Upload
others
View
8
Download
0
Embed Size (px)
Citation preview
Marianne Kolstad - Fluke Industrial – Nordic 1
Color ScopeMetersFluke 190C/200-series
Industrial Bus Health TestFluke 215C and 225C
2
Agenda
• ScopeMeter 190C/200-series highlights
– models and main functions
– practical measurements
• Bus health theory
• Practical measurements bus health
• Typical bus errors
3
Color ScopeMeter 190C/200C-series
•60, 100, 200 MHz bandwidth
•500 MS/s, 1GS/s, 2,5GS/ssampling speed
••1000V CAT II1000V CAT II / 600 V CAT III
•4 hours NiMHNiMH battery
4
Function highlights
• Connect-and-View™ trigger• Automatic save and replay of 100 waveforms• Isolated inputs• Pass/fail• Automatic measurements• TrendPlot™ recorder withcursors
• ScopeRecord™• FFT analysis and othermathematical functions
• bus health•……………..
Focus on Troubleshooting:
5
Intermittent problems are the most difficult ones to findand solve.
These seems to be random catastrophessent to test us, but……
at the bottom there is always a
physical cause!
Troubleshooting
6
How can we find the error without spending hourswaiting for it to show up?
Traditional: experience, intuition and luck
Fluke 190/200-series combines innovativefunctions with high speed and bandwidth
* makes it easier to find the problem* eliminates the need for luck
Fluke ScopeMeter
7
Fluke 190/200-series finds intermittent problems like:
• Errors in the equipment
– bad connections
– defective circuits
– dry weldings
– dust and corrosion
– over heating
• Power problems
– voltage variations
– transients
Intermittent problems
8
1. A multimeter updates the display every time thesignal changes (”auto range”)
The same functionality in a scope saves time andmakes it easier to use
So
lutio
n
Simple troubleshooting
9
Connect-and-ViewTM
• automatically set up andcontinuously adjust time base,voltage/div and trigger
• recognize signal patterns and givesa stable picture of even complexsignals
• two modes:
> 15 Hz for ‘normal’ use
> 1 Hz for ‘slow’ processes
10
ScopeMeter 190C/200C-series
11
2. Oops, what was that……….?
During troubleshooting you can often see abnormalsignals for a short while before they disappear…
Replay
So
lutio
n
cont. Simple troubleshooting
12
Replay
• Automatically stores the last 100 waveforms
13
Replay
REPLAY
14
ScopeMeter 190C/200C-series
15
Signals changing over time
• Long term recording might reveal changes over time
• Several parameters might be interesting to look at:
– voltage, current, frequency, temperature etc.
TrendPlotS
olution
16
TrendPlotTM
• Paperless recorder with date- and time stamp
• 2 readings
• 18000 readings per channel
• Up to 22 days
• min/max/avg foreach reading
• automatic scaling
17
ScopeMeter 190C/200C-series
18
Transients on the main
Power quality can newer be ”trusted”
And, it might be impossible to find one trigger criteria that canbe used for all circumstances
(positive or negative transient? Size? Where in the period? Outage? Dip?Swell?)
Pass/fail
So
lutio
n
19
Pass/fail
190C stores automatically up to 100 pass or 100fail waveforms in memory
A sort of “trigger on all conditions” function
Store reference signal
Compare input signal
How to do it?
20
ScopeMeter 190C/200C-series
21
Different references
Many control systems has a design that requiresmeasurements with different reference.
The control circuit might be driven by 5 V but the powerelectronic is powered from the mains.
Isolated Inputs
Solu
tion
22
• 1000 V CAT II / 600 V CAT III
– Between all inputs and ground
• For applications like
– Power electronics
– UPS’s
– ASD’s
– Power supplies
• Did you ever cut the ground leadto do floating measurement?
This is a safer way!
Isolated inputs
23
• viewing control signals and power electronicsimultaneously
• No worry situation regarding ground connections
• Important: both channel must be grounded!
Isolated inputs
24
Frequency analysis (FFT)
Some times your signals are distorted by signals ofunknown frequencies
FFTS
olution
25
FFT – with “Auto Window”
• Fast Fourier Transformation
• Shows the frequencycomponents in the signal
• …. and amplitude
• Fluke patented ”Autowindow” gives the bestamplitude and less side lobes
More side lobeswith Hamming andHanning filters
26
ScopeMeter 190C/200C-series
27
Fluke 225C and 215CAdding Bus Health Test
Verify the electrical parameters ofindustrial bus systems (“field-buses”):
CAN-bus
AS-i bus
Profibus DP
Profibus PA
Foundation Fieldbus (H1)
MOD-bus (RS-232 & RS485)
Ethernet 10Base2 (coax)
Ethernet 10BaseT (UTP)
Ethernet 100BaseT (UTP)
RS-232 (generic)
RS-485 (generic)
similar list as for 125, yet slightly different !
28
Industrial trend
•Bus systems are automatingtoday's factories and plants
– Data gathering
– Process control
– Production automation
29
Industrial trend
• Utilization of digital networksare increasing
– The traditional 4…20 mA currentloop is getting replaced by digitalcontrol systems
– Intelligence is added to sensorsand actuators
– More standardization of the buses
– Pressure to drive costs down
– Trend towards Industrial Networks,Ethernet-based, e.g. ProfiNet
30
Bus Systems
Consist of two or more components that communicate via acoded signal
Are used to automate, control and monitordifferent kind of sensors, relays, switches, PLC’er,controllers ++++
Advantages: reduce cabling better planning and flexibility conserve energy same component can be used fordifferent tasks save time and money…
31
Industrial Buses and Networks• Within Factory Automation
• Usually Multi-level systems
• Different bus systems
• avg. factory uses 4different bus system,with interfaces
• Top level: Ethernet Link
• management data
• Within machines
• CAN, DeviceNet etc.
• Lowest level: AS-i etc.
– control-lights
– reading switches
– controlling valves
– reading levelsMultiple Bus Systems
Multiple standards
32
• Physical layer measurements
• Disturbances, noise & anomalies
Electrical Signaling
• Copper and fiber cabling
• Certification and troubleshooting
• Communication networks testing
Cabling Infrastructure Network Communications
Portable handheld LAN and WAN testand analysis solutions
Industrial Network Elements
Proper cabling + good signaling + error free network communications = healthy network
Network maintenance tackled from three distinct areas
33
Cable InfrastructureConnecting devices in a network requires several physical elements:
• A suitable cable– UTP/STP– Coax– Wire diameter (AWG 18, 16, 14)
Note: some system standards dictate adedicated cable, e.g. AS-i.
•Terminations
– Some standards demand terminations at both ends of each network segment
• Junction/Terminal blocks– Star or daisy chain devices on a segment
• Each Network Standard defines– Max cable length– Impedance and/or capacitance– Termination resistance– Number of junctions or attached devices
Each element needs to be within spec and functioning properly
34
Physical Layer Signaling
• Analog waveform is used to
convey digital information in the
form of “1’s” and “0’s”
• Serial Data, for all buses
• Encoded
Example: Manchester encoding
0 gives neg. edge
1 gives pos. edge
timing and signal conditioning for AS-i
35
‘Single ended’ vs ‘Differential’ bus wiring
36
‘Single ended’ vs ‘Differential’ bus wiring -2-
37
Physical Layer Signaling
• Serial Data - Single Ended signaling
• ControlNet, RS-232 and others
• Serial Data - Differential signaling
• Single or multiple pairs
• RS-485, CAN, Ethernet etc.
• Uses Manchester or NRZ encoding
• Shared connection (‘bus’) or Point to Point
• Signaling must conform to standard
• EIA, IEC or proprietary specs
• Bus Power, provides DC voltage to transducers,actuators and controllers
• Amplitude Levels
• Signal Timing
• Noise & Distortion
38
Troubleshooting
• Yet, Industrial Buses and Networks donot always operate 100% reliable
– Harsh environments, not an office
EMI, Environmental, Pollutantscausing system failures
* Is it cabling?
* Is it electrical signaling?
* It is the protocol?
* Is it noise from the machinery?
• Need for reliable, easy to use test toolsfor troubleshooting and measurements
• The industrial networks are not part of theoffice infrastructure
maintenance needs to be done by theindustrial maintenance crew
• Many bus-standards are different fromthose used on the office network
39
Test demands
• Verification of cable infrastructure
– Is the cable installed and performing up to spec?
* Segment length, isolation, cable impedance, loaded where needed (nottoo many, not too few!), wire thickness etc.
• Verification of the electrical signals
– Are the signals present, and up to spec?
* Amplitudes, bias voltage, rise/fall-times, pulse width, aberrations etc.
• Can the signals be passed undisturbed?
– Can the signals be properly read?
* Noise pick-up, reflections, common mode signals, grounding errors etc.
Only if these criteria are met, the communication can besuccessful
40
Test of electrical parameters -‘physical layer test’
Can be done as soon as thefirst signals start to flowover the bus or network
i.e. as soon as part of thesystem has been installed
can be done even ifcommunication is difficult orunstable
that is when you need test toolsand find the root cause!
41
225C and 215C• We focus on testing signal integrity of the physical layer – electrical
signals
CAN-Dom.
CAN-Rec.
CAN Dom. highlevel range
CAN Dom. lowlevel range
Rise Time Fall Time
Pulse Width
Bias level
CAN Reces. lowlevel range
42
Industrial Fieldbus Signal
• Distortion or Anomalies
Ringing
Jitter
TransientImpulses
Noise orElectrical
interference
• Ringing•Impedance mismatch
•Jitter• Creates devicesynchronization errors
•Transients•Breaker close/open•Static discharge
• Induced noise fromhigh energy deviceslike:
•VFD’s•Hi Efficiency lightingsystems etc.
43
Bus Health - Manually measurements
• With the ScopeMeter 190 Series, you can do all the measurements toanalyze the physical layer of a fieldbus – manually:
DC bias Peak-Peak levels AC coupling to increase resolution
Use cursors and timebase Manchester encoding pulse is ½ bit Rise time, …. etc
• ….then lookup in the specific bus specification to see if each of themeasured values meets the requirements
• …then repeat the measurements to find the extreme values…
44
Fluke 225C and 215CAutomatic Measurements
Fluke 2x5C make all measurements automatically
….and automatically compare results to relevant standards• Automatically measures the key electrical signal parameters, compares to the
industry standard and records min and max values
• The Bus Health mode quickly displays parameters that need attention
• For buses with a noise levelspecification (e.g. FoundationFieldbus and Profibus PA),the instrument automaticallychecks the noise levels inthree specific bands usingadvanced filtering functions
Save timeEliminate errors
45
Fluke 225C and 215C – Automatic Bus Health testing
• All the major bus standards implemented
– limits stored in the instrument
• Easy selection thru menu
!
46
Toggle between modes
• A picture tells more than 1000 words
• Additional information about the signal quality can be derived from theEye pattern display
toggle back and forth
Quantitative information Qualitative information
47
Bus Test – parameter test screen
• Activity indicatorsmark the presence of ‘live’ signals
• Signal Levelindication
• Data rate
• Quality indicatorsquick indicationof quality for eachparameter tested:good, weak or bad
• Limit values arepreset, yet can bechanged by the user
• Eyepattern modemenu entry to switchover to high resolutionwaveform display
Present measurementvalue
History(statistics)
Limit valuesBus typeeitheruser
definedor
defaultvalues
48
Eyepattern mode
Gives a visible indication of signal quality
•Fast update rate of screen
•User may still change timebase,attenuator and delay settingslike in scope mode
•Adjustable persistence•to select ‘fade out’ time span• menu structure as in scope mode
back toparameter
screen
switch to oscilloscopemode and analyze any
specific signal detail
gives adjustablePersistence, like in
Scope-mode
Save timeGet more information
49
Fluke 225C and 215CConnection guide
The 225C shows ‘wiring’ information on screeni.e. how to make test connections for the specific bus type
50
Fluke 225C vs the Fluke 125Bus Health test
• more measurement detail present value + statistics more parameters, e.g. noise on Fieldbus
• true differential measurements high common mode signal handling
• more convenient, thanks to larger, colordisplay & helper info
• much faster response & update rate
• smaller, monochrome screen• no channel isolation
all measurements refer to ashared, common contact
gives limited ‘common mode’handling
• lower screen update rate
125
225C
(actual screen sizes)
51
Fluke 225C vs the Fluke 125 – 2 –
225C benefits over 125
• isolated channels structure allowsfor true differential measurementson differential buses such as allRS-485 based buses and CAN-bus
• second channel is then used tomeasure common mode voltage orbias over ground
see common mode signal separately
best signal detail of main signal
• larger screen allows for more datato be presented
• higher speed and sample rate ofthe 190-series makes BHTmeasurements and Eyepatternmode much, much faster
• more measurements supported
• wiring information on screen
125
225C
52
Helpful accessory: BHT190
Set of 3 ‘break-out’ adapters• makes each of the pins/wires of the
connector accessible for hooking upa probe
• gives easier and more reliableaccess to wiring of system-under-test
BHT190 is standard with the225C and 215C and
optional for Fluke 125
Get Reliable test connections
53
BHT190
Inserts between existing cable and connection box– DB-9 to DB-9 (male – female)
– M12 to M12 (male – female)
– RJ45 to RJ45, with patch cable
RJ45
M12
54
BHT190
BHT190 seenon Profibus
connector
55
Work smartSave your set up
“Named Set-ups”– store the set-ups using alpha-
numerical names, for easierreference
* name of machine it relates to
* name of test junction
* name of the test carried out
* etc.
56
ScopeMeter 215C/225C
57
Bus Health Troubleshooting
• Often used tool: Profitrace
• Problems:
– difficult to oversee the network just from one place
– needs a notebook connected
• Bus health test with Fluke 225C:
– survive harsh environments
– does not need to connect to a notebook
– easy to carry around
• Just watch this video……….
58
Typical bus errorsWrong termination
• Wrong termination is causing ringing and this can beseen at the overshoot
59
Typical bus errorsMissing termination
• No termination is giving a high data level
60
Typical bus errorsDouble termination
• Double termination is causing a to low level
61
Rockwell Example:Rockwell Example:Impedance MismatchImpedance Mismatch
Check:Check: Waveform transition points and rising/falling edges tospec. Could lead to transmission errors or timing errorsPossible Cause:Possible Cause: Bad connections, faulty input/outputcircuitry on a device connected to the segment
62
RockwellRockwell ExampleExampleConductor Damage, ReflectionsConductor Damage, Reflections
Damage toConductors
causesreflected
waves
Source ofSource ofImpedanceImpedance
CurrentCurrent
ReflectedReflectedCurrentsCurrents
63
Typical bus errorsMany users/dirt or moisture
• Too many users or dirt /moisture in a connection box isdropping the Bias Voltage
CAN-Hi
CAN-Lo
V Bias HiLimit Range
V Bias LoLimit Range
Rise Time Fall Time
Pulse Period
Ground Ref
64
Typical bus errorsAsymmetry
• Asymmetry in balanced line (Profibus, RS485) is effectingthe data pulse– duty-cycle distortion– noise margin– offset
65
Typical bus errorsTemperatur
• Temperature might effect the internal clock of a unit ,andcan change the data rate
66
Typical bus errorsEMC
• EMC might effect the rise time and cause jitter
Immunity Emission
67
Typical bus errorsEMC
• Methods for avoiding this problem:
– use separate cable trench for power and data cables
– use more than one ground connections
– longer cables pick up more noise
– correct termination
– STP cables better than UTP
• Check the integrity by looking at the eye curve
– the eye must be ”clean”
68
Rockwell Example:Rockwell Example:High Energy DischargeHigh Energy Discharge
Check:Check: Record waveform over extended periods or setup scopeto ONLY capture waveforms that exceed a defined envelope.Possible Cause:Possible Cause: Switch failures, inductive discharges in highenergy circuits on same electrical grid, or static.
50KV static discharge induced in DNet CAN lines
69
Typical bus errorsCable length
• To long cables or parasitic capacity will have effect atthe rise/fall time
70
Positioning Fluke 125, 215C and 225C
-Bus Health Connection Diagrams
2.5 GSa.s1 GSa/s25 MSa/s Real time, 5 GSa/sEquivelent time
Sample Rate
Bus Health
-Bus Health with NoiseMeasurements
CATII 1000V, CATIII 600VCATIII 600VSafety Rating
CATII 1000V, CATIII 600VNoIsolated Inputs
4 hr NiMH7 Hr NiMHBattery
-100 Screen Replay
-ScopeRecord ™
TrendPlot ™
200 MHz100 MHz40 MHzBandwidth
2 Scope & DMM2 Scope & Dual DMMInputs
ColorMonochromeDisplay
225C215C125
Electrical, electro-mechanical and electronictroubleshooting, waveform analysis and bushealth analysis and troubleshooting
Electrical and electro-mechanicalsystems & industrial busdiagnostics
Designed For
71
•Download English PDF’s from the web:www.fluke.se – ”applikationer” - ”ScopeMeter”
Application notes
72
Questions