Manual
AriosoSystem
Power(TOC)
Manual
AriosoSystem
Power(TOC)
Inspection Certificate
•Item : Water Purification System
•Model :
•Serial No. :
•Date :
•Origin : Republic of Korea
Authorized Signature
We hereby certify that goods has been inspected before
shipment and found in good order.
General Description 5
8
10
11
12
14
16
22
24
26
33
35
36
37
38
40
42
43
44
System Structure
TOC meter Structure
Installation Condition
Installation Method Ⅰ
Installation Method Ⅱ (Diagram)
Display
Operating
Volume Dispensing ∙ Dispensing Time
Filter Exchange and Display / Default Value
Error Message / Default Value
TOC Meter Error & UV Lamp
Adapter
Using Method of Level Sensor
Using Method of Water Softner
Pre-Treatment system
Accessories
Parts for a instrument
Maintenance
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Index
Manual
General Description
1. Arioso Power (TOC) System
1) Feed Water
3) STD Accessories
2) Included filters
4) Optional Accessories
5) Applications
1
The Power System is the unique water purification system combining
2 systems-RO System & UP System. It can produce simultaneously from the city/ground water.
And, it is a perfect instrument in space budget and efficiency.
: Tap Water / Ground Water (Water Softner required)
: Manual, Power Cord, Level Sensor, Hand Lever
- Pretreatment : Pre filter + 10 ” A/C Filter- RO System : RO Pack
- UP System : ① UP Pack
② 0.2 Capsule filter
- TOC System (option)
① 254/185 UV Lamp
② MW 5000 UF filter
③ Printer
④ Foot-pedal s/w
⑤ Pressure regulater
⑥ Wall-mounting bracket
⑦ Water Tank (20L / 30L / 40L / 60L)
⑧ Frame
① General laboratory tests
② Preparations of stock solution
③ Washing & rinsing for glasswares
④ Analytical Instruments
⑤ Microorganism analysis
⑥ Cell Culture
⑦ DNA/RNA Tests
m
Arioso Power (TOC) System 5
6) Flow Schematic
∙ RO System
∙ UP System
① Pretreatment system
Ⓐ UP pack
④⑥ Pressure sensor
Ⓓ Resistivity sensor
② Temp sensor
Ⓑ 254/186 UV lamp
⑦ RO pack
Ⓔ 0.2 final filter (PF final filter)
③⑧ Conductivity sensor
Ⓒ MW5000 UF filter
⑨ Water sensor
2. Features
∙ Certified CE, ISO 9001, ISO 14001, NRTL
∙ Awarded Korean excellent products
∙ Auto calibraion for water sensor
∙ Extensible point of use
∙ Frost alarm message
∙ 3, 5 steps water level sensor
∙ Filter exchange indicator : Beep & charactor display
∙ Languages : English, Korean, etc
∙ Self-diagnosis functions
Drain RO Product UP Product
① ② ③ ④ ⑤ ⑥ ⑦ ⑧ Ⓐ
Ⓑ
Ⓒ Ⓓ
ⒺTank
⑨
Ts
m
Human C o r p o r a t i o n6
∙ Feed water requirement ; TOC < 50 ppb
3. Specifications
Water Quality
-TOC (ppb) 5~10 5~10 0~5 0~5 5~10 0~5
-Endotoxin (Eu/㎖ ) NA < 0.125 NA < 0.125 < 0.001 < 0.001
-Bacteria (cfu/㎖ ) < 1 < 1 < 1 < 1 < 1 < 1
-Particles ( > 0.22 ㎛/㎖ ) < 1 < 1 < 1 < 1 < 1 < 1
-RNases (ng/㎖) - < 0.01
< 4-DNases (pg/㎖) -DNases (pg/ )
-
-
-
-
-
-
-
< 0.01
< 4
-
-
Dimensions (WxDxH) 320 x 490 x 475 mm
Power 230V / 110V, 50/60Hz, 60W
Standard Accessories Pretreatment System, ROmembrane, UP Pack, Level Sensor, Manual, Power Cord
Optional Accessories· Printer · Foot-Pedal switch · Pretreatment system · Low pressure valve, Wall-mounting bracket· Water tank(20L, 30L, 40L, 60L) · Vent filter for PE water tank · Frame
Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power Arioso PowerⅠⅠ +70 Ⅱ Ⅱ+70 Ⅲ Ⅱ(TOC)
Arioso PowerⅢ(TOC)
Product flow rate(RO, UP) Max. 15 L/Hr Max. RO : 15 L/Hr
UP : 70 L/Hr Max. 25 L/Hr Max. RO : 25 L/Hr UP : 70 L/Hr Max. 25 L/Hr Max. 35 L/HrMax. 35 L/Hr
Display & Controller Large color 70 x 52 touch screen ←
Water quality display
Ambient temp : -10 to 100 ←
Feed water temp : 0 to 100℃ ←
Feed water : 0 to 999㎲/㎝ ←
Pure water (RO) : 0.2 to 250.0㎲/㎝ ←
Ultra Pure (UP) : Up to 18.3㏁·㎝ ←
←←←←←←
- TOC : 0 to 200 ppb
Pressure display Digital Pressure Sensor (0~10kg/㎠) ① Feed water pressure ② RO membrane pressure
Self-check function · Various valves · Auto sensor calibration(RO, UP)· System operation status · Self-diagnosis
Cleaning function · Auto regular flushing · Auto regular cleaning · Auto cleaning lower than setting value
System function· Filter replacement alarm & display (A/C, RO/UP pack, UF, UV lamp, 0.2㎛CF, etc)· Volumetric time setting (RO/UP) · Language(English, Korean, etc)· Auto/Manual Pressure S/W · Extensible point of use
TypeScholar Scholar-UV
Bio Integrate_ PF _ PF
0.2㎛ final filter STD STD STD STD
Pyrogen free final filter STD STD
254/185 UV lamp STD STD STD
MW 5000 UF filter STD STD
< 0.01 < 0.01
< 4 < 4
Arioso Power (TOC) System 7
System Structure
1. Pretreatment System (HMC-PT-3)
2. Main System / Front
2
Air Vent Button(Red Button)
Display
0.2 Capsule filter
Locking system
UP/RESET switch
10 ” Pre Filter10 ” A/C Filter
Output[From this port to main instrument
using Blue 6mm(OD) tubing]Input
[From tap water to this portusing Blue 10mm(OD) tubing]
m
Human C o r p o r a t i o n8
4. Main System / Back Side
3. Main System / Front-inside
① Input Port [6mm(OD)]
Connect to output of Pre-treatment system
using Blue 6mm(OD) tubing
② Pure Water Product Port
Connect to Water tank using White 6mm(OD) tubing
③ Drain Port
Connect to Sink using Blue 6mm(OD) tubing
④ TOC Input (Option)
Connect to Input of TOC Meter using transparent 4mm(OD) tubing
⑤ Printer Port (Option) - 4 pin
Connect to Printer
⑥ TOC Meter Power Port (Option) - 5 pin
Connect to Power of TOC meter
⑦ Level Sensor - 2, 3, 6 pin
After connect level sensor, put it Water Tank
⑧ Power switch
Connect to Power
RO Pack
254/185 nm UV Lamp
MW 5000 UF Filter
UP Pack
(TOC Version)
(General version without TOC Meter)
①
①
②
②
③
③
④ ⑤
⑤
⑥⑦
⑦
⑧
⑧
Arioso Power (TOC) System 9
Product / Drain valve
Control Water volume with Ⓐ valve ,
and then fix Ⓐ valve by turning Ⓑ valve clockwise.
(※Do not fasten too strongly.) (‘Auto drain valve’is installed in Arioso Power I.)
Ⓐ valve
Ⓑ valve
TOC Meter Structure (Option)3
1. Front
2. Back
① Water Flow Controller
Control Water Flow of TOC Meter (15-20㎖/min : approx.
30-35drops / 10 sec) Set the ball in‘② Flow Meter’
between RED lines.
② Flow Meter for TOC
Hole
(Fit for the back-left rubber feet of main system.)
※ If there is air inside Flow Meter, Ball could go down.
RED LineBall
∙ TOC Meter Power Port
Connect to Main system ⑥ TOC Meter Power Port
(Refer to page 09)
∙ TOC Meter Input
Connect to Main system ④ TOC Input using transparent
4mm(OD) tubing
∙ TOC Meter Output
Connect to Sink using transparent 4mm(OD) tubing
※ Make sure that tubing is not twisted and set upside.
Human C o r p o r a t i o n10
Installation Condition 4
1. Environmental Conditions
1) Indoor installation
2) Altitude should be up to 2000m
3) Ambient temperature : 5℃ ~ 40℃
(If the temperature is lower the 5℃, the system can be frozen up)
4) Maximum relative humidity : 80%
5) Power : 230 ± 10%V, 50/60㎐ ± 1㎐, 60W
6) Pollution degree 2
7) Installation Categories Ⅱ
8) The instrument should be placed on steady workbench that is free
from a strong vibration source.
9) The location room should be free from a strong electro magnetic
interference source or harmful or corrosive gases.
CAUTION!! - Refer to accompanying documents
CAUTION!! - ELECTRIC SHOCK
“ To avoid electrical shock, do not open the cover.
Refer servicing to qualified personnel only.”
Arioso Power (TOC) System 11
Installation Method Ⅰ 5
1. Pretreatment System (Refer to Page 8)
2. Main System (Refer to Page 9)
3. Main System (Refer to Page 8)
1) Unscrew the locking system at under of Main body and Open it.
2) Lock the door again when you are using the instrument.
2) Before connect the main system, wash 10 ” A/C filter by opening Tap Water.
[Drain minimum 60 sec.]
※
1) Connect the Input of Pretreatment System into the Tap Water. [Blue 10mm(OD) tubing]
3) Connect the Output of Pretreatment into the Input Port in back side of main system.
[Blue 6mm(OD) tubing]
Human C o r p o r a t i o n12
① Input Port [6mm(OD)]
Connect to output of Pre-treatment system
using Blue 6mm(OD) tubing
② Pure Water Product Port
Connect to Water tank using White 6mm(OD) tubing
③ Drain Port
Connect to Sink using Blue 6mm(OD) tubing
④ TOC Input (Option)
Connect to Input of TOC Meter using transparent 4mm(OD) tubing
⑤ Printer Port (Option) - 4 pin
Connect to Printer
⑥ TOC Meter Power Port (Option) - 5 pin
Connect to Power of TOC meter
⑦ Level Sensor - 2, 3, 6 pin
After connect level sensor, put it Water Tank
⑧ Power switch
Connect to Power
5. TOC Meter Installation Method
4. Connect Front Pack (Refer to Page 9)
1) Connect RO pack Line
2) Connect UP Pack line
3) Connect 0.2 Capsule Filter
4) Connect 254/185 UV Lamp (Option)
5) Connect MW 5000UF filter (Option)
① Put TOC Meter back-right of Main system.
② Insert the back-left rubber feet of main system into the hole of TOC meter.
m
Arioso Power (TOC) System 13
Installation Method Ⅱ (Diagram)6
1. Pre-treatment System Diagram
Air Vent RED button
① Open the tap after connecting lines.
② Drain water for approx. 60sec to wash A/C filter.
③ Connect to input port of Main system after close the tap. (Page 9)
10 ” Pre Filter
Blue 10mm(OD) tubing
10 ” A/C Filter
Human C o r p o r a t i o n14
2. Back side of Main System
TOC Meter
Printer
•Level Sensor Type
⑴ 2 step level sensor ⑵ 3 step level sensor ⑶ 5 step level sensor
s i n k
①
① Blue 6mm(OD) input tubing
② White 6mm(OD) (RO product) tubing
③ Blue 6mm(OD) (RO Drain) tubing
④ Transparent 4mm(OD) (TOC meter Input) tubing
⑤ Printer cable (4pin)
⑥ TOC meter Power (5pin)
⑦ Level Sensor
⑧ Power Cable port & Switch
TOC drain
② ③ ④ ⑤ ⑥ ⑦ ⑧
Arioso Power (TOC) System 15
PretreatmentOutput
Display 7
1. Main Display
2. Operation Display
MODE
R
U
SET
UP/RESET
UP/RESET
A 1.2 bar/F M 1.2 bar/F
: Setting Operating Condition
: Produce / Stop RO water Key
: Produce / Stop UP water Key
: Checking Instrument
or
: ① Press this, Produce / Stop UP water
② Press for 3sec, back to Main display
1)
① A (Auto) : Only works when the Tap water pressure is in 1~5bar(kg/㎠).
It doesn’t work when the Tap water pressure is over 5bar(kg/㎠) or
blow than1bar(kg/㎠). (Default Value : 1~5bar, Modifiable at )
② M (Manual) : Only works regardless of Tap water pressure (Remarks : It works even though there is no supplied water!)
③ 1.2 bar/F : Present Feed Water pressure
④ Setting Method :
MODE
SET ⇨NEXT Auto Manual→ → or
Human C o r p o r a t i o n16
5.2 bar/RO - bar/RO2)
① Pressure of RO Membrane (Default Value : 3 to 7 bar)
Regulate Pressure of RO Membrane, the ratio of RO product / Drain volume
by controlling
② Analog Pressure guage is equipped when it does not have Pressure sensor of RO membrane.
Product / Drain Valve
3)
① RO Product : 0.2 /cm
② Feed water Conductivity : 148 /cm (Default Value : < 350 /cm)
RO148
/cm
feed
0.2
4)
① UP Product : 18.3 cm at 25℃
② TOC : 2 ppb (Option)
UP2 ppb
cm
TOC
18.3
① 15.0℃ : Temperature of Feed Water
② : Beep ON
③ : Beep OFF
④ Setting method :
or5) 15.0 ℃/F 15.0 ℃/F
orSET Beep ON Beep OFF→
① 21.0℃ Amb : The present Ambient Temperature (Default Value : > 5℃)
② Caution: The system can be frozen up if the temperature is lower than 5℃!
6) 21.0 ℃ Amb
∙ Water Tank ( 2step ,3step ,5step )
7)
or (without Pressure sensor)
Arioso Power (TOC) System 17
s
ss s
Ω∙M
Ω∙M
Model Auto Drain Valve생 산 비 율
Arioso Power Ⅰ
Arioso Power Ⅱ
Arioso Power Ⅲ
Product
4
4
5
Drain
6No
6 Built-in
Auto Drain Valve ;
It automatically adjust the ratio of product and drain water
volume.5
(Unit ; /cm)
1)
2)
4)
5)
6)
7)
3)
3. Mode Display
※ For changing Default value Touch display → change the value with key.or
RO / UP
Value Setting
AC / UV
Time Setting
0.2CF / MW5000
Time Setting
Feed
Water µs/cm
Pressure
Value Setting
Running time
RO / UP / AC / UV LAMP
RO / UP
Volume Dispensing
RO / UP
Dispensing Time
: Setting replacement period of RO / UP Pack
① 1st Value : 1st Alert message RO/UP pack replacement
② 2nd Value : 2nd Alarm message RO/UP pack replacement
Setting replacement period of AC Filter, UV Lamp (Unit ; hour)
Set the or first.
Replace period for 0.2 Capsule Filter (or PF Filter) and MW5000 UF Filter
※ In the case of Scholar - PF, Scholar - UV - PF Type Replace 0.2CF →PF Filter!
Setting Max value of Feed water Conductivity.
① FD : Setting Feed Water Pressure (Unit ; kg/㎠, bar)
② RO : Setting RO membrane Pressure (Unit ; kg/㎠, bar)
Display total accumulated running time of
RO Pack, UP Pack, AC Filter, UV Lamp 0.2 CF, UF Filter. (Unit ; hour)
∙ Setting dispensing Volume (Vloume : 0.01 to 99.0 L)
∙ Setting dispensing time
(Option)
or
SET →※
m
m
Human C o r p o r a t i o n18
s
Volume Dispensing Dispensing Time
TOC
UV Time
Temp
Amb / Feed water
TOC
Error No
Language
Running time of UV Lamp for TOC Meter (Unit ; hour)
① Amb : Setting Minimum value of Ambient temperature to prevent freezing (Unit ; ℃)
② Feed Water : Setting Maximum value of Feed Water temperature.
Display Error No. for TOC Meter.
Printing Function
Language selecting function
① : Operating with set value in Mode function
② : Operating regardless default values.
8)
9)
10)
11)
12)
13) SettingValue
On Off
On
Off
(※Pressure Function is maintained.)
Arioso Power (TOC) System 19
① : Check RO Product
② : Check Auto Cleaning
③ : Check TOC Function
④ : Check UP Product
⑤ : Check UV Lamp
4. SET Display
1)
2)
3)
4)
Parts Check
Functional Check RO Product
TOC
UP Product
UV Lamp
Auto Cleaning
① SOL . 1 : No1. valve Operating check
② SOL . 2 : No2. valve Operating check
③ SOL . 3 : No3. valve Operating check
④ SOL . 4 : No4. valve Operating check
⑤ SOL . 5 : No5. valve Operating check
Calibration in case RO sensor (Conductivity Sensor),
UP sensor(Resistivity sensor) has Error
Choose one between above 2 functions.
① Volume Dispensing (Option) : Dispensing RO and UP after putting dispensing volume you want.
(Volume range : 0.01 to 99.0 L)
② Dispensing Time : Setting time for Dispensing RO and UP
※ Please Inquire about Calibration to Head Office.
or
Calibration
Sensor
Volume Dispensing Dispensing Time
Human C o r p o r a t i o n20
Choose one between above 2 functions.
① Volume Dispensing (Option) : Dispensing RO and UP after putting dispensing volume you want.
(Volume range : 0.01 to 99.0 L)
② Dispensing Time : Setting time for Dispensing RO and UP
① : Only works when Feed Water pressure is in 1~5 bar (kg/㎠)
② : Works regardless of feed water pressure (Even works without supplied water)
① Yes : Going back to factory setting values.
② No : Remaining the current setting values.
① Yes : Auto cleaning UP system for 120sec. after Cleaning RO system for 120sec. every 1 hour in standby display.
② No : Cancel upper function.
Calibration to get high accuracy Once equipped the Flow meter
Ex) Calibrate after serveral tests. Set 100ml as dispensing volume
Test the above procedure again till it is calibrated.
produce more than 100ml
Set the above valve as lower than 100
produce less than 100ml
Set the above valve as higher than 100
Range : 0 ~ 200
Range : 0 ~ 200
Using Time Reset
BEEP ON
AUTO
Default Value
Auto Flushing
Volume Calibration
BEEP OFF
Manual
AUTO
Manual
5)
6)
7)
8)
9)
10)
∙ Reset running time of each filters. (Please Inquire it to Head Office)
∙ Setting Alarm or touch sound.
or
or
Arioso Power (TOC) System 21
Caution ; The instrument can be defective because it woks even though there is no supplied water.
Operation8
1)
3)
2)
Power S/W ON
To produce UP,
Touch : RO water Product Key
Touch: UP product key
Main Display
Operation Display
R
U
RO
UP
① Self Test : Auto checking Feed water pressure, storage tank etc.,
② Auto Cleaning : Auto Cleaning inside of instrument. ※ Extension : When RO conductivity isn’t good enough, it automatically extends cleaning for max. 600 sec.
③ RO product : Touch RO to stop producing RO water.
touch UP or right side s/w ON
To stop producing UP, re-touch them.
UP/RESET
Touch Part
Pressure : 3 to 7 bar(Default Valve)
Human C o r p o r a t i o n22
4) Waiting Display
MODE
TOC
S/B
: Switch over MODE display
(Option) : Measuring TOC value, when TOC Meter is equipped.
: ∙ Switch over Standby display.
∙ To release function, re-touch the key
- Function -
※ During Standby, Auto cleaning circulated.
Once UP water (inside of instrument) value is lower than 10 cm
① RO System : Auto cleaning for 120 sec. and Extend max 600sec.
② UP System : Auto cleaning for 60 sec. and Extend max 600sec.
①
③
②
MΩ∙
Arioso Power (TOC) System 23
※Standby function could shorten the filter replacement period!
Volume Dispensing, Dispensing Time9
※ While you use this function,“Filter Exchange message”is not Displayed
1)
※
Volume Dispensing setting method.
ex) set RO product 13.25L
RO key touch (change the color)
select
After touching set 10
set 3
set 0.2
set 0.05
After touching
After touching
After touching
Calibrate volume, when there is difference between
product volume and set volume.
Operating
Volume Calibration put value
( refer to page 21 )
→ → →SET ⇨NEXT
Human C o r p o r a t i o n24
2)
※
Dispensing Time setting method.
Product volume results from Setting time
- Calculate as much as set dispensing volume.
- Setting method
(Product volume can vary depending on water temp, filter condition)
∙ Water temperature : 7℃
∙ Input Conductivity : 185 /㎝
∙ Model : Arioso Power Ⅰ (Scholar Type)
∙
→ Set time →
key
Main Display
Main Display
Dispensing TimeSET
R U
MODE
Main
MainRO / UP
Dispensing Time
→ → → →
→→
or
→
15 sec.
80 sec.
47 sec.
120 sec.
30 sec.
90 sec.
63 sec.
2 min. 26 sec.
3 min. 26 sec.
6 min. 7 sec.
9 min. 15 sec.
5 min. 5 sec.
8 min. 10 sec.
7 min. 15 sec.
10 min. 15 sec.
50 ㎖
250 ㎖
150 ㎖
400 ㎖
100 ㎖
300 ㎖
200 ㎖
500 ㎖
700 ㎖
1.2 L
1.8 L
1 L
1.6 L
1.4 L
2.0 L
Dispensing Time Dispensing TimeProduct volume Product volume
Arioso Power (TOC) System 25
s
Filter replacement and Display 10
1. Pretreatment System
1) Prefilter
2) A/C filter
1) Replacement Period
① Approx. 20 ~ 40days (depends on using time feed water condition.)
② When the filter becomes to brown color (Refer to sticker posted on the housing.)
3) Replacement Method
① Same as Prefilter replacement Method
② After replace a filter, touch screen for 5sec and you can hear“beep”.
Default Value : 400hr
1) Replacement Period
2) Display
∙ Alarm turned on
∙ Message blinking (Message is steadily displayed when it is used over 400 hours.)
After replace it, touch screen for 5sec until you hear the “beep”
2) Replacement Method
① Make the system stop
② Close the Tap Water.
③ Remove Air by pressing the Air Removing Button (Red one) in the housing.
④ Replace PreFilter
※ can disappear by touch screen,
but it appears when you turn on the system again without replacing filter.
Replace A/C Filter
Human C o r p o r a t i o n26
3) RO Filter (Pack)
1) Replacement Period
2) Display
Default Value : ① 1st value : 20 /㎝
② 2nd value : 30 /㎝
∙ Alarm turned on
∙ Message blinking
∙ Alarm turned on
∙ System stop
3) Replacement Method
① Make the system stop
② Close the Tap Water
③ Depressurize the system by touching‘RO’to dispense
water for 20 sec. and stop it.
④ Shut power OFF to the unit and disconnect the electrical power to the unit.
⑤ Open the front door and replace RO pack.
(Connect the Inlet, drain and product tubings.)
⑥ Power it on and open the tap water to operate it again.
•1st value •2nd value
Arioso Power (TOC) System 27
ss
4) UP Pack
2) Display
1) Replacement Period
Default Value : ① 1st value : 13.0 ㎝
② 2nd value : 10.0 ㎝
∙ Alarm turned on
∙ Message blinking
∙ Alarm turned on
∙ Stop producing UP water and be recycled.
① Inlet Line
② Inlet Line
③ Outlet Line
④ Product Line
•2 Line : Scholar, Bio
•4 Line : ∙ Scholar - UV ∙ Integrate
4) Structure of UP Pack
•1st value •2nd value
④③ ② ①
Human C o r p o r a t i o n28
Ω∙M
Ω∙M
3) Replacement Method
① Make the system stop
② Close the Tap Water
③ Depressurize the system by touching‘UP’to dispense
water for 20 sec. and stop it.
④ Shut power OFF to the unit and disconnect the electrical power to the unit.
⑤ Open the front door and replace UP pack.
⑥ Power it on and open the tap water to operate it again.
2) Display
∙ Alarm turned on
∙ Message blinking
∙ Alarm turned on
∙ massage is steadily displayed
3) Replacement Method
∙ Replace 0.2 CF (or PF filter)
•1st value •2nd value
5) 0.2㎛ Capsule Filter or PF (Pyrogen Free) Filter
1) Replacement Period
Default Value : ① 1st value : 280hr
② 2nd value : 300hr
※ Replace‘PF Filter’in case of Scholar-PF, Scholar-UV-PF Type.
m
Arioso Power (TOC) System 29
touch screen for 5sec and you can hear“beep”
※ message can disappear by touching it.
But it appears when you turn on the system again without replacing filter.
Replace 0.2 CF
6) MW5000 UF Filter
1) Replacement Period
2) Display
Default Value : ① 1st value : 280hr
② 2nd value : 300hr
∙ Alaram turned on
∙ Message blinking
∙ Alaram turned on
∙ Message is steadily displayed
3) Replacement Method
•1st value •2nd value
① Make the system stop
② Close the Tap Water ( Remove the Pressure inside)
③ Replace MW 5000 filter
④
Human C o r p o r a t i o n30
touch screen for 5sec and you can hear“beep”
※ message can disappear by touching it.
But it appears when you turn on the system again without replacing filter.
Replace UF Filter
7) UV Lamp
1) Replacement Period
2) Display
Default Value : ① 1st value : 1800hr
② 2nd value : 2000hr
∙ Alarm turned on
∙ Message blinking
∙ Alarm turned on
∙ Message is steadily displayed
•1st value •2nd value
4) Structure and Caution
※ After replacing MW5000 UF Filter or UP Pack, press the‘UP’button and then open the White Cap of MW 5000 UF filter for removing air inside. If the water starts to come out from the vent, close the White Cap. You can use the instrument now.
※ This air removal is also needed for the first use after installation of Bio/Integrate type system.
MW 5000 UF Filter
White Cap
Blue Cap
Applicable for Bio / Integrate Type
Remove the air in the MW 5000 UF Filter!! Or the product volume and purity can be decreased.
Arioso Power (TOC) System 31
3) Replacement Method
4) UV lamp structure
① Close the Tap Water
② Turn the instrument off and disconnect power cord from unit.
③ Replace UV lamp (Remove the Pressure inside)
④ Make sure the UV lamp is again connected to the UV lamp adapter.
⑤
① Outlet Line : Yellow tubing
② Inlet Line : White tubing
③ Adapter
※ Warning!
This unit is equipped with an ultraviolet lamp. Ultraviolet radiation is harmful to the eyes and skin. Do not attempt to observe the lamp directly. Some water will drain from the UV lamp when it is disconnected.
①
②
③
Human C o r p o r a t i o n32
touch screen for 5sec and you can hear“beep”
※ message can disappear by touch screen,
but it appears when you turn on the system again without replacing filter.
Replace UV Lamp
Error Message11
1)
3)
2)
∙ Alarm turned on
∙ Alarm turned on
∙ Alarm turned on
① Adjust the feed water pressure to less than 5kg/㎠
② Install pressure regulator (Ask for it to Head-Office).
Decrease the temperature of feed water. Otherwise, filter can be damaged.
① Use booster pump
② Change feed water pressure mode to Manual mode
Feed Pressure High
Feed Temp Exceeding
Feed Pressure Low
•Solution
•Solution
•Solution
Default Value : 5kg/㎠
Default Value : 35.0℃
Default Value : 1kg/㎠
SET⇨
nextAuto or Manual→ →
※ At manual mode, Instrument can be operated even without Water supply.
Arioso Power (TOC) System 33
4)
5)
6)
∙ Alarm turned on
∙ Alarm turned on
① Install the Pretreatment system.
② If the feed water is ground water, Water softner should be installed.
Freezing Caution
① When Pressure of RO membrane is out of Default value, Purity and Product volume can be influenced by it.
② Adjust Default Value by turning Drain Valve (Page 09)
③ Ratio of Product and Drain volume
Increase the ambient temperature to prevent Freezing.
or
Feed Purity Exceeding
Check drain valve
•Solution
•Solution
•Solution
Default Value : 350 /㎝
Default Value : RO membrane pressure - 3~7㎏/㎠
Default Value : 5℃
Human C o r p o r a t i o n34
s
Model Auto Drain Valve생 산 비 율
Arioso Power Ⅰ
Arioso Power Ⅱ
Arioso Power Ⅲ
Product
4
4
5
Drain
6No
6 Built-in
Auto Drain Valve ;
It automatically adjust the ratio of product and drain water
volume.5
TOC Meter12
1) Error message
2) TOC meter UV lamp replacement method.
No
1
4
7
2
5
8
3
6
9
Error Message
Error 02
Error 08
Error 06
Error 18
Change UV Lamp
Error 03
Error 10
Error 07
System Check Failed
Error Code Error Reason Remark
02 Quality of Sample water is out of the Specification (Below than 1 ㎝)
07 Temp. of sample water is out of the specification (lower than 5℃)
18 UV Lamp doesn’t illuminate. (Disconnection of UV lamp)
01 Replace UV Lamp (After using 4000hr)
06 Temp. of Sample water is over the Specification (higher than 35℃)
10 Value of TOC is out of the range of Specification (Over 200ppb)
03 Temp. Sensor 1 (Error on Thermistor)
08 Temp. Sensor 2 (Error on Thermistor)
19 Message error, measurement is not possible.
TOC Drain Tube
TOC Inlet Tube
TOC UV LampUV Lamp.Slowly lifeit up.
Arioso Power (TOC) System 35
Ω∙M
Adapter13
1. Power Supply
2. UV Lamp Adapter
Human C o r p o r a t i o n36
How to use Level Sensor14
•Level Sensor Type
① 2 step level sensor
② 3 step level sensor
③ 5 step level sensor
When you open the Tank cap,
make sure that level sensor doesn’t
get twisted.
You should Install Level
sensor vertically.
Arioso Power (TOC) System 37
Water Softner for Pure Water System15
1. Name of No.
① Cap
② Main Body
③ Output Line or Drain in Regeneration
④ Input Line (Tap Line)
⑤ Input Valve
Human C o r p o r a t i o n38
①
②
③
⑤
④
Installation
Sink Table
Water Softner
Input
Output
①
③
⑤
④
Pure Water System
3. Regeneration method
1) Close the Tap water.
2) Close the ⑤ valve.
3) Disassemble the ③ Output line and put it in the sink.
4) Open the ① cap.
5) Insert 500~600g salt and close the ① cap.
6) Remain it for 30 minutes and open the ⑤ valve.
And, open the tap water slightly as the water drips drop by drop for 10 minutes.
7) Open the tap water more for 10 minutes to remove the salt all in it.
(flow rate : 1L/min. approx.)
8) Close the tap water and connect ③ Output line to the main instrument.
Now, it is ready to use again!
2. Installation (Refer to Page 38)
1) Connect ④ Input Line to Tap Water (10mm(OD) Blue tubing)
2) Connect ③ Output to Pure Water System or pretreatment system
3) Open the Tap water.
4. Regeneration & Replacement Period
① Available volume of Water : Approx. 2000L (In case of the Hardness of feed water is 1000ppm)
② Regeneration Period : 1 week
③ Replacement period of Cation Resin in Water Softner : Approx. 6 ~ 10 months
Volume of Cation Resin : Approx. 3L
Arioso Power (TOC) System 39
Specifications
Pretreatment System16
21 3
Picture No. Part No.
2) 10 Housing : 1ea
3) Stand
1) 10 Prefilter : 2ea
2) 10 A/C filter : 1ea
1) 10 Prefilter : 1ea
3) Stand
1) 10 Prefilter : 1ea
2) 10 Housing : 2ea
3) 10 Housing : 2ea
4) Stand
Part No. Remark
HMC - PT - 1 In case of plenty of particles and rust in the feed water
HMC - PT - 2 In case of plenty of particles and rust in the feed water
HMC - PT - 3 In case of plenty of particles, rust and organics in the feed water
1
2
3
※ Installation Method
Sink Table Pure Water System
IN : Connect to Tap Water (Input)
OUT : Connect to Pure Water System (Product)
→Input (In)
→Product (Out)
HMC - PT - 1 HMC - PT - 2 HMC - PT - 3
”
”
”
”
”
”
”
Human C o r p o r a t i o n40
54 6
Specifications
Picture No. Part No.
1) Water Softner : 1ea
2) 10 Prefilter : 1ea
3) 10 Housing : 1ea
4) Stand
1) Water Softner : 1ea
2) 10 Prefilter : 1ea
3) 10 Housing : 1ea
4) Stand
10 PrefilterFilters
HMC - 10 PE Replacement Period : Approx. 30 days /
When the filter becomes to brown color
10 A/C filter HMC - 10 AC Replacement Period : Approx. 180 days
1) 10 Prefilter : 1ea
2) Stand
Part No. Remark
HMC - WS 10
① When the feed water is the Ground water.
② In case of plenty of particles and organics and rust in the feed water
③ In case of plenty of Hardness ((Ca++, Mg++) in the feed water
④ When the conductivity of feed water is more than 300~400 g/㎝
HMC - WS 10 - 1
HMC - WS 10 - 2
4
5
6
HMC - WS 10 HMC - WS 10 - 1 HMC - WS 10 - 2
”
”
”
”
”
”
”
Arioso Power (TOC) System 41
20L 30L 40L 60L
•Wall-mounting bracket
•Automatic Regeneration Softner
•Water Tank (Pedestal option)
•Frame
A-type
B-type
Model : RO, UP Series
Model : Power Series
430(W) x 630(D) x 1350(H) mm
520(W) x 650(D) x 1350(H) mm
Accessories17
Model Volume In/Outlet Flow RateFlow RateResin (cm) Salt (cm)
Tank
HMC-AR6 6L 20mm 1000L/Hr 15 x 45 24 x 42 x 72
HMC-AR20 10L 20mm 1500L/Hr 20 x 45 24 x 42 x 72
A B
Human C o r p o r a t i o n42
Parts18
① UP Sensor
② RO Sensor
③ Feed Sensor
④ Sol Valve 1
⑤ Temperature Sensor
⑥ Sol Valve 3
⑦ Sol Valve 4
⑧ Sol Valve 5
⑨ Pressure sensor No.1 (Feed Pressure)
⑩ Pressure sensor No.2
⑪ Main Board
⑫ SMPS
⑬ TOC Board (Option)
⑭ UV Lamp Adapter
⑮ Pump Adapter
⑯ Auto Drain Valve (Installed in Arioso Power Ⅰ only)
Arioso Power (TOC) System 43
①
⑪
⑫
⑮
⑬
④ ⑯
⑤
⑩
⑥⑦
⑧
⑨
⑭
②③
Maintenance19
1) Sol valve Check method
SET Parts Check→ →
•Check method
Sound Pump Symptom Note
SOL .1
SOL .1 + SOL .2
SOL .1 + SOL .3
SOL .1 + SOL .4
SOL .1 + SOL .5
SOL .3
SOL .1 + SOL .2 + SOL .4
O O Produce RO water, Drain
O X -
O O Drain
O X -
O O Increase Drain water
O O Drain
SOL .2
SOL .1 + SOL .2 + SOL .3
SOL .4
SOL .1 + SOL .2 + SOL .4
+ SOL .5
SOL .5
X X -
O O Produce RO water, Drain
O X UV Lamp on Option
O O Drain, UV Lamp on Option
O O Drain
O O Produce RO water, Drain
2) Functional check method
SET Functional Check→ →
•Check method
working valve Pump Symptom Note
RO Product
Auto Cleaning
SOL .1 O Produce RO water, Drain
SOL .1, 4, 5 O Produce UP water, TOC lamp on Option
SOL .1, 3 O Drain
UP Product
UV Lamp
TOC
SOL .1, 4, 5 O Produce UP water, Drain
SOL .4 - UV lamp on Option
Human C o r p o r a t i o n44
•Even after pressing UP → unable to produce UP water
•Over the process of UP Recycling, UP water drips.
•Unable to execute Recycling
•Even after pressing UP → unable to produce UP water
•Over the process of UP Water production, RO water drips.
3) Trouble Shooting Details
Tank Full
Pressure S/W
Pump
Product/Drain Valve
Trouble Parts Problem / Possible Causes Solution
Sol 1
Sol 3
Sol 4
Sol 5
Sol 2
•When the system works, unknown noise occurs.
•Unable to produce RO Water
•No drain comes out
Unable to display
When the system is on following condition ; - Low Pressure - Tank Full - Replace RO Pack
Error on UV lamp lighting
① Even there is no water inside water tank but Tank Full message displays ② When the water Tank is full, unable to display Tank Full
•When RO Water droplets slowly without stopping
the production during‘Auto Drain’.
Even the system works properly but suddenly
the system reboots/resets and initializes to Self Test
Unable to operate the system at all.
Unable to adjust the volume of RO Product / Drain
① Check the Power Cord
② Check the Fuse
③ Check the Power Connector
④ Check the Power Supply
The Valve doesn’t work.
(It is normal.)
① Check the UV Lamp Adapter
② Check the Power Supply
③ Check the Slave Board
④ Check the Relay Board
① Change the location of the
Floater of Level Sensor
② Check the Slave Board
Replace Pressure S/W
Replace Pump
Replace Product / Drain Valve
• Replace Sol 1
• Replace Sol 3
• Replace Sol 4
• Replace Sol 5
Arioso Power (TOC) System 45
1 ) The feed water is not supplied properly from the tap water. - Check
2) Due to defective pump or error, unable to supply the feed water sufficiently to the filters.
- Check the operating status and replace the pump
3) Due to exhausted RO Pack, the filter membrane has been clogged - Replace the Filter
4) When you replace the RO Pack, the cutting parts are not removed.
- Check and remove the Cutting Parts
5) In case the Solenoid Valve No.1 (2-way) is defective. - Check & Replace it
6) When the Procuct/Drain valve has not been adjusted properly or is defective.
- Check & adjust the pressure gauge at front side. If you unable to adjust, please replace to new one.
4) Unable to produce RO Water
5) Unable to produce UP Water1 ) In case unable to produce RO Water properly.
2) In case the Solenoid Valve No.4 (3-way) & Solenoid Valve No.5 (3-way) are defective.
- Check Solenoid Valve No.4 (3-way) first and replace and test again
3) When the UP Pack has been clogged - Replace it
- Due to exhausted filter, the filter has been clogged by particles or dust - Replace the Filter.
- If the cutting parts are not removed properly - Check the removal of Cutting Parts
6) LCD doesn’t work even after main power is switched on
1 ) Main Power Connector Error - Check and Replace
2) Fuse has been disconnected - Replace to spare Fuse
3) Due to power supply error, unable to supply electricity to the controller - Check & Replace
7) When‘ Low Pressure’ Message appears
1 ) In case the feed water (tap water) has not been supplied sufficiently - Check
2) In case the pressure S/W is defective - When the feed water pressure is low, check the water quality/status If necessary / the pretreatment system should be added.
9) When the pressure of RO membrane cannot be adjusted as 3 to 7 bar automatically in Arioso Power Ⅰ
1 ) Adjust it using‘Product/Drain valve’.
2) Check the‘Auto drain valve’and replace it.
8) When the system goes back to initialization repeatedly during running the system. (Repeatedly execute self test, auto drain / Reset system)
1 ) Lack of feed water
2) Auto Pressure S/W - Manual Position
3) Pressure S/W error - Pressure S/W error can be occurred by lack of feed water pressure
4) When the Level Sensor is defective or Sensor has been wrongly placed.
SET ⇨NEXT Auto Manual→ → or
Human C o r p o r a t i o n46