MEMS-based electrochemical gas sensors

and wafer-level methods

HITHESH KUMAR GATTY

Doctoral Thesis Stockholm, Sweden 2015

Page 2: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

ii

Front cover images:

The front cover image on the left shows the photograph of the amperometric gas sensor. Top right TEM image shows the horizontal cross-section of the nanoporous membrane, each nanopore is approximately 40 nm in diameter, and the bottom right image shows the sensor packaged with a custom made chip holder.

TRITA-EE: 2015:030 KTH Royal Institute of Technology ISSN: 1653-5146 School of Electrical Engineering ISBN: 978-91-7595-661-9 Department of Micro and Nanosystems

Akademisk avhandling som med tillstånd av Kungliga Tekniska Högskolan framlägges till offentlig granskning för avläggande av teknologie doktorsexamen i elektrisk mätteknik fredagen den 02 oktober 2015 klockan 10:00 i Q2, Osquldas väg 10, Stockholm.

Thesis for the degree of Doctor of Philosophy at KTH Royal Institute of Technology, Stockholm, Sweden.

© Hithesh K Gatty, Oktober 2015

Tryck: Universitetsservice US AB, Stockholm 2015

Page 3: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

iii

Abstract

This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors and methods of fabrication.

This thesis presents the research in two parts. In the first part, a method to handle a thin silicon wafer using an electrochemically active adhesive is described. Handling of a thin silicon wafer is an important issue in 3D-IC manufacturing where through silicon vias (TSVs) is an enabling technology. Thin silicon wafers are flexible and fragile, therefore difficult to handle. In addressing the need for a reliable solution, a method based on an electrochemically active adhesive was developed. In this method, an electrochemically active adhesive was diluted and spin coated on a 100 mm diameter silicon wafer (carrier wafer) on which another silicon wafer (device wafer) was bonded. Device wafer was subjected to post processing fabrication technique such as wafer thinning. Successful debonding of the device wafer was achieved by applying a voltage between the two wafers. In another part of the research, a fabrication process for developing a functional nanoporous material using atomic layer deposition is presented. In order to realize a nanoporous electrode, a nanoporous anodized aluminum oxide (AAO) substrate was used, which was functionalized with very thin layers (~ 10 nm) of platinum (Pt) and aluminum oxide (Al2O3) using atomic layer deposition. Nanoporous material when used as an electrode delivers high sensitivity due to the inherent high surface area and is potentially applicable in fuel cells and in electrochemical sensing.

The second part of the thesis addresses the need for a high performance gas sensor that is applicable for asthma monitoring. Asthma is a disease related to the inflammation in the airways of the lungs and is characterized by the presence of nitric oxide gas in the exhaled breath. The gas concentration of above approximately 50 parts-per-billion indicates a likely presence of asthma. A MEMS based electrochemical gas sensor was successfully designed and developed to meet the stringent requirements needed for asthma detection. Furthermore, to enable a hand held asthma measuring instrument, a miniaturized sensor with integrated electrodes and liquid electrolyte was developed. The electrodes were assembled at a wafer-level to demonstrate the feasibility towards a high volume fabrication of the gas sensors. In addition, the designed amperometric gas sensor was successfully tested for hydrogen sulphide concentration, which is a bio marker for bad breath. Hithesh K Gatty Micro and Nanosystems, School of electrical engineering KTH Royal Institute of Technology, SE-100 44 Stockholm, Sweden

Page 4: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

iv

Page 5: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

v

To my Parents

Page 6: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

vi

Contents

List of Abbreviations.........................................................................................................ix List of Publications………………………………………………………….…………...xi

1 Introduction

1.1 Microelectromechanical systems (MEMS)………………….…….……………....1 1.2 MEMS based electrochemical sensor....…………………………………….….…2 1.3 Wafer-level fabrication and packaging of electrochemical devices.…..........….....3 1.4 Asthma diagnosis and a vision for hand-held measuring instrument......……........3 1.5 Objectives of this thesis ………………………………………………………..…5 1.6 Thesis structure………………………………………………………………...….5

2 Background to electrochemical sensors and materials

2.1 Introduction ……………………………………………………………………….7 2.2 Amperometric gas sensor ………………………………………………………....9

2.2.1 Working principle………………………………………………………....9 2.2.2 Gas diffusion electrode ….……………………………………………....10 2.2.3 Sensor components ……….…………………………………………..…12 2.2.4 Sensor characteristics …….……………………………………………...15

2.3 An electrochemical adhesive for bonding and debonding ………………………16 2.3.1 An electrochemically active adhesive: ElectRelease™ ………………....16 2.3.2 Principle of bonding and debonding …………………………………….17

2.4 Discussion ……………………………………………………………………….19

Page 7: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

vii

3 A novel method for temporary handling of a thin silicon wafer

3.1 Introduction …………………………………………..……………….…………21 3.2 Temporary wafer handling …………………………………………….………...24 3.3 Wafer bond strength experiment ………………………………………….……..26

3.3.1 Metallized silicon device wafer …………………………………………26 3.3.2 Highly doped silicon device wafer ……………………………………....28

3.4 Discussion ……………………………………………………………………….29

4 Functional nanoporous material using atomic layer deposition

4.1 Introduction.……………………………………………………………………...31 4.2 Fabrication of functional nanopores....…….…………………………………….33

4.2.1 Method to create open ended nanopores ………………………………...33 4.2.2 Dual layer ALD: A predictive approach………………………………....36

4.3 Discussion ……………………………………………………………………….39

5 Wafer-level fabrication of MEMS-based amperometric gas sensors

5.1 Introduction to a design of a nitric oxide sensor..……….....…………….………42 5.2 Fabrication and assembly of the nitric oxide sensor……..………………………43

5.2.1 Measurement set-up and instrumentation …...…………………………..44 5.2.2 Characterization …………………………….…………………………...44

5.3 Integrated and miniaturized nitric oxide sensor ……...……………….…………47 5.3.1 Sensor design and fabrication ……………...……………………………48 5.3.2 Analytical properties ………………………………………………….…50 5.3.3 Operating life time ………….…………………………………………...51

5.4 Integrated and miniaturized hydrogen sulphide sensor………………………….52 5.5 Discussion …………………………………………………………………….....52

6 Conclusions 53

Summary of Appended Papers 55

Acknowledgements 59

References 61

Paper Reprints 77

Page 8: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

viii

Page 9: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

ix

List of Abbreviations

3-D Three Dimension CMOS Complementary metal oxide semiconductor WE Working electrode RE Reference electrode CE Counter electrode ALD Atomic layer deposition PCB Printed circuit board ppb Parts-per-billion mlps mili liter per second PEALD Plasma enhanced ALD IC Integrated circuit DRIE Deep reactive ion etching MEMS Microelectromechanical system Rf-MEMS Radio frequency MEMS AFM Atomic force microscope TSV Through silicon via

SOI Silicon on insulator FET Field Effect Transistor FPGA Field programmable gate array

Chemical formulas

NaOH Sodium hydroxide CsOH Cesium hydroxide LiOH Lithium hydroxide HF Hydrogen fluoride HCl Hydrochloric acid KOH Potassium hydroxide

Page 10: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

x

HNO3 Nitric acid LiCl Lithium chloride LiBr Lithium bromide NaHCO3 Sodium bicarbonate DMSO Dimethly sulphoxide DMF Dimethlyformide THF Tetrahydrofuran PEO-LiX Lithium salt in a polymer matrix PVA/H3PO4 Polyvinylalcohol/phosphoric acid ZrO2 Zirconium oxide NO Nitric oxide H2S Hydrogen sulphide CO Carbon monoxide O2 Oxygen CPT Cyclopentanone H2SO4 Sulphuric acid NH3 Ammonia Bmim NTf2 1-Butyl-3-methylimidazolium

bis(trifluoromethanesulfonyl)imide Emim NTf2 1-ethyl-3-methylimidazolium

bis(trifluoromethylsulfonyl)imide MeCpPtMe3 methylcyclopentadienyl- trimethyl platinum TMA Trimethylaluminum

Page 11: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

xi

List of Publications

The presented thesis is based on the following international reviewed journals:

A. “Temporary wafer bonding and debonding for 3D integration using an electrochemically active polymer adhesive”, H. K. Gatty, S. Schröder, F. Niklaus, N. Roxhed and G. Stemme

ECS Journal of Solid State Science and Technology, vol. 3, pp. 115-121, 2014.

B. “Pt–Al2O3 dual layer atomic layer deposition coating in high aspect ratio nanopores”, G. Pardon, H. K. Gatty, G. Stemme, W. van der Wijngaart and N. Roxhed

Nanotechnology, vol. 24, p. 015602, 2013.

C. “An amperometric nitric oxide sensor with fast response and ppb- level concentration detection relevant to asthma monitoring”, H. K. Gatty, S. Leijonmarck, M. Antelius, G. Stemme and N. Roxhed

Sensors and Actuators B:Chemical, vol. 209, pp 639–644, 2015

D. “A wafer level liquid cavity integrated amperometric gas sensor with ppb level nitric oxide gas sensitivity”, H. K. Gatty, G. Stemme and N. Roxhed Accepted to Journal of Micromachines and Microengineering, 2015

E. “An amperometric hydrogen sulphide sensor applicable for bad breath monitoring”, H. K. Gatty, G.Stemme and N. Roxhed Manuscript, 2015

Page 12: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

xii

The contributions of Hithesh K Gatty to the different publications:

A. Major part of design, all fabrication, Major part of experiments and major part of writing

B. Major part of design, major part of fabrication, major part of experiments and part of writing

C. Major part of design, all fabrication, all experiments and major part of writing

D. Major part of design, all fabrication, all experiments and major part of writing

E. Major part of design, all fabrication, all experiments and major part of writing

Page 13: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

xiii

The work has also been presented at the following international reviewed conferences:

A. “A ppb level, miniaturized fast response amperometric nitric oxide sensor for asthma diagnostics”, H. K. Gatty, S. Leijonmarck, M. Antelius, G. Stemme, and N. Roxhed

Proceedings of the IEEE 26th Int.Conf. on Micro Electro Mechanical Systems(MEMS), Taipei, Taiwan, pp. 381-384, Jan 2013.

B. “Temporary wafer bonding and debonding by an electrochemically active polymer adhesive for 3D integration”, H. K. Gatty, F. Niklaus, G. Stemme, and N. Roxhed

Proceedings of the IEEE 26th Int.Conf. on Micro Electro Mechanical Systems(MEMS), Taipei, Taiwan, pp. 381-384, Jan 2013.

C. “Temporary wafer bonding of thin wafers by electrochemically active polymer”, H. K. Gatty, F. Niklaus, G. Stemme, and N. Roxhed

Proceedings of the Int.Conf. on Wafer Bonding for Microsystems 3D-and Wafer Level Integration, Stockholm, Sweden, pp. 143-144, Dec 2013 (Oral Presentation)

The work has also been presented at the following workshop:

D. “A ppb-level miniaturized amperometric nitric oxide sensor” H. K. Gatty, S. Leijonmarck, M. Antelius, G. Stemme, and N. Roxhed

Medicinteknikdagarna, Stockholm, Sweden, Oct 2013

P.S: Integrated nitric oxide (NO) sensor design (paper D) also appeared in the article titled “MEMS är mums för hälsan” (MEMS are good for your health) under the theme : Medicinteknik published in the newspaper Electroniktidningen, Dated April 2014, written by Anna Wennberg.

Page 14: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

xiv

Page 15: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

1

Chapter 1

Introduction

1.1 Microelectromechanical systems (MEMS)

Today, a mobile phone contains more than ten microelectromechanical system (MEMS) based devices. The number of mobile phones is predicted to increase to approximately two billion by 2020, which would escalate the demand for MEMS devices. The advantage of MEMS devices is their ability to perform complex functions given their small size. MEMS devices are often used in applications such as navigation, image processing and sound processing that are required in mobile phones.

MEMS devices can generally be classified into three categories: sensors, actuators and functional structures. Devices that transduce force, temperature, pressure or any other change in the environment into electrical signals can be classified as sensors whereas actuators essentially convert electrical signals to mechanical motion, heat, or any other form of energy. MEMS based structures are functional elements typically used to assist a device for example microchannels in a micropump, nanoporous structures for high surface area devices, microneedles for drug delivery devices etc. There are typically five main processes that involve the development of MEMS based devices. They are bulk micromachining, surface micromachining, high aspect ratio (HAR) micromachining, wafer bonding and packaging.

Page 16: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

2

The application of MEMS process can be tracked back to the fabrication of a suspended beam structure as a gate electrode in a resonant gate transistor device, by Nathanson et al. in the late 60’s [1]. However, it is only in the early 80’s that MEMS technology started to gain importance with devices such as flow sensors [2, 3], pressure sensor [4], resonator [5] etc. MEMS based actuators such as micromirrors emerged some years later [6]. Recently, functional microstructures have gained importance with devices such as microneedles for painless drug delivery [7] and as microchannels for Lab-on-Chip applications [8]. The success of MEMS technology has given rise to several new interdisciplinary fields such as bio-MEMS [9], optical MEMS [10, 11], microfluidics [8, 12, 13] and RF MEMS [14, 15].

In the future, a comprehensive measurement system such as a personalized health monitoring system using mobile phones could be envisioned in which MEMS devices are likely to play a key role. This calls for the development of novel MEMS based devices in the field of medical systems such as electrochemical sensors, glucose level sensors, heart rate monitor etc. MEMS based electrochemical sensors can provide the paradigm shift necessary for the development of medical sensor systems.

1.2 MEMS based electrochemical sensor

Electrochemical sensors essentially generate electrical signals in response to electrochemical reactions. The reactions typically scale with the surface area of the electrodes i.e., higher surface area provides higher reaction sites are available for an electrochemical reaction. However, increasing the surface area of the electrode more often leads to increased size of the sensor. In order to realize a sensor with a small size while maintaining a high surface area for electrochemical reactions, the ratio of the surface area to volume needs to be increased. In addressing this need, nanoporous structures are becoming popular in realizing a high surface area while maintaining a small size of the sensor. To realize nanoporous structures, MEMS has been identified as a technology to achieve most of the ideal requirements [16]. This is because the MEMS technology allows accommodating the scaling laws, for example fabricating nanoporous electrodes using advanced process techniques such as atomic layer deposition (ALD). Therefore, it is possible to realize high performance electrochemical devices that could be applicable in medical systems such as sensors for asthma detection.

Page 17: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

3

1.3 Wafer-level fabrication and packaging of electrochemical devices

Wafer-level packaging is a method to package a device while it is still at the wafer-level, as compared to dicing the device from the wafer followed by individual packaging. Several methods are available for wafer-level packaging, one of the popular method used in realizing MEMS devices is by bonding a device wafer with a cap wafer [17]. Wafer-level fabricated and packaged electrochemical devices have been studied in realizing glucose sensors [18]. In this method, the electrodes of the sensors are fabricated on a Pyrex wafer and then anodically bonded to a silicon wafer containing microneedles with flow through channels.

1.4 Asthma diagnosis and a vision for hand-held measuring instrument

Figure 1.1: Airways of lungs under normal condition and during an asthmatic attack. An asthmatic patient has a constricted airways and higher production of nitric oxide (NO) along the airways. From [19].

Asthma is a medical condition that is characterized by an inflammation of the airways of the lungs [20-22]. Inflammation of the airways is predominantly caused by air pollution and airborne pollen grains [23, 24]. A pictorial description of a normal airway and an airway during an attack is shown in figure 1.1. Due to the inflammation, the airways in the

Page 18: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

4

lungs release nitric oxide (NO) gas, the concentration of which is used to pre-determine the level of inflammation [21, 25-29]. A concentration of NO above 50 parts-per-billion (ppb) in adults typically indicates the presence of an inflammation [30, 31]. The standardized procedures indicate that the steady state nitric oxide gas concentration from exhaled breath is obtained when the breath is exhaled at a flow rate of 50 mlps and lasts for 10 s [32]. Ideally, a sensor should detect ppb-level nitric oxide gas concentration and respond within 10 s for detecting asthma. In addition, the sensor needs to be selective to interfering gases that are present in the exhaled breath, such as carbon monoxide (CO) and ammonia (NH3). Furthermore, the sensor needs be able to operate at a high humidity since the exhaled breath contains humidity close to 100 %RH. These parameters of the sensor are ideal to be applicable in an asthma measuring instrument.

Figure 1.2: Image of a child breathing into an asthma detecting instrument. The image gives an indication of the size of the instrument. From [33].

Commercially available sensors detect nitric oxide concentration with a ppb-level sensitivity and has response time of the order of 40-60 s [34]. The implication of a slow response on a NO measuring instrument is undesirable. A typical NO measuring instrument that is available in the market is shown in the figure 1.2, which shows a child holding the measuring instrument with his hands [33, 35]. It can be observed that the measuring instrument is considerably large and requires both hands to hold the instrument. One of the reasons, for bigger size of the instrument, is the long response time of the nitric oxide sensor used in the instrument. Due to the long response time, the instrument needs to buffer the exhaled breath sample i.e., the instrument stores the exhaled breath in a reservoir and then pumps to the NO gas sensor at a slower rate. This requires the instrument to have gas sampling and handling systems to buffer and pump the sample, which results in a large

Page 19: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

5

size of the instrument. Other methods of measuring NO using a potentiometric sensor and indirect conversion of NO has been demonstrated [36-38]. Although this has shown promising results for breath samples, the applicability to asthma monitoring is currently not validated with studies on several patients.

In order to realize a hand-held instrument, the size of the measuring instrument should be small so that patients can carry along with them. One of the objectives addressed in this thesis is the development of a miniaturized and fast sensor that eliminates the need for buffering of the breath sample. MEMS technology offers an efficient way to manufacture the sensors at a wafer-level, by using wafer-level fabrication and packaging of sensors. Therefore, a cost effective, hand-held, real time health monitoring system can be envisioned, and exhaling the breath into a mobile phone would enable to determine asthma in a real time mode.

1.5 Objectives of this thesis

This thesis presents research in the field of MEMS and specifically in the area of electrochemical sensors and their method of fabrication. The objective of this thesis is (a) to give a background to the electrochemical sensors and materials (b) to describe a novel method for temporary wafer handling of a thin wafer (c) to introduce fabrication of functional nanoporous membranes relevant for electrochemical sensing, and finally (d) to describe the development of a novel MEMS based gas sensor relevant for asthma and bad breath monitoring.

1.6 Thesis structure

For the sake of coherence the text in the thesis is organized using the general structure as detailed below.

In chapter 2, electrochemical concepts of gas sensing and temporary wafer handling are presented.

In chapter 3, the application of electrically induced adhesive for temporary wafer handling is described.

In chapter 4, fabrication of a highly porous functional material using atomic layer deposition (ALD) is described.

In chapter 5, the design and development of novel MEMS based gas sensor relevant for asthma and bad breath monitoring is described. Furthermore, integration and miniaturization of a novel MEMS based gas sensor is presented and its applicability to hand-held instruments is discussed and finally, chapter 6 concludes this work.

Page 20: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

6

Page 21: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

7

Chapter 2

Background to electrochemical sensors and materials

In this chapter, a brief introduction to amperometric sensors is presented. The working mechanism and the components that are used in an amperometric gas sensor are discussed. In addition, an introduction to the materials and properties of an electrochemically active adhesive is presented.

2.1 Introduction

Detection of gases involves several methods; some of them are listed in table 2.1. Among the various detection methods, amperometric detection has distinct advantages, i.e. high sensitivity, fast response time and high selectivity. In addition, it can be developed using materials that are commercially available which leads to low cost sensors as compared to other type of sensors. Although metal oxide sensors can be fabricated using a simple process method, their slow response, cross interference to other gases and high power consumption constitute considerable drawbacks [39, 40].

Page 22: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

8

Table 2.1: Gas sensing types and principles

One of the first amperometric gas sensor was reported by Qlark in the mid-1950s to monitor dissolved oxygen content in blood [47]. As shown in figure 2.1, the gas sensor consisted of two electrodes, electrolyte and a semi permeable membrane. The oxygen gas diffuses through the semi permeable membrane, and thereafter the electrolyte before reaching the working electrode. The thickness of the semi-permeable membrane controlled the diffusion of oxygen to the working electrode.

Figure 2.1: Schematic representation of the Qlark sensor, an amperometric sensor that was developed to detect oxygen. From [47].

Type Principle

Metal-oxide-semiconductor [39] Measurement of change in the resistance

NDIR sensor [41] Target gas absorption coefficient, comparison

Catalytic gas sensor [42] Resistance change measurement after reaction with the catalytic bead

Fluorescence sensor [43] Detection of wavelength change of the fluorescence

Acoustic wave sensor [44] Change in the resonant frequency

Amperometric sensor [45, 46] Measurement of change in the working electrode current

Nylon cap

Silver wire

Platinum cathode

Polyethylene membrane

Indium solder

Lucite tube

Electrolyte

Page 23: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

9

Following this, amperometric sensors evolved to include a semi permeable membrane as the control element to select specific gases. Modern amperometric gas sensors employ three electrodes with reference electrode as the third electrode that allowed accurate control of the potential to the working electrode [45, 48, 49].

Currently, the research on amperometric gas sensors is typically divided into two fields. One field focusses on increasing the surface to volume ratio of the working electrode in order to realize a small geometric footprint with a high surface area electrode [50-52]. In the second field, the research aims at increasing the operating life time of the sensor by e.g. replacing the conventional acidic/basic electrolyte with ionic liquid electrolytes [46, 53, 54].

2.2 Amperometric gas sensor

In this section, the working principle and a brief overview of the main components of an amperometric sensor are presented. In addition, analytical characteristics of a gas sensor are described.

2.2.1 Working principle

The working principle of an amperometric gas sensor for detection of nitric oxide gas is illustrated in figure 2.2. Detection of the gas involves three major steps. In step 1, the sample gas containing the target gas (NO) is transported to the sensor either by diffusion or by pumping. In step 2, the gas diffuses to the working electrode, where NO undergoes oxidation. The electrons generated by the reaction at the working electrode result in a current that is directly proportional to the gas concentration. In step 3, the reaction products are diffused away from the electrode surface. If the reaction products remain on the electrode surface, it decreases the number of active surface sites for reaction and hence leads to a lower sensitivity. This effect is termed as poisoning of the electrode.

Page 24: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

10

Figure 2.2: Schematic diagram illustrating the steps involved in amperometric detection of nitric oxide gas. The electrochemical reaction of the gas occurs at the three phase point which is formed by the interface of the electrode, electrolyte and NO gas.

Therefore, a major consideration has to be given to the design of the working electrode of the sensor in terms of choice of geometry and composition of materials. A detailed investigation of a type of working electrode, i.e. gas diffusion electrode (GDE) is presented in the next section.

2.2.2 Gas diffusion electrode

An electrochemical reaction (oxidation or reduction) of a gas occurs at the interface between the working electrode, electrolyte and the gas, called as the triple phase point. Typically the working electrode of a sensor is composed of noble metals due to their inertness to chemical reactions or of carbon embedded with noble metals such as Platinum (Pt).

For NO gas, the solubility and diffusivity in liquid electrolytes is poor. For example, in aqueous electrolyte at 20 °C, the solubility of NO is approximately 1.94 mM/atm [55] whereas the solubility of O2 is approximately 0.3 mM/atm. The low solubility of NO severely inhibits its transport through the electrolyte towards the electrode. Hence, an optimum design of the working electrode would provide an efficient transport path of the gas towards the electrode minimizing the diffusion length of the gas in the electrolyte.

The geometry of a gas diffusion electrode (GDE) allows electrochemical reaction to occur at a three phase point minimizing the diffusion length of the gas in the electrolyte.

Page 25: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

11

Figure 2.3 (a) and (b) shows a cross section of a typical GDE and the three phase zone formed in a GDE. It consists of a narrow space in the electrode for the diffusion of gas and interaction with the electrolyte. At the three phase point, the effective concentration of the gas becomes sufficient to sustain a high current density [56]. Porous working electrode provides with large number of three phase points.

Figure 2.3 Three phase points in a gas diffusion electrode. (a) Cross section of a single pore of an electrode showing the electrolyte, gas and the reaction zone.(b) Concept of a three-phase point at an electrode. From [56].

One of the earliest gas diffusion electrode was developed by Schmidt in 1923 who used a porous layer of platinum black that was electrolytically deposited on a hollow piece of carbon [57]. Over the last decade, gas diffusion electrodes have gained renewed interest in terms of design and development, especially for gas sensors and fuel cells [58-61]. In fuel cells, GDEs are used for both working and counter electrodes whereas in gas sensors it is mainly used as a working electrode.

The following section gives a detailed description of the components of an amperometric sensor namely filters, electrodes, electrolytes and potentiostat.

(a) (b)

Page 26: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

12

2.2.3 Sensor components

Figure 2.4 shows a schematic illustration of a three electrode amperometric gas sensor. A brief description of the components that are used in the sensor is given in this section.

Figure 2.4: Schematic illustration of an amperometric gas sensor and its components. The three electrodes together with an electrolyte are sealed inside a plastic casing. The output pins are used for biasing the electrodes and to record the current from the working electrode.

Filter: Filters in gas sensors mainly have the following purposes:

• Decrease poisoning of the electrodes • Enhance the selectivity • Reduce aqueous electrolyte loss • Increase sensor life time In addition, the filter also protects the sensor from an external hazardous environment. Several types of gas permeable filters exist. Typical filters include very thin layer (~20 um thick) of micro-porous Teflon film, PTFE modified porous films [62] or silicone films. Thickness and durability of the filter can be selected depending on the choice of the application. Filters can also be employed to remove unwanted gases by impregnating it with a gas sensitive chemical. For example, NO2 gas can be selectively removed from a mixture of NO and NO2 gas by using a filter dipped with triethanolamine [63].

Page 27: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

13

Moreover, the filter controls the evaporation of the liquid electrolyte, i.e. using a thick filter leads to slow evaporation of the electrolyte. Therefore, an optimally thick filter would allow a low rate of electrolyte evaporation without sacrificing the gas diffusion time.

Working electrode: The electrochemical reaction of nitric oxide gas occurs at the working electrode resulting in a current proportional to the concentration of the gas. Several kinds of NO sensitive electrodes have been reported in the literature and have recently been commercialized [64, 65]. Four different approaches can be identified in developing NO sensitive electrodes. A first and popular approach is using a modified Clark type electrode, in which a thin layer of electrolyte separates a semipermeable membrane and the working electrode. In this approach, the O2 permeable membrane of the Clark type sensor is replaced with an NO permeable membrane such as chloroprene [66], nitrocellulose [67] and silicone rubber [68] to make the sensor NO sensitive. A second approach (GDE approach) consists of coating a NO selective polymer directly on the surface of the electrode. For example, coating of Nafion or cellulose acetate polymer on the electrode surface has been used [69]. Electrocatalysis of NO by a modified iridium or palladium layer can be classified as the third type of NO sensitive electrode [70]. Finally, the fourth type of approach, which is commercially popular, is to use glassy carbon electrodes as the working electrode for NO sensors [71, 72].

Electrolyte: The function of the electrolyte is to support redox reactions, efficiently carry ionic current, and most importantly providing a stable reference potential to the reference electrode. It should also provide temperature stability, wide operating potential range and low viscosity. Four types of electrolytes exist: aqueous, non- aqueous, solid polymers and ionic liquids. A list of different electrolytes that have been used for gas detection is given in table 2.1.

Page 28: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

14

Table 2.1: Few examples of electrolytes. From [73, 74] Type Electrolytes Aqueous electrolytes

Acid: sulphuric, perchloric, phosphoric, HCl, HNO3, acetic acid

Base: NaOH, CsOH, LiOH, KOH

Neutral and salt: KCl, LiCl, LiBr, NaHCO3, acetate.

Non-aqueous Solvents: propylene, carbonate, DMSO, DMF, acetonitrile, ethylene glycol, dichloromethane, propylene glycol, pyridine, THF

Solid polymers

Nafion, PEO*LiX, PVA/H3PO4, PVA/CsOH, and Inorganic Solids, ZrO2

Mixtures: Water-glycol with salts etc.

Ionic liquids [Bmim][NTF2], [Emim][NTF2] etc

Potentiostat: One of the main advantages of an amperometric gas sensor is the low operating voltage and current ranges. Typically gas sensors with aqueous electrolytes operate between 0 and ±1.2 V with a current of few hundred nanoamps. Therefore, a simple printed circuit board (PCB) potentiostat can be used for measuring the current from the working electrode, as well as applying the bias voltage (Vb) between the working electrode (WE) and reference electrode (RE). Figure 2.5 shows a simplified schematic of the potentiostat.

Figure 2.5: Schematic of a PCB based potentiostat connected to a three electrode gas sensor. Two operational amplifiers (OP1 and OP2) are used to measure the current from the working electrode and to bias the reference electrode, respectively.

-+

Rf

Vout

-+

WE

CE

RE

Vb

OP1

OP2

i=0

iWE

Gas sensor

Page 29: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

15

Two operational amplifiers OP1 and OP2 are connected to the electrodes of the gas sensor as illustrated in the figure 2.5. Operational amplifier OP1 is a transimpedance amplifier that converts the working electrode current to an output voltage, i.e. Vout=IWE*Rf, where Rf is a resistor typically of few MΩ. Operational amplifier OP2 regulates the voltage at the reference electrode by maintaining the current, i=0. OP2 also acts as a current sink to the counter electrode. Thus, a constant bias voltage between the WE and RE can be maintained. Further details about the construction of a PCB based potentiostat can be found in [75]. An integrated programmable chip, LMP 91000 from Texas Instruments can also be used to construct a potentiostat [76].

2.2.4 Sensor characteristics

An amperometric gas sensor has the following characteristics:

• The gas concentration is proportional to the working electrode current, and is recorded between the working and the counter electrodes while maintaining a constant potential at the working electrode.

• The combination of the geometry for gas exposure, electrolyte confinement, electrode material and the electrochemical method (potentiometric or amperometric) determine the sensitivity and selectivity of the sensor.

• Low cost, small size and low power consumption are the most interesting characteristics of an amperometric sensor.

The three important parameters that are commonly used to assess the quality of an amperometric gas sensor are the sensitivity, selectivity and response time.

Sensitivity: Sensitivity of a gas sensor is the ratio between the working electrode currents and the respective gas concentrations. In practice, the sensitivity is thus determined from the slope of a plot of the working electrode current as a function of the gas concentration. The sensitivity of the gas sensor can be physically controlled by modifying the surface area of the working electrode. Diffusivity of the gas in the electrolyte affects the sensitivity of the sensor. In addition, filters, membranes and other sealing objects can influence the sensitivity.

Selectivity: Selectivity of a gas sensor is its ability to sense the desired gas while being insensitive to other interfering gases. The selectivity to undesired gases is obtained by choosing the type of electrochemical reaction of the target gas i.e., either oxidation or reduction. For example, in a mixture containing NO and NO2 gas, it is possible to selectively oxidize NO gas. Selectivity based on kinetics depends on the electrode material and its catalytic behavior to the gases. For example, CO gas undergoes fast oxidation on Pt electrode, whereas it is 103 to 106 times slower on the Au electrode [73]. Thus, the sensor

Page 30: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

16

can be made selective by choosing the right electrode material. Selectivity can also be obtained by using appropriate filter membranes designed for a particular gas (See section 2.2.3 for further details).

Response time: The advantage of an amperometric gas sensor is its ability to detect gas within few seconds. The response time of an amperometric sensor is mostly limited by the diffusion of the gas to the working electrode. For sensors applicable to the medical field such as breath analysis (e.g. CO, NO, O2), the response time needs to be few seconds or less [32]. A response time of < 1 s for NO gas using a silicon based sensor was reported by Tierney et al. in which a direct diffusion path for the gas was designed [77]. A comprehensive analysis of the response time of an amperometric gas sensor can be found in [78].

2.3 An electrochemical adhesive for bonding and debonding

This section of the chapter presents the working principle of an electrochemically active adhesive used for the bonding and debonding of two electrically conductive substrates.

2.3.1 An electrochemically active adhesive: ElectRelease™

ElectRelease™ (EIC laboratory, USA) is a thermosetting adhesive that was initially developed for bonding and debonding of aluminum parts typically used in food and beverage packaging industries and for interactive packaging solutions [79-81]. The debonding process also works for a number of metals including aluminum, steel, low alloy steel, stainless steel, copper and titanium [81]. In this work, ElectRelease™ H23 was applied to wafer fabrication processes due to its low viscosity and its ability for slow cross-linking of the adhesive after mixing. The slow cross-linking time allows for several fabrication process steps such as adhesive dilution, spin coating and wafer alignment that are needed for pre-bonding of wafers. A photograph of the ElectRelease™ H23 adhesive is shown in figure 2.6, which contains a base and a hardner in separate tubes.

Page 31: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

17

Figure 2.6: A photograph of an ElectRelease™ H23 adhesive. The cartridge contains a base and a hardner in separate tubes.

2.3.2 Principle of bonding and debonding

In practice, the two parts of the adhesive are mixed in the ratio of 4 (base):1 (hardner). The composition of the H23-base contains silicone fluid, bisphenol-A, epoxy resin, PEG-n-diglycidylether and salt (NH4PF6). The composition of the H23-hardner contains mercaptan and amines [82]. The hardner is an initiator of the cross-linking process of the adhesive. Once the cross-linking occurs, the adhesive mixture has a matrix functionality and an electrolytic functionality. The matrix functionality helps to adhesively bond to a substrate, whereas the electrolytic functionality provides the ionic conductivity between the substrates. A schematic illustration of the adhesive morphology before and after curing is shown in figure 2.7.

Before curing

Epoxy structure before curing

Epoxy structure after curing

After curing

A=PEG

B=PDMS

Figure 2.7: Schematic representation of the adhesive morphology before and after curing. The phase separation of PDMS-graft-PEG comb polymer in nanostructured epoxy resin. From [83].

Page 32: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

18

Three basic steps of bonding and debonding of metal substrates, typically aluminum, using ElectRelease™ adhesive are illustrated in figure 2.8. In step1, the adhesive is applied to one of the substrate, then joined to the other substrate and allowed to cure either at elevated temperature or at room temperature. In step 2, a voltage typically in the range of 10-50 V is applied to the metal substrates. At room temperature, the debonding of the substrates is induced when a voltage of 45 V is applied for 10 s or at 15 V for 20 min [84]. In step 3, the metal substrates are debonded manually. Normally, the adhesive remains in the cathode part leaving the clean anode part [84, 85].

A comprehensive experimental investigation using surface analytical techniques to understand the mechanisms of debonding of the electrochemically active adhesive can be found in [83, 86-88] . In short, four possible processes at aluminum (anode) -adhesive interface that leads to debonding were identified i.e. anodic undermining, blistering debonding, osmotic blistering and anodic oxidation. One or a combination of several processes could be identified as the cause of debonding of substrates.

Figure 2.8: Illustration of the debonding technique using the ElectRelease™ adhesive.

Met

al

Met

al

ElectRelease bond between two metals

ElectRelease debond at anode interface

Met

al

Met

alions

Met

al

Met

al+ -V

e- e-

Step 1 Step 2 Step 3

i+

Page 33: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

19

2.4 Discussion

Electrochemical sensors such as amperometric gas sensors have distinct advantage over other gas sensors in terms of high sensitivity, high selectivity and fast response time. Amperometric sensors can be sensitive to very low (< 25 %RH) or very high humidity (>80 %RH) changes. On the other hand, for most of the operating region, amperometric sensors seem to be promising device to detect and measure gas concentrations. The combination of the MEMS technology and amperometric detection of gas can potentially offer low cost sensors. Further research in the direction of increasing the life time of the sensor would be desirable. The properties of the electrochemically active adhesive and the debonding mechanisms have been presented. Four possible mechanisms were identified for debonding adhesively bonded substrates. The debonding of the substrates could likely be caused by one or more mechanisms involved in the debonding process.

Page 34: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

20

Page 35: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

21

Chapter 3

A novel method for temporary handling of a thin silicon wafer In the previous chapter, a brief introduction to an electrochemically active adhesive and the method for bonding and debonding of two metal substrates were presented. In this chapter, the application of an electrochemically active adhesive to thin silicon wafer handling will be described.

3.1 Introduction

The progress in the packaging of individual IC dies has evolved from a simple ‘wire bonded’ packaging to a complex 3D-IC packaging realized using through silicon vias (TSVs) as shown in figure 3.1. Next generation highly miniaturized and complex IC systems such as MEMS based CMOS image sensors would be developed using 3D-IC integration in which TSVs are a key enabling technology. Emerging applications such as photonics based on silicon interposers are also being developed for products in smart phone and tablets [89]. A silicon interposer is made from a thin silicon wafer in which TSVs are used to connect two or more functionalities in a die. The latest implementation of this technique can be found in Xilinx FPGA virtex 7 SSI (stacked silicon interconnect) device incorporating a thin silicon wafer interposer with TSVs as shown in figure 3.2.

Page 36: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

22

Figure 3.1: Illustration of a roadmap in the progress of packaging of dies. TSV’s would be the enabling technology for the integration of several functionalities on a single die. From [90].

Figure 3.2: Cross section SEM image of Xilinx Virtex 7 SSI device incorporating thin silicon wafer (interposer) with TSVs to connect FPGA with the packaging substrate. From [91, 92].

One of the methods of fabricating TSVs is by etching vias followed by deposition of a conductive material to make the vias conductive, and thereafter thinning the silicon wafer.

Page 37: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

23

Hence, this process involves handling of a thin silicon wafer which is often less than 50 µm in thickness. Thin silicon wafer are highly flexible and thus fragile. Therefore, a reliable solution is needed for handling such thin silicon wafers without causing any damage.

Handling of a silicon wafer can be carried out by adhesively bonding the wafer (device wafer) to be thinned to a thick silicon wafer (carrier wafer). In that way, the carrier wafer is used as a support. Three main process steps are involved in this method. They are bonding, thinning and debonding.

Bonding: Several types of polymers have been investigated for the use in adhesive wafer bonding; including epoxies, thermoplastics and photoresists [93-97]. For temporary bonding in wafer handling applications, both bonding and debonding properties of the adhesive are important. The adhesive should withstand the mechanical and thermal processes that are part of the, for example, TSV fabrication processes [98]. An efficient process of bonding requires the adhesive to be uniform and should allow the device wafer to be released from the carrier wafer without any damage [99, 100].

Wafer thinning: There exists four processes for thinning a wafer: mechanical grinding, chemical mechanical polishing (CMP), wet etching and dry etching [101]. Mechanical grinding is often used for thinning a wafer due to its high rate of thinning and low cost. However, this method of thinning process leads to a rough surface on the wafer [102]. CMP is often used in combination with mechanical grinding to polish the grinded wafers. Isotropic wet etching of silicon is typically done using a HNA solution (HF: Nitric acid: Acetic acid). However, it is difficult to maintain a constant etching rate over the entire surface, hence difficult to get a uniform final thickness [103]. Dry etching (Plasma based Reactive Ion Etching) in combination with CMP allows for a better control of wafer thickness [104]. Combinations of several methods are also prevalent to get a highly uniform surface of the silicon wafer [105].

Debonding: In one of the methods of debonding, the adhesively bonded wafer stack is heated to an elevated temperature for the adhesive to soften and become liquid [106-110]. The device wafer is then released by sliding it off from the carrier wafer. However, one drawback of this method is the high temperature (>200 °C) needed for debonding. Also, sliding a thin device wafer from the carrier wafer may damage the device wafer. In another method, the carrier wafer is coated with two types of adhesives having different bond strengths, as for example in the Zone bond ™ technique [111]. In this technique, one type of adhesive with low bond strength is applied to the center of the wafer and another type of adhesive with higher bond strength is applied the edge of the wafer. The adhesive with higher bond strength is removed by a solvent treatment followed by mechanical delamination of the device wafer. This method requires several time consuming processing

Page 38: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

24

steps. Laser ablation of a polymer sacrificial layer is another method in temporary wafer handling [112-115]. This method was used in controlled transfer of dies between two wafers using, e.g. polyimide or PET as an adhesive for manufacturing low-cost AFM devices [116, 117]. Table 3.1 summarizes few methods and causes of adhesive debonding.

Table 3.1: Methods of adhesive bonding and causes of adhesive debonding. From [108, 114, 118-120].

Methods of bonding Causes of debonding Thermal stability (° C) Thermal release tapes Heat induced 80-130 Thermoplastic adhesive Chemically dissolved 150 Thermosetting adhesive Voltage induced 150-200 Photothermal polymer Laser ablation. 150-200

Thermosetting adhesive is a promising material for adhesive debonding since the bonding and debonding properties can be modified without difficulty. Several articles and patents are available describing the formulations with debonding properties [121]. In the next section, a thermosetting adhesive that enables controlled debonding of a thin wafer will be described.

3.2 Temporary wafer handling

In this work, ElectRelease™ H23 a thermosetting adhesive was utilized for the temporary wafer bonding of a thin silicon wafer. Figure 3.3 illustrates the fabrication process that was developed for temporary wafer handling. In summary, the ElectRelease™ adhesive was diluted with cyclopentanone (CPT) to get a low viscosity adhesive solution. This adhesive solution was spin coated on the carrier wafer in a similar way to a photoresist spin coating. The device wafer was then manually placed above the spin coated carrier wafer. The resulting wafer stack was transferred to the wafer bonding tool for bonding two wafers. Thereafter, the device wafer of the resulting wafer stack was thinned from 100 µm to 40 µm using RIE etching. In order to assess the applicability of this adhesive for temporary wafer handling, bond strength experiments on the prepared wafer stack were carried out and are described in the next section. For more processing details please refer to paper A.

Page 39: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

25

Figure 3.3: Illustration of the fabrication process flow for bonding a device wafer to a carrier wafer with ElectRelease™ H23 electrochemical polymer adhesive. (a) Metal deposition on the device wafer. (b) Spin coating on the carrier wafer with diluted ElectRelease™ H23 adhesive solution. (c) Illustration showing the arrangement of the wafer stack after adhesive bonding of the aluminum coated device wafer. (d) Photograph of the bonded stack showing the electrical contact area of the device wafer which is exposed at the carrier wafer flat. From [paper A].

Page 40: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

26

3.3 Wafer bond strength experiment

Two different types of silicon device wafers were investigated for debonding. One of the device wafers was sputtered with a thin aluminum layer and the other was a highly doped silicon wafer.

3.3.1 Metallized silicon device wafer

To enable the bond strength experiment before and after switching ‘ON’ the voltage, the wafer stack was fixed to a 3 mm thick aluminum plate providing an electrical contact to the carrier wafer. A UV release tape was attached to the device wafer to facilitate debonding and “peel-off” evaluation experiments. To measure the bond strength of the adhesive before switching ON the voltage, a dynamometer was attached to a corner of the UV tape and pulled upwards, thereby exerting a pull force on the attached device wafer as illustrated in figure 3.4(a). The photograph in figure 3.4(b) shows a high bond strength of the electrochemically active adhesive, since the UV tape is peeled off without releasing the device wafer.

Figure 3.4: (a) Illustration of the test set-up for measurement of bond strength. (b) Photograph showing the force that is applied to the UV tape attached to the device wafer, measured with a dynamometer. The inset photograph shows the close-up of peeling of the UV tape from the device wafer. From [paper A].

(a)

p+Si

Device Wafer

Carrier Wafer

Metal contact

Al

Force direction by the dynamometer

UV release tape

ElectRelease

(b)

Page 41: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

27

Thereafter, a constant voltage of +50 V was applied between the device (anode) and the carrier (cathode) wafers as shown in figure 3.5(a). The voltage was switched off when the corresponding current density decreased to 1 mA/cm2. The bond strength was measured again and it was now possible to debond the device wafer from the carrier wafer with a force of only 0.5 N, leaving the ElectRelease™ adhesive on the carrier wafer (figures 3.5(b) and 3.5(c)).

Figure 3.5: (a) Illustration of the evaluation set-up for measurement of bond strength after applying a voltage. (b) Illustration of the debonding procedure after applying a voltage. (c) Photograph showing the debonding of the device wafer when a force is applied to the UV tape attached to the device wafer. From [paper A].

p+Si

Device Wafer

Carrier Wafer

Metal contact

Al

Force direction by the dynamometer

UV release tape

ElectRelease

50 v

p+Si

Device Wafer

Carrier Wafer

Metal contact

ElectRelease

(a) (b)

(c)

Page 42: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

28

3.3.2 Highly doped silicon device wafer

A highly doped silicon wafer was used as the device wafer in order to investigate the ability to debond without a conductive surface coated on the device wafer. A comprehensive description of the wafer bond strength experiment is given in paper A. In this section, shear force measurement on one of the wafer stack before and after voltage application is presented.

Figure 3.6: Illustration of the shear force measurement set-up for shearing of the bonded and released wafer stack, respectively. From [paper A].

Figure 3.6 illustrates a cross-section of the shear test set-up with the mounted wafer stack. The bond strength of the wafer stack was measured before and after applying a voltage using the shear tester. The measured shear force plots of the wafer stack before and after applying the voltage to the stack are shown in figure 3.7. The plots show the measured force on the shear head as a function of the shear head displacement. The shear head displacement represents the distance that the device wafer has moved from its initial position relative to the carrier wafer. Before applying a voltage, the bonded wafer stack could withstand the maximum possible shear force of the shear tester of 490 N. The achieved bond strength of more than 490 N demonstrates once more the high bond strength of the ElectRelease™ adhesive. After a voltage of 50 V was applied to the wafer stack, a second shear test was performed. The device wafer was sheared off at an applied shear force of 163 N.

Shear headShear head

Upper wafer flatUpper wafer flatShear direction

Vacuum fixation

Wafer stack

Lower wafer flat

Fixation recess

Sample mounting stage

Page 43: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

29

Figure 3.7: Shear force measurement for the wafer stack w5w6 before and after the voltage is applied. From [paper A].

3.4 Discussion

For the first time a novel approach for temporary wafer handling of a thin silicon wafer using an electrochemically active adhesive was successfully demonstrated. Two types of device wafers were investigated. In one type of the experiment, a non-conducting device wafer was coated with aluminum layer and in the other type, a highly doped silicon wafer was used as the device wafer. The advantage of using this technique with an aluminum coated device wafer is that the device wafer does not have to be conductive itself, for example a glass wafer can be used instead. In contrast, the advantage of using a highly conductive device wafer is that the post processing step of etching aluminum is avoided. Processing on the device wafer after bonding of the wafer stack is an important issue, which potentially could have an effect on the bond strength of the adhesive and the debonding process. However, the post bonding process, i.e. device wafer thinning by RIE etching, did not show any signs of premature debonding. Damages or cracks in the thin device wafer were not observed after the debonding process concluding that the adhesive withstood the RIE etching of the device wafer.

The presented method could become a breakthrough for several applications. Interactive packaging of IC dies in which, the IC dies can be tested and discarded merely by applying a voltage between the test platforms is an area of application that is gaining interest. In addition, photo-patterning of the adhesive would open up new areas in the field of selective debonding mechanism. In the future, thin wafer handling can also find its use in fabrication of gas sensors in which highly porous working electrodes are fabricated on a thin device wafer that are then transferred by debonding to another wafer containing reference and counter electrodes.

100 101 102

100

200

300

400

500S

hear

For

ce [N

]

Displacement [µm]

w5w6 wafer stackbefore applying the voltage

w5w6 wafer stackafter applying the voltage

max. shear force at 490 N(machine limit)

wafer sheared off at 163 N

Page 44: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

30

Page 45: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

31

Chapter 4

Functional nanoporous material using atomic layer deposition This chapter describes a method to fabricate a functional nanoporous material from a silicon substrate by coating dual layers of platinum (Pt) and aluminum oxide (Al2O3) using atomic layer deposition (ALD). The nucleation delay & film growth properties of Pt and Al2O3 on different types of substrates are discussed. Future prospects and applications of such fabricated functionalized nanoporous materials are presented.

4.1. Introduction

Functional nanoporous materials have been a subject of interest and studied extensively in the last decade [122-126]. In electrochemical sensors, in particular gas sensors, nanoporous materials allow for higher sensitivity. This can be ascribed to the inherent increased surface-to-volume ratio of the nanoporous material leading to an increased line length of the triple-phase point interface [127-132]. Functional nanoporous material can be fabricated by depositing thin layers of desired materials on a nanoporous substrate, e.g. porous alumina. Atomic layer deposition (ALD) is identified as a promising tool to achieve the goal of depositing thin layers of materials on nanoporous susbstrates.

Page 46: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

32

Atomic layer depositon enables to deposit metals and dielectrics on the surface of a substrate with a precise thickness. A schematic illustration of an ALD process is shown in figure 4.1. The process consists of a reaction between the precursor gas (Pulse:A) and reactant gas (Pulse:B) at a given temperature and pressure. During the Pulse: A, the precursor gas flows into the reaction chamber of the ALD and chemisorbs on the surface of the substrate forming a monolayer of molecules. An inert gas such as argon is then purged to remove the excess precursor materials. During the Pulse:B, the reactant gas flows into the reaction chamber and interacts with the monolayer of the precursor molecules. The reactant gas undergoes a reaction with the precursor molecules forming a monolayer of metallic or dielectric material on the surface of the substrate. Finally, an inert gas is purged through the reaction chamber to remove the reaction products and excessive reactant molecules. Thus, one complete ALD cycle results in one monolayer of the material being deposited on the surface of the substrate. By controlling the precursor and reactant gas pressure and time duration, materials can be deposited in narrow crevices and pores that allows for the development of nanoporous materials.

Figure 4.1: Atomic layer depostion process. Pulse A: precursor gas pumped to the ALD chamber. Pulse B reactant gas pumped to the ALD chamber resulting in a reaction product that leaves a monolayer of desried material.

Two methods of ALD are prevalent: thermal ALD and plasma enhanced ALD (PEALD). The thermal ALD process is based on dissociative chemisorption of the precursor by a reactant at the surface of a substrate [133]. On the other hand in PEALD, the surface of the substrate is treated with a plasma during pulse:B cycle in order to oxidize the precursor monolayer as compared to thermal ALD where water or oxygen is often used during the pulse:B. Table 4.1 compares the thermal ALD and PEALD. PEALD is advantageous as compared to thermal ALD due to the increased mass density of the

Precusor Pulse:A

Precursor monolayer

Product

Reactant

Purge

Purge

Pulse:B

Substrate Substrate

Substrate Substrate

Page 47: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

33

deposited material and faster growth per cycle. Hence, the precursor cycle times can be made shorter when PEALD is used. However, most importantly, the conformality of the deposited material is better in thermal ALD due to the formation of smaller grain size. Hence, thermal ALD is useful in depositing materials in narrow corners, crevices and pores. Therefore, in this work, thermal ALD was selected for fabrication of functional nanopores. An extensive study of various precursors for different materials has been carried out by George [134].

Table 4.1: Comparison of thermal and plasma ALD. From [135-137].

Thermal Plasma Mass density of the material

+ ++

Growth per cycle (GPC)

+ ++

Precursor cycle times

Short Shorter

Conformality Better Limited

4.2. Fabrication of functional nanopores

In this section, fabrication of nanoporous material using an anodic aluminum oxide (AAO) substrate is described. In addition, the process of functionalization by depositing two layers of materials using ALD on open ended nanopores is presented.

4.2.1. Method to create open ended nanopores

AAO substrate fabrication: A 525 µm thick single side polished Si wafer with a 400 nm thick SiO2 layer was used as a substrate, on which 1 μm thick aluminium layer was deposited. The aluminium layer was anodized to form a 1.4 µm thick regular pore array with 25 nm diameter and 75 nm pitch (figure 4.2). The anodization was carried out by Synkera Technologies Inc., Logmont CO, USA.

Page 48: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

34

Figure 4.2: SEM image of the cross section of an anodized aluminum oxide susbtrate on a silicon wafer.

Substrate processing: In brief, the back side of the silicon wafer was lithographically developed with 6 mm diameter circular patterns, which was then etched to a depth of approximately 440 μm (figure 4.3(a)). In order to have a supporting grid structure to the AAO layer, a second photolithography step was carried out (figure 4.3(b)). A triangular grid with 20 μm wide and 120 μm long beams was etched from the remaining silicon on the backside wafer cavity (figures 4.3(c) & (d)). The grid structure was necessary to ensure a mechanical stability to the AAO structure. At this point, the wafer was diced to 10 mm × 10 mm chips (figures 4.3(e) & (f)).

Barrier layer etching: The diced chip now consists of three layers: the porous AAO layer, the barrier layer of the AAO layer and the silicon dioxide layer. The silicon dioxide layer below the barrier layer was etched. To create an open ended nanoporous substrate, the barrier layer has to be etched as well. This barrier layer is a thin compact Al2O3 layer at the bottom of the porous AAO. Most commonly used method to remove Al2O3 layer is by using H3PO4 (5 wt%) solution [138]. However, this method would take approximately 50- 90 mins. Etching of the barrier layer should be done with caution since overetching the barrier layer may damage the nanopores. In this work, a recipe was developed to successfully dry etch the Al2O3 barrier layer with an etch gas composition of Cl2 with 20 sccm, BCl3 with 2 sccm and Ar with 20 sccm at 600 W, for approximately 60 s. Thus, an open ended nanoporous AAO membrane suspended on a grid was succesfully

Page 49: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

35

developed (figure 4.3(f)). In the next section, a model and the recipe for functionalization of the nanoporous membrane using ALD are presented.

Figure 4.3: Fabrication flow and SEM images of the AAO substrate at different process steps: a) and b) The AAO film is formed on a Si substrate. Cavities are etched in the substrate from the bottom and the grid is patterned on spray coated photoresist; c) and d) The grid is etched from the backside in the Si until reaching the buried SiO2 layer. The AAO surface is protected with a photoresist layer before dicing. The inset shows the closed end of the AAO pores; e) and f) After dicing, the buried SiO2 and the Al2O3 barrier layer is etched on chip level to open the AAO membrane. The inset shows the pores now open at the bottom. From [paper B].

Page 50: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

36

4.2.2. Dual layer ALD: A predictive approach

Figure 4.4: Schematic illustration of a nanopore during ALD deposition. The spheres represent the precursor gas molecules forming a progressing monolayer until saturation of the pore surface. From [paper B].

Model: Figure 4.4 shows a schematic illustration of a nanopore during ALD deposition in which L indicates the length of the pore and dpore is the diamter of the pore. For a successful ALD deposition in deep nanopores, it is necessary to ensure the saturation of the whole surface, including deep inside the pores with the precursor/reactant molecules. This requires that the exposure times (precursor pulse and the reactant pulse) are sufficiently long. A small pulse duration would result in a partially covered pore wall and if the pulse duration is too long there could be a wastage of precursor/reactant gases [139].

The saturation dose, during the film growth, Ss, is typically 1015 [molecules cm-2] [140-142]. This depends on the material surface, on the pressure applied and on the type of the gas, i.e. the chemisorption of the first few monolayers of gas or vapour molecules depends on weak surface forces. The exposure required for sidewall coverage derived in paper B is given by:

δ S √2πmkT L [Langmuir, L] … (2)

Page 51: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

37

where m is molecular mass of the precursor, k is the Boltzmann constant, T is the temperature in kelvin and p is the precursor partial pressure. This equation is often corrected with a sticking coefficient,α, in the denominator to account for non-ideal saturation of the surface, where α is expressed as the percentage of the surface covered with a monolayer of the precursor. It is also known that, for some materials, there is a delay in the formation of the first monolayer and is called as the nucleation delay. This phenomena can also be corrected by α. This nucleation delay is known to be particularly long for Pt [143], but no exact value of the sticking coefficient has been reported for Pt to date. We note that α was also reported to vary with temperature [144]. The derived model predicts an initial condition for the exposure of the prescursor and reactant gases. The purging gases should ensure that any non-chemisorbed molecules or any reaction products present in the reactor are removed to make the surface suitable for the next reaction and to ensure the self-limiting surface growth process [145, 146].

Recipe: The deposition of Pt results from the reaction of MeCpPtMe3 (precursor) with O2 (reactant). The deposition of Al2O3 is the product of trimethylaluminium (TMA) reacting with H2O. Starting from the exposure value predicted by the theoretical model (see above), the deposition parameters (i.e. exposure and number of cycles) needed to be further tuned to achieve a uniform deposition in these high aspect ratio nanopores (60-80 nm diameter with a length of 1.4 µm). The tuning of the thickness of the deposited layer was achieved by varying the number of cycles. Tables 4.2(a) and (b) give the recipes that were used for the deposition of Pt and Al2O3 in nanopores, respectively. For example, to deposit approximately 10 nm thick platinum on a Al2O3 substrate, 250 cycles were required.

Page 52: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

38

Table 4.2 (a): Recipe parameters for the deposition of Pt in deep pores.

Deep-pore ALD recipe for Pt deposition Process temperature: 250 °C

Precursor source temperature: 35 °C

Cycle steps: Gas: Exposure: (Time × Pressure = Langmuir)

Remarks:

Precursor pulse MeCpPtMe3 1 s By own vapour pressure only (no Ar flow through the source canister)

MeCpPtMe3 1.25 s By own vapour pressure + Ar flow through source canister

Total: 2.25 s × 24 Pa =0.405·106 L

Purge 1 Argon 30 s Reactant pulse Oxygen 1.3 s × 160 Pa =1.56·106 L Purge 2 Argon 30 s Table 4.2 (b): Recipe parameters for the deposition of Al2O3 in deep pores.

Deep-pore ALD recipe for Al2O3 deposition Process temperature: 200 °C

Precursor source temperature: 25 °C

Cycle steps: Gas: Exposure: (Time × Pressure = Langmuir)

Remarks:

Precursor pulse TMA 0.25 s × 160 Pa = 0.3·106 L By own vapour pressure

Purge Argon 10 s Reactant pulse Water vapor 0.7 s × 160 Pa =0.84·106 L Purge Argon 15 s

In order to assess the nucleation delay and film growth properties, deposition of Pt and Al2O3 was also conducted on several planar substrates. It was found that Pt film growth starts after 125 cycles on Al2O3 susbtrate and after 150-160 cycles on SiO2. However, on a bare silicon wafer that was etched with HF and immediately transferred to the reaction chamber, hardly any growth was observed. On the other hand, it was observed that Al2O3 deposition starts immediately on any given surface. These observed differences can be

Page 53: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

39

inferred to different surface adhesion properties. If the substrate is oxidized, the chemisorption properties are enchanced and lead to quick formation of a monolayer of the ALD material. Further explanation of the charaterization process is given in paper B.

To summarize, there are several challenges that should be addresed when developing a recipe for an ALD. One of the main challenge is the estimation of the exposure required for each precursors. In this section, a method to estimate the exposure and number of cycles for deposition of materials in high aspect ratio nanopores was presented.

4.3. Discussion

In recent times, there has been a lack of development of AAO based devices. This could be predominantly due to the bottom-up approach of the technology, i.e. fabrication of a nanostructured template followed by material deposition and finally device development. The top-level is highly complex and is also perhaps very interesting. In this chapter, a successful method to create a wafer-level fabricated nanoporous membrane that is functionalized with two layers of materials using atomic layer deposition was presented. The fabrication of functionalized nanopores at a wafer-level has a vast potential to create novel materials that could be usefull in high density capacitors, fuel cells and electrochemical sensing. In particular, functional nanoporous material with Pt layer could be promising for high sensitivity gas sensing application due to high surface area to volume ratio of the material.

Page 54: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

40

Page 55: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

41

Chapter 5

Wafer-level fabrication of amperometric gas sensors Introduction to asthma and the role of nitric oxide in asthma were presented in chapter 1. In this chapter, the design and the development of a highly sensitive and fast response amperometric NO gas sensor that is potentially applicable for asthma monitoring is described. The sensor consists of three electrodes that act as working, reference and counter electrodes, respectively. The oxidation of nitric oxide gas at the working electrode causes a current flow between the working and the counter electrodes that is proportional to the gas concentration. Furthermore, the developed amperometric sensor was redesigned and successfully tested for hydrogen sulphide gas concentrations that are relevant for bad breath monitoring.

Page 56: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

42

5.1 Introduction to the design of a nitric oxide sensor

The following design parameters were considered when developing the sensor: (1) a highly porous gas diffusion working electrode, (2) ease of fabrication, (3) ease of assembly, (4) ease of integration with a potentiostat and (5) possibility for scaling of the sensor. Based on these design criteria, a novel nitric oxide gas sensor was successfully fabricated and characterized. The design of a three electrode NO sensor that includes working, reference and counter electrodes is shown in figure 5.1. In an attempt to fabricate a high sensitivity sensor, a microporous grid structure was etched in the silicon. This microporous grid was coated with platinum and Nafion™ and then was used as a working electrode. A voltage bias (Vbias) of +0.7 V was applied to the working electrode as compared to the reference electrode using a potentiostat connected to the sensor. An accepted reaction occurring at the anode due to the oxidation of nitric oxide can be written as [147].

2 32 4 3NO H O NO H e

− + −+ → + + ; …………...…………. (3)

Figure 5.1: Schematic of the designed amperometric sensor. The microporous working electrode was fabricated using a 300 µm thick silicon wafer. Silver sputtered on a polycarbonate board was used as the counter and reference electrodes. 5 % wt H2SO4 was used as an electrolyte solution. From [Paper C].

Page 57: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

43

5.2 Fabrication and assembly of the nitric oxide sensor

Among the design criteria listed above, the development of a porous working electrode was of importance. The working electrode of the sensor was fabricated by following a standard MEMS fabrication process. The detailed processing steps can be found in paper C. In summary, a microporous working electrode was first realized by DRIE etching a 300 µm thick silicon wafer. A platinum layer of thickness 10 nm was deposited using ALD on the surface of the microporous grid as described in chapter 4. Thereafter, Nafion™, a semipermeable nanoporous membrane, was dip coated on the microporous grid and dried. This successfully resulted in a highly nanoporous working electrode. To form the counter and reference electrodes, a silver of thickness 500 nm was sputter deposited on a polycarbonate board, which was then demarcated using a milling machine. All the electrodes were assembled using silicone adhesive glue to form an amperometric sensor. A polycarbonate board with meandering channel was realized in order to keep the sensor filled with electrolyte (5 wt % H2SO4) as shown in figure 5.2.

Figure 5.2: Schematic cross section of the assembled amperometric sensor with meandering electrolyte channel. The inset photograph shows an assembled nitric oxide sensor. From [Paper C].

Electrolyte filling port

Meandering electrolyte channel

Electrolyte chamber

Polycarbonate board

Polycarbonate sheet

NafionTM coated working electrode chip

Counter electrode

Reference electrode

Spacer

Working electrode

Reference electrode

Counterelectrode

Page 58: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

44

5.2.1 Measurement set up and instrumentation

A measurement system was set-up for characterizing the amperometric sensor. The schematic diagram of the measurement set up is shown in the figure 5.3. A 200 ppb NO gas concentration was used as a calibration gas for testing the sensor. The calibration gas was diluted with N2 gas to obtain several NO concentrations in order to assess the sensitivity of the sensor. The gases were humidified using a humidifier which consisted of a syringe inserted with a moistened tissue paper. A potentiostat was used to measure the current from the working electrode of the nitric oxide sensor. A Labview™ script was developed to record the experimental parameters, i.e. gas flow, humidity and temperature.

Figure 5.3: Schematic diagram of the measurement set-up used for characterizing the amperometric gas sensor. The data from the flow sensors and the humidity sensor was recorded using a LabVIEW script. From [Paper D].

5.2.2 Characterization

The analytical properties of the sensor such as stability, sensitivity, selectivity, settling time, response to humidity and flow were measured using the measurement set-up described in the previous section. More detailed information about the sensor characterization can be found in paper C. In this section, two important properties of the sensor, namely sensitivity and settling time, together with a model for the settling time will be presented and its relevance to asthma monitoring will be discussed.

ExhaustMixer

Mechanical sealing module

Potentiostat

N2

NO

CO/NH3

W R C

MFC

MFC

Flow sensor

Flow sensor

Humidifier

syringeScrubber

Valve

Valve

Data from connected sensors

Valve

Mixer

Mechanical seali

W R C

Humidifier

syringe

Page 59: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

45

Sensitivity and settling time: In order to assess the sensitivity of the sensor, the output current of the sensor (INO) was measured for different concentrations of NO gas as shown in figure 5.4(a). The sensor was able to detect NO concentrations with sensitivity of 0.045 nA/ppb and in the ppb-level range that corresponds to the concentration range of the NO in asthma monitoring. Jacquinot et al also reported a NO sensor with ppb-level sensitivity. However, to be applicable for hand-held instruments, the large size and discrete components of their sensor were not suitable [148].

Figure 5.4 (a): The distribution of NO sensitivity for five different sensors. One of the sensors with filter is illustrated by the (-.-) straight line. From [Paper C].

Figure 5.4 (b): Response of the sensor when the NO gas was alternately switched between 0 to 200 ppb resulting in a settling time of approximately 8 s. From [Paper C].

The settling time of the sensor was measured by alternately switching the flows of the gases between NO and N2 as shown in figure 5.4(b). The settling time was calculated to be approximately 8 s. In order to predict the sensor settling time in a real measurement condition of asthma monitoring, i.e. with an exhaled breath as the input to the sensor, a sensor model was also developed.

0 20 40 60 80 100 120 140 160 180 2000

2

4

6

8

10

12

14

NO concentration (ppb)

Out

put c

urre

nt, I

NO (n

A)

C 2

C 5

C 5

C 3

C 1

C 4

(with filter)

8 s

Time (s)

Output currentNO FlowN2 Flow

Out

put c

urre

nt, I

NO (n

A)

0 10 20 30 40 50 60

0

2

4

6

8

10

12

14

16

10 % error band

Average final value

Settling time

Page 60: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

46

Sensor model: The transfer function model of the NO sensing system that was used to predict the sensor settling time is shown in figure 5.5. The exhaled breath transfer function was obtained from the NO measurement of exhaled breath reported in [22] as shown in figure 5.6(b). The sensor characteristic transfer function was derived from the experimentally measured response of the sensor (figure 5.4). The response of the sensor was simulated by following the procedure in figure 5.5, with the exhaled breath transfer function as the input to the sensor model. This resulted in a critically damped output response which had a settling time of approximately 7- 8 s (figure 5.7). From this, it was inferred that the sensor could reach a stable output for a given NO concentration within 7-10 s time window. It should be noted that the physiologically relevant NO concentration obtained from the exhaled breath reaches a stable value after approximately 7 s (figure 5.6 (b)) [22, 32]. Therefore, it can be inferred that the experimentally obtained settling time of the NO sensor could be applicable for real time exhaled breath measurements.

Figure 5.5: A transfer function model for NO sensing system. The exhaled breath transfer function serves as the input to the sensor transfer function. The sensor characteristic transfer function allows determining the settling time of the sensor.

Exhaled breath transfer function (sensor input):

Sensor characteristic transfer

function:

Settling time (sensor output)

~ 7-8 s

Page 61: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

47

Figure 5.6: Exhaled NO schematic diagram. (a) The peak concentration of NO occurs when the subject has inhaled through the nose. (b) NO measurement from the exhaled breath. Adapted by permission from Macmillan Publishers Ltd: Immunology and Cell Biology [22], copyright 2001.

Figure 5.7: Sensor output model when exhaled breath is introduced to the sensor. The output from the sensor is found to be stable and lies within the required 10 s duration. It can be inferred that the designed sensor can be used for real time measurements.

5.3 Integrated and miniaturized nitric oxide sensor

In section 5.2, a working prototype of a nitric oxide sensor that measured ppb-level concentration was described. The settling time and the sensitivity were found to be appropriate for the sensor to be applicable in an asthma detecting instrument. However, the sensor lacked a proper design for electrolyte supply which, resulted in its large size (figure 5.2). In addition, the manual assembly of the working, reference and counter electrodes is not well suited for a large scale manufacturing of the sensor. A hand-held gas sensing instrument could be envisioned with a miniaturized and highly sensitive sensor that has fast response time. One of the methods to achieve this would be to fabricate the sensors at a wafer level integrated with liquid electrolyte.

One of the first silicon based nitric oxide amperometric gas sensor with integrated electrodes and electrolyte was reported in [149], a schematic representation is shown in the figure 5.8. It consisted of a gold working and counter electrodes with a silver reference electrode and the electrolyte used was Nafion™ gel. Holes of 2 µm diameter were etched in the silicon using DRIE, on which electrode materials were sputtered to form working, reference and counter electrodes, respectively. Nafion™ gel, which is used as an

a b

Time (s)

FENO(p.p.b)Nasal contamination

Plateau

10 s

0 2 4 6 8 10 12

0

0.2

0.4

0.6

0.8

1

1.2

1.4

Time (seconds)

Am

plitu

de Exhaled breath input

Sensor model

Sensor output

Page 62: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

48

electrolyte, was applied manually on the electrodes. Nitric oxide gas was exposed from the substrate of the device. This design was reported to have a response time of less than 0.5 s. However, the sensor had a detection limit of 5 ppm, which is too high for asthma detection.

Figure 5.8: One of the first amperometric gas sensor fabricated using silicon as the substrate with integrated electrodes and electrolyte. Nafion™ gel was used as the electrolyte. From [149], © 2011 IEEE.

5.3.1 Sensor design and fabrication

The sensor design idea was to integrate all the electrodes on a single chip with a liquid electrolyte packaged inside this chip. To realize such an integrated sensor, a SOI wafer was used for fabrication of the microporous working electrode. The counter and reference electrodes were fabricated on a glass substrate which, was then anodically bonded to the SOI wafer. The cavity formed between the SOI wafer and the glass wafer was used for electrolyte storage. A 3D illustration of the fabricated integrated NO sensor is shown in figure 5.9.

Page 63: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

49

Figure 5.9: A 3D illustration of the amperometric gas sensor showing integrated electrodes. Working, counter and reference electrodes are integrated on SOI and Glass substrates. The cavity formed between the substrates is used for storing electrolyte. The grid structure of the working electrode forms a microporous sensing area. From [Paper D].

The device wafer of the SOI was etched to form a microporous grid structure and the carrier wafer of the SOI was etched to form a cavity. To realize the working electrode platinum was deposited using ALD on the microporous grid structure. The reference and the counter electrode were realized by depositing a 500 nm thick silver layer on the glass wafer. The SOI wafer with working electrode was anodically bonded to the glass wafer containing counter and reference electrodes. Thus, an integrated sensor was developed as shown in figures 5.10(a) and (b). The dimension of the sensor after dicing is approximately 10 × 10 × 1 mm. In order to integrate the electrolyte, the sensor was dip coated with Nafion and then allowed to dry at room temperature followed by filling with 5 %wt H2SO4 electrolyte. A detailed explanation of the fabrication process can be found in paper: D.

Working electrode

Space for electrical contact

Referenceelectrode

Counterelectrode

Page 64: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

50

(a) (b)

Figure 5.10: Final fabricated device. (a) Schematic showing the final assembly of the sensor. (b) Image showing the final fabricated integrated gas sensor device of the dimension 10 × 10 × 1 mm. The microporous electrode foot print is approximately 5 × 5 mm. From [Paper D].

5.3.2 Analytical properties

Sensor characteristics of the integrated sensor were determined by measuring the sensitivity and the response time. The average value of the output current measured at different concentrations is shown in figure 5.11(a). The output current (IO) was found to be linear with the NO gas concentration. From the slope, a sensitivity of 0.04 nA/ppb was calculated, which is in agreement to the sensitivity of the nitric oxide sensor with discrete components.

Counter electrode contact

Workingelectrode

Si Ag

Reference electrode contact

Electrolyte chamber

SiO2Glass PtAg2O

10 mm

Page 65: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

51

(a)

(b)

Figure 5.11 Analytical properties of the sensor. (a) Output current as a function of NO concentration. (b) Response time (t90) of the integrated sensor for 200 ppb NO gas. From [Paper D].

Response time of the sensor was measured in a similar way to the non-integrated sensor by alternately switching NO gas concentration between 0 and 200 ppb. The response time was calculated using t90, of the average final value of the output current which resulted in approximately 12 s as shown in figure 5.11(b).

5.3.3 Operating life time

The operating life time of the sensor is dictated by the volume of the electrolyte present in the sensor. The evaporation of the electrolyte from the sensor was measured to be approximately 2 h when the sensor surface was exposed to the ambient air. In order to further improve the sensor packaging, one of the challenging tasks is to minimize the evaporation of the electrolyte. Several possible methods were investigated including replacing the liquid electrolyte with a non-aqueous electrolyte such as solid polymer electrolyte (Nafion™), ionic liquid (Emim NTF2) and glycerin (100% pure). With ionic liquid and glycerin the ppb-level sensitivity was not achieved. With Nafion as a solid electrolyte, it is well established that the sensor could become humidity dependent, which is undesirable [150]. Currently, commercial sensors employ multiple layers of Teflon™ based membrane to prevent the evaporation of the electrolyte. One of the drawbacks of this method is an

0 20 40 60 80 100 120 140 160 180 2000

1

2

3

4

5

6

7

8

9

NO concentration (ppb)

Out

put c

urre

nt, I

O (n

A)

0 10 20 30 40 50 60 70 80

0

1

2

3

4

5

6

7

8

9

Time (s)

Response time, t90 = 12 s

Out

put c

urre

nt, I

O (n

A)

IOmax

Page 66: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

52

increased response time (> 50 s) of the sensor and hence may not be applicable for real time asthma monitoring.

One of the proposed solutions to prevent the evaporation without sacrificing the response time was to use a mechanical sealing module, described in paper D. Using this module, it was estimated that the sensor would need approximately 30 s of measurement time for one measurement of NO concentration which, would enable approximately 200 measurements until the electrolyte in the sensor evaporates. It could be envisioned that an electrically controlled actuator for the opening and closing of the sealing mechanism can be included for the asthma monitoring application.

5.4 Integrated and miniaturized hydrogen sulphide sensor

The integrated sensor was characterized for hydrogen sulphide (H2S) gas which is a component of the oral breath that causes bad breath. H2S gas is predominantly used as one of the biomarker for bad breath/Halitosis monitoring. The sensor was found to be sensitive to H2S gas in the ppb-level range when the bias voltage between the reference and the working of 1.1 V was applied. Interference to NO and NH3 gases that are commonly found in the oral cavity was experimentally verified and found that the sensor was selective to these gases. Further characterization details can be found in paper E.

5.5 Discussion

Research and development of amperometric gas sensors has spanned over several decades. Amperometric gas sensor has evolved from a two electrodes sensor for sensing oxygen, to a stable three electrode gas sensor that is able to sense several gases. The wafer-level fabricated sensor with integrated electrodes and liquid electrolyte presented in this chapter is a technological leap from a commercially available canister based design to a SOI-glass based design. It is possible to envision a highly sensitive gas sensor that can be manufactured in large volumes. The research presented in this chapter of the thesis is in the direction of an integrated and miniaturized sensor that can applied to a hand held gas monitoring device.

In the near future, a mobile phone carrying a miniaturized gas sensor would allow personalized asthma diagnosis by merely exhaling the breath into the phone. With this, a hand-held health monitoring system that is able to track the medical condition of an individual can be envisioned.

Page 67: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

53

Chapter 6

Conclusions

This thesis has presented research in the field of MEMS and specifically in the area of electrochemical sensors and their method of fabrication. A novel method for temporary handling of a thin device wafer was demonstrated. A successful fabrication process for developing functional nanoporous membranes relevant for electrochemical sensing was described and finally, a brief description of the wafer-level fabricated MEMS based electrochemical sensors that are relevant for medical diagnostics was presented. The following conclusions can be derived from the research presented in this thesis.

• A method for temporary handling of thin wafer using an electrochemically active adhesive was demonstrated. A wafer could be bonded, thinned and debonded without any damage. This method has the potential to facilitate handling of thin wafers with structures such as nanoporous materials or through silicon vias (TSV’s). The metallization process that is required for the nonconductive device wafer opens up new opportunities in handling glass wafers.

• A novel functional nanoporous material with dual layers of materials was successfully designed and fabricated. The fabrication process involved successful deposition of dual layer materials along the side walls of the deep nanopores. Thus fabricated functional nanopores can be potentially applied in electrochemical sensing, fuel cells, as a filter membrane and many other myriads of ubiquitous devices. Furthermore, the fabrication process developed at a wafer-level, demonstrates the ability to manufacture nanoporous membrane in large volumes.

Page 68: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

54

• A MEMS based novel amperometric gas sensor that is sensitive to ppb-level nitric oxide gas concentration was successfully demonstrated. The gas sensor featured a combination of a high surface area microporous electrode and nanostructured Nafion™ coated along the side walls of the micropores. A miniaturized sensor with a dimension of 10 × 10 × 1 mm with integrated electrodes and a liquid electrolyte was designed and developed. A hand-held medical instrument such as asthma monitoring instrument based on the designed sensor can be envisioned. Furthermore, the amperometric sensor was redesigned and successfully tested for ppb-level hydrogen sulphide gas concentrations which are relevant for bad breath monitoring.

To summarize, the electrochemical sensors and fabrication methods presented in this thesis makes a shift in the paradigm of the development of a hand-held monitoring and diagnostic instruments for medical applications. This leads us to conclude that in the future, a comprehensive measurement system such as personalized health monitoring device using mobile phones could be envisioned.

Page 69: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

55

Summary of Appended Papers

Paper A: Temporary wafer bonding and debonding for 3D integration using an electrochemically active polymer adhesive

The use of thin silicon wafers is an enabling technology for 3D integration in the semiconductor industry. However, thin silicon wafers are fragile to handle and reliable solutions are required for thin wafer handling. This paper reports a novel method of bonding and debonding a thin wafer (<50 μm) using an electrochemically active polymer adhesive. In the presented method the carrier wafer is first spin coated with the adhesive and then bonded to the device wafer by applying force and temperature. Debonding of the wafer is realized at room temperature by applying a voltage between the carrier and the device wafer, which substantially reduces the bond strength. The bonding and debonding properties of the adhesive show that temporary wafer bonding using electrochemically active adhesives has the potential to be an attractive approach for temporary wafer bonding for thin wafer handling in 3D integration processes.

Paper B: Pt–Al2O3 dual layer atomic layer deposition coating in high aspect ratio nanopores

Functional nanoporous materials are promising for a number of applications ranging from selective biofiltration to fuel cell electrodes. This work reports the functionalization of nanoporous membranes using atomic layer deposition (ALD). ALD is used to conformally deposit platinum (Pt) and aluminum oxide (Al2O3) on Pt in nanopores to form a metal–insulator stack inside the nanopore. Deposition of these materials inside nanopores allows the addition of extra functionalities to nanoporous materials such as anodic aluminum oxide (AAO) membranes. Conformal deposition of Pt on such materials enables increased performances for electrochemical sensing applications or fuel cell

Page 70: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

56

electrodes. An additional conformal Al2O3 layer on such a Pt film forms a metal–insulator–electrolyte system, enabling field effect control of the nanofluidic properties of the membrane. This opens novel possibilities in electrically controlled biofiltration. In this work, the deposition of these two materials on AAO membranes is investigated theoretically and experimentally. Successful process parameters are proposed for a reliable and cost-effective conformal deposition on high aspect ratio three-dimensional nanostructures. A device consisting of a silicon chip supporting an AAO membrane of 6 mm diameter and 1.3 μm thickness with 80 nm diameter pores is fabricated. The pore diameter is reduced to 40 nm by a conformal deposition of 11 nm Pt and 9 nm Al2O3 using ALD.

Paper C: An amperometric nitric oxide sensor with fast response and ppb-level concentration detection relevant to asthma monitoring

A MEMS-based amperometric nitric oxide (NO) gas sensor is reported in this paper. The sensor is designed to detect NO gas for the purpose of asthma monitoring. The unique property of this sensor lies in the combination of a microporous high-surface area electrode that is coated with Nafion™, together with a liquid electrolyte. The sensor is able to detect gas concentrations of the order of parts- per-billion (ppb) and has a measured NO sensitivity of 0.045 nA/ppb and an operating range between 25 and 65% relative humidity. The settling time of the sensor is measured to 8 s. The selectivity to interfering gases such as ammonia (NH3) and carbon monoxide (CO) was high when placing an activated carbon fiber filter above the sensor. The ppb-level detection capability of this sensor combined with its relatively fast response, high selectivity to CO and NH3 makes the sensor potentially applicable in gas monitoring for asthma detection.

Paper D: A wafer- level liquid cavity integrated amperometric gas sensor with ppb-level nitric oxide gas sensitivity

A miniaturized amperometric nitric oxide (NO) gas sensor based on wafer-level fabrication of electrodes and a liquid electrolyte chamber is reported in this paper. The sensor is able to detect NO gas concentrations of the order of parts-per-billion (ppb) levels and has a measured sensitivity of 0.04 nA/ppb with a response time of approximately 12 s. A sufficiently high selectivity of the sensor to interfering gases such as carbon monoxide (CO) and to ammonia (NH3) makes it potentially relevant for monitoring of asthma. In addition, the sensor was characterized for electrolyte evaporation which indicated a sensor operation lifetime allowing approximately 200 measurements.

Page 71: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

57

Paper E: An amperometric hydrogen sulphide sensor applicable for bad breath monitoring.

A MEMS-based amperometric H2S Sensor relevant for bad breath detection is reported in this paper. The miniaturized wafer level fabricated three electrode amperometric H2S sensor is designed to detect Hydrogen Sulphide (H2S) gas of the order of parts-per-billion level relevant for bad breath measurement. The sensor has a measured sensitivity of 0.65 nA/ppb and a response time of approximately 21 s. The sensor was selective to NO and NH3 gases that are normally present in the oral cavity of an adult.

Page 72: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

58

Page 73: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

59

Acknowledgements I express my gratitude to Niclas Roxhed for the guidance, tolerance and immense

knowledge in the field of MEMS fabrication, and for allowing me to think and act freely. I would like to thank Göran for giving me an opportunity and space to work in the exciting field of MEMS. Thanks for creating an exceptionally innovative, multicultural and transparent research environment. The department is truly an innovation hub!

Thank you for the financial support for the work of this thesis: Swedish Research Council and the Swedish Innovation Agency, the European Research Council (ERC) through the Advanced Grant No: 267528 and Aerocrine AB, producer of handheld diagnostic tools for breath monitoring.

The good thing about a bad story is that it comes to an end and the bad thing about a good story is that it also comes to an end. This journey is also something like a story and would not have been possible without the help of my colleagues. First of all, I would like to thank my co-authors, with whom I had great opportunity of learning. Gaspard, Simon Leijonmarck, Stephan Shröder, Mikael Antelius, Wouter and Frank, you were truly amazing!

This thesis would not have been complete without mentioning the fantastic environment for the research and fun. Here are few tit bits that I would like to share. Carlos, for being a quiet room mate and for his ever dying plants | Mikeal Karlsson’s adventure travel to Varanasi | Tommy: for your humor angle in every situation | Staffan: for an elaborate explanation of how to make ice creams and solar cells | Umer: for your flawless Hindi | pour les connexion francaise:Valentin | Always-in-bright-spirits:Laila, Xiamo Zhou aka Chianty | Floria for giving a thoughtful lesson on pasta, that penne & spaghetti should not be mixed | For opportunity to play badminton: Mina, Bernhard, Miku, Chianti, Federico, Mikhael Asiatici.

To my colleagues for innumerous adventures, Joachim for a wonderful experience on the hands-on-MEMS course | Hans, Kristinn for an optical view of things | Niklas Sandström: for the KT on milling machine | Carl Fredrik Carlborg | Fredrik-knows-everything- Forsberg | Jonas | Gaby | Simon B| Henrik Frid| for organizing several events: Fritzi | Alexander | Reza for the ever ‘Hi Hithesh’| Xuge| Alora and Xiaojing. And to the new phds and postdocs in the department.

Erika and Ulrika, for the help in all the niches and glitches of the administration works. Kjell Noren and Mikeal Berquist for being maverick in building electronic set ups.

Page 74: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

60

To my ex-colleagues, Martin L | unforgettable sailing adventure: Mikeal sterner | Nutapong S| the-serial- Henrik-entrepreneur | Zargham Ba| Farizah | Stephan Braun| Björn samel| Sjoerd hassl and Tomas frisk.

To the people in the clean room Reza Nikpars, Magnus Lindberg and Per Erik, in keeping the clean room running.

I also thank Prof Panduranga Vittal and Prof Navakant Bhat for bringing me into the MEMS field and introducing to semiconductor processing.

Honorable mentions: To restaurang Q: for restoring my faith in cooking | To my Nokia TV: for having least number of channels | To the department coffee machine: that grows in complexity periodically | For the Tuesday-breakfast-only-meetings | To my laptop: for being the slowest laptop ever: so that I could ‘only’ work! | To my apartment: for giving me a beautiful view | To my land lady: for almost never showing up! | To my friends back home: Tanky bois (Mukesh, Gautham, Brian, Shiva, Manoj and Ashwin): you are awesome!.

This thesis would not have been possible without the blessings of my mother and father, for being patient and keeping faith in me. To my brother Santhosh, for keeping my research spirits high | To my youngest brother Abhishek who is still in search of an original idea!| To my beloved wife Melina for being calm, thoughtful and showing me that there is much more to life outside MST and finally to our little bundle of joy, Sohan.

Thank you all!

Hithesh K Gatty

Page 75: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

61

References

[1] H. C. Nathanson, W. E. Newell, R. A. Wickstrom, and J. R. Davis Jr, "The resonant gate transistor," Electron Devices, IEEE Transactions on, vol. 14, pp. 117-133, 1967.

[2] G. N. Stemme, "A monolithic gas flow sensor with polyimide as thermal insulator," Electron Devices, IEEE Transactions on, vol. 33, pp. 1470-1474, 1986.

[3] G. Stemme, "A CMOS integrated silicon gas-flow sensor with pulse-modulated output," Sensors and Actuators, vol. 14, pp. 293-303, 1988.

[4] E. Stemme and G. Stemme, "A balanced resonant pressure sensor," Sensors and Actuators A: Physical, vol. 21, pp. 336-341, 1990.

[5] P. Enoksson, G. Stemme, and E. Stemme, "A silicon resonant sensor structure for Coriolis mass-flow measurements," Microelectromechanical Systems, Journal of, vol. 6, pp. 119-125, 1997.

[6] P. F. Van Kessel, L. J. Hornbeck, R. E. Meier, and M. R. Douglass, "A MEMS-based projection display," Proceedings of the IEEE, vol. 86, pp. 1687-1704, 1998.

[7] N. Roxhed, "A fully integrated microneedle-based transdermal drug delivery system," 2007.

Page 76: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

62

[8] P. S. Dittrich and A. Manz, "Lab-on-a-chip: microfluidics in drug discovery," Nature Reviews Drug Discovery, vol. 5, pp. 210-218, 2006.

[9] W. Wang and S. A. Soper, Bio-MEMS: technologies and applications: CRC Press, 2006.

[10] W. Noell, P.-A. Clerc, L. Dellmann, B. Guldimann, H.-P. Herzig, O. Manzardo, et al., "Applications of SOI-based optical MEMS," Selected Topics in Quantum Electronics, IEEE Journal of, vol. 8, pp. 148-154, 2002.

[11] K. B. Gylfason, C. F. Carlborg, A. Kazmierczak, F. Dortu, L. Vivien, C. A. Barrios, et al., "On-chip temperature compensation in an integrated slot-waveguide ring resonator refractive index sensor array," Optics express, vol. 18, pp. 3226-3237, 2010.

[12] E. Verpoorte and N. F. de Rooij, "Microfluidics meets MEMS," Proceedings of the IEEE, vol. 91, pp. 930-953, 2003.

[13] H. Andersson, W. Van Der Wijngaart, P. Nilsson, P. Enoksson, and G. Stemme, "A valve-less diffuser micropump for microfluidic analytical systems," Sensors and Actuators B: Chemical, vol. 72, pp. 259-265, 2001.

[14] G. M. Rebeiz, RF MEMS: theory, design, and technology: John Wiley & Sons, 2004.

[15] J. Oberhammer and G. Stemme, "Design and fabrication aspects of an S-shaped film actuator based DC to RF MEMS switch," Microelectromechanical Systems, Journal of, vol. 13, pp. 421-428, 2004.

[16] C.-C. Liu, "Applications of microfabrication techniques in electrochemical sensor development," Applied biochemistry and biotechnology, vol. 41, pp. 99-107, 1993.

[17] M. Esashi, "Wafer level packaging of MEMS," Journal of Micromechanics and Microengineering, vol. 18, p. 073001, 2008.

[18] S. Zimmermann, D. Fienbork, A. W. Flounders, and D. Liepmann, "In-device enzyme immobilization: wafer-level fabrication of an integrated

Page 77: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

63

glucose sensor," Sensors and Actuators B: Chemical, vol. 99, pp. 163-173, 2004.

[19] N. M. plus, "Turning discovery into health- Asthma," vol. 6, 2011.

[20] P. Prevention, "Expert Panel Report 3 (EPR-3): Guidelines for the Diagnosis and Management of Asthma-Summary Report 2007," The Journal of allergy and clinical immunology, vol. 120, p. S94, 2007.

[21] F. L. M. Ricciardolo, P. J. Sterk, B. Gaston, and G. Folkerts, "Nitric Oxide in Health and Disease of the Respiratory System," Physiological Reviews, vol. 84, pp. 731-765, 2004.

[22] D. H. Yates, "Role of exhaled nitric oxide in asthma," Immunology and cell Biology, vol. 79, pp. 178-190, 2001.

[23] R. Knox, "Grass pollen, thunderstorms and asthma," Clinical & Experimental Allergy, vol. 23, pp. 354-359, 1993.

[24] J. Schwartz, D. Slater, T. V. Larson, W. E. Pierson, and J. Q. Koenig, "Particulate air pollution and hospital emergency room visits for asthma in Seattle," American review of respiratory disease, vol. 147, pp. 826-831, 1993.

[25] R. M. J. Palmer, A. G. Ferrige, and S. Moncada, "Nitric oxide release accounts for the biological activity of endothelium-derived relaxing factor," Nature, vol. 327, pp. 524-526, 1987.

[26] P. Barnes, "NO or no NO in asthma?," Thorax, vol. 51, pp. 218-220, 1996.

[27] A. Malinovschi, "Nitric oxide exchange in central and pheripheral airways," Ph.D. dissertation, Department of Medical Cell Biology Uppsala Universitet, 2008.

[28] A. D. Smith and D. R. Taylor, "Is exhaled nitric oxide measurement a useful clinical test in asthma?," Current opinion in allergy and clinical immunology, vol. 5, pp. 49-56, 2005.

Page 78: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

64

[29] D. Taylor, M. Pijnenburg, A. Smith, and J. Jongste, "Exhaled nitric oxide measurements: clinical application and interpretation," Thorax, vol. 61, pp. 817-827, 2006.

[30] R. A. Dweik, P. B. Boggs, S. C. Erzurum, C. G. Irvin, M. W. Leigh, J. O. Lundberg, et al., "An official ATS clinical practice guideline: interpretation of exhaled nitric oxide levels (FENO) for clinical applications," American journal of respiratory and critical care medicine, vol. 184, 2012.

[31] J. M. Spergel, M. I. Fogg, and A. Bokszczanin-Knosala, "Correlation of exhaled nitric oxide, spirometry and asthma symptoms," Journal of Asthma, vol. 42, pp. 879-883, 2005.

[32] A. T. S. Documents, "ATS/ERS recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide, 2005," in American journal of respiratory and critical care medicine vol. 171, ed, 2005, p. 912.

[33] Veryday. (2010). A child using Niox Mino. Available: http://www.mynewsdesk.com/se/veryday/images/child-using-niox-mino-32051

[34] Alphasense, "Technical Specification: NO B4 Nitric oxide sensor, 4 electrode," Alphasense, Ed., ed, 2014.

[35] Aerocrine. (2014). About niox mino. Available: http://www.niox.com/en/about-niox-mino/NIOX-MINO/

[36] E. Magori, K. Hiltawsky, M. Fleischer, E. Simon, R. Pohle, O. von Sicard, et al., "Fractional exhaled nitric oxide measurement with a handheld device," Journal of breath research, vol. 5, p. 027104, 2011.

[37] C. Chang, G. Maduraiveeran, J. Xu, G. Hunter, and P. Dutta, "Design, fabrication, and testing of MEMS-based miniaturized potentiometric nitric oxide sensors," Sensors and Actuators B: Chemical, vol. 204, pp. 183-189, 2014.

Page 79: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

65

[38] S. P. Mondal, P. K. Dutta, G. Hunter, B. Ward, D. Laskowski, and R. Dweik, "Development of high sensitivity potentiometric NO x sensor and its application to breath analysis," Sensors and Actuators B: Chemical, vol. 158, pp. 292-298, 2011.

[39] G. Eranna, B. Joshi, D. Runthala, and R. Gupta, "Oxide materials for development of integrated gas sensors—a comprehensive review," Critical Reviews in Solid State and Materials Sciences, vol. 29, pp. 111-188, 2004.

[40] C. Bur, P. Reimann, M. Andersson, A. Schütze, and A. L. Spetz, "Increasing the selectivity of Pt-gate SiC field effect gas sensors by dynamic temperature modulation," Sensors Journal, IEEE, vol. 12, pp. 1906-1913, 2012.

[41] R. Frodl and T. Tille, "A high-precision NDIR gas sensor for automotive applications," Sensors Journal, IEEE, vol. 6, pp. 1697-1705, 2006.

[42] P. Krebs and A. Grisel, "A low power integrated catalytic gas sensor," Sensors and Actuators B: Chemical, vol. 13, pp. 155-158, 1993.

[43] O. S. Wolfbeis, "Materials for fluorescence-based optical chemical sensors," Journal of Materials Chemistry, vol. 15, pp. 2657-2669, 2005.

[44] J. Cheeke and Z. Wang, "Acoustic wave gas sensors," Sensors and Actuators B: Chemical, vol. 59, pp. 146-153, 1999.

[45] J. R. Stetter and J. Li, "Amperometric gas sensors a review," Chemical reviews, vol. 108, pp. 352-366, 2008.

[46] L. Xiong and R. G. Compton, "Amperometric Gas detection: A Review," Int. J. Electrochem. Sci, vol. 9, pp. 7152-7181, 2014.

[47] L. C. J. Qlark, "Monitor and control of blood and tissue oxygen tensions," ASAIO Journal, vol. 2, pp. 41-48, 1956.

[48] H. Bay, K. Blurton, H. Lieb, and H. Oswin, "Electrochemical measurements of blood alcohol levels," Nature, vol. 240, pp. 52-53, 1972.

Page 80: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

66

[49] K. F. Blurton and J. R. Stetter, "Sensitive electrochemical detector for gas chromatography," Journal of Chromatography A, vol. 155, pp. 35-45, 1978.

[50] M. Rashid, T.-S. Jun, Y. Jung, and Y. S. Kim, "Bimetallic Core-Shell Ag@ Pt Nanoparticle-Decorated MWNT Electrodes for Amperometric H< sub> 2</sub> Sensors and Direct Methanol Fuel Cells," Sensors and Actuators B: Chemical, 2014.

[51] S. Rajabzadeh, G. H. Rounaghi, M. H. Arbab-Zavar, and N. Ashraf, "Development of a dimethyl disulfide electrochemical sensor based on electrodeposited reduced graphene oxide-chitosan modified glassy carbon electrode," Electrochimica Acta, vol. 135, pp. 543-549, 2014.

[52] Y. Wang, B. Song, J. Xu, and S. Hu, "An amperometric sensor for nitric oxide based on a glassy carbon electrode modified with graphene, Nafion, and electrodeposited gold nanoparticles," Microchimica Acta, pp. 1-8, 2014.

[53] G. Korotcenkov, Handbook of Gas Sensor Materials: Springer, 2013.

[54] M. T. Carter, J. R. Stetter, M. W. Findlay, and V. Patel, "Amperometric Gas Sensors with Ionic Liquid Electrolytes," ECS Transactions, vol. 58, pp. 7-18, 2014.

[55] I. G. Zacharia and W. M. Deen, "Diffusivity and solubility of nitric oxide in water and saline," Annals of biomedical engineering, vol. 33, pp. 214-222, 2005.

[56] C. H. Hamann, "Electrochemistry. 2nd," Completely Revised and Updated Edition, New York, 2007.

[57] A. Schmid, "The Gas Diffusion Electrodes," Enke-Verlag, Stuttgart, 1923.

[58] N. Furuya, "A new method of making a gas diffusion electrode," Journal of Solid State Electrochemistry, vol. 8, pp. 48-50, 2003.

[59] C. Tran, X.-Q. Yang, and D. Qu, "Investigation of the gas-diffusion-electrode used as lithium/air cathode in non-aqueous electrolyte and the

Page 81: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

67

importance of carbon material porosity," Journal of Power Sources, vol. 195, pp. 2057-2063, 2010.

[60] F. Bidault, D. Brett, P. Middleton, and N. Brandon, "Review of gas diffusion cathodes for alkaline fuel cells," Journal of Power Sources, vol. 187, pp. 39-48, 2009.

[61] L. Cindrella, A. Kannan, J. Lin, K. Saminathan, Y. Ho, C. Lin, et al., "Gas diffusion layer for proton exchange membrane fuel cells—A review," Journal of Power Sources, vol. 194, pp. 146-160, 2009.

[62] W. Cha, Y.-C. Tung, M. E. Meyerhoff, and S. Takayama, "Patterned electrode-based amperometric gas sensor for direct nitric oxide detection within microfluidic devices," Analytical chemistry, vol. 82, pp. 3300-3305, 2010.

[63] J.-S. Do, K.-J. Wu, and M.-L. Tsai, "Amperometric NO gas sensor in the presence of diffusion barrier: selectivity, mass transfer of NO and effect of temperature," Sensors and Actuators B: Chemical, vol. 86, pp. 98-105, 2002.

[64] I. R. Davies and X. Zhang, "Nitric oxide selective electrodes," Methods in enzymology, vol. 436, pp. 63-95, 2008.

[65] X. Zhang, J. Lin, L. Cardoso, M. Broderick, and V. Darley‐Usmar, "A Novel Microchip Nitric Oxide Sensor with sub‐nM Detection Limit," Electroanalysis, vol. 14, pp. 697-703, 2002.

[66] F. Bedioui and N. Villeneuve, "Electrochemical nitric oxide sensors for biological samples–principle, selected examples and applications," Electroanalysis, vol. 15, pp. 5-18, 2003.

[67] A. Ciszewski and G. Milczarek, "Electrochemical detection of nitric oxide using polymer modified electrodes," Talanta, vol. 61, pp. 11-26, 2003.

[68] M. C. Frost, S. M. Rudich, H. Zhang, M. A. Maraschio, and M. E. Meyerhoff, "In vivo biocompatibility and analytical performance of intravascular amperometric oxygen sensors prepared with improved nitric

Page 82: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

68

oxide-releasing silicone rubber coating," Analytical chemistry, vol. 74, pp. 5942-5947, 2002.

[69] F. Pariente, J. Alonso, and H. Abruna, "Chemically modified electrode for the selective and sensitive determination of nitric oxide (NO) in vitro and in biological systems," Journal of Electroanalytical Chemistry, vol. 379, pp. 191-197, 1994.

[70] Y. Xian, W. Sun, J. Xue, M. Luo, and L. Jin, "Iridium oxide and palladium modified nitric oxide microsensor," Analytica chimica acta, vol. 381, pp. 191-196, 1999.

[71] S. Wang and X. Lin, "Electrodeposition of Pt–Fe (III) nanoparticle on glassy carbon electrode for electrochemical nitric oxide sensor," Electrochimica acta, vol. 50, pp. 2887-2891, 2005.

[72] D. Zheng, X. Liu, D. Zhou, and S. Hu, "Sensing of nitric oxide using a glassy carbon electrode modified with an electrocatalytic film composed of dihexadecyl hydrogen phosphate, platinum nanoparticles, and acetylene black," Microchimica Acta, vol. 176, pp. 49-55, 2012.

[73] Z. Cao, W. J. Buttner, and J. R. Stetter, "The properties and applications of amperometric gas sensors," Electroanalysis, vol. 4, pp. 253-266, 1992.

[74] M. C. Buzzeo, C. Hardacre, and R. G. Compton, "Use of room temperature ionic liquids in gas sensor design," Analytical chemistry, vol. 76, pp. 4583-4588, 2004.

[75] T. B. Scheffler and A. A. Schneider, "Potentiostat circuit for electrochemical cells," U.S. Patent No. 5,466,356. , 14 Nov 1995.

[76] T. Instrument, "LMP91000 Sensor AFE system," ed, March 2013.

[77] M. J. Tierney and H. O. L. Kim, "Electrochemical gas sensor with extremely fast response times," Analytical Chemistry, vol. 65, pp. 3435-3440, 1993.

Page 83: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

69

[78] P. R. Warburton, M. P. Pagano, R. Hoover, M. Logman, K. Crytzer, and Y. J. Warburton, "Amperometric gas sensor response times," Analytical chemistry, vol. 70, pp. 998-1006, 1998.

[79] W. Jeffry, H. John, G. Michael, and C. Stuart, "Evaluation of Electrically Disbonding Adhesive Properties for use as Separation Systems," in Structures, Structural Dynamics, and Materials Conference, 2003.

[80] G. Jie, P. Zhibo, C. Qiang, and Z. Li-Rong, "Interactive packaging solutions based on RFID technology and Controlled Delamination Material," in IEEE International Conference on RFID, 2010, pp. 158-165.

[81] D. Haydon, "ElectRelease – electrically disbonding epoxy adhesive," Assembly Automation, vol. 22, pp. 326-329, 2002.

[82] Exonera. (2011). ElectRelease H23. Available: http://www.exonera.com/products.html

[83] C. Jeenjitkaew, Z. Luklinska, and F. Guild, "Morphology and Surface Chemistry of Kissing Bonds in Adhesive Joints Produced by Using ElectRelease™ Adhesive," The Journal of Adhesion, vol. 87, pp. 291-312, 2011/04/29 2011.

[84] ElectRelease E3 Data Sheet Rev1 (TTT. EIC Laboratory, 2007, p.^pp. Pages.

[85] C.-O. Danielsson, "Controlled Delamination Materials-Using Electrochemistry to Break Adhesive Joints in the Packaging Industry," in ECS Meeting Abstracts, 2007, pp. 891-891.

[86] S. Leijonmarck, A. Cornell, C.-O. Danielsson, T. Åkermark, B. D. Brandner, and G. Lindbergh, "Electrolytically assisted debonding of adhesives: An experimental investigation," International Journal of Adhesion and Adhesives, vol. 32, pp. 39-45, 2012.

[87] S. Leijonmarck, A. Cornell, C.-O. Danielsson, and G. Lindbergh, "Electrochemical Characterization of Electrically Induced Adhesive Debonding," Journal of The Electrochemical Society, vol. 158, pp. 109-114, October 1, 2011 2011.

Page 84: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

70

[88] H. K. Gatty, F. Niklaus, G. Stemme, and N. Roxhed, "Temporary wafer bonding and debonding by an electrochemically active polymer adhesive for 3D integration," in IEEE International Conference on MEMS, 2013, pp. 381-384.

[89] P. Garrou, "TSV based memory going to volume production: the era of 3DIC finally begins," in i-micronews, ed, 2014.

[90] epworks. (2014). Global roadmap. Available: http://www.epworks.co.kr/html/roadmap.html

[91] M. Santarini, "Stacked & Loaded: Xilinx SSI, 28-Gbps I/O Yield Amazing FPGAs," Xcell J, vol. 74, pp. 8-13, 2011.

[92] X. D. FPGAs. (2011) IFTLE 73: Xilinx shows 2.5D Virtex 7 at IMAPS 2011. Solid State Technology.

[93] F. Niklaus, G. Stemme, J. Q. Lu, and R. J. Gutmann, "Adhesive wafer bonding," Journal of Applied Physics, vol. 99, pp. 031101-28, 2006.

[94] G. Klink and B. Hillerich, "Wafer bonding with an adhesive coating," in Proc. SPIE 3514, Micromachined Devices and Components IV, 50, 1998, pp. 50-61.

[95] F. Niklaus, H. Andersson, P. Enoksson, and G. Stemme, "Low temperature full wafer adhesive bonding of structured wafers," Sensors and Actuators, A: Physical, vol. 92, pp. 235-241, 2001.

[96] F. Niklaus, P. Enoksson, E. Kälvesten, and G. Stemme, "Low-temperature full wafer adhesive bonding," Journal of Micromechanics and Microengineering, vol. 11, p. 100, 2001.

[97] S. van der Green, M. Rosmeulen, P. Jansen, K. Baert, and L. Deferm, "CMOS compatible wafer scale adhesive bonding for circuit transfer," in International Conference on Solid State Sensors and Actuators, TRANSDUCERS '97 1997, pp. 629-632 vol.1.

Page 85: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

71

[98] S. Pargfrieder, J. Burggraf, D. Burgstaller, M. Privett, A. Jouve, D. Henry, et al., "3D integration with TSV: temporary bonding and debonding," Solid State Technology, vol. 52, pp. 38-+, 2009.

[99] K. Zoschke, M. Wegner, M. Wilke, Ju, x, N. rgensen, et al., "Evaluation of thin wafer processing using a temporary wafer handling system as key technology for 3D system integration," in IEEE Electronic Components and Technology Conference, 2010, pp. 1385-1392.

[100] J. A. Sharpe, M. B. Jordan, S. L. Burkett, and M. E. Barkey, "Analyzing the behavior and shear strength of common adhesives used in temporary wafer bonding," in IEEE Electronic Components and Technology Conference, 2013, pp. 94-100.

[101] M. Reiche and G. Wagner, "Wafer thinning techniques for ultra-thin wafers," Advanced Packaging, vol. 12, pp. 29-30, 2003.

[102] M. Feil, C. Alder, G. Klink, M. König, C. Landesberger, S. Scherbaum, et al., "Ultra thin ICs and MEMS elements: techniques for wafer thinning, stress-free separation, assembly and interconnection," Microsystem technologies, vol. 9, pp. 176-182, 2003.

[103] L. Mauer, H. Itzkowitz, and J. Tadei, "Process control of wet etching for silicon wafer thinning," Solid State Equipment Corporation, 2010.

[104] S. Zhou, C. Liu, X. Wang, X. Luo, and S. Liu, "Integrated process for silicon wafer thinning," in Electronic Components and Technology Conference (ECTC), 2011 IEEE 61st, 2011, pp. 1811-1814.

[105] A. Jourdain, T. Buisson, A. Phommahaxay, A. Redolfi, S. Thangaraju, Y. Travaly, et al., "Integration of TSVs, wafer thinning and backside passivation on full 300mm CMOS wafers for 3D applications," in Electronic Components and Technology Conference (ECTC), 2011 IEEE 61st, 2011, pp. 1122-1125.

[106] S. Pargfrieder, P. Kettner, M. Privett, and J. Ting, "Temporary Bonding and DeBonding Enabling TSV Formation and 3D Integration for Ultra-thin Wafers," in Proc. Electron. Packag. Technol. Conf., 10th, 2008, pp. 1301-1305.

Page 86: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

72

[107] S. Pillalamarri, R. Puligadda, C. Brubaker, M. Wimplinger, and S. Pargfrieder, "High-Temperature Spin-On Adhesives for Temporary Wafer Bonding," Journal of microelectronics and electronic packaging, vol. 4, p. 7, 2007.

[108] R. Puligadda, S. Pillalamarri, W. Hong, C. Brubaker, M. Wimplinger, and S. Pargfrieder, "High-performance temporary adhesives for wafer bonding applications," in MRS Proceedings, 970, 2006.

[109] K. Tamura, K. Nakada, N. Taneichi, P. Andry, J. Knickerbocker, and C. Rosenthal, "Novel adhesive development for CMOS-compatible thin wafer handling," in IEEE Electronic Components and Technology Conference, 2010, pp. 1239-1244.

[110] J. Hermanowski, "Thin wafer handling - Study of temporary wafer bonding materials and processes," in IEEE International Conference on 3D Systems Integration., 2009, pp. 1-5.

[111] A. Phommahaxay, A. Jourdain, G. Verbinnen, T. Woitke, P. Bisson, M. Gabriel, et al., "Ultrathin wafer handling in 3D Stacked IC manufacturing combining a novel ZoneBONDTM temporary bonding process with room temperature peel debonding," in IEEE International Conference on 3D Systems Integration 2012, pp. 1-4.

[112] R. Srinivasan and V. Mayne-Banton, "Self-developing photoetching of poly(ethylene terephthalate) films by far-ultraviolet excimer laser radiation," Applied Physics Letters, vol. 41, pp. 576-578, 1982.

[113] K. Zoschke, T. Fischer, M. Topper, T. Fritzsch, O. Ehrmann, T. Itabashi, et al., "Polyimide based temporary wafer bonding technology for high temperature compliant TSV backside processing and thin device handling," in IEEE Electronic Components and Technology Conference, 2012, pp. 1054-1061.

[114] T. Uhrmann and R. Delmdahl, "Laser debonding enables advanced thin wafer processing," Solid State Technology, vol. 56, pp. 18-22, 2013.

[115] B. Dang, P. Andry, C. Tsang, J. Maria, R. Polastre, R. Trzcinski, et al., "CMOS compatible thin wafer processing using temporary mechanical

Page 87: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

73

wafer, adhesive and laser release of thin chips/wafers for 3D integration," in IEEE Electronic Components and Technology Conference, 2010, pp. 1393-1398.

[116] R. Guerre, U. Drechsler, D. Jubin, and M. Despont, "Selective Transfer Technology for Microdevice Distribution," Journal of Microelectromechanical Systems, vol. 17, pp. 157-165, 2008.

[117] R. Guerre, U. Drechsler, D. Jubin, and M. Despont, "Low-cost AFM cantilever manufacturing technology," Journal of Micromechanics and Microengineering, vol. 18, p. 115013, 2008.

[118] C. Landesberger, S. Scherbaum, and K. Bock, "Carrier techniques for thin wafer processing," in CS Mantech Conference, 2007, pp. 33-36.

[119] P. Tavernier and D. Clarke, "Mechanics of laser-assisted debonding of films," Journal of Applied Physics, vol. 89, pp. 1527-1536, 2001.

[120] T. Lippert, "Laser application of polymers," in Polymers and Light, ed: Springer, 2004, pp. 51-246.

[121] A. Concord, "Debonding technologies for adhesive bonded structures," Universitätsbibliothek, 2012.

[122] S. Polarz and B. Smarsly, "Nanoporous Materials," Journal of Nanoscience and Nanotechnology, vol. 2, pp. 581-612, 2002.

[123] A. Yamaguchi, F. Uejo, T. Yoda, T. Uchida, Y. Tanamura, T. Yamashita, et al., "Self-assembly of a silica-surfactant nanocomposite in a porous alumina membrane," Nature Materials, vol. 3, pp. 337-341, MAY 2004 2004.

[124] G. E. Fryxell, S. V. Mattigod, Y. Lin, H. Wu, S. Fiskum, K. Parker, et al., "Design and synthesis of self-assembled monolayers on mesoporous supports (SAMMS): The importance of ligand posture in functional nanomaterials," Journal of Materials Chemistry, vol. 17, pp. 2863-2874, 2007.

Page 88: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

74

[125] C. Sanchez, C. Boissière, D. Grosso, C. Laberty, and L. Nicole, "Design, synthesis, and properties of inorganic and hybrid thin films having periodically organized nanoporosity," Chemistry of Materials, vol. 20, pp. 682-737, 2008.

[126] A. Stein, Z. Wang, and M. A. Fierke, "Functionalization of porous carbon materials with designed pore architecture," Advanced Materials, vol. 21, pp. 265-293, 2009.

[127] A. Evans, A. Bieberle-Hutter, J. Rupp, and L. Gauckler, "Review on microfabricated micro-solid oxide fuel cell membranes," Journal of Power Sources, vol. 194, pp. 119-129, OCT 20 2009 2009.

[128] A. Liu, H. Geng, C. Xu, and H. Qiu, "A three-dimensional hierarchical nanoporous PdCu alloy for enhanced electrocatalysis and biosensing," Analytica Chimica Acta, vol. 703, pp. 172-178, 2011.

[129] C. Xu, Y. Liu, F. Su, A. Liu, and H. Qiu, "Nanoporous PtAg and PtCu alloys with hollow ligaments for enhanced electrocatalysis and glucose biosensing," Biosensors and Bioelectronics, vol. 27, pp. 160-166, 2011.

[130] H. Razmi and R. Mohammad-Rezaei, "Preparation of tungsten oxide nanoporous thin film at carbon ceramic electrode for electrocatalytic applications," Electrochimica Acta, vol. 56, pp. 7220-7223, 2011.

[131] Q. Li, S. Cui, and X. Yan, "Electrocatalytic oxidation of glucose on nanoporous gold membranes," Journal of Solid State Electrochemistry, pp. 1-6, 2011.

[132] D. D. La, C. K. Kim, T. S. Jun, Y. Jung, G. H. Seong, J. Choo, et al., "Pt nanoparticle-supported multiwall carbon nanotube electrodes for amperometric hydrogen detection," Sensors and Actuators, B: Chemical, vol. 155, pp. 191-198, 2011.

[133] T. Aaltonen, A. Rahtu, M. Ritala, and M. Leskela, "Reaction mechanism studies on atomic layer deposition of ruthenium and platinum," Electrochemical and Solid State Letters, vol. 6, pp. C130-C133, SEP 2003 2003.

Page 89: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

75

[134] S. George, "Atomic Layer Deposition: An Overview," Chemical Reviews, vol. 110, pp. 111-131, JAN 2010 2010.

[135] C. Detavernier, J. Dendooven, D. Deduytsche, and J. Musschoot, "Thermal versus plasma-enhanced ALD: growth kinetics and conformality," ECS Transactions, vol. 16, pp. 239-246, 2008.

[136] J. Van Hemmen, S. Heil, J. Klootwijk, F. Roozeboom, C. Hodson, M. Van de Sanden, et al., "Plasma and Thermal ALD of Al2O3 in a Commercial 200 mm ALD Reactor," Journal of The Electrochemical Society, vol. 154, pp. G165-G169, 2007.

[137] H. Profijt, S. Potts, M. Van De Sanden, and W. Kessels, "Plasma-assisted atomic layer deposition: Basics, opportunities, and challenges," Journal of Vacuum Science & Technology A, vol. 29, p. 050801, 2011.

[138] M. Lillo and D. Losic, "Pore opening detection for controlled dissolution of barrier oxide layer and fabrication of nanoporous alumina with through-hole morphology," Journal of Membrane Science, vol. 327, pp. 11-17, 2009.

[139] M. Ritala, M. Kemell, M. Lautala, A. Niskanen, M. Leskela, and S. Lindfors, "Rapid coating of through-porous substrates by atomic layer deposition," Chemical Vapor Deposition, vol. 12, pp. 655-658, NOV 2006 2006.

[140] R. G. Gordon, D. Hausmann, and E. Kim, "A kinetic model for step coverage by atomic layer deposition in narrow holes or trenches," … Vapor Deposition, 2003.

[141] J. King, A. Wittstock, J. Biener, S. Kucheyev, Y. Wang, T. Baumann, et al., "Ultralow loading Pt nanocatalysts prepared by atomic layer deposition on carbon aerogels," Nano Letters, vol. 8, pp. 2405-2409, AUG 2008 2008.

[142] J. F. O’Hanlon, "A User’s Guide to Vacuum Technology A JOHN WILEY & SONS," INC., PUBLICATION, 2003.

Page 90: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

76

[143] Q. Fang, "Growth of platinum films by atomic layer deposition (ALD)," White Paper, Oxford Instruments.

[144] M. Rose, J. Bartha, and I. Endler, "Temperature dependence of the sticking coefficient in atomic layer deposition," Applied Surface Science, vol. 256, pp. 3778-3782, APR 1 2010 2010.

[145] T. Seidel, J. Dalton, Z. Karim, J. Lindner, M. Daulesberg, and W. Zhang, "Advances in Atomic Level Deposition technologies," in ICSICT-2006: 2006 8th International Conference on Solid-State and Integrated Circuit Technology, Proceedings, 2007, pp. 436-439.

[146] M. Gutsche, H. Seidl, T. Hecht, S. Kudelka, and U. Schroeder, "Atomic layer deposition for advanced DRAM applications," Future Fab Int, vol. 14, p. 1009, 2003.

[147] R. C. Weast, Handbook of Chemistry and Physics 53rd Edition: Chemical Rubber Pub., 1972.

[148] P. Jacquinot, A. W. Hodgson, P. Hauser, B. Müller, and B. Wehrli, "Amperometric detection of gaseous ethanol and acetaldehyde at low concentrations on an Au–Nafion electrode," Analyst, vol. 124, pp. 871-876, 1999.

[149] F. Maseeh, M. J. Tierney, W. S. Chu, J. Joseph, H. O. L. Kim, and T. Otagawa, "A novel silicon micron amperometric gas sensor," in Solid-State Sensors and Actuators, 1991. Digest of Technical Papers, TRANSDUCERS '91., 1991 International Conference on, 1991, pp. 359-362.

[150] E. Aleksandrova, R. Hiesgen, K. A. Friedrich, and E. Roduner, "Electrochemical atomic force microscopy study of proton conductivity in a Nafion membrane," Physical Chemistry Chemical Physics, vol. 9, pp. 2735-2743, 2007.

Page 91: MEMS-based electrochemical gas sensors and wafer-level methods851003/FULLTEXT01.pdf · This thesis describes novel microelectromechanical system (MEMS) based electrochemical gas sensors

77

Paper Reprints