Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
T E C H
A Performance Comparison Between Reprocessed and New Covidien LigaSure™ 5mm Blunt Tip Laparoscopic Sealer/Divider (1637)
E V I D E N C E R E P O R T
The Stryker Sustainability Solutions Reprocessed LigaSure™
5mm Blunt Tip Laparoscopic Sealer/Divider (LF1637) is a
bipolar electrosurgical instrument intended for use with the
ForceTriad™ Energy Platform in general (including urologic,
vascular, thoracic, and thoracoscopic) and gynecologic
laparoscopic surgical procedures where ligation and division
of vessels and lymph is desired. The instrument creates a
seal by application of RF electrosurgical energy to vascular
structures (vessels and lymph) interposed between the jaws
of the instrument. A blade within the instrument is surgeon
actuated to divide tissue.1
This evidence report outlines the extensive testing
performed to demonstrate substantially equivalent functional
performance of Reprocessed LigaSure™ 5mm Blunt Tip
Laparoscopic Sealer/Divider (LF1637) in comparison to
the Original Manufacturer’s (OM) device, which has not
been reprocessed. Wide-ranging bench-top tests were
conducted to validate the following functional attributes:
• Vessel burst pressure
• Thermal spread and jaw temperature
• Blade and jaw functionality
• Electrical resistance and safety
• Overall device reliability and functionality
INTRODUCING THE REPROCESSED LIGASURE™ 5mm BLUNT TIP VESSEL SEALER/DIVIDER (LF1637)
T E C HE V I D E N C E R E P O R T
Electrical resistance testing is performed to verify that each
device provides a continuous conductive pathway, ensuring
proper functionality during each seal. This is performed
as part of the manufacturing process, where each device
(100%) is tested as part of the in-line inspection criteria.
Dielectric withstand testing is also performed on each
device (100%) to ensure integrity of the electrical wire
insulation. The customized testing apparatus and test
method follows industry-accepted criteria referenced in the
electrical safety standard, IEC 60601 (Medical Electrical
Equipment Standard), published by the International
Electrotechnical Commission. Dielectric withstand testing
is important to not only ensure each device functions as
intended after reprocessing, but that it is safe to use on the
patient, as well as safe to handle by the clinical user.
Electrical Resistance and Safety
References1. VSD EL10029 Rev. A 04-2015 RM702135 - Instructions For Use - SSS Reprocessed LigaSure™ Blunt Tip Laparoscopic Sealer/Dividers (Model LF1637)2. Kim F J, Chammas M F Jr., Gewehr E, et al. Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), LigaSureTM V (LV), and plasma trisector (PT).
Surg Endosc 2008, 22:1464-14693. Reports on File. (T14398, T14336)4. Eick S, Loudermilk B, et al. Rationale, bench testing and in vivo evaluation of a novel 5 mm laparoscopic vessel sealing device with homogeneous pressure distribution in
long instrument jaws. Annals of Surgical Innovation and Research 2013, 7:15
Figure 1. Inline electrical testing
To demonstrate the Reprocessed LF1637 device’s
ability to safely and reliably seal a wide range of vessels,
including those larger in diameter, testing was conducted
using porcine carotid and iliac vessels, ranging in size
from 2-7mm diameter.
The Reprocessed LF1637 devices were compared to OM
LF1637 devices in order to evaluate substantial equivalence
of functionality between the two test groups. During the
testing, vessels were first sealed and cut, followed by
pressurizing the vessel by infusing with saline until the vessel
ruptured, while measuring and recording peak pressure.
Before the vessel burst, the maximum internal vessel pressure
was determined using a customized burst pressure fixture.
Vessel burst pressure results were required to demonstrate
substantial equivalency, or better than the OM device.
Larger (elevated) internal vessel pressures indicate a
stronger seal. A Mann-Whitney Test was used to analyze
the median distribution comparison between the groups,
which demonstrated significance with a p-value greater
than (>) 0.05, indicating the distribution’s medians were not
statistically different, and therefore, no statistical difference
between the two groups.
Vessel Burst Pressure
5. Sindram D, Martin K, Meadows JP, et al. Collagen-elastin ratio predicts burst pressure of arterial seals created using a bipolar vessel sealing device in a porcine model. Surg Endosc 2011; 25:2604-2612.
6. Latimer C, Nelson M , Moore C, et al. Effect of collagen and elastin content on the burst pressure of human blood vessel seals formed with a bipolar tissue sealing system. Journal of Surgical Research 2014; 186:73-80.
7. Rothmund R, Schaeller D, Neugebauer AA, et al. Evaluation of thermal damage in a pig model. J Invest Surg. 2012; 25(1):43-50.8. Lamberton GR, Hsi RS, Jin DH, Lindler TU, et al. Prospective comparison of four laparoscopic vessel ligation devices. J Endourol. 2008; 22(10):2307-2312.
Device Type N mmHg P value
Reprocessed (RP) LF1637 59 519.4
0.1112
OM LF1637 30 391.7
Table 1. Vessel Burst Pressure Results Summary
Figure 1. Inline electrical testing
1800
1600
1400
1200
1000
800
600
400
200
0
1800
1600
1400
1200
1000
800
600
400
200
0
Figure 2. Box Plot Comparison – Vessel Burst Pressure
LF1637 Burst Pressure Comparison (OM vs Reprocessed)Seal Quality (Burst of Seal in mmHg Pressure)
LF1637 OM Burst Pressure (mmHg)
Dat
a
LF1637 RP Burst Pressure (mmHg)
Lateral thermal damage can occur as a result of protein
denaturing in tissue, when sealing temperature reaches
above 60°C (140°F).2 This is a critical functional attribute to
consider when advanced energy devices are activated near
vital organs and structures, especially when used in a wide
range of procedures.
Porcine carotid and iliac vessels, ranging in diameters from
2mm to 7mm, were used to evaluate the distance of thermal
spread of tissue after a complete seal for both reprocessed
and OM devices. Infrared thermal images were recorded
using an infrared camera. An infrared camera along with
customized fixturing and validated software were employed
to capture, calculate, and analyze the maximum distance
from the middle of the device jaws to where the tissue
surrounding the jaws reached a temperature that would result
in protein denaturation in tissue (> 60 °C (140°F)).
Additionally, maximum jaw temperature testing was
performed to evaluate the maximum temperature of the
device jaw component during sealing of porcine carotid
and iliac vessels ranging from 2-7mm diameter. Infrared
images of each device’s thermal footprint was recorded
with a FLIR Infrared Camera using customized fixturing
and analyzed using a customized, mathematical software
program. Acceptance criteria dictated that the maximum jaw
temperature during a complete seal cycle shall be equivalent
to or less than the OM.
For both test methods, a total of N=29 OM samples were
compared to N=59 reprocessed samples and were evaluated
using a Mann-Whitney test for significance. Both reprocessed
and OM samples performed substantially equivalent to
each other when comparing medians, with p-values greater
than (>) 0.05, indicating the distribution’s medians were not
statistically different.
Thermal Spread and Jaw Temperature
MKT20150606A
Figure 3. Thermal footprint of distal jaw grasping vessel using FLIR camera (˚F) Figure 4. Image of jaw grasping vessel
As part of the manufacturing process for Reprocessed
LF1637 devices, the cutting blade and some distal tip
subassembly components are removed and discarded
during disassembly of the device prior to the other device
components entering the decontamination and cleaning
steps. These are replaced with brand new, equivalent
components during reassembly of the device, prior to in-line
function testing. The specifications (geometries, dimensional
tolerances, material, etc.) of these replacement components
were derived through reverse engineering the OM devices,
and therefore, are designed to be substantially equivalent to
those originating in an OM device.
As part of the validation activities performed to prove
substantial equivalency of the components, the replacement
blade and distal tip sub-assembly components underwent
extensive testing, in comparison to an OM device. The
following evaluations were performed, and have met the
established acceptance criteria based on bench-top testing
data of an OM device:
• Cut quality - This test evaluated the quality of a cut made
by the replaced blade.
• Blade deployment - This test evaluated the distance that
the blade travels within the jaw assembly when the blade
trigger is fully engaged.
• Blade trigger force - This test assessed the forces
required to engage the blade trigger.
• Blade trigger force through media - This test assessed
the forces required to engage the blade trigger when the
blade is cutting through media.
• Jaw opening angle – This test evaluated the angle
between the upper and lower jaws when the jaw is
completely open.
• Force to open jaws - This test assessed the forces
required to open the jaws.
• Jaw clamp force – This test assessed the forces that
are translated through the shaft of the device to the jaws
when the jaw trigger is engaged.
Blade and Jaw Functionality
Figure 5. Jaw force testing on the Instron (close-up) Figure 6. Jaw force testing on the Instron
R E P R O C E S S E D S U R G I C A L D E V I C E S
T E C HIn order to demonstrate overall device reliability and
functionality of the Reprocessed LF1637 device, extensive
simulated clinical cycle testing was performed, where each
reprocessed device was cycled through 120 simulated-use
actuations, intended to exceed a typical clinical use.
Each device was then evaluated to confirm the following
critical functional attributes:
• Seal cycle completion
▪ Appropriately sized vessels/tissue is adequately sealed.
▪ Device provides audible feedback for the cautery
button activation.
• Blade deployment
▪ When jaws are closed, blade deploys and retracts to
normal position
▪ When jaws are open, blade is not deployed
(safety mechanism)
• Grasping mechanism
▪ Device provides audible and tactile feedback for the jaw
locking/unlocking function
▪ Jaw release mechanism of grasped vessel/tissue when
handle is unlocked
• Unintended electrical energy transmission does not
occur through various feature combinations when
activation button is:
▪ activated, jaws are either open or closed (locked), and no
vessel/tissue is grasped.
▪ not activated, jaws are either open or closed (locked),
and no vessel/tissue is grasped.
▪ not activated, jaws are either closed (locked), and vessel/
tissue is grasped.
• Cable connection and generator recognition
▪ Cable inserts into generator as intended and remains
securely connected to the ForceTriad during use.
▪ Device is recognized by the ForceTriad Generator
software version 3.50 or greater.
▪ Jaw rotation
▪ User can rotate jaws when jaw trigger is unlocked.
• Compatibility with trocar accessory
▪ Device can be inserted and withdrawn through the
cannula of a 5 mm trocar, as intended.
▪ Force required to insert a device through a 5mm trocar is
measured.
Additional testing was performed on the rotator knob and
handle lock mechanisms:
• Rotator Knob
▪ Testing was performed to assess forces required to rotate
the rotator knob throughout its entire range of motion,
in addition to the amount of rotation (measured through
linear displacement) that is possible throughout the entire
range of motion.
• Handle Lock/Unlock Force
▪ Testing was performed to assess the forces required to
engage/disengage and lock/unlock the jaw trigger.
Overall Device Reliability and Functionality
sustainability.stryker.com • 888.888.3433