Upload
samibulut542
View
232
Download
0
Embed Size (px)
Citation preview
7/29/2019 High Pressure Differantial Scanning Calorimetry
1/18
Journal of Food Process Engineering 27 (2004) 359376.All Rights Reserved.
Copyright 2004, Blackwell Publishing 359
HIGH-PRESSURE DIFFERENTIAL SCANNING CALORIMETRY
(DSC): EQUIPMENT AND TECHNIQUE VALIDATION USING
WATERICE PHASE-TRANSITION DATA
S. ZHU1
, S. BULUT2
, A. LE BAIL2
and H.S. RAMASWAMY1,3
1Department of Food Science and Agricultural ChemistryMcGill University
21111 Lakeshore Road,Ste Anne de Bellevue,
Quebec H9X3V9Canada2GEPA-ENITIAA (UMR CNRS 6144-SPI)
Rue de la Graudire BP 82225,F-44322 Nantes Cedex 03, France
Accepted for Publication June 18, 2004
ABSTRACT
Understanding phase transition during high-pressure (HP) processing of
foods is important both with respect to optimizing the process and improve-
ment of product quality, but scientific information available in this area is very
limited. In this study, the phase-transition behavior of water was evaluated
using a HP differential scanning calorimetry (DSC). Tests were carried out
under both isothermal pressure-scan (P-scan) and isobaric temperature-scan
(T-scan) modes with distilled water prefrozen in the sample cell. P-scan was
carried out at 0.3 MPa/min at two temperatures, -10 and-20C, and T-scan
was carried out at 0.15C/min at two pressures, 0.1 and 115 MPa. The
pressure-induced phase transition of water was accurately reproduced by
the P-scan test. Ice melting latent heat during P-scan showed no significant
difference (P>0.05) from the available reference data in literature. The
relationship between P-scan tested (Lm) and reference latent heat wasLm=0.987 L (R2=0.99, n =6) suggesting a mean error less than 2%. T-scan
mode was less appropriate and did not yield promising result. Measured
values were less accurate than P-scan probably because of the influence of
large heat capacity of sample cell. However, reliable and reproducible results
obtained under P-scan mode suggested that the HP DSC can be used for the
calorimetric determination of pressure-dependent water-phase transition in
real food systems during HP freezing/thawing operations.
Blackwell Science, LtdOxford, UKJFPEJournal of Food Process Engineering0145-8876Copyright 2004 by Food & Nutrition Press, Inc., Trumbull, Connecticut.2004275359376Original Article HIGH-PRESSURE DSC: EQUIPMENT/TECH-
NIQUE VALIDATION S. ZHU
ET AL.
3 Corresponding author. TEL: +1-514-398-7919; FAX: +1-514-398-7977; EMAIL:[email protected]
7/29/2019 High Pressure Differantial Scanning Calorimetry
2/18
360 S. ZHUET AL.
INTRODUCTION
High-pressure (HP) process is receiving considerable attention because
of its advantages and potential benefits for the preservation and modification
of foods. Phase transition is one of the major domains that are of potential
interest in HP processing of foods. Pressure-induced phase transitions, such
as freezing/thawing of high-moisture foods and crystallization of lipids, offer
numerous opportunities for process or product development (Knorr 1999).
Water undergoes different types of phase transitions when subjected to differ-
ent pressuretemperature domains (Bridgman 1912). This phenomenon
allows carrying out different freezing/thawing applications such as pressure
shift freezing, HP freezing and HP thawing (Cheftel et al. 2000). HP in this
context generally refers to pressures in excess of several hundred atmospheres
(>50 MPa).Calorimetry is a powerful tool to study thermodynamic properties of
materials and phase-change phenomena. Normally, calorimetric experiments
are carried out with a differential scanning calorimeter (DSC) at a constant
(usually atmospheric) pressure using temperature as the working parameter.
Bridgman (1912) developed experimental apparatus to measure phase transi-
tion and volume change of pure water under pressure, and to estimate latent
heat of water using Clapeyron equation. Hogenboom et al. (1995) developed
a similar apparatus with a novel piezometer to measure volume variation and
phase diagram of an aqueous magnesium sulfate solution.
In HP calorimetry, the cell for holding the sample should be strong
enough to resist high pressure. Thus, the cell becomes large in terms of mass
and heat capacity, which presents a major challenge to obtain accurate results
for temperature scan (T-scan) tests. To overcome this difficulty, an alternative
measure of pressure scan (P-scan) at a constant temperature is adopted in HP
calorimetry. The pressure can be changed either as a step change (Pruzan et al.
1979) or at a constant rate (Randzio et al. 1994). Several studies have been
carried out using P-scan tests for gathering thermal properties of water and
aqueous solution (Chourot et al. 2000; Le Bail et al. 2001) and some organic
liquids such as hexane, butane-1, benzene, toluene (Fuchs et al. 1979; Pruzan
et al. 1979; TerMinassian et al. 1988; Grolier and Randzio 1997) under HP
conditions. Two types of calorimetric apparatus have been used in terms of
working conditions: constant mass or constant volume. Most of the studies in
literature are related to liquids and used constant volume technique (Pruzan
et al. 1979; Randzio et al. 1994; Le Bail et al. 2001).
Most foods and food products contain a high percentage of moisture.
Thermo-physical properties of foods have been generally linked to moisture
content, temperature, phase and pressure. Phase transition of water in these
foods during HP process can be more complicated than that of pure water.
7/29/2019 High Pressure Differantial Scanning Calorimetry
3/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 361
Scientific information related to HP calorimetry of foods is scarce and poorly
documented. The objective of this study was to evaluate the performance of
HP DSC for characterizing the phase-transition behavior of water at high
pressures by employing two modes: (1) the traditional T-Scan and (2) the
more recommended P-scan. A specific objective was to validate the tech-
nique by comparing the experimental results on phase-transition tempera-
tures and latent heat of pure water with published data and thereby pave way
for the use of the technique for gathering data on HP water-phase transition
in foods.
MATERIALS AND METHODS
HP Calorimetric System
A HP DSC (Fig. 1) was used in this study. The experimental system
consisted of a differential calorimetry head, a refrigerated circulator, two HP
cells, a HP compressor and a computer. The calorimetry head (Pass 27, Sceres,
Orsay, France) had two cavities (20 mm diameter 95 mm depth) used forholding the sample and the reference cells. There were 220 thermocouple
junctions installed between the two cavities to amplify the generated temper-
ature differential signal. Two platinum thermometers (Pt 100) were placed
under each of the cavities to monitor/control temperature of the calorimeter.
The temperature of the calorimetry head was controlled through a copper coil
winding around the contour of the head (Fig. 1a) and an oven around the
system. The copper coil was connected to the refrigerated circulator (Huber
CC250, Offenburg, Germany) with a water-glycol medium. During the exper-
iment, the calorimetric head together with copper coil was thermally insulated
in plastic box containing foamed plastics.
The HP cell (made of beryllium copper) contained two plugs with nitrile
O-ring and a threaded bolt for sealing on one side (Fig. 2). It was connected
to a pressure tube (3.2 mm diameter) on the other side with miniature fittings
(M2 Serie, 100 MPa, Harwood Engineering, MA). The pressure within the
system was achieved through the HP compressor (400 MPa, 5 cm3, Nova-
Swiss, Effretikon, CH) driven by a step motor (MO63-LE09, Mijno, Fenwick,
France) and controlled by the computer. A pressure sensor (200 or 400 MPa,
Asco Instruments, Chateaufort, France) was used to monitor and control the
pressure. When the pressure required for a test exceeded 180 MPa, the 400-
MPa sensor was used during the experiment. A software (Labview 6, National
instruments, Austin, TX) was used for system control and data recording
(temperature, pressure and heat flow). Pentane (Sigma, Fallavier, France) was
used as pressurization medium in the system because of its stability on the
7/29/2019 High Pressure Differantial Scanning Calorimetry
4/18
362 S. ZHUET AL.
pressuretemperature domain tested and its properties (low viscosity, no phase
change) (Bridgman 1970; Scaife and Lyons 1980).
Calibration
The pressure sensor was calibrated against a reference pressure gauge
(Bourdon, France). Temperature of the calorimeter was calibrated against a
FIG. 1. EXPERIMENTAL SET-UP OF HP DIFFERENTIAL SCANNING CALORIMETER:
(a) schematic diagram and (b) photograph of the calorimetric head and the cells.
Calorimetric head
Copper coil
HP cell
HP tube
(b)
Refrigerated
circulator
Computer
Cell
HP compressor
Medium
reservoir
HPsensor
Calorimetric head
Coppercoil
(a)
7/29/2019 High Pressure Differantial Scanning Calorimetry
5/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 363
K-type thermocouple (0.1 mm diameter, Omega, Stamford, CT) placed in the
sample cell at selected temperature between -20 and 20C. The calorimeterwas calibrated using a 100 W resistance settled in the sample cell. A voltagebetween 1 and 4.2 V was supplied to this resistance by a power source (220
programmable voltage source, Keithley Instruments SARL, Saint-Aubin,
France) for period between 30 and 90 min. The heat generated from theresistance was evaluated with the voltage and the current measured with a
voltmeter (7055 Voltmeter, Solartron Mobrey SA, Saint-Christophe, France)
and an ampere meter (195A multimeter, Keithley Instruments, SARL, Saint-
Aubin, France). The calorimetry differential temperature signal (mV) attrib-
uted to the electric heating was recorded every 5 s. The peak area was then
integrated to calculate the ratio of heat power to calorimetric voltage at
selected temperatures. The calibration tests were carried out from -20 to 20C.A linear fit of the experimental data was obtained as:
(1)
where q is heat flow rate (mW); Sis calorimetric signal (mV); kis the ratio
of heat flow rate to calorimetric signal (mW/mV); Tis the temperature of the
calibration tests performed (C). Similar calibration procedures have beenused in the study by Chourot et al. (2000) and by Le Bail et al. (2001).
Sample Preparation
All experiments were carried out using distilled water. About 0.5 g of
pure water (weight accurately measured up to milligram level; example
0.5712 g) was vacuum-packaged in a polyethylene bag (80-mm-thick multi-player film) (La Bovida, Nice, France). For the test, the sample was installed
q kS T S R n= = +( ) = =( )37 1 0 080 0 978 82. . . , ,
FIG. 2. SCHEME OF THE HIGH-PRESSURE CELL AND SAMPLE INSTALLATION
7/29/2019 High Pressure Differantial Scanning Calorimetry
6/18
364 S. ZHUET AL.
in the sample cell (see Figs. 1 and 2). The reference cell was prepared in the
same way as the sample cell but without the sample. Air bubble was carefully
removed from the cell during the preparation of the two-calorimeter cells.
Isothermal Pressure-Scan (P-Scan)
After sample installation, the sample and reference cells were placed in
the sample and reference cavities, respectively. The sample was frozen at the
set temperature (-10 or -20C) in the calorimeter. Once the calorimeter signalshowed a stable baseline (close to zero), the pressure was increased from the
atmospheric level (0.1 MPa) at a constant rate (0.3 MPa/min) up to a level
high enough for completion of the melting process (e.g., 160 MPa for the P-scan at -10C) and heat flow signal was recorded every 5 s (Fig. 3). When thepressure reached the corresponding phase-change temperature, ice started
melting (see point 1 in Fig. 3), resulting in a peak of heat flow to compensate
for the temperature drop. A low P-scan rate was employed in this study, as
compared to 1 MPa/min in some previous studies (Le Bail et al. 2001), in
order to obtain a well defined span of calorimetric peak.
Isobaric Temperature-Scan (T-Scan)
T-scan tests were carried out at constant pressure (Fig. 4) by increasing
the temperature at a prescribed rate. Before each T-scan test, the test sample
was frozen in the calorimeter and then stabilized at 0.1 MPa or 115 MPa,
corresponding to a phase-change temperature of about 0 or -10C, respectively(Bridgman 1912). When calorimetric signals approached a balance under a
constant pressure, the temperature was increased from lower than -15C at aconstant rate (about 0.15C/min) and the calorimeter signals were recorded
every 5 s.
RESULTS AND DISCUSSION
Isothermal P-scan
Figure 5 shows a typical experimental differential thermograph of an
isothermal P-scan with pure ice at the calorimetric temperature of-10C. Bycalibration using Eq. (1), calorimetric data is converted into heat flow rate.
Figure 6a demonstrates two peaks of heat flow rate against pressure during P-
scan of the ice samples at calorimetric temperature -10 and -20C, respec-tively. The HP DSC heat flow signal appeared a straight baseline as pressure
was increased at a constant rate. As the pressure level increased, the signal
started deviating from the base line (1 in Figs. 3 and 5). This transition was
7/29/2019 High Pressure Differantial Scanning Calorimetry
7/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 365
FIG. 3. SCHEMATIC DESCRIPTION OF ISOTHERMAL PRESSURE-SCAN: (a) water
phase-change diagram and pressurization and (b) corresponding calorimetric signal (Tc is calorimetric
temperture [C]).
21
0
Tem
perature(C)
Ice-I
Water
TC
2100.1
Pressure (MPa)
P-scan
P1 P3P2
1
2
3
(a)
P1 P2 P30.1
Calorimetricsignal(endodown) 0
Peak
(melting end)
Peak onset
(melting start)
Peak end
Peak baseline
Pressure (MPa)(b)
1 3
2
7/29/2019 High Pressure Differantial Scanning Calorimetry
8/18
366 S. ZHUET AL.
FIG. 4. SCHEMATIC DESCRIPTION OF ISOBARIC TEMPERATURE-SCAN WITH WATER
PHASE-CHANGE DIAGRAM
21
0
T
emperature(C)
Ice-I
Water
2100.1Pressure (MPa)
T-scan
FIG. 5. A TYPICAL MEASUREMENT OF ISOTHERMAL PRESSURE-SCAN (0.3 MPa/min) OF
PURE ICE (0.5712 g) WITH CALORIMETRIC TEMPERATURE AT -10C
1.8
1.5
1.20.9
0.6
0.3
0.0
0.3
0 60 120 180 240 300 360
Time (min)
Heatflo
wsignal(mV)
20
40
6080
100
120
140
160
Press
ure(MPa)
Pressure
Heat flow signal 1 3
2
t1 t2 t3
7/29/2019 High Pressure Differantial Scanning Calorimetry
9/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 367
defined as onset temperature in this study. In the case of P-scan test, this
onset refers to the pressure that is associated with the beginning of the melting
of test sample (ice). For P-scan at -10 and -20C, the onset pressure was111 1.8 (mean SD, n= 3) and 194 1.5 MPa (n= 3) (Fig. 6a), respec-tively, which matched well with phase-change diagram of waterice I
(Bridgman 1912), as empirically expressed by Bridgman (1912) as follows:
(2)
where TP is phase-change temperature of water (C); P is pressure (MPa)(Chourot et al. 1997).
After onset, the temperature of the ice/water mixture decreases with a
further increase in pressure (12 in Fig. 3a), following pressuretemperature
relationship of the phase change of waterice I (Bridgman 1912; Le Bail et al.
2002). This phenomenon enhances the heat transfer between sample and its
surroundings (cell, calorimeter, etc.) because the HP DSC was set at constant
temperature during P-scan tests. Thus the absolute value of calorimetric signal
increases until complete melting of the sample (2 in Figs. 3 and 5) and then
decreases back to the initial baseline (23 in Figs. 3 and 5) representing an
inert or nonevent (melting) phase. Therefore, a pressure-dependent endot-
hermic calorimetric peak can be obtained for each P-scan experiment
(Fig. 6a). For the sample mass and the scan rate (0.3 MPa/min) used, the
endothermic peak (heat flow) spanned over 35 1.4 (n= 3) and 27 1.1 MPa(n= 3) with the summit occurring at 126 2.3 (n= 3) and 205 1.6 MPa(n= 3) for P-scan at -10 and -20C, respectively. The peak at the lowertemperature (-20C) occurred at a higher pressure and also had a lower summitvalue and a shorter bandwidth (Fig. 6a). This is because of the pressure-
dependent characteristics of ice Iwater phase change (Eq. 2) and latent heat
as empirically determined by Bridgman (1912):
(3)
whereL is latent heat of waterice I (J/g) (Chourot et al. 1997).
The shape and size of the P-scan endothermic (heat flow rate) peak
depends on several factors, such as sample mass, P-scan rate, thermo-physical
properties of the cell and pressure medium and so on. As in conventional DSC
experiments, a higher scan rate (either T-scan or P-scan) yields a wider span
of the peak. During a conventional DSC test, the sample span (usually very
small and made of out of light weight aluminum) is the only barrier to the
heat-transfer in the chamber between the heating medium and test/reference
samples. A larger heat capacity or a lower thermal conductivity of test cells
implies a lower heat transfer rate and therefore a longer duration to reach and
T P PP = - -0 0722 0 0001552
. .
L P P= - -333 0 399 0 000388 2. .
7/29/2019 High Pressure Differantial Scanning Calorimetry
10/18
368 S. ZHUET AL.
FIG. 6. PRESSURE-SCAN (0.3 MPa/min) OF PURE ICE WITH CALORIMETRIC
TEMPERATURE AT -10 AND -20C: (A) THAWING HEAT FLOW RATE AND (B) THAWINGLATENT HEAT
100
80
60
40
20
0
50 100 150 200 250
Pressure (MPa)
Heatflowrate(mW/g)
10C 20C
10C 20C
(a)
0
50
100
150
200
250
300
50 100 150 200 250
Pressure (MPa)
Thawinglaten
theat(J/g)
(b)
7/29/2019 High Pressure Differantial Scanning Calorimetry
11/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 369
pass the peak, thus resulting in a wide peak span. Attributed to higher pres-
sures associated with HP DSC, the cell system itself is large and constructed
of more sturdy material. The large heat capacity of the cell system strongly
affects the span of the P-scan signal. Hence an appropriate calibration is vital
for obtaining realistic data from these cells. When using a similar HP colo-
rimeter with 1 g water and P-scan at 1 MPa/min, Le Bail et al. (2001) found
the peak to span over 60 MPa. Randzio et al. (1994) reported a peak span of
70 MPa for solidliquid transition in benzene at a P-scan at 0.3 MPa/min. The
peak span observed was smaller than these reported values because the sample
mass was smaller and further a lower P-scan rate was used in this study.
The onset and summit of the P-scan endothermic peak is related to the
start and end of ice melting, respectively (Le Bail et al. 2002). A weighted-
average pressure ( 1-2) was used to compare results from the P-scan tests
with the reference diagram of Bridgman (1912), which was mathematically
defined as:
(4)
where qb is the baseline of heat flow rate (mW); tis time (s); t1 and t2 are the
time at peak onset and summit (s) (Fig. 5). The average pressure during ice
melting was 119 2.2 (n= 3) and 200 1.6 MPa (n= 3) for P-scan at -10and -20C, respectively (Table 1), as compared with the reference values of112 1.4 and 195 1.0 MPa, respectively (Bridgman 1912) (Fig. 7).
The area of endothermic heat flow rate peak represents the quantity of
the heat transfer during the HP DSC test. The latent heat was evaluated by
P
PP q q t
q q t
t
t1 21
1
2
2-=
-( )
-( )
b
b
d
d
TABLE 1.
CHARACTERIZATION OF CALORIMETRIC SIGNAL PEAK (MEAN SD) FROMISOTHERMAL PRESSURE-SCAN (0.3 MPa/min) EXPERIMENTS USING DISTILLED WATER
Calorimetric temperature -10 0.2C (n= 3) -20 0.2C (n= 3)
Pressure span of peak (MPa) 34.8 1.4 25.9 1.1Pressure at peak onset (P1, MPa) 111 1.8 194 1.5Reference phase-change pressure (MPa)* 112 1.4 195 1.0Average pressure (
1-2, MPa) 119 2.2 200 1.6
Pressure at peak summit (P2, MPa) 126 2.3 205 1.6Measured latent heat (Lm, J/g) 278 1.0 233 1.9Calculated latent heat (L( 1-2), J/g)* 280 1.1 237 1.4
* Values were calculated using Eq. (2) or (3).
P
P
7/29/2019 High Pressure Differantial Scanning Calorimetry
12/18
370 S. ZHUET AL.
integrating the calorimetric peak calibrated with the conversion ratio (Eq. 1)
and subtracting the baseline:
(5)
where Lm is measured latent heat of waterice I (J/g); m is water mass (g).
Figure 6b shows the accumulation of thawing latent heat during the P-scan of
test using pure water sample (ice). The top value of each curve in Fig. 6b
indicates the total amount of absorbed or the latent heat released by the test
sample, which was 278 1.0 (n= 3) and 233 1.9 J/g (n= 3) for P-scan at-10 and -20C, respectively. By substituting the average pressure ( 1-2) intoEq. (3) (Bridgman 1912), the latent heat [L( 1-2) = 333 - 0.399 1-2-0.000388 1-2] as 280 1.1 and 237 1.4 J/g for P-scan at -10 and -20C,respectively, was obtained (Table 1). For all P-scan tests combined, statistical
analysis (t-test: paired two sample for means) indicated no significant differ-
ence (P> 0.05, d.f. = 1/6) between experimentally evaluated latent heat (Lm)from this study and the calculated reference [L( 1-2)] latent heat (Bridgman
1912). Based on the regression relationship between the values from the two
Lm
q q tm b d= -( )1
1000
P
P P
P
P
FIG. 7. PHASE TRANSITION BETWEEN PURE WATER AND ICE I OBSERVED DURING
PRESSURE-SCAN TESTS AS COMPARED TO THE REFERENCE DIAGRAM
(BRIDGMAN 1912)
25
20
15
10
5
0
5
50 100 150 200 250
Pressure (MPa)
Temperature(C)
Average ( 12P )
Ice-I
Water
Reference diagram
Peak summit (melting end)
0.1
7/29/2019 High Pressure Differantial Scanning Calorimetry
13/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 371
methods, the following equation was obtainedLm= 0.987L( 1-2) (R2= 0.99,n= 6) for P-scan tests (Fig. 8).
Isobaric T-scan
Figure 9 is a typical calorimetric signal plot of a T-scan experiment
(0.15C/min, 115 MPa, 0.5712 g ice). Figure 10 shows two peaks of heat flow
rate against temperature during T-scan at atmospheric conditions (0.1 MPa)
and at 115 MPa, respectively. Table 2 summarizes the results from isobaric T-
scan tests. The endothermic peak at the higher pressure occurred at a lower
temperature with a smaller peak area (Fig. 10a) because of the pressure
related thermo-properties of water. The peak-onset temperature was -0.2 0.2and -10.5 0.3C (n= 3) at 0.1 and 115 MPa, respectively (Table 2), as com-pared with calculated phase-change temperatures of 0 0.0 and -10.4 0.1Cby Eq. (2), which correspond very well.
Compared with isothermal P-scan peak, T-scan peak appeared larger in
signal value because ice-melting duration in T-scan (Fig. 9) was shorter than
that in P-scan (Fig. 5). However, T-scan rate (0.15C/min) in this study was
quite small as compared with that used in conventional DSC test with very
tiny sample (usually 1030 mg). HP DSC requires pressure resistant cells to
P
FIG. 8. MELTING LATENT HEAT OF PURE ICE MEASURED BY PRESSURE-SCAN (P-SCAN)
TESTS AS COMPARE TO THE REFERENCE DATA (BRIDGMAN 1912)
200
250
300
350
400
0 50 100 150 200 250
Pressure (MPa)
Latentheat(J/g)
Reference
P-scan
7/29/2019 High Pressure Differantial Scanning Calorimetry
14/18
372 S. ZHUET AL.
hold sample. Thus sample size should be large enough to reduce the ratio of
noise to HP calorimetric signal (Le Bail et al. 2001). A large sample can
strengthen the signal, but tend to increase the span of the peak. Because of
the large mass of the HP cell and sample, a low T-scan rate was selected to
obtain a peak span as narrow as possible. The peak span was 10.2 0.5 (n= 3)and 9.8 0.4C (n= 3), showing peek summit at 3.8 0.3 and -6.8 0.3C,for T-scan (0.15C/min) of the ice sample at 0.1 and 115 MPa, respectively
(Table 2). Although a low T-scan rate was used, the peak was much delayed
FIG. 9. A TYPICAL MEASUREMENT OF ISOBARIC TEMPERATURE-SCAN (0.15C/min) OF
PURE ICE (0.5712 g) AT 115 MPa
3
2
1
0
1
0 20 40 60 80 100 120
Time (min)
Heatflowsignal(mV)
15
10
5
0
5
Calo
rimetrictemperature(C)Heat flow signal
Temperature
TABLE 2.
CHARACTERIZATION OF CALORIMETRIC SIGNAL PEAK (MEAN SD) FROM ISOBARICTEMPERATURE-SCAN (0.15C/min) EXPERIMENTS USING DISTILLED WATER
Calorimeter pressure 0.1 0.0 MPa (n= 3) 115 0.9 MPa (n= 3)
Temperature at peak onset (C) -0.2 0.2 -10.5 0.3Phase-change temperature (C)* 0 0.0 -10.4 0.1Temperature at peak summit (C) 3.8 0.3 -6.8 0.3
Temperature span of peak (C) 10.2 0.5 9.8 0.4Measured latent heat (J/g) 318 2.1 277 2.4Calculated latent heat (J/g)* 333 0.0 282 0.5
* Values were calculated using Eq. (2) or (3).
7/29/2019 High Pressure Differantial Scanning Calorimetry
15/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 373
FIG. 10. TEMPERATURE SCAN (0.15C/min) OF PURE ICE AT 0.1 AND 115 MPa:
(a) thawing heat flow rate and (b) thawing latent heat
200
150
100
50
0
50
Hea
tflowrate(mW/g)
(a)
115 MPa 0.1 MPa
0
50
100
150
200
250
300
350
15 10 5 0 5 10 15 20
Temperature (C)
15 10 5 0 5 10 15 20
Temperature (C)
Thawinglatentheat(J/g)
(b)
115 MPa0.1 MPa
7/29/2019 High Pressure Differantial Scanning Calorimetry
16/18
374 S. ZHUET AL.
because of the large heat quantity required to keep temperature increase in
sample and reference cells. The phase transition of water couldnt be well
demonstrated during T-scan test under current experimental conditions.
The latent heat was estimated using a similar integrating method to the
P-scan peak analysis. The value was 318 2.1 and 277 2.4 J/g for T-scan at0.1 and 115 MPa, respectively (Fig. 10b). Mathematically, the relationship
between measured (Lm) and reference latent heat (L) was expressed as
Lm= 0.961L (R2= 0.95, n= 6). Although the regression agreement was good,and the values seem to be pretty close to each other, statistical analysis (t-test:
paired two sample for means) demonstrated that measured value was signifi-
cantly (P < 0.05, d.f. = 1/6) lower than reference value of Bridgman (1912):318 versus 333 and 277 versus 282 J/g, at atmospheric and HP conditions,
respectively. T-scan test generally showed results that are slightly less accurate
than P-scan probably attributed to influence of large heat capacity of the mass
of the cell and pressure medium during T-scan (temperature increase). While
in P-scan test, the HP DSC worked at constant temperature. Overall the
difference between the values from the two methods and those published in
some classical literature was within a 10% error margin.
CONCLUSIONS
A HP DSC was examined for the evaluation of HP phase-transition
behavior of pure water (in the melting mode). Two techniques (T-scan and P-
scan) were used. Results obtained were statistical differences between the
techniques as well as between the experimental and some published models.
The pressure induced phase transition of ice Iwater was well pictured by P-
scan. The latent heat of ice melting obtained by isothermal P-scan showed no
significant difference (P> 0.05) from reference data. This good agreementvalidated the reliability and accuracy of the HP DSC system during the
isothermal P-scan test. HP calorimetry requires large cell mass to resist pres-
sure, which could reduce the accuracy/efficiency of calorimetric signal during
isobaric T-scan test. T-scan tests produced good and consistent results but were
slightly less accurate than P-scan counterparts or published models. The
results obtained showed that the system could be successfully used for gath-
ering HP phase-transition information, which is essential for the better under-
standing of pressure-dependent phase transition during HP freezing/thawing
of foods.
ACKNOWLEDGMENTS
This study was partially funded by the European Project of Safe Ice
(QLK1-CT-2002-02230) and the Strategic Grants Program of the Natural
7/29/2019 High Pressure Differantial Scanning Calorimetry
17/18
HIGH-PRESSURE DSC: EQUIPMENT/TECHNIQUE VALIDATION 375
Sciences and Engineering Research Council of Canada. Experiments were
carried out at ENITIAA, Nantes, France and analyses were performed at
McGill. The authors wish to thank L. Guihard and J. Laurenceau for their
technical support.
REFERENCES
BRIDGMAN, P.W. 1912. Water in the liquid and five solid forms under
pressure. P. Am. Acad. Arts Sci. 47(13), 441558.
BRIDGMAN, P.W. 1970. P-V-T relations in fluids. In The Physics of the
Pressure, (P.W. Bridgman, ed.) pp. 98148, Dover Publications, New
York.
CHEFTEL, J.C., LEVY, J. and DUMAY, E. 2000. Pressure-assisted freezing
and thawing: Principles and potential applications. Food Rev. Int. 16(4),
453483.
CHOUROT, J.M., BOILLEREAUX, L., HAVET, M. and LE BAIL, A. 1997.
Numerical modeling of high pressure thawing: application to water thaw-
ing. J. Food Eng. 34, 6375.
CHOUROT, J.M., LEBAIL, A. and CHEVALIER, D. 2000. Phase diagram
of aqueous solution at high pressure and low temperature. High Pressure
Res. 19(16), 191199.
FUCHS, A.H., PRUZAN, P. and TERMINASSIAN, L. 1979. Thermal expan-
sion of benzene at high pressure determined by a calorimetric method,
its behavior near melting. J. Phys. Chem. Solids 40, 369374.
GROLIER, J.P.E. and RANDZIO, S.L. 1997. Pressure effects on the thermo-
dynamic properties of (n-hexane plus 1-hexanol) binary mixtures. Fluid
Phase Equilibr. 133, 3544.
HOGENBOOM, D.L., KARGEL, J.S., GANASAN, J.P. and LEE, L. 1995.
Magnesium sulfate-water to 400 MPa using a novel piezometer: Densi-
ties, phase equilibria and planetological implications. Icarus 115, 258
277.
KNORR, D. 1999. Process assessment of high-pressure processing of foods:
An overview. In Processing Foods: Quality Optimization and Process
Assessment, (F.A.R. Oliveira and J.C. Oliveira, eds.) pp. 249267, CRC
Press, Boca Raton.
LE BAIL, A., CHEVALIER, D. and CHOUROT, J.M. 2002. High pressure
calorimetry: Application to phase change under pressure. In Trends in
High Pressure Bioscience and Biotechnology, (R. Hayashi, ed.) pp. 607
613, Elsevier Science, Amsterdam.
LE BAIL, A., CHEVALIER, D., CHOUROT, J.M. and MONTEAU, J.Y.
2001. High pressure calorimetry: Comparison of two systems
7/29/2019 High Pressure Differantial Scanning Calorimetry
18/18
376 S. ZHUET AL.
(differential vs. single cell), application to the phase change of water
under pressure. J. Therm. Anal. Calorim. 66(1), 243253.
PRUZAN, P., TERMINASSIAN, L. and SOULARD, A. 1979. Entropy deter-
mined from heat of compression at high pressure for benzene and carbon
tetrachloride. High-Pressure Sci. Technol. 1, 368378.
RANDZIO, S.L., GROLLIER, J.P. and QUINT, J.R. 1994. An isothermal
scanning calorimeter controlled by linear pressure variations from 0.1 to
400 MPa. Calibration and comparison with the piezothermal technique.
Rev. Sci. Instrum. 65(4), 960965.
SCAIFE, W.G.S. and LYONS, C.G.R. 1980. Dielectric permittivity and pvT
data of some n-alkanes. P. Roy. Soc. Lond. A Mat. 370, 193211.
TERMINASSIAN, L., BOUZAR, K. and ALBA, C. 1988. Thermodynamic
properties of liquid toluene. J. Phys Chem. 92, 487493.