K3 & K3R seriesBenchtop Centrifuges
Instruction Manual
Download from www.centurionscientific.co.ukclick ‘K3 Series’
click ‘Manual Instruction Technical’
Deceleration Rate Set
Key to Displays(See pages 5-9 of this manual for programming instructions - a larger copy of this KEY is included in an appendix)
Program
Last time countdown
Open Lid
Last running RCF
Last RCFt
Start
Screen when runningStart Screen
Time counting downRCFt counting up
RCF actual
User name
Time Set
RCFtNot set
RPM orRCF Set
Programmed Rotor Set
Acceleration Rate Set
GENERAL SAFETYTo ensure that the centrifuge is operated in a safe manner, avoiding the hazard of injury to personnel or damage to material goods, the following safety precautions should be followed at all times: •Thecentrifugemustonlybeopenedbytrainedpersonnelcompetentinitsuse •Neverusethecentrifugeunlesstherotorisproperlymountedandsecure •Neveropen,orattempttoopenthelidwhilethecentrifugerotorisstillturning. •Onlyuseoriginalpartsforthecentrifuge •Neveroperatethecentrifugeifcomponentsaremissingordamaged. •Iftherotorshowsvisiblesignsofwearitsusemustbediscontinuedandreplacementfitted. •Therotormustnotbeoverloaded,themaximumalloweddensityofsamplesatfulloperatingspeed is 1.2g/cm3. •Alwaysendeavourtomakesurethatoppositetubesareofthesameweighttoavoidrotorimbalance.Ifthetubesarefilledwiththesamematerialtheymust,therefore,befilledwiththesame amount. •Nochangesshouldbemadetothemechanicalorelectricalcomponentswithoutpriorconsul-tation with and the written permission of Centurion Scientific. •Thecentrifuge’shasnotbeenmanufacturedfrominertmaterialsortobeexplosionproof.Ensurethatitisnotoperatedwithinanenvironment,orwithmaterialsthatmakesthesearequirement. •Duringoperationa“safe”zoneof30cmmustbemaintainedaroundthecentrifuge.Thiszoneshouldbeclearofpersonnelandhazardousmaterialsatalltimesduringtherun. •Substancesofaradioactive,flammableorexplosivenaturemustnotbecentrifuged. •Substancespronetoreactbrisklywitheachothermustnotbecentrifugedatthesametime. •Toxinsandpathogenicmicro-organismsmustnotbecentrifugedunlesstheproperprecau-tionsfortheirhandlinghavebeentaken.Thesemayinclude,butarenotlimitedto,biologicalseals.Should toxins or pathogens enter the centrifuge or its parts the proper procedures for disinfecting thecentrifugeshouldbecarriedout. •Stronglycorrosivesubstancesthatmaycausedamagetoorimpairthemechanicalstrengthoftherotormayonlybecentrifugedinsideprotectivevessels.Shouldanydoubtexistregardingthesubstanceorthesuitabilityofaparticularvesselforusewithit,consulttheSafetyInspector.
Programmed Rotor Set
Acceleration Rate Set
Deceleration Rate Set
Time Set
RCFtNot set
RCF(g)Set
Stop
User name
ActualTemperature(Refrigerated only)
Temp Set(Refrigerated only)
Temp Set(Refrigerated only)
ActualTemperature(Refrigerated only)
Pre-cool /pre-run(Refrigerated only)
2
Centurion Scientific K3 series instructions PAGE2
INSTALLATION
Theinstrumenthasbeendesignedinsuchawaythatinstallationshouldpresentnoproblemstoatechnicianfamiliarwithequipment of this type.ShouldanyproblemsofquestionsariseduringtheinstallationoftheinstrumentpleaserefertoCenturionScientificforadvice.Duringinstallationtherecordfortheindividualcentrifuge(tobefoundwithourmanual)mustbecompleted.
WORK surfaceTheworksurfacemustbestableand,wherepossible,resonancefree–astandardlaboratorybenchisusuallysuitable.Thelocationofthecentrifugeshouldbewellventilatedandoutofdirectsunlight,havingsufficientroomtoallowairtocirculatearounditfreely.Asafetyzoneextendingfor30cminalldirectionsmustbemaintainedaroundthecentrifuge.Thisspaceshouldbekeptclear,inparticularnohazardousmaterialmaybeplaceintheareaduringcentrifugation.
MAINS connectionBeforeconnectingtheunittothemains,ensurethatthevoltageandfrequencyofsupplymatchthespecificationsontheinstrument.Amainsplug,providedwitha5ampfuse(or13ampforrefrigeratedunits)andofatypesuitableforusewherethecentrifugeistobeinstalled,shouldbefitted.Thisplugmustbewiredupaccordingtotheinternationalconvention.
ROTOR assemblyTherotorassemblylocatesontothetaperedshaftofthedriveshaft,beingheldsecurelyinpositionbyaflatwasherandlockingscrew.Itisimportantthattheseitemsareassembledinthecorrectorderandthatthewasherisusedasintended.Tolocatetherotorfirst,checkthatboththeexteriorsurfaceoftheshaftandthetaperedboreoftherotoritselfarefreedirtoranyforeigndebris.Therotorshouldthenbeloweredintopositiononthetaperedshaft,carebeingtakentoensurethatitseatsfirmlyandevenly.Theflatwasherfollowedbythescrewisthenplacedintothethreadedportionoftheshaft,theassemblythenbeingsecuredbythescrew.Thescrewshouldbehandtightenedwiththekeyprovided.
Rotor screw fixing Rotor bolt fixing
3
Centurion Scientific K3 series instructions PAGE3
OPERATIONFILLING and loading of sample tubes.Thecentrifugefunctionsinthemostefficientmannerwhenunbalanceoftherotorisminimised.Thispreventsunduevibrationperturbingtheseparationzonesofthesubstancebeingcentrifuged.Itisthereforeimportantthatunbalanceduetounequalloadingofthesampletubesisminimisedasfaraspossible.Themaximumpermissibleimbalanceoftherotoris1g.Imbalancemaybeminimisedbyfollowingafewsimpleandquickproceduresasdescribedinthefollowingparagraphs.•Thesampletubesshouldbefilledasevenlyaspossible.Wherepractical,tubescontainingthesamesubstanceshouldbeplacedoppositeeachother.•Therotorshouldbeloadedsymmetrically,thisbeingbestachievedbyusingallofthebores.Whenonlypartiallyloadingtherotoritmustbeensuredthatoppositeboresalwaysreceivetubesofthesameweight.Thefollowingpicturesprovideexamplesofcorrectandincorrectloadingoftherotor.
EMERGENCY lid release in case of lid lock failureIntheunlikelyeventofthelidlockfailingandtheuserbeingunabletoremovetheirsamples,anemergencylidreleaseisprovided.WARNING:OPERATIONOFTHEEMERGENCYLIDRELEASEMUSTONLYBEUNDERTAKENBYQUALIFIEDANDAPPROVEDPERSONNEL.UNDERNOCIRCUMSTANCESMAYTHECENTRIFUGEBEOPERATEDINTHECONDITION.Waituntiltherotorhasstoppedturning(viewthroughthelidport)andswitchoffthepowertothecentrifuge.RemovetheblackplasticsealstofrontundersideofcentrifugeandPULLBOTHCORDSDOWN.WARNING:UNDERNOCIRCUMSTANCESMUSTTHEEMERGENCYLIDRELEASEBEUSEDWHILETHECENTRIFUGEISRUNNING.
Cleaning, Disinfecting and MaintenanceWARNING:ISOLATEANDDISCONNECTTHEPOWERSUPPLYBEFORECARRYINGOUTCLEANINGORMAINTENANCE.
CENTRIFUGE BODY•Regularlywipedowntheexternalsurfaceswithamilddetergent.•Mopupanyspillagewithanabsorbentcloth.Checktoensurethattheliquidhasnotenteredthemainbodyofthecentrifuge.WARNING:IfanyliquidhasenteredthemainbodyofthecentrifugeitMUSTNOTbeused.ContactthesupplierorCenturionScientificLtdimmediatelyforadvice.
ROTOR•Removethelockingscrewwiththekeyprovided.•Removetheflatwasher.•Pulltherotorofftheshaftbyapplyingequalforceofbothsides.NOTE:Ifanystiffnessordifficultyisexperiencedwhenattemptingtoremovetherotorthismaybeovercomebyapplyingasharptaptotherotorshaft.Anon-metallicobject(e.g.plasticmallet)mustbeusedtopreventdamagetotheshaft.•Cleantherotorusingasuitabledetergentandhotwater.Alternativelytherotormaybecleaninginanautoclaveatamaximumtemperatureof120oC•Thoroughlyrinseanddrytherotor.•Inspecttherotorforanyhairlinecracks,damageorotherdefects.IfanyarefoundpleasecontactyoursupplierorCenturionScientificLtd.
CORRECT INCORRECT
4
Centurion Scientific K3 series instructions PAGE4
Start point
Display shown at turn on
PRESS program key(bookemblemtopright)
Display shown when program key pressed
PRESS Page number rectangle
Display shown when Page key pressed
SELECT page number required1-9 pages with 12 names per page (a totalof108programscanbestored)
PRESS Set Name KEY in name requiredUpto14characters,letters&numbers(wehavechosenBiochem8fortheseillustrations)
ROTOR PROGRAMMING
Display shown after pressing Tick on name settings
PRESS Set Parameters
Display shown after pressing Set Parameters
PRESS Rotor rectangle
Display shown after pressing Rotor rectangle
FIRST build a program
Display shown when Set Name pressed
SELECT rotor required by pressing down and releasing
continued...
PRESS tick if correct
Display shown when chosen name is being entered
5
Centurion Scientific K3 series instructions PAGE5
Display shown when Time Numerical rectangle is selected
Display shown after selecting rotor
Correct rotor should be shown in Rotor display
Turn end of run buzzer off by PRESSING Bz(Top right of display)
- Turn ON Shows white square in centre
TIME PROGRAMMING
Display shown
PRESS Time rectangle
Display shown when timer rectangle key pressed
RPM PROGRAMMING
Display shown
PRESS RPM square
Display shown when RPM pressed
SELECT RPM (speed) or RCF (G Force)
Display shown after selecting RPM
PRESS Numerical rectangle next to box
SELECT and PRESS desired timer type:
Time 0-9999minutes
Hold Timer is held on till stop buttonispressed
RCFt SeeW2t document for explanation
Pulse Timer counts up in seconds whenSTARTbuttonheld down.(90secondsmax)
PRESS Rectangle next to Time box
SELECT Time required in minutes & secondsHoldandPulsecannotbechangedOrselectRCFtnumberfrompreviousrun
PRESS Tickafter selection
continued...
Centurion Scientific K3 series instructions PAGE6
6
Display shown after pressing numerical rectangle
SET Speed (RPM) or RCF in numerical form
ACCELERATION and DECELERATION ratesSET Acceleration rate
SELECT Rate 1-10(1isslowest10fastest)
Select rate 1-10(1isoff10fullbraking)
Display shown when Acc pressed
PRESS number required & release
Display after selecting AccSET Deceleration (Brake) rate
PRESS Tick after selection.
Display
If you are satisfied with all parametersCheck Rotor, timer type, RPM or RCFAcc & Dec rates
Display
PRESS number required & release
Display shown when Dec pressed
Then PRESS Tick box (bottomright)
PRESS Acc Square
PRESS Dec Square
Display after selecting Acc and Dec rates
forREFRIGERATED MODELS onlySET Temperature - SEE following page 7A, thenreturntosetAccelerationandDecelerationratesbelow
Centurion Scientific K3 series instructions PAGE7
7continued page 8...
7A
Refrigerated Centrifuges onlyK2015R,K241R and K243RCenturionScientificrefrigeratedcentrifugeshavearangeof- 9oC to 40oC.
Weaccuratelycontrolthistemperaturebyrunningtherefrigerationunitandpulsingheatviaaloopclosetothecoils.PID system is used to control.
Imagine a bath at home, to control the temperature accurately you would run both hot and cold taps together. This is far more effective than filling with cold water and topping up with hot as the temperature difference would be enormous.
Byusingthismethodourchambertemperaturecontrolisunsurpassed.
NoteDuetotheefficientmanagementoftemperaturecontrolyoumayseesomeicepatchesorinturn,noticesomeslightlyhotterareas.Thisisquitenormalanditistheoverallchambertemperaturethatmatters.
ForaccuratetemperaturecontrolwerecommendyouusethePre-cool/pre-runfacility .FityourdesiredrotorandsecurelyfastenasperourInstructions.Afterprogrammingyourdesiredparameters,IEselectionofarotor,speed,time,acceleration&decelerationrates,(asanexample).Leavethecentrifugetoreachthedesiredtemperatureandthenpressthe Pre-coolsymbol.Thiswillrunthecentrifugeat20%ofthemaximumspeedoftherotorfor20minutes,a willshowintheRpmsectiontoshowthePre-coolmodeisinoperation.Itisadvisabletoletthetimerunto20minutes,butyoucanstopatanytime.Oncestoppedloadyoursamples(whichshouldbeatyoursettemperature)andpressthestartarrowbutton.
Foreaseitisadvisabletosetyoursamplestothedesiredcentrifugetemperaturebeforeplacinginthecentrifugeandseparating,otherwiseallowsufficienttimeforthemtoreachthetemperatureinthechamber.
TEMPERATURE PROGRAMMING(Refigeratedmodelsonly-K2015R,K241RandK243R)
Display shown K2015R,K241R and K243R(all our illustrated screens show a chosen and actual temperature of 4oC)
PRESS Temperature numeric square
Display shown when Temp is pressed
SET Temp (-9oC...40oC) in numerical form
return to page 7 to set Acceleration and Deceleration rates...
Centurion Scientific K3 series instructions PAGE7A-applicabletoRefigeratedModelsOnly-
PRESS Tick after selection.
Note: See rotor appendix for minimum possible temperature at maximum speed per rotor.
Note: Displays illustrated elswhere in this manual do not show the Temp square or the Temperature numeric square however they will be present on all refigerated model displays as will set temperature and actual temperature where appropriate (see Key to Displays on page 1)
Pre-cool /pre-run START button.Seeinstructionsabovedetailinguse of this feature - note that this screen will show your selected parameters.
Note:Centurion K3 range centrifuges can storeupto108programs,theseareconvienientlygroupedinto9pageswithupto12namesavailableperpage,eachnamecanconsistofupto14alphanumericcharacters.
HOW TO RETRIEVE SET PROGRAMS
NOW LOAD THE CENTRIFUGE
Display
PRESS Central button on right to open lidLOAD your samples evenly(seeSafetyInstructions,page2)
FIT LID to rotor (if part of rotor type)CLOSE Centrifuge lid pressing both sides down firmly.PRESS Start button (arrowtriangle,bottomright)
Display showing running screen
RUNNING screen displayIf you require early terminationTime,Hold,RCFtPRESSSquarebottomright.
OR in pulse mode RELEASEwhencounter(seconds)hasreacheddesirednumber
Run Complete
Display showing screen when run finished (with Buzzer selected)
PRESS run complete Tick(onlyshowsIFBuzzerisselected)
PRESS lid open button(centre right)
You may now retrieve your samples
Display
PRESS program button(top right)
PRESS Pagenumerical box(9 pages available)
PRESS & SELECTrequired name then
PRESS desired name(12 names availble)
your stored program will now be retrieved.
Centurion Scientific K3 series instructions PAGE8
8
If Display shows Rotor out of balanceAnimbalancehasoccurred
If Display shows Tacho failurePRESS Tick buttonPRESS Lid open buttonCheck rotor for free rotationArethetubestootall?
CALL service department
If Display showsLid Open!Check lid is closed correctly
To check sensors go to parameters page.FirstmakesureCentrifugelidisopen.Press‘di’(toprightsmallbutton)forDiagnosticsscreen.(Shown below)Itreadsasfollows: (either 0 or 1 may show when this screen first appears, follow instructions below to perform diagnostic checks)
When centrifuge has stopped PRESS tick buttonPRESS Lid open buttonCheckfor:Improperlyloadedtubes(see Safety pages)Brokentubesorincorrectlymeasuredtube
Error code shown on display
Error code shown on display
ERROR CODES ON DISPLAYCenturion Scientific K3 series instructions PAGE9
PRESS down firmly on lid (atbothsidesnearfront)
IfnotthenCHECKSENSORasbelowOrCALLServiceDepartment
.... or greater (version number of installed firmware)
.....0or1 .....ByrotatingrotorTachowillshow0or1asrotorturnsifsoTachometerisworking
.....0or1 .....ByrotatingrotorRotorRecwillshow0or1asrotorturnsifsosensorisworking
.....0or1 .....BypressingrotorfirmlyorshakinghardBalancewillshow0or1
.....0or1 .....ByshuttingthelidtheLidswitchwillgofrom0to1showingcorrectsensing
Lid Open! Error code shown on display
CenturionScientificLtd-AllCenturionequipmentisdesignedandmanufacturedunderstrictQualityassuranceconditionsandrigoroustestingandthereforeshouldprovideyouwithyearsoftroublefreeservice.However,shouldyourequirerepairsormaintenanceworkonyourCentrifugepleasecontactourservicedepartmenton+44(0)2392631225.FornoneUKcustomers,pleasecontactyourlocalCenturiondistributorforservice,maintenanceandaftersalessupport.
9
Centurion Scientific K3 series instructions Appendix
Key to Displays(Seepreviouspagesforprogramminginstructions)
Time Set
Programmed Rotor Set
RCFt(Not set)
RPMor
RCFSet
Acceleration Rate Set
Deceleration Rate Set
User name Program
Last timecountdown
Open Lid
Last running RPM
Start
Last RCFt
Start Screen
Screen when running
Time counting down
RCFt counting up
RCF actual
Programmed Rotor Set
Acceleration Rate Set
Deceleration Rate Set
Time Set
RCFt(Not set)
RCF(g)Set
Stop
User name
ActualTemperature(Refrigeratedonly)Temp Set
(Refrigeratedonly)
Time Set
Programmed Rotor Set
RCFt(Not set)
RPMor
RCFSet
Acceleration Rate Set
Deceleration Rate Set
User name Program
Last timecountdown
Open Lid
Last running RPM
Start
Last RCFt
Start ScreenRefrigerated models only
ActualTemperature(Refrigeratedonly)
Temp Set(Refrigerated
only)