IMPROVEMENT OF THE QUALITY IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES IN ASSURANCE IN TRITIUM ANALYSES IN WATERS BY LSC WATERS BY LSC Nuclear and Technological Institute (ITN) Radiological Protection and Safety Unit (UPSR) E.N. 10, 2686-953 Sacavém, PORTUGAL Ana Rita Gomes , João Abrantes, Albertina Libânio, Maria José Madruga, Mário Reis 6th of September, 2010, O42

IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

Embed Size (px)

Citation preview

Page 1: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

IMPROVEMENT OF THE QUALITY IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES IN ASSURANCE IN TRITIUM ANALYSES IN

WATERS BY LSCWATERS BY LSC

Nuclear and Technological Institute (ITN)Radiological Protection and Safety Unit (UPSR)

E.N. 10, 2686-953 Sacavém, PORTUGAL

Ana Rita Gomes, João Abrantes, Albertina Libânio, Maria José Madruga, Mário Reis

6th of September, 2010, O42

Page 2: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

INTRODUCTION

Contamination by Fallout

present in the environment as a result of both natural and anthropogenic sources

Tritium (3H)

Nuclear tests (1945-1963)

produced in the atmosphere through nuclear reactions between fast neutrons resulting from cosmic radiation and nitrogen atoms

Nowadays Levels near the minimumdetectable concentrations

Page 3: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

Radiological Protection and Safety Unit (UPSR)

Environmental Radioactivity Group

Requested by water suppliersto carry out tritium analyses

in drinking waters

INTRODUCTION

Nuclear and Technological Institute (ITN)

Page 4: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

INTRODUCTION

Nuclear and Technological Institute (ITN)Improvement of the method

Until the middle of 2009Portuguese StandardGuideline NP 4362/1997Internal standard for each sample

Samples controlled by volume

DL = 8.0 Bq L-1, 150min

Internal Method ISO standard CD 9698/2008

Control of the quenching interference through a calibration curve

Sample volume started to be controlled by weighing

DL = 4.0 Bq L-1, 120min

Page 5: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

INTRODUCTION

Nuclear and Technological Institute (ITN)Improvement of the methodQuality Assurance Program

Validation of analytical method;

Resources used for the laboratory analysis procedures for sample handling and analysis;

Quality control;

Monitoring and auditing.

Page 6: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

INTRODUCTION

Nuclear and Technological Institute (ITN)Improvement of the methodQuality Assurance Program

Studies ensuring that the method is fit-for-purpose:

Analytical thresholds;

Accuracy (trueness and precision);

Curve fitting and its range;

Budget uncertainties.

Validation of analytical methodology:

Page 7: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

INTRODUCTION

Nuclear and Technological Institute (ITN)Improvement of the methodQuality Assurance Program

Validation of analytical methodology

Submitted for accreditation to the Portugueseaccreditation body (IPAC) in October of 2009.

Page 8: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

EXPERIMENTAL PROCEDURE

The radioactive sample is combined with a liquid scintillation cocktail and the radionuclide decay produces an ionizing particle;

Part of the kinetic energy of this ionizing particle is transferred to the scintillator and converted into light during the radioactive decay process;

Blank Samples

Purification by distillation;

Addition of 0.5g of Na2S2O3 and 1g of Na2CO3 to a volume of about 250 mL;

An aliquot of the distillate (about 8 g) is mixed with 12g of UltimaGold LLT scintillation cocktail in borosilicate glass vials.

3H Determination

Page 9: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

EXPERIMENTAL PROCEDURE

3H Determination Samples are stabilized in the dark;

Tri-Carb 3170 TR/SL (Packard), in low level mode;

DL ≅ 4.0 Bq L-1, 120min (routine analysis).

Page 10: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

Activity Calculation

VncA .ε

= OHmV

2ρ=

Ac Activity, Bq L-1

n Net count rate

ε Efficiency

V Sample volume, L

m Sample mass, kg ρ H2O Water density, kg L-1

Expanded uncertainty

)()()()( 222 Vuunuccu rrrAA ++= ε

Characteristic limits

k Expansion Factor

t Counting time, s

b Background count rate

)(..

2* butb

VkcA +=ε Decision threshold

)..()..(1..

.22222

2*

#

VuuVkVt

kcc

rr

A

A +−

+=

εεε

Detection limit

EXPRESSIONS

Page 11: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Equipment Calibration Method Validation Quality Control

Page 12: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Equipment Calibration 10 standard samples ≅ 175 Bq each

CCl4 (µL): 2,4,6,8,10,12,14,16,18,20

Calibration curve

tSIE....... 31 1013043310430181 −− ±+±=ε

Page 13: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Equipment Calibration Method Validation Detection Limit

24h counting time

Detection Limit = 1.6 Bq L-1

Routine measurements

Page 14: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Equipment Calibration Method Validation

Accuracy (Trueness and Precision)Trueness

Two intercomparison exercises

2 < |Z| ≤ 3 Acceptable

Precision

Precision acceptance criterion of 10.0%

10 sub-samples of a control sample: ≅ 50 Bq L-1

2.4% Acceptable

Detection Limit

Page 15: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Method Validation

Budget Uncertainties

Net count rate

Equipment Calibration

Accuracy (Trueness and Precision)

Detection Limit

Page 16: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

(Least squares method)

Efficiency

Budget Uncertainties

Equipment Calibration

Accuracy (Trueness and Precision)

Detection LimitMethod Validation

Page 17: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Equivalent volume

Budget Uncertainties

Equipment Calibration

Accuracy (Trueness and Precision)

Detection LimitMethod Validation

Page 18: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Budget Uncertainties

Equipment Calibration

Accuracy (Trueness and Precision)

Detection LimitMethod Validation

Page 19: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Equipment Calibration Method Validation Quality ControlInternal QC

Implementation of ISO/IEC 17025;

Storage improvements (tightness of the sample containers and temperature);

Samples and standards prepared by weighing;

Equipment performance tests.

Page 20: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Quality ControlInternal QC (Control Charts)

Warning Limits:

Action Limits:

σ2±X

σ3±X

Trueness of the method

Equipment Calibration Method Validation

Page 21: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Quality ControlInternal QC (Control Charts)

Warning Limits:

Action Limits:

σ2±X

σ3±X

Equipment Calibration Method Validation

Blank Samples

Page 22: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

STEPS FOR QUALITY IMPROVEMENT

Revalidate the method and to check the performance of the technicians

Quality ControlInternal QCExternal QC

Proficiency tests

Equipment Calibration Method Validation

|Z| ≤ 2 satisfactory

Page 23: IMPROVEMENT OF THE QUALITY ASSURANCE IN TRITIUM ANALYSES ... · improvement of the quality assurance in tritium analyses in ... 9dl = 4.0 bq l-1, 120min. ... o-42.ppt author: cd166119

CONCLUSION

The change introduced in the method allowed to improvethe results and consequently analytical data with better

quality are available to the customers.

The requirements for a quality system according to EN ISO/IEC 17025 must be fulfilled in order to get the

accreditation of the analytical techniques.

To keep the analytical process up to date, it is essential to participate in intercomparison

exercises and to maintain internal and external audits.