Upload
franjo
View
212
Download
0
Embed Size (px)
Citation preview
BIOMEDICAL PAPER
“Gold standard” data for evaluation and comparison of 3D/2Dregistration methods
DEJAN TOMAZEVIC, BOSTJAN LIKAR, & FRANJO PERNUS
Faculty of Electrical Engineering, University of Ljubljana, Ljubljana, Slovenia
(Received 15 January 2003; revised version received 26 February 2004; accepted 15 May 2004)
AbstractEvaluation and comparison of registration techniques for image-guided surgery is an important problem that has receivedlittle attention in the literature. In this paper we address the challenging problem of generating reliable “gold standard”data for use in evaluating the accuracy of 3D/2D registrations. We have devised a cadaveric lumbar spine phantom with fidu-cial markers and established highly accurate correspondences between 3D CTand MR images and 18 2D X-ray images. Theexpected target registration errors for target points on the pedicles are less than 0.26 mm for CT-to-X-ray registration and lessthan 0.42 mm for MR-to-X-ray registration. As such, the “gold standard” data, which has been made publicly available onthe Internet (http://lit.fe.uni-lj.si/Downloads/downloads.asp), is useful for evaluation and comparison of 3D/2D image regis-tration methods.
Keywords: 3D/2D image registration, computed tomography, “gold standard”, image-guided therapy, magnetic resonance,spine phantom, X-ray projections
Key link: http://lit.fe.uni-lj.si/Downloads/downloads.asp
Introduction
In image-guided surgery, 3D preoperative medical
data, such as computed tomography (CT) and mag-
netic resonance (MR) images, are commonly used to
plan, simulate, guide, or otherwise assist a surgeon in
performing a medical procedure. The plan, specify-
ing how tasks are to be performed during surgery, is
developed in the coordinate system of a preoperative
image. To monitor and guide a surgical procedure,
the preoperative image and plan must be transformed
into physical space, i.e., a patient-related coordinate
system. This spatial transformation is obtained by
acquiring intraoperative data and registering them
to data extracted from preoperative images [1].
More recent and promising approaches to obtaining
the spatial transformation rely on intraoperative
X-ray projections acquired with a calibrated X-ray
device. The location and orientation of a structure
in a 3D CT or MR image with respect to the geo-
metry of the X-ray device is determined by 3D/2D
registration [2–5]. A necessary step before wide-
spread clinical use of any novel registration technique
is the evaluation and validation of the method [6].
One difficulty in evaluating a registration technique
is the need for a highly accurate “gold standard”.
Motivated by the continuing need for data sets and
“gold standards” to test, validate, and measure the
accuracy of different 3D/2D registration techniques,
we have devised a cadaveric lumbar spine phantom,
because it is very hard to establish a “gold standard”
using real patient data. In this paper we describe the
construction of the phantom to which fiducial
markers were attached. Three-dimensional CT, MR
and 2D X-ray images were acquired, and accurate
“gold standard” rigid registration between 3D and
Correspondence: Prof. Franjo Pernus, Ph.D., University of Ljubljana, Department of Electrical Engineering, Trzaska 25, 1000 Ljubljana,Slovenia. Tel: þ386 1 4768 312. Fax: þ386 1 4768 279. E-mail: [email protected]
Computer Aided Surgery, 2004; 9(4): 137–144
ISSN 1092-9088 print=ISSN 1097-0150 online #2004 Taylor & FrancisDOI: 10.1080=10929080500097687
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
2D images was established by means of fiducial
markers. The accuracy of “gold standard” regis-
tration was assessed by determining the target regis-
tration error (TRE) [7].
Methods and results
Phantom creation
The cadaveric lumbar spine, comprising vertebrae
L1-L5 of an 80-year-old female complete with inter-
vertebral disks and several millimeters of soft tissue,
was placed in a plastic tube and tied with thin nylon
strings (Figure 1, left). The tube was filled with
water to simulate soft tissue and thus enable more
realistic MR, CT and X-ray images to be obtained.
Six fiducial markers (Stryker Leibinger, Freiburg,
Germany) were rigidly attached to the outside of
the plastic tube. Each fiducial marker had two com-
ponents: a base that could be screwed to a rigid
body and a replaceable marker. Different markers
were used for the different imaging modalities:
Markers containing a metal ball (1.5 mm in dia-
meter) were used for CT and X-ray imaging, while
markers with a spherical cavity (2 mm in diameter)
filled with a water solution of Dotarem contrast
agent (Gothia, Sweden) were used for MR.
Image acquisition
The CT image (Figure 1, top center) was obtained
using a General Electric HiSpeed CT/i scanner.
Axial slices were taken with intra-slice resolution of
0.27 � 0.27 mm and inter-slice distance of 1 mm.
For MR imaging, a Philips Gyroscan NT Intera 1.5
T scanner and T1 protocol (flip angle 908,
TR ¼ 3220 ms, TE ¼ 11 ms) was used (Figure 1,
top right). Axial slices were obtained with intra-slice
resolution of 0.39 � 0.39 mm and inter-slice dis-
tance of 1.9 mm. The acquired MR image was retro-
spectively corrected for intensity inhomogeneity
using the information minimization method [8]. X-
ray images (Figure 1) were captured by a Pixium
4600 digital X-ray detector (Trixell, Moirans,
France). The detector had a 429 � 429 mm active
surface, with pixel size of 0.143 � 0.143 mm and
14-bit dynamic range. To simulate C-arm acqui-
sition, the X-ray source and sensor plane were fixed
while the spine phantom was rotated on a turntable
(Figure 2). In this way, mechanical distortion due
to gravitational force and other mechanical imperfec-
tions of the C-arms was avoided. By rotating the
spine phantom around its long axis in 208 steps, 18
X-ray images were acquired. These images were
filtered by a 3 � 3 median filter and then sub-
sampled by a factor of two to remove dead pixel
artifacts.
Finding the centers of fiducial markers
Figure 3 shows close-up views of fiducial markers in
CT, MR and X-ray images. In all 3D and 2D
images, a rough position pm of each fiducial marker
was first defined manually. Next, an intensity
threshold IT that separated the marker from the
surrounding tissues was selected for each marker.
Finally, the center pc of each marker was defined as
pc ¼
Pp[V (I(p)� IT )pPp[V (I(p)� IT )
(1)
Figure 1. The spine fastened in a plastic tube (top left); the final phantom with fiducial markers attached to the outside of the tube (bottom
left); CT image (top center); MR image (top right); and AP (bottom center) and lateral (bottom right) X-ray images of the phantom.
138 D. Tomazevic et al.
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
where I(p) is the intensity at point p and V is a small
neighborhood around point pm. By this method,
centers of markers may be found with sub-pixel or
sub-voxel accuracy [9,10].
Let XCT and XMR be 3 � 6 matrices, each contain-
ing six 3D vectors representing the centers of fiducial
markers found in CT and MR, respectively:
XCT ¼ ½rCT1 , rCT
2 , . . . , rCT6 �
XMR ¼ ½rMR1 , rMR
2 , . . . , rMR6 �
(2)
where r ¼ (x, y, z)T. Similarly, let Xw be a 2 � 6
matrix containing six 2D vectors representing the
centers of markers found in X-ray images obtained
after rotating the phantom through w degrees
(w ¼ 08, 208, . . . , 3408):
Xw ¼ ½pw1, p
w2, . . . , p
w6 � (3)
where p ¼ (x, y)T.
Retrospective calibration of the X-ray setup
To be able to reconstruct the 3D positions of fiducial
markers from their positions in 2D X-ray images, the
acquisition setup (Figure 2) was retrospectively
calibrated. Calibration required the determination
Figure 3. Close-up views of fiducial markers in CT (left), MR (center) and X-ray (right) images.
Figure 2. X-ray image acquisition.
“Gold standard” data for 3D/2D registration 139
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
of 12 geometrical parameters – 3 intrinsic wI and 9
extrinsic wE – denoted by calibration parameter
vector w, w ¼ (wI, wE). The intrinsic parameters
wI ¼ (xs, ys, zs) define the position of the X-ray
source rs in the coordinate systems Ss of the sensor
plane, and therefore define the projection PS(wI) of
any 3D point described in the sensor coordinate
system Ss to the 2D sensor plane. The extrinsic para-
meters wE describe the geometric relation between
the rotating phantom and the X-ray system. Four
parameters define the axis of rotation in the coordi-
nate system Sv of the phantom. We have chosen the
coordinate system of the CT volume for Sv. The
axis of rotation is defined by the point (txv, tyv),
which is the intersection of the axis with the x-y coor-
dinate plane of Sv, and by rotation (vxv, vyv) of the
axis around the x and y of Sv. Similarly, four para-
meters (txs, tys) and (vxs, vys) define the same axis
of rotation in the coordinate system Ss of the X-ray
sensor plane. The last parameter, necessary to deter-
mine the relation between Ss and Sv on the rotation
axis, is distance dvs between the two points of inter-
section (txv, tyv) and (txs, tys). The extrinsic para-
meters wE ¼ (txv, tyv, vxv, vyv, dvs, txs, tys, vxs,
vys)T thus define the transformation TVS(w, wE):
TVS(w, wE) ¼ TRS(txs, tys, vxs, vys)
�T(dvs) �R(w)
�TVR(txv, tyv, vxv, vyv) (4)
This transformation maps, for a given rotation w of
the phantom, any 3D point in coordinate system Sv
to a 3D point in coordinate system Ss (Figure 2).
TVR is the transformation from coordinate system
Sv to the axis of rotation; R(w) is the rotation
around the rotation axis; T(dvs) is the translation
along the rotation axis; and TRS is the final transform-
ation to the coordinate system Ss. For any rotation w,
the projection PVS(w, w) of a 3D point defined in the
coordinate system Sv to the 2D point lying in the
sensor plane defined by Ss is obtained by applying
the projection PS(wI) and the transformation
TVS(w, wE)
PVS(w, w) ¼ PS(wI)TVS(w, wE) (5)
To retrospectively calibrate the X-ray acquisition
system, we thus need to define 12 geometrical
parameters w of the projection PVS(w, w). For this
purpose we have used centers Xw of fiducial
markers found in the X-ray images and the corre-
sponding centers XCT of markers found in the CT
volume. The optimal calibration parameters w are
the ones that bring the fiducial markers XCT in the
CT volume to the best correspondence with the cor-
responding fiducial markers Xw in the X-ray images.
To find the optimal parameters, we projected the
centers of the fiducial markers XCT in the CT
volume to the sensor plane and computed the root
mean squared (RMS) distance Ecalib to the corre-
sponding centers of fiducial markers Xw in the
X-ray images:
Ecalib(w) ¼
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1
M
Xw[F
1
N
XNi¼1
(pwi � PVS(w, w)rCT
i )2
vuut(6)
where N and M stand for the number of fiducial
markers and X-ray images, respectively, and F ¼
fw1, w2, . . . , wMg defines the X-ray images taken at
different phantom rotations. To find the optimal
calibration parameters w, we used nine X-ray
images F ¼ f08, 408, . . . , 3208g and iterative optimiz-
ation, which resulted in a minimum RMS distance
(Ecalib) of 0.31 mm. The small RMS indicates that
calibration was performed well and reflects the
uncertainty of fiducial marker localization in CT
and X-ray images.
Reconstruction of 3D markers
Once the X-ray acquisition system had been cali-
brated, the 3D positions of X-ray fiducial markers
could be reconstructed from their positions in 2D
X-ray images. The least-squares solution is used to
obtain the 3D marker position [11], as described
below. Each point piw, representing the center of the
ith fiducial marker in an X-ray image taken at
rotation w of the phantom, was back-projected to
the X-ray source rs, which yielded the imaginary
line Liw (Figure 4). Line Li
w, which defines the per-
spective projection of a 3D marker to the 2D X-ray
plane, can be expressed in the coordinate system Sv
of the phantom by mapping the X-ray source rs to
point cw:
cw ¼ T�1VS(w, wE)rs (7)
and by expressing the line direction in Sv coordinates
as
vwi ¼
T�1VS(w, wE)(p
wi � rs)
T�1VS(w, wE)(p
wi � rs)
�� �� (8)
where rs and piw are points defined in the sensor coor-
dinate system Ss. A marker’s 3D position riR in the
coordinate system Sv was reconstructed by
140 D. Tomazevic et al.
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
minimizing the RMS distance Erec from point riR to all
lines Liw:
Erec(rRi ) ¼
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1
M
Xw[F
(rRi � cw)� v
wi
�� ��2s(9)
Reconstruction of the 3D positions of six fiducial
markers from the nine X-ray images F ¼ f208,608, . . . , 3408) which were not used for calibration
yielded RMS values of less than 0.06 mm for each
of the six markers. The positions of reconstructed
fiducial markers were incorporated in the 3 � 6
matrix XR ¼ [r1R, r2
R, . . . , r6R].
By using different sets of X-ray images for recon-
struction and calibration, we were able to validate
the calibration procedure. Small RMS values of
0.06 mm indicated that the uncertainty of fiducial
marker localization in X-ray images was less than
that in CT images and that calibration had been
performed well. Therefore, the major source of cali-
bration uncertainty is the uncertainty of fiducial
marker localization in CT images, though its effect
on calibration precision is obviously very small.
“Gold standard” registration and validation
After calibrating the X-ray acquisition system and
reconstructing 3D markers XR from X-ray images,
we were able to establish “gold standard” registration
between X-ray and CT images and between X-ray
and MR images in the coordinate system Sv of the
phantom. This was achieved by a rigid 3D/3D trans-
formation T that minimized the RMS distance Ereg
between the reconstructed fiducial markers XR from
X-ray images and the marker points XCT or XMR
from CT and MR images, respectively:
Ereg(T) ¼
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi1
N
XNi¼1
rRi � Tri
� �2vuut (10)
where ri stands for either riCT or ri
MR. The closed-
form solution of this minimal RMS problem is
known [12,13]. Rigid transformation T can be
decomposed to the rotation component R, rep-
resented by a 3 � 3 matrix, and translation vector t:
Tr ¼ Rrþ t (11)
The optimal solution for the translation com-
ponent is given as:
t ¼ �rR �R�r (12)
where �rR and �r stand for the mean positions of point
sets XR and X, respectively, where set X is either XCT
or XMR. The optimal solution for the rotation com-
ponent is given as
R ¼ BAT (13)
A and B are two orthogonal matrices obtained by
singular value decomposition (SVD) of the matrix
�XR�X
T¼ ADBT (14)
where D is a diagonal matrix and �XR and �X are the
point sets XR and X, centered at corresponding
mean positions �rR and �r, respectively.
Rigid registration of point sets (XCT, XR) and
(XMR, XR) resulted in a minimum RMS distance
Ereg of 0.27 mm for CT and 0.44 mm for MR-to-
X-ray registration. The higher RMS for MR than
for CT can be attributed to three reasons: First,
CT was used in calibration; second, intra- and
inter-slice resolutions of the MR image were lower
than those of the CT image, resulting in higher
fiducial localization uncertainty; and the third, MR
image suffered from non-rigid spatial distortion.
The minimum RMS distance Ereg is also known as
the fiducial registration error (FRE) and can be used
to evaluate the accuracy of point-based rigid regis-
tration [7]. By knowing the FRE, we can determine
the target registration error (TRE), which is the
distance between the true, but unknown, position
of the target and the target position obtained by
registration. The expected TRE of a target point r
Figure 4. Reconstruction of 3D marker position. Due to the
uncertainty of X-ray marker localization and X-ray setup cali-
bration, the projection lines do not cross at the same point.
“Gold standard” data for 3D/2D registration 141
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
can be estimated from FRE [7]:
kTRE2(r)l ¼kFLE2l
N1þ
X3
k¼1
dk
fk
!(15)
where fk is the RMS of the projections of the fiducial
markers to the kth principal axis of marker configur-
ation; dk is the projection of target point r to principle
axis k; N is the number of fiducial markers; and FLE
is the fiducial localization error obtained from the
FRE:
FLE2 ¼N
N � 2FRE2 (16)
Using the above formulation, we validated the
“gold standard” registration by manually defining
eight target points (four per pedicle) on each of the
five vertebrae and computing the mean TRE for
each vertebra. The results of “gold standard” vali-
dation for CT-to-X-ray and MR-to-X-ray regis-
trations are given in Table I. Expected TREs for the
pedicles are less than 0.26 mm for CT-to-X-ray regis-
tration and less than 0.42 mm for MR-to-X-ray
registration.
Publicly available “gold standard” data
The “gold standard” data and detailed information
on how to use it are publicly available from the
Department of Electrical Engineering, University of
Ljubljana, Laboratory of Imaging Technologies web
site (http://lit.fe.uni-lj.si/Downloads/downloads.asp).
The image database consists of 18 X-ray images and
5 CT and 5 MR sub-volumes (Figure 5). In the CT
and MR images, cubic sub-volumes were manually
defined, each containing a single vertebra, approxi-
mately one third of the neighboring vertebrae, and
no fiducial markers. In each sub-volume the areas
occupied by air and the plastic tube were manually
masked and their intensities were replaced by the
average intensity corresponding to water. The edges
between water, the plastic tube, and air were thus
almost eliminated. In this way, any rigid 3D/2D regis-
tration method evaluated using our “gold standard”
data cannot take advantage of markers or edges that
would not be present in actual clinical data. The
markers in X-ray images, which were not excluded
or “airbrushed”, may be considered as outliers,
similar to the way that surgical tools would be in a
clinical setting.
For each CT or MR sub-volume, the “gold stan-
dard” registration position, the coordinates of eight
target points on the pedicles, and 450 randomly
chosen starting positions around the “gold standard”
registration position are also provided in the data-
base. Details of the generation of starting positions
have been previously published in reference 14. Fur-
thermore, the data on our website contain the Matlab
(MathWorks, Natick, MA) source code for analyzing
registration results. The code allows calculation of
TREs and estimation of capturing ranges from the
results of 450 registrations per modality and vertebra.
Figure 5. Transverse and sagittal views of CT (left) and MR (center) sub-volumes, and a lateral X-ray image (right).
Table I. Expected RMS TREs for “gold standard” registration
(in mm).
Vertebrae
L1 L2 L3 L4 L5
CT 0.2 0.15 0.15 0.19 0.26
MR 0.33 0.24 0.24 0.31 0.42
142 D. Tomazevic et al.
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
If different registration methods are to be objectively
compared, it is not only important that the same
image data sets be used, but equally important that the
evaluation protocol, criteria for successful registra-
tion, and error metrics be the same. Otherwise, it will
be difficult to place the results of a novel registration
method in the context of previously published work.
Discussion and conclusion
Before an image-guided therapy (IGT) system is put
into clinical use, its individual components must
undergo rigorous validation. Prerequisites for vali-
dation of 3D/2D image registration, which is the
crucial component of an IGT system, are standardi-
zation of validation methodology (including design of
validation data sets), definition of the corresponding
“gold standard” and its accuracy, a validation proto-
col, and design of validation metrics [14]. A fair com-
parison of different registration techniques is possible
only to a limited degree if standard validation metho-
dology is not publicly available. Motivated by the lack
of publicly available “gold standard” data for evalu-
ation and comparison of different 3D/2D rigid regis-
tration methods, we have devised a lumbar spine
phantom, obtained X-ray-to-CT and X-ray-to-MR
“gold standard” registrations, and established the
accuracy of these registrations. We are aware that
there are surgical interventions and radiosurgical
treatments of the spine for which markers are routi-
nely inserted into the spine of the patient. Marker-
based registered CT, MR and X-ray images of the
spine can serve as a truly clinical “gold standard”,
but are very difficult to obtain. Russakoff et al. [15]
reported on the evaluation of intensity-based 3D/
2D spine image registration using clinical “gold stan-
dard data”. Unfortunately, to the best of our know-
ledge, their data are not publicly available. One
problem with such a “gold standard” is the huge
effort required to remove all traces of the implanted
markers from images to avoid bias, i.e., to ensure
that the evaluated registration method could not
take advantage of markers that would not usually be
present in images acquired for image-guided therapy.
Another approach to obtaining more realistic “gold
standard” data is to overlay features segmented from
clinical images on phantom images [5]. Soft tissue,
surgical instruments and other structures can be
overlaid. Although our “gold standard” is not
equivalent to the clinical “gold standard”, and no
structures have been overlaid on our phantom
images, we believe it is sufficiently realistic to allow
fair evaluation and comparison of registration
methods for image-guided therapy. Our database
also contains MR images, and we believe this to be
very valuable.
The X-ray acquisition system was calibrated retro-
spectively by matching the projections of CT markers
with corresponding markers in X-ray images. Cali-
bration with CT markers is generally superior to cali-
bration with MR markers because CT offers better
resolution and spatial stability. This observation
was confirmed experimentally: CT-based calibration
yielded a smaller calibration error Ecalib of 0.31 mm,
compared with 0.47 mm when calibrating the X-ray
system with MR data. CT-based calibration of the
X-ray image acquisition setup already provides regis-
tration of CT to X-ray images, but does not provide
any indices of registration accuracy. Therefore, we
have reconstructed the 3D positions of markers
from calibrated 2D X-ray images, allowing us to
implement 3D/3D registration between the recon-
structed markers and those found in CT and MR
volumes. The result of such a registration reflected
a) the uncertainty of marker localization in 2D X-ray
images; b) the uncertainty of marker localization in
3D CT or MR images; c) the uncertainty of the
X-ray acquisition calibration; and d) the uncertainty
of marker reconstruction. The FRE of 3D/3D
registration, which reflected all these uncertainties,
was used to evaluate the TRE of the “gold standard”
CT-to-X-ray and MR-to-X-ray registrations accord-
ing to the theory developed in reference 7.
The results presented in Table I indicate that the
“gold standard” registration is highly accurate and
therefore useful for testing 3D/2D registration
methods. However, it should be stressed that the
expected TREs for CT-to-X-ray “gold standard”
registration may possibly be a little larger than
those in Table I. This is because the same CT
markers were used for both X-ray system calibration
and CT-to-X-ray registration, which could have
involved the same bias in calibration and registration.
By acquiring a second CT scan at the time of image
acquisition and using one for calibration and the
other for validation bias could be eliminated. Never-
theless, if we assume that localization errors for CT
markers are much smaller than those for MR
markers, the expected TREs for CT-to-X-ray “gold
standard” registration should be close to those
shown in Table I and certainly not larger than the
TRE values for MR-to-X-ray registration.
We believe that, due to the lack of publicly available
“gold standards” for 3D/2D rigid registration, the
presented “gold standard” data will prove useful for
the evaluation of newly developed methods and the
comparison of existing registration methods.
Acknowledgments
The authors would like to thank L. Desbat,
M. Fleute, R. Martin, F. Esteve and U. Vovk for their
“Gold standard” data for 3D/2D registration 143
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.
generous help and support in the image acquisitions.
This work was supported by the IST-1999-12338
project, funded by the European Commission and
by the Ministry of Education, Science and Sport,
Republic of Slovenia, under grant P2-028.
References
1. Galloway RL. The process and development of image-guided
procedures. Ann Rev Biomed Eng 2001;3:83–108.
2. Lavallee S, Szeliski R. Recovering the position and orientation
of free-form objects from image contours using 3D distance
maps. IEEE Trans Pattern Anal Machine Intell 1995;
17:378–90.
3. Gueziec A, Kazanzides P, Williamson B, Taylor RH. Anatomy-
based registration of CT-scan and intraoperative X-ray images
for guiding a surgical robot. IEEE Trans Med Imag 1998;
17:715–28.
4. Lemieux L, Jagoe R, Fish DR, Kitchen ND, Thomas DGT. A
patient-to-computed-tomography image registration method
based on digitally reconstructed radiographs. Med Phys
1994;21:1749–60.
5. Penney GP, Weese J, Little JA, Desmedt P, Hill DLG,
Hawkes DJ. A comparison of similarity measures for use in
2-D-3-D medical image registration. IEEE Trans Med Imag
1998; 17:586–95.
6. Jannin P, Fitzpatrick JM, Hawkes DJ, Pennec X, Shahidi R,
Vannier MW. White paper: validation of medical image
processing in image-guided therapy. In: Lemke HU, Vannier
MW, Inamura K, Farman AG, Doi K, Reiber JHC, editors.
Computer Assisted Radiology and Surgery. Proceedings of
the 16th International Congress and Exhibition (CARS
2002), Paris, France, June 2002. Berlin: Springer, 2002. p
299–305.
7. Fitzpatrick JM, West JB, Maurer CR Jr. Predicting error in
rigid-body point-based registration. IEEE Trans Med Imag
1998;17:694–702.
8. Likar B, Viergever MA, Pernus F. Retrospective correction of
MR intensity inhomogeneity by information minimization.
IEEE Trans Med Imag 2001;20:1398–1410.
9. Bose CB, Amir I. Design of fiducials for accurate registration
using machine vision. IEEE Trans Pattern Anal Machine
Intell 1990;12:1196–1200.
10. Chiorboli G, Vecchi GP. Comments on design of fiducials for
accurate registration using machine vision. IEEE Trans
Pattern Anal Machine Intell 1993;15:1330–2.
11. Siddon RL, Chin LM. 2-film brachytherapy reconstruction
algorithm. Med Phys 1985;12(1):77–83.
12. Arun KS, Huang TS, Blostein SD. Least-squares fitting of two
3-D point sets. IEEE Trans Pattern Anal Machine Intell
1987;9(5):698–700.
13. Umeyama S. Least-squares estimation of transformation
parameters between two point patterns. IEEE Trans Pattern
Anal Machine Intell 1991;13(4):376–80.
14. Tomazevic D, Likar B, Slivnik T, Pernus F. 3-D/2-D regis-
tration of CT and MR to X-ray images. IEEE Trans Med
Imaging 2003;22(11):1407–16.
15. Russakoff DB, Rohlfing T, Ho A, Kim DH, Shahidi R,
Adler JR, Maurer CR. Evaluation of intensity-based 2D-3D
spine image registration using clinical gold-standard data. In:
Gee JC, Maintz JBA, Vannier MW, editors. Proceedings
of Second International Workshop on Biomedical Image
Registration (WBIR 2003), Philadelphia, PA, June 2003.
Lecture Notes in Computer Science 2717. Berlin: Springer,
2003. p 151–160.
144 D. Tomazevic et al.
Com
pute
r A
ided
Sur
gery
Dow
nloa
ded
from
info
rmah
ealth
care
.com
by
Lau
rent
ian
Uni
vers
ity o
n 10
/02/
13Fo
r pe
rson
al u
se o
nly.