1
Sysmex XN Analysers – A Novel Modular Blood Cell Counting System C Briggs, I Longair, P Kumar & SJ Machin Department of Haematology, University College London Hospitals, 60 Whitfield Street, London W1T 4EU, UK Introduction The newly released Sysmex XN haematology blood cell analyser offers a new con- cept of cell analysis using an integrated modular system which the user can design to their requirement. The instrument has a 34% smaller footprint than previous XE analysers. There is a novel sampler system which allows analysis of samples on both modules at the same time and automatic reflex testing. Figure 1. XN-2000 system Model flexibility. Basic Performance DIFF channel (WDF) WNR channel (WBC plus NRBC on all samples) RBC/PLT & HGB channels Enhanced Options RET channel WPC channel PLT-F channel Body Fluid mode Low WBC mode, counts of less than 0.5 x 10 9 /L extended count for precision and a diff reported. Automatic reflex testing Low WBC, low or abnormal platelets, abnormal lymphs or blast flag can all be auto- maticaly re-tested, no intervention needed. Options include: WPC – White Cell Precursor Channel This further differentiates blasts and abnormal lymphs which were generated by the WDF channel, reducing false positive rate. PLT-F – Platelet count Immature platelet count (IPF) percentage and absolute count. IPF is now only produced by PLT-F channel, no longer RET channel (as on XE-series). If the red cell or impedance platelet histograms are abnormal or if the platelet count is below the numerical cut-off can be set by the user then a fluorescent platelet count is automatically performed. PLT-F is more accurate and precise (compared to the ICSH reference immunological platelet counting method). Low white cell count mode The setting used was < 0.5 x 10 9 /L to trigger an extended count of three times longer provides more precise results with an accurate differential. Aim The aim of this study was to evaluate the performance of the new XN compared to the XE-2100 using both normal (one third) and abnormal specimens. Residual sam- ples from the routine laboratory were selected after all testing had been complete. Samples 390 samples were evaluated on both instruments and by the CLSI reference manual differential reference method. 30 normal samples were analysed on the XN and a manual differential performed and then diluted to a WBC < 0.5 x 10 9 /L and reanalysed on XN using the WPC mode. The diluted blood counts were compared to the manual differential. 185 samples were used to compare the PLT-F count to the ICSH flow cytometric method (67 samples < 20 x 10 9 /L). 1000 samples were analysed on both instruments to determine the time taken for analysis and the number of blood films that would need to be examined, samples were selected to mimic our daily workload. Results 1. Good correlation with the XE-2100 was found for all parameters. 2. The leucocyte differential and NRBC compared very well to the manual 400 cell differential. 3. False positive flags for blasts and abnormal lymphocytes are reduced signifi- cantly (20%) without an increase in false negatives. The atypical lymphocyte flag is also reduced by 20% even without the use of the WPC channel. 4. The low WBC mode produced differentials on WBC < 0.5 x 10 9 /L with no signifi- cant difference from undiluted blood (figure 2). Figure 2. Correlation of leucocyte differentials from the XN compared to a 400 cell manual differential on samples with WBC < 0.5 x 10 9 /L 5. Correlation to the ICSH flow cytometric platelet counting method was impres- sive, particularly on counts at the platelet transfusion threshold (figure 3). Figure 3. XN-Fluorescence platelet [PLT-F] Count Selected by initial Platelet Impedance Count < 20 x 10 9 /L [n=67] PLT-O = Optical Platelet count on the XE-2100 Workflow Study (n=1002) XE-2100 XN with WPC-ch Blood films 199 101 Reflex analysis 75 112 TAT [min] analysis 611 553 TAT [min] analysis + SP 710 604 Table 1. Workflow statistics from 1000 routine samples. TAT = turn around time. SP= Slide preparation unit. Conclusions Excellent general performance, linearity, stability, carry over and reproducibility. PLT-F excellent correlation with ICSH flow reference method. Improved accuracy in thrombocytopenic samples, especially at platelet transfusion threshold values. Reduced analyser TAT time due to reflex operation. Reduced blood films (49%). The flagging on the XN for blasts, abnormal lymphocytes/lymphoblast and atypical lymphocyte flags show marked improvement with fewer false positives with no loss of sensitivity. University College London Hospitals NHS Foundation Trust Neut % y = 0.9426x + 0.8139 R 2 = 0.9683 0 20 40 60 80 100 0 20 40 60 80 100 Manual Neut XN whole blood Neut Lymph % y = 0.9271x + 1.4805 R 2 = 0.948 0 20 40 60 80 100 0 20 40 60 80 100 Manual Lymph XN whole blood Lymph Mono % y = 1.0422x + 1.8556 R 2 = 0.8767 0 10 20 30 40 50 0 10 20 30 40 50 Manual Mono XN whole blood Mono 0 75 150 225 300 XE-2100 XN-2000 Blood smears 500 530 560 590 620 XE-2100 XN-2000 TAT [min] analysis 550 600 650 700 750 XE-2100 XN-2000 TAT [min] analysis + SP

Sysmex XN Analysers – A Novel Modular Blood Cell Counting … · 2016. 11. 3. · Sysmex XN Analysers – A Novel Modular Blood Cell Counting System C Briggs, I Longair, P Kumar

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Sysmex XN Analysers – A Novel Modular Blood Cell Counting … · 2016. 11. 3. · Sysmex XN Analysers – A Novel Modular Blood Cell Counting System C Briggs, I Longair, P Kumar

Sysmex XN Analysers – A Novel Modular Blood Cell Counting System

C Briggs, I Longair, P Kumar & SJ MachinDepartment of Haematology, University College London Hospitals, 60 Whitfi eld Street, London W1T 4EU, UK

Introduction The newly released Sysmex XN haematology blood cell analyser offers a new con-cept of cell analysis using an integrated modular system which the user can design to their requirement. The instrument has a 34% smaller footprint than previous XE analysers. There is a novel sampler system which allows analysis of samples on both modules at the same time and automatic refl ex testing.

Figure 1. XN-2000 system

Model � exibility. Basic Performance

■ DIFF channel (WDF)

■ WNR channel (WBC plus NRBC on all samples)

■ RBC/PLT & HGB channels

Enhanced Options

■ RET channel

■ WPC channel

■ PLT-F channel

■ Body Fluid mode

■ Low WBC mode, counts of less than 0.5 x 109/L extended count for precision and a diff reported.

Automatic re� ex testing

Low WBC, low or abnormal platelets, abnormal lymphs or blast fl ag can all be auto-maticaly re-tested, no intervention needed. Options include:

WPC – White Cell Precursor Channel

This further differentiates blasts and abnormal lymphs which were generated by the WDF channel, reducing false positive rate.

PLT-F – Platelet count

■ Immature platelet count (IPF) percentage and absolute count.

■ IPF is now only produced by PLT-F channel, no longer RET channel (as on XE-series).

■ If the red cell or impedance platelet histograms are abnormal or if the platelet count is below the numerical cut-off can be set by the user then a fl uorescent platelet count is automatically performed. PLT-F is more accurate and precise (compared to the ICSH reference immunological platelet counting method).

Low white cell count mode

The setting used was < 0.5 x 109/L to trigger an extended count of three times longer provides more precise results with an accurate differential.

Aim The aim of this study was to evaluate the performance of the new XN compared to the XE-2100 using both normal (one third) and abnormal specimens. Residual sam-ples from the routine laboratory were selected after all testing had been complete.

Samples

390 samples were evaluated on both instruments and by the CLSI reference manual differential reference method.

30 normal samples were analysed on the XN and a manual differential performed and then diluted to a WBC < 0.5 x 109/L and reanalysed on XN using the WPC mode. The diluted blood counts were compared to the manual differential.

185 samples were used to compare the PLT-F count to the ICSH fl ow cytometric method (67 samples < 20 x 109/L).

1000 samples were analysed on both instruments to determine the time taken for analysis and the number of blood fi lms that would need to be examined, samples were selected to mimic our daily workload.

Results

1. Good correlation with the XE-2100 was found for all parameters.

2. The leucocyte differential and NRBC compared very well to the manual 400 cell differential.

3. False positive fl ags for blasts and abnormal lymphocytes are reduced signifi -cantly (20%) without an increase in false negatives. The atypical lymphocyte fl ag is also reduced by 20% even without the use of the WPC channel.

4. The low WBC mode produced differentials on WBC < 0.5 x 109/L with no signifi -cant difference from undiluted blood (fi gure 2).

Figure 2. Correlation of leucocyte differentials from the XN compared to a 400 cell manual differential on samples with WBC < 0.5 x 109/L

5. Correlation to the ICSH fl ow cytometric platelet counting method was impres-sive, particularly on counts at the platelet transfusion threshold (fi gure 3).

Figure 3. XN-Fluorescence platelet [PLT-F] Count Selected by initial Platelet Impedance Count < 20 x 109/L [n=67]PLT-O = Optical Platelet count on the XE-2100

Work� ow Study

(n=1002) XE-2100 XN with WPC-ch

Blood fi lms 199 101

Refl ex analysis 75 112

TAT [min] analysis 611 553

TAT [min] analysis + SP 710 604

Table 1. Workfl ow statistics from 1000 routine samples. TAT = turn around time. SP= Slide preparation unit.

Conclusions

■ Excellent general performance, linearity, stability, carry over and reproducibility.

■ PLT-F excellent correlation with ICSH fl ow reference method. Improved accuracy in thrombocytopenic samples, especially at platelet transfusion threshold values.

■ Reduced analyser TAT time due to refl ex operation.

■ Reduced blood fi lms (49%).

■ The fl agging on the XN for blasts, abnormal lymphocytes/lymphoblast and atypical lymphocyte fl ags show marked improvement with fewer false positives with no loss of sensitivity.

University College London HospitalsNHS Foundation Trust

Neut %

y = 0.9426x + 0.8139R2 = 0.9683

0

20

40

60

80

100

0 20 40 60 80 100Manual Neut

XN w

hole

blo

od N

eut

Lymph %

y = 0.9271x + 1.4805R2 = 0.948

0

20

40

60

80

100

0 20 40 60 80 100Manual Lymph

XN w

hole

blo

od L

ymph

Mono %

y = 1.0422x + 1.8556R2 = 0.8767

0

10

20

30

40

50

0 10 20 30 40 50Manual Mono

XN w

hole

blo

od M

ono

0

75

150

225

300

XE-2100 XN-2000

Blood smears

500

530

560

590

620

XE-2100 XN-2000

TAT [min] analysis

550

600

650

700

750

XE-2100 XN-2000

TAT [min] analysis + SP