    Evaluation of CDCOM version 1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    2/16

    Problem description

    CDCOM version 1.5.2:

    As a first approximation of the position ofthe gold discs the gridlines of the phantom

    Variations in gridline position led to aninaccurate estimation.

    Combined with the size of the search areathis led to incorrect scores for some

    specific phantom images.

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    3/16

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    4/16

    CDCOM version 1.6

    CDCOM version 1.6.0:

    A template of the phantom with all discpositions is translated, rotated and skewed

    discs on the images.

    For this estimate of the position of the

    discs a relatively large search area is used For the actual detection a similar search

    area compared to version 1.5.2. is used toget similar detection results

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    5/16

    Evaluation of CDCOM 1.6

    Internally two new CDCOM versions have

    been made with different methods toestimate the position of the gold discs

    version 1.5.2

    Debug CDCOM versions have been made

    in which each step in the programme couldbe checked

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    6/16

    Evaluation of CDCOM 1.6

    A CDMAM phantom was made with only

    thick discs

    Using subsets of the visible discs the

    search area could be evaluated

    The three CDCOM versions (1.5.2 and two

    new versions) were compared to eachother using numerous CDMAM series

    Partly in debug mode

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    7/16

    Evaluation of CDCOM 1.6

    Series of images from each manufacturer

    have been used to evaluate CDCOM 1.6(between 8 to 164 images per series)

    been used (each of these comparisonsincorporated 2 to 26 phantoms)

    In total images from approximately 90CDMAM phantoms have been obtainedand used in the evaluation

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    8/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 1: DR system 1

    0.010

    0.100

    0.10 1.00

    CDCOMnew1_1766

    CDCOMold_1766

    CDCOMnew2_1766

    CDCOM1.6

    CDCOM old = CDCOM version 1.5.2.CDCOMnew1 and CDCOMnew2 are the two internally used new versions of CDCOMCDCOM 1.6 = CDCOMnew1 plus some features from CDCOMnew2, thecomplete evaluation has been redone with version 1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    9/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 2: DR system 2

    0.010

    0.100

    0.10 1.00

    CDCOMnew1_1002

    CDCOMold_1002

    CDCOMnew2_1002

    CDCOM1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    10/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 3: DR system 3

    0.010

    0.100

    0.10 1.00

    CDCOMold_1077

    CDCOM1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    11/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 4: CR system

    0.010

    0.100

    0.10 1.00

    CDCOMold_1002

    CDCOM1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    12/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 5: CR system

    0.010

    0.100

    0.10 1.00

    CDCOMnew1_1002

    CDCOMold_1002

    CDCOMnew2_1002

    CDCOM1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    13/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 6: DR system

    0.010

    0.100

    0.10 1.00

    CDCOMnew1_1002

    CDCOMold_1002

    CDCOMnew2_1002

    CDCOM1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    14/16

    Evaluation of CDCOM 1.6

    1.000

    10.000

    Example 7: DR system, phantom comparison

    0.010

    0.100

    0.10 1.00

    CDCOMold_1002

    CDCOMold_1292

    CDCOM1.6_1002

    CDCOM1.6_1292

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    15/16

    Evaluation of CDCOM version 1.6

    CDCOM has been evaluated extensively

    using data from all available systems andphantom comparisons from EUREF,

    The search area is in the correct positionin CDCOM version 1.6

  • 7/28/2019 CDCOM Evaluation [Compatibiliteitsmodus]

    16/16

    CDCOM version 1.6 was released as beta-

    version in September 2011 Potential feedback would be used to

    No feedback with incorrect scores or otherbugs have been reported

    Since January 1st 2012: CDCOM version1.6-beta has been released as officialversion (version 1.6)