Upload
others
View
4
Download
0
Embed Size (px)
Citation preview
Int J Lab Hem. 2017;1–10. wileyonlinelibrary.com/journal/ijlh | 1© 2017 John Wiley & Sons Ltd
Received:28March2017 | Accepted:15July2017DOI: 10.1111/ijlh.12728
O R I G I N A L A R T I C L E
Automated cell counts on CSF samples: A multicenter performance evaluation of the GloCyte system
E. A. Hod1,2 | C. Brugnara3 | M. Pilichowska4 | L. M. Sandhaus5 | H. S. Luu6 | S. K. Forest1,2 | J. C. Netterwald1 | G. M. Reynafarje2 | A. Kratz1,2
1DepartmentofPathologyandCellBiology,ColumbiaUniversityCollegeofPhysiciansandSurgeons,NewYork,NY,USA2NewYork-PresbyterianHospital,NewYork,NY,USA3DepartmentofLaboratoryMedicine,BostonChildren’sHospital,andHarvardMedicalSchool,Boston,MA,USA4TuftsMedicalCenter,Boston,MA,USA5UniversityHospitals,ClevelandMedicalCenter,Cleveland,OH,USA6Children’sMedicalCenterDallas,Dallas,TX,USA
CorrespondenceAlexanderKratz,NewYork-PresbyterianHospital,CoreLaboratory,NewYork,NY,USA.Email:[email protected]
Funding informationThisstudywasfundedbyAdvancedInstruments
AbstractObjectives:Automated cell countershave replacedmanual enumerationof cells inbloodandmostbodyfluids.However,duetotheunreliabilityofautomatedmethodsatverylowcellcounts,mostlaboratoriescontinuetoperformlabor-intensivemanualcountsonmanyorallcerebrospinalfluid(CSF)samples.ThismulticenterclinicaltrialinvestigatediftheGloCyteSystem(AdvancedInstruments,Norwood,MA),arecentlyFDA-approvedautomatedcellcounter,whichconcentratesandenumeratesredbloodcells(RBCs)andtotalnucleatedcells(TNCs),issufficientlyaccurateandpreciseatverylowcellcountstoreplaceallmanualCSFcounts.Methods:TheGloCyteSystemconcentratesCSFandstainsRBCswithfluorochrome-labeled antibodies and TNCs with nucleic acid dyes. RBCs and TNCs are thencounted by digital image analysis. Residual adult and pediatric CSF samples ob-tainedforclinicalanalysisatfivedifferentmedicalcenterswereusedforthestudy.CellcountswereperformedbythemanualhemocytometermethodandwiththeGloCyteSystem following the sameprotocol at all sites. The limits of theblank,detection,andquantitation,aswellasprecisionandaccuracyoftheGloCyte,weredetermined.Results:TheGloCytedetectedasfewas1TNC/μLand1RBC/μL,andreliablycountedaslowas3TNCs/μLand2RBCs/μL.Thetotalcoefficientofvariationwaslessthan20%.Comparisonwithcellcountsobtainedwithahemocytometershowedgoodcor-relation(>97%)betweentheGloCyteandthehemocytometer,includingatverylowcellcounts.Conclusions:TheGloCyteinstrumentisaprecise,accurate,andstablesystemtoob-tain redcell andnucleatedcell counts inCSFsamples. It allows for theautomatedenumerationofevenverylowcellnumbers,whichiscrucialforCSFanalysis.TheseresultssuggestthatGloCyteisanacceptablealternativetothemanualmethodforallCSFsamples,includingthosewithnormalcellcounts.
K E Y W O R D S
automatedcellcounters,cerebrospinalfluidtesting,laboratorytesting,redcellcounts,whitecellcounts
2 | HOD et al.
1 | INTRODUCTION
Until the invention of automated cell counters, hematology labora-tories usedmanual counting chambers to enumerate cells in bloodand body fluids.1 Such manual methods have inherent limitations,especiallyfor lowcellcounts;forexample,atacellcountof5cells/μL, the coefficient of variation (CV) of aNeubauer hemocytometer(HausserScientific,Horsham,PA)is45%.2,3WiththedevelopmentoftheCoultercounterandotherautomatedplatforms,mostlaboratoriesquicklyadoptedautomatedcellcountersfortheenumerationofbloodcells.1 Cell counts in body fluids remainedmanual for severalmoredecades,andonlyinrecentyearshavemanufacturersaddedreliablebodyfluidmodestotheirinstruments.4-12
Today,mostautomatedcellcounterscanprovidereasonablyac-curatecellcountsofalmostallbodyfluids.However,cellularanalysisofcerebrospinalfluid(CSF)samplescontinuestopresentuniquechal-lenges, because reference ranges for total nucleated cells (TNCs) inCSFare0-5cells/μLand“normal”CSFsamplesshouldnotcontainanyredbloodcells (RBCs).2Todistinguishbetweena“normal”CSFwithlessthan5TNCs/μLand0RBCs/μLandan“abnormal”CSFwithele-vatedTNCsand/orRBCs,automatedcellcountersmustbeextremelyaccurateandpreciseatverylowcellcounts.
Several instrumentplatformson themarketofferquantitationofRBCs and TNCs in CSF samples. TheADVIA120i (SiemensMedicalSolutions,Malvern,PA)hasbeenshowntoallowreliableTNCenumer-ationinCSFatverylowcellcounts.However,thepresenceofhighRBCcountscaninfluencetheaccuracyoftheTNCcount.13-15Intheevalu-ationoftheMindrayBC-6800(MindrayNorthAmerica,Mahwah,NJ)bodyfluidmodeincerebrospinalfluid,Buoroetalconcludedthattheinstrumentprovidedaneffectivealternative tomanual cell counts.16 However, they recommendedamicroscopic reviewof all cell countsbetween 4.0 and 7.0cells/μL. The Sysmex XT-4000i and XE-5000cell counters (SysmexAmerica,Lincolnshire, IL)had limitedprecisionat<20TNCs/μL.10,17ThenewerSysmexXN-1000hasalowerlimitofquantitationof5TNCs/μL,andtheminimalreportablenumberofRBCsis1000RBCs/μL.18Recently,Flemingandcolleagueshavedescribedahigh-sensitivityanalysis (hsA) researchmodeon theXN-1000sys-tem.19Thisapplicationhasalowerlimitofquantitationof10RBCs/μLand2TNCs/μL.19,20Buoroandco-workershaveinvestigatedtheuseoftheSysmexUF-1000iBodyFluidModeandfoundgoodcorrelationwithmanualTNCenumeration,withamodestoverestimationofcountsbelow30TNCs/μL.21TheuseoftheIrisiQ200BodyFluidModuleforCSFanalysis (Beckman-Coulter, Indianapolis, IN)was investigatedbyGoubardandcolleaguesaswell asWalkerandco-workers.22,23Bothgroupsfoundgoodcorrelationbetweenthemanualmethodandthere-sultsoftheiQ200.TheCVswereslightlyhigherontheIristhanonthehemocytometer.Finally,Glasserandcolleaguescomparedthe iQ200andtheLH750(Beckman-Coulter)tomanualcountsandfoundunac-ceptableratesoferroratlowercellcounts23(Table1).
Becausetheaccuracyandprecisionat lowCSFcellcounts issoclinically important,andbecausemostautomatedcellcountersmaynothavesufficientaccuracyandprecisionat lowcounts,many lab-oratories still regardmanual chamber counts as the “gold standard”
to obtain cell counts on CSF samples, especially on clear (non-bloody)specimens.20,24TheGloCyteAutomatedCellCounterSystem(AdvancedInstruments,Norwood,MA)isanewplatformthatconcen-tratesandenumeratesTNCsandRBCsusingfluorescentmicroscopyanddigitalimageanalysis.25,26Thepurposeofthismulticenterstudywastodetermine ifGloCyteRBCandTNCcountsweresufficientlyaccuratetoreplacemanualCSFcellcounts.
2 | MATERIALS AND METHODS
2.1 | The GloCyte system
The GloCyte is an automated cell counter system. It concentratesTNCsandRBCsinCSFsamplesandenumeratesthem;30μLofsam-pleand30μLofreagentcontainingfluorochrome-labeledantibodiesagainsthumanRBCsaredispensedintoa0.5-mLtube.Another30μLofsampleand30μLofreagentcontainingnucleicaciddyewithde-tergenttostainTNCsaredispensedintoasecond0.5-mLtube.Aftermixing,thetwosamplesaretransferredintoseparatecartridges.TheTNC stained sample is transferred immediately after mixing, whiletheRBCstainedsampleistransferredaftera3-minuteincubationinthetube.Vacuumsuctionisappliedtothecartridges,removingallliq-uidandpullingallcellsontoamembrane.Asemiconductorlaserandadigitalimageanalysissystemthencaptureandenumeratethefluores-centcellimages.Nomanualcellcountingisnecessary.Theinstrumentdoesnotprovideadditionalqualitativeparametersforthecellcount.Noinstrumentcalibrationisnecessary.
The samples are contained in adisposable cartridgeanddonotcomeincontactwiththeinstrument,eliminatingthepossibilityofcar-ryoverandallowingsafehandlingofpotentiallyinfectiousspecimens.
TABLE 1 PerformanceofpresentlyavailableautomatedCSFcellcountingmethods
InstrumentPerformance Issues in counting cells in CSF References
Advia120i PresenceofhighRBCcountscaninterferewiththeaccuracyoftheWBCcount
15,16
MindrayBC-6800 Microscopicreviewofallcellcountsbetween4.0and7.0recommended
17
SysmexXT-4000iandXE-5000
LimitedPrecisionatWBCcountsof<20cells/μL
11,18
SysmexXN-1000 MinimalreportablenumberofRBCsis1000cells/μL
19
SysmexXN-1000high-sensitivitymode
LowerLimitofquantitationofRBCsis10cells/μL;notFDAclearedintheUS
20,21
SysmexUF-1000i ModestoverestimationoftheWBCcountsbelow30cells/μL
7
iQ200 Unacceptableratesoferroratcountsofless50cells/μL
24
LH750 Unacceptableratesoferroratcountsofless200cells/μL
24
| 3HOD et al.
Wedidnotexperienceanyinstrumentmalfunctionduringthestudythatledtosampleloss.
2.2 | Manual method
Dependingoneachclinicalsite’shemocytometermaterialpreference,manualcountswereperformedinduplicateoneitherglassNeubauerhemocytometers (Hausser Scientific, Horsham, PA) or disposableLevy-Neubauerhemocytometers (INCYTOC-Chip,Seoul,Korea)byadding10μLtoeachchamberandcountingall9squares.27 The mean ofthetwomanualcountswasusedinthestudy.
2.3 | Study protocol
The studywas performed at five sites: Boston Children’s Hospital,Boston, Massachusetts; Columbia University Medical Center, NewYork, New York; Tufts Medical Center, Boston, Massachusetts;UniversityHospitals,ClevelandMedicalCenter,Cleveland,Ohio;andtheUniversityofTexasSouthwesternMedicalCenter,Dallas,Texas.The studywasapprovedby the InstitutionalReviewBoardofeachsite.
ResidualCSFsamplesthathadbeenobtainedforclinicalpurposesand sent for analysis to the clinical laboratory at each participatingsitewereusedforthestudy.Sampleswereincludedinthestudyafterallclinicallynecessarytestinghadbeenperformed.OnlyCSFsamplesthatwere 4hours or less after collection and that still had at least300 μLofsampleremainingqualifiedforstudyinclusion.Thesampleswere de-identified and coded. The coded samples were then splitandanalyzedusingboththeGloCytetestmethodandthegoldstan-dardhemocytometermethod.Eachsamplewasanalyzedfourtimes:twiceusingthehemocytometermethodfollowedbytworunsontheGloCyte.Everysiteanalyzedsamplesfromitsownpatientpopulation.
2.4 | Performance parameters
2.4.1 | Limit analysis
WefollowedCLSIEP17-A2forthedeterminationoftheLOB,LOD,andLOQ.28LoBtestingwasperformedintheAdvancedInstrumentslaboratorywithpurchasedcell-freehumanCSF.LoDtestingwasper-formed at TuftsMedical Center due to availability of CSF sampleswithlowcellcounts.LoQwascalculatedfromLoBandLoD.
• Limit of the Blank (LOB):TheLOBisthehighestnumberofcellsex-pectedtobemeasuredinablanksample.28-30TheLOBwasdefinedas the 95th percentile ofmeasurementsmade on blank samples(samples with no cells detected with the hemocytometer); 240countswereperformedforeachcelltype,withfiveblanksamples,usingtwoGloCytedevices.Countsweremadeover3daysforeachinstrument,withtworeagent lotsforeachblanksampleandfourreplicatesperreagentlotperday.
• Limit of Detection (LOD):TheLOD is the smallestnumberof cellsthatcanreliablybedetectedasdifferentfromzero;itwasdefined
asthevaluewhereatleast95%ofthemeasuredcountsfellabovetheLOB.LODtestingwasperformedusingCSFspecimenswithlowcellcounts(1-2cells/μL).240countswereperformedforeachcelltypeusingtwoGloCytedevices.SixspecimensweretestedforTNCand6weretestedforRBC.Eachspecimenwastested10timeswithonereagentlotandonecartridgelotand10timeswithasecondreagentlotandasecondcartridgelot.
• Limit of Quantitation (LOQ):TheLOQis the lowestcellcount inasamplethatcanbothbereliablydetectedandalsomeetsetguide-linesforprecisionandbias; itwasdefinedasthe lowestcountatwhichtheGloCyteprovidesquantitativemeasurementswithatotalerrorof20%orless.ThiscutoffwasusedbecausesurveysbytheCollegeofAmericanPathologists (CAP) indicateCVsof20%-40%formanualchambercounts.4,31TheLOQwasdeterminedutilizingthesamesamplesusedtodeterminetheLOD,sixsampleswithlowcell counts (LODsamples) for each cell type.The total error (TE)methodwasused in accordancewithCLSIEP17-A2.28 For eachsamplebyinstrumentandreagentlot,thereplicatedmeasurementswere used to estimate the total error (TE) and the percent totalerror (%TE).The“truth”wasassumedtobethemanualreferencecount foreachsample.Therootmeansquare (RMS)approachtocalculatingTEwasemployedutilizingthe%TEoftheLODcounts:
Biaswasdefinedas thedifferencebetween theautomatedcountandthegoldstandardhemocytometermethod.BasedonthedatafromtheassayrunsdetailedfortheLODexperiments,ascatterplotwasgen-eratedwith%TEontheverticalaxisandthemanualreferencecountonthehorizontalaxis.Alinearregressionlinewasfittothe%TEdata.TheLOQwasthendeterminedasthemanualcountforwhichtheaccuracymettheprespecifiedlevelof20%TE.32
2.4.2 | Receiver- operating curve analysis
Receiver-operatingcurveanalysis(ROC)analysiswasconductedusingamanualcountof5cells/μLasthecutoffbetweennormal/abnormal.Anysampleswithamanualcountoffiveorlesswereassignedzero;allotherswereassignedone.AtallGloCyteTNClevels,sensitivityand1-specificitywerecalculatedandplotted,resultingintheROCcurvepresented.
2.4.3 | Precision
• Intrarun Precision: Intrarun precision studies were performed atthreesites,using14samplesrepresentingtheclinically importantrangefortotalnucleatedcells(TNCs)and16samplesrepresentingtheclinicallyrelevantrangeforredbloodcells(RBCs),aswellas6TNCand6RBCGloCytecontrols.Eachsampleandcontrolwererun10timesbyasingleoperatorforRBCand/orTNCcounts.
• Inter-run Precision: Reproducibility was evaluated at three clin-ical sites over 20days. The sites were selected based on theavailability of resources at the sites to perform testing. Testing
TE=
√
bias2+stddev
2and%TE=
TE
manual ref count×100%
4 | HOD et al.
wasperformed twicedaily, using the same sets of controls, for20days.Eachcontrolsetwastestedinduplicate,independentlybytwooperators,ateachsite;160controlsweretestedforeachlevelofthetwocontrolsateachsite.Thetotalnumberofcontrolstestedateachsitewas640.Standarddeviationsand%CVwerecalculatedforeachfactorofinterest(residual,run,day,site,andoperator).Aprespecifiedacceptancecriterionforthisassaywassetat20%.33
2.4.4 | Method comparison (Accuracy)
Duetothelowprevalenceofabnormalsamples,asamplingplanwasdevelopedtoensurecollectionofadequatenumbersofbothnormalandabnormalsamples.EachclinicaltrialsiteinitiallytestedbothTNCand RBC for each specimen. Once a specified number of samplesweretestedforagivenresultrange,eitherTNCorRBCcountsweretested, as needed, so that a valid statistical analysis could be per-formed.Forexample,bytheendofthestudy,sampleswithhighTNCandlowRBCcountsweretestedforTNCcountsonly,asasufficientnumberof lowRBCcountshadalreadybeencollected.TherewerethereforedifferencesinthenumberofspecimenstestedforTNCandRBCcounts.
Accuracy studies were performed with 321 samples for TNCcounts(203adultspecimensand118pediatricspecimens)and422samples for RBC counts (243 adult specimens and 179 pediatricspecimens). Each sample was counted twice manually and twiceusingtheGloCyteSystem.Formethodcomparison,themeanofthetwomanualcounts,bothperformedbythesameoperator,wasusedasthegoldstandard.OnlythefirstGloCytecountwasusedfordataanalysis. Inadditionto theclinical samples,manipulatedCSFsam-ples(31samplesforTNCand41samplesforRBC)werecreatedbydilutinghumanTNCandRBCintopooledblankhumanCSF.Thesemanipulatedsampleswerenecessary toprovidedataat theupperend of the reportable range. Separate pediatric samples (pediatricvenous blood and pediatric CSF) and adult samples (adult venousbloodandadultCSF)werecreatedforseparatecollectionofpedi-atricandadultdata.Twomanual countswereperformedwith theNeubauer hemocytometer, and two countswere performed usingtheGloCyte.
CUSUMstatisticswereused to test theassumptionof linearityforthesamples.AP-valueoflessthan.05wouldhaveindicatedthattherewasasignificantdeviationfromlinearitybetweenthehemocy-tometerandtheGloCytemethod;asallP-valuesinthisanalysisweregreater than .05, therewasnoevidenceagainst theassumed linearrelationship and Passing-Bablok analysiswas appropriate to deter-minebiasbetweenresultsfromthehemocytometerandtheGloCyteSystem.
2.4.5 | Linearity/Reportable range
LinearitywasdeterminedusingmanipulatedCSFsampleswithcon-centrationsspanningthelinearrangeoftheinstrument.Atotalof14
concentration levels ranging from0 to 8000TNCs/μL and 15 con-centration levels ranging from 0 to 800000RBCs/μL were testedonthreeGloCyteinstrumentsbythreeoperators.Sampleswererunintriplicateateachlevel,resultingin126TNCcountsand135RBCcounts.
2.5 | Statistical analysis
Rversion3.3.0(2016-05-03)wasusedforLimitofBlankHistograms,andSPSS(IBM,Armonk,NY)wasusedformixed-modelanalysisforprecision studies. Mixed-model analysis was confirmed with SAS® statisticalsoftwareversion9.4(SASInstitute,Cary,NC)andMinitab(MinitabInc.,StateCollege,PA).SASsoftwarewasusedforallotheranalysis.
3 | RESULTS
3.1 | Limits of blank, detection, quantitation
The LOB for the GloCyte instrument was 0.73cells/μL for RBCsand 0.47cells/μL for TNCs. Thus,when performing cell counts onsampleswithoutcells,theGloCyteSystemprovidesacountof lessthan 1cell/μL. The LOD for theGloCyte Systemwas 0.8RBCs/μLand1.2TNCs/μL,indicatingthattheinstrumentisabletodistinguishsampleswithapproximately1cell/μLfromsampleswithlessthan1cell/μL. Finally, the LOQ for the newdevicewas 2.0RBCs/μL and2.6TNCs/μL(Figure1), indicatingthat it isabletoreliablyandpre-cisely enumerate cells in samples with at least 2RBCs/μL and/or3TNCs/μL(Table2).
3.2 | Precision
For the inter-run precision, the total CV for each control level(N=480),includingwithin-run,between-run,between-day,between-site,andbetween-operatorvariation,was<12%.Thelargestcontribu-tortothetotalCVwasthewithin-runCVof10.1%(Table3).Fortheintrarunprecision,thetotalCVof14TNCand16RBCpatientsam-pleswas<18%(Table4).AllsampleshadCVsbelow20%,whichisthestandardacceptancecriterionforthistypeofassay.4,31SampleswithlowcellcountsalsoshowedlowCVs,indicatinggoodrepeatabilityatlowcellcounts.
3.3 | Accuracy
For adult TNC counts, differences under 4.0% were observed in95% of counts up to 10000cells/μL (Figure 2). For pediatric TNCcounts,differencesupto6.8%wereobserved in95%ofcountsupto 8000cells/μL. The differences between hemocytometer countsandGloCyteenumerationsaremuchsmalleratlowercellcounts,withbothadultandpediatricsamplesshowingdifferencesgroupedcloselyaroundzerowhenmanualcountsare<100cells/μL.Theseverylowcountsincludethedecisionpointsusedbyclinicianswheninterpret-ing CSF results and are therefore key for a correct diagnosis and
| 5HOD et al.
treatmentplan.ForRBCs,differencesbetweentheGloCyteSystemandchambercountswerelargerthanforTNCs.However,thepercentdifferencewasstillsmall(5.6%),asthesemeasurementsrangeintothe900000s. Like forTNCs,differencesbetweenmanual andGloCytecountswereclosetozeroatlowcounts.
Passing-BablokanalysisshowednosignificantbiasofGloCyteTNCcountsvsthemanualmethod(Figure2andTable5).Thedatashowedequivalencebetweenthetwomethodsupto9900TNCs/μL for samples from adults and 7672TNCs/μL for pediatric sam-ples (Figure 3). Manual adult RBC counts were interchangeablewith GloCyte counts up to 901250RBCs/μL. Therewas a smallproportionalbiasinpediatricRBCcounts,undercountingRBCsbyapproximately9%upto817500RBCs/μL.Separateanalysisofpe-diatric samples fromgeneralhospitals (whichanalyzedbothadultandpediatricsamples)showednobias,indicatingthattheGloCytemethodcouldbeusedinterchangeablywiththemanualmethodinthese samples. Samples collected at dedicatedpediatric hospitalsshowed a 13% bias on theGloCyte Systemvs the hemocytome-ter.Thisindicatedthatthedifferencemaybeduetodifferentpro-cedures forhandlingsamples inpediatrichospitalsasopposed togeneralhospitals.Thebiaswasnotpresentatmanualcountsbelow100RBCs/μL,whichincludestheclinicaldecisionthresholdsusedbyclinicians.AsthepresenceofanyRBCsinCSFisconsideredab-normal,biasinhighercountsshouldnotbeanissue,andforlowercounts,GloCytecountswereequivalenttomanualcounting.Bland-Altman plots also showed high accuracy and no significant bias(Figure3).
3.4 | Clinical applications
A cutoff or 5 cells was used for ROC analysis and for sensitivity,specificity, positive predictive value (PPV), and negative predictivevalue(NPV)determination(Table6).AtallGloCyteTNClevels,sen-sitivityand1-specificitywerecalculatedandplotted,resultingintheROC curve presented in Figure4. The area under the curve (AUC)was 0.985, indicating strong agreementwith themanual diagnosis.Sensitivity, specificity, positive predictive value (PPV), and negativepredictivevalue(NPV)wereallgreaterthan0.92,whichalsoindicatesstrongagreement (Table6).Table4 shows the intrarunprecisionattheclinicallyrelevantTNClevelsof5to20cells/μL,andthelowestlevelsofRBCstested.
Cell type Reagent lot LOB: Count/μLLOD: Manual Count/μLa
LOQ: based on total error <20%, assuming manual count (gold standard) as truth
RBC 1 0.33cells/μL 0.6cells/μL 2.0cells/μL
2 0.73cells/μL 0.8cells/μL
Combined 0.53cells/μL 0.8cells/μL
TNC 1 0.33cells/μL 0.7cells/μL 2.6cells/μL
2 0.47cells/μL 1.2cells/μL
Combined 0.40cells/μL 1.2cells/μL
LOB,Limitoftheblank;LOD,limitofdetection;LOQ,limitofquantitation.a100%oftheGloCytecountsweregreaterthantheLOB.
TABLE 2 Limitoftheblank,limitofdetection,andlimitofquantitationoftheGloCytesystem
F IGURE 1 ScatterplotsofLimitofDetection(LOD)Experiments.LimitofQuantitation(LOQ)wasdeterminedasthemanualcountforwhichaccuracymetthelevelof20%totalerror
Tota
l Err
or %
0
20
40
60
80
100
120
140
160
180
200
220
RBC manual reference count (cells/uL)
LOQ Determination for RBC based on LOD samples
LOQ Determination for TNC based on LOD samples
Tota
l Err
or %
0
20
40
60
80
100
TNC manual reference count (cells/uL)
0.00 0.33 0.67 1.00 1.33 1.67 2.00 2.33
0.00 0.33 0.67 1.00 1.33 1.67 2.00 2.33 2.66 3.00
(A)
(B)
6 | HOD et al.
4 | DISCUSSION
AccurateCSFcellcountsarevitalforthediagnosisofseveraldiseases,includingsubarachnoidhemorrhage,meningealinfection,demyelinat-ingdisease,andcentralnervoussystemmalignancies.Thesesamplesrepresentasignificantpercentageofbodyfluidcellcountsperformedinhematologylaboratories.Untilrecently,manymanufacturersofau-tomatedcellcounterswereunabletoachievesufficientprecisionandaccuracyat lowcell counts tovalidate their instruments for clinicaluseforclear (nonbloody)CSFsamples.Flemingandco-workersde-scribedahigh-sensitiveanalysis(hsA)researchmodeontheSysmexXN-1000(Sysmex,Kobe,Japan)specificallyforcountingcellsinflu-idsthatcontain lowcellcounts.19Thelower limitofquantitationofthismethodis10RBCs/μLand2TNCs/μL;thus,thismethodcannotquantitateRBCsaroundthe“normal”CSFrangeofzeroRBCs.UntiltheFDA510Kclearanceof theGloCyteSystem,noautomatedcellcounterwasvalidatedandFDA510Kclearedforverylowcellcounts;thus, laboratorieshadtousemanualchambercountsformanyCSFsamples. Sandhaus and colleagues performed CSF cell counts withtheGloCyteandthehemocytometeronsamplesfrompatientswithacuteleukemia,lymphoma,malignantneoplasmsinvolvingthebrain,meningitis,andencephalitis;resultscorrelatedverywellbetweenthemanualandtheautomatedmethod.25
Thestrengthsofthisstudyaretheparticipationofmultiplemed-icalcenters,makingitpossibletoobtainlargenumbersofsamplesandspanningtheentiretyoftheclinicallysignificantmeasurementranges, including extremely low cell counts,which are crucial forCSF samples.The standardized studyprotocol ensured consistentdatacollectionatallsites.Thismulticenterstudyatfivemajormedi-calcentersshowsthattheGloCyteAutomatedCellCounterSystem
TABLE 4 Intrarunprecisionattheclinicallyrelevantthresholdsof5,10,and20TNCs/μLandat5and10RBCs/μL
Cell type Sample N Mean Median
Standard Deviation %CV
TNC 128 10 7.80 8.00 0.92 11.78
129 10 5.50 5.50 0.85 15.45
199 10 7.90 8.00 0.99 12.59
300 10 4.10 4.00 0.74 18.00
303 10 17.90 18.00 1.10 6.15
306 10 4.90 5.00 0.57 11.58
310 10 11.70 12.00 0.95 8.11
L1103 10 9.80 10.00 0.92 9.38
L1103 10 10.60 11.00 0.84 7.96
L1105 10 10.30 10.50 0.95 9.21
RBC 126 10 5.00 5.00 0.82 16.33
195 10 7.10 7.00 0.99 14.01
307 10 6.30 6.00 0.67 10.71
L1103 10 10.00 10.00 1.05 10.54
L1105 10 10.30 10.50 0.95 9.21
TABLE 3 Overallinterassayprecisionsummarystatistics
With
in ru
nBe
twee
n ru
nBe
twee
n da
yBe
twee
n si
teBe
twee
n op
erat
orTo
tal
Cell
type
Cont
rol
NM
ean
(cel
ls/μL
)SD
%CV
SD%
CVSD
%CV
SD%
CVSD
%CV
SD%
CV
TNC
Low
480
10.56
1.07
10.1
10.
000.
000.
171.58
0.41
3.86
0.25
2.40
1.18
11.2
0
High
480
122.89
7.30
5.94
0.00
0.00
0.00
0.00
3.81
3.10
1.80
1.46
8.43
6.86
RBC
Low
480
11.25
1.03
9.19
0.00
0.00
0.21
1.86
0.43
3.78
0.30
2.68
1.18
10.4
6
High
480
130.
006.
835.25
0.00
0.00
0.87
0.67
2.24
1.72
1.33
1.02
7.36
5.66
| 7HOD et al.
providesresultsinterchangeablewiththecountsobtainedbytrainedtechnologistswith a hemocytometer.With a CV below 20%, theGloCyte’sprecisionmet thestandardacceptancecriterion for thistypeofassay.20ThecorrelationbetweenGloCyteandhemocytom-eterresultswasexcellent.ThesefindingsindicatethattheGloCyte
Systemcanprovide resultswith thesameclinicalvalueasmanualmethodsevenforextremelylowcellcountsthatarewithinornearthereferenceranges.Theseresultscanbeobtainedwithinapprox-imately5minutes,withouttheneedforstaffwhohavethepatternrecognitionskillsnecessaryforcellidentification.Thisshouldallow
F IGURE 2 Passing-Bablok(PB)Correlationplotsfrommethodcomparisonforall(A)adultTNC,(B)pediatricTNC,(C)adultRBC,and(D)pediatricRBCcounts
0
0
200 000
200 000
400 000
400 000
600 000
600 000
800 000
800 000
Passing-Bablok RBC: Adult
Manual count cells/µL
Glo
Cyt
e co
unt c
ells
/µL
0
0
200 000
200 000
400 000
400 000
600 000
600 000
800 000
800 000
Passing-Bablok RBC: Pediatric
Passing-Bablok TNC: Adult Passing-Bablok TNC: Pediatric
Manual count cells/µL
Glo
Cyt
e co
unt c
ells
/µL
Manual count cells/µL
Glo
Cyt
e co
unt c
ells
/µL
Manual count cells/µL
Glo
Cyt
e co
unt c
ells
/µL
(A) (B)
(C) (D)
Passing- Bablok analysis
Cell type
Age Group
CUSUM test for linearity (P- value) Parameter Estimate
95% Confidence Interval
TNC Adult .524 Slope 1.000 (1.00,1.00)
Intercept 0.000 (0.00,0.00)
Pediatric .415 Slope 0.963 (0.91,1.00)
Intercept 0.037 (0.00,0.18)
RBC Adult .107 Slope 1.000 (0.99,1.01)
Intercept 0.000 (0.00,0.01)
Pediatric .788 Slope 0.910 (0.89,0.93)
Intercept 0.000 (-0.05,0.06)
TABLE 5 ResultsofPassing-Bablokanalysis
8 | HOD et al.
0 20 40 60 80 100Manual RBC
–40
–20
0
20
Man
ual R
BC
- G
loC
yte
RB
C 1.96SD = 13.5
–1.96SD = –12.4
Mean = 0.5
Pediatric: RBC
0 200 000 400 000 600 000 800 000Manual RBC
–200 000
–100 000
0
100 000
Man
ual R
BC
- G
loC
yte
RB
C 1.96SD = 66764.4
–1.96SD = –66307.6
Mean = 228.4
Pediatric: RBC
0 20 40 60 80 100Manual TNC
–40
–20
0
20
40
Man
ual T
NC
- G
loC
yte
TNC
1.96SD = 11.9
–1.96SD = –12.7
Mean = –0.4
Pediatric: TNC
0 2000 4000 6000 8000Manual TNC
–1000
0
1000
2000
Man
ual T
NC
- G
loC
yte
TNC
1.96SD = 1096.3
–1.96SD = –876.1
Mean = 110.1
Pediatric: TNC
0 20 40 60 80 100Manual RBC
–200
–100
0
Man
ual R
BC
- G
loC
yte
RB
C
1.96SD = 44
–1.96SD = –50.4
Mean = –3.2
Adult: RBC
0 200 000 400 000 600 000 800 000Manual RBC
–100 000
0
100 000
200 000
300 000
Man
ual R
BC
- G
loC
yte
RB
C
1.96SD = 84475.2
–1.96SD = –76655.5
Mean = 3909.8
Adult: RBC
0 20 40 60 80 100Manual TNC
–100
–50
0
Man
ual T
NC
- G
loC
yte
TNC 1.96SD = 18.5
–1.96SD = –20.9
Mean = –1.2
Adult: TNC
0 2000 4000 6000 8000 10 000Manual TNC
–1000
0
1000
Man
ual T
NC
- G
loC
yte
TNC
1.96SD = 902.5
–1.96SD = –1022.3
Mean = –59.9
Adult: TNC
(A) (B)
(C) (D)
(E) (F)
(G) (H)
F IGURE 3 Bland-AltmanPlotsforPediatricandAdultSamples.Todemonstrateaccuracyandlackofsignificantbiasatlowcellcounts,separategraphsareprovidedforcountsbetweenzeroand100cells/microliter(A,C,E,G)andforallcellcounts(B,D,F,H).Resultsarealsoshownforpediatric(A,B,C,D)andadult(E,F,G,H)specimensseparately,aslabeled
| 9HOD et al.
fortheabilitytoobtainCSFcellcountsonallshifts,includingnightsandholidays.
Ithasrecentlybeensuggestedthathemocytometersmaysoonbeobsolete intheclinical laboratory.20TheavailabilityoftheGloCyteSystem,with the ability to provide accurate andprecise counts atextremely low cell numbers,maymake that prediction a reality inthenearfuture.ThisstudyraisesthehopethattheGloCyteSystem,incombinationwith thebody fluidmodesofotherautomatedcellcountersanddigitalimageanalysis,willallowlaboratoriestodiscon-tinue allmanual cell countingmethods.However, until automatedcell counters can reliably classify all cells that are present in CSFsamples, includingneoplasticcells,therewillstillbeaneedformi-croscopicexamination,eitherbymanualmicroscopyordigitalimageanalysis.
ACKNOWLEDGEMENTS
TheauthorsthankDr.GeorgeHongandJulieMacKenzie(AdvancedInstruments, Norwood, MA) for their input and comments on thepaper,andWarrenHinkle (BioStatSolutions, Inc,FrederickMD)forprovidingthestatisticalanalysis.
REFERENCES
1. Green R, Wachsmann-Hogiu S. Development, history, and futureof automated cell counters. In: KratzA, Brugnara C, eds.Clinics in Laboratory Medicine 35.Philadelphia,PA:Elsevier;2015:1-10.
2. KarcherDS,McPhersonRA.Cerebrospinal,synovial,serousbodyflu-ids, and alternative specimens. In:McPhersonRA,PincusMR, eds.Henry’s Clinical Diagnosis and Management by Laboratory Methods,22ndedn.Philadelphia,PA:Elsevier;2011:480-506.
3. Danise P,MaconiM, RovettiA, et al. Cell counting of body fluids:comparison between three automated haematology analysers andthemanualmicroscopemethod.Int J Lab Hematol.2013;35:608-613.
4. Sandhaus LM. Body fluid cell counts by automated methods. In:Brugnara C, KratzA eds.Automated Hematology Analyzers: State of the Art. Clinics in Laboratory Medicine. Philadelphia, PA: Elsevier;2015:93-103.
5. BoerK,DeufelT,ReinhoeferM.Evaluationof theXE-5000for theautomatedanalysisofbloodcellsincerebrospinalfluid.Clin Biochem. 2009;42:684-691.
6. BuoroS,ApassitiEspositoS,AlessioM,CrippaA,OttomanoC,LippiG.Automatedcerebrospinalfluidcellcountsusingthenewbodyfluidmodeofsysmexuf-1000i.J Clin Lab Anal.2015;30:381-391.
7. ParisA,NhanT,CornetE,PerolJP,MaletM,TroussardX.Performanceevaluation of the body fluid mode on the platform Sysmex XE-5000 series automated hematology analyzer. Int J Lab Hematol. 2010;32:539-547.
8. YangD,ZhouY,ChenB.Performanceevaluationandresultcompari-sonoftheautomatedhematologyanalyzersAbbottCD3700,SysmexXE2100andCoulterLH750forcellcountsinserousfluids.Clin Chim Acta.2013;419:113-118.
9. AulesaC,MainarI,PrietoM,CobosN,GalimanyR.UseoftheAdvia120hematologyanalyzerinthedifferentialcytologicanalysisofbio-logical fluids (cerebrospinal,peritoneal,pleural,pericardial, synovial,andothers).Lab Hematol.2003;9:214-224.
10. LiA,GronlundE,BrattsandG.AutomatedwhitebloodcellcountsincerebrospinalfluidusingthebodyfluidmodeontheplatformSysmexXE-5000.Scand J Clin Lab Invest.2014;74:673-680.
11. BuoroS,AppassitiEspositoS,VavassoriM,etal.ReflextestingrulesforcellcountanddifferentiationofnucleatedelementsinpleuralandasciticfluidsonSysmexXE-5000.J Lab Autom.2016;21:297-304.
12. BuoroS,MeccaT,AzzaraG,etal.CellPopulationDataandreflextesting rules of cell analysis in pleural and ascitic fluids usingbodyfluidmodeonSysmexXN-9000.Clin Chim Acta.2016;452: 92-98.
13. BremellD,MattssonN,WallinF,etal.Automatedcerebrospinalfluidcellcount–newreferencerangesandevaluationofitsclinicaluseincentralnervoussysteminfections.Clin Biochem.2014;47:25-30.
14. TanadaH,IkemotoT,MasutaniR,TanakaH,TakuboT.Evaluationofthe automated hematology analyzerADVIA(R) 120 for cerebrospi-nal fluid analysis and usage of unique hemolysis reagent. Int J Lab Hematol.2014;36:83-91.
15. MahieuS,VertessenF,VanderPlankenM..EvaluationofADVIA120CSFassay(Bayer)vs.chambercountingofcerebrospinalfluidspeci-mens.Clin Lab Haematol.2004;26:195-199.
16. Buoro S, SeghezziM,MeccaT,VavassoriM, CrippaA, LaGioiaA.EvaluationofMindrayBC-6800bodyfluidmodeforautomatedcere-brospinalfluidcellcounting.Clin Chem Lab Med.2016;54:1799-1810.
17. Kleine TO, Nebe CT, Lower C, GeilenkeuserWJ, Dorn-Beineke A.Cellanalysisincerebrospinalfluid(CSF)usingSysmex(R)hematology
TABLE 6 Sensitivity,specificity,positivepredictivevalue,andnegativepredictivevalueoftheGloCyteSystem
Frequency of GloCyte counts
Cell type ≤5 >5 TP TN FN FP Sensitivity Specificity PPV NPV
TNC 176 176 163 164 12 13 0.931 0.927 0.926 0.932
Amanualcountof5wasusedasthecutoffbetweennormalandabnormal.NPV,NegativePredictiveValue;PPV,PositivePredictiveValue.
F IGURE 4 Receiver-operatingcharacteristic(ROC)curvefortotalnucleatedcells(TNC)
0.00.0
0.2
0.2
0.4
0.4Sen
sitiv
ity
1 - Specificity0.6
0.6
0.8
0.8
AUC = 0.9852
1.0
1.0
10 | HOD et al.
analyzersXT-4000iandXE-5000:evaluationwithCSFcontrolsoftheJointGermanSocietyforClinicalChemistryandLaboratoryMedicine(DGKL).Cytometry A.2012;81:255-264.
18. Fleming C, Brouwer R, Lindemans J, de Jonge R.Validation of thebodyfluidmoduleonthenewSysmexXN-1000forcountingbloodcellsincerebrospinalfluidandotherbodyfluids.Clin Chem Lab Med. 2012;50:1791-1798.
19. FlemingC,RusscherH,BrouwerR,LindemansJ,deJongeR.EvaluationofSysmexXN-1000High-SensitiveAnalysis(hsA)ResearchModeforCountingandDifferentiatingCells inCerebrospinalFluid.Am J Clin Pathol.2016;145:299-307.
20. Sandhaus LM. Is theHemocytometerObsolete forBodyFluidCellCounting?Am J Clin Pathol.2016;145:294-295.
21. Buoro S, Apassiti Esposito S, Alessio M, Crippa A, OttomanoC, Lippi G. Automated Cerebrospinal Fluid Cell Counts UsingtheNewBody FluidMode of SysmexUF-1000i. J Clin Lab Anal. 2016;30:381-391.
22. Walker TJ, Nelson LD, Dunphy BW, Anderson DM, Kickler TS.Comparative evaluation of the Iris iQ200 body fluid module withmanualhemacytometercount.Am J Clin Pathol.2009;131:333-338.
23. GlasserL,MurphyCA,MachanJT.Theclinicalreliabilityofautomatedcerebrospinal fluidcell countson theBeckman-CoulterLH750andIrisiQ200.Am J Clin Pathol.2009;131:58-63.
24. DeMott WR, Skikne BS. Cerebrospinal Fluid Analysis. In: JakobsDS, DeMottWR, Oxley DK, eds. Jacobs & DeMott Laboratory Test Handbook.Hudson,Ohio:Lexi-Comp;2001:416-418.
25. SandhausLM,DillmanCA,HinkleWP,MacKenzieJM,HongG.Anewautomatedtechnologyforcerebrospinalfluidcellcounts:comparisonof accuracy and clinical impact of glocyte, sysmex xn, andmanualmethods.Am J Clin Pathol.2017;147:507-514.
26. CSF Cell Counter: Advanced Instruments,Norwood,MA;2017[Availablefrom:http://www.aicompanies.com/product/glocyte/-features.
27. (CLSI) CaLSI. Body Fluid Analysis for Cellular Composition; Approved Guideline; CLSI Document H56-A.Wayne,PA;2006.
28. CLSI. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline, 2nd edition; EP 17-A2. Wayne,PA:CLSI;2012.
29. BournerG,DelaSalleB,GeorgeT,etal.ICSHguidelinesforthever-ificationandperformanceofautomatedcellcountersforbodyfluids.Int J Lab Hematol.2014;36:598-612.
30. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline, 2nd edition. CLSI document EP17-A2,(2012).
31. College of American Pathologists. Hemocytometer Fluid Count, Participant Summary.Northfield,IL:CollegeofAmericanPathologists;2015.
32. MacdonaldR.Qualityassessmentofquantitativeanalyticalresultsinlaboratorymedicinebyrootmeansquareofmeasurementdeviation.J Lab Med.2006;30:111-117.
33. Evaluation of Precision of Quantitative Measurement Procedures (EP05-A3),(2014).
How to cite this article:HodEA,BrugnaraC,PilichowskaM,etal.AutomatedcellcountsonCSFsamples:AmulticenterperformanceevaluationoftheGloCytesystem.Int J Lab Hem. 2017;00:1–10. https://doi.org/10.1111/ijlh.12728