Inspection Certificate

•Item : Water Purification System

•Model :

•Serial No. :

•Date :

•Origin : Republic of Korea

Authorized Signature

We hereby certify that goods has been inspected before

shipment and found in good order.

Page 5: Manual AriosoPower(TOC)

General Description 5

8

10

11

12

14

16

22

24

26

33

35

36

37

38

40

42

43

44

System Structure

TOC meter Structure

Installation Condition

Installation Method Ⅰ

Installation Method Ⅱ (Diagram)

Display

Operating

Volume Dispensing ∙ Dispensing Time

Filter Exchange and Display / Default Value

Error Message / Default Value

TOC Meter Error & UV Lamp

Adapter

Using Method of Level Sensor

Using Method of Water Softner

Pre-Treatment system

Accessories

Parts for a instrument

Maintenance

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

Index

Manual

Page 6: Manual AriosoPower(TOC)
Page 7: Manual AriosoPower(TOC)

General Description

1. Arioso Power (TOC) System

1) Feed Water

3) STD Accessories

2) Included filters

4) Optional Accessories

5) Applications

1

The Power System is the unique water purification system combining

2 systems-RO System & UP System. It can produce simultaneously from the city/ground water.

And, it is a perfect instrument in space budget and efficiency.

: Tap Water / Ground Water (Water Softner required)

: Manual, Power Cord, Level Sensor, Hand Lever

- Pretreatment : Pre filter + 10 ” A/C Filter- RO System : RO Pack

- UP System : ① UP Pack

② 0.2 Capsule filter

- TOC System (option)

① 254/185 UV Lamp

② MW 5000 UF filter

③ Printer

④ Foot-pedal s/w

⑤ Pressure regulater

⑥ Wall-mounting bracket

⑦ Water Tank (20L / 30L / 40L / 60L)

⑧ Frame

① General laboratory tests

② Preparations of stock solution

③ Washing & rinsing for glasswares

④ Analytical Instruments

⑤ Microorganism analysis

⑥ Cell Culture

⑦ DNA/RNA Tests

m

Arioso Power (TOC) System 5

Page 8: Manual AriosoPower(TOC)

6) Flow Schematic

∙ RO System

∙ UP System

① Pretreatment system

Ⓐ UP pack

④⑥ Pressure sensor

Ⓓ Resistivity sensor

② Temp sensor

Ⓑ 254/186 UV lamp

⑦ RO pack

Ⓔ 0.2 final filter (PF final filter)

③⑧ Conductivity sensor

Ⓒ MW5000 UF filter

⑨ Water sensor

2. Features

∙ Certified CE, ISO 9001, ISO 14001, NRTL

∙ Awarded Korean excellent products

∙ Auto calibraion for water sensor

∙ Extensible point of use

∙ Frost alarm message

∙ 3, 5 steps water level sensor

∙ Filter exchange indicator : Beep & charactor display

∙ Languages : English, Korean, etc

∙ Self-diagnosis functions

Drain RO Product UP Product

① ② ③ ④ ⑤ ⑥ ⑦ ⑧ Ⓐ

Ⓒ Ⓓ

ⒺTank

Ts

m

Human C o r p o r a t i o n6

Page 9: Manual AriosoPower(TOC)

∙ Feed water requirement ; TOC < 50 ppb

3. Specifications

Water Quality

-TOC (ppb) 5~10 5~10 0~5 0~5 5~10 0~5

-Endotoxin (Eu/㎖ ) NA < 0.125 NA < 0.125 < 0.001 < 0.001

-Bacteria (cfu/㎖ ) < 1 < 1 < 1 < 1 < 1 < 1

-Particles ( > 0.22 ㎛/㎖ ) < 1 < 1 < 1 < 1 < 1 < 1

-RNases (ng/㎖) - < 0.01

< 4-DNases (pg/㎖) -DNases (pg/ )

-

-

-

-

-

-

-

< 0.01

< 4

-

-

Dimensions (WxDxH) 320 x 490 x 475 mm

Power 230V / 110V, 50/60Hz, 60W

Standard Accessories Pretreatment System, ROmembrane, UP Pack, Level Sensor, Manual, Power Cord

Optional Accessories· Printer · Foot-Pedal switch · Pretreatment system · Low pressure valve, Wall-mounting bracket· Water tank(20L, 30L, 40L, 60L) · Vent filter for PE water tank · Frame

Arioso Power Arioso Power Arioso Power Arioso Power Arioso Power Arioso PowerⅠⅠ +70 Ⅱ Ⅱ+70 Ⅲ Ⅱ(TOC)

Arioso PowerⅢ(TOC)

Product flow rate(RO, UP) Max. 15 L/Hr Max. RO : 15 L/Hr

UP : 70 L/Hr Max. 25 L/Hr Max. RO : 25 L/Hr UP : 70 L/Hr Max. 25 L/Hr Max. 35 L/HrMax. 35 L/Hr

Display & Controller Large color 70 x 52 touch screen ←

Water quality display

Ambient temp : -10 to 100 ←

Feed water temp : 0 to 100℃ ←

Feed water : 0 to 999㎲/㎝ ←

Pure water (RO) : 0.2 to 250.0㎲/㎝ ←

Ultra Pure (UP) : Up to 18.3㏁·㎝ ←

←←←←←←

- TOC : 0 to 200 ppb

Pressure display Digital Pressure Sensor (0~10kg/㎠) ① Feed water pressure ② RO membrane pressure

Self-check function · Various valves · Auto sensor calibration(RO, UP)· System operation status · Self-diagnosis

Cleaning function · Auto regular flushing · Auto regular cleaning · Auto cleaning lower than setting value

System function· Filter replacement alarm & display (A/C, RO/UP pack, UF, UV lamp, 0.2㎛CF, etc)· Volumetric time setting (RO/UP) · Language(English, Korean, etc)· Auto/Manual Pressure S/W · Extensible point of use

TypeScholar Scholar-UV

Bio Integrate_ PF _ PF

0.2㎛ final filter STD STD STD STD

Pyrogen free final filter STD STD

254/185 UV lamp STD STD STD

MW 5000 UF filter STD STD

< 0.01 < 0.01

< 4 < 4

Arioso Power (TOC) System 7

Page 10: Manual AriosoPower(TOC)

System Structure

1. Pretreatment System (HMC-PT-3)

2. Main System / Front

2

Air Vent Button(Red Button)

Display

0.2 Capsule filter

Locking system

UP/RESET switch

10 ” Pre Filter10 ” A/C Filter

Output[From this port to main instrument

using Blue 6mm(OD) tubing]Input

[From tap water to this portusing Blue 10mm(OD) tubing]

m

Human C o r p o r a t i o n8

Page 11: Manual AriosoPower(TOC)

4. Main System / Back Side

3. Main System / Front-inside

① Input Port [6mm(OD)]

Connect to output of Pre-treatment system

using Blue 6mm(OD) tubing

② Pure Water Product Port

Connect to Water tank using White 6mm(OD) tubing

③ Drain Port

Connect to Sink using Blue 6mm(OD) tubing

④ TOC Input (Option)

Connect to Input of TOC Meter using transparent 4mm(OD) tubing

⑤ Printer Port (Option) - 4 pin

Connect to Printer

⑥ TOC Meter Power Port (Option) - 5 pin

Connect to Power of TOC meter

⑦ Level Sensor - 2, 3, 6 pin

After connect level sensor, put it Water Tank

⑧ Power switch

Connect to Power

RO Pack

254/185 nm UV Lamp

MW 5000 UF Filter

UP Pack

(TOC Version)

(General version without TOC Meter)

④ ⑤

⑥⑦

Arioso Power (TOC) System 9

Product / Drain valve

Control Water volume with Ⓐ valve ,

and then fix Ⓐ valve by turning Ⓑ valve clockwise.

(※Do not fasten too strongly.) (‘Auto drain valve’is installed in Arioso Power I.)

Ⓐ valve

Ⓑ valve

Page 12: Manual AriosoPower(TOC)

TOC Meter Structure (Option)3

1. Front

2. Back

① Water Flow Controller

Control Water Flow of TOC Meter (15-20㎖/min : approx.

30-35drops / 10 sec) Set the ball in‘② Flow Meter’

between RED lines.

② Flow Meter for TOC

Hole

(Fit for the back-left rubber feet of main system.)

※ If there is air inside Flow Meter, Ball could go down.

RED LineBall

∙ TOC Meter Power Port

Connect to Main system ⑥ TOC Meter Power Port

(Refer to page 09)

∙ TOC Meter Input

Connect to Main system ④ TOC Input using transparent

4mm(OD) tubing

∙ TOC Meter Output

Connect to Sink using transparent 4mm(OD) tubing

※ Make sure that tubing is not twisted and set upside.

Human C o r p o r a t i o n10

Page 13: Manual AriosoPower(TOC)

Installation Condition 4

1. Environmental Conditions

1) Indoor installation

2) Altitude should be up to 2000m

3) Ambient temperature : 5℃ ~ 40℃

(If the temperature is lower the 5℃, the system can be frozen up)

4) Maximum relative humidity : 80%

5) Power : 230 ± 10%V, 50/60㎐ ± 1㎐, 60W

6) Pollution degree 2

7) Installation Categories Ⅱ

8) The instrument should be placed on steady workbench that is free

from a strong vibration source.

9) The location room should be free from a strong electro magnetic

interference source or harmful or corrosive gases.

CAUTION!! - Refer to accompanying documents

CAUTION!! - ELECTRIC SHOCK

“ To avoid electrical shock, do not open the cover.

Refer servicing to qualified personnel only.”

Arioso Power (TOC) System 11

Page 14: Manual AriosoPower(TOC)

Installation Method Ⅰ 5

1. Pretreatment System (Refer to Page 8)

2. Main System (Refer to Page 9)

3. Main System (Refer to Page 8)

1) Unscrew the locking system at under of Main body and Open it.

2) Lock the door again when you are using the instrument.

2) Before connect the main system, wash 10 ” A/C filter by opening Tap Water.

[Drain minimum 60 sec.]

1) Connect the Input of Pretreatment System into the Tap Water. [Blue 10mm(OD) tubing]

3) Connect the Output of Pretreatment into the Input Port in back side of main system.

[Blue 6mm(OD) tubing]

Human C o r p o r a t i o n12

① Input Port [6mm(OD)]

Connect to output of Pre-treatment system

using Blue 6mm(OD) tubing

② Pure Water Product Port

Connect to Water tank using White 6mm(OD) tubing

③ Drain Port

Connect to Sink using Blue 6mm(OD) tubing

④ TOC Input (Option)

Connect to Input of TOC Meter using transparent 4mm(OD) tubing

⑤ Printer Port (Option) - 4 pin

Connect to Printer

⑥ TOC Meter Power Port (Option) - 5 pin

Connect to Power of TOC meter

⑦ Level Sensor - 2, 3, 6 pin

After connect level sensor, put it Water Tank

⑧ Power switch

Connect to Power

Page 15: Manual AriosoPower(TOC)

5. TOC Meter Installation Method

4. Connect Front Pack (Refer to Page 9)

1) Connect RO pack Line

2) Connect UP Pack line

3) Connect 0.2 Capsule Filter

4) Connect 254/185 UV Lamp (Option)

5) Connect MW 5000UF filter (Option)

① Put TOC Meter back-right of Main system.

② Insert the back-left rubber feet of main system into the hole of TOC meter.

m

Arioso Power (TOC) System 13

Page 16: Manual AriosoPower(TOC)

Installation Method Ⅱ (Diagram)6

1. Pre-treatment System Diagram

Air Vent RED button

① Open the tap after connecting lines.

② Drain water for approx. 60sec to wash A/C filter.

③ Connect to input port of Main system after close the tap. (Page 9)

10 ” Pre Filter

Blue 10mm(OD) tubing

10 ” A/C Filter

Human C o r p o r a t i o n14

Page 17: Manual AriosoPower(TOC)

2. Back side of Main System

TOC Meter

Printer

•Level Sensor Type

⑴ 2 step level sensor ⑵ 3 step level sensor ⑶ 5 step level sensor

s i n k

① Blue 6mm(OD) input tubing

② White 6mm(OD) (RO product) tubing

③ Blue 6mm(OD) (RO Drain) tubing

④ Transparent 4mm(OD) (TOC meter Input) tubing

⑤ Printer cable (4pin)

⑥ TOC meter Power (5pin)

⑦ Level Sensor

⑧ Power Cable port & Switch

TOC drain

② ③ ④ ⑤ ⑥ ⑦ ⑧

Arioso Power (TOC) System 15

PretreatmentOutput

Page 18: Manual AriosoPower(TOC)

Display 7

1. Main Display

2. Operation Display

MODE

R

U

SET

UP/RESET

UP/RESET

A 1.2 bar/F M 1.2 bar/F

: Setting Operating Condition

: Produce / Stop RO water Key

: Produce / Stop UP water Key

: Checking Instrument

or

: ① Press this, Produce / Stop UP water

② Press for 3sec, back to Main display

1)

① A (Auto) : Only works when the Tap water pressure is in 1~5bar(kg/㎠).

It doesn’t work when the Tap water pressure is over 5bar(kg/㎠) or

blow than1bar(kg/㎠). (Default Value : 1~5bar, Modifiable at )

② M (Manual) : Only works regardless of Tap water pressure (Remarks : It works even though there is no supplied water!)

③ 1.2 bar/F : Present Feed Water pressure

④ Setting Method :

MODE

SET ⇨NEXT Auto Manual→ → or

Human C o r p o r a t i o n16

Page 19: Manual AriosoPower(TOC)

5.2 bar/RO - bar/RO2)

① Pressure of RO Membrane (Default Value : 3 to 7 bar)

Regulate Pressure of RO Membrane, the ratio of RO product / Drain volume

by controlling

② Analog Pressure guage is equipped when it does not have Pressure sensor of RO membrane.

Product / Drain Valve

3)

① RO Product : 0.2 /cm

② Feed water Conductivity : 148 /cm (Default Value : < 350 /cm)

RO148

/cm

feed

0.2

4)

① UP Product : 18.3 cm at 25℃

② TOC : 2 ppb (Option)

UP2 ppb

cm

TOC

18.3

① 15.0℃ : Temperature of Feed Water

② : Beep ON

③ : Beep OFF

④ Setting method :

or5) 15.0 ℃/F 15.0 ℃/F

orSET Beep ON Beep OFF→

① 21.0℃ Amb : The present Ambient Temperature (Default Value : > 5℃)

② Caution: The system can be frozen up if the temperature is lower than 5℃!

6) 21.0 ℃ Amb

∙ Water Tank ( 2step ,3step ,5step )

7)

or (without Pressure sensor)

Arioso Power (TOC) System 17

s

ss s

Ω∙M

Ω∙M

Model Auto Drain Valve생 산 비 율

Arioso Power Ⅰ

Arioso Power Ⅱ

Arioso Power Ⅲ

Product

4

4

5

Drain

6No

6 Built-in

Auto Drain Valve ;

It automatically adjust the ratio of product and drain water

volume.5

Page 20: Manual AriosoPower(TOC)

(Unit ; /cm)

1)

2)

4)

5)

6)

7)

3)

3. Mode Display

※ For changing Default value Touch display → change the value with key.or

RO / UP

Value Setting

AC / UV

Time Setting

0.2CF / MW5000

Time Setting

Feed

Water µs/cm

Pressure

Value Setting

Running time

RO / UP / AC / UV LAMP

RO / UP

Volume Dispensing

RO / UP

Dispensing Time

: Setting replacement period of RO / UP Pack

① 1st Value : 1st Alert message RO/UP pack replacement

② 2nd Value : 2nd Alarm message RO/UP pack replacement

Setting replacement period of AC Filter, UV Lamp (Unit ; hour)

Set the or first.

Replace period for 0.2 Capsule Filter (or PF Filter) and MW5000 UF Filter

※ In the case of Scholar - PF, Scholar - UV - PF Type Replace 0.2CF →PF Filter!

Setting Max value of Feed water Conductivity.

① FD : Setting Feed Water Pressure (Unit ; kg/㎠, bar)

② RO : Setting RO membrane Pressure (Unit ; kg/㎠, bar)

Display total accumulated running time of

RO Pack, UP Pack, AC Filter, UV Lamp 0.2 CF, UF Filter. (Unit ; hour)

∙ Setting dispensing Volume (Vloume : 0.01 to 99.0 L)

∙ Setting dispensing time

(Option)

or

SET →※

m

m

Human C o r p o r a t i o n18

s

Volume Dispensing Dispensing Time

Page 21: Manual AriosoPower(TOC)

TOC

UV Time

Temp

Amb / Feed water

TOC

Error No

Print

Language

Running time of UV Lamp for TOC Meter (Unit ; hour)

① Amb : Setting Minimum value of Ambient temperature to prevent freezing (Unit ; ℃)

② Feed Water : Setting Maximum value of Feed Water temperature.

Display Error No. for TOC Meter.

Printing Function

Language selecting function

① : Operating with set value in Mode function

② : Operating regardless default values.

8)

9)

10)

11)

12)

13) SettingValue

On Off

On

Off

(※Pressure Function is maintained.)

Arioso Power (TOC) System 19

Page 22: Manual AriosoPower(TOC)

① : Check RO Product

② : Check Auto Cleaning

③ : Check TOC Function

④ : Check UP Product

⑤ : Check UV Lamp

4. SET Display

1)

2)

3)

4)

Parts Check

Functional Check RO Product

TOC

UP Product

UV Lamp

Auto Cleaning

① SOL . 1 : No1. valve Operating check

② SOL . 2 : No2. valve Operating check

③ SOL . 3 : No3. valve Operating check

④ SOL . 4 : No4. valve Operating check

⑤ SOL . 5 : No5. valve Operating check

Calibration in case RO sensor (Conductivity Sensor),

UP sensor(Resistivity sensor) has Error

Choose one between above 2 functions.

① Volume Dispensing (Option) : Dispensing RO and UP after putting dispensing volume you want.

(Volume range : 0.01 to 99.0 L)

② Dispensing Time : Setting time for Dispensing RO and UP

※ Please Inquire about Calibration to Head Office.

or

Calibration

Sensor

Volume Dispensing Dispensing Time

Human C o r p o r a t i o n20

Page 23: Manual AriosoPower(TOC)

Choose one between above 2 functions.

① Volume Dispensing (Option) : Dispensing RO and UP after putting dispensing volume you want.

(Volume range : 0.01 to 99.0 L)

② Dispensing Time : Setting time for Dispensing RO and UP

① : Only works when Feed Water pressure is in 1~5 bar (kg/㎠)

② : Works regardless of feed water pressure (Even works without supplied water)

① Yes : Going back to factory setting values.

② No : Remaining the current setting values.

① Yes : Auto cleaning UP system for 120sec. after Cleaning RO system for 120sec. every 1 hour in standby display.

② No : Cancel upper function.

Calibration to get high accuracy Once equipped the Flow meter

Ex) Calibrate after serveral tests. Set 100ml as dispensing volume

Test the above procedure again till it is calibrated.

produce more than 100ml

Set the above valve as lower than 100

produce less than 100ml

Set the above valve as higher than 100

Range : 0 ~ 200

Range : 0 ~ 200

Using Time Reset

BEEP ON

AUTO

Default Value

Auto Flushing

Volume Calibration

BEEP OFF

Manual

AUTO

Manual

5)

6)

7)

8)

9)

10)

∙ Reset running time of each filters. (Please Inquire it to Head Office)

∙ Setting Alarm or touch sound.

or

or

Arioso Power (TOC) System 21

Caution ; The instrument can be defective because it woks even though there is no supplied water.

Page 24: Manual AriosoPower(TOC)

Operation8

1)

3)

2)

Power S/W ON

To produce UP,

Touch : RO water Product Key

Touch: UP product key

Main Display

Operation Display

R

U

RO

UP

① Self Test : Auto checking Feed water pressure, storage tank etc.,

② Auto Cleaning : Auto Cleaning inside of instrument. ※ Extension : When RO conductivity isn’t good enough, it automatically extends cleaning for max. 600 sec.

③ RO product : Touch RO to stop producing RO water.

touch UP or right side s/w ON

To stop producing UP, re-touch them.

UP/RESET

Touch Part

Pressure : 3 to 7 bar(Default Valve)

Human C o r p o r a t i o n22

Page 25: Manual AriosoPower(TOC)

4) Waiting Display

MODE

TOC

S/B

: Switch over MODE display

(Option) : Measuring TOC value, when TOC Meter is equipped.

: ∙ Switch over Standby display.

∙ To release function, re-touch the key

- Function -

※ During Standby, Auto cleaning circulated.

Once UP water (inside of instrument) value is lower than 10 cm

① RO System : Auto cleaning for 120 sec. and Extend max 600sec.

② UP System : Auto cleaning for 60 sec. and Extend max 600sec.

MΩ∙

Arioso Power (TOC) System 23

※Standby function could shorten the filter replacement period!

Page 26: Manual AriosoPower(TOC)

Volume Dispensing, Dispensing Time9

※ While you use this function,“Filter Exchange message”is not Displayed

1)

Volume Dispensing setting method.

ex) set RO product 13.25L

RO key touch (change the color)

select

After touching set 10

set 3

set 0.2

set 0.05

After touching

After touching

After touching

Calibrate volume, when there is difference between

product volume and set volume.

Operating

Volume Calibration put value

( refer to page 21 )

→ → →SET ⇨NEXT

Human C o r p o r a t i o n24

Page 27: Manual AriosoPower(TOC)

2)

Dispensing Time setting method.

Product volume results from Setting time

- Calculate as much as set dispensing volume.

- Setting method

(Product volume can vary depending on water temp, filter condition)

∙ Water temperature : 7℃

∙ Input Conductivity : 185 /㎝

∙ Model : Arioso Power Ⅰ (Scholar Type)

→ Set time →

key

Main Display

Main Display

Dispensing TimeSET

R U

MODE

Main

MainRO / UP

Dispensing Time

→ → → →

→→

or

15 sec.

80 sec.

47 sec.

120 sec.

30 sec.

90 sec.

63 sec.

2 min. 26 sec.

3 min. 26 sec.

6 min. 7 sec.

9 min. 15 sec.

5 min. 5 sec.

8 min. 10 sec.

7 min. 15 sec.

10 min. 15 sec.

50 ㎖

250 ㎖

150 ㎖

400 ㎖

100 ㎖

300 ㎖

200 ㎖

500 ㎖

700 ㎖

1.2 L

1.8 L

1 L

1.6 L

1.4 L

2.0 L

Dispensing Time Dispensing TimeProduct volume Product volume

Arioso Power (TOC) System 25

s

Page 28: Manual AriosoPower(TOC)

Filter replacement and Display 10

1. Pretreatment System

1) Prefilter

2) A/C filter

1) Replacement Period

① Approx. 20 ~ 40days (depends on using time feed water condition.)

② When the filter becomes to brown color (Refer to sticker posted on the housing.)

3) Replacement Method

① Same as Prefilter replacement Method

② After replace a filter, touch screen for 5sec and you can hear“beep”.

Default Value : 400hr

1) Replacement Period

2) Display

∙ Alarm turned on

∙ Message blinking (Message is steadily displayed when it is used over 400 hours.)

After replace it, touch screen for 5sec until you hear the “beep”

2) Replacement Method

① Make the system stop

② Close the Tap Water.

③ Remove Air by pressing the Air Removing Button (Red one) in the housing.

④ Replace PreFilter

※ can disappear by touch screen,

but it appears when you turn on the system again without replacing filter.

Replace A/C Filter

Human C o r p o r a t i o n26

Page 29: Manual AriosoPower(TOC)

3) RO Filter (Pack)

1) Replacement Period

2) Display

Default Value : ① 1st value : 20 /㎝

② 2nd value : 30 /㎝

∙ Alarm turned on

∙ Message blinking

∙ Alarm turned on

∙ System stop

3) Replacement Method

① Make the system stop

② Close the Tap Water

③ Depressurize the system by touching‘RO’to dispense

water for 20 sec. and stop it.

④ Shut power OFF to the unit and disconnect the electrical power to the unit.

⑤ Open the front door and replace RO pack.

(Connect the Inlet, drain and product tubings.)

⑥ Power it on and open the tap water to operate it again.

•1st value •2nd value

Arioso Power (TOC) System 27

ss

Page 30: Manual AriosoPower(TOC)

4) UP Pack

2) Display

1) Replacement Period

Default Value : ① 1st value : 13.0 ㎝

② 2nd value : 10.0 ㎝

∙ Alarm turned on

∙ Message blinking

∙ Alarm turned on

∙ Stop producing UP water and be recycled.

① Inlet Line

② Inlet Line

③ Outlet Line

④ Product Line

•2 Line : Scholar, Bio

•4 Line : ∙ Scholar - UV ∙ Integrate

4) Structure of UP Pack

•1st value •2nd value

④③ ② ①

Human C o r p o r a t i o n28

Ω∙M

Ω∙M

3) Replacement Method

① Make the system stop

② Close the Tap Water

③ Depressurize the system by touching‘UP’to dispense

water for 20 sec. and stop it.

④ Shut power OFF to the unit and disconnect the electrical power to the unit.

⑤ Open the front door and replace UP pack.

⑥ Power it on and open the tap water to operate it again.

Page 31: Manual AriosoPower(TOC)

2) Display

∙ Alarm turned on

∙ Message blinking

∙ Alarm turned on

∙ massage is steadily displayed

3) Replacement Method

∙ Replace 0.2 CF (or PF filter)

•1st value •2nd value

5) 0.2㎛ Capsule Filter or PF (Pyrogen Free) Filter

1) Replacement Period

Default Value : ① 1st value : 280hr

② 2nd value : 300hr

※ Replace‘PF Filter’in case of Scholar-PF, Scholar-UV-PF Type.

m

Arioso Power (TOC) System 29

touch screen for 5sec and you can hear“beep”

※ message can disappear by touching it.

But it appears when you turn on the system again without replacing filter.

Replace 0.2 CF

Page 32: Manual AriosoPower(TOC)

6) MW5000 UF Filter

1) Replacement Period

2) Display

Default Value : ① 1st value : 280hr

② 2nd value : 300hr

∙ Alaram turned on

∙ Message blinking

∙ Alaram turned on

∙ Message is steadily displayed

3) Replacement Method

•1st value •2nd value

① Make the system stop

② Close the Tap Water ( Remove the Pressure inside)

③ Replace MW 5000 filter

Human C o r p o r a t i o n30

touch screen for 5sec and you can hear“beep”

※ message can disappear by touching it.

But it appears when you turn on the system again without replacing filter.

Replace UF Filter

Page 33: Manual AriosoPower(TOC)

7) UV Lamp

1) Replacement Period

2) Display

Default Value : ① 1st value : 1800hr

② 2nd value : 2000hr

∙ Alarm turned on

∙ Message blinking

∙ Alarm turned on

∙ Message is steadily displayed

•1st value •2nd value

4) Structure and Caution

※ After replacing MW5000 UF Filter or UP Pack, press the‘UP’button and then open the White Cap of MW 5000 UF filter for removing air inside. If the water starts to come out from the vent, close the White Cap. You can use the instrument now.

※ This air removal is also needed for the first use after installation of Bio/Integrate type system.

MW 5000 UF Filter

White Cap

Blue Cap

Applicable for Bio / Integrate Type

Remove the air in the MW 5000 UF Filter!! Or the product volume and purity can be decreased.

Arioso Power (TOC) System 31

Page 34: Manual AriosoPower(TOC)

3) Replacement Method

4) UV lamp structure

① Close the Tap Water

② Turn the instrument off and disconnect power cord from unit.

③ Replace UV lamp (Remove the Pressure inside)

④ Make sure the UV lamp is again connected to the UV lamp adapter.

① Outlet Line : Yellow tubing

② Inlet Line : White tubing

③ Adapter

※ Warning!

This unit is equipped with an ultraviolet lamp. Ultraviolet radiation is harmful to the eyes and skin. Do not attempt to observe the lamp directly. Some water will drain from the UV lamp when it is disconnected.

Human C o r p o r a t i o n32

touch screen for 5sec and you can hear“beep”

※ message can disappear by touch screen,

but it appears when you turn on the system again without replacing filter.

Replace UV Lamp

Page 35: Manual AriosoPower(TOC)

Error Message11

1)

3)

2)

∙ Alarm turned on

∙ Alarm turned on

∙ Alarm turned on

① Adjust the feed water pressure to less than 5kg/㎠

② Install pressure regulator (Ask for it to Head-Office).

Decrease the temperature of feed water. Otherwise, filter can be damaged.

① Use booster pump

② Change feed water pressure mode to Manual mode

Feed Pressure High

Feed Temp Exceeding

Feed Pressure Low

•Solution

•Solution

•Solution

Default Value : 5kg/㎠

Default Value : 35.0℃

Default Value : 1kg/㎠

SET⇨

nextAuto or Manual→ →

※ At manual mode, Instrument can be operated even without Water supply.

Arioso Power (TOC) System 33

Page 36: Manual AriosoPower(TOC)

4)

5)

6)

∙ Alarm turned on

∙ Alarm turned on

① Install the Pretreatment system.

② If the feed water is ground water, Water softner should be installed.

Freezing Caution

① When Pressure of RO membrane is out of Default value, Purity and Product volume can be influenced by it.

② Adjust Default Value by turning Drain Valve (Page 09)

③ Ratio of Product and Drain volume

Increase the ambient temperature to prevent Freezing.

or

Feed Purity Exceeding

Check drain valve

•Solution

•Solution

•Solution

Default Value : 350 /㎝

Default Value : RO membrane pressure - 3~7㎏/㎠

Default Value : 5℃

Human C o r p o r a t i o n34

s

Model Auto Drain Valve생 산 비 율

Arioso Power Ⅰ

Arioso Power Ⅱ

Arioso Power Ⅲ

Product

4

4

5

Drain

6No

6 Built-in

Auto Drain Valve ;

It automatically adjust the ratio of product and drain water

volume.5

Page 37: Manual AriosoPower(TOC)

TOC Meter12

1) Error message

2) TOC meter UV lamp replacement method.

No

1

4

7

2

5

8

3

6

9

Error Message

Error 02

Error 08

Error 06

Error 18

Change UV Lamp

Error 03

Error 10

Error 07

System Check Failed

Error Code Error Reason Remark

02 Quality of Sample water is out of the Specification (Below than 1 ㎝)

07 Temp. of sample water is out of the specification (lower than 5℃)

18 UV Lamp doesn’t illuminate. (Disconnection of UV lamp)

01 Replace UV Lamp (After using 4000hr)

06 Temp. of Sample water is over the Specification (higher than 35℃)

10 Value of TOC is out of the range of Specification (Over 200ppb)

03 Temp. Sensor 1 (Error on Thermistor)

08 Temp. Sensor 2 (Error on Thermistor)

19 Message error, measurement is not possible.

TOC Drain Tube

TOC Inlet Tube

TOC UV LampUV Lamp.Slowly lifeit up.

Arioso Power (TOC) System 35

Ω∙M

Page 38: Manual AriosoPower(TOC)

Adapter13

1. Power Supply

2. UV Lamp Adapter

Human C o r p o r a t i o n36

Page 39: Manual AriosoPower(TOC)

How to use Level Sensor14

•Level Sensor Type

① 2 step level sensor

② 3 step level sensor

③ 5 step level sensor

When you open the Tank cap,

make sure that level sensor doesn’t

get twisted.

You should Install Level

sensor vertically.

Arioso Power (TOC) System 37

Page 40: Manual AriosoPower(TOC)

Water Softner for Pure Water System15

1. Name of No.

① Cap

② Main Body

③ Output Line or Drain in Regeneration

④ Input Line (Tap Line)

⑤ Input Valve

Human C o r p o r a t i o n38

Page 41: Manual AriosoPower(TOC)

Installation

Sink Table

Water Softner

Input

Output

Pure Water System

3. Regeneration method

1) Close the Tap water.

2) Close the ⑤ valve.

3) Disassemble the ③ Output line and put it in the sink.

4) Open the ① cap.

5) Insert 500~600g salt and close the ① cap.

6) Remain it for 30 minutes and open the ⑤ valve.

And, open the tap water slightly as the water drips drop by drop for 10 minutes.

7) Open the tap water more for 10 minutes to remove the salt all in it.

(flow rate : 1L/min. approx.)

8) Close the tap water and connect ③ Output line to the main instrument.

Now, it is ready to use again!

2. Installation (Refer to Page 38)

1) Connect ④ Input Line to Tap Water (10mm(OD) Blue tubing)

2) Connect ③ Output to Pure Water System or pretreatment system

3) Open the Tap water.

4. Regeneration & Replacement Period

① Available volume of Water : Approx. 2000L (In case of the Hardness of feed water is 1000ppm)

② Regeneration Period : 1 week

③ Replacement period of Cation Resin in Water Softner : Approx. 6 ~ 10 months

Volume of Cation Resin : Approx. 3L

Arioso Power (TOC) System 39

Page 42: Manual AriosoPower(TOC)

Specifications

Pretreatment System16

21 3

Picture No. Part No.

2) 10 Housing : 1ea

3) Stand

1) 10 Prefilter : 2ea

2) 10 A/C filter : 1ea

1) 10 Prefilter : 1ea

3) Stand

1) 10 Prefilter : 1ea

2) 10 Housing : 2ea

3) 10 Housing : 2ea

4) Stand

Part No. Remark

HMC - PT - 1 In case of plenty of particles and rust in the feed water

HMC - PT - 2 In case of plenty of particles and rust in the feed water

HMC - PT - 3 In case of plenty of particles, rust and organics in the feed water

1

2

3

※ Installation Method

Sink Table Pure Water System

IN : Connect to Tap Water (Input)

OUT : Connect to Pure Water System (Product)

→Input (In)

→Product (Out)

HMC - PT - 1 HMC - PT - 2 HMC - PT - 3

Human C o r p o r a t i o n40

Page 43: Manual AriosoPower(TOC)

54 6

Specifications

Picture No. Part No.

1) Water Softner : 1ea

2) 10 Prefilter : 1ea

3) 10 Housing : 1ea

4) Stand

1) Water Softner : 1ea

2) 10 Prefilter : 1ea

3) 10 Housing : 1ea

4) Stand

10 PrefilterFilters

HMC - 10 PE Replacement Period : Approx. 30 days /

When the filter becomes to brown color

10 A/C filter HMC - 10 AC Replacement Period : Approx. 180 days

1) 10 Prefilter : 1ea

2) Stand

Part No. Remark

HMC - WS 10

① When the feed water is the Ground water.

② In case of plenty of particles and organics and rust in the feed water

③ In case of plenty of Hardness ((Ca++, Mg++) in the feed water

④ When the conductivity of feed water is more than 300~400 g/㎝

HMC - WS 10 - 1

HMC - WS 10 - 2

4

5

6

HMC - WS 10 HMC - WS 10 - 1 HMC - WS 10 - 2

Arioso Power (TOC) System 41

Page 44: Manual AriosoPower(TOC)

20L 30L 40L 60L

•Wall-mounting bracket

•Automatic Regeneration Softner

•Water Tank (Pedestal option)

•Frame

A-type

B-type

Model : RO, UP Series

Model : Power Series

430(W) x 630(D) x 1350(H) mm

520(W) x 650(D) x 1350(H) mm

Accessories17

Model Volume In/Outlet Flow RateFlow RateResin (cm) Salt (cm)

Tank

HMC-AR6 6L 20mm 1000L/Hr 15 x 45 24 x 42 x 72

HMC-AR20 10L 20mm 1500L/Hr 20 x 45 24 x 42 x 72

A B

Human C o r p o r a t i o n42

Page 45: Manual AriosoPower(TOC)

Parts18

① UP Sensor

② RO Sensor

③ Feed Sensor

④ Sol Valve 1

⑤ Temperature Sensor

⑥ Sol Valve 3

⑦ Sol Valve 4

⑧ Sol Valve 5

⑨ Pressure sensor No.1 (Feed Pressure)

⑩ Pressure sensor No.2

⑪ Main Board

⑫ SMPS

⑬ TOC Board (Option)

⑭ UV Lamp Adapter

⑮ Pump Adapter

⑯ Auto Drain Valve (Installed in Arioso Power Ⅰ only)

Arioso Power (TOC) System 43

④ ⑯

⑥⑦

②③

Page 46: Manual AriosoPower(TOC)

Maintenance19

1) Sol valve Check method

SET Parts Check→ →

•Check method

Sound Pump Symptom Note

SOL .1

SOL .1 + SOL .2

SOL .1 + SOL .3

SOL .1 + SOL .4

SOL .1 + SOL .5

SOL .3

SOL .1 + SOL .2 + SOL .4

O O Produce RO water, Drain

O X -

O O Drain

O X -

O O Increase Drain water

O O Drain

SOL .2

SOL .1 + SOL .2 + SOL .3

SOL .4

SOL .1 + SOL .2 + SOL .4

+ SOL .5

SOL .5

X X -

O O Produce RO water, Drain

O X UV Lamp on Option

O O Drain, UV Lamp on Option

O O Drain

O O Produce RO water, Drain

2) Functional check method

SET Functional Check→ →

•Check method

working valve Pump Symptom Note

RO Product

Auto Cleaning

SOL .1 O Produce RO water, Drain

SOL .1, 4, 5 O Produce UP water, TOC lamp on Option

SOL .1, 3 O Drain

UP Product

UV Lamp

TOC

SOL .1, 4, 5 O Produce UP water, Drain

SOL .4 - UV lamp on Option

Human C o r p o r a t i o n44

Page 47: Manual AriosoPower(TOC)

•Even after pressing UP → unable to produce UP water

•Over the process of UP Recycling, UP water drips.

•Unable to execute Recycling

•Even after pressing UP → unable to produce UP water

•Over the process of UP Water production, RO water drips.

3) Trouble Shooting Details

Tank Full

Pressure S/W

Pump

Product/Drain Valve

Trouble Parts Problem / Possible Causes Solution

Sol 1

Sol 3

Sol 4

Sol 5

Sol 2

•When the system works, unknown noise occurs.

•Unable to produce RO Water

•No drain comes out

Unable to display

When the system is on following condition ; - Low Pressure - Tank Full - Replace RO Pack

Error on UV lamp lighting

① Even there is no water inside water tank but Tank Full message displays ② When the water Tank is full, unable to display Tank Full

•When RO Water droplets slowly without stopping

the production during‘Auto Drain’.

Even the system works properly but suddenly

the system reboots/resets and initializes to Self Test

Unable to operate the system at all.

Unable to adjust the volume of RO Product / Drain

① Check the Power Cord

② Check the Fuse

③ Check the Power Connector

④ Check the Power Supply

The Valve doesn’t work.

(It is normal.)

① Check the UV Lamp Adapter

② Check the Power Supply

③ Check the Slave Board

④ Check the Relay Board

① Change the location of the

Floater of Level Sensor

② Check the Slave Board

Replace Pressure S/W

Replace Pump

Replace Product / Drain Valve

• Replace Sol 1

• Replace Sol 3

• Replace Sol 4

• Replace Sol 5

Arioso Power (TOC) System 45

Page 48: Manual AriosoPower(TOC)

1 ) The feed water is not supplied properly from the tap water. - Check

2) Due to defective pump or error, unable to supply the feed water sufficiently to the filters.

- Check the operating status and replace the pump

3) Due to exhausted RO Pack, the filter membrane has been clogged - Replace the Filter

4) When you replace the RO Pack, the cutting parts are not removed.

- Check and remove the Cutting Parts

5) In case the Solenoid Valve No.1 (2-way) is defective. - Check & Replace it

6) When the Procuct/Drain valve has not been adjusted properly or is defective.

- Check & adjust the pressure gauge at front side. If you unable to adjust, please replace to new one.

4) Unable to produce RO Water

5) Unable to produce UP Water1 ) In case unable to produce RO Water properly.

2) In case the Solenoid Valve No.4 (3-way) & Solenoid Valve No.5 (3-way) are defective.

- Check Solenoid Valve No.4 (3-way) first and replace and test again

3) When the UP Pack has been clogged - Replace it

- Due to exhausted filter, the filter has been clogged by particles or dust - Replace the Filter.

- If the cutting parts are not removed properly - Check the removal of Cutting Parts

6) LCD doesn’t work even after main power is switched on

1 ) Main Power Connector Error - Check and Replace

2) Fuse has been disconnected - Replace to spare Fuse

3) Due to power supply error, unable to supply electricity to the controller - Check & Replace

7) When‘ Low Pressure’ Message appears

1 ) In case the feed water (tap water) has not been supplied sufficiently - Check

2) In case the pressure S/W is defective - When the feed water pressure is low, check the water quality/status If necessary / the pretreatment system should be added.

9) When the pressure of RO membrane cannot be adjusted as 3 to 7 bar automatically in Arioso Power Ⅰ

1 ) Adjust it using‘Product/Drain valve’.

2) Check the‘Auto drain valve’and replace it.

8) When the system goes back to initialization repeatedly during running the system. (Repeatedly execute self test, auto drain / Reset system)

1 ) Lack of feed water

2) Auto Pressure S/W - Manual Position

3) Pressure S/W error - Pressure S/W error can be occurred by lack of feed water pressure

4) When the Level Sensor is defective or Sensor has been wrongly placed.

SET ⇨NEXT Auto Manual→ → or

Human C o r p o r a t i o n46

Page 49: Manual AriosoPower(TOC)

M E M O

Page 50: Manual AriosoPower(TOC)