8/9/2019 Audiometers audiometric booths.PDF http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 1/116  5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA  Tel +1 (610) 825-6000  Fax +1 (610) 834-1275  Web www.ecri.org  E-mail [email protected] UMDNS Information This Product Comparison covers the following device terms and product codes as listed in ECRI Institute’s Universal Medical Device Nomenclature System™ (UMDNS™):   Audiometers [10-228]   Audiometric Booths [10-229] Audiometers; Audiometric Booths Scope of this Product Comparison This Product Comparison covers pure-tone audiometers used in routine testing and screening for hearing threshold, as well as pure-tone audiometers with speech audiometry capabilities. Also covered are portable and nonportable single- and double-room audiometric booths that have either single- or double-wall construction. Tympanometers, middle-ear analyzers, auditory evoked-potential units, and newborn hearing screening devices are excluded. For more information, see the Product Comparisons titled Recorders, Graphic, Evoked- Potential; Electromyographs; Electronystagmographs and Auditory Function Screening Devices, Newborn These devices are also called: audiometric evoked-potential units, audio spectrum analyzers, tinnitus masker analyzers, acoustic casewalls, acoustic chambers, sound-treated rooms. Purpose Audiometers measure and characterize hearing loss by determining an individual’s hearing threshold (the lowest audible sound to which the individual will respond 50% of the time) for pure tones and speech and then comparing that threshold with a standard range of normal threshold values. These normal audiometric threshold values represent the mean lowest level at which a group of young adults can detect a selected sound. Hearing loss can result from lesions anywhere along the auditory tract. When the outer and/or middle ear is affected, the hearing loss is termed conductive; when the inner ear or auditory nerves are affected, it is termed sensorineural. Abnormal conditions that can cause a conductive hearing loss include obstruction of the outer ear canal (e.g., foreign mass, swelling), neoplasms, congenital malformation, perforated tympanic membrane, disruption or fixation of the ossicles (bones of the middle ear), and fluid and scarring in the middle ear. Abnormal conditions that can cause a sensorineural hearing loss include congenital malformation, head trauma, exposure to intense noise, drug-induced toxicity, genetic factors, aging, infectious diseases, and neural tumors. Audiometry aids in diagnosis of otologic and/or neurologic hearing loss and selection of therapy. Pure-tone audiometry measures hearing sensitivity (hearing threshold) for a series of single-frequency signals chosen from the normal range of human hearing. Although this method can assess the degree of hearing loss, it cannot characterize the extent of impaired speech discrimination (understanding of speech). Speech audiometry measures hearing sensitivity for spoken material. In addition, spoken material can be used to test for speech discrimination. Comprehensive of Biomedical Engineering Site (www.dezmed.com   W   W   W  .   D   E   Z   M   E   D  .   C   O   M

Audiometers audiometric booths.PDF

Embed Size (px)

Citation preview

Page 1: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 1/116

 

5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA  Tel +1 (610) 825-6000  Fax +1 (610) 834-1275  Web www.ecri.org  E-mail [email protected]

UMDNS Information

This Product Comparison covers the followingdevice terms and product codes as listed inECRI Institute’s Universal Medical DeviceNomenclature System™ (UMDNS™): 

  Audiometers [10-228]  Audiometric Booths [10-229]

Audiometers; Audiometric Booths

Scope of this Product Comparison

This Product Comparison covers pure-tone audiometers used in routine testing and screening for hearing

threshold, as well as pure-tone audiometers with speech audiometry capabilities. Also covered are portable and

nonportable single- and double-room audiometric booths that have either single- or double-wall construction.

Tympanometers, middle-ear analyzers, auditory evoked-potential units, and newborn hearing screening

devices are excluded. For more information, see the Product Comparisons titled Recorders, Graphic, Evoked-

Potential; Electromyographs; Electronystagmographs and Auditory Function Screening Devices, Newborn. These devices are also called: audiometric evoked-potential

units, audio spectrum analyzers, tinnitus masker analyzers,

acoustic casewalls, acoustic chambers, sound-treated rooms.

Purpose

Audiometers measure and characterize hearing loss by

determining an individual’s hearing threshold (the lowest

audible sound to which the individual will respond 50% of the

time) for pure tones and speech and then comparing that

threshold with a standard range of normal threshold values.

These normal audiometric threshold values represent the mean

lowest level at which a group of young adults can detect a

selected sound.

Hearing loss can result from lesions anywhere along the

auditory tract. When the outer and/or middle ear is affected, the hearing loss is termed conductive; when the

inner ear or auditory nerves are affected, it is termed sensorineural. Abnormal conditions that can cause a

conductive hearing loss include obstruction of the outer ear canal (e.g., foreign mass, swelling), neoplasms,

congenital malformation, perforated tympanic membrane, disruption or fixation of the ossicles (bones of the

middle ear), and fluid and scarring in the middle ear. Abnormal conditions that can cause a sensorineural hearing

loss include congenital malformation, head trauma, exposure to intense noise, drug-induced toxicity, genetic

factors, aging, infectious diseases, and neural tumors. Audiometry aids in diagnosis of otologic and/or neurologic

hearing loss and selection of therapy.

Pure-tone audiometry measures hearing sensitivity (hearing

threshold) for a series of single-frequency signals chosen from the

normal range of human hearing. Although this method can assess the

degree of hearing loss, it cannot characterize the extent of impaired

speech discrimination (understanding of speech). Speech audiometry

measures hearing sensitivity for spoken material. In addition, spoken

material can be used to test for speech discrimination.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 2: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 2/116

Audiometers; Audiometric Booths

2 ©2007 ECRI Institute. All Rights Reserved.

Figure 1. Examples of pure-tone air-conduction audiograms: ( A ) X-Y graph ( ● = right ear, ▲ = left ear), ( B  ) bar graph (hearing

threshold level in dB, across; frequency in Hz, down), and ( C  ) table format (L/dB, R/dB = hearing threshold level in dB of left andright ears, respectively). Bar graph (B) and table (C) courtesy of Tremetrics, Inc.

Audiometric booths provide a consistent and controlled acoustic environment and keep background noise at

an acceptably low level to provide a suitable test area for clinical audiometry or research.

Principles of operation

Audiometers

In the normal ear, sound waves travel through air in the outer ear and cause bones in the middle ear to vibrate.The inner ear converts these mechanical vibrations to hydraulic energy and ultimately to neural impulses that are

transmitted along the auditory nerve to the brain, where it is perceived as sound.

When a patient has a suspected hearing loss, a physician may order an audiometric analysis; screening and

diagnostic testing are then directed by the audiologist. Baseline and annual screenings may be performed for

workers at risk of hearing loss from occupational noise exposure as part of a hearing conservation program and in

compliance with the Occupational Safety and Health Administration (OSHA) standards. The conventional

procedure for determining hearing threshold begins with familiarizing the patient with the stimulus. The test

proceeds according to a protocol recommended by the American Speech-Language-Hearing Association; the

protocol requires a fixed number of positive responses to stimuli (pure tones or speech) presented in sequences of

increasing intensity.

The weakest sound heard at a selected frequency is the hearing level in decibels (dB HL) for that particularfrequency. This is a relative value; the intensity reference is 0 dB HL, or audiometric 0, which corresponds to the

average threshold response of a group of 18- to 25-year-olds with no otologic pathology. The sensitivity of the

normal ear varies with frequency; therefore, 0 dB HL represents different levels of sound pressure at different

frequencies. (Negative dB HL readings indicate that hearing sensitivity is greater than normal at that p articular

frequency.)

Pure-tone audiometers typically generate a series of frequencies between 125 and 8,000 Hz, depending on the

type of test and audiometer. The American National Standards Institute (ANSI) specifies a number of categories

of audiometers based on the range of frequencies, sound-pressure levels, and testing capabilities available.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 3: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 3/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 3

In a pure-tone audiometer, single-frequency tones are produced by an audio oscillator. An attenuator control is

used to adjust the intensity changes within a specified range, typically from 0 to 90 dB HL. Signals can be

delivered in a continuous or pulsed (interrupted) mode; pulsed tones appear to be more useful when testing

children or patients with tinnitus (ringing in the ear). Microprocessor-based units give the examiner the option of

delivering the signal manually or automatically, using a programmed presentation sequence. The examiner

directs the signal to earphones or to a bone-conduction vibrator. The patient may use hand signals or a

handswitch to respond to a tone that has been heard. Both air- and bone-conduction pure-tone testing must be

performed to determine whether hearing loss is conductive, sensorineural, or both.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 4: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 4/116

Audiometers; Audiometric Booths

4 ©2007 ECRI Institute. All Rights Reserved.

Figure 3. Single-room configuration, double-wall constructionaudiometric booth.

Figure 4. Double-room configuration (control and exam),single-wall construction audiometric booth.

Pure tones are directed to earphones to test air-

conduction hearing, which relies on the action of the

outer, middle, and inner ear and neural pathways. The

electrical signal from the audiometer is transformed

into vibrations within an earphone’s diaphragm that

generate a sound of a frequency and amplitude (dB)proportional to the electrical signal. The examiner can

test the left ear, right ear, or both. The threshold search

protocol is performed for a fixed series of frequencies

in each ear. Results are recorded in a graphic and/or

tabular format called an audiogram (see Figure 1).

Bone-conduction hearing relies solely on the action

of the inner ear and neural pathways. Pure tones are

directed to a bone-conduction vibrator placed on the

forehead or the mastoid process behind the ear. The

electrical signal from the audiometer creates

vibrations within a plastic shell in the vibrator. As the

vibrations are transferred to the skull, the inner ear is

stimulated directly as a result of bone distortion and

indirectly as a result of inputs received from outer-

and middle-ear vibrations. Some instruments record

air- and bone-conduction data on the same graph or

table, while others show the data on separate audiograms.

The procedure for bone-conduction testing is basically the same as that for air-conduction testing. Ideally, both

should be performed in an audiometric booth; however, when this kind of room is unavailable, testing should be

performed in quiet surroundings with minimal background noise. Some instruments feature a two-way

microphone built into the earphone headset to permit communication between the patient and the examiner

when the patient is isolated in an audiometric booth.

In conductive hearing loss, air-conduction thresholds

are higher (indicating more hearing loss) than bone-

conduction thresholds. In sensorineural hearing loss,

 both air- and bone-conduction thresholds are equal and

elevated (hearing is below normal). When hearing loss

is “mixed” (both conductive and sensorineural), air- and

 bone-conduction thresholds are again elevated;

however, air-conduction thresholds are higher than

 bone-conduction thresholds. Bone-conduction testing is

sometimes used in lieu of air-conduction testing for

young children, when reliable pure-tone thresholds

cannot be obtained.Speech audiometry measures the sound levels at

which speech is detected and individual words are

distinguished from one another. Spoken material is

introduced as an input signal to the audiometer from a

microphone, phonograph, tape recorder, or compact

disc (CD). The signal is intensified in an amplifier and

monitored by a volume-units meter. Intensity range is

calibrated in dB HL and is controlled with an attenuator

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 5: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 5/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 5

as in pure-tone audiometry. The examiner directs the signal to earphones or through a second amplifier to

loudspeakers. An ear-selector control may also be used to direct the signal to the left or right earphone only.

Most testing is done through air-conduction

earphones. Sensitivity for speech is determined using a

speech-reception threshold (SRT) test. Threshold search

protocols are similar to those used for pure-toneaudiometry. The second component of speech

audiometry is the speech-discrimination test (SDT),

which assesses the patient’s ability to understand

monosyllabic words at suprathreshold (above

threshold) intensities. Patients with severe inner ear

disease cannot distinguish between similar words, even

at these high sound levels.

When earphones are not practical (e.g., when young

children are being tested), free-field testing is used. The

patient is placed equidistant between two loudspeakers

in a sound-isolation room. Warble tones (tones with

specific modulation rates), narrow-band noise (tones of

several frequencies of equal intensity), or spoken material is used as the auditory signal.

The shape of audiometric curves can be very useful in diagnosing certain types of hearing loss when other

methods of clinical and laboratory analysis are unsuccessful. For example, patients with hereditary dominant

mid- and low-frequency sensorineural loss show no evidence of pathology aside from their characteristic

audiograms (see Figure 2). Audiometric testing of siblings provides confirmation of this diagnosis.

When the nontest ear is suspected of responding to the test signal, it must be masked. One of three types of

masking is applied to the earphone corresponding to the nontest ear: white noise, pink noise, or narrow-band

noise. White noise consists of a wide range of frequencies with uniform loudness and is used to mask tones or

speech. Pink noise is similar to white noise except that it consists of a higher proportion of lower frequencies and

is used to mask speech alone. Narrow-band noise consists only of frequencies close to the ones being tested and is

used to mask pure tones. Proper application of masking is complex, especially when the hearing loss is mixed or

if only one ear has a severe sensorineural loss. Masking intensity is usually quantified in decibels sound-pressure

level (dB SPL).

Audiometric booths

The walls of audiometric booths consist of steel panels on the outside, sound-absorbing perforated steel panels

on the inside, and an incombustible acoustic insulating material (e.g., fiberglass) in between.

Single-wall booths are constructed from a single wall and a layer of sound-absorbing material and are used

primarily for pure-tone air-conduction testing. Double-wall booths employ “booth inside a booth” construction

in which the walls of the inner examination booth are separated from the outer panels by an air space; this design

further reduces outside noise interference and is preferred for bone-conduction, sound-field, and speech testing.Single- or double-wall construction can be used in both single- and double-room configurations.

Single-room booths (see Figure 3) isolate the subject from the environment, but not the examiner. Most

audiometry can be performed in a single-room booth; the exception is live-voice speech testing. Double-room

 booths (see Figure 4) consist of two connecting rooms (a control room and an exam room) that isolate the

examiner and subject from the environment and also from each other. Either a single or double wall (depending

on the booth construction) separates the two rooms and incorporates a window consisting of several layers of

glass. Windows should provide visual contact between examiner and subject but obscure the subject’s view of

any distracting activities in the control room. Windows are also present in the walls of a single-room booth and

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 6: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 6/116

Audiometers; Audiometric Booths

6 ©2007 ECRI Institute. All Rights Reserved.

may be present in the external walls of a double-room booth to allow the inside of the booth to be viewed from

the outside.

Single-wall booths have one door, which opens outward; double-wall booths have a second door, which opens

inward and is mounted on the same frame. Some double-room configurations offer an optional single or double

door between the control and test rooms. Vibration isolators underneath the booth minimize floor vibrations and

sound transmission. A specially designed forced-air system can minimize ventilation noise.Portable booths are single-wall structures and move on locking casters. They may also come with vibration

isolators and ventilation systems. Larger, nonportable booths come in kits that include prewired wall, floor, and

roof sections; wall and corner joiners; acoustic filling material; and a ventilation system—many portable booths

come completely assembled. Some smaller portable booths are designed to be disassembled for transport from

one testing location to another.

Reported problems

Reliable audiometric measurement depends on equipment calibration, the test environment, patient

cooperation, and examiner performance. All audiometric equipment must be calibrated according to international

standards. However, calibration and test results are invalid if the test is not performed in the proper environment.

Background noise levels must not exceed those prescribed by organizations such as ANSI and OSHA, and visualdistractions, ventilation, and temperature must be controlled. Finally, the room should be set up so that the

patient cannot see what the examiner is doing with the audiometer while the test is being performed.

ECRI Institute has received no reports of safety problems unique to audiometric booths. Single-wall booths can

 be constructed of optional fireproof panels and doors; the inner panels and doors of double-wall booths can also

 be made of fireproof material.

Purchase considerations

ECRI Institute recommendations

The accompanying comparison charts include ECRI Institute’s recommendations for minimum performancerequirements for audiometers. The recommended specifications have been divided into two categories:

audiometers used for screening purposes and audiometers used in clinical environments. Each of these categories

is further broken down into two subcategories—one for minimum requirements based on ANSI S3.6-1996, which

will provide the user with a basic audiometer, and the other for recommendations intended for more advanced

functions. These recommendations will be most appropriate for users who commonly perform comprehensive

studies. (See ANSI type designations.)

Screening audiometers should be type 4 as described in ANSI 3.6-1996. Audiometers with these capabilities

should provide the user with all necessary functions to perform under most screening conditions. Selecting a unit

with greater capability may result in more difficult operation and/or higher costs. Ease of operation is important,

particularly in screening programs in which operators may have minimal training and experience. Screening

programs generally do not use speech audiometry, and OSHA does not require it for annual audiograms;

therefore, this feature is not recommended for audiometers used for screening purposes. For screening, pure-tone

audiometers, at minimum, should be capable of testing at frequencies of 500 to 6,000 Hz.

Audiometers used for clinical testing and diagnosis should be at least type 3 as specified by ANSI S3.6-1996.

Type 1 audiometers may be required under certain conditions and are recommended for clinical environments in

which a full range of testing applications must be met. Clinical audiometers may also require speech-testing

capabilities. For this purpose, the audiometer should be designated at least type C. Standardized, specially

prepared speech material from tape or CD input is preferred to microphone input. This ensures consistent testing

and results among subjects.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 7: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 7/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 7

Standard clinical audiometry testing is performed over a range of 250 to 8,000 Hz. OSHA does not require

 bone-conduction testing for annual or baseline audiograms; however, it can be useful in determining diagnosis

(e.g., discovering where a hearing problem may lie) in a clinical setting. For most testing, frequency accuracy

should be within 3%; accuracy within 1% is recommended for some clinical applications. It is also recommended

that half-octave or octave steps be available for both clinical and screening applications. Testing at half-octave

frequencies is often used to verify that the subject’s hearing level does not change significantly within an octave.

Masking noise should be available for clinical testing to ensure that one ear is not compensating for the other

during the test.

Certain considerations for earphones

must be made for any type of audiometric

testing. Those used to deliver the signal to

the subject should include one earphone for

each ear. Although supra-aural earphones

are commonly used, availability of insert

earphones is advantageous, since these

earphones are generally lighter and provide

a better fit than supra-aural earphones. They

also offer greater protection against the

signal being heard by the opposite ear. Free-

field testing is not normally recommended

for screening purposes. However, they are a

good alternative to earphones when testing

small children who may not tolerate

earphones. Free-field testing is also used in

certain hearing-aid assessments.

Earphones should be in proper condition: pads should be soft, pliable, and free of cracks. Cords should also be

checked to make sure that the left and right earphones are connected to the correct earjacks. Earphones should be

calibrated for use with the audiometer to produce the required sound-pressure levels.

Audiometric booth requirements depend on individual circumstances. In some cases, testing may be

performed without a booth; however, when the test area is too noisy, a booth is required. Industry standards for

noise levels should be considered. The amount of sound reduction required will depend on the noise level in the

area, taking into account the noisiest periods of time. Other booth requirements will also depend on individual

circumstances. Size requirements depend on the available space, but the booth should be able to comfortably

accommodate the subject, the examiner, and all necessary equipment. In audiometric booths with two rooms,

there must be a window between the rooms.

Booths with fluorescent lighting must have the light’s starter mounted on the outside of the examination room

 because starters produce a hum that may interfere with testing. In addition, windows must have moisture-

absorbing materials between the panes to prevent fogging. If the booth has a step, a ramp must be provided for

subjects in wheelchairs or on stretchers. Because doors must be solid and close with an acoustic seal, they are

often very heavy. 

Other considerations

Some audiometers feature software that assesses the hearing level by comparing it to age-related values from

the general population. Software is available that allows the user to design a format that can accommodate a large

amount of patient (and company) data or to customize test protocols to fit the individual patient’s needs and 

capabilities. Some audiometers provide a graphic output of results for threshold testing, and several units can

store a limited number of audiograms in nonvolatile memory.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 8: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 8/116

Audiometers; Audiometric Booths

8 ©2007 ECRI Institute. All Rights Reserved.

Multichannel audiometers are available that can examine several people at once. Each channel of a group

audiometer is programmed to administer a complete series of tests to each individual; all channels are linked to a

central control channel through which the examiner can set the test parameters and initiate testing. Special

software programs compare the individual’s hearing threshold to a normal range of values (adjusted for age and

exposure to noise)—for example, from an employee’s occupational group. (Many companies require their

employees to undergo regular audiometric testing as part of a hearing-conservation program. Multichannelaudiometers enable companies to perform time- and cost-effective on-site testing of the workforce.)

Various suppliers include optional parts such as external power amplifiers, computer cables, microphones,

software, memory cards, and service manuals. Check to see if the warranty includes a hotline service for any

technical problems and/or user difficulty.

Stage of development

Microprocessor-based units have automated most test procedures and hearing-threshold calculations. Certain

computerized units have validity checks within their operational programs that repeat the signal for a given

frequency and compare the two responses. Personal-computer-based audiometers that can be used with

notebook, desktop, or networked systems are also available. These systems may have features such as graphical

displays of results, warnings, and referral levels; use of disk drives for storage; and connection to a printer toobtain additional records.

Tympanometers, middle-ear analyzers, and auditory-evoked response units have been developed that test

certain parts of the ear and auditory nerve to gather more specific information about the nature of a hearing

problem. More sensitive and specialized tests have been created for these instruments to enhance diagnosis and

monitor the results of rehabilitative therapy.

Audiometric booths are a widely accepted component of the instrumentation used in clinical diagnosis of

hearing disorders. There have been no recent changes in the technology of these rooms, and none are foreseeable

in the near future.

Bibliography

Burgess RC, ed. Technology and equipment review: audiometers. J Clin Neurophysiol 1994 Sep;11(5):525-8.

Caple GE. Effective hearing programs require assessment, testing, enforcement. Occup Health Saf 1989

 Jan;58(1):48-9, 52.

Durrant JD, Lovrinic JH. Bases of hearing sciences. 3rd ed. Baltimore: Williams & Wilkins; 1994.

Erdreich J. Threshold shift in hearing level stems from non-occupational cause. Occup Health Saf 1989 Jan;58(1):44-

7, 60.

Frank T. ANSI update: specification of audiometers. Am J Audiol 1997 Nov;6(3):29-32.

Kratz IC. Using equipment in unfamiliar clinical settings: audiology screening. J Pediatr Nurs 1997 Oct;12(5):307-

10.

Martin FN. Introduction to audiology. 3rd ed. Englewood Cliffs (NJ): Prentice-Hall; 1986.

Rintelmann W, ed. Hearing assessment. Baltimore: University Park Press; 1979.

Sawyer C. Audiometric monitoring may reduce hearing loss, compensation awards. Occup Health Saf  1989

 Jan;58(1):50-2.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 9: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 9/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 9

Supplier information

Chart A: Audiometers

AMBCO

Ambco Electronics [102475]

15052 Redhill Ave Suite DTustin, CA 92680-6525

Phone: (714) 259-7930, (800) 345-1079 Fax: (714) 259-1688

Internet: http://www.ambco.com 

E-mail: [email protected] 

AMPLIFON

Amplaid Biomedical [408757]

via Gaetano Donizetti 12

Assago MI I-20090

Italy

Phone: 39 (02) 457901 Fax: 39 (02) 48822102

Internet: http://www.amplimedical.com 

E-mail: [email protected] 

AMPLIVOX

Amplivox Ltd [178837]

29-30 Station Approach

Kidlington OX5 1JD

England

Phone: 44 (1865) 842411 Fax: 44 (1865) 841853

Internet: http://www.amplivox.ltd.uk 

E-mail: [email protected] 

CASTLE

Castle Group Ltd [324542]

Salter Road Cayton Low Road Industrial Estate

Scarborough YO11 3UZ

England

Phone: 44 (1723) 584250 Fax: 44 (1723) 583728

Internet: http://www.castlegroup.co.uk 

E-mail: [email protected] 

DANPLEX

GN Otometrics A/S [391083]

2 Dybendalsvaenget Postboks 119

Taastrup DK-2630

Denmark

Phone: 45 72111555 Fax: 45 72111548Internet: http://www.gnotometrics.com 

E-mail: [email protected] 

GN Otometrics North America Inc [426355]

125 Commerce Dr

Schaumburg, IL 60173-5329

Phone: (847) 534-2150, (800) 289-2150 Fax: (847) 534-2151

Internet: http://www.gnotometrics.com 

E-mail: [email protected] 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 10: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 10/116

Audiometers; Audiometric Booths

10 ©2007 ECRI Institute. All Rights Reserved.

DECOS

Decos Medical Systems bv [391295]

Bonnikeplein 24-26

Noordwijk NL-2201 XA

The Netherlands

Phone: 31 (71) 3640712 Fax: 31 (71) 3620833

Internet: http://www.decosmedical.nl 

E-mail: [email protected] 

ELECTRONIC PATZIG

Electronic Patzig Co for Electronic mbH [434007]

Niedersedlitzer Platz 6b

Dresden D-01259

Germany

Phone: 49 (351) 448640 Fax: 49 (351) 4486420

Internet: http://www.electronicpatzig.de 

E-mail: [email protected] 

ENTOMED

Entomed AB [157537]

Bariumgatan 29

Malmo S-213 64

Sweden

Phone: 46 (40) 210090 Fax: 46 (40) 210099

Internet: http://www.entomed.se 

E-mail: [email protected] 

EYMASA

EYMASA [354740]

Maresme 115-117

Barcelona E-08019

SpainPhone: 34 (933) 034356 Fax: 34 (933) 661848

Internet: http://www.eymasa.com 

E-mail: [email protected] 

FRYE

Frye Electronics Inc [101276]

9826 SW Tigard St PO Box 23391

Tigard, OR 97281-3391

Phone: (503) 620-2722, (800) 547-8209 Fax: (503) 639-0128

Internet: http://www.frye.com 

E-mail: [email protected] 

GM INSTRUMENTS

GM Instruments Ltd [177196]

Unit 6 Ashgrove Workshops Ashgrove Road

Kilwinning KA13 6PU

Scotland

Phone: 44 (1294) 554664 Fax: 44 (1294) 551154

Internet: http://www.gm_instruments.com 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 11: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 11/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 11

GRASON STADLER

Grason-Stadler [102263]

5225 Verona Rd Bldg 2

Madison, WI 53711-4495

Phone: (608) 441-2323, (800) 700-2282 Fax: (608) 441-2234

Internet: http://www.grason-stadler.com 

INTERACOUSTICS

Interacoustics A/S [150984]

Drejervaenget 8

Assens DK-5610

Denmark

Phone: 45 63713555 Fax: 45 63713522

Internet: http://www.interacoustics.com 

E-mail: [email protected] 

Interacoustics USA [361066]

7625 Golden Triangle Dr Unit F

Eden Prairie, MN 55344Phone: (952) 941-4200, (800) 947-6334 Fax: (952) 903-4100

Internet: http://www.interacoustics.com 

E-mail: [email protected] 

MADSEN

Madsen Electronics (Denmark) [139621]

Markaervej 2a Postboks 119

Taastrup DK-2630

Denmark

Phone: 45 72111555 Fax: 45 72111348

Internet: http://www.madsen.com 

E-mail: [email protected] 

MAICO

Maico Diagnostic GmbH [236957]

Salzufer 13/14

Berlin D-10587

Germany

Phone: 49 (30) 70714650 Fax: 49 (30) 70714699

Internet: http://www.maico-diagnostic.com 

E-mail: [email protected] 

Maico Diagnostics [348943]

7625 Golden Triangle Dr

Eden Prairie, MN 55344

Phone: (952) 941-4200, (888) 941-4201 Fax: (952) 903-4200, (888) 941-4200

Internet: http://www.maico-diagnostics.com 

E-mail: [email protected] 

MEDIROLL

Mediroll Medico-technical Ltd [287831]

Segner ter 1

Debrecen H-4025

Hungary

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 12: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 12/116

Audiometers; Audiometric Booths

12 ©2007 ECRI Institute. All Rights Reserved.

Phone: 36 (52) 533737 Fax: 36 (52) 534446

Internet: http://www.mediroll.hu 

E-mail: [email protected] 

MICRO AUDIOMETRICS

Micro Audiometrics Corp [108789]

655 Keller RdMurphy, NC 28906-5889

Phone: (828) 644-0771, (866) 327-7226 Fax: (828) 644-0772

Internet: http://www.microaud.com 

E-mail: [email protected]

MONITOR

Monitor Instruments Inc [106938]

437 Dimmocks Mill Rd

Hillsborough, NC 27278-2300

Phone: (919) 732-5400, (800) 853-6785 Fax: (919) 732-6153

Internet: http://www.monitorinstrumentsinc.com 

E-mail: [email protected] 

NAGASHIMA

Nagashima Medical Instruments Co Ltd [155887]

24-1 Hongo 5-chome Bunkyo-ku

Tokyo 113

 Japan

Phone: 81 (3) 38121271 Fax: 81 (3) 38162824

Internet: http://www.nagashima-medical.co.jp 

E-mail: [email protected] 

QUALITONE

Qualitone [103650]

6415 Flying Cloud Dr

Eden Prairie, MN 55344-3305

Phone: (952) 917-7161, (800) 328-3897 Fax: (952) 828-9262

Internet: http://www.qualitonehearing.com 

SIBELMED

Sibel SA [177708]

Rossello 500

Barcelona E-08026

Spain

Phone: 34 (93) 4360007 Fax: 34 (93) 4361611

Internet: http://www.sibelmed.com 

E-mail: [email protected] 

SINGER

Singer Medical Products Inc [106865]

790 Maple Ln

Bensenville, IL 60106-1513

Phone: (630) 860-6500 Fax: (630) 860-3672

E-mail: [email protected] 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 13: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 13/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 13

STARKEY

Starkey Australia Pty Ltd [418605]

74-96 Pyrmont Bridge Road

Camperdown 2050

Australia

Phone: 61 (2) 95502077

Internet: http://www.starkey.com.au 

Starkey Colombia [418609]

Calle 98 No 22-64

Santa Fe de Bogota

Colombia

Phone: 57 (1) 6360960

Internet: http://www.starkey.com 

Starkey France Sarl [418610]

23 rue Claude Nicolas le Doux Europarc Batiment 14

Creteil F-94000

FrancePhone: 33 (1) 49807474

Internet: http://www.starkey.com 

Starkey Laboratories Inc [107526]

6700 Washington Ave S

Eden Prairie, MN 55344

Phone: (952) 941-6401, (800) 328-8602 Fax: (952) 828-9262

Internet: http://www.starkey.com 

Welch Allyn

Welch Allyn GmbH & Co KG [178366]

Zollerstrasse 2-4

 Jungingen D-72417Germany

Phone: 49 (7477) 92710 Fax: 49 (7477) 927190

Internet: http://www.welchallyn.de 

E-mail: [email protected] 

Welch Allyn Inc [101850]

4341 State Street Rd PO Box 220

Skaneateles Falls, NY 13153-0220

Phone: (315) 685-4100, (800) 535-6663 Fax: (315) 685-4091

Internet: http://www.welchallyn.com 

E-mail: [email protected]

Chart B: Audiometric Booths

ACOUSTAIR

Acoustair [321007]

Postbus 1

Moerkapelle NL-2750 AA

The Netherlands

Phone: 31 (79) 5931341 Fax: 31 (79) 5931317

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 14: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 14/116

Audiometers; Audiometric Booths

14 ©2007 ECRI Institute. All Rights Reserved.

Internet: http://www.acoustair.com 

E-mail: [email protected] 

ACOUSTIC

Acoustic Systems [104980]

415 E St Elmo Rd

Austin, TX 78745Phone: (512) 444-1961, (800) 749-1460 Fax: (512) 444-2282

Internet: http://www.acousticsystems.com 

E-mail: [email protected] 

AMPLIFON

Amplaid Biomedical [408757]

via Gaetano Donizetti 12

Assago MI I-20090

Italy

Phone: 39 (02) 457901 Fax: 39 (02) 48822102

Internet: http://www.amplimedical.com 

E-mail: [email protected] 

ECKEL

Eckel Industries Inc [104016]

155 Fawcett St

Cambridge, MA 02138

Phone: (617) 491-3221 Fax: (617) 547-2171

Internet: http://www.eckelacoustic.com 

E-mail: [email protected] 

EYMASA

EYMASA [354740]

Maresme 115-117

Barcelona E-08019

Spain

Phone: 34 (933) 034356 Fax: 34 (933) 661848

Internet: http://www.eymasa.com 

E-mail: [email protected] 

IAC

Industrial Acoustics Co [102440]

1160 Commerce Ave

Bronx, NY 10462

Phone: (718) 430-4523 Fax: (718) 863-1138

Internet: http://www.industrialacoustics.com 

E-mail: [email protected] 

Industrial Acoustics Co GmbH [235118]

Solweg 17

Niederkruechten D-41372

Germany

Phone: 49 (2163) 99910 Fax: 49 (2163) 999123

Internet: http://www.iac-gmbh.de 

E-mail: [email protected] 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 15: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 15/116

Page 16: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 16/116

Audiometers; Audiometric Booths

16 ©2007 ECRI Institute. All Rights Reserved.

Policy Statement

The Healthcare Product Comparison System (HPCS) is published by ECRI Institute, a nonprofit organization.

HPCS provides comprehensive information to help healthcare professionals select and purchase diagnostic and

therapeutic capital equipment more effectively in support of improved patient care.

The information in Product Comparisons comes from a number of sources: medical and biomedical

engineering literature, correspondence and discussion with manufacturers and distributors, specifications from

product literature, and ECRI Institute’s Problem Reporting System. While these data are reviewed by qualified

health professionals, they have not been tested by ECRI Institute’s clinical and engineering personnel and are

largely unconfirmed. The Healthcare Product Comparison System and ECRI Institute are not responsible for the

quality or validity of information derived from outside sources or for any adverse consequences of acting on such

information.

The appearance or listing of any item, or the use of a photograph thereof, in the Healthcare Product Comparison

System does not constitute the endorsement or approval of the product’s quality, performance, or value, or of

claims made for it by the manufacturer. The information and photographs published in Product Comparisons

appear at no charge to manufacturers.

Many of the words or model descriptions appearing in the Healthcare Product Comparison System are

proprietary names (e.g., trademarks), even though no reference to this fact may be made. The appearance of any

name without designation as proprietary should not be regarded as a representation that is not the subject of

proprietary rights.

ECRI Institute respects and is impartial to all ethical medical device companies and practices. The Healthcare

Product Comparison System accepts no advertising and has no obligations to any commercial interests. ECRI

Institute and its employees accept no royalties, gifts, finder’s fees, or commissions from the medical device

industry, nor do they own stock in medical device companies. Employees engage in no private consulting work

for the medical device industry.

About ECRI Institute

ECRI Institute, a nonprofit organization, dedicates itself to bringing the discipline of applied scientific research

in healthcare to uncover the best approaches to improving patient care. As pioneers in this science for nearly 40

years, ECRI Institute marries experience and independence with the objectivity of evidence-based research.

More than 5,000 healthcare organizations worldwide rely on ECRI Institute’s expertise in patient safety

improvement, risk and quality management, healthcare processes, devices, procedures, and drug technology.

ECRI Institute is one of only a handful of organizations designated as both a Collaborating Center of the World

Health Organization and an Evidence-based Practice Center by the U.S. Agency for Healthcare Research and

Quality. For more information, visit http://www.ecri.org. 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 17: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 17/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 17

Product Comparison Charts

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 18: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 18/116

Audiometers; Audiometric Booths

18 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

Clinical Audiometers,

Recommended

Clinical Audiometers,

Required

Screening Audiometers,

Recommended

Screening Audiometers,

RequiredWHERE MARKETED

FDA CLEARANCE

CE MARK (MDD)

TYPE Type 1 pure tone Type 3 pure tone Type 4 pure tone Type 4 pure toneSpeech Type C

CONFIGURATION

PURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125, 250, 500, 750, 1,000,

1,500, 2,000, 3,000, 4,000,6,000, 8,000

250, 500, 1,000, 2,000,3,000, 4,000, 6,000, 8,000

500, 1,000, 2,000, 3,000,4,000, 6,000

500, 1,000, 2,000, 3,000,4,000, 6,000

Bone conduction 250, 500, 750, 1,000, 1,500,2,000, 3,000, 4,000, 6,000

250, 500, 1,000, 2,000,3,000, 4,000

Continuous or step Step StepFrequency accuracy 1% 3% 3% 3%

Intensity, dB HLAir conduction

250 Hz 90 70 90 701,000 Hz 120 @ 500-4,000 Hz 100 @ 500-4,000 Hz8,000 Hz 100 80Graduations 5 5 5 5

Bone conduction250 Hz 70 351,000 Hz 70 @ 1,000-3,000 Hz 60 @ 1,000-3,000 Hz4,000 Hz 60 50Graduations 5 5

Masking, noise type Narrow band, white Narrow bandIntensity, dB SPL ≤115  ≤115

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 19: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 19/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 19

Chart A: Audiometers

MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

Clinical Audiometers,Recommended

Clinical Audiometers,Required

Screening Audiometers,Recommended

Screening Audiometers,Required

SPEECH AUDIOMETRYInputs Tape, CDIntensity, dB HLMasking, noise type Speech spectrum noise

Intensity, dB SPL ≤115  ≤115 EARPHONE TYPE TDH-39, TDH-49, ER3-A

Cushion type MX-41-AR, insertearphones

BONE VIBRATOR YesType

SPECIAL TESTS Hearing-aid testingcapabilities when thesetests are likely to beperformed

POWER REQUIREMENTSLine, VACBattery

Type, Volts

W x D x H, cm (in)

WEIGHT, kg (lb)

PURCHASEINFORMATION

List priceWarrantyDelivery time, AROFiscal year

This is the second of three

pages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 20: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 20/116

Audiometers; Audiometric Booths

20 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

Clinical Audiometers,Recommended

Clinical Audiometers,Required

Screening Audiometers,Recommended

Screening Audiometers,Required

OTHER SPECIFICATIONS Other recommendedfeatures: audible monitoringof test signal; operator-to-subject speechcommunication; signalindicator/VU meter; subjectresponse system.

Other recommendedfeatures: subject responsesystem.

Other recommendedfeatures: subject responsesystem.

LAST UPDATED

Supplier Footnotes 1These recommendationsare the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on

this data.

1These specifications arethe opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on

this data.

1These recommendationsare the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on

this data.

1These recommendationsare the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based on

this data.Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 21: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 21/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 21

Chart A: Audiometers

MODEL AMBCO AMBCO AMBCO AMPLIFON1000 + Otoscreen 2500 Automatic Storage 650 A Amplaid A137

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified YesCE MARK (MDD) Not specified Not specified Not specified YesTYPE Wide range Wide range Limited range Limited range

Speech No No No NoCONFIGURATION Portable Tabletop/portable Portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 250-8,000 (octave,

interoctave bands)250-8,000 (octave,interoctave bands)

250-8,000 (octave,interoctave bands)

125-8,000

Bone conduction NA NA NA 250-8,000Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified >1%

Intensity, dB HLAir conduction

250 Hz 0 to +90 0 to +95 0 to +90 -10 to +100

1,000 Hz 0 to +90 0 to +95 0 to +90 -10 to +1208,000 Hz 0 to +90 0 to +95 0 to +90 -10 to +100Graduations 5 5 5 5

Bone conduction250 Hz NA NA NA -10 to +501,000 Hz NA NA NA -10 to +804,000 Hz NA NA NA -10 to +80Graduations NA NA NA 5

Masking, noise type NA NA NA Narrow band, whiteIntensity, dB SPL NA NA NA -10 to +95 dB HL

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 22: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 22/116

Audiometers; Audiometric Booths

22 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL AMBCO AMBCO AMBCO AMPLIFON1000 + Otoscreen 2500 Automatic Storage 650 A Amplaid A137

SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA

Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-49, EARtone 3A

Cushion type Supra-aural Supra-aural Supra-aural MX-41BONE VIBRATOR

Type NA NA NA B-71SPECIAL TESTS Sequential screening,

accurate, fast screening,digital, printouts

None specified None specified AC-BC steady and pulsedpure tones, warble,automatic threshold,synchronous masking, LCD,noise enclosures

POWER REQUIREMENTSLine, VAC 120/220 120/220 105-125/220 110/240, 50/60 HzBattery None None AB None

Type, Volts NA NA 9 NAW x D x H, cm (in) 24.1 x 29.2 x 16.5 (9.5 x

11.5 x 6.5)24.1 x 29.2 x 16.5 (9.5 x11.5 x 6.5)

22.9 x 33 x 15.2 (9 x 13 x 6) 36 x 28 x 12 (14 x 11 x 5)

WEIGHT, kg (lb) 3.2 (7) 3.2 (7) 1.8 (4) 1.9 (4.3)PURCHASEINFORMATION

List price $795 $1,975 $735  €1,700 (US $2,150) Warranty 5 years 5 years 5 years 1 yearDelivery time, ARO 1 week 1 week 1 week 30 daysFiscal year May to April May to April May to April January to December

This is the second of threepages covering the abovemodel(s). These

specifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 23: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 23/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 23

Chart A: Audiometers

MODEL AMBCO AMBCO AMBCO AMPLIFON1000 + Otoscreen 2500 Automatic Storage 650 A Amplaid A137

OTHER SPECIFICATIONS Digital LED display; front-panel membrane switches;printer available. Meetsrequirements of ANSI.

Microprocessor-controlled;stores 540 tests; fullyautomatic, semiautomatic,or manual testing; computercompatible. Meetsrequirements of ANSI;OSHA compliant.

Carrying case. Meetsrequirements of ANSI.

Built-in talk-forward mic;single insert phone formasking; noise enclosures;LCD, 4 rows x 20characters; external printerand carrying case; datastorage and transfer; built-inmemory for 128 patients,

 AC-BC both ears; RS232port; transfer to Amplaid

 ASA database (Windows);NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.

LAST UPDATED January 2004 January 2004 January 2004 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 24: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 24/116

Audiometers; Audiometric Booths

24 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON Amplaid A177 Plus Amplaid A321 Amplaid A321 HF Amplaid A724

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range, clinical, ANSI

1A-EWide range, clinical, ANSI1A-E

Wide range, clinical, ANSIS3.39

Speech Yes Yes Yes NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 125-8,000 125-20,000 500-4,000Bone conduction 250-8,000 250-8,000 250-8,000 NoContinuous or step Step Step Step StepFrequency accuracy >1% >1% >1% >1%

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +100 No

1,000 Hz -10 to +120 -10 to +120 -10 to +120 -50 to +1208,000 Hz -10 to +100 -10 to +100 -10 to +100 NoGraduations 1, 2, and 5 1, 2, and 5 1, 2, and 5 5

Bone conduction250 Hz -10 to +50 -10 to +50 -10 to +50 NA1,000 Hz -10 to +80 -10 to +80 -10 to +80 NA4,000 Hz -10 to +80 -10 to +80 -10 to +80 NAGraduations 1, 2, and 5 1, 2, and 5 1, 2, and 5 NA

Masking, noise type Narrow band, white Narrow band, white Narrow band, white NAIntensity, dB SPL -10 to +95 dB HL -10 to +95 dB HL -10 to +95 dB HL NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 25: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 25/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 25

Chart A: Audiometers

MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON Amplaid A177 Plus Amplaid A321 Amplaid A321 HF Amplaid A724

SPEECH AUDIOMETRYInputs Mic, tape, CD Mic, tape, CD (2 separate

EXT inputs)Mic, tape, CD (2 separateEXT inputs)

NA

Intensity, dB HL -10 to +105 -10 to +105 -10 to +105 NAMasking, noise type Speech, white Speech, white Speech, white NA

Intensity, dB SPL -10 to +95 dB HL -10 to +95 dB EM -10 to +95 dB EM Not specifiedEARPHONE TYPE TDH-49, EARtone 3A TDH-49, EARtone 3A TDH-49, EARtone 3A Not specified

Cushion type MX-41 MX-41 MX-41 Not specifiedBONE VIBRATOR

Type B-71 B-71 B-71 Not specifiedSPECIAL TESTS  AC-BC-FF pure tones and

speech, warble, automaticthreshold, Stereo Speech,

 ABLB, synchronousmasking, LCD, noiseenclosures

 AC-BC-FF pure tones andspeech, warble, StereoSpeech, multifrequency,

 ABLB (Fowler), MLB(Reger), DLI (Lusher), andSISI (Jerger), tone decay,automatic threshold, tone

and speech Stenger, MHA

 AC-BC-FF pure tones andspeech, warble, StereoSpeech, multifrequency,

 ABLB (Fowler), MLB(Reger), DLI (Lusher) andSISI (Jerger), tone decay,automatic threshold, tone

and speech Stenger, MHA,MLD (Masking leveldifference), high frequencyaudiometry (2 frequencyranges), Bekesy (sweepand fixed frequencies,continuous, pulsed and LOTstimuli, continuous and fixedmasking noise)

Contralateral audiometry,226 Hz tympanometry,acoustic reflex tests,sequential tests includingdecay

POWER REQUIREMENTSLine, VAC 100/240, 50/60 Hz 100/240, 50/60 Hz 100/240, 50/60 Hz 100/240, 50/60 HzBattery None None None None

Type, Volts NA NA NA NAW x D x H, cm (in) 36 x 28 x 12 (14 x 11 x 5) 36 x 28 x 18 (14 x 11 x 7) 36 x 28 x 18 (14 x 11 x 7) 49 x 40 x 20 (19.3 x 15.7 x

7.9)WEIGHT, kg (lb) 1.9 (4.3) 3 (7.7) 3 (7.7) 8.5 (18.8)

PURCHASEINFORMATIONList price  €2,400 (US $3,036)   €3,900 (US$4,933)   €4,850 (US$6,135)   €4,850 (US$6,135) Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 30 days 30 days 30 days 30 daysFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 26: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 26/116

Audiometers; Audiometric Booths

26 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON Amplaid A177 Plus Amplaid A321 Amplaid A321 HF Amplaid A724

OTHER SPECIFICATIONS Built-in talk-forward mic;goose-neck mic; patientmic; single insert phone formasking; free-field stereooutputs; noise enclosures;LCD, 4 rows x 20characters; external printerand carrying case; datastorage and transfer; built-inmemory for 128 patients,

 AC-BC both ears; RS232port; transfer to Amplaid

 ASA database (Windows);NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.

Twin channels; LCDdisplaying right and leftaudiograms; built-in printerand output to externalprinters, PLC5 languagecompatible; built-in talk-forward mic; goose-neckmic; patient mic; singleinsert phone for masking;free-field stereo outputs;noise enclosures; carryingcase; data storage andtransfer; built-in memory forall tests for one patient;RS232 and USB serialports; transfer to Amplaid

 ASA database (Windows);

NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.

Twin channels; LCDdisplaying right and leftaudiograms; built-in printerand output to externalprinters, PLC5 languagecompatible; built-in talk-forward mic; goose-neckmic; patient mic; singleinsert phone for masking;free-field stereo outputs;noise enclosures; carryingcase; data storage andtransfer; built-in memory forall tests for one patient;RS232 and USB serialports; transfer to Amplaid

 ASA database (Windows);

NOAH compatible. Meetsrequirements of EMC andIEC 601-1 Class I Type B.

Diagnostic and highfrequency impedanceaudiometer; built-in thermalprinter; LCD; RS232 datatransfer to Amplaid ASA(Windows); also availableas A724 clinical impedanceaudiometer. Meetsrequirements of EMC andIEC 601-1 Class I Type B.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 27: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 27/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 27

Chart A: Audiometers

MODEL AMPLIFON AMPLIFON AMPLIVOX AMPLIVOX Amplaid A728 Amplaid A756 116 240

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Not specified Not specifiedCE MARK (MDD) Yes Yes Yes YesTYPE Clinical HF, Type 1 ANSI

S3.39 (1987)Screening, Type 3 ANSIS3.39 (1987)

Limited range Wide range

Speech No No No NoCONFIGURATION Tabletop Tabletop/portable Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-12,000 500-4,000 250-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands)

Bone conduction No No NA 250-8,000 (octave,interoctave bands)

Continuous or step Step Step Step StepFrequency accuracy >1% >1% Not specified Not specified

Intensity, dB HL

Air conduction250 Hz -10 to +85 No -10 to +80 -10 to +901,000 Hz -10 to +120 50 to 120 -10 to +100 -10 to +1208,000 Hz -10 to +120 No -10 to +80 -10 to +90Graduations 5 5 5 5

Bone conduction250 Hz NA NA NA -10 to +301,000 Hz NA NA NA -10 to +604,000 Hz NA NA NA -10 to +60Graduations NA NA NA 5

Masking, noise type Narrow band NA NA Narrow band, broadIntensity, dB SPL -10 to +95 dB HL NA NA -10 to +100

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 28: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 28/116

Audiometers; Audiometric Booths

28 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL AMPLIFON AMPLIFON AMPLIVOX AMPLIVOX Amplaid A728 Amplaid A756 116 240

SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA

Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-49 TDH-49 TDH-39 TDH-49

Cushion type MX-41 MX-41 Circumaural or supra-aural Circumaural or supra-auralBONE VIBRATOR

Type NA NA NA B-71SPECIAL TESTS  Automatic and manual AC

threshold without masking,automatic and manualtympanometry (226, 678,800, and 1,000 Hz),eustachian-tube testing(normal and perforated TC),manual and automated

reflexometry, reflex growth,nonacoustic reflex,sequential tymp plus reflexprotocols

Contralateral audiometry,226 Hz tympanometry,acoustic reflex tests,sequential tests

None None

POWER REQUIREMENTSLine, VAC 110/240, 50/60 Hz 110/240, 50/60 Hz 100-110/220-240 100-110/220-240Battery None None Alkaline Alkaline

Type, Volts NA NA 6 6W x D x H, cm (in) 49 x 40 x 20 (19.3 x 15.7 x

7.9)37 x 29 x 18 (14 x 11 x 7) 25 x 17 x 7 (10 x 7 x 2.5) 25 x 17 x 7 (10 x 7 x 2.5)

WEIGHT, kg (lb) 8.5 (18.8) 3.5 (7.7) 0.73 (1.6) 0.85 (1.82)PURCHASEINFORMATION

List price  €6,186 (US $7,825)   €3,000 (US $3,795)  $825 $1,912Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 30 days 30 days 4 weeks 4 weeks

Fiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 29: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 29/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 29

Chart A: Audiometers

MODEL AMPLIFON AMPLIFON AMPLIVOX AMPLIVOX Amplaid A728 Amplaid A756 116 240

OTHER SPECIFICATIONS Diagnostic and high-frequency impedanceaudiometer; built-in thermalprinter; LCD; RS232 datatransfer to Amplaid ASA(Windows); also availableas A724 clinical impedanceaudiometer; ample choice of

 AR tests, includingnonacoustic. Meetsrequirements of EMC andIEC 601-1 Class I Type B.

Screening, Type 3 ANSIS3.39 (1987) withsequential tests and severalreflex tests comparing reflexgrowth.

Carrying case. Meetsrelevant BS and IECrequirements.

Carrying case. Meetsrelevant BS and IECrequirements.

LAST UPDATED July 2005 July 2005 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 30: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 30/116

Audiometers; Audiometric Booths

30 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL AMPLIVOX AMPLIVOX CASTLE DANPLEX260 CA850/3 GA1001 Excalibur DA65

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Limited range Pure tone Wide range

Speech Yes No No YesCONFIGURATION Tabletop/portable Tabletop/portable Portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)500-8,000 (octave,interoctave bands)

125-12,000 250-12,500 (12 frequencies)

Bone conduction 250-8,000 (octave,interoctave bands)

NA NA 250-8,000 (8 frequencies)

Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +90 NA -20 to +120 -10 to +1101,000 Hz -10 to +120 -10 to +90 -20 to +120 -10 to +1208,000 Hz -10 to +90 -10 to +90 -20 to +120 -10 to +100Graduations 5 2-5, 5 5 1, 2.5, 5 dB

Bone conduction250 Hz -10 to +30 NA NA -10 to +401,000 Hz -10 to +60 NA NA -10 to +804,000 Hz -10 to +60 NA NA -10 to +75Graduations 5 NA NA 1, 2.5, 5 dB

Masking, noise type Narrow band, broad NA NA Narrow band, whiteIntensity, dB SPL -10 to +100 NA NA -10 dB to maximum (for

pure tones, -20 dB)

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 31: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 31/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 31

Chart A: Audiometers

MODEL AMPLIVOX AMPLIVOX CASTLE DANPLEX260 CA850/3 GA1001 Excalibur DA65

SPEECH AUDIOMETRYInputs Mic, tape NA NA Tape, CD, micIntensity, dB HL -10 to +100 NA NA -10 to +80Masking, noise type Wide band NA NA Speech, white, ipsilateral,

contralateralIntensity, dB SPL -10 to +100 NA NA -10 to +80

EARPHONE TYPE TDH-49 TDH-39 Telephonics TDH-39Cushion type Circumaural or supra-aural Circumaural or supra-aural TDH-39 Circumaural

BONE VIBRATORType B-71 NA NA B-71

SPECIAL TESTS  ABLB, Stenger, pulse andcontinuous tone,synchronous masking

 Automatic Bekesy andHughson-Westlake graphics

None Stenger, Fowler (ABLB),SISI, master hearing aid

POWER REQUIREMENTSLine, VAC 100-110/220-240 100-110/220-240 220/110 110/220/240 ±10%, 50/60

HzBattery  Alkaline None None None

Type, Volts 6 NA NA NAW x D x H, cm (in) 25 x 17 x 7 (10 x 7 x 2.5) 33.5 x 26.5 x 8.5 (13.2 x

10.4 x 3.3)27.5 x 28 x 7.5 (10.8 x 11 x3)

45 x 29 x 8.5 (17.7 x 11.4 x3.3)

WEIGHT, kg (lb) 0.85 (1.85) 2.9 (6.4) 1.6 (3.5) 4.5 (9.9)PURCHASEINFORMATION

List price $2,100 $3,675 £2,540 (US$3,750) $2,350Warranty 1 year 1 year 4 years 2 yearsDelivery time, ARO 4 weeks 4 weeks ~2 weeks Not specifiedFiscal year January to December January to December Not specified January to December

This is the second of threepages covering the abovemodel(s). These

specifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 32: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 32/116

Audiometers; Audiometric Booths

32 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL AMPLIVOX AMPLIVOX CASTLE DANPLEX260 CA850/3 GA1001 Excalibur DA65

OTHER SPECIFICATIONS Carrying case. Meetsrelevant BS and IECrequirements.

Internal and PC database.Meets relevant BS and IECrequirements.

Unit works with mostnotebook, desktop, ornetworked systems runningon Windows 3.1, 95, 98, or2000; unlimited storage onNT disk; statistical analysisand report letter generationfacility; printout of results in

 A4 format. Meetsrequirements of BS 5966and 601-1.

Optional carrying case, highfrequency to 16,000 Hz.Meets requirements of IECPub. 645. Complies withIEC 601-1.

LAST UPDATED January 2007 January 2007 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 33: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 33/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 33

Chart A: Audiometers

MODEL DECOS ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG Audiology Workstation MA 12 MA 14 MA 15

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No Not specified Not specified Not specifiedCE MARK (MDD) Yes Not specified Not specified Not specifiedTYPE Wide range Narrow range Wide range Wide range

Speech Yes No No YesCONFIGURATION Patient interface, TDH36

stereo, TDH39 contra, B71bone conductor, patientresponse button

Portable Portable Portable

PURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-20,000 250-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands)

125-8,000 (octave,interoctave bands)

Bone conduction 250-8,000 NA 125-8,000 (octave,interoctave bands)

125-8,000 (octave,interoctave bands)

Continuous or step Step Step Step Step

Frequency accuracy 1 Hz Not specified Not specified Not specifiedIntensity, dB HL

Air conduction250 Hz -10 to +100 -10 to +90 -10 to +100 -10 to +1001,000 Hz -10 to +120 -10 to +90 -10 to +100 -10 to +1008,000 Hz -10 to +110 -10 to +90 -10 to +100 -10 to +100Graduations 1 5 5 5

Bone conduction250 Hz -10 to +45 NA -10 to +60 -10 to +601,000 Hz -10 to +75 NA -10 to +70 -10 to +704,000 Hz -10 to +75 NA -10 to +60 -10 to +60Graduations 1 NA 5 5

Masking, noise type Band, white, pink, speech,user-definable

NA Narrow band Narrow band

Intensity, dB SPL -10 to +110 NA -10 to +80 -10 to +80

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 34: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 34/116

Audiometers; Audiometric Booths

34 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL DECOS ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG Audiology Workstation MA 12 MA 14 MA 15

SPEECH AUDIOMETRYInputs CD, mic, hard disk NA NA Mic, tape, CDIntensity, dB HL -10 to +120 NA NA -10 to +110Masking, noise type White, pink, speech, band,

user-definableNA NA Speech

Intensity, dB SPL -10 to +110 NA NA -10 to +110EARPHONE TYPE TDH-39 8105 B2 8105 B2 8106

Cushion type Supra-aural Not specified Not specified Not specifiedBONE VIBRATOR

Type Radioear B71 NA KH 70 KH 70SPECIAL TESTS Tone audiometry, high

frequency, speechaudiometry, ABLB, tonedecay, Weber, speech innoise, directional hearing,visual reinforcementaudiometry, tinnitus, CAT,

tympanometer link

None SISI SISI

POWER REQUIREMENTSLine, VAC 100-110/220-240 230 230 230Battery None 3/R14 None None

Type, Volts NA 1.5 NA NAW x D x H, cm (in) 29 x 25 x 6 (11.4 x 9.8 x

2.4)30 x 22.5 x 5 (11.8 x 8.9 x2)

30 x 22.5 x 5 (11.8 x 8.9 x2)

30 x 22.5 x 5 (11.8 x 8.9 x2)

WEIGHT, kg (lb) 3 (6.6) 1.5 (3.3) 2 (4.4) 2 (4.4)PURCHASEINFORMATION

List price $8,900 $1,218 $1,867 $2,038Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 4 weeks Not specified Not specified Not specifiedFiscal year January to December Not specified Not specified Not specified

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 35: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 35/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 35

Chart A: Audiometers

MODEL DECOS ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG Audiology Workstation MA 12 MA 14 MA 15

OTHER SPECIFICATIONS PC-controlled audiometer;network base links to HL7,SQL server, and Oracle withopen database definition.Custom integration of everyelectronic patient file. Meetsrequirements of ISO 389,645, 4870, 6189, and 8253.

Carrying case included. Carrying case included; freefield via loudspeaker.

Carrying case included; freefield via loudspeaker.

LAST UPDATED January 2007 December 20021  December 20021  December 20021 Supplier Footnotes

Model Footnotes

Data Footnotes 1Model listed is currentlymarketed.

1Model listed is currentlymarketed.

1Model listed is currentlymarketed.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 36: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 36/116

Audiometers; Audiometric Booths

36 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIGMA 22-1 MA 22-2 MA 22-3 MA 32

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Wide range Wide range Wide range Wide range

Speech Yes No Yes YesCONFIGURATION Portable/tabletop Portable/tabletop Portable/tabletop TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements

125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements

125-8,000 (octave,interoctave bands)

Bone conduction 125-8,000 (octave,interoctave bands)

125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements

125-8,000 (octave,interoctave bands); up to16,000 Hz with optionalhigh-tone measurements

125-8,000 (octave,interoctave bands)

Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to 0 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +30 -10 to +130 -10 to +130 -10 to +1308,000 Hz -10 to +10 -10 to +110 -10 to +110 -10 to +110Graduations 5 5 5 5

Bone conduction250 Hz -10 to +60 -10 to +60 -10 to +60 -10 to +601,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +60 -10 to +60 -10 to +60 -10 to +60Graduations 5 5 5 5

Masking, noise type Wide/narrow band Wide/narrow band Wide/narrow band Narrow band, whiteIntensity, dB SPL -10 to +100, -10 to +110 -10 to +100, -10 to +110 -10 to +100, -10 to +110 -10 to +100, -10 to +110

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 37: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 37/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 37

Chart A: Audiometers

MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIGMA 22-1 MA 22-2 MA 22-3 MA 32

SPEECH AUDIOMETRYInputs Mic, tape, CD NA Mic, tape, CD Mic, tape, CDIntensity, dB HL -10 to +110 NA -10 to +110 -10 to +110Masking, noise type Speech NA Speech White, speech

Intensity, dB SPL -10 to +110 NA -10 to +110 -10 to +110EARPHONE TYPE 8106 HDA-200 8106, HDA-200 Not specified

Cushion type Not specified Not specified Not specified EARtone 3ABONE VIBRATOR

Type KH 70 KH 70 KH 70 KH 70SPECIAL TESTS SISI, Stenger, Fowler,

LombardNone specified SISI, Stenger, Fowler,

LombardSISI, Stenger, Luscher,Rainville, Fowler,Langenbeck, Carhart,Lombard, Kingsbury

POWER REQUIREMENTSLine, VAC 230 230 230 230Battery None None None None

Type, Volts NA NA NA NA

W x D x H, cm (in) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5)WEIGHT, kg (lb) 4 (8.8) 4 (8.8) 4 (8.8) 4 (8.8)PURCHASEINFORMATION

List price $2,773 $2,831 $3,130 $3,863Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 38: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 38/116

Audiometers; Audiometric Booths

38 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIG ELECTRONIC PATZIGMA 22-1 MA 22-2 MA 22-3 MA 32

OTHER SPECIFICATIONS RS232 port for PCconnection; CENTRONICSfor printer connection;optional carrying case;monaural and dichoticspeech. Meets requirementsof IEC 645 and and ISO389.

RS232 port for PCconnection; CENTRONICSfor printer connection;optional carrying case.Meets requirements of IEC645 and ISO 389.

RS232 port for PCconnection; CENTRONICSfor printer connection;monaural and dichoticspeech; optional carryingcase. Meets requirements ofIEC 645 and ISO 389.

RS232 port for PCconnection; monaural,dichotic, ipsilateral talkoverspeech system. Meetsrequirements of IEC 645and ISO 389.

LAST UPDATED December 20021  December 20021  December 20021  December 20021 Supplier Footnotes

Model Footnotes

Data Footnotes 1Model listed is currentlymarketed.

1Model listed is currentlymarketed.

1Model listed is currentlymarketed.

1Model listed is currentlymarketed.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 39: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 39/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 39

Chart A: Audiometers

MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ENTOMED ENTOMEDMA 33 MA 33-1 GSI 67 GSI 68

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Not specified Not specified Yes YesTYPE Wide range Wide range Wide range Wide range

Speech Yes Yes No YesCONFIGURATION Tabletop Tabletop Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands)

125-8,000 125-8,000

Bone conduction 125-8,000 (octave,interoctave bands)

250-8,000 (octave,interoctave bands)

250-8,000 250-8,000

Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +130 -10 to +130 -10 to +110 -10 to +1108,000 Hz -10 to +110 -10 to +110 -10 to +100 -10 to +100Graduations 5 5 5 5

Bone conduction250 Hz -10 to +60 -10 to +60 -10 to +35 -10 to +351,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +60 -10 to +60 -10 to +80 -10 to +80Graduations 5 5 5 5

Masking, noise type Narrow band, white Narrow band, white Noise band Noise band, speechIntensity, dB SPL -10 to +100, -10 to +110 -10 to +110 0 to +95 0 to +95

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 40: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 40/116

Audiometers; Audiometric Booths

40 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ENTOMED ENTOMEDMA 33 MA 33-1 GSI 67 GSI 68

SPEECH AUDIOMETRYInputs Mic, tape, CD Mic, tape, CD NA CD, tape and micIntensity, dB HL -10 to +110 -10 to +110 NA -10 to +100Masking, noise type White, speech White, speech NA Noise band, speech

Intensity, dB SPL -10 to +110 -10 to +110 NA -10 to +100EARPHONE TYPE Not specified Not specified TDH-39, EARtone 5A TDH-39, EARtone 5A

Cushion type EARtone 3A EARtone 3A Supra-aural Supra-auralBONE VIBRATOR

Type KH 70 KH 70 B-71 B-71SPECIAL TESTS SISI, Stenger, Luscher,

Rainville, Fowler,Langenbeck, Carhart,Lombard, Kingsbury

SISI, Stenger, Luscher,Rainville, Fowler,Langenbeck, Carhart,Lombard, Kingsbury

Free-field option Free-field option

POWER REQUIREMENTSLine, VAC 230 230 115/230 115/230Battery None None None None

Type, Volts NA NA NA NA

W x D x H, cm (in) 48 x 33 x 9 (18.9 x 13 x 3.5) 48 x 33 x 9 (18.9 x 13 x 3.5) 38.4 x 26.9 x 7.4 (15.1 x10.6 x 2.9)

48 x 26.9 x 7.4 (18.9 x 10.6x 2.9)

WEIGHT, kg (lb) 4 (8.8) 4 (8.8) 2 (4.4) 2.6 (5.7)PURCHASEINFORMATION

List price $4,439 $4,726 Not specified Not specifiedWarranty 1 year 1 year 2 year 2 yearDelivery time, ARO Not specified Not specified Stock, low quantities Stock, low quantitiesFiscal year Not specified Not specified January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue onto

the next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 41: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 41/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 41

Chart A: Audiometers

MODEL ELECTRONIC PATZIG ELECTRONIC PATZIG ENTOMED ENTOMEDMA 33 MA 33-1 GSI 67 GSI 68

OTHER SPECIFICATIONS RS232 port for PCconnection; matrix display;monaural, dichotic, oripsilateral talkover speechsystem. Meets requirementsof IEC 645 and ISO 389.

RS232 port for PCconnection; matrix display;monaural, dichotic, oripsilateral talkover speechsystem. Meets requirementsof IEC 645 and ISO 389.

 Automatic Type 3audiometer; built-in real-time clock and talk-forwardmicrophone; pulsed tone;masking noise tracking thetone level; softwarecalibration; memory for 50audiograms; patient IDnumber entered by numerickeypad; RS232 interface toPC or printer; optionalcarrying case, optional free-field; patient responseswitch supplied; user-programmable setup menu.Meets requirements of ANSIand IEC.

 Automatic Type 3audiometer; built-in real-time clock and talk-forwardmicrophone; pulsed tone;masking noise tracking thetone level; softwarecalibration; memory for 50audiograms; patient IDnumber entered by numerickeypad; RS232 interface toPC or printer; optionalcarrying case, optional free-field; patient responseswitch supplied; talk-backmicrophone supplied;monitor headset supplied;optional live speech monitor

headset; user-programmable setup menu;LED-VU meter in speechsection. Meets requirementsof ANSI and IEC.

LAST UPDATED December 20021  December 20021  January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes 1Model listed is currentlymarketed.

1aModel listed is currentlymarketed.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 42: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 42/116

Audiometers; Audiometric Booths

42 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ENTOMED ENTOMED ENTOMED EYMASASA 201 SA 202/GSI 66 SA 50 Audios 310

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Wide range

Speech No No No YesCONFIGURATION Tabletop/portable Tabletop/portable Handheld TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 125-8,000 500-4,000 125-8,000 (octave bands)Bone conduction NA NA NA 250-6,000 (octave bands)Continuous or step Step Step Step Continuous, step, warbleFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 NA -10 to +1001,000 Hz -10 to +110 -10 to +110 +20 to +60 -10 to +120

8,000 Hz -10 to +100 -10 to +100 NA -10 to +90Graduations 5 5 10 0.5, 1, 5

Bone conduction250 Hz NA NA NA -10 to +451,000 Hz NA NA NA -10 to +704,000 Hz NA NA NA -10 to +70Graduations NA NA NA 0.5, 1, 5

Masking, noise type NA NA NA Narrow band, white, speechIntensity, dB SPL NA NA NA Not specified

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 43: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 43/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 43

Chart A: Audiometers

MODEL ENTOMED ENTOMED ENTOMED EYMASASA 201 SA 202/GSI 66 SA 50 Audios 310

SPEECH AUDIOMETRYInputs NA NA NA Mic, CDIntensity, dB HL NA NA NA -10 to +90Masking, noise type NA NA NA Narrow band, white, speech

Intensity, dB SPL NA NA NA -10 to +90EARPHONE TYPE TDH-39 or Sennheiser

HDA-200, EARtone 5Aseparated outputs

TDH-39 or SennheiserHDA-200, EARtone 5Aseparated outputs

TDH-39 TDH-39

Cushion type Supra-aural (HDA-200circumaural)

Supra-aural (HDA-200circumaural)

Circumaural Circumaural

BONE VIBRATORType NA NA NA B-71

SPECIAL TESTS None None None Weber, SISIPOWER REQUIREMENTS

Line, VAC 115/230 115/230 NA 220, 50/60 Hz; 110, 50/60Hz

Battery None None 6/F22 Not specified

Type, Volts NA NA 9 Not specifiedW x D x H, cm (in) 38.4 x 26.9 x 7.4 (15.1 x

10.6 x 2.9)38.4 x 26.9 x 7.4 (15.1 x10.6 x 2.9)

15.2 x 7.6 x 2.5 (6 x 3 x 1) 46 x 43 x 17 (18.1 x 16.9 x6.6)

WEIGHT, kg (lb) 2 (4.4) 2 (4.4) 0.2 (0.4) 12 (26.4)PURCHASEINFORMATION

List price Not specified Not specified Not specified Not specifiedWarranty 2 year 2 year 1 year 1 yearDelivery time, ARO Stock, low quantities Stock, low quantities Stock, low quantities Not specifiedFiscal year January to December January to December January to December Not specified

This is the second of threepages covering the abovemodel(s). These

specifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 44: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 44/116

Audiometers; Audiometric Booths

44 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL ENTOMED ENTOMED ENTOMED EYMASASA 201 SA 202/GSI 66 SA 50 Audios 310

OTHER SPECIFICATIONS Built-in talk-forwardmicrophone; pulsed tone;software calibration;memory for 1 audiogram;optional carrying case,patient response switchsupplied. Meetsrequirements of ANSI andIEC.

 Automatic Type 3audiometer; built-in real-time clock and talk-forwardmicrophone; pulsed tone;software calibration;memory for 50 audiograms;patient ID number enteredby numeric keypad; RS232interface to PC or printer;optional carrying case,patient response switchsupplied; user-programmable setup menu.Meets requirements of ANSIand IEC.

 Automatic switchoff; low-battery indicator; standardheadset; optional patientresponse switch. Meetsrequirements of ANSI andIEC.

 Air and bone conduction;vocal audiometry; automatictests; modulated signal;patient response control;examiners, headphones,and microphones.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 45: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 45/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 45

Chart A: Audiometers

MODEL EYMASA EYMASA EYMASA EYMASA Audios 320 Audios 330 REDUS 75 REDUS 85

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Limited range

Speech Yes Yes No NoCONFIGURATION Tabletop Tabletop Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-20,000 (octave bands) 250-20,000 (octave bands) 250-8,000 (octave bands) 250-8,000 (octave bands)Bone conduction 250-6,000 (octave bands) 250-6,000 (octave bands) NA NAContinuous or step Continuous, step, warble Continuous pulse, warble Continuous, step Continuous, stepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +120 -10 to +120 -10 to +100 -10 to +100

8,000 Hz -10 to +90 -10 to +90 -10 to +100 -10 to +100Graduations 0.5, 1, 5 0.5, 1, 5 5 5

Bone conduction250 Hz -10 to +45 -10 to +45 NA NA1,000 Hz -10 to +70 -10 to +70 NA NA4,000 Hz -10 to +70 -10 to +70 NA NAGraduations 0.5, 1, 5 0.5, 1, 5 NA NA

Masking, noise type Narrow band, white, speech Narrow band, white, speech NA NAIntensity, dB SPL Not specified Not specified NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 46: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 46/116

Audiometers; Audiometric Booths

46 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL EYMASA EYMASA EYMASA EYMASA Audios 320 Audios 330 REDUS 75 REDUS 85

SPEECH AUDIOMETRYInputs Mic, CD Mic, CD NA NAIntensity, dB HL -10 to +110 -10 to +110 NA NAMasking, noise type Narrow band, white, speech Narrow band, white, speech NA NA

Intensity, dB SPL -10 to +90 NA NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type Circumaural Circumaural Circumaural CircumauralBONE VIBRATOR

Type B-71 B-71 NA NASPECIAL TESTS Weber, SISI, DLI Weber, SISI, DLI, SAL,

Stenger, tone decayNone Screening and abbreviated

ascendant automatic testsPOWER REQUIREMENTS

Line, VAC 220, 50/60 Hz; 110, 50/60Hz

220, 50/60 Hz; 110, 50/60Hz

220, 50/60 Hz; 110, 50/60Hz

220, 50/60 Hz; 110, 50/60Hz

Battery Not specified Not specified Not specified Not specifiedType, Volts Not specified Not specified Not specified Not specified

W x D x H, cm (in) 46 x 43 x 17 (18.1 x 16.9 x

6.6)

46 x 43 x 17 (18.1 x 16.9 x

6.6)

42.9 x 22.1 x 8.4 (16.9 x 8.7

x 3.3)

42.9 x 22.1 x 8.4 (16.9 x 8.7

x 3.3)WEIGHT, kg (lb) 12 (26.4) 12 (26.4) 6 (13.2) 6 (13.2)PURCHASEINFORMATION

List price Not specified Not specified Not specified Not specifiedWarranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 47: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 47/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 47

Chart A: Audiometers

MODEL EYMASA EYMASA EYMASA EYMASA Audios 320 Audios 330 REDUS 75 REDUS 85

OTHER SPECIFICATIONS  Air and bone conduction;vocal audiometry; automatictests; modulated signal;patient response control;examiners, headphones,and microphones.

 Air and bone conduction;vocal audiometry; free field;automatic tests; modulatedsignal; patient responsecontrol; examiners,headphones, andmicrophones; optional highfrequency, tinnitus, insertearphones, externalCD/tape, and vocal.

Optional RS232 interface,printer, carrying case, andPC software. Meetsrequirements of ANSI, IEC601 and 645, and ISO 389.

Optional RS232 interface,printer, carrying case, andPC software. Meetsrequirements of ANSI, IEC601 and 645, and ISO 389.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 48: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 48/116

Audiometers; Audiometric Booths

48 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL EYMASA FRYE FRYE GM INSTRUMENTSREDUS 95 FA-10 FA-12 ASRA

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Limited range Wide range Wide range Wide range

Speech No Yes Yes NoCONFIGURATION Tabletop/portable Tabletop Tabletop PortablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 250-8,000 (octave bands) 125-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands)

125-12,000

Bone conduction 250-6,000 250-6,000 (octave,interoctave bands)

250-6,000 (octave,interoctave bands)

NA

Continuous or step Continuous, step Step Step StepFrequency accuracy Not specified Within 1% Not specified ±0.1%

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +90 -10 to +90 -20 to +901,000 Hz -10 to +100 -10 to +110 -10 to +120 -20 to +1208,000 Hz -10 to +100 -10 to +90 -10 to +90 -20 to +100Graduations 5 5 5 10

Bone conduction250 Hz -10 to +40 -10 to +45 -10 to +45 NA1,000 Hz -10 to +60 -10 to +60 -10 to +70 NA4,000 Hz -10 to +60 -10 to +60 -10 to +60 NAGraduations 5 5 5 NA

Masking, noise type NA Narrow band, white, speech Narrow band, white, speech NAIntensity, dB SPL NA 110 white; others vary 110 white; others vary NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 49: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 49/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 49

Chart A: Audiometers

MODEL EYMASA FRYE FRYE GM INSTRUMENTSREDUS 95 FA-10 FA-12 ASRA

SPEECH AUDIOMETRYInputs NA Mic, external (L, R) Mic, external (L, R) NAIntensity, dB HL NA +95 +95 NAMasking, noise type NA Speech Speech NA

Intensity, dB SPL NA -10 to +95 -10 to +105 NAEARPHONE TYPE TDH-39 TDH only, insert earphones

only, or dual calibrationTDH only, insert earphonesonly, or dual calibration

TDH-39-51

Cushion type Circumaural Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type B-71 Radioear B-71 Radioear B-71 NASPECIAL TESTS Screening and abbreviated

ascendant automatic testsStenger, ABLB, MLB, SISIare free upon request

 ABLB, MLB, SISI, Stengerare free upon request

Manual/semiautomatic andautomatic Hughson-Westlake; pre-/postcomparisons andstatistics

POWER REQUIREMENTSLine, VAC 220, 50/60 Hz; 110, 50/60

Hz

105-132 105-240 100-110/220-240

Battery Not specified None None NoneType, Volts Not specified NA NA NA

W x D x H, cm (in) 42.9 x 22.1 x 8.4 (16.9 x 8.7x 3.3)

46.4 x 34.3 x 14 (18.3 x13.5 x 5.5)

46.4 x 34.3 x 14 (18.3 x13.5 x 5.5)

27 x 6 x 28 (10.63 x 2.36 x11.02)

WEIGHT, kg (lb) 6 (13.2) 5 (11) 5 (11) 1 (2.2)PURCHASEINFORMATION

List price Not specified $3,295 TDH only; $3,400inserts only; $3,575 dualcalibration

$3,295 TDH only; $3,400inserts only; $3,575 dualcalibration

$2,900

Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified 2 weeks 2 weeks 2 weeksFiscal year Not specified April to March April to March January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 50: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 50/116

Audiometers; Audiometric Booths

50 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL EYMASA FRYE FRYE GM INSTRUMENTSREDUS 95 FA-10 FA-12 ASRA

OTHER SPECIFICATIONS Optional RS232 interface,printer, and carrying case.Meets requirements of

 ANSI, IEC 601 and 645, andISO 389.

Frequency-specific free-fieldcalibration; master hearing-aid curves; optional RS232interface, carrying case andfree-field speakers; optional

 ABLB, SISI, MLB package;dual calibration availablewith insert earphones.Meets requirements of ANSI3.42.

Frequency-specific free-fieldcalibration; master hearing-aid curves; optional RS232interface, carrying case andfree-field speakers; optional

 ABLB, SISI, MLB package;dual calibration availablewith insert earphones.Meets requirements of ANSIand ISO.

Baseline and age-relatedassessments; interfaceswith any IBM or compatiblePC running on Windows3.1, 95, 98, NT, 2000, orXP. Meets requirements ofIEC 601-1 and UL.

LAST UPDATED January 2007 July 2005 July 2005 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 51: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 51/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 51

Chart A: Audiometers

MODEL GRASON-STADLER GRASON-STADLER GRASON-STADLER GRASON-STADLERGSI 17 GSI 38 GSI 61 GSI 66

WHERE MARKETED Worldwide Worldwide Worldwide Worldwide, except USAFDA CLEARANCE Yes Yes Yes NoCE MARK (MDD) Yes Yes Yes YesTYPE Limited range Limited range Wide range Wide range

Speech No No Yes NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)125-8,000 125-12,000 (octave,

interoctave bands)125-8,000

Bone conduction NA NA 250-6,000 (octave,interoctave bands)

NA

Continuous or step Continuous, pulsed Step Both StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +70 -10 to +70 -10 to +90 -10 to +1001,000 Hz -10 to +90 -10 to +90 -10 to +120 -10 to +1108,000 Hz -10 to +70 -10 to +70 -10 to +100 -10 to +100Graduations 5 5 1, 2, 5 5

Bone conduction250 Hz NA NA -10 to +45 NA1,000 Hz NA NA -10 to +70 NA4,000 Hz NA NA -10 to +60 NAGraduations NA NA 1, 2, 5 NA

Masking, noise type NA NA Narrow band, white NAIntensity, dB SPL NA NA -10 to +110 (HL) NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 52: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 52/116

Audiometers; Audiometric Booths

52 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL GRASON-STADLER GRASON-STADLER GRASON-STADLER GRASON-STADLERGSI 17 GSI 38 GSI 61 GSI 66

SPEECH AUDIOMETRYInputs NA NA Mic, tape, CD NAIntensity, dB HL NA NA -10 to +105 NAMasking, noise type NA NA White, speech NA

Intensity, dB SPL NA NA -10 to +115 NAEARPHONE TYPE TDH-39 TDH-39 TDH-50P TDH-39, EARtone 3A

Cushion type Supra-aural Supra-aural Supra-aural Circumaural or supra-auralBONE VIBRATOR

Type NA NA B-71 NASPECIAL TESTS None specified None specified ABLB, Doerfler-Stewart,

SISI, Stenger Automatic test

POWER REQUIREMENTSLine, VAC 100, 120, 240 100, 120, 240 100, 120, 240 115/230Battery Optional Ni-Cd None None None

Type, Volts NA NA NA NAW x D x H, cm (in) 33.7 x 35.6 x 9.5 (13.3 x 14

x 3.8)33.4 x 36.8 x 10.9 (13.2 x14.5 x 4.3)

48.3 x 45.7 x 20.3 (19 x 18x 8)

38 x 27 x 7 (15 x 10.6 x 2.8)

WEIGHT, kg (lb) 2.5 (5.6) 4.5 (10) 10 (22) 2 (4.4)PURCHASEINFORMATION

List price $810-929 $2,750-3,550 $5,710-8,000 Not specifiedWarranty 1 year from date of

shipment1 year from date ofshipment

1 year from date ofshipment

1 year

Delivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December Not specified

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 53: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 53/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 53

Chart A: Audiometers

MODEL GRASON-STADLER GRASON-STADLER GRASON-STADLER GRASON-STADLERGSI 17 GSI 38 GSI 61 GSI 66

OTHER SPECIFICATIONS Battery pack. Meetsrequirements of ANSI, IEC,and ISO 9001.

1 sec tympanometry; built-inprinter; ipsilateral andcontralateral reflex testing;many versions available.Meets requirements of

 ANSI, IEC, and ISO 9001.

Optional free-field speakers,built-in amplifiers, computer-interface printer port, high-frequency audiometry andinsert phones, HP DeskJetprinter, sound-field speakerplus booster amplification.Meets requirements of

 ANSI, IEC, and ISO 9001.

Built-in talk-forward mic;optional carrying case, 50-audiogram memory withpatient ID, softwarecalibration, and RS232 port.

LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 54: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 54/116

Audiometers; Audiometric Booths

54 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL GRASON-STADLER GRASON-STADLER INTERACOUSTICS INTERACOUSTICSGSI 67 GSI 68 AA220/AA222 AC33

WHERE MARKETED Worldwide, except USA Worldwide, except USA Worldwide WorldwideFDA CLEARANCE No No Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Wide range

Speech No Yes Yes, AA222 YesCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 125-8,000 125-8,000 125-8,000Bone conduction 250-8,000 250-8,000 250-8,000 250-8,000Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified 1% >±2%

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +110 -10 to +1101,000 Hz -10 to +110 -10 to +110 -10 to +120 -10 to +120

8,000 Hz -10 to +100 -10 to +100 -10 to +110 -10 to +110Graduations 5 5 1, 2, 5 1 or 5

Bone conduction250 Hz -10 to +35 -10 to +35 -10 to +45 -10 to +451,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +80 -10 to +80 -10 to +80 -10 to +80Graduations 5 5 1 or 5 1 or 5

Masking, noise type Narrow band Narrow band Narrow band, white (AA222,speech)

Narrow band, white

Intensity, dB SPL 0 to +95 0 to +95 0 to +110 (dB HL) 0 to +110 (dB HL)

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 55: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 55/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 55

Chart A: Audiometers

MODEL GRASON-STADLER GRASON-STADLER INTERACOUSTICS INTERACOUSTICSGSI 67 GSI 68 AA220/AA222 AC33

SPEECH AUDIOMETRYInputs NA Mic, tape, CD (2-channel) AA222, CD, Mic, tape CD, mic, tapeIntensity, dB HL NA -10 to +100 AA222, -10 to +120 -10 to +120Masking, noise type NA Speech AA, speech noise Speech noise

Intensity, dB SPL NA 0-100 AA222, 0 to +110 (dB HL) 0 to +110 (dB HL)EARPHONE TYPE TDH-39, EARtone 3A TDH-39, EARtone 3A TDH-39 TDH-39

Cushion type Circumaural or supra-aural Circumaural or supra-aural MX-41/AR MX-41/ARBONE VIBRATOR

Type B-71 B-71 B-71 B-71SPECIAL TESTS  Automatic test Automatic test SISI, ABLB, automatic

threshold (HW and Bekesy),Stenger (binaural pure-tonestimulation), AA222 alsoincludes binaural speech

SISI, ABLB, MLD, automaticthreshold (HW), Stenger(binaural pure-tonestimulation), binauralspeech

POWER REQUIREMENTSLine, VAC 115/230 115/230 100-120 or 200-240, 50/60

Hz100-120 or 200-240, 50/60Hz

Battery None None None NoneType, Volts NA NA NA NA

W x D x H, cm (in) 38 x 27 x 7 (15 x 10.6 x 2.8) 48 x 27 x 7 (18.9 x 10.6 x2.8)

35 x 26 x 10 (14 x 10.2 x 4) 50 x 47 x 20 (20 x 19 x 8)

WEIGHT, kg (lb) 2 (4.4) 2.7 (6) 3.3 (7.1) 9 (20)PURCHASEINFORMATION

List price Not specified Not specified Not specified Not specifiedWarranty 1 year 1 year 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified January to December January to December

This is the second of threepages covering the above

model(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 56: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 56/116

Audiometers; Audiometric Booths

56 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL GRASON-STADLER GRASON-STADLER INTERACOUSTICS INTERACOUSTICSGSI 67 GSI 68 AA220/AA222 AC33

OTHER SPECIFICATIONS Built-in talk-forward mic;optional carrying case, 50-audiogram memory withpatient ID, softwarecalibration, and RS232 port.

Built-in talk-forwardmicrophone; optionalcarrying case, 50-audiogram memory withpatient ID, and softwareRS232 port.

Diagnostic impedanceaudiometer; built-in thermalprinter; different headsetand probe holder solutions;clinical probe system;optional carrying case,warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking;RS232C or USBconnection; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring and

printing; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.

Warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking;RS232C or USBconnection; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; NOAH compatible.Meets requirements of ANSIand IEC.

LAST UPDATED January 2004 January 2004 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 57: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 57/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 57

Chart A: Audiometers

MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AC40 AC440/Equinox AD226 AD229e

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Wide range Wide range

Speech Yes Yes Yes YesCONFIGURATION Tabletop Black box Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-20,000 125-8,000 125-8,000 125-8,000Bone conduction 250-8,000 250-8,000 250-8,000 250-8,000Continuous or step Step Step Step StepFrequency accuracy > ±1% >±1% >±1% >±1%

Intensity, dB HLAir conduction

250 Hz -10 to +110 -10 to +110 -10 to +110 -10 to +1101,000 Hz -10 to +120 -10 to +120 -10 to +120 -10 to +120

8,000 Hz -10 to +110 -10 to +110 -10 to +110 -10 to +110Graduations 1 or 5 1 or 5 1 or 5 1 or 5

Bone conduction250 Hz -10 to +45 -10 to +45 -10 to +45 -10 to +451,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +80 -10 to +80 -10 to +80 -10 to +80Graduations 1 or 5 1 or 5 1 or 5 1 or 5

Masking, noise type Narrow band, white Narrow band, white Narrow band, white Narrow band, whiteIntensity, dB SPL 0 to +110 (dB HL) 0 to +110 (dB HL) 0 to +110 (dB HL) 0 to +110 (dB HL)

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 58: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 58/116

Audiometers; Audiometric Booths

58 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AC40 AC440/Equinox AD226 AD229e

SPEECH AUDIOMETRYInputs CD, mic, tape CD, mic, tape NA CD, mic, tapeIntensity, dB HL -10 to +120 -10 to +120 NA -10 to +120Masking, noise type Speech noise Speech NA Speech

Intensity, dB SPL 0 to +110 (dB HL) 0 to +110 (dB HL) NA 0 to +110 (dB HL)EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type MX-41/AR MX-41/AR MX-41/AR MX-41/ARBONE VIBRATOR

Type B-71 B-71 B-71 B-71SPECIAL TESTS SISI, ABLB, MLD, automatic

threshold (HW and Bekesy),Stenger (binaural pure-tonestimulation), binauralspeech, HF and volumescaling

 ABLB, Stenger (binauralpure-tone stimulation),binaural speech, HF, MF,SISI, MLD,, Weber, MasterHearing Aid, Hearing LossSimulation

SISI, ABLB, automaticthreshold (HW and Bekesy),Stenger (binaural pure-tonestimulation)

SISI, ABLB, automaticthreshold (HW and Bekesy),Stenger (binaural pure-tonestimulation), binauralspeech, MHA

POWER REQUIREMENTSLine, VAC 100-120 or 200-240, 50/60

Hz

100-120 or 200-240, 50/60

Hz

100-120 or 200-240, 50/60

Hz

100-120 or 200-240, 50/60

HzBattery None None None None

Type, Volts NA NA NA NAW x D x H, cm (in) 50 x 47 x 20 (20 x 19 x 8) 37 x 43.5 x 7.7 (14.5 x 17 x

3)35 x 26 x 10 (14 x 10 x 4) 48 x 40 x 15 (19 x 16 x 6)

WEIGHT, kg (lb) 13 (29) 5.1 (11.3) 2.1 (4.6) 2.5 (5)PURCHASEINFORMATION

List price Not specified Not specified Not specified Not specifiedWarranty 2 years 2 years 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December

This is the second of three

pages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 59: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 59/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 59

Chart A: Audiometers

MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AC40 AC440/Equinox AD226 AD229e

OTHER SPECIFICATIONS Warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking;RS232C or USBconnection, internal printeror output for laser printer;Windows 98 software;IaBaseII database,PrintView for online PCmonitoring and printing;NOAH compatible. Meetsrequirements of ANSI andIEC.

PC audiometer to beoperated via PC keyboardor audiometric keyboard;warble, talk-forward 0-110dB SPL, talk-back, EARtone5A insert phones (withoutEQFF in speech), HDA280(with EQFF in speech),HDA200 separatelycalibrated, synchronousmasking, and USBconnection; Windows 98software; IaBaseIIdatabase; NOAHcompatible. Meetsrequirements of ANSI andIEC.

Optional carrying case,warble, talk-forward 0-110dB SPL, talk-back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking,RS232C or USBconnection, and output forMTP10 printer, laser printer,and inkjet; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; NOAH andCONNEX compatible.Meets requirements of ANSI

and IEC.

Optional carrying case,warble, talk-forward 0-110dB SPL, talk back, monitorwith built-in loudspeaker,EARtone 3A insert phones,separately calibrated,synchronous masking,RS232C or USBconnection, and output forMTP10 printer, laser printer,and inkjet; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; NOAH andCONNEX compatible.Meets requirements of ANSI

and IEC.LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 60: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 60/116

Audiometers; Audiometric Booths

60 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AS208 AT235 AT235h AZ26

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Limited range Limited range Limited range Limited range

Speech No No No NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 125-8,000 125-8,000 125-8,000Bone conduction NA NA NA NAContinuous or step Step Step Step StepFrequency accuracy >±2% 1% 1% 2%

Intensity, dB HLAir conduction

250 Hz -10 to +90 -10 to +110 -10 to +110 -10 to +901,000 Hz -10 to +100 -10 to +120 -10 to +120 -10 to +100

8,000 Hz -10 to +100 -10 to +100 -10 to +110 -10 to +100Graduations 5 1, 2, 5 1, 2, 5 1, 2, 5

Bone conduction250 Hz NA NA NA NA1,000 Hz NA NA NA NA4,000 Hz NA NA NA NAGraduations NA NA NA NA

Masking, noise type NA NA NA NAIntensity, dB SPL NA NA NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 61: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 61/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 61

Chart A: Audiometers

MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AS208 AT235 AT235h AZ26

SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA

Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type MX-41/AR MX-41/AR MX-41/AR MX-41/ARBONE VIBRATOR

Type NA NA NA NASPECIAL TESTS None Automatic threshold (HW),

tympanometry, automaticand manual reflexometry,decay test, eustachian-tubefunction test

 Automatic threshold (HW),normal and high probe-tonetympanometry, automaticand manual reflexometry,decay test, eustachian-tubefunction test

Not specified

POWER REQUIREMENTSLine, VAC Optional 100-120 or 200-

240, 50/60 Hz

100-120 or 200-240, 50/60

Hz

100-120 or 200-240, 50/60

Hz

100-120 or 200-240, 50/60

HzBattery  AA batteries (4) None None None

Type, Volts 9-12 NA NA NAW x D x H, cm (in) 23 x 18 x 8 (9 x 7 x 3) 36 x 26 x 10 (14.2 x 10.2 x

4)36 x 26 x 10 (14.2 x 10.2 x4)

48 x 40 x 16 (19 x 16 x 6)

WEIGHT, kg (lb) 1.3 (2.9); includes batteriesand headset

2.8 (6.2) 2.8 (6.2) 9.5 (21)

PURCHASEINFORMATION

List price Not specified Not specified Not specified Not specifiedWarranty 2 years 2 years 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 62: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 62/116

Audiometers; Audiometric Booths

62 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS INTERACOUSTICS AS208 AT235 AT235h AZ26

OTHER SPECIFICATIONS Pure, warble, and pulsedtones available; patientsignal button; batteryindicator for testingreliability; optional carryingcase, power supply (100-120 or 200-240 V). Meetsrequirements of ANSI andIEC.

Diagnostic impedanceaudiometer; optionalcarrying case; built-inthermal printer; differentheadset and probe holdersolutions; clinical probesystem; EARtone 3A insertphotos; phones; Windows98 software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.

Diagnostic impedanceaudiometer; optionalcarrying case; built-inthermal printer; differentheadset and probe holdersolutions; clinical probesystem; EARtone 3A insertphones; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.

Clinical impedanceaudiometer; built-in thermalprinter; clinical probesystem; Windows 98software; IaBaseIIdatabase, PrintView foronline PC monitoring andprinting; Interacousticsimpedance module forinterfacing with NOAH.Meets requirements of ANSIand IEC.

LAST UPDATED January 2007 January 2007 January 2007 January 2007

Supplier FootnotesModel Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 63: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 63/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 63

Chart A: Audiometers

MODEL MADSEN MADSEN MADSEN MADSENConera ITERA MM 304 MM 602

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range Limited range Wide range

Speech Yes High frequency No YesCONFIGURATION Portable/desktop, computer

drivenTabletop/portable Portable Tabletop/portable

PURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)250-12,500 (12 frequencies) 250-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands)

Bone conduction 250-8,000 (octave,interoctave bands)

250-8,000 (8 frequencies) NA 250-6,000 (octave,interoctave bands)

Continuous or step Continuous/step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HL

Air conduction250 Hz -10 to +100 -10 to +110 -10 to +90 -10 to +1001,000 Hz -10 to +125 -10 to +120 -10 to +90 -10 to +1258,000 Hz -10 to +100 -10 to +100 -10 to +90 -10 to +100Graduations 1, 5 1, 2.5, 5 5 5

Bone conduction250 Hz -10 to +45 -10 to +40 NA -10 to +451,000 Hz -10 to +75 -10 to +80 NA -10 to +754,000 Hz -10 to +75 -10 to +75 NA -10 to +75Graduations 1, 2, 3, 5, 10 1, 2.5, 5 NA 5

Masking, noise type White, speech Narrow band, white NA Narrow bandIntensity, dB SPL +40 to +90 -10 to +110 NA -10 to +115

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 64: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 64/116

Audiometers; Audiometric Booths

64 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MADSEN MADSEN MADSEN MADSENConera ITERA MM 304 MM 602

SPEECH AUDIOMETRYInputs Mic, tape, CD Tape, CD, mic NA Mic, tape 1, tape 2Intensity, dB HL -10 to +115 -10 to +80 NA -10 to +115Masking, noise type Narrow band, white, speech White, speech, ipsilateral,

contralateralNA White, speech

Intensity, dB SPL -10 to +115 -10 to +80 NA -10 to +115EARPHONE TYPE TDH-39, ER-3A TDH-39 TDH-39 TDH-39

Cushion type Not specified Circumaural Circumaural Supra-auralBONE VIBRATOR

Type B-71 B-71 NA B-71SPECIAL TESTS  Automatic threshold, SISI,

 ABLB, sound-level check,Stenger, Bekesy

Fowler (ABLB), SISI,Stenger, hearing instrumentstimulator

None SISI, Stenger, Fowler

POWER REQUIREMENTSLine, VAC 100/115/127/220/240 110-240, 50/60 Hz NA 100-120/200-240Battery None None Yes None

Type, Volts NA NA 9 NA

W x D x H, cm (in) 38.1 x 35 x 13 (15 x 13.8 x5.1)

45 x 29 x 8.5 (17 x 11.4 x3.3)

30.5 x 21.6 x 6.4 (12 x 8.5 x2.5)

38.1 x 39.4 x 10.9 (15 x15.5 x 4.3)

WEIGHT, kg (lb) 4.6 (10.1) 4.5 (9.9) 1.3 (2.8) 3.9 (8.6)PURCHASEINFORMATION

List price $4,995 $3,995 $795 $2,995Warranty 2 years 2 years 2 years 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue onto

the next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 65: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 65/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 65

Chart A: Audiometers

MODEL MADSEN MADSEN MADSEN MADSENConera ITERA MM 304 MM 602

OTHER SPECIFICATIONS Computer required. Optional carrying case,extended high-frequencyoption to 16,000 Hz.

Standard ME70 headset;optional AC adapter. Meetsrequirements of ANSI andIEC.

Computer memory,calibration, and test setups;audible patient response.Meets requirements of ANSIand IEC.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 66: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 66/116

Audiometers; Audiometric Booths

66 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MADSEN MADSEN MAICO MAICOMM 622 OB 922 MA 25 Series II MA 27

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes No NoTYPE Wide range Wide range, high frequency Limited range Limited range

Speech Yes Yes No NoCONFIGURATION Tabletop/portable Tabletop Portable PortablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)125-20,000 (octave,interoctave bands)

250-8,000 (octave bands) 250-8,000 (octave bands)

Bone conduction 250-6,000 (octave,interoctave bands)

250-8,000 (octave,interoctave bands)

NA NA

Continuous or step Step Step/continuous Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +90 -15 to +901,000 Hz -10 to +125 -10 to +125 -10 to +90 -15 to +908,000 Hz -10 to +100 -10 to +100 -10 to +90 -15 to +90Graduations 1, 5 1, 5 5 5

Bone conduction250 Hz -10 to +45 -10 to +45 NA NA1,000 Hz -10 to +75 -10 to +75 NA NA4,000 Hz -10 to +75 -10 to +75 NA NAGraduations 1, 5 1, 2, 3, 5, 10 NA NA

Masking, noise type Narrow band Narrow band, speech,white, external

NA NA

Intensity, dB SPL -10 to +115 -10 to +115 NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 67: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 67/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 67

Chart A: Audiometers

MODEL MADSEN MADSEN MAICO MAICOMM 622 OB 922 MA 25 Series II MA 27

SPEECH AUDIOMETRYInputs Mic, tape 1, tape 2 Mic, tape, CD NA NAIntensity, dB HL -10 to +115 -10 to +115 NA NAMasking, noise type White, speech Narrow band, white,

speech, externalNA NA

Intensity, dB SPL -10 to +115 -10 to +115 NA NAEARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type B-71 B-71 NA NASPECIAL TESTS SISI, Stenger, Fowler ABLB, DLI, SISI, Stenger,

autothreshold,suprathreshold, Rainville,tone decay, Bekesy

None specified None specified

POWER REQUIREMENTSLine, VAC 100-120/200-240 110/115/127/220/240 Optional 120 adapter 117; 220 optionalBattery None None Yes None

Type, Volts NA NA Not specified NAW x D x H, cm (in) 38.1 x 39.4 x 10.9 (15 x

15.5 x 4.3)52.1 x 41.9 x 15.2 (20.5 x16.5 x 6)

21.6 x 19.1 x 9.1 (8.5 x 7.5x 3.6)

35.6 x 43.2 x 17.8 (14 x 17x 7)

WEIGHT, kg (lb) 3.9 (8.6) 9.1 (20) 1.8 (4) 2.7 (6)PURCHASEINFORMATION

List price $4,500 $5,995 $825 $915Warranty 2 years 2 years 1 year 1 yearDelivery time, ARO Not specified Not specified 10 days 10 daysFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). These

specifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 68: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 68/116

Audiometers; Audiometric Booths

68 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MADSEN MADSEN MAICO MAICOMM 622 OB 922 MA 25 Series II MA 27

OTHER SPECIFICATIONS Computer memory,calibration, and test setups;audible patient response;masking and test signalsgiven simultaneously. Meetsrequirements of ANSI andIEC.

Computer interface;bidirectional software; 2screen languages; clock;12-test memory; optionalprinter. Meets requirementsof ANSI and IEC.

Battery operated withoptional 120 VAC adapter;carrying case. Meetsrequirements of ANSI andIEC.

Built-in carrying case. Meetsrequirements of ANSI andIEC.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 69: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 69/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 69

Chart A: Audiometers

MODEL MAICO MAICO MAICO MAICOMA 39 MA 40 MA 41 MA 42

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) No No No NoTYPE Limited range Wide range Wide range Wide range

Speech No No Yes YesCONFIGURATION Portable Portable Portable Portable/tabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave bands) 125-8,000 (octave bands) 125-8,000 (octave bands) 125-8,000Bone conduction NA 125-4,000 (octave bands) 125-4,000 (octave bands) 125-4,000Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +110 -10 to +110 -10 to +110 -10 to +110

8,000 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +100Graduations 5 5 5 5

Bone conduction250 Hz NA -10 to +40 -10 to +40 -10 to +401,000 Hz NA -10 to +70 -10 to +70 -10 to +704,000 Hz NA -10 to +70 -10 to +70 -10 to +70Graduations NA 5 5 5

Masking, noise type NA Narrow band Narrow band, speech Narrow band, speechIntensity, dB SPL NA -10 to +100 -10 to +100 -10 to +100

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 70: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 70/116

Audiometers; Audiometric Booths

70 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MAICO MAICO MAICO MAICOMA 39 MA 40 MA 41 MA 42

SPEECH AUDIOMETRYInputs NA NA Mic, tape, CD Mic, tape, CDIntensity, dB HL NA NA -10 to +100 -10 to +100Masking, noise type NA NA Narrow band, speech Narrow band, speech

Intensity, dB SPL NA NA -10 to +100 -10 to +100EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type NA B-71 B-71 B-71SPECIAL TESTS None specified None specified None specified None specifiedPOWER REQUIREMENTS

Line, VAC 117/234 117/234 117/234 117/234Battery None None None None

Type, Volts NA NA NA NAW x D x H, cm (in) 32 x 40 x 16 (12.5 x 15.5 x

6.3)32 x 40 x 16 (12.5 x 15.5 x6.3)

32 x 40 x 16 (12.5 x 15.5 x6.3)

32 x 40 x 16 (12.5 x 15.5 x6.3)

WEIGHT, kg (lb) 7.5 (16.5) 7.5 (16.5) 7.5 (16.5) 7.5 (16.5)

PURCHASEINFORMATION

List price $1,235 $1,725 $2,100 $2,495Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 10 days 10 days 10 days 10 daysFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 71: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 71/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 71

Chart A: Audiometers

MODEL MAICO MAICO MAICO MAICOMA 39 MA 40 MA 41 MA 42

OTHER SPECIFICATIONS Digital display. Meetsrequirements of ANSI andIEC.

Digital display. Meetsrequirements of ANSI andIEC.

Digital display. Meetsrequirements of ANSI andIEC.

Master hearing aid with 4commonly used frequencyslopes to imitate hearing-aidacoustics; 2-channel speechtesting; built-in amplifier.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 72: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 72/116

Audiometers; Audiometric Booths

72 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MAICO MAICO MEDIROLL MEDIROLLMA 790 MA 800 Rollingtone AT-61 SA-51

WHERE MARKETED Worldwide Worldwide Not specified Not specifiedFDA CLEARANCE Yes Yes No NoCE MARK (MDD) No No Not specified YesTYPE Limited range Limited range Not specified Wide range

Speech No No Optional NoCONFIGURATION Tabletop Tabletop Tabletop/portable Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 500-8,000 (octave bands) 500-8,000 (octave bands) 125-10,000 250-8,000Bone conduction NA NA 250-8,000 NAContinuous or step NA NA Step StepFrequency accuracy Not specified Not specified <1% <1%

Intensity, dB HLAir conduction

250 Hz NA NA -10 to +90 -10 to +801,000 Hz 0 to +95 0 to +95 -10 to +120 -10 to +100

8,000 Hz 0 to +95 0 to +95 -10 to +100 -10 to +80Graduations 5 5 5 or 1 5

Bone conduction250 Hz NA NA -10 to +45 NA1,000 Hz NA NA -10 to +70 NA4,000 Hz NA NA -10 to +60 NAGraduations NA NA 5 or 1 NA

Masking, noise type NA NA NBN; WN; SN NAIntensity, dB SPL NA NA -10 to +100(AC); -10 to

+60(BC); NBC -10 to+110(AC); -10 to +70(BC);WN -10 to +110(AC);SN -10to +70(BC) SN

NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 73: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 73/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 73

Chart A: Audiometers

MODEL MAICO MAICO MEDIROLL MEDIROLLMA 790 MA 800 Rollingtone AT-61 SA-51

SPEECH AUDIOMETRYInputs NA NA CD or built-in speech test NAIntensity, dB HL NA NA -10 to +110 (AC); -10 to +65

(BC)NA

Masking, noise type NA NA NBN; WN; SN NAIntensity, dB SPL NA NA -10 to +100 (AC); -10 to +60

(BC); NBC -10 to +110(AC); -10 to +70(BC) WN; -10 to +110-(AC); -10 to +70(BC) SN

NA

EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39Cushion type Supra-aural Supra-aural Not specified Not specified

BONE VIBRATORType NA NA B-71 NA

SPECIAL TESTS None specified None specified TTDT; SBLB; SISI; ABLB;Lucher; Bekesy; automatictest; built-in speech test;

"Melody" test

 Automatic test

POWER REQUIREMENTSLine, VAC 117/234, 50/60 Hz 117/234, 50/60 Hz 110-240 V; 50-60 Hz 230, 50 HzBattery None None NA None

Type, Volts NA NA NA NAW x D x H, cm (in) 37.5 x 32.5 x 10 (15 x 13 x

4)37.5 x 32.5 x 10 (15 x 13 x4)

27.5 x 19.5 x 5 (10.8 x 5.9 x2)

19.5 x 15 x 5 (7.7 x 5.9 x 2)

WEIGHT, kg (lb) 4.5 (10) 4.5 (10) 1 (2) 0.45 (1)PURCHASEINFORMATION

List price $3,225 $3,795  €4,100  EUR 1,396Warranty 1 year 1 year 2 years 2 yearsDelivery time, ARO 10 days 10 days 60 days 40 daysFiscal year January to December January to December Not specified Not specified

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 74: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 74/116

Audiometers; Audiometric Booths

74 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MAICO MAICO MEDIROLL MEDIROLLMA 790 MA 800 Rollingtone AT-61 SA-51

OTHER SPECIFICATIONS  Automatic computer ormanual testing; LCD;thermal printouts; RS232 ormodem interface. Meetsrequirements of ANSI andIEC.

 Automatic computer ormanual testing; LCD;thermal printouts; RS232 ormodem interface; stores upto 500 audiograms andbaselines in memory. Meetsrequirements of ANSI andIEC.

 Aplitutde and frequencymodulation; syncronousmasking; 3 seperatechanels, LCD, free-fieldaudiometry with specialamplifier, computerinterface; mains adapter,carrying case; extra smallsize; Meets requirements ofISO 389, EN 60645, andIEC 60601.

Noise-excluding headset;stores audiograms for 99patients; carrying case;LCD; computer interface;built-in mic; built-in watch.Meets requirements of IEC645-4.

LAST UPDATED July 2005 July 2005 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 75: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 75/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 75

Chart A: Audiometers

MODEL MEDIROLL MICRO AUDIOMETRICS MICRO AUDIOMETRICS MICRO AUDIOMETRICSSA-6 Mini Earscan 3 Automatic

Threshold Audiometer

(ES3)

Earscan 3 Manual Audiometer (ES3M)

Earscan 3 Screening Audiometer (ES3S)

WHERE MARKETED Not specified Worldwide Worldwide WorldwideFDA CLEARANCE No Yes Yes YesCE MARK (MDD) Yes No No NoTYPE Wide range Wide range Limited range Wide range

Speech No No No NoCONFIGURATION Portable Handheld, tabletop, portable Handheld, tabletop, portable Handheld, tabletop, portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 250-8,000 125-8,000 250-8,000 125-8,000Bone conduction NA NA NA NAContinuous or step Step Step Step StepFrequency accuracy <1% <1% <1% <1%

Intensity, dB HLAir conduction

250 Hz -10 to +70 -10 to +90 -10 to +80 -10 to +801,000 Hz -10 to +100 -10 to +100 -10 to +80 -10 to +808,000 Hz -10 to +90 -10 to +90 -10 to +80 -10 to +80Graduations 5 5 5 5

Bone conduction250 Hz NA NA NA NA1,000 Hz NA NA NA NA4,000 Hz NA NA NA NAGraduations NA NA NA NA

Masking, noise type NA No No NoIntensity, dB SPL NA NA NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 76: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 76/116

Audiometers; Audiometric Booths

76 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MEDIROLL MICRO AUDIOMETRICS MICRO AUDIOMETRICS MICRO AUDIOMETRICSSA-6 Mini Earscan 3 Automatic

Threshold Audiometer

(ES3)

Earscan 3 Manual Audiometer (ES3M)

Earscan 3 Screening Audiometer (ES3S)

SPEECH AUDIOMETRYInputs NA NA NA NAIntensity, dB HL NA NA NA NAMasking, noise type NA NA NA NA

Intensity, dB SPL NA NA NA NAEARPHONE TYPE TDH-39 TDH-39, EAR-5A TDH-39, EAR-5A TDH-39, EAR-5A

Cushion type Not specified Supra-aural, inserts Supra-aural Supra-aural, insertsBONE VIBRATOR

Type NA NA NA NASPECIAL TESTS  Automatic test; "melody"

test for children (patented) Automatic threshold,binaural presentations

Binaural presentations Automatic screening,binaural presentations

POWER REQUIREMENTSLine, VAC 230, 50 Hz 100-240 100-240 100-240Battery None AA alkaline (4) AA alkaline (4) AA alkaline (4)

Type, Volts NA 5 5 5

W x D x H, cm (in) 12.3 x 7.1 x 3.5 (4.8 x 2.8 x1.4)

9.1 x 3.3 x 14.7 (3.6 x 1.3 x5.8)

9.1 x 3.3 x 14.7 (3.6 x 1.3 x5.8)

9.1 x 3.3 x 14.7 (3.6 x 1.3 x5.8)

WEIGHT, kg (lb) 0.2 (0.4) 0.28 (0.6125) 0.28 (0.6125) 0.28 (0.6125)PURCHASEINFORMATION

List price EUR 1,099 $1,995 $795 $995Warranty 2 years 3 years 3 years 3 yearsDelivery time, ARO 40 days 7-10 days 7-10 days 7-10 daysFiscal year Not specified June to May June to May June to May

This is the second of threepages covering the abovemodel(s). Thesespecifications continue onto

the next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 77: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 77/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 77

Chart A: Audiometers

MODEL MEDIROLL MICRO AUDIOMETRICS MICRO AUDIOMETRICS MICRO AUDIOMETRICSSA-6 Mini Earscan 3 Automatic

Threshold Audiometer

(ES3)

Earscan 3 Manual Audiometer (ES3M)

Earscan 3 Screening Audiometer (ES3S)

OTHER SPECIFICATIONS Noise-excluding headset;stores audiograms for 50patients; carrying case;LCD. Meets requirements ofIEC 645-4.

Talkover; RS-232, USB;optional printer, datastorage; STS and PBIcalculations; batteries andcarrying case included.Meets requirements of ANSIS3.6, IEC, and ISO. MeetsOSHA requirements foraudiometers

Talkover; Batteries andcarrying case included.Meets requirements of ANSIS3.6, IEC, and ISO.

Talkover; RS-323, USB;batteries and carrying caseincluded. Meetsrequirements of ANSI S3.6,IEC, and ISO.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 78: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 78/116

Audiometers; Audiometric Booths

78 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MONITOR MONITOR NAGASHIMA NAGASHIMAMI-5000B MI-7000 AD-98 MT-3A

WHERE MARKETED Canada, USA Canada, USA Worldwide WorldwideFDA CLEARANCE Not specified Not specified Not specified Not specifiedCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Limited range Limited range Not specified Wide range

Speech No No Yes YesCONFIGURATION Tabletop/portable Tabletop/portable Tabletop Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 500-8,000 500-8,000 125-8,000 (octave bands) 125-8,000 (octave bands)Bone conduction NA NA 250-8,000 250-6,000Continuous or step Step Step Step StepFrequency accuracy 1% 1% Not specified Not specified

Intensity, dB HLAir conduction

250 Hz NA NA -10 to +90 -10 to +901,000 Hz 0 to +90 0 to +90 -10 to +110 -10 to +100

8,000 Hz 0 to +90 0 to +90 -10 to +90 -10 to +90Graduations 5 5 5 5

Bone conduction250 Hz NA NA -10 to +55 -10 to +501,000 Hz NA NA -10 to +70 -10 to +604,000 Hz NA NA -10 to +60 -10 to +50Graduations NA NA 5 5

Masking, noise type No No Narrow band, SPN Narrow band, SPNIntensity, dB SPL NA NA -10 to +110 +40 to +100

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 79: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 79/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 79

Chart A: Audiometers

MODEL MONITOR MONITOR NAGASHIMA NAGASHIMAMI-5000B MI-7000 AD-98 MT-3A

SPEECH AUDIOMETRYInputs NA NA Auxiliary AuxiliaryIntensity, dB HL NA NA 0 to +110 -10 to +100Masking, noise type NA NA Narrow band, SPN, white Narrow band, SPN, white

Intensity, dB SPL NA NA -10 to +110 +40 to +100EARPHONE TYPE TDH-49 TDH-49 TDH-49 TDH-49

Cushion type MX-41/AR MX-41/AR 51 51BONE VIBRATOR

Type NA NA BR-572 BR-572SPECIAL TESTS None None SISI, ABLB NonePOWER REQUIREMENTS

Line, VAC 120/240 100-250 100-120, 220-240, 50/60 Hz 100-120, 220-240, 50/60 HzBattery None None None None

Type, Volts NA NA NA NAW x D x H, cm (in) 18.4 x 21.6 x 7 (7.3 x 8.5 x

2.8)24 x 29 x 7.5 (9.5 x 11.5 x3)

42 x 43 x 31.7 (16.5 x 17 x12.5)

21 x 23 x 8.5 (8.3 x 9.1 x3.3)

WEIGHT, kg (lb) 1.4 (3) 2.3 (5) 7.5 (16.5) 3.8 (8.3)

PURCHASEINFORMATION

List price $1,995 $2,795 Not specified Not specifiedWarranty 2 years 2 years 1 year 1 yearDelivery time, ARO 3 weeks 3 weeks 30 days 30 daysFiscal year January to December January to December July to June July to June

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 80: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 80/116

Audiometers; Audiometric Booths

80 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL MONITOR MONITOR NAGASHIMA NAGASHIMAMI-5000B MI-7000 AD-98 MT-3A

OTHER SPECIFICATIONS RS232 interface. RS232 interface; talkover;internal printer.

Type 1 diagnosticaudiometer. Conforms toJIS T1201-1982.

Type 1 diagnosticaudiometer. Conforms toJIS T1201-1982.

LAST UPDATED July 2005 July 2005 April 2001 April 2001Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 81: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 81/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 81

Chart A: Audiometers

MODEL NAGASHIMA QUALITONE QUALITONE QUALITONENSA-100S AAJ-P : C-AAJ AS-110 Screener AS-90 Screener

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Not specified Yes Yes YesCE MARK (MDD) Not specified Yes Yes YesTYPE Not specified Wide range, ANSI Type 2C Wide range, ANSI Type 4 Wide range, ANSI Type 4

Speech No Yes No NoCONFIGURATION Tabletop/portable Portable Portable PortablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 500-4,000 (octave bands) 125-8,000 (octave,

interoctave bands)250-8,000 (octave,interoctave bands)

250-8,000 (octave,interoctave bands)

Bone conduction No 250-4,000 (octave,interoctave bands)

NA NA

Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz NA -10 to +100 -10 to +100 -10 to +901,000 Hz -10 to +70 -10 to +110 -10 to +110 -10 to +908,000 Hz NA -10 to +100 -10 to +100 -10 to +90Graduations 5 2.5, 5 5 5

Bone conduction250 Hz NA -10 to +70 NA NA1,000 Hz NA -10 to +70 NA NA4,000 Hz NA -10 to +70 NA NAGraduations NA 2.5, 5 NA NA

Masking, noise type No Narrow band, white, speech NA NAIntensity, dB SPL NA -10 to +120 NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 82: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 82/116

Audiometers; Audiometric Booths

82 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL NAGASHIMA QUALITONE QUALITONE QUALITONENSA-100S AAJ-P : C-AAJ AS-110 Screener AS-90 Screener

SPEECH AUDIOMETRYInputs No Mic, tape, auxiliary NA NAIntensity, dB HL No -10 to +110 NA NAMasking, noise type No Narrow band, white, speech NA NA

Intensity, dB SPL NA -10 to +120 NA NAEARPHONE TYPE TDR-30 TDH-39 TDH-39 TDH-39

Cushion type 41 Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type NA B-71A, B-72 NA NASPECIAL TESTS None None None NonePOWER REQUIREMENTS

Line, VAC 100-120, 220-240, 50/60 Hz 110/220 110/220 110/220Battery None None None Alkaline

Type, Volts NA NA NA 6; 55 hr life (maximum @ 8hr/day)

W x D x H, cm (in) 26.5 x 22 x 10 (10.4 x 8.7 x4)

48.3 x 35.6 x 22.9 (19 x 14x 9) : 48.3 x 35.6 x 14 (19 x

14 x 5.5)

33 x 33 x 15.2 (13 x 13 x 6) 22.9 x 20.3 x 10.2 (9 x 8 x4)

WEIGHT, kg (lb) 3 (6.6) 11.8 (26) : 9.1 (20.1) 4.1 (9) 1.8 (4)PURCHASEINFORMATION

List price Not specified $3,950; $3840 $865 $790Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 30 days Not specified Not specified Not specifiedFiscal year July to June January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 83: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 83/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 83

Chart A: Audiometers

MODEL NAGASHIMA QUALITONE QUALITONE QUALITONENSA-100S AAJ-P : C-AAJ AS-110 Screener AS-90 Screener

OTHER SPECIFICATIONS Screening audiometer.Conforms to JIS TR01-1982.

Free-field capability; masterhearing aid; 2 channels;built-in cassette deck. Meetsrequirements of ANSI Type2C.

Continuous pulsed-tonewarble (FM); patient signal;reversible tone-interrupterswitch; group-presentationcapability. Meetsrequirements of ANSI Type4.

Continuous, pulsed-tonewarble; patient signal;reversible tone-interrupterswitch. Meets requirementsof ANSI Type 4.

LAST UPDATED  April 2001 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 84: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 84/116

Audiometers; Audiometric Booths

84 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL QUALITONE QUALITONE QUALITONE SIBELMEDCD 2 : CD 3 LWAAJ-P WR-A : WR-B : WR-C AC50-C

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes NoCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Wide range, ANSI Type 2C Wide range Wide range

Speech Yes Yes Yes Yes (optional)CONFIGURATION Portable Portable Portable TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)125-8,000 (octave,interoctave bands)

125-8,000 (octave,interoctave bands)

125-8,000

Bone conduction 250-4,000 (octave,interoctave bands)

250-4,000 (octave,interoctave bands)

250-4,000 (octave,interoctave bands)

250-4,000

Continuous or step Step Step Step StepFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to [+110 : +120] -10 to +110 -10 to [+110 : +120] -10 to +1208,000 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +100Graduations 2.5, 5 2.5, 5 2.5, 5 5

Bone conduction250 Hz -10 to +70 -10 to +70 -10 to +45 -10 to +501,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +70Graduations 2.5, 5 2.5, 5 2.5, 5 5

Masking, noise type Narrow band, white, speech Narrow band, white, speech Narrow band, white, speech Narrow band, whiteIntensity, dB SPL -10 to +120 -10 to +120 -10 to +120 -10 to +100

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 85: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 85/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 85

Chart A: Audiometers

MODEL QUALITONE QUALITONE QUALITONE SIBELMEDCD 2 : CD 3 LWAAJ-P WR-A : WR-B : WR-C AC50-C

SPEECH AUDIOMETRYInputs Mic, CD, auxiliary Mic, auxiliary Mic, auxiliary Tape, CD, mic (optional)Intensity, dB HL -10 to +110 -10 to +110 -10 to +110 -10 to +100Masking, noise type Narrow band, white, speech Narrow band, white, speech White, speech, narrow band Speech

Intensity, dB SPL -10 to +120 -10 to +125 0 to +125 : 0 to +120 : 0 to+120

-10 to +100

EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39Cushion type Supra-aural Supra-aural Supra-aural Supra-aural

BONE VIBRATORType B-71A, B-72 B-71A, B-72 B-71A, B-72 B-71

SPECIAL TESTS None SISI optional None ABLB, Luscher, Stenger,warble, SISI (optional)

POWER REQUIREMENTSLine, VAC 110/220 110/220 110/220 220 (others optional)Battery None None None Not specified

Type, Volts NA NA NA Not specifiedW x D x H, cm (in) 48.3 x 35.6 x 22.9 (19 x 14

x 9)

48.3 x 35.6 x 12.7 (19 x 14

x 5)

33 x 33 x 15.2 (13 x 13 x 6) 47 x 25 x 9.5 (18.5 x 9.8 x

3.7)WEIGHT, kg (lb) 12.2 (27) 7.8 (17.2) 4.5 (10) 4.5 (9.9)PURCHASEINFORMATION

List price $4,750 : $4,995 $3,375 $1,285 : $1,750 : $1,920 $1,400Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified 7 daysFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 86: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 86/116

Audiometers; Audiometric Booths

86 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL QUALITONE QUALITONE QUALITONE SIBELMEDCD 2 : CD 3 LWAAJ-P WR-A : WR-B : WR-C AC50-C

OTHER SPECIFICATIONS Free-field capability; masterhearing aid; 2 channels;channel reverse; patientsignal; electronic frequencyswitching; channel mixing;talkover, talk-back. Meetsrequirements of ANSI.

Free-field capability; masterhearing aid; 2 channels.Meets requirements of ANSIType 2C.

1 channel; reversible toneinterrupter; master hearingaid; patient signal warble.Meets requirements of

 ANSI.

External printer; internaldatabase, hearing-losscalculation; AM; FM; 2channels; optional PCdatabase and free field.Meets requirements of IEC601-1 and 645 and ISO9001.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 87: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 87/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 87

Chart A: Audiometers

MODEL SIBELMED SIBELMED SIBELMED SIBELMED AC50-D AS5-A AS5-AOM SIBELSOUND 400 A

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesTYPE Wide range Limited range Limited range

Speech Yes No No Yes (optional)CONFIGURATION Tabletop Portable Portable TabletopPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 250-8,000 250-8,000 125-8,000Bone conduction 250-4,000 NA 250-4,000 250-4,000Continuous or step Step Step Step BothFrequency accuracy Not specified Not specified Not specified 2%

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +90 -10 to +90 -10 to +1001,000 Hz -10 to +120 -10 to +105 -10 to +105 -10 to +120

8,000 Hz -10 to +100 -10 to +80 -10 to +80 -10 to +100Graduations 5 5 5 1 or 5 (selectable)

Bone conduction250 Hz -10 to +50 NA -10 to +40 -10 to +501,000 Hz -10 to +70 NA -10 to +50 -10 to +704,000 Hz -10 to +70 NA -10 to +50 -10 to +70Graduations 5 NA 5 1 or 5 (selectable)

Masking, noise type Narrow band, white White White Narrow band, white, speechIntensity, dB SPL -10 to +100 +30 to +80 +30 to +80 -10 to +100

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 88: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 88/116

Audiometers; Audiometric Booths

88 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL SIBELMED SIBELMED SIBELMED SIBELMED AC50-D AS5-A AS5-AOM SIBELSOUND 400 A

SPEECH AUDIOMETRYInputs Tape, CD, mic NA NA Tape, CD, mic (optional)Intensity, dB HL -10 to +100 NA NA -10 to +100Masking, noise type Speech NA NA Speech

Intensity, dB SPL -10 to +100 NA NA -10 to +100EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type B-71 NA B-71 B-71SPECIAL TESTS  ABLB, Luscher, Stenger,

warble, SISINone None ABLB, Luscher, Stenger,

warble, SISI (optional)POWER REQUIREMENTS

Line, VAC 220 (other options) 220 (others optional) 220 (others optional) 110-220Battery Not specified Not specified Not specified No

Type, Volts Not specified Not specified Not specified NoW x D x H, cm (in) 47 x 25 x 9.5 (18.5 x 9.8 x

3.7)32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)

32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)

47 x 25 x 9.5 (18.5 x 9.8 x3.7)

WEIGHT, kg (lb) 4.5 (9.9) 2.5 (5.5) 2.5 (5.5) 4.5 (9.9)PURCHASEINFORMATION

List price $1,600 $750 $1,100 $1,400Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 7 days 7 days 7 days 7 daysFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 89: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 89/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 89

Chart A: Audiometers

MODEL SIBELMED SIBELMED SIBELMED SIBELMED AC50-D AS5-A AS5-AOM SIBELSOUND 400 A

OTHER SPECIFICATIONS External printer; internaldatabase; hearing-losscalculation; AM; FM; 2channels; optional PCdatabase and free field.Meets requirements of IEC601-1 and 645 and ISO9001.

Optional masking. Meetsrequirements of ANSI 53.6,IEC 601-1, and ISO 9001.

Meets requirements of ANSI53.6, IEC 601-1, and ISO9001.

USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1-1, ISO 9001, and ISE13485.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 90: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 90/116

Audiometers; Audiometric Booths

90 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL SIBELMED SIBELMED SIBELMED SIBELMEDSIBELSOUND 400 AM SIBELSOUND 400 AO SIBELSOUND 400 AOM SIBELSOUND 400 AOM+

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesTYPE

Speech Yes (optional) Yes (optional) Yes (optional) YesCONFIGURATION Tabletop Portable Portable PortablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 125-8,000 125-20,000 125-20,000Bone conduction 250-4,000 NA 250-4,000 250-4,000Continuous or step Both Both Both BothFrequency accuracy 2% 2% 2% 2%

Intensity, dB HLAir conduction

250 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +1001,000 Hz -10 to +120 -10 to +120 -10 to +120 -10 to +120

8,000 Hz -10 to +100 -10 to +100 -10 to +100 -10 to +100Graduations 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable)

Bone conduction250 Hz -10 to +50 -10 to +50 -10 to +50 -10 to +501,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +704,000 Hz -10 to +70 -10 to +70 -10 to +70 -10 to +70Graduations 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable) 1 or 5 (selectable)

Masking, noise type Narrow band, white, speechnoise

Narrow band, white, speechnoise

Narrow band, white, speechnoise

Narrow band, white, speechnoise

Intensity, dB SPL -10 to +100 -10 to +100 -10 to +100 -10 to +100

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 91: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 91/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 91

Chart A: Audiometers

MODEL SIBELMED SIBELMED SIBELMED SIBELMEDSIBELSOUND 400 AM SIBELSOUND 400 AO SIBELSOUND 400 AOM SIBELSOUND 400 AOM+

SPEECH AUDIOMETRYInputs Tape, CD, mic (optional) Tape, CD, mic (optional) Tape, CD, mic (optional) Tape, CD, micIntensity, dB HL -10 to +100 -10 to +100 -10 to +100 -10 to +100Masking, noise type Speech Speech Speech Speech

Intensity, dB SPL -10 to +100 -10 to +100 -10 to +100 -10 to +100EARPHONE TYPE TDH-39 TDH-39 TDH-39 TDH-39

Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type B-71 B-71 B-71 B-71SPECIAL TESTS  ABLB, Luscher, Stenger,

warble, SISI (optional) ABLB, Luscher, Stenger,warble, SISI (optional)

 ABLB, Luscher, Stenger,warble, SISI (optional)

 ABLB, Luscher, Stenger,warble, SISI (optional)

POWER REQUIREMENTSLine, VAC 110-240 110-240 110-240 110-240Battery No No No No

Type, Volts No No No NoW x D x H, cm (in) 47 x 25 x 9.5 (18.5 x 9.8 x

3.7)32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)

32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)

32.4 x 24.6 x 9.5 (12.8 x 9.7x 3.7)

WEIGHT, kg (lb) 4.5 (9.9) 2.5 (5.5) 2.5 (5.5) 2.5 (5.5)PURCHASEINFORMATION

List price $1,600 $750 $1,100 $1,100Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 7 days 7 days 7 days 7 daysFiscal year January to December January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 92: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 92/116

Audiometers; Audiometric Booths

92 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL SIBELMED SIBELMED SIBELMED SIBELMEDSIBELSOUND 400 AM SIBELSOUND 400 AO SIBELSOUND 400 AOM SIBELSOUND 400 AOM+

OTHER SPECIFICATIONS USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.

USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.

USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM,FM); optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.

USB connection andprinting; internal database,hearing-loss calculation; 2channels (AM, FM); optionalPC database and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 93: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 93/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 93

Chart A: Audiometers

MODEL SIBELMED SINGER SINGER SINGERSIBELSOUND 400 SUPRA MD-1 MD-3 MD-4

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No Yes Yes YesCE MARK (MDD) Yes No No NoTYPE Limited range Limited range Wide range

Speech Yes No No NoCONFIGURATION Portable Portable Tabletop Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-20,000 500-6,000 (octave plus

6,000)500-6,000 (octave,interoctave bands)

125-8,000 (octave,interoctave bands)

Bone conduction 250-4,000 NA NA NAContinuous or step Both Step Step StepFrequency accuracy 2% Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +100 NA NA -10 to +100

1,000 Hz -10 to +120 0 to +100 0 to +100 -10 to +1108,000 Hz -10 to +100 NA NA -10 to +100Graduations 1 or 5 (selectable) 5 5 5

Bone conduction250 Hz -10 to +50 NA NA NA1,000 Hz -10 to +70 NA NA NA4,000 Hz -10 to +70 NA NA NAGraduations 1 or 5 (selectable) NA NA NA

Masking, noise type Narrow band, white, speechnoise

NA NA NA

Intensity, dB SPL -10 to +100 NA NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 94: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 94/116

Audiometers; Audiometric Booths

94 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL SIBELMED SINGER SINGER SINGERSIBELSOUND 400 SUPRA MD-1 MD-3 MD-4

SPEECH AUDIOMETRYInputs Tape, CD, mic NA NA NAIntensity, dB HL -10 to +100 NA NA NAMasking, noise type Speech NA NA NA

Intensity, dB SPL -10 to +100 NA NA NAEARPHONE TYPE TDH-39 TDH-50 TDH-50 TDH-49

Cushion type Supra-aural Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type B-71 NA NA NASPECIAL TESTS  ABLB, Luscher, Stenger,

warble, SISI (optional)None None Pulse

POWER REQUIREMENTSLine, VAC 110-240 105-125; optional 220 105-125; optional 220 105-125; optional 220Battery No Not specified Not specified Not specified

Type, Volts No Not specified Not specified Not specifiedW x D x H, cm (in) 32.4 x 24.6 x 9.5 (12.8 x 9.7

x 3.7)45.7 x 33 x 14 (18 x 13 x5.5)

48.3 x 25.4 x 27.9 (19 x 10x 11)

38.1 x 17.8 x 6.4 (15 x 7 x2.5)

WEIGHT, kg (lb) 2.5 (5.5) 7.7 (17) 11.3 (25) 1.6 (3.5)PURCHASEINFORMATION

List price $1,100 $2,295 $2,695 $695Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 7 days Not specified Not specified Not specifiedFiscal year January to December Not specified Not specified Not specified

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 95: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 95/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 95

Chart A: Audiometers

MODEL SIBELMED SINGER SINGER SINGERSIBELSOUND 400 SUPRA MD-1 MD-3 MD-4

OTHER SPECIFICATIONS USB connection, printing;internal database, hearing-loss calculation; 2 channels(AM, FM), optional PCdatabase and free field.Meets requirements of IEC60645-1:2000, IEC60645.2:1993, EN60601-1,EN60601-1-1, ISO 9001,and ISO 13485.

Includes automatictympanometer; price withoptional printer is $2,700.Meets requirements of

 ANSI.

Includes automatictympanometer; price withoptional printer is $2,995.Meets requirements of

 ANSI.

Model draws audiograms;price with optional printer is$995. Meets requirementsof ANSI.

LAST UPDATED January 2007 January 2004 January 2004 January 2004Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 96: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 96/116

Audiometers; Audiometric Booths

96 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL STARKEY TREMETRICS TREMETRICS TREMETRICS AA 50 RA 300 : 3000 RA 500 RA 650

WHERE MARKETED Not specified Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes No No NoTYPE  ANSI Type 1A Not specified Limited range Limited range

Speech Yes No No NoCONFIGURATION Tabletop/portable Tabletop/portable Tabletop/portable Tabletop/mobile vanPURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 500-8,000 (octave,

interoctave bands)500-8,000 (octave,interoctave bands)

500-8,000 (octave,interoctave bands)

Bone conduction 250-4,000 NA NA NAContinuous or step Step Both Both BothFrequency accuracy Not specified Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +110 NA NA NA

1,000 Hz -10 to +120 0 to +100 -10 to +100 -10 to +1008,000 Hz -10 to +110 0 to +100 -10 to +100 -10 to +100Graduations 1 to 5 5 5 5

Bone conduction250 Hz -10 to +45 NA NA NA1,000 Hz -10 to +70 NA NA NA4,000 Hz -10 to +80 NA NA NAGraduations 1 to 5 NA NA NA

Masking, noise type Narrow band, white, speech NA NA NAIntensity, dB SPL 0 to +100 NA NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 97: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 97/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 97

Chart A: Audiometers

MODEL STARKEY TREMETRICS TREMETRICS TREMETRICS AA 50 RA 300 : 3000 RA 500 RA 650

SPEECH AUDIOMETRYInputs Mic, CD, auxillary NA NA NAIntensity, dB HL 0 to +100 NA NA NAMasking, noise type Speech NA NA NA

Intensity, dB SPL 0 to +100 NA NA NAEARPHONE TYPE TDH-39, 3A, 5A TDH-39 TDH-39 TDH-39

Cushion type Circumaural/insert Supra-aural Supra-aural Supra-auralBONE VIBRATOR

Type B-71 NA NA NASPECIAL TESTS  ABLB, Stenger, SISI,

automatic threshold,Langenbeck, dual intensity

None specified : 1-buttonexpress test mode

None None

POWER REQUIREMENTSLine, VAC 100-240, automatic

switching power supply90-240, selectable 120/240 90-240, selectable

Battery None None None NoneType, Volts NA NA NA NA

W x D x H, cm (in) 44 x 51 x 17 (17.3 x 20 x6.3)

33 x 28 x 11.5 (13 x 11 x4.5)

44.5 x 31.1 x 10.2 (17.5 x12.3 x 4)

16.7 x 31.8 x 22.9 (6.6 x12.5 x 9)

WEIGHT, kg (lb) Not specified 2.7 (6.0) 4.3 (9.5) 2.9 (6.4)PURCHASEINFORMATION

List price $6,500 $1,895 $4,295 Depends on number ofchannels

Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Not specified Not specified Not specified Not specifiedFiscal year Not specified January to December January to December January to December

This is the second of threepages covering the abovemodel(s). These

specifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 98: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 98/116

Audiometers; Audiometric Booths

98 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL STARKEY TREMETRICS TREMETRICS TREMETRICS AA 50 RA 300 : 3000 RA 500 RA 650

OTHER SPECIFICATIONS Master hearing aid;frequency field; printingoption; hearing losssimulator; automaticscoring; talk-forward, talk-back; patient signal; twinmics; external audio input.

 Additional frequencies: 200,2500, 3500, 4500, 5000 Hz.NOAH certified. Meetsrequirements of ANSI S3.6Type 1 and EN 60601-1.

Patented multimediacapability for computer-voice-instructed testing;RS232 port; optionaltalkover. Meetsrequirements of ANSI andISO 9001.

 Automatic, semiautomatic,and manual test; 25customizable test questions;memory for 1,200audiograms;telecommunicationssoftware; dual computer-interface ports. Meetsrequirements of ANSI andISO 9001.

Windows 95, 98, NTcontroller; group testingsystem; optional 1-4channels/module with up to8 channels per PCcontroller; audio/visual errorcodes; built-in talkover; rawdata transfer;telecommunicationssoftware; computer voiceinstructions in 5 languages;patented multimediacapability. Meetsrequirements of ANSI, ISO9001, and OSHA.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 99: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 99/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 99

Chart A: Audiometers

MODEL Welch Allyn Welch Allyn Welch Allyn AM232 AudioScope 3 TM262

WHERE MARKETED Worldwide Worldwide WorldwideFDA CLEARANCE Yes Not specified YesCE MARK (MDD) Yes Not specified YesTYPE Limited range Not specified Limited range

Speech No Not specified NoCONFIGURATION Tabletop/portable Not specified Tabletop/portablePURE-TONEAUDIOMETRY

Frequency, HzAir conduction 125-8,000 (octave,

interoctave bands)500, 1,000, 2,000, 4,000 125-8,000

Bone conduction NA NA NAContinuous or step Continuous, pulsed Continuous StepFrequency accuracy Not specified Not specified Not specified

Intensity, dB HLAir conduction

250 Hz -10 to +70 Not specified -10 to +70

1,000 Hz -10 to +90 20, 25, 40 -10 to +908,000 Hz -10 to +70 NA -10 to +70Graduations 5 Not specified 5

Bone conduction250 Hz NA NA NA1,000 Hz NA NA NA4,000 Hz NA NA NAGraduations NA NA NA

Masking, noise type NA NA NAIntensity, dB SPL NA NA NA

This is the first of threepages covering the abovemodel(s). Thesespecifications continue ontothe next two pages.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 100: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 100/116

Audiometers; Audiometric Booths

100 ©2007 ECRI Institute. All Rights Reserved.

Chart A: Audiometers

MODEL Welch Allyn Welch Allyn Welch Allyn AM232 AudioScope 3 TM262

SPEECH AUDIOMETRYInputs NA NA NAIntensity, dB HL NA NA NAMasking, noise type NA NA NA

Intensity, dB SPL NA NA NAEARPHONE TYPE TDH-39 NA TDH-39

Cushion type Supra-aural NA Supra-auralBONE VIBRATOR

Type NA NA NASPECIAL TESTS None specified None specified None specifiedPOWER REQUIREMENTS

Line, VAC 100, 120, 240 110, 220, 240, 240/260 100, 120, 240Battery Optional Ni-Cad Ni-Cad No

Type, Volts Not specified 3.5 V NAW x D x H, cm (in) 33.7 x 35.6 x 9.5 (13.3 x 14

x 3.8)3.7 x 21.6 (1.4 x 8.5) 33.4 x 36.8 x 10.9 (13.2 x

14.5 x 4.3)WEIGHT, kg (lb) 2.5 (5.6) 0.25 (0.55) 4.5 (10)

PURCHASEINFORMATION

List price $1,083 $678 $3,398-3,942Warranty 1 year 1 year 1 yearDelivery time, ARO 5 days 5 days 5 daysFiscal year January to December January to December January to December

This is the second of threepages covering the abovemodel(s). Thesespecifications continue ontothe next page.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 101: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 101/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 101

Chart A: Audiometers

MODEL Welch Allyn Welch Allyn Welch Allyn AM232 AudioScope 3 TM262

OTHER SPECIFICATIONS Meets requirements of ANSI, IEC, and ISO 9001.

Meets requirements of ANSI, CSA, ETL, IEC.

One tympanometry, built-inprinter, ipsilateral reflextesting, two versionsavailable. Meetsrequirements of ANSI, IEC,and ISO 9001.

LAST UPDATED July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 102: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 102/116

Audiometers; Audiometric Booths

102 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL ACOUSTAIR ACOUSTAIR ACOUSTAIR ACOUSTAIR A-1 A-2 A-3 Audicel

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specified

TYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type Optional Optional Optional OptionalVENTILATION, m³/min(ft³/min)

2.1 (75) 2.1 (75) 2.1 (75) 2.1 (75)

LIGHT PLS-2xgw PLS-2xgw PLS-2xgw PLS-2xgwNumber 1 1 2 1

AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Optional Optional Optional OptionalELECTRICAL OUTLET Optional Optional Optional OptionalW x D x H, m (ft)

Inside, exam 1 x 1 x 2 (3.3 x 3.3 x 6.8) 1.6 x 1 x 2 (5.1 x 3.3 x 6.8) 1.8 x 1.6 x 2 (5.9 x 5.1 x6.8)

1 x 1 x 2 (3.3 x 3.3 x 6.8)

Inside, control Not specified Not specified Not specified Not specified

Outside, overall 1.2 x 1.2 x 2.3 (4 x 4 x 7.6) 1.8 x 1.2 x 2.3 (5.8 x 4 x7.6) 2 x 1.8 x 2.3 (6.6 x 5.8 x7.6) 1.1 x 1.1 x 2.3 (3.6 x 3.6 x7.6)WINDOW

W x H, cm (in) 60 x 60 (24 x 24) 60 x 60 (24 x 24) 60 x 60 (24 x 24) 60 x 60 (24 x 24)WEIGHT, kg (lb) 700 (1,550) 900 (2,000) 1,200 (2,600) 600 (1,300)LIST PRICE Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO On request On request On request On requestFISCAL YEAR Not specified Not specified Not specified Not specifiedOTHER SPECIFICATIONS Radio shielding available;

other sizes possible.Radio shielding available;other sizes possible.

Radio shielding available;other sizes possible.

Radio shielding available;other sizes possible.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 103: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 103/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 103

Chart B: Audiometric Booths

MODEL ACOUSTIC ACOUSTIC ACOUSTIC ACOUSTIC142 Delta 143 Delta RE-125 : RE-126 RE-130

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Optional OptionalJACK PANEL

Number, type 10, stereo phono; 3- and 4-pin CJ; also 2.54 cm (1 in)pass-through

10, stereo phono; 3- and 4-pin CJ; also 2.54 cm (1 in)pass-through

4, stereo phono, pass-through, folding shelf

6, stereo phono

VENTILATION, m³/min(ft³/min)

5.7 (200) 5.7 (200) 2.8 (100) 2.8 (100)

LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2 2 1 1

AUDIOMETER SHELF Optional Optional Yes YesCARPETING Yes Yes Yes YesELECTRICAL OUTLET 6-plug strip mount, external 6-plug strip mount, external None NoneW x D x H, m (ft)

Inside, exam 6 x 6 x 6.5 (15.2 x 15.2 x16.5) (triangular shape)

2.3 x 2.3 x 2 (7.4 x 7.4 x6.5) (triangular shape)

0.7 x 0.9 x 1.5 (2.3 x 3 x 5) 0.6 x 0.9 x 1.6 (2 x 3 x 5.2)

Inside, control NA NA NA NAOutside, overall 6.6 x 6.6 x 7.8 (17.6 x 17.6

x 19.8)2.4 x 2.4 x 2.3 (8 x 8 x 7.8) 0.8 x 1 x 1.8 (2.7 x 3.3 x 6) 1 x 0.8 x 2.01 (3.3 x 2.7 x

6.6)WINDOW

W x H, cm (in) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 45.7 x 61 (18 x 24) 45.7 x 61 (18 x 24)WEIGHT, kg (lb) 1,590 (3,500), shipping 1,633 (3,600), shipping 376 (830) 376 (830), shippingLIST PRICE $9,700 $12,425 $3,200 : $3,200 $4,500WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30-45 days 30-45 days 30 days 30 daysFISCAL YEAR October to September October to September October to September October to SeptemberOTHER SPECIFICATIONS Triangular unit designed for

corner fit in small host room;intended for sound-fieldtesting. Meets requirements

of ANSI.

Unit is triangular anddesigned to fit in a cornerfor sound-field testing in asmall host room. Meets

requirements of ANSI.

 All units preassembled;attenuation values for 125-8,000 Hz (octaves)available. Meets

requirements of ANSI.

 All units preassembled; 3"thick walls; attenuationvalues for 125-8,000 Hz(octaves) available. Meets

requirements of ANSI.LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 104: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 104/116

Audiometers; Audiometric Booths

104 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL ACOUSTIC ACOUSTIC ACOUSTIC ACOUSTICRE-132 RE-140 : RE-141 RE-142 RE-240

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Double wallCONFIGURATION Single room Single room Double room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

VENTILATION, m³/min(ft³/min)

5.7 (200) 2.8 (100) 5.7 (200) 2.8 (100)

LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2 1 2 1

AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 1 None 2; 1 duplex 6-plug strip 6-plug strip mounted

externally under windowW x D x H, m (ft)

Inside, exam 1.8 x 2 x 2.1 (6 x 6.5 x 7) 1 x 0.9 x 2 (3.3 x 3 x 6.5) :1.2 x 0.9 x 2 (4 x 3 x 6.5)

1.9 x 1.8 x 2 (6.3 x 6 x 6.5) 1 x 0.9 x 2 (3.3 x 3 x 6.5)

Inside, control NA NA NA NAOutside, overall 2 x 2.1 x 2.4 (6.5 x 7 x 7.8) 1.2 x 1.1 x 2.3 (4 x 3.7 x

7.7) : 1.4 x 1.2 x 2.3 (4.6 x 4x 7.7)

2.1 x 2 x 2.1 (6.9 x 6.7 x6.9)

1.6 x 1.5 x 2.5 (5.3 x 5 x8.1)

WINDOWW x H, cm (in) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24)

WEIGHT, kg (lb) 1,089 (2,400), shipping 748 (1,650), shipping 1,651 (3,640), shipping 1,814 (4,000), shippingLIST PRICE $8,650 $5,675 $10,700 $14,950WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30 days 30 days 30-45 days 30-45 daysFISCAL YEAR October to September October to September October to September October to SeptemberOTHER SPECIFICATIONS Meets requirements of

 ANSI.

Meets requirements of

 ANSI.

7 other models available in

varied sizes; optional radio-frequency shielding. Meetsrequirements of ANSI.

8 other models available in

varied sizes; optional radio-frequency shielding; largermodels have 2 to 4 lightsand 2 duplex electricaloutlets. Meets requirementsof ANSI.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 105: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 105/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 105

Chart B: Audiometric Booths

MODEL ACOUSTIC ACOUSTIC ACOUSTIC ACOUSTICRM-182 RM-282 RS-142 RS-242

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Double wall Single wall Double wallCONFIGURATION Single room with 4

examination stationsDouble room with 4examination stations

Double room Double room

VIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type NA1  NA1  10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

VENTILATION, m³/min(ft³/min)

5.7 (200) 5.7 (200) 5.7 (200), each room 5.7 (200), each room

LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2 2 2/room 2/room

AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET None None 2/room 2/roomW x D x H, m (ft)

Inside, exam 2.2 x 5.3 x 2 (7.3 x 5 x 6.5) 2.2 x 5.3 x 2 (7.3 x 5 x 6.5) 1.8 x 1.9 x 2 (6 x 6.3 x 6.5) 1.8 x 1.9 x 2 (6 x 6.3 x 6.5)Inside, control NA NA 1.2 x 1.9 x 2 (4 x 6.3 x 6.5) 1.2 x 1.9 x 2 (4 x 6.3 x 6.5)Outside, overall 2.4 x 1.7 x 2.2 (8 x 5.6 x

7.3)2.8 x 2.1 x 2.4 (9.3 x 6.9 x8)

3.5 x 2.1 x 2.3 (11.6 x 7 x7.4)

4 x 2.5 x 2.4 (13 x 8.3 x 8)

WINDOWW x H, cm (in) 61 x 33 (24 x 13) 61 x 45.7 (24 x 18) 76.2 x 61 (30 x 24) 76.2 x 61 (30 x 24)

WEIGHT, kg (lb) 1,706 (3,760), shipping 3,302 (7,280), shipping 2,989 (6,590), shipping 5,489 (12,100), shippingLIST PRICE $13,750 $22,000 $19,975 $35,425WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30-45 days 30-45 days 30-45 days 30-45 daysFISCAL YEAR October to September October to September October to September October to SeptemberOTHER SPECIFICATIONS Group testing; can seat/test

4 people simultaneously;metal stools and divider

curtains available; othermodels seat 2, 6, 8, or 10people. Meets requirementsof ANSI.

Group testing; can seat/test4 people simultaneously;metal stools and divider

curtains included; othermodels seat 2, 6, 8, or 10people. Meets requirementsof ANSI.

7 other models available invaried sizes; optional radio-frequency shielding. Meets

requirements of ANSI.

7 other models available invaried sizes; optional radio-frequency shielding. Meets

requirements of ANSI.

LAST UPDATED July 2005 July 2005 July 2005 July 2005Supplier Footnotes

Model Footnotes

Data Footnotes 1Each test station has anindividual jack box markedfor left and rightheadphones and aresponse button wired to anexternal patchbox.

1Each test station has anindividual jack box markedfor left and rightheadphones and aresponse button wired to anexternal patchbox.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 106: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 106/116

Audiometers; Audiometric Booths

106 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL ACOUSTIC AMPLIFON AMPLIFON AMPLIFONRS-252 AF100 AF111 AF210

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Yes Yes YesTYPE Single-wall control room,

double-wall examinationroom

Single wall Single wall Single wall

CONFIGURATION Double room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 1, Canon3-pin plug; also 2.54 cm (1in) pass-through

Jack and RCA for allaudiometers

Jack and RCA for allaudiometers

Jack and RCA for allaudiometers

VENTILATION, m³/min(ft³/min)

5.7 (200), each room Not specified Not specified Not specified

LIGHT Incandescent Incandescent Incandescent IncandescentNumber 2/room 1 1 1

AUDIOMETER SHELF Optional Optional Optional Optional

CARPETING Yes Yes Yes YesELECTRICAL OUTLET 2/room Yes Yes YesW x D x H, m (ft)

Inside, exam 1.9 x 1.8 x 2 (6.3 x 5.9 x6.5)

0.9 x 1 x 1.9 (3 x 3.3 x 6.2) 1.1 x 1.1 x 2.1 (3.6 x 3.6 x6.9)

2.3 x 1.1 x 2.1 (7.4 x 3.6 x6.9)

Inside, control 1.8 x 2.2 x 2.2 (5.9 x 7.2 x7.2)

NA NA NA

Outside, overall 4.5 x 2.4 x 2.4 (14.7 x 8 x 8) 1 x 1.1 x 2.2 (3.3 x 3.6 x 7) 1.3 x 1.3 x 2.3 (4.1 x 4.1 x7.4)

2.4 x 1.3 x 2.3 (7.8 x 4.1 x7.4)

WINDOWW x H, cm (in) 76.2 x 61 (30 x 24) 71 x 61 (28 x 24) 63 x 54 (25 x 21) 63 x 54 (25 x 21)

WEIGHT, kg (lb) 4,695 (10,350), shipping 300 (662) 260 (573) 380 (838)LIST PRICE $29,650 Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 30-45 days 60 days 60 days 60 daysFISCAL YEAR October to September January to December January to December January to December

OTHER SPECIFICATIONS 6 other models available invaried sizes; optional radio-frequency shielding. Meetsrequirements of ANSI.

None specified. None specified. None specified.

LAST UPDATED July 2005 December 2002 December 2002 December 2002Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 107: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 107/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 107

Chart B: Audiometric Booths

MODEL AMPLIFON AMPLIFON AMPLIFON AMPLIFON AF220 AF90 G1x1 G2x2

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Yes Yes Yes YesTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type Jack and RCA for allaudiometers

Jack and RCA for allaudiometers

Jack and RCA for allaudiometers

Jack and RCA for allaudiometers

VENTILATION, m³/min(ft³/min)

Not specified Not specified Not specified Not specified

LIGHT Incandescent Incandescent Incandescent IncandescentNumber 1 1 1 1

AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET Yes Yes Yes YesW x D x H, m (ft)

Inside, exam 2.3 x 2.3 x 2.1 (7.4 x 7.4 x6.9)

0.8 x 0.8 x 1.8 (2.7 x 2.7 x5.9)

1.1 x 1.1 x 2 (3.6 x 3.6 x6.6)

2.2 x 2.2 x 2 (7.3 x 7.3 x6.6)

Inside, control NA NA NA NAOutside, overall 2.4 x 2.4 x 2.3 (7.8 x 7.8 x

7.4)0.9 x 0.9 x 1.9 (3 x 3 x 6.3) 1.3 x 1.3 x 2.3 (4.1 x 4.1 x

7.4)2.4 x 2.4 x 2.3 (7.8 x 7.8 x7.4)

WINDOWW x H, cm (in) 63 x 54 (25 x 21) 56 x 56 (22 x 22) 55 x 70 (22 x 28) 55 x 70 (22 x 28)

WEIGHT, kg (lb) 560 (1,235) 150 (331) 470 (1,036) 1,310 (2,889)LIST PRICE Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 60 days 60 days 60 days 60 daysFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. Optional custom sizes with

built-in Faraday cage.Optional custom sizes withbuilt-in Faraday cage.

LAST UPDATED December 2002 December 2002 December 2002 December 2002Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 108: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 108/116

Audiometers; Audiometric Booths

108 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL AMPLIFON ECKEL ECKEL ECKELMini Silent Cabin AB-150 AB-200 AB-2000

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Yes Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes No Yes YesJACK PANEL

Number, type Jack and RCA for allaudiometers

4, stereo phono; RS232 port 4, stereo phono 4, stereo phono

VENTILATION, m³/min(ft³/min)

Not specified Passive 30 changes/hr 30 changes/hr

LIGHT Incandescent None Optional OptionalNumber 1 NA Not specified Not specified

AUDIOMETER SHELF Optional None Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET Yes No None NoW x D x H, m (ft)

Inside, exam 0.6 x 0.9 x 1.6 (1.9 x 3 x5.2)

11 x 9.8 x 19.5 (36 x 32 x64)

11 x 8.5 x 17.7 (36 x 28 x58)

11.6 x 7.9 x 20.7 (38 x 26 x68) : 11.6 x 11 x 20.7 (38 x36 x 68)

Inside, control NA NA NA NAOutside, overall 0.7 x 1 x 1.7 (2.3 x 3.3 x

5.6)1 x 0.9 x 1.7 (40 x 36 x 66) 12.2 x 9.8 x 19.8 (40 x 32 x

65)12.8 x 9.1 x 22.9 (42 x 30 x75) : 12.8 x 12.2 x 22.9 (42x 40 x 75)

WINDOWW x H, cm (in) 55 x 70 (22 x 28) 91.4 x 167.6 (36 x 66) 58.4 x 58.4 (23 x 23) 58.4 x 58.4 (23 x 23)

WEIGHT, kg (lb) 220 (485) 99.8 (220), shipping 272 (600), shipping 318 (700), shippingLIST PRICE Not specified $2,300 $2,790 $3,025WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 60 days 4-6 weeks 2-4 weeks 2-4 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS Can be installed on test

vehicles.None specified. None specified. None specified.

LAST UPDATED December 2002 January 2004 January 2004 January 2004

Supplier FootnotesModel Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 109: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 109/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 109

Chart B: Audiometric Booths

MODEL ECKEL ECKEL ECKEL ECKEL AB-4230 : AB-4240 ABC-100 C-11 C-12

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Double wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes No Yes YesJACK PANEL

Number, type 4, stereo phono Pass-through 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

VENTILATION, m³/min(ft³/min)

30 changes/hr Passive 30 changes/hr 30 changes/hr

LIGHT Optional None Single SingleNumber Not specified NA 2 2

AUDIOMETER SHELF Optional No Optional OptionalCARPETING Yes No Yes YesELECTRICAL OUTLET No No 2 duplex 2 duplexW x D x H, m (ft)

Inside, exam 1 x 0.7 x 1.7 (3.2 x 2.2 x5.7) : 1 x 0.9 x 1.7 (3.2 x 3 x5.7)

10.1 x 6.4 x 4.6 (33 x 21 x15)

0.9 x 1.2 x 2 (2.9 x 3.8 x6.6)

1.2 x 1.4 x 2 (3.8 x 4.7 x6.6)

Inside, control NA NA NA NAOutside, overall 1.1 x 0.8 x 1.9 (3.5 x 2.5 x

6.3) : 1.1 x 1 x 1.9 (3.5 x 3.3x 6.3)

11 x 7.3 x 5.5 (36 x 24 x 18) 1.1 x 1.3 x 2.3 (3.5 x 4.4 x7.5)

1.3 x 1.6 x 2.3 (4.3 x 5.3 x7.5) : 1.3 x 2.2 x 2.4 (4.3 x7.1 x 7.8)

WINDOWW x H, cm (in) 58.4 x 58.4 (23 x 23) 38.1 x 76.2 (15 x 30) 61 x 76.2 (24 x 30), double

glazed61 x 76.2 (24 x 30), doubleglazed

WEIGHT, kg (lb) 295 (650) : 344.7 (760),shipping

22.7 (50), shipping 711 (1,568) 1,026 (2,263)

LIST PRICE $3,195 : $3,695 $2,150 $5,900 $6,500WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 2-4 weeks 2-4 weeks 4-6 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to December

OTHER SPECIFICATIONS None specified. None specified. None specified. None specified.LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 110: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 110/116

Audiometers; Audiometric Booths

110 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL ECKEL ECKEL ECKEL ECKELC-13 C-14 C-15 C-16 : C-17

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Single wall Single wallCONFIGURATION Single room Single room Single room Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

VENTILATION, m³/min(ft³/min)

30 changes/hr 30 changes/hr 30 changes/hr 30 changes/hr

LIGHT Single Single Single SingleNumber 2 2 2 2

AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 2 duplex 2 duplex 2 duplex 2 duplexW x D x H, m (ft)

Inside, exam 1.5 x 1.7 x 2 (4.8 x 5.6 x6.6)

1.7 x 1.9 x 2 (5.6 x 6.2 x6.6)

1.9 x 2.3 x 2 (6.5 x 7.5 x6.6)

2.5 x 2.8 x 2 (8.3 x 9.2 x6.6)

Inside, control NA NA NA NAOutside, overall 1.6 x 1.9 x 2 (5.3 x 6.2 x

6.6)1.9 x 2.2 x 2.3 (6.2 x 7.1 x7.5)

2.2 x 2.4 x 2.3 (7.1 x 8 x7.5)

2.4 x 2.5 x 2.3 (8 x 8.1 x7.5) : 2.5 x 2.8 x 2.3 (8.1 x9.1 x 7.5)

WINDOWW x H, cm (in) 61 x 76.2 (24 x 30), double

glazed61 x 76.2 (24 x 30), doubleglazed

61 x 76.2 (24 x 30), doubleglazed

61 x 76.2 (24 x 30), doubleglazed

WEIGHT, kg (lb) 1,163 (2,565) 1,412 (3,114) 1,641 (3,618) 2,036.7 (4,491) : 2,440.8(5,382)

LIST PRICE $8,700 $9,600 $12,200 $13,825 : $14,625WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. None specified. None specified.

LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 111: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 111/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 111

Chart B: Audiometric Booths

MODEL ECKEL ECKEL ECKEL ECKELC-24 : C-25 C-26 : C-27 CL-66 : CL-88 S-121 : S-122

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Double wall Double wall Single wall Single-wall control room,

double-wall examinationroom

CONFIGURATION Single room Single room Single room with cornerbooth

Suite

VIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

VENTILATION, m³/min(ft³/min)

30 changes/hr 30 changes/hr 30 changes/hr 30 changes/hr

LIGHT Single Single Bulb/track SingleNumber 2 2 2 2

AUDIOMETER SHELF Optional Optional Optional Optional

CARPETING Yes Yes Yes YesELECTRICAL OUTLET 2 duplex 2 duplex 2 duplex 2 duplexW x D x H, m (ft)

Inside, exam 1.7 x 1.9 x 2 (5.6 x 6.5 x6.6) : 1.9 x 2.2 x 2 (6.5 x 7.4x 6.6)

2.3 x 2.5 x 2 (7.4 x 8.3 x6.6) : 2.5 x 2.8 x 2 (8.3 x 9.2x 6.6)

1.6 x 1.6 x 2 (5.1 x 5.1 x6.6) : 2.3 x 2.3 x 2 (7.4 x 7.4x 6.6)

2.3 x 1.7 x 2.4 (7.5 x 5.6 x7.9) : 2.5 x 2 x 2.4 (8.2 x 6.6x 7.9)

Inside, control NA NA Not specified 1.8 x 1.9 x 2 (6 x 6.2 x 6.6) :1.8 x 2.3 x 2 (6 x 7.5 x 6.6)

Outside, overall 2.3 x 2.6 x 2.5 (7.5 x 8.5 x8.1) : 2.6 x 2.9 x 2.5 (8.5 x9.4 x 8.1)

2.9 x 3.1 x 2.5 (9.4 x 10.3 x8.1) : 3.1 x 3.4 x 2.5 (10.3 x11.2 x 8.1)

1.9 x 1.9 x 2.3 (6.5 x 6.5 x7.7) : 2.4 x 2.4 x 2.3 (8 x 8 x7.7)

1.9 x 3.9 x 2.3 (6.2 x 12.9 x7.5) : 2.7 x 5 x 2.4 (8.9 x16.4 x 7.9)

WINDOWW x H, cm (in) 61 x 76.2 (24 x 30), double

glazed61 x 76.2 (24 x 30), doubleglazed

61 x 76.2 (24 x 30), doubleglazed

61 x 76.2 (24 x 30), doubleglazed

WEIGHT, kg (lb) 2,923 (6,446) : 3,370.4(7,432)

3,730.5 (8,226) : 4,326.4(9,540)

975 (2,150 ) : 1,292.5(2,850)

4,021.2 (8,865) : 5,041.3(11,114)

LIST PRICE $17,100 : $21,000 $24,150 : $26,500 $7,425 : $9,000 $27,475 : $29,850WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. None specified. None specified.LAST UPDATED January 2004 January 2004 January 2004 January 2004Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 112: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 112/116

Audiometers; Audiometric Booths

112 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL ECKEL ECKEL ECKEL EYMASAS-123 : S-124 S-34 : S-45 S-56 : S-57 CI/CL 40

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified YesTYPE Single-wall control room,

double-wall examinationroom

Single wall Single wall Single wall

CONFIGURATION Suite Suite Suite Not specifiedVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type 9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

9.3, stereo phono; 1, CJ303; 1, CJ 304; 2.5 cm (1 in)pass-through

6, stereo phono

VENTILATION, m³/min(ft³/min)

30 changes/hr 30 changes/hr 30 changes/hr Assisted

LIGHT Single Single Single IncandescentNumber 2 2 2 1

AUDIOMETER SHELF Optional Optional Optional YesCARPETING Yes Yes Yes YesELECTRICAL OUTLET 2 duplex 2 duplex 2 duplex 1W x D x H, m (ft)

Inside, exam 2.8 x 2.3 x 2.4 (9.2 x 7.5 x7.9) : 3 x 2.5 x 2.4 (10.1 x8.2 x 7.9)

1.7 x 1.5 x 2 (5.6 x 4.8 x6.6) : 1.9 x 1.7 x 2 (6.5 x 5.6x 6.6)

2.3 x 1.9 x 2 (7.5 x 6.2 x6.6) : 2.5 x 2.3 x 2 (8.2 x 7.5x 6.6)

1.3 x 1.2 x 2.1 (4.2 x 3.9 x6.8), optional custom sizes

Inside, control 2.2 x 2.6 x 2 (7.1 x 8.6 x6.6) : 2.7 x 2.9 x 2 (8.9 x 9.5x 6.6)

1.7 x 1.9 x 2 (5.6 x 6.5 x6.6) : 1.9 x 2.3 x 2 (6.5 x 7.4x 6.6)

2.3 x 1.9 x 2 (7.5 x 6.2 x6.6) : 2.5 x 2.3 x 2 (8.2 x 7.5x 6.6)

Not specified

Outside, overall 2.4 x 5 x 2.3 (8 x 16.4 x 7.5): 2.5 x 5.5 x 2.3 (8.2 x 18 x7.5)

1.8 x 3.9 x 2.2 (6.2 x 12.9 x7.5) : 2.1 x 4.5 x 2.2 (7.1 x14.6 x 7.5)

2.4 x 5 x 2.3 (8 x 16.4 x 7.5): 2.5 x 5.5 x 2.3 (8.2 x 18.2x 7.5)

1.3 x 1.2 x 2.2 (4.3 x 3.9 x7.2)

WINDOWW x H, cm (in) 61 x 76.2 (24 x 30), double

glazed61 x 76.2 (24 x 30), doubleglazed

61 x 76.2 (24 x 30), doubleglazed

41 x 75 (16 x 30)

WEIGHT, kg (lb) 5,827.9 (12,848) : 6,583.6(14,514)

2,575.4 (5,679) : 3,053(6,732)

3,678.2 (8,109) : 4,478.4(9,873)

460 (1,014.3)

LIST PRICE $37,325 : $42,800 $17,550 : $20,000 $24,150 : $27,800 $2,949WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks 2 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS None specified. None specified. None specified. Optional left- or right-hand

swing door; attenuationvalues available for 125-8,000 Hz (octaves). Meetsrequirements of ANSI.

LAST UPDATED January 2004 January 2004 January 2004 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 113: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 113/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 113

Chart B: Audiometric Booths

MODEL EYMASA EYMASA EYMASA IACCI-35 CI-40 CI-45 : CI-50 120A-1

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Yes Yes Yes Not specifiedTYPE Single wall Single wall Single wall : Double wall Double wallCONFIGURATION Not specified Not specified Not specified Single roomVIBRATION ISOLATORS Yes Yes Yes YesJACK PANEL

Number, type 6, stereo phono 6, stereo phono 6, stereo phono 10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through

VENTILATION, m³/min(ft³/min)

Free Free Assisted 2.8 (100)

LIGHT Incandescent Incandescent Incandescent FluorescentNumber 1 1 1 1

AUDIOMETER SHELF Yes Yes Yes OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 1 1 1 1W x D x H, m (ft)

Inside, exam Not specified Not specified 1.3 x 1.2 x 2.1 (4.2 x 3.9 x6.8)

1.2 x 1.2 x 2 (3.8 x 3.8 x6.5)

Inside, control Not specified Not specified Not specified NAOutside, overall 0.9 x 0.7 x 2 (3 x 2.3 x 6.6) 1.1 x 1.1 x 2.2 (3.6 x 3.6 x

7.2), optional custom sizes1.1 x 1.1 x 2.3 (3.6 x 3.6 x7.5), optional custom sizes

1.8 x 1.8 x 2.5 (5.8 x 5.8 x8.2)

WINDOWW x H, cm (in) 41 x 75 (16 x 30) 41 x 75 (16 x 30) 41 x 75 (16 x 30) 61 x 76.2 (24 x 30)

WEIGHT, kg (lb) 250 (551.2) 360 (793.8) 460 (1,014.3) 1,814 (4,000)LIST PRICE $2,349 $2,599 $3,999 : $4,799 $15,400WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 2 weeks 2 weeks 2 weeks 4-6 weeksFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS Optional left- or right-hand

swing door; attenuationvalues available for 125-8,000 Hz (octaves). Meets

requirements of ANSI.

Optional left- or right-handswing door; attenuationvalues available for 125-8,000 Hz (octaves). Meets

requirements of ANSI.

Optional left- or right-handswing door; attenuationvalues available for 125-8,000 Hz (octaves). Meets

requirements of ANSI.

Optional varied room sizesavailable; optional radio-frequency shielding. Meetsrequirements of ANSI.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 114: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 114/116

Audiometers; Audiometric Booths

114 ©2007 ECRI Institute. All Rights Reserved.

Chart B: Audiometric Booths

MODEL IAC IAC IAC IAC120ACT-1 140ACT-1 160ACT-1 40A-1

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Double wall Single-wall control room,

double-wall examinationroom

Single-wall control room,double-wall examinationroom

Single wall

CONFIGURATION Double room Double room Double room Single roomVIBRATION ISOLATORS Yes Exam room Yes YesJACK PANEL

Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through

10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through

10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through

6, stereo phono; 1, Cinch303; 1, Cinch 304

VENTILATION, m³/min(ft³/min)

5.7 (200), each room 5.7 (200), each room 5.7 (200), each room 2.8 (100)

LIGHT Fluorescent Fluorescent Fluorescent FluorescentNumber 2, each room 2, each room 2, each room 1

AUDIOMETER SHELF Optional Optional Optional OptionalCARPETING Yes Yes Yes YesELECTRICAL OUTLET 6, control room 6, control room 6, control room 1W x D x H, m (ft)

Inside, exam 1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)

1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)

1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)

0.8 x 1.1 x 2 (2.7 x 3.8 x6.5)

Inside, control 1.9 x 1.2 x 2 (6.2 x 3.8 x6.5)

2.3 x 2 x 2.4 (7.5 x 6.7 x7.8)

2.3 x 1.9 x 2 (7.5 x 6.3 x6.5)

NA

Outside, overall 2.5 x 3.9 x 2.5 (8.2 x 12.8 x8.2)

2.5 x 4.6 x 2.5 (8.2 x 15.2 x8.2)

2.5 x 4.7 x 2.5 (8.2 x 15.5 x8.2)1 

1 x 1.4 x 2.3 (3.3 x 4.5 x7.6)

WINDOWW x H, cm (in) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30)

WEIGHT, kg (lb) 5,398 (11,900) 4,627 (10,200) 5,080 (11,200) 816 (1,800)LIST PRICE $33,900 $29,100 $35,500 $5,680WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks 4-6 weeks 4-6 weeks From stock, 9 modelsFISCAL YEAR January to December January to December January to December January to DecemberOTHER SPECIFICATIONS Other models available in

the 120ACT series, whichrange in room size; optionalradio-frequency shielding.Meets requirements of

 ANSI.

Other models available in

the 140ACT series, whichrange in room size; optionalradio-frequency shielding.Meets requirements of

 ANSI.

Other models available in

the 160ACT series, whichrange in room size; optionalradio-frequency shielding.Meets requirements of

 ANSI.

Other models available in

the 40A series, which rangein room size; optional radio-frequency shielding. Meetsrequirements of ANSI.

LAST UPDATED January 2007 January 2007 January 2007 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes 1The 2 rooms have differentheights; specifications arefor the larger room.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 115: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 115/116

  Audiometers; Audiometric Booths

©2007 ECRI Institute. All Rights Reserved 115

Chart B: Audiometric Booths

MODEL IAC IAC NAGASHIMA SIBELMED40ACT-1 Mini Series 250 SN-4A S40

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideCE MARK (MDD) Not specified Not specified Not specified Not specifiedTYPE Single wall Single wall Triple wall Single wallCONFIGURATION Double room Single room Single room Single roomVIBRATION ISOLATORS Yes Optional Yes YesJACK PANEL

Number, type 10, stereo phono; 1, Cinch303; 1, Cinch 304; 2, 2.5 cm(1 in) pass-through

6, stereo phono 4, 3c30 3-pin plug; 1, 3-pinconnector; 1, 10-pin spare;1, relay cord

6, stereo phono

VENTILATION, m³/min(ft³/min)

5.7 (200), each room Not specified Not specified Not specified

LIGHT Fluorescent Fluorescent Incandescent FluorescentNumber 2, each room 1 1 (tungsten), includes

replacement1

AUDIOMETER SHELF Optional Yes Optional YesCARPETING Yes Yes Yes YesELECTRICAL OUTLET 6, control room None 1 1W x D x H, m (ft)

Inside, exam 1.9 x 1.9 x 2 (6.2 x 6.2 x6.5)

0.9 x 0.6 x 1.7 (2.9 x 2 x5.5)

0.7 x 0.7 x 1.8 (2.3 x 2.3 x5.9)

Not specified

Inside, control 1.9 x 1.2 x 2 (6.2 x 3.8 x6.5)

NA NA NA

Outside, overall 2.1 x 3.6 x 2.3 (6.8 x 11.7 x7.6)

1 x 0.7 x 1.9 (3.3 x 2.3 x6.2)

0.8 x 0.8 x 1.9 (2.6 x 2.6 x6.2)

0.9 x 0.9 x 2.15 (3 x 3 x 7.1)

WINDOWW x H, cm (in) 61 x 76.2 (24 x 30) 61 x 76.2 (24 x 30) 53 x 53 (20.9 x 20.9) 45 x 55 (17.7 x 21.7)

WEIGHT, kg (lb) 2,767 (6,100) 295 (650) 290 (639) 265 (584.3)LIST PRICE $21,300 $3,250 Not specified $2,275WARRANTY 1 year 1 year 1 year 1 yearDELIVERY TIME, ARO 4-6 weeks, standard models From stock Not specified Not specifiedFISCAL YEAR January to December January to December Not specified January to DecemberOTHER SPECIFICATIONS Other models available in

the 40ACT series, which

range in room size; optionalradio-frequency shielding.Meets requirements of

 ANSI.

Casters; preassembled;optional intercom system.

Meets requirements of ANSI.

Radio-frequency shielded;attenuation values available

for 125-8,000 Hz (octaves).

Right- or left-hand swingdoor; attenuation values

available for 125-8,000 Hz(octaves). Meetsrequirements of ANSI.

LAST UPDATED January 2007 January 2007 April 2001 January 2007Supplier Footnotes

Model Footnotes

Data Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 116: Audiometers audiometric booths.PDF

8/9/2019 Audiometers audiometric booths.PDF

http://slidepdf.com/reader/full/audiometers-audiometric-boothspdf 116/116

Audiometers; Audiometric Booths

Chart B: Audiometric Booths

MODEL TREMETRICS AR 901

WHERE MARKETED WorldwideCE MARK (MDD) Not specifiedTYPE Not specifiedCONFIGURATION Single roomVIBRATION ISOLATORS OptionalJACK PANEL

Number, type 6, stereo phonoVENTILATION, m³/min(ft³/min)

Not specified

LIGHT Not specifiedNumber Not specified

AUDIOMETER SHELF 16 x 27 in, foldingCARPETING YesELECTRICAL OUTLET YesW x D x H, m (ft)

Inside, exam 0.9 x 0.7 x 1.7 (3 x 2.3 x5.6)

Inside, control Not specifiedOutside, overall 1 x 0.8 x 2 (3.3 x 2.6 x 6.6)

WINDOWW x H, cm (in) 35.6 x 152.4 (17 x 60)

WEIGHT, kg (lb) 347.7 (765)LIST PRICE $3,275WARRANTY 1 yearDELIVERY TIME, ARO Not specifiedFISCAL YEAR January to DecemberOTHER SPECIFICATIONS Portable; casters included;

right- or left-hand doorswing.

LAST UPDATED July 2005Supplier Footnotes

Model Footnotes