45607AT Flash Point Technical Offer 2015.1.docx 1

Technical Offer

Flash Point –2015.1 Measure Flash Point Hydrocarbon Products Quickly, Accurately, and

Completely On-Line!

Advantages:

• Designed to measure Flash Point from 10ºC to 121ºC

• No catalyst used, performance is not impacted by sulfur content in the measured stream

• Microprocessor-controlled instrument diagnostic capabilities

• Analyzer configuration can be changed without opening the explosion proof enclosure

Page 2: 45607AT Flash Point Technical Offer 2015.1 Flash...45607AT Flash Point Technical Offer 2015.1.docx 2 1. System Description The PSPI Model 45607AT Flash Point Monitor is an on-stream

45607AT Flash Point Technical Offer 2015.1.docx 2

1. System Description

The PSPI Model 45607AT Flash Point Monitor is an on-stream process monitor that measures the flash point of hydrocarbon products over a flash range of 10° to 121°C (50° to 250°F) and a viscosity range of up to 1000 SSU (220 CS). The analyzer robust design and measurement principle (no catalyst used in the measurement) delivers reliable measurements over time independently of fuel’s sulfur content. The Flash Point analyzer is completely microprocessor controlled with self-diagnostic capabilities, making it economical and simple to operate and maintain.

Applications Flash Point is one of the key final products specifications in the refineries. By having an online monitoring solution, refineries can optimize plants profitability while at the same time meeting quality requirements.

Methods Test results correlate with ASTM D-56 Tag Closed Cup test method. When measuring flash points above 79°C (175°F), analyzer results should be correlated with a laboratory analysis by ASTM D-93 Pensky-Martins Closed Cup test method or a similar laboratory test method.

Theory of Operation

Sample and air are fed to the measurement enclosure where they are mixed and fed into the Flash Cup Assembly. The mixture flows past a cartridge heater and a liquid space RTD sensor into the base of the flash cup. Excess liquid overflows to drain while the fuel-air vapors rise past a splash shield, vapor-space thermocouple, and high voltage electrodes before venting to drain.The temperature of the fuel-air mixture is slowly increased while a high voltage electrode circuit generates a spark. The vapor-space thermocouple is constantly monitored for a sudden rise in temperature. When flash occurs, the vapor-space thermocouple spikes. The sample temperature at the moment of ignition, as measured by the liquid space RTD sensor, is reported as the flash point temperature and a new measurement cycle is initiated. The flash point temperature is displayed on the monitor’s digital readout and is output as a 4-20 mADC signal.

Page 3: 45607AT Flash Point Technical Offer 2015.1 Flash...45607AT Flash Point Technical Offer 2015.1.docx 2 1. System Description The PSPI Model 45607AT Flash Point Monitor is an on-stream

45607AT Flash Point Technical Offer 2015.1.docx 3

1. Technical Requirements & Specifications Ordering No. Description

45607AT-115V 45607AT-230V

Flash Point 100-120 VAC Flash Point 220-240 VAC

GENERAL INFO

45607 Process Flash Point Monitor

Standard Test Methods Correlates to ASTM D 56 and D 93

OPERATION

Response Time Typically 1 – 7 minutes. Cycle time varies with process operating conditions. Application dependant

APPLICATION RANGE

Ranges 10ºC to 121ºC

Accuracy Correlates to ASTM D 56 and D 93 Repeatability +/- 1.0ºC

SAMPLE CONDITIONS

Sample Phase Liquid

Sample Flowrate 30 to 50 ml/minute filtered sample

Sample Viscosity No greater than 220 cSt at 38ºC

Sample Inlet Pressure 5 – 10 psig

Sample Return Pressure Atmospheric pressure

Sample Temperature At least 15ºC below expected lowest flash point temperature (cooling may be required in sample conditioning system)

SUPPLY & CONNECTIONS

Instrument Air Clean, dry and filtered a 600-1000 cc/minute and regulated at 10 psig

POWER REQUIREMENTS

Voltage 100-120 or 220-240 VAC (+/-10%), 50/60 Hz, single phase

Power 500W

INTERFACE SPECIFICATIONS

Remote Standby Terminals available for customer-supplied dry contact control of instrument. Allows control room to take monitor “off-line”.

Analog Output Isolated 4-20mA for Flash Temperature (standard) Isolated 4-20mA for Sample Temperature (optional)

Alarm Relay SPST fail-safe alarm relays Digital Outputs Modbus over RS-232C, RS-485, RS-422 “Come Read” Contact Dry relay contact for end of measurement cycle notification (optional) ENVIRONMENTAL CONDITIONS

Operating temperature 0°to 40°C; weather protection required; no direct sunlight

Storage temperature Consult Factory

Altitude Sea level to 2,000 meters

SAFETY CHARACTERISTICS

ATEX IECEx: II 2 G ATEX Ex d IIB+H2 T4 Gb IP66

CSA Class I Div I Group B, C, D, T4 Type 13, 4 and 4x, or Class 1 Zone 1, Ex/AEx d IIB+H2 IP66 T4

CE Compliant PHYSICAL SPECS

Dimensions 48x24x13 inches (uncrated) ; 54x29x17 inch (crated)

Weight 250 lbs (uncrated) ; 350 lbs (crated)