19
The world leader in serving science Inge de Dobbeleer Fast Method Development using AutoSRM and autoSIM

Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

Embed Size (px)

Citation preview

Page 1: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

1 The world leader in serving science

Inge de Dobbeleer

Fast Method Development using AutoSRM and autoSIM

Page 2: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

4

RT: 0.00 - 20.00 SM: 7G

0 2 4 6 8 10 12 14 16 18Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

RT: 8.23SN: 35

15.36

14.31

15.52

14.09 16.6718.8517.04

12.42 17.90 19.0113.229.33

12.2610.648.13 8.42 9.756.365.375.214.69RT: 8.23SN: 506

Pentachlorobenzene Masstrace m/z

250 in fullscan acquisition

S/N 35

Pentachlorobenzene m/z 250 in SIM

acquisition S/N 506

Short refresher on SIM benefit

Pesticides in soils

Page 3: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

5

Fullscan TIC

Dichlorvos in SIM @ m/z185

Dichlorvos in SRM 185>109

Dichlobenil in SIM @ m/z171

Dichlobenil in SRM 171>136

Refresher on MS/MS benefit

Pesticides in herbs

Page 4: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

6

Method development using the classical set up with segments or windows

RT: 7.61 - 29.04

8 10 12 14 16 18 20 22 24 26 28Time (min)

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

105

110

115

Rel

ativ

e Ab

unda

nce

15.88

9.3413.8912.06 14.01

19.83

13.24 14.6111.63 17.07

12.8511.149.84 15.59 18.128.60 16.96 23.4010.6919.90

16.27 18.1817.87 20.82 23.17 25.0321.24 23.61

18.67 25.45

22.0819.61 26.9627.32

23.0027.86

Segment this…… 150 pesticides

Page 5: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

7

Method development using the classical set up with segments or windows

RT: 12.95 - 17.25

13.0 13.5 14.0 14.5 15.0 15.5 16.0 16.5 17.0Time (min)

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Relativ

e Abun

dance

15.88

13.89 14.01

13.24 14.6113.6517.07

14.2215.59

16.96

13.4515.07 16.27

16.6116.0014.83 15.3515.81

16.7315.25 15.44 16.44

14.49

14.45

16.1813.03

N1T M_

• In SIM we need 3 ions/comp or 126 ions/segment • In MS/MS we need two transitions or 82 ions/segment • Manual set up leads to errors • Matrix injections leads to RT shift, compounds fall out of the

window, so often methods are made with compounds in two segments

• Many ions/ segment= lower dwell times= lower S/N

42 compounds in a segment

Page 6: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

8

Modern method development for single and triple quads

Fullscan

• Library search • List of compounds and RT

develop

• autoSIM or • autoSRM

GO

• Link to software • Acquire and quantify • Update method

Page 7: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

9

AutoSIM

GC and AS method

Chromatography OK?

no

Yes. Automation can start

FS run + lib search Compound list

Or… Start automation here

Click on the compound; A mass intensity list and S/N is shown; Click on the desired m/z Create a working list

Attach working list to Instrument method; Quantitation software

Presenter
Presentation Notes
AutoSIM allows you to acquire Full Scan data, enter or import from external file (csv, Excel, LIMS database) your list of target analytes, and perform selection of the SIM ions from the FS data in guided, automated fashion. Real time saver and productivity booster.
Page 8: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

10

Productivity Features: t-SIM

Acquisition - Timed

Link the working list; the work list is in fact now a database

Presenter
Presentation Notes
Based on very successful t-SRM feature on TSQ 8000, Timed Acquisition (t-SIM) allows you exactly the same but in SIM mode. It helps optimizing the dwell times in SIM quantitation methods with large number of analytes, especially co-eluting analytes. However, setting up t-SIM can be tedious. This is why AutoSIM compliments (and requires) t-SIM.
Page 9: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

11

AutoSRM

1) Precursor ion selection

2) Product ion selection

3) Collision energy optimization

Presenter
Presentation Notes
AutoSRM is a very simple to use, yet powerful tools that automates the SRM method development process a much as the user desires. It walks the user through (or runs through automatically) the entire method development process for SRM that are fully optimised for the system.
Page 10: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

12

AutoSRM

• Automates the following: • Creation of fullscan, product ion scan and SRM methods • Creation of sample sequences • Creation of data layouts for analyzing results • Selection of precursor, product and collision energies

End result showing optimized transition

Presenter
Presentation Notes
In more detail, it…
Page 11: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

13

Timed-SRM Method Overview

Acquisition - Timed

Link the working list; the work list is in fact now a database

Pre and post times are adjustable; So in case of shift due to heavy matrix the compound is still inside the window

Presenter
Presentation Notes
The timed SRM approach automatically creates windows centered around the retention time of each component in the method. It also allows overlaping of transitions for those compounds that elute closely together. However, these windows are independent of each other.
Page 12: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

14

GC-MS and MS/MS Method Development Workflow

Develop missing compound/ SRM or SIM

Add to database

Multi-component method

Automatic SIM or MRM method creation

Self -customized Auto-optimized

RT Update

My target compound

list Auto SIM Auto SRM

Just ask us for existing databases

Presenter
Presentation Notes
And when utilised in conjunction with the compound database and the instrument method enables a powerful method development workflow This eliminates the vast majority of time and effort required to develop methods from weeks down to days
Page 13: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

15

Analysis of 300 Pesticides – Segmented vs. Timed

Segmented SRM Timed SRM Closest compound to segment break: 5 seconds

Can lead to compounds disappearing in case of RT shift

Closest compound to segment break: 15 seconds

Average number of simultaneous transitions: 55

Average number of simultaneous transitions: 15 (4X higher dwell times- better sensitivity)

Presenter
Presentation Notes
When comparing the performance of these two approaches we can see two real advantages with timed-SRM; First, is the that with timedS, each compound is centered in it’s window. This means that the method is more robust to losing peaks due to matrix shifts than a segmented method when comparing compounds near the edge of the window. More importantly, the timed methods greatly reduces then amount of simultaneous transitions running at one time. This results in more effective use of dwell time and greater sensitivity
Page 14: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

16

AutoSRM User Case

• Created and optimized > 250 transitions for > 80 compounds

• Minimal user interaction over 24 hours

Page 15: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

17

TSQ 8000 Timed-SRM User Case result

Terbacil

Alachlor

Tolylfluanid

Pyridaben

Tea Analysis: 4 pg on-column

Page 16: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

18

TSQ 8000 Timed-SRM User Case result

Spiked ketchup Analysis: 50 ppb Example diazinon

12.5 13.0 13.5 14.0 14.5 15Time (min)

RT: 13.64AA: 1555401SN: 80040

RT: 13.63AA: 1047822SN: 6301

Top: Timed SRM: - S/N 80040 - response: 1555401 Bottom: Segmented SRM: - S/N 6301 - response: 1047822

Page 17: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

19

TraceFinder Method Sync

• Links TraceFinder Method with instrument method

• Enables: • Compound based acquisition setup

• Automated update of acquisition windows

Page 18: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

21

Method Sync – Automated RT Update

…updates both TraceFinder Method and Timed-SRM Method

Page 19: Fast Method Development using AutoSRM and autoSIMtools.thermofisher.com/content/sfs/brochures/Vendor-Semiar-3-ISCC... · Fast Method Development using AutoSRM and autoSIM . 4 . RT:

22

In Conclusion

Image courtesy of digital art FreeDigitalPhotos.net

• Auto SIM and auto SRM:

• Simplifies the method development • Speeds the method development up • Makes adding new compounds easy • Connects the instrument and quantifying method

• Timed SIM and SRM:

• Eliminates the need for cumbersome segmenting • Limits the number of ions per unit of time; • Thus ensuring optimal dwell time and increasing the signal

to noise

Presenter
Presentation Notes
Triple quadrupole is increasingly becoming the technique of choice for many laboratories due to it’s advantages, especially in improving methodologies involving complex matrices that have been traditionally run using single quad instruments There are many perceived barriers to bringing this technology into the laboratory but today this is easier than ever with the latest generation of instrument designed to alleviate these issues