Upload
melvin-garey-franklin
View
219
Download
0
Tags:
Embed Size (px)
Citation preview
Methods of Photosynthesis SpectrometryFor Phytoplankton
Christophe Six
Spectrometry = spectroscopy :
Methods of spectral analysis allowing to understand the composition, the structure of matter and/or the study
of systems transferring energy
Qualitative and quantitative studies of spectra derived from the interaction between the matter and the wavy radiations of different frequences .
Definitions
Spectrophotometry is an analytic, quantitative method that consists in measuring the absorbance (= absorption ~ optical density) of a given chemical substance (or of a whole unicell organism) in solution, function of the light wavelength.
Spectrofluorimetry is an analytic, quantitative method that consists in measuring the emission and excitation levels of fluorescence of a given chemical substance (or of a whole unicell organism) in solution, function of the light wavelength.
Energy and wavelength
E = (h . c) /
E : Photon energyh : Plank constant factorc : Light celerityl : Photon wavelength
X-rays U.V. Visible Infrared Radio wavelengths
400 nm
800 nm
Wavelength in nanometers
Absorbance of molecules and molecular complexes
.Understanding photophysics and photobiology
.Very useful for assays
Using colorimetric assays
Concept of the absorbance measurement
A = log (I0 / I)
I0 I>
Sample PhotomultiplicatorLight Source
Absorbance
T = I / I0
Transmittance
A = -log T
Spectrophotometers
.One or several light source(s)
Extended Visible (350-900 nm) : Tungsten, HalogenUV (<400 nm) : Deuterium
. One monochromator : Selection of wavelengths
. One sample compartment
. One detector : photomultiplicator or photodiode detector
. A result display system
Components :
Single beam spectrophotometers
or monochromator
(nm)
D.O.
400 500
. A simple compartment for a single sample cuvette
. The simplest system
. The reference = blank is measured before the samples for zeroing the device
Blank : all chemical components (buffer, solvent, etc) except the absorbing substance that you want to measure.It is actually rare to be able to use a perfectly true blank but one should approach it as much as possible.
. Instrument useful for simple routine applications (single or few wavelengths)
Various colorimetric assays (proteins, nucleic acids, pigments, etc.)
. Main problems
The decrease of lamp intensity is not compensed
I0 I
In single wavelength mode, one cannot check for artefacts
(fixed) (measured)
Single beam spectrophotometers
The making of these instruments is usually less careful
Double beam spectrophotometers
ReferenceCuvette
SampleCuvette
I0
I
Chopper
Chopper
Monochromator
. Correction of the variations of the light sources
. For each wavelength, one mesures the absorbance of the sample AND the absorbance of the reference (blank)
. Good reliability of the measurements, ideal for absorption spectra(Elimination of solvent absorption)
Double beam spectrophotometers
. Devices generally better than single beam ones
Artefacts
. Other optical phenomenons linked to diffusion, reflexion and diffraction of light may also distort the measurement.
Refraction : deviation of a wave when its speed changes (interface between 2 media)
=> A
Diopter (surface of the cuvette and surface of the sample)
. Other optical phenomenons linked to diffusion, reflexion and diffraction of light may also distort the measurement.
=> A
Artefacts : Light diffusion
. Turbid solutions, cell suspensions
Diffusion occurs when some light is deflected by particules and therefore does not reach the detector
S = F( d, n)
4=Diffusion of Rayleigh
d : Diameter of particulesn : Refraction index : Wavelength
Diffusion also depends on :
- Particule concentration - Particule shape
Impact of diffusion on absorption spectra
0,0E+00
1,0E-11
2,0E-11
3,0E-11
4,0E-11
5,0E-11
350 450 550 650 750 850
Longueur d'onde (nm)
y = 1x4
=> A ok
=> A
Diffusion is -dependent
Ab
sorb
ance
Wavelength (nm)
Spectrum with diffusion Fitting a correction curve Final spectrum
Impact of diffusion on absorption spectra
Example : absorption spectrum of a phycoerythrin I
Measuring absorbance in a diffusing sample
=> A
Bringing the detector nearer to the cuvette
Increasing the surface of the detector
Measuring absorbance in a diffusing sample
Source lumineuse et
monochromateur
Rayon lumineux
Suspension de cellules
Sphèred’intégration
Echantillonhomogène
Détecteur du photomultiplicateur
DO
(nm)
DO
(nm)
DO
(nm)
DO
(nm)
DO
(nm)
A
B
C
Homogeneous sample
Light detector
Cell suspension
Integration sphere
Light beamLight source and
monochromator
If the absorbance of a sample is not stable…
. Sample much colder than the atmosphere of the compartment
Condensation on the cuvette
Gaz formation (diffusion)
. Sample drops on the outside of the cuvette
. There’s not enough sample in the cuvette and the beam passes through the meniscus
. Cuvettes not adapted (micro-cuvettes)
. The sample contains absorbing particules that sink in the cuvette
The Beer-Lambert law
At a given wavelength, the absorbance of a solution is proportional to the concentration of the absorbing chemical species that are present in this solution, and to the optical path
A = . l . C
A : Absorbance (no unit)l : Wavelength (nm)l : Optical path (cm)C : Concentration (mol L-1)
: Extinction coefficient (L mol-1 cm-1)
. The Beer-Lambert law is additive. Pour n chemical species :
A = ,1 . l . C1 + ,2 . l . C2 + 3 . l . C3 + … + ,n . l . Cn
. For l = 1 cm : A = . C => C = A /
A = ,1 . C1 + ,2 . C2 + 3 . C3 + … + ,n . Cn
Fluorescence: what is it ?
Stokes shift
. With fluorescence, there’s no such general relation as the absorbance Beer-lambert law
The measurement depends strongly on : - The nature of the fluorescent system that is studied - The device used to quantify fluorescence (light source intensity, optics configuration, etc.)
Intensity of fluorescence emission
. It is possible to quantify the fluorescence energy when a fluorescence quantum yield Qf :
Energy of fluorescence emitted (If) = Absorbed energy (Ia) x Qf
Qf = f (, T°C, pH, ions, etc.)
Need to use standard curves to quantify molecules by fluorescence (in absolute units)
Spectrofluorimeters
. None photon from the excitation light must be detected by the detector excitation at 90°
On average, there is 106 times less photons that hit the detector of a spectrofluorimeter than in a spectrophotometer
- A light source : Mercury or xenon lamp
- Two monochromators selecting either the emission or excitation precise wavelengths
- A dark compartment with the cuvette in a 90° excitation/emission cuvette holder
Main components :
- A photomultiplicator
Diagrammic representation of a spectrofluorimeter
Xenon lamp
Lens
LensLens
Slit
Entrance Slit Exit slit
Photomultiplicator
Sample
Mirror
shutter Monochromator
Monochromator
Emission and Excitation spectra of fluorescence
Fix monochromator : One given
Excitation
Emission
Sample
Monochromator scanning all wavelengths
Emission spectrum
Quantification of the fluorescence emitted by the excitation of
a given
At which is the maximum of fluorescence emission of the compound ?
600500 700400
15 nm
Fix monochromator : One given
Excitation
Emission
Sample
Monochromateur scanning all wavelengths
600500 700400
Quantification of the fluorescence emitted many wavelengths
Which (s) give(s) rise to the Fluorescence emission at a given ?(Excitation spectra are often similar to absorption spectra)
Excitation spectrum
Fluorescence of marine picocyanobacteria : Synechococcus spp.
600500 700400
Excitation spectrum
Marine phycoerythrins & spectrofluorimetry
600500 700400
Emission spectrum
ExcitationIn the
blue-greenregion,
at 500nm(for instance)
Emission between
560-580 nmdepending on the
type of PE
. There are several types of phycoerythrins (PE)
VariableExcitation between 400 and 550 nm
Emission at 580 nm(for instance)
One or two major maxima
495
545
Phycoerythrin structure and excitation spectra
Phycobiliprotein = Apoprotein + pigment
Pigment = chromophore phycobilin
One or two types of phycobilin are boundto marine phycoerythrins
600500 700400
Excitation spectrum
One or two major maxima
495
545
PAM Fluorimetry and photosynthetic organisms
Diving-PAM©
Monitoring-PAM© Junior-PAM©
Multicolor-PAM©
PAM Fluorimetry and photosynthetic organisms
Objectif : étudier la régulation de l’activité du photosystème II
PAM Fluorimetry and photosynthetic organisms
Objectif : étudier la régulation de l’activité du photosystème II
Absorbed light energy
=
Fluorescence energy+
Photochemistry energy+
Heat energyAntenna
Open/closed PSII centres
Chl a
Antenna
Cen
tre
réac
tio
nn
el
Photochemistry
Fluorescence
Chl a
Antenna
Cen
tre
réac
tio
nn
el
Photochemistry
Fluorescence
Chl a
Antenna
Cen
tre
réac
tio
nn
elPhotochemistry
Fluorescence
Chl a
Antenna
Cen
tre
réac
tio
nn
el
Photochemistry
Fluorescence
Chl a
Cen
tre
réac
tio
nn
el
Photochemistry
Fluorescence
Heat
Chl a
Cen
tre
réac
tio
nn
el
Photochemistry
Fluorescence
Heat
PAM Fluorimeters
. Two types of light : - Modulated light : intermittent, low irradiance non actinic - Actinic light : continuous
Actinic light
Modulated light
Photosystem IIFluorescence
(red light)
Sample
Fiber optics
Conceptual diagram of theJunior-PAM
Signal display
Photo-multiplicator
PAM Fluorimetry : light response curves
Flu
ore
scen
ce (
AU
)
Time
Modulated light ON
F0
Actinic light ON
(Irradiance 1)
(Irradiance 2)(Irradiance 3)
(Irradiance 4)(Irradiance 5)
(Irradiance 6)
Actinic light
FM’ FM’FM’
FM’ FM’ FM’
Flash saturant
FV’
Ft
PAM Fluorimetry : light response curves
When increasing irradiancePSII reaction centres get more and more closed
PSII relative Electron Transfer Rate (rETR) = Irradiance x (FM’-Ft)/FM’
= Irradiance x FV’/FM’
(Irradiance : µmol photons / m² / s)
Flu
ore
scen
ce (
AU
)
Time
Modulatedlight ON
F0
Actiniclight ON
(Irradiance 1)
(Irradiance 2)(Irradiance 3)
(Irradiance 4)(Irradiance 5)
(Irradiance 6)
Actinic light
FM’ FM’FM’
FM’ FM’ FM’
Flash saturant
FV’
Ft
Flu
ore
scen
ce (
AU
)
Time
Modulatedlight ON
F0
Actiniclight ON
(Irradiance 1)
(Irradiance 2)(Irradiance 3)
(Irradiance 4)(Irradiance 5)
(Irradiance 6)
Actinic light
FM’ FM’FM’
FM’ FM’ FM’
Flash saturant
FV’
Ft
PAM Fluorimetry : light response curves
PSII rETR = Irradiance x (FV’/FM’)
PS
II rE
TR
Irradiance (µmol photons/m²/s)
Courbe PSII rETR versus Irradiance
PAM Fluorimetry : light response curvesP
SII
rE
TR
Irradiance (µmol photons/m²/s)
Pas de saturation
PS
II r
ET
R
Irradiance (µmol photons/m²/s)
Saturation du rETR
PSII antenna size : α
PS
II r
ET
R
Irradiance (µmol photons/m²/s)
α > α
ISAT
ISAT < ISAT
PAM Fluorimetry : light response curves
PS
II r
ET
R
Irradiance (µmol photons/m²/s)
Saturation without photoinhibition
PS
II r
ET
R
Irradiance (µmol photons/m²/s)
Saturation and photoinhibition
PAM Fluorimetry : light response curves
Example of application : Prochlorococcus ecotypesP
SII
rE
TR
Irradiance (µmol photons/m²/s)