MEGOHMMETER 65016503
E N G L I S H User Manual
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.
We guarantee that at the time of shipping your instrument has met its published specifications.
An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.
The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2126.51/ 2126.52
Model #: 6501/6503
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com
Table of Contents
1. INTRODUCTION ........................................................... 3
1.1 International Electrical Symbols ......................................41.2 DefinitionofMeasurementCategories ............................41.3 ReceivingYourShipment ................................................51.4 OrderingInformation .......................................................5
1.4.1 AccessoriesandReplacementParts ................. 5
2. PRODUCT FEATURES .................................................. 6
2.1 Description ......................................................................62.2 Features ..........................................................................62.3 Model6501ControlFeatures .........................................82.4 Model6503ControlFeatures .........................................9
3. SPECIFICATIONS ....................................................... 10
4. OPERATION ................................................................ 12
4.1 Auto-ranging .................................................................124.2 SafetyCheck-VoltageTest ..........................................124.3 InsulationResistanceTesting(MΩRange) ...................12
4.3.1 TestVoltage ......................................................134.3.2 SpotTesting ......................................................134.3.3 RatioTesting .....................................................14
4.4 SuccessfulInsulationResistanceTesting .....................144.5 UtilizationoftheGuardTerminal(Model6503) ................... 154.6 InsulationMeasurement-Connections ........................174.7 InsulationResistanceonMotors ...................................21
5. MAINTENANCE .......................................................... 24
5.1 Warning ........................................................................245.2 Cleaning ........................................................................245.3 FuseReplacement ........................................................25
RepairandCalibration ..................................................................26TechnicalandSalesAssistance ...................................................26Limited Warranty ..........................................................................27WarrantyRepairs ..........................................................................27
MegohmmeterModels6501/6503 3
CHAPTER 1
INTRODUCTION
Warning Thesesafetywarningsareprovidedtoensurethesafetyofpersonnelandproperoperationoftheinstrument.• Readtheinstructionmanualcompletelyandfollowall
safetyinformationbeforeoperatingthisinstrument.• Safetyistheresponsibilityoftheoperator!• Testsaretobecarriedoutonlyondeadcircuits!Check
forlivecircuitsbeforemakingresistancemeasure-ments(safetycheck).
• Thesemegohmmetersareratedfor600VCATIIor300VCATIII-PollutionDegree2.
• TheMegohmmeterModels6501/6503aresourcesofhighvoltage,asisthesampleconnectedtothem.Allpersonsperformingorassistinginthetestsmustemployallsafetyprecautionstopreventelectricalshocktothemselvesandtoothers.
• AEMC®Instrumentsconsiderstheuseofrubberglovestobeanexcellentsafetypracticeeveniftheequipmentisproperlyoperatedandcorrectlygrounded.
• Whentestingcapacitancesamples,makesurethattheyhavebeenproperlydischargedandthattheyaresafetotouch.Dielectricinsulationsamplesshouldbeshort-circuitedforatleastfivetimestheamountoftimetheywereenergized.
• Usetheleadssuppliedwiththemegohmmeter.Ifdefectiveorused,replacebeforetesting.
4 MegohmmeterModels6501/6503
1.1 International Electrical Symbols
This symbol signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.
Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
CAT II: For measurements performed on circuits directlyconnectedtotheelectricaldistributionsystem.Examplesaremeasurements on household appliances or portabletools.
CAT III: Formeasurements performed in the building installationatthedistributionlevelsuchasonhardwiredequipmentinfixedinstallationandcircuitbreakers.
CAT IV: For measurements performed at the primary electricalsupply (
MegohmmeterModels6501/6503 5
1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsarecon-sistentwith thepacking list.Notifyyourdistributorofanymissingitems.Iftheequipmentappearstobedamaged,fileaclaimimme-diatelywiththecarrierandnotifyyourdistributoratonce,givingadetailed description of any damage. Save the damaged packingcontainertosubstantiateyourclaim.Donotuseaninstrumentthatappearstobedamaged.
1.4 Ordering Information
Megohmmeter Model 6501 .....................................Cat. #2126.51Includes a soft carrying case, two color-coded leads, two insulated alligator clips, test probe and user manual.
Megohmmeter Model 6503 .....................................Cat. #2126.52Includes a soft carrying case, three color-coded leads, three insulated alligator clips, test probe and user manual.
1.4.1 Accessories and Replacement PartsReplacement soft carrying case ................................Cat. #2126.71
Replacement leads (black and red), alligator clips (black and red) and one safety test probe (6501) ......Cat. #2126.72
Replacement leads (black, red and blue), alligator clips (black, red and blue) and one safety test probe (6503).........Cat. #2126.73
Replacement Fuse - 0.2A, 600V HPC (6501) ............Cat. #2970.95
6 MegohmmeterModels6501/6503
CHAPTER 2
PRODUCT FEATURES
2.1 Description
TheAEMC® Models 6501 and 6503 are compact, self-containedhand-crankedmegohmmeters.Theyarepracticalanddependableinstrumentsdesigned forabroad rangeofplantandfieldserviceapplications,suchasacceptancetestingandpreventivemaintenanceofwiring,cables,switchgear,andmotors.Theeasyhand-crankedoperationprovidesasteadyrectifiedDCvoltageoutputacrosstheentirerangeforconsistentlyreliablereadings.
Thesemegohmmeters incorporateabuilt-ingeneratorandacon-stantDCvoltagecircuit toprovideastableoutputandgivedirectinsulationresistancereading.
TheModels6501/6503haveauniqueauto-ranging featurewhichexpands the scale by x10 and almost doubles the scale length.Whenthepointerreachesneartheendofscale,theauto-rangingfeatureactivates,returningthepointertothebeginningofthescalewiththex10redLEDindicatoron.
Insulation resistance is indicated with a rugged taut-band metermovement on a logarithmic scale. The indicator scale is easy toread.
2.2 Features
• Measures insulationat500V,resistanceandcontinuity (Model6501)
• Threetestvoltages:250V,500V,1000V(Model6503)
MegohmmeterModels6501/6503 7
• Testvoltageconstantacrosstheentiremeasurementrange• AutorangingMΩ,kΩandΩranges(Model6501)• AutorangingMΩranges(Model6503)• Designedforharshenvironments:offshore,mining,heavy-duty
field,industrial,andmilitaryuse• Compact,self-containedpackage;foldingcrank• Largedirectreadingscale• 600Vtestvoltagerange(safetycheck)
8 MegohmmeterModels6501/6503
2.3 Model 6501 Control Features
kΩΩ
MV
x 10
5
01 2
6 710
V
FUS
123451020
1020 30
0.5
50(500V)
(1mA)(5mA)
ΜΩ
2 ! C42300
0100
200 400 500 600ΩκΩ
43 8 9
40
0
Ω
MEGOHMMETERModel 6501INSTRUMENTS
®
600V CAT II / 1000V CAT I !
1 2 3
4
5
6
7
8
9
Figure 2-1
1. Line(-)terminalBlack2. Fuse0.2A3. Earth(+)terminalRed4. AmberLEDilluminateswhenpropercrankspeedisattained,
indicatingpresenceofselectedoutput5. Foldawayhandle6. Mechanicalzeroadjustmentscrew7. Rangeswitch-safetytestfirstonMΩ/Vrangeto600VAC8. Autoranging:RedLEDilluminateswhenreadingneedstobe
multipliedx10toobtainthetruevalue(MΩ,kΩ,Ωranges)9. Securityfingerrestisolatesusershandfromterminalswhile
givingafirmgripduringanoperation
MegohmmeterModels6501/6503 9
2.4 Model 6503 Control Features
MEGOHMMETERModel 6503INSTRUMENTS
®V500V
1kV25
0V
x 10
300
0100
200 400 500 600
V
G
20 15304050100200500
5020 10
5 4 32 1.5
10
1(1kV)
(250V)(500V)ΜΩ
2 ! C42
600V CAT II / 1000V CAT I !
1 2 3
4
5
6
7
8
9
Figure 2-2
1. Line(-)terminal-Black2. Guardterminal-Blue3. Earth(+)terminal-Red4. GreenLED illuminateswhenproper crank speed is attained,
indicatingpresenceofselectedtestvoltage5. Foldawayhandle6. Mechanicalzeroadjustmentscrew7. Rangeswitch-Safetytestfirston600V/V~rangeto600VAC8. Autoranging:RedLEDilluminateswhenreadingneedstobe
multipliedx10toobtainthetruevalue(MΩrange)9. Securityfingerrestisolatesusershandfromterminalswhile
givingafirmgripduringanoperation
10 MegohmmeterModels6501/6503
CHAPTER 3
SPECIFICATIONS
MODELS 6501 6503ELECTRICALInsulation TestsTest Voltage 250V 500V 1000V
−0.5 to 5.5MΩ; 5.5 to 200MΩ
−
1 to 11MΩ; 11 to 500MΩ1 to 11MΩ; 11 to 500MΩ
10 to 110MΩ; 110 to 5000MΩ
Short Circuit Current
5mA (max)
Accuracy 2.5% of Scale LengthResistance TestsMeasurement Range(Auto-Ranging)
0 to 45kΩ; 45 to 500kΩ −
DC Test Current 1mA constant, 500V max −Accuracy 2.5% of Scale Length −Automatic Discharge 5s/µF 8s/µF
Continuity TestsTest Voltage 5mA constant, 10V max −Measurement Range(Auto-Ranging)
0 to 9Ω; 10 to 100Ω −
Accuracy 2.5% of Scale Length −Fuse Protection 0.2A, 600V HPC 6x32mm −Voltage Tests/Safety CheckVoltage Range 0 to 600VACFrequency 45 to 450HzAccuracy 3% of ScaleInput Impedance 100kΩ (approx)
MegohmmeterModels6501/6503 11
MODELS 6501 6503Scale Direct reading on all ranges; 3" (76mm), black on whiteTest Voltage Indicator
Amber LED indicates propertest voltage and crank speed
NOTE: If the LED is not ON during testing, the reading is not valid
Green LED indicates propertest voltage and crank speed
NOTE: If the LED is not ON during testing, the reading is not valid
Auto-Range Indicator
Red LED indicates Auto-Range;multiply MΩ, kΩ & Ω
readings x 10
Red LED indicates Auto-Range;multiply MΩ readings x 10
Power Source Hand-cranked, heavy duty brushless alternator andsolid state rectification and amplification for stable voltage;
test voltage reached at 140rpm (approx)Meter Movement Rugged taut band suspensionMECHANICALDimensions 4.7 x 4.4 x 5.3" (120 x 111 x 135mm) with lidWeight 2.5 lbs (1.1kg)Terminals “+” Line (red); “–” earth (black);
accepts standard 4mm safety plugsENVIRONMENTALTemperature Reference Operating Range Storage Range
68 to 79°F (20 to 26°C), 30 to 50% RH14 to 113°F (-10 to 45°C), 0 to 75% RH-4 to 176°F (-20 to 80°C), 0 to 90% RH
SAFETYSealed IP54 with lid, IP52 without lidRating EN 61010-1, 600V CAT II; 300V CAT III; Pollution Degree 2Double Insulation Yes CE Mark Yes
*Specifications are subject to change without notice
12 MegohmmeterModels6501/6503
CHAPTER 4
OPERATION
4.1 Auto-rangingHow it works:Whenthepointerreachesapproximately90%ofthefullscalelength,theauto-rangeistriggered,theredauto-range(x10)LEDlightsup.Thepointergoesbacktothebeginningofthescaleandresumesitstravel.Thereadingsshouldbemultipliedby10whenthex10LEDison.
4.2 Safety Check - Voltage Test Beforemeasuringinsulationresistance,confirmthatthesampleisfullydischarged(particularlyindielectricandcapacitancesamples)andthatthesampleisnotconnectedtoanenergizedcircuit.Toperformthevoltagetest,settherangeselectiontoVMΩ(Model6501) or 500V MΩ (Model 6503), connect and read on the redvoltage scale (600VACmax).DO NOT CRANK. Ifmeasuring onaDCcircuit, thepointerwilldeflectbut themeasurementmaybeinaccurate.
4.3 Insulation Resistance Testing (MΩ Range)Afterchecking fora livecircuit (seesafetycheck),select the testvoltageandconnectthemegohmmeter.Severalconnectionexam-plesare illustrated further.Crank thehandleuntil theAmberLED(Model6501)orGreenLED(Model6503)ontheupperright-handsidelights,ensuringthattheselectedtestvoltagehasbeenreached.
MegohmmeterModels6501/6503 13
4.3.1 Test VoltageNopublishedstandardtellswhichvoltagetochooseforanygivenwinding.However, published recommendations could be summa-rizedasfollows:
Rated Voltage of Motor Test VoltageBelow 115 250V
115 250 or 500V230 500V460 500 or 1000V
4.3.2 Spot TestingAsageneralruleinspottesting,testvoltageshouldbeapplieduntilnovariationinreadingisnotedfor15secondsorappliedforafixed60seconds.Oncapacitancecircuits,crankingforuptoaminuteormoremaybenecessarytofullychargethesample.Whatminimumvalueshouldbemeasured?TheIEEEstandardN°43-2000statesthatit isimpossibletospecifythevalueofinsulationresistanceatwhichawindingwillfailelectrically,butonmotors,minimumread-ingsgenerallystatedare:
Rated Voltage R minimum 250 or less 2 MΩ 460 2 MΩ
There is no fixed figure for determiningwhat is good and bad inresistancereadings,butagoodguidewouldbe1MΩforeveryonehundredappliedoperatingvolts,asaminimumfigure.Thisappliestomotors and transformers.When the insulation resistance getsthislow,anelectricalbreak-downcanbeexpectedatanytimeandrewindingorreplacingshouldbeconsidered.It is not unusual for awinding to be 10 to 100 times the recom-mendedminimumvalue (IEEEstandard43-2000:Recommended Practice for Testing Insulation Resistance of Rotating Machinery),butthisvarieswithtemperatureandhumidity.
14 MegohmmeterModels6501/6503
4.3.3 Ratio TestingIn time resistance reading (DielectricAbsorption Ratio), readingsare taken at 30 seconds and 60 seconds to obtain the dielectricabsorptionratio.
Insulationresistance@60s
Insulationresistance@30s
Thistestisusefultoincreasetheaccuracyofspottesting.Ingen-eral,aratioof1.25:2orbettershouldberequired.Aratiobelowthisindicatesthatrepairisprobablyneeded.Remember thataDC insulation testmaybeused foracceptancetesting,butismorecommonlyusedtocheckthegradualdeteriora-tionofequipmentoveritslife.Consultyourequipmentmanufacturerforspecifictestortestvoltageifnotknown.Insulationresistancedecreaseswithmoisture,temperatureandageandshouldberecordedovertimeatagiventemperatureandcor-rected.
4.4 Successful Insulation Resistance Testing• Checkwiththeequipmentmanufacturerforfactoryinsula-
tionresistancereadings.• Donotrelyoninsulationresistancetestingaloneasproof
ofwindingconditions.• Donotexpectthesamevalueforallpartsofallmachines.• Observeconsistenttesttimeduration,recognizingthat
totalcurrentthroughinsulationundertestwillvarywithtime.
• Correctallreadingsproperlytoastandardreferencetem-perature(seeIEEEstd#43-2000,Temperature Correction Curve).
• Knowwhatyouaretesting.Isolatethepieceofequipmentfromothercircuitry.
• Watchtrendsratherthanrelyingonsingle“spot”readings.
=DielectricAbsorptionratio(DAR)
MegohmmeterModels6501/6503 15
4.5 Utilization of the Guard Terminal (Model 6503)
G
A
D
BC
A = InsulationB = ShieldC = ConductorD = Exposed surface
Figure 4-1
Guardterminalsareusefulwhenmeasuringhighresistancevaluesandforstabilizingreadings.Whentestingtheinsulationattheendofacable,itisnecessarytoeliminatetheerrorfromsurfaceleakagewhichoccurs,particularlyathighresistancevalues.Thepurposeoftheguardterminal istoprovideathirdterminalwithinthepathofthesurfaceleakage“D”.ConnecttheinstrumentasshowninFigure4-1.Ifthereisnoshieldat“B”,useacopperwirewoundseveraltimesaroundtheexposedsurface“B”.(Note:Ifashield isnotavailableandyoudonotmakeupa shieldaround “B”andconnect to theguardterminal(-),themeasurementwillbeerroneousandleadtoconfusionastothecable’sstate.)
16 MegohmmeterModels6501/6503
If theguard terminal isnotconnectedat“B”, the instrumentmea-sures the current “I” flowing through the insulation anda surfaceleakagecurrent“i1”.SeeFigure4-2.
To EARTHterminal
I
To LINEterminal
NO CONNECTIONto Guard terminal
A
B
C
Ry
Rx
Rz
Rx: resistance through insulation
Ry: surface resistance
Rz: surface resistance
i 1
Figure 4-2
Whentheguardterminalisconnectedat“B”,theinstrumentmea-suresthecurrent“I”andnotthesurfaceleakagecurrent“i1”,whichisnot included in themeasurement.SeeFigure4-3.This typeofmeasurementwillgivethetruevalueoftheresistance“Rx”,provid-ingthe“Ry”and“Rz”arenottoolow.
To EARTHterminal
I
To LINEterminal
To Guard terminal
A
B
C
Ry
Rx
Rz
Rx: resistance through insulation
Ry: surface resistance
Rz: surface resistance
i 1
Figure 4-3
MegohmmeterModels6501/6503 17
4.6 Insulation Measurement - Connections
Figures 4-4 shows the connections to measure the insulation ofone conductor to the other conductors.The cable should be dis-connectedatbothendstoavoidleakagethroughswitchboardsandpanels.For theModel6503, theconnection to theguard terminalisusedtoeliminatetheeffectsofsurfaceleakageacrossexposedinsulationatoneendofthecable.RefertothesectiononUtilizationoftheGuardTerminal(§4.5).
G
Cable Insulation
Conductor Under Test
Cable Insulation
6501
6503
Conductor Under Test
Fus
Figure 4-4
18 MegohmmeterModels6501/6503
Figures 4-5 shows the connections for testing insulation from asupplyconductortoground(motorframe).
Cable Insulation
Conductor Under Test
Figure 4-5
MegohmmeterModels6501/6503 19
Figure4-6showstheconnectionstoatransformer(lightingordis-tribution).Make sure that the switches and/or circuit breakers onbothsidesareopen.Checkthehighvoltagewindingtoground,lowvoltagetoground,andtheresistancebetweenthemwithnowindingground.
Tested Winding
TransformerGrounding Lug
Figure 4-6
20 MegohmmeterModels6501/6503
Figure4-7shows theconnections formeasuring the insulationofa three-phase line togroundby connecting the jumpersbetweenphases.Thisgivesareadingofallconductorsatonce.Ifaloadsuchasamotor,heater,etc.,isattachedtotheotherendoftheline,itwillreadtheloadresistancetogroundatthesametime.Byremovingthejumpers,readingscanbemadebetweentheindividualconduc-torsandground.
~~
~ ~~~
Jumpers
Figure 4-7
MegohmmeterModels6501/6503 21
4.7 Insulation Resistance on Motors
Figure4-8showsreadingtheresistancetogroundofathree-phasemotor winding. Since the three-phase motors are internally con-nected,it isonlynecessarytoconnectoneleadtothemotorleadandtheotherleadtothemotorframeasshown.
Figure 4-8
22 MegohmmeterModels6501/6503
Figure4-9showsthewindingsofathree-phasemotorseparated.Sometimesthiscanbedoneattheleadterminalswhileothertimestheendbellsmustberemovedtogetattheleadwiresofthecoils.By connecting themegohmmeter as shown, thephase insulationresistancevaluecannowbedetermined.Readbetweenphases“A”and“B”,then“B”and“C”,then“C”and“A”.
A
B
C
Figure 4-9
MegohmmeterModels6501/6503 23
Figure4-10showsconnectionsfortestinginsulationfromasupplyconductorinaswitchboxtoground(motorframe).Anidenticaltestmaybecarriedoutfromthemotorstarter.
Motor Side of Switch:Connection to One Leg
GroundedMotor Frame
Starter In
Figure 4-10
24 MegohmmeterModels6501/6503
CHAPTER 5
MAINTENANCE
5.1 Warning
• Formaintenanceuseonlyoriginalfactoryreplacementparts.
• Toavoidelectricalshock,donotattempttoperformanyservic-ingunlessyouarequalifiedtodoso.
• Do not perform any service while the Megohmmeter Model6501/6503isonanycircuit.
• Toavoidelectricalshockand/ordamageto the instrument,donotletwaterorotherforeignagentsintotheelectronicmodule.
Warning: IfthecrankspeedLEDisnotONduringtesting,thereadingisnotvalid.
5.2 Cleaning
Themegohmmetermaybegentlycleanedwithasoftcloth,soapandwater.Dryimmediatelyaftercleaning.Avoidwaterpenetrationintotheelectronicmodule.
Makesure themegohmmeterandall leadsaredrybefore furtheruse.
MegohmmeterModels6501/6503 25
5.3 Fuse Replacement
WARNING: Foryoursafety,foldthecrankanddisconnectalloftheleadsfromthemeter.
• Withaflatscrewdriveroracoin,pushandturnthefuseholderaquarterturncounter-clockwise.
• Replace the fuse with an identical one - 0.2A, 600V HPC6x32mm
26 MegohmmeterModels6501/6503
Repair and Calibration
Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.
For instrument repair and calibration:YoumustcontactourServiceCenterforaCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandpro-cessedpromptly.PleasewritetheCSA#ontheoutsideoftheshippingcontainer.Ifthe instrument is returned for calibration,weneed to know if youwant a standardcalibration,oracalibration traceable toN.I.S.T. (Includescalibrationcertificateplusrecordedcalibrationdata).
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone: (800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail: [email protected]
(Orcontactyourauthorizeddistributor)Costs forrepair,standardcalibration,andcalibrationtraceable toN.I.S.T.areavail-able.NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
If youareexperiencingany technicalproblems,or requireanyassistancewith theproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:
ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone: (800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail: [email protected] www.aemc.com
NOTE: Do not ship Instruments to our Foxborough, MA address.
MegohmmeterModels6501/6503 27
Limited Warranty
TheModels6501and6503arewarrantedtotheownerforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.
For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.
What AEMC® Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.
REGISTER ONLINE AT: www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair: First,requestaCustomerServiceAuthorizationNumber(CSA#)byphoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwiththesignedCSAForm.PleasewritetheCSA#ontheoutsideoftheshippingcon-tainer.Returntheinstrument,postageorshipmentpre-paidto:
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone: (800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail: [email protected]
Caution:Toprotectyourselfagainst in-transit loss,werecommendyouinsureyourreturnedmaterial.
NOTE: You must obtain a CSA# before returning any instrument.
010/15
99-MAN100328v7
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA
Phone:(603)749-6434•Fax:(603)742-2346www.aemc.com
INTRODUCTION1.1International Electrical Symbols1.2Definition of Measurement Categories1.3Receiving Your Shipment1.4Ordering Information1.4.1Accessories and Replacement Parts
PRODUCT FEATURES2.1Description2.2Features2.3Model 6501 Control Features2.4Model 6503 Control Features
SPECIFICATIONSOPERATION4.1Auto-ranging4.2Safety Check - Voltage Test 4.3Insulation Resistance Testing (MΩ Range)4.3.1Test Voltage4.3.2Spot Testing4.3.3Ratio Testing
4.4Successful Insulation Resistance Testing4.5Utilization of the Guard Terminal (Model 6503)4.6Insulation Measurement - Connections4.7Insulation Resistance on Motors
MAINTENANCE5.1Warning  5.2Cleaning5.3Fuse Replacement
Repair and CalibrationTechnical and Sales AssistanceLimited WarrantyWarranty Repairs