Upload
others
View
10
Download
0
Embed Size (px)
Citation preview
1
©2010 Waters Corporation
Paul St. JeanProduct Marketing ManagerGlobal Compliance ProgramsWaters CorporationMilford, MA 01757 USA
Analytical Instrument Qualification (AIQ)Analytical Instrument Qualification (AIQ)
IVT Method Validation ConferenceJuly 28-30, 2010Sir Francis Drake HotelSan Francisco, CA
©2010 Waters Corporation 2
Agenda
Regulatory Expectations for Qualification— Terms— Warning Letters
USP <1058> AIQ Guidance Chapter— How it came about
Developing a roadmap for the AIQ Process— DQ IQ OQ PQ and more
Instrument Qualification Case Study Application— Developing an understanding of what’s involved in qualification— Examining various approaches— Can we be more efficient?
Interactive Exercise— Discuss real challenges in your lab— Group Q and A
2
©2010 Waters Corporation 3
II
Regulatory Expectations for QualificationRegulatory Expectations for Qualification
©2010 Waters Corporation 4
Data Integrity Objective
Data that supports the design, development and production of government regulated products must be accurate and defensible.
3
©2010 Waters Corporation 5
Calibration Requirements
CFR211.68 cGMP Equipment— Automatic, Mechanical and Electronic Equipment.
— “…equipment so shall be routinely calibrated, inspected or checked according to a written program designed to assure proper performance. Written records of these calibration checks and inspections shall be maintained.”
©2010 Waters Corporation 6
Calibration Requirements
CFR 211.160 General Requirements— Laboratory controls shall include…
— “The calibration of instruments, apparatus gauges and recording devices at suitable intervals in accordance with an established written program containing specific directions, schedules and limits for accuracy and precision…”
4
©2010 Waters Corporation 7
Validation versus Qualification
VALIDATION:− Refers to the total life cycle of a product from development through
use and maintenance.
− Customers (Owners) are responsible for Validating Their Processes (personnel, equipment, methods, SOPs) to ensure compliance to CGMP/GLP regulations.
QUALIFICATION: (Inspection,functional testing and documentation review)
− Is a part of the validation process which verifies module and system functional performance prior to being placed on-line and thereafter according to a standard operating procedure.
− Calibration verification is included as part of qualification.
− Vendors can assist customers in meeting their validation requirements but cannot validate customers’ processes.
©2010 Waters Corporation 8
Validation versus Qualification
According to an FDA inspector…..
QUALIFICATION:Answers the question…..
Was the instrument built right?
VALIDATION:Answers the question…..
Was the right instrument built?
• Vendors are in the best position to verify proper instrument operation
• Only users can determine if a system is suitable for their analysis
5
©2010 Waters Corporation 9
Validation versus Qualification
According to an USP 1058 …..
Instruments are QUALIFIED
Processes are VALIDATATED
• It is easiest to qualify instruments using standardized tests
• Procedures and methods are validated for each specific analysis
©2010 Waters Corporation 10
FDA Guideline
Provide Documented Evidence of HPLC System…— Accuracy, Linearity and Precision
o from: • “Validation of Computerized Liquid Chromatographic
Systems” Furman, W.B., Layloff, T.P. & Tetzalff, R.F. August, 1992
o This may be old, but it is still valid and still gets occasional mention in journals
Regulations tell us what to do, but not how to do ito CFR 211.68o CFR 211.160
6
©2010 Waters Corporation 11
Qualification Time Line
Equipment used should be of Appropriate Design, Adequate Capacity and Properly Maintained.
Functional Testing/VerificationFunctional Testing/Verification
CalibrationCalibration
PQPQ
StructurallyStructurallyValidatedValidatedProductsProducts
IQIQ OQOQ PQPQ
InstallationInstallation OperationalOperational PerformancePerformance
QualificationQualification
BeforeBeforePurchasePurchase
Owner'sOwner'sSiteSite
DQDQ
Owner'sOwner'sSiteSite
Vendor'sVendor'sSiteSite
MaintenanceMaintenance
(Initial Calibration)
System Suitability During UseSystem Suitability During UseBefore UseBefore Use After UseAfter Use
©2010 Waters Corporation 12
Method Validation
As you know, Method Validation is another area that regulated users must be concerned with but we should not confuse method validation with system qualification
Methods should be validated on qualified systems…• Qualification shows that the module or system performs as the
vendor intended• Qualification is not intended as a complete re-validation of the
design but rather a way to provide documented evidence of proper system performance in terms of accuracy, linearity (accuracy across a range) and precision
• Qualification does not demonstrate suitability for intended use• Method validation shows suitability of a system to perform a
specific analysis
7
©2010 Waters Corporation 13
ICH
International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use (European Union, Japan and USA)
Purpose:— Harmonization of interpretation / application of technical guidelines
and requirements for product registration
— Reduce or obviate need for duplicate testing during the research and development of new medicines
Objective:— Economical use of human, animal and material resources
— Eliminate unnecessary development delays
Sections of interest related to this talk: — Web site is WWW.ICH.Org
— Q7 Good Manufacturing Practice for Active Pharmaceutical Ingredients
— Q9 Quality Risk Management
©2010 Waters Corporation 14
483 Warning Letters Issued
Personnel Issues (21 CFR 211.25)— Analysts not qualified
— Problems traced to inadequate training
— No training records
— No list of approved computer system users (21 CFR 211.68)
Analytical Method Issues (21 CFR 211.160)— Analytical methods not validated
8
©2010 Waters Corporation 15
483 Warning Letters Issued
Data Audit Trail and Software Issues— No computer system validation report (211.68)— Software not revision controlled (211.68)— No audit trails (21 CFR Part 11)— Lack of audit trails (electronic) proving retention of all
raw data files (211.194 & 21 CFR Part 11)
Equipment Issues— No equipment PM records (211.67)— Un-qualified computer systems (211.68)— Failure to calibrate…no written program (211.68)— Instruments not qualified/calibrated (211.160)— Calibration not conducted at suitable intervals (211.160)
©2010 Waters Corporation 16
IIII
USP <1058>USP <1058>
Analytical Instrument Qualification (AIQ)Analytical Instrument Qualification (AIQ)
Guidance DocumentGuidance Document
9
©2010 Waters Corporation 17
INTRODUCTION
©2010 Waters Corporation 18
Workshop on AIQ
AAPS Workshop co-sponsored with FIP and ISPE “Scientific Approach to Analytical Instrument Validation”,
o Held in Arlington, VA, March 3-5, 2003
Proceedings are published and entered in FDA Docket— Bansal SK, Layloff T, Bush ED, Hamilton M, Hankinson EA, Landy JS, Lowes S, Nasr MM, St. Jean PA,
Shah VP. Qualification of Analytical Instruments for Use in the Pharmaceutical Industry: A Scientific Approach.AAPS PharmSciTech. 2004; 5(1): article 22.
Proceedings form Basis of new USP Guidance —<1058> Analytical Instrument Qualification
o Pharmacopeial Forum Vol. 31(4) July-Aug 2005
10
©2010 Waters Corporation 19
Terminology
Ambiguous terms: Validation or Qualification
Agreed at the workshop that — Processes are validated— Instruments are qualified
AIQ (Analytical Instrument Qualification)— “AIQ Is documented evidence that an instrument performs suitably
for its intended purpose and that it is properly maintained and calibrated”
— Does not include people (training), processes performed on instruments (analytical methods)
— Helps to justify the continued use of instrument, but it alone does not ensure quality of the data
©2010 Waters Corporation 20
Components of Data Quality
Analytical Instrument QualificationAnalytical Instrument Qualification
Analytical Methods ValidationAnalytical Methods Validation
System Suitability TestsSystem Suitability Tests
Quality Control Checks
Quality Control Checks
11
©2010 Waters Corporation 21
Component of Data Quality: Analytical Method Validation
Documented evidence that an analytical method does what it purports to do and addresses the required attributes of the method
Use of a validated method should instill confidence that the method can generate data of acceptable quality
©2010 Waters Corporation 22
Component of Data Quality: System Suitability Tests
Equipment, electronics, software, analytical operations and samples to be analyzed constitute an integral system
SST verifies the system’s performance according to the analyst’s expectation and criteria set forth in the method
H/W, S/W and analysis checks to provide assurance of system integrity before, during and following analysis of unknown samples
System Precision
Separation Parameters
System suitability testing alone is not sufficient, Qualification is still required.
12
©2010 Waters Corporation 23
Component of Data Quality: Quality Control Checks
Most analyses are performed on calibrated or standardized instruments
Single or multi-point calibration performed
Additional quality control check samples used in some analyses (e.g. bioanalyses) provide an in-process assurance of the test’s suitable performance
©2010 Waters Corporation 24
Extent of SST or QC checks
Chemical analyses subject to GMP regulations, requiring tighter precision and accuracy may require more SST
Bioanalyses, largely subject to GLP regulations, requires low level and broad range analysis, is performed with more QC checks during sample analysis
13
©2010 Waters Corporation 25
Components of Data Quality
Analytical Instrument QualificationAnalytical Instrument Qualification
Analytical Methods ValidationAnalytical Methods Validation
System Suitability TestsSystem Suitability Tests
Quality Control Checks
Quality Control Checks
©2010 Waters Corporation 26
IIIIII
Developing a Roadmap for the AIQ ProcessDeveloping a Roadmap for the AIQ Process
14
©2010 Waters Corporation 27
PROCEDURES
©2010 Waters Corporation 28
Performing AIQEssential Parameters
Several options were discussed, but DQ/IQ/OQ/PQ model was preferred
Most widely understood parameters
Need to provide AIQ specific definitions and scope for these parameters
Recommended set of parameters for AIQ:
o Design Qualification (DQ)o Installation Qualification (IQ)o Operational Qualification (OQ)
o Performance Qualification (PQ)
o Change Control
15
©2010 Waters Corporation 29
Design Qualification
Needed when designing new instrument
Most suitably Performed by developer/manufacturer
Not necessary for user to repeat DQ
Users should ensure:— The instrument is fit for their intended use
— Manufacturer has adopted a quality system for development, manufacturing and testing
— Manufacturer has adequate support for installation, service and training
Procedure— Depends on the intended use, complexity of instrument, acceptability of
instrument in industry and prior experience
— Fulfilled by vendor documentation/audits, peer reviews and testing
©2010 Waters Corporation 30
Installation Qualification
IQ is a documented collection of activities needed to install the instrument in the environment where it will operate— System Description
— Delivery: Did we receive what was ordered?
— Facility/Environment/Utilities
— Network and Data Storage
— Assembly and Installation
— Installation Verification
IQ can be performed for new, pre-owned or existing unqualified instruments
16
©2010 Waters Corporation 31
Operational Qualification
OQ Parameters: Test operation of instrument as per specifications in the user’s environment— Test critical parameters to assure required performance
— Parameters based on manufacturer’s recommendation and on user’s intended use
— Secure data storage, backup and archive process
Performance of OQ— Can perform holistic or modular— Modular when applicable— Modular approach will allow mix and match of modules in a system — Use non-method specific testing— Repeat relevant OQ tests when instrument undergoes major repairs
or modifications
©2010 Waters Corporation 32
Performance Qualification
PQ tests are performed on a periodic basis to ensure that the instrument remains in qualified state after IQ/OQ— Run tests to check and verify satisfactory performance of the
instrument
— Preventive maintenance and repairs
— Maintain an SOP for operation/calibration/maintenance
Performance of PQ— Perform at specified intervals (Same tests to generate history of
instrument performance)— Can perform holistic (preferred) or modular— Specifications for tests can be broader than for OQ
17
©2010 Waters Corporation 33
Timing and Applicability
DQ IQ OQ PQ
Prior to purchaseof a new type ofinstrument
At installation ofeach instrument(new, old or existing unqualified)
After installation or major repair of each instrument
Periodically at specified intervals for each instrument
©2010 Waters Corporation 34
Qualification Activities
DQ IQ OQ PQAssurance of vendor’s DQ
System description Fixed parameters Preventive maintenance and repairs
Assurance of adequate support availability from manufacturer
Instrument Delivery SOPs – operation,calibration and maintenance
Instrument’s fitness for use in laboratory
Utilities/facility/environment
Network and data storage Secure data storage, backup and archive
Assembly and installation
Installation verification Instrument Function Tests
Performance Checks
18
©2010 Waters Corporation 35
Software Validation
Software for analytical work classified into the following categories
— Firmware (low-level software for integrated chips in instruments)o Considered a component of the instrument itself o Qualified at the user site simultaneously with qualification of
instrument
— Instrument control, data acquisition and processing softwareo Hardware and software are inextricably intertwined and both are
necessary for the operation of the instrument o Manufacturers perform the validation and provide users a summaryo User perform holistic qualification involving entire instrument and
software system through AIQ process
©2010 Waters Corporation 36
Software Validation (Contd.)
—Stand-alone software (e.g. LIMS)o An authoritative guide for validating software is available:“General Principles of Software Validation”, Final Guidance for
Industry and FDA Staff, Jan. 11, 2002o Software developer specifies the development model appropriate
for the softwareo Validation takes place in a series of activities planned and
executed through various stages of development cycleo User site testing is performed as an essential part of the
software development cycleo User testing, though essential, is only a part of the full validation
19
©2010 Waters Corporation 37
Change Control
Changes are inevitable as new features are added and corrections are madeFollow DQ/IQ/OQ/PQ classification process— DQ: Review the change; adopt only useful or necessary changes— IQ: Install the changes
— OQ/PQ:
o Revise tests and specifications as necessitated by the change
o Change SOP as necessary
o Perform changed OQ or PQ tests
©2010 Waters Corporation 38
AIQ Documentation
Static Documents: — Obtained during DQ, IQ and OQ phases
— For multiple instruments place common documents in one binder or section and instrument specific documents in separate binders or sections
Dynamic Documents:— Obtained during OQ and PQ phase during instrument
maintenance or performance check
— Provide a running record of instrument performance
20
©2010 Waters Corporation 39
Instrument Categories for AIQ
— A single set of principles to qualify the wide variety of instruments is scientifically inappropriate
— Three categories of instruments are suggested
— Users are most qualified to establish the level of qualification needed
— Exact category of an instrument should be determined by the user for their specific instrument or application
— Modern Laboratories typically have a suite of instruments from simple to complex automated
©2010 Waters Corporation 40
Instrument Categories for AIQ :Group A
Conformance to user requirements determined by visual observations – No independent qualification required
Example instruments:
o Magnetic Stirrers
o Nitrogen evaporators
o Ovens
o Vortex mixers
o Spatula
o Glass pipettes
o Mortar and pestle
21
©2010 Waters Corporation 41
Instrument Categories for AIQ :Group B
Conformance to user requirements is generally unambiguous and established according to instrument’s SOPs. Causes of failure are discernible by simple observations.
Example Instruments:
o Balances
o pH meters
o Variable pipettes
o Refrigerator/Freezers
o Thermometers
o Viscometers
o IR spectrometers
o Titrators
o Melting point apparatus
©2010 Waters Corporation 42
Instrument Categories for AIQ :Group C
Conformance is complex, highly method specific and the conformity bounds are determined by their application. A full qualification process applies.
Example Instruments:
o HPLC
o Mass spectrometers
o GC
o Diode-array detectors
o Atomic absorption spectrometers
o Electron Microscope
o Micro-plate readers
o Thermal gravimetric analyzers
o X-ray fluorescence spectrometers
o Elemental analyzers
22
©2010 Waters Corporation 43
Roles and Responsibilities
©2010 Waters Corporation 44
Users
Users include the analysts and management— Are most familiar with their analysis
— Ultimately responsible for AIQ
— Can contract out parts of AIQ, but are still responsible
— Must keep instrument qualified through PQ
23
©2010 Waters Corporation 45
Quality Assurance (QA)
—QA role remains as in any other regulated study
—Ensure that AIQ meets regulatory requirements
—Have users attest to the scientific validity of AIQ
©2010 Waters Corporation 46
Manufacturer
— Perform DQ when designing instrument
— Validate the software and hardware
— Provide summary of their validation work
— Provide results of final test on hardware
— Should provide critical functional test scripts for qualification of instrument/software at the user site
— Make available any software/hardware bugs found after release
— Support installation, training and service
— Be open to audit by users
24
©2010 Waters Corporation 47
AIQ Conclusions
Use the term “Qualification” instead of “validation” for instruments
AIQ provides a “level of confidence” to users that the instrument is suitable for its intended use
AIQ is only one component in delivering reliable and quality data
Use DQ/IQ/OQ/PQ model with specific definitions and scope for each qualification stage
White paper is available on FDA docket, and AAPS Pharm SciTech (AAPS website).
AAPS PharmSciTech. 2004; 5(1): article 22.
©2010 Waters Corporation 48
Affect on Your Laboratory
25
©2010 Waters Corporation 49
What does AIQ mean to You?
AIQ is formalizing a long used process….but:
It is an opportunity to revisit what you are doing— Evaluate using a Scientific Approach
— What value added are you getting from your tests?
— Is there a way to simplify what is being done?
Are you doing enough?— Are you re-qualifying on a regular scheduled basis?
— Are you covering the high risk areas?
Are you doing too much?— Does it take you two years to implement a new Data System?
— Do you take measurements on hardware merely because you can?
— Are you re-qualifying more frequently than needed?
— Are you testing things that need not be tested as part of AIQ?
Risk analysis is a good way to decide
©2010 Waters Corporation 50Questions?
26
©2010 Waters Corporation 51
IVIV
Qualification Case StudiesQualification Case Studies
©2010 Waters Corporation 52
Agenda for Module IV
Determining what is important to test
Some questions asked by equipment users
How various instruments work and the effect on testing
Various approaches in qualifying LC modules and systems
Automating and streamlining the qualification process
27
©2010 Waters Corporation 53
Considerations
As we proceed lets — Consider the importance of the parameters we test
o What value added does the test supply
— Evaluate to what level they need to be tested
o Time to test versus benefit or risk mitigation
— Consider the various approaches we might use to test these
— Gain an understanding how each test and each module under test functions
— Remember that there is no one right way to qualify that an instrument or a system is functioning as intended
©2010 Waters Corporation 54
Considerations (2)
Remember, in the end, the user must be prepared to answer any questions an auditor may have
Understanding how the equipment functions and how it has been qualified leaves you better prepared to support the process in an audit situation
Qualification testing indicates that a module or system performs as intended
Qualification testing does not indicate suitability for a particular application
28
©2010 Waters Corporation 55
What about Part 11 Requirements?
Electronic Records and Signatures
Recent Changes
Risk Based Approach
Predicate Rules Apply— Laboratory Data
— Equivalency to paper based Systems
Can qualification results be made part 11 compliant?— If the results are treated as chromatography data, they can
©2010 Waters Corporation 56
Determining what is important to test (continued)
On LC, we are generally concerned with — Accuracy
— Linearity
— Precision
There are multiple ways to get this information— There is no one right way
— The FDA does not dictate the method to use
29
©2010 Waters Corporation 57
Manual SOPs for Qualification
Traditional/Manual SOPs (Qualification Workbooks) are designed to test key product specificationsGenerally, on Liquid Chromatographs, we verify Accuracy, Linearity and Precision where possible (per FDA guideline)
Workbook tests are derived from factory tests for individual modules adapted for use in the fieldThe workbook tests also allow for typical testing error— Errors in class a volumetric flasks— Errors in tests performed by NIST on the test solutions— Errors in meters and temperature probes— Test errors are additive with the product error
Workbooks provide a “high degree of assurance” that product specifications are metThey test the module as a module…not as it will be used in the system so system error is not be taken into accountThe main disadvantages of this form of testing are the — time required— systems not tested similar to the way they will be used
©2010 Waters Corporation 58
Exploring Ways to Qualify LC Systems
• A new streamlined approach is needed
• A scientific look is taken at the modular tests• Redesigning existing modular tests can simplify the process
• Using the power of the Chromatography Data Software (CDS), the system itself can be used to verify calibration and provide qualification data• Calculations can be performed internally avoiding the need to
deal with external spreadsheets• Automatic reporting and calculations reduces the risk of
human error and interference• Results can be kept on-line and be part 11 compliant
• The concept of Advanced Qualification Technology (AQT) and the Waters Systems Qualification Tool (SystemsQT) is created
30
©2010 Waters Corporation 59
Design Criteria
A design requirement of the SystemsQT was that it could perform all the tests normally performed using a traditional qualification workbook
The system was to be tested the way it will be used
Workbooks were initially developed using tests off the Waters manufacturing floor
SystemsQT was developed with a clean sheet of paper
All procedures were reviewed and new methodologies developed that combined tests wherever possible
New test solutions were incorporated with the aim of minimizing the various chemical kits used
©2010 Waters Corporation 60
Development
Original factory tests and specifications were developed to test a specific module and only that module
Interference from other modules, solvents, columns and other factors were minimized or eliminated when testing at the modular level
The SystemsQT is designed to test the chromatograph as a system the way it will be used, not as individual modules
New pass/fail criteria had to be developed based on all the factors that could affect the outcome
A number of qualified, characterized systems were used to build tolerance windows around procedures
Pass/fail criteria were empirically derived from qualified systems
31
©2010 Waters Corporation 61
UV Detector Qualification
User question:
I have a UV detector (Variable λ or Photo Diode Array)
I want it qualified
How are all the wavelengths tested for accuracy?
Do we need to test all the wavelengths?
Why or why not?
What happens if I change a lamp?
What if an inspector asks how I know λ is right?— Don’t know?
— A better answer
Did you document it?
©2010 Waters Corporation 62
UV Detector Qualification
32
©2010 Waters Corporation 63
Wavelength Accuracy (1)
Example Specification
Wavelength accuracy is one of the most important test we run and inspectors frequently ask about this
It is NOT necessary to test every wavelength in a detector’s range to “provide a high degree of assurance” that an absorbance detector is working properly in terms of wavelength accuracy— Not enough suitable standards are available— It would take lots of time
Replicate gratings are highly reproducible
Design of the wavelength algorithms are validated as part of the instrument design qualification (DQ)
©2010 Waters Corporation 64
Wavelength Accuracy (2)
For an example, if we are qualifying a Waters Tunable UV detector, the unit has on-board diagnostics that can be used to verify proper wavelength accuracy at every power-up— The user is alerted if a wavelength is found out of specification— The cause could be due to something in the cell
The detector uses wavelengths from both the lamp and from an internal filter that is inserted into the light path. A number of choices are available: — Zero Order diffraction grating energy spike (tunable detectors)
o reflects all wavelengths (used in some older detectors)
— 257 nm erbium nitrate filter absorbance maxima— 379 nm erbium nitrate filter absorbance maxima— 486 nm deuterium energy line (used as anchor for 500nm top of range)
— 487nm erbium nitrate filter (generally not used because of 486nm deuterium line)
— 521 nm erbium nitrate filter absorbance maxima— 656 nm deuterium energy line (anchor wavelength peak)
33
©2010 Waters Corporation 65
Wavelength Accuracy (3)
The OQ is used as a means of qualifying not only the detector but also the internal diagnostics
Caffeine can be used for qualification purposes and has peaks at 205nm and 273nm for 1nm Slit Band Width
Testing 205nm with Caffeine is about as far as we can get from the 656nm anchor wavelength provided by the lamp and used in the start up calibration— If a wavelength 452nm away from the calibration wavelength
is accurate… those closer should also be accurate
Using the diagnostics and the caffeine solution, we have verified 6 different wavelengths, — 205, 257, 273, 379, 521 and 656nm
Using the power up diagnostics between qualifications allows us to check 257, 379, 521 and 656nm— It is good practice to power down and back up weekly
©2010 Waters Corporation 66
Wavelength Accuracy (4)
Absorbance detector test procedures are based on ASTM E 1657-94 *— *Standard Practice for Testing Variable-Wavelength
Photometric Detectors Used in Liquid Chromatography
Qualification of wavelength accuracy in this manner is common practice been audited successfully on many occasions
Knowing about the onboard diagnostics in this case and documenting their use allows you to explain to an inspector how you know the wavelengths are accurate after a lamp change between qualifications
34
©2010 Waters Corporation 67
Wavelength Accuracy (5)
The test solutions used were tested by NIST with an accuracy of + or- 0.3nm— Wavelength is fixed by the physics of the test solution— Lambda max can change with detector slit bandwidth
depending upon peak shape of each peak in the solution — Concentration of the accuracy test solutions is not critical
The product specification of 1nm is added to the NIST 0.3nm error and rounded up to a pass/fail criteria of +/-2nm because errors are additive and because the detector “bins” the results to the nearest nanometer — The pass/fail criteria in this case becomes +/- 2nm
©2010 Waters Corporation 68
Wavelength Accuracy (5)
One way of automating this test is by having the Chromatography Data System make successive injections of the caffeine solution, incrementing the wavelength on each injection and determining the lambda max for the two caffeine peaks— Since there are two detector channels, both 205 and 273nm
can be tested at once— Alternating between 2 wavelengths also verifies λ precision
35
©2010 Waters Corporation 69
Wavelength Accuracy (6)
0.00
0.30
0.60
0.90
1.20
1.50
1.80
190 210 230 250 270 290 310
Wavelength (nm)
Abs
orba
nce
204.7 nm
272.0 nm
λ Max’ For Caffeine
©2010 Waters Corporation 70
Absorbance Linearity (1)
Tested by injecting different concentrations of a solution
Caffeine is also a suitable solution for this test— Peaks are symmetrical
Different cell path lengths require different concentration test solutions to test the linear range of the detector— Beer’s law
— The shorter the path length, the more concentrated the solution must be to absorb the same amount of light
A passing detector is generally defined as one where the drop from a theoretical straight line is ≤ 5% at the top end of it’s specified linear range (per ASTM 1657)
36
©2010 Waters Corporation 71
Absorbance Linearity (2)
A line can be drawn from the different injection areas and the straightness of that line or the linearity can be determined by the R2 of the curve
Acceptable pass/fail values are determined by the linearity specification of the detector and the test solutions used
It is impractical to test linearity at every wavelength— Linearity will very at every wavelength due to lamp energy
— Linearity will vary with every mobile phase, diluent and sample
— Qualification provides assurance that the instrument is functioning as intended
— Method validation should be used to test specific samples
o Suitability for intended use
©2010 Waters Corporation 72
Detector Linearity and Sensitivity
R2 = 0.9993
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
1.60
1.80
0 20 40 60 80 100
Amount Injected
Pea
k H
eigh
t (AU
)
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
Pea
k AU
/Am
ount
% RSD = 3.96
Linearity is R2 of curve
Sensitivity is %RSD Of Peak Heights / Amounts
37
©2010 Waters Corporation 73
Injection Accuracy, Linearity and Precision
Injection Accuracy Linearity and Precision are affected by many factors including:— Mobile phase used
— Degassed state of mobile phase
— Sample diluent used
— Viscosity of the sample
— System dead volume
— Sample loop volume
— Syringe size
— Syringe draw rate
— Number of motor steps per microliter (varies with syringe size)
— Vial septa used
— Injection mode used
©2010 Waters Corporation 74
Injection Accuracy
While precision is most important, injection accuracy allows for easier movement of an analysis from one system to the next
One way to test is by weighing the aspirated amount of the sample and translating that to volume— Using water for the sample makes this easier at room temps
— Enough weight must be removed from a vial such that the weighing error is not a significant part of the measurement
38
©2010 Waters Corporation 75
Injection Accuracy
An example injection accuracy specification:
The accuracy specification is at 50uL with specific solvents
— Why was the specification written this way?— Six injections are made to remove enough weight (300uL)— The injector is deemed to meet specification if it passes the test— Apparent performance at different injection volumes, with
different solvents, sample loops, syringes, mobile phases columns and samples will vary
— It is not practical to test at all volumeso Method Validation and System Suitability should be used to
demonstrate suitability for intended use
©2010 Waters Corporation 76
Injection Accuracy
Another way to test, one that can be automated, is by performing actual injections at different volumes to determine the linearity and then using the x intercept as a means to determine the accuracy error
While this may be less metrologic, it demonstrates the injector in the same way that it will be used and provides a high degree of assurance that the injector is operating as intended
The qualification test assures that the syringe drives, needle, valves and injection algorithms are all functioning as intended— Method Validation and System Suitability should be used to
demonstrate suitability for the specific intended use
39
©2010 Waters Corporation 77
Injector Linearity and Accuracy
y = 25000x + 1250
0
200000
400000
600000
800000
1000000
1200000
-1 9 19 29 39 49
volume injected (uL)
area
(uV*
sec)
X-intercept = - 0.05 uL
Linearity = R2 of curve
Accuracy = X Intercept
©2010 Waters Corporation 78
Injection Linearity
Linearity o Linearity, for this example, is specified from 1 to 100uLo Linearity above 100uL is not specifiedo Injections above 150uL will provide best results with the
2500uL syringe so they can be made in one syringe drawo Making injections at 5uL using a large size syringe and large
volume sample loop is not the best practice because a large loop means more air in the system that will stretch before the actual sample is pulled in
o But, a 250uL sample loop is not large enough for accurate 200uL injections. At least a 330uL loop needed for 400uL of total sample loop volume
o One system configuration is not the best for all injection volume sizes (just like columns!)
o Think about how you will be using your system before choosing the configuration
o Use method validation and system suitability to demonstrate suitability for intended use
40
©2010 Waters Corporation 79
Injection Linearity
Injection linearity testing in the OQ is used as a means of determining if the injector is functioning as expected across the typically used analytical range
The algorithms that determine motor drive functionality and hence linearity are validated as part of the design of the product
It is always best to perform linearity tests with specific compounds, mobile phases, wavelengths, sensitivities, sample loops, syringes and columns to be used— Method Validation and System Suitability
— Demonstrates suitability for intended use
©2010 Waters Corporation 80
Injection Linearity
A user Question:— An HPLC injector has been qualified using at 5, 10, 20 and 50uL
— Is it still linear at 200uL?
The answer is “it depends” what you mean— By design, the motor will step the proper amount of steps and the
syringe will move the specified amount so in that sense, it will be linear but:
— Physics will get in the way to make the actual results less linear
o Any injection volume much over one half the sample loop volume will start to disperse out of the loop
o A 200uL injection will require a larger sample loop and that will allow more air in the system. That air stretches when aspirating an injection. This will particularly affect small injection volumes
o A 200uL injection with a standard syringe will require multiple syringe draws and valve movements…this will affect accuracy and linearity
o Mobile phase and degassing used will affect accuracy and linearity
o The column used will affect accuracy and linearity
41
©2010 Waters Corporation 81
Column Effects
A rule of thumb for columns
Do not inject more than 5% of the column void volume
To inject 200uL into a column, the column should have a 4mL void volume— Otherwise the sample diluent becomes the mobile phase
Injecting any more than that alters the chemistry and will make the analysis non-linear
A 4.6 x 150mm column has a void volume of only ≈2.7 mLs
Using a 4mL void volume column will mean waiting about an hour for the peaks to come out or a much higher flow rate
If you are using a system in a non-standard configuration, you can qualify using the standard column and loop, then perform PQ and/or System Suitability with the column and loop to be used for the specific analysis
©2010 Waters Corporation 82
UPLC Flow Rate Linearity
y = 0.8333x + 0.0417
0.0000.200
0.4000.600
0.8001.000
1.2001.400
1.6001.800
-0.1 0.2 0.5 0.8 1.1 1.4 1.7 2
flow rate (mL/min)
1/vo
id ti
me
X-intercept = 0.050 mL/min.Linearity =R2 of curve
42
©2010 Waters Corporation 83
HPLC Flow Linearity and Accuracy
©2010 Waters Corporation 84
Flow Rate Accuracy
HPLC specification example: A column is in place when using the SystemsQTBecause of this, flow rates are limited to the capabilities of the columnFor HPLC, flow rates are tested at 0.5, 0.75, 1 and 1.5, the most common range used by users with analytical columnsFlow rates across the entire range are validated as part of the design qualification (DQ) for the productIf the HPLC passes at the flow rates tested, there is a “high degree of assurance” that it will work within specification at all flow rates (i.e. very low risk of failure)For the most part, flow error is fixed by the check valve size and designThe fixed error of a properly functioning check valve shows up as a higher percentage of total flow at lower flow rates
43
©2010 Waters Corporation 85
System Precision
%RSD of6 ReplicateInjections
©2010 Waters Corporation 86
Gradient Proportioning and Compositional Accuracy
If you use two high pressure pumps (or a binary pump) to create gradients and they use a structurally validated gradient algorithm, so long as the flow rates on both pumps test accurately at multiple flow rates (flow rates are linear), gradients will be accurate and precise.
If you use low pressure mixing on a system that has a structurally validated gradient algorithm, generally any gradient inaccuracy is caused by improper operation or sticking of the Gradient Proportioning Valves (GPVs) so focusing on testing these is one way to assure proper gradient performance— If the valves remain unused for a long time (usually the C or D
valves), they can become sluggish or sticky— Exercising them before use or qualification in such cases is
advised
44
©2010 Waters Corporation 87
Gradient Performance
SD of RetentionTime from each of 7 Peaks overSix Injections
Note: In this UPLC case, there are two high pressure pumps.
SD must be < or = 1.0 sec
©2010 Waters Corporation 88
Compositional Accuracy
Note: In this case, we are testing one pump with low pressure mixing capability of 4 solvents
45
©2010 Waters Corporation 89
Demonstrating “Fit for Intended Use”
A PQ based on specific user and method requirements
Performed by the user AFTER the vendor supplied IQ,OQ and vendor PQ (system level OQ).
Demonstrates that the instrument system can operate reliably under routine, minimum, and maximum operating conditions. Fit for intended use.
Recommended combination of user requirements:— Method specific system suitability
o Precision of quantity of analite injected
o Peak resolution
o Peak tailing
o Column efficiency
— Method specific concentration standards
— Calibration check
©2010 Waters Corporation 90
FDA Requirements
Vendor Supplied IQ, OQ and vendor PQ (system level test)— Works according to manufacturer’s specifications
PQ based on User Requirementso Demonstrates Fit for Intended Use
Ongoing system calibration (qualification) and documented scheduled maintenance
46
©2010 Waters Corporation 91
Advantages of CDS Managed Qualification
Accurate Qualification Testing and Analysis— Less opportunity for human error
— Measures peak areas, peak heights and retention times accurately and consistently
— Custom field calculations and regression analysis
— Testing consistent from system to system
— Reduces time that system is off-line by about half
— Qualifies software and systems in analytical configuration
©2010 Waters Corporation 92
Advantages of CDS Managed Qualification (cont)
Secure and Auditable Qualification Data— 21 CFR Part 11 Compliant Qualification Data
— On-Line Qualification Documentation for Easy Inspection
— Easy tracking and trending of qualification results
— Audit trails and change control part of the data system
— No need for external spreadsheets or third party software
— Secure data environment
47
©2010 Waters Corporation 93
Summary
There are many ways to qualify LC systemsAdvanced technology can save time in your laboratorySystemsQT is the Next Generation in automated Instrument QualificationIt enhances qualification testing and data management in an Empower Chromatography Data System environmentIt offers a whole system test approach for a more accurate indication of overall system-level complianceOperates on a simple wizard built into Empower using Run SamplesBuilt into Empower – no separate softwareto install and uninstallProjects to qualify new modules/systemscan be easily addedDocument review is substantially reduced
©2010 Waters Corporation 94
Questions?