Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
Application of an Optimisation
Algorithm to Configure an Internal
Fixation Device
Salma Ibrahim, B.E. (Medical)
Submitted for the award of the degree of Master of
Engineering in School of Engineering Systems of the faculty
for Built Environment and Engineering, Queensland
University of Technology
2010
Optimisation of an Internal Fixation Device
i
Keywords
Biomechanics, Optimisation Algorithm, Finite Element Modelling,
Fracture Healing, Internal Fracture Fixation
Optimisation of an Internal Fixation Device
ii
Abstract
Project Title: Application of an optimisation algorithm to configure an internal fixation device
Author: Salma Ibrahim
Supervisors: Dr. Sanjay Mishra (Primary)
Dr. Gongfa Chen (Secondary)
Fractures of long bones are sometimes treated using various types of fracture
fixation devices including internal plate fixators. These are specialised plates which
are used to bridge the fracture gap(s) whilst anatomically aligning the bone
fragments. The plate is secured in position by screws. The aim of such a device is to
support and promote the natural healing of the bone.
When using an internal fixation device, it is necessary for the clinician to decide
upon many parameters, for example, the type of plate and where to position it; how
many and where to position the screws. While there have been a number of
experimental and computational studies conducted regarding the configuration of
screws in the literature, there is still inadequate information available concerning
the influence of screw configuration on fracture healing.
Because screw configuration influences the amount of flexibility at the area of
fracture, it has a direct influence on the fracture healing process. Therefore, it is
important that the chosen screw configuration does not inhibit the healing process.
In addition to the impact on the fracture healing process, screw configuration plays
an important role in the distribution of stresses in the plate due to the applied loads.
A plate that experiences high stresses is prone to early failure. Hence, the screw
configuration used should not encourage the occurrence of high stresses.
This project develops a computational program in Fortran programming language to
perform mathematical optimisation to determine the screw configuration of an
Optimisation of an Internal Fixation Device
iii
internal fixation device within constraints of interfragmentary movement by
minimising the corresponding stress in the plate. Thus, the optimal solution suggests
the positioning and number of screws which satisfies the predefined constraints of
interfragmentary movements. For a set of screw configurations
the interfragmentary displacement and the stress occurring in the plate were
calculated by the Finite Element Method. The screw configurations were iteratively
changed and each time the corresponding interfragmentary displacements were
compared with predefined constraints. Additionally, the corresponding stress was
compared with the previously calculated stress value to determine if there was a
reduction. These processes were continued until an optimal solution was achieved.
The optimisation program has been shown to successfully predict the optimal screw
configuration in two cases. The first case was a simplified bone construct whereby
the screw configuration solution was comparable with those recommended in
biomechanical literature. The second case was a femoral construct, of which the
resultant screw configuration was shown to be similar to those used in clinical cases.
The optimisation method and programming developed in this study has shown that
it has potential to be used for further investigations with the improvement of
optimisation criteria and the efficiency of the program.
Optimisation of an Internal Fixation Device
iv
Table of Contents
Keywords ........................................................................................................................................................... i
Abstract ............................................................................................................................................................ ii
Table of Contents ........................................................................................................................................ iv
Figures and Tables ................................................................................................................................... vii
Abbreviations used in the text .............................................................................................................. ix
Statement of Originality............................................................................................................................ x
Acknowledgements.................................................................................................................................... xi
1. Introduction ......................................................................................................................................... 1
1.1. Background .................................................................................................................................. 1
1.2. Problem .......................................................................................................................................... 1
1.3. Aims ................................................................................................................................................. 3
1.4. Significance of the Study ........................................................................................................ 3
1.5. Outline of Thesis ........................................................................................................................ 4
2. Literature Review and Background ............................................................................................... 6
2.1. Treatment of Long Bone Fractures ........................................................................................... 6
2.1.1. Internal Fixators .................................................................................................................... 7
2.1.2. Fracture Healing .................................................................................................................... 9
2.2. Factors Influencing the Strength of the Fixation Construct and Bone Healing11
2.2.1. Stiffness of Fracture Fixation ........................................................................................ 12
2.2.2. Physical Conditions for Fracture Healing ............................................................... 12
2.3. Influence of Working Length and Fracture Gap on Fixation Stability ............... 14
2.4. Screw Positioning ....................................................................................................................... 18
2.5. Limitations of previous studies ............................................................................................ 19
2.6. Summary ......................................................................................................................................... 20
3. Methods - Optimisation .................................................................................................................... 21
3.1. Mathematical Definition of Optimisation ........................................................................ 21
3.2. Types of Optimisation Problems and How to Solve Them ...................................... 22
3.2.1. Constrained/ Unconstrained Optimisation Problems ...................................... 22
3.2.2. Multi-modal Optimisation .............................................................................................. 24
3.2.3. Deterministic Methods .................................................................................................... 26
Optimisation of an Internal Fixation Device
v
3.3. Powell’s method ...........................................................................................................................27
3.3.1. Conjugate Directions .........................................................................................................27
3.3.2. The Algorithm .......................................................................................................................28
3.3.3. Golden Section Search – Search in One Direction. ..............................................32
3.4. Use of optimisation methods in medical engineering................................................33
3.5. Optimisation of Screw Configuration in Internal Fixators ......................................35
3.5.1. Objectives and Constraints.............................................................................................35
3.5.2. Optimisation Criteria ........................................................................................................36
3.5.3. Objective Function ........................................................................................................37
3.5.4. Calculation of Function Value (with the use of FE method)......................37
3.5.5. Data Transfer ...................................................................................................................41
4. Results........................................................................................................................................................44
4.1. Case 1: Simplified Model ..........................................................................................................44
4.1.1. Bone Geometry.....................................................................................................................44
4.1.2. Plate and Screws Geometry ...........................................................................................44
4.1.3. Material Properties ............................................................................................................46
4.1.4. Boundary and Loading Conditions .............................................................................47
4.1.5. Variables to be Optimised ...............................................................................................47
4.1.6. Selection of Values for Optimisation Criteria ........................................................47
4.1.7. Solution for Simplified Model .......................................................................................51
4.2. Case 2: Clinical Model ................................................................................................................55
4.2.1. Clinical Cases .........................................................................................................................55
4.2.2. Additional Cases ..................................................................................................................57
4.2.3. Femoral Bone Geometry ..................................................................................................57
4.2.4. Plate and Screws of Femoral Construct ...................................................................58
4.2.5. Assembly .................................................................................................................................58
4.2.6. Materials..................................................................................................................................58
4.2.7. Loading and Boundary Conditions .............................................................................59
4.2.8. Variables to be Optimised ...............................................................................................59
4.2.9. Selection of Optimisation Criteria...............................................................................62
4.2.10. Solution .................................................................................................................................64
5. Discussion ...........................................................................................................................................72
5.1. Limitations of this Study ......................................................................................................74
Optimisation of an Internal Fixation Device
vi
5.2. Improvements to the Model................................................................................................... 76
5.3. Improvement to the Optimisation Criteria ................................................................ 77
5.4. Future Work - Improvements to the Optimisation Method ............................... 79
6. Conclusions ........................................................................................................................................ 81
7. References .......................................................................................................................................... 83
Appendix ....................................................................................................................................................... 87
Optimisation program including subroutines in Fortran................................................. 87
Python script file to read out values from FEA ...................................................................... 94
Optimisation of an Internal Fixation Device
vii
Figures and Tables
Figure 1 LCP-combination hole allowing conventional plate fixation as well as application of
locked screws (Source: (Perren, 2002)).................................................................................................. 7
Figure 2 Internal fixator used with locked screws. Fixator barely touches the bone as screws
allow reliable maintenance of the initial distance between internal fixator and bone
(Source: (Perren, 2002)). ................................................................................................................................ 8
Figure 3 Direct healing from osteotomy of sheep tibia with compression stabilisation. The
bone fragments are close and compressed and there is no displacement at the site of
the osteotomy. The shape of the osteones do not change when crossing the fracture.
(Source: Perren, 2002) ..................................................................................................................................... 9
Figure 4 Histological images of secondary fracture healing in bone (Source: J. Bone Miner.
Res., 16, 1004– 1014, 2001). ...................................................................................................................... 11
Figure 5 Left: X-ray image of femoral fracture in 35 year old male with flexible fixation.
Right: X-ray image 4 months after fixation, showing obvious signs of callus growth.
Source: (Chen et al., 2010). .......................................................................................................................... 15
Figure 6 Seven week postoperative x-ray showing fracture fixation by placing several
locking screws in main fragments. The screw holes were occupied adjacent to the
fracture site resulting in high stress concentrations occurring in that section of the
plate. Source: (Sommer et al., 2003) ...................................................................................................... 16
Figure 7 Example of contours of an objective function (Source: Rao, S. S.; Engineering
Optimization-Theory and Practice, 3rd Ed. 1996, pp.363) ........................................................ 23
Figure 8 A multi-modal function. Source: (Singh et al., 2006) ............................................................. 25
Figure 9 Conjugate Direction (Source: Rao, S. S.; Engineering Optimization-Theory and
Practice, 3rd Ed. 1996, pp.363) ................................................................................................................. 28
Figure 10 Progress of Powell's Method (Source: Rao, S. S.; Engineering Optimization-Theory
and Practice, 3rd Ed. 1996, pp.363) ....................................................................................................... 31
Figure 11 Illustration of Golden Section Search .......................................................................................... 32
Figure 12 Showing data transfer between different software packages ........................................ 41
Figure 13 Screw positions (variables) to be optimised in the simplified model ........................ 45
Figure 14 Mesh of the simplified cylindrical model .................................................................................. 47
Figure 15 (a) Rigid simplified construct, (b) flexible simplified construct ................................... 48
Figure 16 Nodes used to calculate displacements ..................................................................................... 49
Figure 17 Showing sharp edge (a cause of FE errors) in screw holes of the locking
compression plate ............................................................................................................................................ 51
Figure 18 Optimised solution for simplified model................................................................................... 51
Figure 19 Maximum principal stress distribution in cylindrical construct .................................. 52
Figure 20 (a) treatment of transverse fracture of 73 yr old patient (b) X-ray image showing
failure of implant 7 weeks post-op (c) treatment of fracture of a 35 year old male (d) X-
ray showing successful healing of fracture ......................................................................................... 56
Figure 21 Shows 4 fixed screws (black, 2 at each end of plate) and 6 screw positions
(yellow) to be optimised............................................................................................................................... 60
Figure 22 (a) Fracture healing in patient after 4 months using a flexible screw configuration.
(Source: J Eng Med. Chen et al, 2010); (b) Simulation of the same combination used for
FE analysis ........................................................................................................................................................... 61
file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566205file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566207file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566215file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566215
Optimisation of an Internal Fixation Device
viii
Figure 23 (a) flexible construct, (b) rigid construct .................................................................................. 63
Figure 24 (a) Screw configuration used in clinical case from Chen et al (2010); (b) Resultant
screw configuration from optimisation algorithm .......................................................................... 65
Figure 25 Maximum principal stress distribution in femoral construct of the optimised
solution .................................................................................................................................................................. 67
Figure 26 (a) Flexible construct, (b) Construct with more rigidity due to shorter working
length ...................................................................................................................................................................... 68
Figure 27 Some screw configurations that were tried and tested by the optimisation
algorithm. White represents screws that were chosen by the optimisation algorithm
that were tested. Black represents screws that were fixed throughout the optimisation
process. .................................................................................................................................................................. 73
Figure 28 Illustration of concept of local versus global minimisation ............................................. 75
Figure 29 “Boundaries for optimal healing in the sheep model[s] that lead to timely healing”
(Source: Epari et al, 2007) ........................................................................................................................... 78
Table 1Interfragmentary displacement and maximum principal stress for the most rigid and
the most flexible cylindrical models ....................................................................................................... 49
Table 2 Comparison of displacements from solution and those from constraints for the
cylindrical model .............................................................................................................................................. 52
Table 3 Comparison of displacement and stress resulting from flexible and rigid construct
with that of solution construct for the cylindrical model ............................................................ 53
Table 4 Shear, axial displacement and stress in plate resulting from the configuration from
Chen et al (2010)............................................................................................................................................... 62
Table 5 Interfragmentary displacement and maximum principal stress for most rigid and
most flexible femur models ......................................................................................................................... 64
Table 6 Comparison of displacements from solution and those from constraints .................... 67
Table 7 Comparison of displacement and stress resulting from flexible and rigid construct
with that of solution construct .................................................................................................................. 68
Table 8 Axial and shear displacement resulting from the flexible and rigid constructs from
Figure 27 (a) and (b) ....................................................................................................................................... 69
Table 9 Axial and shear displacements resulting from the removal of pairs of screws from
each side of the fracture gap from the all screws in place construct ..................................... 70
Table 10 Comparison of displacement and stress from optimised solution with that from
screw configuration used in clinical case from Chen et al (2010) .......................................... 70
file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566220file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566220file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566221file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566221file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566221file:///C:/Users/Ibrahim/Desktop/BeginningThesisCloseRevisedDraft%5b1%5dWithoutTrackChanges%5b1%5d.docx%23_Toc279566221
Optimisation of an Internal Fixation Device
ix
Abbreviations used in the text
FE = Finite Element
LCP = Locking Compression Plate
LISS = Less Invasive Stabilising System
DCP = Dynamic Compression Plate
TSP = Travelling Salesman Problem
Optimisation of an Internal Fixation Device
x
Statement of Originality
“The work contained in this thesis has not been previously submitted for a degree or
diploma at any other higher education institution. To the best of my knowledge and
belief, the thesis contains no material previously published or written by another
person except where due reference is made.”
Signature: ________________________________________
Date: ______________________________________________
Optimisation of an Internal Fixation Device
xi
Acknowledgements
I would like to thank my supervisors, Dr. Gongfa Chen and Dr. Sanjay Mishra for
their help, support and guidance throughout this study. I am grateful to the trauma
team at IHBI, my fellow colleagues and friends making the research environment
more enjoyable, and Mr. Mark Barry and the HPC team for their assistance with the
supercomputer.
Optimisation of an Internal Fixation Device
1
1. Introduction
1.1. Background
Severe trauma to the extremities is the leading cause of disability during the wage-
earning period of life. Bone fractures cost the Australian healthcare system one
billion dollars a year. In addition to this cost, the functional loss of limbs impact
significantly on the patients’ quality of life. Studying the impact of fixation devices
on bone healing will fill knowledge gaps and enhance the usefulness of these
devices for the purpose of fracture healing. This will ultimately reduce costs and
improve quality of life for the patient.
High-energy collisions with long bones often result in fractures with significant
misalignments of bone fragments. In these cases it is difficult for the body to pursue
its natural healing course in order to produce a successful healing outcome. For
these instances, surgical fracture treatment is usually required. There are a number
of fracture fixation devices available, including external fixators, intermedullary
nails and internal plate fixators. The need to use any one of them depends on the
physical characteristics of the trauma. Ultimately, the purpose of using these
fixation devices is to restore functionality to the bone and limb.
1.2. Problem
To promote a successful fracture healing outcome, it is necessary to correctly
configure the fracture fixation device according to the physical condition of the
trauma. Some of the configuration parameters that should be decided upon are the
Optimisation of an Internal Fixation Device
2
type of plate, including the length; where to position the plate; how many and
where to position the screws. Each parameter contributes to the progress and
outcome of the healing fracture. If the fixation device is improperly configured, it
can hinder the fracture healing process, resulting in revision surgery. This
increases the burden of the healthcare system and decreases quality of life for the
patient involved.
In internal plate fixation, the screw configuration is one of the vital parameters
decided upon by the orthopaedic surgeon. If the surgeon uses too many screws, the
plate may prematurely fail during treatment of the fracture, in which case, revision
surgery may be required. Furthermore, there may not be sufficient motion at the
fracture gap required for healing. At the other extreme, in the case of using too few
screws, the stress in the plate is decreased at the expense of an increased amount
of motion of the bone fragments. Excess movement causes further complications,
such as a delayed or non-union of the bone fragments. Therefore, the goal is to find
the best screw configurations to be used following the requirement that the
fracture successfully heals, while the implant does not fail.
Previous studies (Tornkvist et al, 1996; Stoffel et al, 2003; Duda et al, 2002) have
used mainly experimental techniques and some finite element analyses to evaluate
the strength and stiffness of certain screw configurations, and to identify trends in
screw placement. The approach taken in this study is to optimise the screw
configuration of the fixation device using mathematical programming, with the
added advantage of simultaneously creating optimum conditions for healing.
Mathematical optimisation techniques have been used successfully for numerous
applications in various fields of engineering. However, they have not been applied
to the topic of fracture healing in the biomedical field. In this study, mathematical
programming is utilised to ultimately optimise the screw configuration with
respect to bone fragment movement constraints in certain directions, ensuring that
the stress in the plate is minimised.
Optimisation of an Internal Fixation Device
3
1.3. Aims
There are two main aims of this study.
1. Develop an optimisation tool.
This involves creating an interface between various softwares used to
develop the optimisation process. Finite element (FE) software is used to do
numerical analysis on a computational fracture model while a Python
program is used to extract information from the FE output database. The
mathematical optimisation algorithm itself is written in Fortran
programming language. It was used to create the software interface.
2. Investigate the potential for the optimisation tool to solve the clinical
problem.
Apply the developed optimisation process to various cases to determine the
optimal screw configuration that enhances bone healing and avoids
mechanical failure of the plate in internal fixation for a particular fracture.
1.4. Significance of the Study
By defining the requirements for timely fracture healing, fracture fixation devices
may be configured in a manner in which they support healing conditions.
In the field of fracture healing, researchers (Epari et al, 2007; Goodship and
Kenwright, 1985) have strived to define the precise conditions required for timely
fracture healing. Goodship and Kenwright applied rigid fixation to one group of
Optimisation of an Internal Fixation Device
4
fractures and controlled axial movement in another group in vivo, in an attempt to
determine the optimal parameters for fracture healing. The results showed that
controlled micro-movement significantly improved healing. Epari et al looked at
the association between strength of healed bones to the stiffness of their respective
fracture fixation configurations. It was found that optimising axial stability and
limiting shear movements was required for timely healing.
In a recent paper, Chen et al (2010) biomechanically analysed two cases presented
to them from the clinical environment from different orthopaedic surgeons. The
internal fixation device that was configured in a rigid manner failed due to a fatigue
fracture and did not heal. The other case which was configured in a more flexible
manner did heal. Although there has been progress in this research field over many
decades, there is still the knowledge gap of selecting the best screw configuration
for a fracture fixation device in a given situation. This project aims to further
research in this area using optimisation mathematical programming.
1.5. Outline of Thesis
Chapter 2 will discuss fixation stability regarding internal fixation devices and
fracture healing. Fracture fixator parameters such as screw positioning and
numbers, and their influence on the strength of the fixator and the stress in the
plate, as well as the influence of the size of the fracture gap will be examined.
Chapter 3 will provide a detailed explanation of the how the mathematical
programming method is interfaced with results from the FE calculations for
optimisation of the screw configuration.
Chapter 4 is the results section which addresses two computational model cases to
which the optimisation method was applied. One is a simplified cylindrical case,
while the other is a femoral ‘clinical’ case.
Optimisation of an Internal Fixation Device
5
Chapter 5 holds a discussion of various aspects of the optimisation method used
and improvements are suggested.
It should be noted that this thesis focuses on the method of optimisation used
rather than the final application.
Optimisation of an Internal Fixation Device
6
2. Literature Review and Background
Severe trauma to the extremities is the leading cause of disability during the wage
earning period of life (BJD, 1998). Over 150,000 Australians are hospitalised with
fractures each year (Welfare, 2006). The socio-economic burden of fractures is
substantial. Loss of working capacity represents over 60% of the total cost of bone
fractures, while less than 20% is due to the direct cost of medical treatment.
Optimal outcomes, therefore, require not only solid bone union but also early and
complete recovery of limb function.
This chapter describes the process of fracture healing and the mechanical
conditions necessary for healing. Fixation stability is vital for fracture healing and is
characterised by the mechanical configuration of the fracture fixator being used.
Although there are numerous mechanical parameters involved in fracture fixation,
this review will focus on one of the mechanical aspects, i.e., screw configurations
and its importance in fracture healing.
2.1. Treatment of Long Bone Fractures
A fracture occurs when a high amount of energy is absorbed by the bone until
failure occurs (Brighton, 1984). For these types of fractures, surgery is often
necessary. There are three main types of fracture fixation treatments involving
surgery. As previously mentioned, they are external fixation, intramedullary nailing
and internal fixation. All fixation devices are designed for the restoration of limb
function, anatomical reduction by stabilisation of bone fragments and promotion of
bone healing.
Internal fixators, i.e. plates and screws, are common in the treatment of shaft
fractures up to the metaphyseal area (Ruedi et al., 2001). A failure rate of 7% is
reported with plate failure, screw loosening or breakage being the causes of failure
(Riemer et al., 1992). In the incidence of failure, due to a large range of possible
Optimisation of an Internal Fixation Device
7
complications, revision surgery is often necessary which decreases the quality of
life of the patient, and increases costs to the healthcare system.
2.1.1. Internal Fixators
Locking plates are internal fracture fixation devices that have been designed to
allow maximal vascularisation to the damaged bones and achieve a minimal
implant-bone interface. Two methods of treatment are available using the Locking
Compression Plate (LCP). This is made possible by the screw combination holes in
which part of the hole allows the fitting of a locked screw, whereas the other part of
the hole allows screws to be positioned at different angles (Figure 1).
Figure 1 LCP-combination hole allowing conventional plate fixation as well as application of locked screws (Source: (Perren, 2002)).
In the compression treatment method, as in conventional plating, anatomic
reconstruction and absolute stability may be achieved. The other treatment method
is called locked splinting, in which the LCP is used to simply bridge the fracture gap,
leaving the defected zone untouched. This method is ideal for the fixation of
comminuted, diaphyseal and metaphyseal fractures (Wagner et al., 2007) (Figure
2). The LCP allows the combination of the compression method and the locked
splinting method.
Optimisation of an Internal Fixation Device
8
Figure 2 Internal fixator used with locked screws. Fixator barely touches the bone as screws allow reliable maintenance of the initial distance between internal fixator and bone (Source: (Perren, 2002)).
With a variety of plates and types of screws available for security of the fracture,
and the large number of different configurations possible, the orthopaedic surgeon,
based on his experience, has to decide upon many mechanical factors regarding
configuration, whilst taking into consideration the biological conditions of the
fracture. The determinants of the fixation method are: which type of plate,
including the length; where to position the plate; how many and where to position
the screws (Wagner et al., 2007).
Chen et al (2010) has undertaken a FE study of comparing the influence of different
numbers of screws on plate failure. The configurations of the screws that were
compared were those that had been used in a clinical case. One was a flexible
fixture (6 screws out of a possible 14) in which successful healing occurred. The
other was a rigid fixture (12 screws out of a possible 14) in which plate failure
occurred without healing of the fracture. In the FE study, it was found that under
physiological loading, the plate that was rigidly fixed experienced significantly
higher stresses than the one fixed in a more flexible manner. In the fatigue analysis
it was found that the plate under rigid fixation fractured at 20 days after surgery,
whilst the plate under flexible fixation was able to endure 2000 days. This study
Optimisation of an Internal Fixation Device
9
has highlighted the major impact of screw configuration for fixation stability for
fracture healing as well as its influence on plate failure.
It is important to describe the fracture healing process to better understand the
implications of mechanical stimulus due to fracture fixation.
2.1.2. Fracture Healing
Naturally, the body has two ways of healing bone fractures. One is primary healing,
which involves direct compression across the bone fragments. In this case there is
no displacement of fragments and there is absolute stability of fixation. Osteones
(functional unit of compact/cortical bone) are able to grow across the bone
fragments. The disadvantage of this process is that the fracture takes an extended
period of time to heal compared to the secondary healing.
Figure 3 Direct healing from osteotomy of sheep tibia with compression stabilisation. The bone fragments are close and compressed and there is no displacement at the site of the osteotomy. The shape of the osteones do not change when crossing the fracture. (Source: Perren, 2002)
The other type of healing is secondary healing, which involves the formation of a
callus around the fracture site. This usually occurs when there is high impact
trauma to the bone, and there is extensive soft tissue damage. Healing of a fracture
Optimisation of an Internal Fixation Device
10
of this calibre involves a number of stages and may take weeks until the formation
of bone is observed.
In such an open fracture, the local bone marrow, periosteum, adjacent soft tissue
and blood vessels are injured. The first course of action of the body is to clot the
vessels around the fracture site, and prevent or fight infection in the area of
trauma. Haematoma and haemorrhage formation results from disruption of
periosteal and endosteal blood vessels at the fracture site (Figure 4, one day after
fracture). Pain and swelling eventually decreases and primary soft callus then
forms (Figure 4, 7 days after fracture). At day 14 the soft callus becomes
mineralised to form new bone. Three weeks post-fracture, the bone fragments are
no longer moving. The stability at this stage is adequate to prevent shortening,
although angulation of the fracture site may still occur. The cells that are stimulated
and sensitised produce new blood vessels, fibroblasts and supporting cells.
Chondroblasts also appear in the callus between bone fragments. Following the
linkage between the bone fragments by the callus, the stage of hard callus begins
until they are firmly united by new bone (Figure 4, days 21 and 28). Bony bridging
of the callus usually occurs at the periphery of the periosteal callus and endosteal
bone preceding the remodelling phase, which continues for several years (Ruedi et
al., 2001).
The previous explanation is the ideal fracture process by secondary healing.
Because the fracture zone is sensitive to mechanical stimulus (Kenwright et al.,
1989) and the tissues differentiate accordingly, it is important to achieve or come
close to achieving adequate mechanical conditions for the stimulation of healing. To
one extreme, there may be too much movement, and the fracture is unstable. In this
case, bone healing will be delayed or will not occur. To the other extreme, there
may be insufficient movement to stimulate any healing. This condition is similar to
primary healing.
Optimisation of an Internal Fixation Device
11
It should be noted that in addition to mechanical stimulation, biological factors
such as hormones, growth factors and blood supply are required for healing.
However, this study will address some of the mechanical influences rather than the
biological aspects.
Figure 4 Histological images of secondary fracture healing in bone (Source: J. Bone Miner. Res., 16, 1004– 1014, 2001).
2.2. Factors Influencing the Strength of the Fixation
Construct and Bone Healing
Bone healing is known to be sensitive to mechanical stability of fixation (Yamagishi
et al., 1955). The strength and stiffness of the fracture callus is related to the degree
of stability of the fixation device (Goodship et al., 1985; Kenwright et al., 1989). The
maturation of the callus is related to the amount of motion between the fracture,
Optimisation of an Internal Fixation Device
12
which depends on the applied loads and fixation stability (Claes et al., 1998; Duda
et al., 2002).
2.2.1. Stiffness of Fracture Fixation
Knowing the amount of stiffness required from a fixator to promote a successful
fracture healing outcome is vital. Epari et al (2007) have achieved this for a variety
of external fixators and intramedullary nails. The study measured firstly the
stiffness of the fixators in vitro, and secondly, the strength and stiffness of healed
tibiae after nine weeks that were treated using the various types of external
fixators and intramedullary nails. Using the experimental technique, a relationship
between the fixation stability and strength of the tibiae was found (Epari et al.,
2007).
A similar study conducted by Woo et al (1984) compared stiffness and strength of
healed femurs using flexible versus rigid internal fixator constructs. The purpose of
the study was to develop concepts for the ideal internal fixation plate, “based on the
mechanical demands of plate stiffness and strength in balance with the
physiological responses of the underlying bone” (Woo et al., 1984). It was found
that in the early stages of healing, plate stiffness in the bending and torsion must be
sufficient to promote union without bone angulation or implant failure. In the later
stages, plate stiffness should be low enough so that the bone may share the
physiological loads.
2.2.2. Physical Conditions for Fracture Healing
As quantitative measurements of the stiffness of internal fixators are unavailable in
the literature, an alternative method of defining the optimal conditions for healing
is required. As aforementioned, the mechanical conditions of the callus are related
to the movements between the fracture gap (interfragmentary movements).
Optimisation of an Internal Fixation Device
13
Goodship and Kenwright (1989) studied the effects of applying 0.5 mm, 1 mm and
2 mm of axial displacement in a 3 mm fracture gap. It was found in the tibiae with
0.5 mm displacement and 1 mm displacement (with 200 N applied force),
increased rates of fracture stiffness and mineralisation was seen. A displacement of
2 mm was detrimental to healing in terms of mineralisation and fracture stiffness.
In the clinical investigation conducted by Goodship and Kenwright, movements
between 0.2 mm and 1 mm were permitted. Movements between these limits
supported healing (Kenwright et al., 1989).
Augat et al (2003) investigated the effects of shear movement at the fracture gap. It
was seen that, in a 3 mm gap size, displacement of 1.5 mm in a shear direction was
detrimental to healing, while that of the same magnitude in the axial direction
supported healing. Shear movements may induce delayed unions and
pseudoarthroses. The type of tissue produced is cartilage and fibrous tissue at the
fracture site (Yamagishi et al., 1955; Augat et al., 2003).
In summary, it is seen that for a 3 mm fracture gap, certain amounts of
displacements in their respective directions are required to promote healing.
Therefore, what is required is a fixation structure that when under an applied load,
creates sufficient motion that promotes healing. As previously mentioned, there are
a number of mechanical determinants contributing to the strength and stiffness of
the internal fixator that may be controlled. This includes which type of plate,
including the length; where to position the plate; how many and where to position
the screws (Wagner et al, 2007). However, from this point, the literature review
will focus mainly on the topic of screw configurations, which is of the scope of the
present study. The arrangement of screws strongly impacts on the loading of the
implant itself, as well as the healing outcome of the fracture.
The distance between the screws and the number of screws in a plate has influence
on the axial, bending and torsional stiffness’ of the fixation construct. Furthermore,
these aspects have great impact on the stress distribution in the plate which is
Optimisation of an Internal Fixation Device
14
important to estimate in order to prevent early plate failure during clinical
treatment. Previous studies (Tornkvist et al, 1996; Duda et al, 2002; Stoffel et al,
2003) have been conducted to investigate the influence of screw arrangement on
the stresses and strains in the plate and screws, rather than their impact on
fracture healing.
2.3. Influence of Working Length and Fracture Gap on
Fixation Stability
In internal fixators, having a large working length (distance between the innermost
screws) greatly dissipates the stress along the length plate under applied loading.
By leaving a space of between 2 and 3 holes across the fracture gap, stress
concentrations may be avoided (Wagner et al., 2007). It was shown by Stoffel et al
(2003) that if a large working length is used (e.g., 10 hole spaces), for example, in
the case of bridging a comminuted fracture under dynamic loading tests, the
construct failed early. Using a large working length will also render the construct to
be too flexible, allowing excessive motion between bone fragments. This movement
will cause a non-union or a delayed union of the bone fragments (Kenwright et al.,
1989; Claes et al., 1998).
With sufficient fixation stability and blood supply, the fracture will heal
successfully. An example of this is from Chen et al (2010) which illustrates an X-ray
image (Figure 5) of a 35 year old male who suffered a femoral fracture treated with
a flexible construct. The surgeon used a moderate working length, with not more
than 3 screws on either side of the fragment.
By using too many screws, large stress concentrations are created in the plate
which lead to premature implant failure. In a study by Sommer et al (2003), this
phenomenon was demonstrated. A 73 year old woman suffered a periprosthetic
fracture in the middle to distal third part of her femoral shaft. The surgeons placed
screws immediately adjacent to the fracture site, inclusive of 12 out of a maximum
Optimisation of an Internal Fixation Device
15
14 holes. The screw combination resulted in high stresses generated in that section
of the plate which led to early failure (7 week post-op) (see Figure 6). Because of
the extreme rigidity of the structure, there was insufficient interfragmentary
movement to promote callus formation. Therefore no healing occurred.
Figure 5 Left: X-ray image of femoral fracture in 35 year old male with flexible fixation. Right: X-ray image 4 months after fixation, showing obvious signs of callus growth. Source: (Chen et al., 2010).
Working length (length between the innermost screws) has been identified as a
major influence on the distribution of stress in the plate, and stiffness and strength
of the bone-fixator construct. Stoffel et al (2003) included in their study an FE
comparison of the stresses experienced in an LCP plate due to the working length,
using gap sizes of 1 mm and 6 mm. It was shown that as the working length
Optimisation of an Internal Fixation Device
16
increased from the distance of 2 holes on the plate to 4 holes, for the 6 mm gap, the
Von Mises stress in the plate increased by 133 %. This was different for the 1 mm
gap model, in which it was demonstrated that the Von Mises stress in the plate
decreased by 10 % (Stoffel et al, 2003).
Figure 6 Seven week postoperative x-ray showing fracture fixation by placing several locking screws in main fragments. The screw holes were occupied adjacent to the fracture site resulting in high stress concentrations occurring in that section of the plate. Source: (Sommer et al., 2003)
In a similar study, Duda et al (2002) used the Less Invasive Stabilisation System
(LISS) plate to secure a ‘worst’ defect of 11 mm representing a comminuted
Optimisation of an Internal Fixation Device
17
fracture. By doubling the working length, i.e. from 2 to 4 hole spaces across the
defect, there was a considerable reduction in the Von Mises stress of the internal
fixator (Duda et al., 2002). Thus, there is a direct contrast in the stresses generated
in the internal fixator due to an increase in working length, i.e. the results given by
Stoffel et al for the 6 mm gap and that of Duda et al for the 11 mm gap. However, for
both cases, the construct became less stiff in compression and bending and the
stresses in the implant were reduced.
In another study investigating the impact of fracture gap sizes, Ellis et al (2001)
used a Dynamic Compression Plate (DCP) to stabilise a no-gap model, 10 mm gap
model and a 40 mm gap model. Plate strain was calculated. For the 10 mm model
and the 40 mm model, placing the screws closest to the fracture site decreased the
strain in the plate. In the no-gap model, placing the screws farthest from the
fracture site minimised the strain in the plate (Ellis et al., 2001).
Claes et al (1998) studied the influence of fracture gap and interfragmentary
strains on biological healing of the fracture gap. Different interfragmentary strains
were applied to the various in-vivo fracture gap size models of 2 mm and 6 mm. As
mentioned previously, it was found that although a large callus formed in the small
gap model due to large interfragmentary strain (31 %), the tissue that was formed
was connective tissue rather than bone. When the 2 mm gap model was subjected
to a smaller strain (7 %), bony bridging occurred which resulted in successful
healing. For larger gap models (6 mm) regardless of the interfragmentary strain,
the tissue type that was found to be produced at the end of the 9 weeks in-vivo
study was connective tissue (Claes et al., 1998).
Optimisation of an Internal Fixation Device
18
2.4. Screw Positioning
Screw positioning is important in determining the loading of the implant itself
(Duda et al., 2002). At least 3 screws should be placed either side of the fracture,
regardless of the quality of the bone (Wagner et al., 2007). More than 3 screws
either side of the fracture site does not increase the axial stiffness of the construct.
In the LCP, by placing additional screws towards the plate ends, the axial stiffness
decreased (Stoffel et al., 2003). This is in contrast to conventional plating where in
the stiffness would increase. Under torsional load, more than 4 screws per
fragment did not have an influence on the rigidity of the construct (Stoffel et al.,
2003).
In a study of compression plate fixation by Cheal et al (1983), in a 3 dimensional FE
model, it was found that in the presence of a fracture gap, the loads on the
innermost screws are increased and are more inclined to static failure during the
early stages of weight bearing. It was also found that the outermost screws are
more vulnerable to fail due to fatigue if the plate is left for a long period.
A study by Field et al (1999) concerned the influence of screw omission on bone
strain. It was found that “certain omission treatments provoked higher levels of
bone strain than would have been obtained if the plate were attached using all
screws” (Field et al., 1999). In an earlier study by Korvick et al (1988) it was shown
that the removal of the inner 2 to 4 screws from a screw-filled 8-hole plate resulted
in significantly higher levels of bone strain. Additionally, it was shown that by
replacing bi-cortical screws (screws that pierce both cortices) by mono-cortical
screws, the strain experienced by the bone was significantly reduced.
Shortening the plate by removing the end screws did not have any major effect on
the rigidity of the construct (Korvick et al., 1988). This is in contrast to the study by
Optimisation of an Internal Fixation Device
19
Sanders et al (2002) who found that the length of the plate was more important
than the position of the screws in providing bending strength (Sanders et al., 2002).
Tornkvist et al (1996) used a dynamic compression plate (DCP) to investigate the
relationship between screw positions and number, and the strengths of the
constructs. It was found that under torsion, strength in the plate was dependent on
the number of screws. It was also found that under bending, as in conventional
plates, strength in the plate was improved by the wider spacing of screws rather
than the increase in the number of screws (Tornkvist et al., 1996).
2.5. Limitations of previous studies
Duda et al (2002) and Stoffel et al (2003) did not take into account the
interfragmentary movement at the fracture site which is important for healing. The
recommendations made by Stoffel et al were based on the maximum Von Mises
stress in the plate and screws disregarding the interfragmentary movements as
well as the stress and strain in the callus.
There is limited information in the literature on the influence of screw
configurations on the physiological responses of the bone, in terms of the stresses
and strains that occur at the callus site. Goodship and Kenwright (1985) did
investigate interfragmentary movements for a fracture gap of 3 mm. However, the
stresses and strains that occur in the callus were not measured.
Stoffel et al (2003) and Tornkvist et al (1996) both conducted experiments to test
screw configurations on the strength of the bone-plate-screw construct or the
stress in the plate and screws. These studies did not test the effect of screw
configurations on the callus stress and strains as it was not possible. By using the
finite element method, it is possible to create a callus material around the fracture
gap. This has been attempted (Claes et al, 1999). However, the problem lies in
defining and validating the callus material as this information is unavailable.
Optimisation of an Internal Fixation Device
20
2.6. Summary
Previous studies (Stoffel et al, 2003; Field et al, 1999; Cheal et al, 1983) show that
the concept of working length cannot be generalised to all types of plates.
It can be observed that the stress in the implant under applied loads is not simply
influenced by the working length and the number of screws. The distribution of
stress in the plate is influenced by other mechanical aspects that have not been
highlighted and specifically addressed in the literature. The size of the fracture gap
plays an important role, as plate stress distribution varies with it. In addition, the
design of the plate influences the stress distribution. Further research needs to be
conducted in these areas. However, this study will not address these issues as it is
not in the scope of the project. A LCP plate will be used with a fracture gap size of 3
mm as there is more information in the literature about these parameters.
For a particular type of fracture (comminuted, oblique, spiral, etc), it is necessary to
find out what configuration of screws (i.e. number and placement) is required to
reduce the stresses in an internal fixator as well as to promote healing of the
fracture. Although some work (experimental and computational) has been done in
this area the most suitable screw configuration for a type of fracture is unknown.
This project attempts to approach the problem using mathematical programming
techniques.
Optimisation of an Internal Fixation Device
21
3. Methods - Optimisation
In broad sense of the term, optimisation is the efficient allocation of limited
resources. The aim is to arrive at the best possible decision in any given set of
circumstances.
3.1. Mathematical Definition of Optimisation
In mathematics, the field of optimisation is dedicated to finding the minimum or
maximum of a function of n real variables, subject to one or more
constraints. Ultimately, the aim is to minimise the effort required or maximise the
benefit desired in a situation, which is often described by a function (Rao, 1996).
Mathematically, the optimisation problem may be stated as follows:
Find
which minimises
Subject to the constraints: , j = 1,2,...,m and
, j = 1,2,...,p
where is known as the objective function, which is the design parameter of
the problem that is wished to be minimised or maximised, with respect to other
design parameters. The constraints, and are inequalities and equalities
respectively. The problem described above is a constrained optimisation problem. A
problem without the constraints is known as an unconstrained optimisation
problem.
Optimisation of an Internal Fixation Device
22
In its simplest form, the objective function will have one variable. This is called a
one-dimensional problem, for which there are a number of mathematical methods
available to solve. Brent’s Method and Golden Section Search are some examples
(Walsh, 1975). As more variables are added to the function, the problem becomes a
multi-dimensional case which is solved using more complex mathematical
procedures, which are described in the following sections.
To illustrate the complexity of the multi-dimensional optimisation problem, refer to
Figure 7. In this case, there are two variables, and . The ellipses represent the
contours of the objective function. The feasibility region, which is bounded by the
constraint functions, is presented. With the addition of more variables, the
objective function surfaces become harder to visualise and have to be solved purely
mathematically (Rao, 1996).
3.2. Types of Optimisation Problems and How to Solve
Them
Depending on the information available about the optimisation problem, there are
a variety of methods available to produce a solution. There are direct, indirect and
gradient methods which make use of different fundamental principles to ultimately
obtain an optimum. These methods are used to solve multi-dimensional problems.
3.2.1. Constrained/ Unconstrained Optimisation Problems
Optimisation methods to solve unconstrained optimisation problems fall in two
categories. One is direct search methods, in which derivatives of the objective
function are not required. The other is descent (gradient methods) methods that
require the derivatives of the function.
Optimisation of an Internal Fixation Device
23
Figure 7 Example of contours of an objective function (Source: Rao, S. S.; Engineering Optimization-Theory and Practice, 3rd Ed. 1996, pp.363)
Constrained minimisation problems may be solved using direct search methods
and indirect methods. A constrained problem becomes replaced by a series of
unconstrained minimisation problems in which penalty functions are used. The
penalty terms represent a measure of violation of the constraint. This is the indirect
search method.
Constraint functions
Optimisation of an Internal Fixation Device
24
The main principles of direct search methods are as follows. An initial guess point
must be selected as to where the location of the minimum (optimal value) is .
This point is checked to determine if it is the optimum. The next step is to generate
a new point .
Direct search methods are unique in the way that they select the new point, as well
as the way they subsequently test the point for optimality. Some examples of direct
search methods are Grid Search Methods, Pattern Directions, Hooke and Jeeves’
Method, Powell’s Method and Simplex Method.
Indirect search (descent) methods are those that utilise the gradient of the
function. Moving in the gradient direction from any point in space will increase or
decrease the function value at the fastest rate. Unfortunately this gradient direction
applies on a local level rather than a global one. Local versus global minima will be
further explained in Section 3.6. All descent methods make use of the gradient
direction to facilitate selection of search directions. Examples of optimisation
methods that use these principles are Steepest Descent (Cauchy) Method, Newton’s
Method, Quasi-Newton Methods and the Davidson-Fletcher-Powell Method.
The problem presented in this study is a constrained optimisation problem.
Although there are a number of optimisation methods available to solve it, there
are certain attributes of one algorithm over another that make it desirable to use.
The following section will discuss the attributes of different types of optimisation
methods used to solve constrained optimisation problems.
3.2.2. Multi-modal Optimisation
Usually the functions dealt with are multi-modal functions (multi-dimensional
problems), which are simply functions with a number of optimums. It may be
assumed that the function in this study is one with multiple optimums, of which the
location of the peak optimum is unknown. An example of a function with 20
optimums is shown in Figure 8. However, one of the prevalent problems with
Optimisation of an Internal Fixation Device
25
optimisation algorithms is that they tend to look for a local optimum rather than
the global one. This means that the algorithm generally tends to find the closest
optimum from its initial point. Hence, there is no guarantee that the optimum
solution found is necessarily the best one.
Figure 8 A multi-modal function. Source: (Singh et al., 2006)
There are algorithms, generally called multi-modal algorithms that have been
created to overcome this problem.
An advantage of using multi-modal algorithms is that they are able to search a
population of points in parallel, rather than just a single point. Any starting point is
permitted as it would not make a significant difference to the number of iterations
necessary to find solutions. The algorithm can provide a number of potential
solutions, as opposed to a single one.
Evolutionary algorithms are examples of multi-modal algorithms that require a
probability distribution function to govern the generation of a new search point.
Unfortunately the present study does not have a probability distribution function,
which is a requirement of this method.
Fitness
Optimisation of an Internal Fixation Device
26
Heuristics are effectively search procedures that move from one solution point to
another with the object of improving the value of the model criterion. They can be
used to develop good (approximate) solutions. This type of algorithm uses the rule
that given a current solution to the model, allow the search of an improved solution
(Taha, 1976).
Simulated Annealing (SA) and Genetic Algorithms (GA) are examples of heuristic
probabilistic methods which are multi-modal algorithms (Singh et al, 2006). The
disadvantage of using these methods is that they are impractical for the
optimisation of structures using the finite element method, which is used in this
study. These methods require a large number of iterations before they would be
able to converge.
3.2.3. Deterministic Methods
Deterministic heuristic methods such as the Simplex method and Powell’s method
are gradient-based mathematical programming methods. These methods have
been used in a number of engineering applications to find the optimal solution for
continuous variables.
They have been known to excel when the gradient of the objective function is
unavailable (Nelder et al., 1965; Del Valle et al., 1988). The Simplex method and
modified Simplex methods have been used in analytical chemistry optimisation
problems. It was observed that using these Simplex methods sometimes there was
lack of convergence, and therefore inefficient. Powell’s method was found to be
more efficient in that it converged quicker than compared to the Simplex method
(Del Valle et al., 1988).
In a study comparing the efficiency of Powell’s method and the Simplex method on
the application of flow injection systems, it was found that Powell’s method
reached optimal conditions with a lower number of experimental evaluations (Del
Valle et al., 1988).
Optimisation of an Internal Fixation Device
27
The optimisation method that is used in this study is Powell’s method. This
algorithm has its advantages and disadvantages. The advantages are that it is a
widely used and tested algorithm which has been used extensively in engineering
and one of the most efficient of those not based on the estimation of the gradient of
the objective function (Del Valle et al., 1988). The disadvantage is that Powell’s
method searches for a local solution rather than a global one. However, the global
optimisation techniques that are available are not well tested and used, under-
developed and inefficient. To reduce the effects of the ‘global issue’ an educated
estimation of the starting point in the search space assists the algorithm in seeking
the optimum.
A description of Powell’s method is provided in the following section.
3.3. Powell’s method
Powell’s method makes use of the properties of conjugate directions. This is
advantageous as convergence is accelerated by minimising along each of a
conjugate set of directions.
3.3.1. Conjugate Directions
Mathematically, conjugate directions may be described as follows. Suppose a
system of linear equations,
Where A is a symmetrical positive definite n-by-n matrix (i.e. , Ax for
all non-zero vectors in and real). Two non-zero vectors u and v are conjugate
(with respect to A) if
Figure 9 is used to illustrate conjugate directions. If X1 and X2 are the minima of the
function, Q obtained by searching along the direction S from 2 different starting
Optimisation of an Internal Fixation Device
28
points Xa and Xb, respectively, the line (X1 - X2) will be conjugate to the search
direction S (Rao, 1996).
Figure 9 Conjugate Direction (Source: Rao, S. S.; Engineering Optimization-Theory and Practice, 3rd Ed. 1996, pp.363)
3.3.2. The Algorithm
Powell discovered a direction set method that produces n mutually conjugate
directions (Walsh, 1975; Press, 1992; Mathews, 2004).
Let be the set of values of variables as the initial guess of the location of the
minimum of the function,
Optimisation of an Internal Fixation Device
29
1. Approximate the minimum of the function to generate the next estimation,
, by proceeding successively to a minimum of f along each of the N
standard base vectors. The process generates a sequence of points,
.
2. Along each standard base vector the function f is a function of one variable.
To minimise each function f requires the application of a one dimensional
minimisation method, such as the Golden Ratio Search.
3. The vector PN – P0 represents the “average” direction moved during each
iteration. It is the average direction moved after trying all N possibilities.
The point X1 is determined to be the point at which the minimum of the
function f occurs along this vector and requires minimisation using, for
instance, the Golden Ratio Search.
4. Since PN – P0 is regarded as a good direction; it replaces one of the direction
vectors in the next iteration. The iteration is then repeated using the new set
of direction vectors to generate a sequence of points.
The algorithm for Powell’s method can be summarized in the following (Press,
1992).
Let be an initial guess at the minimum of the function .
Let be the standard base vectors,
, and let
1. Set , where is the initial guessed point.
2. For find the value of that minimises and set
3. Set .
4. Set for Set .
5. Find the value of that minimises Set
6. Repeat steps 1 to 5.
Optimisation of an Internal Fixation Device
30
A more illustrative explanation of Powell’s method may be explained with
reference to Figure 10.
Powell’s method begins with an initial point and independent search
directions, which are initially the co-ordinate directions.
A search for the minimum is conducted along each of the directions (uni-directional
search) in turn. Successively, on each search, the minimum point obtained in the
previous search is the used as the new departure point. From Figure 10, search is
conducted along the directions then , delivering point 3 as the start for the
next minimisation.
When all directional searches are complete, the total displacement is used as the
new search direction, beginning with a two-fold distance point. If this direction of
expansion is sufficient, it replaces the direction that gave the lower improvement.
Suppose that gave the largest decrease. It is replaced by the new direction and
the unidirectional minimum is found at point 5.
By repeating this procedure, all the initial directions are replaced and a good
estimation of the conjugate directions is obtained. The solution converges when the
difference between the points is sufficiently small.
As mentioned in the previous section, the Powell’s optimisation algorithm includes
a one-dimensional (uni-directional) search method. The search method is used to
find the minimum in each direction initialised by the algorithm. There are a
number of uni-directional search methods available. The one used in this study was
the Golden Section Search method. It is among one of the most efficient region
elimination methods to optimise functions of a single dimension.
Optimisation of an Internal Fixation Device
31
Figure 10 Progress of Powell's Method (Source: Rao, S. S.; Engineering Optimization-Theory and Practice, 3rd Ed. 1996, pp.363)
Optimisation of an Internal Fixation Device
32
3.3.3. Golden Section Search – Search in One Direction.
The golden search method is implemented in Powell’s method. It is used to search
for the minimum in one direction. Each direction is not part of a set of ‘planned’
directions. Rather, it is formed after the evaluation of the result of search in a
previous direction.
Suppose the positions a, b, c and x (Figure 11) are points that lie sequentially on the
x-axis and is a fraction of the way between and . Therefore,
The next trial point is an additional fraction beyond
The next bracketing segment will be either of length relative to the current
one, or else of length . To minimise the worst case possibility, choose to
make these equal.
Figure 11 Illustration of Golden Section Search
b
z
x
w
a c
1-w
b
Optimisation of an Internal Fixation Device
33
The new point is symmetric to point b in the original interval. is equal to
. This implies that the point lies in the larger of the two segments. The scale
similarity implies that x should be the same fraction of the way from b to c as was b
from a to c. In other words,
Therefore, the quadratic equation, , yields
This means that the optimal bracketing interval has its middle point a
fractional distance 0.38197 from one end, say, and 0.61803 from the other end,
say, c. These fractions are called golden-mean or golden section. This optimal
method of function minimisation, the analogue of the bisection method for finding
zeros, is thus called the golden section search.
In summary, the concept of golden search is as follows: given a bracketing triplet of
points, the next point to be tried is that which is a fraction 0.38197 into the larger
of the two intervals. This occurs until a point where the difference between the
current point and the next point is minimal or close to zero. Each new function
evaluation will bracket the minimum to an interval 0.61803 times the size of the
preceding interval (Press, 1996).
3.4. Use of optimisation methods in medical engineering
Mathematical optimisation programming techniques became widely developed in
the 1960s. Since then, optimisation techniques have been used for train scheduling,
optimising design parameters in structural engineering, design of aircraft for
minimum weight, design of wind turbines and pumps for maximum efficiency,
Optimisation of an Internal Fixation Device
34
optimal design of electrical networks, analysis of statistical data and experimental
results to obtain the most accurate representation of the physical phenomenon,
design of optimum pipeline works for process industries, selection of a site for an
industry, and many more applications. In the medical field, there are two main
focuses. One is operating theatre scheduling time, using mainly stochastic dynamic
programming models, and the other is treatment planning in various fields. For
example, the treatment plan for stereotactic radio-surgery using conjugate
gradients and simulated annealing methods.
Many research groups have studied cancer therapy theoretically, clinically and
mathematically. Esen et al (2006) applied an optimisation model called Weapon-
Target Assignment problem (WTA) of military operations research to optimise
cancer therapy. It used mixed-integer nonlinear goal programming models. It had
three objectives: maximise the weighted damage of the cancer cells, minimise the
total weighted side effects and minimise the total dose therapy costs. The model
created facilitates cancer therapists to act in a multi-objective frame. However, the
model created must be clinically validated (Esen et al., 2008). This example alone
demonstrates the power of mathematical programming and its implications.
Maratt et al (2008) investigated the feasibility of an integer programming model to
assist in pre-operative reduction and internal fixation of a distal humerus fracture.
The model aimed at maximising the number of bicortical screws placed while
avoiding screw collisions and favouring screws of grater length over multiple
fracture planes (Maratt et al., 2008).
Rozema et al (1992) used a linear programming technique and a muscle
architecture model to minimise the strains in plate-osteosynthesis devices for
internal fixation of mandibular fractures. The objective was to minimise the strain
in the mandibular bone plate by optimising the position of the plate based on a
number of factors (Rozema et al., 1992). The objective function to be minimised
was an energy function, in which the variables included 3-D displacements and
Optimisation of an Internal Fixation Device
35
rotations that occurred at given external forces and torques on the computational
model. In finding the minimum, the values of the variables became known and were
used as input for the next step, which was to minimise the maximum internal strain
by optimising the placement (co-ordinates) of the bone plates using a deterministic
method called the Simplex Method.
3.5. Optimisation of Screw Configuration in Internal
Fixators
This section explains the application of the optimisation programming tool to find
the best screw configuration in an internal fixation device. It will discuss the
optimisation criteria, the calculation of the function value and the interface
between the FE software and the algorithm itself.
3.5.1. Objectives and Constraints
In this study, the objective function is
Find
which minimises f(S) (maximum principal stress)
subject to the constraints
and
See definitions in the next section.
Optimisation of an Internal Fixation Device
36
3.5.2. Optimisation Criteria
The optimum solution:
1. allows displacement, a, of between mm and mm in the axial direction,
2. allows displacement, d, of between mm and mm in a shear direction, and
3. minimises the maximum principal stress in the plate (objective)
Previous researchers (Stoffel et al, 2003; Duda et al, 2002) have calculated the Von
Mises stress rather than the maximum principal stress. Von Mises stress is the
criterion used to assess yielding of materials whereas maximum principal stress is
used as a failure criterion. Physiologically, the plate is subjected to cyclic loading,
from which fatigue failure results. As failure of the implant is of interest, maximum
principal stress is measured (Chen et al., 2010; Shipley et al., 2002).
There is information in the literature about the required interfragmentary
movements for a certain type and gap of fracture (Kenwright et al., 1989; Claes et
al., 1999). This project uses a transverse cut, distracted 3 mm, as in the
experimental work by Goodship and Kenwright (1989), but it does not take into
account many other biological and mechanical factors used to simulate their model.
Consequently, the quantitative values for interfragmentary movement described in
the abovementioned studies cannot be compared to those of the model used in this
study. Therefore, the range values of and (displacement in the axial and
shear directions) which are important optimisation criteria for this project, are
estimated by comparing the interfragmentary movements from the stiffest and
most flexible models for each computational model to be optimised using the
program.
Optimisation of an Internal Fixation Device
37
3.5.3. Objective Function
The analytical function that is used in a classic optimisation problem is usually
formulated from the available information or data about the situation. The
equation, known as the objective function, is used to obtain a value that is
evaluated at each point in space. This is known as the function value, which is used
as feedback for the optimisation algorithm to evaluate for optimality. It is for
comparison with the previous function value to see if the difference is sufficiently
small in order to determine if a maximum or minimum value has been reached.
This is also known as convergence.
Unfortunately, there is no analyti