71
International Atomic Energy Agency Medical exposure in Medical exposure in radiology: radiology: Optimization of Optimization of protection protection Module VIII.3 - Part 1: Design considerations for the equipment !! CONTINUATION OF FILE VIII.3 Equipment !

International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Embed Size (px)

Citation preview

Page 1: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

International Atomic Energy Agency

Medical exposure in radiology:Medical exposure in radiology:Optimization of protectionOptimization of protection

Module VIII.3 - Part 1: Design considerations for the equipment

!! CONTINUATION OF FILE VIII.3 Equipment !

Page 2: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

International Atomic Energy Agency

Topic 4: Computed Topic 4: Computed TomographyTomography

Page 3: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 3

International Atomic Energy Agency

The CT equipmentThe CT equipment

• The CT scanner

• Hounsfield units

• Difference with projection radiology

• Latest generations of scanners

Page 4: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 4

International Atomic Energy Agency

CT: “Computed” and “Tomography”CT: “Computed” and “Tomography”

CT is a tomographic imaging technique, generating cross-sectional image in the axial planeCT techniques use high kV with heavy filtrationA fan beam is passed through the patientTransmitted radiation is measured by array detectorsCT images or “sections” are maps of µThey are derived by mathematical analysis of multiple projections (use of computer): filtered back projection

Page 5: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 5

International Atomic Energy Agency

Image and “Hounsfield” numberImage and “Hounsfield” number

• Pixel size depends on FOV and matrix• Matrix size 512x512 or 1024x1024

• 12 bits (4096 gray levels)

• Voxel is:• volume element=pixel area x slice thickness

• Relative attenuation coefficient µ is expressed in Hounsfield units or CT numbers

• By definition HUwater = 0, HUair = -1000

• Attention: µ will depend on kV…so will HU

HUt = 1000 x (µt - µwater) / µwater

Page 6: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 6

International Atomic Energy Agency

Image display: windowingImage display: windowing

CT is a DIGITAL IMAGEWindow settings determine HOW tissue attenuation values are displayed

Page 7: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 7

International Atomic Energy Agency

Window settings: optimized for mediastinumWindow settings: optimized for mediastinum

Page 8: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 8

International Atomic Energy Agency

Window settings: optimized for lungWindow settings: optimized for lung

Page 9: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 9

International Atomic Energy Agency

CT scannerCT scanner

• Computed Tomography (CT) was introduced into clinical practice in 1972 and revolutionized X Ray imaging by providing high quality images which reproduced transverse cross sections of the body.

• Tissues are therefore not superimposed on the image as they are in conventional projections

• The technique offered in particular improved low contrast resolution for better visualization of soft tissue, but with relatively high absorbed radiation dose

• Dose distribution different due to rotation

Page 10: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 10

International Atomic Energy Agency

Dose distributions in projection RL and CTDose distributions in projection RL and CT

Conventional: dose decreases from entrance to exit, ratio

1/100 …1/1000

Head exam

Rotational scanning gives dose more concentrated at center of rotation

Also dose distribution in slices, but contribution outside imaged slice

Page 11: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 11

International Atomic Energy Agency

3rd & 4th generation scanners3rd & 4th generation scanners

3rd generation: rotating fan beam and array of

detectors (rotate-rotate)

4th generation: rotating source and fixed ring of

detectors (rotate-stationary)

Page 12: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 12

International Atomic Energy Agency

A look inside a rotate/rotate CTA look inside a rotate/rotate CT

X Ray Tube

Detector Arrayand Collimator

Page 13: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 13

International Atomic Energy Agency

3rd & 4th generation scanners3rd & 4th generation scanners

• Slice (section) acquisition: 1-2 seconds

• Cables limit rotation to 1 revolution

• Modern scanners: slip ring technology

• Beam is highly filtered (-> 10mmAl HVL)

• Heat loading on tube very high

• Collimation • defines section thickness

• reduces scatter

• @ 1mm slice also additional collimation at detector

Page 14: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 14

International Atomic Energy Agency

X Ray beam

Direction of patientmovement

Helical or spiral CTHelical or spiral CT

• Slip ring technology allows continuous rotation

• The patient can then be moved continuously through the beam, making the examination much faster

• Continuous data acquisition and table feed

Page 15: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 15

International Atomic Energy Agency

New possibilities for proceduresNew possibilities for procedures

• The new helical scanning CT units allow a range of new features, such as :• CT fluoroscopy, where the patient is stationary, but

the tube continues to rotate

• multislice CT, where multiple slices can be collected simultaneously

• 3-dimensional CT and CT endoscopy

• All these new technologies and applications:• Constant increase in number of examinations

• High contribution to collective effective dose

Page 16: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 16

International Atomic Energy Agency

CT contribution to number of exams and to CT contribution to number of exams and to collective effective dose collective effective dose (FRG 1990-92)(FRG 1990-92)

% freq X Ray exams FRG 1990-02 % coll effect dose X-ray FRG 1990-92

CT35%

Spine 10%

Chest5%

GI15%

Angio/DSA10%

Urography9%

Misc16%CT

4%Spine 10%

Chest18%

Dental17%

Angio/DSA1%

Extremities20%

Misc30%

FRG: Federal Republic of Germany

Page 17: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 17

International Atomic Energy Agency

Factors influencing dose in CTFactors influencing dose in CT

Page 18: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 18

International Atomic Energy Agency

Equipment related factorsEquipment related factors

• Wave form of generator: ~nil, low ripple

• Range of tube current settings, steps

• Beam filtration

• Beam shaper

• Focus axis distance: • shorter geometry (60 cm instead of 80) gives increase of

wCTDI, but allows less mAs

• Slice collimation• For ~1mm slices: dose increase

• Scan field and scan angle

Page 19: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 19

International Atomic Energy Agency

Dose and spiral CTDose and spiral CT

• With identical protocol settings(Slice thickness, table feed, mAs product per scan or rotation, scan

length)

dose in spiral CT would be same as in sequential scanning

• For interpolation purpose: ~1 additional rotation => slight increase in dose (<10%)

• If pitch is taken >1: less loss of information, but dose is reduced

• BUT often INCREASE:• Scanning over longer distances

• Multi-phase studies, same body scanned repeatedly

Page 20: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 20

International Atomic Energy Agency

Summary on this topic CTSummary on this topic CT

• We learned about the principle of CT scanner, the different generations and the factors influencing this high dose examination

Page 21: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

International Atomic Energy Agency

Topic 5: X Ray beam Topic 5: X Ray beam characteristicscharacteristics

Beam description, factors affecting beam quality and effect on imaging process.

Page 22: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 22

International Atomic Energy Agency

Contents of this topicContents of this topic• Factors affecting X Ray beam and image

• Spectra

• Beam quality

• Tube current

• Influence of mAs

• Filtration

• Wave-form, ripple

• Geometric factors• Inverse square law

• Anode angle

• Heel effect

• Unwanted radiation• Stray radiation, scatter, leakage

• Fighting against scatter

• Enhancing contrast• Lowering kV

• Contrast agents

Page 23: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 23

International Atomic Energy Agency

Factors affecting the X Ray beamFactors affecting the X Ray beam

•Tube current•kVp value•Ripple•Filtration

Page 24: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 24

International Atomic Energy Agency

Photon spectra of X RaysPhoton spectra of X Rays

kVp ≤ charact. line en. kVp > charact. line en.

Page 25: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 25

International Atomic Energy Agency

Beam “Quality”Beam “Quality”

• “Measures” penetrating power

• Subjectively describes the shape of continuous spectrum

• Function of:• kVp

• filtration

• high voltage supply characteristics

• Diagnostic radiology beams are POLYCHROMATIC

• ± Quantification: effective energy

• With moderate filtration: Eeff ≤ 2/3 kVp

• With heavy filtration: Eeff ≤ 1/2 kVp

Page 26: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 26

International Atomic Energy Agency

Changing the tube currentChanging the tube current

Change of QUANTITYNO change of quality

Effective kV not changed

Page 27: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 27

International Atomic Energy Agency

Changing current x time product (mAs)Changing current x time product (mAs)

70 kV- 50 mAs 70 kV- 80 mAs 70 kV- 25 mAs

Page 28: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 28

International Atomic Energy Agency

Effect of changing kVEffect of changing kVpp

Change in QUANTITY&

Change in QUALITY - spectrum shifts to higher energy- characteristic lines appear

Page 29: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 29

International Atomic Energy Agency

Importance of correct choice of kVImportance of correct choice of kV

Right image is made with correct kV; left image with too high kV, leading to loss of contrast.

Page 30: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 30

International Atomic Energy Agency

Remember generator waveformsRemember generator waveforms

100%

13%

4%

Line voltage

Single phase single pulse

Single phase 2-pulse

Three phase 6-pulse

Three phase 12-pulse

0.02 s

0.01 s

kV ripple (%)

Page 31: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 31

International Atomic Energy Agency

Effect of waveform or rippleEffect of waveform or ripple

• During 1/2 mains cycle kV fluctuates between min and max

• At any moment spectrum defined by kV

• The more ripple, the more low energy photons

• Less ripple, beam is harder

• We do not need these low energy photons:• Absorbed in patient or scattered

• Dose to patient increases

• Do not reach the image receptor

• HF or 12-pulse will reduce dose to patient

• And give less noisy images (less scatter)

Page 32: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 32

International Atomic Energy Agency

Ripple increases image noiseRipple increases image noise

The left image is made with an old (high ripple) generator.The increased noise led to loss of detail compared to modern equipment

Page 33: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 33

International Atomic Energy Agency

FiltrationFiltration

Change in QUANTITY&

Change in QUALITY spectrum shifts to higher energy

1- Spectrum out of anode2- After window tube housing

(INHERENT filtration)3- After ADDITIONAL filtration

Page 34: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 34

International Atomic Energy Agency

Inherent filtrationInherent filtration

•Tube housing window 1•Inherent filtration:

–Glass envelope–Cooling oil–Beryllium window (thin)

•“Radioluscent”•You can see the filament: -->1

•Filtration expressed in equivalent Al thickness•Typically: 1mm Al•Housing: lead shield !

Page 35: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 35

International Atomic Energy Agency

Additional filtrationAdditional filtration

Additional filtration: •preferably removes lower energy photons• increases effective energy Eeff

• lowers intensity• reduces patient dose !•but increases tube loading

Page 36: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 36

International Atomic Energy Agency

K-edge filtrationK-edge filtration

kV

Metals with K-edges in diagnostic rangeEx: Erbium (Mo, Rh,Sn)Max absorption above K-edgeAllow modification spectrum: mammoK-edge Tin (Sn) allows lightweight shielding in protective aprons

Relative intensity

Page 37: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 37

International Atomic Energy Agency

Geometrical propertiesGeometrical properties

Inverse square lawAnode angleHeel effect

Page 38: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 38

International Atomic Energy Agency

Inverse square law: dose reductionInverse square law: dose reduction

Dose = 1

Dose = 1/9

Dose = 1/4

Fluency of beam constant

Dose (or doserate) per cm2

proportional to 1/x2

Page 39: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 39

International Atomic Energy Agency

Reduces geometric unsharpnessReduces patient dose

Inverse square lawInverse square law

• Same dose D at image receptor

• Increase at entrance point: • Left: (40/20)2 => 4 D

• Right (120/100)2 => 1.44 D

Front

Image Receptor

Patient

Front

Image Receptor

Patient

Focus20cm

20cm

20cm

100cm

Page 40: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 40

International Atomic Energy Agency

Anode angle and resolutionAnode angle and resolution

Geometrically: size of the effective focal spot directly related to the sine of the angle of the anode. A the angle of the anode is decreased, the effective focal spot is also decreased.Penumbra and geometrical unsharpness decreases

Page 41: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 41

International Atomic Energy Agency

Heel effectHeel effectPlot of intensity shows variation along anode to cathode axis

Heel effect lower by large FID: not the full cone of radiation (C to

B) is used

Page 42: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 42

International Atomic Energy Agency

Unwanted radiationUnwanted radiation

Page 43: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 43

International Atomic Energy Agency

Leakage radiationLeakage radiation

•Radiation transmitted thtrough tube housing• Image taken on film• Cathode side• Need for Radiation protection

•At installation•In use•After maintainance

Leakage

Page 44: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 44

International Atomic Energy Agency

Stray radiationStray radiation

• Is the sum of the leakage and scattered radiation

• Scattered radiation has been deviated after leaving the tube

• Scatter occurs from all material exposed to the radiation: patient, table, filters,room walls

Page 45: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 45

International Atomic Energy Agency

Scattered RadiationScattered Radiation

• Effect on image quality • increasing of noise

• loss of contrast

• Effect on patient dose • increasing of superficial and depth dose

• Possible reduction through :• use of grid

• limitation of the field to the useful portion

• limitation of the irradiated volume (e.g.:breast compression in mammography)

• use of air gap

Page 46: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 46

International Atomic Energy Agency

Anti-scatter gridAnti-scatter grid

Source of X Rays

LeadScattered X Rays

Useful X RaysFilm and cassette

Patient

Page 47: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 47

International Atomic Energy Agency

Anti-scatter gridAnti-scatter grid

• Radiation emerging from the patient• primary beam : contributes to the image

• scattered radiation does reach the detector and contributes to background without information, and contrast is lowered

• scattered radiation also contributes to unnecessary dose

• the grid (between patient and film) eliminates most of scattered radiation on image receptor

Page 48: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 48

International Atomic Energy Agency

Anti-scatter gridAnti-scatter grid

BUT:to keep same doseon image receptor

increase dose

Grid NOT recommended for

extremities orpediatric radiology

Page 49: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 49

International Atomic Energy Agency

Scatter reduction by reducing field sizeScatter reduction by reducing field size

Page 50: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 50

International Atomic Energy Agency

Scatter removal by air gapScatter removal by air gap

Page 51: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 51

International Atomic Energy Agency

Scatter removal by narrow beam geometryScatter removal by narrow beam geometry

Page 52: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 52

International Atomic Energy Agency

Contrast agentContrast agent

• Tissue characteristics and photon energy determine photoelectric absorption

• Basic idea to enhance contrast: lower kV• cf. mammography

• But lower kV, lower penetration

• How to improve contrast?

• Introducing contrast agents: Ba, Iodine• High atomic number : photoelectric ~ Z3

• K-edge falls within diagnostic imaging kV

Page 53: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 53

International Atomic Energy Agency

Summary on Summary on what we learned about the X Ray beam what we learned about the X Ray beam (1)(1)

• Which factors influence the X Ray spectrum?• tube potential

• kVp value

• wave shape of tube potential

• anode track material• W, Mo, Rh etc.

• X Ray beam filtration• inherent + additional

• How to enhance contrast• Lower kV

• Use of contrast agents: Barium, Iodine

Page 54: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 54

International Atomic Energy Agency

Summary on Summary on what we learned about the X Ray beam what we learned about the X Ray beam (2)(2)

• We became familiar with geometrical properties• Inverse square law• Anode angle• Heel effect

• We learned about some unwanted effects• Leakage radiation• Stray radiation• Scattered radiation• How to “remove” unwanted scatter radiation

• Reduce field size• Grid• Air gap• Narrow beam geometry

Page 55: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

International Atomic Energy Agency

Topic 6: Special applicationsTopic 6: Special applications

Dental equipment, Pediatric Equipment

Page 56: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 56

International Atomic Energy Agency

Summary of special applications Summary of special applications

• Pediatric radiology• Specificity

• Requirements on equipment, rooms and accessories

• Positioning and immobilization

• Considerations for the use of the equipment

• Protective shielding

• Dental radiology• Low dose high frequency

• The equipment

• Radioprotection and patient dose

Page 57: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 57

International Atomic Energy Agency

Equipment for pediatric roomEquipment for pediatric room

Page 58: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 58

International Atomic Energy Agency

Specificity of pediatric radiologySpecificity of pediatric radiology

• Longer life expectance• Risk of late detrimental radiation effect greater

• Estimated 2-3 x greater than @ 30-40y or 5-7 x greater than after 50y

• Justification and optimization even more important

• Smaller size require adapted radiographic techniques and exposure factors

• Positioning and lack of co-operation requires short exposures

• Functional differences (e.g. higher heart rate, faster respiration, inability to stop breathing on command, increased intestinal gas etc.)

Page 59: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 59

International Atomic Energy Agency

General, equipment and installation General, equipment and installation considerations considerations (1)(1)

• Short exposure times can improve image quality. So the generator should have enough power to allow short exposure times (3 milliseconds) and the timer should allow short exposure times.

• The use of mobile X Ray units in pediatrics could raise special problems (low power:motion blurring).

• The generator should be of high frequency to improve the accuracy and reproducibility of exposures.

Page 60: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 60

International Atomic Energy Agency

General, equipment and installation General, equipment and installation considerations considerations (2)(2)

• Automatic exposure control (AEC) devices should be used with caution in pediatrics (they should be adapted specifically to pediatrics)..

• Careful manual selection of exposure factors usually results in lower doses

• X Ray rooms for pediatrics should be designed for improving the child’s cooperation (control panel with easy patient visibility and audio communication, etc.).

• Fast film-screen combinations have advantages (reduction of dose)

Page 61: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 61

International Atomic Energy Agency

General, equipment and installation General, equipment and installation considerations considerations (3)(3)

• Low-absorbing materials in cassettes, tabletops, etc. specially important in pediatrics radiology.

• The antiscatter grid in pediatrics gives limited improvement in image quality and increases patient dose given the smaller irradiated volume (and mass) the scattered radiation is less

• Antiscatter grid should be removable in pediatric equipment, particularly fluoroscopic systems

Page 62: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 62

International Atomic Energy Agency

General, equipment and installation General, equipment and installation considerations considerations (4)(4)

• Image intensifiers should have high conversion factors for reducing patient dose in fluoroscopic systems.

• Additional tube filtration may allows dose reductions.

• Pulsed fluoroscopy is recommended since it allows patient dose reduction as in adults

• For CT examinations, the use of specific technical radiographic parameters for CT examinations should be promoted (lower mAs than for adults, and lower kV in some cases)

Page 63: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 63

International Atomic Energy Agency

Dental radiologyDental radiology

Page 64: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 64

International Atomic Energy Agency

Dental radiologyDental radiology

• High frequency (~1/4 of all examinations) but low dose technique

• Wide range of entrance doses:0.5 - 150mGy

• Dental radiology contributed for 1% of collective dose from medical diagnostics

• Image quality often very low

• Many of them in private practice with no medical physicist or RP officer and have no medical physics support available

• Organs at risk: parathyroid, thyroid, larynx, parotid glands

Page 65: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 65

International Atomic Energy Agency

Intra-oral X-ray dental equipmentIntra-oral X-ray dental equipment

Some technical parameters:

-Tube 65kV 7.5 mA-Filtration: 2mm Al

-Focus skin distance: 20cm-Field size: 6cm diameter

Page 66: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 66

International Atomic Energy Agency

Panoramic dental X-ray equipmentPanoramic dental X-ray equipment

Some technical parameters:

-Tube 60-80kVkV -Current 4-10mA

Page 67: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 67

International Atomic Energy Agency

Image receptors in dental radiologyImage receptors in dental radiology

• Small films (2 x 3 or 3 x 4 cm) in light-tight envelopes (no screen)

• Digital intraoral sensors. • Compared with category E film, the radiation dose

is reduced by 60%.

Intraoral Radiology

Panoramic Radiology

• Film-screen combination• Digital sensors. • Compared with film-screen sensitivity class 200,

the radiation dose is reduced by 50-70%.

Page 68: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 68

International Atomic Energy Agency

What we learned about X Ray equipmentWhat we learned about X Ray equipment

• We studied X Ray production (the tube, the generator,…) and the variety of radiological equipment (for plain radiography, fluoroscopy, CT, …)

• We learned about the basic interactions leading to the image formation

• Image intensifiers may drastically improve image quality and reduce staff dose

• and the beam characteristics, including the « undesirable » apects (scatter, leakage)

Page 69: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 69

International Atomic Energy Agency

What we learned about X Ray equipment What we learned about X Ray equipment (ctnd)(ctnd)

• Specific examinations need dedicated equipment (dental, pediatric, mammographic)

• Digital images can enhance contrast and extract regions of interest

• Equipment used by non-radiologists (surgeons, cardiologists,…) require special technical requirements

Page 70: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 70

International Atomic Energy Agency

Where to get more information Where to get more information (1)(1)

• The Physics of Diagnostic Imaging, David J. Dowsett, Patrick A. Kenny and R. Eugene Johnston, Chapman & Hall Medical, ISBN 0-412-40170-1

• Equipment for diagnostic radiology, E. Forster, MTP Press, 1993

• The Essential Physics of Medical Imaging, Williams and Wilkins. Baltimore:1994

• Imaging systems in medical diagnostics, Krestel ed., Siemens, 1990

• The AAPM/RSNA Physics Tutorial for Residents General Overview of Fluoroscopic Imaging, B. A. Schueler, Radiographics Vol 20, 1115-1126,(2000)

Page 71: International Atomic Energy Agency Medical exposure in radiology: Optimization of protection Module VIII.3 - Part 1: Design considerations for the equipment

Module VIII.3-Part 1. Design considerations equipment Continued 71

International Atomic Energy Agency

Where to get more information Where to get more information (2)(2)

• European Guidelines on quality criteria for CT. Report EUR 167262, Office for Official Publications of the European Communities, 1999, Luxemburg

• Radiation exposure in Computed Tomography; 4th revised Edition, December 2002, H.D.Nagel, CTB Publications, D-21073 Hamburg

• Rational use of diagnostic imaging in pediatrics. WHO, 1987

• European guidelines on quality criteria for diagnostic radiographic images. Report EUR16261 (1996)

• Radiation protection and quality assurance in dental radiology. Radiation protection 81. European Commission.(1995) CG-89-95-971-EN-C