62
qN `i^ppf` båÖäáëÜ léÉê~íáåÖ fåëíêìÅíáçåë Operating Instructions 6399765.book Seite 1 Mittwoch, 4. Juli 2012 1:31 13

6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

  • Upload
    others

  • View
    10

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

qN=`i^ppf`

båÖäáëÜ

léÉê~íáåÖ=fåëíêìÅíáçåë

Operating Instructions

6399765.book Seite 1 Mittwoch, 4. Juli 2012 1:31 13

Page 2: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC

2 63 99 765 D3401.201.03.01.02

Table of contents1 Before you begin … ........................................................................................ 5

1.1 Structure of the document ............................................................................................. 71.2 Service life of Sirona instruments.................................................................................. 10

2 Safety instructions .......................................................................................... 113 Technical description ...................................................................................... 13

3.1 Task............................................................................................................................... 133.2 Functionality .................................................................................................................. 133.3 Structure of T1 CLASSIC .............................................................................................. 143.4 Technical data ............................................................................................................... 16

4 Preparation ..................................................................................................... 224.1 Initial start-up and longer breaks in use ........................................................................ 224.2 Prior to starting the work day......................................................................................... 22

6399765.book Seite 2 Mittwoch, 4. Juli 2012 1:31 13

Page 3: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC

63 99 765 D3401.201.03.01.02 3

4.3 Before each patient ....................................................................................................... 225 Operation ........................................................................................................ 24

5.1 Replacing the instrument............................................................................................... 255.2 Inserting and removing burr instruments....................................................................... 265.3 Inserting and removing an EVA/LTA/Per-io-Tor instrument .......................................... 325.4 Inserting and removing a prophylaxis instrument.......................................................... 335.5 Inserting and removing ENDO instruments................................................................... 355.6 Inserting and removing a Kirschner-Meyer instrument.................................................. 375.7 Adjusting the cooling spray ........................................................................................... 385.8 Plugging in and removing the spray clip for NaCI application ....................................... 39

6 Re-processing................................................................................................. 406.1 After each treatment session......................................................................................... 406.2 At the end of the work day............................................................................................. 40

7 Care and maintenance ................................................................................... 41

6399765.book Seite 3 Mittwoch, 4. Juli 2012 1:31 13

Page 4: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC

4 63 99 765 D3401.201.03.01.02

7.1 Spraying the mechanics ................................................................................................ 417.2 Maintaining the pushbutton chuck................................................................................. 447.3 Cleaning the fiber-optic cable surfaces ......................................................................... 467.4 Manually cleaning and disinfecting the surface............................................................. 467.5 Manually cleaning and disinfecting the spray channels ................................................ 487.6 Automated cleaning and disinfecting ............................................................................ 497.7 Cleaning the cooling spray nozzles............................................................................... 517.8 Sterilizing....................................................................................................................... 527.9 Testing the FG clamping system................................................................................... 547.10 Replacing the spray tube on the T1 CLASSIC KM 1.6 L............................................... 56

8 Spare parts and consumables ........................................................................ 579 Storage and transport conditions.................................................................... 5910 Disposal .......................................................................................................... 60

6399765.book Seite 4 Mittwoch, 4. Juli 2012 1:31 13

Page 5: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 1 Before you begin …

63 99 765 D3401.201.03.01.02 5

1 Before you begin …Before you begin …

Intended use TheT1 CLASSIC is used to treat dental diseases and dental trauma. The treatment involves the rotary or oscillatory processing of hard and soft tooth substance, bones, and dental prostheses (crowns, inlays, bridges, etc.).The green, blue, and PROPHY contra-angle handpieces are used for prophylaxis.The green, blue, and Meyer-Kirschner contra-angle handpieces are used for implantology and surgery.The handpiece and the green and blue contra-angle handpieces are rotary prosthetics.The EVA contra-angle handpiece is used for oscillatory prosthetics.The green, blue, and ENDO contra-angle handpieces are used for endodontics.

6399765.book Seite 5 Mittwoch, 4. Juli 2012 1:31 13

Page 6: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

1 Before you begin … Operating Instructions T1 CLASSIC

6 63 99 765 D3401.201.03.01.02

T1 CLASSICcomplies with the latest regulations reflecting state-of-the-art technology. T1 CLASSICmeets standardISO 7785 - 2.1. Read the operating instructions prior to using T1 CLASSIC.2. Only use theT1 CLASSICfor the applications described in the

Operating Instructions.3. Observe the applicable hygiene standards, occupational safety

regulations and accident prevention measures for T1 CLASSIC.Target group This product is intended only for use by trained dental personnel in

dental workstations and laboratories.Scope of validity of Operating Instructions These Operating Instructions apply to the following instruments:

● T1 CLASSIC from serial number 30001

6399765.book Seite 6 Mittwoch, 4. Juli 2012 1:31 13

Page 7: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 1 Before you begin …

63 99 765 D3401.201.03.01.02 7

1.1 Structure of the document1.1.1 Identification of warnings

Warnings ➢ To prevent any personal injury, please observe all warnings.Warnings can be identified as follows:

DANGER! indicates a danger leading to death or serious injury if not avoided.

WARNING! indicates a danger leading to death or serious injury if not avoided.

CAUTION! indicates a danger that may lead to injury if not avoided.Instructions for use ➢ To prevent material damage and additional expenses, please

observe all instructions for use. Instructions for use can be identified as follows:NOTICE! indicates measures for the prevention of material damage.

6399765.book Seite 7 Mittwoch, 4. Juli 2012 1:31 13

Page 8: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

1 Before you begin … Operating Instructions T1 CLASSIC

8 63 99 765 D3401.201.03.01.02

IMPORTANT: indicates important information and information on the avoidance of additional expenses.Tip: indicates information on making work easier.

6399765.book Seite 8 Mittwoch, 4. Juli 2012 1:31 13

Page 9: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 1 Before you begin …

63 99 765 D3401.201.03.01.02 9

1.1.2 Formats and charactersThe symbols and character formats used in the present manual have the following meaning:

✔ Prerequisite1. First action step2. Second action stepor

➢ Alternative action Result

Requests you to do something.

Use of formats and characters [ → 9].

Identifies a reference to another text passage and indicates the relevant page number.

● List Identifies a list.

6399765.book Seite 9 Mittwoch, 4. Juli 2012 1:31 13

Page 10: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

1 Before you begin … Operating Instructions T1 CLASSIC

10 63 99 765 D3401.201.03.01.02

1.1.3 Abbreviations/codes used

1.2 Service life of Sirona instrumentsWhen used as intended:● Non-moving parts of Sirona instruments have a typical service life of

approx. 5 years● Moving parts of Sirona instruments have a typical service life of

approx. 3 yearsNo warranty claim can be inferred here, as wear may occur earlier or later than indicated above depending on use, frequency of sterilization and frequency of maintenance.

FG Friction grip burr instrumentCA Contra-angle handpiece burr instrumentHP Straight handpiece burr instrument

6399765.book Seite 10 Mittwoch, 4. Juli 2012 1:31 13

Page 11: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 2 Safety instructions

63 99 765 D3401.201.03.01.02 11

2 Safety instructionsSafety instructions

Obligations of the user ➢ Only use operating resources in error free condition which do not deviate from the specified data [ → 16].

➢ Protect yourself, patients and others against any foreseeable dangers. In order to do so, follow the safety instructions.

➢ Use the equipment as intended.➢ You should always keep these operating instructions within reach for

further reference.Prevention of infection and cross-contamination Prevent infection and cross-contamination between patients, users, and

third parties as follows: Sterilize equipment after each patient.Take the appropriate hygiene measures e.g. wear protective gloves.

Emitted cooling air The cooling air emitted by the coupling of the motor must have a flow rate of 1.5 - 10 Nl/min.

6399765.book Seite 11 Mittwoch, 4. Juli 2012 1:31 13

Page 12: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

2 Safety instructions Operating Instructions T1 CLASSIC

12 63 99 765 D3401.201.03.01.02

Instrument head overheating If the instrument is defective, the area around the instrument head may heat up, thus creating a risk of burning the patient's oral mucosa.

Malfunction or damage Immediately discontinue use in the case of malfunction or damage. Damaged instruments may cause injury. Notify the dental depot or the manufacturer.

Repair Do not repair the instrument yourself.Spare parts and accessories Use only Sirona original or approved parts. Safe operation is not

guaranteed for parts that have not been approved by Sirona.

If you have any questions, please contact your dental depot or the manufacturer.

6399765.book Seite 12 Mittwoch, 4. Juli 2012 1:31 13

Page 13: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 3 Technical description

63 99 765 D3401.201.03.01.02 13

3 Technical descriptionTechnical description

3.1 TaskThe T1 CLASSIC instrument transmits the driving power and speed of the electric motor to the preparation tool.

3.2 FunctionalityThe transmission ratio of the instrument distorts the operating speed of the preparation tool.The rotary drive movement is converted into a vertical reciprocating movement of 0.4 or 1.1 mm, respectively, with the EVA contra-angle handpiece.The rotary drive movement is converted into a clockwise-counterclockwise movement with the ENDO contra-angle handpiece.

6399765.book Seite 13 Mittwoch, 4. Juli 2012 1:31 13

Page 14: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

3 Technical description Operating Instructions T1 CLASSIC

14 63 99 765 D3401.201.03.01.02

3.3 Structure of T1 CLASSICContra-angle handpiece

FDC

BA

E

A Instrument head (in this case: standard)B PushbuttonC Cooling spray apertureD Opening of chuck systemE Light apertureF Color coding

6399765.book Seite 14 Mittwoch, 4. Juli 2012 1:31 13

Page 15: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 3 Technical description

63 99 765 D3401.201.03.01.02 15

Handpiece

DCB

AA Opening for the clamping systemB Cooling spray apertureC Light apertureD Handpiece center part

6399765.book Seite 15 Mittwoch, 4. Juli 2012 1:31 13

Page 16: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

3 Technical description Operating Instructions T1 CLASSIC

16 63 99 765 D3401.201.03.01.02

3.4 Technical dataContra-angle handpiece

T1 CLASSIC S 200 L/S 200 L mini

S 40 L S 6 L S 1.6 L

Transmission ratio 1:5 1:1 6:1 24:1Color coding for transmission ratio Red Blue Green GreenMaximum motor speed in rpm 40000 40000 40000 40000Maximum operating speed in rpm 200000 40000 6700 1700Pushbutton cover clamping system FG CA CA CAInternal cooling media x x x x

6399765.book Seite 16 Mittwoch, 4. Juli 2012 1:31 13

Page 17: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 3 Technical description

63 99 765 D3401.201.03.01.02 17

Light function x x x xInstrument coupling Sirona Sirona Sirona Sirona

T1 CLASSIC S 200 L/S 200 L mini

S 40 L S 6 L S 1.6 L

6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13

Page 18: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

3 Technical description Operating Instructions T1 CLASSIC

18 63 99 765 D3401.201.03.01.02

Handpiece

T1 CLASSIC S H40 LTransmission ratio 1:1Color coding for transmission ratio BlueMaximum motor speed in rpm 40000Maximum operating speed in rpm 40000Pushbutton cover clamping system CA/HPInternal cooling media xLight function xInstrument coupling Sirona

6399765.book Seite 18 Mittwoch, 4. Juli 2012 1:31 13

Page 19: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 3 Technical description

63 99 765 D3401.201.03.01.02 19

Special contra-angle handpiece

T1 CLASSIC PROPHY EVA 04 L/11 L ENDO L KM 1.6 LTransmission ratio 2.4:1 - - 24:1Maximum motor speed in rpm 12000 14000 15000 40000Maximum operating speed/cycles in rpm 5000 Stroke rate

0.4/1.1 mm/6000

1600 1670

Pushbutton cover clamping system Screw & snap - CA/HP CAInternal cooling media x x - -Light function x x x xInstrument coupling Sirona Sirona Sirona Sirona

6399765.book Seite 19 Mittwoch, 4. Juli 2012 1:31 13

Page 20: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

3 Technical description Operating Instructions T1 CLASSIC

20 63 99 765 D3401.201.03.01.02

Burr instruments

FG CA HPShank diameter in mm 1.590 - 1.600 2.334 - 2.350 2.334 - 2.350Maximum total length in mm 25 34 50Maximum working diameter in mm 2.1 - -Standard ISO 1797-1

ISO 2157ISO 1797-1 ISO 1797-1

6399765.book Seite 20 Mittwoch, 4. Juli 2012 1:31 13

Page 21: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 3 Technical description

63 99 765 D3401.201.03.01.02 21

Dental Treatment Center

T1 CLASSICSpray air pressure in bar 2.7Spray water pressure in bar 2.0Maximum water temperature in °C 40Recommended water content in spray in mL/min > 50

6399765.book Seite 21 Mittwoch, 4. Juli 2012 1:31 13

Page 22: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

4 Preparation Operating Instructions T1 CLASSIC

22 63 99 765 D3401.201.03.01.02

4 PreparationPreparation

4.1 Initial start-up and longer breaks in use➢ Sterilize the instrument and accessories prior to startup.➢ Clean and maintain the instrument after longer breaks in use.

4.2 Prior to starting the work day➢ Purge the water paths and air channels for 20 - 30 seconds.

4.3 Before each patient1. Purge the water paths and air channels for 20 - 30 seconds.2. CAUTION! Stop the motor.

Attach the instrument [ → 25].3. Insert the burr instrument [ → 26].4. Use a sufficient amount of cooling water (> 50 ml/min) [ → 38].

6399765.book Seite 22 Mittwoch, 4. Juli 2012 1:31 13

Page 23: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 4 Preparation

63 99 765 D3401.201.03.01.02 23

5. Operate only with filtered water (< 50 µm). 6. Check the nozzles for blockages and deposits of lime, for example,

and clean the nozzles if necessary [ → 51]. CAUTION! Insufficient cooling leads to overheating of the cavity and

damage to the tooth substance. Ensure that the water content is > 50ml/min.

6399765.book Seite 23 Mittwoch, 4. Juli 2012 1:31 13

Page 24: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

24 63 99 765 D3401.201.03.01.02

5 OperationOperation

NOTICE! Use only sharp and undamaged burrs and diamond grinders. The burrs and diamond grinders must be clean to prevent dirt from being carried into the clamping system.

CAUTION! A loose or partially removed burr instrument can detach itself from the head or break off. This leads to a risk of injury! Therefore, only use the instrument if the burr instrument is at least 10 mm deep and clamped securely in place.

CAUTION! The handpiece must only be operated with the clamping system closed.

CAUTION! Insufficient cooling leads to overheating of the cavity and damage to the tooth substance. Ensure that the water content is > 50ml/min.

6399765.book Seite 24 Mittwoch, 4. Juli 2012 1:31 13

Page 25: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 25

CAUTION! Do not pull the patient's cheek back with the contra-angle handpiece while the motor is running. This would actuate the pushbutton, thus creating a risk of burning the patient's oral mucosa.

CAUTION! Always operate the contra-angle handpieces with cooling spray when marked with a red dot!IMPORTANT: For application details and operating data of the different preparation tools, please refer to the information provided by the manufacturer.

5.1 Replacing the instrument CAUTION! The instrument should only be fitted or removed when

the motor is at standstill.

6399765.book Seite 25 Mittwoch, 4. Juli 2012 1:31 13

Page 26: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

26 63 99 765 D3401.201.03.01.02

Attaching the T1 CLASSIC✔ The motor has come to a stop.1. Insert the instrument until it snaps into place.2. Rotate the instrument until it clicks into place.Removing the T1 CLASSIC✔ The motor has come to a stop.➢ Pull off the instrument. Do not pull on the supply hose while doing

this.

5.2 Inserting and removing burr instrumentsIMPORTANT: Check the pushbutton to make sure it moves freely.NOTICE! Use only sharp and undamaged burrs and diamond grinders. The burrs and diamond grinders must be clean to prevent dirt from being carried into the clamping system.

6399765.book Seite 26 Mittwoch, 4. Juli 2012 1:31 13

Page 27: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 27

CAUTION! A loose or partially removed burr instrument can detach itself from the head or break off. This leads to a risk of injury! Therefore, only use the instrument if the burr instrument is at least 10 mm deep and clamped securely in place.

CAUTION! The handpiece must only be operated with the clamping system closed.

6399765.book Seite 27 Mittwoch, 4. Juli 2012 1:31 13

Page 28: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

28 63 99 765 D3401.201.03.01.02

Inserting the friction grip burr instrument✔ The motor has come to a stop.1. Press the pushbutton and insert the burr instrument up to the stop.2. Then pull the burr instrument to check that it is firmly seated.Removing the friction grip burr instrument✔ The burr instrument has come to a stop.➢ Remove it while pressing and holding down the pushbutton.Contra-angle handpiece burr instrument✔ The motor has come to a stop.1. Insert the burr instrument without pressing the pushbutton.2. Turn the burr instrument back and forth gently until it snaps into

place.3. Then pull and turn the burr instrument to check that it is firmly seated.

6399765.book Seite 28 Mittwoch, 4. Juli 2012 1:31 13

Page 29: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 29

Removing the contra-angle handpiece burr instrument✔ The burr instrument has come to a stop.➢ Remove it while pressing and holding down the pushbutton.Inserting the handpiece burr instrument✔ The motor has come to a stop.1. Turn the center part of the handpiece in the direction of the arrow (A)

up to the stop.2. When using the contra-angle handpiece burr instrument (C): insert

the pin (B) into the handpiece, with the slotted end first. The pin compensates for the difference in length between straight and contra-angle handpiece burr instruments.

3. Insert the burr instrument as far as it will go.

B C

A

6399765.book Seite 29 Mittwoch, 4. Juli 2012 1:31 13

Page 30: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

30 63 99 765 D3401.201.03.01.02

4. Clamp the burr instrument by turning the center part of the handpiece in direction (D) up to the stop. When the two marks (E) are located opposite each other, the burr

instrument is clamped. When using handpiece burr instruments with detached shaft:

ensure that the detached shaft is not lying against the chuck opening (F).

5. Then pull and turn the burr instrument to check that it is firmly seated.

E F

D

6399765.book Seite 30 Mittwoch, 4. Juli 2012 1:31 13

Page 31: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 31

Removing the pin1. Screw in the supplied tool (A).2. Open the clamping system.3. Pull out the pin.A

6399765.book Seite 31 Mittwoch, 4. Juli 2012 1:31 13

Page 32: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

32 63 99 765 D3401.201.03.01.02

5.3 Inserting and removing an EVA/LTA/Per-io-Tor instrumentInserting the EVA/ LTA/ Per-io-Tor instrument➢ Push in the instrument as far as it will go.

In the case of EVA contra-angle handpieces, the instrument can be freely rotated in the first detent position (A). In the second detent position (B), 10 fixed angle positions can be selected.

Removing the EVA/LTA/Per-io-Tor instrument➢ Pull the instrument out by hand.

If this proves difficult: use the ejector (C) included in the scope of supply.BA

C

6399765.book Seite 32 Mittwoch, 4. Juli 2012 1:31 13

Page 33: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 33

5.4 Inserting and removing a prophylaxis instrumentIMPORTANT: Operate the contra-angle handpiece only with clockwise rotation. The contra-angle handpiece is protected by a special seal system against penetration of polishing paste.Inserting a prophylaxis instrument1. Pull the contra-angle handpiece off of the motor.2. Insert the PROPHY spray adapter (D) into the contra-angle

handpiece until it snaps into place. Hold the driver securely in the contra-angle handpiece with the PROPHY spray adapter.

3. Screw the screw plug (A)/screw-in adapter (C) firmly into the head of the contra-angle handpiece.

4. If you use a screw-in adapter, let the tool (B) snap onto it.

A

B

C D

6399765.book Seite 33 Mittwoch, 4. Juli 2012 1:31 13

Page 34: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

34 63 99 765 D3401.201.03.01.02

Removing a prophylaxis instrument1. Pull the contra-angle handpiece off of the motor.2. Insert the PROPHY spray adapter into the contra-angle handpiece

until it snaps into place. Hold the driver securely in the contra-angle handpiece with the PROPHY spray adapter.

3. If you use a snap-on cap, pull the tool from the screw-in adapter.4. Unscrew the screw plug/screw-in adapter.

6399765.book Seite 34 Mittwoch, 4. Juli 2012 1:31 13

Page 35: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 35

5.5 Inserting and removing ENDO instrumentsThe contra-angle handpiece is suitable for holding both straight and contra-angle handpiece instruments.Inserting ENDO instruments✔ The motor has come to a stop.1. When using the contra-angle handpiece instrument: press the

pushbutton and insert the adapter (A) into the head as far as it will go.

2. While pressing and holding down the pushbutton, insert the ENDO instrument as far as it will go.

3. Then pull and turn the ENDO instrument to check that it is firmly seated.

Removing ENDO instruments✔ The motor has come to a stop.

Ø 3,6 – 4 mm

Ø 2,35 mm

B

A

6399765.book Seite 35 Mittwoch, 4. Juli 2012 1:31 13

Page 36: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

36 63 99 765 D3401.201.03.01.02

1. Pull the ENDO instrument out while pressing and holding down the pushbutton.

2. When using the contra-angle handpiece instrument: press down the pushbutton and pull out the adapter by its collar (B).

Attaching the spray clip for rinsing and disinfecting✔ The spray water supply at the motor or treatment center is turned off.1. Attach the silicone hose (D) to the spray clip (C).2. Plug the spray clip into the contra-angle handpiece.Removing the spray clip for rinsing and disinfecting➢ Remove the spray clip from the contra-angle handpiece.

DC

6399765.book Seite 36 Mittwoch, 4. Juli 2012 1:31 13

Page 37: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 37

5.6 Inserting and removing a Kirschner-Meyer instrumentInserting a Kirschner-Meyer instrument✔ The motor has come to a stop.1. Slide the deflection disk (A) onto the shaft of the burr instrument. The

deflection disk prevents the air coming out of the head from entering the open wound.

2. Insert the burr instrument without pressing the pushbutton.3. Turn the burr instrument back and forth gently until it snaps into

place. 4. Push the deflection disk up to the head.5. Then pull and turn the burr instrument to check that it is firmly seated.Removing a Kirschner-Meyer instrument✔ The burr instrument has come to a stop.➢ Remove it while pressing and holding down the pushbutton.

A

6399765.book Seite 37 Mittwoch, 4. Juli 2012 1:31 13

Page 38: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

5 Operation Operating Instructions T1 CLASSIC

38 63 99 765 D3401.201.03.01.02

Connecting the hose for the NaCl solution➢ Attach the silicone hose (B) to the spray tube (A) for the hollow burr.Additionally, you can cool the burr externally with NaCl solution.1. Plug the spray clip (D) in [ → 39].2. Connect the silicone hoses (B and E) using the distributor (C), as

shown in the illustration.

5.7 Adjusting the cooling spray➢ Adjust the flow rate of the cooling water using the control ring (A) (>

50 ml/min).Tip: The cooling water can be measured using a measuring cup and watch.Water flow rateThe maximum water flow is set when the two marks are located opposite one another.

A

D

CB

E

max

A

6399765.book Seite 38 Mittwoch, 4. Juli 2012 1:31 13

Page 39: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 5 Operation

63 99 765 D3401.201.03.01.02 39

5.8 Plugging in and removing the spray clip for NaCI applicationPlugging in the spray clip✔ The spray water supply at the motor or treatment center is turned off.1. Plug the silicone hose (D) into the tubule on the spray clip for saline

solutions (B/C).2. Plug the spray clip into the instrument.Tip:Use the silicone disks (A), to hold back the cooling air emitted from the contra-angle handpiece from the preparation site and to direct it away laterally.Removing the spray clip1. Remove the spray clip from the instrument.2. Flush out the spray clip and silicone hose with clear water.

A B C D

6399765.book Seite 39 Mittwoch, 4. Juli 2012 1:31 13

Page 40: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

6 Re-processing Operating Instructions T1 CLASSIC

40 63 99 765 D3401.201.03.01.02

6 Re-processingRe-processing

6.1 After each treatment session1. Clean and disinfect the instrument and accessories.2. Apply spray to the instrument [ → 41].3. Sterilize the instrument and accessories [ → 52].

6.2 At the end of the work day➢ Apply spray to the instrument [ → 41].NOTICE! Do not leave any instruments on the motor overnight, in order to prevent oil from penetrating into the electric motor. Never lubricate the electric motor.

6399765.book Seite 40 Mittwoch, 4. Juli 2012 1:31 13

Page 41: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 41

7 Care and maintenanceCare and maintenance

Elastomers, e.g. O-rings, must be replaced depending on their degree of wear.For safety and technical reasons, check the clamping system of the contra-angle and straight handpiece burr instruments on an annual basis.Check the clamping system of friction grip burr instruments on a monthly basis.

7.1 Spraying the mechanicsIntervals● At least daily at noon and in the evening● Prior to sterilization● After thermodisinfection

6399765.book Seite 41 Mittwoch, 4. Juli 2012 1:31 13

Page 42: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

42 63 99 765 D3401.201.03.01.02

Required accessories

NOTICE! Using sprays from other manufacturers can reduce the product's service life. Only use Sirona T1 spray.Procedure✔ The motor has come to a stop.1. Remove the burr instrument [ → 26].2. Pull the instrument off the motor [ → 25].3. Fit the spray adapter onto the nozzle of the spray can.

A

BC

A Spray adapter, CLASSICB Spray adapter, PROPHYC T1 Spray

6399765.book Seite 42 Mittwoch, 4. Juli 2012 1:31 13

Page 43: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 43

4. Insert the instrument until it snaps into place and hold it firmly.5. Apply spray to the instrument for 1 - 2 seconds.

IMPORTANT: Hold the spray can upright.Check1. Check whether the liquid escaping from the instrument head is

clean.2. If the liquid is dirty: repeat the spraying procedure. The liquid is still not clean?1. Attach the instrument to the motor and let it run briefly. In this way,

the spray can be distributed more effectively.2. CAUTION! Stop the motor.

Wipe off any leaking oil with a dry cloth.3. Repeat the spraying procedure.

6399765.book Seite 43 Mittwoch, 4. Juli 2012 1:31 13

Page 44: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

44 63 99 765 D3401.201.03.01.02

7.2 Maintaining the pushbutton chuckTo remove deposits and ensure proper functioning of the clamping system, the push-button chuck must be maintained using T1 spray.IntervalCarry out maintenance work on the pushbutton chuck at least once a week.Required accessoriesT1 Spray

6399765.book Seite 44 Mittwoch, 4. Juli 2012 1:31 13

Page 45: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 45

Procedure1. Press the contra-angle handpiece head with the chuck firmly against

the spray can nozzle.2. Spray the chuck for 1 - 2 seconds.

IMPORTANT:Hold the spray can upright.3. Wipe off any leaking protective oil with a lint-free cloth.

6399765.book Seite 45 Mittwoch, 4. Juli 2012 1:31 13

Page 46: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

46 63 99 765 D3401.201.03.01.02

7.3 Cleaning the fiber-optic cable surfaces1. In order to avoid scratching the surfaces (A), blow off any dirt

particles with air using a dry syringe.2. Wipe the surfaces with a Q-tip or a soft cloth and alcohol.

7.4 Manually cleaning and disinfecting the surface✔ Wear appropriate protective clothing.1. Remove the burr instrument [ → 26].2. Automated processing is preferable to manual.

A

6399765.book Seite 46 Mittwoch, 4. Juli 2012 1:31 13

Page 47: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 47

IMPORTANT: Manual cleaning must always be combined with disinfection.

7.4.1 Cleaning the outer surface1. Clean the instrument by brushing it off under running water (< 38°C,

< 100°F, at least drinking water quality).2. Blow the instrument out with max. 3 bar.3. Apply spray to the instrument [ → 41].

6399765.book Seite 47 Mittwoch, 4. Juli 2012 1:31 13

Page 48: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

48 63 99 765 D3401.201.03.01.02

7.4.2 Disinfecting the surface✔ The disinfectants that are approved in your country must have

proven bactericidal, fungicidal and virucidal properties.1. Spray the surface with disinfectant.2. Wipe the disinfectant away using a cloth.

7.5 Manually cleaning and disinfecting the spray channels➢ Clean and disinfect the spray channels using suitable equipment

and the corresponding adapters in accordance with the manufacturer specifications.

6399765.book Seite 48 Mittwoch, 4. Juli 2012 1:31 13

Page 49: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 49

7.6 Automated cleaning and disinfecting ...7.6.1 ... with a DAC UNIVERSAL

We recommend using Sirona DAC UNIVERSAL for automated cleaning, disinfection, and care.For further details, refer to the operating instructions supplied with the unit.➢ Manually maintain the pushbutton chuck [ → 44].

7.6.2 ... with cleaning and disinfection equipmentThe instrument can also be cleaned and disinfected in a suitable piece of cleaning and disinfection equipment.

DAC Universal

DAC Universal

6399765.book Seite 49 Mittwoch, 4. Juli 2012 1:31 13

Page 50: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

50 63 99 765 D3401.201.03.01.02

The cleaning and disinfection equipment used must be approved by its manufacturer for the cleaning and disinfection of dental instruments and comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding time).For further details, refer to the operating instructions supplied with the unit.1. Check whether the instrument is clean after processing.2. If any contamination is present: repeat the procedure. For further processing, the instrument should be dry and free

from any residues.3. Blow the instrument out with max. 3 bar.4. Manually maintain the mechanics [ → 41].5. Manually maintain the pushbutton chuck [ → 44].

95°C

203°F

6399765.book Seite 50 Mittwoch, 4. Juli 2012 1:31 13

Page 51: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 51

7.7 Cleaning the cooling spray nozzlesIf your tap water is very hard, lime deposits may constrict or completely block the cooling spray nozzles.1. Carefully clean the nozzle openings by running a cleaning wire

through them.2. Briefly operate the instrument with cooling spray.

!

6399765.book Seite 51 Mittwoch, 4. Juli 2012 1:31 13

Page 52: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

52 63 99 765 D3401.201.03.01.02

7.8 Sterilizing1. Clean and disinfect the instrument [ → 46].2. Apply spray to the instrument [ → 41].3. Sterilize the instrument in the steam sterilizer with saturated water

vapor.

134°C 274°F

Saturated water vapor: Temperature: 134°C (274°F)Overpressure: 2.04 bar (29.59 psi)

Holding time: 3 min

6399765.book Seite 52 Mittwoch, 4. Juli 2012 1:31 13

Page 53: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 53

Steam sterilizers that fulfill the requirements of either EN 13060, class B (for example, DAC PROFESSIONAL) or EN 13060, class S and are

suitable for the sterilization of straight/contra-angle handpieces are approved.NOTICE! Do not exceed 140°C (284°F), even during the drying phase.After sterilizing1. Remove the instrument from the steam sterilizer immediately.

CAUTION! The instrument will be hot. There is a risk of burning!NOTICE! Do not accelerate the cooling process by dipping the instrument into cold water. This will damage your instrument.

2. Store all instruments so that they are protected from contamination.3. Sterilize again once the storage period has elapsed.

The Hygiene Division of SironaThe Hygiene Division of Sirona

ProfessionalDAC

P

6399765.book Seite 53 Mittwoch, 4. Juli 2012 1:31 13

Page 54: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

54 63 99 765 D3401.201.03.01.02

7.9 Testing the FG clamping systemIntervalTest the FG clamping system at least once a month.Required accessoriesChuck tester

6399765.book Seite 54 Mittwoch, 4. Juli 2012 1:31 13

Page 55: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 7 Care and maintenance

63 99 765 D3401.201.03.01.02 55

Procedure✔ The expiry date of the chuck tester (A) has not elapsed.1. Insert the chuck tester into the FG clamping system [ → 26].2. Pull on the chuck tester until the marking ring appears (withdrawal

force: 22 N).Does the chuck tester slide out of the chuck before the marking ring appears?

CAUTION! The chuck is defective and there can be no guarantee that the burr instrument will be held securely in place. This leads to a risk of injury!1. Take the product out of use. 2. Have the chuck replaced by a technician who is authorized by Sirona

to perform this task.Tip: Keep a record of the time and result of the inspection for your own monitoring purposes.

A

6399765.book Seite 55 Mittwoch, 4. Juli 2012 1:31 13

Page 56: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

7 Care and maintenance Operating Instructions T1 CLASSIC

56 63 99 765 D3401.201.03.01.02

7.10 Replacing the spray tube on the T1 CLASSIC KM 1.6 LRemoving the spray tube1. Disengage the spray tube (B) from the bracket (A).2. Remove the spray tube from the instrument head.Inserting the spray tube1. Plug the spray tube into the instrument head.2. Engage the spray tube in the bracket.

A B

6399765.book Seite 56 Mittwoch, 4. Juli 2012 1:31 13

Page 57: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 8 Spare parts and consumables

63 99 765 D3401.201.03.01.02 57

8 Spare parts and consumablesSpare parts and consumables

Use only Sirona original or approved parts.

Order No.: Order No.:P H40 straight handpiece – Pin for CA burr 89 17 866 NaCl spray tube 89 24 896P H40 – Tool for pin (CA burr) 89 17 874 NaCl distributor 54 55 7333-nozzle service set (with tools) 58 66 749 Silicone disks for contra-angle handpiece,

dia. 2.35 mm41 76 223

3-nozzle spray ring set (with spare parts) 58 66 715 Silicone disks for contra-angle handpiece, dia. 1.6 mm

41 76 215

Tool set for spray ring replacement 62 59 605 Ejector for EVA/ LTA/ Per-io-Tor instruments 41 77 056Chuck tester 33 27 793 ENDO adapter for contra-angle handpiece

instruments58 69 669

6399765.book Seite 57 Mittwoch, 4. Juli 2012 1:31 13

Page 58: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

8 Spare parts and consumables Operating Instructions T1 CLASSIC

58 63 99 765 D3401.201.03.01.02

NaCl spray clip C for contra-angle handpiece 41 74 087 Screw-in adapter, prophylaxis (snap-on) 89 25 679NaCl spray clip H for straight handpiece 41 74 079 T1 Spray (6x 250 ml cans) 59 01 665NaCl silicone hose, thick, 1.8 m long 89 23 658 NaCl coupling for silicone hoses 58 82 618NaCl silicone hose, thin, 200 mm long, 6 pcs. 59 17 419 Cleaning wire for spray nozzles 24 00 232Spray adapter, PROPHY 59 10 984 Spray adapter, T1 CLASSIC 59 53 930

Order No.: Order No.:

6399765.book Seite 58 Mittwoch, 4. Juli 2012 1:31 13

Page 59: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

Operating Instructions T1 CLASSIC 9 Storage and transport conditions

63 99 765 D3401.201.03.01.02 59

9 Storage and transport conditionsStorage and transport conditions

+70°C

-40°C

95%

max.

10%

min.50

106

6399765.book Seite 59 Mittwoch, 4. Juli 2012 1:31 13

Page 60: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

10 Disposal Operating Instructions T1 CLASSIC

60 63 99 765 D3401.201.03.01.02

10 DisposalDisposal

● To the best of our current knowledge, the product does not contain any substances which are hazardous to the environment.

● Disinfect the product prior to disposal.● Please observe the disposal regulations applicable in your country.

6399765.book Seite 60 Mittwoch, 4. Juli 2012 1:31 13

Page 61: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

6399765.book Seite 61 Mittwoch, 4. Juli 2012 1:31 13

Page 62: 6399765 T...Instrument coupling Sirona Sirona Sirona Sirona T1 CLASSIC S 200 L/ S 200 L mini S 40 L S 6 L S 1.6 L 6399765.book Seite 17 Mittwoch, 4. Juli 2012 1:31 13 3 Technical description

0123

tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK

aPQMNKOMNKMPKMNKMO= MTKOMNO ûKJkêKW=MMM=MMM

páêçå~=aÉåí~ä=póëíÉãë=dãÄe in USA: páêçå~=aÉåí~ä=póëíÉãëI=ii`c~Äêáâëíê~≈É=PN QUPR=páêçå~=aêáîÉI=pìáíÉ=NMMaJSQSOR=_ÉåëÜÉáã `Ü~êäçííÉI=k`=OUOTPdÉêã~åó ïïïKëáêçå~KÅçã rp^ ïïïKëáêçå~KÅçã

SP=VV=TSR=aPQMN=

6399765.book Seite 62 Mittwoch, 4. Juli 2012 1:31 13