Upload
doandang
View
227
Download
4
Embed Size (px)
Citation preview
Infant Flow® LP system headgear
Prepare Infant Flow® LP SiPAP driver ...............1Infant Flow LP components ............................2Generator assembly components ...................3Low pressure generator .................................4Nasal prongs ..................................................5Nasal masks ...................................................6Nasal prongs and mask sizing ........................7Headgear sizing .............................................8Generator assembly application .....................9 Headgear application ............................9-11 Check for proper headgear placement .....12 Incorrect headgear placement .............13-14 Attach interface to generator ...................15 Remove the support cradle ......................16 Apply support cradle to headgear ............17 Align generator assembly .........................18
Table of contents
i
Secure the locking strap ...........................19 Attach side straps to generator ................20Final check of generator assembly ................21 Check for correct headgear placement .....21 Check placement of exhalation tube ........22 Check for proper nasal prong position .....23 Incorrect nasal prong position ..................24 Verify proper mask placement ..................25 Incorrect mask placement ........................26 Incorrect mask size—too small .................27 Incorrect mask size—too large .................28Warnings and cautions ................................29Application tips ............................................30Inspect the system .......................................31Notes ......................................................32-33
1
1. Attach the infant circuit to humidifier
and nCPAP driver.
2. Attach generator to the circuit.
3. Perform oxygen sensor calibration
and user verification.
4. Enter desired settings.
Prepare Infant Flow LP SiPAP driver
100
90
80
706050
40
30
21
2
1
2
3
4
5
L/minL/min
NCPAP/Pres Low
Pres High
Infant FlowSiPAP
4
6
8
10
12
14
XDCR PPROX
3 4
1
2
2
A. Prongs and masks
B. Generator
C. Bonnet sizing guide
D. Mask/prong sizing guide
E. Headgear
F. Bonnet
Infant Flow LP components
A
B
E
FD
C
3
Generator assembly components
A. Exhalation tube
B. Securing strap
C. Generator receiver
D. Fixation tab
E. Mask/prong
attachment base
F. Support cradle
G. Pressure and drive lines
C
B
F
E
A
G
D
4
Low pressure generator
• Lowerdrivepressure
– Prevents pressure backup into the
water auto-feed system
• Pressurereliefvalve
Pressure relief valve
5
Nasal prongs
Anatomically designed
A. Flared tip
B. Flexible bellows
C. Base
D. Size indicator
E. Key
A
A
BC
E
B
C
D
D
6
Nasal mask
Anatomically designed
A. Nasal bridge cushion
B. Contour shape
C. Key
D. Flexible bellow
E. Assessment window
F. Size indicator
A
B
C
E
E
D
F
D
• Usesizingguidetodetermineappropriate
mask or prong size
• Nasalprongs:Choosedotsthatfillthe
nare space
• Ifbetweensizes,choosethelargersize
• Nasalmasks:Choosethetrianglethatfits
over the nose
7
Nasal prongs and mask sizing
Headgear sizing
Measure the circumference of the infant’s head; from crown at back of head to brow line
(in centimeters).
8
9
Headgear application
1. Align headgear with the midline of face.
2. Align bottom of the headgear with the
nape of neck.
1.
2
Generator assembly application
Headgear application (continued)
10
3. Place #1 strap just above brow line.
Note:Strapsarelabeled1,2,3.
4. Place #2 strap midline over strap #1.
3.
4.
11
Headgear application (continued)
5. Place #3 strap on top of #1 and
#2 straps.
Note: Strap should be snug—do
not overtighten.
6. Fold #2 strap up and over #3
strap; secure to top strap.
5.
6.
12
Check for proper headgear placement
• Backofheadgearrestsonnapeofneck
• Foreheadstrapsplacedonbrowline
13
Incorrect headgear placement
• Horizontalstraprestsclosetoeye
• Foreheadstrapistoofarabovethe
brow line
14
Incorrect headgear placement (continued)
• Abovethenapeoftheneck,createsa
fold/bulging of headgear
• Topstraptootight,foreheadstrap#3
pulled back
Attach interface to generator
15
Align the notch on the prongs/mask to straddle the pressure lines.
1.
3. 4 .
2.
Remove the support cradle
16
1. Expand the collapsible corrugated tubing.
2. Remove the support cradle from the
generator assembly.
1.
2.
Apply support cradle to headgear
17
1. Apply the support cradle properly
centered on the straps.
2. Secure center tab.
1.
2.
Align generator assembly
1. Center the generator and interface over
patient nose and support cradle.
2.Useagentlerockingmotiontoinsertthe
nasal prongs.
3. Place the drive line and pressure line into
the support cradle with the exhalation tube
resting on top.
18
1. 2. 3.
19
Secure the locking strap
1. Wrap the locking strap over the
exhalation tube and attach to the side
of the support cradle.
2. Position the small slit over the ridge of
the exhalation tube.
1.
2.
20
Attach side straps to generator
1. Wrap the side straps around the
generator tabs.
2. UsetheVelcrotiptosecurethe
strap—loose is better.
3.Usethesidestrapstoadjusttension
on interface—do not over-tighten.
1. 2.
3.
Check for correct headgear placement
21
A. Generator aligned and
centered with nose
B. Side straps properly placed
C. Support cradle tabs secured
to headgear
Note:Strapsshouldbesnug,
but not tight. B
A
C
Final check of generator assembly
Incorrect Correct
22
Check placement of exhalation tube
23
Check for proper nasal prong position
• Bellowsarevisible
• Septumisvisible
• Generatorassemblyhas
no contact with skin
24
Tootight,bellowscompressed
Prong not in the nare
Incorrect nasal prong position
25
Verify proper mask placement
• Maskcoversnasalprofile
• Maskrestsbelowtheeyes
• Bellowsnotcompressed
• Sidestraplowoncheek
Note: Do not over-tighten.
26
Incorrect mask placement
• Toomuchtensionappliedtooneside
• Maskiscollapsed
• Naresarenotvisible
27
Incorrect mask size—too small
• Maskblocksthenares
• Maskunabletocovernose
28
Incorrect mask size—too large
• Highonthebridgeofthenose
• Upagainsttheeyes
• Restontheupperlip
29
Warnings and cautions
WARNING:
• Usethisproductonlyasdirectedintheproductliteraturetoreducetheriskofnasalirritation,
septaldistortion,skinirritationandpressurenecrosis
• Tobeusedbyatrainedpractitioner,underthedirectsupervisionofaqualifiedphysician
• OnlyusetheInfantFlowLPgeneratorwithvariableflownCPAPdrivers
• Donotover-tightenthefixationstraps
30
Application tips
• Selecttheappropriatesizenasalmasktominimizeleaksanddeadspace
• Selecttheappropriatesizenasalprongs;ifbetweensizes,selectthelargersize
• Applicationofanincorrectlysizedprong,mask,bonnetorheadgearwillaffectstabilityof
the generator
• Adjustthestrapstostabilizethegeneratorandmaintainasealatthenoseusingtheleast
tension possible
• Consideralternatingtheuseofprongandmaskinterfacesatsetintervalstochangepressure
points on the infant’s face
• Continuouslymonitorpatient’srespiratorystatus
• Coverbothearsevenly;ensuretheearsarenotfolded
• RecommendhumidificationbeusedwithnCPAPsystems
31
Inspect the system at least every three to four hours to:
• Ensurethepatient’seyesareclearlyvisibleandthenaresarenotblocked,whenusingmasks
• Inspectthefixationdeviceandstrapsforpropertensionandadjustasneededtomaintain
a proper fit
• Monitorthepatientforgastricinsufflationandabdominaldistension
• Monitorforexcessivecondensationincircuitandgenerator
• Alternatetheuseofprongandmaskinterfacestochangepressurepointsonthe
infant’s face
32
Notes
33
Notes
© 2011 CareFusion Corporation or one of its subsidiaries. All rights reserved. Infant Flow is a trademark or registered trademark of CareFusion Corporation or one of its subsidiaries. RC2661-01 (1211/5000) L3414 Rev. B
carefusion.com
CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Hoechberg Germany
+49 931 4972-0 tel +49 931 4972-423 fax
CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887
800.231.2466 toll-free 714.283.2228 tel 714.283.8493 fax
WARNING—U.S. Federal Law restricts this device to sale by or on the order of a physician.