T-BG scaffolds (Fig. 5.3 (B) and Fig. 5.3 (D)), which might be the

result of the used P123 copolymer, considering that P123 copolymer shows an ability to

enhance the rheological property of polymers [298]. It is thus obvious that the polymer

homogeneously covered the entire strut even though some uneven areas could be

observed, as shown in Fig. 5.3 (B). After coating with TCH-loaded Alg and Gel as second

coating layers, as shown in Fig. 5.3 (E and G), even though the color of the scaffolds

became yellow, no morphological changes of the struts were observed by SEM (both T-Alg-

and T-Gel-c-(PL/P123-c-BG) scaffolds) compared to the PL/P123-c-BG scaffolds (Fig. 5.3

(B)). In detail, a fairly homogeneous coating not inducing blocking of pores was observed

(see Fig. 5.3 (E and G)). However, at higher magnification (Fig. 5.3 (F and H)), the rougher

surfaces of the TCH-loaded polymer coatings could be observed in comparison with

PL/P123-c-BG scaffolds (Fig. 5.3 (B)). This observation was the same as reported in the

previous study of Mouriño et al. [299]. It was found that the surface of scaffolds became

rougher after coated with the second layer of Alg [299]. It is probably caused by the fact of

polymer agglomeration during drying. In detail, the alginate coating, for example, took longer

time to be dried compared to the PDLLA/P123 coating. By this fact, the Alg coating may tend

to gather together during drying. However, the obtained rough surface is believed to be

suitable for cell adhesion and proliferation.

Page 93: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

79

Figure 5. 3 SEM images of the scaffolds showing the morphological porous structure and

morphology of a coating surface of: (A, B) PL/P123-c-BG scaffolds, (C, D) T-BG scaffolds,

(E, F) T-Alg-c-(PL/P123-c-BG) scaffolds and (G, H) T-Gel-c-(PL/P123-c-BG) scaffolds.

5.3.3 Mechanical properties

The typical stress-strain curves in Fig. 5.4 (A) as well as the normalized compressive

strength and modulus in Fig. 5.4 (B) illustrate the improvement of the elastic modulus and

the compressive strength of Bioglass®-based scaffolds by layered polymeric coating. In

detail, the scaffolds coating with PDLLA/P123 (curve b in Fig. 5.4 (A)) exhibited an

improvement of the mechanical properties up to 10 times in comparison with the uncoated

scaffolds as depicted as curve (a) in Fig. 5.4 (A). This can be assigned to the formation of a

uniform PDLLA/P123 coating on the struts, as well as to the filling of any cracks on the

Page 94: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

80

surface of the struts, ceasing the crack propagation occurring on Bioglass®-based scaffolds

under load. It is likely that the mechanical strength of polymer coated Bioglass-based

scaffolds in the present study was higher, compared to the values reported in the previous

studies [57,282]. It can be described by the fact that the Bioglass-based scaffolds in the

study of Li et al. [57], for example, were partially coated with polymer, while the polymer was

fully covered the struts of Bioglass-based scaffolds in the present work. Moreover, the

second layer coated with the TCH-loaded Alg and Gel did not enhance further the

mechanical properties of the scaffold. As represented in the compressive stress-strain

curves of both T-Alg-c- and T-Gel-c-(PL/P123-c-BG) scaffolds (curves (c) and (d) in Fig. 5.4

(A)), that their stresses responsed to the strain in the similar trend to the curve of PL/P123-c-

BG scaffolds. The reason for this can be that only a thin layer of Alg and Gel is formed due

to the low polymer concentration used. According to these results, the mechanical properties

of the layered polymer coated Bioglass-based scaffolds predominantly occupied by the

coating layer of PDLLA/P123 blend are confirmed.

Figure 5. 4 Mechanical properties of polymer coated Bioglass-based scaffolds: (A)

representative compressive stress-strain curves and (B) average elastic modulus and

average compressive strength of the scaffolds. * (p 0.05) indicates the statistical

significance of compressive mechanical properties of coated scaffolds, compared to those of

uncoated T-BG scaffolds.

Page 95: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

81

5.3.4 Chemical structure

FTIR analysis was performed on coated scaffolds to confirm the existence of the

polymeric coating and the drug entrapment. First, the spectra of the Bioglass-based

scaffolds before and after drug loading without polymer carrier were considered (Fig. 5.5). In

detail, the spectrum of the T-BG scaffolds presents the characteristic peaks of Bioglass®,

including the double peaks at the wavenumber 1100 - 1040 cm-1

attributed to Si-O-Si

stretching mode [46,180], and the peak at the wavenumber of 458 cm-1

attributed to the Si-

O-Si bending mode [180]. The characteristic peaks of Bioglass were not changed after

loading with TCH, indicating that loaded TCH molecules do not initiate a chemical reaction

with Bioglass. This result means that TCH molecules loaded on the Bioglass scaffolds do

not lose their activity. Moreover, the double peaks at 3482 and 3350 cm-1

in the spectrum of

T-BG scaffold (see the inset (I) in Fig. 5.5), can be understood as an overlapping effect

between the –OH stretching broad peak of Bioglass (3700 - 3000 cm-1

), and double peaks

of TCH (3363 and 3304 cm-1

) and –CH stretching of phenol framework in TCH [287]. In

contrast, the spectrum of T-Alg-c-(PL/P123-c-BG) scaffold presents the peaks at 1620 and

1420 cm-1

assigned to –COO- asymmetric and symmetric stretching of Alg, respectively,

suggesting the existence of Alg in the coated scaffold. Moreover, the peak at 1753 cm-1

observed in the spectrum of T-Alg-c-(PL/P123-c-BG) scaffold is attributed to -C=O stretching

of PDLLA coating. As observed also in the spectrum of T-Gel-c-(PL/P123-c-BG) scaffold, the

-C=O stretching of PDLLA coating exists at the wavenumber of 1759 cm-1

. In addition, the

peak at 3435 cm-1

is assigned to –NH stretching of gelatin. Other detectable peaks at 1642

and 1456 cm-1

are assigned to –C=O and –C-N- stretching, and the peak at 1383 cm-1

is

assigned to –N-H- bending, as shown in the inset (II) in Fig. 5.5, which confirms the

existence of Gel coating. However, the characteristic peaks of TCH in the finger print region

(1500 - 500 cm-1

) are not obvious in the spectra of the coated scaffolds. Also, the change in

the peak position, in either Alg or Gel, is not represented in any spectra of the coated

scaffolds. Therefore, possible molecular interaction between the drug and the polymer

coating cannot be confirmed based on FTIR results.

Page 96: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

82

Figure 5. 5 FTIR spectra of TCH, BG, TCH-loaded Bioglass scaffolds and TCH-loaded

polymer coated Bioglass®-based scaffolds.

5.3.5 In vitro drug release

(i) Release profile

Fig. 5.6 (A) shows the cumulative percentage of TCH release from the Bioglass-

based scaffolds for up to 14 days of immersion in PBS. T-BG scaffolds showed an initial

burst release of 53 % at 1 h, which increased to 99 % in 4 h. This result confirms the low

drug binding affinity of the uncoated Bioglass-based scaffolds. In contrast, in polymer

coated scaffolds, lower initial burst release values (1 h) at 27 % and 22 % for Alg and Gel

coatings, respectively, were measured. Afterwards, both TCH-loaded polymer coated

scaffolds provided almost complete drug release ( 99 %) over 14 days in a sustained

behavior. This drug release kinetic is favorable as it should not only facilitate an effective

initial antibacterial effect but also promote long term protection against infection. Both Alg

and Gel carriers provided a similar release profile, including (i) an initial burst release as a

result of the release of the free drug presenting at the polymer coating surface and (ii) a

further relatively slow release induced by the drug molecules “protected” by the polymer

coating. As described in the literature [216,268,300], drug molecules embedded in the

Page 97: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

83

coating can diffuse through available pathways, i.e. pores and channels, into the medium.

Diffusion pathways can be influenced by the presence of an inhomogeneous coating

accompanied with the intrinsic degradation of the coating. Considering the result of the

degradation study (Fig. 5.6 (B)), Gel coated scaffolds were seen to exhibit slightly faster

degradation rate compared to Alg coated scaffolds. This result can be explained by the fact

that the Alg coating might partly crosslink with Ca ions in PBS [212], which should lead to

slower the degradation of the Alg coating. Also, the initial burst release of T-Alg-c-(PL/P123-

c-BG) scaffolds was slightly lower than that of T-Gel-c-(PL/P123-c-BG) scaffolds. In addition,

the possible interaction of the negatively charged drug (TCH) and the positively charged

polymer (Gel) was not observed in this study, and a superior binding affinity of TCH and Gel

cannot be confirmed by the results of drug release. Therefore, it seems that the key factors

influencing the release of drug from the natural polymer coated Bioglass-based scaffolds

are mainly related to coating homogeneity and dissolution/degradation of the polymer

coating. Compared to previous recent studies [57,282], the initial burst release of the TCH-

loaded Alg coating (22 %) in the present study was significantly lower, e.g. it was 63 % in

vancomycin-loaded CS coating [282] and 33 % in vancomycin-loaded PHBV [57]. However,

the different drug used in the present study should be taken into consideration. On the other

hand, TCH-loaded PCL/HA coated HA scaffolds have released, at the initial stage (1 h), 44

% of the load [291], indicating that the natural coating developed in the present study

enables better release control, reducing the initial burst release. The present approach

using natural polymers such as Alg and Gel as drug carrier seems to lead to superior

performance of the scaffold as drug delivery device in terms of water soluble drug

entrapment and protection of the drug, in comparison with systems based on synthetic

polymers. Another important feature of the present approach is that it is possible to de-

couple the mechanical stability function (provided by the synthetic polymer) from the drug

release function (provided by the natural polymer).

Page 98: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

84

Figure 5. 6 (A) Drug release profile and (B) degradation behavior of TCH-loaded polymer

(Alg and Gel) coated Bioglass scaffolds.

(ii) Morphology after drug release

SEM analysis was used to observe the morphological change of the scaffolds after

14 days of drug release (Fig. 5.7). The polymeric coating was partly maintained on both

scaffolds. In Fig. 5.7 (A)), it can be observed that the surface of the drug-loaded Alg coated

scaffold became rougher by the creation of many holes and channels, which indicates the

degradation of polymeric coating. The appearance of underneath smoother surface was also

observed, which probably is the PDLLA/P123 coating, as depicted by the arrow in Fig. 5.7

(A). It is also possible that the PDLLA/P123 coating started to be eroded during the

degradation of the outer drug-loaded alginate coating. These holes serve as a connection

pathway between the PDLLA/P123 coating and the liquid medium (PBS). The SEM image of

the T-Gel-c-(PL/P123-c-BG) scaffold in Fig. 5.7 (B) shows the generation of holes on the

coating surface, as shown previously on the surface of the T-Alg-c-(PL/P123-c-BG) coating

(Fig. 5.7 (A)). In contrast, the residual polymeric coating (depicted by a solid arrow) was the

PDLLA/P123 layer, while the outer drug-loaded Gel coating could not be distinguished. It is

likely that the Gel coated has degraded after the 14 d immersion in PBS. In addition, the

PDLLA/P123 coating was partly degraded and the surface of the strut could be also

observed (dashed arrow in Fig. 5.7 (B)). The release of TCH-loaded Alg and Gel carriers

Page 99: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

85

predominantly influenced by the degradation of the polymer coating is thus confirmed.

Moreover, it seems that the bioactivity of Bioglass was not inhibited by the layered polymer

coating, since amorphous PDLLA coating used exhibited partly degradation during

experimental study in PBS, as confirmed by the holes observed in Fig. 5.7 (A and B). By this

fact, dissolution of Bioglass can take place and this phenomenon leads to formation of HA

in SBF. As shown in Fig. 5.7 (C and D), formation of possible apatite crystals was observed

on the strut of the both scaffolds after 14 days of immersion in SBF. In particular, it was

obvious in the case of T-Gel-c-(PL/P123-c-BG) scaffold (Fig. 5.7 (D)) that the crystals

covered the strut of the scaffold, while the degradation of polymer coating took place (as

depicted by a solid arrow).

Figure 5. 7 SEM images of TCH-loaded polymer coated Bioglass-based scaffolds after in

vitro release in PBS for 14 days: (A) T-Alg-c-(PL/P123-c-BG); the arrows predicting the

PL/P123 coating and (B) T-Gel-c-(PL/P123-c-BG) scaffolds; dashed arrow depicting the

Bioglass® struts and solid arrow line predicting PL/P123 coating, and SEM images of TCH-

loaded polymer coated Bioglass-based scaffolds after immersion in SBS for 14 days: (C)

T-Alg-c-(PL/P123-c-BG and (D) T-Gel-c-(PL/P123-c-BG) scaffolds; solid arrows depicting

polymer coating.

Page 100: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

86

(iii) Chemical structure after drug release

The FTIR spectra shown in Fig. 5.8 enable to detect the chemical changes of the

scaffolds after 14 days of immersion in PBS. First, a new peak at wavenumber 1475 cm-1

is

observed as a broad double peak close to the peak of -COO- of Alg at 1420 cm-1

, in the

spectra of T-Alg-c-(PL/P123-c-BG) scaffold after immersion (see Fig. 5.8 (A)). The broad

double peak is assigned the overlapping of –COO- stretching band of Alg with –CH2-

bending band of PDLLA [301]. Moreover, the peak at 1620 cm-1

, ascribed –COO- stretching

in Alg, is reduced in intensity, indicating that the Alg content is reduced after immersion.

Moreover, the peak is seen to be shifted to 1643 cm-1

, which is possibly due to protonation

of carboxylate groups [295]. Therefore, these results confirm that Alg coating remains on the

scaffold after 14 days of immersion in PBS. The FTIR spectra of the T-Gel-c-(PL/P123-c-

BG) scaffold are reported in Fig. 5.8 (B). The double peak at wavenumber 1479 and 1424

cm-1

appears after immersion. Similarly to the T-Alg-c-(PL/P123-c-BG) coated scaffold

discussed above. The absorption bands corresponding to PDLLA are stronger detectable in

the spectrum of the scaffold after immersion, which is the result of degradation of the Gel

coating. The degradation of the Gel coating is evidenced by the broader peak of –NH

stretching (3700 - 3000 cm-1

).

Figure 5. 8 FTIR spectra of T-Alg- and T-Gel-c-(PL/P123-c-BG) scaffolds after 14 days of

immersion in PBS.

Page 101: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

87

5.4 Conclusions

Multifunctional layered polymer coated Bioglass-based scaffolds with drug delivery

capability were fabricated by coating Bioglass foams with two different polymer coatings,

including PDLLA/P123 blend and Alg or Gel. The scaffolds exhibited improved mechanical

properties and superior drug delivery function characterized by a low initial burst release and

subsequent controlled drug release. In addition, both Alg and Gel served as a drug carrier

and they did not show significantly different performances in their degradation and release

behaviors. The multifunctional scaffolds fabricated, exhibiting improved mechanical

properties and controlled drug release coupled with high bioactivity characteristic of

Bioglass, belong to a growing family of advanced scaffolds for bone tissue engineering.

Page 102: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

88

Page 103: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 6

Porous Biodegradable Polymer-based Scaffolds for Cartilage Tissue

Engineering Applications

6.1 Introduction

OA affects the musculoskeletal system, especially the joints in the hip, hand, knee

and spin cord [302]. At the joints, the cartilage tissue covering the end of long bones inhibits

direct rubbing of bones against each other [23,302] and it allows the joint to work smoothly

and without causing pain [302]. When OA occurs, the cartilage, which works as a shock

absorber, becomes worn off in some areas, finally leading to loss of elasticity of the cartilage

tissue [125]. In this case, the bones may rub against each other causing very severe pain

[303]. Due to its complex and unique structure, and exposure to high pressure and motion,

the repair of cartilage is one of the most challenging areas in tissue engineering

[23,241,304]. Moreover, the avascular and aneuronal nature of cartilage limits the ability to

deliver signal molecules, growth factors or cellular components for the tissue repair process

[232,241,304]. In order to repair osteochondral defects, a cell-scaffold-based repair strategy

is currently being favored instead of using arthroscopic debridement, microfracture, autograft

and autologous implantation. The reason is that cell-scaffold-based implantation exhibits the

potential to regenerate hyaline cartilage without creating defects at other sites of the joint

[23,241,305]. Therefore, the goal of this approach is to engineer a durable cartilage tissue

that provides a smooth joint surface reconstruction and is resistant to high loads, shear

stresses, combined with good friction properties. Besides the cell type and source that

certainly affect the outcome of the procedure [305], both the scaffold composition and

architecture are relevant for the repair of osteochondral defects. Furthermore, growth factors

and/or cytokine have to be integrated in order to support cell differentiation [24,305]. For

osteochondral defect regeneration, besides the required properties of scaffolds for cartilage,

which were previously mentioned in Chapter 2, cartilage scaffolds must be well connected

Page 104: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

90

with the underlying SB or bone engineered scaffold, in order to enhance in-situ integration of

the osteochondral system [125].

This study is focused on the preparation of distinct layered scaffolds or multilayered

scaffolds, based on bioactive glasses and biodegradable polymers. For the cartilage side of

the scaffold, which will be the focus of this chapter, the elastic properties are a crucial

requirement for mimicking the natural cartilage tissue. Therefore, soft polymeric materials

are materials of choice for cartilage scaffolds in comparison with rigid bioceramics and

metals. In addition, the physical properties (such as structural and architectural) of scaffolds

are crucial for cell behavior. It is well established that a suitable scaffold should exhibit a

similar structure to that of the natural ECM of the tissue to be regenerated, which provides a

suitable microenvironment for cell adhesion, proliferation and differentiation. Based on the

required properties above, Alg was chosen as a scaffold material for cartilage in this study

due to its interesting properties, as mentioned in Chapter 2. In addition, the porous structure

and pore geometry of the Alg-based scaffolds can be tailored by adjusting the parameters of

the fabrication techniques used. Two different fabrication techniques, including freeze-drying

and electrospinning, are focused in the present work. Both techniques provide the ability to

create scaffolds with tailored porous structure, as discussed in Chapter 2.

In the scope of this chapter, Alg was initially fabricated as 3D porous foam by a

combination of gelation and freeze-drying techniques. In order to get more insights on the

scaffold characteristics and the mechanism of the used technique, parameters such as the

polymer and crosslinking agent concentrations are investigated. The first aim is to optimize

processing conditions, so that the fabricated scaffold can achieve the basic requirements as

a suitable cartilage scaffold. The optimized Alg scaffold was further modified in order to

enhance the cell adhesion, proliferation and differentiation based on in vitro chondrocytes

and stem cells culture.

In another approach, fabrication of Alg fibers by using electrospinning technique was

attempted. Few successful studies of electrospun Alg [212,306–308] have been reported

and the electrospinning of Alg aqueous solution is still challenging. Blending solutions of

Alg/PEO [212,306], Alg/PVA [307] and Alg/Gel [308] have been fabricated into fibers, while

Page 105: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

91

individual Alg aqueous solution cannot be used to fabricate fibers by electrospinning

technique due to several reasons. First, Alg is not able to be dissolved in organic solvents,

leading to the lack of conductivity of the solution. Second, the Alg/H2O solution is not able to

be stretched under electrostatic field during electrospinning, due to its weak intermolecular

interaction and subsequently insufficient mechanical properties [173,309,310]. Under these

circumstances, the influence of electrospinning processing conditions and the fundamental

mechanism of the electrospinning technique were investigated in detail. Therefore, a

synthetic biodegradable polymer (poly(L-lactide) (PLLA)) and natural-derived polymer (Alg)

were electrospun and comparatively investigated regarding the basic desired properties for

cartilage regeneration.

Finally, the comparison between the obtained scaffolds is discussed in terms of their

microstructure, physical and mechanical properties. The possible advantages and

disadvantages of the different scaffold types were investigated, in order to seek a suitable

scaffold to be developed further for cartilage regeneration.

6.2 Materials and methods

6.2.1 Fabrication of Alg-foams

Na-Alg (Sigma, = 1.02 g/cm3) with a concentration of 4 wt/v % was prepared in DI

H2O and stirred at room temperature for 2 h. The Na-Alg solution was diluted to 2 and 3 wt/v

%, respectively. 1 ml of each of these Alg solutions was added in a 48 well-plate, followed by

the addition of 100 µl of 0.1, 0.5 and 1 M CaCl22H2O per well, respectively. The mixture was

kept at room temperature for gelation as maximum as 30 min. As next, the Alg-gel was

frozen at - 20 C for 24 h. Then frozen samples were sublimated and a porous structure was

generated after 24 h lyophilizing at temperature of - 50 C under vacuum conditions. As a

result cylindrical 3D porous Alg sponges were obtained, having a diameter of 8 mm and a

height of 8 mm. Finally, for ionic crosslinking, the Alg sponges were immersed in 0.5 M

CaCl22H2O for 4 h and dried at room temperature for at least 24 h.

Page 106: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

92

6.2.2 Fabrication of PLLA fibers and Alg-based fibers

Poly (L-lactide) (PLLA) was dissolved in DCM/methanol (80/20 vol %) with

concentrations of 5 and 7.5 wt/v % at room temperature for 2 h. A PLLA solution was

fabricated as a fibrous mesh by using electrospinning technique. The processing parameters

consisted of 8.5 - 20 kV of applied voltage, 0.5 - 2 ml/h of flow rates, 15 - 18 cm of tip-

collector distances and 2 h of deposition time, as detailed in Table 6.1.

Na-Alg was blended with PVA, P123 and gelatin in DI H2O for different weight ratios

and solvent systems, as detailed in table 6.2. The mixture solution of Alg/Gel blend was kept

at 50 C in order to avoid gel-formation of Gel. During the electrospinning process, hot air

conditions (i.e. a heat gun) were applied to the set-up in order to maintain a constant

temperature of the process at 50 C. The variety of electrospinning processing conditions of

Alg-based solutions is detailed in Table 6.2.

Electrospun Alg/Gel fibers were chemically crosslinked by using glutaraldehyde

(GA) exposure. The GA crosslinking agent was prepared at concentration of 30 vol % in

H2O/EtOH (70/30 vol %) and was transferred to a 1 l glass beaker. The fiber mesh was

attached on the wall of beaker by using double-side adhesive tape. The beaker was

suddenly wrapped with aluminium foil tightly and was kept at room temperature for 48 h for

crosslinking with GA vapor.

6.2.3 Characterization and testing

(i) Porosity and density

The porosity (P) of Alg-foams was calculated by using Eq. 6.1, while the density ()

of the foams was determined by using Eq. 6.2:

P = 1 – [Wfoam/(Alg x Vfoam)] (6.1)

(g/cm3) = Wfoam/Vfoam (6.2)

; where Wfoam is the weight of the Alg freeze-dried foam, Alg is the density of solid Alg (1.02

g/cm3), and Vfoam is the volume of Alg freeze-dried foam, determined from the dimensions of

the foam.

Page 107: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

93

(ii) Microscopy

The cross-section appearance of scaffolds was investigated by using light

microscopy (light microscope Axioplan Zeiss), equipped with a digital camera system and

operated with the software LEICA DFC290.

The microstructure of the freeze-dried foams and electrospun fibers was

characterized by SEM. The SEM (LEO 435 VP) was operated with a tungsten filament, at an

acceleration voltage of 10 kV. Cross-section and plan-view SEM imaging was performed in

order to observe the pore structure. For SEM investigations the samples were sputter-

coated. The pore size of the foams and the diameter of fibers were evaluated from SEM

images using the free available software Image J (version 1.42S).

(iii) Chemical analysis

The chemical structure of the foams and fibers was investigated by using attenuated

total reflectance-FTIR (ATR-FTIR) in transmission mode with the resolution of 4 cm-1

, the

collected signals in 32 scans and the spectrum range of 4000 - 525 cm-1

.

(iv) XRD analysis

The scaffolds were characterized by using XRD (Siemens D500) analysis in order to

assess the crystallinity by employing Cu k radiation. Data were collected over the range of

2 = 10 - 70 using a step size of 0.02 and a counting time of 25 sec per step.

(v) Thermal analysis

Thermal transition features, including glass transition temperature (Tg) and melting

temperature (Tm) of the scaffolds were measured by using differential scanning calorimetry

(DSC; DSC Q 2000). Measurements were performed in the temperature range of (-50) - 200

C with heating rate of 10 C/min and double runs. Tg and Tm were determined by the

minimum value of the first and second endothermic transition peaks, respectively, while cold

crystallization temperature (Tc) was determined by the maximum value of the exothermic

transition peak. The percent crystallinity (Xc) was determined by using Eq. 6.3 [311,312]:

% Xc = [(∆Hm - ∆Hc)/ ∆Hm] × 100 (6.3)

Page 108: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

94

; where ∆Hm is the heat of melting (J/g), ∆Hc is the heat of cold crystallization (J/g)

and ∆Hm is a reference value, representing the heat of melting of 100 % crystalline

polymer (∆Hm of PLLA 93 J/g [312]).

(vi) In vitro biodegradation study

The water absorption of alginate freeze-dried scaffolds was investigated by

immersion in PBS (pH 7.4) for 45 min, 2 h, until 4 months, respectively. Each sample was

placed in a polystyrene bottle containing 50 ml of the PBS medium and incubated in an

orbital shaker at 37 C and 90 rpm. The medium was replaced two times per week. At

interval time, the sample was removed and blotted with filtered paper before weighed. The

obtained weight is wet weight (Wwet). The water absorption at each time point was

determined by using Eq. 6.4 [295]:

Water absorption (%) = [(Wwet – Wdry)/Wwet] x 100 (6.4)

; Wwet is the wet weight (g) of sample after soaking in medium and Wdry is the weight (g) of

the sample before soaking in the medium, in the following referred as the dry weight. Four

samples were investigated and the results are reported as average values ± SD.

The degradation profile and structural stability of the crosslinked Alg-foams were

evaluated by investigating the weight change of the scaffolds in different media, including DI

H2O, PBS and SBF with respect to the immersion times. These experiments were carried

simultaneously to the water absorption experiments. The weight change at each time point

and media was determined by using Eq. 6.5 [219]:

Weight change (%) = [(Wwet – Wdry)/Wdry] × 100 (6.5)

(vii) Mechanical testing

Alg-foams (dimensions of 8 mm in diameter and 8 mm in height) were compressed

by using a universal testing machine (Zwick Z050) for mechanical testing. Thereby, the

cross-head speed during compression testing was 2 mm/min. The samples were pre-loaded

at 0.1 N, while a maximum load of 50 N was used. A stress-strain curve was used to

determine the mechanical properties. The elastic modulus was calculated from the initial

slope of the elastic regime of the stress-strain curves, while the compressive strength was

Page 109: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

95

obtained from the maximum stress before the samples collapse. The results are averaged

from eight specimens that were tested for each condition.

In addition, dynamic mechanical analysis (DMA; Mark IV) was performed in order to

study the mechanical response of the foams under the replication of real forces, which are

expected to dominate in vivo [313]. Alg-foams (dimensions of 8 mm in diameter and 3 mm in

height) were tested in both dry and wet states, under a sinusoidal load with a rate defined by

a frequency (in Hz). Before testing in the wet state, the foams were immersed in PBS

solution until reaching equilibrium conditions. The measurement was performed in

compression mode of deformation, with the cycle frequency varied in the range 0.1 - 10 Hz

(according to the range of typical skeletal movement in vivo [313]) and with the maximum

strain amplitude of 1 %. The viscoelastic properties of the foams are represented as storage

modulus (E’) and loss factor (tan ) as a function of frequency.

The fibrous meshes in the form of thin films (5 mm in width, 30 mm in length and

500 µm in thickness) were tested under tension mode by using a universal testing machine

(Frank) by applying a force at a cross-head speed of 1 mm/min, preload at 0.1 N and

maximum load at 50 N. A tensile stress-strain curve was recorded to determine mechanical

properties. The Young’s modulus was calculated from the initial linear slope of stress-strain

curves and the tensile strength was obtained from the maximum value of stress before

fracture. Cast films fabricated by using the same solution employed for the electrospun

solution were also investigated as control. Eight specimens were tested and the results were

presented as average ± SD.

6.2.4 Statistical analysis

The data were analyzed by using one-way ANOVA analysis and Turkey’s multiple-

comparison test to determine statistical differences. A confidence interval of 95 % (p = 0.05)

was used for all analysis.

Page 110: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

96

6.3 Results and discussion

6.3.1. Effect of processing conditions on the physical and mechanical properties of

the foams

(i) Microstructure

3D porous Alg-foams were fabricated by means of gelation and freeze-drying

technique. Since Na-Alg is naturally water soluble, the structural stability of Alg-foams

cannot be maintained in contact with biological environments. Therefore, it is necessary to

transform soluble Na-Alg into an insoluble Ca-Alg configuration, which can be achieved by

ionic crosslinking, as described in Chapter 2. This crosslinking results in the formation of

Alg-gel, which is water resistant. The slow diffusion of Ca ions containing solution from the

superficial surface toward the bottom of the bulk solution led to the generation of tubular-like

structures along the gel [314], as observed in Fig. 6.1 (B)). A plan-view image in Fig. 6.1 (A)

shows circular-shaped structures spreading around the surface area, whose formation is

attributed to the addition of the CaCl22H2O agent [314]. After freezing the gel, the

propagation of a planar ice front occurred dependent on freezing rate. During lyophilizing,

the ice was sublimated and this process induced a horizontal flat ladder feature, as observed

by plan-view SEM imaging (Fig. 6.1 (C)) and tubular-like structures, as shown in the cross-

section SEM image in Fig. 6.1 (D). The structure of the foams is suitable to match the

required pore structure of scaffolds in order to support cell organization, as observed in

natural cartilage tissue, in which chondrocytes align in a columnar manner [24,125,128,129].

Page 111: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

97

Figure 6. 1 Optical microscopy images showing 3 wt/v % Alg-gel after the introduction of 0.1

M CaCl22H2O into Alg solution: (A) superficial surface (plan-view image) and (B) cross-

section of Alg-gel. SEM images of 3 wt/v % Alg-foam after lyophilized: (C) plan-view and (D)

cross-section SEM image (Reproduced from Nooeaid et al. [145] with the permission of

John Wiley and Sons).

(ii) Pore geometry and structural stability

The pore geometry and the mechanical properties of Alg-foams depend strongly on

concentration of both Alg solution and crosslinking agent used. Alg solution at

concentrations of 2, 3 and 4 wt/v %, and CaCl22H2O at concentrations of 0.1, 0.5 and 1 M

were varied in order to find out optimum conditions, which are used to fabricate the foams

suitable for the application of cartilage regeneration. Therefore, porosity, pore size, water

absorption, physical and mechanical properties, and reproducible efficiency of obtained

forms were investigated.

First, the influence of crosslinking agent concentration on gel-formation and porous

structure of the foams was studied. It was found that at given concentrations of 0.5 and 1 M

CaCl22H2O fast gel-formation (in few min) occurred for each Alg concentration. This

phenomenon led to a hard and inhomogeneous gel. As shown in Fig. 6.2, uncontrolled

shape was obtained after lyophilization. In contrast to this, a lower concentration of

Page 112: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

98

crosslinking agent (0.1 M CaCl22H2O) slowed down the gelation rate and consequently

homogeneous foam were formed exhibiting flexibility and softness when they were handled

with fingers. This result is in an agreement with the previous study that adding 0.075 - 0.1 M

CaCl22H2O into 2 wt/v % Alg solution resulted in a mechanically more stable gels, in

comparison to the one fabricated by using higher concentration of CaCl22H2O [127].

Moreover, it has been shown that the gel, obtained by using lower CaCl22H2O

concentration, remained cohesive upon handling. This result could be due to the fact that

low amount of Ca ions diffuse slowly through the Alg solution and they induce gel formation

at the same time. This phenomenon is known as ionotropic gelation [314]. As a result, the

solution has enough time to form a gel homogeneously. Therefore, 0.1 M CaCl22H2O was

chosen as the crosslinking agent for the gelation of Alg in the present study.

However, obtained Alg-foams did not exhibit sufficient mechanical stability and did

not last for even 1 h long in aqueous solutions. This is supported by the result of weight loss

of Alg-foams with respect to immersion time in DI H2O (at 37 °C) (Fig. 6.3). The foams were

completely decomposed in DI H2O in only 45 min, indicating their insufficient time-dependent

mechanical stability. In order to overcome this limitation, physical crosslinking was applied

by soaking the foams in 0.5 M CaCl22H2O for 4 h. As a result, the foams after crosslinking

lasted longer in DI H2O (i.e. up to 50 % weight loss was observed after 4 weeks of

immersion), indicating their suitability for the intended application in osteochondral tissue

engineering.

Regarding the influence of Alg concentration on foam structure, it was found that

increased Alg concentration led to increased density and relatively reduced porosity of the

foams. The porosity decreased from 97 % down to 93 % with increased concentration from

2 to 4 wt/v % (Fig. 6.4).

Page 113: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

99

Figure 6. 2 Optical photograph showing the appearance of 3 wt/v% Alg-foams obtained with

variation of CaCl22H2O concentrations (0.1 – 1 M).

Figure 6. 3 Comparison of weight loss as a function of immersion time in DI H2O of Alg-

foams without (w/o) and with (w) crosslinking by immersion in 0.5 M CaCl22H2O for 4 h (the

inset represents the weight loss of Alg-foams without crosslinking) (Reproduced from

Nooeaid et al. [145] with the permission of John Wiley and Sons).

Page 114: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

100

Figure 6. 4 The effect of Alg concentrations on the porosity and density of Alg-foams.

SEM images in Fig. 6.5 show the comparison between the microstructure of Alg-

foams fabricated from using different Alg concentrations. 2 and 3 wt/v % Alg-foams exhibited

uniform porous structure with open pores (Fig. 6.5 (A and B) and Fig. 6.5 (D and E)), while

non-uniform porous structure with some closed pores were observed in the case of 4 wt/v %

Alg-foams (Fig. 6.5 (G and H)). Closed pores are not desirable because they inhibit

migration of cells and further limit cellular activity. Moreover, the pore size of foams is an

important factor for cell migration. Appropriate pore size for supporting chondrocytes and

stem cells has been reported in the range of 50 - 300 μm [137,315,316]. In the present work,

Alg-foams, which were fabricated by using different concentrations, exhibited pore sizes in a

wide range of 30 - 500 μm, as shown in Fig. 6.5 (C, F and I). This pore size range overlaps

the required pore size for a suitable cartilage scaffold. In addition, the increase of Alg

concentration resulted in the reduction of pore size. For instance, 4 wt/v % Alg-foams

exhibited pore size in the range of 30 - 325 µm (average pore size 180 ± 47 µm), while 3

wt/v % Alg-foams showed pore size in the range of 125 - 325 µm (average pore size 237 ±

48 µm). In the case of 2 wt/v % Alg-foams, a wide pore size in the range of 150 - 500 µm

(average pore size 305 ± 55 µm) was observed. Thus concerning foam pore size, 3 wt/v %

Page 115: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

101

Alg-foams delivered the closest value to the target pore size, which is desired for suitable

cartilage scaffolds.

Figure 6. 5 SEM images (in plan-view) of (A, B) 2 wt/v %, (D, E) 3 wt/v % and (G, H) 4 wt/v

% Alg-foams, included distribution of pore size: (C) 2 wt/v %, (F) 3 wt/v % and (I) 4 wt/v %

Alg-foams.

(iii) Mechanical properties

In addition to the pore configuration, the mechanical properties of the foams are one

of the important factors, which is necessary to be considered for use in load bearing

applications. The Alg concentration used provides the main impact on the mechanical

properties of the obtained foams. As shown in Fig. 6.4, the increase of the Alg concentration

leads to increased density of Alg-foams, which results in the enhancement of the mechanical

Page 116: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

102

properties under compression load, as shown in Fig. 6.6 (A). In the representative

compressive stress-strain curves of 2, 3 and 4 wt/v % Alg-foams (Fig. 6.6 (A)), all samples

present the same trend, which is defined by sigmoidal shape (a characteristic of polymeric

cellular solids [317]). In addition, the stress-strain curves of 4 wt/v % foam shows drop

features, attributed to the occurrence of inhomogeneous deformation of the foam under

compression load. In contrast, 2 and 3 wt/v % foams exhibited smooth curves, indicating

more homogeneous deformation during compression. The shape of these curves manifests

three deformation mechanisms, including: (I) linear elastic region, in which the foam shows

elasticity and the instant rise of stress represents the deformation of the intact foam; (II)

brittle crushing region, which represents progressive rupture and collapse of the cell walls;

(III) densification region, which shows the densification of the compressed specimen

[317,318]. Elastic modulus was 0.220 ± 0.009 MPa in the case of 3 wt/v % Alg-foams,

while 4 wt/v % Alg-foams exhibited the modulus in the similar range ( 0.22 ± 0.05 MPa)

(Fig. 6.6 (B)). Importantly, 4 wt/v % Alg-foams did not provide an improvement of elastic

modulus. This is probably related to the formation of a non-uniform foam when the

concentration reached 4 wt/v %. The compressive strengths of 2, 3 and 4 wt/v % Alg-foams

were 0.047 ± 0.004, 0.14 ± 0.02 and 0.15 ± 0.02 MPa, respectively. 3 wt/v % Alg-foams

exhibited significant improvement of the elastic modulus and compressive strength

compared to 2 wt/v % Alg-foams, while 4 wt/v % Alg-foams did not lead to further

improvement of the mechanical properties (see Fig. 6.6 (B)). This result suggests that at

higher concentration (above 3 wt/v %) a critical viscosity is reached, which inhibits the

diffusion of Ca ions through Alg solution during the gelation process, leading to formation of

an inhomogeneous gel. Subsequently, an inhomogeneous foam is obtained after

lyophilization, which negatively influences the mechanical properties.

Page 117: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

103

Figure 6. 6 (A) Representative compressive stress-strain curves of 2, 3 and 4 wt/v % Alg-

foams and (B) the mechanical properties, including elastic modulus2 and compressive

strength3 of the foams as a function of concentrations.

(iv) Viscoelastic properties

Since cartilage is exposed to dynamic compression forces, DMA was used to

investigate the viscoelastic properties of the foams in both, dry and wet state. In particular,

the tests of wet foams (by immersion in PBS solution) were performed with the aim to

replicate the realistic conditions in the joint [313]. Fig. 6.7 shows the storage modulus (E’) as

a function of frequency (Hz) in the dry and wet state. In detail, the values of E’ of dry foams

were higher than those of wet foams. This result suggests that water molecules absorbed in

the foam act as a plasticizer in the polymer network, leading to molecular mobility and

subsequently reduced stiffness. E’ values of wet foams ( 0.5 MPa; 0.01 - 10 Hz) were

exhibited in the range of reference values of native cartilage (0.01 - 1.5 MPa; 0.01 - 10 Hz

[313]). In addition, tan values of wet foams were higher than those of dry foams, which is

suggested by the contribution of friction between water molecules and scaffold [319]. Finally,

E’ and tan values of both, dry and wet foams, remained constant along the given

2 Modulus of elasticity was determined at the region of elastic deformation in the stress-strain curve, according to

Hooke’s law ( = E) [383]. 3 Compressive strength was obtained at the yield point, where the onset of plastic deformation takes place.

Therefore, the represented compressive strength in this study is equivalent to yield strength [383].

Page 118: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

104

frequency range. This result indicates that the viscoelastic properties of the foams are in the

steady state within the measured time scale.

Figure 6. 7 Dynamic mechanical properties of 3 wt/v % Alg-foams in compression mode

presenting the storage modulus (E’) and the loss factor (tan ) as a function of frequency, in

both dry and wet state.

(v) Thermal properties

Effect of crosslinking on the thermal transition of 3 wt/v % Alg-foams was evidenced

by DSC results in Fig. 6.8. The thermal transition of both Alg-foams with and without

crosslinking shows only one significant endothermic peak detected from the first heating

ramp (Fig. 6.8 (A)). The peak minimum refers to the melting temperature (Tm) [320]. Tm of

Alg-foams increased from 78.1 °C to 96.6 °C with crosslinking. This result indicates that the

crosslinking enhances intermolecular interaction in Alg network. The increase of enthalpy

(∆Hm) required to break the interaction and to melt the foam was observed after crosslinking.

In detail, ∆Hm required for melting Alg-foam with crosslinking was 560.5 J/g, while ∆Hm

required for melting the foam without crosslinking was 495.5 J/g. In addition, another

important feature obtained from the DSC thermogram in Fig. 6.8 (B) is the glass transition

temperature (Tg), which primarily refers to the softness and flexibility of the polymer chains

Page 119: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

105

[317]. Tg is determined from the second heating ramp in an endothermic transition. It was

found that the introduction of crosslinking did not play a main role in Tg, as shown in Fig. 6.8

(B). Moreover, the detectable Tg of both foams with and without crosslinking at around 23 -

24 °C indicates that the foams might be softer at 37 °C, either in the in vitro or in vivo culture

conditions. However, it can be confirmed that the foams are not decomposed at that

temperature.

Figure 6. 8 DSC thermogram: (A) the 1st heating and (B) the 2

nd heating cycle runs of Alg-

foams with and without crosslinking.

(vi) Swelling and degradation profile

In general, natural cartilage resides in contact with synovial fluid within the synovial

membrane, and the high water content in cartilage provides sufficient softness and lowers

the friction during joint motions [16,19,24]. It has been reported that natural articular cartilage

consists of 80 % water with respect to the wet weight [6]. In the present work, water

absorption of Alg-foams reached a value of 82 ± 2 % (determined after 4 h of immersion in

PBS).

Apart from the ability to absorb water, the structural stability of scaffolds must be

maintained in physiological conditions until the regeneration of new tissue takes place. With

this aim, weight change of the foams in aqueous media (i.e. DI H2O, PBS and SBF, all at 37

Page 120: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

106

°C) with respect to immersion time is presented in Fig. 6.9 (A). At the initial stage ( 4 h), the

weight change of the foams in each medium was drastically increased with increasing

immersion time. This is justified by the fact that Alg-foams tend to absorb water (free and

bulk water) to fill the void region inside the polymer network until reaching the equilibrium

state [295,321–323]. This phenomenon follows the mechanism of hydration initiated by the

hydrophilic groups of Alg [295]. In this case, the foams reached equilibrium state at week 2,

in which water cannot be further gained by the foams. Afterwards, the weight change

remained constant until the end of the experiment (4 months of immersion time). This result

indicates that neither decomposition nor degradation of the foams immersed in DI H2O has

taken place yet. The structural stability of Ca-Alg network can be maintained until the

osmotic pressure overcomes the force of ionic-crosslinking interaction.

In contrast, the weight change of Alg-foams during immersion in PBS is mainly

driven by Ca2+

-Na+ exchange (Ca ions interacting with Alg chains and Na ions in the

medium) [295,324]. The initial phenomenon in alkaline media such as PBS and SBF (both at

pH 7.4) follows Eq. 6.6 [325]:

Ca-Alg2 + 2NaCl 2Na-Alg + CaCl2 (6.6)

When a Ca ion, linking with two Alg chains, exchanges with two Na ions from NaCl in the

medium, Ca-crosslinked Alg network (Ca-Alg2) transforms to Na-Alg. Since this reaction is

reversible, Na-Alg is able to converse to Ca-Alg in contact with excess CaCl2 in the medium.

Alg-foams immersed in PBS showed the highest weight change among all media used (see

Fig. 6.9 (A)). It is considered that more free hydrophilic groups (i.e. COO-) exist in the Alg

network due to loss of Ca-interaction. Thus the network is able to gain more water to fill the

voids. After 4 weeks of immersion, the trend of weight change was significantly reduced,

indicating possible occurrence of decomposition and the foams completely decomposed in

PBS at week 16.

In the case of immersion in SBF, the trend of weight change in the initial stage was

similar to the case of immersion in PBS. However, after 3 weeks in SBF, the weight change

did not vary until the end of the investigation, while the foams in PBS have decomposed.

Page 121: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

107

This observation is suggested to be the result of the reversible reaction which is induced by

the existence of CaCl2 in SBF.

The loss of Ca-interaction inside the Alg network was confirmed by FTIR

spectroscopy (Fig. 6.9 (B)). The characteristic peaks of Alg in the region of 1150 - 900 cm-1

exhibited reduced intensity after 7 days of immersion in all media, indicating deformation of

the Alg network. In particular, the reduced intensity of Ca-O peak at 1009 cm-1

confirmed the

loss of Ca-interaction.

In summary, Alg-foams immersed in different media exhibit slightly different

decomposition rates and therefore different life-times. This result is likely due to the different

compositions of the media used, which directly influences the structural stability of the Alg

network. Therefore, the degradation study of Alg-foams performed in either PBS or SBF in

the present work is only used as a prediction for further in vitro cell culture. In order to

precisely replicate the real body system, degradation studies should also investigate

enzymatic degradation (i.e. with lysozyme and hyaluronidase solution [313]), which play a

fundamental role in the degradation of polysaccharides in vivo.

Figure 6. 9 (A) Weight change of 3 wt/v % Alg-foams as a function of immersion time in

different media, including DI H2O, PBS and SBF. (B) ATR-FTIR spectra of the foam after 7

days of immersion in the media.

Page 122: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

108

6.3.2 Effect of electrospinning conditions on the properties of fibers

(i) Spin-ability of PLLA solution

The electrospinning of biodegradable polymers in the present study was initially

started with a synthetic-derived polymer (PLLA), in order to understand fundamentals of the

influence of polymer solutions and processing conditions using a well-investigated polymer.

First, PLLA fibers were electrospun according to the previous studies of Yunos et al. [85,86].

The optimum conditions have been reported as 5 wt/v % of PDLLA/DMC solution, 8.5 kV of

applied voltage, 15 cm of deposition distance and 0.9 ml/h of feed rate. In our experiment, it

was found that using the solution of 5 wt/v % PLLA/(DCM/MeOH) (80/20 vol %) (Trial 1;

Table 6.1) at the same conditions as Yunos et al. [85], it was not possible to obtain fibers.

With decreasing feed rate (0.5 ml/h; Trial 2), fibers were obtained with the formation of

beads. Even though the applied voltage and deposition distance were adjusted (Trials 3 - 8),

uniform fibers were not achieved by using 5 wt/v % PLLA solution. The presence of beads is

shown in Fig. 6.10 (A), which is a material obtained by using the processing condition in Trial

7. This result can be explained by the fact that the concentration 5 wt/v % PLLA in

DCM/MeOH has insufficient viscosity and low strength in order to be stretched under the

effect of the electrostatic field.

When the concentration was increased to 7.5 wt/v % with the same used feed rate

of 0.5 ml/h and deposition distance of 15 cm, electrospinning at voltages less than 15 kV

(Trials 9, 10) did not lead to fiber formation due to the insufficient strength of electrostatic

field. Under such low strength of electric field, the stable Taylor cone is not able to form and

to accelerate the polymer jet to be drawn from the needle tip to the collector. According to

this, the voltage was raised up (15 - 20 kV; Trials 11 - 15), and in this case the uniform fibers

were obtained with diameters in the range 1 - 1.3 µm, as shown as an example obtained

from using the conditions in Trial 13 (Fig. 6.10 (B)).

In summary, the polymer solution (concentration and solvent) plays the most

important role in the spin-ability. As observed, the PLLA solution started to exhibit suitable

electrospinning behavior at a concentration of 7.5 wt/v %, which is in agreement with the

study of Kenawy et al. [326], who showed that poly(lactic acid) solutions can be electrospun

Page 123: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

109

well at concentrations of around 7 - 8 wt/v %. In addition, the external factors affecting

electrospinning, including applied voltages, deposition distance and feed rate, have an

influence on the formation of electrospinning jet. These factors have an effect on the fiber

morphology and diameter. The applied voltage affects the strength of the electric field, the

acceleration of the electrospinning jet and the stretching ability of the polymer solution. The

deposition distance relates to the flight time of the polymer jet, which plays a role in the

evaporation rate of the solvent. Finally, the feed rate determines the controlled volume of the

polymer solution in order to maintain a stable Taylor cone at the given voltage and it mainly

affects the diameter of obtained fibers.

Table 6. 1 Electrospinning conditions and primary observations of PLLA fibers ( indicates

no fiber formation, indicates beads incorporated into fibers and indicates uniform fibers).

Concentrations

(wt/v %) Trails

Processing parameters Results

Voltage (kV) Distance (cm)) Feed rate (ml/h)

5 1 8.5 15 0.9

2 8.5 15 0.5

3 15 15 0.5

4 18 15 0.5

5 20 15 0.5

6 20 18 0.5

7 15 15 1.15

8 15 15 2

7.5 9 8.5 15 0.5

10 12.5 15 0.5

11 15 15 0.5

12 15 18 0.5

13 18 15 0.5

14 20 15 0.5

15 20 18 0.5

Page 124: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

110

Figure 6. 10 Optical microscopy images of electrospun PLLA fibers obtained by using the

conditions in (A) Trial 4 and (B) Trail 13, according to electrospinning conditions used in

Table 6.1.

(ii) Spin-ability of Alg solution

Up to date, the electrospinning of Alg solution has been studied by considering

variations of the solution systems [173,307–310,327–329]. One of the approaches is to

incorporate water soluble synthetic-derived biodegradable polymers (i.e. PVA and PEO) into

the Alg solution. In the first attempt, Alg was blended with PVA, glycerol, P123 and Gel in

aqueous solution. The series of solution, composition and processing conditions are

reported in Table 6.2. The results of the preliminary study showed that the Alg solutions

blended with PVA and glycerol are not able to form fibers even with variations of

composition and processing conditions (Trials 1 - 4). Regarding the study of Safi et al. [327],

the spin-ability of the Alg solution was seen to improve by increasing hydrogen bonding

between the hydroxyl groups in Alg and either the hydroxyl groups in PVA or the ether

groups in PEO. By this approach, Alg fibers were formed by using a concentration of 7 wt/v

% PVA blended with 2 wt/v % Alg in the volume ratio of 70/30 and 50/50 vol %. The

concentration used in Trail 2 (Table 6.2) was relatively the same as the optimum condition

according to the study of Safi et al. [327]. By using the same electrospinning conditions,

including 12 kV of voltage and 10 cm of needle-collector distance, no fibers were obtained

even with variation of conditions. A reason for this behavior might be the insufficient viscosity

of the used solution, which is caused by the different Mw of the polymers used.

Page 125: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

111

The addition of glycerol in the Alg solution was also attempted (Trials 3 and 4; Table

6.2). Glycerol was aimed to function as a co-solvent, which is expected to enhance the

flexibility of chain conformation and to increase entanglement of Alg molecules [309].

However, fibers were not achieved by this approach, even by the adjustment of processing

conditions. The spin-ability of the Alg solution was further investigated by blending with P123

and the introduction of MeOH as a co-solvent (Trials 5 - 8). P123 copolymer was used in

order to reduce the viscosity of the solution, while MeOH was introduced to enhance the

conductivity of the solution. However, an improvement of spin-ability was not observed. In

detail, as observed from the experiment, no jet was formed during electrospinning. Only

beads were deposited on the collector. This result indicates that the investigated solutions

were not sufficient suitable in terms of interchain interaction to be drawn out and travelled

forward to the collector under the given electrostatic field.

By using the solution of Alg/Gel blending (Trials 9 - 14), a suitable system was found

(Trial 12), which was the only studied system able to form fibers under the investigated

electrospinning conditions (Fig. 6.11 (B)), including voltage of 12 kV, needle-collector

distance of 12 cm and feed rate of 0.1 ml/h. In addition, the set-up was subjected to a heat

gun in order to control the temperature of the process at 50 C. Thus the spin-ability of the

Alg solution can be conveniently improved by the addition of Gel. However, it was still not

possible to achieve high reproducibility, in comparison to the electrospinning of PLLA

solution. Importantly, the electrospinning of Alg/Gel/H2O (1/15/84 wt %) solution was well

performed under controlled temperature conditions. When the content of Gel was decreased

(Trials 9 - 11), the fibers were obtained with the appearance of beads, as shown in Fig. 6.11

(A; Trail 9). It can be concluded that the Gel content plays the key factor in the chain

entanglement of the Alg solution. It is probably due to positively charged Gel forming ionic

interaction with negatively charged Alg. Nevertheless, it is necessary to maintain the

temperature of the solution above the gelling temperature of Gel ( 50 C) in order to avoid

the gelation of the solution during electrospinning. In addition, the applied heat accelerates

the evaporation of the solvent (H2O), leading to skin formation out of the solution jet, which

helps to stabilize the jet being able to travel forward to the collector. However, it was

Page 126: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

112

observed that the Taylor cone was not stably maintained and the jet was not continuously

formed. This is believed to be one of the reasons of the poor spin-ability of the Alg solution.

Therefore, the optimization of electrospinning of Alg solutions requires further research

efforts, in order to enhance both the spin-ability and the reproducibility. As mentioned above,

Alg fibers can be only electrospun with the addition of high amount of Gel. Thus the main

component of the obtained fibers became Gel instead of Alg.

Table 6. 2 Electrospinning conditions and primary observations of alginate fibers ( indicates

no fiber, indicates beads and indicates uniform fibers).

Solutions Compositions

(wt. %) Trials

Electrospinning parameters

Results Voltage

4 (kV)

Distance

(cm)

Feed rate

(ml/h)

Alg/PVA/H2O 1.4/2.1/96.5 1 12 - 20 8 - 15 0.01

3/2/95 2 12 - 20 8 - 15 0.01

Alg/Glycerol/

H2O

2/49/49 3 12 - 20 8 - 15 0.01

4/48/48 4 12 - 20 8 - 15 0.01

Alg/P123/ H2O 3/2/95 5 12 - 20 8 - 15 0.01

1.5/2/96.5 6 12 - 20 8 - 15 0.01

Alg/P123/

H2O/MeOH

3/2/94/1 7 12 - 20 8 - 15 0.01

1.5/2/95.5/1 8 12 - 20 8 - 15 0.01

Alg/Gel/H2O5 1/10/89 9 12 12 0.01

10 12 8 0.01

11 15 12 0.01

1/15/84 12 12 12 0.01

13 12 8 0.01

14 15 12 0.01

4 The applied voltage in the case of Alg solution was fixed at 20 kV because the Alg-based aqueous solutions are

sensitive to high electrostatic field, possibly leading to an occurrence of electric spark during electrospinning. 5 The heat gun was subjected to the electrospinning set-up in order to keep the solution at around 50 °C.

Page 127: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

113

Figure 6. 11 Optical microscopic images of electrospun Alg/Gel fibers obtained by using the

conditions in (A) Trial 9 and (B) Trail 12, according to electrospinning conditions used in

Table 6.2.

(iii) Microstructure

SEM images in Fig. 6.12 show the microstructure of electrospun PLLA and Alg/Gel

fibers, which were obtained by using the optimum conditions (Trail 13; Table 6.1 in the case

of PLLA and Trial 12; Table 6.2 in the case of Alg/Gel). The uniform PLLA fibers, as shown

in Fig. 6.12 (A), were obtained with average diameter of 0.4 ± 0.1 µm, while Alg/Gel fibers

were obtained as a very thin mesh by using the same deposition time as in the case of PLLA

(2 h), as shown in Fig. 6.12 (B). Alg/Gel fibers were uniform without the formation of beads

and they have an average diameter of 0.5 ± 0.2 µm. This result confirms the low spin-ability

of the Alg solution when compared to PLLA solution.

Additionally, when the Alg/Gel fibers were deposited for longer time (9 h) in order to

obtain a thicker mesh, the morphology of Alg/Gel fibers was different from the case of PLLA

fibers. A ribbon-like structure formed in the case of Alg/Gel-electrospinning (see the inset in

Fig. 6.12 (D)), while a common round morphology was found in the case of PLLA fibers (the

inset in Fig. 6.12 (C)). This phenomenon can be described by the reason that the jet of

Alg/Gel solution tends to form a skin on the jet during electrospinning and consequently the

skin collapses into a ribbon as it dried [330,331].

Page 128: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

114

Figure 6. 12 SEM images of PLLA fibers, which were deposited for (A) 2 h and (C) 9 h, and

Alg/Gel fibers, which were deposited for (B) 2 h and (D) 9 h. The distribution of fiber

diameters of both fiber types is included.

Page 129: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

115

(iv) Chemical structure

The presence of both Alg and Gel in the obtained fibers was investigated by using

ATR-FTIR spectroscopy (Fig. 6.13 (A)). The characteristic peaks of Gel were detected at

1637 and 1540 cm-1

, attributed to amide I and II of pure Gel type A, respectively [332]. The

characteristic peaks of Alg were detected as a doublet peak with low intensity at 1082 and

1035 cm-1

, attributed to C-C and C-O stretching of Alg, respectively [332]. Thus the presence

of both Alg and Gel in the electrospun fibers is confirmed.

Moreover, Alg/Gel fibers were chemically crosslinked by exposing to GA vapor, in

order to improve their water resistance and structural stability. The exposure of GA vapor

was chosen in this study instead of soaking in GA solution in order to avoid the potential

toxicity from GA solution. The ATR-FTIR spectrum in Fig. 6.13 (A) of the Alg/Gel fibers after

GA crosslinking is presented in comparison to uncrosslinked fibers to confirm the occurrence

of crosslinking. The detectable peak at 2369 cm-1

in the spectrum of crosslinked fibers

confirmed the occurrence of crosslinking, according to the literature [77]. In addition, the

characteristic peak of amide I (at 1637 cm-1

) in the spectrum of uncrosslinked fibers was

shifted to 1647 cm-1

(see the inset in Fig. 6.13 (A)). This result indicates that a chemical

reaction has taken place by the crosslinking. The effective crosslinking was additionally

confirmed by the detection of ethylene groups (–CH2-) of GA at 2930 and 2880 cm-1

[214],

which exhibited higher intensity in comparison to the spectrum of fibers without crosslinking.

The mechanism of GA crosslinked gelatin is shown in Fig. 6.13 (B). The carbonyl groups in

GA tend to react with reactive amine groups (–NH2) in Gel, forming terminated -C=N- bonds

on a GA molecule.

Page 130: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

116

Figure 6. 13 (A) FTIR spectra of Alg/Gel fibers before and after GA crosslinking (The inset

represents the change of amide I peak before and after crosslinking) and (B) mechanism of

the crosslinking reaction between Gel with GA, according to [78].

(v) Crystalline structure

The crystalline structure of electrospun fibers was analyzed by using XRD analysis

in comparison with the cast film used as control (Fig. 6.14). The -crystalline phase of PLLA

cast film appeared as a sharp and narrow peak at 17.3 and 19.7 2 (as marked by ■ in

Fig. 6.14) and the β-crystalline phase appeared at 30 2 (as marked by in Fig. 6.15),

according to reference [333]. In contrast, the crystalline peaks almost disappeared in the

case of PLLA fibers, indicating that the reduction of crystallinity in PLLA has been taken

place during electrospinning. Consequently, an amorphous phase was dominant in the

fibers. The reason for this behavior is likely the fact that the jet travelling toward the collector

has no enough time for crystal growth during electrospinning [172,238]. As a result, this

phenomenon directly influences the mechanical properties of the fibers. The reduction of

crystallinity leads to the decrease of the mechanical strength, which is discussed in Section

(vii).

Page 131: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

117

Figure 6. 14 XRD patterns of PLLA fibers showing the reduction of crystallinity in

comparison with PLLA cast films (■ and indicate - and β-crystalline phase of PLLA,

respectively).

(vi) Thermal properties

The thermal properties of the electrospun PLLA fibers are reported as the DSC

curves in Fig. 6.15. Tg of PLLA fibers is detected at 63.7 C, which is comparable to the

value of as-received PLLA pellets (67 °C) [334] and of porous PLLA scaffolds fabricated by

phase separation technique [320,335]. This result indicates that the molecular mobility of

PLLA fibers does not occur at the constant temperature (37 C) in body fluids. In contrast,

Alg-foams exhibited Tg values lower than 37 °C. Thus under in vitro culture conditions, it can

be stated that the molecular structure of PLLA fibers is more stable than that of Alg-foams.

An exothermic transition peak is observed at 79.2 C as a cold crystallization. Tm as

endothermic transition peak is shown at 177.6 C, which is not remarkably different from Tm

of PLLA pellets at 170 °C [334]. These results confirm the molecular stability of PLLA fibers

under the physiological conditions (37 °C), without the occurrence of melting or re-

crystallization.

Page 132: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

118

Regarding the XRD results in Fig. 6.14, the reduction of crystallinity of PLLA after

electrospinning is confirmed by quantitative measurement of the degree of crystallinity,

which is based on the DSC result (Fig. 6.15). According to Eq. 6.3, the percent crystallinity

(Xc) is determined by considering the heat energy of melting (∆Hm) and the heat energy of

cold crystalline (∆Hc), which are obtained from the area under the peaks. The crystallinity

(Xc) of PLLA fibers is 28.5 %, while Xc of the cast film has been reported in the range of 35

- 37 % [301]. Therefore, this result confirms the weak crystalline structure found by XRD,

which is the result of the electrospinning process.

Figure 6. 15 DSC thermogram of PLLA fibers, indicating Tg, Tc, and Tm.

Page 133: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

119

(vii) Mechanical properties

Electrospun fibrous scaffolds are interesting for application in cartilage tissue

engineering because the fibers provide a high surface area to volume ratio and allow cells to

grow while providing also sufficient mechanical support [336]. In terms of the mechanical

properties, the high flexibility of the fibers is relevant for achieving low friction and adequate

load transfer, which are characteristic features of cartilage tissue, as discussed in Chapter 2.

In order to evaluate the mechanical properties of the fibers, tensile testing was performed to

compare also with PLLA cast films. As shown in Fig. 6.16 (A), a representative tensile

stress-strain curve of PLLA fibers shows the typical behavior of synthetic polymers, which is

composed of both elastic and plastic regions, as observed also for PLLA cast films (see the

inset in Fig. 6.16 (A)). Importantly, the fibers exhibited higher elongation at break ( 19.4 ±

0.3 %) when compared to the value of the cast films (11 ± 2 %), as reported in Fig. 6.17 (B).

This behavior indicates that PLLA fibers are more flexible than the cast films, which can be

ascribed to the fact that randomly oriented fibers generate highly intersectional interaction

along the mesh. In contrast, the elastic modulus and tensile strength of the fibers ( 18.7 ±

0.6 and 2.3 ± 0.1 MPa, respectively) are lower than those of the cast films ( 47 ± 10 and 21

± 3 MPa, respectively) (Fig. 6.16 (B)), which is likely due to the reduction of the crystallinity

in the fibers. When compared to the reference values reported in Table 2.2; Chapter 2, the

elastic modulus of PLLA fibers is confirmed to be in the range of the values obtained from

the native cartilage tissue (5 - 25 MPa). The tensile strength of PLLA fibers is closer to the

value range of the strength of native cartilage (3.7 - 10.5 MPa).

Page 134: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

120

Figure 6. 16 (A) Representative tensile stress-strain curve of PLLA fibrous meshes in

comparison with the typical curve of PLLA cast films (the inset) and (B) the values of elastic

modulus1, tensile strength

2 and elongation at break

3 of PLLA under tension deformation

mode.

The mechanical properties of crosslinked Alg/Gel fibers could not be tested due to

the very low yield of electrospun Alg/Gel fibers. Since the reproducibility is low, Alg/Gel

fibers were not able to be obtained as a stripe which is generally required as specimen for

tensile testing. Up to date, the mechanical properties of Alg/Gel fibers have not been

reported. Nevertheless, the range of mechanical properties available from the previous

studies with respect to the fiber diameter is summarized in Fig. 6.17, showing the trend of

the mechanical properties of fabricated synthetic and natural polymer-based fibers. In brief,

the elastic modulus of Alg/PEO fibers fabricated in the study of Bhattarai et al. [306] is

significantly lower than the values of synthetic-derived polymer fibers. This result is

suggested by the weak intrinsic strength of Alg itself and the lack of entanglement in the

electrospun Alg solution. In contrast, Gel fibers with and without crosslinking (Huang et al.

[337] and Panzavolta et al. [338]) exhibit the elastic modulus in the same range as the

1 Elastic modulus or Young’s modulus (E), according to tension deformation, was determined by using the same

procedure as compressive deformation. 2 Tensile strength in this case was represented as the values of strength at fracture or strength at break [383].

3 Elongation at break or elongation of fracture, which indicates the ductility of materials, was determined from the

strain at fracture in the stress-strain curve under tension force [383].

Page 135: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

121

values of PLLA fibers (Wang et al. [238] and Cao et al. [278]). In this case, the fibers of

Gel/PLLA (Yan et al. [301]) exhibited a suitable combination of good mechanical properties

and controlled degradation rate of PLLA, in addition to the convenient biological properties of

Gel.

Figure 6. 17 Summary of Young’s modulus values of electrospun fibers with respect to fiber

diameter, which were collected from recent literature reports [238,278,301,306,337–340],

mainly on polyesters, Gel and Alg. The star indicates the position of PLLA fibers obtained in

the present work.

6.4 Conclusions

Porous Alg-foams were successfully fabricated by the combination of gelation,

freeze-drying and physical crosslinking methods. The optimum conditions are composed of

3 wt/v % Alg solution, 0.1 M CaCl22H2O gelation agent and 0.5 M CaCl22H2O crosslinking

agent. The obtained foams exhibited high porosity 95 %, suitable pore size of 237 ± 48

µm and relative high water absorption ( 82 %). Elastic modulus and compressive strength

Page 136: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

122

were 0.22 ± 0.09 MPa and 0.14 ± 0.02 MPa, respectively. The foams showed high

interconnected pore structure and tubular-like porosity, which is suitable for cell arrangement

mimicking the organization of chondrocytes in native articular cartilage. Therefore, Alg-

foams fabricated in the present study provide the potential for use as a cartilage scaffold,

according to porous structure, pore geometry and physico-mechanical properties.

Alg fibers were fabricated by electrospinning of Alg/Gel/H2O solution (1/15/84 wt %).

The spin-ability of Alg-based solution was low in comparison with that of PLLA solution,

leading to low yield of obtained fibers. The reason is likely the low chain entanglement of the

alginate solutions used. The obtained fibers lack suitable structural stability and they

exhibited low water resistance, which was modified by chemical crosslinking (i.e. exposure

of GA vapor). The electrospinning of Alg solution requires further investigation in order to

enhance the spin-ability. For instance, proper solution systems, which could be based on

Gel addition are recommended to be further developed.

Here, Alg scaffolds based on two different structures, namely porous freeze-dried

foams and electrospun fibers, are comparatively studied. Alg-foams fabricated by freeze-

drying technique provide advantages in that they exhibit suitable porosity and pore size for

cartilage regeneration. The pore size in the range of 50 - 300 µm has been suggested to be

feasible for culturing with chondrocytes and for transporting nutrients to cells. In contrast,

fibrous scaffolds exhibit relatively small pore size ( 50 µm in the case of Alg/Gel fibers). In

addition, the columnar pores of Alg-foams can be tailored by the variation of the solution

used and conditions in the process of freeze-drying. On the other hand, Alg/Gel fibers with

fiber diameter of submicron dimension and relative large area per volume ratio exhibit

advantages to structurally mimic the native ECM and to support cell adhesion. In term of

manufacturing, Alg-foams show better reproducibility when compared to Alg/Gel-fibers.

From the materials point of view based on the studies presented in this chapter, it can be

preliminarily concluded that Alg-foams represent a more convenient scaffolds for further

investigations than electrospun structures in the context of their application in cartilage

regeneration strategies , e.g. as the cartilage-side scaffold in a multilayered system for

osteochondral regeneration.

Page 137: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 7

Multilayered Scaffolds Suitable for Osteochondral Tissue Engineering

7.1 Introduction

Osteochondral repair is one of the most challenging areas in the general field of

interface tissue engineering due to the complexity of tissues involved and limitation in self-

repair capability of cartilage [16,19,139,140,261]. Therefore, scaffolds based on biomimetic

approaches for osteochondral tissue engineering are becoming more sophisticated

[20,24,139]. Numerous research groups are concentrated in bi- or multilayered stratified

scaffolds in order to assemble the structure, architecture and functional properties of the

complex cartilage-bone interface tissues with the purpose to achieve the organization of

cells and new tissues by both in vitro and in vivo approaches. In addition, the formation of a

stable interface between cartilage and subchondral bone remains a significant challenge [8,

9], since the interface must exhibit sufficient structural integrity for both cartilage and SB to

become effectively connected. By employing a stratified scaffold approach, suitable scaffold

manufacturing techniques are necessary, representing an important factor to obtain optimal

porous structures and suitable mechanical properties of the rather complex resulting

composite scaffolds. In addition, manufacturing techniques are mainly dependent on the

selected biomaterials as well as on the applications. From a mechanical standpoint, cartilage

and bone are referred to as soft and hard tissues, respectively [261]. Natural polymers (such

as Col and polysaccharides) or water-soluble low molecular weight synthetic polymers are

suitable for fabricating cartilage scaffolds, whereas high molecular weight synthetic

polymers, ceramics, composites and metals are widely used for bone tissue engineering

scaffolds, which has been discussed extensively in Chapter 2. In order to fabricate a suitable

scaffold for the application of osteochondral tissue regeneration, the characteristics and

properties of involved tissues must be initially taken into account. The design of scaffolds,

selection of biomaterials and use of fabrication techniques are a crucial conceptual step. In

Page 138: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

124

the context of the present project, the specific topics related for selected scaffold materials,

the chosen fabrication techniques and suitable scaffold characterization approaches have

been discussed previously in Chapters 4 and 6 for the applications of the scaffolds in SB

and cartilage regeneration, respectively. The main goal of the investigation reported in the

present chapter was the development of multilayered scaffold approaches, which were

based on the optimum processing conditions reported in the previous chapters. The scope

of the work presented thus includes: (i) to develop novel multilayered composite scaffolds

with different approaches, according to biomimetic considerations and (ii) to overcome the

weak and unstable interface of the multilayered composite scaffolds in order to avoid

delamination during in vitro studies and for safe further implantation.

The multilayered scaffolds in the present work were categorized into four different

systems (Fig. 7.1), including (A) monolithic biphasic scaffolds (Alg-foam/Alg-c-BG scaffold),

(B) integrated bilayered scaffolds (Alg-foam/Alg–c-BG scaffold), and (C and D) integrated

electrospun fiber mesh/coated Bioglass® bilayered scaffold (PLLA fiber mesh/PDLLA-c-BG

and Alg/Gel fiber mesh/Alg-c-BG scaffold). It was noted that the polymer coating onto the

Bioglass-based scaffolds for bone phase was the same material as used to fabricate

cartilage scaffold due to the fact of chemical compatibility between dintinct phases. All

investigated systems were mainly different in the utilization of fabrication techniques and

consequently a variety of structural properties of the scaffolds was obtained. In detail, the

interfacial phase was generated by using two different methods, including (i) polymer

infiltration leading to a single biphasic scaffold with continuously formed interface (according

to system A) and (ii) polymeric adhesive serving as an interface, leading to integrated

bilayered scaffold (according to system B). On the other hand, the interface of scaffolds in

the systems C and D was created by directly electrospinning the polymer solution onto

Bioglass scaffolds. The fibers anchored onto the scaffold struts were aimed at enabling a

suitable integration between the two scaffold layers.

Page 139: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

125

Figure 7. 1 The schematic diagram of the four types of multilayered scaffolds for

osteochondral tissue engineering developed in this project.

7.2 Materials and methods

7.2.1 Fabrication of multilayered scaffolds

System A: Single biphasic scaffolds were fabricated as follows: a cylindrical Alg-g-

BG scaffold (fabricated as reported in Chapter 4; with the dimensions of 8 mm width, 8 mm

length and 5 mm height) was placed into a custom cylindrical mold (with the dimensions of 8

mm width, 8 mm length and 10 mm height) and then 800 µl of 3 wt/v % Alg solution was

added on top of the scaffold inside the mold. The Alg solution was allowed to infiltrate into

the porous scaffold for 10 sec, in order to form an intermediate phase. Then, 100 µl of 0.1 M

CaCl22H2O agent was added into the mold, in order to induce the gelation of the Alg

component. After 30 min gelation, the mold was frozen at – 20 C overnight and was

lyophilized at – 50 C under vacuum for 24 h. Finally, the obtained biphasic scaffold was

gently removed from the mold and crosslinked by immersion in 0.5 M CaCl22H2O agent for

4 h before being dried at room temperature for 24 h. The scaffolds were of the same

dimensions as the dimensions of the used mold.

Page 140: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

126

System B: 3 wt/v % Alg-foam (fabricated as reported in Chapter 6) was integrated

with Alg-c-BG scaffold (fabricated as reported in Chapter 4) by incorporation of 2 wt/v %

Alg/45S5 Bioglass® (Alg/BG 1:3 by wt.) solution, acting as an adhesive layer. In detail, a

small amount of Alg/BG adhesive was applied on one side of Alg-c-BG scaffold (with the

dimensions of 8 mm width, 8 mm length and 5 mm height) by using a painting brush

(Perikan No. 23). Then, a cubic shaped Alg-foam (without crosslinking; dimensions of 8 mm

width, 8 mm length and 5 mm thickness) was suddenly placed on the adhesive coated Alg-c-

BG scaffold and pressed manually. The obtained multilayered scaffold was immersed in 0.5

M CaCl22H2O agent for 4 h for crosslinking. In this step, the layers of Alg-foam and the

interface were exposed to the crosslinking agent. Finally, the crosslinked multilayered

scaffold (with the dimensions of 8 mm diameter and 10 mm height) was dried at room

temperature for 24 h.

System C: 7.5 wt/v % PLLA solution was electrospun directly onto PDLLA-c-BG

scaffolds (with the dimensions of 8 mm width, 8 mm length and 5 mm height), which were

adhered on the collector by using double-side adhesive tape. PDLLA-c-BG scaffolds were

fabricated by the same procedure described in Chapter 4. The electrospinning conditions

followed the optimum conditions described in Chapter 6 (18 kV supplied voltage, 15 cm of

needle tip-collector distance, 0.5 ml/h of feed rate and 2 h of deposition time). The PLLA

fiber mesh/PDLLA-c-BG bilayered scaffolds were gently removed from the collector and

were dried at room temperature for 24 h in order to evaporate remained solvent.

System D: The solution of Alg/Gel/DI H2O was prepared in the ratio of 1/15/84 wt %.

The Alg/Gel solution was directly electrospun onto Alg-c-BG scaffolds (with the dimensions

of 8 mm width, 8 mm length and 5 mm height), which were adhered on the collector. The

optimum processing conditions achieved in Chapter 6 were used (12 kV of voltage supplied,

12 cm of needle tip-collector distance and 0.01 ml/h of feed rate). Since the yield of Alg/Gel

electrospinning was very low, the Alg/Gel solution was allowed to be spun for 9 h in order to

obtain a fibrous mesh in the thickness of few hundred microns. Then, the Alg/Gel fiber

mesh/Alg-c-BG bilayered scaffolds were gently removed from the collector and the meshes

Page 141: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

127

were chemically crosslinked by exposing to GA vapor for 48 h. The crosslinking procedure

was detailed in Chapter 6.

7.2.2 Characterization and testing

(i) Microscopy

The appearance of the cross-sectioned multilayered scaffold was characterized

under light microscopy (LEICA M50) with camera operation of LEICA IC80 HD. The

microstructure of the multilayered scaffold was characterized by SEM (LEO 435 VP).

Samples were sputter coated and observed at an accelerating voltage of 10 kV.

(ii) Mechanical testing

Bilayered scaffolds (System B) with the dimensions of 8 mm width, 8 mm length and

10 mm height were tested in compression using a universal testing machine (Zwick Z050) by

applying the compression load at a cross-head speed of 2 mm/min, the preload at 0.1 N and

the maximum load at 50 N. The stress-strain curves were recorded in order to determine the

mechanical behavior of bilayered scaffolds and to compare it with the mechanical properties

of Alg-c-BG scaffolds and Alg-foams. The elastic modulus was calculated from the initial

linear slope in the stress-strain curve and the compressive strength was obtained from the

maximum stress before the sample collapsed. Eight specimens were tested and data were

presented as mean ± SD.

The mechanical strength at the interface of the bilayered scaffolds (System A versus

system B) was quantitatively investigated by using a micro-tensile testing machine

(Zwick/Roell 1120) by applying a maximum force of 100 N and a speed rate of 1 mm/min.

Specimens with dimensions of 3 mm width, 3 mm length, 10 mm height were placed in a

sample holder and fixed with the application of ethyl-cyanacrylate (Loctite® 454) glue and

activated with a glue activator (Loctite® 7455, Loctite Deutschland GmbH, Munich,

Germany). At least eight specimens were tested for each scaffold system and the data were

presented as mean ± SD.

(iii) In vitro acellular bioactivity

The bioactivity and delamination of multilayered scaffolds (Systems C-D) were

qualitatively investigated by immersion in SBF solution (according to Kokubo et al. [297]) (pH

Page 142: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

128

7.4) for 28 days. Each scaffold (System B: the dimensions of 8 mm diameter and 10 mm

height, and Systems C and D: the dimensions of 8 mm diameter and 5 mm height in height)

was put in a polystyrene bottle containing 50 ml SBF solution, which was then incubated in

an orbital shaker (IKA RS 4000i) at 37 C using a rotating speed of 90 rpm. The SBF

solution was replaced twice a week. At every time point, the sample was removed, cleaned

with DI H2O and dried at room temperature for 24 h. The morphology and microstructure of

scaffolds after immersion in SBF were investigated by SEM. The formation of HA was

characterized by using FTIR (Nicolet Nexus 6700, Thermo Scientific, Waltham, MA). The

samples for both, SEM and FTIR spectroscopy, were prepared following the same

procedure as described in Chapter 4 for Bioglass-based scaffolds and in Chapter 6 for Alg-

foams and fibers.

7.3 Results and discussion

7.3.1 Microstructure

Four different approaches for multilayered scaffolds were investigated for

osteochondral tissue regeneration. According to the literature [19,342], two different designs

of multilayered scaffolds suitable for osteochondral tissue engineering have been presented,

namely biphasic but monolithic scaffold and integrated bilayered scaffold. Up to date, it has

not been confirmed which of the designs, either monolithic biphasic scaffold or integrated

bilayered scaffold, is the optimal for this application. Therefore, both scaffold systems were

investigated and comparatively discussed in terms of structure, mechanical integrity at the

interface and in vitro bioactivity.

(i) System A: Monolithic Alg-foam/Alg-c-BG biphasic scaffolds

Fig. 7.2 (A) shows the cross-section of a monolithic Alg-foam/Alg-c-BG biphasic

scaffold, which forms a single continuous macroscopic structure including two distinct

phases. The biphasic scaffold was formed by the infiltration of Alg solution into the porous

Bioglass structure and applying fast gelation process, which leads to fusion of the Alg-gel

with the 3D Bioglass-based scaffold by forming an interconnected interface, as shown in

Fig. 7.2 (a). After lyophilizing, the interface (up to 500 µm in thickness; Fig. 7.2 (a))

Page 143: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

129

generated a porous structure with pore size 100 µm (Fig. 7.3 (A)). Some closed pores

were observed in some regions along the underlying Bioglass-based scaffold (Fig. 7.3 (A

and a)). This phenomenon is related to the relatively low controlled amount of Alg solution

during the infiltration process. It is likely that the excess amount of infiltrated solution can be

avoided by decreasing the infiltration time ( 10 sec), however the interface would not be

sufficiently strong by using shorter time (tested by trial-and-error).

(ii) System B: Integrated Alg-foam/Alg-c-BG bilayered scaffolds

In the case of system B (Fig. 7.2 (B)), Alg-foam served as the scaffold for the

cartilage side and Alg-c-BG scaffold served as the bone scaffold. These scaffolds were

prepared separately and integrated by using Alg/Bioglass adhesive. The intermediate

phase is observed as a dense thin layer in Fig. 7.2 (b). From SEM images (Fig. 7.3 (B and

b)), it was confirmed that Alg-foam and Alg-c-BG scaffold were well integrated by applying

an intermediate layer. It is likely that the intermediate layer provides a suitable connection

between the walls of the Alg-foam and the struts of the underlying Bioglass-based scaffold.

(iii) System C: Integrated PLLA fiber mesh/PDLLA-c-BG bilayered scaffolds

The bilayered scaffolds in system C (Fig. 7.2 (C)) were fabricated by electrospinning

PLLA on top of PDLLA-c-BG scaffolds. During electrospinning, the fibers tended to randomly

align onto the struts of the Bioglass scaffold, used as substrate forming the connecting

points between the two parts (as marked by a dashed circle in the inset; Fig. 7.2 (C)). In

addition, some polymer jets were able to infiltrate through the pores of the Bioglass-based

scaffold, leading to additional integration between both phases, as observed in Fig. 7.2 (c).

As shown in the SEM images, the connection points were obvious between both phases (as

marked by dashed circles in Fig. 7.3 (C)). Moreover, the electrospun PLLA fiber mesh was

obtained as a relatively dense mesh with thickness of around 50 µm (Fig. 7.3 (c)).

(iv) System D: Integrated Alg/Gel fiber mesh/Alg-c-BG bilayered scaffolds

In this approach, random Alg/Gel fibers did not closely pack and became a dense

mesh like in the case of PLLA fibers (Fig. 7.3 (D)). Subsequently, the Alg/Gel fiber mesh

provided larger pore sizes and interconnectivity, which were maintained even after applying

chemical crosslinking, when compared to the PLLA fiber mesh (Fig. 7.3 (d)). In this case in

Page 144: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

130

fact, a fluffy mesh was obtained. In addition, it was found that the fibers did not infiltrate into

the pores of the underlying Bioglass-based scaffold (Fig. 7.3 (D)). This behavior is likely

the result of the intrinsic low strength of the electrospun Alg/Gel solution, as discussed in

Chapter 6. This result might lead to low strength at the interface of the bilayered scaffold.

Figure 7. 2 Optical microscopic images showed the appearance of three different

approaches of multilayered scaffolds, including (A, a) system A: monolithic Alg/Alg-c-BG

biphasic scaffold, (B, b) system B: integrated Alg/Alg-c-BG bilayered scaffold (Reproduced

from Nooeaid et al. [145] with the permission of John Wiley and Sons) and (C, c) system C:

integrated electrospun PLLA fibers/PDLLA-c-BG bilayered scaffold (the inset shows a plan-

view of fibers integrated on the struts of the Bioglass-based scaffold).

Page 145: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

131

Figure 7. 3 SEM images showing cross-sections of four different types of multilayered

scaffolds: (A, a) system A, (B, b) system B (Reproduced from Nooeaid et al. [145] with the

permission of John Wiley and Sons) (C, c) system C and (D, d) system D (the dashed line

marks the interface between the two phases).

Page 146: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

132

7.3.2 Interfacial strength of multilayered scaffolds

Since the interface has a significant impact on the mechanical integrity of bilayered

scaffolds, the interfacial strength of each scaffold system was quantitatively investigated by

application of micro-tensile testing. However, the testing apparatus could not be used on the

scaffolds belong to systems C and D due to the limited 3D structure of the fiber mesh. Thus,

the relative tensile stress-strain curves of bilayered scaffolds in systems A and B were

comparatively investigated in the present work (Fig. 7.4 (A)). Under tension, the interface of

scaffolds exhibited elastic deformation at the initial stage (I) until reaching the yield point,

where the interface-layer starts to break. Afterwards, at stage (II) deformation progresses

continuously until the breaking point (III). Moreover, the stress-strain curves of the interface

followed mainly a typical profile of polymeric materials. The integrated bilayered scaffolds

(System B), which include the adhesive-layer, exhibited higher Young’s modulus (linear

slope at stage (I)) and exhibited higher strength at break, compared to the case of monolithic

layered scaffolds, which have continuous interface. The strength at break of the integrated

bilayered scaffolds was determined in the range of 0.08 - 0.2 MPa, while the strength at

break of the monolithic biphasic scaffolds was lower (up to 0.1 MPa) (Fig. 7.4 (B)). It can be

concluded that the interface formed by the application of an adhesive layer was stronger

than the continuous interface formed in-situ during scaffold fabrication. The reason is likely

the good interconnection between the two distinct phases as a result of the presence of

dense crosslinked Alg/Bioglass adhesive-layer. In contrast, the continuous interface formed

by polymer infiltration through the pores of Alg-c-BG scaffold was not as strong as expected.

This result is explained by the fact that the continuous interface was highly porous, leading

to relative lower strength in comparison with the dense interface-layer formed in the case of

integrated bilayered scaffolds.

Page 147: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

133

Figure 7. 4 (A) Representative stress-strain curves of the bilayered scaffolds (System A vs.

System B) and (B) distribution of the strength at break values of the scaffolds in systems A

and B (the red dashed line is included for the visual aid) (Reproduced from Nooeaid et al.

[145] with the permission of John Wiley and Sons).

7.3.3 Mechanical properties of integrated bilayered scaffolds

In addition to the quantification of the interface strength, the mechanical behavior of

bilayered scaffolds under compressive loading is essential to be investigated. Since joints in

the skeletal system (i.e. knee joint) have to withstand local compression forces during

motion, the compressive stress-strain curves of bilayered scaffolds was monitored in order

to study their mechanical response under compression. According to the better results

achieved from the mechanical interface testing (Fig. 7.4), the system B scaffolds were

chosen to be investigated in this study, in comparison with individual Alg-foam and Alg-c-BG

scaffold. The typical stress-strain curve of the bilayered scaffolds (curve a; Fig. 7.5)

exhibited a sigmoid shape, which represents the same mechanical behavior as the Alg-foam

(curve b; Fig. 7.5). Exceptionally, at the initial stage (in the region up to 20 % deformation)

the curve of bilayered scaffolds showed limited jagged behavior. This is indeed the typical

characteristic of brittle cellular solids (i.e. ceramic foams) [56,317], which was also presented

in the curve of Alg-c-BG scaffold (curve c; Fig. 7.5). These results indicate that the

mechanical response of the system B bilayered scaffolds is predominantly influenced by the

Page 148: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

134

Alg-foam. It can be concluded that Alg-foam mainly acts as a load absorber and load

transferor for the bilayered scaffold, which is supported by the high deformability of the foam

in comparison to brittle Bioglass-based scaffold. Therefore, Alg-foam is confirmed as being

suitable for use as a cartilage scaffold according to the required mechanical performance,

including elasticity and softness [24,343,344].

Figure 7. 5 Representative compressive stress-strain curve of integrated bilayered scaffold

(system B) in comparison with the curves of Alg-foam and Alg-c-BG scaffold (Reproduced

from Nooeaid et al. [145] with the permission of John Wiley and Sons).

7.3.4 In vitro bioactivity

(i) System B: Integrated Alg-foam/Alg-c-BG bilayered scaffolds

It was confirmed that the bilayered scaffold maintained the structural integrity over

28 days in SBF without delamination (Fig. 7.6). This result indicates a strong adhesion

between the two distinct phases. Moreover, the porous structure of the Alg-foam was also

maintained without deformation during immersion in SBF. Importantly, HA did not form on

Alg-foams after 28 days in SBF, which is a desirable behavior avoiding mineralization of the

cartilage side of the scaffold [24,103,125,345]. In the case of Alg-c-BG scaffold, the results

Page 149: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

135

are in an agreement with the bioactivity assessment in Chapter 4. Alg-c-BG scaffold

exhibited the expected ability of HA formation, with HA deposited on the struts of Bioglass-

based scaffold after 28 days of immersion in SBF. However, at day 1 HA formation had not

been observed yet due to the Alg coating, as discussed in Chapter 4. In addition, HA

formation was observed at the interface of the bilayered scaffold (see the inset; Fig. 7.6),

even on the pore walls of the Alg-foam that are in contact with the struts of the Bioglass-

based scaffold. This phenomenon is attributed to the effect of the thin Alg/Bioglass

adhesive layer. It is likely that Bioglass® incorporated into the adhesive improved the

bioactivity of the interface, mimicking the structure of highly mineralized calcified cartilage

[5,8,226].

The no formation of HA on the cartilage side of scaffold (Alg-foams) was confirmed

by ATR-FTIR spectroscopy, as shown in Fig. 7.7. The characteristic peaks of HA were not

detected. In addition, the intensity of the Ca-O peak at 1009 cm-1

was reduced after 28 days

in SBF (the inset in Fig. 7.7), indicating the loss of crosslinking and consequently the

reduction of the mechanical stability of the foam. In contrast, the peak at 1022 cm-1

attributed

to O-H bending increased in intensity after 28 days in SBF. This result indicates that free

negative groups (COO-) are present after losing ionic-interaction during immersion in SBF,

leading to interaction with water molecules and subsequently more O-H groups are

generated, compared to the foam before immersion.

In summary, the desired bioactivity of the bone side of the scaffold was achieved in

addition to the required non-mineralization of the cartilage-side of the scaffold. Moroever, the

bilayered structure of scaffolds in system B was maintained after 28 days of immersion in

SBF, indicating that the fabricated bilayered scaffolds in the present study exhibited the

strong interface and they can avoid delamination during experimental period.

Page 150: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

136

Figure 7. 6 SEM images of integrated bilayered scaffold (system B) after immersion in SBF

for 1 and 28 days (the dashed line indicates the interface between the two phases)

(Reproduced from Nooeaid et al. [145] with the permission of John Wiley and Sons).

Page 151: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

137

Figure 7. 7 ATR-FTIR spectra of Alg-foam after immersion in SBF for 28 days in order to

confirm non-mineralization of the foam (the inset shows the absorption bands in the

wavenumber 1200 - 850 cm-1

).

(ii) System C: Integrated PLLA fiber mesh/PDLLA-c-BG bilayered scaffolds

The bioactivity assessment of system C bilayered scaffolds is presented in Fig. 7.8

by SEM images. After 28 days of immersion in SBF, no delamination was observed in all

investigated scaffolds, indicating suitable adhesion strength at the interface between the

PLLA fiber mesh and the PDLLA-c-BG scaffold. As shown in the inset, some fibers were

adhered to the strut of the PDLLA-c-BG scaffold. In the PDLLA-c-BG layer, HA formation

was not obvious after 1 day in SBF, while HA almost completely covered the struts of

PDLLA-c-BG scaffold after 28 days in SBF. In contrast, HA did not form on the PLLA fiber

mesh after 28 days of immersion in SBF. Hence, PLLA fiber mesh intended for the cartilage

side of the scaffold exhibited no mineralization as required for its use as a cartilage scaffold.

Moreover, the fiber mesh still maintained the interconnectivity during the entire immersion

period in SBF due to the relatively slow degradation of PLLA.

Page 152: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

138

Figure 7. 8 SEM images of integrated bilayered scaffold (system C) after immersion in SBF

for 1 and 28 days.

(iii) System D: Integrated Alg/Gel fiber mesh/Alg-c-BG bilayered scaffolds

In these scaffolds, after 28 days of immersion in SBF, it was observed that the layer

of Alg/Gel fibers became a thin membrane (Fig. 7.9; interface), when compared to the SEM

image of the scaffold before immersion in SBF (Fig. 7.3 (D and d)). It is likely that

interconnection points formed among the fibers after immersion in SBF (as marked by the

dashed circles in Fig. 7.9 (b)). Since the melting temperature of gelatin ( 250 bloom) is

close to the SBF temperature (at 37 °C) [301,346], the Alg/Gel fibers (300 bloom of used

Gel) can be partly melted at the incubation conditions, forming a stronger interconnected

network. Therefore, this effect is suggested to be the reason for the obtained denser layer of

the mesh after 28 days of immersion in SBF. In addition, some cracks were found along the

fiber mesh (Fig. 7.9), which this behavior is likely caused by the breakage of some fibers, as

shown in Fig. 7.9 (b). This result indicates that the Alg/Gel fibers are not sufficiently strong in

Page 153: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

139

order to maintain their structure and to avoid deformation during incubation, even though

chemical crosslinking was applied. In addition, the Alg/Gel fibers did not induce HA

formation, leading to a non-mineralized phase, which is desired for the cartilage side of the

scaffold. In contrast, HA was formed and fully deposited on the strut surfaces of the Alg-c-

BG scaffold after 28 days in SBF, as shown in Fig. 7.9.

Figure 7. 9 SEM image of integrated bilayered scaffold (system D) after immersion in SBF

for 28 days (the dashed line indicates interface between the Alg/Gel fiber mesh and the Alg-

c-BG scaffold and the dashed circles indicate the interconnection between the fibers formed

after immersion in SBF).

7.4 Conclusions

Four approaches of bilayered scaffolds were successfully fabricated for use as a

scaffold in osteochondral tissue engineering applications, according to the porous structure,

mechanical properties and bioactivity. Interestingly, integrated bilayered scaffolds exhibited

higher interfacial strength compared to monolithic but biphasic scaffolds. In addition, the

formation of a controllable interface (by the process of polymer infiltration) of monolithic

biphasic scaffolds was not effectively achieved. In the case of application of electrospinning

of PLLA directly on the Bioglass scaffold, the fibrous layer formed was densely packed.

Page 154: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

140

This leads to small pore size with increasing the thickness of the mesh. In contrast, the

electrospinning of Alg/Gel led to fiber mesh, which was loosely packed. This mesh, however,

exhibited low interconnectivity and subsequently led to low mechanical properties. The GA

crosslinked Alg/Gel fibers were partly deformed after 28 days of immersion in SBF.

Moreover, the Alg/Gel mesh was challenging to produce due to its low spin-ability.

Therefore, the system B – (Alg-foam/Alg-c-BG) bilayered scaffolds are the promising

approach for further investigation and for its use as a suitable scaffold for osteochondral

tissue regeneration. This scaffold type was therefore considered for subsequent studies

discussed in Chapter 9.

Page 155: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 8

Biological Response of Osteoblasts Culturing on Bioglass-based

Scaffolds for Bone Regeneration

8.1 Introduction

An approach in the field of bone tissue engineering requires suitable biomaterial

scaffolds, cells and biomolecules and/or growth factors [55,247]. The 3D scaffold has to

support cell adhesion, proliferation and differentiation, and induce ECM deposition, and

therefore new bone can be grown in 3D [247,262]. The scaffold biomaterials for bone

regeneration are required to provide osteoconductive or even osteoinductive characteristics

in order to accelerate osteogenic differentiation of stem cells [28,347]. 45S5 Bioglass® has

been well established as a bone scaffold due to its osteoconductivity, because of its ability to

bond to bone [31,34,93,276]. In addition, Bioglass can up-regulate the expression of

osteoblast genes in response to dissolution products (i.e. Ca, P and Si), which plays a role in

controlled and enhanced osteogenesis [35,41]. However, highly porous Bioglass®-based

scaffolds are rather brittle and mechanically weak, thus they are not suitable for use in load

bearing applications, as mentioned in Chapter 4. Therefore, biodegradable

polymer/Bioglass® composite scaffolds become a target of interest in bone tissue

engineering approaches [52]. Since the integrated bilayered scaffolds (Alg-foam/Alg-c-BG)

developed in Chapter 7 showed a promising potential for use in osteochondral tissue

engineering, in this chapter Alg-c-BG as SB scaffolds were targeted to study the biological

responses by cultured with bone-like cells. By the fact that Alg has no adhesive sites to cells

and does not adsorb serum proteins due to its high hydrophilicity [61,69], peptides with a cell

adhesive sequence-modified Alg (i.e. Arg-Gly-Asp (RGD) containing amino acid) (RGD-Alg)

was therefore used to enhance cell adhesion on Alg coating in this study. Since amino acid

sequence RGD in fibronectin acts as a primary cell attachment cue, it has been

Page 156: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

142

demonstrated that RGD linear peptide coupling to Alg can enhance osteoblasts adhesion for

several times when compared to unmodified Alg [71].

Therefore, in vitro culture of MG-63 osteoblast-like cells on RGD-Alg and Alg coated

Bioglass®-based scaffolds was investigated in comparison with uncoated Bioglass

®-based

scaffolds, in order to confirm the potential of the developed composite scaffolds for use in

osteochondral tissue engineering applications. Qualitative and quantitative cell adhesion and

proliferation, and cell activity of MG-63 cultured on the coated scaffolds for 3, 7, 14 and 21

days were discussed in this chapter, in comparison with equivalent results on uncoated

Bioglass-based scaffolds.

8.2 Material and methods

8.2.1 Fabrication of Bioglass®-based scaffolds

The fabrication of uncoated Bioglass scaffolds has been described in Chapter 4.

As-fabricated Bioglass scaffolds were sterilized by heat treatment. In detail, the scaffolds

were placed in a furnace (Model B180; Nabertherm GmbH, Germany) and were heated at

160 °C for 7 h. In the case of Alg-c-BG scaffolds, 1.5 wt/v % Alg in DI H2O was prepared as

a coating solution and it was filtered by using a sterilized mesh (pore sizes ~ 0.45 µm). After

that, the sterilized scaffolds were coated with the Alg solution by dipping using a solution of 5

ml per scaffold. The coated scaffolds were then dried under sterilized hood for 24 h. After

drying, Alg-c-BG scaffolds were immersed in 0.5 M CaCl22H2O sterilized solution (1 ml per

scaffold) for 4 h in order to crosslink the Alg coating, as optimized previously in the case of

Alg-foams for cartilage scaffolds. Finally, crosslinked Alg-c-BG scaffolds were washed with

sterilized DI H2O and dried in the sterilized hood for 24 h. RGD-modified Alg coated

Bioglass-based scaffolds (RGD-Alg-c-BG) were prepared by dipping sterilized Bioglass-

based scaffolds in 1.5 wt/v % RGD-Alg solution. Briefly, 100 µM RGD-coupled Alg

(NovaMatrix, FMC Biopolymers, Norway) was mixed with 1 ml Alg solution (1.5 wt/v %).

Then, the RGD-Alg solution was filtered using the same procedure as in the case of the pure

Alg solution. Finally, the sterilized scaffolds were coated and crosslinked as previously

Page 157: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

143

mentioned in the case of Alg-c-BG scaffolds. It is noted that the coating, drying and

crosslinking processes were performed under sterilized hood.

8.2.2 In vitro cell culture

The sterilized scaffolds were pre-treated in culture medium for 24 h, which were

incubated at 37 C, 5 % CO2 and 95 % humidity, in order to stabilize the pH variation due to

the ionic exchange process between scaffolds’ surface and medium. After that, the scaffolds

were gently washed with PBS before cell seeding.

MG-63 osteoblast-like cells were grown in T-150 culture flasks containing culture

medium and kept at 37 C in an atmosphere of 5 % CO2. The cells were trypsinized and

collected as a cell pellet by 5 min of centrifugation at 1200 rpm. Then, the cells were

suspended in culture medium. The culture medium is composed of Dulbecco’s modified

eagle medium (DMEM) with low glucose with L-glutamine supplemented with 10 v/v % fetal

bovine serum (FCS), 1 v/v % penicillin-streptomycin and ascorbic acid. The scaffold was

placed in a 48 well-plate and subsequently seeded with 1 ml cell suspension containing 1

million cells. The cell-seeded scaffolds were cultured in an incubator (at 37 °C, 5 % CO2 and

95 % humidity) for 3, 7, 14 and 21 days. The culture medium was renewed twice a week.

8.2.3 Characterization techniques

(i) Lactate dehydrogenase (LDH) assay

The relative number of cells was evaluated by in vitro toxicology assay kit; lactate

dehydrogenase (TOX7) (Sigma-Aldrich). This assay involves the measurement of the

number of cells via total cytoplasmic LDH [348]. Four scaffolds per each type were placed in

a 48-well plate. Then, 1 ml of lysis buffer was added in each well and incubated at room

temperature for 30 min. After that, 1 ml solution was transferred into a 1.5 ml reaction

vessel. The cell lysis supernatant (140 ml) was obtained after centrifuging the reaction

vessel for 5 min at the speed of 1200 rpm, and it was transferred into a 1 cm cuvette. 60 µl

of LDH substrate mixture (20 µL of each LDH assay dye, LDH assay substrate and LDH

assay growth factor) was added into the cuvette, then the mixture was incubated at room

temperature in dark condition for 30 min. The enzymatic reaction was stopped by addition of

1 N HCl (300 µl per cuvette). Finally, 500 µl of DI H2O was added before the measurement

Page 158: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

144

by UV-Vis spectrophotometer at wavelength 490 and 690 nm was performed. The LDH

activity is presented as the difference between the absorbance at wavelength 490 nm and

the absorbance at wavelength 690 nm.

(ii) Cell imaging

Cell viability, distribution and formation of HA were qualitatively investigated by using

fluorescence microscopy (ZEN with fluorescent lamp HXP120C). DAPI (4, 6-diamino-2-

phenylindole) staining was used to visualize the cell nuclei. The fluorescent bright blue is

observed when the dye binds selectively to double stranded DNA. Encountered OsteoImage

(Lonza) dye was used to visualize HA of the bone-like nodules, which was deposited by

cells, in green fluorescence.

Before staining, the cells-seeded scaffolds were washed with PBS and cultured with

staining reagents starting from OsteoImage mineralization assay for 30 min, followed by

fluorescent fix agent for 15 min. The scaffolds were washed with PBS and were

counterstained with DAPI for 10 min. Finally, the stained scaffolds were kept in PBS before

analyzing.

(iii) Cell morphology

The cell morphology of MG-63 seeded on BG, Alg-c-BG and RGD-Alg-c-BG

scaffolds after 14 days in culture was analyzed by SEM (LEO 435 VP). After the cells-

seeded scaffolds were washed with PBS, they were fixed with 3 v/v % GA in 0.1 M sodium

cacodelate for 1 h and subsequently they were dehydrated by using the series of EtOH

solutions (30, 50, 70, 80, 90, 100 %, respectively) for 30 min. Finally, the fixed samples

were dried by using critical point drying (Leica EM CPD300, Germany). All the samples were

sputter coated and observed under SEM at 10 kV.

(iv) AlamarBlue assay

Cell metabolic activity was quantitatively evaluated by using AlamarBlue (AB) cell

reagent (Invitrogen). AB assay is based on oxidation-reduction reaction, which indicates the

cells’ metabolism activity [349]. Six scaffolds per each type were investigated at 3, 7, 14 and

21 days culture. The scaffolds were placed in a 48-well plate, then 500 µl of 10 v/v % AB dye

in culture medium was added in each well. The samples were placed in the incubator at 37

Page 159: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

145

C and 5 % CO2 atmosphere for 4 h. After that, 500 µl of medium from each well was taken

into a 1 cm cuvette. 500 µl of DI H2O was then added into the cuvette. The absorbance

value of solution was measured at wavelength 570 and 600 nm by using UV-Vis

spectrophotometer (Specord 40; Analytikjena, Germany). The results are presented in % AB

reduction with respect to culture time, which were determined by using Eq. 8.1 [349]:

% AB reduction = [ALW – (AHW × R0)] × 100 (8.1)

; where ALW and AHW are the difference between the absorbance of sample and the

absorbance of medium blank at 570 and 600 nm, respectively, and R0 is correction factor,

which was determined by using Eq. 8.2 [349]:

R0 = AOLW/AOHW (8.2)

; where AOLW and AOHW are the difference between the absorbance of AB mixture and the

absorbance of medium blank at wavelength 570 and 600 nm, respectively.

(v) Alkaline phosphatase (ALP) assay

The ALP assay was used to evaluate the osteoblastic activity, considering that

osteoblasts are very rich in ALP enzyme [350]. The ALP enzyme catalyzes the hydrolysis of

p-nitrophenyl phosphate (p-NPP; transparent) into p-nitrophenol (p-NP; yellow) [275,350].

Four scaffolds of each type were investigated. Briefly, 150 µl of cell supernatant was added

into a 1 cm cuvette. 100 µl of DI H2O was added, then 100 µl of ALP buffer at pH 9.8 (0.1 M

Tris; Mw 12114 g/mole, 2 nM MgCl2; 95.3 g/mole, 9 mM p-NPP) was added. The cuvette

was incubated at 37 C in dark condition. The reaction was stopped by addition of 1 N NaOH

(300 µl), while the reaction time was recorded. The cuvette was filled with DI H2O (350 µl)

and then the absorbance was measured by using UV-Vis spectrophotometer at wavelength

405 and 690 nm. The relative ALP activity was determined by using Eq. 8.3:

ALP activity (OD/min/g) = ∆A/(T × P) (8.3)

; where ∆A is the difference between the absorbance values at 405 nm and at 690 nm, T is

reaction time in min and P is total protein concentration (g/ml). The protein content was

determined by the method of Bradford protein assay with respect to the calibration of

standard protein solutions. The standard solutions were prepared at concentrations of 0,

100, 200, 400, 600, 800 and 1000 µg/ml. The absorbance of the standard solution was

Page 160: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

146

measured by using UV-Vis spectrophotometer at a wavelength 595 nm. The calibration

curve was obtained as Eq. 8.4:

Total protein concentration (µg/ml) = (A595 – 0.014)/0.001; R2 = 0.99 (8.4)

The total protein amount in each sample (25 µl of supernatant and 975 µl of Bradford

solution in a 1 cm cuvette) was obtained by measuring the absorbance at wavelength 595

nm by using UV-Vis spectrophotometer.

8.2.4 Statistical analysis

All data were analyzed by using one-way analysis of variance (ANOVA) and turkey’s

multiple-comparison test to determine statistical differences. A confidence interval of 95 % (p

= 0.05) was used for all analyzes. Mean values and SD are presented.

8.3 Results and discussion

8.3.1 LDH activity

In vitro LDH activity describes the relative number of cells over culture time, which

was determined by measuring relative LDH activity from cell lysates [348]. In Fig. 8.1, it is

seen that the number of cells on uncoated BG scaffolds significantly increased over the

culture time, in particular after cultivation for 14 days (* p 0.05). This result indicates the

proliferation of MG-63 cells cultured on uncoated BG scaffolds. At the initial culture period

(up to 7 days), cells seeded on both Alg-c-BG and RGD-Alg-c-BG scaffolds exhibited higher

LDH activity when compared to cells seeded on uncoated BG scaffolds, in particular cells

seeded on RGD-c-BG scaffolds showed the highest LDH activity. On the other hand, LDH

activity of cells seeded on uncoated BG scaffolds was significantly higher than that of cells

seeded on coated scaffolds after 14 days of culture. This result can be explained that at the

initial time of culture (up to 7 days), intense ionic exchange between the surface of uncoated

BG scaffolds and the culture medium induced increased pH of the medium, leading to

decreased cell proliferation. This response is in agreement with the studies of in vitro cell

culture of PDLLA coated and P(3HB) coated Bioglass-based scaffolds [55,351]. It can be

speculated that polymer coating plays a role in surface reactivity of Bioglass-based

scaffolds, which the coating inhibits the ion exchange at the Bioglass surface in the initial

Page 161: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

147

culture. Furthermore, at day 14 in culture, the cells seeded on uncoated BG scaffolds

exhibited the highest LDH activity compared to the cells seeded on coated scaffolds. The

reason is likely that over a longer period of time, the dissolution of the polymer coating took

place in contact with the culture medium, leading to pH variation and subsequently inhibition

of cell growth on the coated scaffolds. This explanation can be confirmed by low LDH activity

of cells seeded on coated scaffolds at day 21 of culture, compared to cells seeded uncoated

scaffolds. In addition, the result of LDH activity confirms more cells of RGD-Alg-c-BG

scaffolds compared to Alg-c-BG scaffolds. It is likely that RGD-Alg-c-BG scaffolds supported

better the growth of MG-63 cells over culture time compared to Alg-c-BG scaffolds.

Figure 8. 1 Relative LDH activity of MG-63 osteoblast-like cells cultured on uncoated BG,

Alg-c-BG and RGD-Alg-c-BG scaffolds. The results presenting the difference of optical

densities are presented as mean ± SD (n = 4). * (p 0.05) indicates significant difference of

different scaffolds at different culture times.

Page 162: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

148

8.3.2 Cell imaging

Cell adhesion and distribution were qualitatively evaluated by DAPI staining, giving a

blue color by using fluorescence microscopy. Fluorescent DAPI staining images in Fig. 8.2

confirms effective infiltration of seeded MG-63 cells through the porous structure of all

investigated scaffolds and confirms the adhesion of cells on the surface of all the scaffolds

after 3 and 14 days in culture. It is seen that MG-63 cells exhibited quite homogeneous cell

distribution on the struts of all uncoated BG, Alg-c-BG and RGD-Alg-c-BG scaffolds. By this

result, the number of cells was difficult to distinguish among all 3D scaffold types. In

addition, it can be concluded that even though Alg lacks cell adhesion moieties, the present

study proved that porous Alg-c-BG scaffolds can support adhesion of MG-63 cells. Similar to

the previous study of Srinivasan et al. [42] that porous Alg-based scaffolds are

biocompatible with MG-63 cells and can support the growth of cells.

Figure 8. 2 Fluorescent microscopic images of MG-63 osteoblast-like cells-seeded BG, Alg-

c-BG and RGD-Alg-c-BG scaffolds after 3 and 14 days in culture by using DAPI stain for cell

nuclei.

Page 163: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

149

8.3.3 Metabolic activity

The metabolic activity of MG-63 cells seeded on BG, Alg-c-BG and RGD-Alg-c-BG

scaffolds was presented in % AB reduction. Fig. 8.3 shows that in the case of uncoated BG

scaffolds, % AB reduction increased over culture time and in particular % AB reduction

significantly increased (* p 0.05) after 14 days of culture. MG-63 cells maintained their

metabolic activity on uncoated BG scaffolds over 21 days in culture, which is indicated by

the reached values of % AB reduction. This result indicates that uncoated BG scaffolds can

activate metabolic activity of osteoblast-like cells over culture time. This phenomenon is

supported by the effect of dissolution product of Bioglass. In contrast, MG-63 cells seeded

on both Alg-c-BG and RGD-Alg-c-BG scaffolds did not show increased % AB reduction over

the culture time. It confirms the LDH activity (Fig. 8.1) that progressive cell proliferation was

not observed on the coated scaffolds, which is indicated by no changes in cell metabolic

activity over the culture time. In addition, it is likely that both coated scaffolds, Alg-c-BG and

RGD-Alg-c-BG, did not lead to significant difference in cell metabolic activity.

Page 164: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

150

Figure 8. 3 Cell metabolic activity of MG-63 osteoblast-like cells cultured on uncoated BG,

Alg-c-BG and RGD-Alg-c-BG scaffolds. The results in % AB reduction are presented as

mean ± SD (n = 6). * (p 0.05) indicates significant difference of different scaffold types at

different culture times.

8.3.4 Osteoblastic activity

Fig. 8.4 shows ALP activity results of MG-63 cells on the surface of BG, Alg-c-BG

and RGD-Alg-c-BG scaffolds. ALP activity increased in all types of scaffolds with increasing

culture time. After 14 days in culture, all investigated scaffolds promoted significant increase

of ALP activity compared to the culture at day 3 (* p 0.05). At day 7 in culture, increased

osteoblastic activity was not significantly detected. In addition, the results of ALP activity are

in agreement with the results of LDH activity and cell metabolic activity. The increase in LDH

activity cell and metabolic activity leads to an increase in total protein adsorption and

subsequently ALP activity [275]. At day 3 of culture, cells seeded on Alg-c-BG and RGD-Alg-

c-BG scaffolds exhibited significantly lower ALP activity than that detected in cells seeded on

uncoated BG scaffolds (* p 0.05). This result can be explained by the fact that the

dissolution products from uncoated BG scaffolds can activate gene expression in

osteoblasts [42], while the release of ions from the Bioglass surface in the case of coated

Page 165: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

151

scaffolds may be inhibited by the polymer coating at the initial stage of culture, as previously

discussed in the result of LDH activity. However, during 7 and 21 days in culture, there is no

significant difference among all scaffold types, indicating that by longer period of culture

time, the polymer coatings did not negatively affect the osteoblastic activity. It is suggested

that the resorption of the coating layer occurs with increasing culture time and thus ions

released from the coated scaffolds has an effect, as supported by the in vitro bioactive study

of Alg-c-BG scaffolds in Chapter 4. In addition, it is likely that the uncontrollable release of

Ca ions from crosslinked Alg and RGD-modified Alg coatings did not negatively influence the

ALP production of cells cultured on both coated scaffolds. Also, the Ca ions seem to support

the ALP activity of the cells cultured on both coated scaffolds. After 14 days of culture, there

is no further increase of ALP activity in all scaffolds, indicating complete osteoblastic activity.

As also reported by the previous study on Alg/nano Bioglass composite scaffolds seeded

with MG-63 cells, the cells exhibited maximum ALP activity at day 7 and the ALP activity

further decreased after prolong culture, which this phenomenon correlates to the maturation

of the cells [42]. Moreover, at day 14 and day 21 in culture there is no significant difference

of ALP activity between Alg-c-BG and RGD-Alg-c-BG scaffolds. This result indicates that

RGD-Alg-c-BG scaffolds did not improve the activity of the cells in terms of ALP production.

In summary, all scaffold types, including uncoated BG, Alg-c-BG and RGD-Alg-c-BG

scaffolds, can support the activity of ALP production of MG-63 cells. Since ALP is an early

indicator of mineralization of osteoblasts, further tests such as Col production, osteocalcin

and osteopontin should be evaluated in order to confirm the mineralization of MG-63 seeded

on the scaffolds.

Page 166: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

152

Figure 8. 4 ALP activity up to 21 days of MG-63 osteoblast-like cells cultured on uncoated

BG, Alg-c-BG and RGD-Alg-c-BG scaffolds. The results are reported as mean ± SD (n = 4).

* p 0.05 indicates significant difference of results for different scaffold types at different

culture times.

8.3.5 Cell morphology

SEM images in Fig. 8.5 confirm the cell growth of MG-63 cells on all scaffold types

after 14 days in culture. The SEM images (Fig. 8.5 (A-C)) showing overview of porous

scaffolds confirm that the cells were able to infiltrate into the porous structure of all scaffold

types. This observation indicates that the fabricated Bioglass-based scaffolds exhibited

suitable pore size and porosity for supporting the cell seeding and infiltration, and further cell

growth. Polymer coatings (both Alg and RGD-Alg) did not inhibit infiltration of cells, since

open pores still maintained. In addition, the cells are seen to be elongated on the struts of

scaffolds and the cells grew reaching tens of microns in all investigated scaffold types, as

shown in Fig. 8.5 (D-F). The well flattened cells covering the scaffold struts tended to group

and formed a monolayer in all scaffold types. It is likely that there is no significant difference

of cell morphology and cell distribution among all scaffold types. In addition, the crystals

were formed on the struts of all investigated scaffolds, in particular the uncoated scaffolds.

Page 167: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

153

The crystals are likely to be HA crystals that were initiated by ionic exchange between

Bioglass surface and the culture medium. This behavior has been previously reported

[55,351,352]. In the case of coated scaffolds, it can be explained that the dissolution of

polymer coating enables direct contact of the Bioglass surface with the medium, leading to

formation of HA crystals. It can be also seen that the cells grew and covered the crystals, as

obviously shown in Fig. 8.5 (E and F).

The HA formation can be confirmed by fluorescent images of stained OsteoImage

(green color) on MG-63-seeded scaffolds after 3, 7, 14 and 21 days in culture (Fig. 8.6). The

formation of HA was observed in all investigated scaffolds after 3 days in culture and it

tended to increase with culture time.

Figure 8. 5 SEM cross-sectioned images showing MG-63 cells-seeded BG, Alg-c-BG and

RGD-Alg-c-BG scaffolds after cultured for 14 days.

Page 168: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

154

Figure 8. 6 Confocal microscopic images of MG-63 cells cultured on BG, Alg-c-BG and

RGD-Alg-c-BG scaffolds, stained with OsteoImage (green), after 3,14 and 21 days.

All the results indicate high biocompatibility and the ability to stimulate metabolic

activity of MG-63 osteoblast-like cells on BG, Alg-c-BG and RGD-Alg-c-BG scaffolds. The

present study proves that Alg and RGD-Alg coatings did not provide negative effect on cell

activity. In addition, Alg-c-BG and RGD-Alg-c-BG scaffolds could support the adhesion,

growth and activity of osteoblast-like cells similar to previous studies of polymer coated

Bioglass-based scaffolds, for example PDLLA-c-BG and PHB-c-BG scaffolds [55,351].

Even though the polymer coating generally reduces surface roughness of Bioglass-based

scaffolds and therefore inhibits cell adhesion, this phenomenon did not appear in the present

study by Alg coating, as evidenced by SEM images in Fig. 8.5. In addition, negatively

charged Alg coating did not negatively affect cell adhesion. Therefore, it can be speculated

Page 169: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

155

that dissolution products of Bioglass and pH variation of the culture medium in the initial

culture plays the important role in this study. By this fact, the study of pH variation of the

culture medium over culture time should be taken into account.

8.4 Conclusions

From the performed in vitro biological investigation, all scaffolds investigated,

namely uncoated BG, Alg-c-BG and RGD-Alg-c-BG scaffolds, supported MG-63 osteoblast-

like cells adhesion and osteoblastic metabolic activity. The results indicate that all types of

investigated scaffolds are compatible with MG-63 cells and can support their growth. Given

high porosity and macro-pores of both Alg-c-BG and RGD-Alg-c-BG scaffolds promoted

effectively cell infiltration through the porous structure of the scaffolds similar to the case of

uncoated scaffolds. RGD-modified Alg coated onto BG scaffolds led to higher number of

cells over culture time compared to pure Alg coated BG scaffolds, as evidenced by LDH

activity. In addition, cell metabolic activity, proliferation and ALP of all scaffold types were

quantitatively confirmed. The ALP activity results showed that there is no significant

difference of osteoblastic activity among all investigated scaffolds after 14 days in culture.

Moreover, HA increasingly deposited on the surface of all scaffolds with increasing culture

time, which was confirmed by OsteoImage stained fluorescent images. By the present

results, uncoated BG and RGD-Alg-c-BG scaffolds are attractive for bone tissue engineering

applications, while Alg-c-BG scaffolds are limited in their adhesion ability. In addition, further

evaluations should be determined such as Ca content, Col I production, osteocalcin and

osteopontin in order to confirm the mineralization ability of osteoblast-like cells seeded on

the scaffolds. Since the use of MG-63 osteoblast-like cells cultured on the scaffolds in the

present study preliminarily confirmed their osteoblastic activity, MSCs cultured on the

scaffolds should be further studied in order to confirm the osteoblastic differentiation.

Here, the present chapter has confirmed the ability of fabricated Bioglass-based

scaffolds for use in subchondral bone tissue engineering applications, while the porous Alg-

foams fabricated by using freeze-drying technique (as presented in Chapter 6) were aimed

Page 170: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

156

to investigate the potential for use in cartilage regeneration and were presented in the next

chapter.

Page 171: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 9

Biological Response of Chondrocytes and Mesenchymal Stem Cells on

Alginate/Chondroitin Sulfate Scaffolds for Cartilage Regeneration

9.1 Introduction

In tissue engineering approaches, scaffolds, cells and biomolecules are used as

main components for tissue regeneration [19,163,247]. Besides the requirement of a

scaffold material to provide a proper environment for cell growth, the cell source has to be

considered as well in order to develop a specific strategy for tissue regeneration. In the

cartilage engineering approach based on MSCs, MSCs have to be retained within the

scaffold and be capable to undergo chondrogenic differentiation in the specific scaffold

environment [163,242]. In this chapter, porous Alg-foams fabricated by freeze-drying

technique (as detailed in Chapter 6) were aimed to confirm the potential for use as a

cartilage scaffold. Alg scaffolds have been shown to promote proliferation and differentiation

of MSCs to chondrocytes and subsequently to provide their expression of Col II and PGs

[198,234,250,353]. However, Alg allows only a limited cell adhesion, because it does not

exhibit functional groups in order to be recognized by the cells [69]. Therefore, Alg scaffolds

should be modified to enhance the cell adhesion. This functionalization has been done by

grafting with RGD, which facilitates integrin recognition and binding [70,245], grafting with

Gel [134], and blending with fibrin [200], CS [201,354,355] and Col [356,357]. In addition,

chondrocytes and MSCs have been extensively considered to study the optimization of the

scaffold properties and culture conditions. Autologous chondrocytes have a capacity to be

expanded in vitro and match the host’s immunological system; however, chondrocytes tend

to lose their specific phenotype during monolayer expansion [131,234,358]. Another

approach for cartilage tissue engineering is based on MSCs, which have the ability to

differentiate into multiple tissues, including bone and cartilage [242,359]. MSCs exhibit less

donor-site morbidity, they are cost-effective, they can easily be expanded due to their high

Page 172: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

158

proliferation capacity and yield to equal or better long-term outcomes in comparison to

chondrocytes [305]. Moreover, biomolecules and growth factors, which are supplemented

into scaffolds, are crucial for the success of differentiation of MSCs and may influence the

regeneration of cartilage [198,358]. For instance, Chang et al. [360] fabricated

Col/HyA/Chondroitin sulfate (ChS) scaffolds by using the same composition as reported for

the natural cartilage. The scaffolds have shown chondrocytes adhesion and good cell

distribution, and supported the secretion of cartilaginous ECM. Coates et al. [198] have

shown that the incorporation of ChS into HyA and Alg hydrogels has a positive effect on

chondrogenesis, because it up-regulates Sox-9 mRNA and down-regulates Col I. This was

explained by the fact that ChS, which was used for cartilage repair, helps to regulate the

metabolic activity of chondrocytes [198] and stimulates the generation of PGs [361].

Therefore, ChS tends to act as a biological additive in order to regulate chondrocyte

phenotype [198] and increase cell proliferation [361,362]. This can be achieved either by

loading the molecules into the scaffolds [234,361] or by using them as a supplement in the

culture medium [358]. ChS introduced into the culture medium have been shown to increase

the production of sulfated mucopolysaccharides by cultured chondrocytes [363]. In addition,

Steinmetz et al. [364] reported that the presence of ChS in culture medium reduced the

production of Col I during the terminal differentiation of MSCs encapsulated in PEG

hydrogel, in particular when the dynamic culture was applied.

In the present work, in a first attempt ChS was incorporated into the Alg porous

scaffolds, with the aim to improve their cell response and cell activity. First, the impact of

ChS on the physical and mechanical properties of the scaffolds was investigated. Another

aim was to study the influence of ChS on the biological properties of scaffolds by in vitro

porcine chondrocytes and MSCs culturing. Two different culture conditions, including static

and dynamic cultures, were applied in the case of chondrocytes culturing Alg-foams. This

experiment was carried out with the aim to study the effects of different culture conditions on

chondrogenic differentiation. The cell viability, retention of cell phenotype and chondrogenic

differentiation were investigated.

Page 173: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

159

9.2 Materials and methods

9.2.1 Fabrication of Alg/ChS-foams

A mixture of Na-Alg (purchased from Sigma Aldrich) and chondroitin-4-sulfate A

sodium salt (ChS; purchased from Sigma Aldrich) was prepared in DI H2O with the

concentration of 3 wt/v % and the ratio of 85/15 % by wt. of Alg/ChS. The mixture was stirred

at room temperature for 2 h. 1 ml of the Alg/ChS solution was added into a 48 well-plate,

while 100 µl of 0.1 M CaCl22H2O solution was added per well. The gelation reaction was

finished after 30 min at room temperature. After this, the gel was placed into the freezer at -

20 C and it was frozen after 24 h. Then the frozen samples were lyophilized by using the

freeze-drying technique for 24 h at - 50 C under vacuum conditions (as explained in

Chapter 6). Cylindrical 3D porous Alg/ChS-foams were obtained, having dimensions of 8

mm in diameter and 8 mm in height. Then the foams were immersed in 0.5 M CaCl22H2O

solution (pH 2) for 4 h, in order to achieve ionic crosslinking. The crosslinked foams were

dried at room temperature for 24 h. Pure Alg-foams were fabricated by using the same

procedure as the one for the fabrication of Alg/ChS-foams. Thereby a solution of 3 wt/v %

Na-Alg/H2O was used.

9.2.2 Characterization and testing

(i) Porosity

The porosity of Alg/ChS-foams (P) was calculated from Eq. 9.1:

% P = [1 – (Wfoam/(Alg/ChS × Vfoam))] × 100 (9.1)

Here, Wfoam is the weight of the Alg/ChS-foam, Alg/ChS is the density of solid Alg/ChS blend

(Alg/ChS 1.02 g/cm3), and Vfoam is the volume of Alg/ChS-foam, which was determined from

the dimensions of the foam.

(ii) Microscopy

The morphology of Alg/ChS-foams was characterized by SEM (LEO 435 VP), by

using plan-view and cross-section imaging in order to observe the features of the pores.

Samples were sputter-coated and observed at an accelerating voltage of 10 kV. The pore

size of foams was evaluated from SEM images by using the free available software Image J.

Page 174: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

160

(iii) Chemical structure

The existence of ChS in the foams was identified by using ATR-FTIR (Nicolet 6700)

spectroscopy in the transmission mode. Thereby the wavenumber resolution was 4 cm-1

and

the range of 4000 - 525 cm-1

. Moreover, the existence of ChS was confirmed by using

energy-dispersive X-ray (EDX) spectroscopy, being available in the SEM instrument (SEM-

EDX; Inca analyzer, Oxford instruments).

(iv) Thermal properties

The thermal properties of Alg/ChS-foams were analyzed by using DSC (Q2000).

The measurements were performed in the temperature range of (- 50) - 200 C, by using a

heating rate of 10 C/min. The results of Alg/ChS-foams were compared to those of Alg-

foams.

(v) In vitro biodegradation

The biodegradation and water absorption of Alg/ChS- and Alg-foams were

investigated by immersion in PBS solution (pH 7.4, at 37 C) for 6 weeks. Each sample was

placed in a polystyrene bottle containing 50 ml of PBS and incubated in an orbital shaker

(IKA RS 4000i) at 37 C, using a speed of 90 rpm. The PBS solution was replaced twice a

week. At interval of immersion time, the sample was removed, blotted with filtered paper

before the weight was determined. The water absorption and weight change were

determined by using Eq. 9.2 and 9.3, respectively:

% Water absorption = [(Wwet – Wdry)/Wwet] x100 (9.2)

% Weight change = [(Wwet – Wdry)/Wdry] x100 (9.3)

Here, Wwet and Wdry is the weight after and before immersion in PBS, respectively. The

presented results are averaged over four samples.

(vi) Mechanical testing

Compression strength experiments were performed on Alg/ChS- and Alg-foams (8

mm in diameter and 8 mm in height), by using a universal testing machine (Zwick Z050).

The following parameters were used: cross-head speed of 2 mm/min, pre-load of 0.1 N and

maximum load of 50 N. The stress-strain curves were recorded in order to determine the

relevant mechanical properties, including the elastic modulus and compressive strength. The

Page 175: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

161

elastic modulus was calculated from the initial linear slope of the stress-strain curves and the

compressive strength was obtained from the maximum stress before the cell walls of the

foam were collapsed. The presented results are averaged over six samples.

In addition, a DMA (Mark IV) was performed in order to study the mechanical

response of the foams under the application of forces similar to those dominating in vivo

[313] . The foams (8 mm in diameter and 3 mm in thickness) were tested in both, the dry and

wet state, by applying a sinusoidal load with a rate defined by the frequency (in Hz). Before

testing in the wet state, the foams were immersed in PBS solution until reaching equilibrium

conditions. The foams were tested in the compression mode as a function of the frequency

cycle. The used frequency was in the range of 0.1 - 10 Hz, according to the range of a typical

skeletal movement in vivo [313]. The tests were performed under the maximum strain

amplitude of 1 %. The viscoelastic properties of the foams - the storage modulus (E’) and

loss factor (tan = E’’/E’) - are presented by using the log scale and plotting versus the log

scale of frequency.

9.2.3 Release of ChS

The relative ChS content in the foams was determined according to the modified

Dische’s carbazole reaction, which has been modified by Bitter and Muir [365]. In detail,

each of Alg/ChS- and Alg-foams (8 mm in diameter and 3 mm in thickness) was immersed in

5 ml of PBS solution (pH 7.4, 37 C) in a glass vial. At each time point of immersion, 2 ml

of PBS solution was taken out and replaced by the same amount of fresh PBS. Afterwards,

0.2 ml of the obtained solution was given for reaction with 1 ml of 0.025 M sodium

tetraborate decahydrate (Borax; ACS reagent 99.5 %, Sigma-Aldrich) in concentrated

sulfuric acid (H2SO4, Sigma-aldrich), with the aim to initiate degradation of ChS. The mixture

was then cooled down to room temperature and heated up in a boiling water bath for 10 min.

After that, the mixture was again cooled down to room temperature. 40 µl of 0.125 wt/v %

carbazole ( 95 % GC, Sigma Aldrich) in EtOH was added into the mixture and heated up in

the boiling water bath for further 15 min, until pink chromogen was formed. Finally, the

absorbance of the pink colored complex was measured at 530 nm by using the UV

spectrophotometer (Specord 40; Analytikjena, Germany). The amount of the released ChS

Page 176: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

162

was calculated by using a calibration curve of known ChS concentrations (0 - 120 µg/ml), as

shown in equation 9.4:

Absorbance = (0.0018 x ChS concentration) – 0.00497, R2 = 0.99. (9.4)

9.2.4 In vitro culturing of primary porcine chondrocytes and human MSCs

This part of the study was carried out in Laboratory for Experimental Trauma

Surgery, Department of Orthopedic, Trauma and Reconstructive Surgery Charité-University

Medicine Berlin, Campus Benjamin Franklin.

(i) Isolation of porcine chondrocytes

The chondrocytes were isolated according to the method reported by Lohan et al.

[131]. Briefly, porcine cartilage was harvested from the articular cartilage in the knee joint of

pigs (3 - 6 months old). The removed cartilage samples were minced into 1 mm slices. Then,

the samples were enzymatically digested with 0.4 % pronase (7 U/mg, Roche, Basel,

Switzerland) and diluted in Ham’s F-12/Dulbecco’s modified Eagle’s (DMEM) medium (1:1)

(Biochrom AG, Berlin, Germany) for 1 h at 37 C. Then the samples were subsequently

digested with 0.2 wt/v % collagenase ( 0.1 U/mg, SERVA Electrophoresis GmbH,

Heidelberg, Germany) and diluted in the growth medium for 16 h at 37 C. The isolated

chondrocytes were suspended in the growth medium [Ham’s F-12/DMEM (1:1) containing 1

ml/100 ml fetal calf serum (FCS) (Biochrom AG), 25 µg/ml ascorbic acid (Sigma-Aldrich), 50

IU/ml streptomycin, 50 IU/ml penicillin, 2.5 µg/ml amphotericin B and 1 ml/100 ml essential

amino acids (all from Biochrom AG)]. The cell suspension was seeded at a density of 2.8 ×

104 cells/cm

2 in culture flasks.

(ii) Isolation of human bone marrow derived MSCs

Human MSCs were isolated from human femoral head spongiosa (obtained from

patients undergoing joint replacement surgeries of the hip joints) using density gradient

centrifugation with a separating solution, biocoll (Biochrom AG, Berlin, Germany). The

spongiosa of a femoral head was minced and pressed through a cell sieve. Bone spongiosa

fragments were removed and the liquid rest was pressed through a 140 μm pore diameter

filter membrane. To remove the remnants of the particles the isolated cell suspension was

washed with PBS and centrifuged at 200 g in 4 °C. The purified pellet was mixed with the

Page 177: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

163

biocoll solution (Biochrom AG) and centrifuged at 200 g in 4°C. After 20 min, the interphase

containing MSCs was extracted, washed with PBS and centrifuged at 200 g in 4 °C.

Subsequently, MSCs were re-suspended in stem cell growth medium [DMEM 51 ml/100 ml

(Biochrom AG, Berlin, Germany) containing selenium (5 ng/ml; Aldrich), transferring (5

µg/ml; Sigma), linoleic acid (4.7 µg/ml; Sigma), insulin (5 µg/ml; Sigma), ascorbic acid

(1µg/ml; Sigma), dexamethasone (1 µg/ml; Sigma D4902), MCDB 201 with L-glutamine

solution (34 ml/100 ml; Sigma), Foetal calf serum (FCS; 15 ml/100 ml; Biochrom),

streptomycin (50 IU/ml) and penicillin (50 IU/ml)] and seeded in culture flasks (Cell plus

culture flask, Sarstedt, Nümbrecht, Germany). The cultivation proceeded at 37 °C, 90% air

humidity and 5% CO2. The growth medium was changed every 2 - 3 days.

(iii) Static and dynamic cultures

Two types of cells, namely primary porcine articular chondrocytes and human

MSCs, were used. The Alg/ChS- and Alg-foams (both: 8 mm in diameter and 3 mm in

thickness) were sterilized by using a plasma treatment, respectively. The sterilized Alg-

foams were then pre-conditioned for 72 h, while the sterilized Alg/ChS-foams were pre-

conditioned for 24 h, in the basal medium. After that, the foams were soaked in a

suspension of either chondrocytes or MSCs in the basal medium (10 million cells of passage

2 - 3 of chondrocytes and passage 4 - 5 of MSCs). The cell-seeded foams were then placed

in an incubator at 37 C and 5 % CO2. The medium was changed 3 times a week.

The induction of in vitro chondrogenesis of MSCs seeded on the foams was

evaluated by culturing with the presence of TGF-1. MSCs-seeded Alg- and Alg/ChS-foams

were cultured in chondrogenic medium [DMEM with 3.7 g/l NaHCO3 and 4.5 g/l glucose

(Biochrom AG, T041-01) containing 10 µg/ml L-glutamine (Biochrom AG, K0282), 25 µg/ml

HEPES (Biochrom AG, L1613), 10 µg/ml sodium pyruvate (Sigma, S8636), 0.1 µl/ml

dexamethasone (Sigma, D4902), 1.7 µl/ml ascorbic acid (Sigma, A8960), 1 µg/ml prolin

(Sigma, P8865), 1 µl/ml ITS+1 (Sigma, I2521), 50 IU/ml streptomycin, 50 IU/ml penicillin and

10 ng/ml TGF-1 (Petro Tech)], which were placed in an incubator at 37 C and 5 % CO2.

The medium was changed 3 times a week.

Page 178: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

164

During dynamic culture conditions the growth medium and the material degradation

products have to be exchanged after 2-3 days. However, dynamic culturing provides a more

homogeneous cell distribution, compared to static culturing [131,366]. Therefore, two

different culture conditions were comparatively studied in the case of chondrocytes culturing

on the Alg-foams. The dynamic culturing was performed by stirring the cell suspension in a

bioreactor filter tube (TPP, Switzerland) using an orbital shaker at 15 °C and the speed of 12

rpm (digital type roller Struart SRT9D, Bibby Scientific, USA). The medium was changed 3

times a week.

(iv) Cell viability

Cells-cultured for 7 and 14 days in Alg- and Alg/ChS-foams were washed with PBS

and incubated in fluorescein diacetate (FDA; Sigma-Aldrich) (3 µg/ml dissolved in acetone

and diluted 1:1000 in PBS) for 15 min at 37 C. Then, the samples were rinsed three times

with PBS and counterstained with a propidium iodine (PI) solution (1 mg/ml dissolved in PBS

and diluted 1:100 in PBS) (Sigma-Aldrich) for 1 min under dark conditions at room

temperature. The green and/or red fluorescence was visualized by using fluorescence

microscopy (Axioskop 40, Carl Zeiss, Jena, Germany) or confocal laser microscopy (TCS

SPE II, Leica Microsystems, Wetzlar, Germany). Fluorescence images were taken using an

XC30 camera (Olympus Soft Imaging Solution GmbH, Germany).

(v) Histology

For hematoxylin eosin (HE) staining, the samples were stained with a Harris

hematoxylin solution (Sigma-Aldrich) for 4 min, rinsed with water and counterstained in eosin

(Carl Roth GmbH, Karlsruhe, Germany) for 4 min.

For alcian blue (AB) staining, the samples were incubated in a 1 % acetic acid for 3

min and then stained with 1 % AB (Karl Roth, Karlsruhe, Germany) for 30 min. After that, the

stained samples were rinsed with 3 % acetic acid and counterstained with a nuclear fast red

aluminum sulfate solution (Carl Roth) for cell nuclei staining for 5 min. Finally, the samples

were rinsed with water and subsequently dehydrated in an ascending EtOH series. Then,

the samples were observed histologically by using light microscopy (Axioskop 40, Carl

Zeiss) and images were taken using a XC30 camera.

Page 179: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

165

(vi) Immunohistology

The samples were rinsed with Tris buffered saline (TBS: 0.05 M Tris, 0.015 M NaCl,

pH 7.6) and incubated with 5 mg/ml pronase (7 U/mg; Roche, Basel, Switzerland, diluted in

TBS) for 5 min at 37 C. The samples were subsequently rinsed with TBS and blocked with

a protease-free donkey serum (5 % diluted in TBS) for 30 min at room temperature. Then

the samples were rinsed again with TBS and incubated with the polyclonal rabbit anti-type II

or anti-type I collagen antibodies (27.5 µg/ml) (both: Acris antibodies, Herford, Germany), by

using a humidifier chamber overnight at 4 C. After that, the samples were rinsed with TBS

and incubated with a donkey-anti-rabbit-Alexa-Fluor® 488 (10 mg/ml, Invitrogen) secondary

antibody for 30 min at room temperature. Negative controls included omitting the primary

antibody during the staining procedure. Finally, cell nuclei were counterstained with DAPI

(0.1 µg/ml) (Roche). The labeled samples were rinsed several times with TBS, embedded

with Fluoromount G (Southern Biotech, Biozol Diagnostica, Birmingham, USA) and

examined using fluorescence microscopy (Axioskop 40).

9.2.5 Statistical analysis

The data were analyzed by using the one-way analysis of variance (ANOVA) and

the Turkey’s multiple-comparison test in order to determine statistical differences. The

confidence interval of p = 0.05 was used for all analysis. The results are reported as mean

values, with SD.

9.3 Results and discussion

9.3.1 Characterization of Alg/ChS-foams

(i) Morphology

The plan-view SEM image of an Alg/ChS-foam in Fig. 9.1 (A) shows a uniform

distribution of pores, while the cross-section SEM image in Fig. 9.1 (B) reveals a columnar

structure of pores with ladders, which was generated by the same mechanism as discussed

in the case of pure Alg-foams in Chapter 6. Alg/ChS-foams exhibited slightly decreased

porosity ( 93 %) compared to pure Alg-foams (porosity 95 %). The pore size of Alg/ChS-

foams (197 ± 61 µm) was slightly smaller than the one of Alg-foams (237 ± 48 µm) (Fig. 9.3

Page 180: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

166

(C)). In addition, micropores ( 50 µm) were observed on the pore walls of the Alg/ChS-foam

(Fig. 9.1 (B)). This result confirms that the Alg/ChS-foams exhibited a combination of micro-

and macro-pores. Even though only 15 wt % of ChS was incorporated into the foam, ChS

may influence the molecular arrangement of the Alg network. According to the literature

[367,368], the reptation model involves an arrangement of two different anionic

polysaccharides, in which two different types of polymer cannot intersect each other. Alg,

which is the main component of the foam, forms an egg-box network by ionic crosslinking. A

Ca ion interacts with four carboxylate groups of the Alg chains. The crosslinked Alg network

is ordered in a tube-like structure, as shown in Fig. 9.3 (D). The ChS molecules are confined

into the Alg network. Therefore, in a proper environment the ChS molecules move inside the

Alg-network like a snake [367,368]. Fajardo et al. [368] have reported that the ChS

molecules maintain inside the tubular structure of Alg under high acidic conditions (pH 2),

get mobilized and further diffuse through the network, causing an increase of the pH (pH

5).

Page 181: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

167

Figure 9. 1 SEM images of Alg/ChS-foams in (A) plan-view and (B) cross-section; (C)

shows a histogram of the pore size distribution of Alg/ChS-foams and (D) is a scheme

showing the reptation model in the case of Alg/ChS blend, according to [367,368].

(ii) Chemical structure

In order to confirm the incorporation of ChS into the Alg-foams, the chemical

structure of the Alg/ChS-foams was investigated by using ATR-FTIR spectroscopy. The

ATR-FTIR spectrum of an Alg/ChS foam in Fig. 9.2 (A) exhibited peaks at identical positions

as observed for the Alg-foam. The characteristic peaks of Alg (marked by ▲) at 1600 and

1427 cm-1

can be attributed to vibrations of -C=O and -COOH groups [368,369]. The

characteristic peaks of ChS (marked by ■) at 1605 and 1558 cm-1

can be attributed to -C=O

and -N-H- stretching, respectively [370]. The peak of ChS at 922 cm-1

was attributed to the

vibration of C-O-C groups, while the peak at 853 cm-1

assigned the vibration of C-O-S

groups [370], which were both not present in the spectrum of Alg/ChS-foams. This can be

explained by the low content of ChS incorporated into the foam. Only a broad peak at 1258

cm-1

was detected for the Alg/ChS-foam, which can be attributed to the -S=O stretching

Page 182: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

168

vibration of sulfate groups in ChS [368,370,371]. Therefore, it can be concluded that the

incorporation of ChS into the foam was successful.

The EDX results in Fig. 9.2 (B) additionally confirm the presence of ChS in the foam.

The peaks of sulfur (S) were detected in the case of the Alg/ChS-foam, while they were not

pronounced in the case of the pure Alg-foam. The FTIR and EDX results confirm the

presence of ChS within the Alg-foam. In addition, the ionic crosslinking in the foams was

confirmed by strong peaks of calcium (Ca), detected by EDX (see Fig. 9.2 (B)).

Figure 9. 2 (A) ATR-FTIR spectrum and (B) EDX spectra of the Alg/ChS-foam in

comparison to the spectra of pure Alg- and pure ChS-foams, which both results confirm the

existence of ChS in the foam.

(iii) Thermal properties

The thermal properties of the foams were characterized by DSC. Fig. 9.3 (A) shows

the heat transition in endothermic direction, indicating the melting of the material [372,373].

The melting temperature (Tm) of the Alg/ChS-foam is at 83.4 °C, which was lower than the

Tm of the Alg-foam, being at 96.59 °C. At this point, the polymer molecules of the Alg/ChS-

foam needed a higher temperature for melting, compared to the Alg-foam. Regarding

molecular point of view, the Alg/ChS-foams exhibited a lower chain mobility [374–376],

Page 183: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

169

compared to Alg-foams. The reason for this behavior is the presence of ChS, which may

inhibit the interaction between Alg and Ca ions, leading to less intermolecular interactions,

compared to a network without ChS. Nevertheless, this hypothesis has to be proved by the

determination of the crosslinking degree.

Fig. 9.3 (B) shows the endothermic transition of the foams, obtained from the 2nd

heating cycle. This transition results from the energy absorption, required for the change of

the glassy state (hard, brittle) to the rubbery state (soft, flexible), also known as the glass

transition [373]. The Tg of Alg/ChS-foams (29.9 C) was slightly higher than the Tg of Alg-

foams (24.1 °C). The increase in Tg can be the result from the main chain rigidity and

crosslinking. Hence, this result suggests a possible bulk rigidity and brittleness, caused by

the incorporation of ChS.

In summary, even though the incorporation of ChS is believed to influence the

thermal properties of the foams, it can be ensured that the Alg/ChS-foams are not able to be

melted and decomposed under in vitro culture conditions at 37 C. In addition, the results of

the Tg confirm that at 37 C (culture conditions) the foams only become softer, while the

structural integrity is still maintained.

Figure 9. 3 DSC thermograms: (A) the 1st heating and (B) the 2

nd heating cycle of Alg/ChS-

and Alg-foams.

Page 184: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

170

(iv) Water absorption

The water absorption of scaffolds used for cartilage regeneration is an important

factor related to the local pressure resistance of the structure [313,377]. In general, the

negative charges of GAGs in the cartilaginous ECM mainly cause the absorption of water

molecules, which leads to a reduced friction at the osteochondral joint [361,362]. It has been

reported that articular cartilage is composed of 80 % water with respect to the total wet

weight [6]. In the present work, a water absorption of 82 ± 2 % was observed for the Alg-

foams (see Fig. 9.4), which is similar to the reference value of 80 % [43]. The water

absorption of Alg/ChS-foams was measured to be 88 ± 2 % and thus it is higher than the

water absorption value of the Alg-foams. The higher water absorption of the Alg/ChS-foams

is caused by the negative charges of ChS, which enhance the capacity of a foam to absorb

water [362].

In summary, both, the Alg/ChS- and Alg-foams did not exhibit a significant difference

of water absorption from the reference value reported in the literature [43]. Therefore, in

terms of high water absorption, both types of foams are suitable for cartilage regeneration.

Figure 9. 4 Water absorption in % of Alg- and Alg/ChS-foams, in comparison to the water

absorption of natural articular cartilage (*), as referenced in the literature [6].

Page 185: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

171

(v) Biodegradation behavior

In addition to the water absorption, the structural stability of scaffolds has to be

maintained until new tissue grows. Therefore, in order to evaluate the structural stability of

the scaffolds, a biodegradation study was performed. The weight change of the foams,

which is caused by water absorption, was determined during immersion in PBS. The weight

change of Alg/ChS-foams in PBS ( 750 %) was slightly higher than the weight change of

Alg-foams ( 650 %), while the trend of the weight change in both cases is the same (Fig.

9.5). In detail, the foams initially absorb a high amount of the PBS solution and start to

maintain an increased weight after 1 day of immersion. Importantly, after 3 weeks of

immersion, the weight change of the Alg/ChS-foams slightly decreases, while the weight

change of the Alg-foams is maintained until the end of the experiment (6 weeks). It can be

concluded that the Alg/ChS-foams start to lose their stability after 3 weeks of immersion in

PBS, which might be a sign of decomposition.

Figure 9. 5 Degradation profile of Alg/ChS- and Alg-foams in PBS, evaluated by % weight

change with respect to the immersion time.

Page 186: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

172

(vii) Mechanical properties

Representative compressive stress-strain curves of the Alg/ChS- and Alg-foam are

shown in Fig. 9.6 (A). The mechanical properties (elastic modulus and compressive

strength) of the Alg/ChS-foams are significantly higher than those of the Alg-foams (* p

0.05) (Fig. 9.6 (B)). It is believed that the incorporation of ChS into the Alg-foam reinforces

the Alg matrix. According to the reptation model shown in Fig. 9.1 (D), the molecules of ChS,

which are confined in the Alg network, decrease the void space in the network. The

incorporation of ChS significantly increases the elastic modulus of the Alg-foams from 0.22 ±

0.09 to 0.7 ± 0.1 MPa and increases the strength from 0.11 ± 0.01 to 0.8 ± 0.2 MPa.

Especially, the elastic modulus of the Alg/ChS-foams reaches the upper reference value for

healthy human cartilage, which has been reported to be 0.24 - 0.85 MPa [6,12,73,76,141–

143]. Alg/ChS-foams, which exhibit a high elastic modulus (high stiffness), are therefore

suitable in load bearing applications, i.e. cartilage and bone tissue engineering.

Fig. 9.7 presents the storage modulus (E’) and tan of the dry and wet foams as a

function of frequency. It is obvious that the presence of ChS increases the viscoelastic

properties of the dry foams. The E’ values of the dry foams are higher than the values of the

wet foams, for both, Alg- and Alg/ChS-foams. This behavior is caused by the entrapped

water in the wet foams, which acts as a plasticizer and decreases the stiffness of the foams

[313,319,332]. In addition, the E’ values of the wet Alg/ChS-foams are lower than the values

of the Alg-foams. This result is caused by the higher PBS absorption of the Alg/ChS-foams

(see Fig. 9.5) and the incorporation of negatively charged ChS. The results correlate with tan

results, where both, the wet Alg- and Alg/ChS-foams exhibit higher values compared to the

dry foams. Since in the wet foams the pores are filled with PBS, it is difficult to deform them

by applying a dynamic force. Moreover, the Alg/ChS-foams exhibit higher tan values than

the Alg-foams in both, the wet and dry states. This behavior can be explained by the theory

of dynamic mechanical properties of polymers that the loss factor is defined as the loss

modulus (E’’) divided by the storage modulus (E’): tan = E’’/E’. Since the absorption of

water decreases the stiffness (E’ decreases), tan increases at constant E’’, which means

that the internal friction of the material increases. In addition, the E’ values of both, the Alg-

Page 187: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

173

and Alg/ChS-foams, obtained in the wet state are in the range of values reported for the

cartilage tissue (0.1 - 1 MPa, 0.1 - 10 Hz) [313]. These results confirm that the viscoelastic

properties of Alg- and Alg/ChS-foams (either in dry or wet state), do not change with the

variation of frequency and time, which indicates that no deformation occurs during oscillatory

testing [343,378].

Figure 9. 6 (A) Representative compressive stress-strain curves of the Alg/ChS- and Alg-

foam; (B) mechanical properties of Alg/ChS and Alg-foams.

Figure 9. 7 Dynamic mechanical analysis of Alg/ChS- and Alg-foams: storage modulus (E‘)

and tan as a function of frequency; note the logarithmic scaling.

Page 188: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

174

9.3.2 Release profile of ChS molecules

The release of ChS, which is incorporated into the foams, was investigated by

immersion in PBS solution (pH 7.4 at 37 C). By using carbazole reaction described by

Dische [379], a pink colored complex of ChS can be detected at 530 nm wavelength by

using UV-VIS. Fig. 9.8 represents the cumulative release of ChS with respect to the

immersion time in PBS for 14 days. The ChS was gradually released with increasing time in

PBS. However, after day 7 of immersion, no further ChS was released. As confirmed by

immersion for 14 days, the content of the released ChS was not changed compared to day 7

of immersion. This result confirms that at a pH 5 the molecules of the ChS were released

from the tubular network of Alg, as in agreement with [368]. In addition, the ChS release rate

observed in the present work can be used as a guideline for further in vitro cell culture

studies. The ChS delivered to the culture medium is aimed to act as a biological cue, in

order to up-regulate the chondrogenic differentiation.

Figure 9. 8 Cumulative release of ChS from Alg/ChS-foams immersed in the PBS solution

(pH 7.4, 37 °C) measured by using carbazole reaction, as described in Section 9.2.3.

Page 189: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

175

9.3.3 The influence of culturing conditions on the primary porcine chondrocytes

activity

(i) Cell viability

The influence of two types of culture conditions (static and dynamic) on the behavior

of primary porcine chondrocytes-seeded Alg-foams was investigated. The aim was to study

the influence of the culture conditions on the cell activity, in particular the cell viability and

production of cartilaginous ECM. The cell viability was observed in both, the static and

dynamic cultures (see Fig. 9.9 (A) and (B), respectively). In addition, aggregates of

chondrocytes were seen to form in both cases (static and dynamic cultures) after culturing

for 7 days, as exemplarily depicted by white arrows in Fig. 9.9. These observations confirm

the biocompatibility of the scaffolds. The two seeding conditions did not provide a significant

difference in terms of cell viability.

Figure 9. 9 The viability of primary porcine chondrocytes on Alg-foams after 7 days of

culture: (A) static culturing and (B) dynamic culturing (on a rotatory device, 12 rpm at 15 C).

Viable cells are green, dead cells appear red.

Page 190: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

176

(ii) Histological evaluation

The influence of the culturing systems on the production of cartilaginous ECM was

investigated. The results are shown in Fig. 9.10, where the histological evaluation of HE and

AB is depicted. Formation of cell clusters was observed histologically in both cases, static

and dynamic culture conditions (see Fig. 9.10 (A-D)). The results confirm that chondrocytes

still retained a round phenotype over 7 days of culture. In addition, the ECM was formed,

which surrounds the chondrocytes (as depicted by the arrows in Fig. 9.10 (B)), indicating the

deposition of cartilage PGs after 7 days of culture.

Figure 9. 10 Histological evaluation of HE and AB staining of porcine chondrocytes-seeded

Alg-foams after 7 days of static and dynamic culturing. Using AB staining, the foam stains

unspecifically blue. Cell nuclei are violet, sulfated GAGs within the cell clusters are

characterized by a faint blue staining.

Page 191: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

177

(iii) Collagen immunolabeling

The production and deposition of cartilage-specific Col II by chondrocyte-seeded on

Alg-foams was confirmed by the immunohistological evaluation (Fig. 9.11 (B and E)). In both

cases, static and dynamic culture conditions, the cell clusters produced Col II (green-

stained), which is a marker of cartilaginous ECM. Chondrocytes produced a minimum

amount of Col I under both culture conditions (Fig. 9.11 (C and F)).

In summary, the culture conditions (static and dynamic) did not provide significant

differences in terms of cell viability and the production of cartilaginous ECM. The

biocompatibility of the Alg-foams was confirmed by these results. In addition, the Alg-foams

induced the aggregation of chondrocytes and promoted the production of cartilaginous ECM

in both static and dynamic culture conditions. Hence, the results confirmed that given

porosity and pore size of the present Alg-foams exhibited a proper microenvironment in

order to support chondrocyte infiltration and the growth of chondrocytes, and provided an

attractive matrix for cartilage regeneration.

Figure 9. 11 Col II and Col I immunolabeling of porcine chondrocytes-seeded Alg-foams

after culturing for 7 days, observed by confocal microscopy. Cell nuclei are stained in blue,

Col is stained in green color.

Page 192: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

178

9.3.4 The influence of ChS molecules on chondrogenic differentiation of

chondrocytes and MSCs

(i) Cell viability

Since the different culture conditions (i.e. static and dynamic) did not lead to

differences in the growth of chondrocytes, MSCs-seeded scaffolds were cultured under

static conditions. In a first set of experiments, MSCs were simply cultured under non-

chondrogenic conditions for 7 days. By using the live-dead cell assay (see Fig. 9.12), it was

found that mostly viable cells were presented in both, Alg/ChS- and Alg-foams, after 7 days

of culture. MSCs were seen to form clusters (as depicted by the arrows in Fig. 9.12), as

observed in the case of porcine chondrocytes, which resemble pre-cartilage condensation of

MSCs [359]. The cell clusters showed a homogeneous distribution in the entire scaffold (Fig.

9.12). These results confirm the biocompatibility of Alg/ChS-foams. In addition, the highly

porous foams with pore size in the range of 50 - 300 µm support the formation of cell

clusters. However, the presence of ChS did not show any significant improvement of cell

adhesion onto Alg-foams, as originally expected.

Figure 9. 12 Cell viability (FDA/PI live-dead assay) of MSCs-seeded (A, C) Alg- and (B, D)

Alg/ChS-foams after 7 days of culture. Viable cells are green, dead cells appear red.

Page 193: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

179

In the case of chondrocytes culturing, mostly viable cells were obvious after 14 days

in culture. In particular, it is likely that cell adhesion was improved by the incorporation of

ChS into Alg-foam after culturing with chondrocytes for 14 days. Fig. 9.13 shows the viability

of chondrocytes seeded on the foams (as indicated by green color). After 14 days in culture,

Alg/ChS-foams were superior in supporting chondrocyte adhesion compared to Alg-foams.

In addition, chondrocytes were seen to spread on the Alg/ChS-foam (Fig. 9.13 (B)), while

chondrocytes seeded on Alg-foam tended to form clusters inside the pores of the foam after

7 days (as mentioned previously in Fig. 9.9 (A)) and the clusters were maintained even after

14 days in culture (Fig. 9.13 (A)). Therefore, this qualitative observation indicates that the

incorporation of ChS into Alg-foam can enhance the adhesion of chondrocytes after culturing

for 14 days.

Figure 9. 13 Cell viability (FDA/PI live-dead assay) of porcine chondrocytes-seeded (A) Alg-

and (B) Alg/ChS-foams after 14 days of culture. Viable cells are green, dead cells appear

red.

Page 194: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

180

(ii) Histological evaluation

The histological evaluation (Fig. 9.14 (A and B)) shows the formation and deposition

of abundant ECM in both, the Alg/ChS- and Alg-foams, after 7 days of culture. The ECM

was presented in the cell clusters. In addition, the cartilage-specific sulfated PGs containing

ECM was detected by AB assay as a faint blue staining incorporated in the cell clusters (Fig.

9.14 (C and D)). These results confirm the high potential of chondrogenic differentiation of

MSCs in both scaffolds. Moreover, many MSCs acquired a mostly rounded phenotype after

7 days in culture, indicating that both scaffolds might allow further chondrogenic

differentiation of MSCs. Cells revealed no features of hypertrophy such as enlarged lacunes

and bullous cells.

Figure 9. 14 Histological evaluation (HE and AB staining) of MSCs-seeded Alg- and

Alg/ChS-foams after 7 days of culture (without chondrogenic induction). For alcian blue

staining, sulfated cartilage PGs are stained in blue, cell nuclei are stained in red and the

foams appear blue.

Page 195: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

181

(iii) Collagen immunolabeling

MSCs-seeded Alg/ChS- and Alg-foams revealed clearly Col II immumolabeling after

7 days of culture (Fig. 9.15 (A and C)), in comparison to the negative control (Fig. 9.15 (A)).

It is obvious that Col II, which is indicated by green staining, was surrounding the cell nuclei

(stained in blue). This phenomenon is proposed as the onset of chondrogenesis of MSCs

[359]. In particular, the culturing of MSCs on Alg/ChS-foams produces an intensive Col II

(see Fig. 9.15 (C) and Fig. 9.16 (B)). However, Col I could also be detected in both,

Alg/ChS- and Alg- foams (see Fig. 9.15 (E and F)) and supporting result in Fig. 9.16 (C and

D)). A comparable result was observed in the case of chondrocytes, cultured on Alg-foams.

It is likely that the incorporation of ChS plays an important role in the chondrogenic

differentiation of MSCs, which proceeds by the aggregation of MSCs. According to the

literature, the incorporation of ChS into either porous scaffolds or hydrogels enhances both,

the chondrogenic gene expression and cartilage-specific matrix production [359,364,380]. In

addition, ChS has been found to down-regulate the expression of Col X [359,364,380]. In

order to prove this hypothesis, a longer experiment (i.e. 14 days of culture) is desired and

quantitatively biochemical assays are required to confirm the influence of ChS on the

production of Col II and PGs.

Page 196: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

182

Figure 9. 15 Col II and Col I immunolabeling of MSCs-seeded Alg- and Alg/ChS-foams after

7 days of culture, observed by fluorescence microscopy. Col is stained in green and cell

nuclei are stained in blue.

Figure 9. 16 Col II and Col I immunolabeling of MSCs-seeded Alg- and Alg/ChS-foams after

7 days of culture, observed by confocal microscopy. Collagen is stained in green and cell

nuclei are stained in blue.

Page 197: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

183

9.3.5 The influence of chondrogenic induction (TGF-1) on the activity of MSCs

MSCs-seeded Alg- and Alg/ChS-foams were cultured with the induction of TGF-1

growth factor in order to evaluate its effect on chondrogenic differentiation of MSCs.

According to the literature, the TGF- family plays an important role in cartilage regeneration

[246,381,382]. In vitro studies have been demonstrated that TGF-1, which serves as a

chondroinductive factor, induced early Col II expression and PGs accumulation during

chondrogenesis [246,381].

(i) Cell viability

The spreading cells were particularly detected on Alg/ChS-foams after 17 days of

non-induced culture (Fig. 9.17 (B)), while cell clusters were observed on Alg-foam (Fig. 9.17

(A)). Viable cells were mostly visible on both, Alg- and Alg/ChS-foams, appearing in green

color after 14 days of culture with the presence of TGF-1 (Fig. 9.17 (C and D)). In addition,

cell aggregation was detected in higher density with the presence of TGF-1, particularly on

Alg/ChS-foam (Fig. 9.17 (D)), compared with the absence of TGF-1.

Figure 9. 17 Cell viability (FDA/PI live-dead assay) of MSCs-seeded Alg- and Alg/ChS-

foams after culture with and without chondrogenic induction (+TGF-1). Viable cells are

green, dead cells appear red.

Page 198: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

184

(ii) Collagen immunolabelling

With the presence of TGF-1, a superior Col II (green color) surrounding cell nuclei

(blue color) was detected in both MSCs-seeded Alg- and Alg/ChS-foams (Fig. 9.18 (C and

D), respectively), compared to the absence of TGF-1 (Fig. 9.18 (A and B)). It is likely that

the induction of TGF-1 to the culture medium stimulated the chondrogenic differentiation of

MSCs, leading to the production of specific Col II in both Alg- and Alg/ChS-foams.

Figure 9. 18 Col II and Col I immunolabelling of MSCs-seeded Alg- and Alg/ChS-foams after

14 days of culture with and without chondrogenic induction, observed by confocal

microscopy. Col II is stained in green and cell nuclei are stained in blue.

Page 199: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

185

9.4 Conclusions

The Alg- and Alg/ChS-foams provided a good biocompatibility with porcine

chondrocytes and MSCs. The foams fabricated in the present work are confirmed to be a

proper environment for the support of both, porcine chondrocytes and MSCs. Chondrocytes

and MSCs culturing on foams produced abundant Col II and PGs after static culturing for 14

days. Some Col I was also produced. In chondrocytes cultures, Col I appeared to be lower

expressed than Col II. However, in MSCs cultures, substantial amount of Col I was detected

as unexpected in vitro. In particular, the incorporation of ChS into Alg-foams enhanced

chondrocytes and MSCs adhesion, and promoted better spreading of cells compared to Alg-

foams without incorporation of ChS. It can be concluded that the ChS, which was

incorporated into the foams, plays a significant role for the biological properties of the foams,

acting on the response of chondrocytes and MSCs. The results confirmed that an

enhancement of cell adhesion occurs with addition of ChS. In addition, the influence of ChS

on the production of cartilaginous ECM as well as on TGF-1-induced chondrogenesis of

MSCs-seeded foams must be further investigated by quantitative biochemical assays. Also,

the chondroinductive features of Alg/ChS-foams must be confirmed for applications of the

foams in cartilage regeneration and as the cartilage compartment of bilayered scaffolds.

Page 200: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

186

Page 201: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

CHAPTER 10

Summary and Future perspectives

Multilayered scaffolds suitable for interface tissue engineering, i.e. osteochondral

regeneration, were fabricated and intensively discussed in the dissertation. Their

architecture, porous structure, physico-chemical and mechanical properties, and biological

properties were all comprehensively considered.

Bioglass®-based foams were chosen as the scaffold material for the subchondral

bone part. 3D Bioglass-based porous scaffolds exhibiting architecture and porous structure

similar to those of natural cancellous bone were obtained by using foam replication

technique. Moreover, the mechanical strength and the structural stability of the scaffolds

could be improved by coating with biodegradable polymers. Different polymer coatings,

including alginate (Alg), gelatin (Gel), PDLLA and PHBHHx coatings, were investigated

which lead to the enhancement of elastic modulus and compressive strength compared to

uncoated Bioglass-based scaffolds. In addition, such coated scaffolds maintained the

bioactivity of 45S5 Bioglass, which was evidenced by the formation of HA after immersion

in SBF solution. Therefore, all biodegradable polymer coated Bioglass®-based composite

scaffolds developed in the present study can be an appropriate candidate for use in bone

regeneration. In addition, the polymer coated Bioglass-based scaffolds can be used as a

drug/biomolecule carrier, i.e. antibiotic delivery, in bone tissue engineering applications.

Multifunctional scaffolds based on TCH-loaded polymer layers coated Bioglass-based

scaffolds exhibited improved mechanical strength compared to uncoated scaffolds as well as

controlled drug release over 14 days after immersion in PBS. The biological properties of Alg

coated Bioglass-based scaffolds only have been evaluated by culturing with osteoblasts

(MG-63) in order to confirm their biocompatibility, and ability to support cell metabolic activity

and ALP production. When compared to uncoated Bioglass-based scaffolds and RGD-

Page 202: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

188

modified Alg coated Bioglass-based scaffolds, Alg coated scaffolds exhibited good

biocompatibility with MG-63 cells, and promoted the cell growth and osteoblastic activity.

Concerning the cartilage phase in osteochondral tissue, Alg was highly interested

because its chemical structure is similar to hyaluronic acid (HyA), considering that HyA is the

main component in cartilaginous ECM. Columnar porous structure of 3D Alg-foams was

successfully fabricated in order to support the migration and arrangement of cells, and

subsequently organization of new tissue by optimizing polymer concentrations and freeze-

drying conditions. Optimum 3 wt/v % Alg-foams exhibited pore size in the range of 125 - 325

µm, which is suitable for supporting chondrocytes seeding and migration. In addition, the

foams were able to absorb water in the same range as in native cartilage ( 80 %). The

mechanical strength and the structural stability of the foams were improved by the

application of ionic crosslinking. The elastic modulus and compressive strength of the foams

at 0.220 ± 0.009 MPa and 0.14 ± 0.02 MPa, respectively, were in the range of targeted

values of native cartilage (elastic modulus at 0.24 - 0.85 MPa and compressive strength at

0.01 - 3 MPa). Importantly, the foams were non-mineralized, meaning that they are not able

to induce bone formation in contact with body fluids, which is a requirement for cartilage

regeneration. According to the diagram summarizing the related issues in cartilage tissue

engineering approach in Fig. 10.1, the porous foams fabricated in this study have achieved

most of scaffold related criteria. However, the foams still lack cell adhesion, which negatively

affects cell proliferation and differentiation. Therefore, the Alg-foams were modified for the

first time by incorporating a biological cue, i.e. chondroitin sulfate (ChS), in order to improve

cell adhesion and cell behavior. ChS is one of natural glycosaminoglycans in cartilage,

which has a function to stimulate chondrocytes’ metabolism, inducing the synthesis of type II

collagen (Col II) and proteoglycans (PGs). Alg/ChS-foams supported either chondrocytes or

MSCs, and promoted cell proliferation and differentiation. The expression of Col II and PGs

of chondrocytes- and MSCs-seeded on Alg/ChS-foams was characterized as a marker of

cartilaginous regeneration. These results fulfilled the major cell related requirement (Fig.

10.1). In addition, Alg/ChS-foams, in which ChS serves as a biological cue, associated with

introduction of TGF-β1, significantly promoted the enhancement of chondrogenesis of

Page 203: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

189

MSCs. This result indicates that the incorporation of biomolecule (ChS) combined with

growth factor (TGF-β1) plays an influential factor in cartilaginous differentiation and

subsequently matrix production. However, ChS slightly provided the expected enhancement

of cell adhesion. This result is probably due to the low amount of ChS incorporated into the

Alg-foam. This amount might not be sufficiently effective to be recognized by the cells.

Consequently, both chondrocytes and MSCs tended to form clusters inside the pores, but

rarely adhered on the pore wall of the foams. Therefore, some challenging issues

concerning three cornerstones of tissue engineering approach remain to be further studied.

First, it is necessary to modify the foams (i.e. surface functionalization) in order to enhance

cell adhesion. Second, the effect of ChS release rate on the cell differentiation and the

matrix production is suggested to be intensively investigated further in association with the

function of the added growth factor.

Page 204: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

190

Figure 10. 1 Summary of major challenging issues in the area of cartilage tissue

engineering, as investigated in this dissertation, indicating the criteria that have been fulfilled

by the developed scaffolds.

The multilayered scaffolds were designed based on the optimized scaffolds for

subchondral bone and cartilage. Since the ideal scaffold for osteochondral repair does not

exist yet, the development of scaffold strategies, which provide superior long-term outcome,

are gaining more attention and receiving considerable research efforts. Therefore, recent

strategies of bi- or multi-layered scaffolds, including integrated bilayered and monolithic

biphasic scaffolds based on Alg-foam and Alg coated Bioglass® scaffold, were the focus of

this research from the materials point of view. Even though it seems likely that integrated

bilayered scaffolds may exhibit a weak point due to possible delamination at the interface

between the layers, our present study proved that the delamination can be overcome by the

incorporation of an adhesive-intermediate phase, which serves as an interface between

distinct cartilage and bone layers. In contrast, monolithic biphasic scaffolds were hardly

Page 205: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

191

controlled in terms of fabrication parameters and subsequently porous structure in particular

at the interface. In addition, the interfacial testing by using a micro-tensile testing machine

proved that the interfacial strength at break of integrated bilayered scaffolds was higher

compared to that of monolithic biphasic scaffolds. Therefore, it can be preliminarily

summarized in the realm of materials that the integrated bilayered scaffold system

developed in this study is an appropriate approach to be further developed as an

osteochondral construct.

Additionally, fiber meshes (PLLA and Alg/Gel) were electrospun and investigated as

the cartilage phase in bilayered scaffold. The fiber meshes exhibited pore size up to 50 µm

(with up to 500 µm in thickness) and reduction of the pore size was found with increasing

thickness of the mesh. The small pore size is known to be a limitation of electrospun fibers

for use as tissue engineering scaffolds. The small pore size in fiber meshes may inhibit cell

migration and nutrient transfer after implantation. A limited 3-dimensional structure may not

be suitable for regeneration of new cartilage. Nevertheless, the fiber meshes would be an

interesting candidate for use as a calcified cartilage layer at the cartilage-bone interface.

Dense fiber meshes with small pore size can act as a barrier against the infiltration of bone

cells from subchondral bone and to avoid vascularization of the cartilage phase in vivo.

Therefore, this strategy of multilayered scaffolds for osteochondral tissue engineering

applications is promising and it should be developed further in order to achieve more closely

biomimetic structure to the structure of native tissues.

For instance, a novel multilayered scaffold with more sophisticated structure is

offered by the present research as a new perspective focusing on functional osteochondral

scaffolds, as shown in Fig. 10.2. The scaffold is created based on the present work by the

combination of Alg-foam, PLLA fibers and PDLLA-c-BG scaffold for cartilage, interface and

subchondral bone – phases, respectively. Briefly, Alg solution was applied onto PLLA fiber

mesh/PDLLA-c-BG bilayered scaffold and formed gelation by the addition of CaCl2H2O

agent. A porous Alg phase was formed after lyophilization. It was integrated on top of fiber

mesh (Fig. 10.2). In this case, the fiber mesh acted as an intermediate layer between

Bioglass®-based scaffold and Alg-foam, which was aimed as a cartilage-subchondral bone

Page 206: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

192

interface exhibiting dense ECM. This concept was inspired by the previous study of Yunos et

al. [86] and our finding presented in Chapter 7 that the formation of HA at the interface

between Bioglass-based scaffold and PLLA fibers enhanced further the mechanical stability

of the interface. Osteoconductive hybrid fiber meshes (i.e. PLLA/Bioglass hybrid fibers) are

additionally recommended to be utilized as the interface phase instead of single polymeric

fibers. This design mimics calcified cartilage, which provides high ability for local

mineralization at the interface. Subsequently, a strong interface should be formed which is

able to integrate the cartilage and bone-like layers during in vitro cell culture and in in vivo

culture conditions.

It is expected that the scaffold developments and knowledge gained during the

present investigation will lead to advances in the field of osteochondral tissue regeneration

in the near future.

Figure 10. 2 SEM image showing a recommended multilayered scaffold – model for

osteochondral tissue engineering applications, including Alg-foam for cartilage phase, PLLA

fiber mesh for calcified interface phase and PDLLA-c-BG scaffold for bone phase.

Page 207: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

REFERENCES

[1] Smith GD, Knutsen G, Richardson JB. A clinical review of cartilage repair techniques.

J Bone Joint Surg Br 2005;87:445–9.

[2] Nöth U, Steinert AF, Tuan RS. Technology insight: adult mesenchymal stem cells for

osteoarthritis therapy. Nat Clin Pract Rheumatol 2008;4:371–80.

[3] Greene M. Osteoarthritis Diet and Lifestyle for Pain Relief.

http://commonsensehealth.com.

[4] Veronesi F, Giavaresi G, Tschon M, Borsari V, Nicoli Aldini N, Fini M. Clinical use of

bone marrow, bone marrow concentrate, and expanded bone marrow mesenchymal

stem cells in cartilage disease. Stem Cells Dev 2013;22:181–92.

[5] Panseri S, Russo A, Cunha C, Bondi A, Di Martino A, Patella S, et al. Osteochondral

tissue engineering approaches for articular cartilage and subchondral bone

regeneration. Knee Surg Sports Traumatol Arthrosc 2012;20:1182–91.

[6] Izadifar Z, Chen X, Kulyk W. Strategic Design and Fabrication of Engineered

Scaffolds for Articular Cartilage Repair. J Funct Biomater 2012;3:799–838.

[7] Reddi AH, Andrades A. Nanomaterials and Hydrogel Scaffolds for Articular Cartilage

Regeneration 2011;17:301–5.

[8] Milutinovic A. Articular Cartilage Repair : A Review of Surgical , Tissue Engineered ,

and Drug Based Therapies 2007:1–40.

[9] Huey DJ, Hu JC, Athanasiou KA. Unlike bone, cartilage regeneration remains

elusive. Science 2012;338:917–21.

[10] Park HN, Lee JB, Moon H, Yang DH, Kwon IK. The Potential of Scaffolds for Bone

Tissue Engineering. 2011.

[11] Xia Y, Mei F, Duan Y, Gao Y, Xiong Z, Zhang T, et al. Bone tissue engineering using

bone marrow stromal cells and an injectable sodium alginate/gelatin scaffold. J

Biomed Mater Res A 2012;100:1044–50.

[12] Meyer U, Wiesmann HP. Bone and Cartilage Engineering. Springer; 2006.

Page 208: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

194

[13] Rucci N. Molecular biology of bone remodelling. Clin Cases Miner Bone Metab

2008;5:49–56.

[14] Clarke B. Normal bone anatomy and physiology. Clin J Am Soc Nephrol 2008;3

Suppl 3:S131–9.

[15] Schindeler A, McDonald MM, Bokko P, Little DG. Bone remodeling during fracture

repair: The cellular picture. Semin Cell Dev Biol 2008;19:459–66.

[16] Martin I, Miot S, Barbero A, Jakob M, Wendt D. Osteochondral tissue engineering. J

Biomech 2007;40:750–65.

[17] Miao X, Sun D. Graded/Gradient Porous Biomaterials. Materials (Basel) 2009;3:26–

47.

[18] Mikos AG, Herring SW, Ochareon P, Elisseeff J, Lu HH, Kandel R, et al. Engineering

complex tissues. Tissue Eng 2006;12:3307–39.

[19] Nooeaid P, Salih V, Beier JP, Boccaccini AR. Osteochondral tissue engineering:

scaffolds, stem cells and applications. J Cell Mol Med 2012;16:2247–70.

[20] Castro NJ, Hacking SA, Zhang LG. Recent progress in interfacial tissue engineering

approaches for osteochondral defects. Ann Biomed Eng 2012;40:1628–40.

[21] Anatomy of A Joint.

http://www.lpch.org/DiseaseHealthInfo/HealthLibrary/arthritis/0276-Pop.html

[22] Grover CN, Cameron RE, Best SM. Investigating the morphological, mechanical and

degradation properties of scaffolds comprising collagen, gelatin and elastin for use in

soft tissue engineering. J Mech Behav Biomed Mater 2012;10:62–74.

[23] Rodrigues MT, Gomes ME, Reis RL, Rodrigues T. Current strategies for

osteochondral regeneration: from stem cells to pre-clinical approaches. Curr Opin

Biotechnol 2011;22:726–33.

[24] Tampieri A, Sprio S, Sandri M, Valentini F. Mimicking natural bio-mineralization

processes: a new tool for osteochondral scaffold development. Trends Biotechnol

2011;29:526–35.

[25] Syam P. Nukavarapu DLD. Osteochondral tissue engineering: Current strategies and

challenges. Biotechnol Adv 2012;31:706–21.

Page 209: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

195

[26] Keeney M, Pandit A. The Osteochondral Junction and Its Repair. Tissue Eng Part B

2009;15:55–73.

[27] Xu T, Binder KW, Albanna MZ, Dice D, Zhao W, Yoo JJ, et al. Hybrid printing of

mechanically and biologically improved constructs for cartilage tissue engineering

applications. Biofabrication 2013;5:015001.

[28] Chen Q, Roether JA, Boccaccini AR. Tissue Engineering Scaffolds from Bioactive

Glass and Composite Materials. Top Tissue Eng 2008;4:1–27.

[29] Rahaman MN, Day DE. Bal BS, Fu Q, Jung SB, Bonewalde LF, et al. Bioactive glass

in tissue engineering. Acta Biomater 2011;7:2355–73.

[30] Dinarvand P, Seyedjafari E, Shafiee A, Jandaghi AB, Doostmohammadi A, Fathi MH,

et al. New approach to bone tissue engineering: simultaneous application of

hydroxyapatite and bioactive glass coated on a poly(L-lactic acid) scaffold. ACS Appl

Mater Interfaces 2011;3:4518–24.

[31] Hench LL. The story of Bioglass. J Mater Sci Mater Med 2006;17:967–78.

[32] Chen QZ, Thompson ID, Boccaccini AR. 45S5 Bioglass-derived glass-ceramic

scaffolds for bone tissue engineering. Biomaterials 2006;27:2414–25.

[33] Jones JR. Review of bioactive glass: From Hench to hybrids. Acta Biomater

2012;9:4457–86.

[34] Gerhardt L-C, Boccaccini AR. Bioactive Glass and Glass-Ceramic Scaffolds for Bone

Tissue Engineering. Materials (Basel) 2010;3:3867–910.

[35] Hoppe A, Güldal NS, Boccaccini AR. A review of the biological response to ionic

dissolution products from bioactive glasses and glass-ceramics. Biomaterials

2011;32:2757–74.

[36] Hofmann DC, Suh J-Y, Wiest A, Duan G, Lind M-L, Demetriou MD, Johnson WL.

Designing metallic glass matrix composites with high toughness and tensile ductility.

Nature 2008;451:1085–9.

[37] Theodorou G, Goudouri OM, Kontonasaki E, Chatzistavrou X, Papadopoulou L,

Kantiranis N, et al. Comparative Bioactivity Study of 45S5 and 58S Bioglasses in

Organic and Inorganic Environment. Bioceram Dev Appl 2011;1:1–4.

Page 210: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

196

[38] Hupa L. Melt-derived bioactive glasses. Bioactive Glass Materials, Properties and

Applications, Woodhead Publishing Limited; 2011, p. 1–23.

[39] Hench LL. Bioceramics: From Concept to Clinic. J Am Ceram Soc 1991;74:1487–

510.

[40] Hench LL, Splinter RJ, Allen WC, Greenlee TK. Bonding mechanisms at the interface

of ceramic prosthetic materials. J Biomed Mater Res 1971;5:117–41.

[41] Xynos ID, Edgar AJ, Buttery LD, Hench LL, Polak JM. Ionic products of bioactive

glass dissolution increase proliferation of human osteoblasts and induce insulin-like

growth factor II mRNA expression and protein synthesis. Biochem Biophys Res

Commun 2000;276:461–5.

[42] Srinivasan S, Jayasree R, Chennazhi KP, Nair SV, Jayakumar R. Biocompatible

alginate/nano bioactive glass ceramic composite scaffolds for periodontal tissue

regeneration. Carbohydr Polym 2011;87:274–83.

[43] El-gendy R, Yang XB, Newby PJ. Osteogenic Differentiation of Human Dental Pulp

Stromal Cells on 45S5 Bioglass Based Scaffolds. Tissue Eng Part A 2013;19:707–

15.

[44] Price N, Bendall SP, Frondoza C, Jinnah RH, Hungerford DS. Human osteoblast-like

cells (MG63) proliferate on a bioactive glass surface. J Biomed Mater Res

1997;37:394–400.

[45] Silver IA, Deas J, Erecin M. Interactions of bioactive glasses with osteoblasts in vitro :

effects of 45S5 Bioglass, and 58S and 77S bioactive glasses on metabolism,

intracellular ion concentrations and cell viability. Biomaterials 2001;22:175–85.

[46] Peitl Filho O, LaTorre GP, Hench LL. Effect of crystallization on apatite-layer

formation of bioactive glass 45S5. J Biomed Mater Res 1996;30:509–14.

[47] Roohani-Esfahani SI, Nouri-Khorasani S, Lu ZF, Appleyard RC, Zreiqat H. Effects of

bioactive glass nanoparticles on the mechanical and biological behavior of composite

coated scaffolds. Acta Biomater 2011;7:1307–18.

Page 211: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

197

[48] Rezwan K, Chen QZ, Blaker JJ, Boccaccini AR. Biodegradable and bioactive porous

polymer/inorganic composite scaffolds for bone tissue engineering. Biomaterials

2006;27:3413–31.

[49] Chen Q, Zhu C, Thouas GA. Progress and challenges in biomaterials used for bone

tissue engineering: bioactive glasses and elastomeric composites. Prog Biomater

2012;1:2.

[50] Liverani L, Roether JA, Nooeaid P, Trombetta M, Schubert DW, Boccaccini AR. A

Simple fabrication technique for multilayered stratified composite scaffolds suitable

for interface tissue engineering. Mater Sci Eng A 2012;557:54–8.

[51] Blaker JJ, Maquet V, Jérôme R, Boccaccini AR, Nazhat SN. Mechanical properties of

highly porous PDLLA/Bioglass composite foams as scaffolds for bone tissue

engineering. Acta Biomater 2005;1:643–52.

[52] Mohamad Yunos D, Bretcanu O, Boccaccini AR. Polymer-bioceramic composites for

tissue engineering scaffolds. J Mater Sci 2008;43:4433–42.

[53] Metze A, Grimm A, Nooeaid P, Roether JA, Hum J, Newby PJ, et al. Gelatin coated

45S5 Bioglass ® -derived scaffolds for bone tissue engineering. Key Eng Mater

2013;541:31–9.

[54] Newby P, Chatzistavrou X, Roether JA, Hupa L, Boccaccini AR, Erol MM, et al.

Copper-releasing, boron-containing bioactive glass-based scaffolds coated with

alginate for bone tissue engineering. Acta Biomater 2012;8:792–801.

[55] Bretcanu O, Boccaccini AR, Salih V. Poly-dl-lactic acid coated Bioglass® scaffolds:

toughening effects and osteosarcoma cell proliferation. J Mater Sci 2012;47:5661–

72.

[56] Chen QZ, Boccaccini AR. Poly ( D , L -lactic acid ) coated 45S5 Bioglass®-based

scaffolds : Processing and characterization. J Biomed Mater Res Part A 2006:445–

57.

[57] Li W, Nooeaid P, Roether JA, Schubert DW, Boccaccini AR. Preparation and

characterization of vancomycin releasing PHBV coated 45S5 Bioglass®-based glass

– ceramic scaffolds for bone tissue engineering. J Eur Ceram Soc 2014;34:505–14.

Page 212: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

198

[58] Borzacchiello A, Gloria A, Mayol L, Dickinson S, Miot S, Martin I, et al.

Natural/synthetic porous scaffold designs and properties for fibro-cartilaginous tissue

engineering. J Bioact Compat Polym 2011;26:437–51.

[59] Osateerakun P, Kuptniratsaikul S, Bunaprasert T, Kanokpanont S, Itiravivong P.

Feasibility of using alginate/gelatin composite scaffold for human chondrocyte

regeneration. 4th 2011 Biomed Eng Int Conf 2012:290–3.

[60] Frenkel SR, Di Cesare PE. Scaffolds for articular cartilage repair. Ann Biomed Eng

2004;32:26–34.

[61] Rowley JA, Madlambayan G, Mooney DJ. Alginate hydrogels as synthetic

extracellular matrix materials. Biomaterials 1999;20:45–53.

[62] Wang C-C, Yang K-C, Lin K-H, Liu Y-L, Liu H-C, Lin F-H. Cartilage regeneration in

SCID mice using a highly organized three-dimensional alginate scaffold. Biomaterials

2012;33:120–7.

[63] Wang C-C, Yang K-C, Lin K-H, Liu H-C, Lin F-H. A highly organized three-

dimensional alginate scaffold for cartilage tissue engineering prepared by microfluidic

technology. Biomaterials 2011;32:7118–26.

[64] Lee KY, Rowley JA, Eiselt P, Moy EM, Bouhadir KH, Mooney DJ. Controlling

Mechanical and Swelling Properties of Alginate Hydrogels Independently by Cross-

Linker Type and Cross-Linking Density. Macromolecules 2000;33:4291–4.

[65] Patil JS, Kamalapur MV, Marapur SC, Kadam DV. Ionotropic gelation and

polyelectrolyte complexation: The novel techniques to design hydrogel particulate

sustained, modulate drug delivery system: A review 2010;5:241–8.

[66] Lee S-J, Lee IH, Park JH, Gwak S-J, Rhie J-W, Cho D-W, et al. Development of a

hybrid scaffold and a bioreactor for cartilage regeneration. Chinese Sci Bull

2009;54:3608–12.

[67] Chung C, Burdick JA. Engineering Cartilage Tissue. Adv Drug Deliv Rev

2008;60:243–62.

Page 213: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

199

[68] Zhang L, Li K, Xiao W, Zheng L, Xiao Y, Fan H, et al. Preparation of collagen–

chondroitin sulfate–hyaluronic acid hybrid hydrogel scaffolds and cell compatibility in

vitro. Carbohydr Polym 2011;84:118–25.

[69] Mulder L. Cell Adhesion on Alginate Scaffolds for the Tissue Engineering of an Aortic

Valve – A Review. 2002.

[70] Lee KY, Mooney DJ. Alginate: properties and biomedical applications. Prog Polym

Sci 2012;37:106–26.

[71] Schaffner P, Dard MM. Cellular and Molecular Life Sciences Structure and function

of RGD peptides involved in bone biology 2003;60:119–32.

[72] Rubert M, Alonso-Sande M, Monjo M, Ramis JM. Evaluation of Alginate and

Hyaluronic Acid for Their Use in Bone Tissue Engineering. Biointerphases 2012;7:44.

[73] Karageorgiou V, Kaplan D. Porosity of 3D biomaterial scaffolds and osteogenesis.

Biomaterials 2005;26:5474–91.

[74] Erol M, Özyuğuran A, Özarpat Ö, Küçükbayrak S. 3D Composite scaffolds using

strontium containing bioactive glasses. J Eur Ceram Soc 2012;32:2747–55.

[75] Gorgieva S, Kokol V. Biomaterials and Their Biocompatibility : Review and

Perspectives. Biomater. Appl. Nanomedicine, 2011, p. 17–52.

[76] Liu B, Lin P, Shen Y, Dong Y. Porous bioceramics reinforced by coating gelatin. J

Mater Sci Mater Med 2008;19:1203–7.

[77] Azami M, Rabiee M, Moztarzadeh F. Glutaraldehyde Crosslinked Gelatin/

hydroxyapatite Nanocomposite Scaffold , Engineered via Compound Techniques.

Polym Compos 2010:2112–20.

[78] Farris S, Song J, Huang Q. Alternative reaction mechanism for the cross-linking of

gelatin with glutaraldehyde. J Agric Food Chem 2010;58:998–1003.

[79] Desimone D, Li W, Roether JA, Schubert DW, Crovace MC, Rodrigues ACM, et al.

Biosilicate®–gelatine bone scaffolds by the foam replica technique: development and

characterization. Sci Technol Adv Mater 2013;14:045008.

Page 214: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

200

[80] Kim S-M, Yi S-A, Choi S-H, Kim K-M, Lee Y-K. Gelatin-layered and multi-sized

porous β-tricalcium phosphate for tissue engineering scaffold. Nanoscale Res Lett

2012;7:78.

[81] Vroman I, Tighzert L. Biodegradable Polymers. Materials (Basel) 2009;2:307–44.

[82] Jamshidian M, Tehrany EA, Imran M, Jacquot M, Desobry S. Poly-Lactic Acid:

Production, Applications, Nanocomposites, and Release Studies. Compr Rev Food

Sci Food Saf 2010;9:552–71.

[83] Izwan S, Razak A, Fadzliana N, Sharif A, Aizan W, Abdul W. Biodegradable

Polymers and their Bone Applications : A Review. Int J Basic Appl Sci 2012;12:31–

49.

[84] Tokiwa Y, Jarerat A. Biodegradation of poly(L-lactide). Biotechnol Lett 2004;26:771–

7.

[85] Yunos DM, Ahmad Z, Boccaccini AR. Fabrication and characterization of electrospun

poly-DL-lactide (PDLLA) fibrous coatings on 45S5 Bioglass® substrates for bone

tissue engineering applications. J Chem Technol Biotechnol 2009;85:768–74.

[86] Yunos DM, Ahmad Z, Salih V, Boccaccini AR. Stratified scaffolds for osteochondral

tissue engineering applications: Electrospun PDLLA nanofibre coated Bioglass(R)-

derived foams. J Biomater Appl 2011:8–12.

[87] Novak S, Druce J, Chen QZ, Boccaccini AR. TiO2 foams with poly- (D,L -lactic acid)

(PDLLA) and PDLLA /Bioglass coatings for bone tissue engineering scaffolds. J

Mater Sci 2009;44:1442–8.

[88] Verrier S. Bioactive glass containing composites for bone and musculoskeletal tissue

engineering scaffolds. Bioactive Glass: Materials, Properties and Applications,

Woodhead Publishing Limited; 2011, p. 162–88.

[89] Zhao K, Deng Y, Chen JC, Chen GQ, Chun Chen J. Polyhydroxyalkanoate (PHA)

scaffolds with good mechanical properties and biocompatibility. Biomaterials

2003;24:1041–5.

Page 215: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

201

[90] Brandl H, Gross RA, Lenz RW, Fuller RC. Plastics from bacteria and for bacteria:

poly(beta-hydroxyalkanoates) as natural, biocompatible, and biodegradable

polyesters. Adv Biochem Eng Biotechnol 1990;41:77–93.

[91] Shangguan Y-Y, Wang Y-W, Wu Q, Chen G-Q. The mechanical properties and in

vitro biodegradation and biocompatibility of UV-treated poly(3-hydroxybutyrate-co-3-

hydroxyhexanoate). Biomaterials 2006;27:2349–57.

[92] Chen G. Plastics Completely Synthesized by Bacteria : Polyhydroxyalkanoates.

Microbiol Monogr 2010;14:17–38.

[93] Baino F, Vitale-Brovarone C. Three-dimensional glass-derived scaffolds for bone

tissue engineering: current trends and forecasts for the future. J Biomed Mater Res A

2011;97:514–35.

[94] Mouriño V, Cattalini JP, Roether JA, Dubey P, Roy I, Boccaccini AR. Composite

polymer-bioceramic scaffolds with drug delivery capability for bone tissue

engineering. Expert Opin Drug Deliv 2013;10:1353–65.

[95] Francis L, Meng D, Knowles JC, Roy I, Boccaccini AR. Multi-functional P(3HB)

microsphere/45S5 Bioglass-based composite scaffolds for bone tissue engineering.

Acta Biomater 2010;6:2773–86.

[96] Olalde B, Garmendia N, Sáez-Martínez V, Argarate N, Nooeaid P, Morin F, et al.

Multifunctional bioactive glass scaffolds coated with layers of poly(d,l-lactide-co-

glycolide) and poly(n-isopropylacrylamide-co-acrylic acid) microgels loaded with

vancomycin. Mater Sci Eng C Mater Biol Appl 2013;33:3760–7.

[97] Yaylaoğlu MB, Korkusuz P, Ors U, Korkusuz F, Hasirci V. Development of a calcium

phosphate-gelatin composite as a bone substitute and its use in drug release.

Biomaterials 1999;20:711–9.

[98] Kim H-W, Knowles JC, Kim H-E. Development of hydroxyapatite bone scaffold for

controlled drug release via poly(epsilon-caprolactone) and hydroxyapatite hybrid

coatings. J Biomed Mater Res B Appl Biomater 2004;70:240–9.

[99] Uhrich KE, Cannizzaro SM, Langer RS, Shakesheff KM. Polymeric systems for

controlled drug release. Chem Rev 1999;99:3181–98.

Page 216: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

202

[100] Cheng H-W, Luk KDK, Cheung KMC, Chan BP. In vitro generation of an

osteochondral interface from mesenchymal stem cell-collagen microspheres.

Biomaterials 2011;32:1526–35.

[101] Mrosek EH, Schagemann JC, Chung H, Fitzsimmons JS, Yaszemski MJ, Mardones

RM, et al. Porous tantalum and poly-epsilon-caprolactone biocomposites for

osteochondral defect repair: preliminary studies in rabbits. J Orthop Res

2010;28:141–8.

[102] Cao T, Ho K-H, Teoh S-H. Scaffold design and in vitro study of osteochondral

coculture in a three-dimensional porous polycaprolactone scaffold fabricated by

fused deposition modeling. Tissue Eng 2003;9 Suppl 1:S103–12.

[103] Naveena N, Venugopal J, Rajeswari R, Sundarrajan S, Sridhar R, Shayanti M, et al.

Biomimetic composites and stem cells interaction for bone and cartilage tissue

regeneration. J Mater Chem 2012;22:5239.

[104] Dhandayuthapani B, Yoshida Y, Maekawa T, Kumar DS. Polymeric Scaffolds in

Tissue Engineering Application: A Review. Int J Polym Sci 2011;2011:1–19.

[105] Puppi D, Chiellini F, Piras a. M, Chiellini E. Polymeric materials for bone and

cartilage repair. Prog Polym Sci 2010;35:403–40.

[106] Ge Z, Li C, Heng BC, Cao G, Yang Z. Functional biomaterials for cartilage

regeneration. J Biomed Mater Res A 2012;100:2526–36.

[107] Bernhard JC, Panitch A. Synthesis and characterization of an aggrecan mimic. Acta

Biomater 2012;8:1543–50.

[108] Kim H-EH, Knowles JC. Hydroxyapatite and gelatin composite foams processed via

novel freeze-drying and crosslinking for use as temporary hard tissue scaffolds. J

Biomed Mater Res A 2005;72:136–45.

[109] Kuo CK, Ma PX. Ionically crosslinked alginate hydrogels as scaffolds for tissue

engineering: part 1. Structure, gelation rate and mechanical properties. Biomaterials

2001;22:511–21.

Page 217: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

203

[110] Tampieri A, Sandri M, Landi E, Pressato D, Francioli S, Quarto R, et al. Design of

graded biomimetic osteochondral composite scaffolds. Biomaterials 2008;29:3539–

46.

[111] Chang C, Lin F, Lin C-C, Chou C-H, Liu H-C. Cartilage tissue engineering on the

surface of a novel gelatin-calcium-phosphate biphasic scaffold in a double-chamber

bioreactor. J Biomed Mater Res B Appl Biomater 2004;71:313–21.

[112] Bi L, Li D, Liu J, Hu Y, Yang P, Yang B, et al. Fabrication and characterization of a

biphasic scaffold for osteochondral tissue engineering. Mater Lett 2011;65:2079–82.

[113] Ahn J, Lee T, Oh J, Kim S, Kim H, et al. Biphasic Scaffold for the Repair of

Osteochondral Defects in Rabbits. Tissue Eng Part A 2009;15:2595–604.

[114] Chen J, Chen H, Li P, Diao H, Zhu S, Dong L, et al. Simultaneous regeneration of

articular cartilage and subchondral bone in vivo using MSCs induced by a spatially

controlled gene delivery system in bilayered integrated scaffolds. Biomaterials

2011;32:4793–805.

[115] Harley BA, Lynn AK, Wissner-gross Z, Bonfield W, Yannas I V, Gibson LJ. Design of

a multiphase osteochondral scaffold III: Fabrication of layered scaffolds with

continuous interfaces. J Biomed Mater Res A 2010;92:1078–93.

[116] Ohyabu Y, Adegawa T, Yoshioka T, Ikoma T, Uemura T, Tanaka J. Cartilage

regeneration using a porous scaffold, a collagen sponge incorporating a

hydroxyapatite/chondroitinsulfate composite. Mater Sci Eng B 2010;173:204–7.

[117] Shafiee A, Soleimani M, Chamheidari GA, Seyedjafari E, Dodel M, Atashi A, et al.

Electrospun nanofiber-based regeneration of cartilage enhanced by mesenchymal

stem cells. J Biomed Mater Res A 2011;99:467–78.

[118] Cui L, Wu Y, Cen L, Zhou H, Yin S, Liu G, et al. Repair of articular cartilage defect in

non-weight bearing areas using adipose derived stem cells loaded polyglycolic acid

mesh. Biomaterials 2009;30:2683–93.

[119] Alves da Silva ML, Crawford A, Mundy JM, Correlo VM, Sol P, Bhattacharya M, et al.

Chitosan/polyester-based scaffolds for cartilage tissue engineering: assessment of

extracellular matrix formation. Acta Biomater 2010;6:1149–57.

Page 218: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

204

[120] Chen G, Sato T, Tanaka J, Tateishi T. Osteochondral tissue engineering using a

PLGA–collagen hybrid mesh. Mater Sci Eng C 2006;26:118–23.

[121] Ma Z, Gao C, Gong Y, Shen J. Cartilage tissue engineering PLLA scaffold with

surface immobilized collagen and basic fibroblast growth factor. Biomaterials

2005;26:1253–9.

[122] Holland TA, Bodde EWH, Baggett LS, Tabata Y, Mikos AG, Jansen JA.

Osteochondral repair in the rabbit model utilizing bilayered, degradable

oligo(poly(ethylene glycol) fumarate) hydrogel scaffolds. J Biomed Mater Res A

2005;75:156–67.

[123] Ghosh S, Viana JC, Reis RL, Mano JF. Bi-layered constructs based on poly(l-lactic

acid) and starch for tissue engineering of osteochondral defects. Mater Sci Eng C

2008;28:80–6.

[124] Baldwin AD, Kiick KL. Review Polysaccharide-Modified Synthetic Polymeric

Biomaterials. PeptideScience 2010;94:128–40.

[125] Khanarian NT, Haney NM, Burga RA, Lu HH. A functional agarose-hydroxyapatite

scaffold for osteochondral interface regeneration. Biomaterials 2012;33:5247–58.

[126] Williams DF. Biomaterials and tissue engineering in reconstructive surgery

2003;28:563–74.

[127] Stevens M. A rapid-curing alginate gel system: utility in periosteum-derived cartilage

tissue engineering. Biomaterials 2004;25:887–94.

[128] Despang F, Börner A, Dittrich R, Tomandl G, Pompe W, Gelinsky M.

Alginate/calcium phosphate scaffolds with oriented, tube-like pores. Materwiss

Werksttech 2005;36:761–7.

[129] Dittrich R, Tomandl G, Despang F, Bernhardt A, Hanke T, Pompe W, et al. Scaffolds

for Hard Tissue Engineering by Ionotropic Gelation of Alginate?Influence of Selected

Preparation Parameters. J Am Ceram Soc 2007;90:1703–8.

[130] Nguyen LH, Kudva AK, Guckert NL, Linse KD, Roy K. Unique biomaterial

compositions direct bone marrow stem cells into specific chondrocytic phenotypes

Page 219: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

205

corresponding to the various zones of articular cartilage. Biomaterials 2011;32:1327–

38.

[131] Lohan A, Marzahn U, El Sayed K, Haisch A, Kohl B, Müller RD, et al. In vitro and in

vivo neo-cartilage formation by heterotopic chondrocytes seeded on PGA scaffolds.

Histochem Cell Biol 2011;136:57–69.

[132] Wang L, Shelton RM, Cooper PR, Lawson M, Triffitt JT, Barralet JE. Evaluation of

sodium alginate for bone marrow cell tissue engineering. Biomaterials 2003;24:3475–

81.

[133] Oerther S, Le Gall H, Payan E, Lapicque F, Presle N, Hubert P, et al. Hyaluronate-

alginate gel as a novel biomaterial: mechanical properties and formation mechanism.

Biotechnol Bioeng 1999;63:206–15.

[134] Petrenko YA, Ivanov RV, Petrenko AY, Lozinsky VI. Coupling of gelatin to inner

surfaces of pore walls in spongy alginate-based scaffolds facilitates the adhesion,

growth and differentiation of human bone marrow mesenchymal stromal cells. J

Mater Sci Mater Med 2011;22:1529–40.

[135] Kashima K, Imai M. Advanced Membrane Material from Marine Biological Polymer

and Sensitive Molecular-Size Recognition for Promising Separation Technology,

Advancing Desalination. InTech Pub, InTech Pub (Croaria); 2012, p. 7.

[136] Buckley CT, O'Kelly KU. Regular scaffold fabrication techniques for investigations in

tissue engineering. Top Bio-Mechanical Eng., 2004, p. 147–66.

[137] Lien S-M, Ko L, Huang T-J. Effect of pore size on ECM secretion and cell growth in

gelatin scaffold for articular cartilage tissue engineering. Acta Biomater 2009;5:670–

9.

[138] Swieszkowski W, Tuan BHS, Kurzydlowski KJ, Hutmacher DW. Repair and

regeneration of osteochondral defects in the articular joints. Biomol Eng

2007;24:489–95.

[139] St-Pierre J-P, Gan L, Wang J, Pilliar RM, Grynpas MD, Kandel RA. The incorporation

of a zone of calcified cartilage improves the interfacial shear strength between in

vitro-formed cartilage and the underlying substrate. Acta Biomater 2012;8:1603–15.

Page 220: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

206

[140] Spalazzi JP, Doty SB, Moffat KL, Levine WN, Lu HH. Development of controlled

matrix heterogeneity on a triphasic scaffold for orthopedic interface tissue

engineering. Tissue Eng 2006;12:3497–508.

[141] Keaveny TM, Morgan EF, Yeh O. Bone Mechanic. Standard Handbook of Biomedical

Engineering and Design, 2004, p. 1–24.

[142] Miao X, Tan DM, Li J, Xiao Y, Crawford R. Mechanical and biological properties of

hydroxyapatite/tricalcium phosphate scaffolds coated with poly(lactic-co-glycolic

acid). Acta Biomater 2008;4:638–45.

[143] Miao X, Lim W-K, Huang X, Chen Y. Preparation and characterization of

interpenetrating phased TCP/HA/PLGA composites. Mater Lett 2005;59:4000–5.

[144] Wu X, Liu Y, Li X, Wen P, Zhang Y, Long Y, et al. Preparation of aligned porous

gelatin scaffolds by unidirectional freeze-drying method. Acta Biomater 2010;6:1167–

77.

[145] Nooeaid P, Roether JA, Weber E, Schubert DW, Boccaccini AR. Technologies for

Multilayered Scaffolds Suitable for Interface Tissue Engineering. Adv Eng Mater

2013.

[146] Hutmacher DW. Scaffolds in tissue engineering bone and cartilage. Biomaterials

2000;21:2529–43.

[147] Ramay HRR, Zhang M. Biphasic calcium phosphate nanocomposite porous scaffolds

for load-bearing bone tissue engineering. Biomaterials 2004;25:5171–80.

[148] Chen G, Sato T, Tanaka J, Tateishi T. Preparation of a biphasic scaffold for

osteochondral tissue engineering. Mater Sci Eng C 2006;26:118–23.

[149] Luo X, Zhang L, Morsi Y, Zou Q, Wang Y, Gao S, et al. Hydroxyapatite/polyamide66

porous scaffold with an ethylene vinyl acetate surface layer used for simultaneous

substitute and repair of articular cartilage and underlying bone. Appl Surf Sci

2011;257:9888–94.

[150] Oliveira JM, Rodrigues MT, Silva SS, Malafaya PB, Gomes ME, Viegas C a, et al.

Novel hydroxyapatite/chitosan bilayered scaffold for osteochondral tissue-

Page 221: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

207

engineering applications: Scaffold design and its performance when seeded with

goat bone marrow stromal cells. Biomaterials 2006;27:6123–37.

[151] Lien S-M, Chien C-H, Huang T-J. A novel osteochondral scaffold of ceramic–gelatin

assembly for articular cartilage repair. Mater Sci Eng C 2009;29:315–21.

[152] Qu D, Li J, Li Y, Khadka A, Zuo Y, Wang H, et al. Ectopic osteochondral formation of

biomimetic porous PVA-n-HA/PA6 bilayered scaffold and BMSCs construct in rabbit.

J Biomed Mater Res B Appl Biomater 2011;96:9–15.

[153] Kinikoglu B, Rodríguez-Cabello JC, Damour O, Hasirci V. A smart bilayer scaffold of

elastin-like recombinamer and collagen for soft tissue engineering. J Mater Sci Mater

Med 2011;22:1541–54.

[154] Wang Y, Bian Y-Z, Wu Q, Chen G-Q. Evaluation of three-dimensional scaffolds

prepared from poly(3-hydroxybutyrate-co-3-hydroxyhexanoate) for growth of

allogeneic chondrocytes for cartilage repair in rabbits. Biomaterials 2008;29:2858–

68.

[155] Kon E, Delcogliano M, Filardo G, Pressato D, Busacca M, Grigolo B, et al. A novel

nano-composite multi-layered biomaterial for treatment of osteochondral lesions:

technique note and an early stability pilot clinical trial. Inj Int J Care Inj 2010;41:693–

701.

[156] Kang HW, Tabata Y, Ikada Y. Fabrication of porous gelatin scaffolds for tissue

engineering. Biomaterials 1999;20:1339–44.

[157] He L, Liu B, Xipeng G, Xie G, Liao S, Quan D, et al. Microstructure and properties of

nano-fibrous PCL-b-PLLA Scaffolds for Cartilage Tissue Engineering. Eur Cells

Mater 2009;18:63–74.

[158] Sill TJ, von Recum HA. Electrospinning: applications in drug delivery and tissue

engineering. Biomaterials 2008;29:1989–2006.

[159] Dvir T, Timko BP, Kohane DS, Langer R. Nanotechnological strategies for

engineering complex tissues. Nat Nanotechnol 2011;6:13–22.

Page 222: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

208

[160] Li W-J, Tuli R, Huang X, Laquerriere P, Tuan RS. Multilineage differentiation of

human mesenchymal stem cells in a three-dimensional nanofibrous scaffold.

Biomaterials 2005;26:5158–66.

[161] Sisson K, Zhang C, Farach-Carson MC, Chase DB, Rabolt JF. Evaluation of cross-

linking methods for electrospun gelatin on cell growth and viability.

Biomacromolecules 2009;10:1675–80.

[162] Chang J, Lee Y-H, Wu M, Yang M-C, Chien C. Preparation of electrospun alginate

fibers with chitosan sheath. Carbohydr Polym 2012;87:2357–61.

[163] Sundelacruz S, Kaplan DL. Stem cell- and scaffold-based tissue engineering

approaches to osteochondral regenerative medicine. Semin Cell Dev Biol

2009;20:646–55.

[164] Venugopal J, Low S, Choon AT, Ramakrishna S. Interaction of Cells and Nanofiber

Scaffolds in Tissue Engineering 2007:34–48.

[165] Ma Z, Kotaki M, Inai R, Ramakrishna S. Potential of nanofiber matrix as tissue-

engineering scaffolds. Tissue Eng 2005;11:101–9.

[166] Ma G, Fang D, Liu Y, Zhu X, Nie J. Electrospun sodium alginate/poly(ethylene oxide)

core–shell nanofibers scaffolds potential for tissue engineering applications.

Carbohydr Polym 2012;87:737–43.

[167] Cui W, Zhou Y, Chang J. Electrospun nanofibrous materials for tissue engineering

and drug delivery. Sci Technol Adv Mater 2010;11:014108.

[168] Chen G, Sato T, Ushida T, Hirochika R, Shirasaki Y, Ochiai N, et al. The use of a

novel PLGA fiber/collagen composite web as a scaffold for engineering of articular

cartilage tissue with adjustable thickness. J Biomed Mater Res A 2003;67:1170–80.

[169] Milleret V, Simona B, Neuenschwander P, Hall H. Tuning electrospinning parameters

for production of 3D-fiber-fleeces with increased porosity for soft tissue engineering

applications. Eur Cell Mater 2011;21:286–303.

[170] Dahlin RL, Kasper FK, Ph D, Mikos AG. Polymeric Nanofibers in Tissue Engineering.

Tissue Eng Part B 2011;17:349–64.

Page 223: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

209

[171] Ngiam M, Liao S, Patil AJ, Cheng Z, Chan CK, Ramakrishna S. The fabrication of

nano-hydroxyapatite on PLGA and PLGA/collagen nanofibrous composite scaffolds

and their effects in osteoblastic behavior for bone tissue engineering. Bone

2009;45:4–16.

[172] Jeong SI, Ko EK, Yum J, Jung CH, Lee YM, Shin H. Nanofibrous poly(lactic

acid)/hydroxyapatite composite scaffolds for guided tissue regeneration. Macromol

Biosci 2008;8:328–38.

[173] Fang D, Liu Y, Jiang S, Nie J, Ma G. Effect of intermolecular interaction on

electrospinning of sodium alginate. Carbohydr Polym 2011;85:276–9.

[174] Zhang Y, Ouyang H, Lim CT, Ramakrishna S, Huang Z-M. Electrospinning of gelatin

fibers and gelatin/PCL composite fibrous scaffolds. J Biomed Mater Res B Appl

Biomater 2005;72:156–65.

[175] Schek RM, Taboas JM, Hollister SJ, Krebsbach PH. Tissue engineering

osteochondral implants for temporomandibular joint repair. Orthod Craniofac Res

2005;8:313–9.

[176] Hung CT, Lima EG, Mauck RL, Taki E, Leroux MA, Lu HH, et al. Anatomically

shaped osteochondral constructs for articular cartilage repair. J Biomech

2003;36:1853–64.

[177] Huang Q, Goh JCH, Hutmacher DW, Lee EH. In vivo mesenchymal cell recruitment

by a scaffold loaded with transforming growth factor beta1 and the potential for in situ

chondrogenesis. Tissue Eng 2002;8:469–82.

[178] Fedorovich NE, Schuurman W, Wijnberg HM, Prins HJ, van Weeren PR, Malda J, et

al. Biofabrication of Osteochondral Tissue Equivalents by Printing Topologically

Defined. Tissue Eng Part C 2012;18:33–44.

[179] Jayabalan P, Tan AR, Rahaman MN, Bal BS, Hung CT, Cook JL. Bioactive glass 13-

93 as a subchondral substrate for tissue-engineered osteochondral constructs: a pilot

study. Clin Orthop Relat Res 2011;469:2754–63.

Page 224: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

210

[180] Bretcanu O, Chatzistavrou X, Paraskevopoulos K, Conradt R, Thompson I,

Boccaccini AR. Sintering and crystallisation of 45S5 Bioglass® powder. J Eur Ceram

Soc 2009;29:3299–306.

[181] Hollister SJ, Krebsbach PH. Engineered Osteochondral Grafts Using Biphasic

composite Solid Free-Form Fabricated Scaffolds. Tissue Eng 2004;10:1376–85.

[182] Bian W, Li D, Lian Q, Li X, Zhang W, Wang K. Fabrication of a bio-inspired beta-

Tricalcium phosphate / collagen scaffold based on ceramic stereolithography and gel

casting for osteochondral tissue engineering. Rapid Prototyp J 2012;18:68–80.

[183] Peña J, Román J, Victoria Cabañas M, Vallet-Regí M. An alternative technique to

shape scaffolds with hierarchical porosity at physiological temperature. Acta

Biomater 2010;6:1288–96.

[184] Bellucci D, Sola A, Cannillo V. A Revised Replication Method for Bioceramic

Scaffolds. Bioceram Dev Appl 2011;1:1–8.

[185] Boccaccini AR, Chen QZ. Process for preparing a bioactive glass scaffold, 2006.

[186] Dern CD. Freeze-Drying 101: Lyophilization Technology.

www.americanlaboratory.com 2005.

[187] Mujumda AS. Drying technology in agriculture and food sciences. Science Publisher;

2007.

[188] Ho M-H, Kuo P-Y, Hsieh H-J, Hsien T-Y, Hou L-T, Lai J-Y, et al. Preparation of

porous scaffolds by using freeze-extraction and freeze-gelation methods.

Biomaterials 2004;25:129–38.

[189] Lopez-quiroga E, Antelo LT, Alonso AA. Time-scale modelling and optimal control of

freeze-drying. J Food Eng 2012;111:655–66.

[190] Christ M. Smart Freeze Drying with System. www.martinchrist.de

[191] Mackenzie AP. Formation of ice into water vapor in the freeze-drying process. Ann

New York Acad Sci:522–47.

[192] Liang W, Kienitz BL, Penick KJ, Welter JF, Zawodzinski TA, Baskaran H.

Concentrated collagen-chondroitin sulfate scaffolds for tissue engineering

applications. J Biomed Mater Res A 2010;94:1050–60.

Page 225: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

211

[193] Simpson NE, Stabler CL, Simpson CP, Sambanis A, Constantinidis I. The role of the

CaCl2–guluronic acid interaction on alginate encapsulated βTC3 cells. Biomaterials

2004;25:2603–10.

[194] Fang Y, Al-Assaf S, Phillips GO, Nishinari K, Funami T, Williams PA, et al. Multiple

steps and critical behaviors of the binding of calcium to alginate. J Phys Chem B

2007;111:2456–62.

[195] Lee JS, Kim B, Kim MS, Lee SJ, Kim SW, Cho DW, et al. The Effect of 3D

Construction Culture of Human Chondrocytes Using Alginate Sponge. Key Eng

Mater 2006;326-328:883–8.

[196] Wan LQ, Jiang J, Arnold DE, Guo XE, Lu HH, Mow VC. Calcium Concentration

Effects on the Mechanical and Biochemical Properties of Chondrocyte-Alginate

Constructs. Cell Mol Bioeng 2008;1:93–102.

[197] Li Z, Gunn J, Chen M-H, Cooper A, Zhang M. On-site alginate gelation for enhanced

cell proliferation and uniform distribution in porous scaffolds. J Biomed Mater Res A

2008;86:552–9.

[198] Coates EE, Riggin CN, Fisher JP. Matrix molecule influence on chondrocyte

phenotype and proteoglycan 4 expression by alginate-embedded zonal chondrocytes

and mesenchymal stem cells. J Orthop Res 2012;30:1886–97.

[199] Caterson EJ, Li WJ, Nesti LJ, Albert T, Danielson K, Tuan RS. Polymer/alginate

amalgam for cartilage-tissue engineering. Ann N Y Acad Sci 2002;961:134–8.

[200] Ma K, Titan AL, Stafford M, Zheng CH, Levenston ME. Variations in chondrogenesis

of human bone marrow-derived mesenchymal stem cells in fibrin/alginate blended

hydrogels. Acta Biomater 2012;8:3754–64.

[201] Iwasaki N, Yamane S-T, Majima T, Kasahara Y, Minami A, Harada K, et al.

Feasibility of polysaccharide hybrid materials for scaffolds in cartilage tissue

engineering: evaluation of chondrocyte adhesion to polyion complex fibers prepared

from alginate and chitosan. Biomacromolecules 2004;5:828–33.

Page 226: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

212

[202] Miralles G, Baudoin R, Dumas D, Baptiste D, Hubert P, Stoltz JF, et al. Sodium

alginate sponges with or without sodium hyaluronate: in vitro engineering of cartilage.

J Biomed Mater Res 2001;57:268–78.

[203] Yen C, Lin Y, Chang MD, Tien C-W, Wu Y, Liao C-J, et al. Use of porous alginate

sponges for substantial chondrocyte expansion and matrix production: effects of

seeding density. Biotechnol Prog 2008;24:452–7.

[204] Szentivanyi A, Chakradeo T, Zernetsch H, Glasmacher B. Electrospun cellular

microenvironments: Understanding controlled release and scaffold structure. Adv

Drug Deliv Rev 2011;63:209–20.

[205] Teo W, Inai R, Ramakrishna S. Technological advances in electrospinning of

nanofibers. Sci Technol Adv Mater 2011;12:1–19.

[206] Pham QP, Sharma U, Mikos AG. Electrospinning of polymeric nanofibers for tissue

engineering applications: a review. Tissue Eng 2006;12:1197–211.

[207] Li W, Tuli R, Okafor C, Derfoul A, Danielson KG, Hall DJ, et al. A three-dimensional

nanofibrous scaffold for cartilage tissue engineering using human mesenchymal stem

cells. Biomaterials 2005;26:599–609.

[208] Li W-J, Cooper JA, Mauck RL, Tuan RS. Fabrication and characterization of six

electrospun poly(alpha-hydroxy ester)-based fibrous scaffolds for tissue engineering

applications. Acta Biomater 2006;2:377–85.

[209] Li W-J, Danielson KG, Alexander PG, Tuan RS. Biological response of chondrocytes

cultured in three-dimensional nanofibrous poly(epsilon-caprolactone) scaffolds. J

Biomed Mater Res A 2003;67:1105–14.

[210] Lee KY, Jeong L, Kang YO, Lee SJ, Park WH. Electrospinning of polysaccharides for

regenerative medicine. Adv Drug Deliv Rev 2009;61:1020–32.

[211] Bowlin GL, Sell SA, Wolfe PS, Garg K, McCool JM, Rodriguez IA. The Use of Natural

Polymers in Tissue Engineering: A Focus on Electrospun Extracellular Matrix

Analogues. Polymers (Basel) 2010;2:522–53.

[212] Bhattarai N, Zhang M. Controlled synthesis and structural stability of alginate-based

nanofibers. Nanotechnology 2007;18:455601.

Page 227: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

213

[213] Bhattarai N, Edmondson D, Veiseh O, Matsen FA, Zhang M. Electrospun chitosan-

based nanofibers and their cellular compatibility. Biomaterials 2005;26:6176–84.

[214] Skotak M, Noriega S, Larsen G, Subramanian A. Electrospun cross-linked gelatin

fibers with controlled diameter: the effect of matrix stiffness on proliferative and

biosynthetic activity of chondrocytes cultured in vitro. J Biomed Mater Res A

2010;95:828–36.

[215] Klein TJ, Malda J, Sah RL, Hutmacher DW. Tissue engineering of articular cartilage

with biomimetic zones. Tissue Eng Part B Rev 2009;15:143–57.

[216] Kim H-W, Knowles JC, Kim H-E. Porous scaffolds of gelatin-hydroxyapatite

nanocomposites obtained by biomimetic approach: characterization and antibiotic

drug release. J Biomed Mater Res B Appl Biomater 2005;74:686–98.

[217] Zhao H, Wang G, Hu S, Cui J, et al. In vitro Biomimetic Construction of

Hydroxyapatite – Porcine Acellular Dermal Matrix Composite Scaffold for MC3T3-E1

Preosteoblast Culture 2011;17.

[218] Concentrations B, Erisken C. Osteochondral Tissue Formation Through Adipose-

Derived Stromal Cell Differentiation on Biomimetic Polycaprolactone Nanofibrous

Scaffolds with Graded Insulin. Tissue Eng Part A 2011;17:1239–52.

[219] Li J, Sun H, Sun D, Yao Y, Yao F, Yao K. Biomimetic multicomponent

polysaccharide/nano-hydroxyapatite composites for bone tissue engineering.

Carbohydr Polym 2011;85:885–94.

[220] Al-Munajjed AA, Plunkett NA, Gleeson JP, Weber T, Jungreuthmayer C, Levingstone

T, et al. Development of a biomimetic collagen-hydroxyapatite scaffold for bone

tissue engineering using a SBF immersion technique. J Biomed Mater Res B Appl

Biomater 2009;90:584–91.

[221] Zhou J, Xu C, Wu G, Cao X, Zhang L, Zhai Z, et al. In vitro generation of

osteochondral differentiation of human marrow mesenchymal stem cells in novel

collagen-hydroxyapatite layered scaffolds. Acta Biomater 2011;7:3999–4006.

Page 228: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

214

[222] Nagura I, Fujioka H, Kokubu T, Makino T, Sumi Y, Kurosaka M. Repair of

osteochondral defects with a new porous synthetic polymer scaffold. J Bone Joint

Surg Br 2007;89:258–64.

[223] Im G-I, Lee JH. Repair of osteochondral defects with adipose stem cells and a dual

growth factor-releasing scaffold in rabbits. J Biomed Mater Res B Appl Biomater

2010;92:552–60.

[224] Ho STB, Hutmacher DW, Ekaputra AK, Hitendra D, Hui JH. The evaluation of a

biphasic osteochondral implant coupled with an electrospun membrane in a large

animal model. Tissue Eng Part A 2010;16:1123–41.

[225] Marquass B, Somerson JS, Hepp P, Aigner T, Schwan S, Bader A, et al. A novel

MSC-seeded triphasic construct for the repair of osteochondral defects. J Orthop Res

2010;28:1586–99.

[226] Jiang J, Tang A, Ateshian GA, Guo XE, Hung CT, Lu HH. Bioactive stratified polymer

ceramic-hydrogel scaffold for integrative osteochondral repair. Ann Biomed Eng

2010;38:2183–96.

[227] Galperin A, Oldinski RA, Florczyk SJ, Bryers JD, Zhang M, Ratner BD. Integrated Bi-

Layered Scaffold for Osteochondral Tissue Engineering. Adv Healthc Mater 2012:1–

12.

[228] Mohan N, Dormer NH, Caldwell KL, Key VH, Berkland CJ, Detamore MS.

Continuous gradients of material composition and growth factors for effective

regeneration of the osteochondral interface. Tissue Eng Part A 2011;17:2845–55.

[229] Shi X, Zhou J, Zhao Y, Li L, Wu H. Gradient-Regulated Hydrogel for Interface Tissue

Engineering: Steering Simultaneous Osteo/Chondrogenesis of Stem Cells on a Chip.

Adv Healthc Mater 2012:1–8.

[230] Kon E, Delcogliano M, Filardo G, Fini M, Giavaresi G, Francioli S, et al. Orderly

osteochondral regeneration in a sheep model using a novel nano-composite

multilayered biomaterial. J Orthop Res 2010;28:116–24.

Page 229: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

215

[231] Kon E, Delcogliano M, Filardo G, Busacca M, Di Martino A, Marcacci M, et al. Novel

nano-composite multilayered biomaterial for osteochondral regeneration: a pilot

clinical trial. Am J Sports Med 2011;39:1180–90.

[232] Spalazzi JP, Dionisio KL, Jiang J, Lu HH. Osteoblast and Chondrocyte Interactions

During Coculture on Scaffolds. IEEE Eng Med Biol Mag 2003:27–34.

[233] Deng Y, Zhao K, Zhang X-F, Hu P, Chen G-Q. Study on the three-dimensional

proliferation of rabbit articular cartilage-derived chondrocytes on

polyhydroxyalkanoate scaffolds. Biomaterials 2002;23:4049–56.

[234] Lin Y-J, Yen C-N, Hu Y-C, Wu Y-C, Liao C-J, Chu I-M. Chondrocytes culture in three-

dimensional porous alginate scaffolds enhanced cell proliferation, matrix synthesis

and gene expression. J Biomed Mater Res A 2009;88:23–33.

[235] Callahan LAS, Ganios AM, McBurney DL, Dilisio MF, Weiner SD, Horton WE, et al.

ECM production of primary human and bovine chondrocytes in hybrid PEG hydrogels

containing type I collagen and hyaluronic acid. Biomacromolecules 2012;13:1625–

31.

[236] Badami AS, Kreke MR, Thompson MS, Riffle JS, Goldstein AS. Effect of fiber

diameter on spreading, proliferation, and differentiation of osteoblastic cells on

electrospun poly(lactic acid) substrates. Biomaterials 2006;27:596–606.

[237] Gough JE, Jones JR, Hench LL. Nodule formation and mineralisation of human

primary osteoblasts cultured on a porous bioactive glass scaffold. Biomaterials

2004;25:2039–46.

[238] Wang B, Cai Q, Zhang S, Yang X, Deng X. The effect of poly (L-lactic acid) nanofiber

orientation on osteogenic responses of human osteoblast-like MG63 cells. J Mech

Behav Biomed Mater 2011;4:600–9.

[239] Jiang J, Nicoll SB, Lu HH. Co-culture of osteoblasts and chondrocytes modulates

cellular differentiation in vitro. Biochem Biophys Res Commun 2005;338:762–70.

[240] Yang PJ, Temenoff JS. Engineering orthopedic tissue interfaces. Tissue Eng Part B

Rev 2009;15:127–41.

Page 230: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

216

[241] Richardson SM, Hoyland JA, Mobasheri R, Csaki C, Shakibaei M, Mobasheri A.

Mesenchymal stem cells in regenerative medicine: opportunities and challenges for

articular cartilage and intervertebral disc tissue engineering. J Cell Physiol

2010;222:23–32.

[242] Grayson WL, Chao P-HG, Marolt D, Kaplan DL, Vunjak-Novakovic G. Engineering

custom-designed osteochondral tissue grafts. Trends Biotechnol 2008;26:181–9.

[243] Dormer NH, Busaidy K, Berkland CJ, Detamore MS. Osteochondral Interface

Regeneration of Rabbit Mandibular Condyle With Bioactive Signal Gradients. J Oral

Maxillofac Surg 2011:0–7.

[244] Fortier LA, Barker JU, Strauss EJ, McCarrel TM, Cole BJ. The role of growth factors

in cartilage repair. Clin Orthop Relat Res 2011;469:2706–15.

[245] Re’em T, Tsur-Gang O, Cohen S. The effect of immobilized RGD peptide in

macroporous alginate scaffolds on TGFbeta1-induced chondrogenesis of human

mesenchymal stem cells. Biomaterials 2010;31:6746–55.

[246] Toh WS, Liu H, Heng BC, Rufaihah AJ, Ye CP, Cao T. Combined effects of

TGFbeta1 and BMP2 in serum-free chondrogenic differentiation of mesenchymal

stem cells induced hyaline-like cartilage formation. Growth Factors 2005;23:313–21.

[247] Arvidson K, Abdallah BM, Applegate LA, Baldini N, Cenni E, Gomez-Barrena E, et al.

Bone regeneration and stem cells. J Cell Mol Med 2011;15:718–46.

[248] Mano JF, Reis RL. Osteochondral defects : present situation and tissue engineering

approaches. J Tissue Eng Regen Med 2007;1:261–73.

[249] Tuli R, Li W-J, Tuan RS. Current state of cartilage tissue engineering. Arthritis Res

Ther 2003;5:235–8.

[250] Kurth T, Hedbom E, Shintani N, Sugimoto M, Chen FH, Haspl M, et al. Chondrogenic

potential of human synovial mesenchymal stem cells in alginate. Osteoarthritis

Cartilage 2007;15:1178–89.

[251] Shintani N, Hunziker EB. Differential effects of dexamethasone on the

chondrogenesis of mesenchymal stromal cells: influence of microenvironment, tissue

origin and growth factor. Eur Cell Mater 2011;22:302–19; discussion 319–20.

Page 231: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

217

[252] Sharma C, Gautam S, Dinda AK, Mishra NC. Cartilage tissue engineering: current

scenario and challenges. Adv Mater Lett 2011;2:90–9.

[253] Beris AE, Lykissas MG, Papageorgiou CD, Georgoulis AD. Advances in articular

cartilage repair. Injury 2005;36 Suppl 4:S14–23.

[254] Im G-I, Ko J-Y, Lee JH. Chondrogenesis of adipose stem cells in a porous polymer

scaffold: influence of the pore size. Cell Transplant 2012;21:2397–405.

[255] Xie X, Wang Y, Zhao C, Guo S, Liu S, Jia W, et al. Comparative evaluation of MSCs

from bone marrow and adipose tissue seeded in PRP-derived scaffold for cartilage

regeneration. Biomaterials 2012;33:7008–18.

[256] Perets A, Baruch Y, Weisbuch F, Shoshany G, Neufeld G, Cohen S. Enhancing the

vascularization of three-dimensional porous alginate scaffolds by incorporating

controlled release basic fibroblast growth factor microspheres. J Biomed Mater Res A

2003;65:489–97.

[257] Rodrigues MT, Jin S, Gomes ME, Reis RL, Atala A, Yoo JJ, et al. Bilayered

constructs aimed at osteochondral strategies: the influence of medium supplements

in the osteogenic and chondrogenic differentiation of amniotic fluid-derived stem

cells. Acta Biomater 2012;8:2795–806.

[258] Abrahamsson CK, Yang F, Park H, Langer R, et al. Chondrogenesis and

Mineralization During In Vitro Culture of Human Mesenchymal Stem Cells on Three-

Dimensional Woven Scaffolds. Tissue Eng Part A 2010;16:3709–16.

[259] Re’em T, Witte F, Willbold E, Ruvinov E, Cohen S. Simultaneous regeneration of

articular cartilage and subchondral bone induced by spatially presented TGF-beta

and BMP-4 in a bilayer affinity binding system. Acta Biomater 2012;8:3283–93.

[260] Lim JJ, Temenoff JS. The effect of desulfation of chondroitin sulfate on interactions

with positively charged growth factors and upregulation of cartilaginous markers in

encapsulated MSCs. Biomaterials 2013;34:5007–18.

[261] Lu HH, Subramony SD, Boushell MK, Zhang X. Tissue engineering strategies for the

regeneration of orthopedic interfaces. Ann Biomed Eng 2010;38:2142–54.

Page 232: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

218

[262] Salgado AJ, Coutinho OP, Reis RL. Bone tissue engineering: state of the art and

future trends. Macromol Biosci 2004;4:743–65.

[263] Sharma B, Elisseeff JH. Engineering structurally organized cartilage and bone

tissues. Ann Biomed Eng 2004;32:148–59.

[264] Zhao J, Duan K, Zhang JWW, Lu X, Weng J. The influence of polymer

concentrations on the structure and mechanical properties of porous

polycaprolactone-coated hydroxyapatite scaffolds. Appl Surf Sci 2010;256:4586–90.

[265] Singh H, Bandyopadhyay A, Bose S. Polymer Coated Tricalcium Phosphate

Scaffolds For Bone Implants n.d.:99164.

[266] Dorozhkin S, Ajaal T. Toughening of porous bioceramic scaffolds by bioresorbable

polymeric coatings. Proc Inst Mech Eng Part H J Eng Med 2009;223:459–70.

[267] Zhao J, Guo LYY, Yang XBB, Weng J. Preparation of bioactive porous HA/PCL

composite scaffolds. Appl Surf Sci 2008;255:2942–6.

[268] Kim H-W, Knowles JC, Kim H-E. Hydroxyapatite porous scaffold engineered with

biological polymer hybrid coating for antibiotic Vancomycin release. J Mater Sci

Mater Med 2005;16:189–95.

[269] Kim H, Yu H, Lee H. Nanofibrous matrices of poly ( lactic acid ) and gelatin polymeric

blends for the improvement of cellular responses. J Biomed Mater Res Part A

2007:25–32.

[270] Dubnika A, Loca D, Berzina-Cimdina L. Functionalized hydroxyapatite scaffolds

coated with sodium alginate and chitosan for controlled drug delivery. Proc Est Acad

Sci 2012;61:193.

[271] Kokubo T, Ito S. Ca , P-rich layer formed on high-strength bioactive 1990;24:331–43.

[272] Hum J, Luczynski KW, Nooeaid P, Newby P, Lahayne O, Hellmich C, et al. Stiffness

Improvement of 45S5 Bioglass ® -Based Scaffolds Through Natural and Synthetic

Biopolymer Coatings: An Ultrasonic Study. Strain 2013;49:431–9.

[273] Draget KI, Smidsrùd PO, Skjåk-brñk PG. Alginates from Algae. In: Rhee AS and SK,

editor. Polysaccharides Polyam. Food Ind. Prop. Prod. Patents, Wiley-VCH Verlag

GmbH & Co. KGaA, Weinheim; 2005, p. 1–30.

Page 233: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

219

[274] Caridade SG, Merino EG, Martins GV, Luz GM, Alves NM, Mano JF. Membranes of

poly(DL-lactic acid)/Bioglass(R) with asymmetric bioactivity for biomedical

applications. J Bioact Compat Polym 2012;27:429–40.

[275] Meng D, Rath SN, Mordan N, Salih V, Kneser U, Boccaccini AR. In vitro evaluation of

45S5 Bioglass®-derived glass-ceramic scaffolds coated with carbon nanotubes. J

Biomed Mater Res A 2011;99:435–44.

[276] Bellucci D, Sola A, Gentile P, Ciardelli G, Cannillo V. Biomimetic coating on bioactive

glass-derived scaffolds mimicking bone tissue. J Biomed Mater Res A

2012;100:3259–66.

[277] Blaker JJ, Nazhat SN, Maquet V, Boccaccini AR. Long-term in vitro degradation of

PDLLA/bioglass bone scaffolds in acellular simulated body fluid. Acta Biomater

2011;7:829–40.

[278] Cao D, Wu Y-P, Fu Z, Tian Y, Li C-J, Gao C, et al. Cell adhesive and growth

behavior on electrospun nanofibrous scaffolds by designed multifunctional

composites. Colloids Surf B Biointerfaces 2011;84:26–34.

[279] Wang B-Y, Fu S-Z, Ni P-Y, Peng J-R, Zheng L, Luo F, et al. Electrospun

polylactide/poly(ethylene glycol) hybrid fibrous scaffolds for tissue engineering. J

Biomed Mater Res A 2011:1–9.

[280] Sun M, Zhou P, Yang SL, Zhou P, Pan L-F, Liu S, et al. Enhanced cell affinity of the

silk fibroin- modified PHBHHx material. J Mater Sci Mater Med 2009;20:1743–51.

[281] García-García JM, Garrido L, Quijada-Garrido I, Kaschta J, Schubert DW, Boccaccini

AR. Novel poly(hydroxyalkanoates)-based composites containing Bioglass® and

calcium sulfate for bone tissue engineering. Biomed Mater 2012;7:054105.

[282] Yao Q, Nooeaid P, Roether JA, Dong Y, Zhang Q, Boccaccini AR. Bioglass®-based

scaffolds incorporating polycaprolactone and chitosan coatings for controlled

vancomycin delivery. Ceram Int 2013;39:7517–22.

[283] Li Y, Liu T, Zheng J, Xu X. Glutaraldehyde-crosslinked chitosan/hydroxyapatite bone

repair scaffold and its application as drug carrier for icariin. J Appl Polym Sci

2013;130:1539–47.

Page 234: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

220

[284] Balcerzak J, Mucha M. Analysis of Model Drug Release Kinetics From Complex

Matrices of Polylactide-Chitosan. Prog Chem Appl Chitin Abd Its. 2010;15:117–26.

[285] Tigani D, Zolezzi C, Trentani F, Ragaini A, Iafisco M, Manara S, et al. Controlled

release of vancomycin from cross-linked gelatine. J Mater Sci Mater Med

2008;19:1325–34.

[286] Hum J, Boccaccini AR. Bioactive glasses as carriers for bioactive molecules and

therapeutic drugs: a review. J Mater Sci Mater Med 2012;23:2317–33.

[287] Domingues ZR, Cortés ME, Gomes TA, Diniz HF, Freitas CS, Gomes JB, et al.

Bioactive glass as a drug delivery system of tetracycline and tetracycline associated

with β-cyclodextrin. Biomaterials 2004;25:327–33.

[288] Zhao L, Yan X, Zhou X, Zhou L, Wang H, Tang J, et al. Mesoporous bioactive

glasses for controlled drug release. Microporous Mesoporous Mater 2008;109:210–5.

[289] Garg T, Singh O, Arora S, Murthy R. Scaffold: a novel carrier for cell and drug

delivery. Crit Rev Ther Drug Carrier Syst 2012;29:1–63.

[290] Mouriño V, Boccaccini AR, Mourin V. Bone tissue engineering therapeutics :

controlled drug delivery in three-dimensional scaffolds Bone tissue engineering

therapeutics : controlled drug delivery in three-dimensional scaffolds. J R Soc

Interface 2010;7:209–27.

[291] Kim H-W, Knowles JC, Kim H-E. Hydroxyapatite/poly(ε-caprolactone) composite

coatings on hydroxyapatite porous bone scaffold for drug delivery. Biomaterials

2004;25:1279–87.

[292] Zhang Y, Zhang M. Calcium phosphate/chitosan composite scaffolds for controlled in

vitro antibiotic drug release. J Biomed Mater Res 2002;62:378–86.

[293] Kundu B, Lemos A, Soundrapandian C, Sen PS, Datta S, Ferreira JMF, et al.

Development of porous HAp and β-TCP scaffolds by starch consolidation with

foaming method and drug-chitosan bilayered scaffold based drug delivery system. J

Mater Sci Mater Med 2010;21:2955–69.

Page 235: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

221

[294] Prabu P, Kim KW, Dharmaraj N, Park JH, Khil MS, Kim HY. Antimicrobial drug

release scaffolds of natural and synthetic biodegradable polymers. Macromol Res

2008;16:303–7.

[295] Pasparakis G, Bouropoulos N. Swelling studies and in vitro release of verapamil from

calcium alginate and calcium alginate-chitosan beads. Int J Pharm 2006;323:34–42.

[296] Wang QQ, Wan C. The Effect of Porosity on the Structure and Properties of Calcium

Polyphosphate Bioceramics. Ceram – Silikáty 2011;55:43–8.

[297] Kokubo T, Takadama H. How useful is SBF in predicting in vivo bone bioactivity?

Biomaterials 2006;27:2907–15.

[298] Dash S, Murthy PN, Nath PC. Kinetic Modeling on Drug Release from Controlled

Drug Delivery Systems. Acta Polyniae Pharm - Drug Res 2010;67:217–23.

[299] Mouriño V, Newby P, Boccaccini AR. Preparation and Characterization of Gallium

Releasing 3-D Alginate Coated 45S5 Bioglass® Based Scaffolds for Bone Tissue

Engineering. Adv Eng Mater 2010;12:B283–B291.

[300] Lai HL, Abu’Khalil A, Craig DQM. The preparation and characterisation of drug-

loaded alginate and chitosan sponges. Int J Pharm 2003;251:175–81.

[301] Yan S, Xiaoqiang L, Shuiping L, Hongsheng W, Chuanglong H. Fabrication and

Properties of PLLA-Gelatin Nanofibers by Electrospinning. J Appl Polym Sci

2010;117:542–7.

[302] Heijink A, Gomoll AH, Madry H, Drobnič M, Filardo G, Espregueira-Mendes J, et al.

Biomechanical considerations in the pathogenesis of osteoarthritis of the knee. Knee

Surg Sports Traumatol Arthrosc 2012;20:423–35.

[303] Breyner NM, Zonari AA, Carvalho JL, Gomide VS, Gomes D, Góes AM. Cartilage

Tissue Engineering Using Mesenchymal Stem Cells and 3D Chitosan Scaffolds – In

vitro and in vivo Assays. Biomater. Sci. Eng., 2009, p. 211–26.

[304] Lee NK, Oh HJ, Hong CM, Suh H, Hong SH. Comparison of the synthetic

biodegradable polymers, polylactide (PLA), and polylactic-co-glycolic acid (PLGA) as

scaffolds for artificial cartilage. Biotechnol Bioprocess Eng 2009;14:180–6.

Page 236: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

222

[305] Nejadnik H, Daldrup-link HE. Engineering stem cells for treatment of osteochondral

defects. Skeletal Radiol 2012;41:1–4.

[306] Bhattarai BN, Li Z, Edmondson D, Zhang M, Bhattarai N. Alginate-Based

Nanofibrous Scaffolds: Structural, Mechanical, and Biological Properties. Adv Mater

2006;18:1463–7.

[307] Lee YR, Shin DS, Kwon OW, Park WH, Choi HG, Han SS, et al. Preparation of

Atactic Poly ( vinyl alcohol )/ Sodium Alginate Blend Nanowebs by Electrospinning. J

Appl Pol 2007;106:1337–42.

[308] Moon S, Farris RJ. Electrospinning of Heated Gelatin-Sodium Alginate-Water

Solutions. Polym Eng Sci 2009:1616–20.

[309] Nie H, He A, Zheng J, Xu S, Li J, Han CC. Effects of chain conformation and

entanglement on the electrospinning of pure alginate. Biomacromolecules

2008;9:1362–5.

[310] Bonino CA, Krebs MD, Saquing CD, Jeong SI, Shearer KL, Alsberg E, et al.

Electrospinning alginate-based nanofibers: From blends to crosslinked low molecular

weight alginate-only systems. Carbohydr Polym 2011;85:111–9.

[311] Ma PX, Zhang R. Synthetic nano-scale fibrous extracellular matrix. John Wiley Sons,

Inc 1998:27–30.

[312] Cao D, Fu Z, Li C. Heat and compression molded electrospun poly(l-lactide)

membranes: Preparation and characterization. Mater Sci Eng B 2011;176:900–5.

[313] Silva JM, Georgi N, Costa R, Sher P, Reis RL, Van Blitterswijk CA, et al.

Nanostructured 3D constructs based on chitosan and chondroitin sulphate

multilayers for cartilage tissue engineering. PLoS One 2013;8:1–11.

[314] Gelinsky M, Eckert M, Despang F. Biphasic , but monolithic scaffolds for the therapy

of osteochondral defects. Int J Mat Res 2007;98:749–55.

[315] Stenhamre H, Nannmark U, Lindahl A, Gatenholm P, Brittberg M. Influence of pore

size on the redifferentiation potential of human articular chondrocytes in poly (

urethane urea ) scaffolds 2011:578–88.

Page 237: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

223

[316] Woodfield TBF. Interfacial Shear Strength criteria for Tissue-engineering cartilage

anchored to Porous synthetic Scaffolds. 2000.

[317] Nussinovitch A. Production, properties, and applications of hydrocolloid cellular

solids. Mol Nutr Food Res 2005;49:195–213.

[318] Ashby MF. The mechanical properties of cellular solids. Metall Trans A

1983;14A:1755–69.

[319] Ghosh S, Gutierrez V, Fernández C, Rodriguez-Perez MA, Viana JC, Reis RL, et al.

Dynamic mechanical behavior of starch-based scaffolds in dry and physiologically

simulated conditions: effect of porosity and pore size. Acta Biomater 2008;4:950–9.

[320] Planell JA. Development and Characterisation of Completely Degradable Composite

Tissue Engineering Scaffolds 2007.

[321] Pillay V, Dangor CM, Govender T, Moopanar KR, Hurbans N. Drug Release

Modulation from Cross-Linked Calcium Alginate Microdiscs : Swelling , Compression

, and Stability of the Hydrodynamically-Sensitive Calcium Alginate Matrix and the

Associated Drug Release Mechanisms. Drug Deliv 1998;5:35–46.

[322] Xie L, Jiang M, Dong X, Bai X, Tong J, Zhou J. Controlled Mechanical and Swelling

Properties of Poly ( vinyl alcohol )/ Sodium Alginate Blend Hydrogels Prepared by

Freeze – Thaw Followed by Ca 21 Crosslinking. J Appl Polym Sci 2012;124:823–31.

[323] Huang M-H, Yang M-C. Swelling and biocompatibility of sodium alginate/poly( γ -

glutamic acid) hydrogels. Polym Adv Technol 2010;21:561–7.

[324] Bajpai AK, Shukla SK, Bhanu S, Kankane S. Responsive polymers in controlled drug

delivery. Prog Polym Sci 2008;33:1088–118.

[325] Hoffman a S. Hydrogels for biomedical applications. Ann N Y Acad Sci 2001;944:62–

73.

[326] Kenawy E-R, Bowlin GL, Mansfield K, Layman J, Simpson DG, Sanders EH, et al.

Release of tetracycline hydrochloride from electrospun poly(ethylene-co-

vinylacetate), poly(lactic acid), and a blend. J Control Release 2002;81:57–64.

Page 238: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

224

[327] Safi S, Morshed M, Ravandi SAH, Ghiaci M. Study of Electrospinning of Sodium

Alginate , Blended Solutions of Sodium Alginate / Poly ( vinyl alcohol ) and Sodium

Alginate / Poly ( ethylene oxide ). J Appl Polym Sci 2007;104:3245–55.

[328] Li W, Li X, Chen Y, Li X, Deng H, Wang T, et al. Poly(vinyl alcohol)/sodium

alginate/layered silicate based nanofibrous mats for bacterial inhibition. Carbohydr

Polym 2013;92:2232–8.

[329] Jeong SI, Krebs MD, Bonino CA, Khan SA, Alsberg E. Electrospun alginate

nanofibers with controlled cell adhesion for tissue engineering. Macromol Biosci

2010;10:934–43.

[330] Reneker DH, Yarin AL. Electrospinning jets and polymer nanofibers. Polymer (Guildf)

2008;49:2387–425.

[331] Garg K, Bowlin GL. Electrospinning jets and nanofibrous structures. Biomicrofluidics

2011;5:13403.

[332] Saarai A, Kasparkova V, Sedlacek T, Saha P. On the development and

characterisation of crosslinked sodium alginate/gelatine hydrogels. J Mech Behav

Biomed Mater 2013;18:152–66.

[333] Cho A, Shin DM, Jung HW, Hyun JC, Lee JS, Cho D, et al. Effect of Annealing on the

Crystallization and Properties of Electrospun Polylatic Acid and Nylon 6 Fibers. J

Appl Polym Sci 2010;120:752–8.

[334] Ramaswamy S, Clarke LI, Gorga RE. Morphological, mechanical, and electrical

properties as a function of thermal bonding in electrospun nanocomposites. Polymer

(Guildf) 2011;52:3183–9.

[335] Inai R, Kotaki M, Ramakrishna S. Structure and properties of electrospun PLLA

single nanofibres. Nanotechnology 2005;16:208–13.

[336] Nguyen T-H. Fabrication and characterization of cross-linked gelatin electro-spun

nano-fibers. J Biomed Sci Eng 2010;03:1117–24.

[337] Huang Z-M, Zhang YZ, Ramakrishna S, Lim CT. Electrospinning and mechanical

characterization of gelatin nanofibers. Polymer (Guildf) 2004;45:5361–8.

Page 239: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

225

[338] Panzavolta S, Gioffrè M, Focarete ML, Gualandi C, Foroni L, Bigi A, et al.

Electrospun gelatin nanofibers: optimization of genipin cross-linking to preserve fiber

morphology after exposure to water. Acta Biomater 2011;7:1702–9.

[339] Li W-J, Laurencin CT, Caterson EJ, Tuan RS, Ko FK. Electrospun nanofibrous

structure: a novel scaffold for tissue engineering. J Biomed Mater Res 2002;60:613–

21.

[340] Seth D. McCullen, Helene Autefage AC, Eileen Gentleman and MMS. Anisotropic

Fibrous Scaffolds for Articular Cartilage Regeneration. Tissue Eng Part A

2012;18:2073–83.

[341] Rodrigues AA, Batista NA, Bavaresco VP, Baranauskas V, Ceragioli HJ, Peterlevitz

AC, et al. Polyvinyl alcohol associated with carbon nanotube scaffolds for osteogenic

differentiation of rat bone mesenchymal stem cells. Carbon N Y 2012;50:450–9.

[342] Martin I, Schaefer D, Dozin B. Repair of Osteochondral Lesions Current Clinical

Treatments of Osteochondral Lesions. Madame Curie, n.d.

[343] Awad HA, Wickham MQ, Leddy HA, Gimble JM, Guilak F. Chondrogenic

differentiation of adipose-derived adult stem cells in agarose, alginate, and gelatin

scaffolds. Biomaterials 2004;25:3211–22.

[344] Correia CR, Moreira-Teixeira LS, Moroni L, Reis RL, van Blitterswijk CA, Karperien

M, et al. Chitosan scaffolds containing hyaluronic acid for cartilage tissue

engineering. Tissue Eng Part C Methods 2011;17:717–30.

[345] Jin CZ, Cho J, Choi BH, Wang LM, Kim MS, et al. The Maturity of Tissue-Engineered

Cartilage In Vitro Affects the Repairability for Osteochondral Defect. Tissue Eng Part

A 2011;17:3057–65.

[346] Bosch EVD, Gielens C, van den Bosch E. Gelatin degradation at elevated

temperature. Int J Biol Macromol 2003;32:129–38.

[347] Lichte P, Pape HC, Pufe T, Kobbe P, Fischer H. Scaffolds for bone healing:

concepts, materials and evidence. Inj Int J Care Inj 2011;42:569–73.

Page 240: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

226

[348] Tanase CE, Sartoris A, Popa MI, Verestiuc L, Unger RE, Kirkpatrick CJ. In vitro

evaluation of biomimetic chitosan-calcium phosphate scaffolds with potential

application in bone tissue engineering. Biomed Mater 2013;8:025002.

[349] Al-Nasiry S, Geusens N, Hanssens M, Luyten C, Pijnenborg R. The use of Alamar

Blue assay for quantitative analysis of viability, migration and invasion of

choriocarcinoma cells. Hum Reprod 2007;22:1304–9.

[350] Golub EE, Harrison G, Taylor AG, Camper S, Shapiro IM. The role of alkaline

phosphatase in cartilage mineralization. Bone Miner 1992;17:273–8.

[351] Bretcanu O, Misra S, Roy I, Renghini C, Fiori F, Boccaccini AR, Vehid S. In vitro

biocompatibility of 45S5 Bioglass-derived glass – ceramic scaffolds coated with poly

( 3-hydroxybutyrate ). J Tissue Eng Regen Med 2009;3:139–48.

[352] Chen QZ, Efthymiou A, Salih V, Boccaccini AR. Bioglass-derived glass-ceramic

scaffolds: study of cell proliferation and scaffold degradation in vitro. J Biomed Mater

Res A 2008;84:1049–60.

[353] Yang C, Frei H, Rossi FM, Burt HM. Differential in vitro and in vivo responses of bone

marrow stromal cells on novel porous gelatin – alginate scaffolds. J Tissue Eng

Regen Med 2009;3:601–14.

[354] Li Z, Zhang M. Chitosan-alginate as scaffolding material for cartilage tissue

engineering. J Biomed Mater Res A 2005;75:485–93.

[355] Tiğli RS, Gümüşderelioğlu M. Evaluation of alginate-chitosan semi IPNs as cartilage

scaffolds. J Mater Sci Mater Med 2009;20:699–709.

[356] Kim G, Ahn S, Kim Y, Cho Y, Chun W. Coaxial structured collagen–alginate

scaffolds: fabrication, physical properties, and biomedical application for skin tissue

regeneration. J Mater Chem 2011;21:6165.

[357] Lee H, Ahn S-H, Kim GH. Three-Dimensional Collagen/Alginate Hybrid Scaffolds

Functionalized with a Drug Delivery System (DDS) for Bone Tissue Regeneration.

Chem Mater 2012;24:881–91.

Page 241: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

227

[358] Chang C, Kuo T, Lin F, Wang J-H, Hsu Y, Huang H-T, et al. Tissue engineering-

based cartilage repair with mesenchymal stem cells in a porcine model. J Orthop Res

2011;29:1874–80.

[359] Varghese S, Hwang NS, Canver AC, Theprungsirikul P, Lin DW, Elisseeff J.

Chondroitin sulfate based niches for chondrogenic differentiation of mesenchymal

stem cells. Matrix Biol 2008;27:12–21.

[360] Chang C. Gelatin–chondroitin–hyaluronan tri-copolymer scaffold for cartilage tissue

engineering. Biomaterials 2003;24:4853–8.

[361] Muzzarelli RA, Greco F, Busilacchi A, Sollazzo V, Gigante A. Chitosan, hyaluronan

and chondroitin sulfate in tissue engineering for cartilage regeneration: A review.

Carbohydr Polym 2012;89:723–39.

[362] Jerosch J. Effects of Glucosamine and Chondroitin Sulfate on Cartilage Metabolism

in OA: Outlook on Other Nutrient Partners Especially Omega-3 Fatty Acids. Int J

Rheumatol 2011;2011:969012.

[363] Huang D. Effect of Extracellular Chondroitin Sulafte on Cultured Chondrocytes. J Cell

Biol 1974;62:881–6.

[364] Steinmetz NJ, Bryant SJ. Chondroitin sulfate and dynamic loading alter

chondrogenesis of human MSCs in PEG hydrogels. Biotechnol Bioeng

2012;109:2671–82.

[365] Bitter T, Muir HM. A modified uronic acid carbazole reaction. Anal Biochem

1962;334:330–4.

[366] Schulz RM, Bader A. Cartilage tissue engineering and bioreactor systems for the

cultivation and stimulation of chondrocytes. Eur Biophys J 2007;36:539–68.

[367] June RK, Neu CP, Barone JR, Fyhrie DP. Polymer Mechanics as a Model for Short-

Term and Flow-Independent cartilage Viscoelasticity. Mater Sci Eng C MAter Biol

Appl 2011;31:1–18.

[368] Fajardo AR, Silva MB, Lopes LC, Piai JF, Rubira AF, Muniz EC. Hydrogel based on

an alginate–Ca2+/chondroitin sulfate matrix as a potential colon-specific drug

delivery system. RSC Adv 2012;2:11095–103.

Page 242: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

228

[369] Lawrie G, Keen I, Drew B, Chandler-Temple A, Rintoul L, Fredericks P, et al.

Interactions between alginate and chitosan biopolymers characterized using FTIR

and XPS. Biomacromolecules 2007;8:2533–41.

[370] Li A, Xiong S. Preparation and Structure Analysis of Chondroitin Sulfate from Pig

Laryngeal Cartilage. 2010 4th Int Conf Bioinforma Biomed Eng 2010:1–5.

[371] Servaty R, Schiller J, Binder H, Arnold K. Hydration of polymeric components of

cartilage--an infrared spectroscopic study on hyaluronic acid and chondroitin sulfate.

Int J Biol Macromol 2001;28:121–7.

[372] Ciardelli G, Chiono V, Vozzi G, Pracella M, Ahluwalia A, Barbani N, et al. Blends of

poly-(epsilon-caprolactone) and polysaccharides in tissue engineering applications.

Biomacromolecules 2005;6:1961–76.

[373] Gabbott P. Principles and Applications of Thermal Analysis. Oxford, UK: Blackwell

Publishing Ltd; 2008.

[374] Al-rawajfeh AE, Al-salah HA, Al-rhael I, Univesity TT. Miscibility, Crystallinity and

Morphology of Polymer Blends of Polyamide-6/ Poly (β-hydroxybutyrate). Jordan J

Chem 2006;1:155–70.

[375] Rimt PB, Runt JP. Melting Point Depression in Crystalline/ Compatible Polymer

Blends 1984:1520–6.

[376] Manthey JA. mFold , Delta G , and Melting Temperature : What Does it Mean ?

2005:1–8.

[377] Chung C, Yeon J, Yoon I, Hwang H, Balakrishnan P. Colloids and Surfaces B :

Biointerfaces Interpenetrating polymer network ( IPN ) scaffolds of sodium

hyaluronate and sodium alginate for chondrocyte culture. Colloids Surfaces B

Biointerfaces 2011;88:711–6.

[378] Kim IL, Mauck RL, Burdick JA. Hydrogel design for cartilage tissue engineering: a

case study with hyaluronic acid. Biomaterials 2011;32:8771–82.

[379] Zacharias Dische. A New Specific Color Reaction of Hexuronic Acids. J Biol Chem

1947;167:189–98.

Page 243: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

229

[380] Park JS, Yang HJ, Woo DG, Yang HN, Na K, Park K-H. Chondrogenic differentiation

of mesenchymal stem cells embedded in a scaffold by long-term release of TGF-beta

3 complexed with chondroitin sulfate. J Biomed Mater Res A 2010;92:806–16.

[381] Longobardi L, Rear LO, Aakula S, Johnstone B, Shimer K, Chytil A, et al. Effect of

IGF-I in the Chondrogenesis of Bone Marrow Mesenchymal Stem Cells in the

Presence or Absence of TGF-beta Signaling. J Bone Miner Res 2006;21:626–36.

[382] Im G, Jung N, Tae S. Chondrogenic Differentiation of Mesenchymal Stem Cells

Isolated from Patients in Late Adulthood : Tissue Eng 2006;12:527–36.

[383] Callister WD. Materials Science and Engineering: An Introduction. seven. John Wiley

&Sons, Inc.; 2006.

Page 244: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

230

Page 245: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

LIST OF FIGURES

Figure 1. 1 The scheme of knee osteoarthritis (joint degeneration disease), which is the

result of cartilage wearing out in the load bearing joint [3]. ............................. 2

Figure 2. 1 Anatomy of the knee joint, which is the most common case found in joint

degeneration disease (Source: lpch.org [21]), demonstrating arrangement of

ECM and organization of chondrocytes along different zones in cartilage

[19], and showing the structure of cancellous bone as subchondral bone

[12]. .................................................................................................................. 5

Figure 2. 2 Schematic diagram showing the gelation-mechanism of alginate and

calcium cations by the formation of egg-box structure (Courtesy K. Kashima

and M. Imai [135])..........................................................................................16

Figure 2. 3 Schematic diagram of the foam replication technique employed to produce

3D porous bioceramics- and bioactive glass-based scaffolds (according to

[28,184]). .......................................................................................................22

Figure 2. 4 (A) The schematic diagram of the freeze-drying (lyophilazation) process

showing also the phase diagram of water representing the mechanism of

freeze-drying (modified from [186,189]). .......................................................23

Figure 2. 5 Schematic diagram of the electrospinning process in horizontal direction,

which is composed of voltage supply, syringe and needle, syringe pump and

collector; the SEM image shows electrospun PLLA fibers. ...........................26

Figure 3. 1 Schematic diagram of entire tasks carried out in the dissertation thesis. .....41

Figure 4. 1 SEM images of 45S5 Bioglass®-based scaffolds fabricated by foam

replication technique: (A) 3D porous structure and (B) surface of scaffold

struts. .............................................................................................................49

Figure 4. 2 X-ray patterns of as-received Bioglass® and as-sintered 45S5 Bioglass

®-

based scaffolds. The major peaks of the phase Na2Ca2Si3O9 are marked by

■. ....................................................................................................................50

Page 246: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

II

Figure 4. 3 SEM images of Alg coated 45S5 Bioglass®-based scaffolds with variable

concentrations, including (A, B) 1 wt/v %, (C, D) 1.5 wt/v % and (E, F) 2 wt/v

%, the scaffolds were coated once for 5 min soaking time. ..........................54

Figure 4. 4 SEM images of Gel coated 45S5 Bioglass®-based scaffolds with variable

concentrations, including (A, B) 1.5 wt/v %, (C, D) 3 wt/v %, (E, F) 5 wt/v %,

the scaffolds were coated once on for 5 min soaking time and (G, H) 1.5

wt/v % coated for 5 min and 3 dipping cycles. ..............................................55

Figure 4. 5 SEM images of PDLLA coated 45S5 Bioglass®-based scaffolds with variable

concentrations, including (A, B) 3 wt/v %, (C, D) 5 wt/v % and (E, F) 8 wt/v

%, the scaffolds were coated once for 5 min soaking time. ..........................56

Figure 4. 6 PHBHHx coated 45S5 Bioglass®-based scaffolds with variable

concentrations, soaking times and number of dipping times, including (A, B)

5 wt/v %, 5 min of soaking and one coating cycle, and (C, D) 1 wt/v %, 15

sec of soaking time and 30 coating cycles. ...................................................56

Figure 4. 7 (A) Representative compressive stress-strain curves of biodegradable

polymer coated Bioglass®-based composite scaffolds in comparison with the

curve of uncoated Bioglass®-based scaffolds, (B) normalized mechanical

properties (compressive modulus and strength) of coated scaffolds

compared to those of uncoated scaffolds (as reference) and (C) appearance

of coated scaffolds after compression load. * indicates significantly different

mechanical properties of coated scaffolds in comparison with those of

uncoated scaffolds, and # indicates significantly different mechanical

properties of synthetic polymer coated scaffolds in comparison with the

properties of natural polymer coated scaffolds (p 0.05). ............................59

Figure 4. 8 (A) % weight loss of biodegradable polymer coated 45S5 Bioglass®-based

scaffolds after 1, 3, 7, 14 and 28 days of immersion in SBF and (B) variation

of the pH of the SBF solution. .......................................................................62

Page 247: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

III

Figure 4. 9 SEM images of as-sintered 45S5 Bioglass®-based scaffolds after (A) 1, (B)

3, (C) 14, and (D) 28 days of immersion in SBF, showing possible formation

of HCA indicated qualitatively by the morphology of the deposited structures.63

Figure 4. 10 (A) FTIR spectra of as-sintered 45S5 Bioglass®-based scaffolds after

different immersion times in SBF, in comparison with the scaffolds before

immersion. The characteristic peaks of HCA are marked by ▲ and ■ and (B)

XRD patterns of as-sintered 45S5 Bioglass®-based scaffolds after different

immersion times in SBF, in comparison with the scaffolds before immersion.

The major peak of HCA is marked by , while the crystalline peaks of

Na2Ca2Si3O9 were marked by ■. ...................................................................64

Figure 4. 11 SEM images of biodegradable polymer coated 45S5 Bioglass®-based

composite scaffolds after immersion in SBF, showing formation of HCA: (A,

B, C) Alg-c-BG, (D, E, F) Gel-c-BG, (G, H, I), PDLLA-c-BG and (J, K, L)

PHBHHx-c-BG scaffolds for 1, 3 and 28 days, respectively. ........................68

Figure 4. 12 FTIR spectra of biodegradable polymer coated 45S5 Bioglass®-based

scaffolds after 28 days of immersion in SBF: (A) Alg-c-BG, (B) Gel-c-BG, (C)

PDLLA-c-BG, and (D) PHBHHx-c-BG scaffolds. The characteristic peaks of

HCA are marked by ▲ and the characteristic peaks of polymer coatings are

marked by ■. ..................................................................................................69

Figure 5. 1 (A) Scheme of the capillarity test of Bioglass®-based scaffolds, showing the

effect of surface chemistry on the permeability of the porous scaffolds and

(B) photographs representing the coated scaffolds during the capillarity test.77

Figure 5. 2 Contact angles of Bioglass®-based scaffolds showing the surface wettability

of different coatings. * indicates the significant difference (p 0.05) of the

modified coatings on the Bioglass-based scaffolds in comparison with PL-

c-BG scaffolds. ..............................................................................................77

Figure 5. 3 SEM images of the scaffolds showing the morphological porous structure

and morphology of a coating surface of: (A, B) PL/P123-c-BG scaffolds, (C,

Page 248: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

IV

D) T-BG scaffolds, (E, F) T-Alg-c-(PL/P123-c-BG) scaffolds and (G, H) T-

Gel-c-(PL/P123-c-BG) scaffolds. ...................................................................79

Figure 5. 4 Mechanical properties of polymer coated Bioglass scaffolds: (A)

representative compressive stress-strain curves and (B) average elastic

modulus and average compressive strength of the scaffolds. * (p 0.05)

indicates the statistical significance of compressive mechanical properties of

coated scaffolds, compared to those of uncoated T-BG scaffolds. ...............80

Figure 5. 5 FTIR spectra of TCH, BG, TCH-loaded Bioglass scaffolds and TCH-loaded

polymer coated Bioglass®-based scaffolds. ..................................................82

Figure 5. 6 (A) Drug release profile and (B) degradation behavior of TCH-loaded

polymer (Alg and Gel) coated Bioglass scaffolds. .......................................84

Figure 5. 7 SEM images of TCH-loaded polymer coated Bioglass-based scaffolds after

in vitro release in PBS for 14 days: (A) T-Alg-c-(PL/P123-c-BG); the arrows

predicting the PL/P123 coating and (B) T-Gel-c-(PL/P123-c-BG) scaffolds;

dashed arrow depicting the Bioglass® struts and solid arrow line predicting

PL/P123 coating, and SEM images of TCH-loaded polymer coated

Bioglass-based scaffolds after immersion in SBS for 14 days: (C) T-Alg-c-

(PL/P123-c-BG and (D) T-Gel-c-(PL/P123-c-BG) scaffolds; solid arrows

depicting polymer coating. .............................................................................85

Figure 5. 8 FTIR spectra of T-Alg- and T-Gel-c-(PL/P123-c-BG) scaffolds after 14 days

of immersion in PBS. .....................................................................................86

Figure 6. 1 Optical microscopy images showing 3 wt/v % Alg-gel after the introduction of

0.1 M CaCl22H2O into Alg solution: (A) superficial surface (plan-view image)

and (B) cross-section of Alg-gel. SEM images of 3 wt/v % Alg-foam after

lyophilized: (C) plan-view and (D) cross-section SEM image. ......................97

Figure 6. 2 Optical photograph showing the appearance of 3 wt/v% Alg-foams obtained

with variation of CaCl22H2O concentrations (0.1 – 1 M). .............................99

Figure 6. 3 Comparison of weight loss as a function of immersion time in DI H2O of Alg-

foams without (w/o) and with (w) crosslinking by immersion in 0.5 M

Page 249: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

V

CaCl22H2O for 4 h (the inset represents the weight loss of Alg-foams

without crosslinking). .....................................................................................99

Figure 6. 4 The effect of Alg concentrations on the porosity and density of Alg-foams.100

Figure 6. 5 SEM images (in plan-view) of (A, B) 2 wt/v %, (D, E) 3 wt/v % and (G, H) 4

wt/v % Alg-foams, included distribution of pore size: (C) 2 wt/v %, (F) 3 wt/v

% and (I) 4 wt/v % Alg-foams. .....................................................................101

Figure 6. 6 (A) Representative compressive stress-strain curves of 2, 3 and 4 wt/v %

Alg-foams and (B) the mechanical properties, including elastic modulus and

compressive strength of the foams as a function of concentrations. ..........103

Figure 6. 7 Dynamic mechanical properties of 3 wt/v % Alg-foams in compression mode

presenting the storage modulus (E’) and the loss factor (tan ) as a function

of frequency, in both dry and wet state. ......................................................104

Figure 6. 8 DSC thermogram: (A) the 1st heating and (B) the 2

nd heating cycle runs of

Alg-foams with and without crosslinking. ....................................................105

Figure 6. 9 (A) Weight change of 3 wt/v % Alg-foams as a function of immersion time in

different media, including DI H2O, PBS and SBF. (B) ATR-FTIR spectra of

the foam after 7 days of immersion in the media. .......................................107

Figure 6. 10 Optical microscopy images of electrospun PLLA fibers obtained by using the

conditions in (A) Trial 4 and (B) Trail 13, according to electrospinning

conditions used in Table 6.1. .......................................................................110

Figure 6. 11 Optical microscopic images of electrospun Alg/Gel fibers obtained by using

the conditions in (A) Trial 9 and (B) Trail 12, according to electrospinning

conditions used in Table 6.2. .......................................................................113

Figure 6. 12 SEM images of PLLA fibers, which were deposited for (A) 2 h and (C) 9 h,

and Alg/Gel fibers, which were deposited for (B) 2 h and (D) 9 h. The

distribution of fiber diameters of both fiber types is included. .....................114

Figure 6. 13 (A) FTIR spectra of Alg/Gel fibers before and after GA crosslinking (The

inset represents the change of amide I peak before and after crosslinking)

Page 250: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

VI

and (B) mechanism of the crosslinking reaction between Gel with GA,

according to [78]. .........................................................................................116

Figure 6. 14 XRD patterns of PLLA fibers showing the reduction of crystallinity in

comparison with PLLA cast films (■ and indicate - and β-crystalline phase

of PLLA, respectively). ................................................................................117

Figure 6. 15 DSC thermogram of PLLA fibers, indicating Tg, Tc, and Tm. .......................118

Figure 6. 16 (A) Representative tensile stress-strain curve of PLLA fibrous meshes in

comparison with the typical curve of PLLA cast films (the inset) and (B) the

values of elastic modulus, tensile strength and elongation at break of PLLA

under tension deformation mode. ................................................................120

Figure 6. 17 Summary of Young’s modulus values of electrospun fibers with respect to

fiber diameter, which were collected from recent literature reports

[238,278,301,306,337–340], mainly on polyesters, Gel and Alg. The star

indicates the position of PLLA fibers obtained in the present work. ............121

Figure 7. 1 The schematic diagram of the four types of multilayered scaffolds for

osteochondral tissue engineering developed in this project. ......................125

Figure 7. 2 Optical microscopic images showed the appearance of three different

approaches of multilayered scaffolds, including (A, a) system A: monolithic

Alg/Alg-c-BG biphasic scaffold, (B, b) system B: integrated Alg/Alg-c-BG

bilayered scaffold and (C, c) system C: integrated electrospun PLLA

fibers/PDLLA-c-BG bilayered scaffold (the inset shows a plan-view of fibers

integrated on the struts of the Bioglass-based scaffold). ..........................130

Figure 7. 3 SEM images showing cross-sections of four different types of multilayered

scaffolds: (A, a) system A, (B, b) system B, (C, c) system C and (D, d)

system D (the dashed line marks the interface between the two phases). .131

Figure 7. 4 (A) Representative stress-strain curves of the bilayered scaffolds (System A

vs. System B) and (B) distribution of the strength at break values of the

scaffolds in systems A and B (the red dashed line is included for the visual

aid). ..............................................................................................................133

Page 251: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

VII

Figure 7. 5 Representative compressive stress-strain curve of integrated bilayered

scaffold (system B) in comparison with the curves of Alg-foam and Alg-c-BG

scaffold. .......................................................................................................134

Figure 7. 6 SEM images of integrated bilayered scaffold (system B) after immersion in

SBF for 1 and 28 days (the dashed line indicates the interface between the

two phases). ................................................................................................136

Figure 7. 7 ATR-FTIR spectra of Alg-foam after immersion in SBF for 28 days in order

to confirm non-mineralization of the foam (the inset shows the absorption

bands in the wavenumber 1200 - 850 cm-1

). ...............................................137

Figure 7. 8 SEM images of integrated bilayered scaffold (system C) after immersion in

SBF for 1 and 28 days. ................................................................................138

Figure 7. 9 SEM image of integrated bilayered scaffold (system D) after immersion in

SBF for 28 days (the dashed line indicates interface between the Alg/Gel

fiber mesh and the Alg-c-BG scaffold and the dashed circles indicate the

interconnection between the fibers formed after immersion in SBF). .........139

Figure 8. 1 Relative LDH activity of MG-63 osteoblast-like cells cultured on uncoated

BG, Alg-c-BG and RGD-Alg-c-BG scaffolds. The results presenting the

difference of optical densities are presented as mean ± SD (n = 4). * (p

0.05) indicates significant difference of different scaffolds at different culture

times. ...........................................................................................................147

Figure 8. 2 Fluorescent microscopic images of MG-63 osteoblast-like cells-seeded BG,

Alg-c-BG and RGD-Alg-c-BG scaffolds after 3 and 14 days in culture by

using DAPI stain for cell nuclei. ...................................................................148

Figure 8. 3 Cell metabolic activity of MG-63 osteoblast-like cells cultured on uncoated

BG, Alg-c-BG and RGD-Alg-c-BG scaffolds. The results in % AB reduction

are presented as mean ± SD (n = 6). * (p 0.05) indicates significant

difference of different scaffold types at different culture times. ...................150

Figure 8. 4 ALP activity up to 21 days of MG-63 osteoblast-like cells cultured on

uncoated BG, Alg-c-BG and RGD-Alg-c-BG scaffolds. The results are

Page 252: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

VIII

reported as mean ± SD (n = 4). * p 0.05 indicates significant difference of

results for different scaffold types at different culture times. .......................152

Figure 8. 5 SEM cross-sectioned images showing MG-63 cells-seeded BG, Alg-c-BG

and RGD-Alg-c-BG scaffolds after cultured for 14 days. ............................153

Figure 8. 6 Confocal microscopic images of MG-63 cells cultured on BG, Alg-c-BG and

RGD-Alg-c-BG scaffolds, stained with OsteoImage (green), after 3,14 and

21 days. .......................................................................................................154

Figure 9. 1 SEM images of Alg/ChS-foams in (A) plan-view and (B) cross-section; (C)

shows a histogram of the pore size distribution of Alg/ChS-foams and (D) is

a scheme showing the reptation model in the case of Alg/ChS blend,

according to [367,368]. ................................................................................167

Figure 9. 2 (A) ATR-FTIR spectrum and (B) EDX spectra of the Alg/ChS-foam in

comparison to the spectra of pure Alg- and pure ChS-foams, which both

results confirm the existence of ChS in the foam. .......................................168

Figure 9. 3 DSC thermograms: (A) the 1st heating and (B) the 2

nd heating cycle of

Alg/ChS- and Alg-foams. .............................................................................169

Figure 9. 4 Water absorption in % of Alg- and Alg/ChS-foams, in comparison to the

water absorption of natural articular cartilage (*), as referenced in the

literature [6]. ................................................................................................170

Figure 9. 5 Degradation profile of Alg/ChS- and Alg-foams in PBS, evaluated by %

weight change with respect to the immersion time. ....................................171

Figure 9. 6 (A) Representative compressive stress-strain curves of the Alg/ChS- and

Alg-foam; (B) mechanical properties of Alg/ChS and Alg-foams. ...............173

Figure 9. 7 Dynamic mechanical analysis of Alg/ChS- and Alg-foams: storage modulus

(E‘) and tan as a function of frequency; note the logarithmic scaling. ......173

Figure 9. 8 Cumulative release of ChS from Alg/ChS-foams immersed in the PBS

solution (pH 7.4, 37 °C) measured by using carbazole reaction, as

described in Section 9.2.3. ..........................................................................174

Page 253: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

IX

Figure 9. 9 The viability of primary porcine chondrocytes on Alg-foams after 7 days of

culture: (A) static culturing and (B) dynamic culturing (on a rotatory device,

12 rpm at 15 C). Viable cells are green, dead cells appear red. ...............175

Figure 9. 10 Histological evaluation of HE and AB staining of porcine chondrocytes-

seeded Alg-foams after 7 days of static and dynamic culturing. Using AB

staining, the foam stains unspecifically blue. Cell nuclei are violet, sulfated

GAGs within the cell clusters are characterized by a faint blue staining.....176

Figure 9. 11 Col II and Col I immunolabeling of porcine chondrocytes-seeded Alg-foams

after culturing for 7 days, observed by confocal microscopy. Cell nuclei are

stained in blue, Col is stained in green color. ..............................................177

Figure 9. 12 Cell viability (FDA/PI live-dead assay) of MSCs-seeded (A, C) Alg- and (B,

D) Alg/ChS-foams after 7 days of culture. Viable cells are green, dead cells

appear red. ..................................................................................................178

Figure 9. 13 Cell viability (FDA/PI live-dead assay) of porcine chondrocytes-seeded (A)

Alg- and (B) Alg/ChS-foams after 14 days of culture. Viable cells are green,

dead cells appear red. .................................................................................179

Figure 9. 14 Histological evaluation (HE and AB staining) of MSCs-seeded Alg- and

Alg/ChS-foams after 7 days of culture (without chondrogenic induction). For

alcian blue staining, sulfated cartilage PGs are stained in blue, cell nuclei

are stained in red and the foams appear blue. ............................................180

Figure 9. 15 Col II and Col I immunolabeling of MSCs-seeded Alg- and Alg/ChS-foams

after 7 days of culture, observed by fluorescence microscopy. Col is stained

in green and cell nuclei are stained in blue. ................................................182

Figure 9. 16 Col II and Col I immunolabeling of MSCs-seeded Alg- and Alg/ChS-foams

after 7 days of culture, observed by confocal microscopy. Collagen is

stained in green and cell nuclei are stained in blue. ...................................182

Figure 9. 17 Cell viability (FDA/PI live-dead assay) of MSCs-seeded Alg- and Alg/ChS-

foams after culture with and without chondrogenic induction (+TGF-1).

Viable cells are green, dead cells appear red. ............................................183

Page 254: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

X

Figure 9. 18 Col II and Col I immunolabelling of MSCs-seeded Alg- and Alg/ChS-foams

after 14 days of culture with and without chondrogenic induction, observed

by confocal microscopy. Col II is stained in green and cell nuclei are stained

in blue. .........................................................................................................184

Figure 10. 1 Summary of major challenging issues in the area of cartilage tissue

engineering, as investigated in this dissertation, indicating the criteria that

have been fulfilled by the developed scaffolds. ...........................................190

Figure 10. 2 SEM image showing a recommended multilayered scaffold – model for

osteochondral tissue engineering applications, including Alg-foam for

cartilage phase, PLLA fiber mesh for calcified interface phase and PDLLA-c-

BG scaffold for bone phase. ........................................................................192

Page 255: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

LIST OF TABLES

Table 2. 1 Mechanisms of bioactivity and bone bonding of Bioglass®, according to

[29,31,34,39,46]............................................................................................... 9

Table 2. 2 Mechanical properties of natural healthy human osteochondral tissues

[6,12,73,76,141–143]. ...................................................................................17

Table 2. 3 Current 3D scaffold fabrication techniques for polymers and ceramics. .......19

Table 2. 4 Summary of current strategies in osteochondral tissue engineering. ...........28

Table 4. 1 Polymer coating conditions for polymer coated 45S5 Bioglass®-based

scaffolds. An as-sintered rectangular shaped 45S5 Bioglass®-based

scaffold, with the dimensions of 8 mm × 8 mm × 8 mm, was soaked in 5 ml

of each polymer solution. ..............................................................................46

Table 4. 2 Optimized polymer coating conditions for different biodegradable polymer

coated 45S5 Bioglass®-based scaffolds. .......................................................57

Table 6. 1 Electrospinning conditions and primary observations of PLLA fibers (

indicates no fiber formation, indicates beads incorporated into fibers and

indicates uniform fibers). .............................................................................109

Table 6. 2 Electrospinning conditions and primary observations of alginate fibers (

indicates no fiber, indicates beads and indicates uniform fibers). ........112

Page 256: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XII

Page 257: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

ABBREVIATIONS AND SYMBOLS

Abbreviations:

3D Three-dimentional

45S5 BG 45S5 Bioglass®

5S5 BG 5S5 Bioglass®

AB assay Alamar Blue assay

AB Alcian blue

ACI Autologous chondrocytes implantation

ACDC5 Chondrocyte-like cells

ADSCs Adipose-derived stem cells

Alg Alginate

ALP Alkaline phosphatate

ATR-FTIR Attenuated total reflectance-Fourier transform infrared spectroscopy

BCP Biphasic calcium phosphate

BMP-2, -7 Bone morphogenetic protein-2, -7

BMSCs Bone marrow mesenchymal stem cells

Ca-Alg Calcium alginate

CaP Calcium phosphate

Col Collagen

Col I, II, X Type I, II, X collagen

ChS Chondroitin sulfate

CS Chitosan

DAPI 4,6-diamidino-2-phenylindole

DCM Dichloromethane

DMA Dynamic mechanical analysis

DMC Dimethyl carbonate

Page 258: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XIV

DMEM Dulbecco’s modified eagle medium

DSC Differential scanning calorimetry

E´ Storage modulus

E´´ Loss modulus

ECM Extracellular matrix

ESCs Embryonic stem cells

EtOH Ethanol

F127 Pluronic F127 (Blockcopolymer of poly(ethylene glycol))

FDA Fluorescein diacetate

FGF-1 Fibroblast growth factor-1

Fig. Figure

FTIR Fourier transform infrared spectroscopy

GA Glutaraldehyde

GAGs Glycosaminoglycans

Gel Gelatin

HA Hydroxyapatite

HB Hydroxy butyrate

HCA Hydroxycarbonate apatite

HE Hematoxylin eosin

hMSCs Human mesenchymal stem cells

HH Hydroxyl hexanoate

HOBs Human osteoblasts

HOS-TE85 Human osteosarcoma cell line

HyA Hyaluronic acid, Hyaluronan

IGF-1, -2 Insulin-like growth factor-1,-2

LDH Lactate dehydrogenase

MC3T3-E1 Mouse osteoblast cell line

MeOH Methanol

MG-63 Human osteoblast cell line

Page 259: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XV

MSCs Mesenchymal stem cells

Mw Molecular weight

nHA Nanohydroxyapatite

Na-Alg Sodium alginate

OA Osteoarthritis

OD Optical density

P Porosity

P123 Poly(ethylene glycol)-c-poly(propylene glycol)-c-poly(ethylele

glycol)-triblock copolymer

PA6 Polyamide 6

PBS Phosphate buffer saline

PCL Polycaprolactone

PEG Poly(ethylene glycol)

PEO Poly(ethylene oxide)

PGs Proteoglycans

PGA Poly(glycolic acid)

PHA Poly(hydroxyalcanoate)

PHB Poly(3-hydroxybutyrate)

PHBHHx Poly(3-hydroxybutyrate-c-3-hydroxyhexanoate)

PHBV Poly(3-hydroxybutyrate-c-valerate)

PLA Poly(lactic acid)

PLLA Poly(L-lactic acid)

PDLA Poly(D-lactide)

PDLLA Poly(D,L-lactide)

PI Propidium iodine

PLGA Poly(lactic-c-glycolic acid)

p-NP Para-nitrophenol

p-NPP Para-nitrophenylphosphate

PolyHEMA Poly(2-hydroxyethyl methacrylate)

Page 260: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XVI

PVA Poly(vinyl alcohol)

PU Polyurethane

SB Subchondral bone

SEM Scanning electron microscopy

SBF Simulated body fluids

SD Standard deviations

SOX-9 Protein transcription factor, acting during chondrocyte differentiation

TC Cold crystalline temperature

Tg Glass transition temperature

Tm Melting temperature

Tan Loss factor

TBS Tris buffered saline

TCH Tetracycline hydrochloride

TCP Tricalcium phosphate

TGF-β1, 3 Transforming growth factor- beta 1, 3

Xc Degree of crosslinking

XRD X-ray Diffraction

Chemical symbols:

-CH2- Ethyl group

-CH3 Methyl group

-C=O Carbonyl group

-C-O-C- Ether bond

-COO- Carboxylate group

-COOH Carboxyl group

-OH Hydroxyl group

-C=N- Imine bond

-N-H- Amine bond

Page 261: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XVII

Al2O3 Aluminium oxide

Ca2+

Calcium ion

CaCO3 Calcium carbonate

CaCl22H2O Calcium chloride dehydrate

CaO Calcium oxide

Ca10(PO4)6(OH)2 Hydroxyapatite

CO2 Carbon dioxide

CO32-

Carbonate ion

Cu2+

Copper ion

DI H2O Deionized water

Fe2+

Iron ion

H+ Hydrogen ion

HCl Hydrochloric acid

H2SO4 Sulfuric acid

H3O+ Hydronium ion

K+ Potassium ion

KBr Potassium bromide

MgCl2 Magnesium dichloride

Na+ Sodium ion

Na2CO3 Sodium carbonate

NaCl Sodium chloride

Na2O Disodium oxide

NaOH Sodium hydroxide

OH- Hydroxide ion

P2O5 Phosphorous pentoxide

PO43-

Phosphate ion

S Sulfur

Si Silicon

SiO2 Silicon dioxide

Page 262: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XVIII

Si-OH Silanol group

Si-O-Si Silica gel

Si(OH)4 Silicic acid

Sr2+

Strontium ion

Zn2+

Zinc ion

Symbols:

% Percent

wt % Weight in percentage

vol % Volume in percentage

wt/v % Weight per volume in percentage

v/v % Volume per volume in percentage

° Degree

°C Degree Celsius

°C/min Degree Celsius per minute

Density

g Gram

g/mole Grams per mole

g/cm3 Grams per cubic centimeter

ng/ml Nano-grams per milliliter

µg/ml Micro-grams per milliliter

M Molar

mM Millimolar

N Normal

sec Second

min Minute

h Hour

µl Microliter

Page 263: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XIX

ml Milliliter

l Liter

ml/h Milliliters per hour

mm/min Millimeters per minute

cm Centimeter

cm-1

Reciprocal centimeter

µm Micrometer

nm Nanometer

N Newton

kN Kilo Newton

MPa Mega Pascal

kV Kilo Volt

P Porosity

ppi Pores per inch

rpm Rounds per minute

J/g Joules per gram

Hz Hertz

U/mg Units per milligram

Page 264: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XX

Page 265: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

ACKNOWLEDGEMENTS

I would like to thank my funding from the Royal Thai Government Scholarship

granted by the Office of the Civil Service Commission (OCSC), Bangkok, Thailand. Also

thanks to Faculty of Agricultural Product Innovation and Technology, Srinakharinwirot

University, Bangkok, Thailand for providing me a next future journey in the research area.

I also would like to thank all the people who help me along the journey to carry out

this work. My particular gratitude is dedicated to:

First of all, Prof. Dr.-Ing. habil. Aldo R. Boccaccini, who gave me the great

opportunity to work at the Institute of Biomaterials, Department of Materials Science and

Engineering, University of Erlangen-Nuremberg. Thanks for giving me the chance to work

with variety of researches in the field of tissue engineering and in collaboration with many

research groups. I learned a lot of knowledge and achieved plenty of experiences.

Thanks to all collaborations for the kind exchange of knowledge and scientific

discussion, including:

- Prof. Dr. rer. nat. habil. Dirk W. Schubert (Institute of Polymer Materials,

Department of Materials Science and Engineering, University of Erlangen-Nuremberg) for

the close collaboration and for the opportunity to use the facilities during my project.

- PD Dr. med.-vet. Gundula Schulze-Tanzil and her co-workers in Charité-

Universitätsmedizin Berlin, Campus Benjamin Franklin Klinik für Orthopädische for very nice

collaboration and scientific discussion in the field of cell biology for cartilage tissue

engineering. I also would like to thank for her dedication in revising a part of this dissertation.

- Prof. Dr. Ulrich Lohbauer and Dr. Andrea Wagner in the Dental Clinic – Operative

Dentistry and Periodontology, University of Erlangen-Nuremberg, for giving me the

opportunity to use microtensile testing.

- Prof. Dr. Dirk Höfer, Dr. Timo Hammer, and Marina Müller from Hehenstein

Institutes, Institute for Hygiene and Biotechnology, Boennigheim, Germany, who

Page 266: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XXII

collaborated with us on the topic of adipose-derived stem cells culturing on Bioglass-based

scaffolds for vascularization.

- Dr. Subha Narayan Rath (Nikolaus Fiebiger Zentrum for molecular medicine,

University of Erlangen-Nuremberg), and Prof. Dr. med. Ulrich Kneser and his co-workers

(Department of Plastic and Hand Surgery, University Hospital of Erlangen, University of

Erlangen-Nuremberg), who are the close collaboration in the field of osteogenic

differentiation of MSCs on Bioglass-based scaffolds. Special thanks to Dr. Subha Narayan

Rath for kindly discussion about cell biology.

- Both co-authors in my first review paper about osteochondral tissue engineering,

PD Dr. med. Justus Beier (Department of Plastic and Hand Surgery, University Hospital of

Erlangen, University of Erlangen-Nuremberg) and Prof. Dr. Vehid Salih (Eastman Dental

Institute, UCL, London, United Kingdom), who kindly advised and supported during the

process.

- Dr. Beatriz Olalde and co-workers (TECNALIA Health Division and Ciber-BNN,

San Sebastián, Spain) for the fruitful collaboration in the field of drug delivery applications.

Numerous thanks to Dr. Judith A. Roether for her friendly scientific advices and

spiritual support. I am appreciate for her sincerely helps since the first day of my staying in

Erlangen, Germany.

I would like to thank to Dr. Rainer Detsch and Ms. Alina Grünewald for their kindly

help to carry on the bone cell culture on my Bioglass-based scaffolds. I am thankful for their

big attempt. My work would not have been possible without their assistance.

Thanks to Dr. Ing. Joachim Kaschta (Institute of Polymeric Materials, Department of

Materials Science and Engineering, University of Erlangen-Nuremberg) and Dr. Raquel Silva

Lourenço (Institute of Biomaterials) for dynamic mechanical analysis - training. Also thanks

to Dr. Menti Goudouri for the analysis of FTIR spectroscopy and for kindly advising a part of

my thesis.

Special thanks to Dr. Ranjana Rai, who is the first friend and one of my best friends

during my staying in Germany. She has been like my sister, who is always ready to help me.

I would not have passed hard times of my life without her friendship.

Page 267: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XXIII

I also would like to thank to my international friends, Wei Li, Yaping Ding, Qiang

Chen, Qingqing Yao, Bapi Sarker, Preethi Balasubramanian, Fereshteh Zeinab and Kai

Zheng. They are lovely friends who never let me feel lonely during my study abroad. Also

thanks for useful scientific discussion and helping several experimental works.

In addition, I would like to thank to my office mates, Anahi Phillippart, Valentina

Miguez Pacheco, Marwa Tallawi, Dr. Sandra Cabañas Polo, Dr. Judith Bortuzzo and Samira

Tansaz, and also Rama Krishna Chinnam and Luis Cordero for their friendship and their

kindly help in many issues.

Special appreciation and respect are dedicated to Heinz Mahler and Bärbel Wust,

who have helped me a lot concerning my livelihood.

Also thanks to other Biomat-members, Sigrid Seuß, Jasmin Hum and Alexander

Hoppe for experimental assistance. All of my bachelor, master and mini-project students,

Alexander Ritter, Birgitta Carlé, Mani Diba, Anke-Lisa Metze, and Eva Weber, who have

usually brought up fresh ideas and have been willing to learn and to achieve new

experiences together with me.

Especially, I would like to thank to my best friends in Thailand, Wanruedee Temnil,

Patchamon Duangnum, Sudarat Jindabot, Worratai Panichnitinon, Pimpaporn Paebamrung,

Nattaporn Aimampaiwong and Piyachat Chuysrinual, who beside me in all sufferings and

happiness.

Very special thanks are dedicated to Dr. Ing. Mirza Mackovic for his best support.

Thanks for always expecting the best for me, and supporting me in every life situation.

My particular gratitude is dedicated to my parents, Mr. Prasop and Mrs. Theeranuch

Nooeaid. Thanks for their unconditional love, care and support, and thanks for growing me

with gentleness and modesty. They inspire me to keep reaching for my goals and they are

my power behind my every success. Also, I would like to thank to Nooeaid and Suwanarat

families for their care and support.

Page 268: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XXIV

Page 269: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

LIST OF PUBLICATIONS

(1) Nooeaid P., Roether J.A., Weber E., Schubert D.W., Boccaccini A.R. Technologies for

multilayered scaffolds suitable for interface tissue engineering. Adv. Eng. Mater.

2013, DOI: 10.1002/adem.201300072.

(2) Nooeaid P., Salih V., Beier J. P., Boccaccini A. R. Osteochondral tissue engineering:

scaffolds, stem cells and applications. J. Cell. Mol. Med. 2012, 16, 2247-70.

(3) Nooeaid P., Roether J.A., Schubert D.W., Boccaccini A.R. Polymer coated bioactive

glass foams: toughened scaffolds for bone tissue engineering. Biofoams 2011, 21-4.

(4) Nooeaid P., Schulze-Tanzil G., Boccaccini A.R. Stratified scaffolds for osteochondral

tissue engineering. A book chapter in Methods in Molecular Biology. Submitted.

(5) Li W., Nooeaid P., Roether J.A., Schubert D.W., Boccaccini A.R. Preparation and

characterization of vancomycin releasing PHBV coated 45S5 Bioglass-based glass-

ceramic scaffolds for bone tissue engineering. J Eur Cer Society 34, 2014, 505-14.

(6) Subha R., Nooeaid P., Arkudas A., Beier J.P., Strobel L.A., Brandl A., Horch R.E.,

Boccaccini A.R., Kneser U. Adipose-Derived and Bone Marrow-Derived Mesenchymal

Stem Cells Display Different Osteogenic Differentiation Patterns in 3D Bio-active

Bioglass Scaffolds. J Tissue Eng Regen Med 2013, DOI: 10.1002/term.

(7) Yao Q., Nooeaid P., Detsch R., Roether J.A., Dong Y., Goudouri M.O., Schubert

D.,Boccaccini A.R. Bioglass/chitosan-polycaprolactone bilayered composite scaffolds

intended for osteochondral tissue engineering. J. Biomed. Mater. Res. Part A 2014,

In-press.

(8) Yao Q., Nooeaid P., Roether J.A., Dong Y., Zhang Q., Boccaccini A.R. Bioglass®-

based scaffolds incorporating polycaprolactone and chitosan coatings for controlled

vancomycin delivery. Ceramic Inter. 2013, In-press.

(9) Hum J., Luczynski K.W., Nooeaid P., Newby P., Lahayne O., Hellmich C., Boccaccini

A.R. Stiffness improvement of 45S5 Bioglass®-based scaffolds through natural and

synthetic biopolymer coatings: an ultrasonic study. Strain 2013, 49, 431-9.

Page 270: Multilayered Scaffolds for Osteochondral Tissue Engineering … Nooeaid... · Knorpels liegt (E-Modul von 0.24 – 0.85 MPa and Druckfestigkeit von 0.01 – 3 MPa). Wesentlich dabei

XXVI

(10) Liverani L., Roether J.A., Nooeaid P., Trombetta M., Schubert D.W., Boccaccini A.R.

Simple fabrication technique for multilayered stratified composite scaffolds suitable for

interface tissue engineering. Mater. Sci. Eng. A 2012, 557, 54–8.

(11) Metze A., Grimm A., Nooeaid P., Roether J.A., Hum J., Newby P.J., Schubert D.W.,

Boccaccini A.R. Gelatin coated 45S5 Bioglass®-derived scaffolds for bone tissue

engineering. Key Eng. Mater. Vol. 2013, 541, 31-9.

(12) Handel M., Hammer T.R., Nooeaid P., Boccaccini A.R., Hoefer D. 45S5 Bioglass-

based 3D-scaffolds seeded with human adipose tissue-derived stem cells (hASC)

induce in vivo vascularization in the CAM angiogenesis assay. Tissue Eng Part A.

2013, 19, 2703-12.

(13) Olalde B., Garmendia N., Sáez-Martínez V., Argarate N., Nooeaid P., F. Morin,

Boccaccini A.R. Multifunctional bioactive glass scaffolds coated with layers of

poly(D,L-lactide-co-glycolide) and poly(n-isopropylacrylamide-c-acrylic acid) microgels

loaded with vancomycin. Mater. Sci. Eng. C Vol. 2013, 33, 3760-7.