Laparoscopic Placement of the BardPort Intraperitoneal Catheter and Reservoir
Dr. Arlan F. Fuller, Jr.Gillette Center for Women's Oncology
Massachusetts General Hospital
Page 2
Choice of Catheter and Reservoir is the IV Device
Page 3
Titanium Port with 9.6F attached catheter
Page 4
Pre-placement Procedure
Perform laparoscopy in the usual manner with umbilical placement of laparoscope trocar and survey of the abdominal cavity for adhesions
Place a 5 mm operative trocar in a position contralateral to the side of proposed catheter placement
Identify the prospective site in the lower quadrant for entry of the catheter into the peritoneal cavity
Page 5
Mark the site for reservoir placement above the right costal
margin
Page 6
Incise the skin and develop a pocket for the
reservoir
Page 7
Incise the skin and develop a pocket for the
reservoir
Page 8
Check that the reservoir fits snugly into the pocket
Page 9
Place the tunneler on the end of the catheter and start
the passage deep to Scarpa’s fascia
Page 10
Place a 5 mm transverse incision at the target site
Page 11
Bluntly with a hemostat extend the incision beneath Scarpa’s fascia to contact
the tunneler
Page 12
Pass the tunneler out through the target
incision
Page 13
Pass the tunneler out through the target
incision
Page 14
Pass the catheter through the tunnel and the target incision
and cut catheter to remove tunneler
Page 15
Suture the reservoir in place to fascia and confirm catheter
function
Page 16
Place introducer needle through target incision adjacent to
catheter
Page 17
Visualize needle tip at the target site in the peritoneal
cavity
Page 18
Pass the guidewire with “J” tip through the introducer needle
Page 19
Pass the guidewire with “J” tip through the introducer needle
Page 20
Visualize “J” tip of the wire at the target site in the peritoneal
cavity
Page 21
Withdraw the introducer needle leaving the guidewire in place
Page 22
Insert the introducer with the 10F dilator over the guidewire
into the peritoneal cavity
Page 23
Visualize the dilator AND the peel-away introducer in the peritoneal
cavity
Dilator
Dilator
Introducer
Page 24
Withdraw both the wire and the dilator from the
introducer
Page 25
Pass the catheter through the introducer and assist
laparoscopically in advancing the catheter into the peritoneal
cavity
Page 26
Peel away the introducer while holding the catheter in
the peritoneal cavity
Page 27
Advance all the remaining catheter into the peritoneal
cavity
Page 28
Adjust position of catheter tip and confirm function by
injection
Last, subcuticular closure of all incisions