Upload
others
View
9
Download
1
Embed Size (px)
Citation preview
Microbiology Methods for
Drinking Water Laboratories
Erica Fox
Origins of Drinking Water
Bacteriological Testing1854 – Cholera epidemic in London
– Dr. John Snow determined source of Cholera to be public water supply
– Traced nearly all deaths to water from the Broad Street pump, which supplied water from a public well dug 3 feet from an old cesspit
– Cholera epidemic stopped after removal of pump handle
– Showed connection between fecal contamination of drinking water and disease
3
Origins of Drinking Water
Bacteriological Testing
1885 – “Bacterium coli
commune” described by
Theodor Escherich
– Found in higher numbers than
enteric pathogens in fecal
material
– Suggested that detection of B.
coli would indicate presence of
fecal matter
– Later renamed Escherichia coli,
this organism remains the main
indicator of fecal pollution
Regulations
• Safe Drinking Water Act (SDWA), 1974 –
passed by Congress to protect public health by
regulating public drinking water supplies
• National Primary Drinking Water Regulations –
standards set by Environmental Protection
Agency (EPA), as required by SDWA– Found in Code of Federal Regulations (CFR) Part 141
– Establish regulated contaminants and Maximum Contaminant Levels
(MCLs) if applicable, as well as approved methods for analysis
– Microbiological contaminants include Total coliforms (includes Fecal
coliforms and E. coli)
5
Regulations
• National Primary Drinking Water Regulations
– Revised Total Coliform Rule (RTCR) – Establishes MCL for E. coli and uses
E.coli and total coliforms to initiate “find and fix” approach to address fecal
contamination that could enter distribution system.
– Requires public water systems to perform assessments to identify sanitary
defects and take action to correct them
– Surface Water Treatment Rule (SWTR) – HPC are regulated under the SWTR
with a limit of 500 cfu/mL
• Virginia a “Primacy” state – enforces national regulations
– Virginia Waterworks Regulations: 12 VAC 5-590
• Virginia certifies drinking water laboratories through Division of Consolidated
Laboratory Services (DCLS)
– Labs are required to comply with EPA Manual for the Certification of Laboratories
Analyzing Drinking Water
Regulations
• To comply with regulations, certified laboratories must use an approved method. Each of the methods presented today is listed below with the applicable regulation and method reference:
• Surface Water Treatment Rule, CFR 141.74– SimPlate
®
- IDEXX SimPlate®
HPC Test Method for Heterotrophs in Water
• Revised Total Coliform Rule, CFR 141.21– ONPG-MUG - SM 9223 B
• Ground Water Rule, CFR 141.402– ONPG-MUG - SM 9223 B
• Long Term 2 Enhanced Surface Water Treatment Rule, CFR 141.704– Quanti-Tray
®
- SM 9223 B
7
Critical Elements for Microbiology(from EPA Lab Cert Manual, Chapter V)
• Personnel
• Laboratory Equipment and Supplies
• General Laboratory Practices
• Analytical Methodology
• Sample Collection, Handling and Preservation
• Quality Assurance
• Records and Data Reporting
• Action Response to Laboratory Results
Personnel
• Supervisor
– Bachelor’s degree in microbiology, biology or equivalent
– Minimum 2 weeks training at federal/state agency or academic
institution in drinking water microbiological analysis or 80 hours
on-the-job training in water microbiology at certified laboratory
• Analyst
– At least 3 months bench experience in water, milk, or food
microbiology
– Training acceptable to the state in drinking water microbiology
analysis and at least 30 days of on-the-job training in drinking
water microbiology under an experienced analyst
– Demonstrate acceptable results on unknown samples before
analyzing compliance samples
Laboratory Equipment and
Supplies• pH meter
– Accuracy with ± 0.1 units
– Use buffers once
– Standardize before each use with pH 4.0 and 7.0 or 7.0 and 10.0 buffers
– Record slope monthly
– Maintained according to manufacturer’s instructions
• Balance
– Readability of 0.1 g
– Sensitivity of 0.1 g for a load of 150 g and 1 mg for load of 10 g or less
– Readability and Sensitivity checks performed monthly
– Maintenance and calibration performed annually
• Thermometers
– Graduated in 0.5°C increments
• Exception: 1°C for Refrigerator
– Calibration checked annually, discard thermometer if differs more than 1 degree
10
Laboratory Equipment and
Supplies• Incubator
– Must maintain a constant temperature of 35 ± 0.5 °C and provide a source of humidity
– Thermometers on top and bottom shelves of the use area
– Record temperature twice each day in use, separated by at least 4 hours
11
Laboratory Equipment and
Supplies• Autoclave
– Maintain sterilization temperature during cycle
– Complete cycle within 45 minutes when 12-15 minute sterilization used
– Record date, contents, initials,sterilization time & temperature, total time in autoclave each use
– Maintenance performed annually
– Maximum registering thermometer used during each cycle
• Check automatic timer each quarter
with stop watch
• Spore strips or spore ampules should
be used monthly as bioindicators to
confirm sterilization.
12
Laboratory Equipment and
Supplies• Conductivity Meter
– Suitable for checking reagent-grade water
– Readable in micromhos/cm or microsiemens/cm
– Calibrate monthly
• Refrigerator
– Maintain temperature of 1 – 5°C
– Record temperature once each day in use
• Inoculating equipment
– Sterile or presterilized disposable loops and sticks should be used
13
Laboratory Equipment and
Supplies• Pipets
– Presterilized plastic or glass
– 10 mL or less must be accurate to within 2.5%
– Reseal packages between use periods
• Glassware and Plasticware– Borosilicate glass or clear, non-toxic plastic
– Tube closures should be screw caps with non-toxic liners, stainless steel, plastic or aluminum
• Sample Containers– Wide mouth plastic or glass bottles or sterile plastic bags with sodium
thiosulfate
14
General Laboratory Practices
• Sterilization procedures
– Autoclave at 121°C for recommended time
– Media should be removed immediately after end of cycle
• Sample Containers
– Randomly test 1 container from each batch or lot for sterility using 25 mL non-selective broth (such as Tryptic Soy Broth), incubate for 48 hours & check for growth
• Reagent grade water
– Quality should meet criteria for Conductivity (<2.0 micromhos/cm), Metals (Pb, Cd, Cr, Cu, Ni, Zn) less than 0.05 mg/L per contaminant and no greater than 0.1 mg/L collectively, Total Chlorine Residual (<0.1 mg/L), and HPC (<500 CFU/ml)
• Glassware washing
– Use distilled or deionized water in final rinse
– Inhibitory residue test should be performed prior to initial use of detergent or whenever different washing procedure or detergent is used; at the very least, every five years
– pH residue check
15
Analytical Methodology
• Media– Discard media by manufacturer’s expiration date
– Store prepared medium in dark at recommended temperature
– For lab-prepared media, record date prepared, type, lot #, volume, sterilization time & temperature, final pH, expiration date, and initials
– For commercially-prepared media, record date received, type, lot # and pH verification if specified by method, and expiration date
– Each batch of lab-prepared and each lot of commercially-prepared media should be checked before use for sterility and with positive & negative control cultures
– Run positive & negative controls each quarter for commercially-prepared media with shelf-lives longer than 90 days
• Examples of control cultures:– Total coliforms
• Positive: Escherichia coli Negative: Pseudomonas aeruginosa
– Fecal coliforms• Positive: Escherichia coli Negative: Enterobacter aerogenes
– E. coli• Positive: Escherichia coli Negative: Enterobacter aerogenes
16
Analytical Methodology
• Analysis– For compliance samples, use only methods specified in Revised
Total Coliform Rule, Surface Water Treatment Rule and Groundwater Rule
– Laboratory must be certified for all methods used on compliance samples
• Minimum of 1 total coliform method and 1 E. coli method
• Recommend certification in second total coliform method if one method cannot be used due to interference in some drinking water samples
– Shake samples at least 25 times before analyzing
– Tolerance of dilution buffer should be ± 2 mL for volumes of 90 or 99 mL
– RTCR sample volume analyzed must be 100 mL
17
Sample Collection, Handling and
Preservation• Sample collectors should be trained in aseptic sampling
procedures
• Sampling– Locations must be representative of distribution system
– Sample taps must be free of attachments (aerator, strainer, hose)
– Use only cold water taps
– Flush tap until steady temperature
– Collect at least 100 mL, but allow 1 inch air space
• Sample Icing– RTCR: recommended to hold drinking water samples at < 10 °C
during transport
– SWTR: must hold at < 10 °C during transport, sample temperature should be recorded upon receipt at lab
18
Sample Collection, Handling and
Preservation• Sample Holding/Travel Time (all times are from
collection to placing sample in incubator)– Total coliforms in drinking water: 30 hours
– Total coliforms & E.coli in source water: 8 hours
– E. coli under Ground Water Rule: 30 hours
• Sample Information Form– Record information after collection: system name;
sample identification and site location; sample type; date & time of collection; analysis requested; disinfectant residual; sampler name
19
Quality Assurance
• Laboratory should prepare & follow a
Quality Assurance (QA) plan
• Analyze a set of Proficiency Testing (PT)
samples once every 12 months for each
method for which laboratory is certified
• Participate in laboratory audit every three
years
20
Records and Data Reporting
• Laboratory should keep thorough and accurate
records
• QA Plan should describe procedures used for
record retention
• Laboratory should maintain easily accessible
records for 5 years (includes raw data,
calculations and QC data)
• Electronic data should be backed up and should
state how long you will retain records
21
Records and Data Reporting
• Data should be recorded in ink; changes should be lined through and dated & initialed
• Record date & time of sample receipt and name of person receiving sample, along with any comments on sample condition
• Record laboratory sample identification, date & time analysis begins, initials of analyst, method used, media lot number, items noted as QC and results
• Maintain preventive maintenance and repair records for all instruments and equipment for five years
22
Action Response to Laboratory
Results
• RTCR samples: laboratory must promptly notify appropriate authority of positive Total coliform (in excess of 5%) or an E. coli result
• E. coli positive results: water system must notify state by end of day when notified of result
Basic Aseptic Technique
• Disinfect bench tops prior to analysis
• Long hair should be pulled back
• Wash hands prior to analysis
• Take every precaution not to contaminate sterile bottles or media during analysis by talking, coughing, sneezing, etc
• Do not touch any surfaces that are sterile (interiors of bottles or bottle tops, tips of pipets, etc.)
• Keep sterile pipets, bottles, etc. closed until ready to use
• Use care when removing pre-sterilized pipets from bulk bags, to avoid contaminating inside of bag
24
Break
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
26
Coliforms
The term total coliforms refers to a broad group of bacteria that belong to the taxonomic family Enterobacteriaceae. This group includes the genera Enterobacter, Klebsiella, Citrobacter, and Escherichia. Total coliforms are present in the environment and are used as an indicator to determine the efficacy of treatment plant operation and the integrity of the distribution system. In the ONPG-MUG method, total coliforms are defined as any bacteria possessing the enzyme β-D-galactosidase.
27
Coliform GroupTotal Coliforms
Escherichia coli
Klebsiella
Citrobacter
Escherichia
Enterobacter
Klebsiella
Citrobacter
Escherichia
28
Escherichia coli
Escherichia coli (commonly abbreviated as
E. coli) is a member of the fecal coliform
group. E. coli is found in the digestive tract
of warm blooded animals and is considered
an indicator of fecal contamination. In the
ONPG-MUG method, E. coli is defined as
bacteria giving a positive total coliform
response and possessing the enzyme β-
glucuronidase.
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
These methods allow for the simultaneous
detection of total coliform bacteria and E.
coli through the use of nutrient indicators
ortho-nitophenyl-β-D-galactopyranoside
(ONPG) and 4-methyl-umbelliferyl-β-D-
glucuronide (MUG). Coliform bacteria and
E. coli produce enzymes that can
metabolize ONPG and MUG respectively.
Method Summary
30
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
• When ONPG is metabolized, a yellow color change
occurs, indicating the presence of total coliforms.
Method Summary
31
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
• When MUG is metabolized, a fluorescent product is
produced, indicating the presence of E. coli.
Method Summary
32
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
• Colilert®
is a Presence/Absence test.
• Quanti-Tray®
is an enumeration test.
Method Summary
33
Helpful Hints
• Make sure media is not discolored or caking
before opening snap pack.
• If a sample is yellow or fluorescing, but is less
intense than the comparator, it can be
reincubated, but do not exceed a total of 28
hours incubation.
Total Coliforms/E. coli by Colilert® and
Quanti-Tray®
34
Total Coliforms/E. coli by Colilert® and
Quanti-Tray®
Interferences
• Some non-coliform bacteria produce small amounts of
the enzyme β-D-galactosidase. These bacteria are
suppressed, and will not produce a positive response
unless 104
CFU/mL are present.
• A combination of hydrogen sulfide and high levels of
either manganese or iron may cause a greenish-black to
black color change to occur after incubation.
• Some water samples may have background color. If
there is background color, compare the inoculated
sample to a control containing only sample.
35
• Some samples may fluoresce without having the yellow
color. According to the manufacturer, an ONPG(-)/ MUG(+)
test (colorless with fluorescence) is considered a
nonspecific reaction as it does not meet the test definition
of a total coliform or E. coli. Therefore this test would be
interpreted as negative for both total coliforms and E. coli.
Per the method, only samples that test positive for total
coliforms should be checked under UV light for E. coli.
Total Coliforms/E. coli by Colilert® and
Quanti-Tray®
Interferences
36
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
• Colilert®
media QC requirements– Media Fluorescence Check
• Each lot of media must be checked for fluorescence prior to use.
• Media dissolved in sterile reagent grade water should not fluoresce.
– Positive/Negative Control Check
• Each lot of media must be checked for selectivity and proper performance prior to use, and every 3 months while in use. Media dissolved in sterile reagent grade water and inoculated with the following cultures should exhibit the appropriate reactions after a 24 hour incubation at 35 ± 0.5 °C.
Culture Reaction
E. coli Positive yellow color, fluorescence
K. pneumoniae or
E. aerogenesPositive yellow color, no fluorescence
P. aeruginosa No color, no fluorescence
37
Total Coliforms/E. coli by Colilert®
and Quanti-Tray®
• Vessel QC requirements– Vessel sterility check
• Each lot of vessels must be checked for sterility prior to use.
• Add 25mL of single strength Tryptic Soy Broth to a randomly selected vessel and incubate for 48 hours at 35 ± 0.5 °C.
• If no growth is present, the vessel lot may be put into service.
– Vessel volume check• Each lot of vessels must be checked for volume accuracy prior to
use.
• Fill a Class A 100mL graduated cylinder with reagent grade water and transfer to a randomly selected vessel.
• The level of water should be at the 100mL mark on the vessel for the lot to be placed in service.
– Vessel fluorescence check• A randomly selected vessel should be checked for fluorescence
using the UV light box prior to use.
• Vessel should not exhibit any fluorescence.
38
Total Coliforms/E. coli by
Colilert®
and Quanti-Tray®
• Quanti-Tray®
QC requirements
– Quanti-Tray®
sterility check
• Each lot must be checked for sterility prior to use.
• Add 25mL of single strength tryptic soy broth to a randomly
selected Quanti-Tray®
, seal, and incubate for 48 hours at 35 ±
0.5 °C.
• If no growth is present, the Quanti-Tray®
lot may be put into
service.
– Quanti-Tray®
Sealer QC check
• The sealer is checked each month for proper sealing.
• Add 100mL of a solution of water and a dark dye to a Quanti-
Tray®
, and run through the sealer. Check for leaks between
the wells and on edges of the tray.
Total Coliforms/E. coli by
Colilert®
40
Total Coliforms/E. coli by Colilert®
• Materials:
– Sterile sample vessels
– Colilert®
media
– Sterile Reagent Grade
Water
– Sterile pipettes
– Vessel rack
– Incubator, 35 ± 0.5 ºC
– UV light
41
Total Coliforms/E. coli by Colilert®
• Shake sample thoroughly
(at least 25 times)
• Use sterile pipet to draw
volume of sample down to
100 mL.
• Tap Colilert®
media snap
pack on counter to settle
contents. Snap pack
open, and empty media
into sample vessel.
42
Total Coliforms/E. coli by Colilert®
• Replace cap and shake sample vigorously. Media does not have to be completely dissolved in the sample prior to incubation.
• Place samples in a rack, and incubate at 35 ± 0.5 °C for 24-28 hours.
43
Total Coliforms/E. coli by Colilert®
• After incubating for 24
– 28 hours, the
bottles are read using
a Colilert®
Presence/Absence
comparator for
reference.
44
Total Coliforms/E. coli by Colilert®
• Samples that show a
yellow color equal to
or greater than that of
the Comparator are
total coliform positive.
• Samples with no color
are total coliform
negative.
45
Total Coliforms/E. coli by Colilert®
• Total coliform positive (yellow) samples and the Comparator are checked with UV light.
• Samples that show fluorescence equal to or greater than that of the Comparator are E. colipositive.
• Samples with no fluorescence are E. colinegative.
Total Coliforms/E. coli by Colilert®
Questions?
Total Coliform/E. coli by
Quanti-Tray®
48
Total Coliform/E. coli by Quanti-
Tray®
This method is used for the detection of total
coliform bacteria and E. coli in source water and
ground water. This method is approved for the
analysis of samples for Long Term 2 Enhanced
Surface Water Treatment Rule (LT2) and
Ground Water Rule compliance.
49
Total Coliform/E. coli by Quanti-Tray®
• Quanti-Tray®
trays are available in two different
sizes
• The 51-well Quanti-Tray®
contains only large wells,
and is appropriate for samples containing less than
200.5 CFU (colony forming units) per mL.
• The 97-well Quanti-Tray®
/2000 contains 49 large and
48 small wells. This tray is appropriate for samples
containing less than 2419.6 CFU per mL.
• Samples may be diluted as needed
50
Total Coliform/E. coli by Quanti-Tray®
• Materials:– Package of sterile Quanti-
Tray®
trays
– Sterile sample bottles
– Colilert®
media
– Sterile dilution water (90 and 99 mL)
– Sterile pipettes
– Quanti-Tray®
sealer and inserts
– Incubator, 35 ± 0.5 ºC
– UV light
– MPN table
51
Total Coliform/E. coli by Quanti-Tray®
• Preheat Quanti-Tray®
sealer.
• Shake sample thoroughly (at least 25 times)
• Use sterile pipet to draw volume of sample down to 100 mL. Sample may be diluted using sterile deionized water if necessary.
• Tap Colilert®
media snap pack on counter to settle contents. Snap pack open, and empty media into sample bottle.
Procedure
52
Total Coliform/E. coli by Quanti-Tray®
• Replace cap and mix sample until media dissolves. Media needs to be completely dissolved in the sample prior to pouring into the Quanti-Tray
®
.
• Open Quanti-Tray®
by squeezing in at the sides while pulling the foil tab away. Pour sample into tray.
Procedure
53
Total Coliform/E. coli by Quanti-Tray®
• Fit Quanti-Tray®
in
rubber insert and
slide into the
preheated sealer.
Procedure
54
Total Coliform/E. coli by Quanti-Tray®
• The Quanti-Tray®
will
go through the sealer
and the sealed
Quanti-Tray®
will
come out the back.
• Incubate sealed trays
at 35 ± 0.5 °C for 24-
28 hours.
Procedure
55
Total Coliform/E. coli by Quanti-Tray®
• Wells with a yellow color
equal to or greater than
that of the Quanti-Tray®
Comparator are total
coliform positive.
• Count the number of
positive wells, and refer
to the provided Quanti-
Tray®
MPN table to find
the total coliform MPN.
Reading Samples
56
Total Coliform/E. coli by Quanti-Tray®
• Trays with total coliform positive (yellow) wells and the Comparator are checked for fluorescence with a UV light.
• Wells that show fluorescence equal to or greater than that of the Comparator are E. coli positive.
• Count the number of positive wells, and refer to the provided MPN table to find the E. coli MPN.
• The number of positive wells is converted to a Most Probable Number result in the Quanti-Tray
®
or Quanti-Tray®
/2000 MPN Table. The MPN is multiplied by the appropriate dilution factor, if needed.
Reading Samples
57
Total Coliform/E. coli by Quanti-Tray®
MPN table for Quanti-Tray®
/2000
58
Total Coliform/E. coli by Quanti-Tray®
MPN table for Quanti-Tray®
Total Coliform/E. coli by Quanti-
Tray®
Questions?
HPC by SimPlate®
This method is used for the quantification of
heterotrophic bacteria in drinking and raw
water and can be used for raw, in process and
finished water samples. This method is
approved for the analysis of samples for
Surface Water Treatment Rule compliance.
Summary
HPC by SimPlate®
62
Heterotrophs
• Population of microorganisms that
commonly occur in drinking water.
• Include bacteria, yeasts and molds.
• Dependent upon organic carbon source
for growth.
• Growth in distribution system often found
near dead ends or where loss of
disinfectant residual occurs.
63
Why HPC?
• Used to verify adequate disinfection in distribution system in lieu of monitoring for disinfectant residual, or when residual is low.
• Monitoring of distribution system for biofilm growth.
• Monitoring for nitrification in distribution system when chloramines are used.
• Evaluating effectiveness of treatment process.
64
Consideration of SimPlate®
for HPC as
replacement for HPC Pour Plate
method
• Approved by EPA in late 2002 for compliance
monitoring of drinking water samples under
the Surface Water Treatment Rule, as an
alternative to the HPC Pour Plate Method.
• SimPlate® for HPC has the potential to
reduce staff time, while getting similar results.
65
How does it work?
• SimPlate® media contains multiple
enzyme-substrates, each targeted to a
different bacterial enzyme.
• When bacteria metabolize the enzyme
substrates provided by the media, they
produce a fluorescence that is visible
under UV light.
66
SimPlate®
for HPC Analysis
• Materials:– Sleeve of sterile plates
– Bottle of dehydrated SimPlate
®
media
– Sterile reagent grade water
– Sterile pipettes
– Sterile phosphate buffered dilution water
– Incubator, 35 ± 0.5 ºC
– UV light
67
SimPlate®
for HPC Analysis
• Shake sample thoroughly (at
least 25 times)
• Use sterile pipet to transfer 1 mL
total volume of sample into
middle of SimPlate®
, then touch
tip of pipet off on dry portion of
plate. Sample can be diluted
using sterile phosphate buffered
dilution water as needed.
68
SimPlate®
for HPC Analysis
• 9 mL of SimPlate®
media
is aseptically transferred
from the media bottle to
the center of the base
plate.
• Sample and media mix
and begin to spread out
over the surface of the
plate.
69
SimPlate®
for HPC Analysis
• The plate is gently
swirled, dispersing the
sample and media to
the wells.
• Gently tapping the
plate on the counter
helps bring any
remaining air bubbles
to the surface.
70
SimPlate®
for HPC Analysis
• Air bubbles will
frequently remain on
the surface of the
wells after tapping the
plate.
• After waiting a minute
or two for the bubbles
to pop, the excess
media is poured off
the base plate.
71
SimPlate®
for HPC Analysis
• After incubating for
45 – 72 hours, the
plates are read under
a UV light.
• A blank plate is
shown next to a raw
plate, and both are
upside down to
facilitate viewing of
fluorescing wells.
72
SimPlate®
for HPC Analysis
• The number of
positive wells is
converted to a Most
Probable Number
result in the SimPlate®
for HPC MPN Table.
The MPN is multiplied
by the appropriate
dilution factor, if
needed.
73
SimPlate®
for HPC AnalysisLessons learned
• Up to 10 samples and one blank can be run out of one bottle of prepared SimPlate
®media.
• Prepared media that has been refrigerated must be pre-warmed before next use.
• Sterile phosphate buffered dilution water may be used to dilute raw water samples prior to analysis.
• Although air bubbles in the wells do not interfere with the test, we have found that it can make interpretation of fluorescence more difficult. To help prevent this interference, we have found that the majority of bubbles popped if the plates are allowed to sit a minute or two after tapping, prior to pouring off the sample.
74
SimPlate®
for HPC AnalysisLessons learned
• Swirling the plates should be done GENTLY to avoid spilling excess media.
• Gently tapping the plate on the counter can help pop bubbles up out of the wells.
• Use care with the plates while they are inverted. Bumping, dropping, or setting the plates down roughly can result in the sample being knocked out of the wells.
• Any amount of fluorescence greater than the blank is considered positive when counting wells.
• Keeping the lid on or reading plates upside down can help prevent background fluorescence from interfering with reading.
75
SimPlate® for HPC Analysis
• Analysis QC:
– A blank is analyzed with each set of samples. If more than one bottle of media is needed for a set, a blank should be run for each bottle, and the samples that correspond to that bottle/blank should be noted.
• Media QC:
– All lots of media must be checked for proper performance prior to being placed in service. If the lot is in service for more than a quarter, the media check must be repeated.
• Analyze a plate using raw water or inoculated sterile water as the sample. Incubate for 48 ± 3 hours at 35 ± 0.5 °C. The sample must produce growth for the media to be put in service.
• Check for sterility by incubating an uninoculated plate (blank) for 48 ± 3 hours at 35 ± 0.5 °C. No growth should be observed.
• Sterile Reagent Grade Water QC:
– All batches of sterile water must be confirmed sterile prior to being placed in service.
• Add 50mL of sterile water to 50mL of Double Strength Tryptic Soy Broth and incubate for 48 hours. Check for turbidity indicating bacterial growth.
Quality Control
HPC by SimPlate®
Questions?
Thank you for your attention!
Questions?
For more information, please contact:
Erica Fox,