InVue® BT128 RefractometerProcess monitoring for laboratory fluid applications in a benchtop package
ADVANCED MATERIALS HANDLING | USER GUIDE
3
INVUE BT128 REFRACTOMETER
User Guide | Entegris, Inc
TABLE OF CONTENTS—Cautions and Warnings ............................................... 4
Safe Disposal ................................................................. 4
Overview ........................................................................ 4
Work Flow ...................................................................... 4
Unpacking and Hardware Setup ............................... 4
User-Supplied Tools and Equipment ..................... 4
Unpacking .................................................................... 5
Installing BT128 ........................................................... 5
Connecting Power ..................................................... 5
Installing the BT128 Software ..................................... 5
Hardware Device Driver Download ....................... 5
Software Download and Installation ...................... 5
Unit Operation .............................................................. 6
Start Up ......................................................................... 6
Log In to Software ...................................................... 6
Set Up Measurement File (Optional) ...................... 6
Start a Measurement .................................................. 6
Cleaning Sample Chamber ...................................... 7
Reinitializing (Advanced) ........................................... 7
Appendix ........................................................................ 8
Troubleshooting .......................................................... 8
Specifications .............................................................. 9
Dimensions ................................................................ 10
Technical Support ....................................................... 11
4
INVUE BT128 REFRACTOMETER
User Guide | Entegris, Inc
CAUTIONS AND WARNINGS—Before running the InVue® BT128 refractometer, heed
the following:
CAUTION: The InVue BT128 is a sensitive electronic device. Rough handling may cause damage. Do not drop.
WARNING:
• The supplied USB isolator needs to be installed between the BT128 and a computer. This will eliminate potential damage to the BT128 from ground loops.
• Touching live electrical parts can cause fatal shocks and severe burns. Incorrectly installed or improperly grounded equipment is a hazard. Do not touch live electrical parts.
• Properly ground the system before use. Plug the power supply into a properly wired receptacle.
• Follow local electrical codes and the guidelines when installing the BT128. Failure to do so may create an electrical shock hazard. Shock hazards can exist even when equipment is properly installed. The operator should be properly trained and follow established safety practices.
SAFE DISPOSAL—When disposing of any component of the BT128
refractometer, observe the local and national
requirements for the disposal of electrical and
electronic equipment.
OVERVIEW—The Entegris InVue BT128 refractometer is an opti-
cally-based, refractive index metrology solution that
provides exacting and immediate results for single-
sample concentration measurements of solutions
in a benchtop setting. The BT128 (and the included
software) enable user-friendly, real-time information
and can also be utilized to determine process fluid
calibration for use with the Entegris in-line monitoring
devices. Utilizing the index of refraction technology
(IoR) of the BT128 provides a measurement solution
that is not affected by turbidity, diffraction and absor-
ption of light in the sample, or a solution's acidic or
electric properties.
WORKFLOW—
UNPACKING AND HARDWARE SETUP—
USER-SUPPLIED TOOLS AND EQUIPMENT
Supplies required for operation
EQUIPMENT NOTES
Computer Windows® 7 or higher with .Net 4.0 framework
AC power Provided AC-DC power supply requires 100-240 VAC, 47 – 63 Hz input
No-residue cleaning solution
For cleaning sample chamber. Sample dependent — IPA works well for many sample types
Lint-free wipes For cleaning sample chamber
DI water (optional) Used for reinitialization for refractive index reference
Unpacking and hardware
setup
Install 128-connect
software
Operation
Appendix
• User supplied tools and equipment• Unpacking• Installing the BT128• Connecting power
• Start up• Log in to software• Set up measurement file (optional)• Start a measurement• Cleaning sample chamber• Reinitialization
• Troubleshooting• Specifications• Dimensional considerations
• Hardware device driver download• Software download and installation
HANDLEWITH CARE
INVUE BT128 REFRACTOMETER
5User Guide | Entegris, Inc
UNPACKING
• Unpack all components of the BT128
• Ensure the following components are present:
– BT128 unit
– Power supply
– Software
– RS485 isolator
– USB cables (2)
INSTALLING BT128
The sensor is a table-top module and can be
installed on any flat, even surface with the
required available space.
CONNECTING POWER
• Connect the RS485 isolator and USB cables as
shown below.
CRITICAL: The isolator (tan box) helps prevent a ground loop to the unit. It must be connected to prevent damage to the BT128 unit.
• Plug power supply in to BT128 unit.
• Turn the unit on using switch on back of device.
INSTALLING THE BT128 SOFTWARE—
HARDWARE DEVICE DRIVER DOWNLOAD
Communication between the BT128 unit and
128-connect software is established via a RS485
to USB converted connection.
IMPORTANT: Device drivers must be set up or communication issues may arise.
• Power on the BT128.
• Plug the USB cable in to the PC. Wait until Windows
completes the driver installations.
NOTE: If the PC does not have the appropriate drivers, Windows will attempt to download automatically from the internet. If the computer does not have internet ac- cess, the two separate drivers can be downloaded from: http://www.ftdichip.com/Drivers/CDM/CDM21218 _Setup.zip http://accesio.com/files/packages/USB-COM-2SM %20Install.exe
SOFTWARE DOWNLOAD AND INSTALLATION
• Download the software:
– Run the 128-connect installation program found
on www.entegris.com (go to the InVue BT128
product page where you will find a link to access
the software).
NOTE: Prior to downloading the software for the first time, a user profile needs to be completed and approved.
• Run the installer:
– Follow the on-screen instructions for installation
of the software.
– The installer program outputs an icon for “InVue®
BT128 Connect” on the desktop and creates a
default data storage location on your computer’s
hard drive.
6
INVUE BT128 REFRACTOMETER
User Guide | Entegris, Inc
UNIT OPERATION—
START UP
• If not on already, power on the BT128.
• Start the BT128 software by clicking on the
“InVue® BT128 Connect” icon.
• The software will now perform a 15 minute warmup
period of the BT128.
LOG IN TO SOFTWARE
• The software is password protected for various
levels of users.
• Log in to the software by clicking the LOGIN button.
• Use one of the following passwords:
– Operator mode: testop
Allows for data logging and measurement.
– Supervisor mode: testsuper
Adds the ability to recalibrate and reinitialize
the sensor.
SET UP MEASUREMENT FILE (OPTIONAL)
Users can set up a .csv measurement file to log
measurement results.
• To set up a measurement file:
– Click the MENU button.
– Select CREATE MEASUREMENT FILE from the list.
– In the resulting file dialog pop-up, navigate to a
preferred file location and name the file.
• To view the folder location of the measurement file:
– Click the MENU button.
– Select OPEN MEASUREMENT FILE FOLDER from
the list.
START A MEASUREMENT
• After logging, the BT128 is ready for sample
measurements.
• Inject sample to be measured:
– Open the sample chamber lid.
– Sample volume is 1 to 2.5 mL.
IMPORTANT: Ensure sample covers the entire optical window.
– Close the sample chamber lid.
INVUE BT128 REFRACTOMETER
7User Guide | Entegris, Inc
• Start Measurement
– Click the START MEASUREMENT button in
the software.
– Assign a Sample ID
Enter the name of the sample to be measured.
• The BT128 will now proceed with the measurement
and display the results upon completion.
CLEANING SAMPLE CHAMBER
Once measurement is complete, the sample chamber
will need to be cleaned.
• Cleaning sample chamber:
– Using appropriate safety precautions, clear the
sample from the sample chamber using a
lint-free wipe.
– Clean the sample chamber using a lint-free wipe
and a zero-residue cleaning solution.
IMPORTANT: Ensure the optical window in the sample chamber is clean and free of any debris or watermarks.
REINITIALIZING (ADVANCED)
Occasionally, the unit may need to be reinitialized to
correct for any errors.
NOTE: DI H20 will be needed during this procedure as a Refractive Index reference.
• Prepare for reinitialization:
– Clean and dry the sample chamber.
IMPORTANT: Ensure the optical window in the sample chamber is clean and free of any debris or watermarks.
• BT128 Reinitialization Wizard:
– Log in as Supervisor (see Log in to Software
section).
– Click the MENU button.
– Select REINITIALIZE SENSOR from the list.
– Follow the on-screen wizard to complete
reinitialization process.
8
INVUE BT128 REFRACTOMETER
User Guide | Entegris, Inc
APPENDIX—
TROUBLESHOOTING
SYMPTOM PROBABLE CAUSE REMEDY
Unable to take measurement Not logged in Log in as Operator or Supervisor
Unable to communicate with unit
Device drivers not installed Ensure that device drivers are installed and operating correctly
Temporary communication issue Repeat measurement
Restart 128-connect software
Disconnect and reconnect USB cable
Baud rate is not correct In device manger, ensure that the Comm Port baud rate is 9600
BT128 does not have power Ensure the BT128 is connected to the power cord and device switch is on
“Out of Range” error Sample is out of the BT128 measurement range
Ensure liquid is between 1.33000 and 1.40000
Sample volume is too small Ensure the sample covers the entire optical window
“Sensor in Air” error No liquid in sample chamber Ensure 1 mL of liquid is in sample chamber
Erratic measurements Sensor malfunction Reinitialize sensor (see Reinitialization). If problem persists, consult the factory.
Deionized water (DIW) check is not displaying 1.33300 ±0.00090 nD
Sensor drift Reinitialize sensor (see Reinitialization). If problem persists, consult the factory.
INVUE BT128 REFRACTOMETER
9User Guide | Entegris, Inc
SPECIFICATIONS
Functional characteristics
RI range 1.330 – 1.400 nD at room temperature
Sample volume 1.0 – 2.5 mL
RI accuracy ±9.0 × 10e-4 nD at room temperature
RI repeatability 7.0 × 10e-4 nD at room temperature
Sample fluid temperature 4° – 40°C (39° – 104°F)
Ambient temperature 18° – 25°C (64° – 77°F)
Ambient humidity 20 to 30% RH
Storage temperature -15° – 40°C (5° – 104°F)
Control temperature System controls sample to 20°C ±0.5° (68°F ±33°)
Current rating 1.6 amp AC max
Voltage input Provided AC-DC power supply requires 100 – 240 VAC, 47 – 63 Hz input
System start-up time From power on, within 15 minutes
System response time From cycle start, within 5 minutes to display result
Physical characteristics
Wetted materials (bowl) 316L SST, sapphire, PTFE
Weight 3.18 kg (7 lbs)
Dimensions – lid closed 216 mm W × 216 mm D × 116 mm H (8.50" W × 8.50" D × 4.58" H)
Dimensions – lid open Height extends to 235 mm (9.25") when open
Power supply cable length 2.74 meters (9.0')
USB cable length 1.83 meters (6.0')
Communications port USB type B
Power connector 4-pin power DIN
Serial com RS485
USB adapter USB serial converter provided
10
INVUE BT128 REFRACTOMETER
User Guide | Entegris, Inc
DIMENSIONS
216 mm(8.50 ±0.01”)
235 mm(9.25 ±0.09”)
200 mm(7.88”)
116 mm(4.58 ±0.06”)
104 mm(4.08 ±0.06”)
216 mm (8.50 ±0.01”)
Top View Side View
Back View
INVUE BT128 REFRACTOMETER
11User Guide | Entegris, Inc
TECHNICAL SUPPORT—
REGIONAL CUSTOMER SERVICE CENTER NUMBERS
Region Telephone Fax
North America 1 800 394 4083 1 800 763 5820
Germany +49 351 795 97 0 +49 351 795 97 499
France +33 4 76 35 73 50 +33 4 76 35 73 80
Israel +972 73 221 00 00 +972 73 221 00 22
Japan +81 3 5442 9718 +81 3 5442 9738
Malaysia +60 4 427 4200 +60 4 641 3294
Korea +82 31 8065 8300 +82 31 8065 8301
Taiwan +866 3 571 0178 +866 3 572 9520
Singapore +65 6745 2422 +65 6745 4477
China +86 21 8023 6500 +65 21 5080 5598
CORPORATE HEADQUARTERS
Entegris, Inc.
Corporate Headquarters
129 Concord Road
Billerica, MA 01821 USA
Tel. +1 978 436 6500
Fax +1 978 436 6735
For product information or quote request, email [email protected].
129 Concord RoadBillerica, MA 01821 USA
Tel +1 952 556 4181Fax +1 952 556 8022Toll Free 800 394 4083
Corporate Headquarters Customer Service
LIMITED WARRANTY
Entegris' products are subject to the Entegris, Inc. General Limited Warranty. To view and print this information, visit entegris.com and select the Legal & Trademark Notices link in the footer. Entegris does not warranty any failure in the case of customers using unapproved foreign components.
FOR MORE INFORMATION
Please call your Regional Customer Service Center today to learn what Entegris can do for you. Visit entegris.com and select the Contact Us link to find the customer service center nearest you.
TERMS AND CONDITIONS OF SALE
All purchases are subject to Entegris’ Terms and Conditions of Sale. To view and print this information, visit entegris.com and select the Terms & Conditions link in the footer.
www.entegris.comUser Guide | Entegris, Inc
Entegris®, the Entegris Rings Design™ and InVue® are trademarks of Entegris, Inc.
©2017 Entegris, Inc. | All rights reserved. | Printed in the USA | 3220-8157ENT-0517