MEGOHMMETER 6501 6503 ENGLISH User Manual

6501 - Instrumart · 2017. 2. 2. · Megohmmeter Models 6501/6503 5 1.3 Receiving Your Shipment Upon receiving your shipment, make sure that the contents are con-sistent with the

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

  • MEGOHMMETER 65016503

    E N G L I S H User Manual

  • Statement of Compliance

    Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

    We guarantee that at the time of shipping your instrument has met its published specifications.

    An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

    The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

    Serial #: ________________________________

    Catalog #: 2126.51/ 2126.52

    Model #: 6501/6503

    Please fill in the appropriate date as indicated:

    Date Received: _________________________________

    Date Calibration Due: _______________________

    Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com

  • Table of Contents

    1. INTRODUCTION ........................................................... 3

    1.1 International Electrical Symbols ......................................41.2 DefinitionofMeasurementCategories ............................41.3 ReceivingYourShipment ................................................51.4 OrderingInformation .......................................................5

    1.4.1 AccessoriesandReplacementParts ................. 5

    2. PRODUCT FEATURES .................................................. 6

    2.1 Description ......................................................................62.2 Features ..........................................................................62.3 Model6501ControlFeatures .........................................82.4 Model6503ControlFeatures .........................................9

    3. SPECIFICATIONS ....................................................... 10

    4. OPERATION ................................................................ 12

    4.1 Auto-ranging .................................................................124.2 SafetyCheck-VoltageTest ..........................................124.3 InsulationResistanceTesting(MΩRange) ...................12

    4.3.1 TestVoltage ......................................................134.3.2 SpotTesting ......................................................134.3.3 RatioTesting .....................................................14

    4.4 SuccessfulInsulationResistanceTesting .....................144.5 UtilizationoftheGuardTerminal(Model6503) ................... 154.6 InsulationMeasurement-Connections ........................174.7 InsulationResistanceonMotors ...................................21

  • 5. MAINTENANCE .......................................................... 24

    5.1 Warning ........................................................................245.2 Cleaning ........................................................................245.3 FuseReplacement ........................................................25

    RepairandCalibration ..................................................................26TechnicalandSalesAssistance ...................................................26Limited Warranty ..........................................................................27WarrantyRepairs ..........................................................................27

  • MegohmmeterModels6501/6503 3

    CHAPTER 1

    INTRODUCTION

    Warning Thesesafetywarningsareprovidedtoensurethesafetyofpersonnelandproperoperationoftheinstrument.• Readtheinstructionmanualcompletelyandfollowall

    safetyinformationbeforeoperatingthisinstrument.• Safetyistheresponsibilityoftheoperator!• Testsaretobecarriedoutonlyondeadcircuits!Check

    forlivecircuitsbeforemakingresistancemeasure-ments(safetycheck).

    • Thesemegohmmetersareratedfor600VCATIIor300VCATIII-PollutionDegree2.

    • TheMegohmmeterModels6501/6503aresourcesofhighvoltage,asisthesampleconnectedtothem.Allpersonsperformingorassistinginthetestsmustemployallsafetyprecautionstopreventelectricalshocktothemselvesandtoothers.

    • AEMC®Instrumentsconsiderstheuseofrubberglovestobeanexcellentsafetypracticeeveniftheequipmentisproperlyoperatedandcorrectlygrounded.

    • Whentestingcapacitancesamples,makesurethattheyhavebeenproperlydischargedandthattheyaresafetotouch.Dielectricinsulationsamplesshouldbeshort-circuitedforatleastfivetimestheamountoftimetheywereenergized.

    • Usetheleadssuppliedwiththemegohmmeter.Ifdefectiveorused,replacebeforetesting.

  • 4 MegohmmeterModels6501/6503

    1.1 International Electrical Symbols

    This symbol signifies that the instrument is protected by double or reinforced insulation.

    This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.

    Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

    In conformity with WEEE 2002/96/EC

    1.2 Definition of Measurement Categories

    CAT II: For measurements performed on circuits directlyconnectedtotheelectricaldistributionsystem.Examplesaremeasurements on household appliances or portabletools.

    CAT III: Formeasurements performed in the building installationatthedistributionlevelsuchasonhardwiredequipmentinfixedinstallationandcircuitbreakers.

    CAT IV: For measurements performed at the primary electricalsupply (

  • MegohmmeterModels6501/6503 5

    1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsarecon-sistentwith thepacking list.Notifyyourdistributorofanymissingitems.Iftheequipmentappearstobedamaged,fileaclaimimme-diatelywiththecarrierandnotifyyourdistributoratonce,givingadetailed description of any damage. Save the damaged packingcontainertosubstantiateyourclaim.Donotuseaninstrumentthatappearstobedamaged.

    1.4 Ordering Information

    Megohmmeter Model 6501 .....................................Cat. #2126.51Includes a soft carrying case, two color-coded leads, two insulated alligator clips, test probe and user manual.

    Megohmmeter Model 6503 .....................................Cat. #2126.52Includes a soft carrying case, three color-coded leads, three insulated alligator clips, test probe and user manual.

    1.4.1 Accessories and Replacement PartsReplacement soft carrying case ................................Cat. #2126.71

    Replacement leads (black and red), alligator clips (black and red) and one safety test probe (6501) ......Cat. #2126.72

    Replacement leads (black, red and blue), alligator clips (black, red and blue) and one safety test probe (6503).........Cat. #2126.73

    Replacement Fuse - 0.2A, 600V HPC (6501) ............Cat. #2970.95

  • 6 MegohmmeterModels6501/6503

    CHAPTER 2

    PRODUCT FEATURES

    2.1 Description

    TheAEMC® Models 6501 and 6503 are compact, self-containedhand-crankedmegohmmeters.Theyarepracticalanddependableinstrumentsdesigned forabroad rangeofplantandfieldserviceapplications,suchasacceptancetestingandpreventivemaintenanceofwiring,cables,switchgear,andmotors.Theeasyhand-crankedoperationprovidesasteadyrectifiedDCvoltageoutputacrosstheentirerangeforconsistentlyreliablereadings.

    Thesemegohmmeters incorporateabuilt-ingeneratorandacon-stantDCvoltagecircuit toprovideastableoutputandgivedirectinsulationresistancereading.

    TheModels6501/6503haveauniqueauto-ranging featurewhichexpands the scale by x10 and almost doubles the scale length.Whenthepointerreachesneartheendofscale,theauto-rangingfeatureactivates,returningthepointertothebeginningofthescalewiththex10redLEDindicatoron.

    Insulation resistance is indicated with a rugged taut-band metermovement on a logarithmic scale. The indicator scale is easy toread.

    2.2 Features

    • Measures insulationat500V,resistanceandcontinuity (Model6501)

    • Threetestvoltages:250V,500V,1000V(Model6503)

  • MegohmmeterModels6501/6503 7

    • Testvoltageconstantacrosstheentiremeasurementrange• AutorangingMΩ,kΩandΩranges(Model6501)• AutorangingMΩranges(Model6503)• Designedforharshenvironments:offshore,mining,heavy-duty

    field,industrial,andmilitaryuse• Compact,self-containedpackage;foldingcrank• Largedirectreadingscale• 600Vtestvoltagerange(safetycheck)

  • 8 MegohmmeterModels6501/6503

    2.3 Model 6501 Control Features

    kΩΩ

    MV

    x 10

    5

    01 2

    6 710

    V

    FUS

    123451020

    1020 30

    0.5

    50(500V)

    (1mA)(5mA)

    ΜΩ

    2 ! C42300

    0100

    200 400 500 600ΩκΩ

    43 8 9

    40

    0

    MEGOHMMETERModel 6501INSTRUMENTS

    ®

    600V CAT II / 1000V CAT I !

    1 2 3

    4

    5

    6

    7

    8

    9

    Figure 2-1

    1. Line(-)terminalBlack2. Fuse0.2A3. Earth(+)terminalRed4. AmberLEDilluminateswhenpropercrankspeedisattained,

    indicatingpresenceofselectedoutput5. Foldawayhandle6. Mechanicalzeroadjustmentscrew7. Rangeswitch-safetytestfirstonMΩ/Vrangeto600VAC8. Autoranging:RedLEDilluminateswhenreadingneedstobe

    multipliedx10toobtainthetruevalue(MΩ,kΩ,Ωranges)9. Securityfingerrestisolatesusershandfromterminalswhile

    givingafirmgripduringanoperation

  • MegohmmeterModels6501/6503 9

    2.4 Model 6503 Control Features

    MEGOHMMETERModel 6503INSTRUMENTS

    ®V500V

    1kV25

    0V

    x 10

    300

    0100

    200 400 500 600

    V

    G

    20 15304050100200500

    5020 10

    5 4 32 1.5

    10

    1(1kV)

    (250V)(500V)ΜΩ

    2 ! C42

    600V CAT II / 1000V CAT I !

    1 2 3

    4

    5

    6

    7

    8

    9

    Figure 2-2

    1. Line(-)terminal-Black2. Guardterminal-Blue3. Earth(+)terminal-Red4. GreenLED illuminateswhenproper crank speed is attained,

    indicatingpresenceofselectedtestvoltage5. Foldawayhandle6. Mechanicalzeroadjustmentscrew7. Rangeswitch-Safetytestfirston600V/V~rangeto600VAC8. Autoranging:RedLEDilluminateswhenreadingneedstobe

    multipliedx10toobtainthetruevalue(MΩrange)9. Securityfingerrestisolatesusershandfromterminalswhile

    givingafirmgripduringanoperation

  • 10 MegohmmeterModels6501/6503

    CHAPTER 3

    SPECIFICATIONS

    MODELS 6501 6503ELECTRICALInsulation TestsTest Voltage 250V 500V 1000V

    −0.5 to 5.5MΩ; 5.5 to 200MΩ

    1 to 11MΩ; 11 to 500MΩ1 to 11MΩ; 11 to 500MΩ

    10 to 110MΩ; 110 to 5000MΩ

    Short Circuit Current

    5mA (max)

    Accuracy 2.5% of Scale LengthResistance TestsMeasurement Range(Auto-Ranging)

    0 to 45kΩ; 45 to 500kΩ −

    DC Test Current 1mA constant, 500V max −Accuracy 2.5% of Scale Length −Automatic Discharge 5s/µF 8s/µF

    Continuity TestsTest Voltage 5mA constant, 10V max −Measurement Range(Auto-Ranging)

    0 to 9Ω; 10 to 100Ω −

    Accuracy 2.5% of Scale Length −Fuse Protection 0.2A, 600V HPC 6x32mm −Voltage Tests/Safety CheckVoltage Range 0 to 600VACFrequency 45 to 450HzAccuracy 3% of ScaleInput Impedance 100kΩ (approx)

  • MegohmmeterModels6501/6503 11

    MODELS 6501 6503Scale Direct reading on all ranges; 3" (76mm), black on whiteTest Voltage Indicator

    Amber LED indicates propertest voltage and crank speed

    NOTE: If the LED is not ON during testing, the reading is not valid

    Green LED indicates propertest voltage and crank speed

    NOTE: If the LED is not ON during testing, the reading is not valid

    Auto-Range Indicator

    Red LED indicates Auto-Range;multiply MΩ, kΩ & Ω

    readings x 10

    Red LED indicates Auto-Range;multiply MΩ readings x 10

    Power Source Hand-cranked, heavy duty brushless alternator andsolid state rectification and amplification for stable voltage;

    test voltage reached at 140rpm (approx)Meter Movement Rugged taut band suspensionMECHANICALDimensions 4.7 x 4.4 x 5.3" (120 x 111 x 135mm) with lidWeight 2.5 lbs (1.1kg)Terminals “+” Line (red); “–” earth (black);

    accepts standard 4mm safety plugsENVIRONMENTALTemperature Reference Operating Range Storage Range

    68 to 79°F (20 to 26°C), 30 to 50% RH14 to 113°F (-10 to 45°C), 0 to 75% RH-4 to 176°F (-20 to 80°C), 0 to 90% RH

    SAFETYSealed IP54 with lid, IP52 without lidRating EN 61010-1, 600V CAT II; 300V CAT III; Pollution Degree 2Double Insulation Yes CE Mark Yes

    *Specifications are subject to change without notice

  • 12 MegohmmeterModels6501/6503

    CHAPTER 4

    OPERATION

    4.1 Auto-rangingHow it works:Whenthepointerreachesapproximately90%ofthefullscalelength,theauto-rangeistriggered,theredauto-range(x10)LEDlightsup.Thepointergoesbacktothebeginningofthescaleandresumesitstravel.Thereadingsshouldbemultipliedby10whenthex10LEDison.

    4.2 Safety Check - Voltage Test Beforemeasuringinsulationresistance,confirmthatthesampleisfullydischarged(particularlyindielectricandcapacitancesamples)andthatthesampleisnotconnectedtoanenergizedcircuit.Toperformthevoltagetest,settherangeselectiontoVMΩ(Model6501) or 500V MΩ (Model 6503), connect and read on the redvoltage scale (600VACmax).DO NOT CRANK. Ifmeasuring onaDCcircuit, thepointerwilldeflectbut themeasurementmaybeinaccurate.

    4.3 Insulation Resistance Testing (MΩ Range)Afterchecking fora livecircuit (seesafetycheck),select the testvoltageandconnectthemegohmmeter.Severalconnectionexam-plesare illustrated further.Crank thehandleuntil theAmberLED(Model6501)orGreenLED(Model6503)ontheupperright-handsidelights,ensuringthattheselectedtestvoltagehasbeenreached.

  • MegohmmeterModels6501/6503 13

    4.3.1 Test VoltageNopublishedstandardtellswhichvoltagetochooseforanygivenwinding.However, published recommendations could be summa-rizedasfollows:

    Rated Voltage of Motor Test VoltageBelow 115 250V

    115 250 or 500V230 500V460 500 or 1000V

    4.3.2 Spot TestingAsageneralruleinspottesting,testvoltageshouldbeapplieduntilnovariationinreadingisnotedfor15secondsorappliedforafixed60seconds.Oncapacitancecircuits,crankingforuptoaminuteormoremaybenecessarytofullychargethesample.Whatminimumvalueshouldbemeasured?TheIEEEstandardN°43-2000statesthatit isimpossibletospecifythevalueofinsulationresistanceatwhichawindingwillfailelectrically,butonmotors,minimumread-ingsgenerallystatedare:

    Rated Voltage R minimum 250 or less 2 MΩ 460 2 MΩ

    There is no fixed figure for determiningwhat is good and bad inresistancereadings,butagoodguidewouldbe1MΩforeveryonehundredappliedoperatingvolts,asaminimumfigure.Thisappliestomotors and transformers.When the insulation resistance getsthislow,anelectricalbreak-downcanbeexpectedatanytimeandrewindingorreplacingshouldbeconsidered.It is not unusual for awinding to be 10 to 100 times the recom-mendedminimumvalue (IEEEstandard43-2000:Recommended Practice for Testing Insulation Resistance of Rotating Machinery),butthisvarieswithtemperatureandhumidity.

  • 14 MegohmmeterModels6501/6503

    4.3.3 Ratio TestingIn time resistance reading (DielectricAbsorption Ratio), readingsare taken at 30 seconds and 60 seconds to obtain the dielectricabsorptionratio.

    Insulationresistance@60s

    Insulationresistance@30s

    Thistestisusefultoincreasetheaccuracyofspottesting.Ingen-eral,aratioof1.25:2orbettershouldberequired.Aratiobelowthisindicatesthatrepairisprobablyneeded.Remember thataDC insulation testmaybeused foracceptancetesting,butismorecommonlyusedtocheckthegradualdeteriora-tionofequipmentoveritslife.Consultyourequipmentmanufacturerforspecifictestortestvoltageifnotknown.Insulationresistancedecreaseswithmoisture,temperatureandageandshouldberecordedovertimeatagiventemperatureandcor-rected.

    4.4 Successful Insulation Resistance Testing• Checkwiththeequipmentmanufacturerforfactoryinsula-

    tionresistancereadings.• Donotrelyoninsulationresistancetestingaloneasproof

    ofwindingconditions.• Donotexpectthesamevalueforallpartsofallmachines.• Observeconsistenttesttimeduration,recognizingthat

    totalcurrentthroughinsulationundertestwillvarywithtime.

    • Correctallreadingsproperlytoastandardreferencetem-perature(seeIEEEstd#43-2000,Temperature Correction Curve).

    • Knowwhatyouaretesting.Isolatethepieceofequipmentfromothercircuitry.

    • Watchtrendsratherthanrelyingonsingle“spot”readings.

    =DielectricAbsorptionratio(DAR)

  • MegohmmeterModels6501/6503 15

    4.5 Utilization of the Guard Terminal (Model 6503)

    G

    A

    D

    BC

    A = InsulationB = ShieldC = ConductorD = Exposed surface

    Figure 4-1

    Guardterminalsareusefulwhenmeasuringhighresistancevaluesandforstabilizingreadings.Whentestingtheinsulationattheendofacable,itisnecessarytoeliminatetheerrorfromsurfaceleakagewhichoccurs,particularlyathighresistancevalues.Thepurposeoftheguardterminal istoprovideathirdterminalwithinthepathofthesurfaceleakage“D”.ConnecttheinstrumentasshowninFigure4-1.Ifthereisnoshieldat“B”,useacopperwirewoundseveraltimesaroundtheexposedsurface“B”.(Note:Ifashield isnotavailableandyoudonotmakeupa shieldaround “B”andconnect to theguardterminal(-),themeasurementwillbeerroneousandleadtoconfusionastothecable’sstate.)

  • 16 MegohmmeterModels6501/6503

    If theguard terminal isnotconnectedat“B”, the instrumentmea-sures the current “I” flowing through the insulation anda surfaceleakagecurrent“i1”.SeeFigure4-2.

    To EARTHterminal

    I

    To LINEterminal

    NO CONNECTIONto Guard terminal

    A

    B

    C

    Ry

    Rx

    Rz

    Rx: resistance through insulation

    Ry: surface resistance

    Rz: surface resistance

    i 1

    Figure 4-2

    Whentheguardterminalisconnectedat“B”,theinstrumentmea-suresthecurrent“I”andnotthesurfaceleakagecurrent“i1”,whichisnot included in themeasurement.SeeFigure4-3.This typeofmeasurementwillgivethetruevalueoftheresistance“Rx”,provid-ingthe“Ry”and“Rz”arenottoolow.

    To EARTHterminal

    I

    To LINEterminal

    To Guard terminal

    A

    B

    C

    Ry

    Rx

    Rz

    Rx: resistance through insulation

    Ry: surface resistance

    Rz: surface resistance

    i 1

    Figure 4-3

  • MegohmmeterModels6501/6503 17

    4.6 Insulation Measurement - Connections

    Figures 4-4 shows the connections to measure the insulation ofone conductor to the other conductors.The cable should be dis-connectedatbothendstoavoidleakagethroughswitchboardsandpanels.For theModel6503, theconnection to theguard terminalisusedtoeliminatetheeffectsofsurfaceleakageacrossexposedinsulationatoneendofthecable.RefertothesectiononUtilizationoftheGuardTerminal(§4.5).

    G

    Cable Insulation

    Conductor Under Test

    Cable Insulation

    6501

    6503

    Conductor Under Test

    Fus

    Figure 4-4

  • 18 MegohmmeterModels6501/6503

    Figures 4-5 shows the connections for testing insulation from asupplyconductortoground(motorframe).

    Cable Insulation

    Conductor Under Test

    Figure 4-5

  • MegohmmeterModels6501/6503 19

    Figure4-6showstheconnectionstoatransformer(lightingordis-tribution).Make sure that the switches and/or circuit breakers onbothsidesareopen.Checkthehighvoltagewindingtoground,lowvoltagetoground,andtheresistancebetweenthemwithnowindingground.

    Tested Winding

    TransformerGrounding Lug

    Figure 4-6

  • 20 MegohmmeterModels6501/6503

    Figure4-7shows theconnections formeasuring the insulationofa three-phase line togroundby connecting the jumpersbetweenphases.Thisgivesareadingofallconductorsatonce.Ifaloadsuchasamotor,heater,etc.,isattachedtotheotherendoftheline,itwillreadtheloadresistancetogroundatthesametime.Byremovingthejumpers,readingscanbemadebetweentheindividualconduc-torsandground.

    ~~

    ~ ~~~

    Jumpers

    Figure 4-7

  • MegohmmeterModels6501/6503 21

    4.7 Insulation Resistance on Motors

    Figure4-8showsreadingtheresistancetogroundofathree-phasemotor winding. Since the three-phase motors are internally con-nected,it isonlynecessarytoconnectoneleadtothemotorleadandtheotherleadtothemotorframeasshown.

    Figure 4-8

  • 22 MegohmmeterModels6501/6503

    Figure4-9showsthewindingsofathree-phasemotorseparated.Sometimesthiscanbedoneattheleadterminalswhileothertimestheendbellsmustberemovedtogetattheleadwiresofthecoils.By connecting themegohmmeter as shown, thephase insulationresistancevaluecannowbedetermined.Readbetweenphases“A”and“B”,then“B”and“C”,then“C”and“A”.

    A

    B

    C

    Figure 4-9

  • MegohmmeterModels6501/6503 23

    Figure4-10showsconnectionsfortestinginsulationfromasupplyconductorinaswitchboxtoground(motorframe).Anidenticaltestmaybecarriedoutfromthemotorstarter.

    Motor Side of Switch:Connection to One Leg

    GroundedMotor Frame

    Starter In

    Figure 4-10

  • 24 MegohmmeterModels6501/6503

    CHAPTER 5

    MAINTENANCE

    5.1 Warning

    • Formaintenanceuseonlyoriginalfactoryreplacementparts.

    • Toavoidelectricalshock,donotattempttoperformanyservic-ingunlessyouarequalifiedtodoso.

    • Do not perform any service while the Megohmmeter Model6501/6503isonanycircuit.

    • Toavoidelectricalshockand/ordamageto the instrument,donotletwaterorotherforeignagentsintotheelectronicmodule.

    Warning: IfthecrankspeedLEDisnotONduringtesting,thereadingisnotvalid.

    5.2 Cleaning

    Themegohmmetermaybegentlycleanedwithasoftcloth,soapandwater.Dryimmediatelyaftercleaning.Avoidwaterpenetrationintotheelectronicmodule.

    Makesure themegohmmeterandall leadsaredrybefore furtheruse.

  • MegohmmeterModels6501/6503 25

    5.3 Fuse Replacement

    WARNING: Foryoursafety,foldthecrankanddisconnectalloftheleadsfromthemeter.

    • Withaflatscrewdriveroracoin,pushandturnthefuseholderaquarterturncounter-clockwise.

    • Replace the fuse with an identical one - 0.2A, 600V HPC6x32mm

  • 26 MegohmmeterModels6501/6503

    Repair and Calibration

    Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

    For instrument repair and calibration:YoumustcontactourServiceCenterforaCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandpro-cessedpromptly.PleasewritetheCSA#ontheoutsideoftheshippingcontainer.Ifthe instrument is returned for calibration,weneed to know if youwant a standardcalibration,oracalibration traceable toN.I.S.T. (Includescalibrationcertificateplusrecordedcalibrationdata).

    Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone: (800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail: [email protected]

    (Orcontactyourauthorizeddistributor)Costs forrepair,standardcalibration,andcalibrationtraceable toN.I.S.T.areavail-able.NOTE: You must obtain a CSA# before returning any instrument.

    Technical and Sales Assistance

    If youareexperiencingany technicalproblems,or requireanyassistancewith theproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

    ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone: (800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail: [email protected] www.aemc.com

    NOTE: Do not ship Instruments to our Foxborough, MA address.

  • MegohmmeterModels6501/6503 27

    Limited Warranty

    TheModels6501and6503arewarrantedtotheownerforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.

    For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.

    What AEMC® Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.

    REGISTER ONLINE AT: www.aemc.com

    Warranty Repairs

    What you must do to return an Instrument for Warranty Repair: First,requestaCustomerServiceAuthorizationNumber(CSA#)byphoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwiththesignedCSAForm.PleasewritetheCSA#ontheoutsideoftheshippingcon-tainer.Returntheinstrument,postageorshipmentpre-paidto:

    Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone: (800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail: [email protected]

    Caution:Toprotectyourselfagainst in-transit loss,werecommendyouinsureyourreturnedmaterial.

    NOTE: You must obtain a CSA# before returning any instrument.

  • 010/15

    99-MAN100328v7

    Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA

    Phone:(603)749-6434•Fax:(603)742-2346www.aemc.com

    INTRODUCTION1.1International Electrical Symbols1.2Definition of Measurement Categories1.3Receiving Your Shipment1.4Ordering Information1.4.1Accessories and Replacement Parts

    PRODUCT FEATURES2.1Description2.2Features2.3Model 6501 Control Features2.4Model 6503 Control Features

    SPECIFICATIONSOPERATION4.1Auto-ranging4.2Safety Check - Voltage Test 4.3Insulation Resistance Testing (MΩ Range)4.3.1Test Voltage4.3.2Spot Testing4.3.3Ratio Testing

    4.4Successful Insulation Resistance Testing4.5Utilization of the Guard Terminal (Model 6503)4.6Insulation Measurement - Connections4.7Insulation Resistance on Motors

    MAINTENANCE5.1Warning  5.2Cleaning5.3Fuse Replacement

    Repair and CalibrationTechnical and Sales AssistanceLimited WarrantyWarranty Repairs