Upload
others
View
4
Download
0
Embed Size (px)
Citation preview
1
ARIX Volar Distal Radius Locking Plate System
WRIST SYSTEM
2
ARIX WRIST SYSTEM
Table of Contents
Indications
Product Overview
Features & Benefits
Ordering Information - Screws
2.5mm Self-Tapping Cortical Screws (Non-Locking)
2.5mm Self-Tapping Locking Screws
Ordering Information - Plates
ARIX Volar Distal Radius Plates
Ordering Information - Instruments
Set Configuration Suggestion
3
4
7
14
15
16
19
21
3
ARIX Volar Distal Radius Locking Plate System
Indications
ARIX Volar Distal Radius Locking Plate System is intended for use in temporary internal fixation and
stabilization of fractures, osteotomies, and non-unions of the distal radius.
4
ARIX WRIST SYSTEM
Screws : Silver
Plates : Blue
Product Overview
Screw Specifications
Screw TypesOuter
DiameterLength Insertion Types
Screw Head Recess
Color Material
Locking ScrewØ2.5mm 8 ~30mm Self-Tapping Hexagonal Silver Titanium
AlloyCortical Screw(Non-Locking)
Distal Head Types Thickness Length Holes on the shaft Color Material Guide Pin Hole
Standard1.8mm 46mm and
73mm2 holes and
4 holes Blue Pure Titanium Gr.4 Ø1.3mmLarge
Plate Specifications
Color Coding
All plates are made from pure titanium material in conformity with ASTM F67.
All screws are made from titanium alloy material in conformity with ASTM F136.
The titanium is an ideal material for implant.
It is biocompatible, corrosion resistant, and non-toxic in biological environment.
Materials of Implants
5
ARIX Volar Distal Radius Locking Plate System
Self-retaining Hexagonal Recess for connection between Screw Head and Screwdriver Shaft
Product Overview
Screw Head Recess
Screw Options
Self-tapping Locking Screws : 8mm ~ 30mm
Self-tapping Cortical Screws(Non-Locking) : 8mm ~ 30mm
Outer Diameter : 2.5mm
Locking ScrewCortical Screw(Non-Locking)
6
ARIX WRIST SYSTEM
8 Volar Distal Radius Plates for left and right radiuses.
2 types of Distal Head as Standard and Large
2 types of Shaft Length as 2 and 4 locking holes on the shaft
Product Overview
Plate Options
Standard Standard StandardLarge LargeLarge Large
7
ARIX Volar Distal Radius Locking Plate System
Features and Benef its
Anatomical Volar Plate
Anatomical plate contour and screw trajectories for maximum support of
lunate facet and stable fixation of radial styloid
Radial styloid fragmentLunate facet
Minimal soft tissue irritation
Rounded plate edges to minimize soft tissue irritation
Rounded screw head design and minimal screw head prominence in the plate
to reduce the potential for soft tissue irritation.
Low overall profile of the plate and screw construct for minimal soft tissue irritation.
Minimal screw head prominence Rounded screw head
Low overall profile
Anatomical 4 plane cutting for less soft tissue irritation
Rounded plate edges
8
ARIX WRIST SYSTEM
Features and Benef its
Poly-Axial Locking System
Benefits of Locking Plating System
The poly-axial hole of a plate accepts locking or cortical screws as surgical needs.
Locking screws as well as cortical screws can be inserted in ±10° with respect to
the plate in any direction.
Locking plate and screw system provides more stable fixation than conventional plate and screw system.
Locking screws will not loosen from the plate during the healing process.
Locking screws do not rely on bone to plate compression. A fixed angle construct can be created by using
locking screws in osteopenic bone or multifragmentary fractures where secure bone purchase with conventional
screws may be compromised.
10 °
9
ARIX Volar Distal Radius Locking Plate System
Features and Benef its
Oblong hole for accurate plate positioning
Easy Plate Identification
Distal head size information such as standard or large and plate type information
such as left or right are marked on plates for easy plate identification.
The oblong hole in the plate shaft allows distal or proximal adjustment for accurate
plate positioning on the bone. 1mm increment lines are marked
beside oblong hole as a reference for plate positioning.
Distal Head Size
Left of Right type
10
ARIX WRIST SYSTEM
Features and Benef its
Guide Pin holes for temporary plate fixation
The Guide Pin holes of a plate accept up to 1.2mm Guide Pin.
Guide Pin can be placed for temporary maintenance
of bone reduction and provisional plate fixation.
Intraoperative simplicity in choosing screws
2.5mm locking and cortical screws can be used in any holes of a plate
for intraoperative simplicity
Guide Pin holes to approximate the trajectories of the distal screws
with respect to radiocarpal joint
Guide Pin to approximate the trajectories of the ulna
screws with respect to DRUJ
Guide Pin holes for provisional fixation of plate.
2.5mm 2.5mm
11
ARIX Volar Distal Radius Locking Plate System
Drill Guide Block
The drill guide block enables insertion of screws in the nominal trajectory of each locking hole.
Works with scaled drill guide of the variable angle drill guide(111-080)
Total 4 models in standard and large sizes for left and right plates.
The drill guide blocks are made from radiolucent material
Fixed angle screw trajectories by drill guide block[ Large size for right radius ]
Fixed angle screw trajectories by drill guide block[ Standard size for right radius ]
Distal 10° Radial 10°
Proximal 5°Ulnar 3°
Proximal 5°Radial 2°
Proximal 5° Ulnar 2°
Distal 5° Radial 10°
Distal 10° Ulnar1°
Distal 10° Radial 0°
Features and Benef its
Ulnar side Radial side
Distal 10° Radial 10°Proximal 5°
Radial 0°
Proximal 5°Radial 2°
Proximal 5° Radial 2°
Distal 5° Radial 10°
Distal 5° Radial 2° Distal 5°
Radial 1.5°
Ulnar side Radial side
12
ARIX WRIST SYSTEM
Variable Angle Drill Guide
Fixed Angle Drill Sleeve
Helps to drill a pilot hole within the poly-axial range of ±10°.
Scales are marked on drill guide and drill bit for prompt depth measurement of
a pilot hole without the need of depth gauge.
The fixed angle drill sleeve helps to drill a hole in the nominal trajectory of each locking hole.
Scales are marked for prompt depth measurement of a pilot hole.
Features and Benef its
10°
10°
13
ARIX Volar Distal Radius Locking Plate System
Determine the optimal screw length with the depth gauge.
The triangular-prism shaped scale of depth gauge
shows the exact length in any direction.
Depth Gauge
Plate Bender
If necessary, bend distal radius plate to suit the patients’
anatomical conditions using a pair of plate benders.
With the plate benders, hold & bend the plate around the center window.
Do Not bend the plate more than 5°.
Further or Repetitive bending may break the plate.
Features and Benef its
* 3 scales for ease of reading
14
ARIX WRIST SYSTEM
Ordering Information - Screws
2.5mm Self-Tapping Cortical Screws (Non-Locking)
No. Code Head Diameter (A) Outer Diameter (B) Length (L)
1 25-HF-008
Ø3.7 Ø2.5
8
2 25-HF-010 10
3 25-HF-012 12
4 25-HF-014 14
5 25-HF-016 16
6 25-HF-018 18
7 25-HF-020 20
8 25-HF-022 22
9 25-HF-024 24
10 25-HF-026 26
11 25-HF-028 28
12 25-HF-030 30
A
L
B
[Unit: mm]
15
ARIX Volar Distal Radius Locking Plate System
Ordering Information - Screws
2.5mm Self-Tapping Locking Screws
No. Code Head Diameter (A) Outer Diameter (B) Length (L)
1 25L-HF-008
Ø3.8 Ø2.5
8
2 25L-HF-010 10
3 25L-HF-012 12
4 25L-HF-014 14
5 25L-HF-016 16
6 25L-HF-018 18
7 25L-HF-020 20
8 25L-HF-022 22
9 25L-HF-024 24
10 25L-HF-026 26
11 25L-HF-028 28
12 25L-HF-030 30
A
L
B
[Unit: mm]
16
ARIX WRIST SYSTEM
25-DVRA-109-L Volar Distal Radius Locking Plate Left, Standard, Length 46mm, 9 Holes, Thickness 1.8mm, Blue
25-DVRA-109-R Volar Distal Radius Locking Plate Right, Standard, Length 46mm, 9 Holes, Thickness 1.8mm, Blue
Ordering Information – Plates
ARIX Volar Distal Radius Plates
25-DVRA-111-L Volar Distal Radius Locking Plate Left, Standard,Length 73mm, 11 Holes, Thickness 1.8mm, Blue
17
ARIX Volar Distal Radius Locking Plate System
25-DVRA-111-R Volar Distal Radius Locking Plate Right, Standard, Length 73mm, 11 Holes, Thickness 1.8mm, Blue
25-DVRA-209-L Volar Distal Radius Locking Plate Left, Large, Length 46mm, 9 Holes, Thickness 1.8mm, Blue
Ordering Information – Plates
25-DVRA-209-R Volar Distal Radius Locking Plate Right, Large, Length 46mm, 9 Holes, Thickness 1.8mm, Blue
18
ARIX WRIST SYSTEM
25-DVRA-211-L Volar Distal Radius Locking Plate Left, Large, Length 73mm, 11 Holes, Thickness 1.8mm, Blue
25-DVRA-211-R Volar Distal Radius Locking Plate Right, Large, Length 73mm, 11 Holes, Thickness 1.8mm, Blue
Ordering Information – Plates
19
ARIX Volar Distal Radius Locking Plate System
Ordering Information – Instruments
111-082-LDrill Guide Block, Standard, Left
111-082-RDrill Guide Block, Standard, Right
111-083-LDrill Guide Block, Large, Left
111-083-RDrill Guide Block, Large, Right
113-HF-607Hexagonal Screwdriver Shaft
112-25-301Drill Bit for 2.5mm screws, Ø2.0mm,
Drilling length : 30mm, Overall length : 110mm, J-Latch
111-068-2Guide Pin Ø1.1mm for Ø1.3 Plate hole
111-084Fixed Angle Drill Sleeve
for Ø2.0mm drill bit
20
ARIX WRIST SYSTEM
111-075Depth Gauge for 2.5mm screws
111-080Variable Angle Drill Guide for Ø2.0mm drill bit
BD-05Grasping Forceps
26.0240.17Plate Bender
Ordering Information – Instruments
111-092Screwdriver Handle
112-084ARIX Volar Distal Radius Implant/Instrument Container
21
ARIX Volar Distal Radius Locking Plate System
ARIX WRISTØ2.5 Distal Radius Plate
Set Conf iguration Suggestion
NO Code Description Set Quantity
Implants _ Screws
1 25L-HF-008 2.5 Locking Screw (8mm) 5
2 25L-HF-010 2.5 Locking Screw (10mm) 5
3 25L-HF-012 2.5 Locking Screw (12mm) 10
4 25L-HF-014 2.5 Locking Screw (14mm) 10
5 25L-HF-016 2.5 Locking Screw (16mm) 10
6 25L-HF-018 2.5 Locking Screw (18mm) 10
7 25L-HF-020 2.5 Locking Screw (20mm) 10
8 25L-HF-022 2.5 Locking Screw (22mm) 10
9 25L-HF-024 2.5 Locking Screw (24mm) 5
10 25L-HF-026 2.5 Locking Screw (26mm) 5
11 25L-HF-028 2.5 Locking Screw (28mm) 5
12 25L-HF-030 2.5 Locking Screw (30mm) 5
13 25-HF-008 2.5 Cortical Screw (8mm) 5
14 25-HF-010 2.5 Cortical Screw (10mm) 5
15 25-HF-012 2.5 Cortical Screw (12mm) 10
16 25-HF-014 2.5 Cortical Screw (14mm) 10
17 25-HF-016 2.5 Cortical Screw (16mm) 10
18 25-HF-018 2.5 Cortical Screw (18mm) 10
19 25-HF-020 2.5 Cortical Screw (20mm) 10
20 25-HF-022 2.5 Cortical Screw (22mm) 10
21 25-HF-024 2.5 Cortical Screw (24mm) 5
22 25-HF-026 2.5 Cortical Screw (26mm) 5
23 25-HF-028 2.5 Cortical Screw (28mm) 5
24 25-HF-030 2.5 Cortical Screw (30mm) 5
Implants _ Plates
25 25-DVRA-109-R Standard, Length 46mm, Right 3
26 25-DVRA-109-L Standard, Length 46mm, Left 3
27 25-DVRA-111-R Standard, Length 73mm, Right 3
28 25-DVRA-111-L Standard, Length 73mm, Left 3
29 25-DVRA-209-R Large, Length 46mm, Right 3
30 25-DVRA-209-L Large, Length 46mm, Left 3
31 25-DVRA-211-R Large, Length 73mm, Right 3
32 25-DVRA-211-L Large, Length 73mm, Left 3
NO Code Description Set Quantity
Instruments
33 111-092 Screwdriver Handle 2
34 111-075 Depth Gauge 1
35 112-25-301 Drill bit for 2.5mm screws 2
36 111-080 Variable Angle Drill Sleeve for Ø2.0mm drill bit
1
37 111-084 Fixed Angle Drill Sleeve for Ø2.0 drill bit
1
38 BD-05 Grasping Forceps 1
39 113-HF-607 Hexagonal Screwdriver Shaft 2
40 111-068-2 Guide Pin 2
41 111-082-R Drill Guide Block, Standard, Right 1
42 111-082-L Drill Guide Block, Standard, Left 1
43 111-083-R Drill Guide Block, Large, Right 1
44 111-083-L Drill Guide Block, Large, Left 1
45 26.0240.17 Bender 2
22
ARIX WRIST SYSTEM
2.5 Distal Radius Plates and Screws
23
ARIX Volar Distal Radius Locking Plate System
Instruments for 2.5 Distal Radius Plates and Screws
JEIL MEDICAL CORPORATION 702•703•704•705•706•804•805•807•812-ho,
55, Digital-ro34-gil, Guro-gu, Seoul, 152-728, Korea
Tel : 82.2.850.3500 Fax: 82.2.850.3535 www.jeilmed.co.kr
Publication 201501
LBEN-M1601(00)