Inogen One G4 User manual
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
Thank you for choosing an Oxycure concentrator.
More information at w w w .ox ycure.be
The Oxycure range, from left to right
Oxycure portable concentrators Oxycure portable concentrator – Inogen G4 Oxycure portable concentrator – Inogen G3 Oxycure portable concentrator – Zen-O lite Oxycure portable concentrator – Simply Go Mini Oxycure portable concentrator – SimplyGo Oxycure portable concentrator – eQuinox
Oxycure concentratorsOxycure – Kröber concentrator
Oxycure cylinders (gaseous oxygen) Oxycure B2 cylinder – 0.4 m³Oxycure B5 cylinder – 1.0 m³
Oxycure tanks (liquid oxygen) Oxycure liquid oxygen tank – 25.6 m³Oxycure portable liquid oxygen tank (1.0 m³)
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
OXYCURE : ORDERS 081 22 15 90 Oxycure Belgium plc. Business Park Fernelmont L. Génicot, 9B-5380 FernelmontBelgiumTel + 32 (0)81 22 15 90Fax + 32 (0)81 22 15 99
http://www.oxycure.be e-mail : [email protected]
OPENING HOURS From Monday to Friday 09.00 – 12.30 and 13.30 – 18.00
CUSTOMER SERVICE 0800 98 0 68
Our customer service operates 24/7.
It is exclusively to urgent installations or repairs on our medical devices.
Call to guard service Leave a message on the answering machine with your name and phone number and give a brief reason of
your call ; The customer service will call you within 20 minutes ; If the customer service did not call you within 20 minutes, call again.
Notes 1. Technical interventions (repairs) are included in the price of the assistance of the leasing.2. Deliveries of goods, outside the opening hours, will be charged.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
1
USER MANUAL
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
2
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
3
Contents
Chapter 1 1 Intended Use, Contraindications and General Precautions
Chapter 2 3 Description of the Inogen One® G4 Oxygen Concentrator3 ImportantPartsoftheInogenOne®G4OxygenConcentrator4 UserInterfaces5 Input/OutputConnections6 PowerSupplyOptions8 InogenOne®G4Accessories
Chapter 3 11 Operating Instructions11 GeneralInstructions15 AdditionalOperatingInstructions19 BatteryOperatingInstructions20 BatteryCareandMaintenance
Chapter 4 21 Inogen One® G4 Oxygen Concentrator Audible and Visible Signals (including Alerts)
Chapter 5 29 Troubleshooting
Chapter 6 31 Cleaning, Care and Maintenance31 CannulaReplacement31 CaseCleaning31 FilterCleaningandReplacement33 InogenOne®G4ColumnChangeProcedure36 OtherServiceandMaintenance36 DisposalofEquipmentandAccessories
Chapter 7 37 Symbols Used on Concentrator and Accessories
Chapter 8 39 Inogen One® G4 System Specifications
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
4
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
1
1 Intended Use, Contraindications and General Precautions
IntendedUse,ContraindicationsandGeneralPrecautionsChapter1
Intended Use
TheInogenOne®G4OxygenConcentratorisusedonaprescriptivebasisbypatientsrequiringsupplementaloxygen.Itsuppliesahighconcentrationofoxygenandisusedwithanasalcannulatochanneloxygenfromtheconcentratortothepatient.TheInogenOne®G4maybeusedinhome,institution,vehicleandvariousmobileenvironments.
CAUTION USAFederallawrestrictsthisdevicetosalebyorontheorderofaphysician.Mayalsobeapplicableinothercountries.
CAUTION Nasalcannulashouldberatedfor3litersperminute(e.g.SalterLabs16SOFT)toensureproperpatientusageandoxygendelivery.
WARNING Availabilityofanalternatesourceofoxygenisrecommendedincaseofpoweroutageormechanicalfailure.Consultyourequipmentproviderfortypeofback-upsystemrecommended.
CAUTION Itistheresponsibilityofthepatienttomakeback-uparrangementsforalternativeoxygensupplywhentraveling;Inogenassumesnoliabilityforpersonschoosingnottoadheretomanufacturerrecommendations.
TheexpectedlifefortheInogenOne®G4OxygenSystemis5years,withtheexceptionofthesievebeds(metalcolumns)whichhaveanexpectedlifeof1yearandthebatteries,whichhaveanexpectedlifeof500fullcharge/dischargecycles.
! C
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
2
Contraindications
WARNING ThisdeviceisNOTINTENDEDtobelifesustainingorlifesupporting.
CAUTION Undercertaincircumstances,theuseofnon-prescribedoxygentherapycanbehazardous.Thisdeviceshouldbeusedonlywhenprescribedbyaphysician.
CAUTION Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealertsorcommunicatediscomfort.Ifthepatientshowsanysignsofdiscomfort,aphysicianshouldbeconsultedimmediately.
CAUTION TheInogenOne®G4isnotdesignedorspecifiedtobeusedinconjunctionwithahumidifier,nebulizerorconnectedwithanyotherequipment.Useofthisdevicewithahumidifier,nebulizerorconnectedwithanyotherequipmentmayimpairperformanceand/ordamagetheequipment.DonotmodifytheInogenOne®G4Concentrator.Anymodificationsperformedontheequipmentmayimpairperformanceordamageequipmentandwillvoidyourwarranty.
! C
! C
! C
General Precautions
WARNING Thedeviceproducesenrichedoxygengas,whichacceleratescombustion.DoNOTALLOWSMOKINGOROPENFLAMESwithin10feetofthisdevicewhileinuse.
WARNING DonotsubmersetheInogenOne®G4oranyoftheaccessoriesinliquid.Donotexposetowaterorprecipitation.Donotoperateinexposedrain.Thiscouldleadtoelectricalshockand/ordamage.
CAUTION Donotuseoil,grease,orpetroleum-basedproductsonorneartheInogenOne®G4.
CAUTION NeverleavetheInogenOne®G4inanenvironmentwhichcanreachhightemperatures,suchasanunoccupiedcarinhightemperatureenvironments.Thiscoulddamagethedevice.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
3
Important Parts of the Inogen One® G4 Oxygen Concentrator
Description of the Inogen One® G4 Oxygen Concentrator2
On/Off Button
Breath Detection Light
*Display
Audible Alert Button Alert/Alert Light
Flow Control
DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
English
DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
*Actualdisplayappearancemayvary.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
4
User Controls
ON / OFF ButtonPressoncetoturn“ON”;Pressandholdforonesecondtoturn“OFF”.
Audible Alert Button PressingthisbuttonwilltoggletheInogenOne®G4’sbreathdetectionaudiblealertonandoff.
BreathDetectionAlertMode.TheInogenOne®G4willalertwithaudibleandvisualsignalsfor“nobreathdetected”whenthismodeisenabledandnobreathhasbeendetectedfor60seconds.
At60seconds,thedevicewillenterintoautopulsemodeandonceanotherbreathisdetected,thedevicewillexitautopulsemodeanddelivernormallyoninspiration.Thedisplay’smodeindicationareawillshowabellicon,flashingyellowlightanddisplaymessagewhenthealertisenabled.
Ifpowerislost,thebreathdetectionaudiblealertremainssetintheuserpreferredmode.
Flow Setting Control Buttons Usethe–or+flowsettingcontrolbuttonstoselectthesettingasshownonthedisplay.Therearethreesettings,from1to3.
User Interfaces
DisplayThisscreendisplaysinformationregardingflowsetting,powerstatus,batterylifeanderrors.IfyouwouldliketochangethelanguageontheInogenLCDScreencontactyourEquipmentProviderforinstructions.Actualdisplayappearancemayvary.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
5DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
User Interfaces (continued)
Indicator LightsAgreenlightindicatesbreathdetection.Ayellowlightindicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alert).Aflashinglightishigherprioritythannon-flashing.
Audible SignalsAnaudiblesignal(beep)indicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alert).Morefrequentbeepsindicatehigherpriorityconditions.
BacklightAbacklightwillilluminatethescreenfor15secondswhentheon/offbuttonisbrieflypressed.
Input / Output Connections
Particle FilterThefiltermustbeinplaceattheintakeendoftheconcentratorduringoperationtokeepinputairclean.
Cannula Nozzle FittingThenasalcannulaconnectstothisnozzleforInogenOne®G4outputofoxygenatedair.
DC Power InConnectionforexternalpowerfromtheACpowersupplyorDCpowercable.
USB PortForserviceuseonly.
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
6
Power Supply Options
Single and Double Rechargeable Lithium Ion BatteriesThebatterywillpowertheInogenOne®G4withoutconnectiontoanexternalpowersource.Whenfullycharged,asinglebatterywillprovideupto2.7hoursofoperation;adoublebatterywillprovideupto5hoursofoperation.ThebatteryrechargeswhenproperlyinstalledintheInogenOne®G4andtheconcentratorisconnectedtoACorDCpower.Rechargingtimeisupto3hoursforasinglebatteryand5hoursforadoublebattery.Seeinformationinthe“BatteryCareandMaintenance”section.
Power SupplyOverviewTheInogenOne®G4ACpowersupply(BA-401)isusedtopowertheInogenOne®G4concentratorfromanACpowersource.
DescriptionTheInogenOne®G4ACpowersupplyisspecificallydesignedforusewiththeInogenOne®G4OxygenConcentrator(IO-400).TheACpowersupplyprovidestheprecisecurrentandvoltagerequiredtosafelypowertheInogenOne®G4andisdesignedtooperatefromspecifiedACpowersources.WhenusedwithACpowersources,thepowersupplyautomaticallyadaptstoinputvoltagesfrom100Vto240V(50-60HZ)permittingusewithmostpowersourcesthroughouttheworld.
TheACpowersupplywillchargetheInogenOne®G4BatterieswhenusedwithACinputpower.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G4Batterywhenusedonanaircraft.TheACpowersupplyisusedwiththefollowingcomponents:
•PowersupplywithattachedpoweroutputcabletoconnecttotheInogenOne®G4•ACpowerinputcable
TheDCpowercable(BA-306)isspecificallydesignedforusewiththeInogenOne®G4OxygenConcentrator(IO-400).TheDCpowerinputcableconnectsdirectlytotheautomobilecigarettelighterorauxiliaryDCpowersupply.RefertoChapter3foradditionaloperatinginstructionsforusewithDCpowersource.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
7DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
English
*Actualproductappearancemayvary.
AC Power Plug*
Power Output Plug (Plug into Concentrator)
LED Indicator Light*AC Input
InogenOneACPowerSupplyModel#BA-401
InogenOneG4DCPowerCableModel#BA-306
DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft
Power Output Plug (Plug into Concentrator)
WARNING Donotusepowersuppliesorpowercablesotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedpowersuppliesorpowercablesmaycreateasafetyhazardand/orimpairequipmentperformance.Donotwrapcordsaroundpowersupplyforstorage.Donotdrive,dragorplaceobjectsovercord.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.Toavoiddangerofchokingorstrangulationhazard,keepcordsawayfromchildrenandpets.
! C
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
8
Inogen One® G4 Accessories
WARNING Donotusepowersupplies/adaptersoraccessoriesotherthanthosespecified inthisusermanual.Theuseofnon-specifiedaccessoriesmaycreateasafety hazardand/orimpairequipmentperformance.
Nasal CannulaAnasalcannulamustbeusedwiththeInogenOne®G4toprovideoxygenfromtheconcentrator.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.
WARNING Toavoiddangerofchokingandstrangulationhazard,keeptubingawayfromchildrenandpets.
NOTE Increasingthecannulalengthmayreducetheperceivednoiseduringoxygenbolusdelivery.
NOTE Whenusingacannula25ft.inlengthwiththeInogenOne®G4,anincreaseinflowsettingmayberequired.
Inogen One G4 Carry Strap (CA-401)Thecarrystrapisdesignedtoattachdirectlytotheconcentrator.Thereisametalslideforeaseoflengthadjustments,ashoulderpadforcomfortandasewnonvelcrotabtowrapupthatextracannulatubing.
! C
! C
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
9
Carry Bag (CA-400)TheCarryBagprovidesaprotectivecoverwithahandleandshoulderstraptoenableyoutocarrytheInogenOne®G4.TheInogenOne®G4canbeoperatedusingbatterypowerduringtransportwiththeCarryBag.
Inogen One® G4 Optional Accessories
External Battery Charger (BA-403)
TheInogenOne®G4externalbatterychargerwillchargetheInogenOne®G4singleanddoublebatteries.
1.PlugtheExternalBatteryChargerACpowersupplycordintoanelectricaloutlet.
2.PlugtheExternalBatteryChargerACpowersupplyintothebatterycharger.
3.SlideyourchargerontotheInogenOneG4Battery byclickingandlockingintothecharger.4.Whenthebatteryisinthecorrectposition,asolid redlightwillindicatethatthebatteryischarging.5.Whenthegreenlightilluminates,thebatteryis fullycharged.
CAUTION AvoidtouchingtherecessedelectricalcontactsoftheExternalBatteryCharger;damagetocontactsmayaffectchargeroperation.
NOTE Thesecontactsarenotpoweredunlessabatteryisinplaceandcharging.
NOTE TocompletelyremovepowerfromtheExternalBatteryCharger, removetheplug.
DescriptionoftheInogenOne®G4OxygenConcentratorChapter2
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
1010
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
11OperatingInstructionsChapter3
English
Operating Instructions3
! C
General Instructions
1. Place the Inogen One® G4 in a well ventilated location.Airintakeandexhaustmusthaveclearaccess.LocatetheInogenOne®G4insuchawaythatanyauditoryalertsmaybeheard.AlwaysoperatetheInogenOne® G4inanuprightposition(seeimageforproperorientation).
WARNING AvoiduseoftheInogenOne®G4inpresenceofpollutants,smokeorfumes.DonotusetheInogenOne®G4inpresenceofflammableanesthetics,cleaningagentsorotherchemicalvapors.
CAUTION Donotobstructairintakeorexhaustwhenoperatingthedevice.
Blockageofaircirculationorproximitytoaheatsourcemayleadtointernalheatbuildupandshutdownordamagetotheconcentrator.
CAUTION TheInogenOne®G4Concentratorisdesignedforcontinuoususe.Foroptimalsievebed(columns)life,theproductshouldbeusedfrequently.
2. Ensure particle filter is in place.
CAUTION DonotoperatetheInogenOne®G4withouttheparticlefilterinplace.Particlesdrawnintothesystemmaydamagetheequipment.
Intake
ExhaustExhaust
ThissideUP
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
12
3. Install the battery. InserttheInogenOne®G4batterybyslidingbatteryintoplaceuntilthelatchreturnstotheupperposition.
CAUTION TheInogenOne®G4batteryactsasasecondarypowersupplyintheevent ofaplannedorunexpectedlossoftheACorDCexternalpowersupply.When
operatingtheInogenOne®G4fromanACorDCexternalpowersupply,aproperlyinsertedInogenOne®G4batteryshouldbemaintainedintheunit.Thisprocedurewillensureuninterruptedoperationandwilloperateallalertsandalertsintheeventofalossoftheexternalpowersupply.
4. Connect the Power Supply. ConnecttheACinputplugtothepowersupply.ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplugtotheInogenOne®G4.ThegreenLEDonthepowersupplywillbeilluminatedandabeepwillsoundfromtheconcentrator.
CAUTION Ensurethepowersupplyisinawellventilatedlocationasitrelieson aircirculationforheatdissipation.Thepowersupplymaybecomehotduring operation.Makesurethepowersupplycoolsdownbeforehandling.
CAUTION Thepowersupplyisnotwaterresistant.
CAUTION Donotdisassemblethepowersupply.Thismayleadtocomponentfailure and/orsafetyrisk.
CAUTION Donotplaceanythinginthepowersupplyportotherthanthesuppliedwallcord.AvoidtheuseofelectricalextensioncordswiththeInogenOne®G4.
Ifanextensioncordmustbeused,useanextensioncordthathasanUnderwritersLaboratory(UL)Markandaminimumwirethicknessof
18gauge.Donotconnectanyotherdevicestothesameextensioncord.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
13OperatingInstructionsChapter3
NOTE Undercertainconditions(seeTechnicalSpecifications)thepower supplymayshutdown.ThegreenLEDwillblinkorwillnolongerbe
illuminated.Ifthisoccurs,disconnectthepowersupplyforatleast10secondsandreconnect.
NOTE WhenthepowersupplyisdisconnectedfromtheACoutlet,alsodisconnectitfromtheconcentratortoavoidunnecessarybatterydischarge.
5. Connect the nasal cannula tubing to the nozzle fitting.NozzlefittingislocatedonthetopoftheInogenOne®G4.Useofasinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.Additionaltitrationmaybeneededtoensureproperoxygendeliverywhenusingaparticularcannula.
CAUTION Toensureoxygenflow,ensurethatthenasalcannulaisproperlyconnectedtothenozzlefittingandthatthetubingisnotkinkedorpinchedinanyway.
CAUTION Replacethenasalcannulaonaregularbasis.Checkwithyourequipmentproviderorphysiciantodeterminehowoftenthecannulashouldbereplaced.
6. Turn on your Inogen One® G4 by pressing the ON/OFF Button.AsingleshortbeepwillsoundaftertheInogenlogoisdisplayed.“PleaseWait”willappearwhiletheconcentratorstartsup.Thedisplaywillindicatetheselectedflowsettingandpowercondition.Followingabriefstart-upsequence,awarmupperiodupto2minuteswillinitiate.Duringthistimeperiodtheoxygenconcentrationisbuildingtobutmaynothavereachedspecification.AdditionalwarmuptimemaybeneededifyourInogenOne®G4hasbeenstoredinextremelycoldtemperatures.
7. Set the Inogen One® G4 Concentrator to the flow rate prescribed by your physician or clinician. Usethe+or–settingbuttonstoadjusttheInogenOne®G4tothedesiredsetting.Thecurrentsettingcanbeviewedonthedisplay.
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
14
! C
8. Position the nasal cannula on your face and breathe through your nose. TheInogenOne®G4willsensetheonsetofinhalationanddeliveraburstofoxygenataprecisetimewhenyouinhale.TheInogenOne®G4willsenseeachbreathandcontinuetodeliveroxygeninthismanner.Asyourbreathingratechanges,theInogenOne®G4willsensethesechangesanddeliveroxygenonlyasyouneedit.Attimes,ifyouinhaleveryquicklybetweenbreathstheInogenOne®G4mayignoreoneofthebreaths,givingtheappearanceofamissedbreath.ThismaybenormalastheInogenOne®G4sensesandmonitorsthechangesinyourbreathingpattern.TheInogenOne®G4willnormallysensethenextbreathanddeliveroxygenaccordingly.
Agreenlightwillflasheachtimeabreathisdetected.Makecertainthatthenasalcannulaisproperlyalignedonyourfaceandyouarebreathingthroughyournose.
WARNING Ifyoubegintofeelillorareexperiencingdiscomfortwhileusingthisdevice,consultyourphysicianimmediately.
CAUTION TheInogenOne®G4isdesignedtoprovideaflowofhighpurityoxygen.An
advisoryalert,“OxygenLow”,willinformyouifoxygenconcentrationdrops.Ifalertpersists,contactyourequipmentprovider.
GeneralToremovepower,unplugtheinputcordfromitssource(i.e.,ACwalloutlet,DCcarcigarettelighteradapter)anddisconnectitfromtheInogenOne®G4.
CAUTION Ensurethepowersupplyispoweredfromonlyonepowersource(ACorDC)atanygiventime.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
15OperatingInstructionsChapter3
Additional Operating Instructions
For Use In Home with AC PowerForoperationofthepowersupplyusinganACpowersource,followtheseinstructions:
1. ConnecttheACinputtothepowersupply.2. ConnecttheACpowerplugtothepowersourceandconnectthepoweroutputplug
totheInogenOne®G4.ThegreenLEDwillbeilluminated,indicatingthe powersupplyhasinputpower.
Traveling with your Inogen One® G4 System
TheInogenOne®G4Systemmakestravelbyboat,carortrainmoreconvenientforoxygenusersthaneverbefore.Nowyougetthesamequalityperformanceandconveniencewhileonthegothatyou’reusedtoreceivingfromyourInogenOne®G4athome.HerearesomeusefulandimportantinstructionsformaximizingperformanceandconveniencewhenusingyourInogenOne®G4whiletraveling.
Youshouldbeginplanningforatripwithachecklistofitemstoremember.Thischecklistshouldinclude:
✓ ACpowersupplyandDCpowercable✓ ExtraBattery(s)ifrequired✓ Importantphonenumbers,suchasthoseofyourphysicianandhomehealthcare provider,orprovidersintheareayouwillbetraveling✓ Planforbackupoxygenintheeventofaprolongedpoweroutageor mechanicalfailure
For Use In Automobile/RV/Boat/AircraftForoperationusingaDCpowersource,followtheseinstructions:
1.ConnecttheDCpoweroutputplugtotheInogenOne®G4.2.ConnecttheDCpowerplug(CigaretteLighterAdapter)to
thepowersource.3. Theplugshouldinsertintothesocketwithoutexcessive
forceandsecurelyremaininplace.
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
16
4. Checkthedevicedisplayscreentoconfirmthatanexternalpowersupplyisconnected. AniconofabatterywithalightningboltoranACpowerplugwillbedisplayedonthe
screen,indicatingthatexternalpowerisconnectedandfunctioningproperly.
WARNING EnsurethattheautomobilepowersocketisadequatelyfusedfortheInogenOne®G4powerrequirements(minimum15Amp).Ifthepowersocketcannotsupporta15Ampload,thefusemaybloworthesocketmaybedamaged.
WARNING ThetipoftheCigaretteAdapterPlugbecomesHOTwheninuse.Donottouchthetipimmediatelyafterremovalfromanautocigarettelightersocket.
CAUTION Ensuretheautomobilepowersocketiscleanofcigaretteashandtheadapterplugfitsproperly,otherwiseoverheatingmayoccur.
CAUTION Donotusethepowersupplywithacigaretteplugsplitterorwithan extensioncable.ThismaycauseoverheatingoftheDCpowerinputcable.
CAUTION DonotjumpstarttheautomobilewiththeDCpowercableconnected.Thismayleadtovoltagespikeswhichcouldshutdownand/ordamagetheDCpowerinputcable.
CAUTION WhenpoweringtheInogenOne®G4inanautomobileensurethevehicle’sengineisrunningfirst,beforeconnectingDCcableintocigarettelighteradapter.Operatingthedevicewithouttheenginerunningmaydrainthevehicle’sbattery.
CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.
Traveling By AirTheInogenOne®G4conformstoallapplicableFAArequirementsforPOCuseonboardanaircraft.
Planning Your FlightWhenflyingwiththeInogenOne®G4,youmustinformtheairlinethatyouwillbeusingyourInogenOne®G4onboardtheaircraft.YoushouldcontacttheSpecialAssistancedeskoftheairlineonwhichyouwillbetravelingtolearnofanypaperwork,etc.thatmightbeneeded.
! C
! C
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
17OperatingInstructionsChapter3
English
•Someairlinesmayequiptheiraircraftwithonboardelectricalpower.YoumayhaveanopportunitytorequestaseatwithapowerportwhichcanbeusedtopoweryourInogenOne®G4.However,availabilityvariesbyairline,typeofaircraftandclassofservice.Youshouldcheckwithyourairlinesforavailabilityandalwaysplanonhavingsufficientbatterypowerfornolessthan150%oftheexpecteddurationoftheflight.Airlinesmayhavespecificrequirementsforbatterylifeduration,socheckwiththeairlinebeforetraveling.
•YourDCPowerCableisequippedwithacommonlyusedcigarettelighteradapter.However,aircraftsusedifferentpowerportconfigurationsanditisdifficulttodeterminewhichtypeofpowerportcompatibilityyouraircraftmaysupply.It’sagoodideatopurchaseanadapter,foundatelectronicandtravelstores.
Before Your FlightHerearesomethingstokeepinmindthedayyourflightdeparts:
•EnsureyourInogenOne®G4isclean,ingoodconditionandfreefromdamageorothersignsofexcessivewearorabuse.
•BringenoughchargedbatterieswithyoutopoweryourInogenOne®G4fornolessthan150%oftheexpecteddurationoftheflight,plusaconservativeestimateofunanticipateddelays.
•Regional/Commuterairlinesdonotofferonboardelectricalpower.Ifyourtravelplanscallforflightsonregionalairlines,bringenoughchargedbatterieswithyoutopoweryourInogenOne®G4fornolessthan150%oftheexpecteddurationoftheflight,plusaconservativeestimateofunanticipateddelays.
•Arriveattheairportearly.AirportsecurityscreeningpersonnelmayrequireextratimetoinspectyourInogenOne®G4.
•Whilewaitingtoboardyourflight,youmaybeabletoconservebatterypowerbyusingtheACPowerSupplytopoweryourInogenOne®G4fromanelectricaloutletintheairportterminalifavailable.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
18
During Your Flight1. Ifusingairlinepowerport,removetheBatteryfromtheInogenOne®G4Oxygen
Concentrator.Duetoaircraftpowerlimitations,theACPowerSupplycannotbeusedtochargetheInogenOne®G4Batterywhenonboardaircraft.
2. ConnecttheDCPowerPlugforavailableairlinepower.Checkwithairlinepersonneltoensurecompatibility.
•Duringtaxi,takeoffandlanding,stowyourInogenOne®G4undertheseatinfrontofyou.YourInogenOne®G4willfituprightundermostairlineseats.
•ItisnotnecessarytoturnoffyourInogenOne®G4duringtaxi,takeoffandlanding.
CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.TheInogenOne®G4hasbeenverifiedtoprovideoxygentospecificationupto10,000ft.(3048m).Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.
After Your Flight•Remembertorechargeadditionalbatteriesyoumayhaveusedpriortoyournextflight.
Traveling By Bus, Train or BoatContactyourcarriertofindoutaboutpowerportavailability.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
19
! C
English
OperatingInstructionsChapter3
Battery Operating InstructionsEnsurethatthebatteryisinplaceandcharged.DisconnecttheInogenOne®G4fromitspowersource.WhiletheInogenOne®G4isoperatingonbatterypower,thebatterywilldischarge.Thedisplaywillindicatetheestimatedremainingpercentage(%)orminutesofuse.
Whentheconcentratordetectsthatthebatterylifeislow,withlessthan10minutesremaining,alowpriorityalertwillsound.Whenthebatteryisempty,thealertwillchangetoahighpriority.
Whenbatterylifeislow,dooneofthefollowing:• PlugtheInogenOne®G4intoanACorDCpowersourceusingtheACpowersupplyor DCcable.•ReplacethebatterywithachargedbatteryafterturningofftheInogenOne®G4 (bypressingtheON/OFFbutton).Toremovebatterypressandholdthebatterylatch buttonandslidebatteryofftheconcentrator.• Ifthebatteryisdrained,chargethebatteryorremoveitfromtheconcentrator.
IftheInogenOne®G4isbeingpoweredbytheACpowersupplyorDCpower,batterieswillchargeduringoperation.LeavingyourInogenOne®G4pluggedinpastthefullchargetimewillnotharmtheconcentratororthebattery. WARNING Itistheresponsibilityofthepatienttoperiodicallycheckthebatteryand replaceasnecessary.Inogenassumesnoliabilityforpersonschoosingnotto adheretomanufacturersrecommendations.
Normal Battery ChargingToensurethatyourbatteryisproperlycharging,inspectthatthecorrectACandDCpoweroutputplugadapterisbeingusedandthattheadapterisproperlyinsertedintothepoweroutlet.Observethedisplayorlightsthatindicatechargingstatus.
NOTE Whenstartingtochargeafullydischargedbattery,thechargingprocessmay startandstopduringthefirstfewminutes.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
20
Battery Care and MaintenanceYourInogenOne®G4LithiumIonBatteryrequiresspecialcaretoensureproperperformanceandlonglife.UseonlyInogenOne®G4BatterieswithyourInogenOne®G4Concentrator.
Keep DryAlwayskeepliquidsawayfrombatteries.Ifbatteriesbecomewet,discontinueuseimmediatelyanddisposeofbatteryproperly.
Effect of Temperature on Battery PerformanceTheInogenOne®G4singlebatterypowerstheInogenOne®G4Concentratorupto2.7hoursundermostenvironmentalconditions.Toextendtherun-timeofyourbattery,avoidrunningintemperatureslessthan41˚F(5˚C)orhigherthan95˚F(35˚C)forextendedperiodsoftime.
Battery Time Remaining ClockTheInogenOne®G4continuouslydisplaysbatterytimeremaining.Thisdisplayedtimeisonlyanestimateandtheactualtimeremainingmayvaryfromthisvalue.
Please Follow These Important Guidelines to Maximize Battery Performance and Life:• Storebatteryinacool,dryplace.Storewithachargeof40-50%.• Ifusingmultiplebatteries,makesurethateachbatteryislabeled(1,2,3orA,B,C,etc.) androtateonaregularbasis.Batteriesshouldnotbeleftdormantformorethan 90daysatatime.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
21
Display
TheInogenOne®G4displaycontainspowerstatusicons,modeicons,textwithinformationalmessagesanderrornotifications.
Power Status Icons
Theseiconsareexamplesofthoseshowninthedisplay’swindowwhentheInogenOne®G4isoperatingonbatterypower.
Inogen One® G4 Oxygen Concentrator Audible and Visible Signals
AudibleandVisibleSignalsChapter4
! C
! C
! C
! C
Meaning
Batteryisempty.
Batteryhaslessthan10%chargeremaining.Thisiconflashes.
Batteryhasapproximately40%to50%chargeremaining.
Batteryisfull.
Icon
4English
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
22
Power Status Icons (continued)ThemodeiconsbelowareexamplesofthoseshownwhentheInogenOne®G4isoperatingfromanexternalpowersupplyandchargingthebattery.Thelightningboltindicatesthatanexternalpowersupplyisconnected.
! C
! C
! CMeaning
Batteryischargingwithchargelevelbetween60%and70%.
Thebatteryisfullychargedandischargingasnecessarytomaintainitscharge.
Batteryischargingwithchargelevellessthan10%.
TheInogenOne®G4isoperatingfromanexternalpowersourcewithnobatterypresent.
Icon
Mode IconsThesearethemodeiconsshowninthedisplay’swindow.
Display Text
NOTE Whentwoconditionsoccursimultaneously,theconditionwiththehigherprioritywillbedisplayed.
Meaning
Thebreathdetectionaudiblealerthasbeenenabled.
Thebreathdetectionaudiblealertisdisabled.Thisisthedefaultcondition.
Icon
! C
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
23AudibleandVisibleSignalsChapter4
English
Informational MessagesThefollowinginformationdisplaysarenotaccompaniedbyanyaudiblefeedbackoranyvisualchangeintheindicatorlights.
Condition/Action/Explanation
TheInogenlogoisdisplayedatstartup.
Displayedduringwarmup.“X”representstheselectedflowsetting(eg.,Setting2).
Defaultdisplaywhenoperatingonbatterypower.“X”representstheselectedflowsetting(e.g.,Setting2).“HH:MM”representstheapproximatetimeremainingonthebatterycharge(e.g.,1:45).
Defaultdisplaywhenoperatingonanexternalpowersupplyandthebatteryischarging.“xx%”representsthepercentbatterycharge(e.g.,86%).
Defaultdisplaywhenthebatteryisnotchargingorwhenthetimeremainingisnotavailablefromthebattery.
Displayedwhentheconcentratorispluggedinandbeingusedtochargeabattery(notbeingusedforoxygenproduction).Itisnormaltoseeafullychargedbatteryreadbetween95%and100%whenexternalpowerisremoved.Thisfeaturemaximizestheusefullifeofthebattery.
Message Display & Text
SettingXPleaseWait
SettingXBatteryHH:MM
SettingXChargingxx(or)BatteryFull
SettingXBatteryxx%
Chargingxx%(0r) BatteryFull
Notifications
! C
WARNING Audiblenotifications,rangingfrom55dbato65dbadependingontheusersposition,aretowarntheuserofproblems.Toinsurethataudiblenotificationsmaybeheard,themaximumdistancethattheusercanmoveawayfromitmustbedeterminedtosuitthesurroundingnoiselevel.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
24
Low Priority AlertsThefollowinglowpriorityalertmessagesareaccompaniedbyadouble beepandasolid yellow light.
Condition/Action/Explanation
Batterypowerislow,withlessthan10minutesremaining.Attachexternalpowersupplyorpowerdownandinsertafullychargedbattery.
Columnmaintenanceisrequiredwithin30days.Contactyourequipmentprovidertoarrangeforservice.
Message Display & Text
BatteryLow AttachPlug
Replace Columns
Notifications (continued)TheInogenOne®G4monitorsvariousparametersduringoperationandutilizesanintelligentalertsystemtoindicateamalfunctionoftheconcentrator.Mathematicalalgorithmsandtimedelaysareusedtoreducetheprobabilityoffalsealertswhilestillensuringpropernotificationofanalertcondition.Ifmultiplealertconditionsaredetected,thehighestpriorityalertwillbedisplayed.Notethatfailuretorespondtothecauseofanalertconditionforlow,mediumandhighpriorityalertspotentiallywillresultindiscomfortorreversibleminorinjuryonlyanddevelopwithinaperiodoftimesufficienttoswitchtoabackupsourceofoxygen.Thefollowingnotificationmessagesareaccompaniedbyasingle, short beep.
Condition/Action/Explanation
On/Offbuttonhasbeenpressedfortwoseconds.Concentratorisperformingsystemshutdown.
AudibleAlertbuttonhasbeenpressedforfiveseconds.
Message Display & Text
PleaseWait ShuttingDown
HH:MM Vx.x:SerialNumber
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
25AudibleandVisibleSignalsChapter4
English
Low Priority Alerts (continued)Condition/Action/Explanation
Batteryerrorhasoccurred.Checktheconnectionofyourbatteryandensurethatitisproperlyattachedandlatchedonconcentrator.Ifbatteryerrorrecurswithsamebattery,stopusingthebatteryandswitchtoanewbatteryorremovebatteryandoperateconcentratorusingexternalpowersupply.
Concentratorisproducingoxygenataslightlylowlevel(<82%)foraperiodof10minutes.Ifconditionpersists,contactyourequipmentprovider.
Batteryhasexceededitschargingtemperatureandcharginghasstopped.Thebatterywillnotchargewhilethisalertispresentbutwillbegintochargewhenthebatterytemperaturereturnstothenormaloperatingrange.Ifbatterychargingisdesiredsooner,removethebatteryfromtheconcentratorandallowittocoolinanopenareaforapproximately10-15minutes.Then,re-insertthebatteryintotheInogenOne®G4.Iftheproblemstillpersists,contactyourequipmentprovider.
Theconcentratorrequiresservicingattheearliestconvenience.Theconcentratorisoperatingtospecificationandmaycontinuetobeused.Contactyourequipmentprovidertoarrangeforservice.
Theconcentrator’soxygensensorhasmalfunctioned.Youmaycontinuetousetheconcentrator.Iftheconditionpersists,contactyourequipmentprovider.
Message Display & Text
CheckBattery
OxygenLow
RemoveBatterytoCool
ServiceSoon
SensorFail
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
26
Condition/Action/Explanation
Concentratorhasnotdetectedabreathfor60seconds.Checkthatcannulaisconnectedtoconcentrator,therearenokinksintubingandcannulaispositionedproperlyinyournose.
Oxygenoutputconcentrationhasbeenbelow50%for10minutes.Ifconditionpersists,switchtoyourbackupoxygensourceandcontactyourequipmentprovidertoarrangeforservice.
Abreathhasbeenrecognized,butproperoxygendeliveryhasnotbeendetected.
Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Ifpossible,moveconcentratortoacoolerlocationorpowerunitwithanexternalpowersupplyandremovebattery.Ifconditionpersists,contactyourequipmentprovider.
Message Display & Text
NoBreathDetect CheckCannula
OxygenError
O2DeliveryError
BatteryHOT Warning
Medium Priority AlertsThefollowingmediumpriorityalertmessagesareaccompaniedbyatriple beep,repeatedevery25seconds,andaflashing yellow light.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
27
Condition/Action/Explanation
Concentratorhasinsufficientbatterypowertoproduceoxygen.Attachexternalpowersupplyorexchangebattery,thenrestartunitifnecessarybypressingOn/Offbutton.
Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Concentratorhasstoppedproducingoxygen.Ifpossible,moveconcentratortoacoolerlocation,thenturnpoweroffandbackon.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Concentratortemperatureistoohighandoxygenproductionisshuttingdown.Ensureairintakeandoutletventshaveclearaccessandparticlefiltersareclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Message Display & Text
BatteryEmpty AttachPlug
BatteryHOT
SystemHOT
High Priority Alerts
CAUTION IfyouarenotneartheInogenOne®G4youmaynotbeabletohearorseethehighpriorityalerts.MakesuretheInogenOne®G4isinalocationwherethealertsandalertswillberecognizediftheyoccur.
Thefollowinghighpriorityalertmessagesareaccompaniedbyafive beep pattern,repeatedevery10secondsandaflashing yellow light.
AudibleandVisibleSignalsChapter4
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
28
Condition/Action/Explanation
Thismayresultfromtheconcentratorbeingstoredinacoldenvironment(below0ºC(32ºF)).Movetoawarmerenvironmenttoallowtheunittowarmupbeforestartingit.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Concentratorhasstoppedproducingoxygenandisshuttingdown.Youshould:
1. Switchtobackupoxygensource2.Contactyourequipmentprovider
Message Display & Text
SystemCOLD
SystemError
High Priority Alerts (continued)D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
29
Troubleshooting
TroubleshootingChapter5
Useexternalpowersupplyorreplacebatterywithonethatisfullycharged
Checkpowersupplyconnectionandverifygreenlightissolid
CheckDCCableconnectionattheConcentratorandatcigarettelighterorauxiliaryDCpowersource
Contactyourequipmentprovider
Batteryisdischargedornobatteryispresent
ACPowersupplyisnotconnectedproperly
DCCableisnotconnectedproperly
Malfunction
ConcentratordoesnotpoweronwhenOn/Offbuttonispressed
Recommended Solution
RefertoChapter4
Possible Cause
RefertoChapter4
Problem
Anyproblemaccompaniedbyinformationonconcentratordisplay,indicatorlightsand/oraudiblesignals
Solutionstosomepossibleissuesyoumayencounteraredescribedinthissection.
Inogen One® G4 Oxygen Concentrator
5English
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
30
PressOn/Offbuttontopowerconcentrator
Checkcannulaanditsconnectiontoconcentratornozzle
Concentratorisnotpoweredon
Cannulaisnotconnectedproperlyoriskinkedorobstructed
Nooxygen
Recommended SolutionPossible CauseProblem
Troubleshooting (continued)D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
31
Cannula Replacement
Yournasalcannulashouldbereplacedonaregularbasis.Consultwithyourphysicianand/orequipmentproviderand/orcannulamanufacturer’sinstructionsforreplacementinformation.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.
Case CleaningYoumaycleantheoutsidecaseusingaclothdampenedwithamildliquiddetergent(suchasDawnTM)andwater.
WARNING DonotsubmersetheInogenOne®G4oritsaccessoriesinwaterorallowwatertoenterintothecase;thismayleadtoelectricalshockand/ordamage.
WARNING DonotusecleaningagentsotherthanthosespecifiedinthisUserManual.Donotusealcohol,isopropylalcohol,ethylenechlorideorpetroleumbasedcleanersonthecasesorontheparticlefilter.
Filter Cleaning and ReplacementTheparticlefiltermustbecleanedweeklytoensuretheeaseofairflow.Removefiltersfromthefrontofthedevice.Cleantheparticlefilterswithamildliquiddetergent(suchasDawnTM)andwater;rinseinwateranddrybeforereuse.
Cleaning, Care and Maintenance
Cleaning,CareandMaintenanceChapter6
6
! C
! C
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
32
NOTE Itmaybenecessarytocleantheparticlefiltermoreoftenindustyenvironments.
TopurchaseadditionalparticlefilterscontactyourequipmentproviderorInogen.
Output FilterTheoutputfilterisintendedtoprotecttheuserfrominhalationofsmallparticlesintheproductgasflow.TheInogenOne®G4includesanoutputfilterconvenientlylocatedbehindtheremovablecannulanozzlefitting.Inogenrequiresthatthisfilterbereplacedbetweenpatients.
TheoutputfiltermaybereplacedbytheequipmentproviderorbytheownerusingtheOutputFilterReplacementKit(RP-404).
TheInogenOneG4Concentratormustbecleanedanddisinfectedaspertheaboveinstructionsforeachnewpatient.Nospecialmaintenanceneedstobecarriedoutbythepatient.YourequipmentproviderperformsmaintenanceoperationstoassurecontinuedreliableservicefromyourInogenOneG4.Themanufacturer’sinstructionsforthepreventativemaintenanceofthedevicesaredefinedintheservicemanual.Allworkmustbeperformedbytrainedtechnicianscertifiedbythemanufacturer.
DC Input Cable Fuse ReplacementTheCigaretteLighterDCpowerplugcontainsafuse.IftheDCinputcableisbeingusedwithaknowngoodpowersourceandtheunitisnotreceivingpower,thefusemayneedtobereplaced.
Toreplacethefuse,followtheseinstructionsandrefertothephotographbelow.1. Removethetipbyunscrewingtheretainer.Useatoolifnecessary.2. Removetheretainer,tipandfuse.3. ThespringshouldremaininsidetheCigaretteLighterAdapterhousing.Ifthespring
isremoved,makesuretoreplacethespringfirstbeforeinsertingthereplacementfuse.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
33Cleaning,CareandMaintenanceChapter6
English
4. Installareplacementfuse,InogenRP#125(BUSSMDA-12)andreassemblethetip.Ensuretheretainerringisproperlyseatedandtightened.
Inogen One® G4 Column Change Procedure
NOTE Columnchangeinstructionsareonlytobeusedwhenmaintenanceis requiredandarenotintendedforpracticepurpose.
1. TurnofftheInogenOne®G4concentratorbypressingthepowerbuttontoshutdownthedevice.
2.RemovetheInogenOne®G4concentratorfromthecarrycase,ifapplicable.
3.RemovethebatteryfromtheInogenOne®G4concentrator.
4.PlacetheInogenOne®G4concentratoronitssidesothattheundersideisvisible.Themetalcolumnas-semblycanbeseenononesideofthedevice.
FuseCigarette Adapter Plug Retainer Tip Retainer
CAUTION Forcontinuedprotectionagainstriskoffire,onlyusefusespecified.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
34
Inogen One® G4 Column Change Procedure (continued) 5.Unlockthecolumnassemblybypushingthelatch
buttonawayfromthecolumns.
6.Whileholdingthebuttonopen,slidethecolumnassemblyoutofthedevicebypullingonthecolumnpullhandle.
7. RemovethecolumnscompletelyfromtheInogenOne®G4.Bothcolumnsareremovedasonepiece.
Column(metaltube)Installation8.Removedustcapsofnewcolumnassembly.Make
surethereisnodustordebriswherethedustcapswerelocated.
Openandunlocked
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
35Cleaning,CareandMaintenanceChapter6
English
Inogen One® G4 Column Change Procedure (continued)
9.InsertcolumnassemblyintotheInogenOne®G4concentrator.Donotleavethecolumnendsexposed;columnassemblyshouldbeinsertedintotheIno-genOne®G4assoonasthedustcapshavebeenremoved.
10.Pushthecolumnassemblyintothedevicesuchthat thecolumnsarefullyseatedintotheInogenOne®
G4concentrator.Thespringloadedlatchbuttonshouldfullyreturntotheclosedposition.
11. ConnecttheACpowersupplycordtotheInogen One®G4andplugthepowersupplyACcordintoan electricaloutlet.DoNotPowerontheInogenOne® G4concentrator.
12.Pressandholdtheplus(+)and(-)minusbuttonfor5
seconds.Thescreenwilldisplay“sievereset” message.Releasebuttononcemessageisdisplayed
onscreen.
13.Pressthealert buttononceandscreenwill display“sieveresetsuccess”.
14.PressthepowerbuttontoturnontheInogenOne®G4,andusenormally.
WARNING Donotuseanycolumnsotherthanthosespecifiedinthisusermanual.Theuseofnon-specifiedcolumnsmaycreateasafetyhazardand/orimpairequipmentperformanceandwillvoidyourwarranty.! C
Closedandlocked
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
36
Disposal of Equipment and AccessoriesFollowyourlocalgoverningordinancesfordisposalandrecyclingoftheInogenOne®G4andaccessories.IfWEEEregulationsapply,donotdisposeofinunsortedmunicipalwaste.WithinEurope,contacttheEUAuthorizedRepresentativefordisposalinstructions.Thebatterycontainslithiumioncellsandshouldberecycled.Thebatterymustnotbeincinerated.
Maintenance Items List• InogenOne®G4singlebattery(model#BA-400)• InogenOne®G4doublebattery(model#BA-408)• Replacementintakeparticlefilters(model#RP-405)• OutputFilterReplacementKit(model#RP-404)• InogenOne®G4columns(model#RP-406)
Forassistance,ifneeded,insettingup,using,maintaining,ortoreportunexpectedoperationorevents,contactyourequipmentprovider,ormanufacturer.
Other Service and Maintenance WARNING DonotdisassembletheInogenOne®G4oranyoftheaccessoriesor
attemptanymaintenanceotherthantasksdescribedinthisusermanual;disassemblycreatesahazardofelectricalshockandwillvoidyourwarranty.Donotremovethetamperevidentlabel.Foreventsotherthanthosedescribedinthismanual,contactyourequipmentproviderforservicingbyauthorizedpersonnel.
CAUTION DonotuselubricantsontheInogenOne®G4oritsaccessories.
! C
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
37
Symbols Used On Concentrator and Accessories
SymbolsUsedonConcentratorandAccessoriesChapter7
7English
Meaning
Awarningindicatesthatthepersonalsafetyofthepatientmaybeinvolved.Disregardingawarningcouldresultinsignificantinjury.
Acautionindicatesthataprecautionorserviceproceduremustbefollowed.Disregardingacautioncouldleadtoaminorinjuryordamagetoequipment.
SeeUserManualforInstructions.
U.S.FederalRegulationRestrictsthisDevicetoSalebyOrderofPhysician.MayalsobeapplicableinotherCountries.
ACPower
DCPower
NoSmokingwhiledeviceisinuse.
NoOpenFlames(Concentrator);Donotincinerate(Battery).Refertoinstructionmanual/booklet.
Manufacturer
Symbol
WARNING
CAUTION
ONLY
ONLY
ONLY
ONLY
ONLY
! C
EC REP AuthorizedRepresentativeintheEuropeanCommunity
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
38
Meaning
KeepDry
IndoororDryLocationUseOnly,DoNotGetWet
UseNoOilorGrease
DoNotDisassemble(contactyourequipmentproviderforservicingbyauthorizedpersonnel)
DoNotDisposeofInUnsortedMunicipalWaste
TypeBFAppliedPart,NotIntendedforCardiacApplication
ClassIIDevice
ElectricalSafetyAgencyCertificationLogo
CompliesWithApplicableEUDirectivesIncludingMedicalDeviceDirective
Symbol
ONLYONLY
ONLY
ONLYONLY
User Interface Label
Meaning
ON/OFFButton
IncreaseFlowSetting
DecreaseFlowSetting
AudibleAlertButton
Symbol
0123
93-01213-00-00 rev1
0123
COLOR SUMMARY: 1. Black
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
39InogenOne®G4SystemSpecificationsChapter8
English
Inogen One® G4 ConcentratorDimensions:Withsinglebattery
L/W/H:5.91in.(15.01cm.)/2.68in.(6.8cm.)/6.5in.(16.3cm.)L/W/H:5.91in.(15.01cm.)/2.68in.(6.8cm.)/7.2in.(18.2cm.)
Weight: 2.8pounds(1.27Kg.)(includessinglebattery)Noise: 40dBA(onsetting2)Warm-UpTime: 2minutesOxygenConcentration: 90%-3%/+6%atallsettingsFlowControlSettings: 3settings:1to3Power:ACPowerSupply:
DCPowerCable:RechargeableBattery:
ACInput: 100to240VAC50to60Hz
Auto-Sensing: 1.0A
DCInput: 13.5-15.5VDC,10AMax.Voltage: 12.0to16.8VDC
BatteryDuration: Upto2.7hourswithsinglebatteryUpto5hourswithdoublebattery
BatteryChargingTime: Upto3hoursforasinglebatteryUpto5hoursforadoublebattery
EnvironmentalRangesIntendedforUse:
Temperature:41to104˚F(5to40˚C)Humidity:0%to95%,non-condensingAltitude:0to10,000ft(0to3048meters)
EnvironmentalRangesIntendedforShippingandStorage:
Temperature:-13to158˚F(-25to70˚C)Humidity:0%to95%,non-condensingStoreinadryenvironmentAltitude:0to10,000ft(0to3048meters)
Transportation: KeepDry,HandleWithCare
Inogen One® G4 System Specifications8
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
40
Inogen One® G4 Concentrator (continued)
Classifications
ModeofOperation: ContinuousDutyTypeofProtectionAgainstElectricalShock:
ClassII
DegreeofProtectiontoConcentratorComponentsAgainstElectricalShock:
TypeBFNotintendedforcardiacapplication
DegreeofProtectiontoConcentratorComponentsAgainstIngressofWater:
IP22-Verticallydrippingwatershallhavenoharmfuleffect&protectagainstingressofsolidobjects>12.5mm.diameterwhentheenclosureistiltedatanangleupto15˚fromitsnormalposition*
DegreeofSafetyforApplicationinPresenceofAnestheticGases:
Notsuitableforsuchapplication
TestedbyIndependentLaboratory:
Safety:IEC60601-1CAN/CSAC22.2No.60601-1ElectromagneticCompatibility:IEC60601-1-2RTCADO160
*NormalpositionoftheInogenOne®G4isuprightwiththeuserinterfacedisplayfacingupward.
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
41
PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartofthedevice,includingcables,thantherecommendedseparationdistancecalculatedfromtheequationapplicabletothefrequencyofthetransmitter.
Recommendedseparationdistance:d=1.2√P150kHzto80MHzd=1.2√P80MHzto800MHzd=2.3√P800MHzto2.5GHz
WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmeters(m).
FieldstrengthsfromfixedRFtransmitters,asdeterminedbyanelectromagneticsitesurveya,shouldbelessthanthecompliancelevelineachfrequencyrangeb.
Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:
3Vrms150kHz to80MHz
3V/m80MHz to2.5GHz
3Vrms
3V/m
ConductedRFIEC61000-4-6
RadiatedRFIEC61000-4-3
Compliance Level
IEC 60601 Test Level
Immunity Test Electromagnetic Environment - Guidance
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: TheConcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.TheuseroftheConcentratorshouldmakesureitisusedinsuchanenvironment.
Floorsshouldbewood,concreteorceramictile.Iffloorsarecoveredwithsyntheticmaterial,therelativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
Electrostaticdischarge(ESD)
IEC61000-4-2
Electricalfasttransient/burst
EC61000-4-4
Surge
IEC61000-4-5
±6kVcontact
±8kVair
±6kVcontact
±8kVair
±2kVforpowersupplylines
±1kVforinput/outputlines
±2kVforpowersupplylines
±1kVforinput/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
±1kVline(s)toline(s)
±2kVline(s)toearth
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.Iftheuserofthe[MEEQUIPMENTorMESYSTEM]requirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthatthe[MEEQUIPMENTorMESYSTEM]bepoweredfromanuninterruptedpowersupplyorabattery.
Powerfrequencymagneticfieldsshouldbeatlevelscharacteristicofatypicallocationinatypicalhospitalorhomeenvironment.
Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlines
IEC61000-4-11
Powerfrequency(50/60Hz)magneticfield
IEC61000-4-8
<5%UT(>95%dipinUT)for0.5cycle
40%UT(60%dipinUT)for5cycles
70%UT(30%dipinUT)for25cycles
<5%UT(>95%dipinUT)for5sec
<5%UT(>95%dipinUT)for0.5cycle
40%UT(60%dipinUT)for5cycles
70%UT(30%dipinUT)for25cycles
<5%UT(>95%dipinUT)for5sec
3A/m 3A/m
EnglishD.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019
42
NOTE At80MHzand800MHz,thehigherfrequencyrangeapplies.
NOTE Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.
NOTE UTisthea.c.mainvoltagepriortoapplicationofthetestlevel.
a:Fieldstrengthfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetofixedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.IfthemeasuredfieldstrengthinthelocationinwhichtheconcentratorisusedexceedstheapplicableRFcompliancelevelabove,theconcentratorshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasre-orientingorrelocatingthedevice.
b:Overthefrequencyrange150kHzto80MHz,thefieldstrengthsshouldbelessthan3V/m.Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device:
ThisconcentratorisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.TheuseroftheconcentratorcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommuni-cationsequipment(transmitters)andthisconcentratorasrecommendedbelow,accordingtothemaximumoutputpowerofthecommuni-cationsequipment.
150 kHz to 80 MHzd=1.2√P
0.120.381.23.812
0.120.381.23.812
0.230.732.37.323
0.010.11
10100
80 MHz to 800 MHzd=1.2√P
800 MHz to 2.5 GHzd=2.3√P
Separation Distance According to Frequency of Transmitter (M)Rated Maximum Power Output of Transmitter (W)
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.
Guidance and Manufacturer’s Declaration – Electromagnetic EmissionsTheconcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuseroftheconcentratorshouldassurethatitisusedinsuchanenvironment.
NOTE At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE Theguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.
Emissions Test Compliance Electromagnetic Environment - Guidance
RFemissionsCISPR11
Group1 TheconcentratorusesRFenergyonlyforitsinternalfunction.ThereforeitsRFemissionsareverylowandnotlikelytocauseanyinterferenceinnearbyequipment.
RFemissionsCISPR11
ClassB Theconcentratorissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.HarmonicEmissions
IEC61000-3-2ClassA
Voltagefluctuations/flickeremissionsIEC61000-3-3
Complies
D.SPEC.25.4 Ingoen G4 - Mode d'emploi_EN
v. 28/06/2019