Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
4/21/16
1
Concepts, Applications, and Requirements for Quantitative
SPECT/CT Eric C. Frey, Ph.D. ([email protected])
Division of Medical Imaging PhysicsRussell H. Morgan Department of Radiology and
Radiological ScienceJohns Hopkins University
Conflict of Interest DisclosureUnder a licensing agreement between the GE Healthcare and the Johns Hopkins University, Eric Frey is entitled to a share of royalty received by the University on sales of iterative reconstruction software used to obtain some results in this presentation.
Eric Frey is a co-founder of Radiopharmaceutical Imaging and Dosimetry, LLC. This company was founded to provide quantitative imaging and dosimetry service to developers of radiopharmaceutical therapy agents.
These interests have been disclosed and are being managed by the Johns Hopkins University in accordance with its conflict of interest policies.
4/21/16
2
Acknowledgements
• Funding: NIH Grants– R01 EB 000288– R01 CA 109234 – R01 EB 000168– U01 CA 140204
• People– Bin He, Ph.D. (now at New
York Hospital)– Yong Du, Ph.D.– Na Song, Ph.D. (Now at
Montefiore Medical Center)– Lishui Cheng– Xing Rong– Nadège Anizan, Ph.D.– George Sgouros, Ph.D.
Outline
• Introduction to SPECT• Applications of Quantitative SPECT• Requirements for Quantitative SPECT• Obstacles to Achieving Standardization
4/21/16
3
Single-Photon Emission Computed Tomography (SPECT)
γ-rays
Gamma Camera
Imaging Agent
Gam
ma
Cam
era
Gamma CameraTwo-Camera SPECT/CT Systems
X-rayTube
X-rayDetector
4/21/16
4
Gamma Cameras
Matrix Formulation ofImage Reconstruction
3D Activity Distribution Image
ProjectionMatrix
Mean Projection Data Measurement Noise(Quantum, Poisson Distributed)
Reconstructed image (estimated activity distribution image)‘Inverse’ of Projection Matrix
Projection Matrix C is • Large (~4x1012 elements)• Ill-Conditioned• Patient-dependent
4/21/16
5
Computed Tomographyp t,θ( ) = a x,t( )δ ycosθ − xsinθ − t( )dxdy∫∫
θ
y
x
p(t,θ)
t
a(x,y)
This can be inverted analytically. The solution is known as Filtered Backprojection.
Physical Image Degrading Factors
• Attenuation• Scatter• Collimator-Detector Response (CDR)
– Geometric response– Septal penetration and scatter responses
• Partial Volume Effects• Statistical Noise
}Effects of high-energyemissions
4/21/16
6
Ideal Projection from Point Source
Source
Ideal Collimator
Attenuation in Patient
Absorbed
Source
Ideal Collimator
Scattered
4/21/16
7
Effects of Attenuation
• Without attenuation compensation, sources at depth appear dimmer
• Reduces quantitative accuracy
Phantom FBP Reconstruction(no attenuation compensation)
Object Scatter
Scattered
Source
Ideal CollimatorUnscattered
MultiplyScattered
Absorbed
4/21/16
8
Quantitative Effects of Scatter
0.0500.0
1000.01500.02000.02500.03000.0
0 8 16 24 32 40 48 56 64
PhantomPrimary
Primary+Scatter
Pixel Number
Rec
onst
ruct
ed In
tens
ity
Phantom UnscatteredScatter +
Unscattered
Collimator-Detector Response (CDR)
Source
Real Collimator GeometricallyCollimated
Septal Penetration
Septal Scatter
4/21/16
9
Properties of the Full CDR
I-131 Point Source
MEGP Collimator
HEGP Collimator
Distance fromCollimator Face
5 cm 10 cm 15 cm 20 cm
30 cm
Total detected counts are a function of distance
364 keV
Effects of CDR on Spatial Frequencies
• Analagous to spatially varying low-pass filter
0.0
0.2
0.4
0.6
0.8
1.0
1.2
0 0.5 1 1.5 2
LEHRLEGP
GTF
Spatial Frquency (cm-1)
0
0.2
0.4
0.6
0.8
1
0 0.1 0.2 0.3 0.4 0.5
Rel
ativ
e M
agni
tude
Frequency (cycle/pixel)
Rectangular, νm
=0.5
Butterworth, ν
m=0.23, n=6
Hann, νm
=0.5
4/21/16
10
Effect of CDR on SPECT Images
Point SourcePhantom
FBP Reconstructionfrom Projections with
LEHR Collimator
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
0.16
0 50 100 150 200 250
PhantomReconstruction
Imag
e In
tens
ity (A
rbitr
ary
Uni
ts)
Pixel
Partial Volume EffectsPhantom Reconstruction
Spill Out
Spill In
4/21/16
11
Statistical (Quantum) Noise
Mean 2 kcounts 8 kcounts 32 kcounts 128 kcounts
Described by Poisson Distribution
Poisson Noise
• Counts in pixels are independent random variables
• Noise has equal power at all frequencies
• Image has less information at high frequencies due to CDR
0.0
0.2
0.4
0.6
0.8
1.0
1.2
0 0.5 1 1.5 2
LEHRLEGP
GTF
Spatial Frquency (cm-1)
Noise powerSpectrum: leveldepends on countsin image
4/21/16
12
23
Effect of Poisson Noise onSPECT Images
• Ramp filter used in FBP amplifies high frequencies
• Combine with low-pass to reduce high this effect
0
0.1
0.2
0.3
0.4
0.5
0 0.1 0.2 0.3 0.4 0.5
Rel
ativ
e M
agni
tude
Frequency (cycle/pixel)
Rectangular, νm
=0.5
Ramp-Butterworth, ν
m=0.23, n=6
Ramp-Hann, ν
m=0.5
24
Effect of Poisson NoiseFBP Reconstruction
• Ramp filter amplifies high frequencies• Use low pass filter to reduce high
frequency noiseNoise Free FBP Ramp
FBP w/Ramp & Butterworth
4/21/16
13
Reconstruction-Based Compensation
InitialEstimate
NewEstimate
ProjectEachAngle
ComputedProjections
MeasuredProjections
Model
UpdateEstimate
CostFunction
CompareComputed &Measured
Applications of Quantitative SPEC/CT
• Radiopharmaceutical Therapy Treatment Planning (absolute, lateral)
• Diagnosis (relative, lateral)• Response to Therapy (relative,
longitudinal)
4/21/16
14
Radiopharmaceutical Therapy (RPT)n Agents (e.g., monoclonal
antibodies, peptides, microspheres) that target tumors
n Bound to radionuclides whose emissions can kill tumor cellsn Crossfire effectn Bystander effect
n Optimal dose is patient dependent
n Treatment planning to determine administered activity
Common Therapeutic Radionuclides for TRT
RadionuclideHalflife
(hr)β-
Energy(MeV)
γ Energy (keV) (% yield)
I-131 192.5 0.6 0 364 (82), …Y-90 64 .0 2.28 none
Sm-153 46.3 0.81 103 (30), …Lu-177 161.5 0.50 208 (11), …Re-188 17.0 2.12 155 (15), …
4/21/16
15
TRT Treatment Planning Flow Chart
Administer Planning
Agent
Measure Distribution over Time
Calculate Organ and
Tumor Doses
Calculate Therapeutic
Activity
Administer Therapeutic
Quantity
Cumulated Activity and Residence Time
A : Cumulated activity (MBq ⋅sec)A0 : Injected activity (MBq)τ : Residence Time (sec)
A= A t( )dt
t∫ = A0τ
Act
ivity
A(t)
(MBq
)
A
Time t (sec)
where τ = A / A0
4/21/16
16
SPECT/CT VOI Activity Estimation
SPECT Reconstruction
& Convert to Activity
SumActivity in
VOISPECT
Proj.
CT
TotalLiverActivity
Measurement Estimate
SPECT Residence Time Estimation
ResidenceTime
CTSPECT
Proj.
Curve Fitting
SPECTActivity
Estimation
0 hr
4 hr
24 hr
72 hr
144 hr
0
0.02
0.04
0.06
0.08
0.1
0.12
0 50 100 150Time (hours)
( )0
A tA
organ
0
0.02
0.04
0.06
0.08
0.1
0.12
0 50 100 150Time (hours)
( )0
A tA
organ
4/21/16
17
In-111 QSPECT
-5%
-3%
-1%
1%
3%
5%
QSPECT
% E
rror i
n R
esid
ence
Tim
e Es
timat
e(E
stim
ated
-Tru
e)/T
rue
*100
%
Heart Lungs Liver Kidneys Spleen MarrowReconstructed usingOS-EM w/attenuation,scatter, CDR andpartial volumecompensation
50 noise realizations
Error bars showstandard deviations ofactivity estimates due to quantum noise
Precision better than accuracy for most organs
Heart Lungs Liver Kidneys Spleen Marrow
He B, Du Y, Song XY, Segars WP, Frey EC. A Monte Carlo and physical phantom evaluation of quantitative In-111SPECT. Phys Med Biol. 2005;50(17):4169-85.
-20%
-18%
-16%
-14%
-12%
-10%
-8%
-6%
-4%
-2%
0%
0 5 10 15 20 25 30 35 40 45 50# of Iterations (24 subsets/iteration)
% E
rror
in A
ctiv
ity E
stim
ates
Tumor 3 (2.2 cm, ratio 5.2)
Tumor 9 (2.2 cm, ratio 10.5)
• 2.2 cm diameter tumors
Precision for Small Objects
T9 T3
OS-EM w/attenuation, CDR and scatter compensation (no PVC)
4/21/16
18
-90%
-85%
-80%
-75%
-70%
-65%
-60%
-55%
-50%
-45%
-40%
0 5 10 15 20 25 30 35 40 45 50# 0f Iterations (24 subsets/iteration)
% E
rror
in A
ctiv
ity E
stim
ates
Tumor 4 (0.9 cm, ratio 12)
Tumor 2 (0.9 cm, ratio 11)
• 0.9 cm diameter tumors
Quantification of Very Small Objects
T2
T4
OS-EM w/attenuation, CDR and scatter compensation (no PVC)
§ 128 projection views§ Acquisition time: 40s / view
Heart Chamber Myocardium Large
SphereSmall
Sphere Background
Volume (ml) 59.7 115.3 17.5(r =1.61 cm)
5.7(r =1.11 cm)
9580
Activity(mCi) 0.562 0.471 0.136 0.044 8.15
Activity concentration(mCi/μl)
9.38 4.08 7.77 7.72 0.851
I-131 Physical PhantomPhilips Precedence SPECT/CT system with HEGP collimator
4/21/16
19
I-131 QSPECT
(%) Heart Large sphere(r = 1.61 cm
17.5 ml)
Small sphere(r = 1.11 cm
5.7 ml)AGS -15.21 -26.12 -32.72
ADS 4.75 -17.63 -25.77
ADS+Dwn+ -5.20 -21.10 -31.17
ADS+Dwn+PVC* -2.88 -15.49 -19.28
Percent errors of activity estimates for Anthropomorphic torso phantom
50 iterations 24 subsets/iteration
AGS ADS ADS + Dwn ADS+Dwn+PVE+DWN=model-based downscatter compensation*PVC=reconstruction-based PVC compensation
Y-90 QSPECT
38
• Physical phantom experiment– Elliptical phantom with 3 spheres– Philips Precedence SPECT/CT: HEGP– Acquisition time per view: 45s/view– Crystal thickness: 9.525 mm– 128 projection views over 360o
– Matrix size per view: 128*128– Pixel size: 4.664mm– VOIs defined from CT
38
4/21/16
20
Y-90 Physical Phantom Study
39
39Error = (EstimatedActivity – TrueActivity) / TrueActivity×100%
5.5 cm diametersphere
3.3 cm diameter sphere
1.5 cm diameter sphere
% Error -7.0% -9.7% -10.2%
Quantitative SPECT Imaging in Diagnosis and Monitoring of Parkinsonism
• Neurotransmission in dopaminergic systemØ Commercially available agent for SPECT (I-123
FP-CIT)
DAT D2RTatsch, Nucl. Med. Commun. (2001) 22, p819-827.
4/21/16
21
SPECT Imaging in ParkinsonismImaging assessment:Ø Visual inspectionØ Quantitative studies
üDisease degree, progression, etc.
Huang, et. al, Euro. J. Nucl. Med. (2004) 31, p155-161.
Normal Stage I PD Stage II PD
Stage III PD Stage IV PD Stage V PD
Quantitative brain SPECT Imaging
Huang, et. al, Euro. J. Nucl. Med. (2004) 31, p155-161.
Tc99m-TRODAT imaging of Parkinson’s disease in different HYS stages
4/21/16
22
Non-specific background uptake
Left putamenRight putamen
Left caudateRight caudate• GE Millennium VG/Hawkeye
(5/8” thick crystal)• LEHR Collimator• 128 views/360°, 128*128
projection w/ 0.24 cm pixels • CT attenuation maps• Manually defined VOIs using
registered MR Images• Activity concentrations:
• Bkg: 110 kBq/ml• Left Caudate: 212 kBq/ml• Left Putamen: 154 kBq/ml• Right Caudate: 1770 kBq/ml• Right Putamen: 222 kBq/ml
Accuracy of Activity Quantitation:I-123 Brain SPECT
RSD StriatalPhantom
Y. Du, B.M.W. Tsui, and E.C. Frey, "Model-based compensation for quantitative I-123 brain SPECT imaging," Phys Med Biol, 51(5): 1269-1282, 2006
Accuracy of Activity Quantitation:I-123 Brain SPECT
OS-EM w/AttenuationScatter &
CDRF Compensation Post-Reconstruction
pGTM PVC
4/21/16
23
Requirements for Quantitative SPECT/CT
• Quality Control/Calibration• Acquisition• Reconstruction/Processing
Quality Control & Calibration
• Activity meter calibration and QC• Routine Camera and CT QC• Registration of SPECT and CT• Calibration of QSPECT imaging
4/21/16
24
Calibration Factor Measurement• Planar calibration (sensitivity)
– Static image of standard source in air at known distance from camera
– Sensitivity = std. counts/(std. activity * acq. time)• Phantom-based Calibration
– Acquire SPECT study of object with known activity– Reconstruct and compute counts– Scale factor is true phantom activity/image counts– Should be consistent with planar calibration for “ideal”
reconstruction/compensation
Limitations of Planar CalibrationQuantitative Y-90 SPECT
• SPECT Calibration
Scanner Calibration Factor
GE Discovery 670
1.21-1.23
Siemens Symbia
1.15-1.18
Phantom DimensionsLarge Uniform Cylinder
20 cm diameter
Small Uniform Cylinder
4.6 cm diameter
Sphere in cold Elliptical Phantom
5.5 cmdiameter sphere in 32x20 phantom
• Planar CalibrationScanner Calibration
FactorGE Discovery 670
1.14
Siemens Symbia
1.08
4/21/16
25
Variations in Calibration Factor
1.8%
• Largest source of variation (77% of variance) was due to inter-source effects• Suggests that consistent preparation and measurement of source activity is key
Acquisition Parameters
• Collimator selection• Injected activity/acquisition time• Voxel size• Number of views• Energy windows
4/21/16
26
Reconstruction/CompensationFactor Large Object Small Object Commercially
Available
Attenuation Yes Yes Yes
Scatter Yes Yes Energy-based: yes
Model-based: limited
Geometric Response Compensation
No Yes Yes
Full CDR Compensation (High Energy)
Desirable forHE, ME
radionuclides
Desirable forHE, ME
radionuclides
No
Partial volume compensation
No Yes No
Noise Regularization
No Yes? Filtering
Obstacles to Standardization• Radioactivity measurement
– Variety of devices– Variety of radionuclides
• Compensation Methods– Many systems are SPECT-only– Variety of imaging hardware– Variety of image reconstruction and
compensation methods• Clinical Practice
– Variation in protocols– Habits
4/21/16
27
Summary
• Quantitative SPECT/CT is achievable now• There are a number of emerging clinical
applications• Limited commercial availability of state-of-
the-art reconstruction and compensation methods
• There is a need for standardization of protocols and methods