MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

JEREMY DRAKE, PETE RATZLAFF, VINAY KASHYAP AND THE MC CALIBRATION UNCERTAINTIES TEAM

11th IACHEC Meeting, Pune March 2016

Page 3: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

JEREMY DRAKE, PETE RATZLAFF, VINAY KASHYAP AND THE MC CALIBRATION UNCERTAINTIES TEAM

11th IACHEC Meeting, Pune March 2016

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

Page 4: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

OUTLINE

▸ Brief review of our MC uncertainties method

▸Using observations as MC calibration constraints: G21.5-0.6

▸Using observations and MC methods for cross-calibration

IACHEC 11 Pune 2016

Page 5: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

MONTE CARLO APPROACH TO CALIBRATION UNCERTAINTIES

▸ Use brute-force Monte Carlo methods instead:

▸ Simulate 100’s-1000s of response functions that sample nominal response and its uncertainties

▸ Repeat parameter estimation and examine distributions of “best-fit” parameters

▸ Can be used to understand the true accuracy of flux measurements, parameter fits… and refine the calibration itself

Highly correlated - analytical solutions difficult....

IACHEC 11 Pune 2016

Page 6: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CONTEXT WITH PYBLOCXS, STATISTICS APPROACHES

Finesse Brute Force

Sachin TendulkarIan Botham

IACHEC 11 Pune 2016

Page 7: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CONTEXT WITH PYBLOCXS, STATISTICS APPROACHES

Finesse Brute Force

“If you make a team with the No.11’s from all the teams, Hirwani would still bat at No.11” - Harsha Bhogle

Baseball

IACHEC 11 Pune 2016

Page 8: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

TYPICAL UNCERTAINTY CHAIN: CHANDRA ACIS-S

IACHEC 11 Pune 2016

Page 9: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

GENERATING MONTE CARLO EFFECTIVE AREAS

▸ Parameterised instrument models where available; vary parameters, re-compute response, eg:

▸ Mirror trial models

▸ CCD QE, contamination, RMF models

▸ Use a “perturbation function” - a perturbation vs E by which to change subassembly responses between edges

▸ Combine the above into an ARF multiplicative perturbation

IACHEC 11 Pune 2016

[Do MC RMFs too but not discussed today…]

Page 10: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

FOR HRMA WE ALSO USE RAY-TRACE MODEL AREAS

Equal probabilities, except x2 for model f

IACHEC 11 Pune 2016

Page 11: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

PERTURBATION FUNCTION

2014: Added “maxdiff” - the maximum difference allowed between min and max perturbation - controls curvature in function, prevents unrealistic deviations

(Emindev)

(Emaxdev)

(edgeveto)

IACHEC 11 Pune 2016

Page 12: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

PERTURBATION INPUT FILE

▸ Uncertainty data for each instrument subassembly (MM=multi-mirror, OBFM=optical blocking filter medium, etc)

▸ Each line refers to an energy range (in keV) bounded by instrument edges

▸ Format: Emin,Emindev,Emax,Emaxdev,Edgeveto, maxdiff

MM 0.05 0.04 2.291 0.04 0.03 0.04 2.291 0.03 3.425 0.03 0.01 0.03 3.425 0.03 7.000 0.03 0.005 0.03 7.000 0.05 12.0 0.10 0.10 CONTAM 0.05 0.10 0.2838 0.02 0.02 0.10 0.2838 0.02 0.4099 0.02 0.02 0.02 0.4099 0.02 0.532 0.02 0.01 0.02 0.532 0.02 0.6967 0.02 0.02 OBFM 0.05 0.15 0.297 0.07 0.04 0.15 0.297 0.06 0.540 0.03 0.02 0.06 0.540 0.02 1.567 0.02 0.02 0.02 1.567 0.02 12.0 0.02 0.02 EPICPN 0.05 0.20 0.132 0.10 0.11 0.20 0.132 0.15 0.539 0.05 0.03 0.15 0.539 0.04 1.827 0.04 0.03 0.04 1.827 0.04 12.0 0.03 0.04

IACHEC 11 Pune 2016

WE CAN PERFORM THE SAME (DIS)SERVICE FOR YOUR MISSIONS!

Page 13: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

HOW ARE CALIBRATION UNCERTAINTIES DISTRIBUTED?.

▸ Rigorous treatment requires knowledge of how uncertainties are distributed

▸ Unknown!

▸ Assume a truncated normal distribution -1σ to +1σ

▸ Peaked at preferred value

▸ Includes gut feeling!

IACHEC 11 Pune 2016

Page 14: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

RESULTING ACIS-S3 AREAS

Nominal response

IACHEC 11 Pune 2016

These are the basis of the pyBLoCXS calibration uncertainties

Page 15: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

XMM-NEWTON SAMPLE AREAS

10�1 100 101

Energy (keV)

0

200

400

600

800

1000

1200

1400E

ffect

ive

Are

a(c

m2 )

Simulated EPIC-PN ARFS

Nominal area

MC areas

IACHEC 11 Pune 2016

Medium filter

Page 16: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

EXERCISE: LIMITING ACCURACY OF X-RAY TELESCOPES

▸ Method applied to Chandra ACIS-S, XMM EPIC-pn, NuSTAR (see Kristin’s talk):

▸ Simulate spectrum (“fakeit”)

▸ Fit using different effective area realisations a lot of (e.g. 1000) times

▸ XSPEC driven by Perl (Sherpa driven by Python soon…)

▸ Models: blackbody, MEKAL, power-law; all with ISM absorption

▸ Compare with fits to 1000 different “fakeits” using nominal area to probe uncertainties from only counting statistics

IACHEC 11 Pune 2016

Page 17: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

EXAMPLE FITTED PARAMETER DISTRIBUTIONS: EPIC-PN

Different spectrum realisations

Different area realisations

IACHEC 11 Pune 2016

Page 18: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

XMM EPIC-PN LIMITING PRECISION

Different spectrum realisations

Different area realisations

IACHEC 11 Pune 2016

Page 19: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

LIMITING PRECISION SUMMARY

▸ MC analysis using best guess effective area uncertainties finds that the limiting precisions of Chandra and XMM-Newton are reached for about 10,000 counts; ie increasing exposure time to get more counts does not help the accuracy of the fit

▸ BUT:

▸ based only “best guess” uncertainties at subassembly level

▸ how to make sure we do not end up with areas too deviant and to improve uncertainty estimates?

IACHEC 11 Pune 2016

Page 20: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

HOW DO WE IMPROVE UNDERSTANDING OF THE TRUE UNCERTAINTIES?

IACHEC 11 Pune 2016

Page 21: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

G21.5 -0.6 IACHEC 11 Pune 2016

Page 22: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

G21.5 -0.6

▸ Plerionic SNR

▸ Appears to have power-law spectrum

▸ Used as an IACHEC cross-calibration source (Tsujimoto et al 2011)

▸ High NH - relatively insensitive to ACIS contamination model

Tsujimoto et al (2011)

IACHEC 11 Pune 2016

Page 23: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CHANDRA ACIS-S: SIMULTANEOUS FIT TO 8 OBSERVATIONS

IACHEC 11 Pune 2016

Page 24: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

IACHEC 11 Pune 2016MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

RESULTS

Page 25: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CONSTRAINTS ON “GOOD” AND “BAD” AREAS

IACHEC 11 Pune 2016

Page 26: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CONSTRAINTS ON “GOOD” AND “BAD” AREAS

IACHEC 11 Pune 2016

Page 27: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CONSTRAINTS ON “GOOD” AND “BAD” AREAS

IACHEC 11 Pune 2016

Page 28: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CONSTRAINTS ON “GOOD” AND “BAD” AREAS

IACHEC 11 Pune 2016

Page 29: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH
Page 30: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

Calibration

Page 31: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

REFINE TELESCOPE PRECISION ESTIMATES

IACHEC 11 Pune 2016

PoissonCalibrationCalibration best 100

Page 32: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH
Page 33: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

WHY STOP AT JUST CHANDRA?

Page 34: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CHANDRA + NUSTAR: SIMULTANEOUS FIT

IACHEC 11 Pune 2016

Chandra ACIS

NuSTAR FPMs

Page 35: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

“BAD” AREA RATIOS: CHANDRA

IACHEC 11 Pune 2016

Page 36: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

“GOOD” AREA RATIOS: CHANDRA

IACHEC 11 Pune 2016

Page 37: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

SEE KRISTIN’S TALK NEXT FOR DISCUSSION OF NUSTAR CALIBRATION UNCERTAINTIES…

Page 38: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

“BAD” AREA RATIOS: NUSTAR

IACHEC 11 Pune 2016

Page 39: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

“GOOD” AREA RATIOS: NUSTAR

IACHEC 11 Pune 2016

Page 40: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

CHANDRA + NUSTAR + XMM: SIMULTANEOUS FIT

IACHEC 11 Pune 2016

NuSTAR FPMs

Chandra ACIS

XMM pn

Page 41: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH
Page 42: MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION · 2019-10-08 · Sachin Tendulkar Ian Botham IACHEC 11 Pune 2016. MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION CONTEXT WITH

MONTE CARLO CONSTRAINTS ON INSTRUMENT CALIBRATION

SUMMARY

▸ Application of MC effective areas to fitting of fiducial sources with assumptions about the spectral model provides a calibration discriminant

▸ Technique can be applied to multiple missions

▸ Technique can be applied to multiple and diverse sources (perturbation set is common to all)

▸ Needs refinements, e.g. balance between input spectra - “most counts wins”; improved input uncertainties…

IACHEC 11 Pune 2016