Upload
vantuong
View
240
Download
4
Embed Size (px)
Citation preview
Jacqueline Roberts St James University Hospital
Leeds Teaching Hospitals NHS Trust
Calibration & Testing of an
Intego Automatic F-18
Dispensing & Injecting
System
R’nuclide Body dose - point source in air [mSv.m2/(GBq.h)]
Skin dose - contact with 5ml syringe [mSv/(MBq.h)]
Tc-99m 0.02168 0.354
F-18 0.1655 2.88
Ga-68 0.1580 31.4
I-124 0.1745 10.7
Y-90 0 43.5
NMed staff at greater risk of > 500mSv pa skin dose due to increased use of PET tracers & RNT. PET gives ~8x WB γ dose ct Tc99m, skin β+ dose 8 – 89 x greater. Intego > 20% lower wb dose, 95% lower extrem dose
Kemerink et al EJNMMI Mar 2012 39(3) 529-32 / ORAMED study 2011
Automate all or one of dispensing/injecting
Integrated systems do both – direct patient connection to dispensing system
Modular systems do this in separate parts
Different radiopharmacy requirements for both
Many varieties of design
System calibration requirements also differ
Internal calibrator checks & end-point dispensed activity checks needed
Difficulty depends on particular system design
MedRad/Bayer – Intego Integrated F18 Dispensing & Infusion System
Iride
System
Modular
Dispenser;
Injector
Tema Karl 100
Dispenser & Injector
Automated Dose
Dispenser
Automate all or one of dispensing/injecting
Integrated systems do both – direct patient connection to dispensing system
Modular systems do this in separate parts
Different radiopharmacy requirements for both
Many varieties of design
System calibration requirements also differ
Internal calibrator checks & end-point dispensed activity checks needed
Difficulty depends on particular system design
E.g. Modular unit that dispenses to a syringe – can cross-check syringe activity, applying geometrical/volume correction
E.g.2 Modular unit that dispenses in bespoke cartridge & measures with ‘calibrated’ W shield (Trasis)
E.g.3 Integrated unit that infuses direct to patient (Intego) – need to dispense into container & cross-check this with corrections
F18 Stock Vial
Precision Peristaltic Pump
Calibrator Well (Veenstra)
Waste Reservoir
Patient Line
Connection Point Saline Pump
Needles for
F18 Stock Vial
Narrow
Gauge
Tube for
Precision
Pump
Fixed -
Geometry
Plastic Coil
(sits inside
calibrator)
Port
Valve to
Patient
Infusion
Line
Waste F18
Reservoir
Saline Input
from Pump
Setup Considerations Solution
Pharmacy Issues – Sterility of multiple dispensing with one internal tubing set
Local aseptic procedure agreed for r’pharm &
users
Accuracy Calibration Before First Use & Annually – Use 3.7MBq Co57 & Cs137 sealed sources
Achieved but used pen point source of Co57 Activ just about ok
Linearity Check – Before first use & then quarterly
Passed- 5GBq F18 left overnight, auto program
Daily QC incl Constancy Source Used Cs137 source
Dispensing Accuracy & Precision NOT tested by sealed source checks – need to test end point product
Geometrical Checks (Tested to EC61145) Fixed geometry design
Temporal/Power Cycle Stability Repeat testing
Need specific RN (adjusting gain of separate current ranges?) & at least 3.7Mq, with source assay uncertainty <3%
Cs-137 - we had & just about hot enough
Co-57 – only had a penpoint marker source, made temp jig to hold it in position, low activ so gave warning message - but worked !
Important to have accurate reference activity for each source - we remeasured sources
User adjusts gain for each source until difference in measured:expected minimised (<3% to pass)
Before first use then quarterly recommended
User instructions -> 1GBq vial in QC dipper (1.7GBq actually used)
Enter pre-measured vial activity & leave overnight (15h to decay to 3.7MBq)
Automatic program logs data & displays results, relative to reference activity
Results graph displayed & retained in system memory, summary printed
Graph ?exportable via USB or wi-fi
Intego linearity -0.1% to +0.6% Atomlab -0.1 to +0.2% (both wrt 970MBq point)
Intego result quoted wrt first point 0.0 to +0.8%
Zero check
Background check
Bias current adjustment
Constancy Check – your choice of sealed source Accuracy : 1-10% Constancy : 1- 5% (Definition unclear)
Battery / Voltage Check
-5
-4
-3
-2
-1
0
1
2
3
4
5
1 2 3 4 5 6 7 8 9 10 11 12
% D
evia
tio
n fro
m R
efe
ren
ce
Session No
Intego Cs137 Accuracy Deviation (0-10% accepted range)
Talbot, JNM 2009
Dispensed vs presc so relative precision
100 Intego infusions of 555MBq
87 within <1% of presc, 1 over 5%, range -4.7 to +9%
Lecchi, EJNMMI 2012
Intego activity vs external (PET Dose) calibrator
No geom/vol corr ?
12 lots consecutively dispensed
162-405 MBq
-3.1 to +2.7% diff
EANM guidelines ±3%
Observed dose accuracy using
Medrad Intego PET Infusion
System, Talbot et al JNM 2009;
50 (Supp 2):2030
End-Point Tests
Validation of new protocol for FDG infusion
using an automated combined dispenser and
injector, Lecchi et al EJNMMI 2012; 39:1720-
29
Intego activ vs Atomlab 500 calibrator
30-40ml dispensed into bottle (need large bore shield)
Subdispense by weight (calib scales) into known geometry & volume (5ml syringe)
Apply F18 syringe corr & decay correct all data to derive dispensed bottle activity
Range of activities & concentration
Consecutive tests and repeat tests separated by days/month
Leeds End-Point Tests
11 dispensing tests in 4 sessions over 7 weeks, 4th Sept – 23rd Oct 2012
6 tests in 3 sessions before 1st patient use
Used shield for liquid I131 tx (large bore)
Limited by time, available F18 & large dilution factor (difficult to re-use F18 given fixed constraints on total activ & concentration)
Plastic bottle & 40ml vol response estim 0.94-0.96, needs defining accurately wrt vial standard
Intego
Daily variation Cs137 6.19 ± 0.02MBq
Precision ? Can’t rpt same aliquot
Non-lin ±0.6%
Accuracy wrt sealed source calib 0.1% ?
Accuracy wrt F18 =?
Atomlab
Daily varn F18 Mean/ SD 4.98 ± 0.01MBq
Precision of rpt measurement < 1%
Non-lin ±0.2%
Accuracy wrt 20 std – corrected, prec<1%
20 std accuracy wrt F18 standard =?
Redispensing to std geom < 1%
Uncertainties
Intego 100 Vs Atomlab 500 F18 Measurements
0
100
200
300
400
500
600
700
1 2 3 4 5 6 7 8 9 10 11
Dispensing Session
Dis
pe
ns
ed
Ac
tiv
ity
MB
q
Atomlab
Intego
Intego 100 Vs Atomlab 500
F18 Measurements
0
100
200
300
400
500
600
700
0 100 200 300 400 500 600 700
Atomlab Activity (MBq)
Inte
go
Ac
tiv
ity
(M
Bq
)
Intego / Atomlab Ratio
0.60
0.70
0.80
0.90
1.00
1.10
0 1 2 3 4 5 6 7 8 9 10 11 12Measurement Session
Inte
go
: A
tom
lab
Ra
tio
Intego : Atomlab ratio 0.95 to 1 (estim unc <3%) for 11 disp on 4 days over 7 weeks
Good agreement & repeat checks for similar activities consistent over time
Geom bottle corr needed if measuring directly (preferable as max meas activ & cuts dose)
Ongoing tests would assess long term stability
Which machine is closer to ‘true’ ? …
Other systems will have diff ‘features’ to workaround for testing e.g. bespoke dose cartridges need new geom corrections
2008 NIST study : 39 calibrators & traceable calibrated Ge68 source.
Found systematic differences between calib types for F18 of 6-10%, average 7%, consistent across sites
RadQual’s own audit of 100 machines shows some models even exceed 15%
Medrad may modify systems to use Ga68 accuracy source in future
Would help improve accuracy in meas of patient activ but also with QA of PET scanners & PET-based quantification, where these use F18 sources measured on-site
Ge-68 Calibrated
Reference Source
•T1/2 270.95 days
•Source calibrated against
NIST standard (for Ge68 &
F18 equivalent activity)
Belgian Nuclear Research Centre (SCK•CEN), Optimization of Radiation Protection of Medical Staff (ORAMED) http://www.oramed-fp7.eu/en/Presentations.
Extremity doses of nuclear medicine personnel: a concern Kemerink, Vanhavere, Barth & Mottaghy Eur J Nucl Med Mol Imaging. 2012 March; 39(3): 529–532. B.
E. Zimmerman, et al (2001),“Radioassays and experimental evaluation of dose calibrator settings for 18F”, Appl. Radiat. Isot., 54: 113-122.
J. T. Cessna, et al.(2008), “Radionuclide calibrator measurements of 18F in a 3 ml plastic syringe”, Appl. Radiat. Isot., 66: 988-983.
Observed dose accuracy using Medrad Intego PET Infusion System, Talbot, Mountz & Coates J Nucl Med. 2009; 50 (Supplement 2):2030
Validation of a new protocol for (18)F-FDG infusion using an automatic combined dispenser and injector system. Lecchi et al. Eur J Nucl Med Mol Imaging. 2012 Nov;39(11):1720-9.