Contents

Chapter 1 1 Intended Use, Contraindications and General Precautions

Chapter 2 3 Description of the Inogen One® G2 Oxygen Concentrator 3 ImportantPartsoftheInogenOne®G2OxygenConcentrator 4 UserControls 5 UserInterfaces 5 Input/OutputConnections 6 PowerSupplyOptions 9 InogenOne®G2Accessories

Chapter 3 11 Operating Instructions 11 GeneralInstructions 15 AdditionalOperatingInstructions 18 BatteryOperatingInstructions 19 BatteryCareandMaintenanceChapter 4 21 Inogen One® G2 Oxygen Concentrator Audible and Visible Signals (including Alarms)

Chapter 5 29 Troubleshooting Chapter 6 31 Cleaning, Care and Maintenance 31 CannulaReplacement 31 CaseCleaning 32 FilterCleaningandReplacement 34 OtherServiceandMaintenance 34 Disposalofequipmentandaccessories

Chapter 7 35 Symbols Used on Concentrator and Accessories

Chapter 8 37 Inogen One® G2 System Specifications

English

Page 4: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external
Page 5: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

1 Intended Use, Contraindications and General Precautions

IntendedUse,ContraindicationsandGeneralPrecautionsChapter1 1

Intended Use

TheInogenOne®G2OxygenConcentratorisusedonaprescriptivebasisbypatientsrequiringsupplementaloxygen.Itsuppliesahighconcentrationofoxygenandisusedwithanasalcannulatochanneloxygenfromtheconcentratortothepatient.TheInogenOne®G2maybeusedinhome,institution,vehicle,onanairplaneandvariousmobileenvironments.

CAUTION USAFederallawrestrictsthisdevicetosalebyorontheorderofaphysician.Mayalsobeapplicableinothercountries.

CAUTION Useofacannulaotherthanahighflowcannula(e.g.Salter1600Q)mayrestrictoxygendeliveryand/orattachmenttonozzlefittings.

wArNING Availabilityofanalternatesourceofoxygenisrecommendedincaseofpoweroutageormechanicalfailure.Consultyourequipmentproviderfortypeofback-upsystemrecommended.

CAUTION Itistheresponsibilityofthepatienttomakeback-uparrangementsforalternativeoxygensupplywhentraveling;Inogenassumesnoliabilityforpersonschoosingnottoadheretomanufacturerrecommendations.

! C

English

Page 6: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

2

Contraindications

wArNING ThisdeviceisNOTINTENDEDtobelifesustainingorlifesupporting. CAUTION Undercertaincircumstances,theuseofnon-prescribedoxygen

therapycanbehazardous.Thisdeviceshouldbeusedonlywhenprescribedbyaphysician.

CAUTION Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealarmsorcommunicatediscomfort.Ifthepatientshowsanysignsofdiscomfort,aphysicianshouldbeconsultedimmediately.

CAUTION TheInogenOne®G2isnotdesignedorspecifiedtobeusedinconjunctionwithahumidifierornebulizer.Useofthisdevicewithahumidifierornebulizermayimpairperformanceand/ordamagetheequipment.

! C

! C

! C

General Precautions

wArNING Thedeviceproducesenrichedoxygengaswhichacceleratescombustion.DoNOTALLOWSMOKINGOROPENFLAMESwithin10feetofthisdevicewhileinuse.

wArNING DonotsubmersetheInogenOne®G2oranyoftheaccessoriesinliquid.Donotexposetowaterorprecipitation.Donotoperateinexposedrain.Thiscouldleadtoelectricalshockand/ordamage.

CAUTION Donotuseoil,grease,orpetroleum-basedproductsonornearthe

InogenOne®G2.

CAUTION NeverleavetheInogenOne®G2inanenvironmentwhichcanreachhightemperatures,suchasanunoccupiedcarinhightemperatureenvironments.Thiscoulddamagethedevice.

Page 7: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Important Parts of the Inogen One® G2 Oxygen Concentrator

Description of the Inogen One® G2 Oxygen Concentrator

DescriptionoftheInogenOne®G2OxygenConcentratorChapter2 3

2

On/Off Button

Alert/Alarm Light

Flow ControlDisplay

DisplayBacklight Button

Breath Detection Light

Audible Alarm Button

English

Page 8: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

4

User Controls

ON / OFF ButtonPressoncetoturn“ON”;Pressandholdforonesecondtoturn“OFF”.

Audible Alarm Button PressingthisbuttonwilltoggletheInogenOne®G2’sbreathdetectionalertonandoff:

1. DefaultMode.WhentheInogenOne®G2ispoweredup,thebreathdetectionalertisdisabled.Thedisplay’smodeindicationareaisemptyinDefaultMode.

2. BreathDetectionAlertMode.TheInogenOne®G2willalertwithaudibleandvisualsignalsfor“nobreathdetected”whenthismodeisenabledandnobreathhasbeendetectedfor60seconds.Thedisplay’smodeindicationareawillshowabelliconwhenthealertisenabled.

Flow Setting Control Buttons Usethe–or+flowsettingcontrolbuttonstoselectthesettingasshownonthedisplay.Therearefivesettings,from1to5.

Display Backlight ButtonPushon;automaticallyturnsoffafter10seconds.

Page 9: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

DescriptionoftheInogenOne®G2OxygenConcentratorChapter2 5

User Interfaces

DisplayThisscreendisplaysinformationregardingflowsetting,powerstatus,batterylifeanderrors.IfyouwouldliketochangethelanguageontheInogenLCDScreencontactInogenClientServices.

Indicator LightsAredlightindicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Aflashinglightishigherprioritythannon-flashing.Audible SignalsAnaudiblesignal(beep)indicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Morefrequentbeepsindicatehigherpriorityconditions.

Input / Output Connections

Particle FilterThefiltermustbeinplaceattheintakeendoftheconcentratorduringoperationtokeepinputairclean.

Cannula Nozzle FittingThenasalcannulaconnectstothisnozzleforInogenOne®G2outputofoxygenatedair.

DC Power InConnectionforexternalpowerfromtheUniversalPowerSupply.

English

Page 10: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

6

Power Supply Options

Single and Double rechargeable Lithium Ion BatteriesThebatterywillpowertheInogenOne®G2withoutconnectiontoanexternalpowersource.Whenfullycharged,asinglebatterywillprovide2to4hoursofoperation;adoublebatterywillprovide4to8hoursofoperation.ThebatteryrechargeswhenproperlyinstalledintheInogenOne®G2andtheconcentratorisconnectedtoACorDCpower.Rechargingtimeisupto4hoursforasinglebatteryandupto8hoursforadoublebattery.Seeinformationinthe“BatteryCareandMaintenance”section.

UNIVErSAL POwEr SUPPLyOverviewTheInogenUniversalPowerSupply(BA-107/207)isusedtopowertheInogenOne®G2concentratorfromanACpowersourceoraDCpowersource.Thisallowstheusertopowertheconcentratorinthehome,inavehicleorotherplaceswhereACorDCpowerisavailable.

DescriptionTheInogenOne®G2UniversalPowerSupplyisspecificallydesignedforusewiththeInogenOne®G2OxygenConcentrator(IO-200).TheUniversalPowerSupplyprovidestheprecisecurrentandvoltagerequiredtosafelypowertheInogenOne®G2andisdesignedtooperatefromspecifiedACandDCpowersources.WhenusedwithACpowersources,thepowersupplyautomaticallyadaptstoinputvoltagesfrom100Vto240V(50-60HZ)permittingusewithmostpowersourcesthroughouttheworld.

TheUniversalPowerSupplywillchargetheInogenOne®G2BatterywhenusedwithACinputpowerorwhenusedwithaDCpowersupply,suchastheonefoundinyourcar.Duetoaircraftpowerlimitations,theUniversalPowerSupplycannotbeusedtochargetheInogenOne®G2Batterywhenusedonanaircraft.TheUniversalPowerSupplyisusedwiththefollowingcomponents:

•PowerSupplywithattachedpoweroutputcabletoconnecttotheInogenOne®G2•AutomobileDCcigarettelighterPowerInputCable•ACPowerInputCable

Page 11: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

7DescriptionoftheInogenOne®G2OxygenConcentratorChapter2

Power Output Plug

Universal Power Supply

LED Indicator Light

AC Input

DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft

DC Input

AC Power Plug*

wArNING DonotwrapcordsaroundPowerSupplyforstorage.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.

Model#BA-107

! C

*Actualproductappearancemayvary.

English

Page 12: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

8

Power Output Plug

Universal Power Supply

LED Indicator Light

AC Input

DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft

DC Input

AC Power Plug*

Model#BA-207

wArNING DonotwrapcordsaroundPowerSupplyforstorage.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.! C

*Actualproductappearancemayvary.

Page 13: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Inogen One® G2 Accessories

Nasal CannulaAnasalcannulamustbeusedwiththeInogenOne®G2toprovideoxygenfromtheconcentrator.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.

NOTE Increasingthecannulalengthmayreducetheperceivednoiseduringoxygenbolusdelivery.

Carry BagTheCarryBagprovidesaprotectivecoverwithahandleandshoulderstraptoenableyoutocarrytheInogenOne®G2.TheInogenOne®G2canbeoperatedusingbatterypowerduringtransportwiththeCarryBag.

CartTheCarthaswheelsandatelescopinghandletoprovideeasytransportoftheInogenOne®G2.TheInogenOne®G2canbeoperatedusingbatterypowerduringtransport.PlacetheCarryBagovertheCart.MakesuretheCarthandleisinsertedbetweentheelasticstrapinthebackofthecarrybagandthefrontofthecarrybag.

DescriptionoftheInogenOne®G2OxygenConcentratorChapter2 9

English

Page 14: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Inogen One® G2 Optional Accessory

External Battery Charger1. PlugtheExternalBatteryChargerACpowersupply

cordintoanelectricaloutlet.2. PlugtheExternalBatteryChargerACpowersupplyinto

thebatterycharger.3. SlideyourchargerontotheInogenOneG2Batteryby

clickingandlockingintothecharger.4. Whenthebatteryisinthecorrectposition,asolidredlight

willindicatethatthebatteryischarging.5. Whenthegreenlightilluminates,thebatteryis

fullycharged.

CAUTION AvoidtouchingtherecessedelectricalcontactsoftheExternalBatteryCharger;damagetocontactsmayaffectchargeroperation.

NOTE Thesecontactsarenotpoweredunlessabatteryisinplaceandcharging.

NOTE TocompletelyremovepowerfromtheExternalBatteryCharger,removetheplug.

Inogen One G2 BackpackAlternative/optionalwayofcarryingyourInogenOneG2,handsfree,morecomfort,outofyourwaywithextrapocketsforadditionalaccessories.

ToorderpleasecallInogenClientServices.

10

Page 15: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Operating Instructions

OperatingInstructionsChapter3 11

3

! C

Intake

Exhaust

General Instructions

1. Place the Inogen One® G2 in a well ventilated location.Airintakeandexhaustmusthaveclearaccess.LocatetheInogenOne®G2insuchawaythatanyauditoryalarmsmaybeheard.

wArNING AvoiduseoftheInogenOne®G2inpresenceofpollutants,smokeorfumes.DonotusetheInogenOne®G2inpresenceofflammableanesthetics,cleaningagentsorotherchemicalvapors.

CAUTION Donotobstructairintakeorexhaustwhenoperatingthedevice.

Blockageofaircirculationorproximitytoaheatsourcemayleadtointernalheatbuildupandshutdownordamagetotheconcentrator.

2. Ensure the particle filter is in place.

CAUTION DonotoperatetheInogenOne®G2withouttheparticlefilterinplace.Particlesdrawnintothesystemmaydamagetheequipment.

3. Install the battery. InserttheInogenOne®G2batterybyslidingbatteryintoplaceandlockingatbottomofconcentrator.

English

Page 16: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

12

CAUTION TheInogenOne®G2batteryactsasasecondarypowersupplyintheeventofaplannedorunexpectedlossoftheACorDCexternalpowersupply.WhenoperatingtheInogenOne®G2fromanACorDCexternalpowersupply,aproperlyinsertedInogenOne®G2batteryshouldbemaintainedintheunit.Thisprocedurewillensureuninterruptedoperationandwilloperateallalarmsandalertsintheeventofalossoftheexternalpowersupply.

NOTE TheInogenOne®G2BatteryrequiresaninitialfullchargeofuninterruptedchargingfromanemptystateusingtheInogenOne®G2ontheACpowersource.DonotruntheInogenOne®G2onbatterypoweruntilthisinitialcharginghasbeencompleted.Oncetheinitialchargeiscompleted,thebatterymaybeusedwithanyamountofchargelevel.

4. Connect the Universal Power Supply. ConnecttheACPowerplugtotheUniversalPowerSupply.ConnecttheACPowerPlugtothepowersourceandconnectthePowerOutputPlugtotheInogenOne®G2.ThegreenLEDontheUniversalPowerSupplywillbeilluminatedandabeepwillsoundfromtheconcentrator.

CAUTION EnsuretheUniversalPowerSupplyisinawellventilatedlocationasitrelies onaircirculationforheatdissipation.TheUniversalPowerSupplymaybecome hotduringoperation.MakesuretheUniversalPowerSupplycoolsdownbefore handling.

CAUTION TheUniversalPowerSupplyisnotwaterresistant.

CAUTION DonotdisassembletheUniversalPowerSupply.Thismayleadtocomponentfailureand/orsafetyrisk.

CAUTION DonotplaceanythingintheUniversalPowerSupplyportotherthanthesuppliedwallcord.AvoidtheuseofelectricalextensioncordswiththeInogenOne®G2.Ifanextensioncordmustbeused,useanextensioncordthathasanUnderwritersLaboratory(UL)Markandaminimumwirethicknessof18gauge.Donotconnectanyotherdevicestothesameextensioncord.

Page 17: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

OperatingInstructionsChapter3 13

NOTE Undercertainconditions(seeTechnicalSpecifications)theUniversalPowerSupplymayshutdown.ThegreenLEDwillblinkorwillnolongerbeilluminated.Ifthisoccurs,disconnectthePowerSupplyforatleast10secondsandreconnect.

NOTE WhentheUniversalPowerSupplyisdisconnectedfromtheACoutlet,alsodisconnectitfromtheconcentratortoavoidunnecessarybatterydischarge.

5. Connect the nasal cannula tubing to the nozzle fitting.NozzlefittingislocatednexttothehandleoftheInogenOne®G2.Useofasinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.Additionaltitrationmaybeneededtoensureproperoxygendeliverywhenusingaparticularcannula.

CAUTION Toensureoxygenflow,ensurethatthenasalcannulaisproperlyconnectedtothenozzlefittingandthatthetubingisnotkinkedorpinchedinanyway.

CAUTION Replacethenasalcannulaonaregularbasis.Checkwithyourequipmentproviderorphysiciantodeterminehowoftenthecannulashouldbereplaced.

6. Turn on your Inogen One® G2 by pressing the ON/OFF Button. AsingleshortbeepwillsoundaftertheInogenlogoisdisplayed.“PleaseWait”willappearwhiletheconcentratorstartsup.Thedisplaywillindicatetheselectedflowsettingandpowercondition.Followingabriefstart-upsequence,awarmupperiodupto2minuteswillinitiate.Duringthistimeperiodtheoxygenconcentrationisbuildingtobutmaynothavereachedspecification.AdditionalwarmuptimemaybeneededifyourInogenOne®G2hasbeenstoredinextremelycoldtemperatures.

English

Page 18: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

! C

14

7. Set the Inogen One® G2 Concentrator to the flow rate prescribed by your physician orclinician.Usethe+or–settingbuttonstoadjusttheInogenOne®G2tothedesiredsetting.Thecurrentsettingcanbeviewedonthedisplay.

8. Position the nasal cannula on your face and breathe through your nose. TheInogenOne®G2willsensetheonsetofinhalationanddeliveraburstofoxygenataprecisetimewhenyouinhale.TheInogenOne®G2willsenseeachbreathandcontinuetodeliveroxygeninthismanner.Asyourbreathingratechanges,theInogenOne®G2willsensethesechangesanddeliveroxygenonlyasyouneedit.Attimes,ifyouinhaleveryquicklybetweenbreathstheInogenOne®G2mayignoreoneofthebreaths,givingtheappearanceofamissedbreath.ThismaybenormalastheInogenOne®G2sensesandmonitorsthechangesinyourbreathingpattern.TheInogenOne®G2willnormallysensethenextbreathanddeliveroxygenaccordingly.

Agreenlightwillflasheachtimeabreathisdetected.Makecertainthatthenasalcannulaisproperlyalignedonyourfaceandyouarebreathingthroughyournose.

wArNING Ifyoubegintofeelillorareexperiencingdiscomfortwhileusingthisdevice,consultyourphysicianimmediately.

CAUTION TheInogenOne®G2isdesignedtoprovideaflowofhighpurityoxygen.

Anadvisoryalarm,“OxygenLow”,willinformyouifoxygenconcentrationdrops.Ifalarmpersists,contactyourequipmentprovider.

GeneralToremovepower,unplugtheinputcordfromitssource(i.e.,ACwalloutlet,DCcarcigarettelighteradapter).

CAUTION EnsuretheUniversalPowerSupplyispoweredfromonlyonepowersource(ACorDC)atanygiventime.TheUniversalPowerSupplymaynotoperateproperlyifsimultaneouslypoweredusingbothACandDCsources.

Page 19: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

OperatingInstructionsChapter3 15

Additional Operating InstructionsFor Use In Home with AC PowerForoperationofthePowerSupplyusinganACpowersource,followtheseinstructions:

1. ConnecttheACInputtothepowersupply.2. ConnecttheACPowerPlugtothepowersourceandconnectthePowerOutput

PlugtotheInogenOne®G2.ThegreenLEDwillbeilluminated,indicatingtheUniversalPowerSupplyhasinputpower.

Traveling with your Inogen One® G2 System

TheInogenOne®G2Systemmakestravelbyplane,boat,carortrainmoreconvenientforoxygenusersthaneverbefore.Nowyougetthesamequalityperformanceandconveniencewhileonthegothatyou’reusedtoreceivingfromyourInogenOne®G2athome.HerearesomeusefulandimportantinstructionsformaximizingperformanceandconveniencewhenusingyourInogenOne®G2whiletraveling.

Youshouldbeginplanningforatripwithachecklistofitemstoremember.Thischecklistshouldinclude:

✓ UniversalPowerSupply

✓ ExtraBattery(s)ifrequired

✓ Importantphonenumbers,suchasthoseofyourphysicianandhomehealthcare provider,orprovidersintheareayouwillbetraveling

✓ Planforbackupoxygenintheeventofaprolongedpoweroutageormechanicalfailure.

For Use In Automobile/rV/BoatForoperationoftheUniversalPowerSupplyusingaDCpowersource,followtheseinstructions:

1. ConnecttheDCInputtothepowersupply.2. ConnecttheDCPowerPlug(CigaretteLighterAdapter)to

thepowersourceandconnectthePowerOutputPlugtotheInogenOne®G2.ThegreenLEDwillbeilluminated,indicatingthepowersupplyhasinputpower.

English

Page 20: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

3.Theplugshouldinsertintothesocketwithoutexcessiveforceandsecurelyremaininplace.TheCigaretteLighterAdaptercomeswithaslideswitchpre-adjustedtothenarrowposition.Thiswillfitmostautomobilecigarettelightersockets.IftheCigaretteLighterAdapterislooseinthesocket,adjusttheswitchtothe"wide"position.

WARNING EnsurethattheautomobilepowersocketisadequatelyfusedfortheInogenOne®G2powerrequirements(minimum15Amp).Ifthepowersocketcannotsupporta15Ampload,thefusemaybloworthesocketmaybedamaged.

WARNING ThetipoftheCigaretteAdapterPlugbecomesHOTwheninuse.Donottouchthetipimmediatelyafterremovalfromanautocigarettelightersocket.

CAUTION Ensuretheautomobilepowersocketiscleanofcigaretteashandtheadapterplugfitsproperly,otherwiseoverheatingmayoccur.

CAUTION DonotusetheUniversalPowerSupplywithacigaretteplugsplitterorwithanextensioncable.ThismaycauseoverheatingoftheDCpowerinputcable.

CAUTION DonotjumpstarttheautomobilewiththeUniversalPowerSupplyconnected.Thismayleadtovoltagespikeswhichcouldshutdownand/ordamagetheUniversalPowerSupply.

CAUTION WhenpoweringtheInogenOne®G2inanautomobileensuretheengineisrunning.Operatingthedevicewithouttheenginerunningmaydrainthecarbattery.

Traveling By AirTheFAAallowstheInogenOne®G2onboardallU.S.aircraft,hereareafewpointstomakeairtraveleasy.

Planning your FlightWhenflyingwiththeInogenOne®G2,youmustinformtheairlinethatyouwillbeusingyourInogenOne®G2onboardtheaircraft.Youmustalsokeepasignedstatementfromyourphysicianwithyouthatincludes:

•Yourabilitytosee/hearalarmsandappropriatelyrespond.•Whenoxygenuseisnecessary(alloraportionofthetrip).•Maximumflowratecorrespondingtothepressureinthecabinundernormaloperatingconditions.16

! C

! C

Page 21: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

•Anewstatementwillnotbenecessaryeachtimeyoufly,butthestatementshouldbeavailableduringeveryflight.

•Someairlinesmayequiptheiraircraftwithonboardelectricalpower.YoumayhaveanopportunitytorequestaseatwithapowerportwhichcanbeusedtopoweryourInogenOne®G2.However,availabilityvariesbyairline,typeofaircraftandclassofservice.Youshouldcheckwithyourairlinesforavailabilityandalwaysplanonhavingsufficientbatterypowerforthedurationofyourflight,plusaconservativeestimateofunanticipateddelays.

•Youruniversalpowersupplyisequippedwithacommonlyusedcigarettelighteradapter.However,aircraftsusedifferentpowerportconfigurationsanditisdifficulttodeterminewhichtypeofpowerportcompatibilityyouraircraftmaysupply;it’sagoodideatopurchaseanadaptersuchasMagellan’sEmPowerSocketConnectorItem#EA270.ContactMagellan’sat800-962-4943orvisitwww.magellans.comforastorelocationnearyouorpurchaseop-tionsonline.

Before your FlightHerearesomethingstokeepinmindthedayyourflightdeparts:

•EnsureyourInogenOne®G2isclean,ingoodconditionandfreefromdamageorothersignsofexcessivewearorabuse.

•BringenoughchargedbatterieswithyoutopoweryourInogenOne®G2forthedurationoftheflightplusaconservativeestimateofunanticipateddelays.

•Regional/Commuterairlinesdonotofferonboardelectricalpower.Ifyourtravelplanscallforflightsonregionalairlines,youwillneedsufficientbatterypowerforthedurationoftheflight,plusaconservativeestimateofunanticipateddelays.

CAUTION Airlinesmaynotbeequippedtoprovidebackupoxygen.

•Arriveattheairportearly.AirportsecurityscreeningpersonnelmayrequireextratimetoinspectyourInogenOne®G2.

•Whilewaitingtoboardyourflight,youmaybeabletoconservebatterypowerbyusingtheUniversalPowerSupplytopoweryourInogenOne®G2fromanelectricaloutletintheairportterminalifavailable.

•YoushouldinformtheairlineyouwillbeusingyourInogenOne®G2.Haveyourphysicianletterwithyouandreadyforinspectionifrequested.

OperatingInstructionsChapter3 17

English

Page 22: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

During your Flight1. Ifusingairlinepowerport,removetheBatteryfromtheInogenOne®G2Oxygen

Concentrator.Duetoaircraftpowerlimitations,theUniversalPowerSupplycannotbeusedtochargetheInogenOne®G2Batterywhenonboardaircraft.

2. ConnecttheDCPowerPlugforavailableairlinepower.Checkwithairlinepersonneltoensurecompatibility.

•Duringtaxi,takeoffandlanding,stowyourInogenOne®G2undertheseatinfrontofyou.YourInogenOne®G2willfituprightundermostairlineseats.However,ifitdoesn’tfityoumayturnitonitsside.

•ItisnotnecessarytoturnoffyourInogenOne®G2duringtaxi,takeoffandlandingifyourphysician’swrittenstatementrequiresyoureceiveoxygenduringtheseperiods.

CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.TheInogenOne®G2hasbeenverifiedtoprovideoxygentospecificationupto10,000ft(3048m).Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.

After your Flight•Remembertorechargeadditionalbatteriesyoumayhaveusedpriortoyournextflight.

Traveling By Bus, Train or BoatContactyourcarriertofindoutaboutpowerportavailability.

Battery Operating InstructionsEnsurethatthebatteryisinplaceandcharged.DisconnecttheInogenOne®G2fromitspowersource.WhiletheInogenOne®G2isoperatingonbatterypower,thebatterywilldischarge.Thedisplaywillindicatetheestimatedremainingpercentage(%)orminutesofuse.

Whentheconcentratordetectsthatthebatterylifeisbelow10%remaining,alowpriorityalertwillsound.Whenthebatteryisempty,thealertwillchangetoahighpriority.

Whenbatterylifeislow,dooneofthefollowing:•PlugtheInogenOne®G2intoanACorDCpowersourceusingtheUniversalPowerSupply.•ReplacethebatterywithachargedbatteryafterturningofftheInogenOne®G2(bypressingtheON/OFFbutton).•Ifthebatteryisdrained,chargethebatteryorremoveitfromtheconcentrator.18

Page 23: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

OperatingInstructionsChapter3 19

IftheInogenOne®G2isbeingpoweredbytheUniversalPowerSupply,batterieswillchargeduringoperation.LeavingyourInogenOne®G2pluggedinpastthefullchargetimewillnotharmtheconcentratororthebattery.

Normal Battery ChargingToensurethatyourbatteryisproperlycharging,inspectthatthecorrectACandDCpoweroutputplugadapterisbeingusedandthattheadapterisproperlyinsertedintothepoweroutlet.Observethedisplayorlightsthatindicatechargingstatus.

NOTE Whenstartingtochargeafullydischargedbattery,thechargingprocessmay startandstopduringthefirstfewminutes.

Battery Care and MaintenanceYourInogenOne®G2LithiumIonBatteryrequiresspecialcaretoensureproperperformanceandlonglife.UseonlyInogenOne®G2BatterieswithyourInogenOne®G2Concentrator.

Keep DryAlwayskeepliquidsawayfrombatteries.Ifbatteriesbecomewet,discontinueuseimmediatelyanddisposeofbatteryproperly.

Effect of Temperature on Battery PerformanceTheInogenOne®G2singlebatterypowerstheInogenOne®G2Concentratorfrom2to4hoursundermostenvironmentalconditions.Toextendtherun-timeofyourbattery,avoidrunningintemperatureslessthan41˚F(5˚C)orhigherthan95˚F(35˚C)forextendedperiodsoftime.

Battery Time remaining ClockTheInogenOne®G2continuouslydisplaysbatterytimeremaining.Thisdisplayedtimeisonlyanestimateandtheactualtimeremainingmayvaryfromthisvalue.

Please Follow These Important Guidelines to Maximize Battery Performance and Life:•Storebatteryinacool,dryplace.Storewithachargeof40-50%.•Ifusingmultiplebatteries,makesurethateachbatteryislabeled(1,2,3orA,B,C,etc.)androtateonaregularbasis.Batteriesshouldnotbeleftdormantforextendedperiodsoftime.

English

Page 24: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external
Page 25: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Mode

PowerText

Display Icons

TheInogenOne®G2displayisdividedintothreeareas.Theupperleftcornerofthedisplayshowsthebreathdetectionalertstatus.Thelowerleftcornerindicatespowersourceandbatterychargelevel.Therightsideofthedisplaycontainstextinformation,suchasflowsetting,batterytimeremaininganderrornotifications.

Power Status Icons

Theseiconsareexamplesofthoseshowninthedisplay’spowerstatuswindowwhentheInogenOne®G2isoperatingonbatterypower.

Inogen One® G2 Oxygen Concentrator Audible and Visible Signals

AudibleandVisibleSignalsChapter4

! C

! C

! C

! C

Meaning

Batteryisempty.

Batteryhaslessthan10%chargeremaining.Thisiconflashes.

Batteryhasapproximately40%to50%chargeremaining.

Batteryisfull.

Icon

21

4English

Page 26: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Power Status Icons (continued)TheiconsbelowareexamplesofthoseshownwhentheInogenOne®G2isoperatingfromanexternalpowersupplyandchargingthebattery.Thelighteningboltindicatesthatanexternalpowersupplyisconnected.

! C

! C

! CMeaning

Batteryischargingwithchargelevelbetween60%and70%.

Thebatteryisfullychargedandischargingasnecessarytomaintainitscharge.

Batteryischargingwithchargelevellessthan10%.

TheInogenOne®G2isoperatingfromanexternalpowersourcewithnobatterypresent.

Icon

22

Mode IconsThesearetheiconsshowninthedisplay’smodewindow.

Display Text

NOTE Whentwoconditionsoccursimultaneously,theconditionwiththehigherprioritywillbedisplayed.

Meaning

Thebreathdetectionalarmhasbeenenabled.

Theconcentratorisinthestandardbreathdetectmode,andthebreathdetectionalertisdisabled.Thisisthedefaultcondition.

Icon

! C

[blank]

! C

Page 27: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

AudibleandVisibleSignalsChapter4 23

Informational MessagesThefollowinginformationdisplaysarenotaccompaniedbyanyaudiblefeedbackoranyvisualchangeintheindicatorlights.

Condition/Action/Explanation

TheInogenlogoisdisplayedatstartup.

Defaultdisplaywhenoperatingonbatterypower.“X”representstheselectedflowsetting(e.g.,Setting2).“HH:MM”representstheapproximatetimeremainingonthebatterycharge(e.g.,1:45).

Defaultdisplaywhenoperatingonanexternalpowersupplyandthebatteryischarging.“xx%”representsthepercentbatterycharge(e.g.,86%).

Defaultdisplaywhenthebatteryisnotchargingorwhenthetimeremainingisnotavailablefromthebattery.

Displaywhentheconcentratorispluggedinandbeingusedtochargeabattery(notbeingusedforoxygenproduction).Itisnormaltoseeafullychargedbatteryreadbetween95%and100%whenexternalpowerisremoved.Thisfeaturemaximizestheusefullifeofthebattery.

Message Display & Text

SettingXBatteryHH:MM

SettingXChargingxx(or)BatteryFull

SettingXBatteryxx%

Chargingxx%(0r) BatteryFull

NotificationsThefollowingnotificationmessagesareaccompaniedbyasingle, short beep.

Condition/Action/Explanation

On/Offbuttonhasbeenpressedfortwoseconds.Concentratorisperformingsystemshutdown.

AudibleAlarmhasbeenpressedforfiveseconds.

Message Display & Text

InogenOne® ShuttingDown

HH:MM Vx.x:SerialNumber

English

Page 28: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Low Priority AlertsThefollowinglowpriorityalertmessagesareaccompaniedbyadouble beepandasolid red light.

Condition/Action/Explanation

Externalpowersupplyistoolowtoruntheunit,whichisrunningonbatterypowerdespitebeingpluggedin.Checkexternalpowersupplyconnections.Ifconditionpersists,contactyourequipmentprovider.

Externalpowersupplyistoohigh.Checkexternalpowersupplyconnections.Ifconditionpersists,contactyourequipmentprovider.

Batterypowerislow,withlessthan6%remaining.Attachexternalpowersupplyorpowerdownandinsertafullychargedbattery.

Batteryerrorhasoccurred.Switchtoanewbatteryorremovebatteryandoperateconcentratorusingexternalpowersupply.Ifbatteryerrorrecurswithsamebattery,stopusingthebatteryandcontactyourequipmentprovider.

Concentratorisproducingoxygenataslightlylowlevel(<82%).Ifconditionpersists,contactyourequipmentprovider.

Message Display & Text

External PowerLow

External PowerHigh

BatteryLow AttachPlug

BatteryError SeeManual

OxygenLow SeeManual

24

Page 29: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Low Priority Alerts (continued)

Condition/Action/Explanation

Batteryhasexceededitschargingtemperatureandcharginghasstopped.Thebatterywillnotchargewhilethisalertispresentbutwillbegintochargewhenthebatterytemperaturereturnstothenormaloperatingrange.Ifbatterychargingisdesiredsooner,removethebatteryfromtheconcentratorandallowittocoolinanopenareaforapproximately10-15minutes.Then,re-insertthebatteryintotheInogenOne®G2.Iftheproblemstillpersists,contactyourequipmentprovider.

Concentratorisproducingoxygenbutcannotreportbatterystatus.Replacebattery.Ifconditionpersists,contactyourequipmentprovider.

Theconcentratorrequiresservicingattheearliestconvenience.Theconcentratorisoperatingtospecificationandmaycontinuetobeused.Contactyourequipmentprovidertoarrangeforservice.

Theconcentrator’soxygensensorhasmalfunctioned.Youmaycontinuetousetheconcentrator.Iftheconditionpersists,contactyourequipmentprovider.

Message Display & Text

RemoveBatterytoCool

CommError SeeManual

ServiceNeeded

02SensorFailSeeManual

AudibleandVisibleSignalsChapter4 25

English

Page 30: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Condition/Action/Explanation

Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Ifpossible,moveconcentratortoacoolerlocationorpowerunitwithanexternalpowersupplyandremovebattery.Ifconditionpersists,contactyourequipmentprovider.

Concentratorisnotdetectingabreath.•Checkthatcannulaisconnectedtoconcentrator,therearenokinksintubingandcannulaispositionedproperlyinyournose.

Concentratorisexperiencinganerrorbutiscapableofcontinuedoperation.

•Removeandre-insertbatteryand/or•Checkexternalpowersupplyconnections.

Ifconditionpersists,contactyourequipmentprovider.

Oxygenoutputconcentrationisbelow50%.Ifconditionpersists,switchtoyourbackupoxygensourceandcontactyourequipmentprovidertoarrangeforservice.

Message Display & Text

BatteryHOT Warning

NoBreathDetect CheckCannula

SystemError SeeManual

OxygenErrorServiceNeeded

Medium Priority AlertsThefollowingmediumpriorityalertmessagesareaccompaniedbyatriple beep,repeatedevery25seconds,andaflashing red light.

26

Page 31: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Condition/Action/Explanation

Concentratorhasinsufficientbatterypowertoproduceoxygen.Attachexternalpowersupplyorexchangebattery,thenrestartunitifnecessarybypressingOn/Offbutton.

Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Concentratorhasstoppedproducingoxygen.Ifpossible,moveconcentratortoacoolerlocation,thenturnpoweroffandbackon.Ensureairintakeandoutletventshaveclearaccessandparticlefilterisclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Concentratortemperatureistoohighandoxygenproductionisshuttingdown.Ensureairintakeandoutletventshaveclearaccessandparticlefilterisclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Message Display & Text

BatteryEmpty AttachPlug

BatteryHOT ShutDown

SystemHOT ShutDown

High Priority Alerts

CAUTION IfyouarenotneartheInogenOne®G2youmaynotbeabletohearorseethehighpriorityalerts.MakesuretheInogenOne®G2isinalocationwherethealertsandalarmswillberecognizediftheyoccur.

Thefollowinghighpriorityalertmessagesareaccompaniedbyafive beep pattern,repeatedevery10secondsandaflashing red light.

AudibleandVisibleSignalsChapter4 27

English

Page 32: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Condition/Action/Explanation

Thismayresultfromtheconcentratorbeingstoredinacoldenvironment(below0ºC(32ºF)).Movetoawarmerenvironmenttoallowtheunittowarmupbeforestartingit.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.

Concentratorhasstoppedproducingoxygenandisshuttingdown.Youshould:

1. Noteerrornumber2.Switchtobackupoxygensource3.Contactyourequipmentprovider

Message Display & Text

SystemCOLD ShutDown

Error(###) ServiceNeeded

High Priority Alerts (continued)

28

Page 33: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Troubleshooting

TroubleshootingChapter5

Useexternalpowersupplyorreplacebatterywithonethatisfullycharged

Checkpowersupplyconnectionandverifygreenlightissolid

Contactyourequipmentprovider

PressOn/Offbuttontopowerconcentrator

Checkcannulaanditsconnectiontoconcentratornozzle

Batteryisdischargedornobatteryispresent

Powersupplyisnotconnectedproperly

Malfunction

Concentratorisnotpoweredon

Cannulaisnotconnectedproperlyoriskinkedorobstructed

ConcentratordoesnotpoweronwhenOn/Offbuttonispressed

Nooxygen

recommended Solution

RefertoChapter4

Possible Cause

RefertoChapter4

Problem

Anyproblemaccompaniedbyinformationonconcentratordisplay,indicatorlightsand/oraudiblesignals

Solutionstosomepossibleissuesyoumayencounteraredescribedinthissection.

Inogen One® G2 Oxygen Concentrator

29

5English

Page 34: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external
Page 35: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Cannula replacement

Yournasalcannulashouldbereplacedonaregularbasis.Consultwithyourphysicianand/orequipmentproviderand/orcannulamanufacturer’sinstructionsforreplacementinformation.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.

CAUTION Useofacannulaotherthanahighflowcannula(e.g.Salter1600Q)mayrestrictoxygendeliveryand/orattachmenttonozzlefittings.

Case CleaningYoumaycleantheoutsidecaseusingaclothdampenedwithamildliquiddetergent(suchasDawnTM)andwater.

wArNING DonotsubmersetheInogenOne®G2oritsaccessoriesinwaterorallowwatertoenterintothecase;thismayleadtoelectricalshockand/ordamage.

wArNING DonotusecleaningagentsotherthanthosespecifiedinthisUserManual.Donotusealcohol,isopropylalcohol,ethylenechlorideorpetroleumbasedcleanersonthecasesorontheparticlefilters.

Cleaning, Care and Maintenance

31Cleaning,CareandMaintenanceChapter6

6

! C

! C

English

Page 36: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Filter Cleaning and replacementTheparticlefiltermustbecleanedweeklytoensuretheeaseofairflow.Removefilterfromthefrontofthedevice.Cleantheparticlefilterwithamildliquiddetergent(suchasDawnTM)andwater;rinseinwateranddrybeforereuse.

NOTE Itmaybenecessarytocleantheparticlefiltermoreoftenindustyenvironments.

TopurchaseadditionalparticlefilterscontactyourequipmentproviderorInogen.

Output FilterTheoutputfilterisintendedtoprotecttheuserfrominhalationofsmallparticlesintheproductgasflow.TheInogenOne®G2includesanoutputfilterconvenientlylocatedbehindtheremovablecannulanozzlefitting.Inogenrequiresthatthisfilterbereplacedbetweenpatients.

TheoutputfiltermaybereplacedbytheequipmentproviderorbytheownerusingtheOutputFilterReplacementKit(RP-107).

DC Input Cable Fuse replacementTheCigaretteLighterDCPowerPlugcontainsafuse.IftheDCInputCableisbeingusedwithaknowngoodpowersourceandthepowersupplyisnotoperative(nooutputpowerandthegreenLEDisnotilluminated),thefusemayneedtobereplaced.

32

Page 37: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

FuseCigarette Adapter Plug retainer Tip retainer

CAUTION Forcontinuedprotectionagainstriskoffire,onlyusefusespecified.

33Cleaning,CareandMaintenanceChapter6

Toreplacethefuse,followtheseinstructionsandrefertothephotographbelow.1. Removethetipbyunscrewingtheretainer.Useatoolifnecessary.2. Removetheretainer,tipandfuse.3. ThespringshouldremaininsidetheCigaretteLighterAdapterhousing.Ifthespring

isremoved,makesuretoreplacethespringfirstbeforeinsertingthereplacementfuse.

4. Installareplacementfuse,InogenRP#125(BUSSMDA-12)andreassemblethetip.Ensuretheretainerringisproperlyseatedandtightened.

English

Page 38: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

34

Disposal of Equipment and Accessories

FollowyourlocalgoverningordinancesfordisposalandrecyclingoftheInogenOne®G2andaccessories.IfWEEEregulationsapply,donotdisposeofinunsortedmunicipalwaste.WithinEurope,contacttheEUAuthorizedRepresentativefordisposalinstructions.Thebatterycontainslithiumioncellsandshouldberecycled.Thebatterymustnotbeincinerated.

Maintenance Items List• InogenOne®G2singlebattery(model#BA-200)• InogenOne®G2doublebattery(model#BA-224)• Replacementintakeparticlefilters(model#RP-200)• OutputFilterReplacementKit(model#RP-107)

Other Service and Maintenance wArNING DonotdisassembletheInogenOne®G2oranyoftheaccessories

orattemptanymaintenanceotherthantasksdescribedinthetroubleshootingsection;disassemblycreatesahazardofelectricalshockandwillvoidyourwarranty.Donotremovetamperevidentlabel.Foreventsotherthanthosedescribedinthismanual,contactyourequipmentproviderforservicingbyauthorizedpersonnel.

CAUTION DonotuselubricantsontheInogenOne®G2oritsaccessories.

! C

Page 39: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Symbols Used On Concentrator and Accessories

Meaning

Awarningindicatesthatthepersonalsafetyofthepatientmaybeinvolved.Disregardingawarningcouldresultinsignificantinjury.

Acautionindicatesthataprecautionorserviceproceduremustbefollowed.Disregardingacautioncouldleadtoaminorinjuryordamagetoequipment.

SeeUserManualforInstructions

U.S.FederalRegulationRestrictsthisDevicetoSalebyOrderofPhysician.MayalsobeapplicableinotherCountries.

ACPower

DCPower

NoSmokingwhiledeviceisinuse.

NoOpenFlames(Concentrator);Donotincinerate(Battery)

Symbol

WARNING

CAUTION

ONLY

ONLY

ONLY

ONLY

ONLY

SymbolsUsedonConcentratorandAccessoriesChapter7 35

7English

Page 40: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Meaning

KeepDry

IndoororDryLocationUseOnly,DoNotGetWet

UseNoOilorGrease

DoNotDisassemble

DoNotDisposeofInUnsortedMunicipalWaste

TypeBFAppliedPart,NotIntendedforCardiacApplication

ClassIIDevice

ElectricalSafetyAgencyCertificationLogo

CompliesWithApplicableEUDirectivesIncludingMedicalDeviceDirective

Symbol

ONLY

User Interface Label

Meaning

ON/OFFButton

DisplayBacklightButton

IncreaseFlowSetting

DecreaseFlowSetting

AudibleAlarmButton

Symbol

36

ONLY

ONLY

ONLYONLY

0123

93-01213-00-00 rev1

0123

COLOR SUMMARY: 1. Black

Page 41: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Inogen One® G2 ConcentratorDimensions:Withsinglebattery:

L/W/H: 10.7in(27.2cm)/3.9in(9.9cm)/8.75in(22.2cm)L/W/H: 10.7in(27.2cm)/3.9in(9.9cm)/9.5in(24.2cm)

Weight: 7.25pounds(3.3kg)(includessinglebattery)Noise: Lessthan38dBA(aspackaged)onsetting2Warm-UpTime: 2minutesOxygenConcentration: 90%-3%/+6%atallsettingsFlowControlSettings: 5settings:1to5Power: UniversalPowerSupply:

ACInput: 100to240VAC50to60Hz

Auto-Sensing: 1.0A

DCInput: 13.5-15VDC,10ADCOutput: 19VDC,5.0Amax.

RechargeableBattery: Voltage: 12.0to16.8VDC

BatteryDuration: Upto4hourswithsinglebatteryUpto8hourswithdoublebattery

BatteryChargingTime: Upto4hoursforasinglebatteryUpto8hoursforadoublebattery

EnvironmentalRangesIntendedforUse:

Temperature:41to104˚F(4to40˚C)Humidity:0%to95%,non-condensingAltitude:0to10,000ft(0to3048meters)

EnvironmentalRangesIntendedforStorage:

Temperature:-4to140˚F(-20to60˚C)Humidity:0%to95%,non-condensingStoreinadryenvironment

Transportation: KeepDry,HandleWithCare

InogenOne®G2SystemSpecificationsChapter8

Inogen One® G2 System Specifications8

37

English

Page 42: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Inogen One® G2 Concentrator (continued)

ClassificationsModeofOperation: ContinuousDutyTypeofProtectionAgainstElectricalShock:

ClassII

DegreeofProtectionAgainstElectricalShock:

TypeBFNotintendedforcardiacapplication

DegreeofProtectionAgainstIngressofWater:

IP20

DegreeofSafetyforApplicationinPresenceofAnestheticGases:

Notsuitableforsuchapplication

38

ELECTrOMAGNETIC COMPATIBILITy

ThisCEMarkedequipmenthasbeentestedandfoundtocomplywiththeEMClimitsfortheMedicalDeviceDirective93/42/EEC[EN55011ClassBandEN60601-1-2].Theselimitsaredesignedtoprovidereasonableprotectionagainstharmfulinterferenceinatypicalmedicalinstallation.

Theequipmentgeneratesusesandcanradiateradiofrequencyenergyandifnotinstalledandusedinaccordancewiththeinstructions,maycauseharmfulinterferencetootherdevicesinthevicinity.

TestedbyIndependentLaboratory:

Safety:IEC60601-1CAN/CSAC22.2No.60601-1ElectromagneticCompatibility:IEC60601-1-2RTCADO160

Page 43: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

InogenOne®G2SystemSpecificationsChapter8 39

However,thereisnoguaranteethatinterferencewillnotoccurinaparticularinstallation.Ifthisequipmentdoescauseharmfulinterferencewithotherdevices,whichcanbedeterminedbyturningtheequipmentoffandon,theuserisencouragedtotrytocorrecttheinterferencebyoneormoreofthefollowingmeasures:

1. Reorientorrelocatethereceivingdevice.2. Increasetheseparationbetweentheequipment.3. Connecttheequipmentintoanoutletonacircuitdifferentfromthatto

whichtheotherdevice(s)isconnected.4. ConsultInogenoryourequipmentproviderforassistance.

CAUTION PortableandmobileRFcommunicationsequipmentcanaffectmedicalelectricalequipment.

CAUTION Stackingorplacingequipmentadjacenttootherdevicesisnotrecommended,wheresuchconfigurationsarenecessary,allequipmentinquestionshouldbecarefullyobservedtoensurethatEMIdoesnotdegradeperformance.

CAUTION Useofaccessories,transducersandcablesotherthanthosespecified,withtheexceptionoftransducersandcablessoldbythemanufactureroftheequipmentasreplacementpartsforinternalcomponents,mayresultinincreasedemissionsanddecreasedimmunityoftheequipmentorsystem.

English

Page 44: Inogen One @ OxiMedical · 2016-01-13 · CAUTION The Inogen One® G2 battery acts as a secondary power supply in the event of a planned or unexpected loss of the AC or DC external

Guidance and Manufacturer’s Declaration – Electromagnetic EmissionsTheInogenOne®G2OxygenConcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheInogenOne®G2OxygenConcentratorshouldassurethatitisusedinsuchanenvironment.

Emissions Test Compliance Electromagnetic Environment - Guidance

RFemissionsCISPR11

Group1 TheInogenOne®G2OxygenConcentratorusesRFenergyonlyforitsinternalfunction.ThereforeitsRFemissionsareverylowandnotlikelytocauseanyinterferenceinnearbyequipment.

RFemissionsCISPR11

ClassB TheInogenOne®G2OxygenConcentratorissuitableforuseinallestablishmentsincludingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.

HarmonicEmissionsIEC61000-3-2

ClassA

Voltagefluctuations/flickeremissionsIEC61000-3-3

Complies

40