How to Make an Inexpensive, Easy-to-Use Compound Microscope,
Using Mainly PVC Parts
Chris Stewart and John Giannini*
St. Olaf College, Biology Department, 1520 St. Olaf Avenue, Northfield, MN 55057
*Email: [email protected]
http://pages.stolaf.edu/opn-lab/equipment/ 2 January 9, 2017
Introduction
By enabling students to view cells, organelles, and other specimens that would otherwise be
invisible to the human eye, a compound light microscope is a crucial tool for any biology
teaching lab. Unfortunately, the high cost of these instruments (even student models) can
often impose a significant economic burden on many schools – especially if buying a set of 10 to
20 microscopes to cover all of the lab groups in a given course.
Thankfully, a number of educational articles and websites have described various ways to build
inexpensive microscopes for student use as part of a teaching lab or other instructional activity
(Caplan, 2009; Carboni, 2011; Cybulski, Clements, and Prakash, 2014; Drace, Couch, and
Keeling, 2012; Dubowsky, 1996; Mahaffey, 1999; Tsagliotis, 2012; Whibley, 1981). In general,
these sources discuss how to make basic microscopes from supplies that would be available at
most schools or which could be purchased at different types of arts and crafts, hobby, or
hardware stores.
In this same spirit, we explain here how to build a working compound light microscope using
PVC parts, a push-button LED light, inexpensive microscope optics (i.e., an eyepiece and
objective lens), and a few other items that should be available at most schools or which can be
easily purchased at a grocery or hardware store, (e.g., rubber bands, black construction paper,
super glue, etc.).
This design is actually a precursor to a more advanced model that we described in a 2016 article
for The Journal of Chemical Education, which explained how to make a low-cost microscope for
viewing fluorescent cells or other samples (Stewart and Giannini, 2016). We called that
microscope “the OPN Scope” (short for Open Scope) because all of its plans and parts were
designed to be open and accessible to all. That way, others could use or modify the instrument
to fit their own educational or research needs.
Since many of the steps (and several of the parts) for building the microscope described here
are the same as (or similar to) those for making the OPN Scope, we encourage readers to
review the Supporting Information (S4) for our 2016 article, which should be available for free
on the website for The Journal of Chemical Education. The supplement itself describes in detail
how we built a PVC version of the OPN Scope, and we hope that readers will find our
description of the process that we followed there to be helpful.
Finally, to avoid any confusion between the two instruments, we call the microscope described
in this manual the bright-field version of the OPN Scope; and, like other OPN instruments that
we have developed (OPN Lab, 2017), we encourage readers to use or modify this design as
necessary to fit their particular needs.
http://pages.stolaf.edu/opn-lab/equipment/ 3 January 9, 2017
The Workings of a Basic Compound Light Microscope
For those unfamiliar with the instrument, a basic compound light microscope contains two
lenses (Fig. 1): one in the eyepiece (called the “ocular lens”) and the other in the objective
(called, not surprisingly, the “objective lens”). The total magnification of the microscope is the
product of the magnification of the ocular and objective lenses. Thus, for example, using a 5x
eyepiece with a 4x objective will generate a total magnification of 20x (since 5 x 4 = 20).
Figure 1. The lenses in a basic compound light microscope.
In addition, in order for the microscope to maintain a proper focus, there must be a specific
fixed distance between the ocular and objective lenses. This distance is called the “tube
length,” and it is typically measured from the “seat” of the eyepiece (Fig. 1, top) to the
“shoulder” of the objective (Fig. 2, arrow pointing left). Often, the objective will even have the
tube length (in mm) printed on it (Fig. 2, arrow pointing right).
http://pages.stolaf.edu/opn-lab/equipment/ 4 January 9, 2017
Figure 2. An AM Scope 4x objective, which shows the tube length of 160
mm (arrow pointing right) and “shoulder” or raised ring (arrow pointing
left).
Besides maintaining a fixed distance between the ocular and objective lenses, a microscope
typically has a way to make “coarse” and “fine” adjustments in order to bring a sample into
better focus, and readers can visit any number of websites for more information on the design
and use of microscopes, such as: the Bates College compound microscope web page, the
Celestron PDF manual on microscope basics, or the Nikon “Microscopy U” home page.
Materials
We list the raw materials needed to build this microscope in Table 1 below (along with their
estimated cost at January 2017 prices, not including any taxes or shipping costs). We further
include the total and per unit cost to make seven such microscopes, which is approximately $44
per instrument (again, not including any taxes or shipping costs). Of course, because many
schools may already have several of the supplies needed to build this microscope (e.g., large
plastic weigh boats, black construction paper, rubber bands, super glue, etc.), the cost may be
even lower.
http://pages.stolaf.edu/opn-lab/equipment/ 5 January 9, 2017
Table 1. The Parts Needed to Make a Bright-Field Version of the OPN Scope.
Qty Description Unit Price Total Price 1 ¾-inch thick PVC board (5½ inches wide and 8 feet long)
(One board will makes seven microscope bases and PVC disks) $18.00 $18.00
3 ¾-inch thick PVC trim (1½ inches wide and 8 feet long) (Three pieces will make seven sets of support arms and necks)
$5.00 $15.00
1 ¾-inch Schedule 40 (white) PVC pipe (5 feet long) (One pipe will make ten microscope tubes)
$2.00 $2.00
7 ¾-inch Schedule 40 (white) PVC coupling (male adapter) $0.50 $3.50 7 ¾-inch Schedule 80 (dark gray) SPG x SOC bushing $2.00 $14.00 7 3-inch diameter female PVC couplings $3.50 $24.50 7 3-inch diameter male PVC couplings $2.50 $17.50 7 Circular push-button LED lights (less than 3-inch diameters) $2.00 $14.00 4 Sets of two 5x or 10x eyepieces for a compound microscope $18.00 $72.00 7 4x objective lenses for a compound microscope $13.00 $91.00 7 Pieces of black construction paper (3 x 5½ inches) $2.00 $2.00
14 1¼-inch long No. 6 drywall screws (package of 50 screws) $2.00 $2.00 14 Large plastic weight boats (package of 100 available online) $16.00 $16.00 1 Package of large rubber bands $2.00 $2.00 1 Package of sandpaper (60-grit or similar) $1.50 $1.50 1 Bottle of super glue $3.50 $3.50 1 Roll of tape (e.g., masking, labeling, etc.) $1.00 $1.00 1 Roll of electrical tape $1.00 $1.00 1 Package of cork or felt pads $3.00 $3.00 Total Cost: $98.50 $303.50 Cost per Microscope (for 7 models): N/A $43.36
Most of the above items (particularly the PVC board, trim, pipe, and couplings) can be
purchased at any hardware or home improvement store. However, to make the specific PVC
parts for this microscope, readers will need to cut the PVC board, trim, pipe, and couplings to
size (Fig. 3), and we include the dimensions for these various pieces (as well as some basic
instructions for making them) in Table 2 below.
http://pages.stolaf.edu/opn-lab/equipment/ 6 January 9, 2017
Figure 3. The supplies needed to build the bright-field version of the OPN Scope
(including the rectangular base, pieces for the microscope arm and neck, the PVC tube
for the lenses, and the circular disk – all of which need to be cut to size).
While some lumber yards or home improve stores may cut some of these pieces for customers
(if the request is a reasonable one), readers may find that they need to cut other pieces (or,
possibly, all of the parts) themselves. If so, many schools often have a wood shop that should
have the appropriate tools. Otherwise, teachers can see if a friend or colleague has a shop at
home, where these pieces could be cut.
Either way, we have found that using a chop saw works much better than a table saw to cut the
pieces of PVC board and trim as well as the PVC pipe and male PVC coupling. However, please
make sure to bring the saw blade down slowly on these pieces (especially the PVC pipe and
coupling) to avoid nicking or cracking these parts. In addition, we have used a band saw to cut
the circular PVC disk and a drill press with a vice bolted to the platform to drill holes into the
PVC disk and female PVC coupling.
http://pages.stolaf.edu/opn-lab/equipment/ 7 January 9, 2017
Table 2. Dimensions of the PVC Parts for the Bright-Field Version of the OPN Scope.
PVC Piece Microscope Part Instructions
PVC Board Labeled: 4/4” x 6” x 8’ Actually: ¾” x 5½” x 8’
Rectangular Base (¾ x 5½ x 9 inches)
Cut a 9-inch long piece for the base.
Circular Disk (3½ inch diameter)
Cut a 3½-inch square; drill a ½-inch diameter hole in the center; cut the square into a 3½-inch diameter circle; and sand as necessary until it fits into the top of the male PVC coupling.
PVC Trim Labeled: 4/4” x 2” x 8’ Actually: ¾” x 1 ½” x 8’
Arm (¾ x 1½ x 8½ inches – two) (¾ x 1½ x 10 inches – one)
Cut two 8½-inch long pieces and one 10-inch long piece to make the support arm of the microscope.
Neck (¾ x 1½ x 4 inches – two) (¾ x 1½ x 2 inches – one)
Cut two 4-inch long pieces and one 2-inch long piece to make the neck of the microscope, which will hold the PVC tube for the lenses.
PVC Pipe (¾” x 5’)
Microscope Tube (¾ x 5½ inches)
Cut 5½-inch (~140-mm) long pieces to provide a 160-mm tube length (the additional distance will be provided by the PVC coupling that holds the eyepiece and objective, respectively).
3-inch PVC Coupling (Male)
Adjustable Stage – Top Cut ¾ inches off the top of this piece (or add ¾ inches to the length of the support arm and adjust the diameter of the PVC disk as needed).
3-inch PVC Coupling (Female)
Adjustable Stage – Bottom Drill a 1½-inch diameter hole into the bottom half of this piece, so that students can easily turn on and off the push-button LED light.
Also, as explained in our OPN Scope paper (Stewart and Giannini, 2016), working with power
tools can be extremely dangerous, and the significant physical hazards that they pose should
be obvious. Therefore, any readers who are unfamiliar with these types of tools or how to use
them should work with a trained and experienced craftsperson in order to avoid serious injury.
Readers should also make sure to wear the proper protective equipment when working with
these tools, including any necessary eye and ear protection.
http://pages.stolaf.edu/opn-lab/equipment/ 8 January 9, 2017
Readers should further know that, as with the OPN Scope, we recommend using PVC pieces to
make this microscope because, unlike many types of wood, PVC should not warp, split, rot, or
crack over time (especially in places that experience humid summers and dry winters). Also, we
have generally found that warping tends to be the most serious problem. Specifically, if there is
a bow in the rectangular base of the microscope (Cover Image, bottom), then students may find
that the stage “wobbles” a bit, which may cause a “ripple” effect in their images – especially if
using a digital microscope camera to take photographs of their samples.
Of course, PVC board and trim is often much more expensive than wood (especially wood
purchased from the “spare” or “scrap” wood bin at a lumber yard or home improvement store).
As a result, if using wood to help keep costs low, we suggest using plywood for the rectangular
base of the microscope because this type of composite wood may be less likely to warp than
other types of inexpensive wood (e.g., pine). Also, since the arm and neck of the microscope
are made by gluing various pieces of wood together, these pieces are less likely to warp
significantly over time (especially if initially placed in a vice or under a fixed weight to press the
pieces together as the glue dries).
Also, before buying any ¾-inch PVC couplings or bushings to hold the eyepiece and objectives,
we suggest that readers first test out these PVC parts by placing each one of them onto the ¾-
inch PVC pipe that they intend to purchase in order to ensure that the pieces will fit together
smoothly. Otherwise, readers may find that the pipe which they purchased is slightly larger or
smaller than their couplings and bushings. As a result, the ends of each length of PVC pipe
would need to be sanded or wrapped in a few layers of electrical tape (respectively) in order to
provide the appropriate fit.
With respect to the eyepiece and objective lens themselves (Fig. 3, top right), readers can
purchase new or used optics online from websites like Amazon, eBay, or AM Scope or,
alternatively, salvage lenses from any older or broken microscopes that may be at their schools.
For example, in our PVC microscopes, we have successfully tested out new 5x eyepieces (SKU
No. EP5X23H-V299) and 10x eyepieces (SKU No. EP10X23-V299) with 4x objective lenses (SKU
No. A4X-V300) – all manufactured by AM Scope. In addition, we have used 10x eyepieces and
various objectives (4x, 10x, and 40x), which we harvested from a used Carlsan microscope that
we bought on eBay for roughly $15. While both sets of optics required a 160-mm (~6.3-in) tube
length (Figs. 1 and 3), other sets of eyepieces and objectives may require a different fixed
distance. As a result, readers may need to adjust the length of their PVC tubes accordingly.
Also, for simplicity, when buying objective lenses, we recommend selecting ones that have a
raised ring (i.e., shoulder) right below the threads – like the AM Scope 4x objective shown in
Figure 2 above. That way, the objective can fit into the ¾-inch PVC bushing without falling
through the bottom hole.
http://pages.stolaf.edu/opn-lab/equipment/ 9 January 9, 2017
Constructing the Microscope
Once all of the PVC pieces have been cut to size, assembling the microscope is rather straight
forward, and we provide the following general instructions (with supporting figures) to assist
readers in completing these steps.
Step 1. First, lay out all the components for the microscope (Fig. 4) and make sure that all of
the parts listed in Table 1 above are present.
Figure 4. The supplies needed to build the bright-field version of the OPN Scope,
including the rectangular base, pieces for the microscope arm and neck, the PVC tube,
and the circular disk – all of which need to be cut to size.
http://pages.stolaf.edu/opn-lab/equipment/ 10 January 9, 2017
Step 2. Then, to make the support arm and neck for the microscope (Cover Image, left), align
the longer and shorter 1½ inch wide pieces of PVC trim as shown in Figure 5 below.
Figure 5. The pieces of 1½-inch wide PVC trim needed to make the support arm
(left) and neck (right) to the microscope.
Importantly, when arranging the three smaller pieces for the neck, place the middle piece off-
center to create a 1½-inch long slot on one end of the neck and a ½-inch long slot on the other
end (Fig. 5, lower right). The longer slot will be used to hold the neck to the arm of the
microscope (Cover Image, upper left), and the shorter slot will be used to hold the tube for the
eyepiece and objective in place (Cover Image, upper right).
http://pages.stolaf.edu/opn-lab/equipment/ 11 January 9, 2017
Step 3. Next, glue each set of pieces for the arm and neck together with super glue as shown in
Figure 6 below. Make sure to align the pieces correctly at the outset because the glue will dry
quickly once it has been applied.
Figure 6. Gluing the pieces of PVC trim together to make the support arm
(left) and neck (right) for the microscope.
Again, when gluing together the three smaller PVC pieces for the neck, make sure to place the
middle piece off-center, so that it leaves a 1½-inch long slot on one end of the neck and a ½-
inch long slot on the other end (Fig. 6, lower right). The longer slot is needed to hold the neck
to the arm of the microscope (Cover Image, upper left), and the shorter slot will hold the tube
for the eyepiece and objective in place (Cover Image, upper right).
http://pages.stolaf.edu/opn-lab/equipment/ 12 January 9, 2017
Step 4. Next, glue the arm and neck together, making sure that the 1½-inch long slot matches
up with the tall middle piece sticking out of the arm of the microscope (Fig. 7). Also, if need be,
use some rough (e.g., 60-grit) sandpaper to even out the bottom or base of the neck after the
pieces have been glued together, so that the support arm will stand flat on the PVC base of the
microscope. Alternatively, for readers who have access to a wood shop, a belt or disk sander
could be used for this step.
Figure 7. The completed support arm and neck of the microscope.
http://pages.stolaf.edu/opn-lab/equipment/ 13 January 9, 2017
Step 5. Now, glue the support arm to the base of the microscope (Fig. 8). To further secure the
arm to the PVC base, use one or two 1¼-inch long No. 6 drywall screws (Fig. 9). Also, we
suggest first drilling guide holes for these screws using a ⅛-inch bit, so that the drywall screws
will pass through the PVC board and trim more smoothly.
Figure 8. The arm of the microscope glued (and secured) to the base.
http://pages.stolaf.edu/opn-lab/equipment/ 14 January 9, 2017
Readers can also place felt or cork pads on the bottom of the base (in each of the four corners)
to help level the microscope and provide greater stability (Fig. 9).
Figure 9. Using two No. 6 drywall screws to secure the support arm of the microscope
to the base, and placing felt pads on the bottom of the base (in each of the four corners)
to help level the microscope and provide greater stability.
http://pages.stolaf.edu/opn-lab/equipment/ 15 January 9, 2017
Step 6. Next, lay out the pieces for lens tube (Fig. 10) and then put these pieces together. To
do so, first place the objective lens into the ¾-inch Schedule 80 (dark gray) PVC bushing (Fig. 10,
bottom left). As discussed above, the “shoulder” of the objective (i.e., the raised ring just
below the threads) should keep the objective from falling through the bottom hole in the
bushing. Next, insert the PVC tube into the bushing to hold the objective in place. The bottom
of the tube should fit firmly up against the shoulder of the objective (pinning it to the bottom of
the bushing).
Figure 10. The parts for the microscope tube, which holds the eyepiece
and objective (left) as well as an assembled tube (right).
http://pages.stolaf.edu/opn-lab/equipment/ 16 January 9, 2017
As explained above, if the PVC tube is too big (and does not slide far enough down into the
bushing), then wrap a piece of sand paper around the end of the tube to sand down the outside
wall until the tube will fit snugly up against the objective. Alternatively, if the PVC tube is too
small (and does not stay fixed in place when slid into the bushing), wrap a few layers of
electrical tape around the bottom of the tube until it holds the objective firmly in place.
Next, to improve the image seen with the microscope, roll a rectangular piece of black
construction paper (roughly 3 inches x 5½ inches) into a tube and insert it into the PVC pipe
(Fig. 10, center). This will help to block out any ambient light from the surrounding room.
Alternatively, readers can use a thicker piece of Schedule 80 (dark gray) PVC pipe – although
these parts tend to cost more than their Schedule 80 (white) PVC counterparts.
Now, place the ¾-inch Schedule 40 (white) PVC coupling onto the top of the tube and twist the
two pieces together to provide a tight fit. Then, wrap several layers of electrical tape around
the base of the eyepiece, so that it fits snugly into the coupling.
Also, please do not glue either the white PVC coupling or the gray PVC bushing to the PVC tube.
Instead, simply use the natural “compression fitting” of these various pieces to hold them
together, so that the parts can be easily slid on and off. That way, it will be easier to switch
eyepieces, objectives, or even tube lengths in the event that this becomes necessary.
http://pages.stolaf.edu/opn-lab/equipment/ 17 January 9, 2017
Step 7. After assembling the lens tube, cut out a 1-inch wide strip from one of the plastic
weight boats. This strip will help to secure the tube to the neck (Cover Image) while also
allowing the tube to slide up and down in order to allow for “coarse” adjustments in the focus.
Further note that one large plastic weigh boat will make several 1-inch wide strips.
Figure 11. Cutting out a 1-inch wide strip from a plastic weight boat to later hold the
lens tube in place.
http://pages.stolaf.edu/opn-lab/equipment/ 18 January 9, 2017
Step 8. Next, cut out a 3½-inch diameter circle from the other plastic weight boat. Then, trim
this disk as need, so that it will fit into the top of the male PVC coupling. This plastic disk will
serve as a diffuser for the PVC microscope, which will help to soften the intensity of the LED
light striking the sample. Also, depending upon the brightness or intensity of the LED light
used, this diffuser may or may not be necessary.
Figure 12. Cutting out a 3½-inch diameter circle from a plastic weight boat to use as a
diffuser in the adjustable stage of the microscope.
In addition, one simple way to make this plastic circle is to turn the male PVC coupling upside
down and place it on top of the plastic weigh boat. Then, use a permanent marker to trace
around the outside of the coupling. Finally, cut out the plastic circle and trim it to size.
http://pages.stolaf.edu/opn-lab/equipment/ 19 January 9, 2017
Step 9. To make the adjustable stage for this microscope, we follow the same steps described
in the Supporting Information (S4) for our OPN Scope article. Specifically, we cut ¾ of an inch
off the top of the male PVC coupling (Fig. 13, top right), using a chop saw.
Figure 13. Converting a standard set of 3-inch male and female PVC couplings (left) into
the adjustable stage for the PVC microscope (right).
Next, we place the female PVC coupling into in a vice that has been secured to the platform of a
drill press. We then drill a small (e.g., ⅛-inch) guide hole into the bottom half of the coupling
(roughly ¾ inches below the “collar” around the top). Then, using that guide hole, we drill a 1½-
inch diameter hole into the bottom half of the female PVC coupling (Fig. 13, bottom right), so
that students can easily turn on and off the push-button LED light that will be placed in there.
Also, we suggest using a “paddle” bit to drill this large hole into the PVC coupling since this type
bit is specifically design for this type of task, and the long pointed end of the paddle bit should
fit nicely into the guide hole. However, please be careful when using the paddle bit to drill into
the PVC coupling since the long pointed end of the bit can easily pierce the other side of the
coupling once the hole has been finished.
http://pages.stolaf.edu/opn-lab/equipment/ 20 January 9, 2017
Step 10. To assemble the adjustable stage, first lay out the parts (Fig. 14A) and then start
putting the stage together. Specifically, place the 3-inch female PVC coupling over the push-
button LED light and drop the plastic disk into the top of the 3-inch male PVC fitting (Fig. 14B).
Next, place the round PVC disk on top of the plastic disk (Fig. 14C) and, finally, twist the two
couplings together (Fig. 14D). Also, it is this very feature that enables the stage to provide a
“fine” focus for the microscope once the parts have been assembled.
Figure 14. Assembling the adjustable stage for this PVC microscope from its component
parts.
http://pages.stolaf.edu/opn-lab/equipment/ 21 January 9, 2017
Step 11. To finish assembling the microscope, wrap the 1-inch wide strip from the plastic weigh
boat around the PVC tube to hold it to the neck, as shown in Figure 15 below. Then, wrap
labeling (or other) tape around the top, bottom, and sides of the neck to hold the strip from the
weigh boat in place (Fig. 15). Next, wrap the rubber bands around the plastic strip and the back
of the support arm to hold the PVC tube into place (Fig. 15). Finally, place the adjustable stage
(with the push-button LED light) onto the base of the microscope. Now, the bright-field version
of the OPN Scope is ready to use.
Figure 15. A completed bright-field version of the OPN Scope.
http://pages.stolaf.edu/opn-lab/equipment/ 22 January 9, 2017
Using the Microscope
Although using the bright-field version of the OPN Scope is fairly intuitive, we provide some
basic guidance here on how to set up and focus the microscope.
Step 1. Place a slide on the adjustable stage and center it under the objective lens (Fig. 16), and
then turn on the push-button LED light by reaching into the large hole drilled into the female
PVC coupling.
Figure 16. Placing a prepared slide on the adjustable stage made from
PVC couplings.
http://pages.stolaf.edu/opn-lab/equipment/ 23 January 9, 2017
Step 2. To bring a sample into focus, first look through the eyepiece and slowly slide the PVC
tube up or down until the image becomes visible (Fig. 17). This is a coarse adjustment, which
may require fine adjustment later.
Figure 17. Moving the PVC tube up or down as a “coarse” adjustment to
bring an image into focus initially.
http://pages.stolaf.edu/opn-lab/equipment/ 24 January 9, 2017
Step 3. For fine focusing, hold the bottom of the adjustable PVC stage with one hand while
slowly turning the top with the other hand (Fig. 18). The image should become clear and sharp.
Also, while this approach does rotate the image in the field of view, we have not found this to
be an issue at lower magnifications (e.g., 40x or less).
Figure 18. Twisting the male PVC coupling while holding onto the female
coupling in order to provide a “fine” focus.
However, as explained in the Supporting Information (S4) for our OPN Scope paper, at higher
magnifications (e.g., 400x), this technique can rotate a sample out of the field of view. As a
result, students will likely need to use a steady hand when viewing samples at high
magnification (which might be more appropriate for older students or more advanced classes).
http://pages.stolaf.edu/opn-lab/equipment/ 25 January 9, 2017
With respect to the quality of the images, because this PVC microscope uses standard
microscope optics and a fixed tube length, we have found that it can generate images that are
comparable to those of a commercial model (Fig. 19).
Figure 19. Images of a Pinus root taken at 40x magnification using
a Celestron 2-MP digital imager and (A) the bright-field version of
the OPN Scope or (B) an Olympus CH-2 commercial microscope.
http://pages.stolaf.edu/opn-lab/equipment/ 26 January 9, 2017
In addition, there are a number of relatively inexpensive digital microscope cameras available
online that can be used in place of a standard eyepiece and which will enable students to take
photographs of their samples (e.g., AM Scope, Moticam, and OMAX). However, when selecting
a digital camera, please make sure to consider the weight of the device since the rubber bands
that hold the PVC tube in place may not support a heavy camera.
For example, one model that we have successfully used with this version of the OPN Scope is a
2-MegaPixel digital imager made by Celestron, which now provides related software that is
compatible with both Windows and Mac computers (click on the “Support” menu item in the
above link to access these programs). Celestron also makes a 5-MegaPixel digital imager as
well, which readers might find useful.
Converting the PVC Microscope for Overhead Illumination
Besides serving as a bright-field microscope, we have found that this version of the OPN Scope
can also be easily converted to provide overhead illumination for viewing other 3-dimensional
samples. To make this conversion, we simply use a rectangular piece of cardboard
(approximately 5 inches x 3 inches), an inexpensive keychain LED penlight (roughly, $2), and
four Scotch “Extreme” Fasteners ® to hold these parts to the PVC microscope (Fig. 20).
Figure 20. Placing Scotch “Extreme” Fasteners ® on
the cardboard triangle and LED penlight to provide
overhead illumination with the PVC microscope.
http://pages.stolaf.edu/opn-lab/equipment/ 27 January 9, 2017
Specifically, we fold the piece of cardboard into the shape of a triangle and then glue it closed
with super glue (Fig. 20). Note that, on one side, there should be a small amount of overlap
between the pieces, so that the super glue can hold the cardboard together (Fig. 20). Next, we
place Scotch “Extreme” Fasteners ® on the neck of the PVC microscope, the two larger sides of
the cardboard triangle, and the flat side of the LED penlight (Fig. 20).
We then fasten one side of the cardboard triangle to the neck of the PVC microscope and the
LED penlight to the other side of the cardboard triangle (Fig. 21). Students can then turn on the
penlight, place their samples in a Petri dish or on the flat PVC disk, and look at them under the
PVC microscope (Fig. 22).
Figure 21. Providing overhead illumination with the PVC
microscope.
http://pages.stolaf.edu/opn-lab/equipment/ 28 January 9, 2017
Of course, if using a standard 10x eyepiece, this version of the OPN Scope will still provide a
total of 40x magnification (since 10 x 4 = 40), as shown in the images of common items
contained in Figure 22 below. Nevertheless, as previously explained, readers can also purchase
5x eyepieces on the AM Scope website, which would cut the total magnification in half (i.e., to
20x since 5 x 4 = 20). Alternatively, readers could obtain eyepieces or objectives with even
lower magnifications, which would further reduce the total magnifying power of the scope.
Figure 22. Overhead images of (A) 60-grit sandpaper, (B), the edge of a rusty razor
blade, (C) the “W” in “We the People” from a U.S. Quarter, and (D) threads from a gauze
pad (all taken at 40x magnification with a Celestron 2-MP digital imager, using overhead
illumination with the PVC microscope).
Conclusion
Like other OPN instruments that we have developed, we hope that the bright-field version of
the OPN scope will help to promote science education among (and facilitate scientific
exploration by) students at all levels. We further encourage readers to use or modify this
design to fit their particular needs, and we include in Figure 23 below some examples of other
wooden or PVC microscopes that we have made to provide some inspiration on this point.
http://pages.stolaf.edu/opn-lab/equipment/ 29 January 9, 2017
Figure 23. Other wooden and PVC microscopes that we have made.
Hazards
As explained above, working with power tools can pose many significant risks, which are as
dangerous as they are obvious. As a result, readers who are unfamiliar with these types of
tools should work with an experienced craftsperson in order to avoid serious physical injury.
In addition, readers should make sure to wear the proper protective equipment when working
with tools of any kind, including wearing the necessary eye and ear protection.
Disclosures
The authors declare that they have no conflicts of interest related to any product, brand,
company, website, or other item discussed in this manual. Indeed, as with other open source
instruments and equipment that we have developed, http://pages.stolaf.edu/opn-
lab/equipment/, we encourage readers to improve upon the designs set forth in this manual by
using other materials and equipment in addition to the insights and inspirations provided by
their own experiences.
http://pages.stolaf.edu/opn-lab/equipment/ 30 January 9, 2017
References
Caplan, G.M. (2009). Making a home-made microscope. Available online at
https://legacy.owensboro.kctcs.edu/gcaplan/science/13.%20%20Home%20Made%20Microsco
pe.htm. Adapting Curry, A., Grayson, R.F., and Hosey, G.R. (1982). Under the Microscope. Van
Nostrand Reinhold Co., New York.
Carboni, G. (2011). A one-dollar microscope. Available online at
http://www.funsci.com/fun3_en/ucomp1/ucomp1.htm.
Cybulski J.S., Clements J., Prakash M. (2014). Foldscope: Origami-based paper microscope.
PLoS ONE. 9(6): e98781. doi:10.1371/journal.pone.0098781. Available online at
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0098781
Drace, K., Couch, B., and Keeling, P.J. (2012). Increasing student understanding of microscope
optics by building and testing the limits of simple, hand-made model microscopes. J. Microbiol.
& Biol. Educ. 13(1), 45-49.
Dubowsky, N. (1996). Revisiting Antony van Leeuwenhoek: Using a student-constructed
simple microscope to stimulate interest in microscopy. Amer. Biol. Teacher. 58(4), 242-244.
Mahaffey, J.A. (1999). Making Leeuwenhoek proud: Building simple microscopes. Optics &
Photonics News. 10(3), 62-63.
OPN Lab Equipment Web Page. (2016). Available online at http://pages.stolaf.edu/opn-
lab/equipment/.
Stewart, C. and Giannini, J. (2016). Inexpensive, open source epifluorescence microscopes. J.
Chem. Educ. 93(7), 1310-1315. Available online at
http://pubs.acs.org/doi/abs/10.1021/acs.jchemed.5b00984.
Tsagliotis, N. (2012). Build your own microscope: Following in Robert Hooke’s footsteps.
Science in School: Euro. J. for Sci. Teach. 22, 29-35. Available online at
http://www.scienceinschool.org/2012/issue22/microscope.