Indra J. Das, Indra J. Das, PhD, FAAPM, FACR, FASTRO PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Department of Radiation Oncology Indiana University School of Medicine Indiana University School of Medicine Indianapolis, IN, USA Indianapolis, IN, USA Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Indra J. Das, Indra J. Das, PhD, FAAPM, FACR, FASTROPhD, FAAPM, FACR, FASTRODepartment of Radiation OncologyDepartment of Radiation OncologyIndiana University School of MedicineIndiana University School of MedicineIndianapolis, IN, USAIndianapolis, IN, USA

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Page 2: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Stereotactic RadioSurgery

Cranial (Head & Brain) (SRS, SRT) Extra Cranial (lung, liver, pancreas, spine,

prostate); (SBRT)Cranial Lung Liver Spine

Extra-Cranial

Page 3: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Conventional RT vs SBRTCharacteristic Conventional RT SBRTDose / Fraction 1.8 – 3 Gy 6 – 30 Gy

No. of Fractions 10 – 45 1-5

Treatment time 10-15 min Up to 45min

Planning Margin cm mm

Imaging modalities used for Tx planning CT

Multi-modality: CT/MR/PET-CT

Geometric verification Anatomy Stereotactic system

Staff Training Highest Highest + special SBRT Training

Physics / dosimetry monitoring

Indirect Direct

Page 4: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Systems for SBRT

Linear Accelerator Based SBRT Rapid Arc True Beam Body Frame with linear accelerator

Elekta Varian Pro-Lock System

Dedicated Delivery Systems Gamma Knife Novalis, Brain Lab Tomotherapy Cyber-Knife Proton beam

Page 5: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Delivery Systems

Linear Accelerator Novalis Radiosurgery Cyber Knife

Tomotherapy Gamma Knife Proton Beam

Page 6: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

BrainLab Novalis, SRS/SRT/IMRSBrainLab Novalis, SRS/SRT/IMRS

m3 μMLC

BrainLab Cone: 4mm-20mm diameterNovalis Head rings

Page 7: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Cyber Knife Based SRSCyber Knife Based SRS

Cone Size from 5 mm to 60 mm

• Robotic treatment• IGRT• Intra-treatment imaging • Orthogonal kV imaging

Page 8: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Proton Beam SRS/SBRT

CyclotronSynchrotron

Treatment Schematic Harvard/MGH SRS system

Page 9: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Dose Distributions (Liver SBRT)Dose Distributions (Liver SBRT)

Page 10: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Dose Distributions (Lung SBRT)Dose Distributions (Lung SBRT)

3DCRT or IMRT

Page 11: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Dose DistributionsDose Distributions Spine SBRTSpine SBRT

Page 12: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

RecurrentRecurrent--Spine SBRTSpine SBRT

Page 13: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

SBRT at Indiana University A phase I trial of SBRT for medically inoperable

lung cancer was initiated at Indiana University in the late 1990s By Dr Timmerman

Currently SBRT at Indiana University Lung (primary and metastases) Liver (primary Hepatocellular Carcinoma

HCC and metastases) Majority of our patients have been treated utilizing the

Elekta Stereotactic Body Frame SystemFakiris Fakiris et alet al. Stereotactic body radiation therapy for early stage non. Stereotactic body radiation therapy for early stage non--smallsmall--call lung call lung carcinoma: fourcarcinoma: four--year results of a prospective phase II study. Int. J. Radiat. Onyear results of a prospective phase II study. Int. J. Radiat. Oncol. Biol. col. Biol. Phys. V75(3), p677Phys. V75(3), p677--682, 2009682, 2009Cardenes et al. Phase I feasibility trail of stereotactic body radiation therapy for primary hepatocellular carcinoma, Cin Transl Oncol V12(3), p218-24. 2010

Page 14: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

SBRT High dose of radiation delivered in a hypofractionated

regimen (single fraction or 3-5 fractions)

High target accuracy Stereotactic reference - target is localized relative to

a known 3D coordinate system through the use of external fiducial markers

Patient immobilization is important

Rapid dose falloff (sharp dose gradient) Multiple conformal beams

Page 15: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Elekta Stereotactic Body Frame

1

23

46

5

7

8

Z

X

Y

Page 16: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Different Systems

Small box frame system tightly restricts patient motion

Stereotactic reference system to localize target

Can not be indexed to treatment couch

Respiration compression plate

Modular structures highly adaptable to individual patient

Isocenter setup relies on skin marks/tattoos

Indexed to treatment couch

Respiration compression plate and belt

Elekta Stereotactic body frame CIVCO Body Pro-Lok™ system

Page 17: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

CT Simulation

Planning CT obtained from CT-Simulator 2 mm axial slice through area of interest 4D-CT obtained for motion assessment

Fusion Image obtained with patient in frame Positron Emission Tomography PET (lung

and liver) Dual phase CT (liver)

Page 18: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Erdi et. al. Radiotherapy and Oncology 62 (2002) 51-60

Regular CT Dual phase CT

Multi Modality Imaging

CT PET

Page 19: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

4D-CT

CT Raw Data/Images

Respiration Signal

CT Acquisition Software

Spatial X, Y, Z + Temporal => 4D

Page 20: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Motion Assessment – 4DCT

Page 21: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Treatment Planning

Contour volumes Beam placement and beam modifiers Plan evaluation Isocenter stereotactic coordinate

determination Coordinate verification QA

Page 22: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Treatment Planning – Structure Contour

LUNG LIVERGTV – use lung window GTV Spinal Cord Spinal cordWhole lung Whole liverEsophagus Stomach / DuodenumHeart HeartProximal Bronchial Tree Rt and Lt KidneysTrachea Small BowelBrachial plexus, ipsilateral Rt Lung Descending aorta (optional) Bowel Chestwall/ Rib Chestwall/ RibSkin Skin

Structure contoured based on RTOG protocols or internal guidelines

PTV is created per protocol (normally 0.5cm axial margin and 1cmsuperior-inferior margin around GTV)

Page 23: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department
Page 24: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Possible Beam Arrangements

Page 25: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department
Page 26: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Photon

SBRT, Lung

Proton

Page 27: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

CyberKnife Plans

Page 28: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Plan Evaluation Target coverage Target dose heterogeneity Normal tissue constraints Conformity indices High dose spillage Low dose spillage

RTOG 0236 AAPM TG-101Timmerman R. Semin. Radiat. Oncol. 18: p215-222. 2008

Page 29: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Treatment Planning – Isocenter Coordinates

Xiso = Xmeasure + 85

Yiso = Ymeaasure

Ziso = Zmeaasure+n x 100

Page 30: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Treatment Day Patient setup in frame according to

simulation parameters Isocenter setup by stereotactic

reference system Image acquisition and treatment

verification Online correction through couch

adjustment Treat

Page 31: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Image Guided Treatment Verification

Purpose: to verify patient position on the treatment day is the same as the planned and to provide correction guidance if deviation is discovered

IGRT systems Fixed or rotating kV and MV imagers

(2D imaging) 3D volumetric imaging

Page 32: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Cone Beam CT - Lung Tumor MatchBefore Registration (CBCT and Planning CT blended)

After Registration (CBCT and Planning CT blended)

Page 33: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Cone Beam CT - Liver Tissue MatchBefore Registration (CBCT and Planning CT blended)

After Registration (CBCT and Planning CT blended)

Page 34: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Before Registration (CBCT and Planning CT blended)

After Registration (CBCT and Planning CT blended)

Cone Beam CT – Bony Anatomy Verify

Page 35: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Comparison between 2 systems Compared with Elekta body frame, the Pro-lok system

is relatively complex and time consuming in patient simulation and treatment setup

Relatively larger random setup errors were found in Pro-lok system for both SBRT liver and lung treatment setup

Large inter-fractional correction occurs more frequently with Pro-lok system

Image guidance is critical for SBRT patient setup, especially when Pro-lok system is used

Page 36: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Histograms of inter-fractional couch corrections of liver patient setup

0

5

10

15

20

25

0 0.2 0.4 0.6 0.8 1

couch correction [cm]

freq

uenc

y [%

]

VrtLngLat

0

5

10

15

20

25

30

35

40

0 0.2 0.4 0.6 0.8 1

couch correction [cm]fr

eque

ncy

[%]

VrtLngLat

Pro-Lok System Elekta System

For large couch corrections, repeated CBCTs are often performed to verify the correction. For Pro-lok system, 40.7% of all the fractions ended up with repeated CBCT. Only 13.6% of fractions have repeated CBCT when Elekta body frame was used

Page 37: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Results of Two Systems

-0.6

-0.4

-0.2

0.0

0.2

0.4

0.6

0.8

Vrt Lng Lat

Syst

emat

ic e

rror

(cm

)

ProLokElekta

0

0.1

0.2

0.3

0.4

0.5

0.6

Vrt Lng Lat

Ran

dom

err

or (c

m)

ProLokElekta

-0.6

-0.4

-0.2

0.0

0.2

0.4

0.6

0.8

Vrt Lng Lat

Syst

emat

ic e

rror

(cm

)

ProLokElekta

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

Vrt Lng Lat

Ran

dom

err

or (c

m)

ProLokElekta

Liver

Lung

Systematic Setup Error Random Setup Error

Page 38: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Ding et al, Med Phys, 40(5), 051705, 2013

Linac vs CyberKnife

Lung Liver

LinacCyberKnife

Page 39: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

PTV Prescription Gamma Knife 50% CyberKnife 50-70%

Lung (40-48% ) Ding et al, 2013

Liver (67-77%) Ding et al, 2013

Novalis 70-80% Accelerator 80% 3DCRT 95-100% IMRT 100%

Page 40: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Small Field Dosimetry

rel

relf

Qf

Q

fQ

fQff

QQ adingOutput

MDMD

k msrclin

msrclin )(Re)(

//

clin

msr

clin

msr

clin

clin

msr

clin

,w,w

,w,w,,

refmsr

msr

msr

msr

msr

msr

,,,,,

ffQQQoQQoDW

fQ

fQw kkNMD

msrclin

msrclinmsr

msr

clin

clin

msr

msr

msr

msr

clin

clin

clin

clin

msr

msr

clin

clinmsrclin

msrclin

,,

,w

,w

// ff

QQfQ

fQ

fQ

fQ

fQ

fQ

fQ

fQff

QQ kMM

MDMD

MM

msrfmsrairw

fclinfclinairwffQQ PS

PSk msrclin

msrclin

)()(

,

,,,

Page 41: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Why So Much of Fuss?

Reference (ref) conditions cannot be achieved for most SRS devices (cyberknife, gammaknife, tomotherapy etc)

Machine Specific reference (msr) needs to be linked to ref Ratio of reading (PDD, TMR, Output etc) is not the same

as ratio of dose

2

1

2

1

MM

DD

msrclin

msrclin

ffQQk

MM

DD ,

,2

1

2

1

Page 42: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Field Size (cm)

Rel

ativ

e do

se a

t dm

ax

Impact of msrclin

msrclin

,,ffQQk

Page 43: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

What is Common Theme? Multimodality imaging for target delineation Very High dose (20-60 Gy) in few fractions (1-5) Stereotactic immobilization / CBCT/ image based High degree of precision (mm) needed Small Fields

High degree of precision in dosimetry needed Accurate commissioning data Proper selection of detector

Repeated imaging for localization, pretreatment and during treatment

Page 44: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Selection of a SBRT Machine

Hard to compare Advertising for personal gain Only anecdotal and dosimetric data No scientific data on clinical outcome No randomization possible in this era

Page 45: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department

Summary SRS/SRT/SBRT/IMRS is now widely used for hypo-

fractionated treatment world wide with good outcome Image guidance for treatment planning and treatment

verification is very critical to the success of SBRT radiation therapy

Quality assurance and Small Field dosimetry as well as continuous monitoring of the QA and process is critical

For SBRT, image-guided localization techniques shall be used to guarantee the spatial accuracy of the derived dose distribution

AAPM has Task Group report on every modalities that can be referred

There is no superiority of any device that can be recommended

Page 46: Linac or Non-Linac Demystifying And Decoding The Physics ...chapter.aapm.org/pennohio/2013FallSympPresentations/SI9_Indra_Das.… · Indra J. Das, PhD, FAAPM, FACR, FASTRO Department