28
ORTEC ® MÖBIUS Recycler Condensing Liquid Nitrogen HPGe Detector Cooling System User Manual Printed in U.S.A. ORTEC® Part No. 805709 0514 Manual Revision F

Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

ORTEC ®

MÖBIUS Recycler

Condensing Liquid Nitrogen

HPGe Detector Cooling System

User Manual

Printed in U.S.A. ORTEC® Part No. 805709 0514 Manual Revision F

Page 2: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

ii

Advanced Measurement Technology, Inc. a/k/a/ ORTEC

®, a subsidiary of AMETEK

®, Inc.

WARRANTY

ORTEC* warrants that the items will be delivered free from defects in material or workmanship. ORTEC makes no other warranties, express or implied, and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

ORTEC’s exclusive liability is limited to repairing or replacing at ORTEC’s option, items found by ORTEC to be defective in workmanship or materials within one year from the date of delivery. ORTEC’s liability on any claim of any kind, including negligence, loss, or damages arising out of, connected with, or from the performance or breach thereof, or from the manufacture, sale, delivery, resale, repair, or use of any item or services covered by this agreement or purchase order, shall in no case exceed the price allocable to the item or service furnished or any part thereof that gives rise to the claim. In the event ORTEC fails to manufacture or deliver items called for in this agreement or purchase order, ORTEC’s exclusive liability and buyer’s exclusive remedy shall be release of the buyer from the obligation to pay the purchase price. In no event shall ORTEC be liable for special or consequential damages.

Quality Control

Before being approved for shipment, each ORTEC instrument must pass a stringent set of quality control tests designed to expose any flaws in materials or workmanship. Permanent records of these tests are maintained for use in warranty repair and as a source of statistical information for design improvements.

Repair Service

If it becomes necessary to return this instrument for repair, it is essential that Customer Services be contacted in advance of its return so that a Return Authorization Number can be assigned to the unit. Also, ORTEC must be informed, either in writing, by telephone [(865) 482-4411] or by facsimile transmission [(865) 483-2133], of the nature of the fault of the instrument being returned and of the model, serial, and revision (“Rev” on rear panel) numbers. Failure to do so may cause unnecessary delays in getting the unit repaired. The ORTEC standard procedure requires that instruments returned for repair pass the same quality control tests that are used for new-production instruments. Instruments that are returned should be packed so that they will withstand normal transit handling and must be shipped PREPAID via Air Parcel Post or United Parcel Service to the designated ORTEC repair center. The address label and the package should include the Return Authorization Number assigned. Instruments being returned that are damaged in transit due to inadequate packing will be repaired at the sender's expense, and it will be the sender's responsibility to make claim with the shipper. Instruments not in warranty should follow the same procedure and ORTEC will provide a quotation.

Damage in Transit

Shipments should be examined immediately upon receipt for evidence of external or concealed damage. The carrier making delivery should be notified immediately of any such damage, since the carrier is normally liable for damage in shipment. Packing materials, waybills, and other such documentation should be preserved in order to establish claims. After such notification to the carrier, please notify ORTEC of the circumstances so that assistance can be provided in making damage claims and in providing replacement equipment, if necessary.

Copyright © 2014, Advanced Measurement Technology, Inc. All rights reserved.

*ORTEC® is a registered trademark of Advanced Measurement Technology, Inc. All other trademarks used herein are the property of their respective owners.

NOTICE OF PROPRIETARY PROPERTY — This document and the information contained in it are the proprietary property of AMETEK Inc., ORTEC Business Unit. It may not be copied or used in any manner nor may any of the information in or upon it be used for any purpose without the express written consent of an authorized agent of AMETEK Inc., ORTEC Business Unit.

Page 3: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

iii

TABLE OF CONTENTS

1. INTRODUCTION ........................................................................................................................................................... 1

1.1. HOW MÖBIUS WORKS ..................................................................................................................................... 1

1.2. UPS INTEROPERABILITY .................................................................................................................................... 1

1.3. SPECIFICATIONS ............................................................................................................................................... 2

1.3.1. Performance ....................................................................................................................................... 2

1.3.2. Mechanical .......................................................................................................................................... 3

1.3.3. CE ........................................................................................................................................................ 3

1.4. INSPECTION FOR SHIPPING DAMAGE .............................................................................................................. 3

2. SYSTEM OVERVIEW ..................................................................................................................................................... 4

2.1. CONTENTS OF THE SHIPPING CARTON............................................................................................................. 4

2.2. THE MÖBIUS ..................................................................................................................................................... 4

2.3. PREPARING THE MÖBIUS FOR USE .................................................................................................................. 6

3. INSTALLATION ............................................................................................................................................................. 9

3.1. ASSEMBLING THE SYSTEM ............................................................................................................................. 10

3.2. FILLING WITH LN2 .......................................................................................................................................... 11

3.3. SAFETY AND EQUIPMENT WARNINGS ........................................................................................................... 13

3.3.1. Liquid Nitrogen Safety ...................................................................................................................... 13

3.3.2. Other Precautions ............................................................................................................................. 14

3.3.2.1. HANDLING POPTOP DETECTOR/CRYOSTAT ASSEMBLIES ................................................................ 14

3.3.2.2. ELECTRICAL CAUTIONS ............................................................................................................ 14

3.3.2.3. UNINTENDED DETECTOR WARM-UP ......................................................................................... 15

4. MAINTENANCE ......................................................................................................................................................... 16

4.1. CLEANING OR REPLACING THE MÖBIUS FILTER ............................................................................................. 16

4.2. EMPTYING AND WARMING THE MÖBIUS ...................................................................................................... 17

4.3. TRANSPORTING OR SHIPPING THE MÖBIUS .................................................................................................. 18

4.4. DETECTOR WARM-UP .................................................................................................................................... 18

4.5. REMOVING THE DETECTOR/CRYOSTAT ASSEMBLY ........................................................................................ 18

4.6. DECOUPLING A POPTOP CAPSULE FROM THE CRYOSTAT ............................................................................. 19

Page 4: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

iv

ADDITIONAL WARRANTY AND SAFETY NOTICES

There are no serviceable parts in or on the MÖBIUS except those specified in this manual, e.g., the mains power cord, fuses, and filter. Opening the rear panel or otherwise accessing the interior components, attempting to detach the level monitor or remote status display leads, or performing operations other than those described in this manual (e.g., inserting/removing the mechanical cooler’s protective foam block, adding/removing liquid nitrogen, and inserting/withdrawing detector crystostats into/from the dewar) may void your warranty.

Using this equipment in a manner not specified in this manual may impair the protection the equipment provides.

Changes to valving, tubing, or other components of the MÖBIUS system can result in damage to equipment, injury, or death. Modifications to any system components must be made in consultation with ORTEC technical support personnel and authorized in writing in advance of such modifications.

SAFETY INSTRUCTIONS AND SYMBOLS

This manual contains up to three levels of safety instructions that must be observed in order to avoid personal injury and/or damage to equipment or other property. These are:

DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction is not observed.

WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed.

CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed.

Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product.

In addition, the following symbol might appear on the product:

ATTENTION – Consult the manual in all cases where this symbol is marked in order to determine the nature of the potential hazards and any actions that must be taken to avoid them DANGER – Hazardous voltage Protective earth (ground) terminal

Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product.

Page 5: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System

v

[Intentionally blank]

Page 6: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

vi

MÖBIUS-B w/ HJ-style cryostat

MÖBIUS-B w/ PopTop® cryostat

MÖBIUS w/ PopTop® cryostat

Page 7: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

1

ORTEC MÖBIUS™ RECYCLER CONDENSING LIQUID NITROGEN HPGE DETECTOR COOLING SYSTEM

1. INTRODUCTION

The ORTEC MÖBIUS Recycler integrates a Sunpower® CryoTel® Stirling-cycle cryocooler with a specially designed 28 liter liquid nitrogen (LN2) dewar to yield a high-reliability HPGe detector cooling system. This is especially useful for counting laboratories that have periodic power failures. The MÖBIUS fits easily under standard lead shields with the same footprint as a regular dewar. The remote status display can be mounted outside the lead shield for easy visibility.

A continuously operating MÖBIUS requires refilling only once every two (2) years — a major savings in worry, time, money, and exposure to LN2 handling risks. During a power failure, the MÖBIUS performs as a standard dewar. After power is restored, check the LN2 level and top it off if necessary — that’s all it takes to return the system to maximum performance.

The MÖBIUS supports our Streamline and PopTop® detector lines, and horizontal as well as vertical dipstick-type cryostats. In addition, a number of non-ORTEC cryostats and detectors are supported; contact your ORTEC representative for more information. If you have purchased a detector in addition to a MÖBIUS, all detector-specific information is included in the detector’s data sheet and user manual.

1.1. HOW MÖBIUS WORKS

The MÖBIUS has a computer-controlled pressure system that maintains a constant and stable 0.5 psi (3.5 kPa) in the vapor space of the dewar. A pressure transducer feeds the pressure in the dewar to the onboard computer, which controls the rate at which the Sunpower cryocooler re-condenses the nitrogen vapor to liquid. The cryocooler operates only when a cryostat is installed and the two fill/vent valves are closed (i.e., when the system is pressurized). This evaporation/re-condensation process is illustrated in Figure 1.

The fill collar has redundant pressure relief valves designed to open if the pressure exceeds 1 psi (6.9 kPa). In the event of power loss the cryocooler stops, the pressure inside the vapor space starts increasing, and nitrogen vapor slowly begins escaping into the atmosphere via the pressure relief valves. However, if power remains on and the system is operating in an uninterrupted steady state, there is no LN2 loss and the system can operate indefinitely without an LN2 refill.

The LN2 level monitor calculates the level of liquid nitrogen in the dewar, giving you ample opportunity to schedule maintenance.

1.2. UPS INTEROPERABILITY

The MÖBIUS is designed to run on ac mains power. However, during a power failure it acts as a conventional dewar and can keep a detector cold for 200 hours or more, depending on the level of LN2 in the dewar. When power is restored, the unit’s cryocooler restarts and it resumes recycling nitrogen gas boil-off as before.

Figure 1.

Page 8: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

2

The unit can be connected to an uninterruptable power supply (UPS) to sustain operation during short-duration power failures. Fully cooled, it draws about 150 watts of power from a UPS without a “smart” capability. Alternatively, the MÖBIUS can run off of a “smart” UPS with a dry contact relay interface, such as the APC SMT750 UPS equipped with their optional AP9613 dry contact management card. In this configuration, if power fails the UPS signals the MÖBIUS to enter its power-saving mode to conserve UPS battery life. The cryocooler turns off, reducing power consumption to less than 10 watts, while the remote status display continues to provide all status readouts including the liquid nitrogen level in the dewar.

1.3. SPECIFICATIONS

1.3.1. Performance

● Refill Period After the detector is cool and the LN2 is topped off, the MÖBIUS system can run for 2 years or longer without filling. This is based on the sealing technology and assumes system operation in uninterrupted, steady-state conditions.

● Maintenance Wash or replace inlet air filter every 3 months (replacements available from ORTEC). No other maintenance is required.

● Remote Status Display Readout unit with hardwired umbilical cord can be remotely mounted to improve visibility when the detector is placed in a lead shield. The display registers the liquid nitrogen level in the dewar, power level being supplied to the cooler, pressure level within dewar and general status of the system.

● Internal Controller Control electronics, power supply, and cooler electrically isolated from the dewar and protected from LN2 spillage.

● LN2 Level Sensing Capacitance level sensor provides continuous measurement of the LN2 level from 0–100% full.

● Use with UPS The MÖBIUS has a small onboard computer that can receive an input, via the rear-panel I/O connector, in the event of power loss. The input causes the system to enter power saving mode and turn off the cryocooler. With the cryocooler turned off, the onboard computer can be powered by a small UPS.

● LN2 Loss Rate Approximately 3 L/day with non-PopTop cryostats when cooler is off (same as for conventional dewar); loss rate will vary depending on ambient conditions and detector size. With PopTop cryostat typical consumption can be as high as 4L/day when cooler is off.

● Spectroscopic Performance

For new ORTEC detectors purchased with the MÖBIUS: at energies above 15 keV we certify no degradation of the detector resolution warranted on the detector data sheet, and below 15 keV, less than 10% of the warranted resolution.

For existing ORTEC detectors in dipstick cryostats, where the MÖBIUS replaces an existing dewar or X-Cooler®, resolution degradation is based on the existing detector performance. It is expected that the detector will have similar performance compared to using a standard LN2 dewar. Some degradation may occur depending on the detector's age and configuration. Note: Detector in a PopTop cryostat requires a 1.50 inch in diameter dipstick when converting from X-Cooler® to MÖBIUS.

MÖBIUS can be used with existing non-ORTEC detectors in dipstick cryostats having 1.25 or 1.50 inch diameter. It is expected that the detector will have similar performance compared to using a standard LN2 dewar. Some degradation may occur depending on the detector's age and configuration.

Page 9: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

3

1.3.2. Mechanical

● Dewar Capacity 28 L, excluding detector dipstick volume.

● Cooler AMETEK SUNPOWER with passive vibration dampening.

● Auxiliary Cooling Internal fan.

● Audible Noise Less than 60 dB (A) at 1 m fully operational, below 30°C ambient.

● Electrical Supply 100–240 V ac, 50 Hz or 60 Hz.

● Normal Power Consumption 125 W typical, 350 W maximum.

● Dimensions (excluding HPGe detector) 25.8 in. H x 17.2 in. Diam (65.4 cm x 43.7 cm).

● Operating Temperature 0–40°C, 20–90% relative humidity, non-condensing.

1.3.3. CE

● Intended for indoor use.

● Intended for altitude up to 2000 m (6562 ft).

● Mains supply voltage fluctuations up to +/-10% of the nominal voltage.

● Transient overvoltages up to the levels of overvoltage category II.

● Pollution degree 2.

1.4. INSPECTION FOR SHIPPING DAMAGE

If the shipping carton(s) arrives with externally visible damage, do not unpack it. Notify the carrier and make arrangements to file a damage claim. In all cases of shipping damage, it is the customer's responsibility to file a damage claim.

If, during unpacking, you find concealed damage, notify the carrier and file a claim. Be sure to save all packing materials, waybills, and other such documentation to establish claims.

Contact our Global Service Center or your ORTEC representative for further instructions.

Page 10: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

4

2. SYSTEM OVERVIEW

2.1. CONTENTS OF THE SHIPPING CARTON1

When you receive the MÖBIUS, its shipping carton(s) will contain the following components at minimum:

● MÖBIUS cryocooler/dewar main assembly with integrated LN2 fill collar, hardwired remote status display, and adhesive-backed hook-and-loop strip for attachment of remote display.

● Universal power cord kit (contains a universal ac mains power cord and a set of region-specific outlet).

● Fuse kit (contains fuse block and fuses; use the 250 V/10 A time-delay [slo-blo] fuses for 120 V ac mains power, and the 250 V/5 A time-delay [slo-blo] fuses for 220 V ac mains power).

● Acoustic damping foam shim pack with plywood platform and anti-vibration pad.

● User manual and documentation package.

Options will typically be packaged separately and will include their own instructions.

2.2. THE MÖBIUS

Figure 2 through Figure 5 illustrate the main features of the MÖBIUS.

Figure 2. MÖBIUS Top Panel Features (1).

1 All system components and specifications subject to change without notice.

Page 11: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

5

Figure 3. MÖBIUS Top Panel Features (2).

Figure 4. Rear Panel Features.

Page 12: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

6

Figure 5. Top Plug for MÖBIUS-B Models Only.

2.3. PREPARING THE MÖBIUS FOR USE

1) Unpack the dewar and make sure the dewar and vent/fill valve tube stems are empty of any packing material.

2) As shipped from the factory, the cryogenic cooler is protected from damage by a foam block inserted under the black cooler cover beside the fill collar (Figure 6). This foam must be removed before using the MÖBIUS.

Using a #1 Phillips screwdriver, remove the screws from the cooler cover and set them aside.

Lift off the cooler cover, then remove the caution tag.

Remove the foam block and keep it with along with all other shipping materials for transporting the MÖBIUS, returning it to the factory for service, etc.

Replace the cover and tighten the screws just enough that they will not vibrate loose. Do not over-tighten. This seal is intended as an air dam, not a pressure seal, so tighten the screws just enough to establish uniform, snug contact with the gasket.

3) Next, the fuses must be installed. As shipped, the fuse block and fuses are packaged in zippered plastic bags so you can select the proper fuse for the ac mains power at your site. Use the 250 V/10 A time-delay fuses for 120 V ac/60 Hz operation and the 250 V/5 A time-delay fuses for 220 V ac/50 Hz operation. Install the fuses in the block, turn the block until the key tab faces down (Figure 7), and push it into the power entry module until both sides snap positively into place (Figure 8).

CAUTION When the foam block is not in place under the cooler cover, keep the MÖBIUS upright or the cooler may be damaged. See also Section 4.3.

Do not operate the MÖBIUS without the cover installed; it is required for proper air flow to the cooler.

Page 13: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

7

Figure 6. Remove Protective Foam Block and Caution Tag Under Cooler Cover and Keep for Future Use.

Figure 7. Underside of Fuse Block is Keyed.

Figure 8. Insert Keyed Side Down, Snap Both Sides Into Place.

Page 14: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

8

4) The universal power cord adapts for almost any mains power configuration. Simply choose the plug that matches your outlet (Figure 9) and press it onto the power cord. The other end is a standard female IEC connector that plugs into the MÖBIUS power entry module.

Figure 9. Select the Plug That Matches Your Outlet.

Page 15: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

9

3. INSTALLATION

Installation requires at least two people (better still, three) and a step-stool. Specific shield config-urations call for specific installation methods. However, installing the MÖBIUS is essentially no different than installing a standard 30 L dewar because the two are very similar in size. The detector/cryostat assembly should be dry and warm and the MÖBIUS empty.

The mounting surface must be flat and horizontal. The shield and/or MÖBIUS must not be positioned so that it is difficult to disconnect the MÖBIUS from the electrical supply. Orient the dewar so its ventilation openings and filter are not blocked.

Figure 10 is an exploded diagram of a typical spectroscopy system with a PopTop detector, PV4 vertical dipstick cryostat, Model HPLBS shield with split center plug, and MÖBIUS.

Figure 11 shows a typical MÖBIUS installation under an HPLBS shield. Setting the detector height involves the interplay of shield height (adjustment typically requires a hydraulic jack and wrench), the thickness of acoustic damping material and/or other spacers under the dewar, and the insertion depth of the cryostat cooling rod in the dewar.

Note the arrangement of damping and spacer materials under the dewar necessary to bring it to the correct height. Various thicknesses of foam are used on the bottom (i.e., contacting the floor). The foam is followed by the plywood spacer, and the anti-vibration pad is on top. Order is important for the various spacers: Foam on the bottom, followed by the plywood, followed by the anti-vibration mat.

For installations where there is not enough room for all three layers, omit the foam first, then the plywood, then the anti-vibration mat as a last resort.

The detector can also be raised by adjusting the collar position on the cryostat. This is useful when using a center plug to reduce floor shine. Note however that this adjustment lifts the cryostat cooling rod out of the LN2 column, reducing the effective capacity of the dewar. In such cases, the detector begins to warm as soon as the LN2 level

Figure 10.

Page 16: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

10

falls below the cold tip rather than when the dewar is empty. One solution for this problem is to use a cooling rod extender such as the ORTEC Model CRE-1, which adds up 4 inches of cold tip to the cryostat.

Figure 11. The MÖBIUS Installed on Foam, Plywood, and Anti-Vibration Pad.

3.1. ASSEMBLING THE SYSTEM

1) Before installation, you may wish to assemble the the spacers, damping materials, MÖBIUS, and detector/cryostat assembly beside the shield to make preliminary height estimates, cryostat and shield height adjustments, etc. Otherwise, place the MÖBIUS under the shield first with no spacers under it

2) Holding the detector/cryostat assembly by the preamplifier shroud or under the detector capsule base, lower the cooling rod through the bottom of the shield — feeding the detector cabling through first — and through the fill collar into the dewar.

CAUTION Keep the cabling and connectors from being pinched, kinked, deformed, or otherwise damaged as they and the cooling rod pass through the bottom of the shield.

Page 17: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

11

3) If using the split center plug, put the two halves carefully in place before fully lowering the cooling rod through the bottom of the shield. To avoid ground loops, position the detector high enough inside the shield that its base plate does not touch the top of the center plug.

4) Finally, the damping materials and spacers can be inserted under the MÖBIUS. It is easiest to have two helpers lift the MÖBIUS while you slide the spacer stack underneath.

5) Now is a good time to choose a location for the status display and mount it with the hook-and-loop strips provided with the system. Make sure to position it where the readout is clearly visible (for example, on the shield table as in Figure 11).

3.2. FILLING WITH LN2

This procedure requires cryoprotective gloves and eye protection.

Note: pressure from the refill dewar should not exceed 22 psi (152 kPa).

1) Before filling the MÖBIUS with LN2, be sure to read all cautions in Section 3.3.

2) For MÖBIUS-B Models, ensure the supplied top plug is secure (see Figure 5) and the detector is properly installed on the cryostat before filling.

3) The MÖBIUS is now ready for power-up. Connect the mains power cord to a properly grounded electrical

supply and set the power switch to the “On” position. The remote status display will light up and show LN2:

0%, Pwr: 0% indicating that the dewar is completely empty and cooler power is off. If dewar is partially filled then display will show X% depending on the level of LN2 currently in the dewar.

4) With the remote status display clearly visible, open the two fill/vent valves on the front of the dewar (Figure 2). This will depressurize the system. The status display will now show pressure PSIG: 0.0 and fill level LN2: 0% (or X% is dewar is partially filled). Display should look as follows:

5) Connect the transfer hose from the supply dewar to the MÖBIUS. Its two fill/vent valves are identical to one another, hence, interchangeable (i.e., if the supply dewar is connected to the left connection, the right connection becomes the vent, and vice versa). Although not specifically required for protection of the MÖBIUS, an exit line can be connected to the vent valve to route gas and liquid that escapes during filling to a safe location. Suitable transfer and vent hoses are available separately from ORTEC. Pressure from the refill dewar should be between 5 and 22 psi. Pressure from the refill dewar should not exceed 22 psi (152 kPa).

6) Open the valve on the supply dewar to start the flow of LN2 into the MÖBIUS. Cold nitrogen gas will immediately begin blowing out of the vent side and will condense moisture in the air, creating a clearly visible fog. If this fog is not observed, turn the supply dewar valve off immediately and ensure that both valves on the front of the Mobius are open.

7) As the liquid level in the MÖBIUS rises, the liquid nitrogen fill readout on the status display will increase until the dewar is full and the readout indicates 100%. Fill time depends very much on supply dewar pressure, but is typically between 10 and 30 minutes for a completely empty MÖBIUS. Once filled, display would look as

Page 18: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

12

follows:

8) Frost will form on the fill and vent tubing, fittings, and other cold surfaces as the LN2 flows. The frost serves as a warning that these surfaces that are very cold and should not be touched without suitable gloves. It is not unusual for the valve on the supply dewar to frost over during the fill process, so be sure to have gloves handy so you can quickly close the supply valve as soon as the MÖBIUS is full. The frost also provides additional thermal insulation to the lines so do not wipe it off during LN2 transfer.

Note: the fill and vent valves on the MÖBIUS should not be operated (opened or closed) until they are at room temperature.

9) When the MÖBIUS status display indicates LN2: 100%, watch the fog exiting the vent tube very closely for signs of small LN2 droplets first, then large droplets. Immediately close the supply valve as soon as you observe these droplets, otherwise the MÖBIUS system electronics and seals could incur damage.

10) Leave the fill and vent lines connected until all frost on the lines and fittings has thawed (this typically takes just a few minutes). Only when all signs of frost are gone can the lines can be removed safely. Cold gaseous nitrogen will continue to exit the vent as the detector continues to cool. The system should be left undisturbed overnight to allow the detector and MÖBIUS to completely cool down. During cooldown leave the MÖBIUS fill/vent valves open.

11) A warm detector causes a significant amount of LN2 to boil off at first. Once the detector is cold, it is a good idea to top off the LN2 level in the MÖBIUS. After top-off, remove any attached hoses and close the fill/vent valves. The system will begin to pressurize as LN2 in the dewar vaporizes. The MÖBIUS will sense this pressure increase and turn on its cryogenic cooler to begin re-condensing the evaporating LN2. Between PSIG values of 0 and 0.2 system will indicated Vented status. Between PSIG values of 0.45 and 0.55 status will indicate OK status. Between PSIG values of 0.65 and 1.4 status will indicate Cooling status. At first, the cooler and its cooling fan will run at high power. While cooling, MÖBIUS display should look as follows:

where X should have a value between 0.0 and 1.0 psi. Note: LN2 may indicate value slightly below 100% at this point. Note: Two safety valves will vent excessive pressure, if pressure exceeds approximately 1 psi.

12) After about 2 hours the system will stabilize so the cooler and its cooling fan are running just at the power level necessary to maintain a constant pressure of 0.5 psi (3.5 kPa; 104.4 kPa absolute pressure) inside the MÖBIUS dewar. Once the system is stabilized, power consumption level will vary while pressure is maintained around 0.5 psi. It is normal for cooler to turn off occasionally when pressure drops below 0.50 psi. A display example of a pressurized, cooled down MÖBIUS would look as follows:

Page 19: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

13

13) At this point, the MÖBIUS is a closed system that is designed to contain all nitrogen vapor. You are now ready to connect the detector cable to the detector and multichannel buffer.

NOTE: Mobius is designed to operate without liquid nitrogen refilling for up to two years under normal,

continuous operating conditions (e.g. no power loss). The display will blink when the LN2 levels drops below 85%. This value was established to alert the user that the LN2 level in the Mobius is not sufficient to provide cooling for a full seven days upon an extended power outage.

3.3. SAFETY AND EQUIPMENT WARNINGS

3.3.1. Liquid Nitrogen Safety

Before filling the MÖBIUS, be sure to read the following LN2 advisories as well as the information in Section 3.3.2.

DANGER

● The large expansion ratio of LN2 from liquid to gas (1 to 692) can generate high pressures if the dewar does not have adequate venting or pressure relief provisions.

● When filling the MÖBIUS, you may be tempted to use dry nitrogen gas to pressurize the supply dewar. ORTEC STRONGLY RECOMMENDS AGAINST THIS PRACTICE. Instead, only the self-generated pressure of the supply dewar should be used. Compressed gases (even purportedly dry “house” nitrogen) may contain moisture, oil, and oxygen. Any oxygen will liquefy at LN2 temperatures (77 K), creating an explosion hazard. In addition, the moisture and oil can freeze, making the safety valves inoperable. Use a pressure relief of 22 psi (152 kPa) on the supply dewar to avoid over-pressurization in the event of ice blockage. A blockage can rupture the dewar and/or expel the detector/cryostat with explosive force.

If the MÖBIUS dewar becomes contaminated with oil, the unit must be returned to ORTEC for service. Do not use a degreaser or attempt any other remediation unless authorized to do so by our ORTEC Global Service Center.

If water accumulates in the dewar, remove the LN2; obtain a cylinder of high-purity, oil-free, dry nitrogen gas (not “house” nitrogen); and stream the gas at very low pressure through the dewar until all water has evaporated.

● Piping or transfer lines should always be constructed so as to avoid trapping LN2 in the line. Evaporation can result in pressure buildup and eventual explosion of the line. If all lines cannot be emptied, install safety relief valves and rupture discs. Also keep in mind that some materials become brittle and fracture when exposed to LN2 temperatures. For advice on selecting materials for use in storing and transferring LN2, contact ORTEC or your LN2 supplier. Other sources of information include the CRC Handbook of Laboratory Safety and the American Society of Mechanical Engineers’ Boiler and Pressure Vessel Code, Section VIII.

● Nitrogen gas can pose an asphyxiation hazard by displacing air. Transfer LN2 only in a well-ventilated area. Vent storage containers to a well-ventilated area or to the outside to avoid buildup of nitrogen gas in the work area.

WARNING

● On contact with skin, LN2 causes burns similar to thermal burns or frostbite. Eyes are especially vulnerable to this type of exposure. Avoid wearing anything capable of trapping or holding spilled liquid nitrogen close to your skin (for instance, mesh-top running shoes). An impervious apron or coat, cuffless trousers, and high-topped shoes are recommended. Wear safety glasses or, better yet, full-face protection. Remove all watches, rings, bracelets, or other jewelry. When gloves are used to handle containers or cold metal parts, they should be impervious and sufficiently large to be easily tossed off the hand in case of a spill.

● Before removing a detector/cryostat assembly from a MÖBIUS, always vent the system by opening both the fill and vent valves (Figure 2). Failure to do so may result in the entire contents of the dewar spewing

Page 20: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

14

dangerously out of the fill collar as the cryostat is removed. Redundant relief valves on the MÖBIUS fill collar reduce this risk, but it is always best to eliminate the risk by first opening the fill and vent lines.

CAUTION

● The detector electronics must be protected from excessive cold. Avoid spilling LN2 on the cryostat or electronics, and vent cold gas away from the system being filled. A cold detector must not be enclosed in any thermally insulating materials at any time. Following these precautions will prevent damage to the vacuum seals or system electronics from excessive chilling.

● Never attempt to fill the MÖBIUS by manually pouring LN2 through the fill collar. Always install the detector before filling, and use a fill dewar. The recommended filling procedure described in Section 3.2 presents minimal risk to the unit and users.

3.3.2. Other Precautions

3.3.2.1. HANDLING POPTOP DETECTOR/CRYOSTAT ASSEMBLIES

● Grasp the assembly just beneath the base of the detector capsule, where the cable exits the base plate; or by the preamplifier shroud (see Figure 12). Do not touch any cryogenically cooled surfaces (if moving a cryostat that is already cold), and do not hold the detector by the endcap!

Figure 12. Where To Hold a PopTop.

● If the detector is cold, at the same time you lift or lower the assembly, turn it slightly clockwise (as viewed from above the endcap; i.e., turn it in the direction that tightens the PopTop capsule to the cryostat). Applying counterclockwise force to a cold detector will rupture the capsule’s vacuum seal, and the unit must be returned to ORTEC for repair.

3.3.2.2. ELECTRICAL CAUTIONS

● Ensure the unit is fused for the proper voltage — 250 V/10 A time-delay fuses for 120 V ac systems and 250 V/5 A time-delay fuses for 220 V ac systems — before connecting the power cord to ac mains power. Replace blown fuses with the same model and performance specifications. Never short circuit a blown fuse.

● Always disconnect the power cord from the MÖBIUS before performing any maintenance.

Page 21: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

15

● If the mains power cord is damaged or lost, contact ORTEC for the correct replacement cord. Do not replace with an inadequately rated cord.

3.3.2.3. UNINTENDED DETECTOR WARM-UP

If, for any reason, the detector warms up enough that the HV shuts down (e.g., loss of LN2) or if for any other reason the detector warms for more than 10 minutes, it should be warmed to room temperature before re-cooling. Typical warm-up time for most detectors is 24 hours. For higher-efficiency detectors (>70%), allow 36 hours warmup time.

To determine if the detector is cool enough to be operated, attempt to turn on the detector bias (high voltage). If the detector is still too warm, the bias shutdown circuit will prevent high voltage from being applied. If the detector is cold enough, the bias will turn on. Be sure to allow the voltage to stabilize 15–30 minutes before taking data with an HPGe detector.

If the detector remains in shutdown mode after 36 hours of continuous cooling, contact your local ORTEC representative or our service center.

Page 22: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

16

4. MAINTENANCE

4.1. CLEANING OR REPLACING THE MÖBIUS FILTER

The efficiency of the cryogenic cooler depends on the cooling fan’s ability to remove heat from the system. The cooling fan pulls air through a filter on the left side of the unit and exhausts that air on the right side of the unit. As the filter becomes clogged with dust, the cooler becomes less efficient. A less-efficient cooler means the system has to run at higher power levels, which can adversely affect detector performance due to increased vibration. Therefore, ORTEC recommends cleaning the filter every 3 months (more often in areas with higher than normal dust levels).

1) Remove mesh screen and filter by hand, as shown in Figure 13, by grabbing the mesh screen and filter in the center with your fingers and pulling away from the MÖBIUS.

Figure 13. Removing the Mesh Screen and Filter.

2) Clean the air filter thoroughly using dish soap and water. If the air filter is damaged or so dirty that it cannot be well cleaned, discard and replace it with a new one, available from ORTEC (part number - SUNP-300112).

3) Push the air filter back into the bracket on both sides.

Figure 14a. Installing the Filter.

Page 23: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

17

4) Once the air filter is inserted, push the mesh screen back in the bracket as in Figure 15b.

Figure 15b. Installing the mesh screen.

4.2. EMPTYING AND WARMING THE MÖBIUS

Turn the MÖBIUS rear-panel power switch to the “Off” position, open the fill and vent valves to depressurize the dewar, and remove the detector/cryostat assembly according to Section 4.5. Then remove the LN2 with one of the following methods:

● Slower — Obtain a cylinder of high-purity, oil-free, dry nitrogen gas (not “house” nitrogen). Connect to the fill valve and stream at very low pressure to gently boil off the LN2 and evaporate any water from inside the dewar.

● Faster — Use the ORTEC Model WD/30 LN2 withdrawal device (Figure 16) to transfer the LN2 to another dewar. The WD/30 uses the pressure buildup in the dewar vapor space to push the LN2 up the tube and out a transfer line. Attaching the WD/30 requires removing the fill collar. Contact ORTEC for more details.

Figure 16. ORTEC Model WD/30.

CAUTION Do not tip the MÖBIUS and pour the LN2 out of the fill collar! Doing so will damage the cooler and may breach the dewar’s vacuum seal.

Page 24: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

18

4.3. TRANSPORTING OR SHIPPING THE MÖBIUS

1) Empty and warm the MÖBIUS according to Section 4.2.

2) Using a #1 Phillips screwdriver, remove the screws from the cooler cover and set them aside.

3) Grasp the cooler cover firmly and pull up or push sideways on it to break the friction grip of the cover gasket (which is very “sticky”). The gasket is affixed to the top of the MÖBIUS, not the underside of the cover flange. If necessary, carefully insert a thin, flat-blade screwdriver under the cover flange — not under the gasket! — and gently lift the cover until the seal breaks. While doing so, take care not to deform the edge of the flange, damage the gasket, or mar the dewar’s finish.

4) Insert the foam block, recessed side down, on the cooler.

5) Replace the cover and tighten the screws just enough that they will not vibrate loose. Do not overtighten. This seal is intended as an air dam, not a pressure seal, so tighten the screws only enough to establish uniform , snug contact with the gasket.

4.4. DETECTOR WARM-UP

To warm a detector to room temperature, turn the MÖBIUS rear-panel power switch to the “Off” position, open the fill and vent valves to depressurize the system, and remove the detector/cryostat assembly according to Section 4.5. It takes approximately as much time to warm up a detector as it does to cool down, e.g., 12 to 36 hours or more.

4.5. REMOVING THE DETECTOR/CRYOSTAT ASSEMBLY

This procedure requires two people and a step-stool. If the detector is still cold, the person working beneath the shield stand must wear cryogenic gloves and eye protection. If removing a PopTop detector, be sure to read Section 3.3.2.1 and the PopTop user manual first.

● No center plug — As user #1 lifts the detector/cryostat from above, user #2 helps lift the assembly and pulls the transfer collar's RTV piece free of the mouth of the dewar. Optionally, user #2 may use a methanol squirt bottle on the cooling rod at the top of the fill collar, to lubricate the cooling rod is pulled up into the shield. As the detector cable is pulled up through the bottom of the shield, ensure that no connectors are pulled loose or otherwise damaged.

● With center plug — As user #1 lifts the detector/cryostat from above, user #2 helps lift the assembly of the mouth of the dewar (optionally using methanol as in the preceding paragraph). As user #1 continues lifting the detector/cryostat with one hand, user #2 pushes up one of the center plug pieces until user #1 can lift it

NOTE Any time the MÖBIUS is transported or handled in such a way that the unit might accidentally be inverted or placed on its side, it must be emptied of LN2 and the foam block shown in Figure 6 must be reinstalled under the cooler cover; otherwise the cooler may be damaged. Remove the foam when the MÖBIUS is relocated and ready for refilling.

Do not operate the MÖBIUS without the cover installed; it is required for proper air flow to the cooler.

Page 25: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

805709F / 1115 ORTEC

® MÖBIUS™ HPGe Detector Cooling System

19

out or until it falls aside into the bottom of the shield. As the detector cable is pulled up through the bottom of the shield, ensure that no connectors are pulled loose or otherwise damaged.

Place the cold detector/cryostat assembly on an uninsulated flat surface (such as a lab bench or cart), or in a cryostat stand. If you place it on a surface such as a bench, chock it so it cannot roll! If placed in contact with insulating materials such as foam, the o-ring in the detector capsule may freeze and breach the capsule vacuum.

4.6. DECOUPLING A POPTOP CAPSULE FROM THE CRYOSTAT

CAUTION Attempting to remove the PopTop capsule that has not fully warmed to room temperature will result in vacuum bellows damage. The damage may breach the vacuum, causing catastrophic failure of the detector’s crystal. Bellows damage will void your detector warranty. Refer to the PopTop manual for detailed instructions.

Page 26: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced

PROPERTY OF AMETEK, INC. – ORTEC®

ORTEC

® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115

20

[Intentionally blank]

Page 27: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced
Page 28: Condensing Liquid Nitrogen HPGe Detector Cooling System · PROPERTY OF AMETEK, INC. – ORTEC® ORTEC ® MÖBIUS™ Recycler HPGe Detector Cooling System 805709F / 1115 ii Advanced