FX-class - High Flux for Improved Survival

Embed Size (px)

Citation preview

  • 7/23/2019 FX-class - High Flux for Improved Survival

    1/12

    FX-class

    High-Flux Dialysis for Improved Survival

    Haemodialysis

  • 7/23/2019 FX-class - High Flux for Improved Survival

    2/12

    No Copy can Match the Original

    2

    a Fresenius Polysulfone

  • 7/23/2019 FX-class - High Flux for Improved Survival

    3/12

    High-Flux Membranes Improve Patient Outcome

    3

    The benefits of High-Flux membranes the essence of the MPO Study:

    A 37 % reduction of the relative risk of death was observed

    for patients having a serum albumin level 4.0 g/dL.

    Significantly improved survival for diabetes patients.

    The Membrane Permeability Outcome (MPO) Study, a

    clinical trial specifically designed to resolve the effect

    of High- and Low-Flux membranes on patient outcomes,

    has recently been completed.

    The results, officially presented by Prof. Locatel li,

    principal investigator of the study group, at the XLIV

    Congress of the ERA-EDTA, show a survival benefit

    for haemodialysis patients treated with High-Flux

    compared to Low-Flux membranes.*

    The MPO Study finally provides strong evidence in

    favour of using biocompatible High-Flux membranes

    to improve the long-term outcome of patients with

    end-stage renal disease.

    The MPO Study is supported by other studies(3, 4, 5, 6)

    showing a beneficial effect of High-Flux membranes

    for dialysis patients like the recently published data

    by Krane et al. (2007) from the database of the

    German Diabetes and Dialysis (4D) Study (7):

    A beneficial effect of biocompatible High-Flux

    membranes on patients outcome compared to

    cellulosic membranes and Low-Flux membranes

    was shown for patients with type 2 diabetes.41: Evaluation of outcome in patients with type 2 diabetes on maintenancehemodialysis treatment influenced by biocompatibility and flux characteristicsof the dialysis membrane (post hoc analysis of a randomised, double-blindmulticentre study). Data taken from Krane et al., 2007

    As between 56 % and 86 % of dialysis patients

    worldwide have a serum albumin level < 4.0 g/dL,

    the majority of patients on dialysis would benefit from

    High-Flux dialysis (2).

    * Inclusion criteria (1):

    Incident (on HD for 2 months)

    Age 18 80 years

    At risk (serum albumin 4.0 g/dL)

    During the study, additional enrolment of patients with serum albumin > 4.0 g/dL was allowed.

  • 7/23/2019 FX-class - High Flux for Improved Survival

    4/12

    FX-classDialyser Design

    Several state-of-the-art technologies have been

    combined to create the distinctive functional features

    of the FX-classdialysers:

    The fibre bundle geometry, the membrane nanostruc-

    ture, the flow port and the housing design all provide

    advantages in terms of performance, haemodynamics,

    dialysate flow as well as safety and handling.

    Optimised fibre array

    The higher packing density of the fibre bundle

    together with the special wavy fibre structure regulates

    a homogeneous distribution of dialysate over the

    whole cross-section of the dialyser. This is evident in

    the superior clearance values of the FX-class (8).

    4

    Refined haemodynamics

    The lateral blood-inlet port defines a homogeneous

    blood flow-path, avoiding low velocity stagnation

    zones in the header region. Furthermore, the risk

    of accidental twisting of bloodlines is virtually

    eliminated.

  • 7/23/2019 FX-class - High Flux for Improved Survival

    5/12

    The membrane

    Helixone is the advanced High-Flux polysulfone

    membrane of the FX-class dialysers. Helixone has

    been designed specifically to meet the present-day

    demands of High-Flux dialysis and convective therapies:

    Larger average pore size (3.3 nm)

    More even distribution of pores

    High membrane porosity for enhanced hydraulic

    permeability

    The result of these structural refinements is the smooth

    and unrestricted transport of larger uremic toxins across

    the membrane wall, as exemplified by the significantly

    increased sieving coefficient for larger solutes (e. g.

    SC2m

    = 0.8) but with minimal leakage of useful

    proteins like albumin (SCalbumin

    0.001).

    Radial dialysate flow

    The pinnacle structures at both ends of the polypro-

    pylene housing together with the potting technology

    ensure an even, radial flow of the dialysate around the

    individual fibres of the bundle.

    Fibres designed for High-Flux HD

    The reduced inner diameter and wall thickness of the

    fibre increase the internal filtration and minimise the

    diffusion resistance. A significant increase of both

    the diffusive and convective clearances is therefore

    achieved, allowing the efficient removal of a broad

    spectrum of uremic toxins.

    5

  • 7/23/2019 FX-class - High Flux for Improved Survival

    6/12

    steam

    steam

    steam

    sterile air

    sterile water

    The FX-class dialysers are sterilised by the unique

    INLINE steam sterilisation process specifically

    developed by Fresenius Medical Care.

    During the INLINE steam sterilisation process, both

    blood and dialysate compartment of the dialysers are

    rinsed continuously for 15 minutes with steam at a

    temperature of 121 C. 42

    This extensive rinsing of FX-classdialysers with hot

    steam and without chemicals results in extremely lowlevels of residuals.

    Superior Production Process InvolvingINLINE Steam Sterilisation

    Every dialyser is then tested for fibre integrity. Fresenius

    Medical Care carries out its 100 % fibre leak testing

    procedures using a bubble-point test: Air pressure

    is applied to the fibre bundle from one side while

    the other side contains sterile water. If there were

    leakages in the membrane, air would pass the

    membrane and create bubbles, which are then

    detected by automated camera systems. 43

    The dialysers failing the integrity test are discarded.

    Finally, the dialysers are dried with warm, sterile air.

    6

    After INLINE steam sterilisation the dry dialysers

    get labeled, are visually inspected and finally fully

    automatically packed.

    During every production step, all Fresenius Medical

    Care dialysers are undergoing various automated

    in-process controls.

    steam

    sterile water

    sterile air

    42 43

  • 7/23/2019 FX-class - High Flux for Improved Survival

    7/12

    The INLINE steam sterilisation process leads to:

    Highly purified dialysers

    Dialysers free of any toxic chemicals or sterilisation

    by-products

    Low rinsing volumes

    In contrast to INLINE steam sterilisation, gamma-

    irradiation may physically and/or chemically alter the

    membrane as high-energy radiation produces ionisation

    and excitation in polymer molecules such as polysulfone.

    This process may result in physical or chemical cross-

    linking or degradation of the material and cytotoxic

    substances may be generated (9, 10).

    Among others, it has been shown that 4,4'-methylen-

    dianilin, a substance of known carcinogenic potential,

    may be generated in the polyurethane potting material

    of capillary dialysers during gamma-radiation (11).

    Furthermore, chemically active or pyrogenic substances

    and residuals from sterilisation or production may

    remain within the fibre. Intensive priming and rinsing

    procedures are needed with irradiated filters.

    7

    A recent study carried out by the Fraunhofer Institute,

    Germany, shows the effects of test extracts obtained

    from dialysers after undergoing different sterilisation

    procedures on the viability of cells in culture:

    Samples from different irradiated dialysers inhibited

    metabolic activity (determined with a cell proliferation

    assay) of cells by 70 % to 97 %. The samples of

    INLINE steam sterilised FX80 dialysers showed only

    a negligible influence.

    DNA synthesis was determined after incorporation of

    the base analogue BrdU (5-bromo-2'-deoxyuridine):

    The extracts from FME dialysers affected the cells

    only in a non-significant manner, whereas irradiated

    dialysers contain cytotoxic residuals killing a majority

    of the cells. 44

    Therefore, highly intensive rinsing is recommended

    before use of irradiated dialysers.

    Advantages of INLINE Steam Sterilised Dialysers

    Control

    INLINE steam sterilisation(Fresenius Medical Care)

    Gamma-ray sterilisation(different manufacturers)

    DNASyn

    thesisRate(%)

    Extracts from Dialysers

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    44: In vitro cytotoxicity testing of eight dialysers acc. to ISO 10993:Effect of the samples (test extracts) from the dialysers on DNA synthesis(BrdU-Test) on L929 cells. (Fraunhofer Institute, St. Ingbert, Germany;unpublished data)

  • 7/23/2019 FX-class - High Flux for Improved Survival

    8/12

    Advantages of FX-classHigh-Flux Dialysers

    Highly purified dialysers sterile and pyrogen-free

    Excellent haemobiocompatibility, unaffected by

    sterilisation

    Dry packed, light-weight products

    Dialysers without pore-fillers or sterilising agent

    residues

    Safe and comfortable treatment for your patients

    Environmentally friendly sterilisation method

    All production steps from the manufacturing of the

    membrane to the finished dialyser are adjusted to

    each other resulting in constant highest quality.

    The FX-classof dialysers is like all other products

    from Fresenius Medical Care produced with quality

    foremost in mind. Production and quality control

    systems are ISO 9001 and EN 45001 certified;

    the product specifications are also determined

    and controlled according to the acknowledged EN

    standards.

    There has been an increasing interest in the

    development of more efficient haemodialysis

    treatment modalities in recent years. The main

    objective of these efforts has been primarily to remove

    a wide range of uremic retention solutes particularly

    the middle molecules in the most efficient way (12).

    The FX-class possesses outstanding clearances

    for both low-molecular weight solutes but also for

    larger uremic toxins.

    An extended clinical experience worldwide has

    established that the efficient removal of a wide range

    of toxins by High-Flux is a significant contributing

    factor for improved long-term results for dialysis

    patients (EBPG 2.2)(13), for example in terms of

    better control of renal anemia. (14)

    delayed onset of amyloidosis. (15, 16)

    reduced inflammation. (14)

    improved immune response. (17)

    prolonged preservation of residual renal funtion. (18)

    8

  • 7/23/2019 FX-class - High Flux for Improved Survival

    9/129

    0 100 200 300 400 500 600

    0.6

    1.0

    1.4

    1.8

    2.2

    In-VitroPerformance Data

    To utilise a dialyser to its full capacity and achieve

    optimal blood flow conditions in the dialyser, it is

    important to consider the relationship between its

    effective surface area and the achievable blood

    flow rate.

    At low blood flow rates, large dialyser surface areas

    are not exploited to their full extent.

    Optimal application for the FX-classdialysers

    Ultrafiltration coefficient (mL/hxmmHg) 20 33 46 59 73Clearances: Q

    B200

    (mL/min) Urea 170 189 193 197 *

    Creatinine 144 170 182 189

    Phosphate 138 165 177 185

    Vitamin B12

    84 115 135 148

    Inulin 54 76 95 112

    Clearances: QB300

    (mL/min) Urea * 250 261 276 278

    Creatinine 210 230 250 261

    Phosphate 201 220 239 248

    Vitamin B12

    130 155 175 192

    Inulin 81 104 125 142

    Clearances: QB400

    (mL/min) Urea * * 303 326 331

    Creatinine 262 287 304

    Phosphate 248 272 284

    Vitamin B12

    167 190 213

    Inulin 109 133 152

    * Refer to recommended blood flow range.

    The in-vitroperformance data were obtained with QD= 500 mL/min, Q

    F= 0 mL/min and T = 37 C (EN 1283, ISO 8637).

    The ultrafiltration coefficients were measured using human blood, Hct 32 %, protein content 6 %.

    Sieving coefficient QB= 300 mL/min, Q

    F= 60 mL/min Inulin 1

    2-microglobulin 0.8

    Albumin 0.001

    Effective surface area (m2) 0.6 1.0 1.4 1.8 2.2

    KOAurea

    489 824 977 1292 1351

    Wall thickness/inner diameter (m) 35/185

    Blood filling volume (mL) 32 53 74 95 116

    Membrane

    Housing material Polypropylene

    Potting compound Polyurethane

    Sterilisation method INLINE steam

    Treatment mode HD/HDF/HF

    Article number 5008841 5008851 5008861 5008881 5008901

    In-vitroperformance data

    FX 40 FX 50 FX 60 FX 80 FX 100

    FX 100

    FX 80

    FX 60

    FX 50

    FX 40

    SurfaceArea(m2)

    Blood Flow (mL /min)

  • 7/23/2019 FX-class - High Flux for Improved Survival

    10/1210

    Ever since the conception of haemodialysis therapy,

    nephrologists worldwide have addressed the factors

    contributing towards the poor long-term outcome of

    dialysis patients.

    The task has been complicated by the fact that

    many dialysis patients suffer from multiple co-morbid

    conditions, particularly hypertension, diabetes and

    malnutrition all of which contribute to the high

    incidence of cardiovascular disease in this population

    group.

    The central function of haemodialysis therapies is to

    remove a broad range of uremic toxins efficiently

    like the natural kidney.

    Over the last years there has been an increasing body

    of evidence pointing towards a reduced death risk

    in patients undergoing High-Flux dialysis as well

    as advanced treatments such as haemodiafiltration

    (HDF) (13).

    Both High-Flux dialysis and HDF require dialysis

    membranes that are highly permeable to large toxins

    and water. In addition, such membranes must be

    biocompatible and have a high endotoxin retention

    capacity.

    Fresenius Polysulfone and Helixone membranes

    the most widely used dialysis membranes worldwide

    have specially been designed to fulfil these essential

    criteria.

    Together with nephrologists and nurses, Fresenius

    Medical Care continues to contribute towards the

    improved quality of life of HD patients, and ultimately

    their survival.

  • 7/23/2019 FX-class - High Flux for Improved Survival

    11/1211

    The positive results of the MPO Study validate our

    efforts to offer innovative dialysis products such as

    our High-Flux dialysers with Helixone membranes

    so that dialysis patients can look toward the future with

    more confidence. And we are proud that the majority

    of the patients in the studys High-Flux group were

    treated with our dialysers, said Dr. Emanuele Gatti,

    Fresenius Medical Cares Chief Executive Officer for

    Europe, Latin America, Middle East and Africa.

    High-Flux Dialysis Powered by Fresenius Helixone

    Literature

    1. Locatelli F et al., The effect of membrane permeability on

    ESRD: design of a prospective randomised multicentre trial.

    J Nephrol (1999); 12(2): 85-8.

    2. Pisoni RL et al., Anemia management and outcomes from 12

    countries in the Dialysis Outcomes and Practice Patterns

    Study (DOPPS). Am J Kidney Dis. (2004); 44: 94-111.

    3. Hornberger JC et al., A multivariate analysis of mortality and

    hospital admissions with High-Flux dialysis. J Am Soc Nephrol

    (1992); 3: 1227-1237.

    4. Woods HF and Nandakumar M, Improved outcome for haemo-

    dialysis patients treated with High-Flux membranes. Nephrol

    Dial Transplant (2000); 15 (S1): 36-42.

    5. Chauveau P et al., Dialyzer membrane permeability and survival

    in hemodialysis patients. Am J Kidney Dis (2005); 45: 565-571.

    6. Delmez JA et al., Cerebrovascular disease in maintenance hemo-

    dialysis patients: results of the HEMO study. Am J Kidney Dis

    (2006); 47: 131-138.

    7. Krane V et al., Dialyzer membrane characteristics and outcome

    of patients with type 2 diabetes on maintenance hemodialysis.

    Am J Kidney Dis (2007); 49: 267-275.

    8. Wizemann V et al., Efficacy of haemodiafiltration. Nephrol Dial

    Transplant (2001);16 Suppl 4: 27-30.

    9. Hemmerich KJ, Polymer Materials Selection for Radiation-

    Sterilized Products, MDDI (2000); 2: 78-90.

    10. Takesawa S et al., Varying methods of sterilisation, and their

    effects on the structure and permeability of dialysis membranes.

    Trans Am Soc Artif Intern Organs (1987); 33: 584-587.

    11. Shintani H et al., Analysis of a carcinogen, 4,4'-methylene-

    dianiline, from thermosetting polyurethane during sterilization.

    J Anal Toxicol (1989); 13: 354-357.

    12. Vanholder R et al., Review on uremic toxins: classification,

    concentration and interindividual variability. Kidney Int (2003);

    63: 1934-1943.

    13. Tattersall, J et al., EBPG guideline on dialysis strategies.

    Nephrol Dial Transplant (2007); 22 Suppl 2: ii5-21.

    14. Merello Godino JI et al., Results from EuCliD (European Clinical

    Dialysis Database): impact of shifting treatment modality. Int

    J Artif Organs. (2002); 25(11): 1049-60.

    15. Koda Y et al., Switch from conventional to High-Flux membrane

    reduces the risk of carpal tunnel syndrome and mortality of

    hemodialysis patients. Kidney Int (1997); 52: 1096-1101

    16. Locatelli F et al., Comparison of mortality in ESRD patients

    on convective and diffusive extracorporeal treatments. Kidney

    Int (1999); 55: 286-293

    17. Lonnemann G et al., A switch to High-Flux helixone membranes

    reverses suppressed interferon-gamma production in patients

    on Low-Flux dialysis. Blood Purif. (2003); 21(3): 225-31.

    18. McKane, W et al., Identical decline of residual renal function

    in High-Flux biocompatible hemodialysis and CAPD. Kidney Int.

    (2002); 61(1): 256-65.

  • 7/23/2019 FX-class - High Flux for Improved Survival

    12/12

    7327011/4G

    B(2,0V&B04.08)

    Copyright2008FreseniusMedicalCareDeutschlandGmbH

    Fresenius Medical Care Deutschland GmbH 61346 Bad Homburg v. d. H. Germany Phone: +49 (0) 6172-609-0 Fax: +49 (0) 6172-609-2191

    Head office: Else-Krner-Strae 1 61352 Bad Homburg v. d. H.

    www fmc-ag com