Disclosures and Support Donation from Medtronic (this work is investigator initiated), and other gift funds Support of NSF Engineering Research Center for Sensorimotor Neural Engineering NDSEG Fellowship NIH Training Grant (Computational Neuroscience) Some of this work involves off-label use of DBS Work conducted under FDA IDEs (and institutional IRBs)

Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Disclosures and Support

• Donation from Medtronic (this work is investigator initiated), and other gift funds

• Support of NSF Engineering Research Center for Sensorimotor Neural Engineering

• NDSEG Fellowship• NIH Training Grant (Computational

Neuroscience)• Some of this work involves off-label use of DBS• Work conducted under FDA IDEs (and

institutional IRBs)

Page 2: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

What is Deep Brain Stimulation?

• A “Brain Pacemaker”

• Constant electrical stimulation to deep brain structures

• Used to treat a variety of neurological disorders

• Stimulation target varies by disorder being treated

• Movement Disorders:

Essential Tremor, Parkinson’s Disease

5

Page 3: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Clinical DBS

• Originally based off of cardiac

pacemaker technology

– Battery and stimulation generator

implanted in chest

• “Open-Loop”

– Delivers therapy constantly

• Lead targets determined by

disorder and symptoms

6

Page 4: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

DBS for Essential Tremor

[Vaillancourt, 2003]

8

Page 5: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Current Issues:• Battery replacement

requires surgery

• Clinicians manually balance:– Minimizing power

– Minimizing side-effects

– Minimizing tremor

• An Open Loop Solution to a Closed Loop Problem

DBS Shortcomings

Deep Brain Stimulation

Image Credits: Medtronic and Microsoft Research

Motor Commands

9

Page 6: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

A Solution: Closed-Loop DBS

Sensors allow for

adaptive DBS

algorithms that

control when and how

stimulation will occur.

AdaptiveDeep Brain Stimulation

WearableSensing

Image Credits: Medtronic and Microsoft Research

NeuralSensing

Motor Commands

10

Page 7: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Closed-loop DBS will:

• Increase battery life

• Reduce side-effect

exposure

• Provide an important

research tool

• Require new hardware

platforms

ImprovedMotor

Commands

AdaptiveDeep Brain Stimulation

Sensing

Image Credits: Medtronic and Microsoft Research

Page 8: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Computational

Control Algorithm

Neurostimulator

Ne

ura

lC

om

man

ds

Signal ProcessingPhysiological

SensorsComm Link

Closed-Loop DBS System Design

12

Page 9: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Our Approach

Leveraging Investigational Devices

from Medtronic

– Shortcut to fully implanted systems in humans

– Gives a straightforward regulatory process to

use

Activa PC+S: Sensing and stimulation

Nexus-D: Real-time communication

PC

UI

Ap

pli

cati

on

Nexus-API

COM PORT

Nexus-D

USB-IR Bridge

Nexus SPTM

Activa PC+S

USB IR Coil

13

Page 10: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

The Nexus-D Extension

Designed and built a new Nexus adapter

• Enables a wireless & ambulatory Nexus-D

with Bluetooth

• Microcontroller expands command set

for ease of mobile development

Nexus-D Extension

Blu

eto

oth

Embedded Microcontroller

IRDA

Nexus-D

Nexus SPTM

Activa PC+S

IR Coil

14

Page 11: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Enables quick and iterative prototyping

without worrying about abstracted hardware

Nexus C# DLL

Computational

Host Application

Control Algorithm

Activa PC+S

Signal Processing

Sensors

Data

Nexus C# DLL

Stimulator

User Interface

Desktop PC

16

Page 12: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Now designing the closed-loop

algorithm and how to make use

of these components

System Integration

Publications:J. Herron, T. Denison, and H.J. Chizeck, “Closed-Loop DBS with Movement Intention”, In the Proceedings of the 7th International IEEE EMBS Neural Engineering Conference, Montpellier, France, April 2015

J. Herron and H. J. Chizeck, “Prototype Closed-Loop Deep Brain Stimulation Systems Inspired by Norbert Wiener”, In the Proceedings of the IEEE 2014 Conference on Norbert Wiener in the 21st Century, Boston, MA, June 2014.

Page 13: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Wearable Closed-Loop DBS

ImprovedMotor

Commands

AdaptiveDeep Brain Stimulation

Sensing

Image Credits: Medtronic and Microsoft Research

Use wearable sensors to

demonstrate closed-loop

DBS

• Evaluate stimulation power efficiency improvements

• Investigate side-effect mitigation

19

Page 14: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

• Inertial Sensing Watch

9 axis IMU: accelerometer, gyroscope, and

compass

• Electromyography

Muscle activity sensed via electrodes on the

arm

Worn systems allow for ambulatory data

collection for long-term monitoring and

diagnosis

Wearable Sensors

Image: Thalmic Labs Myo Aarmband

Image: LG G Watch

20

Page 15: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Wearable CLDBS for Tremor

1. Patient begins to experience

tremor

2. Wearable sensor senses the

arm’s tremor motions.

3. Sensor wirelessly transmits data

to computer

4. Experimental algorithm

determines new stimulation

parameters

5. Wirelessly updates stimulation

parameters to Activa PC+S using

the NexDEx and Nexus-D.

Image Credit: UW BRL

Page 16: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

CLDBS with Wearables for ET

Two algorithms tested

1. Tremor-modulated IMU Stimulation

• (direct) presence/absence of tremor to changes stimulation

2. Movement-triggered EMG Stimulation

• (indirect) sensing of movement or rest changes stimulation

In collaboration with Andrew L. Koat UW Medical Center:

Submitted Publications:J. Herron, M. Thompson, T. Brown, H.J. Chizeck, J. Ojemann, A. L. Ko, “Chronic ECoG Sensing for Movement Intention and Closed-Loop DBS with Wearable Sensors in an Essential Tremor Patient”, submitted to Journal of Neurosurgery.

Page 17: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

25

ET Wearable CLDBS Trial

IMU-Triggered EMG-Triggered

Tremor Strength

Page 18: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

26

Experimental Results

Both approaches resulted in a decrease in stimulation power and an increase in tremor band power when compared to open-loop stimulation

• IMU system had lower power• EMG system had lower tremor

Need a means of comparing algorithms ….

Page 19: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

27

Performance Metrics

• Tremor Difference Termo Use total tremor band power in with:

No stim (BPNO), Open-Loop (BPOL), and Closed-Loop (BPCL)o Final value represents the increase in tremor due to closed-loop

algorithm, normalized against open-loop tremor

• Stim. Power Reduction Termo Assuming impedance is constant, power is proportional to amplitude squaredo Square CLDBS amplitude (VCL) area under the curve, divide by total CLDBS

duration (T), divide by OLDBS amplitude squaredo Final value is the stimulation power reduction that the closed-loop system

provides compared to open-loop DBS

Page 20: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

28

Performance Metrics

Can use these two terms to determine the algorithm’spower-tremor tradeoff

Metric value represents the algorithm’s percentage decrease in stimulation power per 1% increase in tremor

Page 21: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Experimental Algorithm Performance

29

IMU Tremor-Modulated System• Tremor Difference: 36%• Power Reduction: 84%• Trade-Off: 2.3%

EMG Movement-Trigger System• Tremor Difference: 8.2%• Power Reduction: 53%• Trade-Off: 6.5%

IMU-Modulated EMG-Modulated

Page 22: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Wearable Closed-Loop DBSFor PD Rest Tremor Patients

30

Two algorithms tested – both using the IMU1. Tremor-Modulated Stimulation

- Same algorithm as ET-IMU tremor-modulated trial

2. Moving-Baseline StimulationTwo control loops based off of thresholding logic:-“Fast Loop” treats tremor quickly as it is sensed-“Slow Loop” inhibits return of tremor over long-term

Work done in collaboration with Dr. Helen Bronte-Stewart at Stanford University.

I built experimental systems for their use, participated in data collection, analyzed data

Publications:M. Malekmohammadi, J. Herron, A. Velisar, Z. Blumenfeld, M.Trager, H. J. Chizeck, H. Brontë‐Stewart, “Kinematic Adaptive Deep Brain Stimulation for Resting Tremor in Parkinson's Disease”, Journal for Movement Disorders, January 2016

Submitted Publications:J. Herron, A. Velisar, M. Malekmohammadi, H. Bronte-Stewart, and H. J. Chizeck, “Metrics for Comparing Closed-Loop Deep Brain Stimulation Algorithms”, Submitted to Journal of Neural Engineering.

Page 23: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Tremor Modulated: Patient 1, Limb 1

Cal Closed Loop Open Loop Closed-Loop

31

Tremor Diff: 12% Power Red: 83% Trade-Off: 7.2%

Page 24: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Cal Closed Loop Open Loop Closed-Loop

32

Tremor Modulated: Patient 1, Limb 2

Tremor Diff: 26.7% Power Red: 72% Trade-Off: 2.7%

Page 25: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Wearable Summary

• Demonstrated dynamic closed-loop DBS using different sensors and algorithms (in PD)

• Algorithm performance can vary across patients and across limbs in same patient

– Need to develop patient-specific algorithms

– Requires patient system identification and modelling

• Ongoing work by PhD Student Andrew Haddock

35

Page 26: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

36

Neural Sensing for BCI

Page 27: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Why Neural Sensing?

37

Deep Brain Stimulation

Sensing

Neural Sensing for DBS

• Neural sensing may be useful for determining

when a patient is suffering from symptoms

• Can determine when a patient is moving to

deliver therapy when needed

Neural Sensing for Brain-Computer

Interfacing

• Neural sensing may allow implanted patients to

control devices with their mind

Page 28: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Cortical Strip Electrodes

• Using a four-electrode strip for

cortical sensing

• Placement is flexible, can target

motor cortexElectrode Image

Credits: Medtronic

DBS Electrodes

• Standard DBS Electrode with

four contacts

• Deep brain structures targeted

for therapy

Sensing Electrodes

Page 29: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Closed-Loop DBS for ET: Prompted

Movement Task

Houston, Thompson, Ojemann, Ko, Chizeck, IEEE NER’17

With CL DBS

Herron, Thompson, Brown, Chizeck, Ojemann, Ko,

In press, IEEE TNSRE, 2017.

Page 30: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Closed-Loop DBS for ET: Therapeutic Effect

OFF ON CL

Houston, Thompson, Ojemann, Ko, Chizeck, IEEE NER’17

Activa PC+S, and cortical strip electrodes

3 ET subjects at UW, ~12 months

Herron, Thompson, Brown, Chizeck, Ojemann,

Ko, In press, IEEE TNSRE, 2017.

Page 31: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

• With Stanford, collected

first Nexus-D streamed

data from a human

• From PD STN during

rest and movement

Streamed Data Locally recorded on Activa PC+S

Lines from 6 dropped packets

PD STN Data with the Nexus-D

41

Beta Band

Page 32: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Neural Sensing and Brain-Computer Interfaces

in ET Patients

42

Collaborating with Dr. Andrew Ko and Dr. Jeff Ojemann at UW Medical Center

Also done in collaborating with PhD students Margaret Thompson and Tim Brown

Two Experimental Tasks1. Cortical sensing during overt and imagined movement

-Using neural signals to determine when patient is moving

2. Constant Velocity Brain Computer Interface Experiment-Training patient to volitionally manipulate cursors with his brain

Upcoming Publications:J. Herron, M. Thompson, T. Brown, H.J. Chizeck, J. Ojemann, A. L. Ko, “Chronic ECoGSensing for Movement Intention and Closed-Loop DBS with Wearable Sensors in an Essential Tremor Patient”, submitted to Journal of Neurosurgery.

Page 33: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Activa PC+S Cortical Sensing

Overt Movement Imagined Movement

43

Overt Movement Imagined Movement

Page 34: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Stability of Cortical Recordings

In hand motor area, beta-band across 1-4 months

Publications:M. Thompson, J. Herron, A. Ko and H. J. Chizeck, “Demonstration of a Chronic Brain-Computer Interface using a Deep

Brain Stimulator” in the 6th International Brain-Computer Interface Meeting, May 2016, Asilomar, CA.

Rest Overt Movement

Page 35: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Neural Sensing of Movement Summary

• Sensing with the Activa PC+S with cortical strips on the motor cortex can easily distinguish hand and arm movements from rest

• This has been stable over one year ofsessions, demonstrating the chronic sensing possibilities

• Imagined movements are more difficult with only a small change in a small band.

45

Page 36: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Implanted BCI

Image Credit: Phys.org

Utilize implanted electrodes

as signal sources for a

Brain-Computer Interface

(BCI)

• Demonstrate use of brain signals for command and control of devices

• Compare efficacy of different signal locations for BCI

Page 37: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Integration with Unity

• Proof of concepts for Activa PC+S BCI games

prototyped in Unity

• Allows easy configuration of trial geometries and tasks

Desktop Computer

Host ApplicationActiva PC+S

Signal Processing PhysiologicalSensors

DataNexus C# DLL

Unity Environment

Page 38: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

BCI Training: Overt Movement Trials

48

Page 39: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

BCI: Imagined Movement Trials

49

Page 40: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Neuroethics

Page 41: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Patients with DBS have reported problems post-op

(Agid, et. al. 2006):

Changes in self-perception.

o “I don’t seem to recognize myself without the problems I had before.”

o “I’m an electronic doll”—“I feel like a ‘Robocop’”—“I’m under remote

control”

o “Even though I know my motor state is perfectly normal, I just can’t get

rid of the thoughts I used to have, when I was ill.”

Changes to interpersonal relationships.

o “Since the operation, he wants to live like a young man […] I prefer

him as he was before, always nice and quiet.”

o “During all the years he was ill, I had to grin and bear it […] but I can’t

stand it any more. […] He just sits there and expects me to do

everything.”

Neuroethics

The Psychosocial Impact of DBS

Page 42: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Theories from the literature—DBS may:

o Make patients feel alienated from or inauthentic to

themselves. (Kraemer, 2011 & 2013)

o Challenge or clash with the patient’s self-narratives. (Baylis,

2011)

o Impair the user’s ability to be autonomous (Mackenzie 2014).

Two questions:

o What are the psychosocial effects of closed-loop DBS?

o Would closed-loop control make DBS easier to accept

from any of these perspectives?

Explaining Psychosocial Effects

Page 43: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Closed-Loop DBS User Interviews

• Semi-structured interviews—an on-going conversation

with each patient.

• Open-ended questions on a wide range of topics:

✓ “Has ET changed how you get along with people?”

✓ “Do you ever forget that you have a brain

implant?”

✓ “How often to you turn your stimulator on or turn it

off?”

• Interviews conducted while the patient tests the

experimental platform.

Submitted Publications:T. Brown, J. Herron, M. Thompson, A. Ko, H. Chizeck, S. Goering. “Controlling Our Brains — On the Possibility of Brain Computer Interface-Triggered Deep Brain Stimulation for Essential Tremor”, Submitted to Brain-Computer Interfaces.

Page 44: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Preliminary Results (1st ET Subject)

Initial body image issues (post-op):

o Initial worries about others being able to

see implanted leads.

o “I feel like I’ve got a conduit going down

the back of my neck. I’m always rubbing

on it.”

o He is constantly aware that the implant is

there.

T. Brown, P. Moore, J. Herron, M. Thompson, T. Bonaci, H. Chizeck, S. Goering. “Personal Responsibility in the Age of User-Controlled Neuroprosthetics,” 2016 IEEE International Symposium on Ethics in Engineering, Science, and Technology

Page 45: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Preliminary Results (1st ET Subject)

Battery conservation is a high priority:

o He keeps his stimulator turned off “95% of the

time.”

o Why conserve battery?

✓Avoid surgery.

✓Medical costs after retirement.

✓ Looking ahead: a thinner stimulator unit.

Mixed reactions to closed-loop control:

o Excited to have it

o Worries about hard-to-use system.

Page 46: Disclosures and Support › presentation › 46b9 › ...Now designing the closed-loop algorithm and how to make use of these components System Integration Publications: J. Herron,

Acknowledgements

Jeff Ojemann

Professor, Neurosurgeon

Andrew Ko

Neurosurgeon

This work is supported by AwardNumber EEC-1028725 from the NationalScience Foundation. The content is thesole responsibility of the authors anddoes not necessarily represent theofficial views of the National ScienceFoundation.

Collaborators:

Funding:

Helen Bronte-Stewart

Professor, Director of Stanford

Movement Disorders Center

Timothy Brown

Philosophy PhD Student

University of Washington

StanfordAnca Velisar

Research Scientist

Zack Blumenfeld

Research Scientist

Margaret Thompson

EE PhD StudentTamara Bonaci

EE PhD

Brady Houston

Neuro PhD Student

Thanks all the

members of the UW

BioRobotics Lab!

Katherine Pratt

EE PhD Student

Andrew Haddock

EE PhD Student

57

Jeffrey Herron

EE PhD Student