UMDNS Information
This Product Comparison covers the followingdevice term and product code as listed inECRI Institute’s Universal Medical DeviceNomenclature System™ (UMDNS™):
Arthroscopes [10-198]
Arthroscopes
Scope of this Product Comparison
This Product Comparison covers diagnostic and operating arthroscopes; the chart also lists compatible light
sources. Other endoscopes are covered in the following Product Comparisons:
Bronchoscopes, Flexible; Video
Colonoscopes; Sigmoidoscopes
Duodenoscopes; Gastroscopes; Choledochoscopes; Video Systems, Endoscopic, IngestibleCameras
Laparoscopes
Video Endoscopy Systems
These devices are also called: articular cavity endoscopes.
Purpose
Arthroscopes are telescopic devices that enable visualization and
therapeutic treatment of the interior of a joint (e.g., knee, shoulder,
wrist, elbow, hip, ankle). They are used by physicians to facilitate the
diagnosis and/or treatment of a wide variety of joint diseases and
injuries. As a minimally invasive surgical procedure, arthroscopy is
less traumatic than arthrotomy (open-joint surgery) and usually
results in a shorter recovery time.
Principles of operation
An arthroscope consists of a magnifying-lens system sealed within a rigid tube and is available in two basic
optical designs: the rod-lens system and the single-image fiber system. The rod-lens system employs long, rodlike
lenses separated by small spaces, permitting small scope diameters and providing a brighter image than that with
the single-image fiber system. Single-image fiber systems are even narrower in diameter (1 to 2.7 mm versus 4 to
6.5 mm) and consist of one image-transmitting fiber bundle rather than separate lenses. The smaller diameter of
single-image fiber systems enables the arthroscopist to view hard-to-reach parts of the knee and to examine small joints such as the wrist or ankle.
In both designs, the lens system is surrounded by optical fibers
that transmit light to the interior of the joint from a distant source. In
most cases, the intensity of the light source can be adjusted for direct
viewing or viewing with video systems. (For more detailed
information, see the Product Comparison titled Light Sources,
Fiberoptic.)
Arthroscopes generally provide a field of view similar to that of the
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 2/27
Arthroscopes
2 ©2007 ECRI Institute. All Rights Reserved.
Figure 1. Triangulation method of arthroscopy
human eye— between 70° and 100°. Most manufacturers offer arthroscopes with a number of viewing angles
(directions of view). Many arthroscopists prefer a 30° direction of view because it increases the area that can be
visualized while maintaining forward orientation as the scope is rotated about its longitudinal axis. Most
arthroscopes use video cameras that attach to the eyepiece to view and document the procedure. The camera
houses charge-coupled devices that convert the image produced at the eyepiece to a digital electronic signal.
Once the arthroscope is inserted into the joint (often through a working sleeve or
sheath), irrigation valves, which may be
components of the sleeve or the scope, allow
the joint to be distended and/or lavaged with
saline solution, washing out blood and debris
and allowing for better viewing.
There are two basic techniques for
performing arthroscopic surgery. One method
uses an operating arthroscope with an
instrument channel, which allows tools to be
passed down the optical line of the
arthroscope. This technique is useful when it is
difficult to enter a joint from a second or third portal. The other method, called the triangulation method, involves
inserting a diagnostic arthroscope through one portal while the operating instrument is inserted through a second
portal. This method allows for a greater variety of approaches than is possible with an operating arthroscope.
Reported problems
As with any surgical procedure, there is a risk of infection or introduction of foreign particles. Skin
microorganisms can enter the joint, especially during trocar insertion, and subsequent joint infections could
present serious complications.
Major contraindications to arthroscopy include sepsis (generalized infection), blood-clotting disorders, and
restricted joint motion. Hemarthrosis (blood entering the joint or its synovial cavity) is a complication that can betreated by aspirating the blood from the joint. Other complications include breakage of surgical instruments
inside the joint, lethal air embolism, subcutaneous emphysema, effusions, adhesions, reflex sympathetic
dystrophy, and cardiovascular and neurologic problems.
Use of optical catheter systems for arthroscopic evaluation and treatment of the knee in an office setting may
compromise visual acuity, resulting in missed or incorrect diagnoses (Meister et al. 1996). The optical catheter has
been a popular choice due to its compact size and relatively low cost.
Purchase considerations
ECRI Institute recommendations
Included in the accompanying comparison chart are ECRI Institute’s recommendations for minimum
performance requirements for arthroscopes; recommended specifications have been categorized into one group—
general arthroscopes. The general category covers both diagnostic and operating arthroscopes.
Compatibility with existing and future equipment should be taken into account. Standard arthroscopes should
permit connection to all common models of camera heads, and video arthroscopes should preferably be
compatible not only with the facility’s current video processors but with more recent processor models offered by
the same manufacturer. Additionally, the manufacturer should make available a sufficient variety of light guides
to permit connection to all common light source designs.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 3/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 3
Hospitals should carefully consider the reprocessing methods approved by the scope manufacturer. The wider
the variety of options, the better the hospital may be able to effectively manage its inventory. While
glutaraldehyde and EtO are commonly approved agents, hospitals may wish to strongly consider scopes that are
also compatible with autoclaving and other fast sterilization methods.
Assessing factors that affect image usefulness (e.g., resolution, color reproduction, field of view) is highly
subjective. Therefore, ECRI Institute recommends trialing arthroscopes to allow surgeons to directly compareperceived performance. When performing such trials, hospitals should standardize all other equipment,
including monitors, video processors, light sources and guides, and signal format, to ensure that the only variable
under consideration is the scope’s performance.
Other considerations
Accessory equipment includes electrosurgical and shaver accessories as well as various mechanical cutting and
grasping instruments. Purchasers should consider buying arthroscopes and accessories from the same
manufacturer to ensure compatibility.
Stage of development
With the incorporation of fiberoptic light cables and improvements in lens systems, the arthroscope has
become more streamlined and refined, enabling less traumatic and more advanced procedures; these
improvements have also enabled the arthroscopist to examine smaller joints such as the temporomandibular joint.
New instruments allow for fewer complications and higher success rates with current techniques. Increased
miniaturization of the arthroscope and refinement of instruments and techniques are continuing.
Arthroscopy is being performed more widely in the outpatient arena as surgeons become more comfortable
with arthroscopic techniques and as instruments become smaller and less expensive.
Bibliography
Altman RD, Gray R. Diagnostic and therapeutic uses of the arthroscope in rheumatoid arthritis and osteoarthritis. Am J Med 1983 Oct 31;75(4B):50-5.
American Academy of Orthopaedic Surgeons. Symposium on arthroscopy and arthrography of the knee. St.
Louis: CV Mosby; 1978.
Dandy DJ. Arthroscopic surgery [annotation]. J Bone Joint Surg Br 1984 Nov;66(5):627-8.
Harrison C, Upson K. Arthroscopy instrumentation and time-savers. AORN J 1984 Jun;39(7):1230-3.
Ike RW. Diagnostic arthroscopy. Baillieres Clin Rheumatol 1996 Aug;10(3):495-517.
Jackson RW. Current concepts review: arthroscopic surgery. J Bone Joint Surg Am 1983 Mar;65(3):416-20.
Johnson LL, Shneider DA, Austin MD, et al. Two per cent glutaraldehyde: a disinfectant in arthroscopy and
arthroscopic surgery. J Bone Joint Surg Am 1982 Feb;64(2):237-9.
Meister K, Harris NL, Indelicato PA, et al. Comparison of an optical catheter office arthroscope with a standard
rigid rod-lens arthroscope in the evaluation of the knee. Am J Sports Med 1996 Nov-Dec;24(6):819-23.
Sanders B, Buoncristiani R. Diagnostic and surgical arthroscopy of the temporomandibular joint: clinical
experience with 137 procedures over a 2-year period. J Craniomandib Disord 1987 Fall;1(3):202-13.
Shibuya T, Kino K, Yoshitake H, et al. Use of a new instrument for lateral release in arthroscopic surgery of the
tempomandibular joint: a preliminary study. BR J Oral Maxillofac Surg 2004 Apr;42(2):166-9.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 4/27
Arthroscopes
4 ©2007 ECRI Institute. All Rights Reserved.
Takahashi T, Yamamoto H. Development and clinical application of a flexible arthroscopy system. Arthroscopy
1997 Feb;13(1):42-50.
Villar RN. Hip arthroscopy. Br J Hosp Med 1992 May 20-Jun 2;47(10):763-6.
Supplier information
AED
Advanced Endoscopy Devices Inc [360637]
22134 Sherman Way
Canoga Park, CA 91303
Phone: (818) 227-2720 Fax: (818) 227-2724
Internet: http://www.aed.md
E-mail: [email protected]
AESCULAP
Aesculap AG & Co KG [139214]
Am Aesculap-Platz Postfach 40
Tuttlingen D-78532
Germany
Phone: 49 (7461) 950 Fax: 49 (7461) 952600
Internet: http://www.aesculap.de
E-mail: [email protected]
Aesculap Inc [101658]
3773 Corporate Pkwy
Center Valley, PA 18034
Phone: (610) 797-9300, (800) 258-1946 Fax: (610) 791-6880
Internet: http://www.aesculapusa.com
E-mail: [email protected]
Aesculap Japan Co Ltd [186629]
Hongo TS Building 2-38-16 Hongo Bunkyo-ku
Tokyo 113-0033
Japan
Phone: 81 (3) 38142525 Fax: 81 (3) 56847080
Internet: http://www.bbraun.co.jp
E-mail: [email protected]
B Braun Medical Ltd [305573]
Thorncliffe Park
Sheffield S35 2PW
EnglandPhone: 44 (114) 2259000 Fax: 44 (114) 2259111
Internet: http://www.bbraun.co.uk
E-mail: [email protected]
CANADA ENDOSCOPE
Canada Endoscope Corp [377686]
160 Konrad Crescent Unit 4
Markham ON L3R 9T9
Canada
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 5/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 5
Phone: (905) 479-2277, (800) 563-2907 Fax: (905) 479-6495
Internet: http://www.canadaendoscope.ca
E-mail: [email protected]
EFER
EFER [177016]
Zone Industrielle Athelia 1La Ciotat Cedex F-13705
France
Phone: 33 (4) 42715101 Fax: 33 (4) 42833325
Internet: http://www.efer.com
E-mail: [email protected]
ENDO OF SHENYANG
Endoscope Ltd Co of Shenyang University Endoscope [291543]
No 176 Lianhe Road Dadong District
Shenyang 110044
People’s Republic of China
Phone: 86 (24) 88908788 Fax: 86 (24) 88325910
Internet: http://www.china-endoscope.com E-mail: [email protected]
HANS HERMANN
Hans Hermann GmbH [332067]
Wuerttemberger Strasse 26
Fridingen D-78567
Germany
Phone: 49 (7463) 99670 Fax: 49 (7463) 996750
Internet: http://www.hermann-medizintechnik.de
E-mail: [email protected]
HI-TEC
Hi-TEC Medical GmbH [248698]
Industrie-Strasse 7
Tuttlingen D-78532
Germany
Phone: 49 (7461) 94690 Fax: 49 (7461) 946916
Internet: http://www.hitecmedical.de
E-mail: [email protected]
KARLHEINZ HINZE
Karl-Heinz Hinze Optoengineering GmbH & Co [282461]
Kieler Strasse 464-470
Hamburg D-22525
GermanyPhone: 49 (40) 542590 Fax: 49 (40) 543573
Internet: http://www.hinze-opto.de
E-mail: [email protected]
LINVATEC
Linvatec Corp [226512]
11311 Concept Blvd
Largo, FL 33773
Phone: (727) 392-6464, (800) 237-0169 Fax: (727) 399-5256
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 6/27
Arthroscopes
6 ©2007 ECRI Institute. All Rights Reserved.
Internet: http://www.conmed.com/linvatec.pht
E-mail: [email protected]
MAHE\RG Medical
MAHE Barthelmes Medizintechnik GmbH [286788]
Friedrich-Woehler-Strasse 10
Emmingen D-78576Germany
Phone: 49 (7465) 92750 Fax: 49 (7465) 927529
Internet: http://www.mahe-med.de
E-mail: [email protected]
MAHE International Inc [298390]
468 Craighead St
Nashville, TN 37204
Phone: (615) 269-7256, (800) 294-7946 Fax: (615) 269-4605
Internet: http://www.maheinternational.com
E-mail: [email protected]
MAHE Medical Technology Co Ltd [340682]1/412 Gardenhome Village Phaholyothin Rd Kukod Lamlukke
Pathumthani 12130
Thailand
Phone: 66 (2) 9936769 Fax: 66 (2) 9936778
OLYMPUS
Olympus America Inc [276750]
2 Corporate Center Dr PO Box 9058
Melville, NY 11747-9058
Phone: (631) 844-5000, (800) 645-8160 Fax: (631) 844-5442
Internet: http://www.olympusamerica.com
E-mail: [email protected]
Olympus (Europa) GmbH [155981]
Wendenstrasse 14-18 Postfach 104908
Hamburg D-20097
Germany
Phone: 49 (40) 237735401 Fax: 49 (40) 23773656
Internet: http://www.olympus-europa.com
E-mail: [email protected]
Olympus Optical Co (UK) Ltd [320553]
2-8 Honduras Street
London EC1Y O7X
EnglandPhone: 44 (20) 72532772 Fax: 44 (20) 72516330
Internet: http://www.olympus.co.uk
E-mail: [email protected]
PRECISION OPTICS
Precision Optics Corp [169370]
22 E Broadway
Gardner, MA 01440-3338
Phone: (978) 630-1800, (800) 447-2812 Fax: (978) 630-1487
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 7/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 7
Internet: http://www.poci.com
E-mail: [email protected]
REMA
REMA Medizintechnik GmbH [160182]
In Breiten 10
Duerbheim D-78589Germany
Phone: 49 (7424) 4064 Fax: 49 (7424) 501590
Internet: http://www.rema-surgery.com
E-mail: [email protected]
SMITH NEPHEW
Smith & Nephew Endoscopy (UK) [344552]
Endoscopy House Cardinal Park Godmanchester
Huntingdon PE29 2SN
England
Phone: 44 (1480) 423200 Fax: 44 (1480) 423201
Internet: http://www.smith-nephew.com/uk
Smith & Nephew Inc Endoscopy [270765]
150 Minuteman Rd
Andover, MA 01810-1031
Phone: (978) 749-1000, (800) 343-5717 Fax: (978) 749-1599
Internet: http://www.endo.smith-nephew.com
E-mail: [email protected]
Smith & Nephew Ltd (Hong Kong) [289996]
13/Fl Grandtech Centre Unit 1318 8 On Ping Street Siu Lek Yuen Shatin
New Territories
People’s Republic of China
Phone: 852 26487700 Fax: 852 26487282Internet: http://www.smith-nephew.com
E-mail: [email protected]
SOPRO-COMEG
SOPRO-COMEG [399024]
Zone d'Activites Athelia IV avenue des Genevriers
La Ciotat Cedex F-13705
France
Phone: 33 (4) 42980101 Fax: 33 (4) 42717690
Internet: http://www.sopro.acteongroup.com
SOPRO-COMEG [452355]
124 Gaither Dr Suite 140Mount Laurel, NJ 08054
Phone: , (866) 717-6776 Fax: (856) 222-4726
Internet: http://www.sopro-comeg.com
E-mail: [email protected]
KARL STORZ
Karl Storz Endoscopy-America Inc [102159]
600 Corporate Pointe
Culver City, CA 90230-7600
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 8/27
Arthroscopes
8 ©2007 ECRI Institute. All Rights Reserved.
Phone: (310) 338-8100, (800) 421-0837 Fax: (310) 410-5527, (800) 321-1304
Internet: http://www.karlstorz.com
E-mail: [email protected]
Karl Storz GmbH & Co KG [152361]
Postfach 230
Tuttlingen D-78503Germany
Phone: 49 (7461) 7080 Fax: 49 (7461) 708105
Internet: http://www.karlstorz.com
E-mail: [email protected]
STRYKER ENDOSCOPY
Stryker Endoscopy [107536]
5900 Optical Ct
San Jose, CA 95138
Phone: (408) 754-2000, (800) 624-4422 Fax: (408) 754-2503, (800) 729-2917
Internet: http://www.strykerendo.com
E-mail: [email protected]
Stryker (S) Pte Ltd [307324]
1 Maritime Square #10-39 Harbourfront Centre
Singapore 099253
Republic of Singapore
Phone: 65 63776515 Fax: 65 62722464
Internet: http://www.stryker.com
E-mail: [email protected]
RICHARD WOLF
Richard Wolf Medical Instruments Corp [102126]
353 Corporate Woods Pkwy
Vernon Hills, IL 60061-3110Phone: (847) 913-1113, (800) 323-9653 Fax: (847) 913-1488
Internet: http://www.richardwolfusa.com
E-mail: [email protected]
Richard Wolf GmbH [161114]
Postfach 1164
Knittlingen D-75434
Germany
Phone: 49 (7043) 350 Fax: 49 (7043) 35300
Internet: http://www.richard-wolf.com
E-mail: [email protected]
Richard Wolf UK Ltd [366841]Waterside Way Wimbledon
London SW17 0HB
England
Phone: 44 (20) 89447447 Fax: 44 (20) 89441311
Internet: http://www.richardwolf.uk.com
E-mail: [email protected]
Richard Wolf France Sarl [366842]
rue Daniel Berger Zone d'Activites la Neuvillette
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 9/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 9
Reims F-51100
France
Phone: 33 (3) 26780289 Fax: 33 (3) 26876033
Internet: http://www.richard-wolf.com
E-mail: [email protected]
Note: The data in the charts derive from suppliers’ specifications and have not been verified through
independent testing by ECRI Institute or any other agency. Because test methods vary, different products’
specifications are not always comparable. Moreover, products and specifications are subject to frequent changes.
ECRI Institute is not responsible for the quality or validity of the information presented or for any adverse
consequences of acting on such information.
When reading the charts, keep in mind that, unless otherwise noted, the list price does not reflect supplier
discounts. And although we try to indicate which features and characteristics are standard and which are not,
some may be optional, at additional cost.
For those models whose prices were supplied to us in currencies other than U.S. dollars, we have also listed the
conversion to U.S. dollars to facilitate comparison among models. However, keep in mind that exchange rates change
often.
Need to know more?
For further information about the contents of this Product Comparison, contact the HPCS Hotline at +1 (610) 825-
6000, ext. 5265; +1 (610) 834-1275 (fax); or [email protected] (e-mail).
Last updated July 2007
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 10/27
Arthroscopes
10 ©2007 ECRI Institute. All Rights Reserved.
Policy Statement
The Healthcare Product Comparison System (HPCS) is published by ECRI Institute, a nonprofit organization. HPCS
provides comprehensive information to help healthcare professionals select and purchase diagnostic and
therapeutic capital equipment more effectively in support of improved patient care.
The information in Product Comparisons comes from a number of sources: medical and biomedical engineering
literature, correspondence and discussion with manufacturers and distributors, specifications from product
literature, and ECRI Institute’s Problem Reporting System. While these data are reviewed by qualified health
professionals, they have not been tested by ECRI Institute’s clinical and engineering personnel and are largely
unconfirmed. The Healthcare Product Comparison System and ECRI Institute are not responsible for the quality or
validity of information derived from outside sources or for any adverse consequences of acting on such
information.
The appearance or listing of any item, or the use of a photograph thereof, in the Healthcare Product Comparison
System does not constitute the endorsement or approval of the product’s quality, performance, or value, or of
claims made for it by the manufacturer. The information and photographs published in Product Comparisons
appear at no charge to manufacturers.
Many of the words or model descriptions appearing in the Healthcare Product Comparison System are proprietary
names (e.g., trademarks), even though no reference to this fact may be made. The appearance of any name
without designation as proprietary should not be regarded as a representation that is not the subject of
proprietary rights.
ECRI Institute respects and is impartial to all ethical medical device companies and practices. The Healthcare
Product Comparison System accepts no advertising and has no obligations to any commercial interests. ECRI
Institute and its employees accept no royalties, gifts, finder’s fees, or commissions from the medical device
industry, nor do they own stock in medical device companies. Employees engage in no private consulting work
for the medical device industry.
About ECRI Institute
ECRI Institute, a nonprofit organization, dedicates itself to bringing the discipline of applied scientific research in
healthcare to uncover the best approaches to improving patient care. As pioneers in this science for nearly 40
years, ECRI Institute marries experience and independence with the objectivity of evidence-based research.
More than 5,000 healthcare organizations worldwide rely on ECRI Institute’s expertise in patient safety
improvement, risk and quality management, healthcare processes, devices, procedures, and drug technology.
ECRI Institute is one of only a handful of organizations designated as both a Collaborating Center of the World
Health Organization and an Evidence-based Practice Center by the U.S. Agency for Healthcare Research and
Quality. For more information, visit http://www.ecri.org.
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 11/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 11
Product Comparison Chart
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 12/27
Arthroscopes
12 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1
AED AESCULAP AESCULAP
General Arthroscopes 5700A : 5700B : 5700C :
5700W
PE 182A : PE 202A PE 484A : PE 524A
WHERE MARKETED Worldwide Worldwide, except USA Worldwide, except USAFDA CLEARANCE Yes No NoCE MARK (MDD) Not specified Yes YesTYPE Diagnostic/operating Diagnostic Diagnostic Diagnostic
Diameter, mm 5 : 4 : 2.7 : 1.9 2.7 4Insertion length withsleeve, mm
90 : 140 : 155 : 180 Not specified 92
OPTICSLens configuration Rod lens Rod lens Rod lensDirection of view, ° 0, 30, or 70 0, 12, 25, 30, 45, 70, 90 0 : 30 0 : 70Field of view, ° 70-100 70, 90, 105, 120, 155 92 : 97 90 : 90
SLEEVE Yes Yes YesOutside diameter, mm 2.2 : 3 : 4 : 5.5 3.8 6Irrigation valves 1, 2 2 : 1 2 : 1Locking system Bayonet mounting Bayonet mounting Bayonet mounting, 1-click
couplingIRRIGATION FLOW,mL/min
250-350 Not specified 360
COMPATIBLE LIGHTSOURCES
All common models ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
DOCUMENTATIONOPTIONS
Yes Yes Yes
RECOMMENDEDREPROCESSING
All common methods;autoclavable preferred
Glutaraldehyde, EtO,STERIS, STERRAD,autoclave
Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,autoclave
PURCHASEINFORMATION
List price $800 Not specified Not specifiedWarranty 1 year 3 years 6 months 6 monthsDelivery time, ARO 2-4 weeks Not specified Not specifiedNumber sold Not specified Not specified Not specified
Year first sold Not specified 1997 Not specifiedFiscal year January to December Not specified Not specified
OTHER SPECIFICATIONS Adapts to all sheathsystems; C-mount, othervideo mounts, and universalarthroscopes available.
None specified. Sleeves with 1 or 2 rotatingstopcocks, autoclave tipsheath available;repair/exchange program.
LAST UPDATED July 2007 August 2005 August 2005Supplier Footnotes 1These recommendations
are the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based onthis data.
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 13/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 13
Product Comparison Chart
MODEL AESCULAP AESCULAP AESCULAP AESCULAPPE 485A PE 505A PE 514A PE 525A
WHERE MARKETED Worldwide, except USA Worldwide, except USA Worldwide, except USA Worldwide, except USAFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 4 4 4 4Insertion length withsleeve, mm
115 115 92 115
OPTICSLens configuration Rod lens Rod lens Rod lens Rod lensDirection of view, ° 0 30 30 70Field of view, ° 100 102 93 102
SLEEVE Yes Yes Yes YesOutside diameter, mm 6 6 6 6Irrigation valves 2 : 1 2 2 : 1 2 : 1Locking system Bayonet mounting, 1-click
couplingBayonet mounting, 1-clickcoupling
Bayonet mounting, 1-clickcoupling
Bayonet mounting, 1-clickcoupling
IRRIGATION FLOW,
mL/min
360 360 360 360
COMPATIBLE LIGHTSOURCES
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,autoclave
PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 6 months 6 months 6 months 6 monthsDelivery time, ARO Not specified Not specified Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified 1995 1987 Not specifiedFiscal year Not specified Not specified Not specified Not specified
OTHER SPECIFICATIONS Sleeves with 1 or 2 rotatingstopcocks, autoclave tipsheath available;repair/exchange program.
Sleeves with 2 rotatingstopcocks, autoclave tipsheath available;repair/exchange program.
Sleeves with 1 or 2 rotatingstopcocks, autoclave tipsheath available;repair/exchange program.
Sleeves with 1 or 2 rotatingstopcocks, autoclave tipsheath available;repair/exchange program.
LAST UPDATED August 2005 August 2005 August 2005 August 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 14/27
Arthroscopes
14 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL AESCULAP CANADA ENDOSCOPE EFER ENDO OF SHENYANGPE 544A : PE 545A CE 0004-A, 2504-A, 3004-
A, 4504-A, 7004-A, 0027-A,
3027-A
AW 2000P : AW 3000P : AY4030P
ShenDa J0230 : J0270 :J02000
WHERE MARKETED Worldwide, except USA Worldwide Worldwide ChinaFDA CLEARANCE No Not specified Yes NoCE MARK (MDD) Yes Not specified Yes NoTYPE Diagnostic Diagnostic Diagnostic Diagnostic/operating
Diameter, mm 4 2.7, 4 2 : 3 : 4 4Insertion length withsleeve, mm
92 : 115 120, 180 76 : 85 : 183 175
OPTICSLens configuration Rod lens Rod lens Rod lens : Self-focus : Self-
focusRod lens
Direction of view, ° 45 0, 25, 30, 45, 70 0 : 0 : 30 30 : 70 : 0Field of view, ° 93 : 95 80-106 120 65
SLEEVE Yes Yes Optional YesOutside diameter, mm 6 3.5, 5.5 2.6 : 3.8 : 5.5 5.1Irrigation valves 2 : 1 1.2 2 2
Locking system Bayonet mounting, 1-clickcoupling
Bayonet mounting Bayonet mounting Self-locking collar
IRRIGATION FLOW,mL/min
360 Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
ACMI, Aesculap, BritishStandard, HSW, Olympus,Storz, Winter, Wolf
ACMI, Applitech, Dyonics,Olympus, Pilling, Storz,Stryker, Wolf, Zimmer
Compatible with all light-source types
Storz, Wolf
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,STERRAD
EtO
PURCHASEINFORMATION
List price Not specified Not specified Not specified $580 : $580 : $500Warranty 6 months 1 year 1 year 1 yearDelivery time, ARO Not specified 1 week Not specified 40 days
Number sold Not specified Not specified Not specified Not specified Year first sold Not specified 1998 Not specified 1982Fiscal year Not specified November to October Not specified January to December
OTHER SPECIFICATIONS Sleeves with 1 or 2 rotatingstopcocks, autoclave tipsheath available;repair/exchange program.
None specified. None specified. None specified.
LAST UPDATED August 2005 August 2005 July 2007 March 2001Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 15/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 15
Product Comparison Chart
MODEL HANS HERMANN HI-TEC KARL STORZ KARL STORZH84-017-000 : H84-040-070 Arthroscope 28005 BA/CA 28031 BA
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic/operating Diagnostic/operating Diagnostic/operating
Diameter, mm 4 : 2.7 : 1.7 2.7, 4; varies 4 6.5Insertion length withsleeve, mm
175 Varies 300 350
OPTICSLens configuration Rod lens Rod lens Rod lens Rod lensDirection of view, ° 0 : 30 : 70 0, 30, 70 30, 70 30Field of view, ° 95, 105 Not specified 90-110 Full screen
SLEEVE Optional Yes Yes YesOutside diameter, mm 5.5 : 3.5 : 2.5 Varies 4 8Irrigation valves 1 or 2 Not specified 1 2Locking system Bayonet mounting Not specified Bayonet mounting Continuous-flow sheath,
bayonet mountingIRRIGATION FLOW,
mL/min
Not specified Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf
All ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,autoclave
Autoclave Glutaraldehyde, EtO,autoclave
Glutaraldehyde, EtO,autoclave
PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 1 year 1 year 3 years 3 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified Not specified Not specifiedFiscal year January to December Not specified January to December January to December
OTHER SPECIFICATIONS None specified. None specified. Hip arthroscopy scope. Hip arthroscopy scope.LAST UPDATED November 2002 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 16/27
Arthroscopes
16 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL KARL STORZ KARL STORZ KARL STORZ KARL STORZ28208 BA 28300 BA, CA 28305 AA, BA 28728 BWA
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic/operating Diagnostic/operating Diagnostic/operating Diagnostic/operating
Diameter, mm 2.7 2.4 1.9 4Insertion length withsleeve, mm
110 100 65 120
OPTICSLens configuration Rod lens Rod lens Rod lens Rod lensDirection of view, ° 30 30, 70 0, 30 30Field of view, ° 75 75 90-110 90-110
SLEEVE Optional Optional Optional OptionalOutside diameter, mm 3.5 3.2 2.5 6Irrigation valves 1 1 1 2Locking system Bayonet mounting Bayonet mounting, strobel
lockingBayonet mounting Autolock sheath
IRRIGATION FLOW,
mL/min
Not specified Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,autoclave
STERIS, STERRAD, EtO,autoclave
STERIS, STERRAD, EtO,autoclave
Glutaraldehyde, EtO,autoclave
PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 3 years 3 years 3 years 3 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified Not specified Not specifiedFiscal year January to December January to December January to December January to December
OTHER SPECIFICATIONS Extended eyepiece. None specified. None specified. Forward oblique; 12 cmscope.LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 17/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 17
Product Comparison Chart
MODEL KARL STORZ KARL STORZ KARLHEINZ HINZE KARLHEINZ HINZE28731 AWA, BWA, CWA,FWA, DA
28834 AWA, BWA, CWA,DA
MEDICASCOPE Micro- Arthroscope
MEDICASCOPE Standard Arthroscope
WHERE MARKETED Worldwide Worldwide Worldwide, except USA Worldwide, except USAFDA CLEARANCE Yes Yes No NoCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic/operating Diagnostic/operating Diagnostic Diagnostic
Diameter, mm 4 4 1.7 4 : 2.7Insertion length withsleeve, mm
180 180 107 175 : 187
OPTICSLens configuration Rod lens Rod lens Self-focus lens Rod lensDirection of view, ° 0, 30, 45, 70, 90 0, 30, 70, 90 0 0, 30, 70Field of view, ° 90-110 90-110 105 95, 110 : 90
SLEEVE Optional Optional Optional OptionalOutside diameter, mm 5.5, 6 5.5, 6 2.2 5 : 3.5Irrigation valves 1, 2 1, 2 Not specified Not specifiedLocking system Bayonet mounting, autolock
sheathBayonet mounting, autolocksheath
Not specified Not specified
IRRIGATION FLOW,mL/min
Not specified Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
ACMI, Dyonics, Olympus,Pilling, Storz, Stryker, Weck,Wolf, Zimmer
All All
DOCUMENTATIONOPTIONS
Yes Yes Video Video
RECOMMENDEDREPROCESSING
STERIS, STERRAD, EtO,autoclave
STERIS, STERRAD, EtO,autoclave
Liquid, EtO Liquid, EtO, autoclave
PURCHASEINFORMATION
List price Not specified Not specified Not specified Not specifiedWarranty 3 years 3 years 1 year 1 yearDelivery time, ARO Not specified Not specified 1 week 1 weekNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified 1992 1992
Fiscal year January to December January to December Not specified Not specifiedOTHER SPECIFICATIONS None specified. C-mounted scopes fortricam camera.
Sleeves with 1 or 2 fixed orrotatable stopcocks.
Sleeves with 1 or 2 fixed orrotatable stopcocks;repair/exchange program.
LAST UPDATED July 2007 July 2007 April 2004 April 2004Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 18/27
Arthroscopes
18 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL LINVATEC LINVATEC LINVATEC LINVATEC1.9 mm Autoclavable Series 2.5 mm Autoclavable Series 2.9 mm Series 4 mm Series
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 1.9 2.5 2.9 4Insertion length withsleeve, mm
65 72 185 180
OPTICSLens configuration Rod lens Rod lens Rod lens Rod lensDirection of view, ° 0, 30 0, 30, 70 0, 30, 44, 70 0, 30, 70Field of view, ° 60 80 105 105
SLEEVE Yes Yes Yes YesOutside diameter, mm 2.75 3.5 4, 4.7, 5.2 5.5, 5.9Irrigation valves 1 1 1, 2 1, 2Locking system Bayonet mounting J-lock QuickLatch QuickLatch
IRRIGATION FLOW,mL/min
600-800 600-800 800-1,000; 1,500-1,700maximum
800-1,000; 1,500-1,700maximum
COMPATIBLE LIGHTSOURCES
Dyonics, Linvatec, Olympus,Storz, Stryker, Wolf, others
Dyonics, Linvatec, Olympus,Storz, Stryker, Wolf, others
Dyonics, Linvatec, Olympus,Storz, Stryker, Wolf, others
Dyonics, Linvatec, Olympus,Storz, Stryker, Wolf, others
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,autoclave, STERIS,STERRAD
Glutaraldehyde, EtO,autoclave, STERIS,STERRAD
Glutaraldehyde, EtO,autoclave, STERIS,STERRAD
Glutaraldehyde, EtO,STERIS, STERRAD,autoclave
PURCHASEINFORMATION
List price $3,895 $3,850 $3,680 $3,410Warranty 1 year 1 year 1 year 2 yearsDelivery time, ARO Not specified Not specified Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specified
Year first sold 2000 2000 2001 1992Fiscal year January to December January to December January to December January to December
OTHER SPECIFICATIONS Eyecup or cartridge
designs.
Eyecup or cartridge
designs.
Eyecup or cartridge
designs.
Eyecup or cartridge
designs; T1400, T1430, andT1470 couple directly to C-mounted camera head.
LAST UPDATED August 2005 August 2005 August 2005 August 2005Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 19/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 19
Product Comparison Chart
MODEL MAHE\RG Medical OLYMPUS OLYMPUS OLYMPUS Arthroscope A70940A1 : A70941A1 :
A70942A1 A70960A1 : A70961A1 : A70962A1
A70963A1
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 4 : 2.7 4 2.7 2.7Insertion length withsleeve, mm
176 : 145 140 140 50
OPTICSLens configuration Rod lens Rod lens Rod lens Rod lensDirection of view, ° 0, 30, 70 0, 30, 70 0, 30, 70 30Field of view, ° 85-105 115 70 70
SLEEVE Yes Optional Optional OptionalOutside diameter, mm 5.7 : 4 Not specified Not specified Not specifiedIrrigation valves 1.2 1, 2 1, 2 1, 2Locking system Self-locking collar Autolock connection Autolock connection Autolock connection
IRRIGATION FLOW,
mL/min
Not specified 560-700 560-700 560-700
COMPATIBLE LIGHTSOURCES
ACMI, Dyonics, Olympus,Storz, Wolf
All; Olympus recommended All; Olympus recommended All; Olympus recommended
DOCUMENTATIONOPTIONS
Yes Video and still images(JPEG, TIFF)
Video and still images(JPEG, TIFF)
Video and still images(JPEG, TIFF)
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,autoclave
Steam autoclavingrecommended, EtO
Steam autoclavingrecommended, EtO
Steam autoclavingrecommended, EtO
PURCHASEINFORMATION
List price $1,490-2,890 $4,570 $4,690 $4,570Warranty 1 year 5-year Goldtip autoclavibility
guarantee in U.S.5-year Goldtip autoclavibilityguarantee in U.S.
5-year Goldtip autoclavibilityguarantee in U.S.
Delivery time, ARO 1-4 weeks 3-5 days 3-5 days 3-5 daysNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified 2001 2001 2001Fiscal year January to December April to March April to March April to March
OTHER SPECIFICATIONS Sleeves with 1 or 2 rotatingstopcocks available;repair/exchange program.
QuickLock cameraconnection, repair/exchangeprogram.
QuickLock cameraconnection, repair/exchangeprogram.
QuickLock cameraconnection, repair/exchangeprogram.
LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes 1TrueView II 1TrueView II 1TrueView IIData Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 20/27
Arthroscopes
20 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL OLYMPUS OLYMPUS OLYMPUS OLYMPUS A7504A1 : A7505A1 A7506A1 : A7507A1 WA70000A1 : WA70005A1 :
WA70010A1 WA70015A1 : WA70020A1 :WA70025A1
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 1.9 1.9 4 2.7Insertion length withsleeve, mm
45 95 140 140
OPTICSLens configuration Fiber telescope Fiber telescope Rod lens Rod lensDirection of view, ° 0, 30 0, 30 0 : 30 : 70 0 : 30 : 70Field of view, ° 90 90 115 70
SLEEVE Optional Optional Optional OptionalOutside diameter, mm Not specified Not specified Not specified Not specifiedIrrigation valves 1, 2 1, 2 1, 2 1, 2Locking system Autolock connection Autolock connection Autolock connection Autolock connection
IRRIGATION FLOW,
mL/min
560-700 560-700 560-700 560-700
COMPATIBLE LIGHTSOURCES
All; Olympus recommended All; Olympus recommended All; Olympus recommended All; Olympus recommended
DOCUMENTATIONOPTIONS
Video and still images(JPEG, TIFF)
Video and still images(JPEG, TIFF)
Video and still images(JPEG, TIFF)
Video and still images(JPEG, TIFF)
RECOMMENDEDREPROCESSING
Steam autoclavingrecommended, EtO
Steam autoclavingrecommended, EtO
Steam autoclavingrecommended, EtO
Steam autoclavingrecommended, EtO
PURCHASEINFORMATION
List price $4,970 $4,970 $4,720 $4,930Warranty 5-year Goldtip autoclavibility
guarantee in U.S.5-year Goldtip autoclavibilityguarantee in U.S.
5-year Goldtip autoclavibilityguarantee in U.S.
5-year Goldtip autoclavibilityguarantee in U.S.
Delivery time, ARO 3-5 days 3-5 days 3-5 days 3-5 daysNumber sold Not specified Not specified Not specified Not specified
Year first sold 2001 2001 2005 2005Fiscal year April to March April to March April to March April to March
OTHER SPECIFICATIONS QuickLock cameraconnection, repair/exchangeprogram.
QuickLock cameraconnection, repair/exchangeprogram.
One-hand control, directcamera connection withintegrated light guide cable,repair/exchange program.
One-hand control, directcamera connection withintegrated light guide cable,repair/exchange program.
LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes 1TrueView Mini 1TrueView Mini 1TrueView Direct 1TrueView DirectData Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 21/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 21
Product Comparison Chart
MODEL OLYMPUS PRECISION OPTICS REMA RICHARD WOLFWA70021A1 5132-812 : 5132-814 :
5204-8064 mm : 2.7 mm : 1.7 mm 8476.43X
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Not specified Diagnostic
Diameter, mm 2.7 4 4 : 2.7 : 1.7 2.7Insertion length withsleeve, mm
95 176 Not specified 100, 115
OPTICSLens configuration Rod lens Rod lens Rod lens Panoview PlusDirection of view, ° 30 0 : 30 : 70 0 : 30 : 70 25, 70Field of view, ° 70 66 Not specified 96
SLEEVE Optional Not specified Yes YesOutside diameter, mm Not specified Not specified 2.7 : 3.2 : 5.5 4Irrigation valves 1, 2 Not specified Not specified 1Locking system Autolock connection Not specified Self-locking collar Self-locking spring-action
collar
IRRIGATION FLOW,mL/min
560-700 NA Not specified 350
COMPATIBLE LIGHTSOURCES
All; Olympus recommended Most brands ACMI, Aesculap, HSW,Olympus, Storz, Winter,Wolf, Zimmer
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
DOCUMENTATIONOPTIONS
Video and still images(JPEG, TIFF)
Video Yes Yes
RECOMMENDEDREPROCESSING
Steam autoclavingrecommended, EtO
EtO, cidex Glutaraldehyde, EtO,autoclave
Autoclave, EtO,glutaraldehyde, liquid,STERIS
PURCHASEINFORMATION
List price $4,930 $1,490-1,690 Not specified $3,979Warranty 5-year Goldtip autoclavibility
guarantee in U.S.1 year 2 years 1 year
Delivery time, ARO 3-5 days Immediate 4 weeks Immediate
Number sold Not specified Not specified Not specified Not specified Year first sold 2005 Not specified Not specified Not specifiedFiscal year April to March Not specified January to December Not specified
OTHER SPECIFICATIONS One-hand control, directcamera connection withintegrated light guide cable,repair/exchange program.
Couplers and adapters areoptional.
Sleeves with 1 or 2 fixed orrotatable stopcocks;repair/exchange program.
Uses rod lens system.
LAST UPDATED July 2007 August 2005 July 2007 July 2007Supplier Footnotes
Model Footnotes 1TrueView DirectData Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 22/27
Arthroscopes
22 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL RICHARD WOLF RICHARD WOLF RICHARD WOLF RICHARD WOLF8672.4XX 8860.43X 8871.402 8880.5XX
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 2.7 1.9 2.4 4Insertion length withsleeve, mm
180 120 78 170
OPTICSLens configuration Panoview Plus Panoview Plus Panoview Plus Panoview PlusDirection of view, ° 0, 30, 70 5, 25 25 0, 25, 45, 70Field of view, ° 96 96 96 96
SLEEVE Yes Yes Yes YesOutside diameter, mm 4 2.8 3, 3.5 5.6, 6.0, 6.3, 6.6Irrigation valves 1 2 1 1, 2Locking system Self-locking spring-action
collarSelf-locking spring-actioncollar
Self-locking spring-actioncollar
Qui-Connect, self-lockingspring-action collar
IRRIGATION FLOW,
mL/min
350 Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Autoclave, EtO,glutaraldehyde, liquid,STERIS
Autoclave, EtO,glutaraldehyde, liquid,STERIS
Autoclave, EtO,glutaraldehyde, liquid,STERIS
Autoclave, EtO,glutaraldehyde, liquid,STERIS
PURCHASEINFORMATION
List price $4,779 $4,195 $4,075 $3,708Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Immediate Immediate Immediate ImmediateNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified
OTHER SPECIFICATIONS Uses rod lens system. Uses rod lens system. Uses rod lens system. Uses rod lens system.LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 23/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 23
Product Comparison Chart
MODEL RICHARD WOLF RICHARD WOLF RICHARD WOLF RICHARD WOLF8882.4XX 8883.4XX 8883.X73 8885.4XX
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 5 4 4 4Insertion length withsleeve, mm
180 170 170 180
OPTICSLens configuration Panoview 3 Direct couple Direct couple Panoview 3Direction of view, ° 25, 70 30, 70 30 0, 25, 70Field of view, ° 115 100 85 115
SLEEVE Yes Yes Yes YesOutside diameter, mm 6.3, 6.6 5.6, 6.0, 6.3, 6.6 4 5.6, 6.0, 6.3, 6.6Irrigation valves 2 1, 2 1, 2 1, 2Locking system Qui-Connect, self-locking
spring-action collarQui-Connect, self-lockingspring-action collar
Self-locking spring-actioncollar
Qui-Connect, self-lockingspring-action collar
IRRIGATION FLOW,
mL/min
Not specified Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
Olympus, Pilling, Smith &Nephew/Dyonics, Storz
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
Autoclave, EtO,glutaraldehyde, liquid,STERIS
Autoclave, EtO Autoclave, EtO Autoclave, EtO,glutaraldehyde, liquid,STERIS
PURCHASEINFORMATION
List price $3,947 $4,191 $4,191 $4,310Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO Immediate Immediate Immediate ImmediateNumber sold Not specified Not specified Not specified Not specified
Year first sold Not specified Not specified Not specified Not specifiedFiscal year Not specified Not specified Not specified Not specified
OTHER SPECIFICATIONS Uses rod lens system. None specified. None specified. Uses rod lens system.LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 24/27
Arthroscopes
24 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL SMITH & NEPHEW SMITH & NEPHEW SMITH & NEPHEW SMITH & NEPHEW Arthroscope 1.9 : 2.7 : 4 Autoclavable Arthro 2.7 : 4 Autoclavable HD
VideoArthro 4 mm Autoclavable VideoArthro 4
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic/operating Diagnostic/operating Diagnostic/operating Diagnostic/operating
Diameter, mm 1.9 : 2.7 : 4 2.7 : 4 4 4Insertion length withsleeve, mm
65 : 90 : 155 90 : 155 155 115
OPTICSLens configuration Rod lens, 4 mm Rod lens, sapphire distal
and proximal windowsRod lens, sapphire distaland proximal windows, 4mm
Rod lens, sapphire distaland proximal windows
Direction of view, ° 0, 30, 70 0, 30, 70 0, 30, 70 30, 70Field of view, ° 75 : 90 : 115 90 : 115 90 : 115 115
SLEEVE Optional Yes Optional YesOutside diameter, mm 2.9, 3.5, 4.5, 5.5 2.9, 4.5, 5.5 2.2, 2.9, 3.6, 4.5, 5.5 2.9, 4.5, 5.5Irrigation valves 1, 2 1, 2 1, 2 1, 2
Locking system J-lock, Storz J-lock J-lock J-lockIRRIGATION FLOW,mL/min
250 (4.5 sheath) 250 (4.5 sheath) 250 (4.5 sheath) 250 (4.5 sheath)
COMPATIBLE LIGHTSOURCES
ACMI, Olympus, Smith &Nephew, Storz, Stryker,Weck, Wolf, Zimmer
ACMI, Olympus, Smith &Nephew, Storz, Stryker,Weck, Wolf, Zimmer
ACMI, Olympus, Smith &Nephew, Storz, Stryker,Weck, Wolf, Zimmer
ACMI, Olympus, Smith &Nephew, Storz, Stryker,Weck, Wolf, Zimmer
DOCUMENTATIONOPTIONS
Yes Yes Yes Yes
RECOMMENDEDREPROCESSING
EtO, STERIS, STERRAD EtO, STERIS, STERRAD,steam
EtO, STERIS, STERRAD,Steam
EtO, STERIS, STERRAD,steam
PURCHASEINFORMATION
List price $3,338 $3,600 $4,500 $4,174Warranty 1 year 1 year 1 year 1 yearDelivery time, ARO 24 hr 24 hr 24 hr 24 hrNumber sold Not specified Not specified Not specified Not specified
Year first sold 1978 1994 2007 2002Fiscal year Not specified Not specified Not specified Not specifiedOTHER SPECIFICATIONS Interchangeable cannula
system; repair/exchangeprogram; stainless steelconstruction with sapphiredistal window.
Direct view. Arthroscope couples directlyto camera; optimized for usewith HD video camera.
Compatible withautoclavable camera,couples directly to camerahead.
LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 25/27
Arthroscopes
©2007 ECRI Institute. All Rights Reserved 25
Product Comparison Chart
MODEL SMITH & NEPHEW SOLOS ENDOSCOPY SOPRO-COMEG SOPRO-COMEGVideo Arthroscope 1.9 : 2.7: 4
AR8400 : AR8430 : AR8470 161 191 040 161 240 [980 : 982]
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic/operating Diagnostic Diagnostic Diagnostic
Diameter, mm 1.9 : 2.7 : 4 4 1.9 2.4Insertion length withsleeve, mm
75 : 90 : 155 Not specified 73 69
OPTICSLens configuration Rod lens, 4 mm Rod lens, sapphire distal
windowRod lens Rod lens
Direction of view, ° 0, 30, 70 0 : 30 : 70 0 0 : 30, WAField of view, ° 75 : 90 : 115 105 45 98-106
SLEEVE Optional Yes Not specified Not specifiedOutside diameter, mm 2.2, 2.9, 3.6, 4.5, 5.5 Not specified 2.5 3.2, 3.8 (high flow)Irrigation valves 1, 2 1, 2 1 1Locking system J-lock Bayonet mounting Standard Standard
IRRIGATION FLOW,mL/min
250 (4.5 sheath) Not specified Not specified Not specified
COMPATIBLE LIGHTSOURCES
ACMI, Olympus, Smith &Nephew, Storz, Stryker,Weck, Wolf, Zimmer
ACMI, Olympus, Solos,Storz, Stryker, Wolf
All common models All common models
DOCUMENTATIONOPTIONS
Yes Yes Not specified Yes
RECOMMENDEDREPROCESSING
EtO, STERIS, STERRAD Glutaraldehyde, steam,autoclave, EtO,
Glutaraldehyde, STERIS,STERRAD, EtO, autoclave
Glutaraldehyde, STERIS,STERRAD, EtO, autoclave
PURCHASEINFORMATION
List price $3,795 $2,500 $2,052 $1,946 : $1,993Warranty 1 year 1 year 2 years 2 yearsDelivery time, ARO 24 hr 1 day Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specified
Year first sold 1987 1987 2005 1990
Fiscal year Not specified April to March January to December January to DecemberOTHER SPECIFICATIONS Arthroscope couples directlyto camera; no independentcoupler needed; stainlesssteel construction withsapphire distal windows.
Repair in 3 days;autoclavable with renewedwarranty.
None specified. None specified.
LAST UPDATED July 2007 August 2005 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 26/27
Arthroscopes
26 ©2007 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL SOPRO-COMEG SOPRO-COMEG SOPRO-COMEG STRYKER ENDOSCOPY161 271 [770 : 772 : 775] 161 411 [800 : 802 : 803 :
805]161 412 255 342
WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE Yes Yes Yes YesCE MARK (MDD) Yes Yes Yes YesTYPE Diagnostic Diagnostic Diagnostic Diagnostic
Diameter, mm 2.7 4 4 2.3Insertion length withsleeve, mm
119 119 169 74
OPTICSLens configuration Rod lens Rod lens Rod lens Rod lensDirection of view, ° 0 : 30 : 70, WA 0 : 30 : 45 : 70, WA 70 85-120Field of view, ° 98-106 98-106 98-106 120
SLEEVE Not specified Not specified Not specified YesOutside diameter, mm 4.0 5.6, 6.5 (high flow) 5.6 3.5Irrigation valves 1, 2 1, 2 2 1, 2Locking system Standard Standard Standard Self-locking collar, speed
lock
IRRIGATION FLOW,mL/min
Not specified Not specified Not specified 200
COMPATIBLE LIGHTSOURCES
All common models All common models All common models ACMI, Smith and Nephew,Storz, Stryker, Weck, Wolf
DOCUMENTATIONOPTIONS
Yes Yes Not specified Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, STERIS,STERRAD, EtO, autoclave
Glutaraldehyde, STERIS,STERRAD, EtO, autoclave
Glutaraldehyde, STERIS,STERRAD, EtO, autoclave
Glutaraldehyde, EtO,peracetic acid, STERRAD;peracetic acid is used in theSTERIS SYSTEM 1, withanticorrosive chemicals
PURCHASEINFORMATION
List price $1,946 : $2,088 : $2,088 $1,790 : $1,922 : $1,922 :$1,922
$1,701 $4,606
Warranty 2 years 2 years 2 years 90 days
Delivery time, ARO Not specified Not specified Not specified Not specifiedNumber sold Not specified Not specified Not specified Not specified Year first sold 1981 1981 2005 1999Fiscal year January to December January to December January to December Not specified
OTHER SPECIFICATIONS None specified. Ready Reserverepair/exchange available.
None specified. C-mount; direct-coupleddevice; eyepiece scopes.
LAST UPDATED July 2007 July 2007 July 2007 July 2007Supplier Footnotes
Model Footnotes
Data Footnotes
Comprehensive of Biomedical Engineering Site (www.dezmed.com
W
W W . D
E Z M E D
. C O M
8/9/2019 Arthroscope s
http://slidepdf.com/reader/full/arthroscope-s 27/27
Arthroscopes
Product Comparison Chart
MODEL STRYKER ENDOSCOPY STRYKER ENDOSCOPY502 Series 502-477, -577 Series
WHERE MARKETED Worldwide WorldwideFDA CLEARANCE Yes YesCE MARK (MDD) Yes YesTYPE Diagnostic Diagnostic
Diameter, mm 1.9, 2.7, 2.7 4Insertion length withsleeve, mm
65, 140 140, 166
OPTICSLens configuration Rod lens Rod lensDirection of view, ° 0, 30, 70 0, 30, 45, 70Field of view, ° 80, 120, 80, 120 105
SLEEVE Yes YesOutside diameter, mm 2.8, 4 5, 5.8, 6.5Irrigation valves 1, 2 1, 2Locking system J-lock, speed lock Self-locking collar, speed
lockIRRIGATION FLOW,
mL/min
100, 200 200, 500, 900
COMPATIBLE LIGHTSOURCES
ACMI, Smith and Nephew,Storz, Stryker, Weck, Wolf
ACMI, Smith and Nephew,Storz, Stryker, Weck, Wolf
DOCUMENTATIONOPTIONS
Yes Yes
RECOMMENDEDREPROCESSING
Glutaraldehyde, EtO,peracetic acid, autoclavable;peracetic acid is used in theSTERIS SYSTEM 1, withanticorrosive chemicals
Glutaraldehyde, EtO,peracetic acid, STERRAD,autoclave; peracetic acid isused in the STERISSYSTEM 1, withanticorrosive chemicals
PURCHASEINFORMATION
List price $5,182-5,190 $4,991-5,183Warranty 90 days 90 daysDelivery time, ARO Not specified Not specified
Number sold Not specified Not specified Year first sold 2002 1995Fiscal year Not specified Not specified
OTHER SPECIFICATIONS None specified. Autoclavable.LAST UPDATED July 2007 July 2007Supplier Footnotes