WORKING PROCEDURE Doc. No: WP/QA- 04-01
He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4
COVER PAGE
ClassificationUncontrolled
Controlled
Copy No. (If Controlled):
This Procedure is the property of Heavy Mechanical Complex-3 and the contents are
confidential. This procedure shall not be copied, reproduced or distributed in any form
without the prior approval of the Management Representative.
Heavy Mechanical Complex-3
Hattar Road
Taxila
Prepared By Reviewed By Approved By
Signature
Name
Designation
1
7/31/2019 Analysis(Xrf)Wpqa0401
2/4
WORKING PROCEDURE Doc. No: WP/QA- 04-01
He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4
TABLE OF CONTENTS
1. Purpose
2 Scope
3. Responsibilities.
4. Definitions/ Abbreviations.
5. References.
6. Prerequisites.
7. Limitations.
8. Precautions.
9. Procedure for Chemical Analysis.
10. Acceptance criteria.
11. Results and Records
2
7/31/2019 Analysis(Xrf)Wpqa0401
3/4
WORKING PROCEDURE Doc. No: WP/QA- 04-01
He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4
1.0 Purpose:
This procedure describes the methods used to determine the chemical composition of
different types of materials with handheld XRF Metal Analyzer (X-MET3000TX).
2.0 Scope:
This working procedure covers methods to perform Chemical analysis test at
temperature 0oC to 300oC according to material specifications.
3.0 Responsibilities:
3.1 Engineer (Material Testing) is responsible for preparation of this procedure.
3.2 Manager (QA) and Sr. Manager (QA) are responsible to review and approve this
procedure.
4.0 Definitions / Abbreviations:
HMC -3 Heavy Mechanical Complex -3.
QAD Quality Assurance Division.
Sr. Manager. Senior Manager.
QA Quality Assurance
MT Materials Testing
ASTM American Society for Testing and Material
5.0 References
ASTM E1476
ASTM E305
ASTM E322
ASTM E572
6.0 Prerequisites:
6.1 The sample should have proper identification.
6.2 Test specimen should be ground/polished and cleaned to remove dust, oil and corrosion
from the surface of specimen.
6.3 The machine should have valid Calibration.
3
7/31/2019 Analysis(Xrf)Wpqa0401
4/4
WORKING PROCEDURE Doc. No: WP/QA- 04-01
He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4
7.0 Limitations.
7.1 The analyzer determines only those elements which have atomic number greater than 21.
8.0 Precautions
8.1 X-Ray fluorescence metal analyzer generates X-ray radiation, so unnecessary exposure
to the radiation should be avoided.
8.2 Check that infrared safety sensor should work properly before analysis.
8.3 Use safety shield for small samples.
9.0 Procedure: Turn the power key of analyzer to ON position. The yellow light indicator is switched on.
Wait 1-2 minutes for the peltier cooler and X-rays tube to stabilize.
Push the PDA power switch ON.
From the Start menu tap, selectX-MET to start the program.
Wait for warm-up time completion automatically.
After appearing the measurement screen, name the sample to be measured, select the mode
of operation, and define LOG file which saves all the measurement data, name and
directory.
Modify the way of presenting the measurement and set the measurement time.
Touch the nose of the analyzer to the sample and press the trigger.
Result will be displayed after measuring time lapsed or when trigger is released( if
Measuring time set to 0).
9.1 Acceptance Criteria
Acceptance is made according to the applicable code.
9.2 Results and Records
Chemical analysis test results are recorded on format no, CA01-01.