INTRODUCTION

Designed by Godfrey N. Hounsfield to overcome the visual representation challenges in radiography and conventional tomography by collimating the X-ray beam and transmitting it only through small cross-sections of the body

Page 3: Basics in ct

G.N.HOUNSFIELD ALLAN M. CORMACK

In 1979, G.N. Hounsfield shared the Nobel Prize in Physiology & Medicine with Allan MacLeod Cormack, Physics Professor who developed solutions to mathematical problems involved in CT.

Page 4: Basics in ct

Important events

YEAR EVENTS

1969 G.N. Hounsfield developed first clinically useful CT head scanner

1971 First clinically useful CT head scanner was installed at Atkinson-Morley Hospital (England)

1972 First paper on CT presented to British Institute of Radiology by Hounsfield and Dr. Ambrose

1974 Dr. Ledley introduced the whole body CT scanner (ACTA scanner)

1979 G.N. Hounsfield shared the Nobel Prize with Allan MacLeod Cormack

Page 5: Basics in ct

C.T. scan

• Computed tomography (CT) scan machines uses X-rays, a powerful form of electromagnetic energy.

• CT combines X radiation and radiation detectors coupled with a computer to create cross sectional image of any part of the body.

Page 6: Basics in ct

Cross-sectional slicesThink like looking into a loaf of bread by cutting it into thin slices and then viewing the slices individually.

Page 7: Basics in ct

BASIC PRINCIPLE

• The internal structure of an object can be reconstructed from multiple projections of the object.

• CT scanning is a systematic collection and representation of projection data.

Page 8: Basics in ct

Comparison of CT with Conventional Radiography

• Conventional radiography suffers from the collapsing of 3D structures onto a 2D image

• CT gives accurate diagnostic information about the distribution of structures inside the body

Page 9: Basics in ct

This is the basic idea of computer aided tomography. In a CT scan machine, the X-ray beam moves all around the patient, scanning from hundreds of different angles.

Comparison of CT with Conventional Radiography

A conventional X-ray image is basically a shadow.Shadows give you an incomplete picture of an object's shape.

Page 10: Basics in ct

Comparison of CT with Conventional Radiography

Page 11: Basics in ct

Comparison of CT with Conventional Radiography

Radiographic procedure is qualitative and not quantitative

Page 12: Basics in ct

Comparison of CT with Conventional Tomography

Limitations of Conventional tomography:

1. Image blurring persists

2. Degradation of contrast due to scatter radiation

3. Problems with Film/screen combination

Page 13: Basics in ct

Comparison of CT with Conventional Radiography and Tomography

Although spatial resolution is lower in CT, it has extremely good low contrast resolution, enabling the detection of very small changes in tissue type

MATTER

LINEAR ATTENUATION COEFFICIENT

( µ )

FAT 0.194

WATER 0.222

CSF 0.227

PLASMA 0.230

RED BLOOD CELLS 0.247

Page 14: Basics in ct

GENERATIONS

• Data gathering techniques have developed in stages termed Generations.

• Scan time reduction is the predominant reason for introducing new configurations

Page 15: Basics in ct

FIRST GENERATION

• Narrow pencil beam

• Single detector per slice

• Translate –Rotate movements of Tube- detector combination

• Scan time-5min

• Designed only for evaluation of brain

Page 16: Basics in ct

I gen. CT

•The axis of rotation passed through the centre of the patient’s head

•The total number of transmission measurements = No. of linear measurements X No. of rotatory steps

= 160 X 180 = 28,800

Matrix size: 80 x 80

Scan time: 5min

Grey levels: 8

Over night image reconstruction

Page 17: Basics in ct

I Generation CT Scanner

• Head kept enclosed in a water bath

• Two side-by-side detectors• A reference detector

Page 18: Basics in ct

SECOND GENERATION

• Narrow fan beam(30-100)

• Linear detector array(30)• Translate-Rotate movements of

Tube-Detector combination• Fewer linear movements are

needed as there are more detectors to gather the data.

• Between linear movements, the gantry rotated 30o

• Only 6 times the linear movements got repeated

• Scan time~20secs

Page 19: Basics in ct

THIRD GENERATION

• Rotate(tube)-Rotate(detectors) Translatory motion is completely eliminated

• Pulsed wide fan beam(500-550)• Arc of detectors(600-900)• Detectors are perfectly aligned

with the X-Ray tube• Both Xenon and scintillation

crystal detectors can be used• Scan time< 5secs

Page 20: Basics in ct

III gen. CT scanners

Page 21: Basics in ct

FOURTH GENERATION

• Continuous wide fan beam(500-550)• Ring of detectors(> 2000)• Rotate(tube)-Fixed(detector)• X-ray tube rotates in a circle inside the

detector ring• When the tube is at predescribed

angles, the exposed detectors are read.

• Scan time< 2 secs

Page 22: Basics in ct

III Vs IV gen. CT scanners

Page 23: Basics in ct

CT Data Acquisition Components

Page 24: Basics in ct

DATA ACQUISITION

The scanning process begins with data acquisition.

Data Acquisition refers to a method by which the patient is systematically scanned by the X ray tube and detectors to collect enough information for image reconstruction.

A basic data acquisition scheme consists of • X ray tube • Filters• Collimators• Detectors

Page 25: Basics in ct

CT Gantry

Page 26: Basics in ct

CT gantry internal components

1.X-ray tube & collimator

2.Detector assembly

3.Tube controller

4.High freq. generator

5.Onboard computer

6.Stationary computer

Page 27: Basics in ct

CT gantry internal components

Page 28: Basics in ct

CT Patient Couch

Page 29: Basics in ct

CT

Page 30: Basics in ct

X-RAY TUBE

• Rotating anode type

• More heat loading and heat dissipation capabilities

• Small focal spot size (0.6mm) to improve spatial resolution

Page 31: Basics in ct

FILTERS

Compensation filter is being used

• To absorb low energy x rays

• To reduce patient dose

• To provide a more uniform beam

Page 32: Basics in ct

COLLIMATORS

• To decrease scatter radiation

• To reduce patient dose • To improve image quality• Collimator width determines

the slice thickness

Page 33: Basics in ct

DETECTORS• The detectors gather information by measuring the x-

ray transmission through the patient.• Two types:

Scintillation crystal detector

(Cadmium tungstate+ Si Photodiode) Can be used in third and fourth generation scanners

Xenon gas ionisation chamber Can be used in third generation scanners only

Page 34: Basics in ct

Scintillation crystal detector used in I & II gen. CT scanners

Page 35: Basics in ct

Scintillation crystal detector used in III and IV gen. CT scanners

Page 36: Basics in ct

Detector Cross-talk

• Detector cross talk occurs when a photon strikes a detector, is partially absorbed and then enters the adjacent detector and is detected again.

• Crosstalk produces two weak

and signals coming from two different detectors.

• Crosstalk is bad because it

decreases resolution. • Crosstalk is minimized by using

a crystal that is highly efficient in absorbing X-rays (high stopping power).

Page 37: Basics in ct

Xenon gas ionization chamber

Page 38: Basics in ct

Gas filled detector’s efficiencyGas filled detectors are less efficient than solid state detectors.

The problem can be partially overcome by the following 3 ways.• By using Xenon (z=54), the heaviest of the inert gases • By compressing the Xenon 8 to 11 atmospheres to increase its density

• By using a long chamber to increase the number of atoms along the

path of the beam.

Page 39: Basics in ct

Why are Xenon gas detectors not used in IV generation CT scanners?

• Typical size of a chamber is 1-2 mm wide, 10mm high and 8-10cm deep

• These 10mm long side plates are the reason why Xenon detectors are not in IV generation CT scanners.

Page 40: Basics in ct

Disadvantage of Xenon gas detector

Efficiency - 50 to 60%

This low efficiency is caused by two factors.• Low density of the absorbing material • Absorption of X-rays by the front window, which

is needed to contain the high pressure gas

Page 41: Basics in ct

OTHER SCAN CONFIGURATIONS

Interest in faster scan times evolves from a desire to image moving structures such as the wall of the heart and contrast material in blood vessel and heart chambers and to overcome motion artifacts due to cardiac rhythm and patient breathing .

• Dynamic Spatial Reconstructor(DSR)

• Electron beam computed tomography

Page 42: Basics in ct

DYNAMIC SPATIAL RECONSTRUCTOR

• 28 X-ray tubes• X-ray tubes are aligned with 28

light amplifiers and TV cameras that are placed behind a single curved fluorescent screen

• The gantry rotates about the patient at a rate of 50 RPM

• Data for an image acquired in about 16 ms.

• Reconstruct 250 C.S. images from each scan data

Page 43: Basics in ct

DSR

Page 44: Basics in ct

DSR

Page 45: Basics in ct

Disadvantages of DSR

• High Cost

• Mechanical motion is not eliminated

Page 46: Basics in ct

Electron Beam Computed Tomography

• Electron gun• Large Arcs of tungsten targets• Detector ring• 17 slices per second

Page 47: Basics in ct
Page 48: Basics in ct
Page 49: Basics in ct

• What is the radiation dose with EBT?• An advantage of the EBT scanner over

conventional scanners is instead of exposing the entire circumference of the body to the X-ray beam, the EBT X-ray beam enters from the back. Thus, anterior structures such as the breast and thyroid are subjected to a lesser dose of radiation (17% of the entrance skin dose). EBT scanning is usually 1/5th to 1/10th the radiation exposure as Spiral CT scanning.

Page 50: Basics in ct

DSR• The Dynamic Spatial Reconstructor •

The seminal scanner for dynamic volumetric imaging is the x-ray CT scanner known as the Dynamic Spatial Reconstructor (DSR) designed and installed at the Mayo Clinic.[1, 2] The DSR has the ability to obtain up to 240 contiguous 0.9mm thick sections in a short a time period as 1/60 second and to repeat this acquisition rate 60 times per second. In practice, the data rate is somewhat reduced from these numbers. The DSR consists of a gantry weighing approximately 17 US tons with a length of 20.5 feet and a diameter of 15 feet. 14 x-ray guns reside in a hemicylindrical configuration and aim at a juxtaposed hemicylindrical fluorescent screen. The images produced on the fluorescent screen by the firing of the x-ray guns are recorded by a bank of 14 television cameras which, until recently, were image isocons which sent analogue signals to be recorded on a bank of 8 video disc recorders. Since american television is comprised of 240 usable lines, and video rates are 60 per second, it is possible to reconstruct a cross sectional image of the body by digitizing each of the 240 television line for each of the 14 cameras and produce 240 cross sections representing 1/60 second resolution. Each stop action stack of slices in actuality represents approximately 0.1second which is the time in which all 14 x-ray guns are sequentially pulsed on. Because there were only 8 video disc recorders, every two television line were averaged to reduce the lines per camera to 120, and the data from two cameras were recorded interlaced on a single video channel. In addition, as the images are produced the gantry rotates at 15 rotations per minute. Thus, in 1/60 second, the gantry moves a degree and a half such that if the organ system of interest moves slow enough to allow for 2/60 second of scanning, it is possible to generate 28 angles of view to use in the reconstruction process. Up to a point, the more angles of view used in the reconstruction process, the better the images. Typically, at least 4/60 seconds of scanning are used to generate a good quality reconstruction. This can not only be accomplished by utilizing contiguous 1/60 second data sets, but it is possible to retrospectively gate the data together by selecting the same time point from within several physiologic cycles such as the cardiac or respiratory cycle. Physiologic signals are recorded in with the video signals to allow for this retrospective gating process. The designers of the DSR, to achieve the ability to obtain dynamic, volumetric image data sets made compromises in the image resolution such that grey scale resolution was sacrificed. To improve the image quality, the video imaging chains have been converted from image isocon cameras to charged coupled device (CCD) cameras. [76] Much larger lenses were ground by Old-Delf and tapered fiberoptics were pulled to take the images from the lens to a microchannel plate intensifier which then transmits the image to the CCD chips. The process of pulling the tapered fiberoptics introduces some twisting of the fiberoptics and therefor custom warping algorithms had to be developed for each camera. The images are digitized on each, camera, the images are unwarped, and the data is sent now to digital tape running at video rates. Although the DSR has remained a one of a kind system and represents a true tour de force, much of the current image manipulation and display associated with the massive data sets generated have served as the vanguard for data handling of images coming off of the currently commercially available scanners.

Page 51: Basics in ct

Dynamic Spatial Reconstructor

• The Dynamic Spatial Reconstructor (DSR) is a high-temporal resolution, three-dimensional (3-D) X-ray scanning device based on computed tomography (CT) principles. It was designed for investigation of some problems inherent in current diagnostic imaging techniques, and to allow quantitative studies of cardiovascular structure and function. One of the research protocols in which DSR is currently used involves studying selected pediatric patients with complex congenital heart disease. Initial results show that 3-D dynamic images can be obtained from these patients with minimal invasiveness and that these images may provide useful diagnostic information.

Page 52: Basics in ct

Dynamic Spatial Reconstructor

• Dynamic Spatial Reconstructor• The first three-dimensional volume scanning (simultaneous acquisition

of multiple contiguous slices) CT scanner with high temporal resolution (scan repetition rate of up to 60 times/sec), called the dynamic spatial reconstructor (DSR), was presented in 1980 [8].

• This machine allowed to examine the renovascular anatomy, detecting arterioles as small as 1 mm in diameter [9] and providing detailed dilution curves that showed the transit of contrast in the four zones of the renal cortex (superficial, middle, inner and juxtamedullary) and in two different areas (outer and inner) of the renal medulla. These dilution curves allowed a precise calculation of intrarenal RBF. Despite the great potential of the DSR, it was not extensively used because of its limited availability, and its high operating and maintenance costs. However, further studies on intrarenal hemodynamics were made possible when Imatron, a California company, marketed the first commercially available EBCT, which is described in the next section.

Page 53: Basics in ct

Dynamic scanning • the acquisition of the same physical image or set of images in rapid succession

such that time dependent changes (e.g. contrast enhancement, motion) can be studied. The term was originally used and most often refers to computed tomography CT scanning. Very rapid dynamic CT acquisitions have been accomplished by cine CT, which has scan times as short as 50 msec. Recently, the development of slip ring technology has resulted in sub-second scan times on conventional X-ray tube based CT scanners. With these faster scan times, many dynamic processes can be monitored. Ultimately, the number of images that can be obtained during a dynamic image study is limited by the heat loading of the X-ray tube.

• There are a variety of applications for dynamic CT scanning. These include dual-phase abdominal studies after contrast administration, cardiac imaging studies, studies of perfusion using iodinated contrast media, and measurement of cerebral blood flow using administered xenon as a diffusible agent (see xenon CT scanning and perfusion measurements). Sometimes, the term dynamic scanning is used for rapid scanning of a volume even if only a single time point is sampled. This is in some respects a misuse of the term since the images do not portray dynamic events. It derives from the increased scanning rates made available by the advent of dynamic scanning (see incremented dynamic scanning).

Page 54: Basics in ct

Dynamic Spatial Reconstructor

• The Dynamic Spatial Reconstructor (DSR) is an experimental apparatus that deserves

• mention, being the only method besides RT3D ultrasound capable of real time 3D cardiac

• imaging. It was constructed for research (and recently decommissioned) at the Mayo Clinic

• and was too expensive for general clinical use. The DSR incorporated aspects of fluoroscopy

• and CT, using multiple X-ray sources and fluoroscopic screens to gathered 3D data in real-time

• (Robb 1983). During its many years of operation, the DSR gathered unique and valuable data

• for research in cardiac dynamics and for validation of other methods of measurement.

Page 55: Basics in ct

• Cardiovascular Computed Tomography (CVCT)• Also called as ultrafast CT / Electron beam CT (EBCT)• Motion of the parts of the machine is completely eliminated• Electron gun (320cm long, 130keV)• Focusing and deflecting coils• Four 180cm diameter tungsten target arcs• 2 Rings of detectors• Electron beam scans the large target, X-rays are produced and

collimated into a 2cm wide fan beam by a set of circular collimators• The X-ray beam passes through the patient and is detected by an array

of luminescent crystals• Both the tungsten targets and the detector array cover an arc of 2100• One scan can be obtained in 50Mts• Without moving the patient, 8 contiguous tomographic images can be

obtained.

Page 56: Basics in ct

Electron Beam Computerized Tomography (EBCT)

• This instrument represents a novel concept in the use of x-ray to obtain fast tomographic scanning. In contrast to the DSR and conventional CT, EBCT has no mechanical parts (x-ray tubes and/or TV cameras) moving around the patients, resulting in lower heat production and enabling fast scanning. An electron beam, originating from an electron gun located behind the patient is magnetically deflected sequentially onto four tungsten target rings, producing eight fan beams (two from each target ring) of x-ray radiation that pass through the patient. Eight almost simultaneous renal tomographic sections can thereby be obtained, that are thicker (8 mm) than those produced by the DSR. Alternatively, consecutive 1.5, 3, or 6 mm thick tomographic slices can be obtained by using a single target ring and moving the patient table at pre-determined increments. Although its temporal resolution is lower than that offered by the DSR (50 or 100 msec/image), it is nonetheless sufficient to obtain adequate evaluation of renal function. Furthermore, because of the slightly longer scan duration and lower image noise compared to the DSR, its spatial resolution is superior [10].

• Jaschke, et al. [11, 12] were the first to demonstrate the potential of the EBCT in measuring RBF, and establish the basic principles for that calculation which correlated highly with measurements obtained with radioactive microspheres. Subsequent validation studies demonstrated the accuracy of EBCT-derived measurements of renal, cortical and medullary (compared to their in vitro) volumes [13] and perfusion (compared to electromagnetic flowmetry) within a wide range of RBF values [14], as well as changes in blood flow distribution

Page 57: Basics in ct

• The Imatron Electron Beam Tomography (EBT) Scanner• Imatron's Electron Beam Tomography (EBT) scanner combines

advanced science and technology to create an innovative diagnostic imaging system. Dramatically different from conventional (mechanical) CT scanners, Imatron's patented electron beam technology and unique design offer diagnosticians scan times as fast as 50 and 100 milliseconds. In addition to routine cross sectional imaging of all body organs, EBT is able to evaluate physiology and blood flow and to perform any other examination where speed is essential.

Utilizing proprietary EBT technology, a powerful electron beam is generated and then focused onto one of four tungsten target rings positioned beneath the patient. Each 210 degree sweep of the electron beam produces a continuous 30 degree fan beam of x-rays that pass through the patient to a stationary array of detectors which generates cross-sectional images.

Page 58: Basics in ct

• Engineering• Imatron's engineering department is an accomplished group of

scientists and engineers working to create the world's most advanced computed tomography (CT) scanner. Together, they harness and shape a powerful (650 mA at 130 kV) electron beam, then guide it through a high-vacuum chamber and around multiple tungsten-coated target rings.

These patented image acquisition techniques are at the heart of the Electron Beam Tomography (EBT) scanner, creating a scanner so fast that it can capture stop-action images of the human heart.

The vast amounts of data captured in a single breath-hold are transferred at speeds of 200 Mbs to 1 Gbs over a DICOM network to our near real-time reconstruction systems. Sophisticated programming techniques enable us to create 2D and 3D images, including virtual angiograms (fly-throughs of the arteries of the heart) and virtual colonoscopies (fly-throughs of the colon).

Page 59: Basics in ct

Radiation dose from EBT scans compared to other sources of

radiation • EBT Coronary Calcium

ScanNon-invasive Coronary AngiogramEBT Low-dose Lung ScanEBT Abdominal/Pelvis ScanBackground Sunshine Radiation in 1 yearCross country airplane tripStandard Chest X-rayStandard Abdominal X-rayStandard Spine X-ray seriesStandard Coronary AngiogramStandard Lower G.I. X-ray seriesSpiral CT Whole body scan

50 to 70 mrem80 to 120 mrem100 to 150 mrem100 to 300 mrem300 mrem2 mrem8 to 10 mrem48 mrem300 mrem500 to 1000 mrem600 mrem600 to 1000 mrem

0.5 to 0.70 mSv0.8 to 1.2 mSv0.75 to 1.5 mSv0.75 to 3.0 mSv3.0 mSv0.02 mSv0.08 to 0.1 mSv0.48 mSv3.0 mSv6.0 to 10.0 mSv6.0 mSv6.0 to 10.0 mSv

Mrem and mSv exposure varies with body size.Recommended safety limits for radiation exposure is less than 5,000 mrem per year.

Page 60: Basics in ct

Electron Beam CT

• The latest foray of technology in CT is the electron beam CT (EBCT) scanner. In EBCT an electron beam is electro-magnetically steered towards an array of tungsten X-ray anodes that are positioned circularly around the patient. The anode that was hit emits X-rays that are collimated and detected as in conventional CT. The use of an electron beam allows for very quick scanning because there are no moving parts. An entire scan can be completed in 50 to 100 milliseconds. This quick scan time makes this the only CT method which can scan the beating heart. At the present time, these machines are installed in only a few sites world-wide.

Page 61: Basics in ct

Ultrafast / Electron Beam CT Scan

• What is an ultrafast/electron beam CT (computed tomography) scan?• In conventional x-rays, a beam of energy is aimed at the body part being studied.

A plate behind the body part captures the variations of the energy beam after it passes through skin, bone, muscle, and other tissue. While much information can be obtained from a regular x-ray, specific detail about internal organs and other structures is not available. With computed tomography (also called CT or CAT scan), the x-ray beam moves in a circle around the body. This allows many different views of the same organ or structure, and provides much greater detail. The x-ray information is sent to a computer which interprets the x-ray data and displays it in two-dimensional form on a monitor. A new technology called ultrafast CT (also known as electron-beam tomography, or EBT) is now used, in some cases, to diagnose heart disease. Ultrafast CT can take multiple images of the heart within the time of a single heartbeat, thus, providing more detail about the heart's function and structures, and also greatly decreasing the amount of time required for a study.

• A three-dimensional (3D) version of ultrafast CT may be used to assess the pulmonary arteries and veins in the lungs.

• Ultrafast CT scan may also be used to evaluate selected heart defects after birth.

Page 62: Basics in ct

Electron beam computed tomography (CT)

• Exam Overview• Electron beam computed tomography (CT) scanning is

a new test that can be used to detect calcium buildup in the lining of arteries.

• Electron beam CT scanning is much faster than standard CT scanning. Electron beam CT scanning can produce an image in a fraction of a second and can take an accurate picture of an artery even while the heart is beating. Standard CT scanning is not fast enough to take pictures of a pumping heart. In standard CT, several pictures, or "slices," of the heart are taken from different angles. These pictures are then analyzed using a computer to create a three-dimensional view of the heart.

Page 63: Basics in ct

EBCT

• Why Is It Done?

• This test is used to identify calcium buildup in heart arteries, which can be a risk factor for coronary artery disease (CAD). It may be used as a screening tool to detect hardening of the arteries in people who are at high risk of developing atherosclerosis.

Page 64: Basics in ct

EBCT

• Results• Fat and calcium buildup may be seen in the arteries during an

electron beam CT scan. Aggressive treatment for atherosclerosis may be appropriate.

• If electron beam CT scanning does not show the presence of calcium buildup in the arteries, then the chances of having CAD are low. 1 Most people who have a negative angiography test result also have a negative electron beam CT scan test result. 1

• A high calcium score on an electron beam CT scan indicates a greater risk of having cardiovascular problems within the next 2 to 5 years, especially when a person also has multiple risk factors for developing coronary artery disease. 1 However, the scan does have a fairly high rate of false-positive results.

Page 65: Basics in ct

Electron Beam CT

• Advanced Body Scan of Newport is a physician owned facility that opened in June 2001 with a mission to provide the community with cutting edge diagnostic and screening medical imaging procedures. Electron Beam Tomography (EBT) is an innovative outgrowth of Computed Tomography (CT) technology specifically developed to acquire images fast enough to freeze the beating heart. This 'Ultrafast' cat scanner is open and non-claustrophobic, and the X-ray radiation used to acquire the body images is only about 20% of what you would receive from a conventional CT scanner.  The full body EBT scan delivers not much more radiation than your total annual radiation exposure from natural causes.

Page 66: Basics in ct

Electron Beam CT

• When you visit us at Advanced Body Scan of Newport, you can be assured that you will be made to feel welcome and nurtured from your initial greeting to your final medical consultation. You stay fully clothed for the exam which takes only a few minutes to complete. No shots or needles are involved.  You simply lie down on your back, hold your breath for a few seconds while the images are taken, and within minutes, you can see the results yourself on the screen.

• Advanced Body Scan of Newport offers discount scan packages to corporate customers. Our sales staff will be happy to answer any questions and customize any multiple scan package to meet your needs.

• Only GE-Imatron™ Electron Beam CT Scanners have FDA approval for Coronary Artery Calcium Scoring and CT Coronary Angiography.

Page 67: Basics in ct

Electron Beam CT

• Heart Scan: Coronary Artery Calcium Score) the EBT Heart Scan is a breakthrough in Early Detection and Cardiac Risk Assessment. Coronary artery calcium is a marker for plaque in a blood vessel. These calcium deposits form years before the first symptoms of heart disease, such as chest pain or shortness of breath, appear.

• Full Body Scan: The Body Scan includes the Heart Scan, the Lung Scan, a CT scan of your Abdomen and Pelvis and a Bone Scan, allowing us to detect liver tumors, gallstones, kidney stones, kidney tumors, aneurysms and abnormalities of the prostate or ovaries.  Before you leave you will have had a personal consultation with our health professional.

• Electron Beam CT Coronary Angiography (EBA):   Approved by the FDA in November 1999, the Electron Beam CT scanner provides an opportunity to examine your coronary arteries without the risks of conventional angiography.  Coronary angiography done on the Electron Beam CT Scanner takes only about half an hour and is reimbursed by insurance plans

Page 68: Basics in ct

Electron Beam CT

• Lung Scan: Lung Cancer strikes 150,000 Americans each year. Despite treatment advances, five-year survival for lung cancer patients remains only 14%. Only CT scans can find lung cancer at an earlier and potentially more curable stage. All current or former smokers or those exposed to carcinogens such as asbestos could benefit from this test.

• 3D QCT Bone Scan: Bone density provides the critical information you need to identify osteoporosis, a  debilitating disease that can lead to bone fractures. An essential exam for all post-menopausal women, 3D QCT Bone Densitometry is more accurate than conventional DEXA studies.

• Virtual Colonoscopy: Colorectal Cancer is the second leading cause of cancer death. This test is a new technique for visualizing colon polyps that could become cancerous. This alternative to standard Colonoscopy does not require a hospital or outpatient stay, or anaesthesia. Utilizing air and advanced image rendering, a virtual voyage through your colon is generated from the scan images.

Page 69: Basics in ct

CVCT• Cardiovascular CT• • Integrating knowledge from Philips' leadership in cardiac imaging and monitoring

systems, Brilliance CT is uniquely designed to adapt to your patients and overcome the many challenges that unpredictable heartbeat rhythms can present. Philips-patented Rate Responsive™ image acquisition technology adapts to your patients - rather than the other way around. The many intelligent technologies built into the Brilliance CT configurations not only improve departmental workflow, but the early detection of cardiac disease as well.

•  •

• Features and Benefits• Redefining patient care through automatic adjustment to variations in patient

heart rate and rhythm during the scan and subsequent image reconstruction• Least invasive, most reliable solution for assessment through accurate

determination of the most stable cardiac phase for each heart region• Achieve temporal resolutions as low as 53ms to visualize coronary anatomy and

determine stenosis as well as plaque constituents.

Page 70: Basics in ct

Toshiba CVCT

• Expanding the Limits of Cardiovascular CT

Delivering 16 0.5mm isotropic slices with 400msec rotation times, the Aquilion™ 16 CFX images the smallest vessels with confidence in the widest range of patients.

• Temporal resolution as low as 400 msec for clear, detailed cardiac evaluation

• Flexible scan speeds and SURECardio software ensure optimal imaging for variable and high heart rates

• Industry’s longest couch supports whole-body CTA without patient repositioning

• Ideal for use in cardiology, trauma, neurology and oncology • A suite of applications, called SURETechnologies, specifically

designed to provide the highest productivity and best image quality at the lowest possible dose

Page 71: Basics in ct

Cine ct

• Cine ct, • a very fast computed tomography CT scanner which can

acquire images from a single or small number of image locations in a small fraction of a second. The scanner, also named millisecond CT, ultrafast CT or electron beam CT, requires no mechanical motion to acquire the projection data. Instead of rotating an X ray tube around the object, the cine CT system sweeps an intense electron beam across a large, stationary anode target which surrounds the patient. X-rays are emitted from the point where the electrons strike the target. The X-rays transmitted through the object are measured by a stationary array of detectors. The time needed for collecting data for a single slice is 50-100 ms, although a longer scan time can be used to reduce image noise.

Page 72: Basics in ct

Cine CT• In some scanning modes, two detector rings allow the imaging of two

sections in a single scan of the X-ray target. Also, multiple target rings, scanned sequentially, allow a volume to be imaged rapidly. The system is very well suited to dynamic studies or studies where motion artefacts might be a problem (e.g. cardiac applications), and for imaging of uncooperative patients (e.g. children).

• Ideally, the X-ray target would completely surround the object, thereby allowing the source of X-rays to rotate through 360. In the ideal system, the detectors would also be on a complete ring. Mechanical constraints do not allow this arrangement. As a result, the source and detectors are not actually on the same plane, introducing complications to the reconstruction process and the possibility of artefacts. Another limitation in image quality results from the limit on the electron beam current. Although the current is high (~ 1 000 mA), the very short imaging time leads to a low mAs. Cine CT systems generally have higher noise level and lower spatial resolution than conventional CT systems, but are ideal when their high scanning speed is well suited to the clinical application. Among the possible uses are cardiac imaging with and without the use of contrast agents, lung imaging, and paediatric studies.

Page 73: Basics in ct

Cine CT

• The cine CT system has no mechanical scanning motion. In this system both the X-ray detector and the X-ray tube anode are stationary. The anode, however, is a very large semicircular ring that forms an arc around the patient scan circle, and is part of a very large, non-conventional X-ray tube. The source of X-rays is moved around the same path as a fourth generation CT scanner by steering an electron beam around the X-ray anode. Because the electron beam can be moved very rapidly, this scanner can attain very rapid image acquisition rates. In the literature, this system has been referred to variably as fifth generation and sixth generation. It has also been described as a stationary-stationary scanner. The terms millisecond CT, ultrafast CT and electron beam CT have also been used, although the latter can be confusing since the term suggests that the patient is exposed to an electron beam.

Page 74: Basics in ct

CVCT

Page 75: Basics in ct

CVCT

Page 76: Basics in ct

Applications of CVCT

• CardioVascular CT is an effective, non-invasive imaging technique for patients with congenital and acquired heart disease. Various forms of aortic disease - including aortic malpositions, dissections, and coronary aneurysms - are well suited for investigation by CV CT.

The ability of CV CT to acquire volumetric axial images in a short breath-hold enables clinicians to build accurate 3D anatomical models of the entire heart and vascular structures. In addition, CV CT can be used to screen pericardial lesions, and accurately show the extent and location of intracardiac masses and tumors adjacent to the heart.

Page 77: Basics in ct

Formation of CT images

Page 78: Basics in ct

CT Sinogram

The data acquired for one CT slice can be displayed before reconstruction. This type of

display is called a Sinogram.

Page 79: Basics in ct

What are we measuring?

The average linear attenuation coefficient (µ), between tube and detectors

Attenuation coefficient reflects the degree to which the X-ray intensity is reduced by a material

Page 80: Basics in ct

PIXEL & VOXEL

• Cross sectional layer of the body is represented as an image matrix.

• Each square of the image matrix is called pixel(picture element) and it represents tiny block of tissue called voxel (volume element)

Page 81: Basics in ct

• The linear attenuation coefficient () of each pixel is determined by :

1. Composition of the voxel

2. Thickness of the voxel

3. Quality of the radiation beam

Linear attenuation coefficient

µ

Page 82: Basics in ct
Page 83: Basics in ct

Algorithms for image reconstruction

• An algorithm is a mathematical method for solving a problem

• The linear attenuation coefficients of all the pixels in the image matrix should be determined by solving thousands of equations.

• All algorithms attempt to solve the equations as rapidly as possible without compromising accuracy.

Page 84: Basics in ct

IMAGE RECONSTRUCTION

Process of generating an image from the raw data or set of unprocessed measurements made by the imaging system.

Image reconstruction algorithms

• Back projection• Iterative methods• Analytic methods : 2D Fourier analysis

Filtered Back Projection

Page 85: Basics in ct

Simple back projection method

• The image profiles look like steps. The height of the steps is proportional to the amount of radiation that passed through the block. The steps are then assigned to a gray scale density that is proportional to their height. When the rays from the two projections are super imposed, or back projected, they produce a crude reproduction of the original object. In practice, many more projections would be added to improve image quality.

• All points in the back projected image receive density contributions from neighboring structures.

• The radio dense material severely attenuates the beam and produces localized spikes in the image profiles. Back projection of the rays from these spikes produces a star pattern. A great number of projections will obscure the star pattern, but the back ground density remains as noise to deteriorate the quality of the CT image.

Page 86: Basics in ct

Simple back projection method

Page 87: Basics in ct

Iterative method

• Assumption (for eg. all points in the matrix have same value)

• Comparison (with the measured values)• Correction (to bring the two into agreement)• Repetition (of the process until the assumed

and measured values are the same or within acceptable limits)

Page 88: Basics in ct

Iterative method

Simultaneous reconstruction• All projections for the entire matrix are calculated at the

beginning of the iteration, and all corrections are made simultaneously for each iteration.

Ray by Ray correction• One ray sum is calculated and corrected and these corrections

are incorporated into future ray sums, with the process being repeated for every ray in each iteration.

Point – by – point correction• The calculations and corrections are made for all rays passing

through one point, and these corrections are used in ensuring calculations, again with the process being repeated for every point.

Page 89: Basics in ct

Two – dimensional Fourier analysis:

• Basis: Any function of time [f(t)] or space [f(x)] can be represented by the sum of various frequencies and amplitudes of sine and cosine waves.

Page 90: Basics in ct

2D FOURIER ANALYSIS

Page 91: Basics in ct

Filtered back projection

• The image is filtered or modified to counter balance the effect of sudden density changes, which causes blurring (star – pattern). Those frequencies responsible fro blurring are eliminated to enhance more desirable frequencies.

• The density of the projected rays is adjusted that is, the inside margins of dense areas are enhanced while the centers and immediately adjacent areas are repressed. The net effect is an image more closely resembling the original object.

Page 92: Basics in ct

FILTERED BACK PROJECTION

Page 93: Basics in ct

FILTERED BACK PROJECTION

Page 94: Basics in ct

Comparison of Mathematical methods

• In terms of speed, Analytical methods are faster than iterative methods

• But with incomplete data, iterative methods are faster than analytical methods. Analytical methods does time – consuming interpolations to fill in missing data, whereas iterative methods simply average adjacent points.

Page 95: Basics in ct

CT NUMBER & HOUNSFIELD UNITThe computer calculates a relationship between the linear attenuation coefficients of the pixel and water which is given as CT number.

To image materials with higher than dense bone CT number larger than 1000 should be available

CT numbers based on a magnification constant of 1000 are Hounsfield units

Page 96: Basics in ct

Relationship between CT numbers & Gray scale

Page 97: Basics in ct

WINDOWING

• Windowing is the process in which the gray level can be manipulated using the CT numbers to provide optimum demonstration of different structures seen on the image.

• Window Width(WW) is the range of CT numbers for the gray scale & WW control alters the image contrast

• Window Level(WL) is the centre of the gray-scale image & WL control alters the image density/CT number of the tissue to be displayed

Page 98: Basics in ct

Graphic illustration of the effect of different WW & WL settings on the CT image

Page 99: Basics in ct

Image Quality in CT

Image quality is the visibility of diagnostically important structures in the CT image.

The factors that affect CT image quality are • Quantum mottle (noise)• Resolution : Spatial and contrast• Patient exposure.

The factors are all interrelated

Page 100: Basics in ct

Quantum mottle (Noise)

• Quantum mottle is the statistical fluctuations of X-photons absorbed by the detector

• The only way to decrease noise is to increase the number of photons absorbed by the detector.

• The way to increase the number of photons absorbed is to increase x-ray dose to the patient.

• Mottle becomes more visible as the accuracy of the reconstruction improves.

Page 101: Basics in ct

RESOLUTION

i) Spatial resolution

• Spatial resolution is the ability of the CT scanner to display separate images of two objects placed close together.

Page 102: Basics in ct

ii) Contrast resolution

• Contrast resolution is the ability of the CT scanner to display an image of a relatively large (2 or 3mm) object that is only slightly different in density from its surroundings.

Page 103: Basics in ct

RADIATION DOSE

• Even distribution of radiation dose to the tissues as exposures are from almost all angles.

• No overlapping of scan fields takes place.

• Exposure factors used are higher to improve spatial and contrast resolutions and to reduce noise.

Page 104: Basics in ct

Comparison of CT with Conventional Radiography

Page 105: Basics in ct

CT ARTIFACTS

Artifacts are distortions or errors in the image that are unrelated to the object scanned .

Most common artifacts in CT are• Motion artifacts• Streak artifacts• Beam hardening artifacts • Partial volume averaging artifacts• Ring artifacts

Page 106: Basics in ct

EFFECTS OF ARTIFACTS

• DETERIORATE IMAGE QUALITY

• SUBJECT INFORMATION IS LOST

• PATHOLOGICAL DETAILS ARE LOST

Page 107: Basics in ct

MOTION ARTIFACTS

Cause : Patient movement

Appearance: Blurred / streaks / ghost images

Rectification: • reduction in scan time• Clear and concise instruction to the patient • proper patient immobilization• if needed,administration of

sedatives/antiperistaltic drugs

Page 108: Basics in ct

Motion artifact

Page 109: Basics in ct

STREAK ARTIFACTS

Cause: Presence and movements of objects of very high density(contrast media, metallic implants,surgical clips)

Appearance: Streaks

REMEDY:-•Remove the offending object if possible. Use a smoothing algorithm. e.g. Standard algorithm.

Page 110: Basics in ct

StreakingStreaking occurs due to inconsistency in a small group of readings

•Partial volume

•Photon starvation

•Metal artifacts

•Patient movement

Page 111: Basics in ct

DENTURES PRODUCING

STREAK ARTIFACT

SURGICAL CLIP IN HEART

PRODUCING STREAK ARTIFACT

Page 112: Basics in ct
Page 113: Basics in ct
Page 114: Basics in ct

BEAM HARDENING ARTIFACTS

Cause :

Polyenergetic X ray spectrum(25-120kV)

APPEARANCE:-

Wide dark streak

Rectification :

• Beam hardening correction algorithm

Page 115: Basics in ct

PARTIAL VOLUME AVERAGING ARTIFACTS

• Cause: presence of tissues with highly varying absorbtion properties in a voxel.

• Rectification : Usage of Thinner CT slices

Page 116: Basics in ct
Page 117: Basics in ct

OUT OF FIELD ARTIFACT

 CAUSE:-

Scan FOV not covering the entire anatomy

APPEARANCE:-

Shading/streaks

REMEDY:- Ensure that scan field of view is larger than the object to be scanned

Page 118: Basics in ct

RING ARTIFACTS

• CAUSE : Detector failure or miscalibration of a detector

• APPEARANCE:-

Ring

• Rectification : regular quality assurance checks

Page 119: Basics in ct

RING APPEARANCE

Page 120: Basics in ct
Page 121: Basics in ct