76
86 6.1 PREFORMULATION STUDIES Characterization of drug A) Appearance LFX was found to be pale yellow crystalline, odourless powder; OFX as white to pale yellow powder and NFX as pale yellow crystalline, odourless powder as reported in literature. B) Solubility LFX was found to be freely soluble in water and in ethyl alcohol. It was also soluble in 2-propanol and acetone; OFX was found to be freely soluble in acetic acid, slightly soluble in water, methanol, ethanol and acetone; NFX was found to be slightly soluble in water and alcohol. C) Thermal analysis i) Melting point Melting point of LFX, OFX and NFX was found to be in the range of 214-215 o C, 252-254ºC and 220-221 o C respectively as reported in the literature, thus indicating purity of the drug samples. ii) Differential scanning calorimetry (DSC) The DSC thermogram of LFX, OFX and NFX showed a sharp endothermic peak at 214.5°C, 251.5°C and 220.5°C respectively corresponding to their melting points, as shown in Fig. 6.1, 6.2 and 6.3 respectively.

6.1 PREFORMULATION STUDIES Characterization of drug A ...shodhganga.inflibnet.ac.in/bitstream/10603/16358/12/12_chapter 6.pdf · 6.1 PREFORMULATION STUDIES ... Thermal analysis i)

  • Upload
    others

  • View
    20

  • Download
    0

Embed Size (px)

Citation preview

86

6.1 PREFORMULATION STUDIES

Characterization of drug

A) Appearance

LFX was found to be pale yellow crystalline, odourless powder; OFX as white to

pale yellow powder and NFX as pale yellow crystalline, odourless powder as reported

in literature.

B) Solubility

LFX was found to be freely soluble in water and in ethyl alcohol. It was also

soluble in 2-propanol and acetone; OFX was found to be freely soluble in acetic acid,

slightly soluble in water, methanol, ethanol and acetone; NFX was found to be

slightly soluble in water and alcohol.

C) Thermal analysis

i) Melting point

Melting point of LFX, OFX and NFX was found to be in the range of 214-215oC,

252-254ºC and 220-221oC respectively as reported in the literature, thus indicating

purity of the drug samples.

ii) Differential scanning calorimetry (DSC)

The DSC thermogram of LFX, OFX and NFX showed a sharp endothermic peak

at 214.5°C, 251.5°C and 220.5°C respectively corresponding to their melting points,

as shown in Fig. 6.1, 6.2 and 6.3 respectively.

87

Fig. 6.1 DSC of LFX sample

Fig. 6.2 DSC of OFXsample

88

Fig. 6.3 DSC of NFX sample

D) Spectral analysis

i) Infrared spectroscopy

The IR spectrum of the pure LFX, OFX and NFX sample recorded by FTIR

spectrometer is shown in Fig.6.4 to 6.6 which were compared with standard

functional group frequencies of LFX, OFX and NFX as shown in Table 6.1 to 6.3.

89

Fig. 6.4 IR spectra of LFX sample

Table 6.1 IR spectral data of LFX

Sr.No. Group assignment Wave number (cm-1

)

1 COOH (O-H stretching) 3271

2 -CH3

2935

3 C=O (C=O stretching) 1724

4 H3C-N (C-N stretching) 1292

5 F (C-F stretching) 1089

6 Benzene ring 1495

40060080010001200140016001800200024002800320036004000

1/cm

0

15

30

45

60

75

90

%T

30

66

.61

30

47

.32

29

62

.46

26

23

.01

23

64

.57

17

24

.24

17

04

.96

16

20

.09

15

66

.09

15

27

.52

14

85

.09

14

50

.37

14

19

.51

14

00

.22

13

61

.65

13

11

.50

12

53

.64

11

95

.78

11

68

.78

11

34

.07

11

18

.64

10

87

.78

10

60

.78

97

2.0

6

93

3.4

8

87

5.6

2

80

6.1

9

75

6.0

4

70

9.7

6

48

6.0

3

40

5.0

2

D2

90

Fig. 6.5 IR spectra of OFX sample

Table 6.2 IR spectral data of OFX

Sr.No. Group assignment Wave number (cm-1

)

1 N-H stretch 3124

2 C=O Stretch 1716

3 CH Plane deformation 1245

4 CH out of deformation 813

5 C –F stretch 1028

6 Benzene ring 1495

91

Fig. 6.6 IR spectra of NFX sample

Table 6.3 IR spectral data of NFX

Sr.

No.

Group assignment Wave number (cm-1

)

1 C=O stretch of carboxylic group 1624.73

2 C-N stretch of tertiary group 1269.9

3 C-H stretch of aliphatic group 2369.12

4 N-H stretch of secondary group 2932.23

5 C out of plane 786.90

6 C-F stretching 1091.63

7 C=C stretching 1452.72

40060080010001200140016001800200024002800320036004000

1/cm

-15

0

15

30

45

60

75

90

%T

34

38

.84

34

21

.48

33

61

.69

29

68

.24

29

20

.03

28

87

.24

28

44

.81

28

39

.02 2

77

3.4

5

26

78

.94

25

51

.65

23

58

.78

18

41

.89

17

93

.68

17

30

.03

16

18

.17

15

85

.38

15

54

.52

15

37

.16

15

00

.52

14

88

.94

14

54

.23

13

86

.72

13

44

.29

12

72

.93

12

40

.14

11

82

.28

11

30

.21

10

91

.63

10

31

.85

96

2.4

1

93

1.5

5

90

2.6

2

82

3.5

5

78

4.9

77

69

.54

73

8.6

9

70

0.1

1

66

1.5

4

62

2.9

65

92

.11

56

7.0

35

47

.75

52

0.7

4

44

9.3

84

20

.45

S 13

92

ii) UV Spectroscopy

The solutions of LFX, OFX and NFX sample were scanned from 200-400 nm, a

wavelength maxima was found to be 288 nm, 293nm and 271nm resp. in simulated

tear fluid (Fig. 6.7 to 6.9)

Fig. 6.7 Spectra of LFX in simulated tear fluid

Fig. 6.8 Spectra of OFX in simulated tear fluid

93

Fig. 6.9 Spectra of NFX in simulated tear fluid

Compatibility studies

Preformulation studies were carried out to study the compatibility of pure drug

LFX, OFX and NFX with the polymers like PXM, sodium alginate, gellan gum, CP

974P, chitosan and HPMC K4M/HEC etc. prior to the preparation of ophthalmic IS

hydrogel. The individual IR spectra of the polymers (Fig. 6.10 to 6.17), as well as the

combination spectra of drug and polymer are shown (Fig. 6.18 to 6.29). The results

indicate no interaction between drug and polymers when compared with infrared

spectrum of pure drug (Fig.6.4 to 6.6). It is confirmed by absence of additional peaks

and all principal absorption peaks are retained in the combination spectra. Hence all

excipients were found to be compatible with the drug. DSC of physical mixtures of

optimized thermosensitive formulations of each drug showed a characteristic

endothermic peak of drug (Fig. 6.30 to 6.32).

94

Fig. 6.10 IR spectra PXM 188

Fig. 6.11 IR spectra of PXM 407

40060080010001200140016001800200024002800320036004000

1/cm

0

15

30

45

60

75

90

%T

35

62

.28

35

41

.06

35

21

.78

34

96

.70

34

81

.27

34

42

.70

34

21

.48

33

98

.34

29

20

.03

28

73

.74

23

58

.78

17

33

.89

17

14

.60

16

97

.24

16

47

.10

16

29

.74

16

22

.02

15

56

.45

15

39

.09

14

56

.16

14

15

.65

13

92

.51

13

73

.22

13

52

.01

12

98

.00

12

51

.72

10

95

.49

10

22

.20

94

8.9

1

84

6.6

9

66

9.2

5 65

1.8

9

57

8.6

0

54

9.6

7

46

8.6

74

53

.24

42

2.3

8

S 7

95

Fig. 6.12 IR spectra of chitosan

Fig. 6.13 IR spectra of HPMC K4M

40060080010001200140016001800200024002800320036004000

1/cm

0

15

30

45

60

75

90

%T

35

00

.56

34

77

.42

34

42

.70

34

19

.56

23

58

.78

17

12

.67

16

97

.24

16

66

.38

16

49

.02 1

63

3.5

91

62

0.0

9 15

56

.45

15

39

.09 1

51

9.8

0

14

69

.66

14

56

.16

14

17

.58

13

92

.51 1

33

8.5

1

13

17

.29

10

22

.20

94

5.0

5

90

0.7

0

71

9.4

0

66

9.2

56

53

.82 62

1.0

45

95

.96

58

8.2

5 55

5.4

6

43

9.7

4

42

0.4

5

S 11

40060080010001200140016001800200024002800320036004000

1/cm

0

15

30

45

60

75

90

%T

35

62

.28

35

39

.13

35

21

.78

35

00

.56

34

79

.34

34

40

.77

34

21

.48

29

21

.96

23

58

.78

16

64

.45

16

47

.10 16

33

.59

16

20

.09

15

39

.09

14

56

.16 1

41

7.5

81

39

2.5

1

13

73

.22

13

38

.51

13

15

.36

10

22

.20

94

6.9

8

85

2.4

8

66

9.2

56

49

.97

61

9.1

15

97

.89

56

8.9

65

49

.67

45

1.3

14

41

.67

42

2.3

8

S 9

96

Fig. 6.14 IR spectra of gellan gum

Fig. 6.15 IR spectra of CP 974P

40060080010001200140016001800200024002800320036004000

1/cm

60

65

70

75

80

85

90

95

100

%T

36

68

.36

36

02

.78 2

76

9.5

9

27

19

.44

25

22

.72

24

80

.29

24

34

.00

23

99

.28

23

52

.99

23

06

.71

22

52

.70

21

98

.70

21

40

.84

20

44

.40

19

86

.54

18

86

.25

18

66

.97

16

77

.95

16

16

.24

14

85

.09

14

54

.23

13

92

.51

13

57

.79

12

99

.93

10

87

.78

10

56

.92

10

29

.92

95

6.6

3

93

7.3

4

84

0.9

1

78

6.9

0

72

5.1

8

70

2.0

4

64

8.0

4

60

5.6

1

S17

97

Fig. 6.16 IR spectra of HEC

Fig. 6.17 IR spectra of sodium alginate

40060080010001200140016001800200024002800320036004000

1/cm

-10

0

10

20

30

40

50

60

70

80

90

%T

29

27

.74

28

79

.52

23

58

.78

23

39

.49

17

16

.53

16

45

.17

16

29

.74

14

65

.80

14

52

.30

14

29

.15

13

80

.94

13

61

.65

13

40

.43

13

15

.36

10

20

.27

93

9.2

7

89

2.9

8

83

3.1

9

66

7.3

26

51

.89

61

7.1

85

94

.03

57

4.7

5

46

6.7

44

45

.53

42

6.2

44

03

.09

DS 9

40060080010001200140016001800200024002800320036004000

1/cm

-10

0

10

20

30

40

50

60

70

80

90

%T

37

39

.72

35

81

.56

35

60

.35

35

42

.99

35

19

.85

34

98

.63

34

75

.49

34

42

.70

34

25

.34

33

98

.34

23

58

.78

23

35

.64

18

65

.04

18

26

.46

17

93

.68

17

93

.68

17

66

.67

17

39

.67

16

43

.24

16

29

.74

16

20

.09

15

71

.88

15

58

.38

15

39

.09 1

51

4.0

21

49

0.8

71

45

4.2

31

41

7.5

8

13

94

.44 1

36

7.4

41

33

8.5

11

31

5.3

6

10

22

.20

94

6.9

8

90

4.5

5

67

1.1

86

49

.97

61

9.1

15

95

.96

58

4.3

95

76

.68

54

9.6

7

47

0.6

04

49

.38

42

4.3

1

S 10

98

Fig. 6.18 IR spectra of P1

Fig. 6.19 IR spectra of S2

99

Fig. 6.20 IR spectra of G3

Fig. 6.21 IR spectra of C3

100

Fig. 6.22 IR spectra of P4

Fig. 6.23 IR spectra of S4

101

Fig. 6.24 IR spectra of G5

Fig. 6.25 IR spectra of C4

102

Fig. 6.26 IR spectra of P7

Fig. 6.27 IR spectra of S7

40060080010001200140016001800200024002800320036004000

1/cm

0

10

20

30

40

50

60

70

80

90

%T

29

70

.17

28

89

.17

28

69

.88

28

50

.59

28

08

.16

27

58

.02

27

38

.73

27

00

.16

25

49

.72

23

33

.71

19

63

.40

16

20

.09

15

81

.52

14

69

.66

14

54

.23

14

11

.80

13

80

.94

13

42

.36

12

76

.79

12

42

.07

11

14

.78

10

60

.78 1

03

3.7

71

01

0.6

3

96

0.4

8

94

5.0

5

84

4.7

6

73

6.7

6

65

5.7

5

60

5.6

1

50

9.1

7

48

6.0

3

42

0.4

5

S18

40060080010001200140016001800200024002800320036004000

1/cm

10

20

30

40

50

60

70

80

90

100

%T

35

14

.06

34

90

.92

34

54

.27

31

36

.04

29

74

.03

29

10

.38

28

91

.10

28

64

.09

28

33

.24

27

73

.45

26

77

.01

25

53

.58

23

58

.78

23

35

.64

22

73

.92

22

02

.56

21

58

.20

21

09

.98

17

26

.17

16

77

.95

16

18

.17

15

85

.38

15

68

.02

15

31

.37

14

81

.23

14

65

.80

14

46

.51

14

07

.94

13

84

.79 13

46

.22

12

69

.07

12

07

.36

11

82

.28

11

41

.78

10

99

.35

10

33

.77

93

3.4

8

90

2.6

2

82

5.4

8

78

4.9

77

69

.54

73

4.8

3

70

2.0

4

65

9.6

1

62

2.9

6

59

9.8

2

56

5.1

0

52

4.6

04

95

.67

46

0.9

64

32

.03

40

8.8

8

S 15

103

Fig. 6.28 IR spectra of G7

Fig. 6.29 IR spectra of C7

104

Fig. 6.30 DSC of P1

Fig. 6.31 DSC of P4

Temp Cel350.0300.0250.0200.0150.0100.050.0

DS

C m

W30.00

20.00

10.00

0.00

-10.00

-20.00

-30.00

DD

SC

mW

/min

52.8Cel

184.8Cel

217.5Cel

Temp Cel250.0200.0150.0100.050.0

DS

C m

W

4.00

2.00

0.00

-2.00

-4.00

-6.00

-8.00

-10.00

DD

SC

mW

/min

50.6Cel

129.4Cel

105

Fig. 6.32 DSC of P7

6.2 SELECTION OF VEHICLE

The solubility of FQ viz. LFX, OFX and NFX were tested in various buffers at

the dosage level desired 0.5, 0.3, 0.3%w/v respectively. It was found to be in the order

as given below:

Solubility of LFX ranked as acetate > distilled water > citrophosphate >

phosphate. Solubility of OFX ranked as acetate > STF > citrophosphate > distilled

water > phosphate buffer and solubility of NFX ranked as acetate > STF >

citrophosphate > phosphate buffer > distilled water.

Temp Cel250.0200.0150.0100.050.0

DS

C m

W

20.00

10.00

0.00

-10.00

-20.00

-30.00

DD

SC

mW

/min

55.9Cel

155.3Cel

221.6Cel

106

6.3 STANDARD CALIBRATION CURVE

Table 6.4 shows the absorbance of LFX, OFX and NFX standard solutions in

simulated tear fluid. Fig. 6.33, 6.34 and 6.35 shows a representative standard

calibration curve with slope, regression coefficient and intercept. The curve was

found to be linear at λmax of 288 nm, 293nm and 271nm resp. in simulated tear fluid.

The calculation of the drug content, in vitro drug release and stability studies are

based on this calibration curve.

Table: 6.4 Preparation of calibration curve data of in STF

LFX OFX NFX

Concentratio

n (µg/ml)

Absorbanc

e

Concentratio

n (µg/ml)

Absorbanc

e

Concentratio

n (µg/ml)

Absorbanc

e

0 0 0 0 0 0

2 0.150 2 0.222 2 0.232

4 0.299 4 0.362 4 0.425

6 0.419 6 0.531 6 0.641

8 0.544 8 0.694 8 0.867

10 0.709 10 0.857 10 1.027

12 0.828 12 1.013

14 1.028

107

Fig. 6.33 Std. calibration curve of LFX at 288nm

Fig. 6.34 Std. calibration curve of OFX at 293nm

y = 0.071x+0.045

R² = 0.9989

0

0.2

0.4

0.6

0.8

1

1.2

0 2 4 6 8 10 12 14 16

Ab

sorb

an

ce

Conc.(μg/ml)

y = 0.086x

R² = 0.997

0

0.2

0.4

0.6

0.8

1

1.2

0 2 4 6 8 10 12

Ab

sorb

an

ce

Conc.(μg/ml)

108

Fig. 6.35 Std. calibration curve of NFX at 271nm

y = 0.105x

R² = 0.997

0

0.2

0.4

0.6

0.8

1

1.2

0 2 4 6 8 10 12

Ab

sorb

an

ce

Conc.(μg/ml)

109

6.4 PHYSICOCHEMICAL CHARACTERIZATION

6.4.1 Placebo Formulations

6.4.1.1 Phase transition based on temperature

Table 6.5 Formulations of PXM 407 and PXM 188 combinations

Sr.

No.

PXM

407

PXM

188

Chitosan

HPMC

K4M

Gelling

capacity

Gelation

temp.

Clarit

y

(%w/v) (oC)

1 15 4 - - + 55.0±0.58 +++

2 16 4 - - + 48.3±0.40 +++

3 17 4 - - ++ 42.0±0.50 +++

4 18 4 - - ++ 38.24±0.36 +++

5 18 4 - 0.2 +++ 34.0±0.25 +++

6 18 4 - 0.4 +++ 33.50±0.58 +++

7 18 4 0.25 0.2 +++ 33.0±0.5 +++

8 18 4 0.5 0.2 +++ 33.1±0.62 +++

Mean ± SD, n = 3

Note: 1) (+) Phase transition within 60 sec, collapse of gel structure within 1-2

hr, (++); Phase transition within 60 sec, collapse of gel structure within 3-4 hr, (+++)

Phase transition within 60 sec and gel structure stable for more than 6 hr.

2) (-) turbid, (+) slightly turbid, (++) clear solution, (+++) clear and transparent.

110

6.4.1.2 Phase transition based on ion

6.4.1.2.1 Sodium alginate and HPMC K4M or HEC

Table 6.6 Sodium alginate and HPMC K4M or HEC

Sr. No. Sodium alginate

HPMC K4M

HEC

Gelling capacity Clarity

(%w/v)

1 1 - - + ++

2 1.5 - - ++ ++

3 2 - - ++ -

4 1.5 0.2 - +++ ++

5 1.5 0.4 - +++ ++

6 1.5 - 1 +++ ++

7 1.5 - 1.5 +++ ++

8 1.5 - 2 +++ ++

6.4.1.2.2 Gellan gum

Table 6.7 Gellan gum ISGS

Sr. No. Gellan gum

(% w/v)

Gelling capacity Clarity

1 0.1 ++ +++

2 0.2 ++ +++

3 0.3 +++ +++

4 0.4 +++ +++

5 0.5 +++ +++

111

6.4.1.3 Phase transition based on pH

Table 6.8 Formulations of CP 974P and HPMC K4M

Sr.

No.

CP

974P

HPMC

K4M

Citrophosphate

buffer pH 6

Gelling

capacity

Clarity Ph

(%w/w)

1 0.1 1 q.s. +++ ++ 4.5

2 0.2 1 +++ ++ 4.5

3 0.3 1 +++ ++ 4.5

4 0.4 1 +++ ++ 4.5

5 0.5 1 +++ ++ 4.5

112

6.4.2 ISG medicated formulations

Table 6.9 Physicochemical characterization of medicated ISG formulations

Sr.

No.

Formulation

code

Clarity Gelling

capacity

pH Mucoadhesive

force (dyne/cm2)

Gelation

temp. (oC)

1 P1 +++ +++ 5.5 5444.4±6.92 34.0±0.75

2 P2 +++ +++ 5.5 6125±3.32 33.50±0.12

3 P3 +++ +++ 6 4144.44±7.75 34.0±0.53

4 P4 +++ +++ 6 5444.44±7.03 33.0±0.41

5 P5 +++ +++ 6 6125±3.33 35.1±0.46

6 P6 +++ +++ 5.5 4144.44±7.75 34.0±0.48

7 P7 +++ +++ 5.5 5444.44±7.03 33.0±0.5

8 P8 +++ +++ 5.5 6125±3.33 33.1±0.66

9 S1 ++ +++ 6 3402.76±1.951 -

10 S2 ++ +++ 6 3763.89±4.64 -

11 S3 ++ +++ 6 4324.4±1.16 -

12 S4 ++ +++ 6 3402.76±1.951 -

13 S5 ++ +++ 6 3763.89±4.64 -

14 S6 ++ +++ 6 4324.4±1.16 -

15 S7 ++ +++ 6 5525±7.28 -

16 S8 ++ +++ 6 6125±9.89 -

17 G1 +++ +++ 6.8 3402.76±7.34 -

18 G2 +++ +++ 6.9 4402.76±7.170 -

19 G3 +++ +++ 7.0 5202.76±1.38 -

20 G4 +++ +++ 6.8 3402.76±7.34 -

21 G5 +++ +++ 6.9 4402.76±7.170 -

22 G6 +++ +++ 7.0 5202.76±1.38 -

23 G7 +++ +++ 6.9 3402.76±7.34 -

24 G8 +++ +++ 6.9 4402.76±7.170 -

25 C1 ++ +++ 4.5 4402.76±0.961 -

26 C2 ++ +++ 4.5 5083.33±1.80 -

27 C3 ++ +++ 4.5 6644.4±7.884 -

113

Sr.

No.

Formulation

code

Clarity Gelling

capacity

pH Mucoadhesive

force (dyne/cm2)

Gelation

temp. (oC)

28 C4 ++ +++ 4.5 4402.76±0.961 -

29 C5 ++ +++ 4.5 5083.33±1.80 -

30 C6 ++ +++ 4.5 6644.4±7.884 -

31 C7 ++ +++ 4.5 4402.76±0.961 -

32 C8 ++ +++ 4.5 5083.33±1.80 -

33 C9 ++ +++ 4.5 6644.4±7.884 -

Mean ± SD, n = 3

6.4.2.1 Appearance and clarity

All formulations containing PXMs and chitosan with non-ionic polymer, HPMC

K4M were found to be very clear without any precipitation. All formulations

containing CP 974P were found to clear. With increase in concentration of CP 974P,

clarity was found to be decreased. Formulations containing gellan gum were found to

be very clear, whereas formulations containing sodium alginate were found to be

satisfactory. As the concentration of sodium alginate was increased above 1.5% w/v,

turbidity also increased (Table 6.5 to 6.8).

6.4.2.2 pH

The pH of all thermosensitive IS gel was found to be in the range of 6-7. For ion

sensitive IS gel containing sodium alginate, pH was in the range of 5-6, similarly

gellan gum IS gel showed pH in the range of 6-7. The pH of all CP 974P based IS gel

was found to be in the range of 4.5-5. These pH values were considered to be

acceptable since the ophthalmic pH ranges between 4.5-7.0. Hence no discomfort or

excessive tear flux might occur on instillation (Table 6.9).

114

6.4.2.3 Gelation temperature

The liquid–gel conversion temperatures are considered to be ideal for eye in the

range of 25-34oC. If it is lower than 25

oC, a gel forms at room temperature and if

higher than 34oC, does not form gel when instilled in the eye. This results in the

drainage of the solution from the eyes. PXM solutions undergo thermoreversible

gelation, depending on the type, concentration and other ingredients. Gelation

temperature of PXM 407 can be adjusted within the range of 33-34°C by modifying

cross-linking agents, by mixing the different combinations of PXM (PXM 188).

Below the transition temperature, PXM solutions allow a comfortable and precise

carrier by the patient to the cul-de-sac, where thermogelation occurs. The results

obtained for gelation temperature study (Table 6.5) for different PXM 407

concentrations (15-18%w/v) along with PXM 188 (4.0%w/v) confirms that it is

dependent on polymer concentration; as the PXM 407 concentration increases

gelation temperature decreases. Formulation containing PXM 407 at concentration

15%w/v and 16% w/v did not form gel at physiological conditions in simulated tear

fluid (temperature 34oC and pH 7.4). Concentration 17%w/v and 18%w/v shows

gelation temperature of 42.0±0.50 and 38.24±0.36. Further addition of mucoadhesive

polymer, HPMC K4M (0.2-0.4%w/v) reduces the liquid to gel conversion

temperature of the ISGS. The gelation temperature reducing effect of mucoadhesive

polymers could be explained by their ability to bind to polyoxyethylene chains present

in the PXM molecules. This will promote dehydration, causing an increase in

entanglement of adjacent molecules and extensively increasing intermolecular

hydrogen bonding which will lead to gelation at lower temperature. Further addition

of chitosan (0.25%w/v) a mucoadhesive and penetration enhancer lowered the

gelation temperature to 33.0±0.5. Increase in chitosan concentration to 0.5%w/v

115

causes gelation at 33.1±0.62. Finally Combination of 18%w/v PXM 407 and 4%w/v

PXM 188 with 0.2%w/v HPMC K4M and 0.25%w/v chitosan was used for further

study.

6.4.2.4 Gelling capacity

Combination of 18%w/v PXM 407 and 4%w/v PXM 188 was found to be gel

forming quickly within 60 sec. and gel structure stable for more than 3hr. When

HPMC K4M (0.2-0.4%w/v) and chitosan (0.25%w/v) was incorporated in PXM

combination, the gel formulation was quick and gel structure retained for longer time

ie more than 6 hr (Table 6.5). This suggests increase in gel strength with addition of

dual mucoadhesive polymer. Chitosan is more mucoadhesive. The formulation of

sodium alginate at concentration 1-2%w/v was found to be liquid at non-physiological

conditions. When it was added to simulated tear fluid, in presence of Ca++

ions, it was

transformed to gel. This might be attributed to interaction of Ca++

ions with the G

entity of sodium alginate resulting into 3 dimensional gel structures. Alginates are

good gel forming but as the concentration of sodium alginate increases from 1.0 to

2.0%, turbidity increases (Table 6.6). So sodium alginate in concentration of 1.5%w/v

was further combined with non ionic mucoadhesive polymers HPMC K4M and HEC,

at different concentrations. The resultant formulations retained gelling ability and

after addition to ionic environment formed stable gel structure. All the formulations

containing different concentrations of gellan gum (0.1%-0.5%) showed the gelling

ability in the presence of simulated tear fluid. It is attributed to formation of a three

dimensional network by its complex formation with Ca++

ions and hydrogen bonding

with water (Table 6.7). Gellan gum forms a clear gel on contact with cations in the

tear fluid and once gelled, the formulation resists the natural drainage process from

the precorneal area. CP 974P showed concentration dependent gelation in simulated

116

tear fluid at lower concentration than PXM (Table 6.8). We found that on addition of

STF, developed formulations showed instant gelation. The texture of the gel solely

affected by the amount of polymers used or added in respective formulations. As the

concentration of CP increases, its acidic nature stimulates eye tissues. To improve the

gelling properties, a mucoadhesive HPMC K4M (1.0%w/v) was added.

Fig. 6.36 Representative for sol to gel transition

6.4.2.5 Drug content

Table 6.10 shows the percent drug content for all medicated formulations. The

drug content was found to be in acceptable range.

Table 6.10 % drug content for all medicated ISG formulations

Sr.No. Formulation

code

% Drug

content

Sr.No. Formulation

code

% Drug content

1. P1 99.51±0.608 17. G1 98.93±0.208

2. P2 99.73±0.264 18.. G2 99.96±0.057

3. P3 99.83±0.057 19.. G3 99.71±0.608

4. P4 99.76±0.40 20. G4 99.81±0.360

5. P5 98.93±0.208 21. G5 98.8±0.250

6. P6 99.53±0.288 22. G6 99.51±0.360

7. P7 99.93±0.115 23. G7 99.9±0.15

117

8. P8 99.26±0.513 24. G8 98.60±0.258

9. S1 98.93±0.208 25. C1 99.85±0.70

10. S2 99.61±0.057 26. C2 99.51±0.360

11. S3 99.78±0.355 27. C3 99.65±0.217

12. S4 99.39±0.265 28. C4 99.9±0.18

13. S5 98.93±0.208 29. C5 99.96±0.40

14. S6 99.10±0.057 30. C6 99.45±0.70

15. S7 98.93±0.208 31. C7 99.71±0.360

16. S8 99.81±0.057 32. C8 99.65±0.217

33. C9 99.85±0.70

Mean ± SD, n = 3

6.4.2.6 Mucoadhesion test (Bioadhesion potential)

Ocular mucoadhesion relies on the interaction of a polymer and the mucin coat

that is present on conjunctival and corneal surface of the eye. This mucus is secreted

by goblet cells of conjunctiva. Chemically, mucin consists of a protein or polypeptide

core with carbohydrate side chains branching off the core. The polymer with many

hydrophilic functional groups (eg: -CO, -OH, -NH2 and SO4) can establish

electrostatic interactions and lipophilic interactions and hydrogen bonding with the

underlying surface. Among non-covalent forces, hydrogen bonding appears to be the

most important. PXM 407 gives colorless and transparent gels but weak mechanical

strength and rapid erosion. It is concluded that as the polymer concentration increases

mucoadhesion increased. This might be due to increased entanglement of polymer

chains with mucin at increased concentration of polymer. The addition of

mucoadhesive polymer, HPMC K4M has increased mucoadhesive force. The

118

reinforcement of the mucoadhesive forces of PXM solutions by the used

mucoadhesive polymer could be explained by the fact that, secondary bond forming

groups (e.g. hydroxyl, ether oxygen and amine) are the principle source of

mucoadhesion. HPMC K4M and HEC have an ample of -OH and R-O-R groups

along their chain length. CP is known to be excellent mucoadhesive polymers. Hence,

satisfactory bioadhesion was observed in case of all ophthalmic gel formulations. The

bioadhesion is determined by the availability of carboxyl groups. The % of carboxyl

group present in CP 974P is high. These groups simultaneously bind to the sugar

residue in oligosaccharide chains present in the mucus membrane, resulting in a

strong bond between the polymer and mucus membrane. With increase in the density

of hydrogen bonding groups interaction increase with the glycoproteins of the mucin.

Also CP may adopt conformation that has more favorable macromolecular

accessibility of its functional groups for hydrogen bonding. Thus increase in the

mucoadhesive force will lead to increase in retention time and thus bioavailability.

Sodium alginate, an anionic mucoadhesive have shown less bioadhesion as compared

to CP. Gellan gum also form good clear gel without addition of mucoadhesive. Once

gelled, the formulation resists the natural drainage process from the precorneal area.

119

6.4.2.7 In vitro drug release studies

The release profile of a drug gives valuable insight into its in vivo performance.

All the developed ISGS were subjected to in vitro release studies (Table 6.11 to 6.14.

and Fig. 6.37 to 6.40).

Fig. 6.37 In vitro release profile of all temperature sensitive IS gel

Mean ± SD, n = 3

0

10

20

30

40

50

60

70

80

90

100

0 2 4 6 8

Cu

mu

lati

ve

% d

rug r

elea

sed

Time (hr)

P1

P2

P3

P4

P5

P6

P7

P8

120

Table 6.11 Average cumulative % release of all temperature sensitive IS gel

Time(

hr)

LFX OFX NFX

P1 P2 P3 P4 P5 P6 P7 P8

0 0 0 0 0 0 0 0 0

1 25.59±

0.04

25.3±

0.07

9.09±

0.21

19.0±

0.42

21.49±

0.88

9.7±

0.92

12.61±

0.95

8.72±

0.95

2 46.54±

0.04

46.25±

0.55

18.25±

0.75

32.30±

1.3

38.88±

1.83

15.33±

1.67

20.99±

1.0

12.30±

0.92

3 64.1±

0.05

63.52±

0.34

21.62±

0.27

51.53±

0.96

53.42±

0.56

23.55±

0.9

30.3±

1.36

18.15±

0.83

4 76.92±

0.06

76.87±

0.11

27.24±

1.5

61.45±

0.51

63.53±

1.13

28.83±

083

38.52±

0.76

20.057±

0.57

5 89.58±

0.03

84.49±

0.32

29.65±

0.63

69.33±

0.79

66.55±

1.02

30.22±

0.57

45.8±

1.0

23.50±

0.72

6 97.38±

0.06

90.08±

0.15

36.25±

0.43

77.44±

0.72

70.43±

0.12

36.3±

1.52

51.05±

1.52

26.99±

0.12

7 - 97.38±

0.08

38.44±

0.25

88.28±

0.88

72.18±

0.15

41.45±

1.02

61.57±

1.57

29.28±

0.15

8 - - 41.34±

0.91

99.77±

0.73

75.92±

2.42

45.58±

1.1

68.23±

1.52

32.02±

2.42

Mean ± SD, n = 3

121

Fig. 6.38 In vitro release profile of ion sensitive IS gel (Sodium Alginate)

Mean ± SD, n = 3

0

10

20

30

40

50

60

70

80

90

100

0 2 4 6 8

Cu

mu

lati

ve

% d

rug r

elea

sed

Time (hr)

S1

S2

S3

S4

S5

S6

S7

S8

122

Table 6.12 Average cumulative % release of all ion sensitive

IS gel (Sodium Alginate)

Time

(hr)

LFX OFX NFX

S1 S2 S3 S4 S5 S6 S7 S8

0 0 0 0 0 0 0 0 0

1 58.55±

0.45

24.96±

0.51

38.77±

0.44

25.06±

0.78

24.7±

0.57

36.39±

1.39

16.08±

0.23

14.52±

0.69

2 89.59±

0.4

52.68±

0.74

58.92±

0.62

52.57±

1.16

34.99±

0.91

63.27±

1.22

26.7±

0.73

18.95±

0.44

3 93.02±

0.05

54.45±

0.4

81.04±

0.63

54.29±

0.58

47.29±

0.88

74.04±

0.34

35.26±

0.5

24.78±

0.23

4 - 60.19±

0.15

85.53±

0.81

60.01±

0.17

51.2±

1.25

81.7±

0.79

42.22±

0.9

29.78±

0.37

5 - 66.9±

0.43

89.14±

0.46

66.0±

0.8

55.07±

0.66

89.16±

1.39

45.34±

1.0

33.12±

0.77

6 - 71.4±

0.67

90.56±

0.98

71.21±

0.31

60.68±

0.86

94.58±

0.69

47.56±

0.63

36.79±

0.76

7 - 85.77±

0.12

- 76.13±

0.67

69.91±

0.4

95.75±

0.43

57.57±

0.54

38.3±

0.63

8 - 90.36±

0.44

- 85.48±

0.7

71.23±

0.92

99.92±

0.9

62.81±

0.88

40.3±

0.5

Mean ± SD, n = 3

123

Fig. 6.39 In vitro release profile of all ion sensitive IS gel (Gellan gum)

Mean ± SD, n = 3

0

10

20

30

40

50

60

70

80

90

100

0 1 2 3 4 5 6 7

Cu

mu

lati

ve

% d

rug r

elea

sed

Time (hr)

G1

G2

G3

G4

G5

G6

G7

G8

124

Table 6.13 Average cumulative % release of all ion sensitive IS gel (Gellan

gum)

Time(

hr)

LFX OFX NFX

G1 G2 G3 G4 G5 G6 G7 G8

0 0 0 0 0 0 0 0 0

1 55.93±

1.1

29.39±

0.76

25.39±

1.74

36.2±

0.47

62.18±

1.81

47.4±

2.18

42.89±

0.707

41.21±

1.41

2 64.62±

1.03

50.13±

0.65

32.81±

1.13

46.53±

0.58

80.92±

2.89

60.78±

1.7

62.68±

1.41

57.78±

1.42

3 73.41±

1.06

58.81±

2.72

38.87±

1.11

58.94±

0.8

95.52±

0.61

78.2±

0.32

74.2±

2.21

71.62±

1.41

4 88.65±

2.58

70.4±

0.71

50.74±

0.97

75.12±

1.5

- 90.24±

0.31

84.46±

1.41

82.74±

1.42

5 100.8±

1.49

84.14±

1.13

58.08±

0.28

86.3±

0.47

- - 85.26±

2.12

96.18±

1.4

6 - 93.92±

0.48

65.67±

0.68

95.52±

0.61

- - 90.24±

0.31

97.34±

0.7

7 - - 75.2±

0.32

- - - 100±

0.45

99.71±

0.79

Mean ± SD, n = 3

125

Fig. 6.40 In vitro release profile of all pH sensitive IS gel

Mean ± SD, n = 3

0

10

20

30

40

50

60

70

80

90

100

0 2 4 6 8

Cu

mu

lati

ve

% d

rug r

elea

sed

Time (hr)

C1

C2

C3

C4

C5

C6

C7

C8

C9

126

Table 6.14 Average cumulative % release of all pH sensitive IS gel

Time

(hr)

LFX OFX NFX

C1 C2 C3 C4 C5 C6 C7 C8 C9

0 0 0 0 0 0 0 0 0 0

1 49.5±

0.88

38.8±

0.93

25.0±

0.75

25.0±

0.75

22.2±

0.41

12.7±

0.57

12.7±

0.42

14.5±

0.46

14.5±

1.29

2 75.5±

0.86

61.26±

1.54

43.6±

0.61

43.6±

0.61

38.2±

0.87

33.6±

0.48

28.8±

0.25

28.1±

0.25

21.4±

1.03

3 91.7±

1.0

81.32±

1.02

60.8±

0.45

60.8±

0.45

61.7±

0.44

52.7±

0.45

39.6±

0.21

42.0±

0.39

34.9±

1.28

4 96.9±

0.66

87.04±

0.67

68.2±

0.28

68.2±

0.28

62.5±

0.13

60.2±

0.34

42.7±

0.47

43.5±

0.83

42.2±

0.83

5 - 96.09±

1.33

70.66

±0.35

70.66

±0.35

65.8±

1.89

62.1±

0.49

60.6±

0.6

56.2±

0.3

61.1±

0.21

6 - 100.86

±1.24

78.5

±0.7

71.2

±0.7

70.84

±1.18

64.37

±1.03

75.36

±0.42

67.91

±0.29

66.03

±0.85

7 - - 88.58

±0.4

88.58

±0.4

75.88

±0.29

65.74

±0.33

89.54

±0.47

81.5

±0.58

80.93

±0.82

8 - - 97.92

±0.6

96.88

±0.1

78.0±

0.43

71.84

±0.58

90.45

±0.53

85.51

±0.47

84.56

±0.67

Mean ± SD, n = 3

127

The in vitro drug release profile of all ISGS is affected by the intrinsic properties

(solubility and permeability) of the FQ drugs viz. LFX, OFX (BCS Class I); and

NFX. Also use of mucoadhesive polymers and permeation enhancer has shown

impact on drug release profile. LFX is more water soluble than OFX and OFX is more

soluble than norfloxacin. Solubility of respective drug has shown effect on release

profile. Release of LFX is rapid as compared to OFX and NFX. Formulation P1, P3

and P6 shows cumulative percent drug release of 46.54±0.04, 18.25±0.75 and

15.33±1.67 respectively after 2 hr. Addition of chitosan (0.25%w/v) as a penetration

enhancer increases release of OFX and NFX from their formulation. Formulation P4

and P7 shows cumulative percent drug release of 99.77±0.73 and 68.23±1.52

respectively after 8 hr. Chitosan plays dual role of penetration enhancer as well as

mucoadhesive. Mucoadhesive properties of chitosan (0.5%w/v) further sustained the

drug release. A similar release profile was observed from alginate based ion sensitive

ISGS. Formulation S1 and S4 with 1.0% HEC shows cumulative percent drug release

of 89.59±0.4 and 52.57±1.16, while S7 with 0.2% HPMC K4M shows cumulative

percent drug release of 26.7±0.73 after 2 hr. Gellan based ion sensitive ISGS of FQ

drugs shows rapid release than other developed systems. Release was sustained as

polymer concentration increased. Polymer concentration dependant release was

observed from pH sensitive ISG system. As concentration of CP 974 P was increased

from 0.1 to 0.3% drug release becomes more sustained.

Following the results obtained for bioadhesion force, gelling capacity and in vitro

drug release profile of all the formulations, selected formulations were further

subjected to rheological behavior and transcorneal permeation profile.

128

6.4.2.8 Rheological behavior

Table 6.15 to 6.17 shows the viscosity of ISGS of FQ drugs viz. LFX, OFX and

NFX using Brookfield viscometer LVDV-II + Pro model with T- bar spindle code S

93. Formulations posses thinning nature on shearing as, shear stress increases with

increase in angular velocity. The results obtained from the rheological study of

prepared ISGS suggest that the viscosity decreases following ascending order of

angular velocity. Generally viscosity values in the range of 15-150 cps significantly

improve the contact time. The instillation of formulation should influence minimally

pseudoplastic character. Since the ocular shear rate is very large ranging from 0.03 S-1

during interblinking periods to 4250 – 28,500 S-1

during blinking, viscoelastic fluids

with a viscosity that is high under conditions of low shear rate and low under

conditions of high shear are preferred. The rheological profile of prepared ISGS is

shown in Fig. 6.41 to 6.43.

Fig. 6.41 Rheological behavior of LFX ISGS

0

100

200

300

400

500

600

700

800

0 10 20 30 40 50 60 70 80 90 100

Vis

cosi

ty C

ps

Angular Velocity (rpm)

P1

S2

G3

C3

129

Table 6.15 Rheological profile of LFX ISGS

Angular

Velocity (rpm)

Viscosity (cps)

P1 S2 G3 C3

0.5 790 220 395 239

2.5 162 132 135 144

5 95.3 69.569 78.4 73.478

10 51.9 43.78 43.5 50

20 30.1 16.945 23.4 18

50 14.2 7.896 12.35 9

100 7.23 4.567 5.3 5.12

Fig. 6.42 Rheological behavior of OFX ISGS

0

100

200

300

400

500

600

700

800

0 10 20 30 40 50 60 70 80 90 100

Vis

cosi

ty C

ps

Angular Velocity (rpm)

P4

S4

G5

C4

130

Table 6.16 Rheological profile of OFX ISGS

Angular

Velocity (rpm)

Viscosity (cps)

P4 S4 G5 C4

0.5 800 220 356 210

2.5 219 101 123.17 104

5 96 68.7 70 36.1

10 55.6 42 35.89 21.3

20 29.2 13.5 21.5 12.2

50 15.3 6.7 12 5.5

100 6.99 4.298 4.798 4.9

Fig. 6.43 Rheological behavior of NFX ISGS

131

Table 6.17 Rheological study of NFX ISGS

Angular

Velocity (rpm)

Viscosity (cps)

P7 S7 G7 C7

0.5 810 457 322 210

2.5 212 123 112.67 104

5 95.3 88.9 63.4 36.1

10 51.9 42.5 32.45 21.3

20 30.1 21.5 19 12.2

50 14.2 11.9 9.7 5.5

100 7.23 6.2 4.578 4.9

6.4.2.9 Transcorneal permeation studies

Fig. 6.44 Transcorneal permeation of LFX ISGS Mean ± SD, n = 3

Table 6.18 Average cumulative % drug permeated from LFX ISGS

0

20

40

60

80

100

0 2 4 6 8

Cu

mu

lati

ve

% d

rug p

erm

eate

d

Time (hr)

P1

S2

G3

C3

Eye

drop

132

Time(hr) P1 S2 G3 C3 Marketed eye

drop

0 0 0 0 0 0

1 45.99±0.633 48.38±0.95 28.67±1.5 8.21±0.36 61.82±0.36

2 50.79±0.57 55.97±0.7 47.92±1.15 40.92±0.74 69.12±1.0

3 61.13±0.51 65.3±1.34 56.97±1.73 49.32±0.85 74.21±0.7

4 72.71±1.16 83.6±0.89 67.83±1.0 57.92±0.8 86.08±1.5

5 80.79±0.55 90.98±1.29 75.86±1.0 67.76±0.69 95.04±1.0

6 94.46±0.52 95.22±0.8 89.94±1.0 75.0 ±0.67 -

7 97.26±1.35 98.28±0.52 96.96±1.0 89.23±0.34 -

8 99.18±0.6 - - - -

Mean ± SD, n = 3

Fig. 6.45 Transcorneal permeation of OFX ISGS

Mean ± SD, n = 3

0

20

40

60

80

100

0 2 4 6 8

Cu

mu

lati

ve

% d

rug p

erm

eate

d

Time (hr)

P4

S4

G5

C4

Eye

drop

133

Table 6.19 Average cumulative % drug permeated from OFX ISGS

Time(hr) P4 S4 G5 C4 Marketed eye

drop

0 0 0 0 0 0

1 18.49±0.65 37.91±1.05 47.4±0.54 29.68±2.18 36.58±0.78

2 26.75±0.78 42.51±0.63 60.78±0.53 40.85±1.7 37.49±1.41

3 37.5±1.0 56.5±0.68 73.2±0.86 47.85±2.21 47.291±0.63

4 52.71±0.74 64.45±0.61 87.46±0.72 56.38±1.41 49.43±0.78

5 72.55±1.07 71.74±0.5 93.24±0.52 72.15±0.31 60.64±1.41

6 92.26±0.84 90.48±0.59 98.96±0.52 85.23±0.78 73.98±0.63

7 - - - 94.27±0.78 92.17±0.63

Mean ± SD, n = 3

Fig. 6.46 Transcorneal permeation of NFX ISGS

Mean ± SD, n = 3

0

20

40

60

80

100

0 2 4 6 8

Cu

mu

lati

ve

% d

rug p

erm

eate

d

Time (hr)

P7

S7

G7

C7

Eye

drop

134

Table 6.20 Average cumulative % drug permeated from NFX ISGSs

Time

(hr)

P7 S7 G7 C7 Marketed eye

drop

0 0 0 0 0 0

1 39.96±0.7 48.29±1.26 63.49±0.57 22.68±0.57 91.829±1.41

2 49.9±1.41 59.15±1.83 67.5±0.49 40.85±0.7 96.18±0.5

3 67.75±0.7 71.26±0.85 72.57±0.5 47.85±0.6 97.08±1.53

4 71.6±1.41 79.36±1.01 86.53±0.24 56.38±3.29 100.15±1.83

5 75.16±0.24 84.92±1.29 96±0.55 72.15±0.31 -

6 78.25±1.93 98.33±0.52 99.37±0.4 85.23±1.53 -

7 82.21±0.24 99.1±1.53 - 94.27±1.53 -

Mean ± SD, n = 3

135

Table 6.21 Model fitting for release study of the formulations

Code

Zero order Higuchi matrix Korsmeyer peppas

R2

K R2

K R2 K N

P1 0.9169 19.892 0.9969 41.879 0.9681 2.103 0.5302

S2 0.9767 17.283 0.9842 39.032 0.9567 0.2982 0.6199

G3 0.9479 14.70 0.9885 37.569 0.9432 2.097 0.7210

C3 0.8891 17.027 0.9857 40.333 0.9923 10.459 0.5967

P4 0.9119 21.196 0.9914 44.61 0.9635 16.49 0.5198

S4 0.9315 35.778 0.9922 57.55 0.8418 16.47 0.6189

G5 0.9065 10.704 0.9912 31.334 0.9825 12.073 0.5810

C4 0.8891 17.027 0.9857 40.333 0.9923 10.459 0.5967

P7 0.9801 49.766 0.9993 65.446 0.9127 9.672 0.6026

S7 0.9662 34.486 0.9988 54.652 0.9322 15.783 0.5034

G7 0.9119 21.196 0.9914 44.61 0.9635 16.49 0.5198

C7 0.9664 16.197 0.9924 36.924 0.9577 12.346 0.6523

The transcorneal permeation study of all ISGS is affected by the intrinsic

properties of the FQ drugs viz. LFX, OFX and NFX. Also use of mucoadhesive

polymers and permeation enhancer has shown impact on drug permeation study

(Fig.6.44 to 6.46). Higuchi matrix diffusion mechanism was observed from all ISG

formulation. The release of drug from the formed matrix is influenced by diffusion

and/or erosion. The overall diffusion-controlled release kinetics was found. The best

fit kinetic model was Higuchi matrix model. When compared using student t test

ANOVA followed by Dunnet’s test was done to study transcorneal permeation after 2

136

hr. LFX and NFX IS gel formulations showed sustained release as compared to

marketed eye drop. For OFX containing IS gel formulation, only P4 showed sustained

release (Fig. 6.47 to 6.49).

Fig. 6.47 Comparison of transcorneal permeation of LFX IS gel and

marketed eye drop

Fig. 6.48 Comparison of transcorneal permeation of OFX IS gel and

marketed eye drop

0

10

20

30

40

50

60

70

P4 S4 G5 C4 Eye drop

Cu

mu

lati

ve

% d

rug p

erm

eate

d

***

137

Fig. 6.49 Comparison of transcorneal permeation of NFX IS gel and

marketed eye drop

The results are expressed as mean + SEM. Comparison between the groups was

made by one way analysis of variance (ANOVA) followed by Tukey-Kramer

Multiple Comparisons Test *-P<0.05, **-P<0.01, ***-P<0.001; *-Comparison of IS

gel formulations against marketed eye drop.

6.4.2.10 Drug polymer interaction studies

All the ISGS of FQ drugs viz. LFX, OFX and NFX did not show any change in

clarity, pH, gelling ability, drug content and gelation temperature as studied after

autoclaving (Table 6.22). Thin layer chromatography was carried out to check the

drug polymer interaction after autoclaving. Rf values were found to be nearly same

for standard drug and its formulation after autoclaving (Table 6.23 to 6.25 and Fig.

6.50 to 6.52). Absence of additional spots in chromatogram before and after

autoclaving indicates stability of drug in given system.

0

20

40

60

80

100

120

P7 S7 G7 C7 Eye drop

Cu

mu

lati

ve

% d

rug p

erm

eate

d

***

******

***

138

Table 6.22 Drug polymer interaction studies

Sr.

No.

Formulation

code

Clarity Gelling

capacity

pH Drug

contents

Gelation

temp.

(oC)

1 P1 +++ +++ 5.5 99.9±0.1 34.0±0.2

2 P4 +++ +++ 6 98.80±0.258 33.0±0.35

3 P7 +++ +++ 5.5 99.78±0.70 33.0±0.35

4 S2 ++ +++ 6 99.71±0.360 -

5 S4 ++ +++ 6 99.75±0.217 -

6 S7 ++ +++ 6 99.9±0.1 -

7 G3 +++ +++ 7.0 99.86±0.40 -

8 G5 +++ +++ 6.9 99.85±0.70 -

9 G7 +++ +++ 6.9 99.71±0.360 -

10 C3 ++ +++ 4.5 99.9±0.1 -

11 C4 ++ +++ 4.5 98.60±0.258 -

12 C7 ++ +++ 4.5 99.65±0.70 -

Mean ± SD, n = 3

Note: 1) (+) Phase transition within 60 sec, collapse of gel structure within 1-2

hr, (++); Phase transition within 60 sec, collapse of gel structure within 3-4 hr, (+++)

Phase transition within 60 sec and gel structure stable for more than 6 hr.

2) (-) turbid, (+) slightly turbid, (++) clear solution, (+++) clear and

transparent.

139

P1 S3 G5 C3

Fig. 6.50 TLC Chromatogram of LFX formulations

Table 6.23 TLC data of LFX formulations

Formulation

code

Solvent

front

Drug (R)

LFX

Sample Rf

Drug

Rf

Sample

P1 6.9 2.2 2.1 0.31 0.32

S2 6.5 2.1 2 0.32 0.30

G3 5.5 1.6 1.5 0.290 0.272

C3 6.8 1.9 1.8 0.279 0.265

P4 S4 G5 C4

Fig. 6.51 TLC Chromatogram of OFX formulations

140

Table 6.24 TLC data of OFX formulations

Formulation

code

Solvent

front

Drug (R)

OFX

Sample Rf Drug Rf

Sample

P4 7.0 1.8 1.9 0.26 0.27

S4 6.8 1.67 1.7 0.25 0.25

G5 6.5 1.8 2.0 0.28 0.30

C4 7.0 1.8 1.9 0.27 0.28

P7 S7 G7 C7

Fig. 6.52 TLC Chromatogram of NFX formulations

Table 6.25 TLC data of NFX formulations

Formulation

code

Solvent

front

Drug (R)

NFX

Sample Rf Drug Rf

Sample

P7 6.8 1.9 1.8 0.279 0.264

S7 6.5 1.5 2.0 0.23 0.307

G7 6.4 1.5 1.3 0.23 0.20

C7 6.9 1.6 1.5 0.231 0.217

141

All the selected formulations are good in viscosity. The systems are shear

thinning. Also transcorneal permeation study data suggests overall diffusion

controlled release kinetics. Amongst the four formulations developed, based on three

systems, for FQ drugs viz. LFX, OFX and NFX for further study two systems were

considered viz. gellan based ion sensitive ISG system and PXM based

thermosensitive ISG system. These two systems were found to be very well in clarity.

Also they give good film on the eye surface as observed on separated goat cornea.

142

6.4.2.11 Antimicrobial efficacy studies

The result of antimicrobial study shown that there were no changes in the

antimicrobial activity of FQ drugs viz. LFX, OFX and NFX due to formulation

ingredients and working conditions as compared to reference formulation (marketed

eye drop formulation), Table 6.26 and Fig.6.53.

Table 6.26 Antimicrobial efficacy of ophthalmic ISG formulations

Formulation Code

Pseudomonas A Staphylococcus A

% Efficacy % Efficacy

Standard LFX 100 100

P1 97.14 92.10

G3 98.23 93.34

Standard OFX 100 100

P4 94.28 96.84

G5 94.22 96.03

Standard NFX 100 100

P7 97.14 92.10

G7 98.00 94.00

143

Standard LFX P1 G3

Standard OFX P4 G5

Standard NFX P7 G7

Fig. 6.53 Antimicrobial efficacy of ophthalmic formulations

144

6.4.2.12 Isotonicity

Fig. 6.54 shows that ISGS of FQ drugs viz. LFX, OFX and NFX exhibited no

change in the shape of blood cells (bulging and shrinkage) which reveals the isotonic

nature of the formulation and compared with that of marketed LFX eye drop.

Normal RBC Marketed eye

drop

P1

P4

P7 G3 G5 G7

Fig. 6.54 Isotonicity of formulations

6.4.2.13 Precorneal clearance study using gamma

scintigraphy

In vivo precorneal clearance of radionuclide was studied using single photon emission

computing tomography (SPECT LAB). It was chosen as it emits low energy gamma

rays which do not lead to serious health hazards. Six IS gel formulations as well as

marketed eye drops were assessed in terms of their ocular retention time. Both,

thermosensitive and gellan

145

based ion sensitive optimized formulations containing LFX, OFX and NFX were

assessed on a group of four rabbits with a minimum washout period of 3 days.

Recording was started immediately after instillation at a rate of 15 seconds per image

for 10 min. and more using SIEMENS ECAM gamma camera (SPECT LAB Pune,

INDIA). Region of interest (ROI) was choose (Fig.6.58 and 6.59) and time activity

curve was plotted and rate of drainage from eye was calculated (Fig.6.55 to 6.57) upto

10 min. A single whole body static image also was taken after 120 min. of instillation

(Fig.6.60).

Fig. 6.55 Time activity curve of LFX formulations (P1and G3) and marketed

eye drop solution (mean±SD; n=2)

0

20

40

60

80

100

0 1 2 3 4 5 6 7 8 9 10

Cou

nts

/S

ec

Time (min)

P1

G3

Eye

drop

146

Fig. 6.56 Time activity curve of OFX formulations (P4 and G5) and

marketed eye drop solution (mean±SD; n=2)

Fig. 6.57 Time activity curve of NFX formulations (P7and G7) and marketed

eye drop solution (mean±SD; n=2)

0

20

40

60

80

100

0 1 2 3 4 5 6 7 8 9 10

Cou

nts

/Sec

Time (min)

P4

G5

Eye

drop

0

20

40

60

80

100

0 1 2 3 4 5 6 7 8 9 10

Cou

nts

/S

ec

Time (min)

P7

G7

Eye

drop

147

Fig.6.58 Representative ocular contact of an IS gel with sequential pictures

148

Fig.6.59 Representative ocular contact of an IS gel with sequential pictures

and time activity curve

149

Fig.6.60 Representative static complete body image (A) marketed eye drop

(B) developed ISG system

For scintigraphic studies, during prelabeling efficiency, labeling parameters like

SnCl2 concentration and pH were optimized (Table 5.12). The acquired gamma

camera images showed that both developed ISGS form good clear gel over the

corneal surface immediately after instillation. Marketed eye drop solutions were drain

very rapidly from the corneal region whereas; all ISGS were cleared at slow rate with

improved contact for longer duration.

150

IS gel forming abilities of the developed systems significantly controls precorneal

drainage. Thus, increased residence time in eye would help to increase ocular

bioavailability. The period of drug absorption is short because the activity gradient

decreases rapidly owing to precorneal solution drainage and conjunctival systemic

absorption. A minimum of 5-10 min of ocular contact time was determined to be

necessary for significantly reducing systemic drug absorption [38].

Superficial cornea1 opacity has been observed with gellan based systems on the

rabbit eye after gamma scintigraphy study. Thermosensitive system does not show

any opacity on the eyes of rabbits. A similar corneal change has been noticed with

carbomer, contained in Pilopine HS ophthalmic gel [151]. Looking towards the data

supported, further study was conducted on optimized thermosensitive ISG medicated

formulation.

6.4.2.14 Ocular irritation studies

Optimized thermosensitive ISG medicated formulation containing LFX, OFX and

NFX was found to be well tolerated and non-irritant at used combinations of PXM

showing mucomimetic properties as well as optical clarity. Excellent ocular tolerance

was noticed for all the three formulations. No signs of redness, watering of the eye

and swelling were observed throughout the study (Fig. 6.61 to 6.63).

151

1hr 24 hr 48 hr 72hr

Right eye untreated

P1

Fig. 6.61 Ocular irritation test of optimized IS LFX formulation in rabbit eye

1 hr 24 hr 48 hr 72 hr

Right eye untreated

P4

Fig. 6.62 Ocular irritation test of optimized IS OFX formulation in rabbit eye

1 hr 24 hr 48 hr 72 hr

Right eye untreated

P7

Fig. 6.63 Ocular irritation test of optimized IS NFX f ormulation in rabbit eye

6.4.2.15 Ocular pharmacokinetic study

Ocular pharmacokinetic study

The optimized thermosensitive ISGS and eye drops were subjected to in vivo

studies to determine drug levels in aq. humor. The calibration curve profile of LFX,

OFX and NFX is depicted in Fig. 6.64, 6.66 and 6.68 resp. Fig.6.65, 6.67 and 6.69

shows HPLC chromatograms of Levofloxacin hemihydrates,

152

OFX and NFX in rabbit aq. humor, retention time was found to 2.48 min, 3.11 min

and 2.66 min resp. Table 6.27 shows the aq. concentration of the drug at each

sampling interval for eye drops of each drug and P1, P4 and P7 formulations. The plot

of the concentration in aq. humor vs. time is shown in Fig. 6.70.

Fig. 6.64 Calibration curve of LFX by HPLC

Fig. 6.65 Chromatogram of LFX (P1) IS gel in aq. humor by HPLC

-41.000

-10.600

19.800

50.200

80.600

111.000

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00

[mAU]

[min]

2.4

8

Ite

m 2

153

Fig. 6.66 Calibration curve of OFX by HPLC

Fig. 6.67 Chromatogram of OFX (P4) IS gel in aq. humor by HPLC

-3.000

0.200

3.400

6.600

9.800

13.000

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00

[mAU]

[min]

3.1

1

Ite

m 6

154

Fig. 6.68 Calibration curve of NFX by HPLC

Fig. 6.69 Chromatogram of NFX (P7) IS gel in aq. humor by HPLC

-7.000

2.200

11.400

20.600

29.800

39.000

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00

[mAU]

[min]

2.6

6

155

Table 6.27 Aq. humor concentration vs time

Sr.

No.

Time

(Hr)

Concentration (µg/ml)

P1 Marketed

LFX eye

drops

P4 Marketed

OFX eye

drops

P7 Marketed

NFX eye

drops

1 0 0 0 0 0 0 0

2 0.5 2 2 1.8 1.5 1 0.9

3 1 4 2.5 3.3 2.2 1.4 1.1

4 2 3.8 2 3.1 1.8 1.2 0.9

5 3 3.5 1.5 3 1.2 0.9 0.5

6 4 3 0.8 2.5 0.6 0.7 0.3

7 5 2 0.5 2.2 0.3 0.6 0.1

8 6 1.8 0.3 2 0.1 0.5 0.07

Fig. 6.70 Plot of aq. humor concentration vs time

156

Data Analysis:

Pharmacokinetic parameters were determined by non-compartmental

analysis.

The maximum aq. humor concentration (Cmax) and time (Tmax) of it

occurrence were directly computed from the plasma concentration vs. time

plot.

The elimination rate constant (Kel) was determined from the terminal

phase of the log aq. humor concentration vs. time study and was calculated as

Kel = 2.303 x slope.

The elimination half-life was calculated using the formula 0.693/Kel.

The area under the curve (AUC) was calculated from the aq. humor

concentration vs. time study by trapezoidal method.

Table: 6.28 Pharmacokinetic parameters of optimized IS gel and maketed

eye drops

Formulation

Code

Tmax

(hr)

Cmax

(μg/ml)

t1/2

(hr)

AUC

(µg/ml/hr)

Kel (hr-1

)

P1 1 4.0 3.709 17.2 0.1868

Marketed LFX eye drops 1 2.5 1.46 7.82 0.4743

P4 1 3.3 3.820 14.62 0.1814

Marketed OFX

eye drops

1 2.2 0.959 6.35 0.7226

P7 1 1.4 3.166 5.2 0.2188

Marketed NFX

eye drops

1 1.1 1.0933 3.1 0.6383

157

Fig. 6.71 Comparison of area under curve (µg/ml/hr) in aq. of optimized IS

gel formulations and marketed eye drop

The pharmacokinetic parameters are shown in Table 6.28. The Cmax of marketed

eye drop formulation of LFX, OFX and NFX was found be 2.5, 2.2 and 1.1 μg/ml

respectively. The corresponding Tmax was 1.0 hr. The Cmax of P1, P4 and P7

formulations were 4.0, 3.3 and 1.4 μg/ml respectively and the corresponding Tmax was

1.0 hr for all the three formulations. Drainage is very rapid and generally limits ocular

contact at the site of absorption to about 3-10 minutes. However, the lag time (the

time for drug to transverse the cornea and appear in the aq. humor) is sufficiently long

to extend time to maximal concentration in the aq. to between 20 and 60 minutes for

most drugs. The MIC90 of LFX, OFX and NFX is in the range of ≤0.25-2µg/ml, for

most of the susceptible microorganisms [152]. As shown in Fig. 6.70, the MIC90 of

drug in aq. humor was obtained by ISG formulation and was maintained upto study

duration of 6 hr. In the marketed eye drops solution, there is rapid increase in drug

concentration and then drop down after some time. Cmax of ISGS viz. P1, P4 and P7

0 5 10 15 20

P1

LFX Marketed eye drops

P4

OFX Marketed eye drops

P7

NFX Marketed eye drops

Area (µg/ml/h)

158

was found to be 1.6, 1.5 and 1.3 times higher than marketed eye drops solution

respectively at the similar Tmax of 1 hr. Fig. 6.71 shows comparison of area under

curve in aq. of optimized thermosensitive IS gel formulations and marketed eye drop.

The AUC0-360min of LFX is more than OFX and NFX. The results indicate the

significant permeation of LFX than OFX and NFX. Also the IS gel formulation

showed more AUC0-360min than their respected marketed eye drop formulations. The

more AUC0-360min of IS gel formulations is because of increased contact time in the

eye. The developed IS gel formulations improved contact time, there by improved

bioavailability of drug as proved by drug aq. humor concentrations.

6.4.2.16 Stability Studies

Stability of the prepared formulation from the period of manufacturing to its

usage is very important. Selected sterilized formulations viz. P1, P4 and P7 were

stored at 5±3oC and 30±2

oC/65% RH ±5% RH for duration of 90 days. The

formulations were evaluated at periodic intervals for assay, clarity, pH, liquid–gel

conversion and transcorneal permeation study (Table 6.29). No significant change

was observed. The optimized systems of LFX, OFX and NFX are ISG based on

thermogelation, gels at 33-34oC. The formulation should be stored at cool conditions

or below 25oC. At these storage conditions (cool place) the developed systems

remains in the form of clear solution. As degradation is less than 5 percent,

approximate shelf life of 24 months can be allotted to the optimized formulations

[126].The rate constant of decay was determined by plotting the log of % drug

remaining vs time for P1, P4 and P7 respectively (Fig. 6.72 to 6.74) using Arrhenius

plot according to first order kinetics. The degradation rate constants were calculated

from slopes of the straight line.

159

Table: 6.29 Drug content and permeation data during stability studies

For

mul

atio

n

Storage

conds.

Parameters evaluated

15 days 30 days 60 days 90 days

Drug

Conte

nt

%CD

R

(6Hr)

Drug

Conten

t

%CD

R

(6Hr)

Drug

Conten

t

%CD

R

(6Hr)

Drug

Conten

t

%CD

R

(6Hr)

P1 5±3oC 99.9±

0.20

96.46±

0.52

99.6±

0.55

95.40±

0.32

99.5±

0.7

94.79±

0.72

99.0±

0.68

94.50±

0.72

30°C,

65%RH

±5% RH

99.90±

0.60

95.00±

0.74

99.5±

0.55

94.40±

0.60

99.0±

0.68

94.00±

0.38

98.0±

0.36

93.00±

0.58

P4 5±3oC 99.50±

0.40

94.26±

0.44

99.0±

0.40

93.06±

0.84

98.19±

0.10

92.76±

0.34

97.19±

0.10

92.26±

0.39

30°C,

65%RH

±5% RH

99.00±

0.90

93.00±

0.88

98.50±

0.10

93.00±

0.84

97.00±

0.30

92.89±

1.04

96.00±

0.50

90.66±

0.30

P7 5±3oC 99.93±

0.15

80.55±

1.03

99.03±

0.11

79.25±

1.33

98.93±

0.15

79.05±

0.93

97.73±

0.35

78.25±

0.83

30°C,

65%RH

±5% RH

99.53±

0.15

77.95±

1.03

98.03±

0.12

77.25±

1.0

97.73±

0.18

76.65±

0.90

97.00±

0.15

75.85±

0.53

Table: 6.30 Degradation rate constant (K) data during stability studies

Temperature

Degradation rate constant (K)

P1 P4 P7

5±3oC 1.5 ×10

-4 3.0 ×10

-4 2.6 ×10

-4

30°C and

65%RH±5% RH

2.3 ×10 -4

5.3 ×10 -4

4.6×10 -4

160

Fig. 6.72 Log of % drug remaining for P1 vs time

Fig. 6.73 Log of % drug remaining for P4 vs time

1.99

1.992

1.994

1.996

1.998

2

0 20 40 60 80 100

Log o

f %

dru

g r

emain

ing

Time (days)

5 degree C

30 degree C

1.975

1.98

1.985

1.99

1.995

2

2.005

0 20 40 60 80 100

Log o

f %

dru

g r

emain

ing

Time (days)

5 degree C

30 degree C

161

Fig. 6.74 Log of % drug remaining for P7 vs time

1.984

1.986

1.988

1.99

1.992

1.994

1.996

1.998

2

2.002

0 20 40 60 80 100

Lo

g o

f %

dru

g r

ema

inin

g

Time (days)

5 degree C

30 degree C