    WORKING PROCEDURE Doc. No: WP/QA- 04-01

    He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008

    Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

    COVER PAGE

    ClassificationUncontrolled

    Controlled

    Copy No. (If Controlled):

    This Procedure is the property of Heavy Mechanical Complex-3 and the contents are

    confidential. This procedure shall not be copied, reproduced or distributed in any form

    without the prior approval of the Management Representative.

    Heavy Mechanical Complex-3

    Hattar Road

    Taxila

    Prepared By Reviewed By Approved By

    Signature

    Name

    Designation

    1

  • 7/31/2019 Analysis(Xrf)Wpqa0401

    2/4

    WORKING PROCEDURE Doc. No: WP/QA- 04-01

    He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008

    Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

    TABLE OF CONTENTS

    1. Purpose

    2 Scope

    3. Responsibilities.

    4. Definitions/ Abbreviations.

    5. References.

    6. Prerequisites.

    7. Limitations.

    8. Precautions.

    9. Procedure for Chemical Analysis.

    10. Acceptance criteria.

    11. Results and Records

    2

  • 7/31/2019 Analysis(Xrf)Wpqa0401

    3/4

    WORKING PROCEDURE Doc. No: WP/QA- 04-01

    He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008

    Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

    1.0 Purpose:

    This procedure describes the methods used to determine the chemical composition of

    different types of materials with handheld XRF Metal Analyzer (X-MET3000TX).

    2.0 Scope:

    This working procedure covers methods to perform Chemical analysis test at

    temperature 0oC to 300oC according to material specifications.

    3.0 Responsibilities:

    3.1 Engineer (Material Testing) is responsible for preparation of this procedure.

    3.2 Manager (QA) and Sr. Manager (QA) are responsible to review and approve this

    procedure.

    4.0 Definitions / Abbreviations:

    HMC -3 Heavy Mechanical Complex -3.

    QAD Quality Assurance Division.

    Sr. Manager. Senior Manager.

    QA Quality Assurance

    MT Materials Testing

    ASTM American Society for Testing and Material

    5.0 References

    ASTM E1476

    ASTM E305

    ASTM E322

    ASTM E572

    6.0 Prerequisites:

    6.1 The sample should have proper identification.

    6.2 Test specimen should be ground/polished and cleaned to remove dust, oil and corrosion

    from the surface of specimen.

    6.3 The machine should have valid Calibration.

    3

  • 7/31/2019 Analysis(Xrf)Wpqa0401

    4/4

    WORKING PROCEDURE Doc. No: WP/QA- 04-01

    He a v y M e c h a n i c a l C o m p l e x - 3 Rev. No : 00 Date: 09/06/2008

    Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

    7.0 Limitations.

    7.1 The analyzer determines only those elements which have atomic number greater than 21.

    8.0 Precautions

    8.1 X-Ray fluorescence metal analyzer generates X-ray radiation, so unnecessary exposure

    to the radiation should be avoided.

    8.2 Check that infrared safety sensor should work properly before analysis.

    8.3 Use safety shield for small samples.

    9.0 Procedure: Turn the power key of analyzer to ON position. The yellow light indicator is switched on.

    Wait 1-2 minutes for the peltier cooler and X-rays tube to stabilize.

    Push the PDA power switch ON.

    From the Start menu tap, selectX-MET to start the program.

    Wait for warm-up time completion automatically.

    After appearing the measurement screen, name the sample to be measured, select the mode

    of operation, and define LOG file which saves all the measurement data, name and

    directory.

    Modify the way of presenting the measurement and set the measurement time.

    Touch the nose of the analyzer to the sample and press the trigger.

    Result will be displayed after measuring time lapsed or when trigger is released( if

    Measuring time set to 0).

    9.1 Acceptance Criteria

    Acceptance is made according to the applicable code.

    9.2 Results and Records

    Chemical analysis test results are recorded on format no, CA01-01.

    4